Microscopy Buffer Tablets PH 7.2 Buffer Tablets PH 6.4 Buffer Tablets PH 6.8
Microscopy Buffer Tablets PH 7.2 Buffer Tablets PH 6.4 Buffer Tablets PH 6.8
0100
1.11373.0100 Reagents
1.11374.0100 Cat. No. 109468
Buffer tablets pH 7.2
for preparing buffer solution acc. to WEISE
100 tabs
Also required:
Buffer tablets pH 6.4 for Giemsa stain:
Cat. No. 106009 Methanol for analysis EMSURE® 1 l, 2.5 l, 5 l
for preparing buffer solution acc. to WEISE ACS,ISO,Reag. Ph Eur
for the staining of blood smears Cat. No. 109204 Giemsa’s azur eosin methylene blue 100 ml, 500 ml,
solution 1 l, 2.5 l
for microscopy
Buffer tablets pH 6.8 for May-Grünwald stain:
Cat. No. 101424 May-Grünwald’s eosine-methylene blue 100 ml, 500 ml,
for preparing buffer solution acc. to WEISE solution modified 1 l, 2.5 l
for staining of blood smears for microscopy
for Wright stain:
Cat. No. 101383 Wright’s eosin methylene blue solution 100 ml, 500 ml,
for microscopy 2.5 l
For professional use only
for Leishman stain:
Cat. No. 105387 Leishman’s eosin methylene blue solution 500 ml
modified
In Vitro Diagnostic Medical Device for microscopy
Sample preparation
The sampling must be performed by qualified personnel.
Intended purpose
Please use thin, air-dried blood or bone-marrow smears, as well as cytologi-
These cal materials, that have been stored not longer than three days.
Buffer tablets pH 6.4 - for preparing buffer solution acc. to WEISE for the The smears must be dried in air for at least 30 minutes and be fixed ac-
staining of blood smears cording to the relevant instructions prior to the actual reaction.
Buffer tablets pH 6.8 - for preparing buffer solution acc. to WEISE for the
staining of blood smears All samples must be treated using state-of-the-art technology.
Buffer tablets pH 7.2 - for preparing buffer solution acc. to WEISE for stain- All samples must be clearly labeled.
ing of blood smears Suitable instruments must be used for taking samples and their prepara-
are used to prepare buffer solutions according to WEISE and pH-buffered tion. Follow the manufacturer’s instructions for application / use.
staining solutions. These solutions are used in conjunction with other in When using the corresponding auxiliary reagents, the corresponding in-
vitro diagnostic products from our portfolio for the reproducible staining of structions for use must be observed.
sample materials of human origin, serving to make target structures evalu-
able for diagnostic purposes (by fixing, embedding, staining, counterstain-
Reagent preparation
ing, mounting) in haematological and clinico-cytological specimen material,
such as whole blood and bone-marrow smears. Buffer solution acc. to WEISE pH 6.4, 6.8, or 7.2
The type of buffer tablet that is appropriate for use from case to case is de- For preparation of approx. 1000 ml of solution, add and dissolve:
scribed in the corresponding instructions for use for our staining solutions,
solid dyes, and test kits. Buffer tablet
Cat. No. 111373 (pH 6.4) or
Phosphate buffer solutions acc. to WEISE are used for Giemsa’s, May-
Cat. No. 111374 (pH 6.8) or 1
Grünwald’s, combination of Giemsa’s/May-Grünwald’s (“Pappenheim stain”),
Cat. No. 109468 (pH 7.2)
Wright’s, Leishman’s and Hemacolor® stainings (Cat. No.111661 and
depending on the required reaction color
111674) are used as solutions for rinsing and for dilution if necessary.
Distilled water 1000 ml
Principle
Neutral distilled water is required for blood smear staining to obtain repro- The buffer solution can be used for at least 4 weeks when kept in a tightly
ducible staining results. It is necessary both for preparing dilute Giemsa’s closed glass bottle.
solution and for rinsing stained slides.
Freshly distilled water can be stabilized by adding a buffer. The buffer mix-
Dilute Giemsa’s staining solution for manual staining
ture acc. to WEISE is an effective means for this purpose. Upon addition of
this buffer mixture the pH is automatically adjusted to the required value. For preparation of approx. 200 ml solution mix:
Fluctuations in the pH of distilled water are caused on the one hand by
Giemsa’s azur eosin methylene blue solution 10 ml
atmospheric carbon dioxide and on the other by detergent residues on
equipment. Acidic distilled water produces a red-tinted stain, and alkaline Buffer solution acc. to WEISE pH 6.4, 6.8, or 7.2 190 ml
distilled water a blue-tinted stain.
Mix well, leave to stand for 10 min, and filter if necessary
Buffered solutions of pH 6.4, 6.8, or 7.2 acc. to WEISE reliably prevent
discolorations.
Buffer solution acc. to WEISE pH 6.4 shows a result with reddish
stained erythrocytes and red-violet stained cells with nuclei, Procedure
buffer solution acc. to WEISE pH 6.8 shows reddish stained erythro- The stated times should be adhered to in order to guarantee an optimal
cytes and more intense red-violet stained cells with nuclei, staining result.
buffer solution acc. to WEISE pH 7.2 shows reddish-brownish stained
erythrocytes and even more intensive red-violet stained cells with nuclei. Giemsa stain on the staining rack
Air-dry (e. g. over night or at 50°C in the drying cabinet) neutrophilic light violet light violet light violet light violet
granula
Wright stain on the staining rack eosinophilic reddish to red- brick-red brick-red to brick-red
granula brown red-brown
Slide with air-dried smear basophilic dark violet dark violet to dark violet to dark violet
granula black black
Wright’s eosin methylene 1 ml cover completely 1 min
blue solution Thrombocytes violet violet violet violet
Buffer solution acc. to 1 ml mix 4 min Erythrocytes reddish reddish reddish reddish
WEISE
Buffer solution acc. to rinse 1 min Buffer solution acc. to WEISE pH 7.2
WEISE
Giemsa May-Grün- Wright Leishman
Buffer solution acc. to rinse 1 min stain wald stain stain
WEISE stain
Air-dry (e. g. over night or at 50°C in the drying cabinet) Nuclei resp. violet violet blue-violet red-violet
chromatin
Diagnostics
Result
Diagnoses are to be made only by authorized and qualified personnel.
Buffer solution acc. to WEISE pH 6.4
Valid nomenclatures must be used.
This method can be supplementarily used in human diagnostics.
Giemsa May-Grün- Wright Leishman
stain wald stain stain Further tests must be selected and implemented according to recognized
stain methods.
Suitable controls should be conducted with each application in order to
Nuclei resp. red to violet red-violet red-violet red-violet avoid an incorrect result.
chromatin
Capacity
1 buffer tablet is sufficient for the preparation of 1 l buffer solution acc. to
WEISE.
Additional instructions Literature
For professional use only. 1. H
istological & Histochemical Methods, J. A. Kiernan,1990, Pergamon
In order to avoid errors, the application must be carried out by qualified Press, Second Edition
personnel only. 2. R
omeis - Mikroskopische Technik, Editors: Maria Mulisch, Ulrich Welsch,
National guidelines for work safety and quality assurance must be followed. 2015, Springer Spektrum, 19. Auflage
Microscopes equipped according to the standard must be used.
If necessary use a standard centrifuge suitable for medical diagnostic labo- 3. K
linische Hämatologie, Herbert Begemann, 1975, Georg Thieme Verlag,
ratory. 2. Auflage
Auxiliary reagents
Cat. No. 100579 DPX new 500 ml
non-aqueous mounting medium
for microscopy
Cat. No. 101383 Wright’s eosin methylene blue solution 100 ml, 500 ml,
for microscopy 2.5 l
Cat. No. 101424 May-Grünwald’s eosine-methylene blue 100 ml, 500 ml,
solution modified 1 l, 2.5 l
for microscopy
Cat. No. 103699 Immersion oil Type N acc. to ISO 8036 100-ml drop-
for microscopy ping bottle
Cat. No. 104699 Immersion oil 100-ml drop-
for microscopy ping bottle,
100 ml, 500 ml
Cat. No. 105387 Leishman’s eosin methylene blue 500 ml
solution modified
for microscopy
Cat. No. 106009 Methanol for analysis EMSURE® 1 l, 2.5 l, 5 l
ACS, ISO, Reag. Ph Eur
Cat. No. 107961 Entellan® new 100 ml, 500 ml,
rapid mounting medium 1l
for microscopy
Cat. No. 109016 Neo-Mount® 100-ml drop-
anhydrous mounting medium ping bottle,
for microscopy 500 ml
Cat. No. 109204 Giemsa’s azur eosin methylene blue 100 ml, 500 ml,
solution 1 l, 2.5 l
for microscopy
Hazard classification
Cat. No. 109468
Cat. No. 111373
Cat. No. 111374
Please observe the hazard classification printed on the label and the infor-
mation given in the safety data sheet.
The safety data sheet is available on the website and on request.
blood smear
staining set for microscopy