You are on page 1of 20

chapter

7
Radiation Detectors
When radiations from a radioactive material radiation and secondary ionization from δ
pass through matter, they interact with rays (see Chapter 6, Section A.1). The elec-
atoms and molecules and transfer energy to trons produced by ionization are attracted to
them. The transfer of energy has two effects: the positive electrode and the ionized atoms
ionization and excitation. Ionization occurs to the negative electrode, causing a momen-
when the energy transferred is sufficient to tary flow of a small amount of electrical
cause an orbital electron to be stripped away current.
from its parent atom or molecule, thus creat- Gas-filled detectors include ionization
ing an ion pair (a negatively charged electron chambers, proportional counters, and Geiger-
and a positively charged atom or molecule). Müller (GM ) counters. The use of these detec-
Excitation occurs when electrons are per- tors in nuclear medicine is somewhat limited
turbed from their normal arrangement in an because their stopping power and detection
atom or molecule, thus creating an atom or efficiency for x rays and γ rays are quite low;
molecule in an excited state. Both of these however, they find some use for applications
processes are involved in the detection of in which detection efficiency is not a major
radiation events; however, ionization is the factor and for detection and measurement
primary event, and hence the term ionizing of nonpenetrating, particle-type radiations.
radiation is used frequently when referring Some of their applications are discussed in
to the emissions from radioactive material. Chapters 12 and 23.
Radiation interactions were discussed in
detail in Chapter 6. In this chapter, we 2. Ionization Chambers
describe the basic principles of radiation In most ionization chambers, the gas between
detectors used in nuclear medicine. the electrodes is air. The chamber may or may
not be sealed from the atmosphere. Many dif-
ferent designs have been used for the elec-
trodes in an ionization chamber, but usually
A. GAS-FILLED DETECTORS
they consist of a wire inside of a cylinder or a
pair of concentric cylinders.
1. Basic Principles For maximum efficiency of operation, the
Most gas-filled detectors belong to a class of voltage between the electrodes must be suf-
detectors called ionization detectors. These ficient to ensure complete collection of ions
detectors respond to radiation by means of and electrons produced by radiation within
ionization-induced electrical currents. The the chamber. If the voltage is too low, some of
basic principles are illustrated in Figure 7-1. the ions and electrons simply recombine with
A volume of gas is contained between two one another without contributing to electrical
electrodes having a voltage difference (and current flow. Figure 7-2 shows the effect of
thus an electric field) between them. The voltage difference between the electrodes on
negative electrode is called the cathode, the the electrical current recorded by an ioniza-
positive electrode the anode. The electrodes tion chamber per ionizing radiation event
are shown as parallel plates, but they may detected. Recombination occurs at low volt-
be a pair of wires, concentric cylinders, and ages (recombination region of the curve). As
so forth. Under normal circumstances, the the voltage increases there is less recombina-
gas is an insulator and no electrical current tion and the response (electrical current)
flows between the electrodes. However, radi- increases. When the voltage becomes suffi-
ation passing through the gas causes ioniza- cient to cause complete collection of all of the
tion, both direct ionization from the incident charges produced, the curve enters a plateau
87
88 Physics in Nuclear Medicine

Voltage source
 

 Anode
Current
Air or measuring
Incident other device
ionizing e
e gas
radiation e
 e e I





 Cathode

FIGURE 7-1 Basic principles of a gas-filled detector. Electrical charge liberated by ionizing radiation is collected by
positive (anode) and negative (cathode) electrodes.

Saturation
region
Recombination
region
Amplitude of output pulse

Saturation
voltage, Vs

Applied voltage
FIGURE 7-2 Voltage response curve (charge collected vs. voltage applied to the electrodes) for a typical ionization
chamber. In usual operation, applied voltage exceeds saturation voltage Vs to ensure complete collection of liberated
charge.

called the saturation region. The voltage at ionization event in air is approximately
which the saturation region begins is called 34 eV.* Thus a 1-MeV β particle, for example,
the saturation voltage (Vs). Typically, Vs ≈ causes approximately (106/34) ≈ 3 × 104 ioniza-
50-300 V, depending on the design of the tions in air and releases a total amount of
chamber. Ionization chambers are operated at electrical charge of only approximately 3 ×
voltages in the saturation region. This ensures 10−15 coulombs.
a maximum response to radiation and also
that the response will be relatively insensi- * The average energy expended in producing a single
tive to instabilities in the voltage applied to ionization event is symbolized by W. This is not the same
the electrodes. as the average energy required to ionize an air molecule,
The amount of electrical charge released in but is the average energy expended per ionization by the
ionizing particle, including both ionization and excitation
an ionization chamber by a single ionizing effects. This is discussed in detail in Chapter 6, Section
radiation event is very small. For example, A.4. Values of W for some detector materials are listed in
the energy expended in producing a single Table 7-1.
7 • Radiation Detectors 89

TABLE 7-1
SOME PROPERTIES OF DETECTOR
MATERIALS USED AS IONIZATION
DETECTORS

Ge(Li)
Si(Li) or Ge CdTe* Air
ρ(g/cm )3
2.33 5.32 6.06 0.001297
Z 14 32 48 & 52 ~7.6

W(eV) 3.6 2.9 4.43 33.7
CdTe, cadmium telluride; Ge, germanium; Li, lithium; Si,
silicon.
*Cadmium zinc telluride (CZT) is CdTe in which some of
the Te atoms (typically 20%) are replaced by zinc atoms.
CZT has properties similar to CdTe.

Average energy expended per electron-hole pair created
or per ionization.

FIGURE 7-3 A battery-powered radiation survey meter.


An ionization chamber is contained in the base of the
Because of the small amount of electrical unit, with the entrance window on the bottom face of the
charge or current involved, ionization cham- device (not shown). The meter indicates radiation level.
bers generally are not used to record or count The rotary switch is used to select different scale factors.
(Courtesy Ludlum Measurements, Inc., Sweetwater, TX.)
individual radiation events. Instead, the total
amount of current passing through the
chamber caused by a beam of radiation is
measured. Alternatively, the electrical charge
released in the chamber by the radiation are discussed in Chapter 23. A typical survey
beam may be collected and measured. meter can measure exposure rates down to
Small amounts of electrical current are approximately 1 mR/hr or air kerma rates
measured using sensitive current-measuring down to approximately 10 µGy/hr.
devices called electrometers. Two devices con- Dose calibrators are used to assay activity
sisting of ionization chambers and electrom- levels in syringes, vials, and so forth contain-
eters in nuclear medicine are survey meters ing materials that are to be administered to
and dose calibrators. A typical ionization patients. Unlike other types of ionization
chamber survey meter is shown in Figure 7-3. chambers discussed in this section, dose cali-
The survey meter is battery operated and por- brators employ sealed and pressurized cham-
table. The ionization chamber consists of an bers filled with argon gas. This eliminates the
outer cylindrical electrode (metal or graphite- effect of changing barometric pressure on
coated plastic) with a wire electrode running output readings. Dose calibrators typically
down its center. There is often a pro­tective are calibrated to read directly in units of
cap on the end of the chamber for most mea- activity (becquerels or curies), with switches
surements; however, it is removed for mea- to set the display for different radionuclides.
surement of nonpenetrating radiations such Dose calibrators are discussed in detail in
as α particles, β particles, and low-energy Chapter 12, Section D.1.
(10 keV) photons. A device that records total charge collected
Survey meters are used to monitor radiation over time is the pocket dosimeter. The basic
levels for radiation protection purposes (see principles are illustrated in Figure 7-4. The
Chapter 23, Section E). Ionization current is ionization chamber electrodes are a central
displayed on a front-panel meter. Many older charging electrode and the outside case of the
units are calibrated to read traditional units dosimeter. They are insulated electrically
of exposure rate in roentgens per hour (R/hr) from one another and form an electrical capac-
or mR/hr. Newer units are calibrated to read itor. The capacitor is first charged to a refer-
Systeme International units of air kerma in ence voltage V by connecting the charging rod
grays per hour (Gy/hr), mGy/hr, and so forth, to a separate charging unit. If the capacitance
or have a switch-selectable option for choosing between the charging electrode and the case
between the two systems of units. The defini- is C, the charge stored on the capacitor is Q =
tions and relationships between these units V × C. When the chamber is exposed to
90 Physics in Nuclear Medicine

Outside
case

C
 Charging
electrode

 Insulator

FIGURE 7-4 Schematic representation of a pocket dosimeter.

radiation, electrical charge ΔQ is collected by dosimeters are suitable for measuring radia-
the electrodes, discharging the capacitor. The tion exposures down to approximately 10 mR
voltage change across the capacitor is mea- (air kerma of 0.1 mGy) to an accuracy of
sured and is related to the amount of electrical approximately 20%.
charge collected by the ionization chamber A basic problem with ionization chambers
electrodes (ΔQ = ΔV × C). is that they are quite inefficient as detectors
Pocket dosimeters are used in nuclear for x rays and γ rays. Only a very small per-
medicine to monitor radiation levels for radi- centage (< 1%) of x rays or γ rays passing
ation protection purposes. A typical system is through the chamber actually interact with
shown in Figure 7-5. The ionization chamber and cause ionization of air molecules. Indeed,
is contained in a small metal or plastic cyl- most of the electrical charge released in an
inder (~1.5 cm diameter × 10 cm long) that ionization chamber by photon radiations
can be clipped to a shirt pocket or collar. comes from secondary electrons knocked loose
Electrodes recessed into one end of the from the walls of the chamber by the incident
chamber are used to connect the dosimeter radiations rather than by direct ionization of
to a separate charger unit to charge up the air molecules. The relatively low detection
capacitor to the reference voltage. Voltage on efficiency of ionization chambers is not a
the capacitor causes a fine wire within the serious limitation in the applications described
chamber to be deflected. The position of the earlier; however, it precludes their use for
wire changes as the voltage on the capacitor most other applications in nuclear medicine,
changes. The wire is observed through a such as imaging.
viewing window at one end of the chamber. Two additional problems with ionization
Its position is read against a scale that has chambers should be noted. The first is that for
been calibrated in terms of the total radia- x rays and γ rays, their response changes with
tion recorded by the chamber, usually in photon energy because photon absorption in
units of air kerma (gray) or exposure (roent- the gas volume and in the chamber walls (i.e.,
gens) (see Chapter 23, Section E). Pocket detection efficiency) and relative penetration

FIGURE 7-5 Pocket dosimeter with charging system. (Courtesy Ludlum Measurements Inc., Sweetwater, Tx.)
7 • Radiation Detectors 91

1.2
End-cap off

1.0

Exposure rate (indicated/actual)

0.8

0.6

End-cap on
0.4

0.2

0.0
10 100 1000
Photon energy (keV)
FIGURE 7-6 Energy response curve for a typical ionization chamber survey meter with and without a removable
protective end cap.

of photons through the chamber walls are increased to a sufficiently high value, the
both energy-dependent processes. Figure 7-6 electrons liberated by radiation gain such
shows a typical energy-response curve for a high velocities and energies when accelerated
survey meter. A second problem is that in toward the positive electrode that they cause
unsealed chambers the density of the air in additional ionization in collisions with other
the chamber, and hence its absorption effi- atoms in the gas. These electrons in turn can
ciency, changes with atmospheric pressure (ρ cause further ionization and so on. This
∝ P ) and temperature (ρ ∝ 1/T ). Most cham- cascade process is called the Townsend ava-
bers are calibrated to read accurately at sea- lanche or the gas amplification of charge. The
level pressure (Pref = 1.013 N/m2 = 760 mm Hg) factor by which ionization is increased is
and average room temperature (Tref = 22°C = called the gas amplification factor. This factor
295K). For other temperatures T and pres- increases rapidly with applied voltage, as
sures P the chamber reading must be cor- shown in Figure 7-7. The gas amplification
rected (multiplied) by a temperature-pressure factor may be as high as 106, depending on
correction factor the chamber design and the applied voltage.
Detectors that operate in the ascending
CTP = ( Pref × T ) /( P × Tref ) (7-1)
portion of the curve shown in Figure 7-7 are
Temperature must be expressed on the Kelvin called proportional counters. In this region,
scale in this equation (K = °C + 273). The the ionization caused by an incident radiation
correction is significant in some cases, for event is multiplied (amplified) by the gas
example, at higher elevations (P ≈ 0.85 N/m2 amplification factor. The total amount of
≈ 640 mm Hg at 1600-meter elevation). Note charge produced is equal to the number of
that temperature-pressure corrections are not ionizations caused by the primary radiation
required with sealed chambers, such as in event (at 34 eV/ionization in air) multiplied
most dose calibrators. A defective seal on such by the amplification factor. Thus the total
an instrument obviously could lead to errone- charge produced is proportional to the total
ous readings. amount of energy deposited in the detector by
the detected radiation event.
3. Proportional Counters Actually, proportional counters are not
In an ionization chamber, the voltage between simply ionization chambers operated at high
the electrodes is sufficient only to collect voltages but are specially constructed cham-
those charges liberated by direct action of the bers designed to optimize the gas amplifica-
ionizing radiations. However, if the voltage is tion effect, both in terms of the amount of
92 Physics in Nuclear Medicine

Amplitude of output pulse

Ionization
chamber
region
Proportional
counter
region

Applied voltage
FIGURE 7-7 Voltage response curve for a proportional counter. With increasing applied voltage, the charge collected
increases because of the gas amplification effect.

amplification and the uniformity of this are shown in Figure 7-8. The center wire
amplification within the chamber. In particu- (anode) is maintained at a high positive
lar, proportional counters are filled with gases voltage relative to the outer cylindrical elec-
that allow easy migration of free electrons, trode (cathode). The outer electrode may be a
because this is critical for the amplification metal cylinder or a metallic film sprayed on
effect. Common fill gases are the noble gases, the inside of a glass or plastic tube. Some GM
with argon and xenon being the most popular. counters have a thin radiation entrance
The major advantage of proportional coun- window at one end of the tube. The cylinder
ters versus ionization chambers is that the of the tube is sealed and filled with a special
size of the electrical signal produced by an gas mixture, typically argon plus a quenching
individual ionizing radiation event is much gas (discussed later).
larger. They are, in fact, useful for detecting When ionization occurs in a GM counter,
and counting individual radiation events. electrons are accelerated toward the center
Furthermore, because the size of an individ- wire. Gas amplification occurs in the GM
ual current pulse is proportional to the counter as in a proportional counter. In addi-
amount of energy deposited by the radiation tion to ionizing gas molecules, the accelerat-
event in the detector, proportional counters ing electrons also can cause excitation of gas
can be used for energy-sensitive counting, molecules through collisions. These excited
such as to discriminate between radiation gas molecules quickly (~10−9 sec) return to
events of different energies on the basis of the ground state through the emission of
electrical pulse size (see Chapter 10). They photons at visible or ultraviolet (UV) wave-
are still inefficient detectors for higher energy lengths. If a UV photon interacts in the
x rays and γ rays. Consequently, they find gas, or at the cathode surface by photoelec-
very limited use in nuclear medicine. Propor- tric absorption (see Chapter 6, Section C.2),
tional counters are used mostly in research this releases another electron, which can
applications for measuring nonpenetrating trigger a further electron avalanche as it
radiations such as α particles and β particles. moves toward the anode (see Fig. 7-8). In
A practical application is discussed in Chapter this way, an avalanche ionization is propa-
12, Section D.2. gated throughout the gas volume and along
the entire length of the center wire.
4. Geiger-Müller Counters As the avalanche progresses, the electrons,
A Geiger-Müller (GM) counter is a gas-filled being relatively light, are quickly collected,
detector designed for maximum gas amplifi- but the heavy, slow-moving positive ions are
cation effect. The principles of a GM counter not. Eventually, a “hose” of slow-moving
7 • Radiation Detectors 93

Positive ions

Primary electron
Secondary electrons

Incident
radiation
Anode

Insulator
RL Signal

Outer cylinder
Thin window (cathode) V
 

Primary electron track

Secondary electron

Electron avalanche

Ultraviolet radiation

FIGURE 7-8 Operating principles of a Geiger-Müller counter. The incoming radiation produces ion pairs by direct
ionization and through secondary fast electrons (δ rays) created in the ionization process. These ion pairs are then
multiplied by an avalanche process that in turn triggers further avalanches through the emission of ultraviolet radia-
tion. This process is terminated when a sufficient number of positive ions collect around the anode, effectively reducing
the electric field experienced by the electrons owing to charge buildup at the anode.

positive charges is formed around the center Once the avalanche has terminated in a
wire. The avalanche then terminates because GM counter, an additional problem arises.
the positive ions reduce the effective electric The positive ion cloud moves toward the outer
field around the anode wire, eventually drop- electrode. When the ion cloud is very close to
ping it below the level required for gas the outer electrode, electrons are pulled out
multiplication. from it to neutralize the positive ions. Some
The avalanche ionization in a GM tube of these electrons enter higher-energy orbits
releases a large and essentially constant of the positive ions; when they eventually
quantity of electrical charge, regardless of drop into the lower-energy orbits, UV radia-
voltage applied to the tube (Fig. 7-9) or the tion is emitted. This can cause the release of
energy of the ionizing radiation event. The more electrons from the outer wall and set off
gas amplification factor may be as high as another avalanche. Thus if no precautions are
1010. The large electrical signal is easily taken, a single ionizing radiation event can
detected with electronic circuits. Thus a GM cause the GM counter to go into a pulsating
counter, like a proportional counter, can be series of discharges.
used to detect and count individual ionizing This problem is prevented by the introduc-
radiation events. However, because the size tion of a quenching gas into the GM counter
of the electrical signal output is constant, gas mixture. Such GM counters are called
regardless of the energy of the radiation self-quenched. Effective quenching gases have
detected, a GM counter cannot be used to three properties: First, they tend to give up
distinguish between radiation events of dif- electrons easily. When the positive ion cloud
ferent energies. is formed, molecules of the quenching gas
94 Physics in Nuclear Medicine

Geiger-Müller
region
Amplitude of output pulse

Ionization
chamber
region

Proportional
counter
region

Applied voltage
FIGURE 7-9 Voltage response curve (pulse amplitude vs. applied voltage) for a Geiger-Müller counter.

neutralize other ions by donating electrons to applied to the counter electrodes. Figure 7-10
them. The ion cloud is thus converted into shows the results of such an experiment. This
ionized molecules of quenching gas. Second, curve is called the counting curve or plateau
when the quenching gas molecules are neu- curve of the GM counter. As the high voltage
tralized by electrons entering higher energy is increased, the counting rate increases
orbits, they deenergize themselves by dissoci- rapidly as more and more of the output pulses
ating into molecular fragments rather than exceed the counter threshold. When the
by emitting UV photons. Third, the quench- voltage is sufficient that essentially all pulses
ing gas molecules are strong absorbers of UV are above the threshold and are counted, a
radiation. Thus the few UV photons that are plateau region is reached. The point at which
released during neutralization of the positive the plateau begins is called the knee of the
ion cloud are quickly absorbed before they can curve. Further increases in voltage may still
set off another avalanche. increase the amplitude of the output pulses;
Commonly used quenching gases include however, the counting rate remains constant
heavy organic vapors (e.g., alcohol) and as the radiation source is constant.*
halogen gases (e.g., Cl2). The organic vapors When the voltage is increased to a very
are more effective quenching agents but have high value, the counting rate again begins to
the disadvantage that their molecular frag- increase. This happens when the voltage is so
ments do not recombine after dissociation. high that spontaneous ionization begins to
Thus an organic quenching gas eventually is occur in the chamber. The curve then enters
used up, typically after approximately 1010 the spontaneous discharge region. GM coun-
radiations have been detected. Halogen gas ters should not be operated in the spontane-
molecules recombine after dissociation and ous discharge region because no useful
thus have an essentially unlimited lifetime in information can be obtained there. Further-
a GM counter. more, if the counter contains an organic
A certain minimum voltage is required quenching gas, it is rapidly used up by the
between the electrodes of a GM counter to spontaneous discharges, thus shortening the
sustain an avalanche ionization and to raise
the amplitude of the pulses to the threshold
of the counting system. This voltage can be
determined by exposing the GM counter to a *Actually, for most GM counters the counting rate
constant source of radiation and observing increases by 1% to 2% per 100 volts in the plateau region.
the counting rate as a function of voltage This is of no practical consequence in nuclear medicine.
7 • Radiation Detectors 95

Spontaneous
discharge

Knee
Counting rate

Plateau

Threshold

Applied voltage
FIGURE 7-10 Counting curve (counting rate from a fixed radiation source vs. applied voltage) for a Geiger-Müller
counter. As voltage increases, pulse amplitude increases above threshold of counting system electronics. When all
events produce a signal above the threshold, a plateau is reached. At very high voltages, spontaneous discharge events
occur within the chamber. These are not caused by radiation events but by electrical breakdown in the gas.

life of the counter. The proper operating


voltage is the plateau region, about one third
the distance from the knee to the spontaneous
discharge region.
GM counters are simple, rugged, and rela-
tively inexpensive radiation detectors. Much
of the early (pre-1950s) work in nuclear medi-
cine was done with GM counters; however,
they have since been replaced for most appli-
cations by other types of detectors. The major
disadvantages of GM counters are low detec-
tion efficiency (< 1%) for γ rays and x rays
and an inability to distinguish between radia-
tion events of different energies on the basis
of pulse size for energy-selective counting
(because all pulses from a GM counter are the
same size).
FIGURE 7-11 Radiation survey meter with an external
GM counters are used mostly in survey Geiger-Müller (GM) “pancake” counter radiation detec-
meters for radiation protection purposes. An tor attachment. In addition to the external detector
example is shown in Figure 7-11. The detector shown, some units have built-in GM counters. (Courtesy
in this survey meter is of the pancake type. Ludlum Measurements, Inc. Sweetwater, TX.)
The entrance window at the end of the counter
tube is a thin layer of mica (0.01-mm thick)
that is sufficiently thin to permit passage of
particles and low-energy photons into the
counter. The rather fragile window is pro- for example, 0.1-mm-thick aluminum or
tected by a wire screen. GM counters designed stainless steel. Many GM counters are pro-
for counting only relatively penetrating radia- vided with removable covers on the entrance
tions, such as γ rays and high-energy β window that can be used to distinguish
particles, have thicker, sturdier windows, between penetrating and nonpenetrating
96 Physics in Nuclear Medicine

radiations by observing the difference between In spite of their apparent advantages, semi-
counting rates with and without the cover in conductor detectors have a number of prob-
place. GM survey meters are more sensitive lems that have limited their use in nuclear
than ionization chamber survey meters, typi- medicine. The first is that both Si and Ge
cally by a factor of approximately 10. (especially Ge) conduct a significant amount
of thermally induced electrical current at
room temperature. This creates a background
B. SEMICONDUCTOR DETECTORS “noise current” that interferes with detection
of radiation-induced currents. Therefore Si
Semiconductor detectors are essentially solid- detectors (usually) and Ge detectors (always)
state analogs of gas-filled ionization cham- must be operated at temperatures well below
bers. Because the solid detector materials room temperature.
used in semiconductor detectors are 2000 to A second problem is the presence of impuri-
5000 times more dense than gases (see Table ties even in relatively pure crystals of Si and
7-1), they have much better stopping power Ge. Impurities (atoms of other elements)
and are much more efficient detectors for x enter into and disturb the regular arrange-
rays and γ rays. ment of Si and Ge atoms in the crystal matrix.
Semiconductor detectors normally are poor These disturbances create “electron traps”
electrical conductors; however, when they are and capture electrons released in ionization
ionized by an ionizing radiation event, the events. This results in a substantial reduction
electrical charge produced can be collected by in the amount of electrical signal available
an external applied voltage, as it is with gas- and limits the thickness of a practical detec-
filled detectors. This principle could not be tor to approximately 1 cm. Because of the
applied using a conducting material for the relatively low atomic numbers of Si and Ge,
detector (e.g., a block of metal) because such this restricts their efficiency for detection of
a material would conduct a large amount of γ rays.
current even without ionizing events. Insula- Two approaches have been used to solve
tors (e.g., glass) are not suitable detector the impurity problem. One is to prepare very
materials either, because they do not conduct pure samples of the detector material. This
even in the presence of ionizing radiation. has been accomplished only with Ge [high-
Hence only semiconductor materials can func- purity germanium (HPGe)] and is, unfortu-
tion as “solid ionization chambers.” nately, quite expensive. Also, the size of pure
The most commonly used semiconductor crystals is limited to approximately 5 cm in
detector materials are silicon (Si) and germa- diameter by 1 cm thick. Detectors made of
nium (Ge). More recently, cadmium telluride HPGe are sometimes called intrinsic Ge detec-
(CdTe) or cadmium zinc telluride (CZT) have tors. A second approach is to deliberately
been used as the detector material in small introduce into the crystal matrix “compensat-
nuclear medicine counting and imaging ing” impurities that donate electrons to fill
devices. Characteristics of these semiconduc- the electron traps created by other impuri-
tor materials are listed in Table 7-1. One ion- ties. Lithium (Li) is commonly used in Si and
ization is produced per 3 to 5 eV of radiation Ge detectors for this purpose. Detectors made
energy absorbed. By comparison, this value of “lithium-doped” materials are called
for gases (air) is approximately 34 eV per ion- lithium-drifted detectors, or Si(Li) or Ge(Li)
ization. Thus a semiconductor detector not detectors. Unfortunately, the process of pre-
only is a more efficient absorber of radiation paring Si(Li) or Ge(Li) crystals is time con-
but produces an electrical signal that is suming and expensive. Crystal sizes are
approximately 10 times larger (per unit of limited to a few centimeters in diameter by
radiation energy absorbed) than a gas-filled approximately 1 cm thick for Si(Li) and
detector. The signal is large enough to permit approximately 5 cm diameter by 5 cm thick
detection and counting of individual radiation for Ge(Li).
events. Furthermore, the size of the electrical An additional problem is that Li ions tend
signal is proportional to the amount of radia- to “condense” within the crystal matrix at
tion energy absorbed. Therefore semiconduc- room temperature, especially in Ge. There-
tor detectors can be used for energy-selective fore Si(Li) and Ge(Li) not only must be oper-
radiation counting. For reasons discussed ated at low temperatures (to minimize
in Chapter 10, Section C.1, they are in fact thermally induced background currents) but
the preferred type of detector for this Ge(Li) detectors must and Si(Li) detectors
application. should also be stored at low temperatures.
7 • Radiation Detectors 97

Liquid nitrogen (T = 77K or −196°C) is used evaporates and the container needs periodic
for detector cooling. Ge(Li) detectors can refilling—typically every 2 to 3 days, de­
be ruined by only an hour or so at room tem­ pending on container size and insulation
perature. Si(Li) detectors can tolerate ele- characteristics.
vated temperatures, but they provide optimum CdTe and CZT (which has properties very
performance if they also are stored at liquid similar to CdTe) are more recently developed
nitrogen temperatures. semiconductor materials that overcome two of
Because of the difficulties inherent in the major disadvantages of Si and Ge: (1) they
Li-drifted detectors, HPGe has become the can be operated at room temperature without
detector material of choice for γ-ray spectros- excessive electronic noise, and (2) their high
copy applications (see Chapter 10), and most atomic number means that even relatively
manufacturers have now stopped producing thin detectors can have good stopping effi-
Ge(Li) detectors. Si(Li) finds applications in ciency for detecting γ rays. Although CdTe
low-energy x-ray and β-particle spectroscopy, and CZT are now being used in some nuclear
in which its low atomic number is not a medicine counting and imaging devices, their
disadvantage. use has generally been restricted to small
Figure 7-12 shows schematically a typical detectors, or detectors comprising multiple
semiconductor detector assembly. The detec- small elements, because of the difficulty and
tor consists of a thin, circular disc of the expense of growing large pieces of CdTe or
detector material [Si(Li), Ge(Li), or HPGe] CZT with the required purity. Additional dis-
with electrodes attached to its opposite faces cussion of their properties for pulse-height
for charge collection. One electrode is a thin spectrometry is presented in Chapter 10,
metal foil fastened to the front surface Section C.1.
(“entrance window”), whereas the other is a
wire or set of wires embedded in the opposite
surface of the crystal. Other detector shapes C. SCINTILLATION DETECTORS
and electrode configurations also are used.
Figure 7-12 also shows in cross-section an
apparatus used to cool the crystal with liquid 1. Basic Principles
nitrogen. A “coldfinger” extends from the liq- As indicated earlier in this chapter, radiation
uid nitrogen container (a Dewar flask) to cool from radioactive materials interacts with
the detector. Some of the preamplifier elec- matter by causing ionization or excitation of
tronic circuitry also is cooled to reduce atoms and molecules. When the ionized or
electronic noise levels. Liquid nitrogen excited products undergo recombination or
deexcitation, energy is released. Most of the
energy is dissipated as thermal energy, such
as molecular vibrations in gases or liquids or
lattice vibrations in a crystal; however, in
some materials a portion of the energy is
released as visible light.* These materials are
called scintillators, and radiation detectors
made from them are called scintillation
detectors.
Liquid The scintillator materials used for detec-
nitrogen tors in nuclear medicine are of two general
types: inorganic substances in the form of
solid crystals and organic substances dis-
solved in liquid solution. The scintillation
mechanisms are different for these two
types and are described separately in later
“Coldfinger” sections.
Vacuum A characteristic common to all scintillators
is that the amount of light produced following
Detector
*For simplicity, the term visible light is used to describe
FIGURE 7-12 Schematic representation of a typical scintillation emission. In fact, the emissions from many
semiconductor detector assembly. “Coldfinger” is a scintillators extend into the UV portion of the spectrum
thermal conductor for cooling the detector element. as well.
98 Physics in Nuclear Medicine

the interaction of a single γ ray, β particle, or compounds are commonly used for this mate-
other ionizing radiation, is proportional to the rial. The photoemissive surface is called the
energy deposited by the incident radiation in photoca­thode, and electrons ejected from it
the scintillator. The amount of light produced are called photoelectrons. The conversion effi-
also is very small, typically a few hundred to ciency for visible light to electrons, also known
a few thousand photons for a single γ-ray as the quantum efficiency, is typically 1 to 3
interaction within the energy range of inter- photoelectrons per 10 visible light photons
est for nuclear medicine imaging (70-511 keV ). striking the photocathode. The dependence of
In the early days of nuclear physics, it was quantum efficiency on the wavelength of the
common to study the characteristics of par- light is shown for a conventional bialkali pho-
ticles by observing and counting, in a dark- tocathode in Figure 7-14.
ened room, the scintillations produced by A short distance from the photocathode is
these particles on a zinc sulfide scintillation a metal plate called a dynode. The dynode is
screen. The obvious limitations on counting maintained at a positive voltage (typically
speed and accuracy with this system have 200-400 V) relative to the photocathode and
been eliminated in modern application with attracts the photoelectrons ejected from it.
the introduction of ultrasensitive electronic A focusing grid directs the photoelectrons
light detectors called photomultiplier (PM ) toward the dynode. The dynode is coated with
tubes. a material having relatively high secondary
emission characteristics. CsSb also can be
2. Photomultiplier Tubes used for this material. A high-speed photo-
PM tubes (also called phototubes and some- electron striking the dynode surface ejects
times abbreviated PMT ) are electronic tubes several secondary electrons from it. The elec-
that produce a pulse of electrical current tron multiplication factor depends on the
when stimulated by very weak light signals, energy of the photoelectron, which in turn is
such as the scintillation produced by a γ ray determined by the voltage difference between
or β particle in a scintillation detector. Their the dynode and the photocathode.
basic principles are illustrated in Figure 7-13. Secondary electrons ejected from the first
The inside front surface of the glass dynode are attracted to a second dynode,
entrance window of the PM tube is coated which is maintained at a 50-150 V higher
with a photoemissive substance. A photo­ potential than the first dynode, and the elec-
emissive substance is one that ejects electrons tron multiplication process is repeated. This
when struck by photons of visible light. occurs through many additional dynode
Cesium antimony (CsSb) and other bialkali stages (typically 9 to 12 in all), until finally a

Photocathode Photoelectron
Glass
Dynodes envelope Vacuum
Entrance Focusing
window grid

Anode

Light
photon C R
Output
signal

High voltage
supply

FIGURE 7-13 Basic principles of a photomultiplier tube.


7 • Radiation Detectors 99

UV
30

25
Quantum efficiency (%)

20

15

10

0
300 350 400 450 500 550 600 650 700
Wavelength (nm)
FIGURE 7-14 Quantum efficiency as a function of wavelength for a typical bialkali photocathode. The peak occurs at
approximately 400 nm, which is well-matched to the emission wavelength of many scintillators. UV, ultraviolet light.

shower of electrons is collected at the anode. PM tubes are sealed in glass and evacu-
Typical electron multiplication factors are ×3 ated. Electrical connections to the dynodes,
to ×6 per dynode. The total electron multipli- the photocathode, and the anode are made
cation factor is very large—for example, 610 through pins in the tube. The focusing of the
(~6 × 107) for a 10-stage tube with an average electron beam from one dynode to the next
multiplication factor of 6 at each dynode. can be affected by external magnetic fields.
Thus a relatively large pulse of current is Therefore PM tubes often are wrapped in
produced when the tube is stimulated by even metal foil for magnetic shielding. “Mu-metal,”
a relatively weak light signal. Note that the an alloy composed of iron, nickel, and small
amount of current produced is proportional to amounts of copper and chromium, is com-
the intensity of the light signal incident on monly used for this purpose. PM tubes come
the photocathode and thus also to the amount in various shapes (round, square, and hexago-
of energy deposited by the radiation event in nal) and sizes (Fig. 7-15). Most of those used
the crystal. in nuclear medicine have photocathodes in
PM tubes require a high-voltage supply. the range of 1- to 7.5-cm diameter. There are
For example, as shown in Figure 7-13, if the also position-sensitive and multichannel PM
tube has 10 dynodes, with the first at +300 V tubes available that have the ability to deter-
relative to the photocathode and the remain- mine the location of incident light on the
ing 9 dynodes and the anode at additional photocathode.
+100 V increments, a voltage of +1300 V is
needed. Furthermore, the voltage supply 3. Photodiodes
must be very stable because the electron mul- In some applications, the PM tube may be
tiplication factor is very sensitive to dynode replaced by a light-sensitive semiconductor
voltage changes. Typically a 1% increase in detector, such as a Si photodiode. Note that
high voltage applied to the tube increases the in this case, the semiconductor is not being
amount of current collected at the anode by used to detect the γ rays directly but to detect
approximately 10%. This is of considerable the visible light emitted from a scintillator
importance in applications where pulse size material in which a γ ray has interacted. The
is being measured, such as in pulse-height photons from the scintillator have sufficient
spectrometry to determine γ-ray energies (see energy to cause ionization within Si, and the
Chapter 10). total charge produced is proportional to the
100 Physics in Nuclear Medicine

FIGURE 7-15 Assortment of photomultiplier tubes illustrating their wide variety of shapes and sizes. (Courtesy
Hamamatsu Corp., Bridgewater, NJ.)

number of scintillation light photons incident large and independent of the energy of the
on the photodiode. These photodiode detec- incident radiation. The gain of such devices
tors have the advantage that they can be can be as high as 107. Light-sensitive detec-
made very small in area and that they are tors consisting of a large number of tiny
typically only a few millimeters thick, includ- (20-50 µm) geiger-mode APDs that are incor-
ing the packaging. They also have signifi- porated into a single device are being devel-
cantly higher quantum efficiency than PM oped for use in scintillation detectors.
tubes, with values ranging typically between
60% and 80%. However, conventional Si
photodiodes have unity gain (compared with 4. Inorganic Scintillators
106 to 107 for a PM tube), requiring very low- Inorganic scintillators are crystalline solids
noise electronics for readout. that scintillate because of characteristics of
A related device, the Si avalanche photo­ their crystal structure. Individual atoms and
diode (APD) uses a very high internal electric molecules of these substances do not scintil-
field such that each electron produced within late. They are scintillators only in crystalline
the device gains enough energy between colli- form. Table 7-2 summarizes the properties of
sions to create further ionization. This is anal- a number of inorganic scintillators of interest
ogous to the proportional region for gas-filled for nuclear medicine applications.
detectors that was discussed previously in Some inorganic crystals are scintillators in
Section A.3. APD detectors can reach gains of their pure state; for example, pure NaI crys-
102 to 103 but still require low-noise electronics tals are scintillators at liquid nitrogen tem-
for successful operation. The gain of these peratures. Most are “impurity activated,”
devices also is a very strong function of bias however. These are crystals containing small
voltage and temperature, and these parame- amounts of “impurity” atoms of other ele-
ters therefore must be very carefully controlled ments. Impurity atoms in the crystal matrix
for stable operation. These types of solid-state cause disturbances in its normal structure.
light detectors are used presently only in Because they are responsible for the scin­
specialized nuclear medicine systems, such tillation effect, the impurity atoms in the
as small animal scanners (see Chapter 17, crystal matrix are sometimes called activator
Section A.4, and Chapter 18, Section B.5) and centers. Some impurity-activated scintillators
dual-modality positron emission tomography– that have been used in radiation detectors
magnetic resonance imaging (PET-MRI) include sodium iodide [NaI(Tl)] and cesium
systems (see Chapter 19, Section F). iodide [CsI(Tl)]. In each case, the element in
APDs also can be operated at higher bias parentheses is the impurity that is added to
voltages in geiger mode. This is analogous to create activator centers in the crystal.
the gas-filled GM counter (Section A.4), and The most commonly used scintillator for
the output signal of the APD becomes very detectors in nuclear medicine is NaI(Tl). Pure
7 • Radiation Detectors 101

TABLE 7-2
PROPERTIES OF SOME SCINTILLATOR MATERIALS USED IN NUCLEAR MEDICINE

Property NaI(Tl) BGO LSO(Ce) GSO(Ce) CsI(Tl) LuAP(Ce) LaBr3(Ce) Plastic*


3
Density (g/cm ) 3.67 7.13 7.40 6.71 4.51 8.34 5.3 1.03
Effective atomic number 50 73 66 59 54 65 46 12
Decay time (nsec) 230 300 40 60 1000 18 35 2
Photon yield (per keV) 38 8 20-30 12-15 52 12 61 10
Index of refraction 1.85 2.15 1.82 1.85 1.80 1.97 1.9 1.58
Hygroscopic Yes No No No Slightly No Yes No
Peak emission (nm) 415 480 420 430 540 365 358 Various
*Typical values—there are many different plastic scintillators available.
BGO, Bi3Ge4O12; GSO(Ce), Gd2SiO5(Ce); LSO(Ce), Lu2SiO5(Ce); LuAP(Ce), LuAlO5(Ce)

NaI crystals are scintillators only at liquid optical window at one end to permit the exit
nitrogen temperatures. They become efficient of scintillation light from the crystal to the PM
scintillators at room temperatures with the tube. A transparent optical “coupling grease”
addition of small amounts of thallium. Single is used between the crystal and the PM tube
crystals of NaI(Tl) for radiation detectors are to minimize internal reflections at this inter-
“grown” from molten sodium iodide to which face. The crystal and PM tube are hermeti-
a small amount of thallium (0.1-0.4 mole cally sealed in a light-tight jacket to keep out
percent) has been added. Crystals of rela- moisture and extraneous light and for mechan-
tively large size are grown in ovens under ical protection. The inside surface of the radia-
carefully controlled temperature conditions. tion entrance window and sides of the crystal
For example, crystals for gamma cameras are coated with a highly reflective diffuse
(see Chapter 13) use NaI(Tl) crystals that are material to maximize the light collected by the
typically 30-50 cm in diameter by 1-cm thick. PM tube photocathode. With efficient optical
Figure 7-16 shows the construction of a coupling, good reflective surfaces, and a crystal
typical scintillation detector consisting of a free of cracks or other opacifying defects,
NaI(Tl) crystal and PM tube assembly. The approximately 30% of the light emitted by
crystal is sealed in an aluminum or stainless- the crystal actually reaches the cathode of the
steel jacket with a transparent glass or plastic PM tube. Some NaI(Tl) detectors have very

Scintillation
Photoelectrons light
Reflector

PM Tube NaI(Tl)

 ray

First
Photocathode Interaction
dynode
Mu-metal
shield
FIGURE 7-16 Arrangement of NaI(Tl) crystal and photomultiplier (PM) tube in a typical scintillation detector
assembly.
102 Physics in Nuclear Medicine

thin aluminum or beryllium foil “entrance 5. The scintillation light is well-matched in


windows” to permit detection of radiations wavelength to the peak response of the
having relatively low penetrating power, such PM tube photocathode (see Fig. 7-14).
as low-energy x rays and γ rays (E  10 keV) The emission spectrum of light from
and β particles; however, most NaI(Tl) detec- NaI(Tl) is shown in Figure 7-18.
tors have thicker entrance windows of alumi- Some disadvantages of NaI(Tl) detectors are
num or stainless steel and are best suited for the following:
detecting higher-energy γ rays (E  50 keV). 1. The NaI(Tl) crystal is quite fragile and
Figure 7-17 shows some typical integral easily fractured by mechanical or
NaI(Tl) crystal and PM tube assemblies. thermal stresses (e.g., rapid tempera-
Some reasons for the usefulness of NaI(Tl) ture changes). Fractures in the crystal
scintillation detectors include the following: do not necessarily destroy its usefulness
1. It is relatively dense (ρ = 3.67 g/cm3) and as a detector, but they create opacifica-
contains an element of relatively high tions within the crystal that reduce the
atomic number (iodine, Z = 53). Therefore amount of scintillation light reaching
it is a good absorber and a very efficient the photocathode.
detector of penetrating radiations, such 2. Sodium iodide is hygroscopic. Exposure
as x rays and γ rays in the 50- to 250-keV to moisture or a humid atmosphere
energy range. The predominant mode of causes a yellowish surface discoloration
interaction in this energy range is by that again impairs light transmission to
photoelectric absorption. the PM tube. Thus hermetic sealing is
2. It is a relatively efficient scintillator, required.
yielding one visible light photon per 3. At higher γ-ray energies ( 250 keV),
approximately 30 eV of radiation energy the predominant mechanism of interac-
absorbed. tion is by Compton interaction, and
3. It is transparent to its own scintillation larger volumes of NaI(Tl) are required
emissions. Therefore there is little loss for adequate detection efficiency.
of scintillation light caused by self- Other types of detectors have advantages
absorption, even in NaI(Tl) crystals of over NaI(Tl) detectors in certain areas. For
relatively large size. example, gas-filled detectors are cheaper (but
4. It can be grown relatively inexpensively have much lower detector efficiency), and
in large plates, which is advantageous semiconductor detectors have better energy
for imaging detectors. resolution (but are expensive to use in large-
area imaging cameras). However, the overall
advantages of NaI(Tl) have made it the detec-
tor of choice for almost all routine applica-
tions in nuclear medicine involving γ-ray
detection in the 50-250-keV energy range.
At higher energies, particularly for detec-
tion of the 511-keV emissions from positron
emitters, denser scintillators generally are
preferred. Bismuth germanate (Bi4Ge3O12;
BGO) is a commonly used scintillator in PET
imaging, because of its excellent detection
efficiency at 511 keV. Lutetium oxyortho­
silicate [Lu2SiO5(Ce); LSO] is slightly less effi-
cient at 511 keV than BGO, but is brighter
and faster and may sometimes offer advan-
tages over BGO when the counting rate
on the detector is high, when fast timing
information is needed, or when small scintil-
lator elements are to be decoded in an imaging
system (see Chapter 18, Section B). However,
LSO is rather expensive to grow, because of
its high melting point and its raw material
FIGURE 7-17 NaI(Tl) crystal and photomultiplier tube
costs. A related material, lutetium yttrium
assemblies. (Courtesy Crystals and Detectors Division, Saint- orthosilicate (LYSO—LSO in which a small
Gobain Ceramics and Plastics, Inc., Newbury, OH.) fraction of the lutetium atoms are replaced
7 • Radiation Detectors 103

UV

1.00

Intensity (arbitrary units) 0.80

0.60

0.40

0.20

0.00
300 350 400 450 500 550 600 650 700
Wavelength (nm)
FIGURE 7-18 The emission spectrum of NaI(Tl) scintillator at room temperature. UV, ultraviolet light. (Data courtesy
Kanai Shah and Jarek Glodo, Radiation Monitoring Devices Inc., Watertown, MA.)

with yttrium) has scintillation properties photoelectric component dominates in high-Z


very similar to LSO. BaF2 and CsF also have materials over much of that range. As well,
historically been used in positron cameras high-Z materials tend to be denser than low-Z
because of their very fast decay time (which materials, thus increasing their linear atten-
is important for timing of γ-ray interactions uation coefficients and enhancing their stop-
in coincidence detection) (see Chapter 18, ping power per unit of detector thickness.
Section A.2). However, their low detection Thus, high-Z detector materials generally are
efficiency compared with BGO and LSO has more efficient than low-Z materials for detect-
prevented any widespread application. ing high-energy γ rays.
New scintillator materials continue to be Most detector materials consist of a mixture
discovered and developed. Among the more of elements. The effective atomic number, Zeff,
pro­mising recent candidates for nuclear medi- is a useful and convenient parameter for com-
cine applications are LuAP [LuAlO3(Ce)], lan- paring the ability of such mixtures to stop
thanum bromide [LaBr3(Ce)], and lanthanum high-energy γ rays. It is given by
chloride [LaCl3(Ce)].
x
Zeff = w1 Z1x + w2 Z2x +  wn Znx (7-2)
5. Considerations in Choosing an
Inorganic Scintillator where wi is a weighting factor proportional to
The ability of a scintillator to stop high-energy the fractional number of electrons per gram
γ rays ( 100 keV) is of particular importance for element i and can be calculated as
in nuclear medicine, especially for imaging
applications. As discussed in Chapter 6, mi Zi
Section D.1, the photoelectric component wi = (7-3)

n
of the mass-attenuation coefficient increases i =1
mi Zi
strongly with the atomic number, Z, of a mate-
rial, whereas the Compton component is where mi represents the number of atoms of
essentially independent of atomic number. element i present in the compound or mixture.
As illustrated in Fig. 6-18, although the The power x depends on the energy of the
Compton component is significant over much γ rays. For γ rays in the 100- to 600-keV range,
of the nuclear-medicine energy range, the x is typically taken to be between 3 and 3.5.1
104 Physics in Nuclear Medicine

handle is approximately 1/2τ. Faster scintil-


EXAMPLE 7-1
lators detectors therefore can handle higher
Calculate the effective atomic number of BGO event rates on the detector.
(Bi4Ge3O12). Compare this with the value for The efficiency of the scintillator in convert-
NaI(Tl). (Ignore the contribution of the trace ing a γ ray into visible light photons (photon
amounts of thallium.) yield) is important in determining the preci-
sion with which the energy of the interacting
Answer γ ray can be determined. This becomes rele-
There are three elements contributing to vant in many counting and imaging systems
BGO: Bi (Z = 83), Ge (Z = 32), and O (Z = 8). in which it is important to distinguish between
The denominator for the weighting factors is γ rays that have Compton scattered in the
(83 × 4) + (32 × 3) + (8 × 12) = 524. The weight- sample (and therefore lost energy; see Chapter
ing factor for Bi is (83 × 4) / 524 = 0.634, for 6, Section C.3) and those that remain unscat-
Ge is (32 × 3) / 524 = 0.183, and for O is (8 × tered. Higher photon yield also is important
12) / 524 = 0.183. These three weighting in determining the positioning accuracy in
factors add to 1, as they should. Then Zeff can many imaging systems in which it is common
be calculated as to share the limited number of scintillation
Zeff = (0.634 × 833.5 + 0.183 photons among multiple PM tubes to deter-
mine the location of an interaction (see
× 323.5 + 0.183 × 83.5 )1 / 3.5
Chapter 13 and Chapter 18, Section B).
= 73.1 The index of refraction of a scintillator plays
NaI(Tl) consists of Na (Z=11) and I (Z=53) a role in determining how efficiently scintilla-
atoms that are present in equal quantities. tion light can be coupled from a scintillator
The weighting factor for Na is 11/(11 + 53) = crystal into a PM tube. The index of refraction
0.172 and for I is 53/(11 + 53) = 0.828. There- of the glass entrance window on a PM tube is
fore Zeff for NaI(Tl) is ~1.5. Therefore for best transmission of light
from the scintillator into the PM tube with
Zeff = (0.172 × 113.5 + 0.828 × 533.5 )1 / 3.5 minimal internal reflection at the scintillator
= 50.2 crystal and PM tube interface, the scintillator
should have an index of refraction as close to
1.5 as possible. In practice, most scintillators
In addition to choosing a scintillator that of interest have indices of refraction signifi-
has sufficient stopping power for the efficient cantly higher than 1.5 and this is one reason
detection of γ rays with a particular energy, why only a fraction of the scintillation light
the other properties listed in Table 7-2 also produced actually reaches the PM tube.
influence the choice of scintillator materials Lastly, it is important that the emission
for a specific application. The decay time of spectrum of the light produced by the scintil-
the scintillator is important in two respects. lator is a good match for the quantum effi-
Firstly, it determines the precision with which ciency of the photodetector that is used to
the time of γ-ray interaction in the scintillator convert the scintillation light into an elec-
can be determined. Faster light production tronic pulse. In the case of a PM tube with
within the scintillator (faster decay time) a standard bialkali photocathode (see Fig.
results in better timing precision. This is 7-14), it is apparent that scintillators that
important in nuclear medicine applications in have their peak light emission in the range of
which timing is important, most notably in 350-475 nm are optimal. NaI(Tl) is an example
PET in which coincident 511 keV annihila- of a scintillator that has an emission spec-
tion photons are detected (see Chapter 18, trum (see Fig. 7-18) that matches the quantum
Section A). Secondly, the decay time of the efficiency of PM tubes very well. The trans-
scintillator is a limiting factor in how many mission of the scintillation light through the
γ-ray interactions a detector can process per glass used in the PM tube entrance window
unit time. To unambiguously detect two inter- also must be considered. Many glasses are
actions, they should be separated by roughly efficient absorbers of UV light at wavelengths
2-3 times the decay time; otherwise events significantly shorter than ~350 nm.
“pile up” on top of each other, leading to dead
time (see Chapter 11, Section C). As a rough 6. Organic Scintillators
rule of thumb, if the scintillator has a decay In contrast with inorganic scintillators,
time of τ, the maximum event rate that a the scin­tillation process in organic scintilla-
detector made using that scintillator can tors is an inherent molecular property. The
7 • Radiation Detectors 105

scintillation mechanism is one of molecular emits light. Some common primary scin-
excitation (e.g., by absorbing energy from a γ tillators include p-bis-(omethylstyryl)
ray or β particle) followed by a deexcitation benzene (abbreviated as bis-MSB) and
process in which visible light is emitted. 2,5-diphenyloxazole (also known as
These substances are scintillators whether PPO).
they are in solid, liquid, or gaseous forms. 3. The emissions of the primary solute are
Certain plastics (e.g., see Table 7-2) are not always well matched to the response
organic scintillators and have been used for characteristics of PM tubes. Therefore a
direct detection of β particles emitted from secondary solute, or waveshifter, is some-
radionuclides, particularly in compact probes times added to the solution. The function
designed for surgical use (see Chapter 12, of this material is to absorb emissions of
Section F.2). A more common application for the primary solute and reemit photons
organic scintillators, however, is to employ of longer wavelength, which are better
them in liquid form for liquid scintillation matched to the PM tube response.
(LS) counting. In these systems, the scintilla- 1,4-di-(2-5-pheny­loxazole) benzene (also
tor is dissolved in a solvent material in a glass known as POPOP) is a commonly used
or plastic vial and the radioactive sample is secondary scintillator.
added to this mixture. The vial is then placed 4. LS solutions frequently contain addi-
in a light-tight enclosure between a pair of tives to improve some aspect of their
PM tubes to detect the scintillation events performance, such as the efficiency of
(Fig. 7-19). energy transfer from the solvent to
LS solutions consist of four components: the primary solute. Solubilizers (e.g.,
1. An organic solvent comprises most of the hyamine hydroxide) are sometimes
solution. The solvent must dissolve not added to improve the dissolution of
only the scintillator material but also added samples such as blood.
the radio­active sample added to it. The The precise “cocktail” of solvent, primary
solvent actually is responsible for most and secondary solutes, and additives depends
of the direct absorption of radiation on the sample type that is being measured. A
energy from the sample. High-speed wide variety of different LS cocktails opti-
electrons generated by ionizing radia- mized for different applications are available
tion events in the solvent transfer energy commercially.
to the scintillator molecules, causing the Because of the intimate relationship
scintillation effect. Commonly used sol- between sample and detector, LS counting is
vents include di-iso-propylnapthalene the method of choice for efficient detection of
(DIN) and phenylxylylethane (PXE), particles, low-energy x rays and γ rays, and
which are replacing traditional, more other nonpenetrating radiations. It is widely
environmentally harsh solvents such as used for measurement of 3H and 14C. In
toluene and xylene. medical applications, it is used primarily for
2. The primary solute (or primary fluor) sensitive assay of radioactivity in biologic
absorbs energy from the solvent and specimens, such as blood and urine.

Liquid scintillator solution


containing dissolved
radioactive sample

Photomultiplier tube Photomultiplier tube

Scintillation
light
FIGURE 7-19 Arrangement of sample and detector for liquid scintillation counting. The sample is dissolved in a liquid
scintillator solution in a glass or plastic vial.
Although well suited for counting non- to reduce the concentration of primary
penetrating radiations in biologic samples, and secondary solutes in the final
LS counters have numerous drawbacks as solution, thus reducing the scintillator
general-purpose radiation detectors. They are output efficiency.
inefficient detectors of penetrating radiations Quenching can be minimized in various
such as γ rays and x rays of moderate energy ways. For example, dissolved oxygen may be
because the detector solution is composed purged by ultrasound, and hydrogen peroxide
primarily of low-density, low-Z materials. In may be added for color bleaching. However,
addition, liquid scintillators generally have there are no convenient ways to eliminate
low light output, only about one third that of all causes of quenching; therefore a certain
NaI(Tl). This problem is worsened by the rela- amount must be accepted in all practical
tively inefficient light coupling from the scin- LS counting. Quenching becomes a serious
tillator vial to the PM tubes as compared with problem when there are wide variations in its
NaI(Tl) integral detectors. extent from one sample to the next. This
For sample counting, special sample prepa- causes unpredictable variations in light
ration may be required to dissolve the sample. output, for the same amount of radiation
Problems in sample preparation are discussed energy absorbed, from one sample to the next.
in Chapter 12, Section C.6. Also, the sample Quench correction methods are employed in
itself is “destroyed” when it is added to the LS counters to account for this effect (Chapter
scintillator solution. 12, Section C.5).
Finally, all LS counting suffers from the
problem of quenching. Quenching in this
context refers to any mechanism that reduces REFERENCE
the amount of light output from the sample 1. Johns HE, Cunningham JR: The Physics of Radiology,
(not to be confused with the electrical ed 4, Springfield, IL, 1983, Charles C Thomas, pp
241-243.
quenching that occurs in GM counters; see
Section A.4). There are basically three types
of LS quenching: BIBLIOGRAPHY
1. Chemical quenching is caused by sub- A comprehensive reference for many different
stances that compete with the primary radiation detectors is the following:
fluor for absorption of energy from the Knoll GF: Radiation Detection and Measurement, ed 4,
solvent but that are themselves not scin- New York, 2010, John Wiley.
tillators. Dissolved oxygen is one of the
A detailed reference for inorganic scintillator
most troublesome chemical quenchers. mechanisms, properties, growth, and applications
2. Color quenching is caused by substances is the following:
that absorb the emissions of the primary Lecoq P, Annenkov A, Getkin A, Korzhik M, Pedrini C:
or secondary solute. Blood and other Inorganic Scintillators for Detector Systems: Physical
colored materials are examples. Fogged Principles and Crystal Engineering. Berlin, 2006,
or dirty containers can also produce a Springer.
type of color quenching. A detailed general reference for scintillation
3. Dilution quenching occurs when a rela- detectors is the following:
tively large volume of sample is added Birks JB: The Theory and Practice of Scintillation Count-
to the scintillator solution. The effect is ing. London, 1967, Pergamon Press.

You might also like