Plant Procedure
Calcium Based Deposit Determination
Preliminary Test
SCOPE
HANDLING & SAFETY
• This test procedure is designed to determined the presence of calcium based
deposit on typical deposit sample. • As the procedure includes
the use of mildly corrosive
EQUIPMENTS & APPARATUS REQUIRED chemicals, normal care in
the handling of these
• Pipettes for adding acid chemicals must be
• Evaporating dish adhered to. Minimum
requirements are:
• Drying oven
• Wear safety glasses, PVC
gloves and a PVC apron
CHEMICALS REQUIRED or safety overalls.
• HCl 1:1 • Ensure safety shower
including eye wash station
MAIN PROCEDURE is accessible and located
close to work area.
• Collect deposit samples in an evaporating dish.
• For complete safe
• Put the samples into drying oven, raise the temperatures to 60 to 110 oC to remove handling instructions refer
moisture or evaporate liquid for 10 – 15 minutes. to SDS provided by the
chemical supplier for each
• Take out the samples from the oven and let it reach room temperature.
chemical.
• Using the pipette, add several drops of HCl and observe closely for gas release from
the samples.
• Gas release is indication of the presence of calcium based deposit in the sample.

All statements, information and data presented herein are believed to be accurate and reliable but are not to be taken as a guarantee, express warranty or implied warranty of merchantability
or fitness for a particular purpose, or representation, express or implied, for which seller assumes legal responsibility, and they are offered solely for your consideration, investigation and
verification. Statements or suggestions concerning possible use of this product are made without representation or warranty that any such use is free of patent infringement and are not
recommendations to infringe on any patent.