You are on page 1of 32

TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 1

Turbidimetry in Particle Size 8 On-line Particle Size Determination


Using Turbidimetry 26
Analysis 9 Other Considerations 27
9.1 Choice of the Methods 27
Theodora Kourti 9.2 Corroboration of Results 28
Department of Chemical Engineering, McMaster 10 Conclusion 28
University, Hamilton, Ontario, Canada L8S 4L7 Abbreviations and Acronyms 29
Related Articles 29
References 29
1 Introduction 2
2 Theoretical Background 5
2.1 Turbidity and Particle Size 5 The size of particles in a suspension can be estimated
2.2 The Scattering Coefficient 6 by measuring the turbidity of the suspension. Turbidity
2.3 Turbidity and Moments of the measures the attenuation of a beam of light traveling
Particle Size Distribution 7 through the suspension caused by the scattering and
3 Particle Size Determination in Mono- absorption of light by the particles. The amount of
disperse Suspensions 7 scattering and absorption depends on the size of the
3.1 Specific Turbidity 8 particles and their concentration in the suspension. Most of
3.2 Turbidity Ratio 8 the reported applications in the literature use turbidimetry
4 Particle Size Distribution Determination for the estimation of the particle size distribution (PSD) or
in Polydisperse Suspensions 9 an average size for particles with diameters ranging from
4.1 Specific Turbidity 9 0.035 to 50 µm. Turbidimetric techniques are low cost,
4.2 Turbidity Ratio 11 experimentally simple, fast, nondestructive and require
4.3 The Wavelength Exponent no calibration. They are robust methods with very good
Technique 12 reproducibility and can be used either off- or on-line, in a
4.4 Resolution of a Conflict in the laboratory or a plant environment. They have been used for
Literature 13 particle or droplet size determination in a variety of diverse
4.5 Solving for the Particle Size systems such as latex particles, silver bromide sols and
Distribution 13 cow’s milk. Most of the time common ultraviolet/visible
(UV/VIS) spectrophotometers are utilized for the turbidity
5 Estimation of Average Diameters 16
measurements. Certain modifications may be necessary on
5.1 Turbidity-average Diameters 17
these instruments depending on the size of the particles.
5.2 Volume – Surface-average Diameters 17
The turbidity of the suspension is measured and the size
5.3 Weight-average Diameters 18
of the particles or the PSD is estimated utilizing light
5.4 Other Applications of Turbidimetry 18
scattering theory. For spherical isotropic homogeneous
6 Procedural Details 19 particles and absence of multiple scattering this theory
6.1 Instrumentation and Forward is well established. The extension to nonspherical and
Scattering 19 inhomogeneous particles is a field of active research
6.2 Refractive Indices of the Particles yielding numerous publications.
and the Medium 20 A controversy existed in the literature regarding the
6.3 The Choice of the Wavelengths 22 capability of turbidimetry to provide the full PSD but
6.4 Preparation of Samples 22 has now been resolved. For polydisperse suspensions,
6.5 Particle Volume Fraction for Spe- for certain combinations of particle sizes and optical
cific Turbidity 23 properties, the method cannot provide the full PSD,
7 Effect of Errors on the Estimated but only an average of the PSD. This limitation is
Particle Size and Particle Size simply a result of the light scattering patterns and it is
Distribution 23 expected theoretically. In some cases it can be overcome
7.1 Error in Measurements 23 by simply altering the optical properties of the suspension
7.2 Error in the Value of the Refrac- (i.e. choosing the proper wavelengths for the turbidity
tive Index Ratio, m 23 measurements or a medium with a different refractive
7.3 Error in the Assumed Distributional index to suspend the particles). To make the best use of
Form 24 the method and interpret the results in a meaningful way

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
2 PARTICLE SIZE ANALYSIS

one should be aware of this limitation and work with particle size using on-line turbidimetry has also attracted a
it. When this is done, excellent results can be achieved. lot of attention..6,9 – 11,15,21 – 24/ Some of the work reported
The advantages of the method and a good knowledge in the literature for particle size determination in poly-
of its capabilities make it a very robust choice for many disperse systems (full PSD or an average diameter) for
applications. The numerous reports from the successful different materials is given in Table 1.
application of the method are a testimony to this. A controversy in the literature.11,12,31/ regarding the
capability of turbidimetry to provide the full PSD has
been resolved by a theoretical analysis and experimental
verification reported in Kourti et al..15,32/ The results can
1 INTRODUCTION
be summarized as follows.
The capability of turbidimetric techniques to provide an
Turbidimetry refers to a group of techniques that utilize average particle size or the PSD of a suspension and their
either turbidity measurements at different wavelengths, sensitivity to experimental error cannot be generalized
or a combination of measurements of turbidity and across refractive indices and particle size ranges. It
of the particle volume fraction, to estimate the size depends on the optical properties of the suspension and
of particles in suspensions. Turbidity measures the the size of the particles relative to the wavelength and
attenuation of a beam of light when it passes through conclusions cannot be extrapolated from one optical
a suspension of particles. Using light scattering theory regime and particle size to another and from one
this attenuation can be theoretically related to the size technique to another. For monodisperse suspensions the
and concentration of particles in the suspension if the particle size can always be estimated using turbidimetry,
optical properties (refractive indices of particles and provided that two turbidity measurements at widely
medium) are known. For spherical particles and absence separated wavelengths together with a measurement of
of multiple scattering this theory is well established. the particle volume fraction in the suspension is used. This
Extensions to nonspherical particles have been reported. is irrespective of the optical properties of the suspension
Simple UV/VIS spectrophotometers can be modified (refractive indices of particles and medium, wavelength
(and in some cases used as bought) for the turbidity of light) and the size of the particles. In some simpler
measurements. Wavelengths in the near-infrared (NIR) cases only two turbidity measurements or measurement
range can also be utilized. Particle sizes from the of one turbidity and the volume fraction are sufficient.
lower submicrometer range to many micrometers can For polydisperse suspensions, turbidity measurements
be measured with the proper choice of wavelength for the at sufficient wavelengths (the number depends on the
turbidity measurements. Applications of turbidimetry for detail of the PSD one wishes to resolve) together with a
particle size determination have been reported mainly for measure of the particle volume fraction are required for
particles with diameters between 0.035 and 50 µm. The the determination of the full PSD. However, for certain
simplicity of this fast, noninvasive method, its low cost, combinations of particle sizes and optical properties,
good reproducibility and the wide range of particle sizes the method cannot provide the full PSD, but only
that it covers have made it very attractive for particle an average of the PSD. This limitation is simply a
size determination. Mainly owing to its simplicity, the result of the light scattering patterns and it is expected
method is robust and suitable for industrial production theoretically. More specifically, it stems from the fact
environments as an on-line particle size determination that for some combinations of optical properties and
method. particle size, a light scattering function (the wavelength
The first attempts to use turbidimetry for particle size exponent) changes very slowly with the relative size of
determination started in the 1950s. At the time work- the particles to the wavelength. When this happens, in
ers tried to use approximation theories.1/ or tabulate the presence of experimental noise, estimation of the full
turbidity functions.2,3/ mainly owing to a lack of power- PSD is not possible. In some cases this limitation can
ful computing machines. Ever since, there has been an be overcome by simply altering the optical properties of
abundance of reports on the use of turbidimetric tech- the suspension (i.e. choosing proper wavelengths for the
niques for the determination of particle size in a variety turbidity measurements or a medium different than water
of systems, such as silver bromide sols in water,.4 – 6/ to suspend the particles). Therefore, to make the best use
polystyrene latexes,.7,8/ poly(vinyl acetate) latexes,.9 – 15/ of the method and interpret the results in a meaningful
homogenized milk and oil-in-water emulsions,.16 – 18/ cou- way, one should be aware of the regimes where this
pler dispersions.19/ and minerals..20/ A wide range of limitation occurs.
wavelengths in the UV/VIS region have been utilized for Because of the behavior of the wavelength exponent,
the turbidity measurements but also work in the NIR turbidimetric techniques that utilize the measurement of
region has been reported..6,16,17,19/ The determination of the particle volume fraction (which is a measurement of

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
Table 1 Examples of applications of turbidimetry in particle sizinga

System m Particle size, Wavelengths Turbidimetric Assumed PSDb Commentsb Other methodsb
D (µm) (nm) method
Minerals.20/ : TiO2 1.9 0.1 < D < 10 436, 546, Specific Log-normal on weight Estimated PSD Centrifugal
Al(OH)3 1.86 650, 800 turbidity basis agrees very well sedimentation
Al2 O3 1.33 with measured
SiO2 1.163 PSD from
centrifugal
sedimentation

DOI: 10.1002/9780470027318.a1517
TiO2 pigments..25/ – Mean crystal 190 – 1100 44 turbidities No shape assumed. Study involves TEM
Oval-shaped crystals. sizes: and volume Estimate 26 heights of empirical modeling
Estimate crystal 0.086 – 0.237 fraction histogram in the range with latent

This article is © 2006 John Wiley & Sons, Ltd.


size-area equivalent 0.012 – 0.6375 nm variables – PLS
diameter
Stable silver bromide sols 1.7 D < 0.25 (1) 436, 546 Specific (1) Apparent diameters Important EM
in water.5/ (2) 450, 800 turbidity (2) Log-normal discussions on
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS

(ZOLD) apparent
diameters. See
section 5
Ostwald ripening process: 1.7 D < 0.8 2.22 µm Specific Apparent diameter (Dt ) On-line monitoring EM
silver bromide grains.6/ turbidity of grain growth.

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
See section 8
Variety of materials.26/ – Weight average: 900 Specific Assumed g PSD with Some particles too Sieve
PVC, 138 turbidity known spread. large. Only D32 Coulter counter
sand, 116, 196 Reported average size expected to be
protein, 20 – 30 correct.
KCl 92 – 284

Polystyrene latexes.7/ 1.2 0.65 < D < 1.3 440 – 610 basis: Turbidity ratio Positively skewed with For polystyrene see EM
546 maximum also Haseler..19/

This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
polydispersity 1.5 See sections 4.2
and 7.3
Polystyrene.8/ 1.2 Synthesized 300 – 800 Turbidities No assumption on shape Solve directly for the
bimodal and only heights of the PSD
trimodal PSDs
Styrene – butadiene 1.16, 1.14 0.04 < D < 0.14 300 – 600 Specific (1) Apparent Dt Only Dt correct SEM
rubber and acryloni- (101 points) turbidity (2) Log-normal DLS
trile – butadiene (3) No shape
rubber.27/ assumption
Copolymer latexes: 1.17, 1.15 D < 0.4 350 – 1000 Specific Apparent diameters See section 5.1 EM
poly(butadiene – turbidity
styrene).28/
Photographic 1.166, 1.21 0.05 < D < 1.0 400 – 800 Specific Log-normal on weight Conclude that EM
dye-forming coupler turbidity and basis (verified by EM) success of method
dispersions.29/ wavelength depends on values
exponent of a
3

(continued overleaf )
4
Table 1 (continued)
System m Particle size, Wavelengths Turbidimetric Assumed PSDb Commentsb Other methodsb
D (µm) (nm) method
Ektacolor coupler 1.15 D < 1.0 400, 700 Specific Log-normal See sections 4.1, 4.2, –
dispersions.19/ – droplet 400, 2.22 turbidity 5.2 and 7.3
size distribution
Cis-polyisoprene 1.136 0.6 < D < 1.1 650, 800, Specific Log-normal on weight Broad PSDs.15/ See Coulter counter
Latexes.58/ 950, 1100 turbidity basis also section 9.2 EM

DOI: 10.1002/9780470027318.a1517
Copolymer latexes: – 12 < DN < 50 400 – 600 Specific Apparent diameter See section 5.2. Some EM
divinyl – benzene/ turbidity (D32 ) coincidental

This article is © 2006 John Wiley & Sons, Ltd.


styrene.66/ observations are
discussed by
Kubota et al..68/
Latexes.68/ – 2 < D < 40 400, 700 Specific Apparent diameter See section 5.2 EM
turbidity (D32 )
Octacosane 1.15 D < 3.0 436, 546 Turbidity 1. Log-normal Approach has serious EM
suspensions.48/ ratio and 2. One parameter limitations..50/ See
wavelength PSD, positively section 4.3.
exponent skewed
Poly(vinyl acetate).16/ 1.1 0.1 < D < 1.0 320 – 720 Turbidity ratio Positively skewed See sections 4.2 and EM

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
4.4
Poly(vinyl acetate) 1.1 1. Dw : 0.20 300, 600 Specific Log-normal See sections 4.4 and Disc centri-
latex.11,12/ 2. Dt : 0.11 turbidity 8.0 fugation
Angular light
scattering
Poly(vinyl acetate).15,44/ 1.095 0.035 < D < 0.4 400 – 600 Specific Log-normal (Dap D Dw ) Experimental & HDC
turbidity theoretical analysis DLS
demonstrates the

This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
limitations of
turbidimetry in
some (m, a)
regions
Poly(vinyl acetate).23,24/ 1.095 D < 0.4 400 Specific Apparent diameter, Dw First successful DLS
turbidity on-line application
Poly(vinyl acetate) 1.1 D < 0.4 350 – 700 Method of Moments of PSD See section 4.5 HDC
latex.10/ moments
Paraffin oil in (water C 1.1, 1.05 0.1 < D < 10.0 300 – 1700 Specific (1) Log-normal See also other Coulter counter
glycerol/ C milk.16/ 0.1 < D < turbidity (2) An upper limit work.17,30/ for
40.0 function similar systems.
(3) Exponential type Successful
application
a The systems have been ordered based on the optical properties of the suspension (value of m).
b Abbreviations (other than those already defined in the text): PLS, partial least squares; TEM, transmission electron microscopy; ZOLD, zeroth-order logarithmic distribution; EM,
electron microscopy; SEM, scanning electron microscopy; DLS, dynamic light scattering; HDC, hydrodynamic chromatography.
PARTICLE SIZE ANALYSIS
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 5

the particle size and number independent of the scattering on the wavelength) absorption of light by the particles
properties) provide reliable estimates for the particle size and/or the medium. The amount of light scattered and
(the PSD or an average diameter) in a wider range of absorbed by the particles depends on their size, their
sizes and optical properties than techniques that utilize concentration in the suspension, the size of the particles
turbidity measurements only. relative to the wavelength of the incident beam and
In this article, a brief overview of light scattering the refractive indices of the particles and the medium.
theory is first presented followed by a description of Therefore, theoretically, turbidity can be related to the
the turbidimetric techniques that are most commonly particle size. For a monodisperse suspension (particles of
used for particle size determination. For each technique equal diameter, D) of spherical, isotropic particles, in the
we include its theoretical background, comments on its absence of multiple scattering (Equation 2):
advantages and limitations and references to its successful
applications in the literature. We also include discussions pD2
tl0 D N Kext .2/
regarding the regions of particle sizes and optical 4
properties where the technique may present difficulties
where N is the number of particles per cm3 and Kext is
in providing reliable estimates of the particle size or
the extinction coefficient at wavelength l0 . Kext is equal
the PSD. We then present cases where turbidimetry was
to the sum of the absorption Kabs and the scattering Kscat
used to provide estimates of the average diameters of
coefficients. The extinction coefficient can be calculated
the PSD and discuss the advantages and pitfalls of such
from the rigorous Mie theory and the equations for
uses. We follow with details on the procedure such as
this calculation can be found in several sources..3,33 – 36/
choice of wavelengths, calculation of refractive index
In the work presented in this article we assume no
and instrumentation. The sensitivity of the turbidimetric
absorption (Kext  Kscat ) and this is the case for most
methods to experimental error is extensively discussed.
of the work reported in the literature. In most of the
Applications on the use of the method for on-line
reported applications workers have chosen to avoid
particle size measurements follow. We conclude with
wavelengths where the medium or the particles absorb,
a discussion on how users should proceed with choosing
for practical reasons, explained in section 6.3. However,
the appropriate method for their application and with
should readers decide to work with absorbing particles,
comparing results from other particle sizing techniques.
they should use Kext for Equations (3), (4), (12), (13),
Throughout the article several applications reported in
(15) and (18). The validity of the discussions in this article
the literature are critically discussed in order to point
does not change with the nonabsorption assumption.
out some approaches with serious limitations and some
For a monodisperse suspension of spherical, nonab-
coincidental observations that may result in misleading
sorbing, isotropic particles, in the absence of multiple
conclusions. In this way the readers can judge if the
scattering (Equation 3):
approaches reported in the literature are suitable for
their use.  
pD2 pD2 D np
tl0 D NRscat D N Kscat D N K , .3/
4 4 lm nm

2 THEORETICAL BACKGROUND where Rscat , the scattering cross-section, is the total energy
scattered by one sphere when the incident light is of
2.1 Turbidity and Particle Size unit intensity and either plane polarized or unpolarized.
The scattering coefficient, Kscat , or K.D/lm , np /nm /, is
Turbidity, tl0 (cm 1 ), gives a measure of the attenuation dimensionless and it is an involved function of two
of the intensity of a beam of light traversing through a parameters, a and m. a D p.D/lm / is defined as the
suspension of particles (Equation 1): relative size of the particle diameter to the wavelength
  of the light in the medium (lm ) and m D np /nm is the
1 I0
tl0 D ln .1/ ratio of the refractive index of the particles to the
l I
refractive index of the medium. lm D l0 /nm , where l0
where I0 and I represent intensities of the incident and is the wavelength of the incident beam in vacuo. Both
transmitted beams, respectively, and l is the length of np and nm are evaluated at l0 . The estimation of the
the optical path (i.e. cell length, cm). The value of the refractive indices for homogeneous and inhomogeneous
turbidity changes with the wavelength of the incident light particles is discussed in section 6.2. In cases where the
beam, l0 . I is measured at 180° to I0 . particles absorb, the refractive index has both real and
The attenuation of the intensity of light traversing imaginary parts.
through a particle suspension is due to scattering of For a polydisperse suspension of spherical nonab-
light by the particles and in some cases (depending sorbing, isotropic particles, in the absence of multiple

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
6 PARTICLE SIZE ANALYSIS

scattering turbidity is given by Equation (4): here in some detail. The scattering coefficient is defined
Z 1   as the ratio of the scattering to the geometrical cross-
pD2 D np section. The scattering coefficient exhibits an oscillatory
tl 0 D N K , f .D/ d.D/ .4/
0 4 lm nm dependence on a, with a series of successive minima and
where f .D/ is the normalized PSD. maxima (Figure 1). The first maximum occurs at lower
In general, because Kscat is an involved nonlinear a values the larger is the value of m; the numerical
function of a (i.e. D/lm ), it is possible from turbidity value of Kscat at the first maximum is larger the larger
measurements at a sufficient number of wavelengths is m. At large values of a, m has virtually no effect
(or a combination of turbidity measurements and other on Kscat , in agreement with the theory of diffraction
properties of the particle suspension) to estimate the for objects very large compared with the wavelength
size of the particles in a monodisperse suspension or used, and Kscat approaches the value of 2. Values of
the PSD in a polydisperse suspension. However, it has Kscat for various large a values are shown in Table 2.
been shown.32/ that for certain combinations of (m, a) Note that for m D 1.2 and 200 < a < 220 the value of
values the estimation of the particle size or the PSD the oscillating Kscat can be found between 2.12 and 1.99,
of the suspension may present difficulties with some whereas for 2500 < a < 2700, Kscat oscillates between
turbidimetric techniques. Being aware of these (m, a) 2.016 and 2.006. When plotted against a, the scattering
regimes will help the user choose the proper turbidimetric coefficient also exhibits secondary fluctuations (ripples)
technique and proper wavelengths to work with in order that become increasingly irregular for higher values of m.
to obtain the most reliable estimate of particle size of a Another parameter which is often used in light scattering
suspension. is r D 2ajm 1j; r is the phase shift suffered by a wave
passing through the center of a sphere..37/ When Kscat is
plotted against r, the r position of the extremes (minima
2.2 The Scattering Coefficient and maxima) of the main oscillation is independent of
Certain features in the behavior of the scattering the value of m..33,34/ Recall that this discussion is for
coefficient are important in explaining the capability of spherical, isotropic spheres. Scattering for nonspherical
turbidimetric techniques to provide the particle size or particles is an area of active research.38/ and publications
the PSD of a suspension. These features will be discussed are available..39 – 41/
As mentioned earlier, in the general case the scattering
4.0 coefficient can be calculated from the rigorous Mie
m =1.2 theory..3,33 – 36/ In certain (m, a) regimes the equations for
3.5 Kscat can be approximated with simple expressions which
m =1.1 give Kscat within 1% accuracy from its true value..37/
3.0
These expressions are discussed here because they help
2.5 in elucidating some of the problems encountered by
researchers when converting turbidity measurements to
2.0 particle size.
Kscat

1.5 1. When the suspended particles are very small com-


1.0 pared with the wavelength both outside (a − 1) and
inside (jmaj − 1), the particle,.34,37,42/ the Rayleigh
0.5 scattering theory gives Equation (5):

0.0    2
0 20 40 60 80 100 D np 8 m2 1
K , D a4 .5/
α = πD/λm lm nm 3 m2 C 2

Figure 1 Scattering coefficient, Kscat , plotted against a for 2. For particles very large compared with the wave-
m D 1.1 and 1.2. length, the scattering coefficient approaches a

Table 2 Range of values (maximum – minimum) observed for the oscillating Kscat in different ranges of a values for m D 1.1 and 1.2

m 20 < a < 40 80 < a < 100 140 < a < 160 200 < a < 220 800 < a < 1000 1500 < a < 1700 2500 < a < 2700
1.1 3.45 – 1.64 2.39 – 1.88 2.24 – 1.99 2.175 – 1.98 2.051 – 1.998 2.03 – 2.002 2.0195 – 2.002
1.2 2.7 – 1.7 2.26 – 1.94 2.16 – 1.98 2.12 – 1.99 2.0365 – 2.0065 2.023 – 2.006 2.016 – 2.006

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 7

constant value independent of m.3,33,34,37/ (Equa- 6.0 4


tion 6):
  5.5
D np
K , D2 .6/
lm nm

Exponent z
5.0 3

Exponent y
Walstra.37/ has reported that this is true for r D 4.5
2ajm 1j > 300 and a > 2000. Table 2 gives the
range of Kscat values for different a values and m D 1.1 m =1.05
4.0 2
and 1.2. Note that for m D 1.2 and a values between m =1.1
200 and 220 the values of the oscillating Kscat can be 3.5
found between 2.12 and 1.99. Assuming a constant m =1.15
m =1.25 m =1.2
value of 2.0 may introduce a maximum of 6% error in 3.0 1
0 2 4 6 8 10 12
Kscat in this range. For very large values of a, assuming
Kscat D 2.0 gives Kscat within 1% accuracy from the α = πD/λm
true value.
Figure 2 Exponents y and z for m D 1.05, 1.1, 1.15, 1.2 and 1.25.
Equations (5) and (6) are the two extremes of a
general expression where for monodisperse suspensions where both k and y are functions of D. When the particle
of nonabsorbing spheres, the dependence of the scattering sizes in the suspension are such that k and y do not change
coefficient on the diameter at a given wavelength is significantly with D, then (Equation 10):
approximated by Equation (7):.4,5/ Z 1
   z tl0 D Nk Dy f .D/ d.D/ D Nk Jy .10/
D np D
K , Dk 0
D k0c az .7/ 0
lm nm lm where k and y are constants. In these cases, turbidity is
where k0 and k0c are proportionality constants. Note proportional to the yth moment, Jy , of the PSD. The value
that z D 4 in the Rayleigh regime and z D 0 for large of y (i.e. the order of the moment) changes with a and m
particles (compare Equation 7 with Equations 5 and 6, (Figure 2). For a < 0.5, y approaches the limiting value
respectively). When plotting Kscat against the wavelength of 6 (z D 4, Rayleigh regime) and therefore turbidity is
for constant D, we have Kscat D k00 l0 z , and therefore z proportional to the sixth moment of the PSD. For values
has been termed ‘‘the wavelength exponent’’..43/ When of m between 1.1 and 1.2, and for a < 1.6, the value of y is
plotting Kscat against a at constant lm , we can write almost independent of the value of m. Kourti et al..15,32,44/
Equation (8): discussed that for small values of m (m D 1.05, 1.1, 1.15)
the value of y is approximately 4.0 (4.2  y  3.8) for a
ln.Kscat / D z ln a C ln k0c .8/ very wide range of a values (for m D 1.05, 3 < a < 14; for
m D 1.15, 3 < a < 6). For suspensions with PSDs in the
The value of z can be obtained for any a at any m value above (m, a) regions, turbidity would be approximately
from the slope of ln Kscat vs ln a. proportional to the fourth moment of the PSD. Finally,
for very large particles, y D 2 and turbidity is proportional
to the second moment of the PSD.
2.3 Turbidity and Moments of the Particle Size
Distribution

In regimes where Kscat can be approximated by Equa- 3 PARTICLE SIZE DETERMINATION IN


tion (7), the turbidity of a monodisperse suspension at a MONODISPERSE SUSPENSIONS
given wavelength l0 can be written as t D NkDy , where
y D z C 2 and k is a constant. Several authors.5,43/ have The turbidity of a suspension of spherical particles
tabulated values of y or z for several pairs of (m, a) values as given in Equation (3) is a function of both the
(in the notation of Heller et al.,.43/ z  n). Values of y number concentration of particles (N) and the particle
and z are plotted in Figure 2 for several m values and diameter (D). The estimation of these two unknowns
a  12.0. From Equations (4) and (7), for a polydisperse requires sufficient measurements of turbidity at different
system, at a given l0 , we can write Equation (9): wavelengths or measurements of turbidity and another
property that is also related to particle concentration
Z 1
and diameter. Two approaches have been used in the
tl 0 D N k.D/Dy.D/ f .D/ d.D/ .9/
0
literature:.33/ the turbidity at a given wavelength is divided

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
8 PARTICLE SIZE ANALYSIS

either by the particle volume fraction or by the turbidity particle diameter. Note that owing to the oscillatory
value at another wavelength, such that the unknown behavior of the specific turbidity, more diameters may
number of particles is eliminated, and the new quantity correspond to one specific turbidity value. For example,
is related to particle size only. Once the particle size is for polystyrene a specific turbidity reading of 15 000 cm 1
determined, substitution into Equation (3) will give N. at 400 nm in the range shown in Figure 3 corresponds to six
different diameters (0.15, 1.7075, 1.855, 1.865, 1.9225 and
3.1 Specific Turbidity 2.56 µm; four due to the main oscillating pattern of Kscat
and two due to a secondary rippling effect). Therefore,
The volume fraction of the spherical particles in a if the size range is not a priori known, measurements at
monodisperse suspension is given by Equation (11): two (preferably widely spaced) wavelengths are required
pD3 to extend the applicability of the method to a wide
fDN .11/ range of particle sizes..15/ Once the particle size has
6
been estimated, the concentration of the particles in
The particle volume fraction can be measured (see the suspension (N) can be calculated from Equation (3).
section 6.5) using gravimetric analysis or densito- Note that in the Rayleigh regime, where Kscat is given
metry..21,23,45/ The specific turbidity, defined as the ratio by Equation (5), the specific turbidity at given lm is
of the turbidity of the suspension to the volume fraction directly proportional to the volume of the particles
of the suspended particles, is independent of the particle (t/f / D3 ), and that for very large particles (Kscat D 2.0) it
number concentration (Equation 12): is proportional to the inverse of the diameter (t/f D 3/D).
 
D np
tl 0 3 K ,
lm nm
D .12/ 3.2 Turbidity Ratio
f 2 D
The ratio of two turbidity readings taken at two
Specific turbidity is plotted against diameter in Figure 3 wavelengths is a quantity independent of the particle
for two different types of particles suspended in water: number concentration (Equation 13):
poly(vinyl acetate) with m ³ 1.1 and polystyrene with  
m ³ 1.2. The particle diameter D can be calculated D np
tl01 K ,
lm nm l
from one turbidity measurement and the particle volume D   01 .13/
fraction when the optical properties of the suspension tl02 D np
K lm , nm
l02
are known and the expected particle size range is known.
For example, if it is known that we have submicrometer Subscripts l01 and l02 denote that the quantities t, np ,
particles then one specific turbidity measurement for nm and lm are evaluated at these wavelengths; l01 ,
poly(vinyl acetate) at 600 or 400 nm would give the l02 are the wavelengths in vacuo. Figure 4(a) shows
the turbidity ratio (at 400/600 nm) for polystyrene and
8 poly(vinyl acetate) particles suspended in water. The
Specific turbidity × 10−4 (cm−1)

particle diameter could be estimated from one turbidity


7 ratio (two turbidity measurements) if the ratio was a
PS-400 nm
monotonic function of particle size or a. However, this is
6
not the case, as can be seen in Figure 4(a). An indentation
5 exists in the turbidity ratio vs a curve, at the lower a
PS-600 nm
regime, owing to the first inflection point of the Kscat
4 vs a curve (in polydisperse suspensions, this indentation
vanishes). Also, for larger particles, the turbidity ratio
3 PVAc-400 nm vs a curve exhibits an oscillatory behavior. Therefore,
2
PVAc-600 nm unless the size range is a priori known, a reading at a third
wavelength (two turbidity ratios) is advised, to define
1 uniquely the particle size. Again once the particle size is
estimated the number of particles can be calculated from
0 Equation (3).
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0
An example where only one turbidity ratio is
Diameter (µm)
not sufficient to determine the particle size was
Figure 3 Specific turbidity plotted against particle diameter for reported by Kourti..15/ The turbidity ratio measure-
monodisperse poly(vinyl acetate) and polystyrene latexes at 400 ment at l01 D 400 nm and l02 D 500 nm for a 60 : 40
and 600 nm. poly(styrene – butadiene) latex was repeated several times

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 9
7 and t500 /t600 ) plotted against diameter for monodis-
perse polystyrene suspensions. The diameters are in
6 a range around the first indentation in the turbidity
ratio – diameter curve. Note that suspensions with two
τ400 nm/τ600 nm

5
very different particle sizes (0.2 and 0.25 µm) will give the
4
same turbidity ratio at two different wavelengths (1.72
for t500 /t600 and 3.54 for t400 /t600 ). In fact, for this exam-
3 ple, with an experimental error as small as 2.0% in the
turbidity ratios, any solution between 0.2 and 0.25 µm is
PVAc possible. It would be preferable if the three wavelengths
2
PS
at which the measurements were taken were more widely
1 separated.
The turbidity ratio method practically gives no informa-
0
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 tion on the particle size for (i) very large particles, where
for all the wavelengths used the scattering coefficient is
(a) Diameter (µm)
constant (substitution of Kscat D 2 in Equation 13 gives a
4.0
turbidity ratio equal to unity), and (ii) very small particles,
where for all the wavelengths used the Rayleigh Equa-
τ400 / τ600 tion (5) applies and turbidity ratio becomes independent
3.5 of the particle size.
Turbidity ratio

3.0
4 PARTICLE SIZE DISTRIBUTION
DETERMINATION IN POLYDISPERSE
2.5 SUSPENSIONS

2.0 Specific turbidity and turbidity ratio measurements can


τ500 / τ600
also be used to estimate the PSD in polydisperse suspen-
sions. Numerical approaches that can be used to solve
1.5 for the PSD from turbidity and specific turbidity mea-
0.10 0.15 0.20 0.25 0.30
surements are discussed in detail in section 4.5. However,
(b) Diameter (µm) the estimation of the full PSD is not always possible with
turbidimetry..15,32/ Some turbidimetric techniques are not
Figure 4 (a) Turbidity ratio at 400/600 nm plotted against
particle diameter for polystyrene with m ³ 1.2 and poly(vinyl theoretically expected to provide the full PSD of a suspen-
acetate) with m ³ 1.1 latexes. (b) Turbidity ratios for polysty- sion if, because of its optical properties (refractive indices
rene, m ³ 1.2. Note that owing to the oscillation in the curves, of medium and particles) and the sizes of the particles, the
particle suspensions with two different diameters (0.2 and (m, a) values of the suspension are within certain ranges.
0.25 µm) give the same turbidity ratio values at two different In such cases the specific turbidity technique will provide
wavelengths.
an estimate of an average diameter while techniques sim-
ilar to the turbidity ratio provide no information on the
particle size. The (m, a) regimes where the estimation
(for four different dilutions of the same sample). Each of the full PSD is practically possible are discussed for
time, two estimates (two solutions) of particle size were each one of the turbidimetric methods presented in this
obtained. In one of the solutions the estimates of the section.
diameter were between 209 and 228 nm and in the other
between 157 and 165 nm. Results from other methods had 4.1 Specific Turbidity
shown that the true diameter range was around 200 nm
Examples of successful applications of the method for
[specific turbidity (208 – 221 nm), HDC (228 nm) and DLS
particle and droplet size distribution determination can
(194 nm)] and therefore the first solution could be picked
be found in Meehan and Beattie,.5/ Walstra.16/ and Yang
as the correct one.
and Hogg..20/ For a polydisperse suspension of spherical
Caution should be exercised when diameters are in
particles, the volume fraction is (Equation 14):
a range around the first indentation in the turbidity
Z 1
ratio– diameter curve when using the turbidity ratio pD3 p
fDN f .D/ d.D/ D N J3 .14/
method. Figure 4(b) shows turbidity ratios (t400 /t600 0 6 6

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
10 PARTICLE SIZE ANALYSIS

where J3 is the third moment of the PSD. The specific Equation (16):
turbidity is given by Equation (15a): Z 1
Z 1   D6 f .D/ d.D/
n tl 0 J6
D2 K lDm , nmp f .D/ d.D/ D kl0 Z0 1 D kl0 D kl0 D3t .16/
tl 0 3 0 f J3
D Z 1 .15a/ D3 f .D/ d.D/
f 2 3 0
D f .D/ d.D/
0
where Dt (or D63 ) is the turbidity-average diam-
Alternatively, the specific turbidity of a suspension can eter and kl0 D 4p4 lm4 [.m2 1//.m2 C 2/]2 and the
be written as Equation (15b): subscript l0 denotes that t, lm and m are evalu-
Z 1 ated at l0 . Llosent et al..27/ tried to estimate the
tl0 t
D .D, l0 /F.D/ d.D/ .15b/ full PSD in the Rayleigh regime using either an
f 0 f assumed shape of the PSD or using a regulariza-
where t/f.D, l0 / is the specific turbidity corresponding tion method (discussed later) to solve at discrete
to particles with diameter D and wavelength l0 and F.D/ points for the PSD. Their results confirmed the
is the normalized PSD on a weight basis. suggestions of Kourti et al..32/ that only the turbidity-
For a suspension of spheres, when the optical properties average diameter can be correctly estimated in these
of the suspension are known, a minimum of P specific regimes.
turbidity measurements are required for the estimation 2. For any m value and very large diameters (where
of a P parameter PSD from Equation (15a) or (15b), Kscat ³ 2), we have Equation (17):
as discussed in the next section. For example, if we Z 1
assume that the PSD is known to be log-normal described D2 f .D/ d.D/
by the mean diameter and the variance, we need at tl 0 J2 3
D 3Z 1
0
D3 D .17/
least two specific turbidity measurements. If the form of f 3 J 3 D 32
D f .D/ d.D/
the PSD is not assumed and we attempt to recover the 0
frequency distribution at P given diameters (i.e. P discrete
D32 , the volume – surface average, is also sometimes
points), then at least P specific turbidity measurements
referred to as the Sauter mean diameter.
are required.
In the presence of experimental error, the estimation of 3. (m, a) regimes where the exponent y is close to
the full PSD from Equation (15a) or (15b) is impossible 4.0. In an extensive study with theoretical analy-
if all the turbidities (i.e. at all the wavelengths required sis and experimental validation, Kourti et al..15,32/
for the estimation) are approximately proportional to the showed that for suspensions with PSDs in (m, a)
same moment of the PSD, Jy . Given that f is always regimes where the exponent y is very close to 4.0
proportional to the third moment J3 of the PSD, in these (3.8 < y < 4.2), specific turbidities result in apparent
cases specific turbidity becomes proportional to Jy /J3 , diameters numerically very close to the weight aver-
that is proportional to the (y 3)rd power of the Dy3 age (Dw or D43 ) of the true PSD. The regions of (m, a)
average diameter..32/ Kourti et al..32/ showed that when values where this happens can be read from a y vs a
this happens for all of the P wavelengths used, many plot, such as that in Figure 2. From such a plot we can
alternative solutions (PSDs) correspond to the same see that y ³ 4, for m D 1.05 at 3.0 < a < 14 and for
specific turbidity measurements and it is impossible to m D 1.15 at 3.0 < a < 6.0. If the whole distribution
isolate the true solution unless there is some a priori covers a values in this regime for all the wavelengths
knowledge. However, all these alternative solutions have used for the turbidity measurements, then within
their Dy3 average diameters numerically very close to experimental error many alternative solutions are
each other. In that case it was suggested that only this possible and only the weight-average of the distri-
average diameter Dy3 can be correctly estimated from bution can be correctly estimated. This is true for
specific turbidities.15,32/ in this range of wavelengths. any type of distribution (multimodal or unimodal
The (m, a) regimes where the full PSD is practically continuous distribution, or simply the mixture of
impossible to estimate for nonabsorbing particles are as two populations). Table 3 gives an illustration of the
follows: effect of small measurement errors on the estimation
of the PSD in these regimes, for m D 1.1. In case A,
1. For any m value and very small a (Rayleigh regime). the true distribution is log-normal (see Equation 21)
If for all the wavelengths used for the turbidity with Dg D 175 nm and s D 0.2 and weight-average
measurements we are in the Rayleigh regime, then diameter Dw D 201 nm. The turbidities for such a
substitution of Equation (5) in Equation (15a) gives suspension were theoretically calculated at three

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 11

Table 3 Illustration of effect of small random measurement errors on the estimated PSD for m D 1.1a

Case True PSD Different random Estimated PSD


error for each test
Dg (nm) s DN (nm) DW (nm) Dg (nm) s DN (nm) DW (nm)
A 175 0.2 178 201 A1 116 0.39 125 199
A2 198 0.08 198 202
B 455 0.2 464 523 B1 520 0.002 520 520
B2 336 0.362 359 533
a Two cases with different true PSDs are examined: (A) Dg D 175, s D 0.2 and (B) Dg D 455, s D 0.2. A maximum error of š1.2% was introduced
randomly in the specific turbidities at several wavelengths. Then the PSD was re-estimated. For each case two different random errors were tried,
and two very different PSDs were obtained, both far from the true solution, but with Dw close to the true solution.

wavelengths. An error of less than 1.0% was then this information and used specific turbidity to monitor
introduced in the three turbidity values and the the progress of polymerization reactions with on-line
parameters of the PSD were re-estimated (case A1). measurements. The a range where specific turbidity mea-
Note that the estimated Dw D 199 is within 1% of the surements give Dw may be extended further when the
value of the true weight average of the PSD whereas distribution is log-normal. For example, for suspensions
Dg and s are significantly different. A different with m D 1.1 and log-normal distributions of submicrom-
random error less than 1% was introduced in the three eter particles, it has been observed that the mean value
turbidities for case A2. Again, only the weight aver- of the apparent diameters, estimated from specific tur-
age of the estimated distribution Dw D 202 is within bidities at 400 and 600 nm, is within 5% the true weight
0.5% of the true value and the other parameters are average diameter of the PSD..15,32/ Monitoring these two
significantly different from the true distribution but apparent diameters gives information on the weight aver-
also from case A1. Similar results were obtained for age of the PSD.
case B. For different random errors of š1.2% on the It should be noted that in all the regimes where the
turbidity values, cases B1 and B2 were obtained with estimation of the full PSD is impossible, specific turbidity
weight averages within 2% of the true value but with gives good estimates of meaningful average diameters.
all the other parameters significantly different from Compared with other turbidimetric techniques, where
the true distribution. information only from turbidity measurements is used,
the specific turbidity has the advantage that an additional
Reports in the literature confirm that the PSD cannot be
measurement is used, the solids volume fraction, which
estimated with specific turbidity in some cases. Haseler.19/
is always proportional to the third moment of the PSD.
working with Ektacolor suspensions (m D 1.15) reported
Therefore, whenever turbidities at two or more wave-
that experimental errors in specific turbidities (measured
lengths are proportional to the yth moment of the distri-
at 400 and 700 nm) would lead to large errors in the
bution, although the estimation of the true PSD is impossi-
calculated distribution. These results were observed
ble, a meaningful average of the PSD, the Dy3 average, can
because of the small (m, a) values with which he was
always be obtained from specific turbidity measurements.
working. Based on the above discussion, only Dw could
be correctly estimated in those (m, a) regimes. Walstra,.16/
trying to estimate the globule size distribution in cow’s 4.2 Turbidity Ratio
milk (m D 1.1) from specific turbidity measurements,
The ratio of two turbidity readings taken at different
reported that for very fine emulsions (D32 < 200 nm) with
wavelengths for a polydisperse suspension is given by
rather high polydispersity the reliability of the method
Equation (18):
decreases and showed that a set of experimental data
Z 1  
could be fitted to many different types of distributions n
with the same Dw . D2 K lDm , nmp f .D/ d.D/
tl01 l01
The observations of Kourti et al..15,32/ revealed a very D Z0 1   .18/
tl02 n
useful property of the specific turbidity: the weight- D2 K lDm , nmp f .D/ d.D/
0 l02
average diameter of suspensions can be correctly esti-
mated even though the determination of the PSD is Again, subscripts l01 and l02 denote that the quantities
impossible. This property of specific turbidity can be uti- t, np , nm and lm are evaluated at the corresponding
lized for practical purposes where a fast estimate of the wavelengths; l01 , l02 are wavelengths in vacuo. For a
weight-average diameter is required and is sufficient to PSD described by P parameters, a minimum of (P C 1)
monitor the progress of a process. Gossen.23,24,46/ utilized turbidity measurements at different wavelengths (one of

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
12 PARTICLE SIZE ANALYSIS

which is chosen as reference) are required, in principle, with particle diameters between 650 nm and 1.3 µm from
for the estimation of these parameters. The method gives turbidity measurements at l01 D 440 nm, l02 D 610 nm
no information on the particle size in the following cases, and l03 D 546 nm. The values of m and a in their system
for nonabsorbing particles: were such as to allow for successful PSD determination.
However, when Haseler.19/ tried the turbidity ratio
1. For any value of m and very large particles where method for a system with smaller m and a values (m D
Kscat D 2, the turbidity ratio equals 1.0. 1.15, l0 in the range 400 – 700 nm and submicrometer
2. For any value of m and very small particles, where particles), he reported that within experimental error a
the Rayleigh Equation (5) applies, the turbidity ratio large number of different distributions gave the same
equals a constant value, independent of the particle turbidity spectra. Similarly, Maxim et al..31/ reported that
size. the ‘‘turbidity ratio leads to multivalued solutions and
3. When, for all the wavelengths used, turbidities are unless prior estimates of the answer by some other
approximately proportional to the same moment method are available there are no criteria for choosing
of the PSD. This case is observed for very small between alternative solutions’’. They were working with
m values (m ³ 1.15, 1.1, 1.05) and for a very wide small m values (m D 1.1), submicrometer particles and
range of a values below the first Kscat maximum, wavelengths of 320 – 720 nm with a reference wavelength
where the exponent y changes very slowly with a. of 546 nm, that is an (m, a) area where Kourti.15/ showed
This in turn means that even for broad distributions that even a 0.5% error in the turbidity ratio measurements
with sizes falling in these a regimes, turbidities at means that the PSD estimation is practically impossible.
widely separated wavelengths will be proportional to The results of both Haseler and Maxim et al. were
approximately the same moment of the PSD, and expected based on the (m, a) values of their systems.
their ratio will be almost independent of the particle It is clear from the above examples that one cannot
size. Kourti et al..15,32/ showed that for m D 1.1 generalize conclusions on the capability of turbidimetric
and submicrometer particles, the estimation of the techniques to provide the PSD across (m, a) values.
true PSD from turbidity measurements at 350 and
700 nm (with reference l0 D 546 nm) is impossible for 4.2.1 Methods Similar to Turbidity Ratio
measurement errors as small as 0.5%. One example is
Alternatively to Equation (18), some authors.47,48/ use
reported in which the effect of error on the estimated
the ratio given in Equation (19):
PSD is studied for a turbidity ratio error as small
as š0.35% in the above three wavelengths. The .tl/f/l01
PSD was re-estimated for different errors (less than .19/
.tl/f/l02
š0.35%) each time. Several very different log-normal
distributions were obtained; for example, one of the Notice that, numerically, f is eliminated from Equa-
estimated PSDs had a geometric mean Dg D 500 nm tion (19), hence there is no reason to measure it. Other
and standard deviation s D 0.15, Dw D 541 nm and authors use the DQ (dispersion-quotient) method (Equa-
polydispersity 1.05, while another had Dg D 285 nm, tion 20):.33/
s D 0.4, Dw D 499 nm and polydispersity 1.6.
tl01 l202
DQ D .20/
The turbidity ratio method should therefore be avoided tl02 l201
for the estimation of the PSD for systems of very small or
Both of the above approaches have the same limitations
very large particles and for systems with small m values
as the turbidity ratio method for the estimation of the
(m < 1.15) and a values below approximately 10 (for a
PSD.
short wavelength of 400 nm and water medium, these a
values cover all the submicrometer size range). In these
4.3 The Wavelength Exponent Technique
regimes the method is so sensitive to experimental error
that it offers practically no information on the size of the In this approach, the particle size can be estimated only
suspended particles (i.e. neither the PSD nor an average from turbidity measurements as in the turbidity ratio
can be estimated). technique (i.e. the particle volume fraction does not need
Owing to the behavior of the turbidity ratio in the to be known). The change of the turbidity of a suspension
ranges of (m, a) values described above, controversial with the wavelength is recorded and related to the size
reports on the capabilities of the turbidity ratio to of the suspended particles. The technique is described in
provide the full PSD have appeared in the literature. detail in Bateman et al.,.49/ Heller et al..43/ and Kourti..15/
Wallach and Heller.7/ obtained estimates very close to the It has been shown.15,50/ that the turbidity ratio and
true distribution for polystyrene suspensions (m D 1.2) the wavelength exponent are theoretically equivalent

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 13

methods for the determination of the parameters of for a specific turbidity method and re-estimation of
the PSD. The functions used in the formulations of the the parameters with no experimental error. Both the
two techniques to convert the turbidity measurements to workers were correct for the types of cases they were
PSDs can be derived from each other, and in case of studying. However, their findings cannot be extrapolated
no measurement error the use of either method gives over turbidimetric methods and (m, a) regimes. This
identical results. Consequently, in the (m, a) regions extrapolation caused the controversy.
where the estimation of the particle size in a suspension
is not possible using the turbidity ratio method, the 4.5 Solving for the Particle Size Distribution
wavelength exponent method also fails. It should also be
noted that although the two methods are theoretically Equations (15a) and (18) give the value of the specific
equivalent for the determination of the PSD, the turbidity or turbidity ratio, respectively, at given wave-
formulation of the turbidity ratio method is less complex lengths if the PSD, f .D/, is known. The inverse problem
than that of the wavelength exponent. Consequently, is to determine the PSD given the turbidity or the specific
the turbidity ratio method requires less computation to turbidity measurements at several wavelengths, l0i . If a
calculate the PSD and the error propagation is smaller. continuous set of measurements of tl0 vs l0 was avail-
Melik and Fogler.48/ suggested an approach in which a able, one could theoretically deconvolute these equations
combination of the wavelength exponent and the turbidity to yield f .D/. However, in practice, one usually has tur-
ratio method is used and claimed that with this approach bidity measurements at a finite number of wavelengths,
the two parameters of the PSD can be estimated from only so we have to reconstruct f .D/ from only a number of dis-
two turbidity measurements. However, since these are not crete values of tl0i . This is a typical problem in science..51/
two different methods but are theoretically equivalent, at We routinely measure ‘‘enough points’’ and then ‘‘draw a
least three turbidity measurements are required for the curve through them’’. In doing so for this problem, we are
determination of a two-parameter PSD using either of making some assumptions either about the underlying
the methods or their combination. Otherwise (i.e. using function f .D/ or about the nature of Kscat , or both.
two turbidity measurements only), an infinite number Two approaches have been taken in the literature
of solutions will be obtained. The claims of Melik and to solve for f .D/. When a small number of turbidity
Fogler.48/ were shown.50/ to be unfounded. measurements are available, a type of distribution that
can be described by a few parameters is assumed for
4.4 Resolution of a Conflict in the Literature f .D/ (i.e. Gaussian, log-normal, etc.) and we solve for
these parameters. When many turbidity measurements
The analysis presented by Kourti et al..15,32/ which are available it is possible in some cases to solve for fj
showed that the capability of a turbidimetric technique [the discretized f .D/] at a number of selected diameters,
to provide the PSD is strongly related to the (m, a) Dj . In general, the second approach requires many more
values of the suspension helped to resolve a controversy turbidity measurements than the first.
that existed in the literature. This conflict, on whether
the true PSD of a suspension can be estimated from
4.5.1 Approximate the Particle Size Distribution
turbidimetric techniques, arose between Maxim et al..31/
with a Known Distributional Form
and Zollars..11,12/ As discussed earlier, Maxim et al. used
the turbidity ratio method to estimate the full PSD When a small number of turbidity measurements are
in commercial suspensions, with m D 1.1 and particles available, a convenient approach is to approximate f .D/
with diameters smaller than 1.0 µm. A high correlation with a distributional form that can be described by a
between the parameters of the distribution was observed. few parameters. This form is then substituted into Equa-
The authors stated that a valid estimate of the true tion (15a) or (18) and its parameters are estimated from
distribution is impossible in the presence of experimental the turbidity measurements using non-linear parameter
error. According to the analysis presented by Kourti estimation procedures such as the Levenberg – Marquardt
et al..32/ these observations are justified (and were algorithm..52/ The assumed distributional form is one
expected) for the (m, a) values of the suspension and the that is known to describe similar systems (e.g. for latex
turbidity ratio method. Zollars used a simulation study particles the log-normal distribution described by two
with no experimental error to re-estimate the parameters parameters is frequently used).
of log-normal PSDs from theoretically calculated specific Extensive reviews on the distributional forms that
turbidities. Of course, the parameters were correctly re- have been used to describe the PSD in various sus-
estimated. The author stated that ‘‘turbidimetry does pensions can be found in several sources..15,33,53/ Some
not exhibit multiple solutions as reported by other workers.1/ calculated turbidity for a Gaussian distri-
investigators’’. His observations.11,12/ are also justified bution of spheres. However, most naturally occurring

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
14 PARTICLE SIZE ANALYSIS

populations are skewed and not symmetrical. The log- since the contribution of tails of very large particles to the
arithmic normal distribution.11,15,20,33/ has been used to theoretically calculated specific turbidity is eliminated.
described such populations. In this distribution ln D is Note that if Equation (22) can be directly used, then
normally distributed. The normalized PSD, f .D/ is given the polydispersity of a suspension with a log-normal
by Equation (21): distribution is given by Dw /DN D exp.3s2 / and the
  particle volume fraction is given by Equation (26):
1 .ln D ln Dg /2
f .D/ D p exp .21/
sD 2p 2s2 f D ND3g exp.4.5s2 / .26/

where ln Dg is the mean of ln D and Dg is both the median Some workers.4,5/ had named as log-normal the
and the geometric mean diameter; s is the standard distribution with the frequency function in Equation (27):
deviation of ln D and it is called the geometric mean
standard deviation. For small values of s (s < 0.14) f 0 .D/ D Df .D/ .27/
the frequency curve can be closely approximated by a
normal distribution..33/ The nth moment of the log-normal The two distribution functions are related by Equa-
distribution is given by Equation (22):.54/ tion (28):.55 – 57/
  f .D/ dD D f 0 .D/ d ln D .28/
n .ns/2
Jn D E.D / D exp n.ln Dg / C .22/
2
Therefore, if f 0 .D/ is used, then Equation (4) should be
The number- (DN ), volume – surface- (D32 ), weight- (Dw ) integrated with respect to ln D.29/ and not with respect to
and turbidity- (Dt ) average diameters can therefore be D. This distribution described by f 0 .D/, has been called a
calculated as (Equation 23):.15/ ZOLD.
When applying Equation (15b), f .D/ should be
 1/3
J1 J3 J4 J6 expressed on a weight basis. A log-normal distribution on
DN D ; D32 D ; Dw D ; Dt D .23/ a weight basis can be used..20,29,58/ If the number distri-
J0 J2 J3 J3
bution is log-normal, then the weight distribution is also
The calculation of the moments of the estimated PSD log-normal with the same geometric standard deviation s
from Equation (22) are correct only if the integration is (Equation 29):.54,59/
done from Dmin D 0 to Dmax D 1 or, in practice, if the  
upper limit of the PSD, D0max , is such that when using 1 .ln D ln Dwm /2
F.D/ D p exp .29/
f .D/ from Equation (21), we have Equation (24): sD 2p 2s2
Z D0max where F.D/ is the corresponding weight distribution for
f .D/ d.D/ D 1.0 .24/ f .D/ of Equation (21). Dwm is both the median and the
0
geometric mean diameter on a weight basis. Dwm and Dg
Sometimes it may be known a priori (either from are related by Equation (30):
another method such as EM, DLS or particle chromatog-
raphy, or by filtering of the suspension) that particle Dwm D Dg exp.3s2 / .30/
sizes smaller than Dmin and/or larger than Dmax do not
exist. In these cases if we wish to estimate a log-normal Heller et al..7,47,53/ used another type of distribution
envelope we can do so. In the parameter estimation all that can be described by two parameters. However, the
the integrations should be done between Dmin and Dmax . maximum polydispersity for that distribution is DW /DN D
A log-normal envelope, characterized by Dg and s, can 1.5..15/ Therefore, even though the distributional form
be estimated; however, the PSD should be normalized can describe positively skewed distributions, it cannot
before each iteration in the parameter estimation. Fur- successfully describe highly polydisperse systems.
thermore, the calculation of the various moments should
be done using the explicit equation for the moments 4.5.2 No Assumptions on Shape of Particle Size
(Equation 25): Distribution
Z Dmax If no distributional form is assumed for f .D/, then an
Jn D Dn f .D/ dD .25/ alternative is to solve for f .D/ at a finite number of
Dmin
points..8,60 – 64/ In this approach, in the space of diameters
and not Equation (22). The log-normal distribution from Dmin to Dmax we define P points (i.e. P diameters Dj ,
estimated with previous knowledge of Dmin and Dmax is j D 1 : P) and we try to estimate fj at the corresponding
expected to be a more accurate estimate of the true PSD, diameter Dj . The number of P points must be less than

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 15

or equal to the number of specific turbidity or turbidity (Equation 38):


ratio measurements made.
Note that if the integrand of Equation (4) is discretized X
P 1
q.fO / D b2 .fO12 C fO22 / C .2fOj fOj 1 fOjC1 /2 .38/
at P 1 intervals, then turbidity at a given wavelength l0i
jD2
can be approximated by Equation (31):
with b × 1. The solution now is:
X
jDPDDmax
tl0i ³ fj gij .31/ f D .0T 0 C xH/ 1 0T t∗ .39/
jD1DDmin
where (Equation 40):
where fj ³ Nf .Dj / D Nj is the discretized PSD and gij
depend on the kernel and the quadrature equation 1 C b2 2 1 0 0 0 0 ...
used..27,61/ gij is evaluated at given Dj from the Mie 2 5 4 1 0 0 0 ...
theory. Note that for the simple case that the suspension 1 4 6 4 1 0 0 ...
is a mixture of P monodisperse suspensions with diam- HD 0 1 4 6 4 0 1 ...
eters D1 , D2 , . . . , DP , with relative number of particles ... ... ... ... ... ...
... ...
N1 , N2 . . . , NP , we have Equations (32) and (33): 0 0 0 0 1 2 4 5
0 0 0 0 0 1 C b21 2
p 2 .40/
gij D D K.l0i , Dj / .32/
4 j Note that in this method one has to choose the location
X
P of Dmin and Dmax and the number of intervals but also the
Nj D N .33/ value of x. The choice of these parameters was discussed
jD1 in detail by Eliçabe and Garcia-Rubio..61,62/
This technique was tested with simulations.61,62/ for
When turbidity is calculated for L wavelengths then
the estimation of unimodal and bimodal PSDs of
Equation (31) can be written as Equation (34): varying breadth and mean diameter. Satisfactory results
were reported.61/ for broad distributions with particle
t ³ 0f .34/
diameters in the micrometer range (distributions with
where t D [tl01 , tl02 , . . . , tl0L ]T , 0 D fgij g, f D [f1 , f2 , . . . , Dmax D 1.45 or 2.25 µm). For m D 1.2 the results were
fP ]T and T denotes transpose. not satisfactory for distributions covering sizes from 50 to
In the inverse problem, we have to estimate f from 325 nm, that is, a regime where a large number of latex
turbidity measurements t∗ (Equation 35): particle diameters are expected to be found.
More attempts to resolve the PSD with no assumptions
t∗ D 0f C e .35/ were reported later by other workers..63,64/ Finally, an
empirical method that relates turbidity measurements to
where e is the vector of errors (measurement errors particle size using latent variable methods has also been
plus errors due to discretization of the true distribution). reported..25/
Solving Equation (35) is a linear algebra problem..65/ The Gulari et al..8,60/ tried to estimate PSDs in polystyrene
least-squares solution is Equation (36): suspensions (m ³ 1.2), from a large number of turbidity
measurements at the wavelength range between 300 and
fls D .0T 0/ 1 0T t∗ .36/ 800 nm. In each case, the unknown PSD was assumed
to be the sum of 20 equally spaced delta functions
However, because the rank of 0 is incomplete, small (gij from Equation 32). The height (mass fraction) and
errors in t∗ result in large errors in the estimated position of each diameter was optimized using a non-
parameters. For this reason, regularization methods.65/ linear least-squares minimization procedure. During the
should be used. A penalty function is introduced to the analysis, the insignificant parameters are zeroed and the
original least-squares problem (Equation 37): parameter matrix collapses until convergence is reached.
The suspensions tested in that study were synthesized by
minfj0f̂ t∗ j2 C xq.f̂ /g .37/

mixing two or three monodisperse latexes, with particle
diameters approximately 1 µm separated from each other.
where x is a nonnegative parameter and j j indicates the The distributions were resolved on a weight basis. In each
modulus. q.f̂ / could be the sum of squares, or the first or distribution the weight fraction of the small particles
second differences, of the estimates. Eliçabe and Garcia- was significant. Users should be careful when using this
Rubio.61,62/ added another term to the sum of squares of approach (i) when a very small weight fraction of the
the second differences in order to force f1 and fp to zero small particles is present (see discussion on bimodal

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
16 PARTICLE SIZE ANALYSIS

distributions, section 7.3) and (ii) for suspensions with in the particular application.9,10/ were very large (a as
(m, a) values where y does not change significantly with large as 9.5). For larger m and a values the approximating
a. Readers should also keep in mind that the resolution equations for the calculation of Kscat include more
of the method depends on the location of the assumed involved functions.37/ of a [or r D 2a.m 1/] than the
delta functions (number and width of D intervals) and simple power series expansion attempted in the method
of course on the number of measurements (observations) of moments. As discussed earlier, the term Kscat D2 can
available. be set proportional to Dy for larger values of a, and the
In general, the estimation of f .D/ at a finite number value of y changes as a function of the particle size. For
of points is expected to give satisfactory results only poly(vinyl acetate) latexes in particular, and for a very
for certain suspensions (i.e. corresponding to certain wide range of a values, the value of y is very close to
m and a values). The results will be questionable in 4.0. For most of the diameters used for the calculation
(m, a) regimes where y does not change significantly of the above term in that study.9,10/ at each wavelength,
with a. Kourti and MacGregor.44/ emphasized that more the a values were such that y was close to 4.0. It is clear
sophisticated deconvolution methods cannot overcome that a reasonable value from the parameter estimation
the limitations that turbidity ratio and specific turbidity is expected only for C4l , since all the other coefficients
have for particles in the (m, a) regimes where turbidity is are practically equal to zero. From the results reported
simply proportional to a moment of the PSD. Several by Kiparissides,.9/ only C4l seems to have a consistent
years later, Llosent et al..27/ confirmed this by trying value for all the wavelengths; the estimated values of
several deconvolution approaches to estimate the PSD all the other coefficients change, both in sign and value,
in the Rayleigh regime; only turbidity-average diameters randomly..15/ It should be clear that with meaningless
could be correctly estimated and not the PSD. estimates of the Cil coefficients the correct estimation of
the moments of the PSD is impossible.
4.5.3 The Method of Moments
This method was presented.9,10/ as an alternative tech-
nique to solving for the PSD, but has some limitations for
5 ESTIMATION OF AVERAGE DIAMETERS
the particular application it was used. In this approach the
term K.D/lm , np /nm /pD2 /4 in Equation (4) is approxi- Several workers have tried to use turbidimetry to estimate
mated with a polynomial (Equation 41): only an average of the PSD of a polydisperse system. To
do this they treat the system as monodisperse. That is,
  X P for specific turbidity, instead of using Equation (15a) they
D np pD2
K , ³ Cil Di .41/ use Equation (12) and estimate a single, average diameter
lm nm 4 iD0 which is termed an ‘‘apparent diameter’’ (Equation 42):
Turbidity can thus be expressed as a sum of P moments tl0 3 K.aav , m/
of the unknown PSD. Using P turbidity ratios (P C 1 D .42/
f 2 Dav
turbidity measurements) one can estimate the P moments
(J0 of the normalized distribution is unity). The PSD is where aav D pDav /lm . Dav is the apparent diameter that
then represented analytically using a set of its moments, always corresponds to the Dy3 average of the PSD.
by modifying a gamma distribution which serves as a basis. The concept of the ‘‘apparent’’ diameter was intro-
The method was applied for particle size determination in duced and theoretically explained in very careful work
poly(vinyl acetate) latexes..9,10/ First, the Cil coefficients by Meehan and Beattie..4,5/ They commented that the
at each wavelength were estimated from Equation (41) apparent diameter obtained by treating a polydisperse
by calculating the value of Kscat pD2 /4 exactly from the system as monodisperse corresponds to an average of
Mie theory for a set of 20 diameters (between 0.04 and the unknown PSD. They also concluded that different
0.8 µm) and setting the value of P equal to 6. Six Cil kinds of averages are yielded by different light scattering
were estimated for each wavelength. The procedure was methods, and that the kind of average obtained by a light
repeated for seven wavelengths (350 – 800 nm). scattering method depends on the value of m of the sus-
As discussed in a critical review,.15/ the expansion of the pension, the type of the PSD and the actual size of the
scattering coefficient to a power series of a (or for constant particles in the suspension. Their comments have been
lm , to a power series of D) is an approximation applicable corroborated by numerous reports.3,4,7,15,16,29,66/ where it
for small a and ma values. In these cases the coefficients of is either discussed and/or shown experimentally that the
ai (or Di ) are known functions of the optical constants.37/ type of average obtained from the apparent diameter
(i.e. no estimation of Cil is necessary). The a values depends on the turbidimetric technique used and the m
corresponding to the diameters and wavelengths reported and a values of the suspension.

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 17

It has been suggested.7/ that the ratio of the apparent This characteristic can be exploited to provide a fast
diameters estimated at two wavelengths can provide and inexpensive tool to monitor particle growth in a
a rough idea of the polydispersity of the system. suspension where the value of m is not accurately known.
However, this is not always true. It has already been By simply measuring specific turbidity (turbidity and
shown that, sometimes, especially for small m values solids fraction) at each time interval, the relative increase
(m ³ 1.1), the apparent diameters corresponding to very in the turbidity-average diameter as a function of time
broad distributions or to bimodal distributions do not can be recorded. From a measurement of the diameter or
change significantly with the wavelength, leaving the the PSD with another technique (which is independent
impression that the system is monodisperse. Note that for of m) at the end of the process, we can back-calculate the
monodisperse systems, the apparent diameters obtained turbidity-average diameter at each instant in time.
from two different turbidimetric methods (i.e. specific A report.28/ that a correct turbidity-average can be
turbidity, turbidity ratio) should be numerically very close obtained for a values as large as 4.74 from specific
to each other (i.e. equal, within experimental error). turbidity measurements should be treated with caution.
Specific turbidity has been the only turbidimetric These observations were rather coincidental with the
method used in the literature for the determination polydispersities of the systems investigated, as explained
of averages of the PSD. This is because, as already in detail elsewhere..15/ In the same work.28/ the term
discussed, it is the only turbidimetric method that can ‘‘weight’’ average was used for what in fact is the
provide estimates of meaningful averages of the PSD ‘‘turbidity’’ average, as can be deduced from the definition
(Dy3 averages) at some well-defined (m, a) regimes. in the original publication,.67/ where EM results for the
The following averages may be obtained from specific latex samples used in the later study.28/ were presented.
turbidity: (i) the turbidity-average for very small particles
(a < 0.5) and any value of m, (ii) the volume – surface 5.2 Volume – Surface-average Diameters
average for very large particles and any m value and
(iii) weight-average for a combination of (m, a) values Volume – surface-average diameters have been estimated
and m < 1.2 which can be identified very easily for the in the literature from specific turbidity measurements
system under investigation. In this section we discuss of suspended particles that were very large compared
the use of specific turbidity for the estimation of with the wavelength (Kscat ³ 2). It should be noted that,
meaningful average diameters. A critical discussion of although the value of the oscillating Kscat stays very close
several applications reported in the literature is presented, to 2.0 for large a values, the error by assuming it to be
so that the readers can identify with situations closer to equal to 2.0 is small only for very large a values (i.e.
their needs. the error is within 3% for a > 800 and within 1% for
a > 2000) (see Table 2). For a values around 200 – 260,
where some of the cases reported in the literature are, the
5.1 Turbidity-average Diameters
error of assuming Kscat D 2.0, depending on the spread of
These averages can be obtained from specific turbidity the distribution, could be as large as 6% for m D 1.2 and
measurements, for suspensions with any value of m, 9% for m D 1.1 (at these a values Kscat still depends on m).
provided that all the particles in the suspension are small Therefore, depending on the range of a values covered by
compared with the wavelength at which the turbidity the true distribution, the estimated D32 average diameter
measurements are taken (Rayleigh regime). could be within 10% error of the true value.
A practical application, where the growth of sil- Bagchi and Vold.66/ showed experimentally that for
ver bromide grains is monitored on-line by following very large particles, the specific turbidity varied linearly
turbidity-average diameters, was reported by Haseler with the inverse of the volume – surface-average diameter
and Parkin.6/ for an Ostwald ripening process. They were (1/D32 ). However, a linearity observed with respect to the
careful to choose long wavelengths [infrared (IR) regime, inverse of other averages was coincidental and attributed
2.22 µm] so that the Rayleigh scattering theory could be to the fact that the degrees of polydispersity of the latexes
applied even for the largest particles in their system..15/ investigated were very similar..15/ Kubota et al..68/ dis-
Note that in the Rayleigh regime, if two particle cussed this coincidental linearity and presented more
suspensions have the same value of m, the ratio of their experimental results from suspensions of various poly-
specific turbidities measured at the same wavelength (see dispersities, verifying that only the specific turbidity vs
Equation 16) gives the ratio of their turbidity-average (1/D32 ) curves are linear, as was theoretically expected.
diameters, independent of the value of m (Equation 43): Kourti.15/ showed that a claim by Dobbins.69/ and
  Dobbins and Jizmagian.70/ that the volume – surface-
t1 /f1 Dt1 3 average diameter can be obtained from one specific
D .43/
t2 /f2 Dt2 turbidity measurement in any particle size regime cannot

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
18 PARTICLE SIZE ANALYSIS

be supported. Their observations, that ‘‘the specific diameters from specific turbidities at 600 and 400 nm,
turbidity is primarily dependent on the D32 diameter and obtained for a variety of polydisperse poly(vinyl acetate)
only weakly dependent on the type of the distribution’’, latexes (m ³ 1.1) with different distributions (unimodal,
were shown to be coincidental with the types of multimodal) in the submicrometer regime, is numerically
distributions that they used for their investigation. These very close to the true weight average of the distribution.
distributions had also their weight averages close to each That means that for m D 1.1 the weight average from
other and all of them included particle sizes in the same submicrometer particle suspensions can be estimated very
range. Similar claims by Haseler.19/ were also discussed simply, from two specific turbidity measurements at 400
in Kourti..15/ It was shown that in the examples reported and 600 nm. Similar heuristic rules could be derived for
by Haseler,.19/ the estimated D32 was close to the true other suspensions with small m values.
value because the location of the whole distribution was Cheesman,.72/ for example, suggested a technique in
correctly estimated, and not because specific turbidity is which the specific turbidity is measured at several wave-
always expected to give a correct D32 as claimed..69,70/ lengths and the apparent diameter (Dav ) is estimated
It has been suggested.71/ that the specific turbidity of a for each wavelength. Then the corresponding aav values
polydisperse system can be expressed as Equation (44): are calculated (aav D pDav /lm ) at each wavelength. The
apparent diameter that gives aav near the a value of the
tl 0 3 K.aav , m/
D .44/ first point of inflection of the Kscat vs a curve is chosen to be
f 2 D32 the weight-average diameter (for m D 1.1, this a value was
where (Equation 45): chosen 2.7 and for m D 1.2, it was 2.2). The apparent diam-
Z 1 Z eter selected this way in the polydisperse systems he stud-
1
K.aav , m/ D2 f .D/ d.D/ D D2 K.aav , m/f .D/ d.D/ ied was within 4% of the true weight-average diameter.
0 0 The observations of Cheesman can be explained
.45/ theoretically. At the a value at the inflection point (see
where aav corresponds to an average diameter Dav . It was peaks of the first ripple of the y vs a curve in Figure 2)
suggested that the two parameters aav and D32 can be used the value of y, for almost any value of m, is close to 4.5. If
to interpret turbidity data over a wavelength range. The the weight-average diameter corresponds to this a value,
D32 diameter will correspond to the true PSD whereas aav then the a values for the individual particle diameters of
will correspond to the average diameter ‘‘seen’’ by the the distribution are located around it and the value of y
measurements ‘‘averaged over the wavelength range’’. varies between 4 and 4.5. Hence in this technique, the
However, by definition, the only correct form for the wavelength is chosen such that the entire distribution is
calculation of an average diameter from specific turbidity brought to an a regime where y is close to 4.0 and the
measurements is that of Equation (42). Equation (44) is apparent diameter is the weight-average diameter.
valid only for very large diameters, where K.aav , m/ D 2, As has been pointed out,.44/ the fact that the weight-
independent of particle size. In any other case if we average diameter of suspensions can be correctly esti-
try to estimate the two parameters (aav and D32 ) in mated even though the PSD determination is impossible
Equation (44), ‘‘over a range of wavelengths’’, these is a very useful property of the specific turbidity. It
estimates will be highly correlated. Their numerical was suggested.44/ that this property could be utilized
values will depend on the wavelength range used and for practical applications, in situations where following
the D32 estimate will not, in general, correspond to the an average diameter is sufficient to monitor the progress
volume – surface-average of the true PSD. of a process. An example would be on-line particle size
measurements to monitor the progress of polymerization
5.3 Weight-average Diameters reactions. Gossen.23,46/ and Gossen and MacGregor.24/
used this suggestion and applied specific turbidity on-
For suspensions with small values of m (usually m < 1.15) line to monitor weight-average diameters of poly(vinyl
and particle sizes corresponding to certain a values acetate) particles during latex production. The results
(i.e. for m D 1.05, 3 < a < 14; for m D 1.15, 3 < a < 6), reported.23,24,46,73/ show good agreement between the
the apparent diameters obtained from specific turbidity weight-average diameters obtained by turbidimetry and
measurements are numerically close to the weight- those obtained by on-line DLS.
average diameter, Dw . This average is essentially the only
correct estimate of the particle size that can be obtained
5.4 Other Applications of Turbidimetry
for polydisperse systems in these (m, a) regimes. If users
know the value of m of their suspensions they can devise The dependence of turbidity on particle size can also be
simple ways to obtain this average. For example, it has exploited in other ways than its conventional use in par-
been observed.15/ that the mean value of the apparent ticle size determination. Kourti.15/ suggested that simple

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 19

turbidity measurements at one wavelength can be used comparable to the wavelength the difference between the
to monitor polymerization processes in order to detect true and apparent turbidity should be minute. Bohren
the onset of several events, as for example the begin- and Huffman.75/ suggested that although conventional
ning of the reaction in latex systems (i.e. the end of spectrophotometers are not usually designed to reject
induction period) or the onset of agglomeration. By fol- scattered light at small angles, such instruments may be
lowing turbidity measurements at one wavelength during adequate for scattering measurements if the particles
the emulsion polymerization of vinyl acetate in a con- are not too large. For monodisperse polystyrene latexes
tinuous laboratory-scale reactor, Kiparissides et al..9,10/ (m D 1.2) with diameters of 109, 312 and 1050 nm and
could detect whether the reactor was operating under turbidity measurements at wavelengths between 300 and
steady-state conditions or under oscillations. Turbidity 600 nm, they used a conventional spectrophotometer,
measurements at 400 nm were used by Kourti.15/ to fol- suggesting that the forward scattering should not be
low qualitatively the effect of several variables on particle significant. Walstra.76/ suggested that sometimes errors
size during vinyl acetate emulsion polymerizations. Tur- attributed to forward scattering.18/ were in fact due to
bidity measurements at one or more wavelengths (and extremely high optical densities that the investigators had
reduced to the same degree of dilution) can also be used included in their results.
for routine operations to monitor the progress of the The brief analysis that follows illustrates the conse-
reaction and detect any deviations from a known desir- quences of the error added to the value of the true
able path. Finally, Crawely et al..74/ discussed the use of turbidity due to forward scattering. The effect that an
turbidity measurements at several wavelengths in studies error in the measured turbidity value has on the estimated
of crystallization, agglomeration and attrition. particle size depends on the turbidimetric technique used
and the (m, a) values of the system and it cannot be
generalized. The turbidity ratio utilizes two turbidity
6 PROCEDURAL DETAILS measurements at different wavelengths, therefore the
error due to forward scattering is introduced twice in
6.1 Instrumentation and Forward Scattering this method, but only once in the specific turbidity. In
Figure 5 we plot.15/ as a function of a, for two different m
The turbidity of a suspension can be measured very values, the percentage error in the estimated diameter in
easily in a laboratory or a plant environment. Very often monodisperse suspensions for a 3% error in the measured
spectrophotometers in the UV/VIS range are utilized for specific turbidity. Note that the effect changes with the
this purpose. A dilute suspension is inserted in the cell of (m, a) combination. For m D 1.2 and a values near the
the spectrophotometer. We have (Equation 46): specific turbidity maximum (which for m D 1.2 occurs at
  a D 7.33), the error becomes very large. However, for a
A 1 100
tl0 D 2.303 D ln .46/
l l T
14
where A is the absorbance reading, T is the percent
transmittance at wavelength l0 and l is the optical path 12
(i.e. width of cell). There are also reported applications
Error in diameter (%)

where turbidity measurements at longer wavelengths, 10


in NIR regime,.6,16,17,19,26/ were used for particle size
determination. Sometimes, certain modifications may 8
m =1.2
be necessary.19/ in the instruments used to measure
turbidity so that forward scattering is rejected and 6
theoretical calculations of turbidity can be legitimately
compared with the measured values. This is because 4
for the determination of particle size from turbidity
measurements, it is assumed that the detector measures 2 m =1.1
the attenuation of the beam of the source light. However,
all the spectrophotometers measure the attenuated light
0 1 2 3 4 5 6 7
from the source plus a certain amount of light scattered at
small angles. Forward scattering will add error to the value α = πD/λm
of the true turbidity. Reports in the literature.34/ mention
Figure 5 Percentage error in the estimated particle diameter of
that the larger the particles the greater is the possible monodisperse suspensions plotted as a function of a for m D 1.1
discrepancy between the true and the measured (or and 1.2 when there is a 3% error in the specific turbidity
apparent) turbidity, but for small particles and particles measurements.

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
20 PARTICLE SIZE ANALYSIS

values less than 4.0 the error in the diameter is the same 400 nm and suspensions with m D 1.1 and 1.2: a Hewlett-
for m D 1.1 and 1.2, and it is approximately 4.0%. For Packard Model 8450A at wavelengths of 350 – 700 nm
water medium (nm ³ 1.33), a value of a D 4 corresponds and a bandwidth of 2 nm and a Bausch and Lomb (now
to D/l0 ³ 1 (a D 4 D pDnm /l0 ), which means that for Spectronic Instruments) Spectronic 20 at wavelengths of
particles smaller and comparable to the wavelength the 380 – 580 nm and a bandwidth of 20 nm. The accuracy of
error is small. Therefore, if the error due to forward the method was tested by estimating the size of monodis-
scattering introduces a 3 – 4% error in specific turbidity, perse suspensions with known diameters and also by
then for particles comparable to l0 the effect on particle comparing the size estimated by turbidimetry with that
size should not be significant, and this would explain the measured by EM and DLS. For numerous unknown sam-
observations reported by various workers. A final note on ples, the diameters estimated by turbidimetry were in
Figure 5: it demonstrates that a 3% error in the specific very good agreement with those measured by DLS and
turbidity measurements may cause a very large error in EM. Forward scattering did not introduce any problems.
the estimated size for a values near the peaks or the Gossen and MacGregor.24/ used for on-line measure-
valleys of the Kscat vs a curve. This is due to the light ments in a pilot-plant environment a Bausch and Lomb
scattering behavior and has nothing to do with the source single-beam Spectronic 20 equipped with a cylindrical
of the error (i.e. this 3% could be due to either a wrong Pyrex flow-through cell with a 1 cm pathlength, without
solids fraction measurement or to forward scattering, and any modifications for forward scattering. The incident
would have the same effect on the estimated size). beam had a bandwidth of 20 nm and the wavelength was
Errors in turbidity measurements due to forward 400 nm. Weight-average diameters up to 200 nm obtained
scattering are discussed in a number of references, during latex production were in very good agreement with
where suggestions are given for the modification of weight-average diameters obtained by on-line DLS,.73/
the spectrophotometers..6,18,20,30,49,76 – 78/ Also, several and a change in Dw of the order of 10 nm was detectable.
empirical approaches have been suggested to account for The accuracy of the results was considered satisfactory
the difference between true and apparent turbidity..20,66/ for their objective to monitor changes in the particle size
For example, a correction factor (R) can be estimated.20/ due to process disturbances.
for several suspensions with known particle diameters; R
is equal to the ratio of the measured to the theoretically 6.2 Refractive Indices of the Particles and the Medium
expected turbidity. This factor can then be used for
unknown suspensions, but with (m, a) values in the same Turbidity is a very strong function of the optical properties
range for which it was estimated. In another approach, of the particle suspension, that is, the ratio of the refractive
a calibration was used.66/ to relate apparent specific index of the particles to that of the medium, m, and the
turbidity with the true particle size for large latex particles refractive index of the medium, nm . The latter is used for
where forward scattering is significant. the calculation of lm and consequently the calculation of
Users can assess the accuracy of the turbidity mea- a. For the correct determination of the particle size the
surements of their spectrophotometers by comparing the values of these properties must be accurately known. The
experimental values with theoretically calculated values values of the refractive indices for different materials can
for monodisperse suspensions with known diameters, cov- be found in relevant handbooks..79/ Refractive indices
ering the range of (m, a) values that they plan to analyze. can also be directly measured, or estimated using a
Before proceeding to any modifications one should judge combination of techniques as described below.
whether the accuracy is satisfactory for the purpose of the Llosent et al..27/ used a Bellingham Abbé 60 refrac-
study to be carried out using the turbidimetric results. For tometer to measure the refractive index of latex particles,
example, if the purpose of the study is to use the method as np , required to calculate m for their work. The refractive
a reproducible and inexpensive way to monitor particle index of the polymer film was measured at several wave-
size during a reaction, then the accuracy of the spec- lengths. They then used the measured values to estimate
trophotometer with no modifications may be satisfactory, the constants for a Cauchy.80/ equation that expresses np
and similarly if the objective is to detect the qualitative as a function of the wavelength l0i (Equation 47):
effect of various parameters (emulsifier, impurities) on B C
the particle size of the latex. For other applications that np .l0i / D A C 2
C 4 .47/
l0i l0i
require more accurate determinations of larger particles,
one may consider modifications of the spectrophotometer This equation was subsequently used to calculate np at
and the use of a correction equation.76/ that accounts for other wavelengths.
the angle of acceptance of the detector for the calculation If the value of the refractive index of the medium nm
of the scattering coefficient. Kourti.15/ used two spec- is known but np and therefore m are not known, it may
trophotometers without modifications for diameters up to be possible to estimate m directly using a combination of

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 21

techniques. Kourti.15/ suggested that, since turbidimetry For monomer swollen particles, or particles produced
is very sensitive to the value of m, it is probably the best by core and shell polymerizations, one approach is to
method to ‘‘estimate’’ the values of m as a function of calculate the value of m by treating the particles as coated
wavelength for a specific suspension. Suppose we wish to spheres. For coated spheres the refractive index ratio can
estimate the value of m for suspensions of polybutadiene be calculated as m D m1 v1 C m2 v2 , where m1 is the ratio
particles in water. Using a monodisperse suspension of of the refractive index of the core to the medium and
polybutadiene particles in water with known diameters v1 is the volume fraction of the core; m2 and v2 are the
(estimated by EM or DLS), the scattering coefficient and corresponding values for the coating. (v1 C v2 D 1.0). This
therefore m can be calculated from the specific turbidity approximation is not very accurate for thin coatings. In
measurements at one wavelength (provided nm is known). the above approach, for monomer swollen polymers, the
By measuring specific turbidities at more wavelengths one volume fractions of the monomer and polymer are treated
can estimate the dependence of m on wavelength. The as core and coating respectively. It should be emphasized
advantage of this approach is that the value of m for that this is an approximation, and the true morphology
a particular suspension is measured in the environment of monomer swollen particles may be different than a
in which it will be used. This may be more meaningful core and coating approximation. Bohren and Hoffman.75/
than, say, using a latex film to measure the refractive also presented an algorithm to calculate the scattering
index of particles and dividing by the refractive index of functions of coated spheres (consisted of homogeneous
medium. Once m has been measured it can be used for core and homogeneous coating of uniform thickness) and
suspensions of the same material (i.e. polybutadiene in they discuss the limitations of this algorithm, in general,
this example), but of course for any particle sizes or PSD. for anisotropic spheres. For very accurate calculations,
DLS is an excellent method for this purpose because these approximations to estimating the value of m for
the particle sizes can be measured in suspension and core and shell particles must be checked experimentally.
for monodisperse particles DLS is independent of the This can be done as mentioned earlier by using another
refractive index. Another method to determine optical method (i.e. DLS) to estimate the particle size and then
constants would be angular light scattering. using turbidimetry to estimate the value of m for particles
Sometimes the medium may not be a clear substance with known polymer composition. The above calculations
but may have additives. In latex samples, for example, should be performed on monodisperse suspensions. A
emulsifier may be added to the water medium, or calibration curve could then be constructed to give the
monomer may be dissolved in it if the samples are value of m as a function of the polymer composition in
withdrawn at low monomer conversions. In these cases the particle. If DLS is used to verify the particle size of
the refractive index of clear water cannot be used. The monomer swollen particles, the same degree of dilution
must be used to make sure that the same particle size is
refractive index of the medium for a given concentration
measured (because monomer is dissolved in water and
of water-soluble monomer (or other substance present
extracted from the particle with high dilutions).
in it) can be read directly from a calibration curve
For monomer swollen particles and carboxylated
of refractive index as a function of the concentration
particles that swell in water, both the density and the
of monomer in water. Such calibration curves can be
refractive index depend on the degree of swelling and
easily constructed from refractive index – concentration
appropriate corrections should be made. In monomer
increments obtained from differential refractometers. In
swollen particles, the amount of monomer associated
high dilutions the effect of monomer on the refractive
with the particles (and therefore the refractive index of
index of the medium is usually negligible.
the particles and the medium) depends on the conversion,
the solubility of monomer and the degree of dilution of
6.2.1 Inhomogeneous Particles/Latex Particle the original sample.
Applications Latex produced by multicomponent polymerization is
also a case where the calculation of the refractive index of
Sometimes the particles in the suspension may not the particles may present difficulties. The refractive index
be homogeneous. Examples from latex systems are depends on the polymer composition and may change
carboxylated particles that swell in water, particles during the reaction due to composition drift. Therefore,
swollen with monomer in latex samples withdrawn at calibration may be needed to relate the refractive index
low conversions and particles produced by core and of the particles to their composition (by taking into
shell polymerizations. There are several issues related to account the conversion of the monomers at the time of
determining the optical properties of such particles and sampling). When monitoring particle growth with on-line
research is active on light scattering by inhomogeneous turbidimetry during multicomponent polymerizations,
particles..81 – 83/ on-line gas chromatography can be utilized to provide

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
22 PARTICLE SIZE ANALYSIS

information on the composition of the particles. The determination is attempted at these short wavelengths
refractive index can then be calculated from the particle without accounting for the distribution of the absorbing
composition utilizing some type of calibration. On-line monomer between the medium and the particles, the
densitometry combined with on-line gas chromatography results may be questionable. However, calculating the
can provide information on the particle volume fraction, monomer distribution is tedious and requires a knowledge
f. More details on this topic can be found in Kourti..15/ of conversion, partition coefficient of the monomer
between the water and the polymer phase and the degree
6.3 The Choice of the Wavelengths of dilution. It is clear that using short wavelengths is
not practical if the objective is to follow particle growth
As discussed earlier, the successful determination of the during the reaction.
PSD or of an average particle size of a suspension
depends on the m and a values of the suspension and 6.4 Preparation of Samples
the turbidimetric technique used. The refractive index of
the medium and therefore the value of m of a suspension In particle size determination by light scattering methods,
can be altered by adding other substances to the medium it is usually assumed that the scattering by an array
[e.g. glycerol.5,16/ ] provided that these substances do not of particles is incoherent (the particles are randomly
affect the particle size. The range of a values covered positioned in space to avoid mutual polarization) so
by the PSD can also be manipulated by changing the that the scattering functions corresponding to an isolated
wavelengths at which the turbidity measurements are particle may be used, and also that there is no significant
taken. As a general rule, wavelengths widely separated multiple scattering. These conditions are practically
from each other should be used when the estimation met.33/ when turbidity measurements are taken with
of the full PSD is required. In this way the range of a sufficiently dilute suspensions. In practice this is achieved
values covered by the PSD changes significantly from one by diluting the suspension and working at a concentration
wavelength to another (the y exponent changes) and the range where the turbidity, at the shortest wavelength to
probability of the turbidity being proportional to the same be used, varies linearly with the particle concentration.
moment of the PSD at both wavelengths is reduced. It Figure 6 shows how turbidity changes with concentration.
should be noted, however, that when wavelengths widely The shorter the wavelength, the lower is the concentration
separated from each other are used, the value of m at each of suspended solids required to avoid multiple scattering.
wavelength must be accurately known; using a constant The fact that the sample needs to be diluted may
m may result in false estimates..15/ If only one value for have an effect on the particle size in certain cases. For
m, estimated at, say, l00 , is available and must be kept example, in latex samples withdrawn at low conversion,
constant for all the wavelengths, then wavelengths on polymer particles swollen with monomer may change size
either side of l00 (i.e. both shorter and longer) and as close by changing the degree of dilution, and this should be
to l00 as possible must be used. taken into account. Also, for latex particles stabilized
To keep wavelengths widely separated, several workers
chose one wavelength in the UV/VIS range and one in 3.0
the IR region. The first (1.7 µm).16,17/ and the second
(2.22 µm).19/ IR transmission windows for water have 2.5 460 nm
been the choices for long wavelengths. Most workers
Turbidity (cm−1)

have avoided wavelengths below 400 nm where some 2.0


particles may absorb light and wavelengths where the 360 nm
medium absorbs light. 500 nm
Eliçabe and Garcia-Rubio.62/ used wavelengths in 1.5
regimes where particles absorb to resolve the PSD for
a computer-simulated polystyrene latex example, where 1.0
560 nm
clear water was assumed to be the medium. The resolution
of the PSD was improved. However, in practical situations 0.5
for latex particle size analysis, it is preferable to avoid
working at short wavelengths where absorption occurs
0.0
(e.g. the styrene ring absorbs at 269.5 nm). In latex 0 1 2 3 4 5 6
samples withdrawn at low monomer conversions, residual Solids concentration (mg mL−1)
monomer which absorbs light at certain short wavelengths
may be present in the particles and extracted owing to Figure 6 Turbidity plotted against solids concentration in the
dilution in the medium. As a consequence, if particle size sample: linearity check.

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 23

by ionic or nonionic emulsifiers or other groups, the The sensitivity of the estimated particle size to such
user must be careful to dilute the sample without risking errors has been investigated by several workers. The
destabilization. Some emulsifier (the same as the one used results are briefly summarized in the following sections.
in the latex production) must be added to the diluting
medium to prevent destabilization. Samples prepared in 7.1 Error in Measurements
this way remained stable for a long time;.15/ turbidity
The effect of measurement errors on the estimated
measurements taken 3 years after the samples were
particle size depends on the technique used (specific
prepared showed no difference within experimental error
turbidity or turbidity ratio) and on the (m, a) values of
with measurements taken for those samples when they
the system and cannot be generalized, as illustrated in the
were freshly prepared. Kourti.15/ discussed in detail other
following. Kourti.15/ reported an average error of 0.4% in
considerations for latex particle size determination, for
the repeatability of transmittance readings. Such an error
both off-line and on-line applications.
is translated to an average 2% error in specific turbidity.
For the (m, a) values used, with m D 1.1 and a < 3.5
6.5 Particle Volume Fraction for Specific Turbidity this error would translate to a maximum error of 2% in
the particle diameter for monodisperse suspensions. If a
Particle volume fraction is measured by gravimetric
different technique, say turbidity ratio, was used for the
analysis. The weight of the dry solids per cubic centimeter
same (m, a) regime, then from a small bias of C0.5%
(c, g cm 3 ) is measured and converted to volume fraction
on the transmittance readings at both wavelengths (the
using the density of the particles (r, g cm 3 ); f D c/r.
same bias would be the case due to forward scattering)
Particle volume fraction is usually measured on undiluted
the error in the particle size for monodisperse particles
samples. The particle volume fraction corresponding to
can be 2 – 3%. For the same size regime, if instead of the
the diluted sample on which the turbidity measurement
bias the error was random (š0.5%), then the error in the
is taken is calculated by taking into account the degree of
particle size can be as large as 9%. Because the effects
dilution. Therefore, the degree of dilution must be known
of turbidity measurements on particle size cannot be
very accurately. Particle volume fraction for on-line
generalized over (m, a) values, a good practice for users
applications can be measured by densitometry..21,23,24/
would be to construct a curve such as that in Figure 5
for the m value of the suspensions with which they plan
to work. The effect of a certain percent error in specific
7 EFFECT OF ERRORS ON THE ESTIMATED turbidity (or turbidity ratio) on the estimated particle
PARTICLE SIZE AND PARTICLE SIZE diameter can be calculated with simulations for a wide a
DISTRIBUTION range. Then, by changing the value of a used (by changing
the wavelength), they can move the system to a regime
The determination of the particle size or PSD of a where measurement errors would not affect the particle
suspension using turbidimetry involves measurements size significantly.
and assumptions. Errors in the measurements and The effect of measurement errors on the estimated
uncertainty in the assumptions will have a minor or a PSD in polydisperse systems was the subject of a detailed
major effect on the estimated size. The sources of errors study.15,32/ and it has been shown to depend on the
can be grouped as follows: technique used (specific turbidity or turbidity ratio) and
on the (m, a) values of the suspension. As discussed in
ž Experimental errors in measurements: these involve section 4, the estimation of full PSD is impossible within
random measurement errors and errors due to for- experimental error in some (m, a) regimes. However, in
ward scattering (bias) in the turbidity measurements. these regimes, if specific turbidity is used, an average
For specific turbidity methods there are also errors in diameter can be correctly estimated and the effect of
the calculation of the solids volume fraction due to measurement error on this average diameter is not large.
dilution, error in gravimetric analysis and uncertainty In regimes where only the weight-average diameter can be
in the value of density of the particles. estimated, for a random error of š3% in specific turbidity
ž Uncertainty in the m values and the dependence at three wavelengths, the maximum error was for a weight
of the m on the wavelength: m is required for the average 5%. As a comparison the maximum error for the
theoretical calculation of the scattering functions. number average in these regimes was 40%..15/

ž For polydisperse samples, error in the assumptions


7.2 Error in the Value of the Refractive Index Ratio, m
related to the PSD: either the form of the PSD has
to be assumed or the number of discrete points, their Kourti.15/ studied the effect that errors in the value of
range and the value of the smoothing parameters. m have on the estimated PSD. It was concluded that

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
24 PARTICLE SIZE ANALYSIS

turbidimetry is sensitive to such errors but the magnitude body of the distribution is therefore the most satisfactory
of the effect on the estimated PSD depends on the result expected in this case.
(m, a) values of the suspension. In certain (m, a) regimes, The error introduced in the estimated PSD when
the effect that the error in m has on the estimated the assumed distributional form is different from the
parameters of the PSD is comparable to the effect of true one has been studied either experimentally or with
turbidity measurement errors, whereas in others it is simulated cases by several workers. Wallach and Heller.7/
much higher. In regimes where the y exponent does not tried to estimate the PSD of three suspensions with
change significantly with the particle size, the errors in the positively skewed, negatively skewed and symmetrical
value of m had a larger effect on the estimated size than distributions, using the turbidity ratio method and
turbidity measurement errors. Therefore, observations assuming a distributional form which can describe only
reported in the literature on the magnitude of this positively skewed distributions. In the cases of negatively
effect cannot be generalized for any (m, a) values. For skewed and symmetrical PSDs, the estimates of the
example, the observation of Wallach and Heller.7/ that parameters chosen to describe the distribution had a
‘‘the error of the disregard of the dispersion of m is significant error; however, Kourti.15/ reported that a
less that experimental error’’ was simply a coincidence careful examination of those results showed that the
of the (m, a) values of the system with which they were estimated PSD enveloped (i.e. recovered correctly)
working. Whenever there is uncertainty in the value the main body of the true distribution. This behavior
of m of the suspension to be analyzed, a sensitivity was expected since the polydispersities of the three
analysis should accompany the results, to indicate the synthesized distributions and of the assumed distribution
degree of confidence. Alternatively, the results should be were very small; no distributions with very long tails to
corroborated with another method. small or large particles were studied.
Sometimes, the dependence of the refractive index of In an extensive simulation study, Kourti.15/ investigated
the medium, nm on the wavelength is known, but only one the error introduced in the estimated PSD when the shape
measurement for the refractive index of the particles, np , is assumed to be log-normal. Specific turbidities were
at one wavelength l00 is available. If this is the case, it is calculated (at 400, 500 and 600 nm) for several polystyrene
recommended that a constant m (calculated at l00 ) is used suspensions of submicrometer particles with several types
at all the wavelengths, and not a variable m calculated of continuous distributions: positively skewed, negatively
with constant np and wavelength-dependent nm . For skewed, symmetrical, bimodals with broad peaks and
water medium, a wavelength-dependent nm decreases overlapping tails. The polydispersities were from 1.04
with longer wavelengths and using constant np will force to 1.9. Then, from these calculated specific turbidities
m to increase with longer wavelengths (when it is expected and assuming a log-normal form, the PSD of the
to decrease); this in turn means significant errors when suspension was re-estimated. The following conclusions
analyzing real data since the estimated particle size moves were drawn that are valid for the estimation of the PSD
in the opposite direction to the true value. Finally, when (using the specific turbidity method) of suspensions of
a constant value of m is used, the error is smaller if the submicrometer particles, excluding the Rayleigh regime
turbidity measurements are taken at wavelengths both where only Dt can be estimated and the (m, a) regimes
shorter and longer and as close as possible to the one at where specific turbidity can only give Dw :
which m was measured.
ž When the true distribution is continuous with no long
In all cases, however, if the wavelength-dependent nm
tails to either small or large particles, the estimated
is known, it should be used to calculate the value of lm
distribution with a log-normal assumption will cover
and therefore a for the scattering coefficient calculation.
the main body of the true PSD when plotted on a
number basis.
7.3 Error in the Assumed Distributional Form ž On a weight basis the main body of the true
The main criticism in the literature about approaches distribution will be covered by the estimated log-
that attempt to recover the PSD by assuming a type of normal distribution, in all cases, even when the true
the distribution (i.e. log-normal, Gaussian, etc.) is that PSD has long tails.
the assumed distribution may be very different from ž The weight-average diameter of the distribution
the true one. When a distributional form is assumed, a estimated with a log-normal assumption is very close
constraint (i.e. the shape of the distribution) is forced to the true weight average of the suspension, even
upon the solution. The expected result is a PSD with the in cases that the true distribution is significantly
assumed shape that will envelope the true distribution; different from a log-normal shape. For the cases
the details of the true PSD are not expected to be studied the weight-average diameter of the PSD
resolved. The estimation of the location of the main estimated when assuming a log-normal form (on a

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 25

number or weight basis) was, in general, within 1.0%


of the weight average of the true PSD.

These conclusions are corroborated by experimental I


results reported by Haseler.19/ for submicrometer
polystyrene suspensions. They also agree with the obser-

f (D )
vations of Yang and Hogg..20/ These authors also studied II
experimentally the error on the estimated PSD when
the shape of the true distribution is different from the
assumed log-normal form. The estimated distribution
was compared with the true one on a weight basis; the
distributions studied were broad, covering particle sizes
from 0.1 to 10.0 µm. They found that ‘‘the estimated dis-
tribution conformed closely to the central portion of the D0 D0 + 0.2 D0 + 0.4 D0 + 0.6 D0 + 0.8
true distribution (on a weight basis) and was not strongly Diameter (µm)
biased towards either of the tails’’.
It should be noted that when using Equation (31) to Figure 7 Bimodal distribution as resolved by a log-normal
estimate fj for diameter Dj .j D 1, 2, . . . , P/, instead of assumption. Case I: the estimated log-normal distribution
assuming a form for the PSD, there are also assumptions (continuous line) and the true distribution (histogram) are
involved. These include the choice of the parameter in located between 0 and 0.8 µm (D0 D 0.0). Case II: the estimated
log-normal distribution (dashed line) and the true distribution
the regularization algorithm and the number of points (histogram) are located between 0.4 and 1.2 µm (D0 D 0.4).
to which the PSD will be discretized and the location
of the Dj s. The number of points depends on the
available number of turbidity measurements at different in Figure 7 shows the distribution (on number basis)
wavelengths. For a study on the effect that the number estimated with a log-normal assumption for polystyrene
of discrete points at which fj is estimated has on the suspensions (m ³ 1.2) where the true distribution (shown
estimated PSD, readers are referred to Eliçabe and by the histogram) was in the size range between 0.0 and
Garcia-Rubio..62/ In that study the true distributions were 0.8 µm (D0 D 0.0). Distribution II is the log-normal PSD
all simulated continuous distributions. estimated when the size range for the true distribution
There is also the question of whether there is an advan- was between 4.0 and 1.2 µm (D0 D 0.4). Note that
tage in proceeding with collecting more wavelengths and although the two histogram distributions have the same
trying to estimate the distribution using Equation (31). In breadth (0.8 µm), in case II the distribution has a smaller
the case of continuous unimodal distributions, the reso- polydispersity (since the average size is larger). Note that
lutions of the two techniques under experimental error in case I the estimated log-normal distribution tends to
seem to be comparable; an assumption of a log-normal envelope only the larger particles of the true distribution,
shape that will provide the main body of the true PSD and whereas in case II it uncovers the correct range but of
the true Dw may be sufficient for most routine applications course cannot resolve the bimodal shape.
in the submicrometer and lower micrometer region. When dealing with bimodal distributions one should
keep in mind that the contribution of small particles to
the value of the specific turbidity could be totally obscured
7.3.1 The Resolution of Bimodal Distributions
by the contribution of the large ones, depending on their
Regularization methods have the potential to recover the a values (relative to the a value location where the peak
PSD without assuming any shape, if the true distribution of Kscat occurs). As an example, assume a polystyrene
is a bimodal one with two broad peaks, such as that suspension (m ³ 1.2) with a bimodal distribution which
shown by the histogram in Figure 7. Eliçabe and Garcia- consists of two monodisperse populations with D1 D
Rubio.62/ resolved such a simulated distribution where the 0.6 µm and D2 D 0.1 µm, and where the number of
particle diameters covered a very wide range from 50 to small particles is 10 times the number of the large ones
2550 nm; in that study wavelengths in the absorption range (N2 /N1 D 10). The contribution of the small particles to
were included for the turbidity measurements. For such the total specific turbidity at 400 nm is only 0.46%. Hence
bimodal distributions, the regularization methods have the presence of small particles will cause less disturbance
the capability (provided that there are enough turbidity than the experimental error even though there are
measurements) to uncover the true shape whereas the 10 times more small than large particles. In such a case
log-normal assumption will always produce an envelope it is not expected that the regularization methods will
around the main body, as shown in Figure 7. Distribution I produce the correct solution within experimental error.

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
26 PARTICLE SIZE ANALYSIS

Wallach and Heller.7/ suggested using combinations latexes by Kiparissides et al..9,10/ and Zollars.11/ but no
of two distributions with the same form to describe quantitative results from on-line particle size determi-
distributions that are negatively skewed, or bimodals. nation were reported by these workers. Haseler and
A combination of two-parameter distributions (say two Parkin.6/ used on-line turbidity measurements to moni-
log-normal distributions) requires the estimation of five tor the growth of silver bromide in a ripening process.
parameters (D1g , D2g , s1 , s2 , N2 /N1 ). At least five specific The ripening suspension was circulated continuously in
turbidity measurements at wavelengths far from each an undiluted state through a thin cell.
other are required for this estimation if the particles In all of the applications mentioned above only
are in the submicrometer range. These parameters are turbidity was measured on-line. An experimental set-
expected to be very highly correlated and therefore the up to combine on-line determination of the particle
solution to be sensitive to experimental error. A good volume fraction with on-line turbidity measurements,
solution may be obtained in cases where the contribution required for on-line specific turbidity measurements, was
of the distribution of the smaller particles to the overall first suggested by Pollock..86/ However, no results were
specific turbidity is large; otherwise the large particles will reported on particle sizing, probably because he faced
always obscure the small particles. some problems with that set-up. A simulation for on-line
Readers should keep in mind that in general, bimodal latex PSD determination in continuous latex reactors was
distributions (except those where the distance between presented by Bradolin and Garcia-Rubio..22/
the modal diameters of the two populations is large Raphael and Rohani.26/ described a device that could
compared with the breadth of the two populations) are be used on-line to measure the mean particle size
very difficult to resolve with most particle sizing meth- in continuous precipitators or crystallizers. The solids
ods. Using EM, for example, a large number of particles concentration for this set-up is assumed to be known
must be counted, sometimes at different magnifications and no measurements on the particle volume fraction are
in order to resolve the distributions successfully. DLS will taken on-line. This device was not tested on-line, however,
usually result in distributions weighted towards larger for continuous production; the measurements reported
particles. For bimodals where the distance of the two were from spot samples. These samples were diluted in a
modal diameters is large compared with the breadth of vessel from which the diluted sample is pumped through
the distributions, one can usually determine very success- the device and turbidity measurements are taken.
fully the particle sizes involved in the two populations (by A serious disadvantage of some of the set-ups suggested
EM and sometimes by DLS), but it is difficult to estimate for on-line latex particle size determination.9,10,11,22,86/
the relative number of particles in the two populations. In is that they require a continuous flow of diluted latex
such cases, if the diameters involved are known very accu- through the spectrophotometer. For specific turbidity
rately from EM (D1 , D2 , . . . , Dn ), then using a sufficient measurements this means that a known amount of latex
number of turbidity measurements one could estimate the is continuously mixed with a known amount of diluent.
relative number of particles (f1 , f2 , . . . , fn ) (from Equa- Both the latex and diluent have to be pumped at constant
tions 31 and 39). Again, in order to obtain reasonable flow rates. This in turn requires that a very small amount
estimates, the small particles should weigh significantly of latex be pumped at the exit of the reactor, at flow
and the distribution must be in an (m, a) regime where rates as low as 10 mL min 1 , for long periods of time.
specific turbidity measurements do not simply result in A peristaltic pump tried by Gossen.21/ for this set-up
the same apparent diameters (otherwise all combinations failed owing to coagulation of the latex due to high
resulting in the same average are possible solutions). shear and chemical attack of the tube by the monomer.
Other types of pumps tried in the same laboratory also
failed..23/ Another problem with continuous dilution
of the latex is that an enormous amount of waste
8 ON-LINE PARTICLE SIZE (very dilute latex) is created that requires treatment
DETERMINATION USING before disposal. This would cause problems in large-scale
TURBIDIMETRY operations. The set-up suggested earlier from the same
laboratory.85/ utilizes discrete sampling, but it has serious
As early as 1973, Hamielec and Wright.84/ suggested disadvantages..23/
using turbidimetry for on-line latex particle size determi- A system suggested by Gossen.23,24/ measures both
nation. The configuration of a sampling device for on-line solids fraction and turbidity on-line, utilizes discrete
measurements of the PSD of poly(vinyl acetate) latexes sampling, is mechanically simple, has a short cycle time
using turbidimetry was suggested from that laboratory and high accuracy and it is set up such that thorough
in 1976..85/ On-line turbidity measurements in fully auto- cleaning of the sampling element is performed after each
mated systems have been reported for poly(vinyl acetate) sample. This sampling device was used for the on-line

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 27

application of specific turbidity to monitor particle growth the shape is smooth and the log-normal assumption is
during poly(vinyl acetate) production in continuous pilot- sufficient, then this assumption can be used for subsequent
plant reactors. Samples were taken at discrete intervals routine analysis. DLS could also be used to corroborate
of 5 min. Results were reported for a period of 10 h. the results for submicrometer particles. Alternatively,
The weight-average diameters estimated by turbidimetry sedimentation field fractionation.87 – 89/ could be used;
showed excellent agreement with results from on-line this method would be especially useful in cases where
DLS..73/ These results, obtained by Gossen, represent EM cannot be used either because the particles change
the first published successful effort.24,46,73/ to combine size when they are measured dry or they need special
on-line turbidity and density measurements and to treatment for EM measurements.
monitor, on-line, weight-average diameters during latex Sometimes only routine-type checks are needed and
production in pilot-scale reactors, in a fully automated an average of the PSD or the log-normal assumption
manner. Details on the set-up and an error propagation that will provide the main body of the distribution
analysis study can be found in Gossen..23,24/ may be sufficient. Turbidimetry could be used as a fast,
inexpensive technique for such applications.
In some cases only qualitative information is required,
9 OTHER CONSIDERATIONS for example when studying qualitatively the effect of
certain process parameters on the particle size. In such
cases almost any technique can be utilized provided that
9.1 Choice of the Methods
the same quantity is followed and that this quantity
It is a well-established perception in particle size analysis is sensitive to small changes in the particle size. For
that there is no method which can be successfully example, specific turbidity can be used to follow a
utilized for every application. The effectiveness of each weight-average diameter or DLS to follow the intensity-
particle sizing method depends on the special needs average diameter. It is important that the same method
applicable to each case. The selection of the method is a be used to follow the same average throughout the
compromise between accuracy, cost, speed of analysis and study because, as discussed in the next section, different
robustness of the instrument in the plant or the laboratory methods are theoretically directly related (and therefore
environment. This section gives some examples that might are more sensitive) to different averages of the PSD.
help users decide if turbidimetry is suitable for their Other applications where the dependence of turbidity
applications. on particle size can be qualitatively exploited is to
There are cases where an accurate estimate of the use turbidimetry as a detection system (as discussed
full PSD is required. In such cases the time required in section 5.4) to detect the onset of agglomeration or
for analysis and the cost of the method are not usually reactor oscillations in latex production.
an issue. Examples of such cases are the evaluation of When the product is suspected to contain two pop-
a new process and the PSD of the new product, the ulations of particles with different sizes (e.g. large and
study of the effects of various parameters on particle small particles due to a secondary nucleation), the small
size in a set of designed experiments and the detailed particle population will be clearly detected by EM or
study of a nucleation process where the PSD reflects the chromatographic (i.e. size separation) techniques. Size
mechanisms that take place. For such cases more time separation techniques are very powerful in detecting
and effort are usually spent than for routine particle size small-particle populations. If the contribution of the
analysis and more than one particle sizing methods and/or small particles to the overall turbidity is significant
alternative approaches to the same method are used. EM then the bimodal distribution can be resolved by tur-
will provide measurements of the sizes involved in the bidimetry using regularization algorithms (i.e. solving
distribution, and will give an indication of the type of the for PSD without assuming a fixed shape). Otherwise,
distribution (i.e. detect the presence of two populations, specific turbidity will only provide the weight-average
etc.). However, resolving the full f .D/ with EM is a diameter for a bimodal suspension of submicrometer
tedious task because a very large number of particles need particles.
to be photographed and counted from representative When choosing a method for on-line measurements,
parts of the sample. Specific turbidity can be used to one must consider several factors: the type of process
obtain a correct estimate of the weight-average diameter (batch or continuous), the purpose of the measurement
of submicrometer particles if the (m, a) regimes do not (is it routine analysis where we are satisfied with an
allow for the full PSD determination. If the (m, a) values average size, or do we need accurate PSD determination),
allow for the full PSD determination, then regularization the required sampling frequency and analysis time of
methods, where no fixed shape of the distribution is the method, the cost relative to the expected accuracy
assumed, should be tried first. If these methods show that and the reliability and robustness of the method in the

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
28 PARTICLE SIZE ANALYSIS

plant environment. Turbidimetric methods utilize the measures an ‘‘average diameter’’ that corresponds to
fastest measurements (light is used as a probe). As a the mean diffusion coefficient of the suspension. With
comparison, fractionation (separation) techniques cannot the assumption that the distribution of the diffusion
provide any estimate of the particle size or the PSD until coefficients is Gaussian, a PSD and its corresponding
the fractionation is completed. weight-average diameter can be calculated. However,
It is clear that users must consider several factors before although the estimated mean diffusion coefficient is cor-
choosing the technique or a combination of techniques rect, this weight-average diameter is not always equal
that are most suitable for their needs. This requires a to the true weight average of the suspension. In con-
good knowledge of the process, the product and the trast, the weight average measured by turbidimetry is
capabilities and limitations of the technique. Finally, numerically close to the true weight average for most
quoting Groves,.90/ ‘‘the validity of the data need have no submicrometer suspensions. In another example, some
direct relation to the cost of instrumentation’’. DLS instruments utilize the Rayleigh – Gans approxima-
tion to convert distributions from an intensity basis to
9.2 Corroboration of Results a weight basis. This approximation is not valid for very
large particles and may cause serious errors in the calcu-
When another particle sizing method is used to corrobo- lation of the PSD on a weight basis for broad distributions
rate results from turbidimetry, this second method must or widely spaced bimodal distributions, even if the distri-
be chosen carefully. In general, when comparing results bution estimated on an intensity basis is correct.
on the full PSD from different methods, the size range
to which each of the methods is applicable, their res-
olution, the effects of errors on the estimated size and 10 CONCLUSION
any assumptions introduced in each technique should be
considered. This holds even when comparing results from Turbidimetric particle sizing methods are experimentally
two different turbidimetric techniques (i.e. turbidity ratio simple, robust and have very good reproducibility and
and specific turbidity). Erroneous conclusions may be potential for on-line applications. They have been applied
obtained, as illustrated with the following examples. for the estimation of the PSD or an average size to a
The size range to which each of methods is applica- variety of particle suspensions (Table 1) with diameters
ble should be considered. Wales.58/ used particle size ranging from 0.035 to 50 µm.
measurements from a Coulter Counter instrument to cor- Specific turbidity can be used for reliable estimates of
roborate results from turbidimetry. Some discrepancies the average particle size and of the PSD of a suspension
were observed between the two methods. From a detailed for a wider range of particle sizes than the turbidity ratio
discussion.15/ of his work it became clear that in particle or the wavelength exponent techniques. This is because
size regions where both Coulter Counter and turbidime- specific turbidity involves two independent measurements
try are capable of providing a valid estimate of the PSD of the particle size via two different properties of
the two methods compared favorably. The discrepancies the suspension (turbidity and solids volume fraction),
observed for suspensions with PSDs in the lower sub- whereas the other turbidimetric techniques are based on
micrometer range were not due to forward scattering as turbidity measurements only. Therefore, if the volume
originally claimed..58/ Rather, they were due to the fact fraction of particles in the suspension can be measured,
that the Coulter Counter instrument could not detect specific turbidity should be the turbidimetric technique
very small particles (in the lower submicrometer region) of choice. The turbidity ratio or the wavelength exponent
and specific turbidity for the (m, a) values of those sus- method should be used only if the particle volume fraction
pensions could only provide a correct Dw and not the full is unknown.
PSD. Discrepancies between turbidimetry and Coulter Because of the light scattering behavior, turbidimetric
Counter results were also observed by Walstra..16/ The methods are not theoretically expected to provide infor-
Coulter Counter instrument should not be used to cor- mation on the full PSD in many situations. In order to
roborate results from other techniques when dealing with decide the type of information about the particle size
distributions of small particle sizes (i.e. lower submicrom- (average size or full PSD) that can be extracted from any
eter particles) or PSD with tails of small particles. turbidimetric method, one must consider both the value of
Any assumptions involved in the methods should be the refractive index ratio (m) of the suspension and the a
considered. Some average diameters obtained by various values covered by the distribution. The same PSD that can
techniques can be directly derived from the underly- be successfully estimated for one system may be impossi-
ing theory whereas others depend on the validity of ble to estimate for another system with a different m value.
certain assumptions. Only quantities estimated directly There are a few critical combinations of (m, a) values
from theory should be compared. For example, DLS in turbidimetric analysis for nonabsorbing particles:

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 29

ž suspensions of submicron particles and values of m Size Distributions ž Ultrasonic Measurements in Particle
smaller than 1.15; Size Analysis ž Velocimetry in Particle Size Analysis
ž suspensions with any value of m and very small
particles (a < 0.5);
ž suspensions of very large particles (a > 150). REFERENCES
In these regimes, the turbidity ratio and the wavelength
1. E. Atherton, R.H. Peters, ‘Some Aspects of Light Scat-
exponent method should be avoided and only specific tering from Polydisperse Systems of Spherical Particles’,
turbidity should be used. If the particle volume fraction is Br. J. Appl. Phys., 4, 344 – 349, 366 – 369 (1953).
not known, then for suspensions in the above regimes it is 2. W. Heller, W.J. Pangonis, ‘Theoretical Investigations on
better to choose a sizing method other than turbidimetry. the Light Scattering of Colloidal Spheres. I. The Specific
Furthermore, reliable information on the full PSD Turbidity’, J. Chem. Phys., 26, 498 – 506 (1957).
may be very difficult to obtain in the above regimes, 3. W. Heller, ‘Theoretical Investigations on the Light Scat-
by any turbidimetric technique, within experimental tering of Colloidal Spheres. III. Analytical Expressions
error. Specific turbidity, however, will give a very good for the Turbidity Approximating the Performance of the
estimate of meaningful averages of the PSD: the weight- Mie Equations Prior to the First Maximum’, J. Chem.
average diameter in the first case (from the mean value Phys., 26, 1258 – 1264 (1957).
of the apparent diameters at 400 and 600 nm), the 4. W.H. Beattie, ‘Determination of Particle Size in Silver
turbidity-average diameter for suspensions of very small Bromide Sols by Light Scattering’, PhD Thesis, University
particles and the surface – volume-average diameter for of Minnesota, 1958.
suspensions of very large particles. 5. E.J. Meehan, W.H. Beattie, ‘Determination of Particle
Size in Silver Bromide Sols by Light Scattering’, J. Phys.
Chem., 64, 1006 – 1016 (1960).
6. S.C. Haseler, E.M.L. Parkin, ‘Silver Halide Particle Size
ABBREVIATIONS AND ACRONYMS Measurement by Infrared Thin Cell Turbidimetry’, in
Particle Size Analysis 1981, eds. N.G. Stanley Wood,
DLS Dynamic Light Scattering T. Allen, Wiley Heyden, Chichester, 403 – 413, 1982.
EM Electron Microscopy 7. M.L. Wallach, W. Heller, ‘Experimental Investigations
HDC Hydrodynamic Chromatography on the Light Scattering of Colloidal Spheres. VI. Determi-
IR Infrared nation of Size Distribution Curves by Means of Turbidity
NIR Near-Infrared Spectra’, J. Phys. Chem., 68, 924 – 930 (1964).
PLS Partial Least Squares 8. Es. Gulari, G. Bazzi, Er. Gulari, A. Annapragada, ‘Latex
PSD Particle Size Distribution Particle Size Distributions from Multiwavelength Turbid-
ity Spectra’, Part. Charact., 4, 96 – 100 (1987).
SEM Scanning Electron Microscopy
9. C. Kiparissides, ‘Modeling and Experimental Studies
TEM Transmission Electron Microscopy
of a Continuous Emulsion Polymerization Reactor’,
UV/VIS Ultraviolet/Visible
PhD Thesis, McMaster University, Hamilton, Ontario,
ZOLD Zeroth-order Logarithmic Distribution 1978.
10. C. Kiparissides, J.F. MacGregor, S. Singh, A.E. Hamielec,
‘Continuous Emulsion Polymerization of Vinyl Acetate,
RELATED ARTICLES Part III. Detection of Reactor Performance by Turbidity
Spectra and Liquid Exclusion Chromatography’, CJChE,
58, 65 – 71 (1980).
Food (Volume 5)
11. R.L. Zollars, ‘Turbidimetric Method for On-line Deter-
Particle Size Analysis in Food
mination of Latex Particle Number and Particle Size
Distribution’, J. Colloid Interface Sci., 74, 163 – 172 (1980).
Particle Size Analysis (Volume 6) 12. R.L. Zollars, ‘Turbidimetric Method for Determination
Particle Size Analysis: Introduction ž Centrifugation in of Latex Particle Size Distribution’, J. Dispersion Sci.
Particle Size Analysis ž Diffraction in Particle Size Anal- Technol., 2(2 – 3), 331 – 344 (1981).
ysis ž Electrozone Sensing in Particle Size Analysis ž 13. T. Kourti, A. Penlidis, A.E. Hamielec, J.F. MacGregor,
Field-flow Fractionation in Particle Size Analysis ž Filtra- ‘Measurement of Particle Size Distribution of Latex
tion in Particle Size Analysis ž Light Scattering, Classical: Particles in the Submicron Range Using Size Exclusion
Size and Size Distribution Characterization ž Optical Chromatography and Turbidity Spectra’, in Am. Chem.
Particle Counting ž Photon Correlation Spectroscopy in Soc. PMSE Div. Prepr., 53, 147 (1985).
Particle Sizing ž Sedimentation in Particle Size Analysis ž 14. T. Kourti, A. Penlidis, J.F. MacGregor, A.E. Hamielec,
Sieving in Particle Size Analysis ž Surface Area and Pore ‘Measuring Particle Size Distribution of Latex Particles

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
30 PARTICLE SIZE ANALYSIS

in the Submicrometer Range Using Size Exclusion 31. L.D. Maxim, A. Klein, M.E. Meyer, C.H. Kuist, ‘Particle
Chromatography and Turbidity Spectra’, Am. Chem. Soc. Size Determination by Turbidimetry’, J. Polym. Sci.
Symp. Ser., 332, 242 – 255 (1987). Part C, 27, 195 – 205 (1969).
15. T. Kourti, ‘Polymer Latexes: Production by Homoge- 32. T. Kourti, J.F. MacGregor, A.E. Hamielec, ‘Turbidime-
neous Nucleation and Methods for Particle Size Deter- tric Techniques: Capability to Provide the Full Particle
mination’, PhD Thesis, McMaster University, Hamilton, Size Distribution’, Am. Chem. Soc. Symp. Ser., 472, 2 – 19
Ontario, 1989. (1991).
16. P. Walstra, ‘Estimating Globule-size Distribution of Oil 33. M. Kerker, The Scattering of Light and Other Electromag-
in Water Emulsions by Spectroturbidimetry’, J. Colloid netic Radiation, Academic Press, New York, 1969.
Interface Sci., 27, 493 – 500 (1968). 34. H.C. Van de Hulst, Light Scattering by Small Particles,
17. A.T. Shaw, D.F. Darling, ‘Fat Droplet Size Analysis: John Wiley & Sons, New York, 1957.
the Application of Spectroturbidimetry’, in Particle Size 35. W.J. Wiscobe, ‘Mie Scattering Calculations. Advances
Analysis 1981, eds. N.G. Stanley Wood, T. Allen, Wiley in Technique and Fast Vector Speed Computer Codes’,
Heyden, Chichester, 355, 1982. National Center for Atmospheric Research, Boulder, CO,
18. W. Heller, R.M. Tabibian, ‘Experimental Investigation of NCAR Technical Note Tn-140 str, 1979.
the Light Scattering of Colloidal Spheres. II. Sources of 36. W. Wiscobe, ‘Improved Mie Scattering Algorithms’,
Error in Turbidity Measurements’, J. Colloid Sci., 12, Appl. Opt., 1505 – 1509 (1980).
25 – 39 (1957). 37. P. Walstra, ‘Approximation Formulae for the Light
19. S.C. Haseler, ‘Particle Size Distribution Measurement Scattering Coefficient of Dielectric Spheres’, Br. J. Appl.
by Automated Dual Wavelength Turbidimetry’, Part. Phys., 15, 1545 – 1552 (1964).
Charact., 2, 14 – 19 (1985). 38. ‘Conference on Light Scattering by Nonspherical Parti-
20. K.C. Yang, R. Hogg, ‘Estimation of the Particle Size cles: Theory, Measurements and Applications’, Goddard
Distributions from Turbidimetric Measurements’, Anal. Institute for Space Studies, New York, 29 September – 1
Chem., 51, 758 (1979). October 1998.
21. P.D. Gossen, ‘On-line Particle Size Measurement and 39. M.I. Mishchenko, ‘Light Scattering by Randomly Ori-
Mathematical Modeling of Continuous Poly(vinyl ace- ented Axially Symmetric Particles’, J. Opt. Soc. Am., A:
tate) Emulsion Polymerization’, Master Thesis, McMaster Opt. Image Sci., 8(6), 871 – 882 (1991).
University, Hamilton, Ontario, 1988. 40. N.G. Khlebtsov, S. Yu Shchegolev, ‘Allowance for Non-
22. A. Bradolin, L.H. Garcia-Rubio, ‘On-line Particle Size spherical Particles in Determining the Parameters of
Distribution Measurements for Latex Reactors’, Am. Dispersed Systems by Turbidity Spectrum Method.
Chem. Soc. Symp. Ser., 472, 64 – 85 (1991). II. Characteristic Scattering Functions for Systems of
23. P.D. Gossen, ‘On-line Instrumentation for Latex Pro- Nonspherical Particles with Random Orientation in the
duction’, PhD Thesis, McMaster University, Hamilton, Van de Hulst Approximation’, Opt. Spektrosk., 42(6),
Ontario, 1991. 1152 – 1157 (1977).
24. P.D. Gossen, J.F. MacGregor, ‘On-line Particle Diameter 41. M.I. Mishchenko, L.D. Travis, D.W. Mackowski, J.
for Poly-vinyl Acetate Latex Using Specific Turbid- Quant. Spectrosc. Radiat. Transfer, 55, 535 – 575 (1996).
ity Method’, J. Colloid Interface Sci., 160, 24 – 38 42. W. Heller, ‘Theoretical Investigations on the Light Scat-
(1993). tering of Spheres. XVI. Range of Practical Validity of the
25. V.-M. Taavitsainen, J.-P. Jalava, ‘Soft and Harder Mul- Rayleigh Theory’, J. Chem. Phys., 42, 1609 – 1615 (1965).
tivariate Modelling in Developing the Properties of 43. W. Heller, H.L. Bhatnagar, M. Nakagaki, ‘Theoretical
Titanium Dioxide Pigments’, Chemom. Intell. Lab. Syst., Investigations on the Light Scattering of Spheres. XIII’,
29, 307 – 319 (1995). J. Chem. Phys., 36, 1163 – 1170 (1962).
26. N. Raphael, S. Rohani, ‘On-line Estimation of the Solids 44. T. Kourti, J.F. MacGregor, ‘Particle Size Determination
Concentrations and Mean Particle Size Using a Tur- Using Turbidimetry: Capabilities, Limitations and Eval-
bidimetry Method’, Powder Technol., 89, 157 – 163 (1996). uation for On-line Applications’, Am. Chem. Soc. Symp.
27. M.A. Llosent, L.M. Gugliotta, G.R. Meira, ‘Particle Size Ser., 472, 34 – 63 (1991).
Distribution of SBR and NBR Latexes by UV/VIS 45. P.D. Gossen, T. Kourti, A. Penlidis, J.F. MacGregor,
Turbidimetry Near the Rayleigh Region’, Rubber Chem. A.E. Hamielec, ‘On-line Particle Size and Conversion
Technol., 69(4), 696 – 712 (1996). Determination in Continuous Latex Reactor Trains’, in
28. S.H. Maron, P.E. Pierce, I.N. Ulevitch, ‘Determination of Preprints, 6th International IFAC/IFIP/IMEKO Confer-
Latex Particle Size by Light Scattering’, J. Colloid Sci., ence on Instrumentation and Automation in the Paper,
18, 470 – 482 (1963). Rubber, Plastics and Polymerization Industries, Akron,
29. R.J. Gledhill, ‘Particle Size Distribution Determination OH, 1986.
by Turbidimetry’, J. Phys. Chem., 66, 458 – 463 (1962). 46. P.D. Gossen, Presented at the Symposium on Particle Size
30. J.D.S. Goulden, ‘Spectroturbidimetry of Emulsions’, Br. Analysis in Polymer Science, Organized by the PMSE
J. Appl. Phys., 12, 456 – 460 (1961). Division, 191st ACS Meeting, Boston, MA, April 1990.

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
TURBIDIMETRY IN PARTICLE SIZE ANALYSIS 31
47. M.L. Wallach, W. Heller, A.F. Stevenson, ‘Theoretical 64. M. Bertero, C. De Mol, E.R. Pike, ‘Particle Size Dis-
Investigations on the Light Scattering of Colloidal tributions from Spectral Turbidity: a Singular System
Spheres. XII. The Determination of Size Distribution Analysis’, Inverse Problems, 2, 247 – 258 (1986).
Curves from Turbidity Spectra’, J. Chem. Phys., 34, 65. L.C. Lawson, R.J. Hanson, Solving Least Squares Prob-
1796 – 1802 (1961). lems, SIAM – Society for Industrial and Applied Mathe-
48. D.H. Melik, H.S. Fogler, ‘Turbidimetric Determination matics, Philadelphia, PA, 1995.
of Particle Size Distributions of Colloidal Systems’, J. 66. P. Bagchi, R.D. Vold, ‘A Simple Method for Determina-
Colloid Interface Sci., 92, 161 – 180 (1983). tion of the Average Particle Size of Coarse Suspensions
49. J.B. Bateman, E.J. Weneck, D.C. Eshler, ‘Determination from Measurements of Apparent Specific Turbidity’, J.
of Particle Size and Concentration from Spectrophoto- Colloid Interface Sci., 53, 194 (1975).
metric Transmission’, J. Colloid Sci., 14, 308 – 329 (1959). 67. S.H. Maron, C. Moore, A.S. Powell, ‘Electron Micros-
50. T. Kourti, J.F. MacGregor, A.E. Hamielec, ‘Minimum copy of Synthetic Latices’, J. Appl. Phys., 23, 900 – 905
Number of Turbidity Measurements Required for the (1952).
Determination of Particle Size Distributions’, J. Colloid 68. K. Kubota, S. Hayashi, H. Mandai, ‘A Few Comments on
Interface Sci., 120, 292 – 295 (1987). Simple Optical Measurements for Determining Sizes of
51. W.H. Press, S.A. Teukolsky, W.T. Vettering, B.P. Flan- Latex Particles Suspended in Water’, J. Colloid. Interface
ney, Numerical Recipes in C. The Art of Scientific Com- Sci., 98, 581 (1984).
puting, Cambridge University Press, Cambridge, 1992. 69. R.A. Dobbins, ‘Measurement of a Mean Particle Size in
52. D.W. Marquardt, ‘An Algorithm for Least Squares a Gas Particle Flow’, AIAA J., 1, 1940 – 1942 (1963).
Estimation of Non-linear Problems’, J. Soc. Ind. Appl. 70. R.A. Dobbins, G.S. Jizmagian, ‘Optical Scattering Cross
Math., 11(2), 431 – 441 (1963). Sections for Polydispersions of Dielectric Spheres’, J. Opt.
53. A.F. Stevenson, W. Heller, M.L. Wallach, ‘Theoretical Soc. Am., 56, 1345 – 1350 (1966).
Investigation on the Light Scattering of Colloidal Spheres. 71. L.H. Garcia-Rubio, ‘Averages from Turbidity Measure-
XI. Determination of Size Distribution Curves from ments’, Am. Chem. Soc. Symp. Ser., 332, 161 – 178
Spectra of the Scattering Ratio or from Depolarization (1987).
Spectra’, J. Chem. Phys., 34, 1789 – 1795 (1961). 72. G.C.N. Cheesman, ‘A Rapid Method for the Measure-
54. R.R. Irani, C.F. Callis, Particle Size: Measurement, Inter- ment of the Average Particle Size of Non-uniform
pretation and Application, John Wiley & Sons, New York, Latices and Dispersions’, in Particle Size Analysis, ed.
1963. M.J. Groves, Heyden, London, 160 – 166, 1978.
55. E.P. Honig, ‘Logarithmic Distribution Functions for Col- 73. D.F. Nicoli, T. Kourti, P.D. Gossen, J.-S. Wu, J.F. Mac-
loidal Particles’, J. Phys. Chem., 69, 4418 – 4419 (1965). Gregor, Y.-J. Chang, ‘On-line Latex Particle Size Deter-
56. J.F. Yan, ‘Mass Log-normal Distribution Function for mination by Dynamic Light Scattering: Design for an
Particle Size Analysis’, J. Colloid Interface Sci., 49, Industrial Environment’, Am. Chem. Soc. Symp. Ser.,
152 – 153 (1974). 472, 86 – 97 (1991).
57. W.F. Espenscheid, M. Kerker, E. Matijevic, ‘Logarithmic 74. G. Crawley, M. Cournil, D. Benedetto, ‘Size Analysis
Distribution Functions for Colloidal Particles’, J. Phys. of Fine Particle Suspensions by Spectral Turbidimetry:
Chem., 68, 3093 – 3096 (1964). Potential and Limits’, Powder Technol., 91, 197 – 208
58. M. Wales, ‘Particle Size Distribution in Rubber Latex’, J. (1997).
Phys. Chem., 66, 1768 – 1772 (1962). 75. C.F. Bohren, D.R. Huffman, Absorption and Scattering of
59. G. Herdan, Small Particle Statistics, 2nd edition, Butter- Light by Small Particles, John Wiley & Sons, New York,
worths, London, 1960. 1983.
60. Er. Gulari, A. Annapragada, Es. Gulari, B. Jawad, ‘De- 76. P. Walstra, ‘Discussion of Errors in Turbidimetry’, Br. J.
termination of Particle Size Distribution using Light Appl. Phys., 16, 1187 – 1192 (1965).
Scattering Techniques’, Am. Chem. Soc. Symp. Ser., 332, 77. R.O. Cumprecht, C.M. Sliepcevich, ‘Scattering of Light
133 – 145 (1987). by Large Spherical Particles’, J. Phys. Chem., 57, 90 – 95
61. G.E. Eliçabe, L.H. Garcia-Rubio, ‘Latex Particle Size (1953).
Distribution from Turbidimetry Using a Combination 78. R.M. Tabibian, W. Heller, J.N. Eppel, ‘Light Scattering
of Regularization Techniques and Generalized Cross- of Colloidal Spheres. (I) Specific Turbidity’, J. Colloid
validation’, Adv. Chem. Ser., 227, 83 – 104 (1990). Sci., 11, 195 – 213 (1956).
62. G.E. Eliçabe, L.H. Garcia-Rubio, ‘Latex Particle Size 79. J. Brandrup, E.H. Immergut, Polymer Handbook, 2nd
Distribution from Turbidimetry Using Inversion Tech- edition, John Wiley & Sons, New York, 1975.
niques’, J. Colloid Interface Sci., 129(1), 192 – 200 80. J. Morgan, Introduction to Geometric and Physical
(1989). Optics’, McGraw-Hill, New York, 1953.
63. J. He, S.-M. Wang, J.-C. Cheng, S.-Y. Zhang, ‘Inversion 81. U. Comberg, T. Wriedt, ‘Scattering by Inhomogeneous
of Particle Size Distribution from Light Scattering Particles, Exact Theories and Effective Medium Theo-
Spectrum’, Inverse Problems, 12, 633 – 639 (1996). ries’, in Electromagnetic and Light Scattering – Theory and

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517
32 PARTICLE SIZE ANALYSIS

Applications III, eds. T. Wriedt, Yu. Eremin, University Acetate’, PhD Thesis, McMaster University, Hamilton,
of Bremen, Bremen, 1998. Ontario, 1984.
82. V.I. Ivakhnenko, J.C. Stover, Yu.A. Eremin, ‘Model- 87. J.C. Giddings, G. Karaiskakis, K.D. Caldwell, M.N. Mye-
ing of Light Scattering by Nonspherical In- rs, ‘Colloid Characterization by Sedimentation Field-flow
homogeneous Particles’, in Electromagnetic and Fractionation. I. Monodisperse Populations’, J. Colloid
Light Scattering – Theory and Applications III, eds. Interface Sci., 92, 66 – 80 (1983).
T. Wriedt, Yu. Eremin, University of Bremen, Bremen, 88. F.-S. Yang, K.D. Caldwell, J.C. Giddings, ‘Colloid Char-
1998. acterization by Sedimentation Field-flow Fractionation.
83. J. Sinzig, M. Quinten, ‘Scattering and Absorption by II. Particle Size Distribution’, J. Colloid Interface Sci., 92,
Spherical Multilayer Particles’, Appl. Phys. A: Solids 81 – 91 (1983).
Surf., A58, 157 – 162 (1994). 89. F.-S. Yang, K.D. Caldwell, M.N. Myers, J.C. Giddings,
84. A.E. Hamielec, J.D. Wright, Chem. Can., December, ‘Colloid Characterization by Sedimentation Field-flow
27 – 29 (1973). Fractionation. III. Emulsions,’ J. Colloid Interface Sci.,
85. S.W. Ray, ‘Minicomputer Control of a Latex Reactor’, M. 93, 115 – 125 (1983).
Eng. Thesis, McMaster University, Hamilton, Ontario, 90. M.J. Groves, ‘Particle Size Characterization in Disper-
1976. sions’, J. Dispersion Sci. Technol., 1, 97 – 124 (1980).
86. M.J. Pollock, ‘Modeling and Control of Sustained Oscilla-
tions in the Continuous Emulsion Polymerization of Vinyl

Encyclopedia of Analytical Chemistry, Online © 2006 John Wiley & Sons, Ltd.
This article is © 2006 John Wiley & Sons, Ltd.
This article was published in the Encyclopedia of Analytical Chemistry in 2006 by John Wiley & Sons, Ltd.
DOI: 10.1002/9780470027318.a1517

You might also like