Technical Information
Deltabar FMD71, FMD72
Level measurement with electronic differential
pressure
Application
The Deltabar FMD71/FMD72 is used to measure the differential pressure or level,
volume or mass of liquids in pressurized or vacuum vessels
Your benefits
The electronic differential pressure system eliminates traditional mechanical
issues resulting in greater process availability and reliability
• Safety risks are minimized with the electronic differential pressure system
architecture and design
• Lowest total cost of ownership due to reduced installation time, maintenance,
downtime and spare requirements, downtime and spare requirements.
• Multivariable level measurement: HART-based differential pressure, head pressure
and sensor temperatures from one system
• Continuous health indication of the entire system via HART-based diagnostic
• High reproducibility and long-term stability
• Overload-resistant and function-monitored
Deltabar FMD71, FMD72
Table of contents
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Mounting location . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
FMD71: measuring range of individual sensors . . . . . . . . . 11 Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
FMD72: measuring range of individual sensors . . . . . . . . . 12 General installation instructions . . . . . . . . . . . . . . . . . . . 26
Thermal insulation - FMD71 high-temperature version . . . 26
Installing the sensor modules . . . . . . . . . . . . . . . . . . . . 27
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Mounting sensor modules with PVDF installation
Output Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Signal range 4 to 20 mA . . . . . . . . . . . . . . . . . . . . . . . . 13
Installing the transmitter . . . . . . . . . . . . . . . . . . . . . . . 27
Signal on alarm 4 to 20 mA . . . . . . . . . . . . . . . . . . . . . . 13
Sensor and transmitter cable . . . . . . . . . . . . . . . . . . . . . 28
Maximum load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Dead time, time constant . . . . . . . . . . . . . . . . . . . . . . . 14
Dynamic behavior, current output . . . . . . . . . . . . . . . . . . 14 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Dynamic behavior, HART . . . . . . . . . . . . . . . . . . . . . . . 14 Ambient temperature range . . . . . . . . . . . . . . . . . . . . . 29
Damping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Storage temperature range . . . . . . . . . . . . . . . . . . . . . . 29
alarm current setting . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Climate class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Firmware version . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Wireless HART data . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Protocol-specific data . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . 29
2 Endress+Hauser
Deltabar FMD71, FMD72
Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Operation concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Operating languages . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Ordering information . . . . . . . . . . . . . . . . . . . . . . . 69
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Measuring point (TAG) . . . . . . . . . . . . . . . . . . . . . . . . . 69
Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Field of Activities . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Technical Information . . . . . . . . . . . . . . . . . . . . . . . . . 70
Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . . 70
Brief Operating Instructions . . . . . . . . . . . . . . . . . . . . . 70
Safety Instructions (XA) . . . . . . . . . . . . . . . . . . . . . . . . 70
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Service-specific accessories . . . . . . . . . . . . . . . . . . . . . . 71
Registered trademarks . . . . . . . . . . . . . . . . . . . . . . 71
HARTâ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Endress+Hauser 3
Deltabar FMD71, FMD72
Electrical symbols
Symbol Meaning
Direct current
Alternating current
Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a
grounding system.
Symbol Meaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation
A Reference to page
Reference to graphic
Visual inspection
4 Endress+Hauser
Deltabar FMD71, FMD72
Symbols in graphics
Symbol Meaning
1. , 2. , 3. … Series of steps
A, B, C, ... Views
Brief Operating Instructions (KA): getting the 1st measured value quickly
KA01105P:
The Brief Operating Instructions contain all the essential information from incoming acceptance to
initial commissioning.
Endress+Hauser 5
Deltabar FMD71, FMD72
p
0
LRL LRV URV URL MWP OPL
A0029505
1 OPL OPL: The OPL (over pressure limit = sensor overload limit) for the measuring
device depends on the lowest-rated element, with regard to pressure, of the
selected components, i.e. the process connection has to be taken into
consideration in addition to the measuring cell. Take pressure-temperature
dependency into consideration.
2 MWP MWP: The MWP (maximum working pressure) for the sensors depends on the
lowest-rated element, with regard to pressure, of the selected components, i.e. the
process connection has to be taken into consideration in addition to the
measuring cell. Take pressure-temperature dependency into consideration. The
MWP may be applied at the device for an unlimited period of time. The MWP can
be found on the nameplate.
p - Pressure
A0029545
1 Calibrated/adjusted span
2 Zero point-based span
3 Upper range limit
6 Endress+Hauser
Deltabar FMD71, FMD72
Example
URL
TD =
|URV - LRV|
Endress+Hauser 7
Deltabar FMD71, FMD72
1 2 3 4
p
A0020465
1
2
3
4
p
A0016448
8 Endress+Hauser
Deltabar FMD71, FMD72
Measuring system The FMD71/FMD72 consists of 2 sensor modules and one transmitter. One sensor module measures
the hydrostatic pressure (high pressure) and the other one the head pressure (low pressure). The
level (electronic differential pressure) is calculated in the transmitter using these two digital values.
Process connections
• Thread
• EN and ASME flanges
• Flush mount hygienic connections
MWP (depends on the lowest-rated element, with regard to pressure, of the selected
components)
FMD71:
to 40 bar (600 psi)
FMD72:
to 100 bar (1 500 psi)
Supply voltage
• Version for non-hazardous areas: 12 (13) to 45 V DC
• Ex ia: 12 (13) to 30 V DC
Output
4 to 20 mA with superimposed HART protocol
Endress+Hauser 9
Deltabar FMD71, FMD72
Options
• NACE-compliant materials
• Software pre-configurations
Product design Level measurement (level, volume and mass) with Deltabar:
LP 1
p2
p1
PLP = p2 HP
PHP = p1 + p2
A0016449
The FMD71/FMD72 is best suited to level measurement in vessels with pressure overlay or in
vacuum vessels and tanks, high distillation columns and other vessels with changing ambient
temperatures.
The sensor module HP is mounted on the lower measuring connection and the sensor module LP is
mounted above the maximum level. The transmitter can be mounted on pipes or walls with the
mounting bracket.
The sensor signal is transmitted digitally. In addition, sensor temperatures and the individual process
pressures present at the respective sensor modules can be individually evaluated and transmitted.
NOTICE
Incorrect sizing/order of sensor modules
‣ In a closed system, please note that the sensor module is affected by the superimposed head
pressure (p2) in addition to the hydrostatic pressure (p1). This must be taken into account when
sizing the sensor module on the high-pressure side (HP).
System integration The device can be given a tag name (max. 8 alphanumeric characters).
Measuring point (TAG), see additional specifications.: Product Configurator order code for
"Identification", option "Z1"
10 Endress+Hauser
Deltabar FMD71, FMD72
Input
Measured variable Measured process variables
• Pressure HP and Pressure LP
• Sensor temperature HP and sensor temperature LP
• Transmitter temperature
FMD71: measuring range of The maximum span of the differential pressure corresponds to the URL of the HP sensor.
individual sensors
Gauge pressure
Sensor Maximum sensor measuring range MWP OPL Vacuum resistance Option 1)
[bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] [barabs (psiabs)]
100 mbar (1.5 psi) –0.1 (–1.5) +0.1 (+1.5) 2.7 (40.5) 4 (60) 0.7 (10.5) 1C
250 mbar (4 psi) –0.25 (–4) +0.25 (+4) 3.3 (49.5) 5 (75) 0.5 (7.5) 1E
400 mbar (6 psi) –0.4 (–6) +0.4 (+6) 5.3 (79.5) 8 (120) 0 1F
Absolute pressure
Sensor Maximum sensor measuring range MWP OPL Vacuum resistance Option 1)
[barabs (psiabs)] [barabs (psiabs)] [barabs (psiabs)] [barabs (psiabs)] [barabs (psiabs)]
Endress+Hauser 11
Deltabar FMD71, FMD72
FMD72: measuring range of The maximum span of the differential pressure corresponds to the URL of the HP sensor.
individual sensors
Gauge pressure
Sensor Maximum sensor measuring range MWP OPL Vacuum resistance 1) Option 2)
[bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] [barabs (psiabs)]
400 mbar (6 psi) –0.4 (–6) +0.4 (+6) 4 (60) 6 (90) 0.01 (0.15) 1F
1 bar (15 psi) –1 (–15) +1 (+15) 6.7 (100) 10 (150) 0.01 (0.15) 1H
2 bar (30 psi) –1 (–15) +2 (+30) 13.3 (200) 20 (300) 0.01 (0.15) 1K
4 bar (60 psi) –1 (–15) +4 (+60) 18.7 (280.5) 28 (420) 0.01 (0.15) 1M
10 bar (150 psi) –1 (–15) +10 (+150) 26.7 (400.5) 40 (600) 0.01 (0.15) 1P
40 bar (600 psi) –1 (–15) +40 (+600) 100 (1500) 160 (2400) 0.01 (0.15) 1S
1) The vacuum resistance applies for the measuring cell under reference operating conditions. (see "Reference operating conditions" section)
2) Product Configurator order code for "Sensor range"
Absolute pressure
Sensor Maximum sensor measuring range MWP OPL Vacuum resistance 1) Option 2)
[barabs (psiabs)] [barabs (psiabs)] [barabs (psiabs)] [barabs (psiabs)] [barabs (psiabs)]
10 bar (150 psi) 0 +10 (+150) 26.7 (400.5) 40 (600) 0.01 (0.15) 2P
40 bar (600 psi) 0 +40 (+600) 100 (1500) 160 (2400) 0.01 (0.15) 2S
1) The vacuum resistance applies for the measuring cell under reference operating conditions. (see "Reference operating conditions" section)
2) Product Configurator order code for "Sensor range"
12 Endress+Hauser
Deltabar FMD71, FMD72
Output
Output Signal 4 to 20 mA with superimposed digital communication protocol HART 6.0, 2-wire
1)
Software operation Internal + LCD External + LCD Internal
on Display on Display
off off
R R
SW /min
SW /min
HART HART
damp.
damp.
FIELD COMMUNICATION PROTOCOL FIELD COMMUNICATION PROTOCOL
Sensor Sensor
– + E – + E
A0021280
2)
Order code option 4 5 8
1) In the case of the T17 transmitter housing, the operating keys are always arranged internally on the electronic insert.
2) Product Configurator order code for "Display, operation:"
Maximum load In order to guarantee sufficient terminal voltage in two-wire devices, a maximum load resistance R
(including line resistance) must not be exceeded depending on the supply voltage U0 of the supply
unit. In the following load diagrams, observe the position of the jumper and the explosion protection:
R Lmax A R Lmax B
[W ] [W ]
1435 1391
Test
1217 1174 Test
783 739
348 304
12 20 30 40 45 U 13 20 30 40 45 U
[V] [V]
1 1
2 2
3 RLmax £ U – 12 V 3 RLmax £ U – 13 V
23 mA 23 mA
A0017533
When operating via a handheld terminal or via a PC with an operating program, a minimum
communication resistance of 250 Ω must be taken into account.
Endress+Hauser 13
Deltabar FMD71, FMD72
Dead time, time constant Presentation of the dead time and the time constant:
I
100 %
90 %
63 %
t1 t2 t
t3
A0019786
Dynamic behavior, Dead time (t1) [ms] Time constant (T63), t2 Time constant (T90), t3
current output
max. 120 120 280
Dynamic behavior, HART Dead time (t1) [ms] Dead time (t1) [ms] + Dead time (t1) [ms] +
Time constant T63 (= t2) [ms] Time constant T90 (= t3) [ms]
Reading cycle
• Acyclic: max. 3/s, typical 1/s (depends on command # and number of preambles)
• Cyclic (Burst): max. 3/s, typical 2/s
The Deltabar FMD71/FMD72 offers BURST MODE functionality for cyclical value transmission via
the HART communication protocol.
Response time
• Acyclic: min. 330 ms, typically 590 ms (depending on command # and number of preambles)
• Cyclic (burst): min. 160 ms, typically 350 ms (depending on command # and number of
preambles)
alarm current setting Adjusted min. alarm current: Product configurator order code for "Service", option "IA"
Starting time 10 s
14 Endress+Hauser
Deltabar FMD71, FMD72
Multidrop current 4 mA
Endress+Hauser 15
Deltabar FMD71, FMD72
Device description files (DTM, DD) Information and files available at:
• www.endress.com
• www.fieldcommgroup.org/registered-products
16 Endress+Hauser
Deltabar FMD71, FMD72
Power supply
LWARNING
Electrical safety is compromised by an incorrect connection!
‣ In accordance with IEC/EN61010 a separate circuit breaker must be provided for the device .
‣ When using the measuring device in hazardous areas, installation must comply with the
corresponding national standards and regulations and the Safety Instructions or Installation or
Control Drawings.
‣ All explosion protection data are given in separate documentation which is available upon
request. The Ex documentation is supplied as standard with all devices approved for use in
explosion hazardous areas.
‣ Devices with integrated overvoltage protection must be grounded.
‣ Protective circuits against reverse polarity, HF influences and overvoltage peaks are integrated.
Terminals • Supply voltage and internal ground terminal: 0.5 to 2.5 mm2 (20 to 14 AWG)
• External ground terminal: 0.5 to 4 mm2 (20 to 12 AWG)
Terminal assignment
1 2
Test
3
Test
7
6
A0019989
1 Housing
2 Supply voltage
3 4 to 20 mA
4 Devices with integrated overvoltage protection are labeled "OVP" (overvoltage protection) here.
5 External ground terminal
6 4 to 20 mA test signal between positive and test terminal
7 Internal ground terminal, minimum supply voltage = 12 V DC, jumper is set as illustrated in the diagram.
8 Jumper for 4 to 20 mA test signal
Supply voltage Electronic version Jumper for 4 to 20 mA test signal in Jumper for 4 to 20 mA test signal
"Test" position (delivery status) in "Non-test" position
Endress+Hauser 17
Deltabar FMD71, FMD72
Test • Measurement of 4 to 20 mA test signal via the positive and test terminal:
possible. (Thus, the output current can be measured without interruption via
the diode.)
• Delivery status
• Minimum supply voltage: 13 V DC
A0019992
Test • Measurement of 4 to 20 mA test signal via positive and test terminal: not
possible.
• Minimum supply voltage: 12 V DC
A0019993
Cable entries Explosion protection Cable gland Permitted cable diameter Permitted wire cross-sections
• Standard Plastic M20x1.5 5 to 10 mm (0.2 to 0.39 in) 0.5 to 2.5 mm2 (20 to 14 AWG)
• Ex ia
• Ex ic
Cable specification for • Endress+Hauser recommends using twisted, shielded two-wire cables.
transmitter connection • Terminals for core cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
• The cable outer diameter depends on the cable entry used.
Residual ripple Without influence on 4 to 20 mA signal up to ±5 % residual ripple within the permitted voltage
range [according to HART hardware specification HCF_SPEC-54 (DIN IEC 60381-1)]
NOTICE
Device could be destroyed!
‣ Devices with integrated overvoltage protection must be grounded.
18 Endress+Hauser
Deltabar FMD71, FMD72
90
°
A
°
90
B
A0016465
Process isolating diaphragm axis is horizontal (A) Process isolating diaphragm pointing Process isolating diaphragm pointing
upwards (B) downwards (C)
Calibration position, no measuring error < +0.2 mbar (+0.003 psi) < –0.2 mbar (–0.003 psi)
This effect can be corrected using the function to adjust the position (position adjustment) for the
differential pressure. Additional position adjustments for individual pressure signals are not
available.
A position-dependent zero shift can be corrected on the device.
IEC 61298-3 ≤ Reference accuracy up to 10 to 60 Hz: ±0.35 mm (±0.01 in); 60 to 500 Hz: 2 g
Application limits A high ratio between the level and head pressure or between the differential pressure and static
pressure can result in large measured errors. A maximum ratio of 1:10 is recommended. For
calculation purposes, please use the free "Applicator" calculation tool, which is available online at
"www.endress.com/applicator" or on CD-ROM.
Endress+Hauser 19
Deltabar FMD71, FMD72
Reference accuracy The reference accuracy contains the non-linearity [DIN EN 61298-2 3.11] including the pressure
hysteresis [DIN EN 61298-23.13] and non-repeatability [DIN EN 61298-2 3.11] in accordance with
the limit point method as per [DIN EN 60770].
Measuring cell Sensor Reference accuracy (A) Calculated reference accuracy (ADiff)
[%URL for every sensor] of the differential pressure
Standard Platinum
100 mbar (1.5 psi) Gauge pressure A= ±0.075 - Calculation (mbar, bar or psi):
A= ±0.15 1)
2 2
250 mbar (3.75 psi) Gauge pressure A= ±0.075 - ADiff = (AHP. URLHP) + (ALP . URLLP)
A= ±0.15 1)
100 100
400 mbar (6 psi) Gauge pressure A= ±0.075 - A0016468
A= ±0.15 1)
Percentage calculation of URL dP:
1 bar (15 psi) Gauge pressure/ Absolute pressure A= ±0.075 A = ±0.05
2 bar (30 psi) A= ±0.15 1) ±0.075 1)
ADiff . 100
4 bar (60 psi) ADiff [%]=
10 bar (150 psi) PDiff
40 bar (600 psi) A0016469
Ordering Information
Description Option 1)
Platinum D
Standard G
Measuring cell –10 °C (+14 °F)to ≤ –20 to –10 °C (–4 to +14 °F) Calculated thermal change (TDiff)
+60 °C (+140 °F) > +60 to +125 °C (+140 to +257 °F) of the differential pressure
100 mbar (1.5 psi) Ttotal = ±0.176 Ttotal = ±0.276 Calculation (mbar, bar or psi):
250 mbar (4 psi)
400 mbar (6 psi) 2 2
TDiff = (THP . URLHP) + (TLP . URLLP)
1 bar (15 psi) Ttotal = ±0.092 Ttotal = ±0.250
2 bar (30 psi) 100 100
A0016474
4 bar (60 psi)
10 bar (150 psi) Percentage calculation of URL dP:
40 bar (600 psi)
TDiff . 100
TDiff [%]=
PDiff
A0016475
20 Endress+Hauser
Deltabar FMD71, FMD72
Measuring cell Sensor –10 °C (+14 °F)to ≤ > Calculated thermal change (TDiff)
+60 °C (+140 °F) +60 to +150 °C of the differential pressure
(140 to +302 °F)
100 mbar (1.5 psi) Gauge Ttotal = ±0.176 T = ±0.75 Calculation (mbar, bar or psi):
250 mbar (4 psi) pressure TTotal = ±0.352 1) T = ±1.25 1)
400 mbar (6 psi) 2 2
TDiff = (THP . URLHP) + (TLP . URLLP)
1 bar (15 psi) Gauge Ttotal = ±0.092 T = ±0.5
2 bar (30 psi) pressure Ttotal = ±0.184 1) T = ±0.75 1) 100 100
A0016474
4 bar (60 psi)
10 bar (150 psi) Percentage calculation of URL dP:
40 bar (600 psi)
TDiff . 100
1 bar (15 psi) Absolute Ttotal = ±0.092 T = ±0.75 TDiff [%]=
2 bar (30 psi) pressure Ttotal = ±0.184 1) T = ±1.25 1) PDiff
4 bar (60 psi) A0016475
Total performance The "Total performance" specification comprises the non-linearity including hysteresis, non-
reproducibility as well as the thermal change in the zero point. All specifications apply to the
temperature range –10 to +60 °C (+14 to +140 °F).
Measuring cell % of URL for every % of URL for every % of URL for every Calculated total performance (TPDiff)
sensor - standard sensor - high- sensor - hygienic of the differential pressure
version temperature version version
100 mbar (1.5 psi) TP = ±0.2 TP = ±0.46 TP = ±0.575 Calculation (mbar, bar or psi):
250 mbar (4 psi)
400 mbar (6 psi) 2 2
TPDiff = (TPHP. URLHP) + (TPLP. URLLP)
1 bar (15 psi) TP = ±0.15 TP = ±0.46 TP = ±0.5
2 bar (30 psi) 100 100
A0016470
4 bar (60 psi)
10 bar (150 psi) Percentage calculation of URL dP:
40 bar (600 psi)
TPDiff . 100
TPDiff [%]=
PDiff
A0016471
The "Applicator Sizing Electronic dp" selection tool, available free of charge on the Endress
+Hauser web site (www.endress.com/applicator), enables detailed calculations for your
respective applications.
Endress+Hauser 21
Deltabar FMD71, FMD72
100 mbar (1.5 psi) Gauge pressure L = ±0.1 L = ±0.2 Calculation (mbar, bar or psi):
250 mbar (4 psi) L = ±0.25 1) L = ±0.45 1)
400 mbar (6 psi) 2 2
Absolute pressure L = ±0.3 L Diff = (L HP . URLHP) + (L LP . URLLP)
L = ±0.55 1)
100 100
1 bar (15 psi) Gauge pressure L = ±0.05 L = ±0.2 A0016463
Total error The total error comprises the total performance and long-term stability. All specifications apply to
the temperature range –10 to +60 °C (+14 to +140 °F).
Measuring cell % of URL for every % of URL for every % of URL for every Calculated total error (TEDiff)
sensor - standard sensor - high- sensor - hygienic of the differential pressure
version temperature version version
100 mbar (1.5 psi) TE = ±0.25 TE = ±0.51 TE = ±0.925 Calculation (mbar, bar or psi):
250 mbar (4 psi)
400 mbar (6 psi) 2 2
TE Diff = (TE HP . URLHP) + (TE LP . URLLP)
1 bar (15 psi) TE = ±0.2 TE = ±0.51 TE = ±0.7
2 bar (30 psi) 100 100
A0016472
4 bar (60 psi)
10 bar (150 psi) Percentage calculation of URL dP:
40 bar (600 psi)
TE Diff . 100
TE Diff [%]=
PDiff
A0016473
22 Endress+Hauser
Deltabar FMD71, FMD72
90
°
A
°
90
B
A0016465
Process membrane axis is Process membrane pointing upwards Process membrane pointing
horizontal (A) (B) downwards (C)
Sensor with 1/2" thread and Calibration position, no < +4 mbar (+0.06 psi) < –4 mbar (–0.06 psi)
silicone oil measured error
Sensor with thread > 1/2" and < +10 mbar (+0.145 psi) < –10 mbar (–0.145 psi)
flanges The value is doubled for inert oil. The value is doubled for inert oil.
This effect can be corrected using the function to adjust the position (position adjustment) for the
differential pressure. Additional position adjustments for individual pressure signals are not
available.
A position-dependent zero shift can be corrected on the device .
IEC 61298-3 ≤ Reference accuracy up to 10 to 60 Hz: ±0.35 mm (±0.01 in); 60 to 500 Hz: 2 g
Application limits A high ratio between the level and head pressure or between the differential pressure and static
pressure can result in large measured errors. A maximum ratio of 1:10 is recommended. For
calculation purposes, please use the free "Applicator" calculation tool, which is available online at
"www.endress.com/applicator" or on CD-ROM.
Endress+Hauser 23
Deltabar FMD71, FMD72
Reference accuracy The reference accuracy contains the non-linearity [DIN EN 61298-2 3.11] including the pressure
hysteresis [DIN EN 61298-23.13] and non-repeatability [DIN EN 61298-2 3.11] in accordance with
the limit point method as per [DIN EN 60770].
Measuring cell Sensor Reference accuracy (A) Calculated reference accuracy (ADiff)
[%URL for every sensor] of the differential pressure
Standard Platinum
400 mbar (6 psi) Gauge pressure A= ±0.15 - Calculation (mbar, bar or psi):
±0.3 1)
2 2
1 bar (15 psi) Gauge pressure/ A= ±0.075 A= ±0.05 ADiff = (AHP. URLHP) + (ALP . URLLP)
2 bar (30 psi) Absolute pressure A= ±0.15 1) A= ±0.075 1)
ADiff . 100
ADiff [%]=
PDiff
A0016469
Ordering information
Designation Option 1)
Platinum D
Standard G
Measuring cell –10 to +60 °C (+14 to +140 °F) –40 to –10 °C (–40 to +14 °F) Calculated thermal change (TDiff)
+60 to +80 °C (+140 to +176 °F) of the differential pressure
400 mbar (6 psi) Ttotal = ±0.215 Ttotal = ±0.43 Calculation (mbar, bar or psi):
Tspan = ±0.2 Tspan = ±0.4
Tzero point = ±0.015 Tzero point = ±0.03 2 2
TDiff = (THP . URLHP) + (TLP . URLLP)
1 bar (15 psi) Ttotal = ±0.101 Ttotal = ±0.42
2 bar (30 psi) Tspan = ±0.1 Tspan = ±0.4 100 100
A0016474
4 bar (60 psi) Tzero point = ±0.01 Tzero point = ±0.02
10 bar (150 psi) Percentage calculation of URL dP:
40 bar (600 psi)
TDiff . 100
TDiff [%]=
PDiff
A0016475
24 Endress+Hauser
Deltabar FMD71, FMD72
Total performance The "Total performance" specification comprises the non-linearity including hysteresis, non-
reproducibility as well as the thermal change of the zero point. All specifications apply to the
temperature range –10 to +60 °C (+14 to +140 °F) .
Measuring cell % of URL for every sensor Calculated total performance (TPDiff) of the differential pressure
TPDiff . 100
TPDiff [%]=
PDiff
A0016471
The "Applicator Sizing Electronic dp" selection tool, available free of charge on the Endress
+Hauser web site (www.endress.com/applicator), enables detailed calculations for your
respective applications.
400 mbar (6 psi) L = ±0.035 L = ±0.14 L = ±0.32 Calculation (mbar, bar or psi):
L = ±0.25 1)
2 2
1 bar (15 psi) L = ±0.020 L = ±0.08 L = ±0.180 L Diff = (L HP . URLHP) + (L LP . URLLP)
L = ±0.1 1)
100 100
2 bar (30 psi) L = ±0.025 L = ±0.05 L = ±0.075 A0016463
Total error The total error comprises the total performance and long-term stability. All specifications apply to
the temperature range –10 to +60 °C (+14 to +140 °F).
Measuring cell % of URL/year for every sensor Calculated total error (TEDiff) of the differential pressure
TE Diff . 100
TE Diff [%]=
PDiff
A0016473
Endress+Hauser 25
Deltabar FMD71, FMD72
Installation
• Moisture must not penetrate the housing when mounting the device, establishing the electrical
connection and during operation.
• When measuring in media containing solids, such as dirty liquids, installing separators and drain
valves is useful for capturing and removing sediment.
• Point the cable and connector downwards where possible to prevent moisture from entering (e.g.
rain or condensation water).
Mounting location The FMD71/FMD72 is best suited to level measurement in vessels with pressure overlay or in
vacuum vessels and tanks, high distillation columns and other vessels with changing ambient
temperatures.
The sensor module HP is mounted on the lower measuring connection and the sensor module LP is
mounted above the maximum level. The transmitter can be mounted on pipes or walls with the
mounting bracket.
General installation Mounting the sensor modules and transmitter is very easy
instructions • The housings of the sensor modules can be rotated up to 360°.
• The transmitter is freely rotatable in the mounting bracket.
The sensor modules and transmitter can be easily aligned when mounted.
Your benefits
• Easy mounting due to optimum alignment of housing
• Easily accessible device operation
• Optimum readability of the onsite display (optional)
• Easy pipe installation due to optional alignment of the modules.
Thermal insulation - FMD71 The FMD71 high-temperature version may only be insulated up to a certain height. The maximum
high-temperature version permitted insulation height is indicated on the devices and applies to an insulation material with a
heat conductivity ≤ 0.04 W/(m x K) and to the maximum permitted ambient and process
temperature. The insulation height is not indicated on hygienic connections.
• Ambient temperature (TA): ≤ 70 °C (158 °F)
• Process temperature (TP): ≤ 150 °C (302 °F)
The data were determined under the most critical application "quiescent air".
A 1 2
B
A0021075
A Ambient temperature
B Process temperature
1 Insulation height
2 Insulation material
26 Endress+Hauser
Deltabar FMD71, FMD72
The mounting bracket can be installed on pipes with a diameter of 1¼" to 2" or on walls.
A0017514
Installing the transmitter The transmitter is installed with the mounting bracket supplied. The mounting bracket can be
installed on pipes with a diameter of 1¼" to 2" or on walls.
A0021145
Endress+Hauser 27
Deltabar FMD71, FMD72
28 Endress+Hauser
Deltabar FMD71, FMD72
Environment
Ambient temperature range • Without onsite display: –40 to +80 °C (–40 to +176 °F)
• With onsite display: –20 to +70 °C (–4 to +158 °F)
Extended temperature operation range with limitations in optical properties, such as display speed
and contrast. –40 to +80 °C (–40 to +176 °F)
For devices for use in hazardous areas, see Safety Instructions .
The device can be used in this temperature range. The values of the specification, such as thermal
change, may be exceeded.
Climate class Class 4K4H (air temperature: –20 to +55 °C (–4 to +131 °F), relative humidity: 4 to 100 %) fulfilled
as per DIN EN 60721-3-4 (condensation possible)
Electromagnetic • Electromagnetic compatibility as per EN 61326 Appendix A and NAMUR Recommendation EMC
compatibility (NE21). For details, refer to the Declaration of Conformity.
• Maximum deviation: < 0.5 % of span
• All EMC measurements were performed with a turn down (TD) = 2:1.
Endress+Hauser 29
Deltabar FMD71, FMD72
Process
Process temperature range • Thread and flanges: –25 to +125 °C (–13 to +257 °F)
for devices with ceramic • Hygienic connections:–25 to +130 °C (–13 to +266 °F) , 150 °C (302 °F) for max. 60 minutes
process membrane FMD71 • High-temperature version: –15 to +150 °C (+5 to +302 °F); see ordering information, order code
610, option "NB".
• For saturated steam applications, use a device with a metal process membrane, or provide a siphon
for temperature isolation when installing.
• Pay attention to the process temperature range of the seal. See also the following table.
1)
Seal Notes Process temperature range Option
FKM FDA 3), 3A Class I, USP –5 to +125 °C (+23 to +257 °F) –5 to +150 °C (+23 to +302 °F) B
Class VI
FFKM Perlast G75LT - –20 to +125 °C (–4 to +257 °F) –20 to +150 °C (–4 to +302 °F) C
The process temperature ranges indicated here refer to the permanent operation of the FMD71. In the case of devices with hygienic process
connections, a higher temperature (max. 150 °C (302 °F)) may be applied for a short period (max. 60 min.) for cleaning purposes.
Hygienic process connections with flush-mounted process –40 to +130 °C (–40 to +266 °F)
isolating diaphragm For a maximum of 60 minutes: +150 °C (+302 °F)
30 Endress+Hauser
Deltabar FMD71, FMD72
Endress+Hauser 31
Deltabar FMD71, FMD72
Mechanical construction
Device height The device height is calculated from
• the height of the housing and
• the height of the individual process connection.
The individual heights of the components are listed in the following sections. To calculate the device
height simply add up the individual heights of the components. Where applicable also take into
consideration the installation distance (space that is used to install the device). You can use the
following table for this purpose:
B
D
C
A0021292
Device height
32 Endress+Hauser
Deltabar FMD71, FMD72
220 (8.66)
A0016478
1)
Material Degree of protection Cable entry Weight kg (lbs) Option
Stainless steel IP66/68 NEMA 4x/6P • M20 2.6 (5.73) 2.5 (5.51) B
• G ½"
• NPT ½"
A0021293
1)
Material Degree of protection Cable entry Weight kg (lbs) Option
Endress+Hauser 33
Deltabar FMD71, FMD72
1
2
115 (4.53)
A0016480
1)
Material Degree of protection Cable entry Weight Option
kg (lbs)
Stainless steel IP66/68 NEMA 4x/6P • M20 cable entry 1.35 (2.98) B
• NPT ½"
Selecting the electrical There is a dependency between the cable entries of the transmitter housing and those of the sensor
connection module housing.
Different cable entries are available in the sensor module housing depending on the cable entry
selected for the transmitter housing (see the following table):
1)
Cable entry Transmitter housing Cable entry Sensor module housing Option
34 Endress+Hauser
Deltabar FMD71, FMD72
H
ø8 (0.31)
20 (0.79)
17 (0.67)
ø3 (0.12)
3 (0.12)
ø6 (0.24)
M20x1.5
A0020945
1)
Description Material Weight Option
kg (lbs)
Description Height H
Endress+Hauser 35
Deltabar FMD71, FMD72
H
ø8 (0.31)
20 (0.79)
13 (0.51)
17 (0.67)
20 (0.79)
17 (0.67)
ø3 (0.12) G ¼"
3 (0.12)
ø6 (0.24) ø17.5 (0.24)
G ½" G ½"
A0020935 A0020936
H
20 (0.79)
ø11.4 (0.45) 17 (0.67)
ø17.5 (0.24)
G ½"
A0020937
kg (lbs)
A Thread ISO 228 G ½" A EN 837 AISI 316L (CRN) 0.63 (1.39) CRN GCJ
PVDF - GCF
• Mount only with mounting bracket
• MWP 10 bar (150 psi), OPL max. 15 bar (225 psi)
• Process temperature range:
+10 to +60 °C (+14 to +140 °F)
B Thread ISO 228 G ½" A, G ¼" (female) AISI 316L (CRN) CRN GLJ
C Thread ISO 228 G ½" A, bore 11.4 mm (0.45 in) AISI 316L (CRN) CRN GMJ
Description Height H
36 Endress+Hauser
Deltabar FMD71, FMD72
H
25 (0.98)
ø11.4 (0.45)
25 (0.98)
NPT ¼"
NPT ½" NPT ½"
A0020939
A0020938
C D
H
25 (0.98)
ø3 (0.12) NPT ½"
25 (0.98)
NPT ½" ø26.5 (1.04)
A0020940 A0020943
kg (lbs)
A ANSI ½" MNPT, ¼" FNPT AISI 316L (CRN) 0.63 (1.39) CRN RLJ
B ANSI ½" MNPT, bore hole11.4 mm (0.45 in) AISI 316L (CRN) CRN RKJ
D ANSI ½" FNPT 11.4 mm (0.45 in) AISI 316L (CRN) CRN R1J
Description Height H
Endress+Hauser 37
Deltabar FMD71, FMD72
H
33.5 (1.32)
NPT 1½" NPT 2"
27 (1.06)
A0020948
A0020949
kg (lbs)
A ANSI 1 ½" MNPT AISI 316L (CRN) 0.63 (1.39) CRN U7J
38 Endress+Hauser
Deltabar FMD71, FMD72
H
ø40.5 (1.59)
22 (0.87)
11 (0.43)
M44x1.25
A0020950
kg (lbs)
Description Height H
Endress+Hauser 39
Deltabar FMD71, FMD72
H
33.5 (1.32)
G 1½" G 2"
25 (0.98)
30 (1.18)
28 (1.1)
ø55 (2.17) ø68 (2.68)
A0020946
A0020947
kg (lbs)
40 Endress+Hauser
Deltabar FMD71, FMD72
A B
13 (0.51)
72 (2.83)
72 (2.83)
20 (0.79)
ø61 (2.4) ø77 (3.03)
ø52 (2.05) ø52 (2.05)
A0026376
C D
72 (2.83)
12 (0.47)
49 (1.93)
42 (1.65)
ø42 (1.65)
20 (0.79)
12 (0.47)
ø38 (1.5)
A0026377 ø66 (2.6)
A0026378
kg (lbs)
3)
A Tri-Clamp ISO 2852 DN 40 – DN 51 (2"), DIN 32676 DN50, EHEDG, 3A (CRN) AISI 316L (1.4435) 0.7 (1.54) CRN TDJ
B Tri-Clamp ISO 2852 DN76.1 (3"), EHEDG, 3A, with FDA seal 0.9 (1.98) CRN TFJ
3)
C Varivent Type N for pipes 40 – 162, PN40, EHEDG, 3A (CRN) 1 (2.21) CRN TRJ
D Varivent Type F for pipes DN25-32 PN40, 316L, EHEDG, 3A, with FDA seal 0.46 (1) - TQJ
Endress+Hauser 41
Deltabar FMD71, FMD72
E F
72 (2.83)
49 (1.93)
8 (0.31)
ø61 (2.4) ø61 (2.4)
ø52 (2.05) ø52 (2.05)
42 (1.65)
ø48 (1.89)
20 (0.79)
11 (0.43)
ø42 (1.65)
ø68 (2.68)
10 (0.39)
A0026380
ø38 (1.5)
ø56 (2.2)
A0026379
G H
72 (2.83)
49 (1.93)
8 (0.31)
ø48 (1.89)
23 (0.91)
12 (0.47)
ø42 (1.65)
ø67 (2.64)
10 (0.39)
A0026386
ø38 (1.5)
ø55 (2.17)
A0026381
kg (lbs)
3)
E DIN 11851 DN40 PN25, EHEDG, 3A (CRN) AISI 316L (1.4435) 0.7 (1.54) CRN MZJ
3)
F DIN 11851 DN50 PN25, EHEDG, 3A (CRN) 0.9 (1.98) CRN MRJ
G DIN 11864-1 A DN40 PN16 pipe DIN11866-A, slotted nut, 316L, EHEDG, 3A 1 (2.21) - NCJ 3)
H DIN 11864-1 A DN50 PN40 pipe DIN11866-A, slotted nut, 316L, EHEDG, 3A 1 (2.21) - NDJ 3)
42 Endress+Hauser
Deltabar FMD71, FMD72
68 (2.68)
72 (2.83)
ø60 (2.36) ø61 (2.4)
ø52 (2.05) ø52 (2.05)
12 (0.47)
20 (0.79)
ø42 (1.65)
16 (0.63)
ø42 (1.65)
ø69 (2.72)
ø65 (2.56)
A0026387
A0026388
K ø90 (3.54)
ø70 (2.76)
ø52 (2.05)
39 (1.54)
ø38 (1.5)
17 (0.67)
53 (2.09)
ø50 (1.97)
A0026389
kg (lbs)
I APV inline DN50 PN25, 316L, 3A, with FDA seal AISI 316L (1.4435) 1.2 (2.65) CRN TMJ
J DRD DN50 (65 mm) PN25, slip-on flange AISI 304 (1.4301) 0.9 (1.98) - TIJ
Endress+Hauser 43
Deltabar FMD71, FMD72
1
2
68.5 (2.7)
3
6
4
5
dM = ø26 (1)
ø43.5 (1.71)
A0023744
• The roughness of the surface in contact with the medium is Ra < 0.76 µm (30 µin).
• Silicone molded seal: FDA 21CFR177.2600/USP Class VI, order number: 52023572
• EPDM molded seal: FDA, USP Class VI; 5 pcs, order number: 71100719
Material of the molded seal Material of the measuring cell seal Approval of measuring cell seal Option 1)
(exchangeable seal) on ceramic sensor (seal not
exchangeable)
Silicone EPDM FDA 2) 3A Class II, USP Class VI. DVGW, KTW, W270, WRAS, K
ACS, NSF61
44 Endress+Hauser
Deltabar FMD71, FMD72
62
b
26
2
g
k
D
A0026336
D Diameter of flange
b Thickness
g Raised face
k Pitch diameter
g2 Diameter of hole
Engineering unit mm
mm mm mm mm mm kg (lbs)
Endress+Hauser 45
Deltabar FMD71, FMD72
FMD71 process connections, ASME flanges, connection dimensions as per ASME B 16.5, raised face RF
flush membrane
g2
b
2.44
0.08
1.024
g
k
D
A0026337
D Diameter of flange
b Thickness
g Raised face
k Pitch diameter
g2 Diameter of hole
Engineering unit in
1)
Flange Boltholes Weight Approval 2) Option 3)
in lb./sq.in in in in in in kg (lbs)
4)
1 150 AISI 316/316L 4.25 1.18 2 4 0.62 3.12 0.9 (1.98) - ACJ
4)
1 300 AISI 316/316L 4.88 1.18 2 4 0.75 3.5 1.4 (3.09) - ANJ
4)
1½ 150 AISI 316/316L 5 0.69 2.88 4 0.62 3.88 2.1 (4.63) CRN AEJ
4)
1½ 300 AISI 316/316L 6.12 0.81 2.88 4 0.88 4.5 2.6 (5.73) CRN AQJ
4)
2 150 AISI 316/316L 6 0.75 3.62 4 0.75 4.75 3.0 (6.62) CRN AFJ
5)
2 150 ECTFE 6 0.75 3.62 4 0.75 4.75 2.4 (5.29) CRN AFN
1) AISI 316L
2) CSA approval: Product Configurator order code for "Approval"
3) Product Configurator order code for "Process connection"
4) Combination of AISI 316 for required pressure resistance and AISI 316L for required chemical resistance (dual rated)
5) ECTFE coating on AISI 316/316L. When using in hazardous areas: avoid electrostatic charge on the plastic surfaces.
46 Endress+Hauser
Deltabar FMD71, FMD72
FMD71 process connections, JIS flanges, connection dimensions in accordance with JIS B 2220 BL, raised face RF
flush-mounted process
isolating diaphragm
g2
62
b
26
2
g
k
D
A0026336
D Diameter of flange
b Thickness
g Raised face
k Hole circle
g2 Diameter of hole
Engineering unit mm
1) 2)
Flange Boltholes Weight Option 3)
mm mm mm mm mm kg (lbs)
Endress+Hauser 47
Deltabar FMD71, FMD72
20 (0.79)
20 (0.79)
38 (1.5)
38 (1.5)
ø11.4 (0.45)
13 (0.51)
17 (0.67)
3 (0.12)
3 (0.12)
ø3.2 (0.13)
ø6 (0.24) G ¼"
ø17.5 (0.69) ø17.5 (0.69)
G ½" A G ½" A
A0026391 A0026392
20 (0.79)
38 (1.5)
ø11.4 (0.45)
17 (0.67)
ø17.5 (0.69)
G ½" A
A0026393
kg (lbs)
A Thread ISO 228 G ½" A EN 837 AISI 316L 0.63 (1.39) GCJ
B Thread ISO 228 G ½" A, G ¼" (female) EN 837 AISI 316L GLJ
C Thread ISO 228 G ½" A EN 837, bore 11.4 mm (0.45 in) AISI 316L GMJ
48 Endress+Hauser
Deltabar FMD71, FMD72
38 (1.5)
38 (1.5)
ø21.4 (0.84) ø21.4 (0.84)
NPT ¼"
25 (0.98)
15 (0.59)
25 (0.98)
ø11.4 (0.45)
NPT ½"
NPT ½"
A0026394 A0026395
C
26 (1.02)
7 (0.28)
10 (0.39)
20 (0.79)
NPT ½"
8 (0.31)
ø38 (1.5)
A0026396
kg (lbs)
A ANSI ½" MNPT, ¼" FNPT AISI 316L (CRN) 0.63 (1.39) CRN RLJ
B ANSI ½" MNPT, bore hole11.4 mm (0.45 in) AISI 316L (CRN) CRN RKJ
D ANSI ½" FNPT 11.4 mm (0.45 in) AISI 316L (CRN) CRN R1J
Endress+Hauser 49
Deltabar FMD71, FMD72
A B
3 (0.12)
38 (1.5)
3 (0.12)
34 (1.34)
1
ø18.4 (0.72)
27
3.5 (0.14)
18 (0.71)
ø28 (1.1)
2 (0.08)
3.5 (0.14)
ø30 (1.18)
14 (0.55)
ø18 (0.71)
2 (0.08)
G ½" A G 1" A
ø23.6 (0.93) ø36 (1.42)
ø26.5 (1.04) ø39 (1.54)
A0021613 A0021614
C D
25 (0.98)
1 (0.04)
24 (0.94)
5 (0.2)
ø26 (1.02)
30 (1.18)
ø26 (1.02)
26 (1.02)
2 (0.08)
2 (0.08)
kg (lbs)
A Thread ISO 228 G ½" A, DIN 3852 FKM gasket seal (item 1) pre-installed AISI 316L 0.4 (0.88) GRJ
50 Endress+Hauser
Deltabar FMD71, FMD72
24 (0.94)
28 (1.1)
23 (0.91)
27 (1.06)
ø26 (1.02)
ø26 (1.02)
NPT 1"
NPT 1½"
A0026400
A0026402
24 (0.94)
31 (1.22)
ø26 (1.02)
NPT 2"
A0026403
kg (lbs)
A ANSI 1" MNPT AISI 316L (CRN) 0.7 (1.54) CRN U5J
C ANSI 2" MNPT AISI 316L (CRN) 1.3 (2.87) CRN U8J
Endress+Hauser 51
Deltabar FMD71, FMD72
FMD72 process connections, EN/DIN flanges, connection dimensions in accordance with EN 1092-1/DIN 2527
flush-mounted process
isolating diaphragm
g2
35
dM = ø28
b
2
60
g
k
D
A0026405
D Diameter of flange
b Thickness
g Raised face
k Hole circle
g2 Diameter of hole
dM Max. diameter of the process isolating diaphragm
Engineering unit mm
1) 2)
Flange Boltholes Option 3)
1) The roughness of the surface in contact with the medium including the raised face of the flanges (all standards) is <Ra 0.8 μm (31.5 μin). Lower
surface roughness on request.
2) AISI 316L
3) Product Configurator, order code for "Process connection"
4) Description as per DIN 2527 provided in brackets
5) The sealing surface in these process connections is smaller than what is specified in the standard. Due to the smaller sealing surface, a special
seal must be used. Contact a seal manufacturer or your local Endress+Hauser Sales Center.
52 Endress+Hauser
Deltabar FMD71, FMD72
FMD72 process connections, ASME flanges, connection dimensions as per ASME B 16.5, raised face RF
flush membrane
g2
1.42
b
dM = ø 1.1
0.08
2.36
g
k
D
A0026406
D Diameter of flange
b Thickness
g Raised face
k Pitch diameter
g2 Diameter of hole
dM Max. diameter of process membrane
Engineering unit in
1) 2)
Flange Boltholes Approval 3) Option 4)
1 300 4.88 0.69 0.06 6) 1.3 (2.87) 4 0.75 3.5 CRN ANJ
6)
1½ 150 5 0.69 2.88 1.5 (3.31) 4 0.62 3.88 CRN AEJ
6)
1½ 300 6.12 0.81 2.88 2.6 (5.73) 4 0.88 4.5 CRN AQJ
1) The roughness of the surface in contact with the medium is <Ra 0.8 μm (31.5 μin). Lower surface roughness on request.
2) Material AISI 316/316L (combination of AISI 316 for required pressure resistance and AISI 316L for required chemical resistance (dual rated))
3) CSA approval: Product Configurator order code for "Approval"
4) Product Configurator order code for "Process connection"
5) Screws must be 15 mm (0.59 in) longer than the standard flange screws.
6) The sealing surface in the case of these process connections is smaller than specified in the standard. Due to the smaller sealing surface, a special
seal must be used. Please contact a seal manufacturer or your local Endress+Hauser sales organization.
Endress+Hauser 53
Deltabar FMD71, FMD72
A B
H
dM dM
d d
D1 D1
A0021801
A0021802
H Height H Height
D1 Diameter D1 Diameter
d Diameter of seal d Diameter of seal
dM Max. diameter of process membrane dM Max. diameter of process membrane
C D
H
H
m
m
dM dM
D1 D1
D3 D3
A0021803
A0021804
H Height H Height
m Height m Height
D1 Diameter D1 Diameter
D3 Diameter D3 Diameter
dM Max. diameter of process membrane dM Max. diameter of process membrane
PN kg (lbs)
A Clamp DN18-22, 40 34 (1.34) - 27.5 17.2 max. 40 - AISI 316L 0.5 (1.10) TBJ
3A (1.08) (0.68) (1.57) (1.4435)
1) The roughness of the surface in contact with the medium is Ra <0.76 µm (30 µin).
2) With CRN approval
54 Endress+Hauser
Deltabar FMD71, FMD72
E F
H
H
m
m
dM dM
D1 D1
D3 D3
A0021805 A0021806
H Height H Height
m Height m Height
D1 Diameter D1 Diameter
D3 Diameter D3 Diameter
dM Max. diameter of process membrane dM Max. diameter of process membrane
G H
H
H
dM dM
D2 D2
D1 D1
A0021800 A0021807
H Height H Height
D1 Diameter D1 Diameter
D2 Diameter D2 Diameter
dM Max. diameter of process membrane dM Max. diameter of process membrane
PN kg (lbs)
E SMS 1", 3A 25 35.5 (1.4) - 51 (2.01) 21.65 max. 40 20 (0.79) AISI 316L 0.7 (1.54) T6J
(0.85) (1.57) (1.4435)
1) The roughness of the surface in contact with the medium is Ra <0.76 µm (30 µin).
Endress+Hauser 55
Deltabar FMD71, FMD72
I R J
2 (0.08)
H
H
dM
D2
k
dM D1
D2 A0021809
k H Height
D1 D1 Diameter
A0021808
k Pitch diameter
H Height D2 Diameter
D1 Diameter dM Max. diameter of process membrane
k Pitch diameter
D2 Diameter
dM Max. diameter of process membrane
K L
H
H
15,5 (0.6)
dM dM
D2 D2
D1 k
k D1
105 (4.13)
A0021811
A0021810
H Height
H Height D1 Diameter
D1 Diameter k Pitch diameter
k Pitch diameter D2 Diameter
D2 Diameter dM Max. diameter of process membrane
dM Max. diameter of process membrane
PN kg (lbs)
I Neumo D25, 3A 16 64 (2.52) 30.4 50 (1.97); 4 x, R 3.5 21.65 max. 40 AISI 316L 0.8 (1.76) S1J
(1.2) mm (0.14 in) (0.85) (1.57) (1.4435)
L APV Inline 25 99.5 64 (2.52) 82 (3.23); 6 x ø 8.6 28 (1.10) 1.2 (2.65) TMJ
(3.92) mm (0.34 in) + 2 x
M8
1) The roughness of the surface in contact with the medium is Ra <0.76 µm (30 µin).
56 Endress+Hauser
Deltabar FMD71, FMD72
131 (5.16)
52 (2.05)
ø6,4 (0.25)
ø52 (2.05)
90 (3.54)
90 (3.54)
ø40 (1.57)
140 (5.51)
ø6 (0.24)
122 (4.8) 175 (6.89)
86 (3.39)
70 (2.76)
122 (4.8)
158 (6.22)
ø42...60 (1.65...2.36)
A0016495
→ 33 0.5 (1.1) PA
Endress+Hauser 57
Deltabar FMD71, FMD72
Process connections
• Endress+Hauser supplies DIN/EN flanges and threaded connection in stainless steel as per AISI
316L (DIN/ EN material number 1.4404 or 1.4435). With regard to their stability-temperature
property, the materials 1.4404 and 1.4435 are grouped together under 13E0 in EN 1092-1: 2001
Tab. 18. The chemical composition of the two materials can be identical.
• "Clamp connections" and "Hygienic process connections": AISI 316L (DIN/EN material number
1.4435)
• Some process connections are also available in alloy C276 (DIN/EN material number 2.4819). For
this purpose see the information in the "Mechanical construction" section.
58 Endress+Hauser
Deltabar FMD71, FMD72
3 15 10
2 6 14 11
1 5 13
4 12
A0016496
1 T14 housing, RAL 5012 (blue) Die-cast aluminum with protective powder-coating on
polyester base
11 Screw A4
13 Round strand rope for nameplates AISI 304 (1.4301) / AISI 316 (1.4401)
Endress+Hauser 59
Deltabar FMD71, FMD72
T17 1
2
16 16 3
15
17
7
8
14 9 5 4 3
6 2
13
12 11 10
A0020021
Cover
4 Nameplates Lasered on
10 Screw A2-70
13 Screw A4-50
Thread coating: Heat-curing lubricant
varnish
60 Endress+Hauser
Deltabar FMD71, FMD72
Sensor modules
3
5
2 4
6
7
1
A0021295
1 Sensor module housing and cover Aluminum with protective powder-coating on polyester base
3 Blind plug PBT-GF30 FR, for dust ignition-proof: AISI 316L (1.4435)
Connecting parts
1
4 2
4
6 3
5
A0016497
Endress+Hauser 61
Deltabar FMD71, FMD72
Filling oil
Description Option 1)
silicone oil 1
62 Endress+Hauser
Deltabar FMD71, FMD72
Operability
Operation concept Operator-oriented menu structure for user-specific tasks
• Commissioning
• Software operation
• Diagnostics
• Expert level
Reliable operation
• Local operation possible in up to two languages
• Standardized operation at the device and in the operating tools
• Parameters relating to measured values can be locked/unlocked using the device's write protection
switch, using the device software or via remote operation
Device reset
Endress+Hauser 63
Deltabar FMD71, FMD72
Functions:
• 8-digit measured value display including sign and decimal point, bargraph for 4 to 20 mA HART as
current display.
• Simple and complete menu guidance due to breakdown of parameters into several levels and
groups.
• Each parameter is given a 3-digit ID number for easy navigation.
• Option for configuring the display according to individual requirements and preferences, such as
language, alternating display, display of other measured values such as sensor temperature,
contrast setting.
• Comprehensive diagnostic functions (fault and warning message, peak-hold indicators, etc.).
• Quick and safe commissioning
Overview
4 5
3
2
– + E
1
A0016498
1 Operating keys
2 Bargraph
3 Symbol
4 Header
5 Parameter ID number
A0016499
The operating keys on the outside of the device make it unnecessary to open the housing. This
guarantees:
• Complete protection against environmental influences such as moisture and contamination
• Simple operation without any tools
• No wear.
64 Endress+Hauser
Deltabar FMD71, FMD72
6 7 8
SW/min
1 2 3
on Display
t 4 5 off
on R
SW /min
HART
damp.
FIELD COMMUNICATION PROTOCOL
off
Sensor
A0016500
Operating languages You can also choose another language in addition to the standard language "English":
1)
Description Version
English (standard) AA
German AB
French AC
Spanish AD
Italian AE
Portuguese AG
Chinese simplified AK
Japanese AL
Endress+Hauser 65
Deltabar FMD71, FMD72
RoHS FMD72:
The measuring system complies with the substance restrictions of the Restriction on Hazardous
Substances Directive 2011/65/EU (RoHS 2).
RCM-Tick marking The supplied product or measuring system meets the ACMA (Australian Communications and Media
Authority) requirements for network integrity, interoperability, performance characteristics as well
as health and safety regulations. Here, especially the regulatory arrangements for electromagnetic
compatibility are met. The products are labelled with the RCM- Tick marking on the name plate.
A0029561
Ex approvals • ATEX
• FM
• CSA
• IECEx
• NEPSI
All explosion protection data are given in separate documentation which is available upon request.
The Ex documentation is supplied as standard with all Ex-systems .
Suitable for hygiene For information on installation and approvals, see documentation SD02503F "Hygiene approvals".
applications
For information on 3-A and EHEDG-tested adapters, see documentation TI00426F "Weld-in adapter,
process adapter and flanges".
Pressure Equipment Pressure equipment with allowable pressure ≤ 200 bar (2 900 psi)
Directive 2014/68/EU (PED)
Pressure equipment (having a maximum allowable pressure PS ≤ 200 bar (2 900 psi)) can be
classified as pressurized equipment in accordance with Pressure Equipment Directive 2014/68/EU. If
the maximum allowable pressure is ≤ 200 bar (2 900 psi) and the pressurized volume of the pressure
equipment is ≤ 0.1 l, the pressure equipment is subject to the Pressure Equipment Directive (cf.
Pressure Equipment Directive 2014/68/EU, Article 4, point 3). The Pressure Equipment Directive
only requires that the pressure equipment shall be designed and manufactured in accordance with
the "sound engineering practice of a Member State".
Reasons:
• Pressure Equipment Directive (PED) 2014/68/EU Article 4, point 3
• Pressure equipment directive 2014/68/EU, Commission's Working Group "Pressure", Guideline
A-05 + A-06
Note:
A partial examination shall be performed for pressure instruments that are part of safety equipment
for the protection of a pipe or vessel from exceeding allowable limits (equipment with safety
function in accordance with Pressure Equipment Directive 2014/68/EU, Article 2, point 4).
Manufacturer declaration Depending on the desired configuration, the following documents can be ordered additionally with
the device:
• FDA conformity
• TSE-free, materials free from animal origin
• Regulation (EC) No. 2023/2006 (GMP)
• Regulation (EC) No. 1935/2004 on materials and articles intended to come into contact with food
66 Endress+Hauser
Deltabar FMD71, FMD72
Other standards and The applicable European guidelines and standards can be found in the relevant EU Declarations of
guidelines Conformity. The following were also applied:
EN 60529:
Degrees of protection by housing (IP code)
DIN 16086:
Electrical pressure measuring instruments, pressure sensors, pressure transmitters, pressure
measuring instruments, concepts, specifications in data sheets
Procedure for writing specifications in data sheets for electrical pressure measuring instruments,
pressure sensors and pressure transmitters.
EN 61326-X:
EMC product family standard for electrical equipment for measurement, control and laboratory use.
CRN approvals Some device versions have CRN approval. A CRN-approved process connection with a CSA approval
must be ordered for a CRN-approved device. The CRN-approved devices are assigned a registration
number.
Ordering information: Product Configurator, order code for "Process connection" (the CRN process
connections are indicated appropriately in the "Mechanical construction" section.)
Classification of process Endress+Hauser devices are designed in accordance with ANSI/ISA 12.27.01. allowing the user to
sealing between electrical waive the use and save the cost of installing external secondary process seals in the conduit as
systems and (flammable or required by the process sealing sections of ANSI/NFPA 70 (NEC) and CSA 22.1 (CEC). These
combustible) process fluids instruments comply with the North-American installation practice and provide a very safe and cost-
in accordance with ANSI/ISA saving installation for pressurized applications with hazardous fluids. Please refer to the following
12.27.01 table for the seal class assigned (single seal or dual seal):
Further information can be found in the control drawings of the relevant devices.
Endress+Hauser 67
Deltabar FMD71, FMD72
Level range; % A
PMI test (XRF), internal procedure, metal parts in contact with the medium, inspection certificate KG
68 Endress+Hauser
Deltabar FMD71, FMD72
Ordering information
Detailed ordering information is available as follows:
• In the Product Configurator on the Endress+Hauser website:www.endress.com → Click "Corporate"
→ Select your country → Click "Products" → Select the product using the filters and search field →
Open product page → The "Configure" button to the right of the product image opens the Product
Configurator.
• From your Endress+Hauser Sales Center:www.addresses.endress.com
Product Configurator - the tool for individual product configuration
• Up-to-the-minute configuration data
• Depending on the device: direct input of information specific to the measuring point, such as
measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic generation of the order code with its breakdown in PDF or Excel output format
• Ability to order directly in the Endress+Hauser Online Shop
Endress+Hauser 69
Deltabar FMD71, FMD72
Documentation
Field of Activities Pressure measurement, powerful instruments for process pressure, differential pressure, level and
flow:
FA00004P/00/EN
Safety Instructions (XA) Depending on the approval, Safety Instructions (XA) are supplied with the device. The Safety
Instructions are an integral part of the Operating Instructions.
Information on the Safety Instructions (XA) that are relevant for the device is provided on the
nameplate.
70 Endress+Hauser
Deltabar FMD71, FMD72
Accessories
Service-specific accessories Accessories Description
DeviceCare SFE100 Configuration tool for HART, PROFIBUS and FOUNDATION Fieldbus field devices
Field Xpert SMT70, The Field Xpert SMT70 tablet PC for device configuration enables mobile plant
SMT77 asset management in hazardous (Ex Zone 2) and non-hazardous areas. It is
suitable for commissioning and maintenance staff. It manages Endress+Hauser
and third-party field instruments with a digital communication interface and
documents the progress of the work. The SMT70 is designed as a complete
solution. It comes with a pre-installed driver library and is an easy-to-use, touch-
enabled tool for managing field devices throughout their entire life cycle.
The Field Xpert SMT77 for device configuration enables mobile plant asset
management in areas categorized as Ex Zone 1. It is suitable for commissioning
and maintenance staff for easy management of field instruments with a digital
communication interface. The touch-enabled tablet PC is designed as a complete
solution. It comes with comprehensive pre-installed driver libraries and offers users
a modern software user interface to manage field instruments throughout the
entire life cycle.