Professional Documents
Culture Documents
STS
SERVICE DATA BULLETIN
LAS Technical Support Specialist Group
710 Bridgeport Avenue, MS255
Shelton, CT 06484-4794
PURPOSE:
MATERIAL REQUIRED:
GENERAL:
1
Optical Alignment AAnalyst 600/700/800
Index
1 REQUIRED TOOLS 5
2
3 SAMPLE BEAM ADJUSTMENT 25
4 TURRET CALIBRATION 33
6 DETECTOR ADJUSTMENT 38
6.1 D2 ADJUSTMENT 40
7 WAVELENGTH CALIBRATION 41
3
7.1 CONDITIONS FOR THE WAVELENGTH CALIBRATION 41
8 FINAL TEST 45
4
1 Required Tools: B312-0322 Optical Alignment kit
Parts of the kit:
5
B312-0340 Standard Lamp Compartment (looks like a lamp, has a target)
(Place the target closer to the front of the instrument, needs to be the same
depth as the cathode in the lamp is. )
B312-0342 Standard Swiveling Mirror Fixture, (this is placed in the holes that
are used for the cover support). The image on this fixture has to be looked at
from the back of the instrument.
6
B085-0115 Grip 1/4"
In the SE4200 base module, the icons from left to right are:
1. Save
2. Print
3. Preview
4. Calibrate Instrument (picture of instrument)
5. Align Lamps (lamp with cable)
6. Slit Parameter (rainbow colors)
7. Lamp Parameter (light bulb)
8. Gain Functions (display with energy bar)
9. Wave Drive (blue waves)
10. Read (stop light)
11. Peak Results (peak shape)
12. Continuous Display (blue display with reading)
13. Stop (stop sign)
14. Optic Alignment (wrench)
15. Direct Command
16. Instrument Info (Letter “I”)
7
2 Before You Start
To complete the optical alignment procedure described below, the atomizer must
be in the uninstall position for the AAnalyst 700/800 (use the ACU base
module to do this), the THGA must be swung out in the AAnalyst 600 (there
are two latch positions) and the instrument must be powered on. In addition, take
out all of the HCL/EDL lamps. The optical alignment must be performed in a dark
area.
The base modules are located in: C:\Program Files\WinLab32-
AA\Diagnostics\AAnalyst 600 or 700 or 800 (choose the instrument you are
using). To close a base module, click in the upper left hand corner of the window.
8
Warning! Risk of 450 Volts DC shock on the HCL lamp connectors
when the instrument is powered on!
. • Remove the four screws shown in the figure above to remove the
detector (use the short Torx driver found in the alignment kit).
.
. • Take off the detector plate with detector board. Be careful; make
sure the quartz window of the detector is not damaged. Place Kim
Wipes or other similar material around it and store in a safe place.
9
• To change the operation mode from single to double beam mode while
the instrument is initialized, go to the TOOLS pull down menu and
open the OPERATION MODE window and change to Flame.
10
• Place target fixture (B015-6118) in front of the slit assembly.
• Move the slit motor with single steps until the 70 mm height line on the
target fixture matches the 70 mm height mark on the slit assembly.
There is a “hatch” line on the slit assembly at the 70 mm position. Look at
the fixture at a low position near the optical bench to get accurate optical
height.
Note: Clicking on the right arrow moves the slit down toward the optical
base.
Take a close look at the figure below! Notice that there are two rectangles for
each slit size (sample & reference). The position of the sample and reference
rectangles will reverse between the exit and entrance slits. The exit and
entrance slits are also not at the same height, but the sample portion for
the entrance & exit slits are centered on the 70 mm line.
11
View from Parabola to Slit Assy.
• Press the Save button to store the value into the FLASH RAM on the
Lamp Controller Board. The number previously saved is printed on the
label that is underneath the instrument cover. (See the last page)
Note: There is no need to adjust the Hall sensor board to set the reference
position of the slit, just write down the number of steps for the
reference position. You will need this number if a new Lamp
Controller Board is installed.
Note: If the above test fails it may be that the slit gear is pushed up against
the slit housing too tightly causing it to bind and miss steps. If this
occurs, loosen the set screws on the stepper motor slit drive and
reposition the motor to reduce the contact pressure and provide easier
movement.
12
• Repeat going back and forth between slit positions to reconfirm
proper operation.
13
• In the Wavedrive window (left side), press Find Ref. Pos.
After pressing Find Ref. Pos., the wavelength drive moves back until the Hall
sensor enters the magnetic field and stops. In that position, the grating arm
must be parallel to the sine bar.
• Place the feeler gauge between the sine bar and grating arm. The
blade should just fit through the gap if aligned correctly and have the
same fit over the entire length of the bar.
14
If the grating arm looks like the diagram below:
Grating Arm
• Loosen the Torx screw that attaches the Hall sensor, and move the
sensor closer to the stepper motor.
• Loosen the Torx screw of the Hall sensor and move the Hall sensor
closer to the back of the spectrometer.
• If the grating arm looks like the figure below, and the gap is 0.5 mm, it
does not need to be adjusted.
15
2.7 Install the Mercury Lamp
Remove the fiber optic from the slit assembly. (It is put back after
focus is finished.)
If the fiber optic is in place, the top section of the slit will appear as if the light is
not going through the slit. (Be careful not to damage the tip of the fiber optic.)
. • Remove the fiber optic from the slit assembly board. Loosely put
Kim Wipes around the fiber optic. Use a T9 Torx.
.
. • Install the lens onto the mercury lamp assembly.
.
. • Mount the mercury lamp on the side of the instrument frame in front
of the exit slit.
.
. • Plug in the power cord of the mercury lamp into the output socket of
the mercury lamp power supply (B001-4672).
.
. • Connect the mercury lamp supply power cord to the power outlet
on the rear of the instrument. Do not use the COOLING SYSTEM
outlet because this is software controlled.
.
. • Align the position of the mercury lamp to produce a bright, uniform
light on the parabola surface. Use a piece of white paper (i.e.
business card) in front of the parabola P1 (it is not necessary to
remove the parabola mask, but it must stay level or the image will
be distorted). Make sure the light is bright, even, and passes
through only one slit, otherwise the light will not be seen in the
sample path. It is very important to get a bright image.
16
sample slit is on the bottom. Center the sample slits to the 70
mm line. See picture on the next page.
.
. • Remove the optical square (mirror).
17
Using the wavelength entry (left side of window), of the Wavedrive window,
type in 40 nm and press the Set button.
5 mm
The height and tilt of the grating must be aligned to ensure that both the white
and green light are at exactly the same height as the entrance slit. Use the
2.0nm Low slit position.
18
2.8.1 Height Adjustment of the Grating
. • Using the wavelength entry field of the Wavedrive window, drive the
wavelength drive to approximately 0 nm. Change the wavelength
drive position by typing in different wavelengths (close to 0 nm) or
by manually moving the wavelength drive (best access point:
damping wheel on the shaft of the stepper motor) so that the light
image is half overlapping the slit.
.
. • Use a 2 mm Allen key. With white light (0 nm) adjust the height of
the grating such that the light image has the same height as the
entrance sample slit. Match the 70 mm line to the middle of the
sample portion.
.
. • If the image is too low, carefully turn the height screw counter-
clockwise.
.
. • If the image is too high, carefully turn the height screw clockwise.
19
2.8.2 Tilt of Grating
. • Find the wavelength reference point by clicking on the Find Ref.
Pos. button in the Wavedrive window.
.
. • Use the wavelength entry field in the Wavedrive window to drive the
wavelength to 546.1nm. Change the wavelength drive position by
typing in different wavelengths (± 546.1nm) or by manually moving
the wavelength drive (best access point: damping wheel on the
shaft of the stepper motor), so that the green light image is half
overlapping the entrance sample slit.
.
. • With green light (546.1nm) adjust the tilt of the grating, so that the
light image has the same height as the entrance slit
.
. • If the image is too low, carefully turn Tilt screw counter-clockwise
.
. • If the image is too high, carefully turn Tilt screw clockwise.
• If a large tilt adjustment was required in the previous step, than the
white light image may need to be readjusted for height.
20
Note: The parabola is not symmetrical to its axis, a sharp rectangular image at
the entrance slit can be obtained only if the hash mark appears on the side
of the furnace compartment.)
To monitor the focus point, it is necessary to have the E4 flat optic mounted at
the entrance slit.
The focus of the Parabola can be adjusted with either white or green light using
the 0.2 nm Low slit! DO NOT USE A 2.0 SLIT, IT MUST BE 0.2 NM!
. • Carefully cover the mercury lamp to prevent stray light from
entering the slit area. Also, make sure the heat from the hot
mercury lamp remains adequately ventilated. Try using aluminum
foil to prevent stray light.
.
. • Using the wavelength entry of the Wavedrive window, drive the
wavelength drive to 546.1 nm (green light).
.
. • Place a white card between E4 and T2. (Or where the baffle was
removed from, to the right of the sample compartment).
.
. • Slightly loosen the two set screws locking the parabola in position.
.
. • By moving the wavelength drive manually clockwise and counter-
clockwise (use the damping wheel), the light image can be moved
over the entrance slit. This creates a light image on the white card
at the baffle.
. • One of the following 4 images will appear on the white card:
21
.
2. Focus maladjusted:
A more or less narrow light image is moving from the left side to the right
side (or vice versa).
22
3. Rotation maladjusted:
A more or less narrow light image is moving from the top to the bottom (or
vice versa).
4. Correctly adjusted:
Over the entire area, the light appears smoothly, reaching a maximum intensity
and then disappearing. The image appears to collapse on itself, but there is no
movement from left to right, up and down, or diagonally.
23
Turn the large back center screw slowly to correct side to side
movement, this moves the mirror in and out. Rotate the mirror
to correct up and down movement, but keep the “hash” mark
close to its 70 mm position.
At this point, make sure the height, side-to-side and zero order adjustments
of P1 are still OK.
24
3 Sample Beam Adjustment
The sample beam of the spectrometer optics will be aligned in the following
steps. White light, with the 2.0 nm L slit at the Ref. Position of the wavedrive
assembly, must pass through the entrance slit.
• Take out the baffle at the right hand side of the burner compartment. (This
was previously done with the monochromator adjustments.)
25
. • Now make the height adjustment (screw next to the top arrow) on
E4 using the cross-hair and 70mm target fixtures.
.
. • Place the 70 mm cross-hair fixture (B015-6119) in the
monochromator between the wavelength drive assembly and the
parabola. Adjust the cross hair fixture position to get a sharp
image of the hair line on the target fixture at T2. Try to use the
same position for the cross-hair fixture for height adjustments of
other optics.
.
. • Adjust the height screw on E4 to superimpose the cross-hair image
onto the lined fixture in front of T2.
.
. • Remove the 70 mm cross hair fixture that is in the monochromator.
.
. • Remove the target fixture from the front of Torroid T2.
26
. • If the focus is not correct (image is not clear), than loosen the
screws holding Torroid T2 (right side of sample compartment) to
the optical bench and move the optic forward and backward.
.
. • If the shape is not square, than the rotation is incorrect and the
rotation screws need to be adjusted.
.
. • After alignment is completed tighten the lock down screws.
.
. • Remove the furnace compartment target.
27
The position of the E2 mirror is fixed by two set pins.
Since there is a fixed 8 position lamp housing used in the Analyst 600/700/800,
the linear mirror E1 (moves from side to side), and the tilting mirror T1 is used to
select the up/down lamp positions. The linear mirror is positioned in front of the
selected HCL/EDL lamp position, and the T1 tilting mirror chooses the upper or
lower lamp.
28
• In the Turret-Adjust section of Lamp Parameter, press Ref. Pos.
The stepper motor for the linear mirror E1 moves to the left side Hall sensor
position.
The stepper motor for the swivel mirror T1 moves to the Hall sensor position. In
this Ref. position the distance between the movable casting and the optic
mount should be 0.5 mm. Use a feeler gauge to check for proper spacing.
29
• Reposition the Hall sensor to change this distance. Press the
Ref.Pos. button after each Hall sensor adjustment.
These adjustments ensure that the linear mirror (E1), and the Torroid mirror
(T1), stop their movement by the Hall sensor before they reach a physical
stop.
30
3.4.3 Side Adjustment of T1 (Tilting Torroid)
• In the Lamp Parameter Window type in the Pos. hor entry field 8000
and press the Set A. Pos button.
This will allow adequate room to place the standard linear mirror (B312-0338)
(flat metal target fixture) on the surface of E1.
. • Loosen the two screws for the Side and the Main Fixing Screw
(shown in the picture on page 33, Fixing Screws for Focus and Side
of T1 Mirror Assy.). Center the light image of the reflected beam on
the mask by means of moving the complete T1 mirror assembly.
.
. • Tighten the Side screws and the Main Fixing Screw.
In the Lamp Parameter Window type in the Pos. hor. entry field 10,000 and the
previously determined steps for the Pos. Vert., and press the Set A. Pos
button. This causes the E1 mirror to move to the casting and touch it and
provides enough room to make the following height adjustment.
. • Place the target fixture (B015-6118) on the pads with the two
holes on top of the optical base in front of the E1 mirror (where the
support bracket was removed from).
.
. • Place the cross-hair fixture in the monochromator and move it until
you see a sharp image of the height line.
.
. • Turn the height screw of the E1 (linear) mirror until the line from the
height jig matches the marked line of the target fixture.
.
. • Remove the cross-hair fixture from the monochromator.
.
. • Remove the target fixture B015-6118.
31
3.4.5 Side Adjustment of E1 (Linear Mirror)
In the Lamp Parameter Window type in the Pos. hor entry field 10,000 and the
previously determined steps for the Pos. Vert. and press the Set A. Pos
button. This causes the E1 mirror to move to the casting and provides enough
room to make the following side-to-side adjustment.
32
image should be visible near the center of the lamp fixture.
.
. • Manually enter step numbers for vertical and horizontal to
center the light image on the lamp fixture.
.
. • Loosen the Main Fixing Screw and the screws for Focus that
hold the mirror to the optical bench.
.
4 Turret Calibration
Since there is no longer a rotating turret, the swivel and linear mirrors must
move to optimize the light throughput for each lamp position. The combination of
step values must be obtained separately for each position using the following
calibration procedure and those values are then stored into the flash prom on the
lamp controller board.
The approximate number of steps for the horizontal (stepper motor for
mirror E1) and vertical (linear stepper motor for mirror T1) are given in the
table below:
33
Lamp 1 2 3 4 5 6 7 8
Pos. hor. 300 300 3200 3200 6040 6040 8900 8900
5. With the Up/Down & Left/Right buttons center the light image onto
the center hole of the lamp fixture.
7. Store the values found for the first lamp position by pressing the
Set L. Pos. button.
9. Remove the lamp fixture from position one and place in position
two. Repeat steps three to eight for all lamp positions.
Press the Save button in the Turret- Adjust field to store all of the values into
the flash prom of the lamp controller board.
34
5.1 Beam Splitter Height and Side Adjustments
35
centered in the entrance window of the D2 lamp assembly.
36
.
. • In the lamp current entry field type in 15 mA for Cu and click on the
Set button.
.
. • Select Beam Splitter In (AAnalyst 800).
Note: The final D2 lamp adjustments will be performed using the detector
and Read window.
37
6 Detector Adjustment
The detector adjustment is performed with the primary wavelength from the Cu
HCL using the Sample Beam.
38
Read AA-BG
• Open the fixing screw for the horizontal adjustment with the
short Torx screwdriver, put in the excentric tool into the point of
horizontal adjustment and carefully move the detector plate to
get the maximum number of counts for the sample beam.
Note: The maximum for the Ref. Beam will now be adjusted with the 0.7
nm H slit position.
. • Open the slit parameter window and select the 0.7 nm H slit
position.
.
. • In the Read Window press the Start button.
The numbers of counts for the Reference and Sample Beam are displayed.
(If there are only Sample counts, make sure Flame is selected in Operation
Mode.)
. • Carefully adjust the side-to-side and height of the flat mirror E3 to
39
get maximum counts for the reference beam (reference counts).
.
. • Block the sample beam.
.
. • Open the fixing screw for the vertical adjustment , put in the
excentric tool into the point of vertical adjustment and carefully
move the detector plate with the excentric tool to zero the counts
for the sample beam (analytical counts).
Not being able to get below 60 counts probably means the focus adjustment of
the parabola is not correct.
6.1 D2 Adjustment
Note: The final D2 adjustment should be performed after the detector
adjustment.
40
7. Wavelength Calibration
. • Close the 4200BM and switch off the AAnalyst 600/700/800
spectrometer.
.
. • Re-install all optics covers.
.
. • Reinstall the burner compartment.
Make sure the light beam is not blocked by the furnace. Remove the furnace
from the light path with the uninstall command of the ACU base module!
41
instrument calibration. ( Zero Order )
.
. • Close the Align Lamp window.
.
. • Click on the wrench button in the tool bar of the SE BM.
• In the left hand side of the Optic Alignment window select the radio
button Wave.
42
cycle 6 Pos. 5 No lamp installed
cycle 7 Pos. 6 No lamp installed
cycle 8 Pos. 7 No lamp installed
cycle 9 Pos. 8 No lamp installed
The more sequences that are run the better the corrected values. It is
recommended to use 27 cycles yielding 3 sequences.
• press the Start button in the Optic Alignment window.
The Lamps are switched on one after the other.
Below is a print out of the STDOUT.LOG file from a successful wavelength
calibration:
------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------
0 0 1
1 19488 19499
2 33216 33238
3 45990 46019
4 peaks, switch = 1
109: instrument error: "Undefined header" at command CNTRL:SENS:REV?
'
* Instrument connected *
Operation Mode = Flame
Optics = double beam
BGCorrection = D2
LampTable = Lamp table Flame
Start Wavelength Drive Test Cycles
43
12 2 553.97 553.97 0.00 1461 12.0 255 0.2H 07.10.1999 08:51:50
13 3 767.03 767.03 0.00 7096 12.0 255 0.7H 07.10.1999 08:52:07
19 0 0.00 0.00 0.00 5989 32.0 0 0.2L 07.10.1999 08:53:17
20 1 324.98 324.98 0.00 3940 20.0 255 0.7H 07.10.1999 08:53:42
21 2 553.97 553.97 0.00 1510 12.0 255 0.2H 07.10.1999 08:54:01
22 3 767.03 767.03 0.00 7109 12.0 255 0.7H 7.10.1999 08:54:20
4: no correction result
5: no correction result
6: no correction result
7: no correction result
8: no correction result
* points found : 4***
Calibration results ***
lamp : num nominal -real- peak delta
steps nm steps nm steps nm
0 : 3 0 0.00 0 0.00 0 0.00
1 : 3 19488 324.80 19499 324.98 11 0.18
2 : 3 33216 553.60 33238 553.97 22 0.37
3 : 3 45990 766.50 46022 767.03 32 0.53
4: no correction result
5: no correction result
44
6: no correction result
7: no correction result
8: no correction result
* points found : 4
store wavelength calibration to nvram
-------------------------------------------------------------------------------------------------------------
-------------------------------------------------------------------------------------------------------------
** Press Set WL-Table in the Optic Alignment window. (left
hand side of the window). The corrected values for
wavelength calibration are stored in the nvram (Zero Power
ram) on the 4200 Controller Board.
8 Final Test
0 0 0
45
1 19488 19495
2 33216 33224
3 45990 45992
4 peaks, switch = 1
* Instrument connected *
• Open the Align Lamp window to setup HCLs for different elements to
make sure, the wavedrive is correctly peaking the line!
46
.
. • Select Y-Axis: Counts; X-Axis: nm.
.
. • Check off DARK and LINE in Legend.
.
. • Apply Start and see Wavelength scan.
.
. • Put cursor over upper left corner of X-Y GRAPH for Ni Triplett
.
. • Depress right mouse button -Hold – and drag over the Ni Triplet
and release button.
.
. • Click right mouse button
A count comparison test is to perform to make sure, that the sample- and the
reference-beam are aligned at their maximum position. If the maximas are not
matching each other, repeat chapter 6, Detector Adjustment on page 32.
47
1. Select 0.7 H slit.
5. Write down the Analytical Counts and Reference counts in: PEAKED with
REF. BEAM. Column.
7. Open the Test pull down menue and disable the Reference beam.
9. Open the Test pull down menue and enable the Reference beam.
11. Write down the Analytical Counts and Reference counts in: PEAKED with
SAMPLE BEAM
0.7H
48
49
Numbers from Optical Alignment
(These need to be recorded & placed on the inside of the top cover)
Date:
Serial Number:
Slit Position (2.0 L):
Date:
Serial Number:
Slit Position (2.0 L):
50