Professional Documents
Culture Documents
| User Manual
SYSTEM VERSION
COPYRIGHT
Copyright 2021, Human Gesellschaft für Biochemica und Diagnostica mbH, Wiesbaden,
Germany. All rights reserved.
No part of this documentation may be reproduced in any form, nor processed, copied or
distributed by means of electronic systems, without prior permission of HUMAN in writing.
Since all precautionary measures were taken into account in producing these operating
instructions, the manufacturer accepts no responsibility for any errors or omissions. This
includes any liability for damage that could arise from possible incorrect operation based on this
information. Subject to changes without notice as result of technical development.
TABLE OF CONTENTS
1 SAFETY INSTRUCTIONS 5
1.1 INTRODUCTION 5
1.2 USER WARRANTY 5
1.3 INTENDED USE OF THE INSTRUMENT 5
1.4 GENERAL SAFETY WARNINGS 6
1.5 DISPOSAL MANAGEMENT CONCEPT 6
1.6 BIOHAZARD WARNING 7
1.7 INSTRUMENT DISINFECTION 7
2 SYSTEM DESCRIPTION 9
2.1 UNPACKING 9
2.2 PRODUCT FEATURES 9
2.3 PRODUCT STRUCTURE AND COMPOSITION 10
2.4 INSTALLATION 10
2.4.1 Selecting a proper location for the Instrument 10
2.4.2 Power supply requirement 10
2.4.3 Connecting the instrument to power supply 10
2.4.4 Installing the flow cell 11
2.5 MAIN STRUCTURE 12
2.5.1 Front view 12
2.5.2 Keypad 13
2.5.3 Back view 14
2.5.4 Internal printer 14
2.5.5 Flow cell and cuvette 14
2.5.6 Lamp and peristaltic pump 15
2.5.7 Schematic system diagram (for reference only) 16
3 OPERATION 17
3.1 POWERING ON 17
3.2 INTRODUCTION TO MAIN MENUS FUNCTIONS 17
3.3 SOFTWARE USAGE 18
3.4 SYSTEM SETTINGS 18
3.5 SAMPLE TEST 21
3.5.1 Test parameter settings 22
3.5.2 Item test 31
4 REPORT 41
4.1 SAMPLE REPORT 42
4.2 ITEM REPORT 43
4.3 QC REPORT 45
4.4 DELETING RECORDS 46
5 POWER OFF 47
6 MAINTENANCE 49
6.1 OVERVIEW 49
6.2 INSTRUMENT CLEANING 49
6.2.1 Cleaning the surface of the instrument 49
6.2.2 Cleaning the flow cell 49
6.3 INSTRUMENT MAINTENANCE 50
6.3.1 Changing the fuse 50
6.3.2 Adjusting the peristaltic pump tube 50
6.3.3 Changing the aspiration tube 51
6.4 BASIC TROUBLESHOOTING 51
6.5 HUMAN SOFTWARE AND SETTINGS UPDATE 53
7 APPENDIX 55
7.1 TECHNICAL SPECIFICATION 55
CONTENTS
Safety Instructions 5
1 SAFETY INSTRUCTIONS
1.1 Introduction
This manual is considered part of the instrument and must be available to the
operator and the maintenance personnel. For accurate installation, use and
maintenance, please read the following instructions carefully.
In order to avoid damage to the instrument or personal injury, carefully read
the General safety warnings, describing the appropriate operating procedures.
Please contact your HUMAN authorised local Technical Service in the event of
instrument failure or other difficulties with the instrument.
Figure 1
Biological hazard
symbol
2 SYSTEM DESCRIPTION
The product is intended for the in vitro diagnostic [IVD] clinical testing of serum,
plasma, urine and CSF samples. The HumaLyzer Primus is a semi-automated
Chemistry Analyzer.
2.1 Unpacking
Remove all the parts from their package. When unpacking the instrument,
please make sure that the following items are conained in the packing. In case
of damage or missing item, please contact the supplier immediately.
2.4 Installation
! Warning:
AC power must be well earthed (near-ground voltage <5V). The AC power
supply must be stable. High-power equipment may not be used on the same
power supply. A voltage-regulated power supply is preferable. In areas with
unstable voltage, please use a UPS power supply to ensure proper functioning
of the instrument. When removing the power cable, pull on the plug instead of
on the power cable. In the event that smoke, abnormal smell or unusual sound
emanate from the instrument, immediately switch off the power and contact
the distributor.
Figure 2
Opening the lamp cover
! Warning:
Do not touch the surface of the light inlet window on the flow cell by hand to
avoid transferring grease from your hands onto the window, which will affect
transmittance. Ensure that the aspiration tube is in the socket. Otherwise, it will
cause pressure build-up and bend the tube, which will affect the test results.
This can easily cause significant bubble formation and increase the water blank.
Figure 3
Front view
1
1 Print cover
2 Lamp cover 2
3 Aspiration tube
4 Aspiration key
5 Keypad 3
7 LCD
5 6 7
2.5.2 Keypad
Figure 4
Keypad
Paper feed key: Manually feed paper; is also the shortcut key
for parameter printing.
Rinse key: Clean the pipeline.
Figure 5
1 2 3 4
Back view
Figure 6
Opening the lamp cover
Figure 7
Internal components of the
colorimetric system
1 1 Flow Cell
2 2 Thermostatic bath
3 Aspiration key
• Thermostatic bath
• Flow cell
2 4 5
Figure 8
Light source system 1
1 Fixing screw
2 Adjusting screws 1
3 Lens
4 Lamp bracket 3
5 Lamp socket
Figure 9
5
Peristaltic pump system 1
6
2
1 Peristaltic pump head
2 Pump tube
3 7
3 Water inlet
4 Water outlet 4
8
5 Fixing block
6 Adjustment screw
7 Pump installation board
8 Pump generator
3 OPERATION
3.1 Powering on
Flip the power switch on the back of the instrument. After several seconds, a
menu will appear in the screen as shown in Figure 11:
Figure 11
Temperature control
After the temperature control process runs for approximately three minutes (if
the Esc key is pressed on the keypad, the system will stop temperature control
and directly enter the software main menu), as shown in Figure 12.
Figure 12
Main menu
2. Report
Perform data administration for the test system. The functions include
printing by test and patient number, deleting the test results, and searching
the quality control information for the current month.
18
3. Sys. Setup
Set up the basic system functions. It includes printer settings, power on and
off, time settings, and an external administration system to transmit all test
data from the current day.
4. Power off
This will request to wash the tubes before power is switched OFF. Users can
turn off the power supply directly to prevent tubes from being washed (it is
recommended that tubes be washed before shutdown).
in the menu indicates that the up and down keys on the keypad can be
used to select the desired menu item. Also date and time as well as the current
software version is displayed in the lower part of the display.
Figure 13
System settings menu
1. Printer settings
Press key 1 in the system settings menu. The instrument will enter the
following menu as shown in Figure 14.
Figure 14
Printer settings
2. Time settings
In the system settings menu, press key 2. The instrument will enter the
following menu as shown in Figure 15.
Figure 15
Time settings
Use the ◄► key on the keypad to select the item you wish to change, press
the appropriate number keys to change the current date or time, press the
Enter key to save and press the Esc key to return to the system settings menu.
3. Data transfer
In the system settings menu, press key 3 (before this operation, ensure that
the communication cable is connected between the instrument and the
external PC and that the serial interface on the external PC is ready to receive
data. The instrument will enter the following menu as shown in Figure 16.
When the menu shown in Figure 17 appears, it indicates that data transfer is
complete. A special software for receiving data is required.
20
Figure 16
Data transferring
Figure 17
Data transfer finished
4. Optics settings
In the system settings menu, press key 4 to enter the optics settings menu as
shown in Figure 18. Wavelength X corresponds to the installation position of
the corresponding filters as shown in Figure 19.
Figure 18
AD (analog to digital)
display menu
Figure 19
Layout of filter wheel (facing the
installation side)
First, press the Rinse key to aspirate distilled water (the pipeline must be filled)
then press the left or right key ◄► to switch the wavelength. After the stable
AD value has been determined, press the Enter key to save the AD value for the
current wavelength. The Save? prompt will disappear
! Note:
This operation will affect test performance; it is only used to determine
whether to replace the lamp or if Redo is shown repeatedly during water blanking.
Ensure sufficient water is present in the flow cell during this operation; aspirate
using the Rinse key as appropriate. Pressing the Enter key will only store the AD
value for the current filter; other filters will remain unchanged.
Figure 20
Test selection menu
Figure 21
Parameters confirmation menu
Press the ◄► key to switch between QC, STD, Edit and OK, then press the Enter
key to enter the corresponding programmes.
24
Figure 22
Parameter
editing menus
Basic procedure:
1. Select the test which you wish to edit; move the cursor by pressing
on the
keypad; the selected test will be highlighted on the display.
5. Press the Enter key on the last page of the test settings to automatically save
the parameter settings and enter the test process.
3. Unit
Select the unit for the test result. The following 9 units are available: mg/
dl, mg/l, g/l, µmol/l, mmol/l, mol/l, U/l, IU/l and blank. Select the unit using
the ◄► key.
4. Temperature
Select test temperature, 4 settings are available: ambient temperature, 25°C,
30°C and 37°C. Choose the temperature using the ◄► key.
5. First wavelength
Choose the required wavelength in the test according to the kit requirements.
5 wavelengths are available: 340nm, 405nm, 500nm, 546nm and 620nm.
Choose the required wavelength using the ◄► key.
6. Second wavelength
Select the secondary wavelength in the test according to the kit
requirements, and the system will flexibly implement the dual wavelength
method by setting the secondary wavelength. It includes 6 selectable
secondary wavelengths: 340nm, 405nm, 500nm, 546nm, 620nm and none.
When a single wavelength is used for the test, the secondary wavelength
must be set as none. Choose the secondary wavelength using the ◄► key.
7. Blank
Select the blank testing, including none, reagent blank and sample blank.
Choose the blank type using the ◄► key. In the end-point method
test, reagent blank and sample blank can be selected according to user
requirements. Reagent blank can be selected for the two-points method test,
and no blank settings can be selected for the kinetics method.
8. Factor
Enter the calculation factor; negative and positive values can be switched
using the ◄► key. The system supports the direct input of factors and the
determination of calculation factors by system calibration (standard test).
9. Norms high
The entered result is the value of norms high based on empirical evidence.
! Note:
A certain time delay is necessary in order to ensure that the test samples are
at a stable temperature and to eliminate bubbles in the flow cell. The time delay
is 5 seconds for most end-point method tests. The time delay should not be less
than 10 seconds on kinetics method tests below 25°C. The time delay should
not be less than 12 seconds on kinetics method tests below 30°C. The time delay
should not be less than 15 seconds on kinetics method tests below 37°C.
After setting all the parameters, press the Enter key to automatically save the
parameters and return to the test parameters confirmation menu. Press the Esc
key if you wish to return without saving. Therefore, correct standard parameters
! Note: Generally, an aspirate
volume of 500µl is sufficient
to ensure a cross-contamination
must be entered. ratio below 1%. The aspirate vol-
ume can be increased or reduced
by number entry on the keypad.
28
Method
The calculation method for standard tests is linear regression or non-linear
regression. Select using the ◄► key.
No. of Std
The number of standard preparations used in the standard test process. The
maximum is 8 for calibration.
Repeat Times
The number of times the test is repeated for each standard preparation.
Concentration
Standard preparation concentration used in the standard test process.
Figure 23
Standard parameter setup menu
! Note:
When non-linear regression is selected, the number of standards should be
more than 3; otherwise, calibrate according to the linear regression method.
When more than two standards are selected, the number of repetitions is
changed to a single measurement for all standards time.
When more than three standard concentrations are entered, they must be
entered using the progressive or degressive method; otherwise, the calibration
will fail. When it processes standard parameters, the system will indicate the
third started standard concentration range as shown in Figure 24. The system
will evaluate the progressive or degressive tendency automatically in sequence,
and the third started concentration range will be indicated at the right side.
Figure 24
Standard concentration menu
After setting all the parameters, press the Enter key to automatically save the
parameters and return to the item parameters confirmation menu. Press the Esc
key if you wish to return without saving. Therefore, correct standard parameters
must be entered.
3.5.1.5 QC settings
In the parameters confirmation menu, select QC and press the Enter key to enter
the settings menu for quality control parameters according to the indications as
shown in Figure 25.
Control
The system can set two quality control preparations. Select using the ◄► key
(in the settings menu for quality control parameters, press the ◄► key to select
the quality control preparations in cycle).
Average Value
Standard concentration of the quality control preparation.
SD
Standard deviation of the quality control preparation.
Batch number
Corresponding batch number for quality control.
Figure 25
Quality control parameters menu
After setting all the parameters, press the Enter key to automatically save the
parameters and return to the menu for item parameters confirmation. Press the
30
Esc key if you wish to return without saving. Therefore, correct quality control
parameters must be entered.
Figure 26
Test switch indication
After finishing the test switch and information loading, the system will enter
temperature control as shown in figure Figure 27. The temperature control
time is set on the basis of the temperature difference before and after the item
switch and typically lasts from several seconds to three minutes. Press the Esc
key if you wish to cancel temperature control and enter the blank test.
Figure 27
Temperature control state
32
Figure 28
Blank test
As shown in Figure 28, a water blank is performed first and Aspirate Please
is shown in the water blank column. Prepare distilled water and press the
aspiration key. The system will automatically perform the test and output the
results of the water blank.
When the value of the water blank is too high, the display will show Aspirate
Please Retest and the real-time absorbance value will be shown below as shown
in Figure 29. In this case, the water blank must be repeated. If Aspirate Please
Retest is still shown after several repetitions, please check whether the light
path is blocked.
Figure 29
Water blank abnormality
After testing the water blank, the system will automatically enter the reagent
blank, aspirate the reagent according to the indications, test the system, and
output the results. After testing the reagent blank, no key needs to be pressed,
as the system will automatically enter the test selection menu and wait for
several seconds, as shown in Figure 30.
The cursor points to Test sample with no selection; select the test sample using
the
key. Press the Enter key to enter the corresponding test process, and press
the Esc key to return to this test selection menu for the other tests.
Figure 30
Test selection menu
1. Test selection
In the test selection menu, move the cursor to Test Standard; the text will
then be highlighted. The system determines whether the settings for the
standard parameters are correct. If the settings are wrong, the system will
display None!, which indicates that the standard test is either unnecessary
or incorrect as shown in Figure 31.
34
If the settings are correct, it displays OK? as shown in Figure 32. Confirm
whether the standard test is to be performed. If it should be performed,
please press the Enter key to enter the test procedure. Otherwise, move the
cursor to select other tests or press the Esc key to return to the main menu.
Figure 31
Standard test selection
Figure 32
2. Test operation
After entering the standard test procedure, the standard test menu will
appear as shown in Figure 33. Test displays the item name for the current
test. STD displays the serial number of the current standard fluid. The
corresponding concentration value is shown on the right side. For example,
with Con.: 133.00 as shown, please ensure that solution of that concentration
is introduced before aspirating the standard fluid.
Status shows the test process schedule and test results. Aspirate Please,
Testing… and Save? are displayed in sequence in the lower-middle area of
the menu. The specific test procedure is the same as the item test process;
please refer to the specific operation of the item test.
Figure 33
Standard test menu
3. Test saving
After aspirating the corresponding standard fluid according to the procedure,
the system will calculate the standard factors in real time. The K and b
calibration values will be displayed in the Result column, and Save? will
be shown in the system’s status column. Press the Enter key to save the
calibration result.
All subsequent item tests will be calculated on the basis of this calibration
result until the standard factors are reset or recalibrated and saved as shown
in Figure 34. Press the Esc key to cancel saving and directly enter the item
test process. In this case, all the subsequent item tests will be calculated
according to the previous calibration settings. If recalibration is required,
please select Standard Test in the test selection menu to perform calibration
again as shown in Figure 30.
Figure 34
Standard test saving
36
This is selected using the ◄► key, which highlights the selected quality
control preparation as shown in Figure 35. To select double quality control,
press and hold the ◄► key until two of the quality control preparations are
highlighted together as shown in figure Figure 36. Confirm that the selected
quality control preparation is correct and press the Enter key to enter the
menu for the quality control test as shown in Figure 37.
Figure 35
Quality control test selection
Figure 36
2. Test operation
After entering the process for the quality control test, the menu for the
quality control test will appear as shown in Figure 37. Test displays the item
name of the current test. QC shows the serial number of the current quality
control solution. The corresponding quality control batch number is shown
on the right side.
For example, with Batch: 070691, please ensure that quality control
preparations with the same batch number are introduced before aspirating
the quality control solution. Status displays the progress of the test process
and the test results. Aspirate Please, Testing… and Save? are displayed in
sequence in the lower-middle of the menu. The specific test procedure is the
same as the item test process; Please refer to the specific operation of the
item test.
Figure 37
Quality control test menu
3. Test saving
After aspirating the corresponding quality control solution according to the
procedure, the system will calculate the results in real time. Quality control
results are displayed, and quality control conditions are evaluated in the
Result column (C>2SD indicates that the quality control result is larger than
the value of two SD; otherwise, this information is not displayed), and Save?
will be shown in the system’s status column.
After confirming that quality control was successful, press the Enter key to
save the quality control results. When double quality control is selected, test
the two quality control preparations in sequence. Aspirate the corresponding
quality control solution according to the indications and save the valid
quality control results (only one quality control result for the same item and
same batch number is saved each day) as shown in Figure 38.
38
Press the Esc key to repeat quality control. After the quality control test is
complete, the system will automatically skip to the test selection menu and
point to the item test as shown in Figure 30.
Figure 38
Saving quality control test
1. Test operation
When sample blank is selected in the parameter settings, a sample blank is
performed first before the sample test as shown in Figure 39. When no blank
or reagent blank is selected, this means it was already finished during the
previous blank test, and the system will directly enter test menu as shown
in Figure 40. Blank and Test will display the item name for the current test.
>Print On or >Print Off on the upper-right corner of the window allow the online
print function to be switched on or off, respectively, using the ► key. If this
function is switched on, the results are printed immediately after the test
is complete. It can only be changed while the system is waiting for sample
aspiration.
Sample displays the next sample number to be tested and can be changed
before the test. A change will be indicated by highlighting Change No. on the
right side of the corresponding column. It is advisable to check whether the
number is correct before starting the test, as the previous test result would
otherwise be automatically overlapped by the new test result with the same
item and same number.
Status displays the progress of the test process, and Result indicates the test
absorbance and concentration.
Aspirate Please, Testing… and Printing… are displayed in sequence in the lower-
middle part of the status column. Please aspirate the sample or wait for the
test according to the indications. When Overlapped? appears after testing, it
indicates that this item already has a test result with the same sample number
on the same day. Press the Enter key if you want to overlap the existing result. If
you want to save the result as a new sample, enter a new sample number then
press the Enter key.
Figure 39
Sample test menu
Sample blank
Figure 40
Sample test
2. Test process
The procedure and related operations for this test process are the same as
for no sample blank (no blank or reagent blank setting) and detailed in that
section. If sample blank is selected, a sample blank will be performed before
each item test and a sample test will be automatically entered to wait for
the sample to be aspirated. Its operation is completely the same as that of
the sample test.
• Confirm the sample number (the sample number displayed is the num-
ber of the sample to be tested) and activate or deactivate the online print
function using the ► key as shown in Figure 39 and Figure 40.
40
• When the system displays Aspirate Please, prepare the test sample and
press the aspiration key to imbibe the sample preparation. The system
will imbibe the fluid according to the set aspirate volume and will auto-
matically enter the test process. Testing... will be displayed in the status
column, and the test progress will be indicated in seconds in the Status
column as shown in Figure 41.
Figure 41
• After the test is complete, the test results are automatically saved (if this
item already has a test result with the same sample number on the same
day, Overlapped? will appear and the system will wait for a response.
Press the Enter key if you want to overlap the existing result. If you want
to save the result as a new sample, enter a new sample number then
press the Enter key). The sample number will be increased by one for the
next test sample and can be modified by the user.
• Print the results: If the online print function is available, activate or deac-
tivate it using the ► key while the system is waiting for fluid aspiration.
• When the current sample test and print are complete, the sample num-
ber will be increased by one and the system will wait for fluid aspiration
for the next sample test.
• If operated according to instructions, the system can continuously test
samples. If quality control, calibration or cancellation are required during
the test procedure, press the Esc key to return to the test selection menu
and select the relevant operations.
4 REPORT
The report functions deal mainly with data from test results and quality control
tests for the current month. They also provide basic functions for data search,
print and delete, etc. In the system’s main menu, press the 2 key to enter the
administration menu as shown in Figure 42. The available functions are sample
report, item report and control report. The corresponding function modules are
accessed using number keys 1, 2 and 3 respectively.
Sample report
Search the test results according to the date and sample number.
Item report
Search the test results according to the date and item number.
QC report
Search the quality control results according to the month, item number and
batch number from quality control.
Figure 42
Administration function selection
! Note:
The HumaLyzer Primus has storage capacity for 2200 test results. The system
can automatically manage its storage. When the quantity of test data exceeds
the systems storage capacity, the system will automatically delete parts of old
data. The system will also issue corresponding notifications during this process.
Please wait a moment while the system clears old data, after which it will
automatically return to normal operation and can continue to perform tests.
42
After entering the time range and sample number, press the Enter key to search
for records which meet the search criteria. If no matching record is found,
None will be displayed at the lower-right of the window as shown in Figure
44. If matching records are found, all records which satisfy the criteria will be
displayed as shown in Figure 45.
Figure 44
Sample search criteria
No record
Figure 45
Several shortcut keys are defined in the display window for search results as
shown in Figure 45:
- Page up/ down
! Note: The „remark“ in the dis-
play and printed report is not
displayed if status is normal; „H“
Only 3 records can be displayed per page; press the
key to display more indicates the test result is above
records. the maximum reference value;
- Print „L“ indicates that the test result
Press the Enter key to print all search results. is below the minimum reference
- Delete value.
Press the Del key to delete all search results.
- Return
Press the Esc key to return to the search window.
The item name corresponding to the number will be displayed on the right side
of the screen in real time. Alternately, you may enter characters in character
mode when ABC appears on the upper-right side of the screen. The “ . “ key is
used to switch between number and character input mode. As shown in Figure
46, the system will search for all results for the ALT item tested on November 9,
2009.
Figure 46
After entering the time range and item number, press the Enter key to search
for records which meet the search criteria. If no matching record is found,
None will be displayed at the lower-right of the window as shown in Figure
47. If matching records are found, all records which satisfy the criteria will be
displayed as shown in Figure 48.
44
Figure 47
Item search condition
No record
Figure 48
4.3 QC Report
Enter the month, item number and quality control batch number to search for in
quality control results. Enter the test date, and enter the item number you wish
to search for in Item No.. The item name corresponding to the number will be
displayed on the right side in real time (if necessary, switch the character input
mode using the “ . ” key). Enter the quality control batch number in “Batch No”
as shown in Figure 49. The system will search It will search for quality control
results for the ALT item with batch number 070691 in November 2009.
Figure 49
Control report search
After entering the month, item number and quality control batch number, press
the Enter key to search for records which meet the criteria. If no matching record
is found, None will be displayed at the lower-right of the window as shown in
Figure 50. If matching records are found, all records which satisfy the criteria
will be displayed as shown in Figure 51.
Figure 50
Quality control search condition
No record
46
Figure 51
Several shortcut keys are defined in the display window for search results as
shown in Figure 51:
- Page up/ down
Only 3 records can be displayed per page; press the
key to display more
Figure 52
Quality control search
This will delete all test results or quality control results. Individual records can be
deleted as described in sections 4.2 and 4.3.
5 POWER OFF
After completing all tests, return to the main menu. Press key 4 to execute the
power-off program. When the screen displays as in Figure 53, press the Enter
key (if you press the Esc key, the system will not power off and you will return
to the main menu).
Figure 53
Confirmation of power off
When the system displays as in Figure 54, it indicates that the system needs to
aspirate distilled water to clean the flow cell.
Figure 54
Cleaning the flow cell
Prepare a sufficient amount of distilled water and press the Aspiration key or
the Rinse key to clean the flow cell as shown in Figure 55.
Figure 55
Indication of cleaning procedure
After the flow cell has been cleaned, the system displays Power off please! as
shown in Figure 56.
Figure 56
Power off
48
6 MAINTENANCE
6.1 Overview
HumaLyzer Primus is a clinical precision analyzer. Daily maintenance work
must be performed properly to keep the instrument in good working condition.
Maintenance work on the HumaLyzer Primus is very easy to perform, but it must
be done thoroughly and carefully.
! Note: Please refer to the material safety datasheet of the Flow Cell Cleaner
([REF] 18222) for precautionary statements.
Problem Resolution
1. Analyser cannot be - Check whether the instrument is switched
started. on.
- Check whether the power plug is loose.
- Check the fuse.
- Check the voltage.
2. The photometer lamp does - Check the power supply before changing
not shine. lamps.
- If the lamp is damaged, replace.
3. Internal printer does not - Check whether the plug is properly connect-
print. ed.
4. There is no fluid in the - Check whether the peristaltic pump is func-
flow cell. tioning properly.
- Check the connection of the interfaces for
the aspiration tube and flow cell, etc.
- Aspiration tube is too long or too short.
- Aspiration tube may be blocked, and should
be cleared and purged.
- Flow cell is very dirty.
52
7. The result reproducibility is - There are bubbles inside the flow cell, and it
low. needs to be washed.
- Check the aspiration status of the flow cell.
- Too little reagent is being used; the dosage
needs to be increased.
- Change the bulb of photometer.
- The reaction fluid is contaminated.
- The aspiration tube is inserted relatively
deep in the flow cell and pulls it out slightly.
8. The aspirate flow volume - Check whether the aspiration tube is
of the flow cell is not blocked.
constant. - The peristaltic pump tube may need to be
replaced.
9. Quality control results - Check the expiration date of the reagent.
! Note: Different versions of
the instrument may have
slightly different procedures.
are not within the target
value range.
- Check whether the settings are correct and
whether they need to be edited.
- Ensure the quality control system is not con-
None of these operations should
taminated.
negatively affect the functions
- Repeat the measurement using other meth-
and operation of the instrument; ods.
they can be performed easily by - Check the flow in the flow cell, and use an-
the user. other reagent and quality control system to
repeat the measurement.
https://www.human.de/sw-hlp
Figure 58
If the information are not accessible via internet, it can be obtained free of
charge from your local distributor. If the installed software or settings are not
the latest one, contact your local distributor .
54
7 APPENDIX