environmental conditions in prepartum and postpartum sheep
Katlin J. Hornig, BS; Stacey R. Byers, DVM, MS; Robert J. Callan, DVM, PhD; Timothy Holt, DVM; Megan Field, PhD; Hyungchul Han, PhD
Objective—To compare β-hydroxybutyrate (BHB) and glucose concentrations measured
with a dual-purpose point-of-care (POC) meter designed for use in humans and a laboratory biochemical analyzer (LBA) to determine whether the POC meter would be reliable for on- farm measurement of blood glucose and BHB concentrations in sheep in various environ- mental conditions and nutritional states. Animals—36 pregnant mixed-breed ewes involved in a maternal feed restriction study. Procedures—Blood samples were collected from each sheep at multiple points through- out gestation and lactation to allow for tracking of gradually increasing metabolic hardship. Whole blood glucose and BHB concentrations were measured with the POC meter and compared with serum results obtained with an LBA. Results—464 samples were collected. Whole blood BHB concentrations measured with the POC meter compared well with LBA results, and error grid analysis showed the POC values were acceptable. Whole blood glucose concentrations measured with the POC me- ter had more variation, compared with LBA values, over the glucose ranges evaluated. Re- sults of error grid analysis of POC-measured glucose concentrations were not acceptable, indicating errors likely to result in needless treatment with glucose or other supplemental energy sources in normoglycemic sheep. Conclusions and Clinical Relevance—The POC meter was user-friendly and performed well across a wide range of conditions. The meter was adequate for detection of pregnancy toxemia in sheep via whole blood BHB concentration. Results should be interpreted with caution when the POC meter is used to measure blood glucose concentrations. (Am J Vet Res 2013;74:1059–1065)
P regnancy toxemia, also known as pregnancy ketosis,
is a metabolic disorder that can develop as a result of relative undernourishment or excessive body condition BHB ABBREVIATIONS β-Hydroxybutyrate CV Coefficient of variation in late-gestation ewes, typically those bearing twin or dGA Days of gestational age multiple fetuses.1 A suspicion of clinical pregnancy tox- LBA Laboratory biochemical analyzer emia is supported by a ewe’s history and signalment as POC Point of care well as a progression of signs such as anorexia, lethargy, weakness, and neurologic abnormalities.2 When unde- tected or left untreated, the condition usually results in detection is delayed, resulting in an increase in the se- death of the ewe and its fetuses. verity of clinical signs. Therefore, a rapid and accurate Treatment of pregnancy toxemia can have disap- diagnosis of ketosis could increase the possibility of a pointing results and be expensive and time-consuming, successful treatment outcome while decreasing the like- particularly when multiple sheep are affected or when lihood of illness and death of affected ewes and lambs and lessening economic costs associated with delayed Received December 1, 2012. treatment and sheep loss. Accepted April 2, 2013. Sheep producers and veterinarians typically use From the Department of Clinical Sciences, College of Veterinary Med- urine ketone test strips to evaluate ketone status in late- icine and Biomedical Sciences (Hornig, Byers, Callan, Holt), and gestation ewes. The urine ketone strips can be used to the Department of Animal Sciences, College of Agriculture Sciences semiquantitatively measure the amount of acetoacetate (Field, Han), Colorado State University, Fort Collins, CO 80523. Supported with start-up funds provided to Dr. Byers. in urine but not of BHB, which is the primary circu- Presented in abstract form at the American College of Veterinary In- lating ketone body.3 Accuracy of the urine ketone test ternal Medicine Forum, New Orleans, June 2012. strips in dairy cows is reportedly variable, and results Address correspondence to Dr. Byers (stacey.byers@colostate.edu). can be affected by environmental conditions (eg, ambi-
AJVR, Vol 74, No. 8, August 2013 1059
Unauthenticated | Downloaded 06/26/23 02:07 PM UTC
12-12-0397r.indd 1059 7/18/2013 2:41:11 PM
ent humidity or temperature), user characteristics (eg, 22 dGA, 75 through 80 dGA, and 88 through 90 dGA; nature of handling and interval from testing to reading), and weekly thereafter from 91 dGA through parturition some medications, and pigmenturia.4,5 Urine collection (at approx 147 to 152 dGA). Postpartum samples were can be difficult and is stressful for ewes because urina- collected once per day until serum BHB concentrations tion is most reliably induced by restraining the ewe and were < 1.0 mmol/L, then every 2 weeks for 6 weeks, and holding off the nares to prevent breathing.6 The ewe again when lambs were weaned (approx 11 to 12 weeks eventually becomes distressed and postures to urinate. after parturition). For sample collection, each ewe was A POC meter designed to measure blood glucose restrained within its pen by 1 person while a second and BHB concentration in humans has been evalu- person obtained the blood sample via jugular veni- ated for detection of ketosis in dairy cattle and dairy puncture with an 18-gauge, 2.8-cm needle and 12-mL sheep.7–9 The only other method for measuring blood syringe. BHB concentrations in prepartum animals is with an LBA. A POC glucometer is available for veterinary use; BHB and glucose concentration measurement— however, it has not been evaluated or marketed for use Immediately after collection, blood samples were ana- in sheep. lyzed with each of 2 dual-purpose POC metersa: one Although blood glucose concentrations vary in sheep calibrated for BHB measurement and the other cali- with pregnancy toxemia, they can be used to assist in brated for glucose measurement. Meter calibration was evaluating response to treatment. Therefore, a reliable performed in accordance with the manufacturer’s in- POC meter capable of measuring both BHB and glucose structions by insertion of a calibrator key specific to the concentrations would be beneficial and economical for BHB or glucose test strips into the meter. A lot code sheep producers and veterinarians. The objective of the was matched between the calibrator key and the test study reported here was to compare blood BHB and glu- strips. Calibration was performed prior to each blood cose concentrations measured with a dual-purpose POC collection day and when lot numbers changed between meter with those of an LBA in various environmental con- test strips. The POC meters were designed to measure ditions and throughout gestation and lactation in healthy whole blood BHB concentrations between 0.0 and 8.0 and nutritionally stressed ewes at high risk for developing mmol/L10 and glucose concentrations from 20 to 500 pregnancy toxemia. The hypothesis was that the POC me- mg/dL.11 Samples were immediately retested with a new ter would yield results of sufficient accuracy to support its test strip whenever the meter displayed an error code or use for measurement of blood BHB and glucose concen- provided a result of high or low. trations in sheep on farms. Before the study began, the precision of the POC meter had been tested by use of whole blood samples collected from a nonpregnant sheep and a ketotic cow. Materials and Methods A whole blood sample (3 mL) collected by jugular or Animals—Thirty-six adult mixed-breed white-face coccygeal venipuncture had been immediately tested ewes were used for the study. Each ewe was identified 10 times with the 2 POC meters calibrated for study with a unique ear tag number. The sheep were concur- BHB and glucose measurements, and the CV had been rently involved in an unreported study of maternal nu- calculated for each analyte. tritional energy restriction and were assigned equally The POC meters were evaluated in various con- into 3 study groups. Group 1 (control group) was fed ditions, which included environmental temperatures 100% of their nutrient requirements, group 2 was fed ranging from –10° to 25°C, dusty conditions, and wet 50% of their nutrient requirements from 28 through (rain and snow) conditions given the open barn design. 78 dGA, and group 3 was fed 50% of their nutrient re- The 2 meters and test strips were stored in small, insu- quirements from 28 dGA through parturition. Nutri- lated coolers with chemical hand warmersb when en- ent requirements were based on the National Research vironmental temperatures were lower than the meters’ Council’s recommendations for twin-bearing ewes in operating temperature range (10° to 50°C). Extra test early and late gestation. The ewes were housed in in- strips were stored in their original packaging between dividual sand pens with individual automatic waterers 10° and 25°C. and feed pans to control their feed intake. All pens were The remainder of each blood sample was trans- located within a 3-sided barn that was not temperature ferred into evacuated serum tubesc and stored on ice regulated. prior to transport to the Colorado State University Ovulation in the ewes was synchronized, after Clinical Pathology Laboratory. Within 3 hours after which they were exposed to a ram at 12-hour intervals sample collection, the serum tubes containing clotted for a 2-day period. The second day of exposure to the blood were centrifuged for 5 minutes at 1,000 X g, se- ram was used as the breeding date for the study. Trans- rum was harvested, and serum samples were refriger- abdominal ultrasonography was performed on ewes ob- ated overnight. The serum BHB and glucose concen- served to have been bred at 45 days after breeding (45 trations were measured with an LBAd within 24 hours dGA), and ewes confirmed to be pregnant with twins after blood sample collection. The LBA was calibrated were selected for the maternal malnutrition study. The in accordance with the manufacturer’s instructions. study protocol was approved by the Colorado State Uni- The analyzer measured serum BHB concentration with versity Institutional Animal Care and Use Committee. a spectrophotometric method and glucose concentra- tions with a hexokinase method. Blood sample collection—Blood samples were col- lected from each ewe throughout gestation at the fol- Statistical analysis—Data were analyzed with the lowing intervals: once at 3 time points from 20 through aid of a commercial software package.e The D’Agostino-
1060 AJVR, Vol 74, No. 8, August 2013
Unauthenticated | Downloaded 06/26/23 02:07 PM UTC
12-12-0397r.indd 1060 7/18/2013 2:41:11 PM
Pearson test indicated that the data from blood sample presumption that treatment would be deemed neces- testing were not normally distributed, so nonparamet- sary in a clinical setting when the BHB concentration ric analyses were used. The Wilcoxon test was used to was ≥ 0.8 mmol/L. No lower limit for intervention was determine whether the POC meter results for blood deemed necessary for this study. BHB and glucose concentrations were significantly dif- For the glucose concentration error grid plot, zone ferent from the LBA results for serum BHB and glucose A represented the POC meter glucose concentrations concentrations. The Spearman rank correlation coeffi- that were within 20% above or below the LBA concen- cient (ρ) was calculated to assess correlations between trations as well as when the POC concentration was ≤ POC meter and LBA results. The percentage difference 50 mg/dL when the LBA concentration was also ≤ 50 between the POC meter and LBA results was calculated mg/dL. Zone B contained the POC meter glucose con- for each paired whole blood and serum BHB or glucose centrations that were > 20% of LBA concentrations but value. Percentages of POC meter results for BHB and would result in a benign treatment decision or no treat- glucose that were within 5%, 10%, and 20% above or ment decision. Zone C contained the POC glucose con- below the LBA values were then calculated. centrations that would result in treating a sheep when Agreement between results of the POC meter and the LBA glucose concentration was within the accept- the LBA was assessed via Bland-Altman methods for able range, and zone D contained POC glucose concen- multiple observations per test subject,12 and plots were trations that would result in not treating a sheep when constructed so that the identities of sheep could be dis- the LBA concentration was lower than the acceptable cerned by feeding group. Bland-Altman plots describe range. The zones were established with the presump- the limits of agreement between 2 methods as the mean tion that treatment would be deemed necessary when difference between measurements ± 1.96 SD. Through the glucose concentration was ≤ 50 mg/dL. No upper examination of the plotted data, one can detect system- limit for treatment intervention was deemed necessary atic biases (mean difference) and proportional biases for glucose concentrations. The meter would be con- (positive or negative slope in the data). The LBA was sidered acceptable for on-farm use if at least 95% of the considered the gold standard method in these analyses. BHB or glucose concentration were within zones A and Error grid plots were developed to theoretically B. The BHB and glucose concentration intervention evaluate the results of clinical decisions regarding guidelines were not implemented in this study. Values whether the POC meter BHB or glucose concentration of P ≤ 0.05 were considered significant. Device results should be used rather than the LBA concentration. To are summarized as mean ± SD. do this, the Clarke error grid analysis method and ac- ceptance criteria were modified because the critical lim- Results its pertinent to sheep are different from those used in humans.13 On the BHB plot, zone A represented POC Blood samples—Four hundred sixty-four blood meter BHB concentrations within 20% above or below samples were obtained from the 36 ewes over the study the LBA concentrations as well as when the POC BHB period. Three blood samples were not used in BHB or concentration was < 0.8 mmol/L and the correspond- glucose concentration statistical analyses because of ing LBA BHB concentration was also < 0.8 mmol/L. gross hemolysis that occurred during collection. Addi- Zone B contained POC meter concentrations that were tionally, 16 blood glucose concentrations were reported > 20% of LBA concentrations but would result in a be- as low (< 20 mg/dL) by the POC meter and remained nign treatment decision. Zone C contained POC meter low despite retesting. The corresponding serum LBA BHB concentrations that would result in not treating glucose concentrations ranged from 10 to 71 mg/dL. a sheep when the LBA BHB concentration was higher These values were omitted from the statistical analy- than the acceptable range, and zone D contained POC ses. Error codes were randomly displayed by the meter concentrations that would result in treating a sheep but were not documented. The manufacturer’s instruc- when the LBA BHB concentration was within the ac- tions recommended verification that the meter had ceptable range. These zones were established with the been properly calibrated and retesting of samples with Table 1—Median (interquartile range) blood concentrations of BHB and glucose as measured with a POC meter and serum concentrations as measured with an LBA in 36 pregnant mixed-breed ewes with different feeding regimens* (12 sheep/group).
BHB (mmol/L) Glucose (mg/dL)
No. of No. of samples samples Animals tested POC LBA tested POC LBA All 461 0.30 (0.10–0.70) 0.42 (0.28–0.73) 446 42 (34–54) 54 (48–61) Group 1* 161 0.30 (0.20–0.70) 0.44 (0.27–0.74) 158 43 (35–54) 55 (50–62) Group 2* 200 0.30 (0.10–0.70) 0.38 (0.27–0.69) 193 41 (33–53) 52 (47–60) Group 3* 100 0.30 (0.20–0.75) 0.47 (0.30–0.76) 95 40 (33–55) 53 (45–62) *Sheep in group 1 were fed 100% of their nutrient requirements, those in group 2 were fed 50% of their nutrient requirements from 28 through 78 dGA, and those in group 3 were fed 50% of their nutrient require- ments from 28 dGA through parturition. The LBA was considered the more accurate method. All median values differ significantly (P < 0.001) between devices.
AJVR, Vol 74, No. 8, August 2013 1061
Unauthenticated | Downloaded 06/26/23 02:07 PM UTC
12-12-0397r.indd 1061 7/18/2013 2:41:11 PM
a new test strip. Following these recommendations led ± 0.14 mmol/L), and the low BHB concentration CV, to resolution of the error code display and obtainment obtained through use of ovine blood samples, was not of valid measurements, which were used in the statis- relevant because the mean was close to 0 mmol/L (0.12 tical analyses. ± 0.042 mmol/L). The CV reported by the meter manu- facturer was 3.8% for the mid BHB concentration range Precision testing—Precision testing of the POC (2.36 ± 0.09 mmol/L) and 3.1% for the high range (6.32 meters revealed a smaller CV for BHB concentrations ± 0.2 mmol/L), and a CV was not reported for the low than for glucose concentrations, and the CVs were BHB concentration range (0.34 ± 0.03 mmol/L). The similar to those reported by the meter manufacturer.10 CV for the glucose concentration (sheep blood sample) The high BHB concentration CV, obtained through use was 5.8% (52.5 ± 3.0 mg/dL), compared with the man- of bovine blood samples, was 3.56% (mean ± SD, 3.92 ufacturer’s reported CV of 6% for the low glucose con- centration range (43.9 ± 2.3 mg/dL).11 POC meter and LBA comparisons—The POC me- ter results for blood BHB and glucose concentrations differed significantly (P < 0.001) from the LBA results for serum concentrations overall and by feeding group (Table 1). The correlation between the POC meter and LBA values was 0.934 (P < 0.001) for BHB concentra- tion and 0.620 (P < 0.001) for glucose concentration. The percentage difference in values obtained with the POC meter and LBA indicated the POC meter re- sults for blood BHB and glucose concentrations were quite different from the LBA results. For BHB values, 13%, 21%, and 37% of POC meter concentrations were within 5%, 10% and 20% above or below the LBA values, respectively. Because meter performance criteria have not been established for blood BHB concentrations, samples were analyzed further when the LBA-measured BHB concentrations were > 0.8 mmol/L, which would be more clinically relevant for making treatment deci- sions. In that further analysis, 37%, 48%, and 71% of POC meter values were within 5%, 10%, and 20% of the LBA values, respectively. Patterns were similar to those of BHB for blood glucose concentrations as measured with the POC meter, for which 11%, 23%, and 44% of the POC meter glucose values were within 5%, 10%, and 20% above or below the LBA values. The Bland-Altman difference plot revealed that the mean difference in BHB concentrations between de- vices was small, with narrow limits of agreement (Fig- ure 1). Visual examination of the plot revealed a small negative constant bias in the BHB data and an increase in data scatter as BHB concentration increased. Eleven upper and 5 lower outlier values were identified, and Figure 1—Error grid analysis (A) and Bland-Altman plot (B) com- the blood samples to which these pertained had been paring blood BHB concentrations measured with a dual-purpose collected across multiple dates; however, 3 were from POC meter with serum BHB concentrations measured with an the same collection date and 3 others were from the LBA in samples obtained from pregnant mixed-breed ewes with same sheep over 3 weeks. When the plot was examined different feeding regimens. Sheep in group 1 (triangles; n = 12) were fed 100% of their nutrient requirements, those in group 2 by feeding group, a small positive bias was evident in (squares; 12) were fed 50% of their nutrient requirements from the data points for the group fed 50% of their nutri- 28 through 78 dGA, and those in group 3 (circles; 12) were fed ent requirements from 28 through 78 dGA and a nega- 50% of their nutrient requirements from 28 dGA through parturi- tion. A—The error grid comprises 4 zones. Zone A represents tive bias was identified in the data points for the group POC meter BHB concentrations within 20% of the LBA concen- fed 50% of their nutrient requirements from 28 dGA trations as well as when the POC BHB concentration was < 0.8 through parturition. The difference plot for glucose mmol/L and the corresponding LBA BHB concentration was also < 0.8 mmol/L. Zone B contains POC meter values that were concentration showed a larger mean difference with > 20% of LBA values but would result in a benign treatment deci- wide limits of agreement, negative constant bias, and sion. Zone C contained POC meter BHB concentrations that would more scatter in the data than was evident for BHB (Fig- result in not treating a sheep when the LBA BHB cocentration was above the acceptable range, and zone D contained POC concen- ure 2). Twelve upper and 3 lower outliers were identi- trations that would result in treating a sheep when the LBA BHB fied; 6 values pertained to blood samples collected on concentration was within the acceptable range. B—Bland-Altman the same day, and 3 values pertained to samples col- difference plots were prepared, accounting for multiple observa- lected on a day when severe hemolysis was noticed in 3 tions per test subject. The dashed lines indicate the 95% limits of agreement (± 1.96 SD of the mean difference), and the solid line discarded samples. Outlier data points represented ap- indicates the mean difference between methods. proximately 3% of the samples collected.
1062 AJVR, Vol 74, No. 8, August 2013
Unauthenticated | Downloaded 06/26/23 02:07 PM UTC
12-12-0397r.indd 1062 7/18/2013 2:41:11 PM
provided an opportunity to evaluate the POC meter in various environmental conditions and ewe nutritional states, which would be difficult to do in a nonexperi- mental setting because most producers would inter- vene to reduce the effects observed in these ewes. The POC meter was more appropriate for measuring blood BHB concentrations than blood glucose concentra- tions in sheep. The results were poorer than those in a pre- viously reported study,8 but the differences between stud- ies may have been attributable to the larger sample size in the present study, ewe metabolic effects from malnutrition during gestation, environmental conditions, differences in sample handling, statistical techniques, or LBAs used. An- other explanation is potential variations between the POC meters used, given that they are marketed with different names depending on the country of origin.8 The LBA and sample handling details were not described for the earlier study,8 and blood samples in that study were obtained at only 1 point during gestation or lactation. In addition, the use of the Bland-Altman plots and error grid plots in the study provided more clinically relevant results regarding meter performance, compared with the other study.12–14 The precision of the POC meter was compared with the manufacturer’s reported precision because we were unable to identify any reports of studies in which the meter had been evaluated. The CV was similar to the manufacturer’s reported results.10,11 Glucose meter performance criteria have been established by several testing and health organizations and range from 5% to 20%.15 Similar criteria are not available for POC meter performance when measuring blood BHB concentra- tion. When used as a POC glucometer, the results for the device did not meet the criteria established by the testing organizations. This was not surprising because Figure 2—Error grid analysis (A) and Bland-Altman plot (B) com- paring blood glucose concentrations measured with a dual-pur- the meter is manufactured for measurement of analytes pose POC meter with serum glucose concentrations measured in human blood samples. The BHB performance results with an LBA in samples from pregnant mixed-breed ewes with were best when compared with the LBA results in a different feeding regimens. A—The error grid comprises 4 zones. more clinically relevant range (> 0.8 mmol/L) rather Zone A represents POC meter glucose concentrations within 20% of the LBA concentrations as well as when the POC con- than across the entire range. Some of the variation at centration was ≤ 50 mg/dL when the LBA concentration was also the lower ranges is likely attributable to reporting dif- ≤ 50 mg/dL. Zone B contains POC meter concentrations that ferences, considering that the LBA results were provid- were > 20% of LBA concentrations but would result in a benign treatment decision or no treatment decision. Zone C contained ed to 2 decimal places, whereas the POC meter results POC meter glucose concentrations that would result in treating were provided to only 1 decimal place. a sheep when the LBA glucose concentration was within the The Bland-Altman difference plots also showed acceptable range, and zone D contained POC glucose concentra- tions that would result in not treating a sheep when the LBA was that the POC meter compared more favorably with the lower than the acceptable range. See Figure 1 for remainder of LBA when used to measure BHB concentration rather key. than glucose concentration. The systematic bias evi- Modified error grid plots developed to hypotheti- dent with both meters used could be addressed through cally evaluate results of clinical decision making based application of a correction factor to the results or es- on use of the POC meter versus the LBA results showed tablishment of meter-specific reference limits. However, that 97% of BHB concentrations were within zones A the significant proportional bias in the blood glucose (93%) and B (4%), 2% were within zone C, and 1% was results would not be correctable. The increase in scatter within zone D (Figure 1). For glucose concentrations, at higher BHB and glucose concentrations was likely 72% were within zones A (68%) and B (4%), 27% were due to the smaller number of blood samples in these within zone C, and 1% was within zone D (Figure 2). ranges. Modified error grid plots revealed that the POC Discussion meter was clinically acceptable for use according to our criteria for measuring blood BHB concentrations in The present study was conducted to evaluate the sheep, whereas the measurement function for glucose performance of a dual-purpose POC meter for mea- concentrations was not acceptable. No differences were surement of blood BHB and glucose concentrations in 3 observed in values between nutritional status groups for groups of sheep during gestation and lactation. The ma- either analyte. The plots were useful in identifying risk ternal malnutrition study from which they originated for over- or undertreatment or inappropriate treatment
AJVR, Vol 74, No. 8, August 2013 1063
Unauthenticated | Downloaded 06/26/23 02:07 PM UTC
12-12-0397r.indd 1063 7/18/2013 2:41:12 PM
decisions. Use of the POC meter would result in a large was not measured; however, an unreported preliminary number of decisions resulting in overtreatment when nutritional energy restriction study revealed no signifi- used to measure blood glucose concentrations. The risk cant effect on Hct in the sheep. All the ewes used in the of overtreatment would not likely increase illness sever- study were visibly examined for signs of anemia (pale ity but would increase the cost to the producer because mucous membranes and sclera) during sample collec- low POC meter glucose concentrations would most tion, and no gross evidence of anemia was observed; commonly lead to oral or parenteral administration of however, the presence of mild anemia may have been glucose substrates to increase blood glucose concentra- overlooked. Some serum samples had evidence of mild tion. Diagnosing hyperglycemia rather than normogly- to moderate lipemia, but there was no apparent rela- cemia would not likely result in different intervention tionship with nutritional status or gestational stage be- strategies because it would be difficult to identify the cause concentrations would fluctuate among blood col- cause of an isolated high glucose concentration. An in- lection points and sheep. Similar factors have not been crease in blood glucose concentration can result from reported as affecting blood BHB concentration, and the stress induced by capture and blood collection, actual manufacturer reports that Hct, various medications, meter characteristics, medical disorders, or typical in- and high triglycerides concentration do not affect BHB traindividual variation. results.10 The POC meter was robust and easy to use in vari- The POC meter used in the present study was ous conditions; however, the low end of the operating designed for use in humans, in which blood glucose range of 10°C would prevent its use by producers and is known to be distributed evenly between the RBCs veterinarians working with pregnant ewes in colder cli- and plasma; however, in sheep, 83% of the glucose mates. We found that keeping the meter and test strips is distributed in the plasma and the remainder in the in a small cooler with a chemical hand warmer main- RBCs.19,20 Although glucose utilization by the RBCs tained the proper operating temperature. Because of the continues until the serum is separated from the cells,21 large number of samples tested at each time point, we the low serum glucose concentrations observed in the used 2 identical POC meters to reduce the need to re- study sheep were surprising. Blood samples were kept calibrate between BHB and glucose testing. One POC on ice to limit cell glucose utilization until the samples meter would be adequate to measure both analytes in could be centrifuged, and serum was removed from the most farm or veterinary settings. blood clot within 3 hours after collection. We identified The choice to evaluate the meters in various con- a significant difference in LBA-measured glucose con- ditions that many sheep producers would experience centrations between the control sheep fed 100% of their resulted in many potential sources of variation in the nutrient requirements and those fed a restricted amount present study. Operator variables included differences (data not shown); however, serum glucose concentra- in sample collection and handling technique and risk tions in the control group were still quite lower than of test strip contamination from dirt, lanolin, blood, or the lower reference limit of our laboratory (70 mg/dL). other factors. Despite the use of warming containers Given the study findings, the laboratory’s reference lim- for the meters and test strips, the temperature and en- its for adult sheep may not be appropriate for pregnant vironmental variations potentially could have affected ewes. Our methods for sample collection and storage meter and strip performance. Multiple lots of test strips after serum separation should not have dramatically al- were required, adding another potential source of varia- tered the blood glucose and BHB concentrations.16 tion in results. To reduce this variation, the meters were The 3 groups of sheep that differed in nutritional calibrated prior to every collection day and before new status were used to determine whether feed restriction boxes of test strips were used. In addition, the expira- would sufficiently alter ewe metabolism to affect blood tion date of the test strips was heeded and extra test glucose and BHB concentrations. No evidence of such strips were stored in a climate-controlled environment an alteration was observed; however, the feed-restricted in their original packaging. sheep had BHB and glucose results that would typically The ewes were not accustomed to being handled require treatment intervention to prevent development nor confined, contributing to the difficulty in obtaining of pregnancy toxemia. blood samples and increasing the risk of hemolysis or Overall, the dual-purpose POC meter appeared to blood clots forming in samples prior to testing. Hemo- be adequate for measurement of blood BHB concentra- lysis is easy to produce in sheep blood samples and is tion in sheep and useful in the diagnosis of ketosis and known to increase serum glucose concentrations.16 The pregnancy toxemia. Its use could lead to earlier disease study blood samples had various degrees of hemoly- detection and intervention as well as lower treatment sis, which may have affected POC meter performance; and management costs for producers. The meter was however, the effect of hemolysis on the results was not not as reliable for measurement of blood glucose con- evaluated. centration. Therefore, glucose results should be inter- preted with caution to avoid inappropriate medical Other factors such as anemia, hypertriglyceride- management of suspected hypoglycemic or hypergly- mia, hyperproteinemia, handling technique, ambient cemic sheep. temperature, altitude, and pregnancy-induced blood volume expansion could have also affected blood glu- a. Precision Xtra Blood Glucose & Ketone Monitoring System, Ab- cose results obtained with the POC meter.17,18 The ef- bott Diabetes Care, Abbott Park, Ill. fects of these factors on blood glucose concentration b. Grabber Hand Warmers, YSC Inc, Grand Rapids, Mich. have not been evaluated in combination with gesta- c. Serum Vacutainer glass blood collection tubes, Becton, Dickin- tional stage and fetal numbers in sheep. Hematocrit son and Co, Franklin Lakes, NJ.
1064 AJVR, Vol 74, No. 8, August 2013
Unauthenticated | Downloaded 06/26/23 02:07 PM UTC
12-12-0397r.indd 1064 7/18/2013 2:41:12 PM
d. Roche Hitachi 917 Blood Chemistry Analyzer, Hitachi Ltd, To- 10. Precision Xtra blood β-ketone test strips [package insert]. Al- kyo, Japan. ameda, Calif: Abbott Diabetes Care Inc, 2006. e. MedCalc, version 12.3, MedCalc Software, Mariakerke, 11. Precision Xtra blood glucose test strips [package insert]. Ala- Belgium. meda, Calif: Abbott Diabetes Care Inc, 2009. 12. Bland J, Altman D. Agreement between methods of measure- ment with multiple observations per individual. J Biopharm Stat References 2007;17:571–582. 1. Brozos C, Mavrogianni VS, Fthenakis GC. Treatment and con- 13. Clarke W, Cox D, Gonder-Frederick L, et al. Evaluating clinical trol of peri-parturient metabolic diseases: pregnancy toxemia, accuracy of systems for self-monitoring of blood glucose. Diabe- hypocalcemia, hypomagnesemia. Vet Clin North Am Food Anim tes Care 1987;10:622–628. Pract 2011;27:105–113. 14. Bland JM, Altman DG. Statistical methods for assessing agree- 2. Bulgin M. Diseases of the periparturient ewe. In: Youngquist RS, ment between two methods of clinical measurement. Lancet Threlfall WR, eds. Current therapy in large animal theriogenol- 1986;327:307–310. ogy. 2nd ed. St Louis: WB Saunders Co, 2007;695–700. 15. Tonyushkina K, Nichols JH. Glucose meters: a review of techni- 3. Federici MO, Benedetti MM. Ketone bodies monitoring. Diabe- cal challenges to obtaining accurate results. J Diabetes Sci Tech- tes Res Clin Pract 2006;74(suppl 2):S77–S81. nol 2009;3:971–980. 4. Laffel L. Ketone bodies: a review of physiology, pathophysiology 16. Morris JD, Fernandez JM, Chapa AM, et al. Effects of sample and application of monitoring to diabetes. Diabetes Metab Res handling, processing, storage, and hemolysis on measurements Rev 1999;15:412–426. of key energy metabolites in ovine blood. Small Ruminant Res 5. Oetzel GR. Monitoring and testing dairy herds for metabolic 2002;43:157–166. disease. Vet Clin North Am Food Anim Pract 2004;20:651–674. 17. Ginsburg BH. Factors affecting blood glucose monitor- 6. Jones M, Miesner MD, Baird AN, et al. Diseases of the urinary ing: sources of errors in measurement. J Diabetes Sci Technol system. In: Pugh DG, Baird AN, eds. Sheep and goat medicine. 2009;3:903–913. 2nd ed. St Louis: WB Saunders Co, 2012;325–360. 18. Hytten F. Blood volume changes in normal pregnancy. Clin Hae- 7. Iwersen M, Falkenberg U, Voigtsberger R, et al. Evaluation of matol 1985;14:601–612. an electronic cowside test to detect subclinical ketosis in dairy 19. Coldman MF, Good W. The distribution of sodium, potassium cows. J Dairy Sci 2009;92:2618–2624. and glucose in the blood of some mammals. Comp Biochem 8. Panousis N, Brozos C, Karagiannis I, et al. Evaluation of Pre- Physiol 1967;21:201–206. cision Xceed meter for on-site monitoring of blood [beta]- 20. MacKay E. The distribution of glucose in human blood. J Biol hydroxybutyric acid and glucose concentrations in dairy sheep. Chem 1932;97:685–689. Res Vet Sci 2012;93:435–439. 21. Zhang DJ, Elswick RK, Miller WG, et al. Effect of serum-clot 9. Voyvoda H, Erdogan H. Use of a hand-held meter for detecting contact time on clinical chemistry laboratory results. Clin Chem subclinical ketosis in dairy cows. Res Vet Sci 2010;89:344–351. 1998;44:1325–1333.
AJVR, Vol 74, No. 8, August 2013 1065
Unauthenticated | Downloaded 06/26/23 02:07 PM UTC