FOR YOUR SAFETY, PLEASE READ CAREFULLY!
DO NOT attempt to operate this regulator unless you have been trained in its proper use or are under competent supervision.
Zu Ihrer eigenen Sicherheit ,Bitte sorgfältig lesen!
!
Vor Inbetriebnahme der Armatur ist die Beachtung der Bedienungsanleitung und insbesondere der Sicherheitshinweise erforderlich.
1. General
1.1 Contents
1.1.1 Safety instructions
1.1.2 Applications
1.2 Preface
2. Safety instructions
2.1 These pressure control panel are made to the demands of the existing standards and regulations.
2.2 Changes or modifications to the pressure control panel are prohibited without the prior consent of the manufacturer.
2.3 Pressure control panel are to be handled by trained personnel.
2.4 Improper handling and use can evolve risks for the user and other persons as well as damage to the device.
2.5 All parts in contact with oxygen must be kept free of oil and grease!
2.6 The valve has always to be opened slowly!
2.7 Pressure control panel should only be used with gases indicated in the labeling.
2.8 The pressure control panel must not be exposed to ambient temperatures below -30°C and above +74°C
3. Applications
3.1 Usages
VIGOUR VSP Series Pressure control panels described are intended for use with non-corrosive gases up to a
purity grade of 6.0.The permissible gases and pressure ranges are specified on the type plate.
Pressure control panels reduce a variable inlet pressure to an outlet pressure that is as constant as possible.
4.Technical Information
4.1 Labeling
1
4.2 Model Information
VSP-3AE S -300-10-P-P-N2
4.3 Scheme
Waste gas Process gas Waste gas Process gas Waste gas Waste gas Process gas Waste gas
Waste gas Process gas Waste gas Process gas Waste gas
Waste gas
2
VSP-1/1E control panel are Single bank of gas supply design with a regulator complete purging features, it ensure
process gas purity are maintained during changing of gas cylinders.
VSP-2/2E control panel are Dual bank of gas supply design with a regulator complete purging features, it ensure process
gas purity are maintained during changing of gas cylinders.
VSP-3A/3AE/5A/6A/8AE Series Automatic Changeover Control panel are design with two pressure regulators which are
connected on the both outlet pressure side and are set to different outlet pressures. The changeover occurred
automatically by pressure differential.
First, the regulator with the higher outlet pressure (= working pressure) delivers the process gas. When the cylinder
pressure drops below the outlet pressure set on the other side (changeover pressure), this cylinder takes over the supply
of the process gas without interrupting the continuity of supply. Before changing a cylinder, the position of the handle
must be reversed in order to give priority to the cylinder which is now delivering the process gas (outlet pressure
increase).
VSP-3A/3AE/6A Series
VSP-8AE Series
3
5. Installation
5.2 Slide the mounting plate into the upper slot of the
basic frame. Insert it in the basic frame and lowered it
into the bottom slot..
Now it is held in position.
4
5.7.1 Push the tubes as far as it will go into the compression ring fitting.
5.7.2 Tighten union nut by hand
5.7.3 Tighten union nut with a wrench by ¼ turns while applying a counter grip with a second wrench.
Note: ¼ turns when the compression rings a pre-assembled on the pipe. ¼ turns when they are not
preassembled.
5.8 When using corrosive, toxic, self-igniting or flammable gases, ensure that the purge gas is charged
safety. The waste gas line must be connected to the waste gas valve and to the relief valve of the regulator
6.Start-up
6.1 Before 1st usages, the complete pressure control panel must be purged . If corrosive gas is to be used
as the process gas, dry inert gas must be used for purging.
6.2 Close all valves and unscrew the adjusting knob of the regulators
6.3 Gradually open the cylinder valve.
6.4 Gradually open the process gas valve.
6.5 Set the regulator to the required outlet pressure with the adjusting knob, observe and make sure no
unusual vibration in the regulator.
6.6 Check the complete control panel and all threaded connections for leaks.
7. Shut-down
7.1 During short work break / interruption, shut off the gas supply from the pressure regulator.
7.2 Work finished, shut off regulator and also gas cylinder valve.
7.3 When removing pressure regulator from gas cylinder, always check pressure gauges indicate zero.
8. Changing cylinder
*IMPORTANT:
When process valve needs to be closed although no flow at the downstream.
Turn the knot of the regulator anti-clockwise to release the pressure fully.
5
9. Repair
9.1 Repairs may only be carried out in authorized workshops by expert persons.
9.2 Only original spare parts may be used. The materials have been adapted to the gas type in each instance.
9.3 In case of unauthorized repairs or use of non-original spare parts, any form of liability for resulting damages will
expire as well as the manufacturer’s warranty.
9.4 After being repaired, the pressure regulator must be checked with respect to proper function, leak-tightness and
cleanliness of the gas-wetted surfaces.
10.This Document
We reserve the right to make technical alterations which improve the product. All rights to the manufacturer.
6
BedienungsanleitungDruckminderer
Serie VSP
Zu Ihrer eigenen Sicherheit ,Bitte sorgfältig lesen!
Vor Inbetriebnahme der Armatur ist die Beachtung der Bedienungsanleitung und insbesondere der Sicherheitshinweise erforderlich.
7
1. Anwendung
2. Sicherheitshinweise
2.1 Diese Entspannungsstation entspricht dem Stand und den anerkannten Regeln der Technik.
2.2 Ohne Genehmigung des Herstellers dürfen keine Änderungen oder Umbauten an der
Entspannungsstation vorgenommen werden.
2.3 Entspannungsstationen dürfen nur von eingewiesenem Personal bedient werden.
2.4 Bei unsachgemäßer Behandlung und nicht bestimmungsgemäßer Verwendung können
Gefahren für den Verwender und andere Personen sowie eine Beschädigung des Gerätes eintreten.
2.5 Zu beachtende Vorschriften
2.6 Unfallverhütungsvorschriften
BGV 1 "Allgemeine Vorschriften"
BGV 4 "Elektrische Anlagen und Betriebsmittel"
BGV D1 "Schweißen, Schneiden und verwandte Verfahren"
BGV D34 "Flüssiggas"
BGV B6 "Gase"
BGV B7 "Sauerstoff”
2.7 Gesetze, Verordnungen, Technische Regeln
Gesetze über Technische Arbeitsmittel, (Gerätesicherheitsgesetz, FTA) mit allgemeiner
Verwaltungsvorschrift, sowie Verzeichnisse A und B zum GtA.
8
Verordnung über Druckbehälter, Druckgasbehälter und Füllanlagen,
(Druckbehälterverordnung) und allgemeine Verwaltungsvorschrift sowie Anhänge zur Druckbehälterverordnung.
Technische Regeln Druckgase TRG, insbesondere TRG 280
Technische Regeln Rohrleitungen TRR, insbesondere TRR 100
Verordnung über elektrische Anlagen in explosionsgefährdeten Räumen (ElexV).
Errichten elektrischer Anlagen in explosionsgefährdeten Bereichen VDE 0170/DIN 57 165
Elektrische Betriebsmittel für explosionsgefährdete Bereiche VDE0170/DIN 57 165
Blitzschutzanlagen VDE 0185/DIN 57 185
Bestimmungen für das Einbeziehen von Rohrleitungen in Schutzmassnahmen von Starkstromanlagen mit
Nennspannungen bis 1000 V, VDE 0190
2.8 Merkblätter und Richtlinien
Richtlinie 97/23/EG Artikel 3; Absatz 3
Richtlinien für die Vermeidung von Zündgefahren infolge elektrostatischer Aufladungen.
Richtlinien für Laboratorien ZH1/119
Merkblatt Umgang mit Druckgasflaschen (DVS 0212)
Sicherheitsdatenblätter der jeweiligen Gasart
2.9 Rauchen oder offenes Feuer (z. B. Kerzen) in der Nähe Ihrer Gasversorgung ist streng verboten!
BRAND BZW: EXPLOSIONSGEFAHR!
3. Inbetriebnahme
Vor Aufnahme der Arbeit sind die Angaben dieser Betriebsanleitung, insbesondere die Sicherheitshinweise zu
lesen und während der Arbeit zu befolgen.
Prüfen, ob Gasflaschenventil- Anschluss sauber und ohne Beschädigung ist
(evtl. mit sauberer, d.h. öl,- fett- und wasserfreier Druckluft ausblasen).
Bei Beschädigung darf der Druckminderer nicht angeschlossen werden.
Druckminderer anschließen.
Verbindung zwischen Druckregler und Verbrauchsgerät herstellen.
Bei brennbaren und giftigen Gasen eine Abblasleitung an das Abblaseventil
anbringen. Alle Verbindungen auf Dichtheit prüfen.
Druckregler entspannen, Gasflaschenventil langsam öffnen (Hochdruckmanometer zeigt Flaschendruck an) dann
den gewünschten Hinterdruckdruck am Einstellgriff einstellen.
Druckeinstellung bei Druckabfall möglicherweise nochmals korrigieren.
Außerbetriebnahme
3.1 Alle Ventile schließen.
3.2 Eingestellten Druckregler komplett druckentlasten; Manometer zeigen 0 bar an. Handrad des Druckreglers
herausdrehen.
4. Reparatur
4.1 Reparaturen dürfen nur von sachkundigen Personen in autorisierten Reparaturwerkstätten ausgeführt werden.
4.2 Es dürfen nur Original-Ersatzteile verwendet werden. Die Werkstoffe sind auf die jeweilige
Gasart abgestimmt, deshalb immer Gasart beachten.
4.3 Nach Reparatur muss die Entspannungsstation komplett geprüft werden. Bei
Entspannungsstationen für Sauerstoff sind die Auflagen der Bauartanerkennung zu beachten.
4.4 Bei eigenmächtigen Reparaturen, oder Verwendung von Nicht-Original- Ersatzteilen, erlischt jegliche Haftung
der daraus entstehenden Schäden, sowie die Gewährleistung des Herstellers.