You are on page 1of 62

Humalyzer Primus

| User Manual

Cat No. 18200100


REVISION LIST OF THE MANUAL
Rev. /DATE. REVISION DESCRIPTION
01/2010-01 First edition
02/2010-07 Correction of typing errors
03/2010-11 Removal of software version statement
04/2011-09 Correction dimension
05/2013-08 Correction of specifications, delivery, minor errors
06/2021-03 QR code for software and settings updates added in chapter 6.5;
optimizing description of Two-Point and Kinetic method in chapter
3.5.1.3 Editing Test Parameters

SYSTEM VERSION

COPYRIGHT

Copyright 2021, Human Gesellschaft für Biochemica und Diagnostica mbH, Wiesbaden,
Germany. All rights reserved.

No part of this documentation may be reproduced in any form, nor processed, copied or
distributed by means of electronic systems, without prior permission of HUMAN in writing.
Since all precautionary measures were taken into account in producing these operating
instructions, the manufacturer accepts no responsibility for any errors or omissions. This
includes any liability for damage that could arise from possible incorrect operation based on this
information. Subject to changes without notice as result of technical development.

SERVICE AND SUPPORT


CONTENTS

TABLE OF CONTENTS

1 SAFETY INSTRUCTIONS 5
1.1 INTRODUCTION 5
1.2 USER WARRANTY 5
1.3 INTENDED USE OF THE INSTRUMENT 5
1.4 GENERAL SAFETY WARNINGS 6
1.5 DISPOSAL MANAGEMENT CONCEPT 6
1.6 BIOHAZARD WARNING 7
1.7 INSTRUMENT DISINFECTION 7

2 SYSTEM DESCRIPTION 9
2.1 UNPACKING 9
2.2 PRODUCT FEATURES 9
2.3 PRODUCT STRUCTURE AND COMPOSITION 10
2.4 INSTALLATION 10
2.4.1 Selecting a proper location for the Instrument 10
2.4.2 Power supply requirement 10
2.4.3 Connecting the instrument to power supply 10
2.4.4 Installing the flow cell 11
2.5 MAIN STRUCTURE 12
2.5.1 Front view 12
2.5.2 Keypad 13
2.5.3 Back view 14
2.5.4 Internal printer 14
2.5.5 Flow cell and cuvette 14
2.5.6 Lamp and peristaltic pump 15
2.5.7 Schematic system diagram (for reference only) 16

3 OPERATION 17
3.1 POWERING ON 17
3.2 INTRODUCTION TO MAIN MENUS FUNCTIONS 17
3.3 SOFTWARE USAGE 18
3.4 SYSTEM SETTINGS 18
3.5 SAMPLE TEST 21
3.5.1 Test parameter settings 22
3.5.2 Item test 31
4 REPORT 41
4.1 SAMPLE REPORT 42
4.2 ITEM REPORT 43
4.3 QC REPORT 45
4.4 DELETING RECORDS 46

5 POWER OFF 47

6 MAINTENANCE 49
6.1 OVERVIEW 49
6.2 INSTRUMENT CLEANING 49
6.2.1 Cleaning the surface of the instrument 49
6.2.2 Cleaning the flow cell 49
6.3 INSTRUMENT MAINTENANCE 50
6.3.1 Changing the fuse 50
6.3.2 Adjusting the peristaltic pump tube 50
6.3.3 Changing the aspiration tube 51
6.4 BASIC TROUBLESHOOTING 51
6.5 HUMAN SOFTWARE AND SETTINGS UPDATE 53

7 APPENDIX 55
7.1 TECHNICAL SPECIFICATION 55
CONTENTS
Safety Instructions 5

1 SAFETY INSTRUCTIONS

1.1 Introduction
This manual is considered part of the instrument and must be available to the
operator and the maintenance personnel. For accurate installation, use and
maintenance, please read the following instructions carefully.
In order to avoid damage to the instrument or personal injury, carefully read
the General safety warnings, describing the appropriate operating procedures.
Please contact your HUMAN authorised local Technical Service in the event of
instrument failure or other difficulties with the instrument.

1.2 User warranty


HUMAN warrants that instruments sold by one of its authorised representatives
shall be free of any defect in material or workmanship, provided that this
warranty shall apply only to defects which become apparent within one year
from the date of delivery of the new instrument to the purchaser. The HUMAN
representative shall replace or repair any defective item within this warranty
period at no charge, except for transportation expenses to the point of repair.

This warranty excludes the HUMAN representative from liability to replace


any item considered as expendable in the course of normal usage, e.g.: Lamps,
valves, syringes, glassware, fuses, tubing etc. The HUMAN representative
shall be relieved of any liability under this warranty if the product is not used
in accordance with the manufacturer‘s instructions, altered in any way not
specified by HUMAN, not regularly maintained, used with equipment not
approved by HUMAN or used for purposes for which it was not designed.

1.3 Intended use of the instrument


The instrument must be used for its intended purpose (see paragraph 2). It must
[IVD]
be operated in perfect technical conditions, by qualified personnel, in such
working conditions and maintained as described in this manual, in the General
safety warnings. This manual contains instructions for qualified professional
operators.
6

1.4 General safety warnings


Use only chemical reagents and accessories specified and supplied by
HUMAN and/or mentioned in this manual. Place the product so that it has
proper ventilation. The instrument should be installed on a flat, stationary
working surface, that is free of vibrations.
Do not operate in area with excessive dust. Operate at temperature and at a
humidity level in accordance with the specifications listed in this manual. Do not
operate this instrument with covers and panels removed. Use only the power
cord specified for this product, with the grounding conductor of the power cord
connected to earth ground.
Use only the fuse type and rating specified by the manufacturer for this
instrument. The use of fuses with improper ratings may pose electrical and fire
hazards. To avoid fire or shock hazard, observe all ratings and markings on the
instrument. Do not power the instrument in environments that are potentially
explosive or at risk of fire.
Prior to cleaning and/or performing maintenance on the instrument, switch
off the instrument and remove the power cord. Only cleaning materials
described in this manual may be used, as other materials may damage parts. It
is recommended to always wear protective clothing and eye protection while
using this instrument. All warning symbols that appear in this manual must be
carefully observed.

1.5 Disposal management concept


The applicable local regulations governing disposal must be observed. It
is the user‘s responsibility to arrange for proper disposal of the individual
components.
All parts which may contain potentially infectious materials must be
disinfected by suitable, validated procedures (autoclaving, chemical treatment)
prior to disposal. Applicable local regulations for disposal must be carefully
observed. The instruments and electronic accessories (without batteries, power
packs etc.) must be disposed of according to the applicable local regulations
for the disposal of electronic components.
Batteries, power packs and similar power sources must be removed from
electric/electronic parts and disposed of in accordance with applicable local
regulations.

HumaLyzer Primus | User manual


Safety Instructions 7

1.6 Biohazard warning


Analytical instruments for in vitro diagnostic application involve the handling of
human samples and controls which should be considered at least potentially
infectious. Therefore every part and accessory of the respective instrument
which may have come into contact with such samples must equally
be considered as potentially infectious. The „BIOHAZARD“ warning label must
be affixed to the instrument prior to first use with biological material!

Figure 1
Biological hazard
symbol

1.7 Instrument disinfection


Before performing any servicing on the instrument it is very important to
thoroughly disinfect all possibly contaminated parts. Before the instrument is
removed from the laboratory for disposal or servicing, it must be
decontaminated. Decontamination must be performed by authorised well-
trained personnel, and in observance of all necessary safety precautions.
8

HumaLyzer Primus | User manual


SYSTEM DESCRIPTION 9

2 SYSTEM DESCRIPTION
The product is intended for the in vitro diagnostic [IVD] clinical testing of serum,
plasma, urine and CSF samples. The HumaLyzer Primus is a semi-automated
Chemistry Analyzer.

2.1 Unpacking
Remove all the parts from their package. When unpacking the instrument,
please make sure that the following items are conained in the packing. In case
of damage or missing item, please contact the supplier immediately.

HumaLyzer Primus 1 [REF]:18200


Flow Cell 1 [REF]:18200/14
User Manual 1 [REF]:18200/100
Power Cord 1 [REF]:18200/75
Spare Fuse 2 [REF]:16670/38
Aspiration Tube 1 [REF]:18200/18
Waste tube [REF]:18200/19
Pump Tube 1 [REF]:18200/20
Lamp assembly 1 [REF]:18200/13
Dust Cover 1 ---
Waste Bottle 1 [REF]:18200/192
Cable Tie 4 ---
Printer Paper 2 rolls [REF]:18144/5
RS-232 Cable 1 [REF]:16670/43

2.2 Product features


Monochrome LCD, English menu (Spanish, Russian and French software available
on request). Two test modes for single wavelength and double wavelength.
The following test methods are available:
- End-point
- Fixed-time
- Kinetic
- Bichromatic
- Absorbance
- Multi-point calibration
- It can perform sample analysis of serum, plasma, whole blood and urine;
- Compatible with reagents without any specific restrictions;
- Internal printer can print reports and programming parameters, etc.
10

2.3 Product structure and composition


The product is composed of an optical system, a sipper system, a computer
control system and software.

2.4 Installation

2.4.1 Selecting a proper location for the Instrument


Find a place that is not exposed to direct sunshine. The surface of the selected
installation location should be flat and provide sufficient space to accommodate
the HumaLyzer Primus. The front edge of the instrument should be near the
edge of the working table. Avoid significant vibrations on the working table (for
example, do not place a centrifuge on the working table).

! Note: The temperature of


the instruments operating
environment must be between
In order to ensure proper functioning of the instrument, it may not be placed in
the following areas:
- Areas subject to extreme temperature changes;
10…30°C with a relative humidity - Very hot or cold areas;
below 70%. - Very dusty areas;
- Areas near electromagnetic devices that produce a magnetic field;
- Areas exposed to direct sunshine.

2.4.2 Power supply requirement


- 110/ 220Vac ± 10%
- 50/ 60 ± 1Hz
- 80VA

2.4.3 Connecting the instrument to power supply


1. Insert one end of the power cable into the instrument’s power plug.
2. Insert the other end of the power cable into the power supply’s outlet.

HumaLyzer Primus | User manual


SYSTEM DESCRIPTION 11

! Warning:
AC power must be well earthed (near-ground voltage <5V). The AC power
supply must be stable. High-power equipment may not be used on the same
power supply. A voltage-regulated power supply is preferable. In areas with
unstable voltage, please use a UPS power supply to ensure proper functioning
of the instrument. When removing the power cable, pull on the plug instead of
on the power cable. In the event that smoke, abnormal smell or unusual sound
emanate from the instrument, immediately switch off the power and contact
the distributor.

2.4.4 Installing the flow cell


Open the cover of the light source as shown in the following Figure 2.

Figure 2
Opening the lamp cover

1. Remove the adhesive tape used to secure the flow cell.


2. Remove the wrapping materials.
3. Place the flow cell into thermostatic bath.
4. Carefully insert the flow cell. Run the aspiration tube through the fixing hole
on the machine; allow it to extend a few centimetres.
5. Close the lamp cover.
12

! Warning:
Do not touch the surface of the light inlet window on the flow cell by hand to
avoid transferring grease from your hands onto the window, which will affect
transmittance. Ensure that the aspiration tube is in the socket. Otherwise, it will
cause pressure build-up and bend the tube, which will affect the test results.
This can easily cause significant bubble formation and increase the water blank.

2.5 Main structure


This part presents the structure of all major components of the HumaLyzer
Primus.

2.5.1 Front view

Figure 3
Front view
1

1 Print cover
2 Lamp cover 2

3 Aspiration tube
4 Aspiration key
5 Keypad 3

6 Power supply indicator lamp 4

7 LCD

5 6 7

HumaLyzer Primus | User manual


SYSTEM DESCRIPTION 13

2.5.2 Keypad

Figure 4
Keypad

Direction keys: Move the cursor in four directions on the


screen: up, down, left and right.

Paper feed key: Manually feed paper; is also the shortcut key
for parameter printing.
Rinse key: Clean the pipeline.

Exit key: Exit the menu and cancel the operation.

Enter key: Confirm the input and save.


14

2.5.3 Back view

Figure 5
1 2 3 4
Back view

1 Parallel port (not used)


2 RS-232 serial port
3 RS-232 serial port
4 Drain port
5 Power switch
6 Fuse
7 AC input
8 Fan
5 6 7 8 9 10
9 Backplate label
10 Peristaltic pump

2.5.4 Internal printer


HumaLyzer Primus has a built-in thermal printer.

2.5.5 Flow cell and cuvette

! Note: To open the light source


cover, first press the knobs on
both sides of the cover firmly in-
1. Open the lamp cover as shown in the following Figure 6.

ward and ensure that they recess,


then open the cover in the direc-
tion shown in the figure above.

Figure 6
Opening the lamp cover

HumaLyzer Primus | User manual


SYSTEM DESCRIPTION 15

2. Internal components of the colorimetric system

Figure 7
Internal components of the
colorimetric system

1 1 Flow Cell
2 2 Thermostatic bath
3 Aspiration key

• Thermostatic bath
• Flow cell

3. Using the cuvette system


• Remove flow cell from thermostatic bath.
• Place the cuvette with samples into the thermostatic bath.
• All test methods are the same as those in a test using the flow cell sys-
tem.

2.5.6 Lamp and peristaltic pump


1. Open the lamp cover.
2. When changing the lamp, remove the connecting line for the lamp bracket
and power supply, unscrew the fixing screws on the lamp bracket, remove
the lamp bracket, replace the old lamp with a new one and place it in the
same position, then tightly screw in the tightening screws on the lamp
bracket. Reconnect the connecting line for the light bracket and the power
supply firmly.
16

3. Lamp and peristaltic pump as shown in the following figure:

2 4 5

Figure 8
Light source system 1

1 Fixing screw
2 Adjusting screws 1

3 Lens
4 Lamp bracket 3
5 Lamp socket

Figure 9
5
Peristaltic pump system 1
6
2
1 Peristaltic pump head
2 Pump tube
3 7
3 Water inlet
4 Water outlet 4
8
5 Fixing block
6 Adjustment screw
7 Pump installation board
8 Pump generator

2.5.7 Schematic system diagram (for reference only)


Keyboard
Figure 10 
Schematic system diagram Optic and Signal A/D LCD
 colorimetric  amplification  transform  Computer  display
system

Printer

HumaLyzer Primus | User manual


OPERATION 17

3 OPERATION

3.1 Powering on
Flip the power switch on the back of the instrument. After several seconds, a
menu will appear in the screen as shown in Figure 11:
Figure 11
Temperature control

After the temperature control process runs for approximately three minutes (if
the Esc key is pressed on the keypad, the system will stop temperature control
and directly enter the software main menu), as shown in Figure 12.

Figure 12
Main menu

3.2 Introduction to main menus functions


1. Test
Edit parameter settings, calibration settings and quality control settings
for the test item selected. It will save the parameters automatically after
confirmation and start the test process to perform the test on that item.

2. Report
Perform data administration for the test system. The functions include
printing by test and patient number, deleting the test results, and searching
the quality control information for the current month.
18

3. Sys. Setup
Set up the basic system functions. It includes printer settings, power on and
off, time settings, and an external administration system to transmit all test
data from the current day.

4. Power off
This will request to wash the tubes before power is switched OFF. Users can
turn off the power supply directly to prevent tubes from being washed (it is
recommended that tubes be washed before shutdown).

3.3 Software usage


The number in front of the item name corresponds to the number key on the
keypad, for example, if you press number key 1 on the keypad in the main menu,
the system will enter the test menu); ◄► to the right of the name indicates
that the left and right keys on the keypad can be used to change parameter
settings.

 in the menu indicates that the up and down keys on the keypad can be

used to select the desired menu item. Also date and time as well as the current
software version is displayed in the lower part of the display.

3.4 System settings


System parameters are used to set the fundamental parameters of the
instrument such as date and time, etc. Press key 3 and enter the system settings
menu as shown in Figure 13. The ID: Indicates the instrument serial number.

Figure 13
System settings menu

HumaLyzer Primus | User manual


OPERATION 19

1. Printer settings
Press key 1 in the system settings menu. The instrument will enter the
following menu as shown in Figure 14.

Figure 14
Printer settings

The instrument is equipped with an internal thermal printer. It can be


activated or deactivated using the ◄► key on the keypad. Instant printing
can be reactivated during the test procedure (please refer to the printer
settings for detailed settings during the test procedure).

2. Time settings
In the system settings menu, press key 2. The instrument will enter the
following menu as shown in Figure 15.

Figure 15
Time settings

Use the ◄► key on the keypad to select the item you wish to change, press
the appropriate number keys to change the current date or time, press the
Enter key to save and press the Esc key to return to the system settings menu.

3. Data transfer
In the system settings menu, press key 3 (before this operation, ensure that
the communication cable is connected between the instrument and the
external PC and that the serial interface on the external PC is ready to receive
data. The instrument will enter the following menu as shown in Figure 16.
When the menu shown in Figure 17 appears, it indicates that data transfer is
complete. A special software for receiving data is required.
20

Figure 16
Data transferring

Figure 17
Data transfer finished

4. Optics settings
In the system settings menu, press key 4 to enter the optics settings menu as
shown in Figure 18. Wavelength X corresponds to the installation position of
the corresponding filters as shown in Figure 19.

Figure 18
AD (analog to digital)
display menu

Figure 19
Layout of filter wheel (facing the
installation side)

HumaLyzer Primus | User manual


OPERATION 21

1. Checking the AD (analog to digital) value


Press the Rinse key to aspirate a certain volume of distilled water (the sipper
system must be filled). Press the left or right key ◄► to switch between
wavelengths and check the AD value of the current wavelength.

2. Saving the AD value


Because the AD value will be taken as a fiducial value for calculation, it
cannot be saved arbitrarily. You only need to save the AD again after filter
replacement or if the system displays a constant Retest prompt during a
water blank test.

First, press the Rinse key to aspirate distilled water (the pipeline must be filled)
then press the left or right key ◄► to switch the wavelength. After the stable
AD value has been determined, press the Enter key to save the AD value for the
current wavelength. The Save? prompt will disappear

! Note:
This operation will affect test performance; it is only used to determine
whether to replace the lamp or if Redo is shown repeatedly during water blanking.
Ensure sufficient water is present in the flow cell during this operation; aspirate
using the Rinse key as appropriate. Pressing the Enter key will only store the AD
value for the current filter; other filters will remain unchanged.

3.5 Sample test


The sample test includes parameter settings and the selected test. Parameter
settings include the test method, test name, unit, temperature, wavelength,
blank, delay time, test time, and aspirate volume, range of the normal reference
value, factor, standard and quality control, etc. The test includes a blank test,
standard test, quality control test and sample test, etc.
22

3.5.1 Test parameter settings

3.5.1.1 Selecting test item


Figure 20 shows the main menu for test selection. The first line is the
status column, and the following three lines indicate the test name and its
corresponding serial number in sequence. You can use the direction keys,
number keys or name search to select tests. The details are as follows:

Figure 20
Test selection menu

1. Select using direction keys


Select tests using the up, down, left or right keys. The selected test will be
highlighted and its number will be shown at the top-left of the screen.

2. Select by name search


The input methods are displayed at the top-right of screen as 123 and ABC.
Use the “ . “ key to switch between the two. 123 indicates numerical input,
which allows numbers to be entered directly; the test will be automatically
selected according to its serial number and it will be highlighted on the
display. ABC indicates character input, which allows characters to be entered
as indicated on the keypad; the characters will be displayed at the middle-
top of the screen. After you finish entering the search term, press the Enter
key to automatically select the corresponding test (there is no response if
the search fails to find a corresponding test).

HumaLyzer Primus | User manual


OPERATION 23

3.5.1.2 Confirming test parameters


Select a test and press the Enter key to enter the parameters confirmation menu
test as shown in Figure 21, Press the ◄► key to switch between QC, STD, Edit
and OK, then press the Enter key to enter the corresponding programmes.

Figure 21
Parameters confirmation menu

The meaning of each parameter is as follows:

1st row: Test name, Test method, Unit, Temperature


2nd row: First/ Second wavelength, Blank type, Factor, Aspirate vol.,
Decimal digits
3rd row: Delay Time, Test Time, Norms high, Norms Low
4 row:
th
QC settings, Standard settings, Parameter settings, Enter test
menu

Press the ◄► key to switch between QC, STD, Edit and OK, then press the Enter
key to enter the corresponding programmes.
24

3.5.1.3 Editing test parameters


In the parameters confirmation menu, select Edit, press the Enter key, and
the parameter setting for the corresponding test will be opened for editing as
shown in Figure 22.

Figure 22
Parameter
editing menus

Basic procedure:
1. Select the test which you wish to edit; move the cursor by pressing 
 on the
keypad; the selected test will be highlighted on the display.

2. To edit the settings


If the right side of the selected test displays test ◄► , it can be edited using
the ◄► key. If the right side displays ► , it indicates that the next editing
menu can be entered using the ► key. If the selected test has a decimal point,
it indicates that decimal points may be entered; otherwise, only integers can
be entered.

3. Quick page turning


If the current page does not need to be edited, press the Enter key to move
to the next page or test process.

4. Test parameter print


In the test parameter settings menu press Paper to print this test parameter
once the parameter has been set.

5. Press the Enter key on the last page of the test settings to automatically save
the parameter settings and enter the test process.

HumaLyzer Primus | User manual


OPERATION 25

Set all parameters:


1. Test name
The system allows 60 tests to be set, including approximately 39 fixed tests.
Users can define the remaining tests.

2. Select test method


End-point method, two-points method or kinetics method. Choose the
method using the ◄► key.
• End-point method: In the end-point method, the reaction reaches the
end point after a period of time by mixing reagents and sample. The end
point is indicated when, for example, the reaction fluid no longer chang-
es colour or absorbance. At that time, the absorbance value of the reac-
tion fluid and concentration of the reaction solution can be obtained on
the basis of the Lambert-Beer Law. The kits for some tests require that
tests use dual wavelength, which allows some interference factors to be
eliminated. The instrument supports the dual wavelength test method.

• Kinetics method: In the kinetics method, the instrument continuously


measures the change of absorbance of the reaction fluid after mixing
reagents and sample. The change of absorbance should be measured in
the linear segment of the reaction curve only. The delay time is used to
exclude the non-linear segment of the reaction curve, no measurement
points are taken here. During the measuring time, the instrument records
one measurement point per second and calculates the average change of
absorption (ABS/min). The concentration result can be obtained on the
basis of the Lambert-Beer Law by multiplication of the average ABS/min
with the calibration factor.

• Two-points method: The two-points method test is a special type of the


kinetics method test. In this method, the difference in the absorbance
value at two points in the linear section is measured once the absorb-
ance value begins to change linearly. One measurement point is taken
after the delay time and at the beginning of the measuring time, and the
second point at the end of the measuring time. This allows the concen-
tration of the reaction solution to be obtained.
26

3. Unit
Select the unit for the test result. The following 9 units are available: mg/
dl, mg/l, g/l, µmol/l, mmol/l, mol/l, U/l, IU/l and blank. Select the unit using
the ◄► key.

4. Temperature
Select test temperature, 4 settings are available: ambient temperature, 25°C,
30°C and 37°C. Choose the temperature using the ◄► key.

5. First wavelength
Choose the required wavelength in the test according to the kit requirements.
5 wavelengths are available: 340nm, 405nm, 500nm, 546nm and 620nm.
Choose the required wavelength using the ◄► key.

6. Second wavelength
Select the secondary wavelength in the test according to the kit
requirements, and the system will flexibly implement the dual wavelength
method by setting the secondary wavelength. It includes 6 selectable
secondary wavelengths: 340nm, 405nm, 500nm, 546nm, 620nm and none.
When a single wavelength is used for the test, the secondary wavelength
must be set as none. Choose the secondary wavelength using the ◄► key.

7. Blank
Select the blank testing, including none, reagent blank and sample blank.
Choose the blank type using the ◄► key. In the end-point method
test, reagent blank and sample blank can be selected according to user
requirements. Reagent blank can be selected for the two-points method test,
and no blank settings can be selected for the kinetics method.

8. Factor
Enter the calculation factor; negative and positive values can be switched
using the ◄► key. The system supports the direct input of factors and the
determination of calculation factors by system calibration (standard test).

9. Norms high
The entered result is the value of norms high based on empirical evidence.

10. Norms low


The entered result is the value of norms low based on empirical evidence.

HumaLyzer Primus | User manual


OPERATION 27

11. Delay time


Enter the delay time for the test process. The delay time is the time between
aspirating the sample to the flow cell and the actual test start.

! Note:
A certain time delay is necessary in order to ensure that the test samples are
at a stable temperature and to eliminate bubbles in the flow cell. The time delay
is 5 seconds for most end-point method tests. The time delay should not be less
than 10 seconds on kinetics method tests below 25°C. The time delay should
not be less than 12 seconds on kinetics method tests below 30°C. The time delay
should not be less than 15 seconds on kinetics method tests below 37°C.

12. Test time


Enter the reaction time for the test procedure (the end-point method does
not require this parameter).

13. Decimal digits


The reserved digits after the decimal point in the result; the maximum is 4.

14. Aspirate volume


The amount of solution aspirated by the peristaltic pump. This parameter
indicates the amount of imbibed fluid which will be tested each time. In
order to ensure test accuracy, the aspirate volume is generally greater than
400µl. (General it should be set at 500µl; the aspirate volume of reagents
with significant contamination can be increased to 700µl).

After setting all the parameters, press the Enter key to automatically save the
parameters and return to the test parameters confirmation menu. Press the Esc
key if you wish to return without saving. Therefore, correct standard parameters
! Note: Generally, an aspirate
volume of 500µl is sufficient
to ensure a cross-contamination
must be entered. ratio below 1%. The aspirate vol-
ume can be increased or reduced
by number entry on the keypad.
28

3.5.1.4 Standard settings


In the parameters confirmation menu, select STD and press the Enter key to
enter the settings menu for standard parameters according to the indications
as shown in Figure 23.

Method
The calculation method for standard tests is linear regression or non-linear
regression. Select using the ◄► key.

No. of Std
The number of standard preparations used in the standard test process. The
maximum is 8 for calibration.

Repeat Times
The number of times the test is repeated for each standard preparation.

Concentration
Standard preparation concentration used in the standard test process.

Figure 23
Standard parameter setup menu

! Note:
When non-linear regression is selected, the number of standards should be
more than 3; otherwise, calibrate according to the linear regression method.
When more than two standards are selected, the number of repetitions is
changed to a single measurement for all standards time.

When more than three standard concentrations are entered, they must be
entered using the progressive or degressive method; otherwise, the calibration
will fail. When it processes standard parameters, the system will indicate the
third started standard concentration range as shown in Figure 24. The system
will evaluate the progressive or degressive tendency automatically in sequence,
and the third started concentration range will be indicated at the right side.

HumaLyzer Primus | User manual


OPERATION 29

Figure 24
Standard concentration menu

After setting all the parameters, press the Enter key to automatically save the
parameters and return to the item parameters confirmation menu. Press the Esc
key if you wish to return without saving. Therefore, correct standard parameters
must be entered.

3.5.1.5 QC settings
In the parameters confirmation menu, select QC and press the Enter key to enter
the settings menu for quality control parameters according to the indications as
shown in Figure 25.

Control
The system can set two quality control preparations. Select using the ◄► key
(in the settings menu for quality control parameters, press the ◄► key to select
the quality control preparations in cycle).

Average Value
Standard concentration of the quality control preparation.

SD
Standard deviation of the quality control preparation.

Batch number
Corresponding batch number for quality control.

Figure 25
Quality control parameters menu

After setting all the parameters, press the Enter key to automatically save the
parameters and return to the menu for item parameters confirmation. Press the
30

Esc key if you wish to return without saving. Therefore, correct quality control
parameters must be entered.

3.5.1.6 Parameter print


Under the menu for parameters confirmation or the menu for parameter
settings as shown in Figure 21 or Figure 22, press the Paper key to print the
parameters of the current item. After you finish setting and checking all the
parameters, select OK in the menu for parameters confirmation and press the
Enter key to enter the item test process. It must therefore be ensured that the
correct parameters are entered. Pressing the Esc key will return to the item
selection menu.

3.5.1.7 Item user-definition


The instrument can save 60 test items, including approximately 39 fixed items
and 21 user-defined items; user-defined items can also be implemented directly
on the basis of the 47 fixed items. The specific user-defined methods are as
follows:
1. Move the cursor to the Name column for parameter settings. The item name
is highlighted or displayed as a cursor symbol (no item name).
2. Directly enter the item name according to the character index on the key-
pad; the name is limited to a maximum of 5 characters. As you enter the
name, the system will automatically determine characters at a 3-second in-
terval each, and will indicate when to enter the next character.
3. If necessary, press the Del key to delete the characters you have entered.
4. Set item parameters, then press Enter to automatically save the user-de-
fined item name and parameters and enter the test process.

HumaLyzer Primus | User manual


OPERATION 31

3.5.2 Item test


After setting test parameters, press the Enter key to automatically save the
parameter settings and enter the test process. First, the system will switch
the light filter according to the set wavelength and will load the required
information. The information output is as shown in Figure 26. The switching
time is very short (typically few seconds). Temperature control is then started
automatically.

Figure 26
Test switch indication

After finishing the test switch and information loading, the system will enter
temperature control as shown in figure Figure 27. The temperature control
time is set on the basis of the temperature difference before and after the item
switch and typically lasts from several seconds to three minutes. Press the Esc
key if you wish to cancel temperature control and enter the blank test.

Figure 27
Temperature control state
32

3.5.2.1 Blank test


After finishing the test switch and temperature control, the system automatically
enters the blank test process. Based on the parameters’ settings, the system
will select whether to perform a reagent blank. When the blank settings in the
parameter settings are set to perform no blank or a sample blank, the blank test
will simply be performed as a water blank. When a reagent blank is selected, the
blank test will be performed as a water blank and a reagent blank.

Figure 28
Blank test

As shown in Figure 28, a water blank is performed first and Aspirate Please
is shown in the water blank column. Prepare distilled water and press the
aspiration key. The system will automatically perform the test and output the
results of the water blank.

When the value of the water blank is too high, the display will show Aspirate
Please Retest and the real-time absorbance value will be shown below as shown
in Figure 29. In this case, the water blank must be repeated. If Aspirate Please
Retest is still shown after several repetitions, please check whether the light
path is blocked.

Figure 29
Water blank abnormality

HumaLyzer Primus | User manual


OPERATION 33

After testing the water blank, the system will automatically enter the reagent
blank, aspirate the reagent according to the indications, test the system, and
output the results. After testing the reagent blank, no key needs to be pressed,
as the system will automatically enter the test selection menu and wait for
several seconds, as shown in Figure 30.

The cursor points to Test sample with no selection; select the test sample using
the 
 key. Press the Enter key to enter the corresponding test process, and press
the Esc key to return to this test selection menu for the other tests.

Figure 30
Test selection menu

3.5.2.2 Standard test


The system can obtain the standard factors through two methods. One method
is the entry of factors when setting parameters. K=xxx will then be displayed in
the test selection menu as shown in Figure 30. The other method is obtaining
the standard factors from a standard test. This section will explain the basic
process of the standard test in detail.

1. Test selection
In the test selection menu, move the cursor to Test Standard; the text will
then be highlighted. The system determines whether the settings for the
standard parameters are correct. If the settings are wrong, the system will
display None!, which indicates that the standard test is either unnecessary
or incorrect as shown in Figure 31.
34

If the settings are correct, it displays OK? as shown in Figure 32. Confirm
whether the standard test is to be performed. If it should be performed,
please press the Enter key to enter the test procedure. Otherwise, move the
cursor to select other tests or press the Esc key to return to the main menu.

Figure 31
Standard test selection

Standard parameter has errors

Figure 32

Standard parameter is correct

2. Test operation
After entering the standard test procedure, the standard test menu will
appear as shown in Figure 33. Test displays the item name for the current
test. STD displays the serial number of the current standard fluid. The
corresponding concentration value is shown on the right side. For example,
with Con.: 133.00 as shown, please ensure that solution of that concentration
is introduced before aspirating the standard fluid.

HumaLyzer Primus | User manual


OPERATION 35

Status shows the test process schedule and test results. Aspirate Please,
Testing… and Save? are displayed in sequence in the lower-middle area of
the menu. The specific test procedure is the same as the item test process;
please refer to the specific operation of the item test.

Figure 33
Standard test menu

3. Test saving
After aspirating the corresponding standard fluid according to the procedure,
the system will calculate the standard factors in real time. The K and b
calibration values will be displayed in the Result column, and Save? will
be shown in the system’s status column. Press the Enter key to save the
calibration result.

All subsequent item tests will be calculated on the basis of this calibration
result until the standard factors are reset or recalibrated and saved as shown
in Figure 34. Press the Esc key to cancel saving and directly enter the item
test process. In this case, all the subsequent item tests will be calculated
according to the previous calibration settings. If recalibration is required,
please select Standard Test in the test selection menu to perform calibration
again as shown in Figure 30.

Figure 34
Standard test saving
36

3.5.2.3 Quality control


1. Test selection
In the test selection menu, move the cursor to Test QC. The text will then be
highlighted. On the right side, the system displays that the quality control
preparation can be selected. Select the preparation using the ◄► key on the
extreme right side. The system supports single quality control and double
quality control.

This is selected using the ◄► key, which highlights the selected quality
control preparation as shown in Figure 35. To select double quality control,
press and hold the ◄► key until two of the quality control preparations are
highlighted together as shown in figure Figure 36. Confirm that the selected
quality control preparation is correct and press the Enter key to enter the
menu for the quality control test as shown in Figure 37.

Figure 35
Quality control test selection

Select single quality control

Figure 36

Select double quality control

HumaLyzer Primus | User manual


OPERATION 37

2. Test operation
After entering the process for the quality control test, the menu for the
quality control test will appear as shown in Figure 37. Test displays the item
name of the current test. QC shows the serial number of the current quality
control solution. The corresponding quality control batch number is shown
on the right side.

For example, with Batch: 070691, please ensure that quality control
preparations with the same batch number are introduced before aspirating
the quality control solution. Status displays the progress of the test process
and the test results. Aspirate Please, Testing… and Save? are displayed in
sequence in the lower-middle of the menu. The specific test procedure is the
same as the item test process; Please refer to the specific operation of the
item test.
Figure 37
Quality control test menu

3. Test saving
After aspirating the corresponding quality control solution according to the
procedure, the system will calculate the results in real time. Quality control
results are displayed, and quality control conditions are evaluated in the
Result column (C>2SD indicates that the quality control result is larger than
the value of two SD; otherwise, this information is not displayed), and Save?
will be shown in the system’s status column.

After confirming that quality control was successful, press the Enter key to
save the quality control results. When double quality control is selected, test
the two quality control preparations in sequence. Aspirate the corresponding
quality control solution according to the indications and save the valid
quality control results (only one quality control result for the same item and
same batch number is saved each day) as shown in Figure 38.
38

Press the Esc key to repeat quality control. After the quality control test is
complete, the system will automatically skip to the test selection menu and
point to the item test as shown in Figure 30.

Figure 38
Saving quality control test

3.5.2.4 Sample test


In the test selection menu, the cursor points to Sample test with no selection
and Sample Test is highlighted. Press the Enter key to enter the sample test
procedure.

1. Test operation
When sample blank is selected in the parameter settings, a sample blank is
performed first before the sample test as shown in Figure 39. When no blank
or reagent blank is selected, this means it was already finished during the
previous blank test, and the system will directly enter test menu as shown
in Figure 40. Blank and Test will display the item name for the current test.

>Print On or >Print Off on the upper-right corner of the window allow the online
print function to be switched on or off, respectively, using the ► key. If this
function is switched on, the results are printed immediately after the test
is complete. It can only be changed while the system is waiting for sample
aspiration.

Sample displays the next sample number to be tested and can be changed
before the test. A change will be indicated by highlighting Change No. on the
right side of the corresponding column. It is advisable to check whether the
number is correct before starting the test, as the previous test result would
otherwise be automatically overlapped by the new test result with the same
item and same number.

HumaLyzer Primus | User manual


OPERATION 39

Status displays the progress of the test process, and Result indicates the test
absorbance and concentration.

Aspirate Please, Testing… and Printing… are displayed in sequence in the lower-
middle part of the status column. Please aspirate the sample or wait for the
test according to the indications. When Overlapped? appears after testing, it
indicates that this item already has a test result with the same sample number
on the same day. Press the Enter key if you want to overlap the existing result. If
you want to save the result as a new sample, enter a new sample number then
press the Enter key.

Figure 39
Sample test menu

Sample blank
Figure 40

Sample test

2. Test process
The procedure and related operations for this test process are the same as
for no sample blank (no blank or reagent blank setting) and detailed in that
section. If sample blank is selected, a sample blank will be performed before
each item test and a sample test will be automatically entered to wait for
the sample to be aspirated. Its operation is completely the same as that of
the sample test.
• Confirm the sample number (the sample number displayed is the num-
ber of the sample to be tested) and activate or deactivate the online print
function using the ► key as shown in Figure 39 and Figure 40.
40

• When the system displays Aspirate Please, prepare the test sample and
press the aspiration key to imbibe the sample preparation. The system
will imbibe the fluid according to the set aspirate volume and will auto-
matically enter the test process. Testing... will be displayed in the status
column, and the test progress will be indicated in seconds in the Status
column as shown in Figure 41.

Figure 41

• After the test is complete, the test results are automatically saved (if this
item already has a test result with the same sample number on the same
day, Overlapped? will appear and the system will wait for a response.
Press the Enter key if you want to overlap the existing result. If you want
to save the result as a new sample, enter a new sample number then
press the Enter key). The sample number will be increased by one for the
next test sample and can be modified by the user.
• Print the results: If the online print function is available, activate or deac-
tivate it using the ► key while the system is waiting for fluid aspiration.
• When the current sample test and print are complete, the sample num-
ber will be increased by one and the system will wait for fluid aspiration
for the next sample test.
• If operated according to instructions, the system can continuously test
samples. If quality control, calibration or cancellation are required during
the test procedure, press the Esc key to return to the test selection menu
and select the relevant operations.

HumaLyzer Primus | User manual


REPORT 41

4 REPORT
The report functions deal mainly with data from test results and quality control
tests for the current month. They also provide basic functions for data search,
print and delete, etc. In the system’s main menu, press the 2 key to enter the
administration menu as shown in Figure 42. The available functions are sample
report, item report and control report. The corresponding function modules are
accessed using number keys 1, 2 and 3 respectively.

Sample report
Search the test results according to the date and sample number.

Item report
Search the test results according to the date and item number.

QC report
Search the quality control results according to the month, item number and
batch number from quality control.

Figure 42
Administration function selection

! Note:
The HumaLyzer Primus has storage capacity for 2200 test results. The system
can automatically manage its storage. When the quantity of test data exceeds
the systems storage capacity, the system will automatically delete parts of old
data. The system will also issue corresponding notifications during this process.
Please wait a moment while the system clears old data, after which it will
automatically return to normal operation and can continue to perform tests.
42

4.1 Sample report


Enter the time range and sample number to be searched for in the test results.
Enter the start date and end date, respectively. The end date must be on or later
than the original day. Enter the sample number you wish to search for in Sample
No. “ - ” stands for number A to number B; B must not be less than A. When A is
equal to B, a single sample will be searched. As shown in Figure 43, the system
will search the test results of samples 1, 2 and 3 tested on November 9, 2009.
Figure 43
Sample report search

After entering the time range and sample number, press the Enter key to search
for records which meet the search criteria. If no matching record is found,
None will be displayed at the lower-right of the window as shown in Figure
44. If matching records are found, all records which satisfy the criteria will be
displayed as shown in Figure 45.

Figure 44
Sample search criteria

No record

Figure 45

Condition satisfied records

HumaLyzer Primus | User manual


REPORT 43

Several shortcut keys are defined in the display window for search results as
shown in Figure 45:
- Page up/ down
! Note: The „remark“ in the dis-
play and printed report is not
displayed if status is normal; „H“
Only 3 records can be displayed per page; press the 
 key to display more indicates the test result is above
records. the maximum reference value;
- Print „L“ indicates that the test result
Press the Enter key to print all search results. is below the minimum reference
- Delete value.
Press the Del key to delete all search results.
- Return
Press the Esc key to return to the search window.

4.2 Item report


Enter the time range and item number to be searched for in the test results.
Enter the start date and end date, respectively. The end date must be later than
the original date or on the same day as the start date. Enter the item number
you wish to search for in Item No.

The item name corresponding to the number will be displayed on the right side
of the screen in real time. Alternately, you may enter characters in character
mode when ABC appears on the upper-right side of the screen. The “ . “ key is
used to switch between number and character input mode. As shown in Figure
46, the system will search for all results for the ALT item tested on November 9,
2009.

Figure 46

After entering the time range and item number, press the Enter key to search
for records which meet the search criteria. If no matching record is found,
None will be displayed at the lower-right of the window as shown in Figure
47. If matching records are found, all records which satisfy the criteria will be
displayed as shown in Figure 48.
44

Figure 47
Item search condition

No record

Figure 48

Records matching criteria

! Note: The „remark“ in the dis-


play and printed report is not
displayed if status is normal; „H“
Several shortcut keys are defined in the display window for search results as
shown in Figure 48:
- Page up/ down
indicates the test result is above Only 3 records can be displayed per page; press the 
 key to display more
the maximum reference value; records.
„L“ indicates that the test result - Print
is below the minimum reference Press the Enter key to print all search results.
value. - Delete
Press the Del key to delete all search results.
- Return
Press the Esc key to return to the search window.

HumaLyzer Primus | User manual


REPORT 45

4.3 QC Report
Enter the month, item number and quality control batch number to search for in
quality control results. Enter the test date, and enter the item number you wish
to search for in Item No.. The item name corresponding to the number will be
displayed on the right side in real time (if necessary, switch the character input
mode using the “ . ” key). Enter the quality control batch number in “Batch No”
as shown in Figure 49. The system will search It will search for quality control
results for the ALT item with batch number 070691 in November 2009.

Figure 49
Control report search

After entering the month, item number and quality control batch number, press
the Enter key to search for records which meet the criteria. If no matching record
is found, None will be displayed at the lower-right of the window as shown in
Figure 50. If matching records are found, all records which satisfy the criteria
will be displayed as shown in Figure 51.

Figure 50
Quality control search condition

No record
46

Figure 51

Records meet conditions

Several shortcut keys are defined in the display window for search results as
shown in Figure 51:
- Page up/ down
Only 3 records can be displayed per page; press the 
 key to display more

! Note: The „remark“ in display


and print report: Normal con-
dition is no display. „H“---Test
records.
- Print
Press the Enter key to print all search results.
result higher than 2SD, „L“ ---Test - Delete
result lower than 2SD. Press the Del key to delete all search results.
- Return
Press the Esc key to return to the search window.

4.4 Deleting records


The delete command can be used to clear all test results and/ or quality control
results stored on the device as shown in Figure 52.

Figure 52
Quality control search

This will delete all test results or quality control results. Individual records can be
deleted as described in sections 4.2 and 4.3.

HumaLyzer Primus | User manual


POWER OFF 47

5 POWER OFF
After completing all tests, return to the main menu. Press key 4 to execute the
power-off program. When the screen displays as in Figure 53, press the Enter
key (if you press the Esc key, the system will not power off and you will return
to the main menu).

Figure 53
Confirmation of power off

When the system displays as in Figure 54, it indicates that the system needs to
aspirate distilled water to clean the flow cell.

Figure 54
Cleaning the flow cell

Prepare a sufficient amount of distilled water and press the Aspiration key or
the Rinse key to clean the flow cell as shown in Figure 55.

Figure 55
Indication of cleaning procedure

After the flow cell has been cleaned, the system displays Power off please! as
shown in Figure 56.

Figure 56
Power off
48

HumaLyzer Primus | User manual


MAINTENANCE 49

6 MAINTENANCE

6.1 Overview
HumaLyzer Primus is a clinical precision analyzer. Daily maintenance work
must be performed properly to keep the instrument in good working condition.
Maintenance work on the HumaLyzer Primus is very easy to perform, but it must
be done thoroughly and carefully.

6.2 Instrument cleaning

6.2.1 Cleaning the surface of the instrument


- Keep the instrument’s working environment clean;
- The surface of the instrument may be cleaned using a neutral cleaner and a
! Note: Do not let solvent, liq-
uids or corrosive materials
contact the instrument.
wet cloth;
- Use only a soft cloth to clean the liquid crystal display.

6.2.2 Cleaning the flow cell


The flow cell must be kept clean in order to ensure the accuracy and reliability
of test results.
1. Cleaning the exterior of the flow cell
• The flow cell must be installed according to instructions (see section
2.4.4).
• If the exterior of the flow cell is dirty, wipe clean using a soft cloth mois-
tened with dehydrated alcohol.

2. Cleaning the interior of the flow cell


• Place the container with distilled water under the aspiration tube, press
the Rinse key, and start the continuous washing function. Press the Rinse
key again to finish the washing procedure. 30 seconds of continuous
washing are typically sufficient.
• Flow Cell Cleaner 18222 is used to clean the flow cell. Press the Rinse key
to aspirate the cleaner then press the Rinse key again to stop the rotation
of peristaltic pump. Leave the cleaner in the flow cell for 5 minutes, then
wash continuously with distilled water for 1 minute. Cleaner can be used
to wash again if the flow cell is still not clean.
50

! Note: Do not leave reaction


fluid or other contaminants in
the flow cell for an extended peri-
The flow cell must be cleaned in the following cases:
- The value of the water blank is too high;
- Test items have been changed (use water only);
od of time. - Before power off (see above).

! Note: Please refer to the material safety datasheet of the Flow Cell Cleaner
([REF] 18222) for precautionary statements.

6.3 Instrument maintenance

6.3.1 Changing the fuse

! Note: A fuse of the type in-


dicated above must be used.
The user must ensure that the
The fuse is installed in the fuse box beside the power switch on the rear of the
instrument. Open the box cap to conveniently access and change the fuse.
Fuse specification: 3.15AL250V.
power cable is disconnected while Users are recommended to use a voltage-regulated power supply.
changing the fuse.

6.3.2 Adjusting the peristaltic pump tube

! Note: The instrument will not


aspirate if the pump tube is
not installed properly.
After the instrument has been in use for six months, the position of peristaltic
pump may have to be adjusted.
Procedure:
- Rotate the lock catch on the peristaltic pump tube clockwise, and open the
back plate on the pump tube.
- Remove the peristaltic pump tube.
- Loosen the fixing steel wire on the pump tube connecter, rotate the pump
tube 180°, and then retighten the steel wires.
- Install the pump tube properly and lock it (rotate the lock catch on the peri-
staltic pump counter-clockwise).

HumaLyzer Primus | User manual


MAINTENANCE 51

6.3.3 Changing the aspiration tube


If the aspiration tube for the flow cell becomes blocked by objects, it can be
cleaned and purged using a syringe. The aspiration tube can be replaced if it
becomes seriously damaged or blocked.
- Remove the flow cell;
- Remove the aspiration tube at the inlet, and replace with a new aspiration
tube. Sheath the compass tube (middle section) at one end of the new aspi-
ration tube, then sheath the fixing tube (thick section). Be sure to leave 1cm
outside the compass tube (as shown in Figure 57);
- Insert the tube into the flow cell inlet.
Figure 57
Aspiration tube connector
at the flow cell inlet

6.4 Basic troubleshooting

Problem Resolution
1. Analyser cannot be - Check whether the instrument is switched
started. on.
- Check whether the power plug is loose.
- Check the fuse.
- Check the voltage.
2. The photometer lamp does - Check the power supply before changing
not shine. lamps.
- If the lamp is damaged, replace.
3. Internal printer does not - Check whether the plug is properly connect-
print. ed.
4. There is no fluid in the - Check whether the peristaltic pump is func-
flow cell. tioning properly.
- Check the connection of the interfaces for
the aspiration tube and flow cell, etc.
- Aspiration tube is too long or too short.
- Aspiration tube may be blocked, and should
be cleared and purged.
- Flow cell is very dirty.
52

5. There is no reading on the - Check whether the photometer bulb works


photometer. properly.
- Try another reading wavelength.
- Flow cell is not inserted in the bottom.

6. The difference value in wa- - Clean flow cell.


ter blank is too high. - Replace distilled water.
- Check the photometer bulb.

7. The result reproducibility is - There are bubbles inside the flow cell, and it
low. needs to be washed.
- Check the aspiration status of the flow cell.
- Too little reagent is being used; the dosage
needs to be increased.
- Change the bulb of photometer.
- The reaction fluid is contaminated.
- The aspiration tube is inserted relatively
deep in the flow cell and pulls it out slightly.
8. The aspirate flow volume - Check whether the aspiration tube is
of the flow cell is not blocked.
constant. - The peristaltic pump tube may need to be
replaced.
9. Quality control results - Check the expiration date of the reagent.
! Note: Different versions of
the instrument may have
slightly different procedures.
are not within the target
value range.
- Check whether the settings are correct and
whether they need to be edited.
- Ensure the quality control system is not con-
None of these operations should
taminated.
negatively affect the functions
- Repeat the measurement using other meth-
and operation of the instrument; ods.
they can be performed easily by - Check the flow in the flow cell, and use an-
the user. other reagent and quality control system to
repeat the measurement.

HumaLyzer Primus | User manual


MAINTENANCE 53

6.5 Human Software and Settings Update


Information about the latest software and settings are accessible via

https://www.human.de/sw-hlp

or by scanning the QR Code with a mobile device supporting QR Codes. This QR


code can also be found as a label on the device.

Figure 58

If the information are not accessible via internet, it can be obtained free of
charge from your local distributor. If the installed software or settings are not
the latest one, contact your local distributor .
54

HumaLyzer Primus | User manual


APPENDIX 55

7 APPENDIX

7.1 Technical specification

Resolution 0.001 Abs (display) , 0.0001 Abs (internal calcu-


lation)
Light source: Halogen tungsten lamp (6V, 10W)
Wavelength: There are 5 filters inside: 340, 405, 500, 546,
620nm;
Two more wavelengths can be added optionally.
Temperature control: Ambient temperature, 25°C, 30°C, 37°C
Flow cell : Micro flow cell (32µl)
Interface: RS-232 communication
Display: Liquid crystal display
Printer: Internal printer
Operating environment: 10…30°C; relative humidity ≤ 70%
Storage environment: 10...30°C; relative humidity ≤ 80%
Dimensions ( W x D x H): Instrument without any components:
36 x 35 x 16cm
Space required for routine use: 53 x 55 x 29cm
Packaging: 56 x 43 x 32cm
Weight: Gross 6.8kg, Net 5.9kg
Power supply: 110/ 220 V AC ± 10%, 50/ 60Hz ± 1Hz
Input power: 80VA
Fuse type: Φ5 × 20
Fuse specification: 3.15AL250V
56

HumaLyzer Primus | User manual


 57
HUMAN
Gesellschaft für Biochemica und Diagnostica mbH
Max-Planck-Ring 21 • 65205 Wiesbaden • Germany
Tel.: +49 6122/9988 0 • Fax: +49 6122/9988 100
eMail: human@human.de • www.human.de

You might also like