Professional Documents
Culture Documents
KI Test1
KI Test1
1. Calibration Standards:
If needed, standard solutions can be prepared by dilution, 100 ppm and 10 ppm, from
1000 ppm standard.
On preparation add to each standard 0.5 nil of NaNO3 5 M.
2. Electrode Preparing:
n Ion-selective Electrode: take the protection cap off the electrode (black cap).
n Use syringe to add filling electrolyte solution into the electrode via the small
opening on the side of the electrode (just above the white marking). Make sure that no
spilling is done while filling.
n I bolding the electrode vertically, apply a light down pressure on the white cap (do
not turn the white cap), allowing a very small amount of electrolyte to exit from the
bottom junction to enable the sensor wetting.
3. Calibration Procedures:
1 /2