Objectives
z Temperature Monitoring is
essential to maintain cold chain of
a vaccine
z Different temperature monitoring
system
z Introduction of electronic access
control of cold rooms.
Temperature monitoring tools
z Chart recorders
z Computerized temperature
monitoring system : Multilog,
Wireless system
Internatio Primary
In-country Intermediate In-country
Device nal vaccine Service level
transport vaccine store transport
transport store
Electronic
temperature
monitor with LCD
screen
Vaccine cold
chain monitor
Vaccine vial
monitor
Freeze indicator
Multi-channel
computerized
temperature
recording sys.
Thermometer
Chart
Recorder
Fridge tag
33 44
Vaccine storage
Shelves
Side view of
shelves 55 66
WIC
WIC
Door of cold
22 room
11
Legends
Sensor probe
Sensor probe
Sensor cable
Sensor cable
11 Sensor number
Sensor number
Location of Thermostat sensors
z The location of thermostat sensor should be decided after temperature mapping. The
temperature mapping study is performed in empty, partial loaded and fully loaded
condition. Plot the graph and identify the coldest and hottest locations of the
WIC/WIF
z To set temperature, just use common sense. Determine how space will be utilized
and the sizes on the containers within the space. Optimally, there should be enough
sensors evenly spaced throughout the unit, especially where vaccines will be
located. However, No WICs/WIFhave this luxury, and are limited to one thermostat
sensor.
z The WIC stores freeze sensitive vaccines and always place sensor near the coldest
place of the WIC.
In general The coldest part would be front space of the evaporator of the cooling
unit.
z The WIF stores most heat sensitive vaccine OPV and used for preparation of Ice
packs hence in this case the thermostat should be placed near the hottest place of
the WIF, just on opposite wall from the evaporators (see figures).
z The air temperature fluctuates rapidly in comparison to liquid, and can give out-of-
spec data if the temperature range is narrow.
z It is recommended to place thermostat rod immersed in a liquid (water or mercury).
z Though the vaccines safe temperature range for WICs is +2 to 8 deg. Celsius, but
the range of the thermostat should be set for +4 to 8deg. Celsius.
Data Loggers for Storage/ Cold Rooms
Sensor
Radio Receiver
17
Installation of Cobalt System at GMSD Mumbai
18
Proposed System
WIF -18°C
WIF -18°C
Computer with
WIC WIC Receiver
21
Daily Temperature Graph
22
Weekly Temperature Graph
23
Evidence of exposure to freezing
temperatures
10
0 20
-10 30
-20 40
WIC/WIF Access Control