You are on page 1of 7

MINI-REVIEW

Interference in Ion-Selective Electrodes Due to


Proteins and Lipids
Sudip Kumar Datta and Parul Chopra*

Background: Ion-selective electrodes (ISE) have become the mainstay of electrolyte measurements in the clini-

Downloaded from https://academic.oup.com/jalm/article/7/2/589/6424217 by guest on 06 October 2023


cal laboratory. In most automated analyzers used in large diagnostic laboratories, indirect ISE (iISE) -based electro-
lyte estimation is done; whereas direct ISE (dISE) -based equipment are mostly used in blood gas analyzers and in
the point-of-care (PoC) setting.
Content: Both the techniques, iISE as well as dISE, are scientifically robust; however, the results are often not in-
terchangeable. Discrepancy happens between the two commonly due to interferences that affect the two measur-
ing principles differently. Over the last decade, several studies have reported discrepancies between dISE and iISE
arising due to abnormal protein and lipid contents in the sample.
Summary: The present review endeavors to consolidate the knowledge accumulated in relation to interferences
due to abnormal protein and lipid contents in sample with the principal focus resting on probable solutions
thereof.

IMPACT STATEMENT

• Interference due to abnormal protein and lipid concentrations on electrolyte measurements by Ion-
selective electrodes (ISE) is widely reported.
• This predominantly affects indirect ISE leading to discrepancy between indirect and direct ISE results.
• Several researchers have suggested solutions for correction of indirect ISE results using sample free
water content, algorithms using protein and lipid concentration, or otherwise.
• These solutions have their own limitations and are not universally applicable. Moreover, reports studying
the effects of these interferents simultaneously are few and lack consensus.
• This review aims to consolidate the existing knowledge and probable solutions thereof.

Department of Laboratory Medicine, All India Institute of Medical Sciences, New Delhi, India.
*Address correspondence to this author at: Department of Laboratory Medicine, All India Institute of Medical Sciences, New Delhi, India.
E-mail parul6588@yahoo.co.in.
Received July 13, 2021; accepted September 13, 2021.
DOI: 10.1093/jalm/jfab125
C American Association for Clinical Chemistry 2021. All rights reserved.
V
For permissions, please email: journals.permissions@oup.com.

...................................................................................................
March 2022 | 07:02 | 589–595 | JALM 589
MINI-REVIEW Interference in Ion-Selective Electrodes

INTRODUCTION to erroneously high or low results. Discrepancy is


the difference in results arising between different
Ion-selective electrodes (ISE) are potentiometric analytic techniques measuring the same analyte.
sensors used to detect the potential difference Interferents in a specimen may lead to discrep-
between two half cells within an electrochemical ancy in measurement due to their variable effect
cell. When this potential difference is produced at on the two measuring principles. During estima-
a selective membrane and solution interface, it is tion of electrolytes by iISE, interference has often
proportional to the logarithm of ionic activity or been reported due to high or low concentration
concentration of the ion being measured. The of proteins and lipids, but dISE methods are

Downloaded from https://academic.oup.com/jalm/article/7/2/589/6424217 by guest on 06 October 2023


membrane may be composed of glass, polymeric known to be relatively spared (2–4). This often
material, or crystalline material; however, it is se- leads to discrepancies in values obtained from the
lective for the ion of interest. Measurements by two ISE methods. It is important to identify these
ISEs are rapid, have high accuracy and precision, interferences and take appropriate steps or es-
low cost, use small sample volumes, and provide tablish procedures to eliminate the effect of such
measurements over a wide range. Thus, they have interferences by either removing the interferent
become the mainstay of electrolytes measure- or using appropriate testing methods.
ment in the clinical laboratory.
ISEs may be direct and indirect. Direct ISEs DISCREPANCIES DUE TO ELECTROLYTE
(dISE) are present mostly in arterial blood gas EXCLUSION EFFECT
(BGA) and point-of-care (PoC) testing analyzers. In
dISE, the membrane is directly exposed to the A common factor affecting the accuracy of the
sample, and it measures electrolyte activity. iISE method is displacement of plasma water by
Measurement by direct potentiometry is depen- proteins and lipids in cases of hyperproteinemia
dent only on molality and therefore not affected or hyperlipidemia, by a phenomenon recognized
by variations in the concentration of protein or lip- as the “electrolyte exclusion effect” (5). In iISE,
ids in the sample. Indirect ISEs (iISE) are commonly samples are premixed with diluent, varying in ra-
used in high throughput automated chemistry tios from 1:5 to 1:46 allowing measurement with
analyzers. In iISEs, samples are diluted before test- lower sample volumes and expanding the mea-
ing to allow measurement with lower sample vol- surable concentration range (6). The concentra-
umes and to expand the measurable tion of the electrolytes is proportional to molality
concentration range. In the last 2 decades several and mass concentration of water in kg/L in the
studies have reported the discrepancies arising plasma, which is normally taken as 0.93 kg/L for
between dISE and iISE values due to abnormal calculation purposes. However, in samples with el-
protein and lipid concentrations in the sample. evated lipids and/or proteins the value may be as
Dimeski et al. has discussed several interferences low as 0.8 kg/L. When the dissolved proteins or lip-
in his grand review published in 2010 (1). Here, we ids increase, iISE output that is standardized for a
focus on the discrepancies arising out of abnor- predilution plasma water mass concentration of
mal protein and lipid content in the samples and 0.93 kg/L gives falsely low results. This phenome-
consolidate the possible corrective measures. non of “pseudohyponatremia” has been reported
Interference in measurement may occur when in the literature due to the effects of proteins as
a different substance present in the specimen well as lipids (7). Techniques requiring sample dilu-
leads to a false alteration in measurement leading tion, such as in flame photometry or iISE, are

........................................................................................................
590 JALM | 589–595 | 07:02 | March 2022
Interference in Ion-Selective Electrodes MINI-REVIEW

reported to be affected by the presence of in- that study, produced significant pseudohyperna-
creased proteins and/or lipids in the sample be- tremia in hypoproteinemic samples (<40 g/L) (13).
cause the calculation algorithms assume mass Similar findings were reported earlier in another
concentration of water to be 0.93 kg/L (8, 9). study, where 34 out of 80 ICU patients (43%)
Similarly, studies showing erroneous hypokalemia exceeded a difference of 4 mmol/L between iISE
and decreased chloride (Cl) concentrations are and dISE (6). Interestingly, different automated
also available and are thought to be of similar analyzers use different dilutions for electrolytes
magnitude; however, they are less evident in view measurement. Surprisingly, the equipment using
of their lower concentrations in serum compared the least dilution factor was observed to have the

Downloaded from https://academic.oup.com/jalm/article/7/2/589/6424217 by guest on 06 October 2023


to sodium (Naþ) (7). On the other hand, in dISE, highest differences from the dISE measurements;
the sample comes in direct contact with the selec- hence the interplay of other factors contributing
tive membrane and electrolyte estimation is not to the process needs to be investigated.
affected by any variation in the nonaqueous Several studies also report the effect of lipemia
phase giving a true estimate of the plasma electro- on values Naþ, Kþ, and Cl concentrations mea-
lytes despite any changes in nonaqueous plasma sured by iISE vis-à-vis dISE (9, 15). Hyperlipidaemia
components. Direct ISE provides measurement of decreases the serum electrolyte levels when mea-
activity, which is the concentration of free un- sured by iISE in a similar way to hyperproteinemia.
bound ions in solution, and it depends on the ac- Dimeski et al. (9) observed that there is a de
tivity coefficient of the sample. Activity coefficient crease of Naþ and Cl by 1 mmol/L and Kþ by
of the sample, in turn, depends on the ionic 0.04 mmol/L, respectively, per 10 mmol/L increase
strength. The charge of protein and other macro- in total lipids i.e., cholesterol plus triglycerides
and micromolecules contribute to the ionic (TG). Case reports are also available for patients
strength of the sample. This fundamental differ- with obstructive jaundice having hypercholesterol-
ence between the two apparently similar technol- emia who presented with pseudohyponatraemia
ogies often gives rise to discrepancies in (16, 17). These articles report patients with chole-
electrolyte measurement. stasis secondary to bone marrow transplant with
The opposite effect of pseudohypernatremia graft versus host disease, primary biliary cirrhosis,
has also been reported in literature due to low hepatitis C, or drug or secondary to structural ab-
protein or albumin concentration (10–12). In a re- normality in the case of pancreatic cancer; all hav-
cently reported study, almost half of the samples ing hyperlipidemia and erroneous electrolyte
analyzed using iISE, which were wrongly classified values (10, 18, 19). In a case report, a 43-year-old
as normonatremic, were found to be of pseudo woman with refractory primary biliary cirrhosis
normonatremia, i.e., a normal value obtained and cholestasis had extreme hypercholesterol-
when there is true hyponatremia (13). However, emia from elevations of lipoprotein-X particles
not enough literature is available on this issue. that resulted in pseudohyponatremia. After multi-
Dimeski et al. (14) have defined the clinically rele- ple plasmapheresis and normalization of choles-
vant difference between indirect and direct so- terol level from 2415 mg/dL (62.45 mmol/L) to
dium results based on analytical and biological 200 mg/dL (5.17 mmol/L), her falsely decreased
variation as 4 mmol/L (14). Using this cutoff, one serum sodium level of 121 mmol/L normalized to
study reported that, all five different routine 139 mmol/L when measured by indirect potenti-
chemistry instruments using iISE, compared in ometry (18).

.....................................................................................................
March 2022 | 07:02 | 589–595 | JALM 591
MINI-REVIEW Interference in Ion-Selective Electrodes

MITIGATION STRATEGIES has thus been suggested as a tool to recognize


pseudohyponatremia. An observed difference be-
Several studies have compared the BGA and tween calculated and measured osmolality may
PoC electrolyte analyzers with fully automated help the laboratory to identify suspected cases of
chemistry analyzers for Naþ, potassium (Kþ), and pseudohyponatremia. However, this would mean
Cl estimation. Most authors have advocated the requirement of additional cost, time, and
use of a single technology for serial monitoring of expertise.
electrolytes for management of patients, or Osm can be determined directly by measure-
recommended the use of dISE in critical care ment of colligative properties of the solution,

Downloaded from https://academic.oup.com/jalm/article/7/2/589/6424217 by guest on 06 October 2023


cases (3, 20–22). However, several others have most commonly, using freezing point depression
suggested methods to correct the pseudo or by measurement of vapor pressure. Since Naþ,
hyponatremia. Cl, glucose, and urea are the major osmotic sub-
stances in normal plasma, osmolality can also be
Consideration of Serum Free Water Fraction calculated using the equation
One such approach used the calculation of se- mOsm=kg ¼ 1:86½Naþ ðmmol=LÞ þ glucose ½mmol=L
rum free water fraction as initially suggested by þ urea ðmmol=LÞ

Waugh (23) and described in detail later by Levels of sodium ion are required in the formula
Fortgens and Pillay (7). The Naþ measured by iISE for Osm, hence, erroneous measurement of so-
can be multiplied by the normal serum water dium can cause erroneous calculation of Osm.
value of 93% and divided by its calculated value to The same has been studied by looking at the ef-
obtain an accurate estimate of serum Naþ. fect of lipemia on calculated and measured Osm,
However, in the formula which revealed that lipemia does not affect Osm,
serum water % ¼ 99:1  ½1:1  10  5  lipid ðTGÞ concentration as measured by freezing point depression.
in mmol=L – ð0:07  protein concentration in g=LÞ However, interference of lipemia with respect to
Naþ and Kþ concentrations measured using the
only the TG concentration and total protein con- indirect ISE method causes low calculated Osm
centrations are considered. The cholesterol com- values and may cause an erroneous increase in
ponent of serum lipids is conspicuously absent, the osmolal gap, especially in cases of toxic alco-
however, high cholesterol is recognized as a cause hol poisoning (26).
of pseudohyponatremia although uncommon
(24). Calculation of Protein-Corrected Sodium
This problem of pseudohypernatremia
Consideration of Osmolality
becomes all the more relevant in small children
In a recent elaborate review on hyponatremia in who commonly have hypoalbuminemia; and
hematologic diseases, the authors discussed at timely identification of hyponatremia is crucial for
length the different hematological conditions that proper management. Therefore, use of dISE has
are associated with pseudohyponatremia. These been recommended by a couple of studies on
conditions with underlying hyperglobulinemia, pseudohypernatremia or pseudonormonatremia
hypertriglyceridemia, hypercholesterolemia, or in sick infants (27, 28). In a study on 450 preterm
other causes of dyslipidemia mostly have isotonic infants (weight <2500 g) admitted to the neonatal
hyponatremia (serum osmolality being 280– intensive care unit, low plasma protein level
295 mOsm/kg) (25). Measurement of osmolality (<4.3 g/dL) was found to be an independent risk

........................................................................................................
592 JALM | 589–595 | 07:02 | March 2022
Interference in Ion-Selective Electrodes MINI-REVIEW

factor for a Naþ level difference of >4 mmol/L be- interference due to lipemia for accurate estima-
tween the two methods (27). However, since dISE tion of electrolytes in lipemic samples. Although
instruments mean additional cost to a hospital, the interference was studied only in the case of
Stove et al. suggested using a protein-corrected iISE previously, a study by Sen et al. (32) compared
sodium result as an acceptable alternative (29). In two different dISE based equipment, namely
their study, they identified that for change of every Vitros 250 (Raritan, New Jersey, USA) and HDC-
10 g/L in total protein concentration, a deviation Lyte (Kolkata, India). In their experimental study,
of 1.3 mmol/L is observed with Naþ indirect result. Intralipid 10% (which is a sterile fat emulsion con-
They proposed the following equation: taining soya oil, egg lecithin, and glycerol) was

Downloaded from https://academic.oup.com/jalm/article/7/2/589/6424217 by guest on 06 October 2023


protein  corrected Naþ indirect ¼ Naþ indirect  10:53 added in escalating concentrations to induce lipe-
þ ð0:1316  total proteinÞ mia in serum samples. Both the Naþ and Kþ val-
ues for the normal, hyper-, and hypoconditions
On application of this equation on a prospective
were found to have interference due to high TG
cohort of 4006 patient samples the percentage of
levels. The interference was observed beyond TG
conflicting data between iISE and dISE results sig-
level of 650 mg/dL (7.35 mmol/L) and was found
nificantly decreased.
to be increasing as the TG levels rose to
Another recent study, used the equation
1550 mg/dL (17.52 mmol/L) and beyond. The in-
fNaþ þ
adjusted ¼NaindirectISE [6.2  (0.16  albumin)]g
terference was statistically significant, and the
proposed by Story et al (12) to adjust serum so-
maximum drift was seen with TG levels 1050 mg/
dium concentrations obtained from iISE for serum
dL (11.86 mmol/L). They concluded that dISE, like
albumin concentrations. They reported that the
iISE at high TG concentrations, may influence elec-
differences between the direct and indirect meth-
trolyte measurements (32). However, the rationale
ods became negligible after use of the equation
for the same has not yet been understood.
(30).
We have recently published a report with the in-
However, the use of corrective formulas, devel-
oped based on regression model-based algo- tention of studying the simultaneous effect of pro-
rithms often seem impractical and tedious. teins and lipids on electrolyte measurement by
Besides, they do not hold true universally at all iISE (3). Although the previously reported results
ranges of measurement or for all interfering sub- by different authors were corroborated, we could
stance. Jones et al. derived regression equations not establish a combined formula for correcting
for the relationship between both the absolute iISE results using protein, TG, and cholesterol.
and relative difference between dISE and iISE and
the total protein concentration. But due to the CONCLUSION
large spread of data around the regression line,
the authors disapproved the use of such equa- It is important in clinical decision making to dis-
tions and recommended dISE for electrolyte esti- tinguish dyselectrolytemia from pseudodyselec-
mation instead of iISE in routine clinical trolytemia due to lipid or protein interference. A
laboratories (31). correlation with clinical findings of dyselectrolyte-
mia is often therefore the first step to identify an
Removal of Lipid Interference analytical error. However, laboratories need to be
Some studies suggest using either dISE or proactive to prevent release of erroneous reports.
methods such as ultracentrifugation or lipid clear- Mitigation strategies described by various
ing agents (e.g., Lipoclear) for removal of authors have been discussed here. In case

.....................................................................................................
March 2022 | 07:02 | 589–595 | JALM 593
MINI-REVIEW Interference in Ion-Selective Electrodes

discrepancy between clinical and laboratory find- both in large, automated laboratories and as BGA
ings is found, the sample may be rerun on dISE. or PoC devices because they provide rapid, inex-
However, dISE is not indicated in all situations. It is
pensive, accurate, and precise results. In spite of
necessary only when the sample is visibly lipemic,
high selectivity of the membranes that form an in-
or there is discrepancy between calculated and
measured Osm or, in cases with hyperglycemia. In tegral part of these electrodes, iISE are subject to
case instrument for measurement of dISE or Osm a lot of interference occurring due to abnormal
is not available, formulas for calculation of cor- levels of proteins and lipids in serum. Therefore,
rected sodium by free serum water or for protein-

Downloaded from https://academic.oup.com/jalm/article/7/2/589/6424217 by guest on 06 October 2023


discrepancies arise between the dISE- and iISE-
corrected-sodium may be used. These formulas,
based methods in these settings. Although several
however, have not been validated on large num-
ber of samples and for their robustness (33). attempts have been made to determine correc-
Another option is removal of lipids from a lipemic tive formulas, so far, no universally acceptable so-
sample but it requires extra reagents or instru- lution to the problem has appeared. Hence, dISE
mentation. Use of all these strategies would ulti- may be the safer bet in critical care settings and
mately demand additional cost, time, and
interchangeable use of electrolyte results from dif-
expertise.
ferent technologies should be avoided during se-
The ISEs play an indispensable role in clinical
chemistry testing for measurement of electrolytes rial monitoring.

Author Contributions: All authors confirmed they have contributed to the intellectual content of this paper and have met the follow-
ing 4 requirements: (a) significant contributions to the conception and design, acquisition of data, or analysis and interpretation of
data; (b) drafting or revising the article for intellectual content; (c) final approval of the published article; and (d) agreement to be ac-
countable for all aspects of the article thus ensuring that questions related to the accuracy or integrity of any part of the article are ap-
propriately investigated and resolved.

Conceptualization: S.K. Datta.


Writing, review & editing: P. Chopra, S.K. Datta.
Approval of final version: P. Chopra, S.K. Datta.
Both authors read and approved the final manuscript.
Authors’ Disclosures or Potential Conflicts of Interest: No authors declared any potential conflicts of interest.

REFERENCES
1. Dimeski G, Badrick T, John AS. Ion selective electrodes 5. Burtis CA, Ashwood ER, Bruns DE, editors. Tietz
(ISEs) and interferences—a review. Clin Chim Acta 2010; fundamentals of clinical chemistry and molecular
411:309–17. diagnostics. 7th Ed.St. Louis (MO):Elsevier/Saunders,
2. Goldwasser P, Ayoub I, Barth RH. Pseudohypernatremia 2012.
and pseudohyponatremia: a linear correction. Nephrol 6. Joosen AMCP, Boonen K, Hulsman N, Schuitemaker FJ,
Dial Transplant 2015;30:252–7. Thelen MHM. Hypernatraemia in disguise. Clin Chem Lab
3. Chopra P, Datta SK. Discrepancies in electrolyte Med 2014;52:e231–234.
measurements by direct and indirect ion selective 7. Fortgens P, Pillay TS. Pseudohyponatremia revisited: a
electrodes due to interferences by proteins and lipids. J modern-day pitfall. Arch Pathol Lab Med 2011;135:
Lab Physicians 2020;12:084–91. 516–9.
4. Katrangi W, Baumann NA, Nett RC, Karon BS, Block DR. 8. Langelaan MLP, Kamp L, Zandijk E, Raijmakers MTM.
Prevalence of clinically significant differences in sodium Prevalence of pseudonatremia in a clinical laboratory -
measurements due to abnormal protein concentrations role of the water content. Clin Chem Lab Med 2017;55:
using an indirect ion-selective electrode method. J Appl 546–53.
Lab Med 2019;4:427–32. 9. Dimeski G, Mollee P, Carter A. Effects of hyperlipidemia

........................................................................................................
594 JALM | 589–595 | 07:02 | March 2022
Interference in Ion-Selective Electrodes MINI-REVIEW

on plasma sodium, potassium, and chloride Plasma sodium measurements by direct ion selective
measurements by an indirect ion-selective electrode methods in laboratory and point of care may not be
measuring system. Clin Chem 2006;52:155–6. clinically interchangeable. J Clin Monit Comput 2017;31:
10. Chow E, Fox N, Gama R. Effect of low serum total protein 1103–9.
on sodium and potassium measurement by ion-selective 22. van den Ancker W, Haagen IA, van der Voort PHJ. Direct
electrodes in critically ill patients. Br J Biomed Sci 2008; sodium measurement prevents underestimation of
65:128–31. hyponatremia in critically ill patients. Intensive Care Med
11. Lang T, Prinsloo P, Broughton AF, Lawson N, Marenah CB. 2015;41:553–4.
Effect of low protein concentration on serum sodium 23. Waugh WH. Utility of expressing serum sodium per unit
measurement: pseudohypernatraemia and of water in assessing hyponatremia. Metabolism 1969;
pseudonormonatraemia!. Ann Clin Biochem 2002;39: 18:706–12.
66–7. 24. Le Riche M, Burgess LJ, Marais AD.

Downloaded from https://academic.oup.com/jalm/article/7/2/589/6424217 by guest on 06 October 2023


12. Story DA, Morimatsu H, Egi M, Bellomo R. The effect of Pseudohyponatraemia in a patient with obstructive
albumin concentration on plasma sodium and chloride jaundice. Clin Chim Acta Int Acta 2006;366:357–60.
measurements in critically ill patients. Anesth Analg 2007; 25. Koumpis E, Florentin M, Hatzimichael E, Liamis G.
104:893–7. Hyponatremia in patients with hematologic diseases. J
13. Tel-Karthaus N, Salet GAM, Jacobs LHJ, Nabbe KCAM, Clin Med 2020;9:3721.
Schoenmakers CHH, van der Doelen RHA, et al. 26. Alpdemir M, Alpdemir MF, Eryilmaz M, Yucel D. Effect of
Instrument dependent erroneous sodium lipemia on measured and calculated osmolality in native
measurements in hypoproteinemic critically ill patients lipemic samples and intravenous lipid emulsion-added
are causing significant misclassification of dysnatremias.
pools. Natl J Lab Med 2020;9:BO01–5.
Clin Chem Lab Med 2019;57:e222–5.
27. Kim H, Kim JK, Cho SC. Comparison of sodium ion levels
14. Dimeski G, Morgan TJ, Presneill JJ, Venkatesh B.
between an arterial blood gas analyzer and an
Disagreement between ion selective electrode direct and
autoanalyzer in preterm infants admitted to the neonatal
indirect sodium measurements: estimation of the
intensive care unit: a retrospective study. BMC Pediatr
problem in a tertiary referral hospital. J Crit Care 2012;27:
2016;16:101.
326.e9–e16.
28. King RI, Mackay RJ, Florkowski CM, Lynn AM. Electrolytes
15. Ladenson JH, Apple FS, Koch DD. Misleading
in sick neonates - which sodium is the right answer? Arch
hyponatremia due to hyperlipemia: a method-dependent
Dis Child Fetal Neonatal Ed 2013;98:F74–76.
error. Ann Intern Med 1981;95:707–8.
29. Stove V, Slabbinck A, Vanoverschelde L, Hoste E, De
16. Igbinedion SO, Pandit S, Mavuram MS, Boktor M.
Paepe P, Delanghe J. How to solve the underestimated
Pseudohyponatraemia secondary to hyperlipidaemia in
obstructive jaundice. BMJ Case Rep 2017;2017: problem of overestimated sodium results in the
bcr2017221984. hypoproteinemic patient. Crit Care Med 2016;44:e83-88.
17. Ravella S, Lefavour GS, Carayannopoulos MO, Parikh A. 30. den Elzen WP, Cobbaert CM, Arbous MS, Elzo Kraemer
Hyponatremia in a patient with obstructive jaundice. CV, Schoe A, de Jonge E, et al. Interchangeability of
Kidney Int 2015;88:921–2. sodium and chloride measurements by indirect and
18. Hussain I, Ahmad Z, Garg A. Extreme direct ISE assays: stakeholders, take responsibility! Pract
hypercholesterolemia presenting with Lab Med 2019;16:e00126.
pseudohyponatremia - a case report and review of the 31. Jones BJ, Twomey PJ. Relationship of the absolute
literature. J Clin Lipidol 2015;9:260–4. difference between direct and indirect ion selective
19. Sivakumar T, Chaidarun S, Lee HK, Cervinski M, Comi R. electrode measurement of serum sodium and the total
Multiple lipoprotein and electrolyte laboratory artifacts protein concentration. J Clin Pathol 2008;61:645–7.
caused by lipoprotein X in obstructive biliary cholestasis 32. Sen S, Ghosh P, Ghosh TK, Das M, Das S. A study on
secondary to pancreatic cancer. J Clin Lipidol 2011;5: effect of lipemia on electrolyte measurement by direct
324–8. ion selective electrode method. J Biomol Res Ther 2016;
20. Refardt J, Sailer CO, Chifu I, Winzeler B, Schnyder I, 5:142.
Fassnacht M, et al. The challenges of sodium 33. Nguyen MK, Ornekian V, Butch AW, Kurtz I. A new
measurements: indirect versus direct ion-selective method for determining plasma water content:
method. Eur J Endocrinol 2019;181:193–9. application in pseudohyponatremia. Am J Physiol Renal
21. Weld BA, Morgan TJ, Presneill JJ, Weier S, Cowley D. Physiol 2007;292:F1652–6.

.....................................................................................................
March 2022 | 07:02 | 589–595 | JALM 595

You might also like