GASTRIC-LAVAGE MACHINE
USER’S MANUAL
CONTENTS
Introduction ………………………………………………………2
Specification………………………………………………………2
Principle Diagram…………………………………………………2
Installation of Tubes……………………………………………7
Indicators……………………………………………………………8
Contraindication……………………………………………………10
Accessories……………………………………………………………10
After-sale service…………………………………………………10
5
-1-
Introduction
Specification
1. Working condition
Ambient Temperature: +5~+40℃
Relative Humidity: ≤80%
Atmospheric pressure: 860 hPa~1060 hPa
Supply Power: AC220V±22V、50Hz±1Hz
Input power: ≤80VA
Patient position: 60 to 100 cm higher than the bucket
Principle Diagram
F1-F4:
Solenoid valve AC24V
Dashed parts:
The pressure value
-2-
Function and instruction
The display indicates the status
of in and out of the stomach to
When the feeding stomach
facilitate to master the status at
liquid volume more than that of
any time.
the liquid volume out of the
stomach is observed, can, press
Press the working switch
the liquid volume balance key to
down, the work can be started
carry out outside point action or halted
under the automatic programming.
The connection process between the
For each press, the efficient
tubes and Input mouthpiece,
result to reduce the incoming
Input stomach, Output stomach
stomach liquid volume and to
and Output mouthpiece. For the
increase liquid volume out of the
details, please find figure 2.
stomach can be played in each
cycle with a difference of
200ml.After start of the liquid
volume balance function, the
liquid balance pilot light
twinkling. Please make
observation, if the single
efficiency is not apparent,
point-action once again can be
made after completing the one
cycle.
-3-
﹙1﹚When the machine is working, to set the
pressure value, you can press “positive
pressure” button or “negative
pressure ”button. Then the screen will flash,
and the machine will work under the factory
settings or the previous settings.
Figure 1
-4-
Exterior structure
Output stomach display
Counting reset
key
Pressure value increase
Liquid button
volume
balance Pressure value decrease
indicator button
Liquid volume
balance key
-5-
Operating instruction and process
step 1 step 2 step 3
Use a φ6 pipe of suitable Connect the “Input Use a φ 8 pipe of
length to connect the stomach”, the “Output suitable length to
Pipeline
“Input mouthpiece” and stomach” by two pieces connect the “Output
connection(
the other side of the pipe φ6 pipe separately. And mouthpiece” and the
Fig.1)
connects to the water connect the other side of other side connects to
container. the two φ 6 pipes by a the waste liquid
Y-shaped connector. container.
Adjustment Put the three-piece tube(twoφ6 pipes and Y-shaped tube ) into 1000ml measuring
glasses(filled with water) or “the test liquid”(optional accessory). Switch on
the machine and turn on the working switch. Check if all the tubes connect tightened
and the machine work normally.
-6-
Working 1. Put the three-piece tube(twoφ6 pipes and Y-shaped tube ) into clean water. Switch
process on the machine and keep it working after two working cycles or more times till
the pipes are free air.
2. Switch off the machine and check if all the pipes are well connected.
3. Put “counting reset key”, “counting display” shows zero.
4. Press “Working switch”, then the machine will work automatically.
5. When you find the amount of input stomach liquid is much more than the amount
of output stomach liquid, please press “Liquid volume balance key”.
6. When the machine stops, make sure it has powered off and take out the stomach
tube.
Attention When the machine is working, please observe the patient, the amount of
input stomach liquid and temperature of the liquid, the amount of output
stomach liquid. Once you have found unmoral situation, please stop the
machine immediately.
Installation of Tubes
-7-
连接示意图
-8-
Cleaning and Disinfection
After using, the machine and all the reusable parts should be clean and
disinfected completely.
a. Put the stomach tube into the 5000ml clean water. Make sure there is enough
clean water in the container which is connected with “Input mouthpiece”.
Switch on the machine and keep it work 4-5 cycles to empty all the waste
of the tubes.
b. Put the “Input mouthpiece” into the 5000ml clean water. Then put the stomach
tube into the mixed solution of disinfectant and oil cleaning agent (1.5
concentration of the water). And put the “Output mouthpiece” into another
water container. Then keep the machines working more than 10 cycles.
c. Instead of the mixed solution by clean water, repeat the above process and keep
the machine 3-5 cycles. Then empty all the water of the tubes.
d. Dip a duster cloth into disinfectant and use it to clean the housing of the
machine.
The recommend disinfectant:75% medicinal alcohol
Attention: To avoid pollute the stomach tube, please don’t put the “Input
mouthpiece” together with the “Output mouthpiece”, the three-piece tube into
the same container.
Indicators
-9-
General Fault Shooting
- 10 -
Contraindication
Our machine should not be used for people who swallow corrosive material, for
instance strong acid, strong base, or who have upper gastrointestinal tract ulcer
or upper gastrointestinal tract cancer. Manufacturers recommend people with
esophagus blockage, esophageal variceal bleeding and thoracic aortic aneurysm.
Accessories
PVC tube 3 sets
Stomach tube 3 pieces
Dust covers 7 pieces
Connector between stomach tube and PVC tube 2 pieces
Countersunk head 2 pieces
Power wire 1 piece
Fuse T2L 250V(Φ5mm×Φ20mm) 2 pieces
After-sale service
The SC-IIIB gastric lavage machine is packed and the stated requirements for storage and
the instruction for uses are well followed, if there is any fault happens to the machine and causes
failure to work normally within one year, We provide a free-of-charge warrant. But the other
accessories, such as pipes, stomach tubes are not included in the warranty.
- 11 -