Service manual
Vorwort
Please carefully read each step of the procedure that is to be carried out
before beginning the servicing of the unit. Always use the correct tools
and the indicated measuring instruments. Any non-compliance with the
instructions and/or recommendation found in this technical
documentation can lead to a malfunctioning of the equipment or damage
to it.
Use only original replacement parts as supplied by FRITZ STEPHAN
GMBH.
56412 Gackenbach
Table of Contents
Table of Contents
Table of Contents .......................................................................................3
1 General Information ...........................................................................5
1.1 Equipment Designation and Manufacturer ..............................5
1.2 Technical safety inspections ....................................................5
1.3 Maintenance.............................................................................5
1.4 Intended use .............................................................................6
3 Maintenance of Akzent Color ..........................................................17
3.1 Maintenance Sets ...................................................................17
3.2 Maintenance of basic unit ......................................................18
3.2.1 Preparation ................................................................18
3.2.2 Maintenance of the gas inlet block ............................19
3.2.2.1 Changing the filter elements of the water trap .20
3.2.3 Replacing / Emptying the Water Trap .......................23
3.2.4 Maintenance of the Akzent Color vaporizer bracket .24
3.2.5 Changing filter of power unit fan ..............................24
3.2.6 Maintenance of connection for external PNT ...........26
3.2.7 Maintenance of operating mode selection switch .....27
3.2.8 Checking O2-Flush ....................................................28
3.2.9 Checking the power cable .........................................28
3.3 Pneumatic coupling ...............................................................29
3.3.1 Maintenance of pneumatic coupling .........................29
3.3.2 Maintenance check ....................................................31
3.4 Maintenance of Akzent Color ventilation system..................32
3.4.1 Procedure...................................................................32
3.4.1.1 APL valve ........................................................35
3.4.1.2 Internal PNTs ...................................................36
3.4.1.3 Valve covers.....................................................37
3.4.1.4 Valve lid ...........................................................38
3.4.1.5 CO2 absorbent lime container ..........................39
3.4.1.6 Bellows container .............................................40
3.4.1.7 CO2 absorbent lime sieve .................................41
3.4.1.8 Pressure gauge..................................................42
1 General Information
1.1 Equipment Designation and Manufacturer
1.3 Maintenance
Before each maintenance work or returning the unit for repair work,
please clean and disinfect all components.
Caution
Contraindication The AKZENT COLOR anesthesia unit is not suitable for patients with a
tidal volume outside the specified limits of 5- 1500 ml (limited by max
70 mbar airway pressure).
Fusing, 2 x 6.3 AT
Feeder circuit socket
Battery Rated voltage 12 V DC
Rated capacity 7,5 Ah
Fusing 10 AT
Operating time minimum 30 min
Gas supply Gas supply AIR 2,8 – 6,5 bar max. 110 L/min
pressure (kPa x 100) (without separate air
exit)
O2 2,8 – 6,5 bar max. 45 l/min
(kPa x 100) max. 90 l/min
(substituted)
N2O 2,8 – 6,5 bar max. 25 l/min
(kPa x 100)
The gases must be dry, and free of oil and dust. Use only pipelines which
match ISO 5359 and/or ISO 73961-1.
Pressure Ppeak
PEEP
Pmean
Volume MV
VTe
Gases InO2 18 - 100%
InCO2 0 – 20%
EtCO2 0 – 20%
InAGA 0 – 21,9%
(Sevo, Desf, Enfl, (dependent on AGA)
Isof, Halo,)
InN2O 0 – 85%
Optional measuring Timer 0 – 10h
value
Countdown 10 - 0h
Pmean 5 – 70 mbar
EtAGA 0 – 21,9%
(Sevo, Desf, Enfl, (dependent on AGA)
Isof, Halo)
EtN2O 0 – 85%
2 Safety Instructions
This symbol draws attention to important conditions and/or
circumstances.
The following symbols are later repeated at their relevant place in the
Operating Instructions that refer to safety information, and are to be
strictly adhered to.
Danger
Caution
The considerable force used when coupling the ventilation system can
cause pinching of the finger tips or skin folds.
Warning
Warning
Use only dry, purified gases (AIR, O2, N2O). Moisture in the gas lines
can lead to technical failure of the unit.
Warning
There must be no objects between the coupling device and the ventilation
system. Leakage or even a disabling of the inhalation anesthesia unit
function can be the result.
Warning
When moving the unit on its mobile stand, it must be ensured that the
wheel brakes have been released, and when setting up the unit again that
the brakes have been proper locked.
Warning
Warning
The bellows as well as its mounting device at the ventilation system must
be dry and grease-free during installation.
Warning
Warning
The protective cover for the SD card slot must always be in place.
Warning
The following maintenance sets for AKZENT COLOR are available from
Fritz Stephan GmbH.
3.2.1 Preparation
There are two different types of adapter (DIN or NIST). The maintenance
kit contains both O-rings (NIST) and sealing washers (DIN).
The O-rings for the NIST connections AIR, O2 and N2O are identical.
For the DIN connections, there are two types of sealing washers: for O2
and N2O (19516006) and for AIR (195160007).
When dismantling the water trap unit, make sure that the transparent
plastic containers are not damaged. Always use the supplied mounting
tool to unscrew the containers. Check that O-ring 195060077 fits
Caution
properly.
After changing the filter element, check the transparent plastic container
for any signs of cracks or damage.
Caution
O-ring 195060077
Emptying the water The water trap must be emptied when the full-mark has been reached.
trap
3. Turn the operation mode switch to the „Standby“ position.
4. Hold the water trap by the corrugated surface and remove it from the
holder.
5. Insert an empty syringe in the tubule in the connection at least 20 ml.
6. Draw out the water and remove the syringe.
7. Dispose of the syringe, in accordance with local hospital regulations,
in an environmentally manner.
8. Insert the water trap again and push until it stops.
Warning
Fig. 12: Plastic housing including filter mat of AKZENT COLOR power
unit
1. Replace filter of power unit fan. Pull plastic housing of filter element
away from frame by hand.
2. Press new plastic housing including new filter mat onto the frame
until you can clearly hear it click into position.
Before inserting new O-rings, grease these as well as the O-ring adapter
lightly with oxygen grease (Article No.: 170260600).
Always check that setting of the operating mode switch complies with
mode shown on AKZENT COLOR monitor. If this does not comply,
AKZENT COLOR must not be used until the mode switch has been
Caution
adjusted.
Check the following points once again after replacing the O-rings:
Fig. 26: Detailed view: containers for bellows and CO2 absorbent lime,
bellows and CO2 absorbent lime sieve
Fig. 27: Detailed view: APL valve, inspiration and expiration valves,
valve covers, internal PNTs
The two O-rings of the APL valves differ in size. Grease the O-rings
lightly with oxygen grease (170260600) before fitting in position.
Caution
The four O-rings of the internal PNTs differ in size. Grease O-rings
lightly with oxygen grease (170260600) before fitting in position.
Caution
The O-rings of the valve covers differ in size. Grease O-rings lightly with
oxygen grease (170260600) before fitting in position.
Caution
Fig. 32: Replacing O-rings at the lid of inspiration and expiration valves
Ensure that O-ring of CO2 absorbent lime sieve remains in position when
fitting the sieve itself later.
Caution
Ensure that O-rings fit firmly in place and are not crushed.
If the function check is negative, check all work steps again and repeat
the test. If this still fails to produce any results, contact Customer Service
of FRITZ STEPHAN GMBH, stating the serial numbers for AKZENT
Caution
COLOR and the ventilation system. Please also provide Customer Service
with a printout from the hyper-terminal.
4.1 Checklist
Pre-conditions for The last technical safety inspection has already been carried out as
testing scheduled. Visual check of the self-inspection sticker.
The unit is completely assembled and connected.
- Measured gas recirculation and test hose of the water trap.
- Anesthesia gas lead-off system.
The built-in rechargeable battery must be completely charged.
4.1 Checklist
Test OK Not OK
Visually inspect the equipment to ensure that it is in good condition:
Correct and complete assembly
Hygienic cleanliness
No recognisable external damage
Use of suitable and approved accessories
Test seal for regular technical inspection
Check for presence and proper operation of a separate manual ventilation
bag
Test connection to power source
(Turn on main power switch)
Power indicator lights up green
Battery indicator lights up green
Connection to gas supply
(central gas system pressure between 2.8 and 6.5 bar for all gases)
Connection to potential equalisation
Test OK Not OK
Connection to anesthesia gas transmission line
If present:
Check reserve compressed gas tank for sufficient residual pressure
4.1 Checklist
Test OK Not OK
Gas type test
Open the O2 mixing valve → O2 rises continuously to 100 V% (-5%) / N2O
= 0 V%
Open the air mixing valve → O2 approximately 21 V% / N2O = 0 V%
Open the N2O mixing valve → N2O rises continuously to 75 V% (± 5%)
(Note: If an internal anesthesia gas module for measuring N2O is not
present, an external device must be used.)
Using a test lung, check manual ventilation for proper operation
Using a test lung, check the respirator for proper operation
(measured values for pressure and volume are shown on the monitor)
Check the pressure limiting (APL) valve and its pop-off function for proper
operation
Check the inspiration and expiration valves in the respiration system
(Note: Test using manual ventilation (mode).)
If necessary, calibrate the electrochemical O2 cell, if present
Check alarm limit settings
Check alarm operation
Alarm triggered when O2 mains supply is unplugged
Alarm triggered when test lung is removed
Check secretion suction (condition and operation) if present
Table 3: Checklist
Should one of the tests not be passed, the equipment may not be put into
operation!
Danger
Test OK Not OK
The equipment check must be performed between successive narcosis
sessions.
Visually inspect the equipment to ensure that it is in good condition
Visually inspect the vaporizer(s)
Visually inspect the CO2 absorber
Drain condensation from breathing tubes and water traps if necessary.
Check secretion suction (condition and operation), and if necessary drain
and clean
Subset of functional tests that must be performed by the anesthetist before
or when narcosis equipment is connected to each patient:
Check the gas mixing unit (GMU)
Check correct connection of the hoses of the respiration system and manual
ventilation
Check the respiration system for leaks
Check manual ventilation for proper operation
Check the pressure limiting (APL) valve for proper operation
Check the inspiration and expiration valves in the respiration system
(Note: Test using manual ventilation (mode).)
Table 4: Check between successive narcosis sessions
Check that the different supply gases O2, Air and, depending on the
version, N2O, are connected correctly.
To do so, close all dosage valves of the gas mixing unit. Then open
the O2 dosage valve of the gas mixing unit to the maximum. The O2
concentration shown on the AKZENT COLOR monitor must be 95-99
%.
When this test is successful, close the O2 dosage valve. Then open the
Air dosage valve to the maximum. The O2 concentration shown on
the AKZENT COLOR monitor must be 21 %.
When using an AKZENT COLOR with N2O, close the Air dosage
valve, open the N2O valve completely and adjust O2 valve to 1 liter.
Check the N2O concentration (at the ratio of N2O:O2 /3:1) with
internal (ILCA) or external anesthesia gas module.
Check the function of the O2 flush button. The O2 concentration
shown on the monitor of the AKZENT COLOR must be 95-99 %.
If all tests are successful, the AKZENT COLOR is ready to use. If one of
the tests failed, check the supply gas connections again and repeat the
tests.
Connection to Akzent Connect RS232 interface of AKZENT COLOR via connecting cable
Color (Ref. Nr. 180960235) and USB adapter (Ref. Nr. 180160392) to USB
port of PC.
Start program Hyperterminal and make settings as described above.
By pressing the keys “CONTR” and “X” the connection is enabled. Then
insert password „56412“.
Enter Submenu
Type „st“ and press System test
„Return“
Type „Rcs“ and press Configuration status
„Return“
Type „r“ and press Display of measured values
„Return“
5.1 Systemtest
ValveTest started
Check V1
Current: 38
Check V2
Current: 38
Check VpCAL
Current: 77
P Offset Cal done
Check V3
Current: 39
Check Vflow
Current: 149
Check V5
Current: 38
Check V6
Current: 32
Check Vext_pnt
Current: 77
Result: 0x 0
ValveTest done
OffsetCalibration started
return result
OffsetCalibration done
Start Heater Check
Heater Check done
Start Temperaturesensor Check
Temperaturesensor Check done. O.K.
LeakageTest started
wait for user input
user confirmation
trying to increase pressure
measuring pressure
pressure ok
The values shown in system test can vary slightly due to different
component tolerances.
The values shown in measured value display can vary slightly due to
different component tolerances.
6 Firmware Update
6 Firmware Update
Switch off AKZENT COLOR
Remove cover from SD card slot at the rear of AKZENT COLOR
Firmware Update
Type in Password
Password:
11
Confirm
Once the password has been entered correctly the firmware can be started
Firmware Update
Start
7 Pneumatics
7 Pneumatics
8 Circuit diagram
8 Circuit diagram
8 Circuit diagram
Hardware: Failure MA high Electronic failure Start unit again and check
Valve failure according to user manual
Contact FRITZ STEPHAN
GMBH Customer Service
Upper P-limit! PA high Ppeak is above the pre-set Check pre-set upper limit
upper limit for PAW PAW and if necessary
change
Check patient hose system
for stenosis
PEEP >PAW, PA high PEEP setting lies above Check patient’s respiratory
lower limit! the pre-set lower limit for tract including tubes and
PAW hoses
Check pre-set PEEP and if
necessary change the pre-
set ventilation mode
Check pre-set lower limit
PAW and if necessary
change
Check hose system for
blockage, water droplets
High MV! PA high Measured MV above the Check pre-set upper limit
pre-set upper limit value for MV and if
necessary change
Compare the pre-set values
for VTe and Frq with that
of the measured MV value