BeneFusion VP5

Infusion Pump

W×H×D: 245 x 87 x174 (mm)

Classification:

Screen:
Brightness:
Information: Delivery rate, current infusion, VTBI, total volume, pressure limit, battery capacity, drugs,

remaining time, alarms

Infusion sets
Automatic recognition of infusion set brands after calibration, or BeneFusion exclusive IV sets

Diameter Φ 3.5-4.5mm, Thickness 0.8-1.2mm


Basic parameters
Mode: 6 modes: Rate mode, Time mode, Body weight mode, Ramp up/down mode,
Sequential mode, Micro-infusion mode
0.1-2000ml/h
Increment: 0.01ml(0.1-99.99ml/h); 0.1ml(100-999.9ml/h); 1ml(1000-2000ml/h)
0.10-9999.99ml(increment 0.01ml)
00:00:01-99:59:59( h:m:s )adjustable
0-9999.99 ml
0.1-5.0ml/h adjustable, step 0.1ml/h, default is 1.0 ml/h
Purge: 0.10-2000ml/h

Automatic/Manual Bolus: 0.1-2000ml/h, default is 800ml/h;


Preset automatic Bolus volume: 0.10-9999.99ml

Drug library up to 2000 drugs, “ON” and “OFF” selectable; use-defined drug list available
History records up to 2000, data could be transmitted to PC and saved as EXCEL format to print out.

Language
11 languages: English, Spanish, French, Russian, Turkish, Portuguese, Bulgarian, Czech, Polish, Romanian, Chinese

4 pressure unit: mmHg/kPa/bar/psi selectable, auto converted in different units, default is mmHg.
12 levels selectable, 75-975 mmHg, default is 450 mmHg.
Auto data lock :
Air detector: 5 levels selectable: 50ul, 100ul, 250ul, 500ul, 800ul, default is 100ul;
Single bubble or accumulated bubble volume in 15 min.
Anti-bolus:

Alarm type:
High: Occlusion, battery empty, VTBI Done, air in line, door-open, KVO finish, system malfunction
Middle: Standby time expired, system abnormal
Low: VTBI near done, Reminder, no battery, battery low, AC Power Disconnection

3 minutes before battery is completed; 30 minutes as low battery


Reminder :

Battery:

Power supply 100-240V~, 50/60Hz, 0.53-0.28A ; DC Voltage:10V-16V, 2.25A-1.5A

Humidity:
Air pressure:

>5 years under continuous duty conditions

Weight: 0.1-300.0kg or 0.1-660.8lb, kg/lb selectable;


Drug-Amount: 0.1-999.9, g/mg/ug/ng/IU selectable, default is mg;

Dose unit: mg,ug,ng,IU/kg/h; mg,ug,ng,IU/kg/min selectable, default is mg/kg/h;

are registered trademarks or trademarks owned by Shenzhen Mindray Bio-medical Electronics Co., LTD.
Mindray Building, Keji 12th Road South,
2013 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved. Specifications subject to changes without prior notice.
High-tech Industrial Park, Nanshan, Shenzhen 518057, P.R. China
P/N:ENG-BeneFusion VP5 Infusion Pump-210285x2-20160722
Tel: +86 755 8188 8998 Fax: +86 755 26582680
E-mail: intl-market@mindray.com www.mindray.com

You might also like