Professional Documents
Culture Documents
Introduction
Instead, detectors usually measure the secondary products arising from the
interactions of the radiation with the detector material.
But, ionizing radiation such as gamma photons and neutrons must first
undergo interactions in the detector material that produce secondary
charged particles, recoil atoms or electrons that, in turn, produce charge
pairs as they slowdown.
348
In some detectors the average current can be used as a measure of the
amount of ionization or energy deposition, per unit mass of detector
material, caused by incident radiation.
The outer tube contains the gas and defines the active volume of the
chamber.
If the electrons released from the atoms in the wall material escape from
the wall and enter the active gas volume of the chamber,
The potential difference between the anode wire and the cathode
establishes an electric field that causes positive and negative charges to
move in opposite directions.
Electrons rapidly drift toward the anode and positive ions migrate more
slowly toward the cathode.
The motion of these ion pairs causes a flow of current in the external
circuit and establishes a voltage across the load resistance.
Geiger-Mueller counters.
All three are known as ionization chambers, but they each have a unique
process for forming the total number of ion pairs that are collected at the
electrodes.
All three operate by forming initial ion pairs from the incident radiation.
Once these ion pairs are formed, it is important that they do not recombine
and thereby fail to contribute to the electrical signal.
the negative electrons will move to the anode and the positive ions to the
cathode,
At very low voltages, the ion pairs do not receive enough acceleration to
reach the electrodes and therefore may combine together to form the
original molecule instead of being collected by the electrodes.
Ion chambers operating in this region are not useful radiation detectors.
In this region II, the ionization chamber Region, only the primary ion
pairs formed by the initial radiations are collected.
the electrons and positive ions gain such high velocities and energies
during their acceleration toward the electrodes that they cause secondary
ionization.
The latter will increase the measured current. This process is referred to as
the gas amplification.
This factor can be as high as 106 per individual primary event depending
on the design of the gas detector and the applied voltage.
Proportional counters are usually filled with 90% argon and 10% methane
at atmospheric pressure.
These counters, however, are not commonly used for y-and x-ray counting
because of poor counting efficiency (<1%).
This limited proportionality region (region III) where the number of ion-
electron pairs collected is relatively independent of the initial number of
ion-electron pairs created by the incident radiation.
When highly accelerated electrons strike the anode with a great force,
ultraviolet (UV) light is emitted, which causes further emission of
photoelectrons by gas ionization and from the chamber walls.
The circuit in this case produces a large electrical pulse in response to each
radiation event, regardless of the applied voltage (within the Geiger Muller
region) or the energy of the radiation.
The pulse size is essentially same for all radiation events, no matter the
type of radiation or the amount energy transfer.
If the tube voltage is too large, the tube undergoes continuous avalanches
around the central anode, one leading to another.
Ionization chamber
It only uses the discrete charges created by each interaction between the
incident radiation and the gas,
and does not involve the gas multiplication mechanisms used by other
radiation instruments, such as the Geiger counter or the proportional
counter.
Ion chambers have a good uniform response to radiation over a wide range
of energies and are the preferred means of measuring high levels of
gamma radiation.
They are widely used in the
and the resultant positive ions and dissociated electrons move to the
electrodes of the opposite polarity under the influence of the electric field.
The electric field also enables the device to work continuously by mopping
up electrons, which prevents the fill gas from becoming saturated, where
no more ions could be collected, and by preventing the recombination of
ion pairs, which would diminish the ion current.
This has the effect of ensuring the gas in the chamber is acting as though it
were a portion of an infinitely large gas volume, and increases the
accuracy by reducing interactions of gamma with the wall material.
The higher the atomic number of the wall material, the greater the chance
of interaction.
The wall thickness is a trade-off between maintaining the air effect with a
thicker wall, and increasing sensitivity by using a thinner wall.
These chambers often have an end window made of material thin enough,
so that beta particles can enter the gas volume.
Gamma radiation enters both through the end window and the side walls.
The efficiency of the chamber can be further increased by the use of a high
pressure gas.
Typically a pressure of 8-10 atmospheres can be used, and various noble
gases are employed.
The higher pressure results in a greater gas density and thereby a greater
chance of collision with the fill gas and ion pair creation by incident
radiation.
Proportional Counter
An ionizing particle entering the gas collides with an atom of the inert gas
and ionizes it to produce an electron and a positively charged ion,
commonly known as an "ion pair".
As the ionizing particle travels through the chamber it leaves a trail of ion
pairs along its trajectory, the number of which is proportional to the energy
of the particle if it is fully stopped within the gas.
Typically a 1 MeV stopped particle will create about 30,000 ion pairs.
The chamber geometry and the applied voltage is such that in most of the
chamber the electric field strength is low and the chamber acts as an ion
chamber.
This is the "ion drift" region. In the immediate vicinity of the anode wire,
the field strength becomes large enough to produce Townsend
avalanches.
This avalanche region occurs only fractions of a millimetre from the anode
wire, which itself is of a very small diameter.
A key design goal is that each original ionizing event due to incident
radiation produces only one avalanche.
For this reason, the applied voltage, the geometry of the chamber and the
diameter of the anode wire are critical to ensure proportional operation.
If avalanches start to self-multiply due to UV photons as they do in a
Geiger-Muller tube, then the counter enters a region of "limited
proportionality" until at a higher applied voltage the Geiger discharge
mechanism occurs with complete ionization of the gas enveloping the
anode wire and consequent loss of particle energy information.
Therefore, it can be said that the proportional counter has the key design
feature of two distinct ionization regions:
1. Ion drift region: in the outer volume of the chamber – the creation
of number ion pairs proportional to incident radiation energy.
2. Avalanche region: in the immediate vicinity of the anode – Charge
amplification of ion pair currents, while maintaining localized
avalanches.
[( ) ]
V ln2 V
ln M ln = ln - ln K
lln b ∆ Vλ b
ln ( ) paln( )
a a
where a is the anode wire radius, b is the radius of the counter, p is the
pressure of the gas, and V is the operating voltage. K is a property of the
gas used and relates the energy needed to cause an avalanche to the
pressure of the gas. The term ∆ V λ gives the change in voltage caused by an
avalanche.
This large pulse from the tube makes the Geiger counter relatively cheap
to manufacture, as the subsequent electronics are greatly simplified
(Fig.5).
The electronics also generate the high voltage, typically 400–900 volts,
that has to be applied to the Geiger-Müller tube to enable its operation.
There are two types of detected radiation readout: counts or radiation dose.
The counts display is the simplest and is the number of ionizing events
detected displayed either as a count rate, such as "counts per minute" or
"counts per second", or as a total number of counts over a set time period
(an integrated total).
The counts readout is normally used when alpha or beta particles are being
detected.
A Geiger-Müller tube can detect the presence of radiation, but not its
energy, which influences the radiation's ionizing effect.
The readout can be analog or digital, and modern instruments are offering
serial communications with a host computer or network.
Principle of operation:
The walls of the tube are either metal or have their inside surface coated
with a conducting material or a spiral wire to form the cathode, while the
anode is a wire-mounted axially in the center of the chamber.
When ionizing radiation strikes the tube, some molecules of the fill gas are
ionized directly by the incident radiation, and if the tube cathode is an
electrical conductor, such as stainless steel, indirectly by means of
secondary electrons produced in the walls of the tube, which migrate into
the gas.
This creates positively charged ions and free electrons, known as ion pairs,
in the gas.
The strong electric field created by the voltage across the tube's electrodes
accelerates the positive ions towards the cathode and the electrons towards
the anode.
Close to the anode in the "avalanche region" where the electric field
strength rises exponentially as the anode is approached, free electrons gain
sufficient energy to ionize additional gas molecules by collision and create
a large number of electron avalanches.
These spread along the anode and effectively throughout the avalanche
region.
This is the "gas multiplication" effect which gives the tube its key
characteristic of being able to produce a significant output pulse from a
single original ionizing event.
If there were to be only one avalanche per original ionizing event, then the
number of excited molecules would be in the order of 106 to 108.
For common sizes of tubes the complete ionization of the gas around the
anode takes just a few microseconds.
This short, intense pulse of current can be measured as a count event in the
form of a voltage pulse developed across an external electrical resistor.
This can be in the order of volts, thus making further electronic processing
simple.
These ions have lower mobility than the free electrons due to their higher
mass and move slowly from the vicinity of the anode wire.
This creates a "space charge" which counteracts the electric field that is
necessary for continued avalanche generation.
Too low a pressure and the efficiency of interaction with incident radiation
is reduced.
Too high a pressure, and the “mean free path” for collisions between
accelerated electrons and the fill gas is too small, and the electrons cannot
gather enough energy between each collision to cause ionisation of the
gas.
For alpha particles, low energy beta particles, and low energy X-rays, the
usual form is a cylindrical end-window tube.
This type has a window at one end covered in a thin material through
which low-penetrating radiation can easily pass.
Mica is a commonly used material due to its low mass per unit area. The
other end houses the electrical connection to the anode.
Used for gamma radiation detection above energies of about 25 KeV, this
type generally has an overall wall thickness of about 1-2 mm of chrome
steel.
Because most high energy gamma photons will pass through the low
density fill gas without interacting, the tube uses the interaction of photons
on the molecules of the wall material to produce high energy secondary
electrons within the wall.
Some of these electrons are produced close enough to the inner wall of the
tube to escape into the fill gas.
As soon as this happens the electron drifts to the anode and an electron
avalanche occurs as though the free electron had been created within the
gas.
The avalanche is a secondary effect of a process that starts within the tube
wall with the production of electrons that migrate to the inner surface of
the tube wall, and then enter the fill gas.
Thin walled:
Thin walled tubes are used for high energy beta detection, where the beta
enters via the side of the tube and interacts directly with the gas, but the
radiation has to be energetic enough to penetrate the tube wall.
The lower energy photons interact better with the fill gas so this design
concentrates on increasing the volume of the fill gas by using a long thin
walled tube and does not use the interaction of photons in the tube wall.
The transition from thin walled to thick walled design takes place at the
300–400 keV energy levels.
Above these levels thick walled designs are used, and beneath these levels
the direct gas ionization effect is predominant.
Neutron detection:
G–M tubes will not detect neutrons since these do not ionize the gas.
The neutrons interact with the boron nuclei, producing alpha particles, or
directly with the helium-3 nuclei producing hydrogen and tritium ions and
electrons.
Gas mixtures:
The components of the gas fill mixture are an inert gas such as helium,
argon or neon which is ionised by incident radiation, and a "quench" gas of
5–10% of an organic vapor or a halogen gas to prevent spurious pulsing by
quenching the electron avalanches.
The Geiger plateau is the voltage range in which the G-M tube operates in
its correct mode, where ionization occurs along the length of the anode.
This is the "starting voltage". Increasing the voltage still further results in
rapidly rising counts until the "knee" or threshold of the plateau is reached,
where the rate of increase of counts falls off.
However, the plateau has a slight slope mainly due to the lower electric
fields at the ends of the anode because of tube geometry.
At the end of the plateau the count rate begins to increase rapidly again,
until the onset of continuous discharge where the tube cannot detect
radiation, and may be damaged.
Depending on the characteristics of the specific tube (manufacturer, size,
gas type, etc.) the voltage range of the plateau will vary.
Dead time and recovery time in a Geiger Muller tube. The tube can
produce no further pulses during the dead time, and only produces pulses
of lesser height until the recovery time has elapsed.
The ideal G–M tube should produce a single pulse for every single
ionizing event due to radiation.
It should not give spurious pulses, and should recover quickly to the
passive state, ready for the next radiation event.
However, when positive argon ions reach the cathode and become neutral
atoms by gaining electrons, the atoms can be elevated to enhanced energy
levels.
These atoms then return to their ground state by emitting photons which in
turn produce further ionization and thereby spurious secondary discharges.
The G.M.counter can account for about 500 particles per second.
The G.M. counter will not register those particles that pass through it in
the dead time.
Dead time refers to the time taken by the tube to recover between counts.
It requires about 200 µs for the tube to recover.
If lot of particles enter the G.M. tube at a rapid rate, the tube will not have
time to recover and some particles may not be counted.
Chemical quenching:
If a poor gas quencher is introduced to the tube, the positive argon ions,
during their motion toward the cathode, would have multiple collisions
with the quencher gas molecules and transfer their charge and some energy
to them.
Thus, neutral argon atoms would be produced and the quencher gas ions in
their turn would reach the cathode, gain electrons therefrom, and move
into excited states which would decay by photon emission, producing tube
discharge.
However, effective quencher molecules, when excited, lose their energy
not by photon emission, but by dissociation into neutral quencher
molecules.
No spurious pulses are thus produced.
Even with chemical quenching, for a short time after a discharge pulse
there is a period during which the tube is rendered insensitive and is thus
temporarily unable to detect the arrival of any new ionizing particle (the
so-called dead time; typically 50–100 microseconds).
This causes a loss of counts at sufficiently high count rates and limits the
G.M. tube to an effective (accurate) count rate of approximately 10 3 counts
per second even with external quenching.
External quenching:
Efficiency:
Limitations:
Because the output pulse from a Geiger-Müller tube is always of the same
magnitude (regardless of the energy of the incident radiation), the tube
cannot differentiate between radiation types.
Secondly, the inability to measure high radiation rates due to the "dead
time" of the tube.
This is an insensitive period after each ionization of the gas during which
any further incident radiation will not result in a count, and the indicated
rate is, therefore, lower than actual.
Typically the dead time will reduce indicated count rates above about
104 to 105 counts per second depending on the characteristic of the tube
being used.
While some counters have circuitry which can compensate for this, for
accurate measurements ion chamber instruments are preferred for high
radiation rates.
Scintillation Detectors
This pulse carries information about the energy of the original incident
radiation.
The number of such pulses per unit time also gives information about the
intensity of the radiation.
They include
The iodine provides most of the stopping power in sodium iodide (since it
has a high Z = 53).
They exhibit high efficiency for detection of gamma rays and are capable
of handling high count rates.
In this case a “light pipe” is used: light pipes can be cut to any size or
shape out of any transparent material such as Lucite.
Photocathode:
Dynodes:
The next dynode has a higher voltage which makes the electrons released
from the first to accelerate towards it.
At each dynode 3-4 electrons are released for every incident electron, and
with 6 to 14 dynodes the total gain, or electron amplification factor, will
be in the range of ~104-107 when they reach the anode.
This pulse carries information about the energy of the original incident
radiation.
The number of such pulses per unit time also gives information about the
intensity of the radiation.
Quantum Efficiency:
The sensitivity of a photocathode is usually quoted in terms of
the quantum efficiency.
The standard for quotation is the number of photoelectrons per keV energy
loss by fast electrons in a NaI(Tl) scintillator.
The PMT has been the main choice for photon detection ever since due to
the fact that they have high quantum efficiency and high amplification.
Lately however semiconductors have begun to compete with the PMT, the
photodiode for example which has higher quantum efficiency in the visible
range and above, lower power consumption and smaller size.
The quantum efficiency for the photodiode is high (60-80%) compared to
the PMT (20-30%) which gives a higher energy resolution.
They are produced by the decay of nuclei as they transition from a high
energy state to a lower state.
A detailed analysis of this spectrum is typically used to determine
the identity and quantity of gamma emitters present in a sample, and is a
vital tool in radiometric assay.
the electronic devices that analyse the detector output signals, such as a
pulse sorter (i.e., multichannel analyzer).
signal amplifiers,
rate meters,
Through these processes, the energy of the gamma ray is absorbed and
converted into a voltage signal by detecting the energy difference before
and after the interaction (or, in a scintillation counter, the emitted photons
using a photomultiplier).
Data Acquisition:
The voltage pulses produced for every gamma ray that interacts within the
detector volume are then analyzed by a multichannel analyzer (MCA).
From this shape, the signal is then converted into a digital form.
the number of detected signals for each channel represents the spectral
intensity of the radiation in this energy range.
MCAs are typically interfaced with via USB, RS-232 or Ethernet, but can
use PCI also.
Operation:
Thus the screen display of the number of counts vs. channel number is
really a histogram of the number of counts vs. pulse height, i.e. a pulse
height spectrum.
The range of pulse heights to be analyzed can be set via upper and lower
level discriminators at the input.
Detector performance:
Detector resolution:
The most common figure used to express detector resolution is full width
at half maximum (FWHM).
This is the width of the gamma ray peak at half of the highest point on the
peak distribution.
Not all gamma rays emitted by the source that pass through the detector
will produce a count in the system.
The probability that an emitted gamma ray will interact with the detector
and produce a count is the efficiency of the detector.
The same units are used (i.e., percentages); therefore, the spectroscopist
must take care to determine which kind of efficiency is being given for the
detector.
Relative efficiency values are often used for germanium detectors, and
compare the efficiency of the detector at 1332 keV to that of a 3 in × 3 in
NaI detector (i.e., 1.2×10 −3 cps/Bq at 25 cm).
Relative efficiency values greater than one hundred percent can therefore
be encountered when working with very large germanium detectors.
The energy of the gamma rays being detected is an important factor in the
efficiency of the detector.
The pulse height distribution of the gamma rays emitted by the radioactive
decay of 24Na as measured by a NaI(Tl) scintillation detector is shown in
Figure.11.
Na emits only two gamma rays of energies 1.369 and 2.754 MeV.
24
The single and double escape peaks arise from the escape of one and both
0.511MeV annihilation photons generated in the NaI crystal by the
annihilation of positrons created in pair-production interactions.
The backscatter peak arises from photons scattered from the source
material into the NaI detector.
Fig: Pulse height distribution of the gamma rays emitted by the
radioactive decay of Na as measured by a NaI(Tl) scintillation
detector.
Semiconductor detector:
Detection mechanism:
Under the influence of an electric field, electrons and holes travel to the
electrodes, where they result in a pulse that can be measured in an outer
circuit, as described by the Shockley-Ramo theorem.
The holes travel in the opposite direction and can also be measured.
As the amount of energy required to create an electron-hole pair is known,
and is independent of the energy of the incident radiation, measuring the
number of electron-hole pairs allows the intensity of the incident radiation
to be determined.
As the electrons travel fast, the time resolution is also very good, and is
dependent upon rise time.
Si(Li) detectors with thin entrance windows are commonly used in alpha
and beta particle spectrometers.
Since Si(Li) detectors have a much lower atomic number than CZT.
NaI(Tl), and hpge, their relative efficiency per unit thickness is
significantly lower for electromagnetic radiation.
However, for x-ray or gamma ray energies less than about 30 keV,
commercially available Si(Li) detectors are thick enough to provide
performance which is superior to CZT, NaI(Tl), and HPGe.
Based upon the fact that a majority of the applications require a thin
window, Si(Li) detectors are often manufactured with very thin beryllium
windows.
As charged particles pass through these strips, they cause small ionization
currents that can be detected and measured.
The drawback is that silicon detectors are much more expensive than these
older technologies and require sophisticated cooling to reduce leakage
currents (noise source).
They also suffer degradation over time from radiation, however this can be
greatly reduced thanks to the Lazarus effect.
(1) Ge(Li) a germanium crystal doped with lithium ions to cancel the
effect of natural impurities in the germanium crystal, and
(2) the more recent (HPGe) high purity germanium crystal in which
impurity atom concentration are less than 1010 atom/cm3.
The more expensive HPGe detectors have replaced the older Ge(Li)
technology since they can be kept at room temperature when not in use
whereas Ge(Li) crystals must always be kept at liquid nitrogen
temperatures (-196 oC).
Their efficiency ranges from excellent for low energy x-rays too good for
medium to high-energy gamma rays over an energy range of 1 keV to 10
MeV.
The performance of these detectors is often compared to NaI(Tl) and
Cd/Zn telluride (CZT) detectors.
Because of the higher atomic number arid larger size, NaI(Tl) detectors
often have a higher efficiency for high energy gamma rays than do
germanium detectors, but a much poorer energy resolution.
Impurities in the crystals trap electrons and holes, ruining the performance
of the detectors.
At higher temperatures, the electrons can easily cross the band gap in the
crystal and reach the conduction band, where they are free to respond to
the electric field, producing too much electrical noise to be useful as a
spectrometer.
Ge(Li) crystals could never be allowed to warm up, as the lithium would
drift out of the crystal, ruining the detector.
The full energy absorption peaks of both NaI(Tl) and Ge detectors can be
approximated as Gaussian shapes, and the width is characterized by the
parameter σ in the general form of Gaussian distribution.
Generally the width is specified in terms of the full width at half maximum
(FWHM), that is, the distance ∆E between the two points E1 and E2 where
F(E1) = F(E2)=A/ 2 where A is a normalization constant, manipulation of
which results in ∆E=2σ√2ln2≅2.35σ.
The absolute FWHM (the width of the full absorption peak) increase with
energy, roughly as E1 /2, but the ratio ∆E/E decreases like E−1/ 2.
For a Ge detector at typical value for a good detector is 1.7 keV at 1332
keV ( the energy of the 60Co gamma ray taken as standard for energy
measurement), which would correspond to about 1.2 keV at 662 keV if the
E1/2 dependence is valid.
Fig 14:Comparison of the energy resolution of Ge(Li) and NaI(Tl)
detectors. The gamma-ray source is a mixture of 108mAg and 110mAg
The high density of these materials mean they can effectively attenuate X-
rays and gamma-rays with energies of greater than 20keV that traditional
silicon-based sensors are unable to detect.
The wide band gap of these materials also mean they have high resistivity
and are able to operate at, or close to, room temperature (~295K) unlike
germanium-based sensors.
A reasonable degree of cooling for the detector and the directly coupled
preamplifier enhances detector system performance.
CZT detectors offer an excellent option for low energy x-ray spectroscopy
where cooling is not possible.
These detectors are not available in large sizes. Their small size diminishes
the possibility of making detectors with large efficiencies for high-energy
electromagnetic radiation.
Cluster detectors
Gamma-Detector Arrays
The back and the cross-sectional view of the detectors and shielding
system of a typical gamma detector array.
Solved Problems
Problem 1
The 662 keV photon in the decay of 137Cs is observed by a NaI(Tl) detector
with an energy resolution of 53 keV. What will be the resolution for a
measurement of the 1.836 MeV photon in the decay of 90Y?
Solution
42.85
Therefore, 25.7
=1.667
Problem 2
Problem 3
Solution
Solution
Problem 5
Solution
Therefore, n=3.16
Problem 6
Solution
Problem 7
p eEτ
The electrons acquire an average velocity v dr = =
2me 2 me , in the electric field
E, where τ is the average time-interval between two consecutive collisions.
l l
As τ = v ∝
σ v dr , where l is the mean free path of the electrons in the gas and
dr
σ is the interaction cross section, we have
E E
v dr ∝ ∝
σp p
If σ is independent of velocity.
Problem 8
Solution
V 107 V
E (r ) = =1.51 x =151 kV/cm
Electrical field at the wire = rln
R0 m
Ri
V 4 V
E (r ) = =7.55 x 10 =755 V/cm
R0 m
rln
Ri
Problem 9
A proton with total energy 1.4 GeV traverses two scintillator counters 10
m apart. What is the time of flight? If the particle is an electron what is its
time of flight?
Solution
The proton has rest mass m p=0.938 GeV and hence
√
E 1.4 1
γ= = =1.49 , β = 1- =0.74
m p 0.938 γ2
Problem 10
Solution
n
1
The mean of the counts is N= ∑ N =1.05 x 10 4.
n 1 i
The standard deviation of a reading is
√
n
1
σ= ∑ ( N i - N )2 =0.037 x 10 4
n-1 1
If the Al foil is uniform, the counts taken at various locations should
follow the Poisson distribution with a standard deviation
∆N= √ N = √1.05 x 10 =0.01 x 10
4 4
Since the standard deviation of the readings (0.01 x 10 4) is more than
three times ∆ N , the foil cannot be considered uniform.
Write N=N 0 e as ln N =ln N 0−µ t .
−µt
dN
As N
=−µ t , we have
∆N
=µ∆t
N
∆N ∆t
Or N =µ t( t )
∆t 1 ∆N
As t =
µt N ,
For a given set of data, the smaller t is, the larger is the fractional
error o t.
Problem 11
Solution
= ( 1.19 x 10 ) x ( 1.6 x 10
6 -19
)
=1.904 x 10 -13 A
Problem 12
Charge ( 5 x 10 )( 1.6 x 10 )
10 -19
= =1.33 x 10-10 Amp.
second 60
−10
Average current=1.33 x 10 Amp .