Flow Meter Calibration


by Editorial Staff

You can calibrate flow measurement devices by three methods :

1. Wet calibration using actual fluid flow


2. Measurement check of physical dimensions and use of empirical gables relating flow rate to
these dimensions
3. Dry calibration using flow simulation by electronic or mechanical means.

Flow Meter Calibration

It is common to make measurements using a combination of several instruments acting together. It is


also appropriate to calibrate different portions of such a system by different methods.

In the case of a flowmeter with a transmitter and an indicator, a common practice is to wet calibrate the
flowmeter and separately dry calibrate the transmitter and indicator. You can often further separate the
flowmeter itself into a wetted primary (the portion generating the flow signal) and a flow secondary (the
portion providing the signal processing).

Whenever separately calibrating portions of a system, maintain adequate accuracy rations for each
calibration and calculate the overall system accuracy by the previously described method.

Wet Calibration

A wet calibration, using actual fluid flow, normally provides the highest calibration accuracy for a
flowmeter and sees use where accuracy is a prime concern or when the form of the meter does not lend
itself to other methods. Precision flowmeters usually see wet calibration at the time of manufacture.
After service, the user often removes and recalibrates them. The actual interval between these
calibrations varies depending on the user’s experience with the meter and the importance of the
measurement.

You can perform the recalibration at the flowmeter manufacturers’ facility, in an independent laboratory,
or at the user’s site when adequate facilities are available. When performing the calibration at an outside
facility, specify the flow range you want to calibrate, the number of points you are taking, and the fluid
viscosity (if you are using something other than water at room temperature).

If you are using the flowmeter for billing or other accounting purposes, you might want to obtain an as-
received and after-adjustment calibration. The as-received results enable evaluating what magnitude of
error may have existed in the meter during use, prior to recalibration. If this error is large, evaluate the
impact of the error on prior measurements and determine whether additional action is required.

The procedures outside laboratories should follow in calibrations are normally left to the discretion of the
laboratory. You can specify a system calibration, where you calibrate all elements of the flow measuring
system together, when accuracy is critical.

We usually perform wet calibrations with water, air, or hydrocarbon fuels using one or more of four basic
standards: weight tanks, volumetric tanks, pipe provers, or master flowmeters.

Weight, volume standards

Three methods of calibration see use when a weight or volume tank is the standard. In the static
calibration method, the flow starts quickly to begin the test, holds steady during the test, and then shuts
off at the end of the test. Compare the totalized reading from the flowmeter with the weight or volume
you collect and calculate the meter performance.

A static system operates best with flowmeters that have minimum sensitivity to low flow rates. It does not
give optimum results with high-performance digital output meters, such as vortex flowmeters, because
of errors obtained during the short periods of low flow at the beginning and end of the test. For this
reason, static symptoms are no longer in general use.

A weight or volume standard can also be used in a completely dynamic mode. In this method, the flow is
set at a steady rate before the start of the test. The flowmeter reading and initial weight or volume are
read simultaneously to start the test and again, after the desired collection period, to end the test.
Completely dynamic systems are limited by speed or response consideration and the general difficulties
used.

Hybrid dynamic start-and-stop


Hybrid dynamic start-and-stop, static reading weight and volume systems generally provide more
accurate liquid calibrations than purely static or dynamic systems. In such systems, obtain the desired
test flow rate while the flow is diverted past the weight or volume flow standards.

Initiate the test run by diverting the flow into the standard and complete it by diverting it out of the
system. The flow signal, which should be digital, gates on and off by the diversion. Read the weight or
volume after an appropriate settling time, and compare it with the totalized flowmeter reading.

The key to the dynamic start-and-stop reading system is the design of the flowdiverter valve that
switches the flow in and out of the standard. In a well-designed system, the actual diversion time is much
smaller than the collection time, and the flow pattern through the diverter is independent of flow rate.

Under these conditions, the limiting factor in system uncertainty is the basic accuracy and resolution of
the weight or volume standards. With care, you can reduce errors to less than 0.1% of reading under
laboratory conditions. Few calibration systems of this type see field use because of their size and cost.

Master Flow meters

Perhaps the most common method of wet calibration is comparing the output of the meter under test
with one or more flowmeters that have been certified against high-accuracy standards, known as master
flowmeters. You can construct a small master meter comparison test stand at less cost than other wet
calibration methods.

If properly designed and maintained, it is capable of calibrating flowmeters to a tolerance of ±1%


reading or less. If you are selecting flowmeters as masters, consider the desired range of the test stand,
the required accuracy, the properties of the test fluid, and the allowable head loss.

Author : Thomas Burgess

Articles You May Like :

Pressure Transmitter Calibration


Area Velocity Flow Meter Principle

Read Circular Chart Recorder

Doppler Flow Meter Advantages

Rotary Flow Meters Principle

Don't Miss Our Updates


Be the first to get exclusive content straight to your email.

Enter your name here...

Enter your email here...

We promise not to spam you. You can unsubscribe at any time.

Subscribe Now

Calibration, Flow Measurement


Control Valve Failure Rate Calculation
Bathtub Curve

Recommended Articles
1. Calibration and Troubleshooting of Oxygen Analyzer

2. Pressure Transmitter Calibration at the bench

3. DP Level Transmitter with Zero Suppression Calibration

4. Difference between Repeatability and Reproducibility

5. Instrument Zero and Span Calibration

6. Capacitance Level Switch Calibration

7. DP Transmitter Interface Level Measurement Principle, Limitations, Selection, Installation, Design &
Calibration

8. Installation and Calibration of Pressure Transmitter

9. Temperature Sensor calibration procedure

10. Uncertainty Calculations of Pressure Calibration


1 thought on “Flow Meter Calibration”

Dianne Miller
June 28, 2022 at 10:05 am

Thank you for helping me understand flow calibration. I like that you discuss how it works. There
is an element called a flow measuring system. That is the process of measuring combination
materials for calibration purposes. These are helpful information in securing balancing the
chemicals. These are important to the metrology laboratory and food and chemical industry.

Reply

Leave a Comment

Name *

Email *

Website

Post Comment

Categories

Select Category
Recent Comments

sreekanth on Various Communication Protocols in PLC

Thara bhai joginder on PLC Logic Functions

dinfortune@gmail.com on Basics of Alarms and Trips

Saurabh on Troubleshooting Ultrasonic Flow Meter Common Problems

Amir on Accelerometer Vibration Probe Principle

Reagan Nyero on Hot Well Level Control System in Condensing Turbine

DLER on How to Configure a Smart Transmitter Using a HART Communicator?

More Articles

1. What is Sensor Trim and Output Trim in Smart Transmitter?

2. Pressure Gauge Calibration Procedure

3. RTD Calibration

4. Calibration : Up-tests and Down-tests

5. Uncertainty Calculations of Pressure Calibration

6. Functional Testing of Control Valve Positioners

7. Calibration Procedure of Voltmeter and Ammeter

8. Open Tank DP Level Transmitter Calibration

9. Hydrogen Sulfide Gas Detector Calibration Procedure

10. Calibration of Weight using the ABBA Method

2022 © Reproduction without explicit permission is prohibited. - PLC SCADA Courses - Engineers Community

You might also like