You are on page 1of 8

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/27223501

Model za usporedivost rezultata dobivenih s dva različita biokemijska


analizatora u laboratoriju akreditiranom prema normi ISO 15189

Article · January 2009


Source: OAI

CITATIONS READS

0 77

7 authors, including:

Marijana Miler Ana-Maria Simundic


University Clinical Hospital Center "Sestre Milosrdnice" Klinička bolnica "Sveti Duh"
55 PUBLICATIONS 460 CITATIONS 285 PUBLICATIONS 6,802 CITATIONS

SEE PROFILE SEE PROFILE

Mario Štefanović Elizabeta Topic


University Clinical Hospital Center "Sestre Milosrdnice" Dubrovnik International University
85 PUBLICATIONS 571 CITATIONS 114 PUBLICATIONS 1,235 CITATIONS

SEE PROFILE SEE PROFILE

Some of the authors of this publication are also working on these related projects:

Croatian society of medical biochemistry and laboratory medicine, Working group for Preanalytical Phase View project

Preanalytics View project

All content following this page was uploaded by Elizabeta Topic on 21 May 2014.

The user has requested enhancement of the downloaded file.


Izvorni stručni članak Original professional article

Model za usporedivost rezultata dobivenih s dva različita biokemijska


analizatora u laboratoriju akreditiranom prema normi ISO 15189
A model for results comparison on two different biochemistry analyzers in
laboratory accredited according to the ISO 15189
Marijana Miler, Ana-Maria Šimundić, Mario Štefanović, Dragica Ferenec-Ružić, Marina Kvaternik, Elizabeta Topić, Nada Vrkić
Klinički zavod za kemiju, Klinička bolnica Sestre milosrdnice, Zagreb
University Department of Chemistry, Sestre milosrdnice University Hospital, Zagreb, Croatia

Sažetak Abstract
Uvod: Laboratoriji akreditirani prema normi ISO 15189 koji imaju dva ili više Introduc tion: Laboratories accredited according to ISO 15189 standard
različitih analizatora ili postupaka ispitivanja za iste analite, moraju definirati which have two or more different analyzers or procedures for examinations,
mehanizam kojim osiguravaju usporedivost rezultata. Cilj ovog rada je pred- should define mechanism for comparison of results. The objective of this stu-
staviti model usporedbe rezultata hitnih biokemijskih pretraga s dva različita dy was to present the model for comparison of emergency tests on two diffe-
tipa automatiziranih analizatora za opće biokemijske pretrage. rent types of automated general biochemistry analyzers.
Materijali i metode: Koncentracije općih biokemijskih hitnih parametara Materials and methods: General biochemistry emergency tests were mea-
mjerene su u serumu i mokraći na dva automatizirana analizatora: Olympus sured in serum and urine on two automated analyzers Olympus AU2700 and
AU2700 i Olympus AU640 (Olympus, Hamburg, Germany) tijekom 60 dana. Olympus AU640 (Olympus, Hamburg, Germany) during 60 day period. Olym-
Kao referentni analizator služio je Olympus AU2700, a Olympus AU640 bio je pus AU2700 was used as the reference analyzer and Olympus AU640 as a bac-
pomoćni sustav. Za usporedbu rezultata korišteni su kontrolni uzorci na dvije kup system. Control samples on two levels were used for result comparison.
koncentracijske razine. Results: Average 60-day period bias was within acceptance criteria for all
Rezultati: U razdoblju od 60 dana prosječno odstupanje za sve pretrage bi- analyzes included in this model of comparability assessment. Only low con-
lo je unutar dopuštenih granica. U manje od 2% slučajeva kontrolni uzorci s trol concentrations for some tests were exceeding allowable biases in less than 2%
niskim vrijednostima odstupali su više od dopuštenih granica, dok uzorci viših cases and high level controls were never exceeding allowable biases.
koncentracija nisu niti jednom odstupali od dopuštenih kriterija. Conclusions: This model of results comparison ensures continuous analyti-
Zaključak: Ovaj model usporedbe rezultata osigurava neprekidnu kvalitetu cal quality of laboratory reports and improves overall efficiency of laboratory
analitičke faze izdavanja laboratorijskih nalaza te poboljšava ukupnu učinko- services.
vitost laboratorija. Key words: accreditation; analyzer comparison; internal quality control; la-
Ključne riječi: akreditacija; usporedba analizatora; unutarnja kontrola kvali- boratory error
tete; laboratorijska pogreška

Pristiglo: 13. kolovoza 2009. Received: August 13, 2009


Prihvaćeno: 4. rujna 2009. Accepted: September 4, 2009

Uvod Introduction
Norma ISO 15189 za medicinske laboratorije temelj je za The ISO 15189 standard for medical laboratories ser ves as
osiguravanje zahtjeva za sposobnošću i kvalitetom u bio- an inevitable tool for providing requirements for compe-
kemijskom laboratoriju. tence and quality in biochemistry laboratory.
U svakom je laboratoriju od neizmjerne važnosti osigurati In every laboratory it is of outstanding importance to
unutarnju kontrolu kvalitete (engl. internal quality control, ensure internal quality control (IQC) as well as to be in-
IQC). Laboratorij mora biti uključen i u vanjsku procjenu cluded in external quality assurance schemes (EQAS). If
kvalitete (engl. external quality assurance schemes, EQAS). laboratories accredited according to ISO 15189 standard

Biochemia Medica 2009;19(3):287–93


287
Miler M. i sur. Usporedivost biokemijskih analizatora - model
Miler M. et al. Comparability of biochemistry analyzers – a model

Ukoliko akreditirani laboratoriji imaju dva ili više analiza- have two or more different analyzers or procedures for
tora ili postupka za određenu pretragu, moraju definirati examinations, they should define mechanism for compa-
mehanizam kojim osiguravaju usporedivost rezultata (1). rison of those results (1). The comparability of different
Usporedivost različitih analitičkih sustava smanjuje ana- analytical systems minimizes analytical bias and enables
litičko odstupanje i omogućava uspoređivanje pojedinih comparison of individual test results with the population
rezultata pretrage s vrijednostima u populaciji u odnosu values based reference inter val or with cut-off value (2). It
na referentni inter vale ili granične vrijednosti (2). Uspore- also provides fast and reliable results reporting in labora-
divost ujedno omogućava brzo i pouzdano izdavanje na- tory, especially in emergency department, regardless to
laza, posebice u hitnom laboratoriju, a neovisno o anali- the analytical system which is used and functional. This
tičkom sustavu koji se upotrebljava ili je funkcionalan. Na ensures the best patient care and is key point for conti-
taj način omogućava se najbolja skrb za bolesnike, što je nuous improvement of laboratory accredited according
ključno za proces neprekidnog poboljšavanja u laborato- to ISO 15189 standard.
rijima akreditiranim prema ISO 15189. The objective of this study was to describe the model for
Cilj ovog rada bio je opisati model za usporedbu rezultata comparison of emergency tests on two different types of
hitnih pretraga dobivenih s dva različita automatizirana automated analyzers used in laboratory on daily basis.
analizatora na kojima se svakodnevno radi u laboratoriju.
Materials and methods
Materijali i metode
Setting
Ustanova University Department of Chemistry is the laboratory ac-
Klinički zavod za kemiju akreditiran je prema normi ISO credited according to ISO 15189 standard since 2007. It is
15189 od 2007. godine. Zavod je dio Kliničke bolnice „Ses- the department of Sestre milosrdnice University Hospital
tre milosrdnice” u Zagrebu, koja je ustanova tercijarne in Zagreb, Croatia, which is tertiary care medium-sized
zdravstvene zaštite, srednje veličine. U laboratorij se ti- hospital. In laboratory, approximately 30.000 samples per
jekom mjesec dana zaprimi oko 30,000 uzoraka od čega month is admitted, out of which 11.000 emergency sam-
je otprilike 11,000 hitnih uzoraka s općim biokemijskim ples with general biochemistry tests. On daily basis the-
pretragama. Dnevno se u hitni laboratorij zaprimi oko 350 re are about 350 emergency samples with biochemistry
biokemijskih uzoraka. tests.
Laboratorij posjeduje 2 tipa analizatora za određivanje Laboratory has two types of clinical chemistry analyzers
općih biokemijskih pretraga: Olympus AU2700 i Olympus for per forming general biochemistry tests: Olympus
AU640 (Olympus, Hamburg, Germany). Kao referentni AU2700 and Olympus AU640 (Olympus, Hamburg, Ger-
analizator služi Olympus AU2700, dok je drugi, Olympus many). Olympus AU2700 was used as the reference ana-
AU640, pričuvni sustav. lyzer while the other, Olympus AU640, was used as a con-
tinuous backup system for the reference analyzer.
Uzorci
Usporedivost je osigurana svakodnevno za sljedeće pret- Samples
rage u serumu: ukupni i konjugirani bilirubin, glukoza, Comparability is assured on daily basis for the following
kreatinin, ureja, ukupni proteini, enzimi (AST, ALT, LDH, serum analyzes: total and conjugated bilirubin, glucose,
CK, CK-MB i α-amilaza), kalij, natrij, kloridi, kalcij, C-reak- creatinine, urea, total protein, enzymes (AST, ALT, LDH,
tivni protein (CRP). Uz navedene pretrage u serumu, us- CK, CK-MB and α-amylase), potassium, sodium, chloride,
poređuje se i koncentracija glukoze te aktivnost amilaze calcium, C-reactive protein (CRP). Comparability is also as-
u mokraći. sured for amylase and glucose in urine.
Tijekom razdoblja od 60 dana neprekidno se provodila During the period of 60 days internal quality control pro-
unutarnja kontrola kvalitete na dvije razine i na dva razli- cedures were continuously performed on two levels with
čita analizatora. Koncentracija CRP određena je iz komer- two different analyzers. For determination of CRP concen-
cijalno dostupnih kontrolnih uzoraka ITA Control serum tration commercially available controls ITA Control serum
level 1 i 2 (Olympus, Hamburg, Njemačka), aktivnost ami- level 1 and 2 (Olympus, Hamburg, Germany) were used,
laze i koncentracija glukoze u mokraći određena je iz ko- for urine amylase and glucose commercially available Uri-
mercijalno dostupnih kontrolnih uzoraka Urine chemistry ne chemistry control 1 and 2 (Bio-Rad Laboratories, Ir vi-
control 1 i 2 (Bio-Rad Laboratories, Ir ving, CA, USA), a ak- ng, CA, USA) and for CK-MB commercially available CK-
tivnost CK-MB iz CK-MB Kontrola (Herbos Dijagnostika, Si- MB Kontrola (Herbos Dijagnostika, Sisak, Croatia). For all
sak, Hr vatska). Sve druge pretrage mjerene su iz Control other tests Control serum 1 and 2 (Olympus, Hamburg,
serum 1 i 2 (Olympus, Hamburg, Njemačka). Ciljne vrijed- Germany) were used. Target control values for all control
nosti kontrolnih uzoraka navedene su u tablici 1. samples are listed in Table 1.

Biochemia Medica 2009;19(3):287–93


288
Miler M. i sur. Usporedivost biokemijskih analizatora - model
Miler M. et al. Comparability of biochemistry analyzers – a model

TABLICA 1. Ciljne vrijednosti kontrolnih uzoraka i dopušteni raspon. TABLE 1. Control sample target values with acceptable range.

Target value (± 2SD range)


Test (Units)
L H
Total bilirubin (µmol/L) 25.6 (18.9–32.3) 115 (85.0–145)
Conjugated bilirubin (µmol/L) 19.4 (14.4–24.4) 68.0 (50.3–85.7)
Glucose (mmol/L) 5.26 (4.42–6.10) 12.5 (10.5–14.5)
Creatinine (µmol/L) 116 (90.5–142) 452 (353–551)
Urea (mmol/L) 6.34 (4.95–7.73) 27.0 (21.0–32.9)
Total protein (g/L) 39.9 (35.5–44.3) 74.3 (66.1–82.5)
AST (U/L 37°C) 44.2 (34.0–54.4) 128 (98.6–157)
ALT (U/L 37°C) 41.7 (32.1–51.3) 123 (94.5–151)
LDH (U/L 37°C) 151 (123–178) 537 (440–633)
CK (U/L 37°C) 150 (120–180) 387 (309–464)
CK–MB (U/L 37°C) 35 (27–45)
α–amylase (U/L 37°C) 84.5 (67.6–101) 241 (193–289)
Potassium (mmol/L) 3.75 (3.41–4.09) 6.44 (5.86–7.02)
Sodium (mmol/L) 118 (111–125) 149 (140–159)
Chloride (mmol/L) 90.2 (82.1–98.3) 112 (102–122)
Calcium (mmol/L) 2.26 (2.01–2.51) 3.30 (2.94–3.66)
CRP (mg/L) 18.3 (14.7–22.0) 32.0 (25.6–38.5)
Urine amylase (U/L 37°C) 61.5 (49.2–73.8) 172 (138–206)
Urine glucose (mmol/L) 1.94 (1.55–2.33) 16.3 (13.0–19.6)
* H – control with concentration above reference inter val; L – control with concentration within or below reference inter val.

Procjena usporedivosti Comparability assessment


Usporedivost je procijenjena izračunom odstupanja iz- Comparability was assessed by calculation of the test
mjerenih koncentracija prema sljedećoj jednadžbi: concentration bias according to the following equation:

conc. AU2700 – conc. AU640 conc. AU2700 – conc. AU640


Bias = × 100 Bias = × 100
conc. AU2700 conc. AU2700

u kojoj bias predstavlja mjeru slaganja između referen- where bias represents a measure of agreement between
tnog i pričuvnog analizatora, conc. označava izmjerenu reference and backup analyzer, conc. means concentra-
koncentraciju određene pretrage, a AU2700 ili AU640 je tion of the particular test, AU2700 or AU640 indicates
tip analizatora (Olympus AU2700 ili Olympus AU640). Re- type of analyzers: Olympus AU2700 or Olympus AU640.
zultati su pomnoženi sa 100 kako bi se dobila vrijednost Result was multiplied with 100 to obtain the percentage
u postotku. value.
Najveća dopuštena odstupanja za sve pretrage određe- Maximum allowable biases for all tests were defined ac-
na su prema vrijednostima koeficijenata varijacije koje je cording to the values of the coefficient of variation (CV)
odredilo Hr vatsko društvo medicinskih biokemičara u kri- for particular parameter given by the Croatian Society of
terijima vanjske procjene kvalitete. Medical Biochemists external quality assessment criteria.

Postupnik za usporedbu rezultata Workflow for the results comparison


Kontrolni uzorci mjereni su svakodnevno na oba anali- All control samples were measured on both analyzers (Ol-
zatora (Olympus AU2700 i AU640). Svaki dobiveni rezul- ympus AU2700 and AU640) on daily basis. Every control

Biochemia Medica 2009;19(3):287–93


289
Miler M. i sur. Usporedivost biokemijskih analizatora - model
Miler M. et al. Comparability of biochemistry analyzers – a model

tat upisan je na Levey-Jenningsov graf i provjerena mu je sample result was plotted on the Levey-Jennings chart
prihvatljivost prema Westgardovim pravilima (ciljna vri- and checked for acceptability according to Westgard ru-
jednost ± 2 standardne devijacije (SD)) (3,4). Ukoliko su les (target value ± 2 standard deviations (SD)) (3,4). If con-
rezultati za oba kontrolna uzorka bili prihvatljivi na obje trol sample test results were acceptable on both levels,
koncentracijske razine, uneseni su u model za usporedbu they were entered in the model for analyzer comparison.
analizatora. Ukoliko rezultati nisu bili unutar prihvatljivog If results were not within acceptable range, IQC procedu-
raspona, unutarnju kontrolu kvalitete trebalo je proves- re should be re-run with fresh control samples. Analyzes
ti sa svježim kontrolnim uzorcima. Pretrage s rezultatima with results outside acceptable range should be calibra-
izvan dopuštenih raspona, potrebno je kalibrirati te po- ted and IQC procedure per formed again. When control
novno provesti IQC. Ukoliko su rezultati kontrolnih uzo- values on both analyzers fulfilled required IQC criteria
raka na oba analizatora unutar zadanih odstupanja, ru- and are within acceptable bias, routine work could be
tinski rad s uzorcima bolesnika može se nastaviti. Ukoliko proceeded. If control values are not comparable, medical
vrijednosti na dva analizatora nisu usporedive, medicinski biochemists should check if their difference is clinically
biokemičar treba provjeriti je li razlika među koncentra- significant in comparison to inter-individual biological
cijama klinički značajna u usporedbi s inter-individualnim variations according to Westgard (5). If the difference is
biološkim varijacijama prema Westgardu (5). Ukoliko razli- not clinically significant, routine work may be continued.
ka nije klinički značajna, rutinski rad može se nastaviti. Za For tests with difference above inter-individual coefficie-
pretrage koje se razlikuju više od vrijednosti inter-indivi- nt of variation (CV), control samples should be re-run or
dualnih koeficijenata varijacije (CV), kontrolni uzorci mo- test calibrated (Figure 1).
raju se pustiti ponovno ili je testove potrebno kalibrirati
(slika 1).

Run controls on
AU2700 and
AU640
Not acceptable

Calibration of
unacceptable
controls

Check acceptability Not acceptable


of IQC

Acceptable

Check if bias of AU2700


Outside limits
and AU640 is within
limits of acceptance

Clinically
Within YES
significant
limits

NO
Proceed with
routine work

SLIKA 1. Postupnik za provođenje unutarnje kontrole kvalitete i za us- FIGURE 1. Workflow for internal quality control and results compara-
poredbu rezultata dobivenih s dva različita analizatora. AU2700 – Olym- bility on two different analyzers. AU2700 – Olympus AU2700; AU640
pus AU2700; AU640 - Olympus AU640; IQC – internal quality control - Olympus AU640; IQC – internal quality control

Biochemia Medica 2009;19(3):287–93


290
Miler M. i sur. Usporedivost biokemijskih analizatora - model
Miler M. et al. Comparability of biochemistry analyzers – a model

Rezultati Results
Srednje odstupanje tijekom razdoblja od 60 dana bilo je Average 60-day period bias was within acceptance crite-
unutar dopuštenih granica za sve pretrage koje su uklju- ria for all analyzes included in this model of comparability
čene u ovaj model procjene usporedivosti (tablica 2). Sa- assessment (Table 2). Only low control concentrations for
mo su kontrole niskih koncentracija za bilirubin, kloride, bilirubin, chloride, glucose, potassium, sodium, creatini-
glukozu, kalij, natrij, kreatinin i ureju u manje od 2% slu- ne and urea were exceeding allowable biases in less than
čajeva odstupali više od dopuštenih granica. Vrijednosti 2% cases, though those exceeding values were not clini-
koje su prelazile dopušteno odstupanje nisu bile klinički cally significant and as such were disregarded. Control
značajna odstupanja te su, kao tak ve, zanemarene. Kon- samples with high concentration were never exceeding
trolni uzorci s visokim koncentracijama niti jednom nisu allowable biases. Conjugated bilirubin concentration in
bile izvan dopuštenog odstupanja. Koncentracija konju- the low control level had the largest 60-day period ave-
giranog bilirubina u kontrolnom uzorku s niskim vrijed- rage (4.17%) and maximum bias (16.48%). The minimum
nostima imala je najveće srednje odstupanje (4,17%) kao average bias was obser ved for sodium (1.16%).
i najveće apsolutno odstupanje (16,48%). Najmanje sred-
nje odstupanje izračunato je za natrij (1,16%).

TABLICA 2. Prosječna izmjerena odstupanja za kontrolne uzorke TABLE 2. Average measured biases for low and high concentra-
s visokim i niskim koncentracijama analita s dopuštenim odstu- tion control samples with allowable bias values.
panjima.

Average measured bias (%)


Test Allowable bias** (%)
L* H*
Total bilirubin 1.42 ± 1.37 1.44 ± 1.28 12
Conjugated bilirubin 4.17 ± 3.48 3.90 ± 2.95 12
Glucose 2.25 ± 1.32 1.36 ± 1.25 5
Creatinine 2.60 ± 2.06 2.55 ± 2.10 8
Urea 3.67 ± 2.90 2.79 ± 2.30 10
Total protein 1.56 ± 1.70 1.41 ± 1.13 5
AST 2.22 ± 3.00 1.84 ± 1.57 15
ALT 3.63 ± 3.00 2.93 ± 2.10 15
LDH 2.58 ± 2.40 2.10 ± 1.66 12
CK 1.83 ± 1.73 2.56 ± 2.02 12
CK-MB 3.62 ± 2.90 4.59 ± 2.60 12
α-amylase 2.11 ± 1.71 2.02 ± 1.49 15
Potassium 1.50 ± 1.29 1.86 ± 1.30 5
Sodium 1.16 ± 0.95 1.14 ± 0.81 3
Chloride 1.37 ± 1.34 1.03 ± 0.96 4
Calcium 1.42 ± 1.18 1.32 ± 0.91 5
CRP 1.44 ± 1.18 2.04 ± 1.78 24
Urine amylase 3.41 ± 2.64 2.77 ± 2.21 15
Urine glucose 3.28 ± 3.40 3.03 ± 2.17 22
* H, control with concentration above reference inter val; L, control with concentration within or below reference inter val.
** According to Croatian Society of Medical Biochemists external quality assessment criteria.

Biochemia Medica 2009;19(3):287–93


291
Miler M. i sur. Usporedivost biokemijskih analizatora - model
Miler M. et al. Comparability of biochemistry analyzers – a model

Rasprava Discussion
Laboratoriji akreditirani prema normi ISO 15189 mora- Laboratories accredited according to ISO 15189 standard
ju osigurati kvalitetne rezultate za sve pretrage. Isto tako have to provide quality test results and also test repor-
moraju osigurati izdavanje nalaza u preporučenom vre- ts in assigned turn-around time (TAT) which is 1 hour for
menu (engl. turn-around time, TAT). TAT se može postići emergency tests. This TAT could be accomplished by the
pomoću dva analizatora koji su istovremeno dostupni i two analyzers both available and functional with compa-
funkcionalni, a rezultati su im međusobno usporedivi. rable results between themselves.
Cilj ovog rada bio je opisati model koji se provodi u la- The aim of this study was to describe the model imple-
boratoriju, a služi kao mehanizam za neprekidnu provje- mented in laboratory which is a mechanism for conti-
ru usporedivosti rezultata na dva različita automatizirana nuous verification of results comparability on two diffe-
analizatora. Provjera usporedivosti provodi se svakodnev- rent automated analyzers. This verification was per for-
no, što je određeno značajkama hitnih pretraga. med on daily basis, defined due to characteristics of the
Prema našim saznanjima, ovo je pr vi objavljeni primjer emergency test procedure.
modela za osiguravanje usporedivosti dva automatizira- To the best of our knowledge this is the first published
na analitička sustava. example of the model for ensuring the comparability of
Kao dopuštena odstupanja pričuvnog od referentnog two automated analytical systems.
analizatora preuzete su vrijednosti CV iz kriterija za vanj- For allowable biases of backup toward reference analyzer,
sku procjenu kvalitete Hr vatskog društva medicinskih CVs defined by the Croatian Society of Medical Bioche-
biokemičara. Kriteriji uključuju biološku varijaciju, a pred- mists external quality assessment criteria were used. Tho-
stavljaju i ukupno dopuštena odstupanja od srednje vri- se criteria include biological variation and represent the
jednosti (5). total allowed bias from the mean value (5).
Ovim se radom pokušalo dokazati da se na jednostavan The key finding of this study is that a simple algorithm
način uvođenjem postupnika može osigurati usporedi- could ensure continuous comparability and reliability
vost i pouzdanost rezultata hitnih pretraga te stalna pri- of emergency test results and constant preparedness of
pravnost dva analizatora. Model osigurava brzi prelazak s two analyzers. The presented model provides fast swit-
jednog na drugi analitički sustav ukoliko na referentnom ch from one to another analytical system due to possible
sustavu dođe fatalne greške. turndown episode of reference analyzer.
Jedan od glavnih indikatora kvalitete u akreditiranom la- One of the principal quality indicator in accredited labo-
boratoriju je osiguravanje pouzdanih i točnih rezultata s ratory is ensuring reliable and accurate results from the
prikladnim tumačenjem. Ovim modelom može se mjeriti samples with appropriate interpretation. This model cou-
učinkovitost laboratorija te istovremeno osigurati najtoč- ld measure laboratory per formance and also provide the
nije moguće rezultate (6), neovisno o tome na kojem su most accurate results (6), regardless on which automated
automatiziranom sustavu određivane koncentracije. Na system samples are determined. In this way, laboratory
ovaj način, laboratorij može uvelike doprinijeti postavlja- can greatly contribute to patient diagnosis and provide
nju dijagnoze bolesnika kao i osigurati važan udio najbo- the important part of the patients’ best health care. If la-
lje zdravstvene zaštite svakog pojedinog bolesnika. Kada boratory results are not comparable from different ana-
laboratorijski rezultati dobiveni s dva različita analitička lytical systems, possibility for incorrect diagnosis and
sustava ne bi bili usporedivi, povećavala bi se mogućnost poor quality of patient care is increased (7,8).
za postavljanje pogrešne dijagnoze, a ujedno bi se sma- This model also provides a trustful basis for ensuring lon-
njila kvaliteta brige za bolesnika (7,8). Model osigurava i gitudinal patients follow up. Thereby, model is one of the
pouzdanu podlogu za osiguravanje longitudinalnog pra- way for laboratory process and emergency test reports
ćenja pojedinih bolesnika. Prema tome, model može pos- standardization (9).
lužiti i kao jedan od načina standardizacije laboratorijskih Although the most laboratory errors are preanalytical or
postupaka i izdavanja nalaza hitnih pretraga (9). postanalytical phase (10), analytical errors due to system
Iako je većina laboratorijskih grešaka iz prijeanalitičke ili failure could be avoided with two equally ready and fun-
poslijeanalitičke faze (10), greške u analitičkoj fazi uzroko- ctional analyzers. The described model for comparability
vane kvarom sustava mogu se izbjeći pomoću dva jedna- could reduce analytical phase errors occurred due to so-
ko spremna i funkcionalna analizatora. Opisani model us- me failure of the reference analyzer.
poredivosti može smanjiti analitičke greške nastale zbog This model of results comparison ensures continuous
kvara referentnog analizatora. analytical quality of laboratory reports and improves ove-
Ovaj model usporedbe rezultata osigurava kvalitetu la- rall efficiency of laboratory ser vices. Reliable and compa-
boratorijskih nalaza te poboljšava ukupnu učinkovitost rable results on daily basis provide high quality of exami-
laboratorija. Pouzdani i usporedivi rezultati svakodnevno nation procedures and reporting of results in recommen-

Biochemia Medica 2009;19(3):287–93


292
Miler M. i sur. Usporedivost biokemijskih analizatora - model
Miler M. et al. Comparability of biochemistry analyzers – a model

osiguravaju visoku kvalitetu postupaka ispitivanja i izda- ded and appropriate time to the best of the patient ca-
vanje nalaza u preporučenom i prikladnom vremenu, a re.
sve za što bolju skrb bolesnika.

Adresa za dopisivanje: Corresponding author:


Marijana Miler Marijana Miler
Klinički zavod za kemiju University Depar tment of Chemistry
KB “Sestre milosrdnice” Sestre milosrdnice University Hospital
Vinogradska 29 Vinogradska 29
10 000 Zagreb 10 000 Zagreb, Croatia
e-pošta: marijana.miler@gmail.com e-mail: marijana.miler@gmail.com

Literatura/References
1. Medical laboratories – Particular requirements for quality and compe- 6. Simundic AM, Topic E. Quality indicators. Biochem Med 2008;18:
tence (ISO 15189:2003). 311-9.
2. Ricós C, Perich C, Minchinela J, Álvarez V, Simón M, Biosca C, et al. Appli- 7. Panteghini M, Forest JC. Standardization in laboratory medicine: New
cation of biological variation. A review. Biochem Med 2009;19:250-9. challenges. Clin Chim Acta 2005;355:1-12.
3. Barry PL. QC - The Levey-Jennings Control Chart. Available at: http:// 8. Guzel O, Guner EI. ISO 15189 Accreditation: Requirements for quality
www.westgard.com/lesson12.htm. Accessed August 1st, 2009. and competence of medical laboratories, experience of a laboratory I.
4. Kazmierczak SC. Laboratory quality control: Using patient data to as- Clin Biochem 2009;42:274-8.
sess analytical performance. Clin Chem Lab Med 2003;41:617-27. 9. Myers GL. Introduction to standardization of laboratory results. Steroi-
5. Ricós C, García-Lario JV, Alvarez V, Cava F, Domenech M, Hernández A, ds 208;73:1293-6.
et al. Biological variation database, and quality specifications for im- 10. Plebani M. Laboratory errors: How to improve pre- and post-analytical
precision, bias and total error. The 2008 update. Available at: http:// phases? Biochem Med 2007;17:5-9.
www.westgard.com/guest36.htm. Accessed August 7th, 2009.

Biochemia Medica 2009;19(3):287–93


293
View publication stats

You might also like