You are on page 1of 1

EUROPEAN PHARMACOPOEIA 11.0 2.1.3.

Ultraviolet ray lamps for analytical purposes

2.1. APPARATUS 01/2008:20102

01/2008:20101

2.1.2. COMPARATIVE TABLE OF


POROSITY OF SINTERED-GLASS
2.1.1. DROPPERS FILTERS(1)
The term ‘drops’ means standard drops delivered from a Table 2.1.2.-1
standard dropper as described below. Porosity number Maximum Germany France United
Standard droppers (Figure 2.1.1.-1) are constructed of (Ph. Eur.)(2) diameter of pores Kingdom
in micrometres
practically colourless glass. The lower extremity has a circular 1.6 – –
less than 1.6 5f
orifice in a flat surface at right angles to the axis.
– 1 - 2.5 5 – 5
4 1.6 - 4 – – –
– 4-6 – 5 –

10 4 - 10 4f – 4
16 10 - 16 4 4 –

40 16 - 40 3 3 3
– 40 - 50 – – 2
100 40 - 100 2 2 –
– 100 - 120 – – 1
160 100 - 160 1 1 –
– 150 - 200 0 0 –

250 160 - 250 – – –


– 200 - 500 – 00 –

Special Uses
Diameters in micrometres
< 2.5 Bacteriological filtration
4 - 10 Ultra-fine filtration, separation of micro-organisms of large
diameter
10 - 40 Analytical filtration, very fine filtration of mercury, very fine
dispersion of gases
40 - 100 Fine filtration, filtration of mercury, fine dispersion of gases
100 - 160 Filtration of coarse materials, dispersion and washing of gases,
support for other filter materials
160 - 500 Filtration of very coarse materials, dispersion and washing of
gases.

01/2023:20103

2.1.3. ULTRAVIOLET RAY LAMPS FOR


ANALYTICAL PURPOSES
Mercury vapour in quartz lamps is used as a source of
Figure 2.1.1.-1. – Standard dropper ultraviolet (UV) light. A suitable filter may be fitted to
eliminate the visible part of the spectrum emitted by the
Dimensions in millimetres
lamp. When the Pharmacopoeia prescribes in a test the use
Other droppers may be used provided they comply with the of ultraviolet light of wavelength 254 nm or 365/366 nm, an
following test. instrument consisting of a mercury vapour lamp and a filter
20 drops of water R at 20 ± 1 °C flowing freely from the that gives an emission band with maximum intensity at about
dropper held in the vertical position at a constant rate of 254 nm or 365/366 nm may be used.
1 drop per second weighs 1000 ± 50 mg. Alternatively, lamps fitted with fluorescent tubes emitting UV
The dropper must be carefully cleaned before use. Carry out 3 light at 254 nm (narrow bandwidth) or with fluorescent tubes
determinations on any given dropper. No result may deviate by emitting broadband UVA light (315-380 nm ; long-wave UV
more than 5 per cent from the mean of the 3 determinations. light) can be used.
(1) The given limits are only approximate.
(2) The European Pharmacopoeia has adopted the system proposed by the International Organization for Standardization (ISO).

General Notices (1) apply to all monographs and other texts 17


www.webofpharma.com

You might also like