You are on page 1of 6

Available online at www.sciencedirect.

com

Surface & Coatings Technology 202 (2008) 1668 – 1673


www.elsevier.com/locate/surfcoat

Surface characterization of AISI 316L biomaterials obtained


by electropolishing in a magnetic field
T. Hryniewicz a,⁎, R. Rokicki b , K. Rokosz a
a
Politechnika Koszalinska, Division of Surface Electrochemistry, Raclawicka 15-17, PL 75-620 Koszalin, Poland
b
ElectroBright, Macungie PA, USA
Received 12 April 2007; accepted in revised form 16 July 2007
Available online 7 August 2007

Abstract

The paper deals with comparison of the results obtained by electropolishing AISI 316L surfaces in an externally applied magnetic field with those
of a standard electropolishing process. All electrochemical investigation methods, electrochemical impedance spectroscopy (EIS), polarization curve
measurements, and surface analytical methods such as surface roughness measurements and scanning electron microscopy were carried out to
determine the efficacy of electropolishing in a magnetic field.
© 2007 Elsevier B.V. All rights reserved.

Keywords: Magnetoelectropolishing; Modes of treatment; Stainless steels; Surface roughness comparison; Corrosion behaviour

1. Introduction and standard electropolished metallic surfaces. The most prom-


inent difference, which will likely have the most profound
Electropolishing (ECP) [1–7] is one of the most important implications on bio- and haemocompatibility, is difference in
surface treatment processes used for stainless steels. The ECP static water contact angle as reported previously by the authors
process is governed by many different mechanisms and depends [8]. The magnetoelectropolished surfaces show a remarkable
on many parameters [1–4,6]. Given its variability, no single reduction of water contact angle, which in case of NitinolT,
theory for attaining the best electropolish covers all the com- resembles the water contact angle of pure titanium oxide.
plexities. A modified technology [8–10] is proposed to obtain
bare metal surfaces for implants, stents and medical devices. 2. Metallic biomaterials surface characteristics
The effects of an externally applied magnetic field during ECP
(magnetoelectropolishing) of biomaterials, such as 316L vm In recent years, interest in metallic biomaterials such as
stainless steel, on their surface properties were investigated. The medical grade austenitic stainless steels, tantalum, and NitinolT
results have been compared with standard electropolished bio- applications has been steadily growing [8–14]. The main reason
materials in the same electrolyte under the same process con- for this is their remarkable haemo- and biocompatibility prop-
ditions. To evaluate surface properties, the following techniques erties and high corrosion resistance. The good biocompatibility
were used: SEM (Scanning Electron Microscopy), CAM (Contact originates from their chemical surface film composition. The
Angle Measurement), AES (Auger Electron Spectroscopy), medical grade AISI 316L vm stainless steel after electrolytic
polarization curves, EIS (Electrochemical Impedance Spectros- polishing forms very thin, of a few-nanometer, compact oxide
copy), and surface roughness measurements. These analyses film resistant against corrosion in the presence of physiological
showed significant differences between magnetoelectropolished human body fluids environments. A content of around 44%
titanium in NitinolT is responsible for the spontaneous creation
of very titanium rich oxide, which is also present on pure tita-
⁎ Corresponding author. nium [7] and other titanium containing alloys. The oxide films
E-mail addresses: Tadeusz.Hryniewicz@tu.koszalin.pl (T. Hryniewicz), formed on metallic biomaterials are of paramount importance for
info@electrobright.com (R. Rokicki). its biocompatibility in orthopaedic, dental and biliary [11–20]
0257-8972/$ - see front matter © 2007 Elsevier B.V. All rights reserved.
doi:10.1016/j.surfcoat.2007.07.067
T. Hryniewicz et al. / Surface & Coatings Technology 202 (2008) 1668–1673 1669

applications as well as haemocompatibility in cardiovascular study, the ECP process has been carried out, first, on the plateau
and peripheral stents and heart valves. level, and then under oxygen evolution regime. The purpose
Whether it is a bare stent or with a coating material, 316L SS of the work was to achieve an electropolished surface of the
is the most commonly used metal for stents. It has well-suited sample exhibiting advantages of the treatment under different
mechanical properties and excellent corrosion resistance (carbon conditions, concerning reduced microroughness, better surface
content below 0.030 wt.%) making it the preferred material for wetting and increased surface energy, reduced and more
this application [21–26]. According to Mani et al. [26], out of the uniform corrosion resistance, minimization of external surface
eight coronary stents approved by the US Food and Drug soiling and improved cleanability in shorter time periods.
Administration (FDA), seven are made from 316L SS. However,
even if the supremacy of 316L SS platforms for making stents is 3. Material and experimental procedures
evident, biocompatibility is an issue with bare SS stents.
Surface characteristics of a stent material, which influence 3.1. Materials
thrombosis and neointimal hyperplasia, include surface energy,
surface texture, surface potential, and the stability of the surface The medical grade AISI 316L vm stainless steel, as the most
oxide layer [26,27]. The surface properties of a material usually frequently used biomaterial, was used for the study. Two sets of
depend on the surface treatment. Different surface treatments AISI 316L vm stainless steel samples punched of a sheet-steel of
can produce different surface textures and surface chemistries on about 2 mm thick and 18 mm in diameter have been used for the
the same material surface [27]. This can provide different surface investigations of mass loss vs. magnetic field strength, surface
energies to the material surface. To improve the corrosion re- film, roughness, wettability, corrosion behaviour, Auger and
sistance, different films (Ti or Ta) are coated on 316L SS other surface examinations. Also the standard ECP experiments
substrate by physical deposition sputtering [28]. However, when were carried out in parallel under the same conditions on the
the material is plastically deformed, cracks appear in the coating same material cut to evaluate the differences between these two
[28]. This kind of acute change (cracking) in surface mor- processes. Mass loss study results, concerning the effect of a
phology might pose a serious threat when the material is exposed magnetic field strength have been reported elsewhere by the
to in vivo conditions. The nature of the coating may be bio- authors [8,9].
compatible, but, if the coating loses its integrity during the stent
placement and expansion, it may cause adverse effects. 3.2. Electropolishing treatment
The stability of the surface oxide layer directly influences the
biocompatibility of a material as the surface layer acts as a barrier The studies of electropolishing were carried out under two
to the release of ions from the bulk materials underneath the modes: (1) on the plateau level, and (2) beyond plateau, under
surface. Also the endothelial cell damage caused by the release oxygen evolution regime (Fig. 1).
of very low concentration metal ions may be considered as a The electrolytic polishing was performed both in the absence
potentially toxic effect [27]. The stability of oxide layer is also a and in the presence of a magnetic field. For processing, both
key to several of the surface characteristics. It influences surface rigid or flexible permanent magnets and electromagnets may be
energy by providing hydrophilicity to a material surface [8,9] applied. The effective magnetic field is assumed to be in the
and surface potential by preventing the release of electrons. range between 0.1 T and up to 1 T. For the experiments, a
Since the stability of the natural surface oxide layer in 316L SS constant external magnetic field below 1 T was applied to the
(and NiTi) is not very high, the possibility of metal ion release is ECP system by neodymium ring magnets (Fig. 2). For both
enhanced. processes, standard ECP and MECP, the same type of a pro-
The quest to develop totally passive metallic biomaterials, with prietary electrolyte was applied, being a mixture of sulphuric
improved fatigue resistance properties continues. Among the new and orthophosphoric acids. The bath was unstirred during the
approaches used to achieve this goal are: electrolytic polishing, process.
anodization, acidic and/or basic chemical etching, chemical
passivation, heat treatment in different gas atmospheres, ion
implantation, etc. [14–36]. All of these processes concentrate on
creating almost 100% Ni free amorphous, thin oxide film, which,
at least theoretically, should not create negative body response,
should be mechanically compatible with bulk constituents, and
should minimize Ni release to the host environment. Here we
would like to present some advantages and specific features
obtained after magnetoelectropolishing (MECP) [8–10], in
comparison with the surface after a standard electropolishing
(ECP).
The process of ECP in the presence of magnetic field has
been investigated by the authors previously [8–10,37–39]. The Fig. 1. Anodic polarization curve for AISI 316L stainless steel and the
ECP system with an externally applied magnetic field has been modes of electropolishing: (a) on the plateau level, (b) in the range of oxygen
utilized to enhance, or retard, the dissolution process. In this evolution.
1670 T. Hryniewicz et al. / Surface & Coatings Technology 202 (2008) 1668–1673

Fig. 2. Magnetoelectropolishing system schematic: A — anode, K — cathode,


M — magnet or electromagnet, W — steel wire of cylindrical shape.

Fig. 4. Results of surface roughness parameters Ra and Rz on AISI 316L vm


3.3. Roughness measurements stainless steel samples, after: ECP — electropolishing, and MECP — magneto-
electropolishing (the oxygen evolution mode).
A computerized HOMMEL TESTER T800 system of
Hommelwerke GmbH for roughness measurement was used represented as a vector Z(ω) whose real part Re(Z) is plotted on
for the study of surface roughness. The comparison of the the x-axis and the imaginary part Im(Z) on the y-axis of the
roughness after both standard ECP and MECP were carried out complex plane whereby the vector Z(ω) forms an angle ϕ with
on the samples in two perpendicular directions. the real axis. A straight line parallel to the y-axis indicates a
perfect surface coating with zero corrosion whereas any cor-
3.4. Corrosion measurements roding surface displays a semicircle as the geometrical locus of
the impedance Z(ω) at one and only one frequency ω. Hence
Corrosion studies after electropolishing on the plateau level the radius of the semicircle provides a measure of the corrosion
and beyond in the oxygen evolution regime were carried out in a rate, i.e. a small radius is indicative of a high corrosion rate and
3% NaCl and in Ringer's solutions at 25 °C, both in the vice versa a larger radius signifies an increase in corrosion
presence (MECP) and absence (ECP) of a magnetic field. resistance.
During the study, both electrochemical impedance spectroscopy
(EIS) and measurement of polarization curves were performed 3.5. SEM surface examinations
on the electropolished 316L samples surfaces. The electro-
chemical system used for the corrosion measurements consisted Scanning electron microscopy was used to investigate the
of the potentiostat ATLAS 98 with the software IMP98, current surfaces of the stainless steel samples after ECP and MECP. The
platinum electrode Ept-101, and the saturated calomel electrode apparatus used was Jeol ISN-550LV.
EK-101P used as a reference. The EIS results were obtained
each time after holding the samples at an open circuit potential 4. Experimental results
for 60 min.
Electrochemical impedance spectra can be evaluated using The magnetic field used in the electrolytic polishing is
so-called Nyquist plots in which the impedance Z is usually affecting the surface roughness. Results of surface roughness

Fig. 3. Results of surface roughness parameters Ra and Rz on AISI 316L vm Fig. 5. EIS characteristics: corrosion results of 316L in Ringer's body fluid.
stainless steel samples, after: ECP — electropolishing, and MECP — magneto- Electropolishing on the plateau level: a — after standard electropolishing, b —
electropolishing (the plateau level mode). after magnetoelectropolishing.
T. Hryniewicz et al. / Surface & Coatings Technology 202 (2008) 1668–1673 1671

Fig. 8. Polarisation curves: corrosion characteristics of 316L SS in 3% aqueous


soln. NaCl after electropolishing under the oxygen evolution mode: a — after
standard electropolishing, b — after magnetoelectropolishing.
Fig. 6. Polarisation curves: corrosion characteristics of 316L in Ringer's body
fluid. Electropolishing on the plateau level: a — after standard electropolishing,
b — after magnetoelectropolishing. EIS results coming from the ECP on the plateau level appear to
be totally independent of the magnetic field applied during
measurements, comprising two modes of electropolishing, on electropolishing.
the plateau level (a), and in the oxygen evolution regime (b), are Hence more attention was paid to the process of electro-
shown in Figs. 3 and 4, respectively. polishing beyond the anodic curve plateau, under oxygen
The roughness measurements were performed on three pairs evolution regime. The corrosion studies were carried out in two
of stainless steels samples electropolished in the absence (ECP), solutions: Ringer's physiological fluid substitute solution, and
and in the presence of magnetic field (MECP). Comparison of aqueous 3% NaCl solution, in both cases at 25 °C. The EIS
the surface roughness results obtained after electropolishing corrosion results of 316L vm sample, performed at a stationary
indicate no improvement in case of magnetic field used in the potential, are presented in Fig. 7.
process carried out on the plateau level (Fig. 3). On the other The EIS results obtained for two different surface treatments
hand, all surface roughness parameters (Ra, Rz, etc) have been reveal a considerable difference in the corrosion behaviour of
reduced after MECP beyond the plateau level, under oxygen the same 316L vm stainless steel sample surface. In case of 3%
evolution mode of treatment (Fig. 4). NaCl solution, the Nyquist plots obtained indicate a better
The corrosion behaviour of 316L vm sample surface was corrosion resistance after ECP in the absence of magnetic field
also studied concerning two modes of electropolishing, on the (Fig. 7a,b), but, what is more important for biomaterials, a better
plateau level (a) and beyond (b), in the absence and in the corrosion resistance after ECP in the presence of magnetic field
presence of magnetic field. The corrosion investigation results (Fig. 7c,d), is observed in the Ringer's solution.
obtained in Ringer's solution are presented in Figs. 5 and 6, and Two pairs each of polarization curves, in 3% NaCl solution,
show that there is no effect of the magnetic field, neither in the and in the Ringer's solution, are presented in Figs. 8 and 9,
EIS characteristics nor the polarization curves. Moreover, the respectively. A very slight change in the course of curve b
against curve a is observed in Fig. 8, meaning no basic corrosion

Fig. 7. EIS characteristics after electropolishing under the oxygen evolution


mode: a, b — corrosion results of 316L in 3% aqueous NaCl soln.: a — after Fig. 9. Polarisation corrosion characteristics of 316L SS in Ringer's body fluid
MECP, b — after ECP; c,d — corrosion results of 316L in Ringer's body fluid: after electropolishing under the oxygen evolution mode: a — after standard
c — after ECP, d — after MECP. electropolishing, b — after magnetoelectropolishing.
1672 T. Hryniewicz et al. / Surface & Coatings Technology 202 (2008) 1668–1673

the dissolved material or the composition of the electrolyte used.


Instead, it depends on the strength of the externally applied
magnetic field. It is important to note, however, that this effect
holds only when the electropolishing process is performed
below the oxygen evolution regime (plateau region), (see Fig. 1).
In the case of fixed-potential electropolishing performed beyond
the plateau current densities, in the oxygen evolution regime, the
experimental data show the process to be completely opposed to
the one described above.
The use of an externally applied magnetic field in the ECP
process provides the treated surface with some new properties
and better characteristics, concerning microroughness, hydro-
philicity, corrosion resistance, and oxide film morphology. The
higher level of metal finishing seems to be very interesting for
multiple applications, such as e.g. medical stents and implant
devices. The addition of the external magnetic field also dras-
tically minimizes microtopography by lowering microroughness
(see Fig. 4) and minimizing actual surface area in micro and
nanoscales of the various metallic materials [8].
From a practical point of view the more important features of
influence of a magnetic field used during an ECP process are the
alteration of morphology and chemistry of the finished surface.
The main reason for utilizing an ECP process is to improve the
quality of the treated surface and the incorporation of a magnetic
field during the ECP process provides an enhanced opportunity
to accomplish the desired results. These modified features of the
process can be observed in Fig. 4 and Fig. 10. It improves
corrosion resistance (Fig. 7) and prevents Ni ion leaching from
the bulk material to the host environment, which can be a
potential cause of allergic reaction.
Fig. 10. SEM micrographs of 316L vm samples: (a) after ECP, magnification Some other important findings were the corrosion results
1000×, (b) after MECP, magnification 1000×.
obtained on 316L SS samples in aqueous 3% NaCl solution in
comparison with the results in the Ringer's body fluid. It appears
improvement of 316L steel in 3% NaCl aqueous solution after its that the ECP carried out in magnetic field improves corrosion
magnetoelectropolishing. On the other hand, the same study resistance of 316L SS surface in the Ringer's solution, but
results in Ringer's body fluid (Fig. 9) show a considerable shift unfortunately it decreases the corrosion resistance in aqueous
on the potential axis, from − 1500 mV to about − 1000 mV 3% NaCl solution. It is worth to find out what is the reason of
(curve b in Fig. 9), meaning there is an increase in corrosion such a change in magnetic field effects. This fact requires further
resistance of 316L after magnetoelectropolishing in comparison study.
with the same 316L SS after a standard electropolishing. These The studies excluded any reason for conducting the ECP
polarization curves results well coincide with the results process on the plateau level. Much bigger discrepancy in the
obtained on Nyquist plots presented in Fig. 7. surface roughness in two perpendicular directions has been noted
Fig. 10 show SEM micrographs of 316L surfaces after (see Fig. 3). On the other hand, no change in corrosion behaviour
treatment in the oxygen evolution regime during ECP (Fig. 10a) between standard ECP and MECP was observed (see Fig. 5). That
and MECP (Fig. 10b). A slightly smoother and darker 316L SS may mean no change in the surface film obtained after MECP
sample surface has been obtained after magnetoelectropolishing process on the plateau level (without oxygen evolution).
(Fig. 10b). The darker colour suggests (Fig. 10b) increased It is known [40] that the surface layer of 316L SS after
surface energy as a result of a change in surface film composition electrolytic polishing (ECP) consists of oxides and hydroxides
after MECP [39]. of chromium and iron. The atomic ratio of Cr: Fe in the 316L SS
bulk equals 1: 3.63, the same ratio after abrasive polishing is 1:
5. Discussion 2.10, whereas after standard electropolishing it equals 1: 1.71
[39,40]. Our own investigation of the 316L SS surface layer after
The effect of an external magnetic field in ECP in two modes magnetoelectropolishing indicates on the ratio Cr: Fe to be 1:
of treatment was investigated. In the MECP process [8–10], an 0.71 [39]. It is much better than after a standard ECP but the
externally applied magnetic field works in two distinctive ways: thickness of the surface film has changed drastically. It is well
either enhancing or retarding the rate of dissolution. The change known [1,2,4] that the thickness of the surface layer after many
in dissolution rate does not depend on the magnetic properties of mechanical and chemical treatments are about two orders higher
T. Hryniewicz et al. / Surface & Coatings Technology 202 (2008) 1668–1673 1673

(∼10− 7 m) than after electrolytic polishing (∼10− 9 m). After a [6] P. Dettner, Electrolytic and Chemical Polishing of Metals, Ordentlich
standard ECP it is still two to three times higher than the Publishers Second Printing, May 1988.
[7] R. Rokicki, Met. Finish. 88 (2) (1990) 69.
thickness after MECP. It seems to be not only the composition of [8] T. Hryniewicz, R. Rokicki, K. Rokosz, Met. Finish. 104 (12) (2006) 26.
the surface layer, that is much better after MECP with higher [9] R. Rokicki, Med. Dev. & Des. Ind. 1 (3) (2006) 1.
contents of Cr2O3, to have a decisive influence on the 316L SS [10] R. Rokicki, US Patent Appl. #2006124472 (USPTO), 15 June, 2006.
corrosion behaviour. [11] J. Lévesque, D. Dubé, M. Fiset, D. Mantovani, Adv. Mater. Process. 162
(9) (2004) 45.
A considerable change in corrosion behaviour of biomater-
[12] W. Fender, R. Brown, Adv. Mater. Process. 163 (4) (2005) 36.
ials electropolished in the presence of a magnetic field is [13] C. Wang, S. Chen, J. Serb. Chem. Soc. 66 (7) (2001) 477.
meaningful. This improved corrosion behaviour well coincides [14] C.C. Shih, S.J. Lin, Y.L. Chung, Y.Y. Su, J. Biomed. Mater. Res. 52 (2)
with the Auger results of AISI 316L SS presented elsewhere (2000) 323.
[39]. [15] C.C. Shih, C.M. Shih, Y.Y. Su, S.J. Lin, J. Biomed. Mater. Res. 67A (2003)
1320.
[16] E. Krasicka-Cydzik, K. Kowalski, I. Glazowska, J. Achiev. in Mater.
6. Conclusion Manuf. Eng. 18 (1–2) (2006) 147.
[17] A.W. Hassel, Min. Invas. Ther Allied Technol. 13 (4) (2004) 240.
This study shows the difference in physical and chemical [18] S.A. Shabalovskaya, Bio-Med. Mater. Eng. 12 (2002) 69.
properties between standard electropolished and magnetoelec- [19] M.F. Maitz, I. Tsyganov, M.T. Pham, E. Wieser, J. Biomater. Appl. 17 (4)
tropolished biomaterials surfaces. All the above discussed (2002) 303.
[20] A. Taylor, Metals, in: U. Sigwart (Ed.), Endoluminal stenting, W.B.
changes were achieved by applying a magnetic field to an ECP Saunders Company Ltd, London, 1996, p. 28.
process. The obtained results indicate more uniform surface [21] H.T. Spijker, R. Graaff, P.W. Boonstra, H.J. Busscher, W. van Oeveren,
roughness and better corrosion resistance of 316L SS in the Biomaterial 24 (26) (2003) 4717.
Ringer's body fluid. These improvements may have a pro- [22] W. Kajzer, A. Krauze, W. Walke, J. Marciniak, J. Achiev. Mater. Manuf.
found influence on bio- and haemocompatibility of metallic Eng. 18 (1–2) (2006) 115.
[23] P.N. Sawyer, Nature 206 (1965) 1162.
biomaterials. [24] P.N. Sawyer, Ann. N.Y. Acad. Sci. 146 (1968) 49.
An improved surface technology has been developed which [25] J.W. Arnold, G.W. Bailey, Poultry Sci. 79 (1) (2000) 1839.
is the use of a magnetic field in the process of electropolishing [26] G. Mani, M.D. Feldman, D. Patel, C.M. Agrawal, Biomaterial 28 (2007)
much beyond the plateau level. The promising results have been 1689.
presented concerning the basic biomaterial, the medical grade [27] J. Palmaz, Tex. Heart Instr. J. 24 (3) (1997) 156.
[28] F. Macionczyk, B. Gerold, R. Thull, Surf. Coat. Technol. 142–144 (2001)
stainless steel of type 316L. Since the general mechanisms 1084.
underlying the interaction between a metal and tissue/blood are [29] I.H. Lee, H.B. Lee, J. Biomed. Mater. Res. 41 (8) (1998) 3104.
still not completely understood, the biocompatibility and haemo- [30] R.A. Silva, I.P. Silva, B. Rondot, J. Biomater. Appl. 00 (2006) 1.
compatibility of metallic stents still remains an issue. These tasks [31] H. Zhao, J.V. Humbeeck, J. Sohier, I.D. Scheerder, J. Mater. Sci., Mater.
Med. 13 (10) (2002) 911.
are expected to become the next step of the studies.
[32] E. Ruckenstein, S.V. Gourisankar, J. Colloid Interface Sci. 101 (2) (1984)
436.
Acknowledgement [33] J. Seeger, M. Ingegno, E. Bigatan, N. Klingman, D. Emery, C. Widen-
house, et al., J. Vasc. Surg. 22 (3) (1995) 327.
Mr Zbigniew Kuklinski is acknowledged for providing SEM [34] J. Palmaz, A. Benson, E. Sprague, J. Vasc. Interv. Radiol. 10 (4) (1999) 439.
micrographs of the biomaterials surface examined (Fig. 10). [35] A. Colombo, F. Airoldi, J. Invas. Cardiol. 15 (10) (2003) 566.
[36] J. Hecker, L. Scandrett, J. Biomed. Mater. Res. 19 (4) (1985) 381.
[37] T. Hryniewicz, R. Rokicki, K. Rokosz, Metal Surface Modification by
References Magnetoelectropolishing, Proc. of 16th International Metallurgical &
Materials Conference METAL 2007, Surf. Eng., vol. D2, Hradec nad
[1] T. Hryniewicz, Fizykochemiczne i technologiczne podstawy procesu elek- Moravici, Czech Republic, 22–24 May 2007, p. 1.
tropolerowania stali, Monografie, vol. 26, WSI Koszalin, 1989, 161 pages. [38] T. Hryniewicz, R. Rokicki, K. Rokosz, Investigation of Corrosion
[2] T. Hryniewicz, R.H. Muller, C.W. Tobias, A Study of Electropolishing of Resistance of 316L Austenitic Steel after Electropolishing/Badanie
Ferrous Alloys Using Rotating Disk Electrodes, Materials and Molecular odpornoœci korozyjnej stali austenitycznej 316L po polerowaniu elek-
Research Division, Lawrence Berkeley Laboratory, University of California, trolitycznym, Forum Motoryzacji, Słupsk, 25 May 2007, Mater. X Słupsk.
Berkeley CA 94720, LBL-12879, UC-25, June 1981, (122 pages), prep. for [39] T. Hryniewicz, S.J. Fernandes, R. Rokicki, Comparative Auger Studies of
the US Department of Energy under Contract W-7405-ENG-48. 316L Stainless Steel, PK + IST + EBM Project, November, 2006 Lisboa.
[3] T. Hryniewicz, Surf. Coat. Technol. 64 (2) (1994) 75. [40] G. Selvaduray, S. Trigwell, Effect of Surface Treatment on Surface
[4] T. Hryniewicz, in: T. Hryniewicz (Ed.), Elektrochemia dla Inżynierii Characteristics of 316L Stainless Steel, Proc. of the Confer. Materials and
Powierzchni, Koszalin University of Technology, Koszalin, 2005, 386 pages. Processes for Medical Devices, Nov. 14–18, 2005, Boston MA.
[5] A. Baron, W. Simka, G. Nawrat, D. Szewieczek, A. Krzyżak, J. Achieve.
Mater. Manuf. Eng. 18 (1–2) (2006) 55.

You might also like