Chapter 6 Troubleshooting

6.1 Introduction .................................................................................................................................1

6.2 Basic Concept on Error Processing ............................................................................................1

6.3 Error Recovery Processing on Sampler Analysis .......................................................................2

6.3.1 Error Recovery processing to be Common to Single Model and Twin Model .................2
6.3.1.1 Sampler Analysis Continue .....................................................................................2
6.3.1.2 Sampler Analysis Suspend ......................................................................................2
6.3.1.3 Sampler Analysis Stop ............................................................................................2

6.3.2 Error Processing on the Twin Model ...............................................................................4


6.3.2.1 Conditions for continuing the sampler analysis when using Twin Model .................4

6.4 Categories of Error Monitoring ....................................................................................................5

6.4.1 Classifications of Error ....................................................................................................5

6.5 Processing Details for Each Errors .............................................................................................7

6.5.1 Pressure/ Vacuum Error Messages ................................................................................7


6.5.1.1 Abnormal Pressure Loss .........................................................................................7
6.5.1.2 [ 0.25MPa Pressure Error ] ......................................................................................8
6.5.1.3 [ 0.16MPa Pressure Error ] ......................................................................................9
6.5.1.4 [ 0.07 MPa Pressure Error ] ...................................................................................10
6.5.1.5 [-0.04 MPa Pressure Error ] ...................................................................................11

6.5.2 Temperature Errors .......................................................................................................12


6.5.2.1 [ 41°C reagent heater temperature is high ] ..........................................................12
6.5.2.2 [ 41°C reagent heater temperature is low ] ............................................................12
6.5.2.3 [ 34°C reagent heater temperature is high ] ..........................................................13
6.5.2.4 [ 34°C reagent heater temperature is low ] ............................................................13
6.5.2.5 [ 41°C FCM reaction chamber temperature is high ] .............................................14
6.5.2.6 [ 41°C FCM reaction chamber temperature is low ] ...............................................14
6.5.2.7 [ 34°C FCM reaction chamber temperature is high ] .............................................15
6.5.2.8 [ 34°C FCM reaction chamber temperature is low ] ...............................................15
6.5.2.9 FCM detector temperature is high .........................................................................16
6.5.2.10 FCM detector temperature is low ..........................................................................16
6.5.2.11 FCM sheath temperature is high ...........................................................................17
6.5.2.12 FCM sheath temperature is low .............................................................................17
6.5.2.13 Environment temperature is high ...........................................................................18
6.5.2.14 Environment temperature is low ............................................................................18
6.5.2.15 [ 41°C reagent heater thermistor error ] .................................................................19
6.5.2.16 [ 34°C reagent heater thermistor error ] .................................................................20
6.5.2.17 [ 41°C FCM reaction chamber thermistor error ] ...................................................21
6.5.2.18 [ 34°C FCM reaction chamber thermistor error ] ...................................................22
6.5.2.19 FCM detector thermistor error ...............................................................................23
6.5.2.20 FCM sheath thermistor error .................................................................................24

XN-1000/2000 S/M March 2015


6.5.2.21 Environment temperature thermistor error ............................................................25
6.5.2.22 APD thermistor error ..............................................................................................26
6.5.2.23 Temperature stabilizing error .................................................................................27

6.5.3 Reagent and Chamber Error .......................................................................................28


6.5.3.1 Out of CELLPACK DCL .........................................................................................28
6.5.3.2 Out of CELLSHEATH(C) .......................................................................................29
6.5.3.3 Out of SULFOLYSER ............................................................................................30
6.5.3.4 Out of Lysercell WNR ............................................................................................31
6.5.3.5 Out of Fluorocell WNR ...........................................................................................32
6.5.3.6 Out of Lysercell WDF ............................................................................................33
6.5.3.7 Out of Fluorocell WDF ...........................................................................................34
6.5.3.8 Out of Lysercell WPC ............................................................................................35
6.5.3.9 Out of Fluorocell WPC ...........................................................................................36
6.5.3.10 Out of CELLPACK DFL .........................................................................................37
6.5.3.11 Out of Fluorocell RET ............................................................................................38
6.5.3.12 Out of Fluorocell PLT .............................................................................................39
6.5.3.13 Out of CELLPACK DCL .........................................................................................40
6.5.3.14 Out of CELLSHEATH(C) .......................................................................................41
6.5.3.15 Out of SULFOLYSER ............................................................................................42
6.5.3.16 Out of Lysercell WNR ............................................................................................43
6.5.3.17 Out of Lysercell WDF ............................................................................................44
6.5.3.18 Out of Lysercell WPC ............................................................................................45
6.5.3.19 Out of CELLPACK DFL .........................................................................................46
6.5.3.20 Reservoir tank is empty (CELLSHEATH(C)) .........................................................47
6.5.3.21 Reservoir tank is empty (SULFOLYSER) ..............................................................48
6.5.3.22 Reservoir tank is empty (Lysercell WNR) ..............................................................49
6.5.3.23 Reservoir tank is empty (Lysercell WDF) ..............................................................50
6.5.3.24 Reservoir tank is empty (Lysercell WPC) ..............................................................51
6.5.3.25 Reservoir tank is empty (CELLPACK DFL) ...........................................................52
6.5.3.26 Out of diluted CELLPACK DST .............................................................................53
6.5.3.27 RU has stopped supplying reagent .......................................................................54
6.5.3.28 CELLPACK DCL aspiration error ..........................................................................55
6.5.3.29 FCM sheath aspiration error ..................................................................................56
6.5.3.30 RBC sheath fluid aspiration error ..........................................................................57
6.5.3.31 Fluorocell WNR aspiration error ............................................................................58
6.5.3.32 Fluorocell WDF aspiration error .............................................................................59
6.5.3.33 Fluorocell WPC aspiration error ............................................................................60
6.5.3.34 Fluorocell RET aspiration error ..............................................................................61
6.5.3.35 Fluorocell PLT aspiration error ..............................................................................62
6.5.3.36 Waste chamber 1 not draining ...............................................................................63
6.5.3.37 Waste chamber 2 not draining ...............................................................................64
6.5.3.38 RBC/HGB chamber not draining ...........................................................................65
6.5.3.39 Waste container is full ...........................................................................................66
6.5.3.40 Water leak detected ...............................................................................................67
6.5.3.41 Water leak detected (analysis not possible) ..........................................................68
6.5.3.42 Water leak sns error ..............................................................................................69

6.5.4 Motor Error ....................................................................................................................70


6.5.4.1 WB aspiration motor error .....................................................................................70

XN-1000/2000 S/M March 2015


6.5.4.2 FCM sheath motor error ........................................................................................71
6.5.4.3 RBC sheath motor error ........................................................................................72
6.5.4.4 Aspiration unit up-down motor error ......................................................................73
6.5.4.5 Aspiration unit left-right motor error ......................................................................74

6.5.5 Sample Aspiration Error ................................................................................................75


6.5.5.1 Insufficient blood volume .......................................................................................75
6.5.5.2 Insufficient blood volume (short sample) ...............................................................76
6.5.5.3 Blood cannot be aspirated .....................................................................................78
6.5.5.4 Aspiration Sensor error ..........................................................................................79

6.5.6 Sampler Operation Errors .............................................................................................80


6.5.6.1 Rack feed-in error ..................................................................................................80
6.5.6.2 Rack feed-in home position error ..........................................................................81
6.5.6.3 Rack move error(Front belt) ..................................................................................82
6.5.6.4 Rack move mechanism initialization error(Front belt) ...........................................83
6.5.6.5 Rack ejection error ................................................................................................84
6.5.6.6 Rack ejection home position error .........................................................................85
6.5.6.7 Tube presence verification home position error ....................................................86
6.5.6.8 Feed-in table stopper position error .......................................................................87
6.5.6.9 Ejection table stopper position error ......................................................................88
6.5.6.10 Sampler belt error ..................................................................................................89
6.5.6.11 Rack not placed on feed-in table. ..........................................................................90
6.5.6.12 Ejection table is full ................................................................................................91
6.5.6.13 Completed sampler analysis stop ..........................................................................92

6.5.7 Hand Unit and Tube Holder Error ..................................................................................93


6.5.7.1 Hand up-down error ...............................................................................................93
6.5.7.2 Hand open/close error ...........................................................................................94
6.5.7.3 Mixing error ............................................................................................................95
6.5.7.4 Tube pickup error ..................................................................................................96
6.5.7.5 Tube holder move error .........................................................................................97
6.5.7.6 Tube return error ...................................................................................................98
6.5.7.7 Tube remains in tube holder ..................................................................................99
6.5.7.8 Two tubes are in tube holder ...............................................................................100
6.5.7.9 The sample must be remixed. .............................................................................101

6.5.8 Sample Number / Rack Number Error ........................................................................102


6.5.8.1 Failed to read sample number during sampler analysis (Sampler analysis). ......102
6.5.8.2 Failed to read rack number ..................................................................................103
6.5.8.3 Failed to read sample number during sampler analysis. .....................................104
6.5.8.4 Positive ID check error ........................................................................................105
6.5.8.5 Sample number not input ....................................................................................106

6.5.9 Order Error ..................................................................................................................107


6.5.9.1 Analysis item not specified ..................................................................................107
6.5.9.2 Invalid analysis item is specified ..........................................................................108
6.5.9.3 Invalid analysis item is specified (Sampler analysis) ...........................................109
6.5.9.4 No analyzer is ready. ...........................................................................................110

6.5.10 Data Analysis Error .....................................................................................................111

XN-1000/2000 S/M March 2015


6.5.10.1 Background check error ......................................................................................111
6.5.10.2 RBC sampling error .............................................................................................112
6.5.10.3 PLT sampling error ..............................................................................................113
6.5.10.4 WNR sampling error ............................................................................................114
6.5.10.5 WDF sampling error ............................................................................................115
6.5.10.6 WPC sampling error ............................................................................................116
6.5.10.7 RET sampling error .............................................................................................117
6.5.10.8 PLT-F sampling error ...........................................................................................118
6.5.10.9 RBC detector clog ...............................................................................................119
6.5.10.10 Bubbles in RBC detector .....................................................................................120
6.5.10.11 Low count error ....................................................................................................121
6.5.10.12 HGB error ............................................................................................................122
6.5.10.13 RBC channel error ...............................................................................................123
6.5.10.14 PLT channel error ................................................................................................124
6.5.10.15 WNR channel error ..............................................................................................125
6.5.10.16 WDF channel error ..............................................................................................126
6.5.10.17 WPC channel error ..............................................................................................127
6.5.10.18 RET channel error ..............................................................................................128
6.5.10.19 PLT-F channel error ...........................................................................................129
6.5.10.20 < Data Errors > ....................................................................................................130
6.5.10.21 Analysis result is high ..........................................................................................131

6.5.11 Cover Error ..................................................................................................................132


6.5.11.1 Front cover open error .........................................................................................132
6.5.11.2 Front cover is open ..............................................................................................133
6.5.11.3 RBC detector cover is open. ................................................................................134
6.5.11.4 FCM detector cover is open. ...............................................................................135
6.5.11.5 Fluorocell WNR cover is open .............................................................................136
6.5.11.6 Fluorocell WDF cover is open .............................................................................137
6.5.11.7 Fluorocell WPC cover is open .............................................................................138
6.5.11.8 Fluorocell RET cover is open ..............................................................................139
6.5.11.9 Fluorocell PLT cover is open ...............................................................................140

6.5.12 Laser Error ..................................................................................................................141


6.5.12.1 Laser life ..............................................................................................................141
6.5.12.2 Laser output error ................................................................................................142

6.5.13 System Error ...............................................................................................................143


6.5.13.1 Analyzer barcode reader communication error ...................................................143
6.5.13.2 Sampler barcode reader communication error ....................................................144
6.5.13.3 Conveyor communication error ...........................................................................145
6.5.13.4 Communication error during sampler analysis. ...................................................146
6.5.13.5 RFID communication error ..................................................................................147
6.5.13.6 Instrument communication error .........................................................................148

6.5.14 QC Error ......................................................................................................................149


6.5.14.1 < X-barM control error > ......................................................................................149
6.5.14.2 < L-J Control Error > ............................................................................................150
6.5.14.3 < X-bar control error > ........................................................................................151
6.5.14.4 < Control has expired. > ....................................................................................152
6.5.14.5 < Control is not entered. > ..................................................................................153

XN-1000/2000 S/M March 2015


6.5.14.6 < QC not executed. > ..........................................................................................154
6.5.14.7 <WNR Scattergram sensitivity error> ..................................................................155
6.5.14.8 <WDF Scattergram sensitivity error> .................................................................156
6.5.14.9 <WPC Scattergram sensitivity error> ..................................................................157
6.5.14.10 <RET Scattergram sensitivity error> ..................................................................158
6.5.14.11 <PLT-F Scattergram sensitivity error> .................................................................159
6.5.14.12 <Check Analysis Mode> ......................................................................................160

6.5.15 User Maintenance / Warning .......................................................................................161


6.5.15.1 Piercer replacement is required. ..........................................................................161
6.5.15.2 Cleaning is required. ...........................................................................................162
6.5.15.3 Cleaning is required (Warning) ............................................................................163
6.5.15.4 CELLPACK DCL has expired. .............................................................................164
6.5.15.5 SULFOLYSER has expired .................................................................................165
6.5.15.6 Lysercell WNR has expired ................................................................................166
6.5.15.7 Fluorocell WNR has expired ................................................................................167
6.5.15.8 Lysercell WDF has expired .................................................................................168
6.5.15.9 Fluorocell WDF has expired ................................................................................169
6.5.15.10 Lysercell WPC has expired .................................................................................170
6.5.15.11 Fluorocell WPC has expired ................................................................................171
6.5.15.12 CELLPACK DFL has expired ..............................................................................172
6.5.15.13 Fluorocell RET has expired .................................................................................173
6.5.15.14 Fluorocell PLT has expired ..................................................................................174
6.5.15.15 Press Start SW ....................................................................................................175

6.5.16 Dye Holder Error .........................................................................................................176


6.5.16.1 Wrong reagent installed in Fluorocell WNR holder ..............................................176
6.5.16.2 Wrong reagent installed in Fluorocell WDF holder ..............................................177
6.5.16.3 Wrong reagent installed in Fluorocell WPC holder ..............................................178
6.5.16.4 Wrong reagent installed in Fluorocell RET holder ...............................................179
6.5.16.5 Wrong reagent installed in Fluorocell PLT holder ................................................180
6.5.16.6 Fluorocell WNR is not installed ............................................................................181
6.5.16.7 Fluorocell WDF is not installed ............................................................................182
6.5.16.8 Fluorocell WPC is not installed ............................................................................183
6.5.16.9 Fluorocell RET is not installed .............................................................................184
6.5.16.10 Fluorocell PLT is not installed ..............................................................................185
6.5.16.11 Fluorocell WNR has already been used ..............................................................186
6.5.16.12 Fluorocell WDF has already been used ..............................................................187
6.5.16.13 Fluorocell WPC has already been used ..............................................................188
6.5.16.14 Fluorocell RET has already been used ...............................................................189
6.5.16.15 Fluorocell PLT has already been used ................................................................190
6.5.16.16 Cannot recognize Fluorocell WNR information ...................................................191
6.5.16.17 Cannot recognize Fluorocell WDF information ....................................................192
6.5.16.18 Cannot recognize Fluorocell WPC information ....................................................193
6.5.16.19 Cannot recognize Fluorocell RET information .....................................................194
6.5.16.20 Cannot recognize Fluorocell PLT information .....................................................195
6.5.16.21 Fluorocell WNR RFID tag error ...........................................................................196
6.5.16.22 Fluorocell WDF RFID tag error ............................................................................197
6.5.16.23 Fluorocell WPC RFID tag error ............................................................................198
6.5.16.24 Fluorocell RET RFID tag error .............................................................................199

XN-1000/2000 S/M March 2015


6.5.16.25 Fluorocell PLT RFID tag error .............................................................................200

6.6 SA-10/SA-20 Sensors .............................................................................................................201

6.6.1 Sensor List (Sampler Sonsor Display) ........................................................................201

6.6.2 SA-10 Sensor List (Other than Sampler Sonsor Display) ...........................................202

6.6.3 SA-20 Sensor List (Other than Sampler Sensor Display) ...........................................203

6.7 Sampler (Unidiractional Sampler SA-01) Operation Errors ...................................................204

6.7.1 Rack Feed In error ......................................................................................................204

6.7.2 Rack feed-in home position error ................................................................................205

6.7.3 Rack move error ..........................................................................................................206

6.7.4 Rack move home position error ..................................................................................207

6.7.5 Rack ejection error ......................................................................................................208

6.7.6 Rack ejection home position error ...............................................................................209

6.7.7 Sampler belt error ........................................................................................................210

6.7.8 Rack Removed ............................................................................................................211

6.7.9 Rack not placed on feed-in table .................................................................................212

6.7.10 Ejection table is full ......................................................................................................213

XN-1000/2000 S/M March 2015


Chapter 6 Troubleshooting

6.1 Introduction
This troubleshooting describes the error monitoring and error recovery processing of the analysis
instruments (analyzer and sampler) on the XN-1000/2000 software specifications. The errors are
categorized based on the basic concept in the next section, and error recovery processing is
described.

6.2 Basic Concept on Error Processing


The basic concept on XN-1000/2000 error processing is as follows.
Errors are classified into 2 categories: “Hardware Error” and “Analysis Error”

Hardware error is an error which occurring by the instrument status such as


Hardware Error
the analyzer and the sampler.

Analysis error is an error related to result data. This error affects the analysis
Analysis Error
data.

Table 1: Error Categories

1. Reset error methods are classified into 3 categories:


a. Only accepting by the operator.
Performing the individual maintenance sequence by the operator.
b. Operating the hardware. (Refer to the Table 2: Dealing with the error occurrences by the
operators.
The operation flow from the error occurrence to the error reset is shown in below.

RESET ERROR HELP Screen Analysis Instrument Action

The error clears by clicking


Only accepting by the operator [Accept] key is available.
[Accept] key.

Executing the individual


maintenance sequence by
Performing the individual
clicking [Execute] key. The
maintenance sequence by the [Execute] key is available.
error clears when the
operator
maintenance sequence is
done.

The error clears when the


Operating the Hardware [Accept] key is gray out necessary processing such as
power ON/OFF is done.

Table 2: Dealing with the Error Occurrence by the Operator

2. On Manual Analysis, the analysis channels which are not affected by errors can be analyzed by
limiting the discrete even though the error occurs. However, because Sampler Analysis follows
“work list setting “, it cannot be analyzed by Sampler Analysis.

XN-1000/2000 S/M 6-1 March 2015


6.3 Error Recovery Processing on Sampler Analysis
The error recovery processing on Sampler Analysis of XN-1000/2000 is as follows:

6.3.1 Error Recovery processing to be Common to Single Model and Twin Model

Recovery processing on an error occurrence while analyzing by Single Model and/or Twin Model is
classified into “Sampler Analysis continue”, “Sampler Analysis suspend” and “Sampler Analysis
stop. “
When multiple errors occur at the same time, give priority to processing in order of “Sample
Analysis is stop”, “Sampler Analysis suspend”, and “Sampler Analysis continue.” Details of each
processing are as follows:

6.3.1.1 Sampler Analysis Continue


<Outline>
When the error as a purpose of calling the operator attention occurs, help message appears
on the HELP screen. Sampler Analysis continues.
<Processing when error occurs>
Sampler Analysis continues.
< Processing to clear the error >
None.

6.3.1.2 Sampler Analysis Suspend


<Outline>
Sampler Analysis is suspended when process or check is needed by the operator (out of
reagents, analyzer barcode reader communication error).
<Processing when error occurs>
1. The analyzer returns the sample tubes to the rack when sample tubes are removed from
the rack to the analyzer.
2. After all sample tubes are returned to the rack; the analyzer transports the rack to the start
point and to end point on the analysis line, and then suspends temporarily Sampler
Analysis.
<Processing to clear the error>
1. The operator clears all suspended error.
2. The analyzer restarts Sampler Analysis automatically.

6.3.1.3 Sampler Analysis Stop


Sampler analysis is stopped when an impossible Sample Analysis error occurs. Depending on the
error occurring when Sampler Analysis is stopped, the errors are categorized into “Sampler
Analysis Stop 1”, “Sampler Analysis Stop 2, and “Sampler Analysis Stop 3”. When multiple errors
occur at the same time, give priority to processing in order of “Sampler Analysis Stop 3”, “Sampler
Analysis stop 2” and “Sampler Analysis stop 1”. Details of each processing are as follows.

6.3.1.3.1 Sampler Analysis Stop 1


<Outline>
When the error which cannot be recovered by the operator (thermistor error or other) occurred,
Sampler Analysis is stopped and the rack is automatically ejected to the rack ejecting table.

XN-1000/2000 S/M 6-2 March 2015


<Processing when error occurred>
1. The analyzer returns the sample tube to the rack when the sample tube is sent to the
analyzer from the rack.
2. The analyzer drains the rack to the rack ejection table after all sample tubes are returned
to the rack.
3. The analyzer initializes the sampler after all racks on the analysis line are ejected.
4. The error message “Complete sampler analysis stop” appears after the analyzer
initialized the sampler.
<Processing when error cleared>
The analyzer becomes ready to start analysis when the operator accepts the error message
“Completed sampler analysis stop”, and also there are no other errors.

6.3.1.3.2 Sampler Analysis Stop 2


<Outline>
When the error relating to the sample tubes transport (hand unit error or other), or error except
analysis line (rack feed-in error or other) occurred, the analyzer stops the sample tube and
rack transport. After error clears, the analyzer automatically drains the rack to the rack
ejection table.
<Processing when error occurred>
1. The analyzer stops transporting the sample-tube to the analyzer, and also it stops
transporting the rack to the sampler.
2. The analyzer which has not occurred errors is transition to Manual mode. (Only Twin
Mode)
<Processing when error cleared>
1. All errors are cleared by the operator.
2. The analyzer returns the sample tubes to the rack when returning the sample tube to the
rack from the analyzer is possible.
3. The analyzer drains the rack to the rack ejection table after processing 2 is completed.
4. The analyzer initializes the sampler after all racks are drained from the analysis line.
5. The error message “Completed Sampler Analysis Stop” appears after the analyzer
completed initializing the sampler.
6. The analyzer becomes ready to start analysis when the operator accepts the error
message “Completed sampler analysis stop”, and also there are no other errors.

6.3.1.3.3 Sampler Analysis Stop 3


<Outline>
The analyzer stops transporting the sample tubes and rack when the error relating to the
analysis line of the sampler (rack move error or others) occurred. The operator drains the rack
manually.
<Processing when error occurred>
1. The analyzer stops sampler transport at the analyzer, and also, stops the rack transport at
the sampler.
2. The analyzer is transition to Manual Mode.
<Processing when error cleared>
1. The operator removes all sample tubes and rack from the analysis line and the analyzer.
2. The operator clears all errors.
3. The analyzer initializes the sampler after all errors are cleared.
4. The error message “Completed sampler analysis stop” appears after the analyzer
completed initializing the sampler.
5. The analyzer becomes ready to start analysis when the operator accepts the error
message “Completed sampler analysis stop”, and also, there are no other errors.

XN-1000/2000 S/M 6-3 March 2015


6.3.2 Error Processing on the Twin Model

6.3.2.1 Conditions for continuing the sampler analysis when using Twin Model
On the sampler analysis using Twin Model, if sampler analysis is possible by the other analyzer,
sample analysis is continued, even if one of the analyzers is incapable of analyzing (including
switching to Manual Analysis Mode).
The sampler analysis suspends when all analyzers connecting to the sampler are incapable of
analyzing.
After suspending, when one of the analyzers becomes capable of sample analysis, sampler
analysis restarts.
However, the sampler analysis suspends without depending on the status of analyzer if the
following conditions are congruent with the errors.

1. When error is congruent with the suspending condition by user setting. (sampler ID read
error or the others)
2. When error is congruent with the suspending condition by system setting. (replacement of
common reagents or others)

Also, the sampler analysis and transport suspend without depending on the status of the analyzer if
the following conditions are congruent with the errors.

1. When the error relating to the Hand Unit (including pressure error) occurs.
2. When the sampler mechanism error occurs.

XN-1000/2000 S/M 6-4 March 2015


6.4 Categories of Error Monitoring
The incidence of errors on the XN-1000/2000 is classified.

6.4.1 Classifications of Error

The errors are classified into 4 categories as follows by sources of error monitoring. Depending on
the sampler and analyzer, there are related or unrelated categories to each instrument. (○=related
categories)
Instrument
Category Details of error
Analyzer Sampler

The error occurring due to an instrument


malfunction or a limit monitoring level is over the
specified value.
Hardware Error ○ ○
<Restoring Process>
It becomes ready status after the error is cleared
by executing necessary sequences.

The error occurring by analysis results


Analysis Error ○ <Restoring Process>
Re-analyze to check that the error cleared.

The error as a purpose of calling the operator


attention
Caution ○ ○ <Restoring Process>
Deleting the error message when the error is
cleared by executing necessary sequences.

The error as a purpose of calling the operator


maintenance
Maintenance ○ <Restoring Process>
Deleting the error message when the error is
cleared by executing necessary sequences.
Table 6-4-1 Error Categories

XN-1000/2000 S/M 6-5 March 2015


Process Details and Description on the next section

Indicating abnormal analysis result in hyphen mask “- - - -“.


On the next chapter, describing the “Data Mask” as follows
Data Mask
✔ : Indicating in hyphen mask.
NA: Not displayed with hyphen mask.

Disable the analyzer after an error occurred.


On the next chapter, describing “invalid Analysis” as follows:
Invalid Analysis ✔: Analysis disable
(✔): Analyzable with conditions. (Only analyzable discrete or only manual mode)
NA: Analyzable

Indicating the process of sampler analysis after the error occurred.


On the next chapter, describing the “Sample Analysis” as follows:
Sampler Analysis Continue: Processing “sampler analysis continue.”
Sampler Analysis Suspend: Processing “sampler analysis suspend.”
Sampler Analysis Continue / Suspend (Only Twin Model):
According to "3.2: Twin Model Error Processing", processing either "Sampler Analysis
Continue" or "Sampler Analysis Suspend."
Sampler Analysis
Sampler Analysis Suspend:
Processing “sampler analysis suspend” only sampler suspend condition is set.
Processing “sampler analysis continue” when sampler suspend condition is not set.
Sampler Analysis Stop
Processing “sampler analysis stop” if removing racks by an operator is necessary,
mention it in an important notice.
NA: Indicating the error not occurred while operating sample analysis

An alarm sounds from IPU when the error occurred.


On the next chapter, describing “alarm sound” as follows:
NA: Alarm does not sound.
Alarm Sound Warning: A warning alarm sounds.
Analysis stop: An analysis stop alarm sounds.
(Analysis stop): An analysis stop alarm sounds when suspending sampler analysis.
A warning alarm sounds when continuing sampler analysis.

Reanalysis when the data mask error occurred while operating sample analysis.
On the next chapter, describing “re-analysis” as follows: (Changeable by reanalysis rule
setting)
Block: Not reanalyzing whether Repeat error occurs.
Reanalysis
Repeat: Reanalyzing when Block error does not occur.
None: Reanalyzing depending on whether Block / Repeat errors occur or not.
NA: Errors inapplicable reanalysis Judgment.

Table 6-4-2 Processing when errors occurred (describing details on the next chapter)

XN-1000/2000 S/M 6-6 March 2015


6.5 Processing Details for Each Errors
The processing details for each error on the XN-1000/2000 describe as follows.

6.5.1 Pressure/ Vacuum Error Messages

6.5.1.1 Abnormal Pressure Loss

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process when error
occurred Sampler Single Model: Sampler Analysis Stop 2
Measurement Twin Model: Sampler Analysis Stop 2

Important Notice Stop the analyzer sequence immediately.

Meaning The pneumatic pressure abnormally low.

Purpose of error
To ensure the valve performing.
detection

Method of error
Check 0.25MPa pressure sensor while analysis operating (always)
detection

Conditions of error
When the pressure is less than 0.15Mpa for 3 seconds continuously.
detection

Error Message Abnormal Pressure Loss

Operation could not be continued due to pressure loss. Remove rack and tubes from
Action Message
analyzer and press [Execute] key. Analyzer initialize will be executed.

HELP [Execute] Restart the analyzer.


Process Clear the error after restarting the analyzer.

Error Code 119010 Judgment Analyzer

HELP Screen Key Execute


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-7 March 2015


6.5.1.2 [ 0.25MPa Pressure Error ]

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Repeat


Process when error
occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/Suspend

Important Notice None

Meaning 0.25MPa pressure is outside of the specified range.

Purpose of error
To ensure the solenoid Valve, Valve, and DP performing.
detection

Method of error
Check 0.25MPa pressure sensor when analyzer starts and while operating analysis.
detection

Pneumatic Unit Operating (except operating Analysis)(not monitoring while adjusting


pressure) :
Conditions of error Outside 0.25MPa ±0.04MPa range.
detection
Operating Analysis:
Outside range of 0.18Mpa-0.29MPa on the timing of the Timing Chart.

Error Message 0.25MPa Pressure Error

0.25 MPa pressure is outside the required range. Press [Execute] key. A pressure
Action Message
adjustment will be executed.

HELP [Execute] Execute pressure adjustment.


Process The error clears when the pressure is inside the specified range.

Error Code 113025 Judgment Control Unit

HELP Screen Key Execute


Detailed Information
Maintenance Menu Pressure Adjustment

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-8 March 2015


6.5.1.3 [ 0.16MPa Pressure Error ]

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error2

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/Suspend

Important Notice None

Meaning 0.16MPa pressure is outside of the specified range.

Purpose of Error
To ensure the Sheath pressure used on the FCM.
Detection

Method of Error
Check 0.16MPa pressure sensor when analyzer starts and while operating analysis.
Detection

Pneumatic Unit Operating (except operating Analysis)(not monitoring while adjusting


pressure) :
Conditions of Error Outside 0.16 MPa ±0.016 MPa range.
Detection
Operating Analysis:
Outside range of 0.16 MPa ±0.016 MPa on the timing of the Timing Chart.

Error Message 0.16 MPa Pressure Error

0.16 MPa pressure is outside the required range. Press [Execute] key. A pressure
Action Message
adjustment will be executed.

HELP [Execute] Execute pressure adjustment.


Process The error clears when the pressure is inside the specified range.

Error Code 113016 Judgment Control Unit

HELP Screen Key Execute


Detailed Information
Maintenance Menu Pressure Adjustment

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-9 March 2015


6.5.1.4 [ 0.07 MPa Pressure Error ]

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔


Alarm Sounds Analysis Stop Reanalysis Repeat
Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/Suspend

Important Notice None

Meaning 0.07 MPa pressure is outside of the specified range.

Purpose of Error
To ensure the Sheath pressure used on the RBC/PLT counting, DP, and rinse.
Detection

Method of Error
Check 0.07MPa pressure sensor when analyzer starts and while operating analysis.
Detection

Pneumatic Unit Operating (except operating Analysis)(not monitoring while adjusting


pressure) :
Conditions of Error Outside 0.07 MPa ±0.01 MPa range.
Detection
Operating Analysis:
Outside range of 0.07 MPa ±0.01 MPa on the timing of the Timing Chart.

Error Message 0.07 MPa Pressure Error

0.07 MPa pressure is outside the required range. Press [Execute] key. A pressure
Action Message
adjustment will be executed.

HELP [Execute] Execute pressure adjustment.


Process The error clears when the pressure is inside the specified range.

Error Code 113007 Judgment Control Unit


HELP Screen Key Execute
Detailed Information
Maintenance Menu Pressure Adjustment

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-10 March 2015


6.5.1.5 [-0.04 MPa Pressure Error ]

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/Suspend

Important Notice None

Meaning -0.04 MPa pressure is outside of the specified range.

Purpose of Error To ensure charging volume of dilution sample, draining reaction chamber, DP
Detection operations.

Method of Error
Check -0.04MPa pressure sensor when analyzer starts and while operating analysis.
Detection

Pneumatic Unit Operating (except operating Analysis)(not monitoring while adjusting


pressure) :
Conditions of Error Outside range of -0.037 MPa -(-0.054) MPa.
Detection
Operating Analysis:
Outside range of -0.037 MPa - (-0.054) MPa on the timing of the Timing Chart.

Error Message -0.04 MPa Pressure Error

-0.04 MPa pressure is outside the required range. Press [Execute] key. A pressure
Action Message
adjustment will be executed.

HELP [Execute] Execute pressure adjustment.


Process The error clears when the pressure is inside the specified range.

Error Code 114004 Judgment Control Unit

HELP Screen Key Execute


Detailed Information
Maintenance Menu Pressure Adjustment

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-11 March 2015


6.5.2 Temperature Errors

6.5.2.1 [ 41°C reagent heater temperature is high ]


6.5.2.2 [ 41°C reagent heater temperature is low ]

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

Meaning 41°C reagent heater temperature is outside the specified range.

Purpose of Error To ensure temperatures of WNR Lysing, WDF Lysing, and RET/PLT-F Diluent.
Detection To ensure the reaction of samples.

Method of Error
Check 41°C Reagent Heater Thermister when in ready status and starting analysis.
Detection

Target temperature = -0.0699t + 43.98°C (t=environment temperature)


1. Ready Status:
Conditions of Error When the temperature is outside range of [ Target Temperature ] ±2°C for 3 seconds
Detection continuously.
2. Start analyzing:
When the temperature is outside range of [ Target Temperature ] ±3°C

41°C reagent heater temperature is high


Error Message
41°C reagent heater temperature is low

41°C reagent heater temperature is outside the required range. Press [Execute] key. The
Action Message
temperature sensor screen will appear.

HELP [Execute] The temperature sensor screen appears.


Process The error is cleared when temperature becomes inside the specified range.

High:125040
Error Code Judgment Control Unit
Low: 125045

Detailed Information HELP Screen Key Execute

Maintenance Menu Temperature sensor appears

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-12 March 2015


6.5.2.3 [ 34°C reagent heater temperature is high ]
6.5.2.4 [ 34°C reagent heater temperature is low ]

Analyzer Type A
Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

Meaning 34°C reagent heater temperature is outside the specified range.

Purpose of Error To ensure temperature of WPC Lysing.


Detection To ensure the reaction of samples.

Method of Error
Check 34°C Reagent Heater Thermister when in ready status and starting analysis.
Detection

Target temperature = -0.0292t + 34.906°C (t=environment temperature)


[Environment Temperature < 34°C]
1. Ready status:
When the temperature is outside range of [ Target Temperature ] ±2°C for 3 seconds
continuously.
2. Start analyzing:
Conditions of Error
When the temperature is outside range of [ Target Temperature ] ±3°C
Detection
[Environment Temperature ≥ 34°C]
1. Ready status:
When the temperature is outside range of 31°C - [ Target Temperature ] for 3 seconds
continuously.
2. Start analyzing:
When the temperature is outside range of 31°C - [ Target Temperature ]

34°C reagent heater temperature is high


Error Message
34°C reagent heater temperature is low

34°C reagent heater temperature is outside the required range. Press [Execute] key. The
Action Message
temperature sensor screen will appear.

HELP [Execute] The temperature sensor screen appears


Process The error is cleared when temperature becomes inside the specified range.

High:125050
Error Code Judgment Control Unit
Low: 125055

Detailed Information HELP Screen Key Execute

Maintenance Menu Temperature sensor appears

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-13 March 2015


6.5.2.5 [ 41°C FCM reaction chamber temperature is high ]
6.5.2.6 [ 41°C FCM reaction chamber temperature is low ]

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔


Alarm Sounds Analysis Stop Reanalysis Repeat
Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

The temperatures of WNR reaction chamber, WDF reaction chamber, RET reaction
Meaning
chamber and PLT-F reaction chamber are outside the specified range.

To ensure temperatures of WNR reaction chamber, WDF reaction chamber, RET reaction
Purpose of Error chamber and PLT-F reaction chamber.
Detection
To ensure the stainlyzing reactions of WNR, WDF, RET and PLT-F.

Method of Error
Check 41°C Reaction Thermister when in ready status and starting analysis.
Detection

Target temperature = 41°C


1. Ready Status:
Conditions of Error When the temperature is outside range of [ Target Temperature ] ± 2°C for 3 seconds
Detection continuously.
2. Start analyzing:
When the temperature is outside range of [ Target Temperature ] ±3°C

41°C FCM reaction chamber temperature is high


Error Message
41°C FCM reaction chamber temperature is low

41°C FCM reaction chamber temperature is outside the required range. Press [Execute]
Action Message
key. The temperature sensor screen will appear.

HELP [Execute] The temperature sensor screen appears


Process The error is cleared when temperature becomes inside the specified range.

High:123250
Error Code Judgment Control Unit
Low: 123255

Detailed Information HELP Screen Key Execute

Maintenance Menu Temperature sensor appears

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-14 March 2015


6.5.2.7 [ 34°C FCM reaction chamber temperature is high ]
6.5.2.8 [ 34°C FCM reaction chamber temperature is low ]

Analyzer Type A
Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

Meaning The temperature of WPC reaction chamber is outside the specified range.

Purpose of Error To ensure temperatures of WPC reaction chamber.


Detection To ensure the reaction of WPC samples.

Method of Error
Check 34°C Reaction Thermister when in ready status and starting analysis.
Detection

[Environment temperature < 34°C]


Target Temperature = 34°C
1. Ready status:
When the temperature is outside range of [ Target Temperature ] ±2°C for 3 seconds
continuously.
2. Start analyzing:
Conditions of Error When the temperature is outside range of [ Target Temperature ] ±3°C
Detection [Environment temperature ≥ 34°C]
Target Temperature = t + 3°C (t = environment temperature)
1. Ready status:
When the temperature is outside range of 31°C - [ Target Temperature ] for 3 seconds
continuously.
2. Start analyzing:
When the temperature is outside range of 31°C - [ Target Temperature ]

34°C FCM reaction chamber temperature is high


Error Message
34°C FCM reaction chamber temperature is low

34°C FCM reaction chamber temperature is outside the required range. Press [Execute]
Action Message
key. The temperature sensor screen will appear.

HELP [Execute] The temperature sensor screen appears


Process The error is cleared when temperature becomes inside the specified range.

High:123260
Error Code Judgment Control Unit
Low: 123265

Detailed Information HELP Screen Key Execute

Maintenance Menu Temperature sensor appears

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-15 March 2015


6.5.2.9 FCM detector temperature is high
6.5.2.10 FCM detector temperature is low

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

Meaning The temperature of FCM detector is outside the specified range.

Purpose of Error
To ensure temperatures of FCM detector.
Detection

Method of Error
Check FCM detector thermister when in ready status and starting analysis.
Detection

[ Target Temperature ] = 40°C


1. Ready Status
Conditions of Error when the temperature is outside range of [ Target temperature ] ± 2°C for 3 seconds
Detection continuously.
2. Starting Analysis
when the temperature is outside range of [ Target temperature ] ± 2°C

FCM detector temperature is high


Error Message
FCM detector temperature is low

FCM detector temperature is outside the required range. Press [Execute] key. The
Action Message
temperature sensor screen will appear.

HELP [Execute] The temperature sensor screen appears.


Process The error is cleared when temperature becomes inside the specified range.

High:121250
Error Code Judgment Control Unit
Low: 121255

Detailed Information HELP Screen Key Execute

Maintenance Menu Temperature sensor appears

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-16 March 2015


6.5.2.11 FCM sheath temperature is high
6.5.2.12 FCM sheath temperature is low

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔


Alarm Sounds Analysis Stop Reanalysis Repeat
Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

Meaning The temperature of FCM sheath is outside the specified range.

Purpose of Error
To ensure temperatures of FCM sheath.
Detection

Method of Error
Check FCM sheath thermister when in ready status and starting analysis.
Detection

[ Target Temperature ] = 5°C - 45°C


1. Ready Status
Conditions of Error WWhen the temperature is outside range of [ target temperature ] for 3 seconds
Detection continuously.
2. Starting Analysis
When the temperature is outside range of [ target temperature ]

FCM sheath temperature is high


Error Message
FCM sheath temperature is low

FCM sheath temperature is outside the required range. Press [Execute] key. The
Action Message
temperature sensor screen will appear.

HELP [Execute] The temperature sensor screen appears.


Process The error is cleared when temperature becomes inside the specified range.

High:124110
Error Code Judgment Control Unit
Low: 124115

Detailed Information HELP Screen Key Execute

Maintenance Menu Temperature sensor appears

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-17 March 2015


6.5.2.13 Environment temperature is high
6.5.2.14 Environment temperature is low

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔


Alarm Sounds Analysis Stop Reanalysis Repeat
Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

Meaning The environment temperature of the instrument is outside the specified range.

Purpose of Error Check the temperature of use environment for the instrument.
Detection Check that the environment temperature thermistor is not defective.

Method of Error
Check the Environment Thermister when in ready status and starting analysis.
Detection

[ Target Temperature ] = 5°C - 38°C


1. Ready Status
Conditions of Error when the temperature is outside range of [ target temperature ] for 3 seconds
Detection continuously.
2. Starting Analysis
when the temperature is outside range of [ target temperature ]

Environment temperature is high


Error Message
Environment temperature is low

Environment temperature is outside the required range. Press [Execute] key. The
Action Message
temperature sensor screen will appear.

HELP [Execute] The temperature sensor screen appears.


Process The error is cleared when temperature becomes inside the specified range.

High:122510
Error Code Judgment Control Unit
Low: 122515

Detailed Information HELP Screen Key Execute

Maintenance Menu Temperature sensor appears

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-18 March 2015


6.5.2.15 [ 41°C reagent heater thermistor error ]

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔


Alarm Sounds Analysis Stop Reanalysis Block (Fixed)
Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 1
Measurement Twin Model: Sampler Analysis Stop 1

Important Notice None

41°C Reagent Heater Thermistor is defective or 41°C reagent heater thermistor wire is
Meaning
broken.

Purpose of Error
To ensure that 41°C reagent heater thermistor performs properly.
Detection

Method of Error
Check the error information of the driver board in ready status or operating.
Detection

1. Ready Status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.

Error Message 41°C reagent heater thermistor error

Analysis cannot be continued because the 41°C reagent heater thermistor wire is broken.
Action Message
Remove the rack and tubes from the analyzer and turn off the analyzer power.

HELP [Execute] None.


Process Clear the error after turning off the analyzer power.

Error Code 127440 Judgment Control Unit


HELP Screen Key [Accept] (grayout)
Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-19 March 2015


6.5.2.16 [ 34°C reagent heater thermistor error ]

Analyzer Type A
Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Block (Fixed)


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 1
Measurement Twin Model: Sampler Analysis Stop 1

Important Notice None

34°C Reagent Heater Thermistor is defective or 41°C reagent heater thermistor wire is
Meaning
broken.

Purpose of Error
To ensure that 34°C reagent heater thermistor performs properly.
Detection

Method of Error
Check the error informationof the driver board in ready status or operating.
Detection

1. Ready Status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.

Error Message 34°C reagent heater thermistor error

Analysis cannot be continued because the 34°C reagent heater thermistor wire is broken.
Action Message
Remove the rack and tubes from the analyzer and turn off the analyzer power.

HELP [Execute] None.


Process Clear the error after turning off the analyzer power.

Error Code 127450 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-20 March 2015


6.5.2.17 [ 41°C FCM reaction chamber thermistor error ]

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Block (Fixed)


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 1
Measurement Twin Model: Sampler Analysis Stop 1

Important Notice None

41°C Reaction Chamber Thermistor is defective or 41°C reaction chamber thermistor wire
Meaning
is broken.

Purpose of Error
To ensure that 41°C reaction chamber thermistor performs properly.
Detection

Method of Error
Check the error information of the driver board in ready status or operating.
Detection

1. Ready status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.

Error Message 41°C FCM reaction chamber thermistor error

Analysis cannot be continued because the 41°C FCM reaction chamber thermistor wire is
Action Message
broken. Remove the rack and tubes from the analyzer and turn off the analyzer power.

HELP [Execute] None.


Process Clear the error after turning off the analyzer power.

Error Code 127610 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-21 March 2015


6.5.2.18 [ 34°C FCM reaction chamber thermistor error ]

Analyzer Type A
Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Block (Fixed)


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 1
Measurement Twin Model: Sampler Analysis Stop 1

Important Notice None

34°C Reaction Chamber Thermistor is defective or 34°C reaction chamber thermistor wire
Meaning
is broken.

Purpose of Error
To ensure that 34°C reaction chamber thermistor performs properly.
Detection

Method of Error
Check the error information of the driver board in ready status or operating.
Detection

1. Ready status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.

Error Message 34°C FCM reaction chamber thermistor error

Analysis cannot be continued because the 34°C FCM reaction chamber thermistor wire is
Action Message
broken. Remove the rack and tubes from the analyzer and turn off the analyzer power.

HELP [Execute] None.


Process Clear the error after turning off the analyzer power.

Error Code 127620 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-22 March 2015


6.5.2.19 FCM detector thermistor error

Analyzer Type A
Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Block (Fixed)


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 1
Measurement Twin Model: Sampler Analysis Stop 1

Important Notice None

Meaning FCM Detector Thermistor is defective or FCM detector thermistor wire is broken.

Purpose of Error
To ensure that FCM detector thermistor performs properly.
Detection

Method of Error
Check the error information of the driver board in ready status or operating.
Detection

1. Ready Status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.

Error Message FCM detector thermistor error

Analysis cannot be continued because the FCM detector thermistor wire is broken.
Action Message
Remove the rack and tubes from the analyzer and turn off the analyzer power.

HELP [Execute] None.


Process Clear the error after turning off the analyzer power.

Error Code 127350 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-23 March 2015


6.5.2.20 FCM sheath thermistor error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔


Alarm Sounds Analysis Stop Reanalysis Block (Fixed)
Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 1
Measurement Twin Model: Sampler Analysis Stop 1

Important Notice None

Meaning FCM sheath thermistor is defective or FCM sheath thermistor wire is broken.

Purpose of Error
To ensure that FCM detector thermistor performs properly.
Detection

Method of Error
Check the error information of the driver board in ready status or operating.
Detection

1. Ready Status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.

Error Message FCM sheath thermistor error

Analysis cannot be continued because the FCM sheath thermistor wire is broken. Remove
Action Message
the rack and tubes from the analyzer and turn off the analyzer power.

HELP [Execute] None.


Process Clear the error after turning off the analyzer power.

Error Code 127600 Judgment Control Unit


HELP Screen Key [Accept] (grayout)
Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-24 March 2015


6.5.2.21 Environment temperature thermistor error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔


Alarm Sounds Analysis Stop Reanalysis Block (Fixed)
Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 1
Measurement Twin Model: Sampler Analysis Stop 1

Important Notice None

Meaning Environment Thermistor is defective or its wire is broken.

Purpose of Error
To ensure that temperature adjustment of each heater performs properly.
Detection

Method of Error
Check the error information of the driver board in ready status or operating.
Detection

1. Ready Status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.

Error Message Environment Temperature thermistor error

Analysis cannot be continued because the environment temperature thermistor wire is


Action Message
broken. Remove the rack and tubes from the analyzer and turn off the analyzer power.

HELP [Execute] None.


Process Clear the error after turning off the analyzer power.

Error Code 127900 Judgment Control Unit


HELP Screen Key [Accept] (grayout)
Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-25 March 2015


6.5.2.22 APD thermistor error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Block (Fixed)


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 1
Measurement Twin Model: Sampler Analysis Stop 1

Important Notice None

Meaning APD Thermistor for Temperature Control is defective or its wire is broken.

Purpose of Error
To ensure that APD thermistor for temperature control performs properly.
Detection

Method of Error
Check the error information of the driver board in ready status or operating.
Detection

1. Ready Status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.

Error Message APD thermistor error

Analysis cannot be continued because the APD thermistor wire is broken. Remove the
Action Message
rack and tubes from the analyzer and turn off the analyzer power.

HELP [Execute] None.


Process Clear the error after turning off the analyzer power.

Error Code 127630 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-26 March 2015


6.5.2.23 Temperature stabilizing error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔


Alarm Sounds Reanalysis Reanalysis NA
Process When Error
Occurred Sampler Single Model: NA
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

Meaning Temperature of each unit is not stabilized.

Purpose of Error
To ensure optimal temperature for analysis
Detection

Method of Error Monitor temperature of each unit, when performing temperature stabilization sequence of
Detection the analyzer.

When the temperature of each unit does not reach the following temperatures within 15
minutes after starting temperature stabilization.
1. 41°C Reaction Chamber: 41°C
Conditions of Error 2. 41°C Reagent Heater: 42°C
Detection
3. 34°C Reaction Chamber (only Type A) : 34°C
4. 34°C Reagent Heater (only Type A): 34°C
5. FCM detector Heater: 42°C

Error Message Temperature stabilizing error

Failed to stabilize temperature in each unit. Remove the rack and tubes from the analyzer
Action Message
and turn off the main analyzer power.

HELP [Execute] None.


Process Clear the error after turning off the analyzer power.

Error Code 129040 Judgment Control Unit


HELP Screen Key [Accept] (grayout)
Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-27 March 2015


6.5.3 Reagent and Chamber Error

6.5.3.1 Out of CELLPACK DCL

Type A / Type B (Only when Reserve Tank and/or RU is not


Analyzer
Applicable connected.)

Sampler NA

Error Category Maintenance

Data Mask ✔ Invalid Analysis ✔


Alarm Sounds 3 Reanalysis NA
Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Suspend

Important Notice None

Meaning Reagent container is empty.

Purpose of Error
Prevent from in flowing air bubbles into the reagent chamber.
Detection

Method of Error Check the Air Bubble Sensor of CELLPACK DCL Aspiration Line during aspirating
Detection CELLPACK DCL.

Conditions of Error
When detecting air bubbles
Detection

Error Message Out of CELLPACK DCL

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 131029 Judgment Sequence


HELP Screen Key [Execute]
Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key Reagent replacement procedure appears.

201435

XN-1000/2000 S/M 6-28 March 2015


6.5.3.2 Out of CELLSHEATH(C)

Type A / Type B (Only for XN-21/XN-11. When Reserve Tank is


Analyzer
Applicable not connected.)

Sampler NA

Error Category Maintenance

Data Mask ✔ Invalid Analysis ✔


Alarm Sounds 3 Reanalysis NA
Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Suspend

Important Notice None

Meaning Reagent container is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the Air Bubble Sensor of CELLSHEATH(C) Aspiration Line during aspirating
Detection CELLPACK DCL.

Conditions of Error
When detecting air bubbles
Detection

Error Message Out of CELLSHEATH(C)

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 132070 Judgment Sequence


HELP Screen Key [Execute]
Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key Reagent replacement procedure appears.

XN-1000/2000 S/M 6-29 March 2015


6.5.3.3 Out of SULFOLYSER

Type A / Type B (Only when Reserve Tank and/or RU is not


Analyzer
Applicable connected.)

Sampler NA

Error Category Maintenance

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds 3 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Suspend

Important Notice None

Meaning Reagent container is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the Air Bubble Sensor of SULFOLYSER Aspiration Line(SLS) during aspirating
Detection SULFOLYSER.

Conditions of Error
When detecting air bubbles.
Detection

Error Message Out of SULFOLYSER

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 134000 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-30 March 2015


6.5.3.4 Out of Lysercell WNR

Type A / Type B (Only when Reserve Tank is not connected.)


Analyzer
Applicable
Sampler NA

Error Category Maintenance

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds 3 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Suspend

Important Notice None

Meaning Reagent container is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the Air Bubble Sensor of Fluorocell WNR Aspiration Line(WNR-2) during aspirating
Detection Fluorocell WNR.

Conditions of Error
When detecting air bubbles.
Detection

Error Message Out of Lysercell WNR

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 133200 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-31 March 2015


6.5.3.5 Out of Fluorocell WNR

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis ✔

Alarm Sounds 3 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Suspend

Important Notice None

Meaning Reagent container is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the number of remaining tests which is recorded in RFID Tag of Fluorocell WNR
Detection during aspirating Fluorocell WNR.

Conditions of Error
The number of remaining tests is zero.
Detection

Error Message Out of Fluorocell WNR

Action Message Open the Fluorocell WNR cover and install new Fluorocell WNR.

None
HELP [Execute]
The error clears when Fluorocell WNR with remaining 1 or number of tests set to the dye
Process
holder.

Error Code 135200 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-32 March 2015


6.5.3.6 Out of Lysercell WDF

Analyzer Type A / Type B (Only when Reserve Tank RU is not connected.)


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis (✔)

Alarm Sounds 3 Reanalysis NA


Process When Error Sampler Single Model: Sampler Analysis Suspend
Occurred Measurement Twin Model: Sampler Analysis Suspend

Analyzable when only CBC, RET, and PLT-F are specified as


Important Notice
discrete and also with manual mode

Meaning Reagent container is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check Air Bubbles Sensor of Lysercell WDF Aspiration Line(WDF-2) during aspirating
Detection Lysercell WDF.

Conditions of Error The number of Lysercell WDF diaphragm pump operations is over 175 times after the
Detection Float Switch of WDF chamber moves to lower position.

Error Message Out of Lysercell WDF

Discrete except DIFF can be analyzed in manual mode. Press [Execute] key. The reagent
Action Message
replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 133140 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-33 March 2015


6.5.3.7 Out of Fluorocell WDF

Analyzer Type A / Type B (Only when Reserve Tank is not connected.)


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis (✔)

Alarm Sounds 3 Reanalysis NA


Process When Error Sampler Single Model: Sampler Analysis Suspend
Occurred Measurement Twin Model: Sampler Analysis Suspend

Analyzable when only CBC, RET, and PLT-F are specified as


Important Notice
discrete and also with the manual mode

Meaning Reagent container is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the number of remaining tests recorded in RFID Tag of Fluorocell WDF during
Detection aspirating Fluorocell.

Conditions of Error
The number of remaining tests is zero.
Detection

Error Message Out of Fluorocell WDF

Discrete except DIFF can be analyzed in manual mode. Open the Fluorocell WDF cover
Action Message
and install new Fluorocell WDF.

None.
HELP [Execute]
Process The error clears when Fluorocell WNR with 1 or more tests remaining set to the dye
holder.

Error Code 133150 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-34 March 2015


6.5.3.8 Out of Lysercell WPC

Analyzer Type A (Only when Reserve Tank is not connected.)


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis (✔)

Alarm Sounds 3 Reanalysis NA


Process When Error Sampler Single Model: Sampler Analysis Suspend
Occurred Measurement Twin Model: Sampler Analysis Suspend

Analyzable when only CBC, RET, and PLT-F are specified as


Important Notice
discrete and also with the manual mode

Meaning Reagent container is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check Air Bubbles Sensor of Lysercell WPC Aspiration Line(WPC-2) during aspirating
Detection Lysercell WPC.

Conditions of Error
When detecting air bubbles
Detection

Error Message Out of Lysercell WPC

Discrete except WPC can be analyzed in manual mode. Press [Execute] key. The reagent
Action Message
replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 133230 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent Replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-35 March 2015


6.5.3.9 Out of Fluorocell WPC

Analyzer Type A
Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis (✔)

Alarm Sounds 3 Reanalysis NA


Process When Error Sampler Single Model: Sampler Analysis Suspend
Occurred Measurement Twin Model: Sampler Analysis Suspend

Analyzable when only CBC, DIFF, RET, and PLT-F are specified
Important Notice
as discrete and also with the manual mode

Meaning Reagent container is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the number of remaining tests recorded in RFID Tag of Fluorocell WPC during
Detection aspirating Fluorocell.

Conditions of Error
The number of remaining tests is zero.
Detection

Error Message Out of Fluorocell WPC

Discrete except WPC can be analyzed in manual mode. Open the Fluorocell WPC cover
Action Message
and install new Fluorocell WPC.

None.
HELP [Execute]
Process The error clears when Fluorocell WPC with 1 or more tests remaining set to the dye
holder.

Error Code 135230 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-36 March 2015


6.5.3.10 Out of CELLPACK DFL

Type A / Type B (Only when Reserve Tank and/or RU is not


Analyzer connected. and when the configurations of only RET & PLT
Applicable analysis.)

Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis (✔)

Alarm Sounds 3 Reanalysis NA


Process When Error Sampler Single Model: Sampler Analysis Suspend
Occurred Measurement Twin Model: Sampler Analysis Suspend

Analyzable when only CBC,DIFF and WPC are specified as


Important Notice
discrete and also with the manual mode

Meaning Reagent container is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check Air Bubbles Sensor of CELLPACK DFL Aspiration Line (DFL) during aspirating
Detection CELLPACK DFL.

Conditions of Error
When detecting air bubbles
Detection

Error Message Out of CELLPACK DFL

CBC and DIFF can be analyzed in manual mode. Press [Execute] key. The reagent
Action Message
replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 131210 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent Replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-37 March 2015


6.5.3.11 Out of Fluorocell RET

Type A / Type B
Analyzer
Applicable (Occurring only with the configuration of RET analysis)

Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis (✔)

Alarm Sounds 3 Reanalysis NA


Process When Error Sampler Single Model: Sampler Analysis Suspend
Occurred Measurement Twin Model: Sampler Analysis Suspend

Analyzable when only CBC, DIFF, PLT-F and WPC are specified
Important Notice
as discrete and also with the manual mode

Meaning Reagent container is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the number of remaining tests recorded in RFID Tag of Fluorocell RET during
Detection aspirating Fluorocell RET.

Conditions of Error
The number of remaining tests is zero.
Detection

Error Message Out of Fluorocell RET

Discrete except RET can be analyzed in manual mode. Open the Fluorocell RET cover
Action Message
and install new Fluorocell RET.

HELP [Execute] None.


Process The error clears when Fluorocell RET with 1 or more tests remaining set to the dye holder.

Error Code 135210 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-38 March 2015


6.5.3.12 Out of Fluorocell PLT

Type A / Type B
Analyzer
Applicable (Occurring only with the configuration of PLT-F analysis)

Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis (✔)

Alarm Sounds 3 Reanalysis NA


Process When Error Sampler Single Model: Sampler Analysis Suspend
Occurred Measurement Twin Model: Sampler Analysis Suspend

Analyzable when only CBC, DIFF, RET and WPC are specified
Important Notice
as discrete and also with the manual mode

Meaning Reagent container is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the number of remaining tests recorded in RFID Tag of Fluorocell PLT during
Detection aspirating Fluorocell PLT.

Conditions of Error
The number of remaining tests is zero.
Detection

Error Message Out of Fluorocell PLT

Discrete except PLT-F can be analyzed in manual mode. Open the Fluorocell PLT cover
Action Message
and install new Fluorocell PLT.

HELP [Execute] None.


Process The error clears when Fluorocell PLT with 1 or more tests remaining set to the dye holder.

Error Code 135220 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-39 March 2015


6.5.3.13 Out of CELLPACK DCL

Analyzer Type A / Type B (Only when Reserve Tank is not connected.)


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds 2 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reserve tank is almost empty.

Purpose of Error
Prompt the user to exchange the reagent.
Detection

Method of Error Check Air Bubbles Sensor of CELLPACK DCL Aspiration Line (DCL) and the double float
Detection switch of CELLPACK DCL reserve tank during aspirating CELLPACK DCL.

Occrred when either of following case is satisfied.


Conditions of Error (1) When the Upper Switch of the double float switch is in the lower position, and the
Detection Lower Switch of the double float switch is in the upper position.
(2) When detecting air bubbles on Reserve Tank.

Error Message Out of CELLPACK DCL

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 131035 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-40 March 2015


6.5.3.14 Out of CELLSHEATH(C)

Type A / Type B (Only for XN-21/XN-11. Only when Reserve Tank


Analyzer
Applicable is not connected.)

Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds 2 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reserve tank is almost empty.

Purpose of Error
Prompt the user to exchange the reagent.
Detection

Method of Error Check Air Bubbles Sensor of CELLPACK DCL Aspiration Line (DCL) and the double float
Detection switch of CELLPACK DCL reserve tank during aspirating CELLPACK DCL.

Occrred when either of following case is satisfied.


Conditions of Error (1) When the Upper Switch of the double float switch is in the lower position, and the
Detection Lower Switch of the double float switch is in the upper position.
(2) When detecting air bubbles on Reserve Tank.

Error Message Out of CELLSHEATH(C)

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 132071 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-41 March 2015


6.5.3.15 Out of SULFOLYSER

Analyzer Type A / Type B (Only when Reserve Tank is not connected.)


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA


Alarm Sounds 2 Reanalysis NA
Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reserve tank is almost empty.

Purpose of Error
Prompt the user to exchange the reagent.
Detection

Method of Error Check the float switch of SULFOYLYSER reserve tank and the cycle count of
Detection SULFOYLYSER diaphragm pump operations during aspirating SULFOLYSER.

Conditions of Error The SLS chamber float switch is in the lower position and the number of SULFOLYSER
Detection diaphragm pump operations is lower than 60 after the Float Switch goes down.

Error Message Out of SULFOLYSER

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 134005 Judgment Sequence


HELP Screen Key [Execute]
Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-42 March 2015


6.5.3.16 Out of Lysercell WNR

Analyzer Type A / Type B (Only when Reserve Tank is not connected.)


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds 2 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reserve tank is almost empty.

Purpose of Error
Prompt the user to exchange the reagent.
Detection

Method of Error Check the Float Switch of WNR chamber and the cycle count of Lysercell WNR
Detection diaphragm pump operations during aspirating Lysercell WNR.

Conditions of Error The Float Switch is in the lower position and the number of Lysercell WNR diaphragm
Detection pump operations is lower than 175 after the float switch goes down.

Error Message Almost out of Lysercell WNR

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 133205 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information Maintenance Menu Reagent replacement

The reagent replacement procedure appears


Show Detailed Procedure Key

XN-1000/2000 S/M 6-43 March 2015


6.5.3.17 Out of Lysercell WDF

Analyzer Type A / Type B (Only when Reserve Tank is not connected.)


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds 2 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reserve tank is almost empty.

Purpose of Error
Prompt the user to exchange the reagent.
Detection

Method of Error Check the Float Switch of WDF chamber and the cycle count of Lysercell WDF diaphragm
Detection pump operations during aspirating Lysercell WNR.

Conditions of Error The Float Switch is in the lower position and the number of Lysercell WDF diaphragm
Detection pump operations is lower than 175 after the float switch goes down.

Error Message Out of Lysercell WDF

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 133145 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-44 March 2015


6.5.3.18 Out of Lysercell WPC

Analyzer Type A (Only when Reserve Tank is not connected.)


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds 2 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reserve tank is almost empty.

Purpose of Error
Prompt the user to exchange the reagent.
Detection

Method of Error Check the Float Switch of WPC chamber and the cycle count of Lysercell WPC diaphragm
Detection pump operations during aspirating Lysercell WPC.

Conditions of Error The Float Switch is in the lower position and the number of Lysercell WPC diaphragm
Detection pump operations is lower than 175 after the float switch goes down.

Error Message Out of Lysercell WPF

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 133235 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-45 March 2015


6.5.3.19 Out of CELLPACK DFL

Analyzer Type A/Type B (Only when Reserve Tank is not connected.)


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds 2 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reserve tank is almost empty.

Purpose of Error
Prompt the user to exchange the reagent.
Detection

Method of Error Check the Float Switch of CELLPACK DFL chamber and the cycle count of CELLPACK
Detection DFL diaphragm pump operations during aspirating CELLPACK DFL.

Conditions of Error The Float Switch is in the lower position and the number of CELLPACK DFL diaphragm
Detection pump operations is lower than 60 after the float switch goes down.

Error Message Out of CELLPACK DFL

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 131215 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-46 March 2015


6.5.3.20 Reservoir tank is empty (CELLSHEATH(C))

Type A/Type B (Only for XN-21/XN-11. Only when Reserve Tank


Analyzer
Applicable is not connected.)

Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds 3 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reserve tank is empty.

Purpose of Error
Prevent from in flowing air bubbles into the reagent chamber.
Detection

Method of Error
Check the Air Bubble Sensor of CELLSHEATH(C) Aspiration Line during aspirating c.
Detection

When either of following condition is met.


Conditions of Error - The Float Switch is in the lower position
Detection - When detecting air bubbles
- When completing CELLSHEATH(C) draining

Error Message Reservoir tank is empty (CELLSHEATH(C))

Press [Execute]. The reagent replacement screen will appear. About 8 minutes will be
Action Message
required for CELLSHEATH(C) reagent replacement.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 132072 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-47 March 2015


6.5.3.21 Reservoir tank is empty (SULFOLYSER)

Analyzer Type A/Type B (Only when Reserve Tank is not connected.)


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds 3 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reserve tank is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the Float Switch of SULFOLYSER chamber and the cycle count of SULFOLYSER
Detection diaphragm pump operations during aspirating SULFOLYSER.

When either of following condition is met.


- The Float Switch is in the lower position
Conditions of Error
Detection - The number of SULFOLYSER diaphragm pump operations is over 60 times after the
Float Switch of SULFOLYSER chamber moves to lower position.
- When completing SULFOLYSER draining

Error Message Reservoir tank is empty (SULFOLYSER)

Action Message Press [Execute]. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 134006 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-48 March 2015


6.5.3.22 Reservoir tank is empty (Lysercell WNR)

Analyzer Type A/Type B (Only when Reserve Tank is not connected.)


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds 3 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reserve tank is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the Float Switch of Lysercell WNR chamber and the cycle count of Lysercell WNR
Detection diaphragm pump operations during aspirating Lysercell WNR.

When either of following condition is met.


- The Float Switch is in the lower position
Conditions of Error
Detection - The number of Lysercell WNR diaphragm pump operations is over 175 times after the
Float Switch of Lysercell WNR chamber moves to lower position.
- When completing Lysercell WNR draining

Error Message Reservoir tank is empty (Lysercell WNR)

Action Message Press [Execute]. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 133206 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-49 March 2015


6.5.3.23 Reservoir tank is empty (Lysercell WDF)

Analyzer Type A/Type B (Only when Reserve Tank is not connected.)


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds 3 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reserve tank is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the Float Switch of Lysercell WDF chamber and the cycle count of Lysercell WDF
Detection diaphragm pump operations during aspirating Lysercell WDF.

When either of following condition is met.


- The Float Switch is in the lower position
Conditions of Error
Detection - The number of Lysercell WDF diaphragm pump operations is over 175 times after the
Float Switch of Lysercell WDF chamber moves to lower position.
- When completing Lysercell WDF draining

Error Message Reservoir tank is empty (Lysercell WDF)

Action Message Press [Execute]. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 133146 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-50 March 2015


6.5.3.24 Reservoir tank is empty (Lysercell WPC)

Analyzer Type A/Type B (Only when Reserve Tank is not connected.)


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds 3 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reserve tank is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the Float Switch of Lysercell WPC chamber and the cycle count of Lysercell WPC
Detection diaphragm pump operations during aspirating Lysercell WPC.

When either of following condition is met.


- The Float Switch is in the lower position
Conditions of Error
Detection - The number of Lysercell WPCdiaphragm pump operations is over 60 times after the
Float Switch of Lysercell WPC chamber moves to lower position.
- When completing Lysercell WPC draining

Error Message Reservoir tank is empty (Lysercell WPC)

Action Message Press [Execute]. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 133236 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-51 March 2015


6.5.3.25 Reservoir tank is empty (CELLPACK DFL)

Analyzer Type A/Type B (Only when Reserve Tank is not connected.)


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds 3 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reserve tank is empty.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the Float Switch of CELLPACK DFL chamber and the cycle count of CELLPACK
Detection DFL diaphragm pump operations during aspirating CELLPACK DFL.

When either of following condition is met.


- The Float Switch is in the lower position
Conditions of Error
Detection - The number of Lysercell WPCdiaphragm pump operations is over 60 times after the
Float Switch of Lysercell WPC chamber moves to lower position.
- When completing CELLPACK DFL draining

Error Message Reservoir tank is empty (CELLPACK DFL)

Action Message Press [Execute]. The reagent replacement screen will appear.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 131216 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key The reagent replacement procedure appears

XN-1000/2000 S/M 6-52 March 2015


6.5.3.26 Out of diluted CELLPACK DST

Type A / Type B (Only when Reserve Tank is not connected.)


Analyzer
Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds 3 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reagent container in RU is empty.

Purpose of Error
To ensure the necessary reagent volume from RU
Detection

Method of Error
Check the reagent remaining volume information from RU.
Detection

Conditions of Error When diluted CELLPACK DST in RU is empty. (Error will not occur when “Ignore RU-20
Detection error” is selected in service setting/

Error Message Out of diluted CELLPACK DST

Action Message Check RU. Please wait if RU is preparing reagent.

HELP [Execute] None.


Process The error clears when diluted CELLPACK DST is sufficient volume.

Error Code 131036 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None.

XN-1000/2000 S/M 6-53 March 2015


6.5.3.27 RU has stopped supplying reagent

Type A / Type B (Only when Reserve Tank is not connected.)


Analyzer
Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds 3 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning Reagent container in RU is empty.

Purpose of Error
To ensure the necessary reagent volume from RU
Detection

Method of Error
Check the reagent information from RU.
Detection

Conditions of Error When reagetnt supply from RU is impossible. (Error will not occur when “Ignore RU-20
Detection error” is selected in service setting/

Error Message RU has stopped supplying reagent

Action Message Check RU.

HELP [Execute] None.


Process The error clears when reagent supply from RU is possible.

Error Code 131037 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None.

XN-1000/2000 S/M 6-54 March 2015


6.5.3.28 CELLPACK DCL aspiration error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔

Alarm Sounds 3 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the float-switch in the Reagent Chamber in the Analyzer right after the reagent
Detection replenishing during analyzing.

Conditions of Error
When the float switch is in the lower position.
Detection

Error Message CELLPACK DCL aspiration error

Action Message Check the reagent level and press Execute key. Reagent replenishment will be executed.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 131020 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key None.

XN-1000/2000 S/M 6-55 March 2015


6.5.3.29 FCM sheath aspiration error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔


Alarm Sounds Analysis Stop Reanalysis NA
Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the float-switch in the FCM Sheath Chamber in the analyzer right after the reagent
Detection replenishing during analyzing.

Conditions of Error
When the float switch is in the lower position.
Detection

Error Message FCM sheath aspiration error

Action Message Check the reagent level and press [Execute] key. Reagent replacement will be executed.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 132019 Judgment Sequence


HELP Screen Key [Execute]
Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key None.

XN-1000/2000 S/M 6-56 March 2015


6.5.3.30 RBC sheath fluid aspiration error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔


Alarm Sounds Analysis Stop Reanalysis NA
Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the float-switch in the RBC Sheath Chamber in the analyzer right after the reagent
Detection replenishing during analyzing.

Conditions of Error
When the float switch is in the lower position.
Detection

Error Message RBC sheath fluid aspiration error

Action Message Check the reagent level and press [Execute] key. Reagent replacement will be executed.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 132020 Judgment Sequence


HELP Screen Key [Execute]
Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key None.

XN-1000/2000 S/M 6-57 March 2015


6.5.3.31 Fluorocell WNR aspiration error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the Air Bubble Sensor of Fluorocell WNR Aspiration Line(WNR-2) and Cover
Detection Sensor of Fluorocell WNR during aspirating Fluorocell WNR.

Conditions of Error 1. When detecting air bubbles.


Detection 2. When detecting the Fluorocell WNR cover opened

Error Message Fluorocell WNR aspiration error

Action Message Check if sufficient reagent remains and press [Execute] key. Replace the reagent.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 135300 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key None.

XN-1000/2000 S/M 6-58 March 2015


6.5.3.32 Fluorocell WDF aspiration error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis (✔)


Alarm Sounds Analysis Stop Reanalysis NA
Process When Error Sampler Single Model: Sampler Analysis Suspend
Occurred Measurement Twin Model: Sampler Analysis Continue / Suspend

Analyzable when CBC, RET and PLT-F are specified as discrete


Important Notice
and also in manual mode.

The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the Air Bubbles Sensor of the Fluorocell WDF Reagent Line(WDF-1) and the Cover
Detection Sensor of the Fluorocell WDF during aspirating Fluorocell WDF.

Conditions of Error 1. When detecting air bubbles.


Detection 2. When detecting the opened Fluorocell WDF cover.

Error Message Fluorocell WDF aspiration error

Discrete except DIFF can be analyzed in manual mode. Check the reagent level and
Action Message
press [Execute] key. Reagent replacement will be executed.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 131100 Judgment Sequence


HELP Screen Key [Execute]
Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key None.

XN-1000/2000 S/M 6-59 March 2015


6.5.3.33 Fluorocell WPC aspiration error

Analyzer Type A
Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis (✔)


Alarm Sounds Analysis Stop Reanalysis NA
Process When Error Sampler Single Model: Sampler Analysis Suspend
Occurred Measurement Twin Model: Sampler Analysis Continue / Suspend

Analyzable when CBC, DIFF, RET and PLT-F are specified as


Important Notice
discrete and also in manual mode.

The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the Air Bubbles Sensor of the Fluorocell WPC Reagent Line(WPC-1) and the
Detection Cover Sensor of the Fluorocell WPC during aspirating Fluorocell WPC.

Conditions of Error 1. When detecting air bubbles.


Detection 2. When detecting the opened Fluorocell WPC cover.

Error Message Fluorocell WPC aspiration error

Discrete except WPC can be analyzed in manual mode. Check the reagent level and
Action Message
press [Execute] key. Reagent replacement will be executed.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 135330 Judgment Sequence


HELP Screen Key [Execute]
Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key None.

XN-1000/2000 S/M 6-60 March 2015


6.5.3.34 Fluorocell RET aspiration error

Type A / Type B (Occurring only with the configuration of RET


Analyzer
Applicable analysis.)

Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis (✔)


Alarm Sounds Analysis Stop Reanalysis NA
Process When Error Sampler Single Model: Sampler Analysis Suspend
Occurred Measurement Twin Model: Sampler Analysis Continue / Suspend

Analyzable when CBC, DIFF, PLF-T and WPC are specified as


Important Notice
discrete and also in manual mode.

The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the Air Bubble Sensor of the Fluorocell RET Reagent Line(RET) and the Cover
Detection Sensor of the Fluorocell RET during aspirating the Fluorocell RET.

Conditions of Error 1. When detecting air bubbles


Detection 2. When detecting the opened Fluorocell RET cover.

Error Message Fluorocell RET aspiration error

Discrete except RET can be analyzed in manual mode. Check the reagent level and press
Action Message
[Execute] key. Reagent replacement will be executed.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 135310 Judgment Sequence


HELP Screen Key [Execute]
Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key None.

XN-1000/2000 S/M 6-61 March 2015


6.5.3.35 Fluorocell PLT aspiration error

Type A / Type B (Occurring only with the configuration of PLT-F


Analyzer
Applicable analysis.)

Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis (✔)


Alarm Sounds Analysis Stop Reanalysis NA
Process When Error Sampler Single Model: Sampler Analysis Suspend
Occurred Measurement Twin Model: Sampler Analysis Continue / Suspend

Analyzable when CBC, DIFF, PLT-F and WPC are specified as


Important Notice
discrete and also in manual mode.

The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.

Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection

Method of Error Check the Air Bubble Sensor of the Fluorocell PLT Reagent Line(PLT) and the Cover
Detection Sensor of the Fluorocell RET during aspirating the Fluorocell PLT.

Conditions of Error 1. When detecting air bubbles


Detection 2. When detecting the opened Fluorocell RET cover.

Error Message Fluorocell PLT aspiration error

Discrete except PLT-F can be analyzed in manual mode. Check the reagent level and
Action Message
press [Execute] key. Reagent replacement will be executed.

HELP [Execute] The reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 135320 Judgment Sequence


HELP Screen Key [Execute]
Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key None.

XN-1000/2000 S/M 6-62 March 2015


6.5.3.36 Waste chamber 1 not draining

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None.

Meaning The waste chamber 1 is unable to drain.

Purpose of Error
To ensure that the fluid used in the analysis has been drained.
Detection

Method of Error
Check the float switch position of the waste chamber 1 after the draining is completed.
Detection

Conditions of Error
When the Float Switch is in upper position.
Detection

Error Message Waste chamber 1 not draining

Check the drain tube and press [Execute] key. Execute the waste chamber drain
Action Message
sequence.

HELP [Execute] Executing the waste chamber draining.


Process The error clears when waste chamber draining completed properly.

Error Code 141001 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu The waste chamber drain sequence

Show Detailed Procedure Key None.

XN-1000/2000 S/M 6-63 March 2015


6.5.3.37 Waste chamber 2 not draining

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None.

Meaning The waste chamber 2 (for sheath) is unable to drain.

Purpose of Error
To ensure that the fluid used on the analysis has been drained.
Detection

Method of Error
Check the Float Switch position of the waste chamber 2 after the draining is completed.
Detection

Conditions of Error
When the float switch is in lower position.
Detection

Error Message Waste chamber 2 not draining

Check the drain tube and press [Execute] key. Execute the waste chamber drain
Action Message
sequence.

HELP [Execute] Executing the waste chamber draining.


Process The error clears when waste chamber draining completed properly.

Error Code 141002 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu The waste chamber drain sequence

Show Detailed Procedure Key None.

XN-1000/2000 S/M 6-64 March 2015


6.5.3.38 RBC/HGB chamber not draining

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Suspend Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None.

Meaning Draining from the RBC/HGB Chamber does not operate properly.

Purpose of Error
To ensure draining from the RBC/HGB chamber.
Detection

Check the blank converted values and the following sample converted values when
draining from RBC/HGB chamber.
Method of Error Sample converted value: The converted value of the chamber transmitted light after
Detection draining residual liquid from the RBC/HGB chamber.
Sample converted value: The converted value of the chamber transmitted light after
draining rinse solution from the RBC/HGB chamber.

If any of the following conditions are met, it becomes an error. (When the reaction
chamber drain for maintenance executes, judging with condition 1 only.)
1. Sample converted value 1 < 15000
Conditions of Error
Detection 2. Sample converted value 2 < 15000
3. Sample converted value 1 - Blank Convert value < 12000
4. Sample converted value 2 - Blank Convert value < 12000

Error Message RBC/HGB chamber not draining

Action Message Press [Execute] key. Execute the reaction chamber drain sequence.

HELP [Execute] Executing the reaction chamber draining.


Process The error clears when the reaction chamber is completed draining properly.

Error Code 141010 Judgment Control Unit

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu The waste chamber drain sequence

Show Detailed Procedure Key None.

XN-1000/2000 S/M 6-65 March 2015


6.5.3.39 Waste container is full

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

Meaning The drain container is full.

Purpose of Error
To ensure draining.
Detection

Method of Error Check the waste tank with the filled sensor (float switch) in ready status and when draining
Detection completed.

In ready status: When the float switch remains in the upper position for continuously 3
seconds
Conditions of Error
Detection Draining complete
d: When the float switch is in upper position.

Error Message Waste container is full

Action Message Replace the waste container and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 141190 Judgment Sequence

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-66 March 2015


6.5.3.40 Water leak detected

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔

Alarm Sounds 2 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 2
Measurement Twin Model: Sampler Analysis Stop 2

Important Notice None

Meaning Water leaks inside of the analyzer

Purpose of Error
Detect water leak early.
Detection

Method of Error
Check the Water Leak Sensor all the time
Detection

Conditions of Error
When detecting the water leak for continuously 3 seconds
Detection

Error Message Water leak detected

Analysis cannot be continued because water leak detected. Check the analyzer and turn
Action Message
the main power of the analyzer off.

HELP [Execute] None.


Process The error clears when turn off the analyzer.

Error Code 429200 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-67 March 2015


6.5.3.41 Water leak detected (analysis not possible)

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔

Alarm Sounds 3 Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 2
Measurement Twin Model: Sampler Analysis Stop 2

Important Notice None

Meaning Water leaks inside of the analyzer

Purpose of Error
Detect water leak early.
Detection

Method of Error
Check the Water Leak Sensor all the time
Detection

Conditions of Error When “Continue using analyzer when leak is detected” check is OFF and detecting the
Detection water leak for continuously 3 seconds

Error Message Water leak detected (analysis not possible)

Action Message Water leak detected in analyzer. Turn off the main power and check the analyzer.

HELP [Execute] None.


Process The error clears when turn off the analyzer.

Error Code 429200 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-68 March 2015


6.5.3.42 Water leak sns error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA

Process When Error Sampler Single Model: Sampler Analysis Stop 2


Occurred Measurement Twin Model: Sampler Analysis Stop 2

Depending on the service setting, the error not detected.


Important Notice
Stop the analyzer sequence immediately.

Meaning Water leak sensor does not operate properly.

Purpose of Error
To ensure that the water leak sensor works properly.
Detection

Method of Error
Check the water leak sensor error from the driver board always.
Detection

Conditions of Error
When detecting the water leak sensor error for continuously 3 seconds
Detection

Error Message Water leak sns error

Analysis cannot be continued because water leak sns error. Remove the rack and tubes
Action Message
from the analyzer and turn the main power of the analyzer off.

HELP [Execute] None.


Process The error clears when turn off the analyzer.

Error Code 429201 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-69 March 2015


6.5.4 Motor Error

6.5.4.1 WB aspiration motor error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Suspend

Important Notice None

Meaning The motor does not rotate smoothly because the load of WB aspiration pump is too large.

Purpose of Error
To ensure the aspiration / discharge volume of WB samples.
Detection

Check the status of Upper Position Sensor (Photo-Interrupter) when the WB aspiration
Method of Error
motor operation is completed. Only on test operation, check the number of remaining
Detection
pulses.

Conditions of Error
When the upper position sensor (photo-interrupter) does not shade the light.
Detection

Error Message WB aspiration motor error

Action Message Press [Execute] key. WB aspiration motor test will start.

HELP [Execute] Executing WB aspiration motor test.


Process The error clears when the WB aspiration test is completed properly.

211000

Error Code parameter 1: The Judgment Sequence


number of remaining
pulses
Detailed Information
HELP Screen Key [Execute]

Maintenance Menu WB aspiration motor test

Show Detailed Procedure Key none

XN-1000/2000 S/M 6-70 March 2015


6.5.4.2 FCM sheath motor error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Suspend

Important Notice None

The motor does not rotate smoothly because the loads of pushing out of samples and a
Meaning
volumetric syringe on the FCM sheath are too large.

Purpose of Error
To ensure the volume of FCM analysis samples.
Detection

Method of Error Check the status of the Photo-Interrupter Home Position and the number of remaining
Detection pulses when sheath motor operation is completed.

(1) When the sheath motor does not shaded the light with the photo-interrupter of the
Conditions of Error home position.
Detection
(2) Remaining pulses > 4800

Error Message FCM sheath motor error

Action Message Press [Execute] key. Sheath motor test will start.

HELP [Execute] Executing sheath motor test.


Process The error clears when the sheath motor test is completed properly.

212010

Error Code parameter 1: The Judgment Sequence


number of remaining
pulses
Detailed Information
HELP Screen Key [Execute]

Maintenance Menu Sheath motor test

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-71 March 2015


6.5.4.3 RBC sheath motor error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Suspend

Important Notice None

The motor does not rotate smoothly because the loads of pushing out of samples and a
Meaning
volumetric syringe on the RBC sheath are too large.

Purpose of Error
To ensure the volume of RBC series analysis samples.
Detection

Method of Error Check the status of the Photo-Interrupter Home Position and the number of remaining
Detection pulses when sheath motor operation is completed.

(1) When the sheath motor does not shaded the light with the photo-interrupter of the
Conditions of Error home position.
Detection
(2) Remaining pulses > 4800

Error Message RBC sheath motor error

Action Message Press Execute key. Sheath motor test will start.

HELP [Execute] Executing sheath motor test.


Process The error clears when the sheath motor test is completed properly.

212000
Error Code parameter 1: The number of Judgment Sequence
remaining pulses

Detailed Information HELP Screen Key [Execute]

Maintenance Menu Sheath motor test

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-72 March 2015


6.5.4.4 Aspiration unit up-down motor error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 2
Measurement Twin Model: Sampler Analysis Stop 2

Important Notice The analyzer sequence stops immediately as detecting errors.

The motor does not rotate smoothly because the load of the aspiration unit up-down
Meaning
operation is too large.

Purpose of Error
To ensure the aspiration unit up-down operation.
Detection

Check the status of the Upper Position of the Photo-Interrupter when the aspiration unit
Method of Error
up-down motor ascends. Only when operating with the test mode, check the number of
Detection
remaining pulses while the motor stops.

Conditions of Error When the aspiration unit up-down motor does not shaded the light with the upper position
Detection of the photo-interrupter.

Error Message Aspiration unit up-down motor error

Operation could not be continued because aspiration unit up-down motor stopped.
Action Message Remove rack and tubes from analyzer and press [Execute] key. Analyzer initialize will be
executed.

HELP [Execute] Restart the analyzer.


Process The error clears when the analyzer restarts.

212020
Error Code parameter 1: The number of Judgment Sequence
remaining pulses

Detailed Information HELP Screen Key [Execute]

Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-73 March 2015


6.5.4.5 Aspiration unit left-right motor error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

The motor does not rotate smoothly because the load of the aspiration unit left-right
Meaning
operation is too large.

Purpose of Error
To ensure the aspiration unit left-right operation.
Detection

Check the status of the Photo-Interrupter Home Position when the aspiration unit left-right
Method of Error
motor operation is completed. Only when operating with the test mode, check the number
Detection
of remaining pulses when the motor operation completed.

Conditions of Error
When the upper position of the aspiration unit sensor (photo interrupter) is not shaded.
Detection

Error Message Aspiration unit left-right motor error

Action Message Press [Execute] key. Aspiration unit motor test will start.

HELP [Execute] Execute the aspiration unit motor test.


Process The error clears after the aspiration unit motor test is completed properly.

212021

Error Code parameter 1: The Judgment Sequence


number of remaining
pulses
Detailed Information
HELP Screen Key [Execute]

Maintenance Menu Aspiration unit motor test

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-74 March 2015


6.5.5 Sample Aspiration Error

6.5.5.1 Insufficient blood volume

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Caution

Data Mask ✔ Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis Block (Fixed)


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None

Meaning Blood volume is insufficient for the sample analysis.

Purpose of Error
The sampler analysis stops due to insufficient blood volume.
Detection

Method of Error
Check the blood sensor when checking the tube presence during the sampler analysis.
Detection

Conditions of Error When judging that there are tubes by the Sample Tube Detection Sensor, and also when
Detection the blood volume is insufficient by the Sample Blood Detection Sensor.

Error Message Insufficient blood volume

Action Message Insufficient blood volume. Check the sample and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 161200 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-75 March 2015


6.5.5.2 Insufficient blood volume (short sample)

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask ✔ Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis Block (Fixed)


Process When Error Sampler Single Model: (Sampler Analysis Suspend)
Occurred Measurement Twin Model: (Sampler Analysis Suspend)

The error might not be detected when the Aspiration Sensor is


Important Notice
set to “not used.

Meaning The specified volume of blood samples cannot be aspirated during aspirating whole blood.

Purpose of Error
To ensure the WB volume.
Detection

By the blood aspiration sensor, compare between the following conversion value and
blank conversion value while analyzing with WB mode (excluding the background check).
Sample Conversion Value 1:
The value that converted (detected) the turbidity by diluting the piercer tip disposal blood
with EPK before discharging RBC samples
Method of Error
Detection Sample Conversion Value 2:
The value that converted (detected) the turbidity by diluting the whole blood pump
disposal blood with EPK after discharging WBC samples
Sample Conversion Value 3:
The value that converted (detected) the turbidity by diluting the RBC sample blood with
CELLPACK DCL

A = Sample Conversion Value 1 – Blank Conversion Value


B = Sample Conversion Value 2 – Blank Conversion Value
C = A/B x 100
D = Sample Conversion Value 3 - Blank Conersion Value
E = D/B x 100
Conditions of Error F = A/D x 100
Detection When matching with any of following conditions, this error will occur.
- A is 3400 or less
- C is 70 or less ; C is 200 or more
- D is 3400 or less
- E is 70 or less; E is 200 or more
- F is 200 or more

Error Message Insufficient blood volume (short sample)

Insufficient blood volume. Check the sample and press [Accept] key. If the Aspiration
Action Message
Sensor is set to “not used”, this error will not occur.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

XN-1000/2000 S/M 6-76 March 2015


Error Code 161010 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

: 313B037

XN-1000/2000 S/M 6-77 March 2015


6.5.5.3 Blood cannot be aspirated

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask ✔ Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis Block (Fixed)


Process When Error Sampler Single Model: (Sampler Analysis Suspend)
Occurred Measurement Twin Model: (Sampler Analysis Suspend)

The error might not be detected when the Aspiration Sensor is


Important Notice
set to “not used.

Meaning The specified volume of blood samples cannot be aspirated during aspirating whole blood.

Purpose of Error
To ensure the WB volume.
Detection

By the blood aspiration sensor, compare between the following conversion value and
blank conversion value during analyzing with WB mode (excluding the blank check).
Method of Error
Detection Sample Conversion Value 2:
The value that converted (detected) the turbidity by diluting the whole blood pump
disposal blood with EPK after discharging WBC samples

Conditions of Error B = Sample Conversion Value 2 – Blank Conversion Value


Detection When B is 3400 or less

Error Message Blood cannot be aspirated

Failed to correctly aspirate blood. Check the sample and press [Accept] key. If the
Action Message
Aspiration Sensor is set to “not used”, this error will not occur.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 161000 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

A: 201218

XN-1000/2000 S/M 6-78 March 2015


6.5.5.4 Aspiration Sensor error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error Sampler Single Model: Sampler Analysis Suspend
Occurred Measurement Twin Model: Sampler Analysis Continue / Suspend

The error might not be detected when the Aspiration Sensor is


Important Notice
set to “not used.

Meaning The blood aspiration sensor does not operate properly.

Purpose of Error
To ensure that the blood aspiration sensor operates properly.
Detection

Method of Error By the blood aspiration sensor, compare between the following conversion values and
Detection blank conversion values while analyzing with WB mode (excluding the blank check).

When matching either of the following conditions, it becomes error.


Conditions of Error
Blank Convert value < 50
Detection
Blank Convert value > 15000

Error Message Aspiration Sensor error

There is a problem in the Aspiration Sensor. Press [Accept]. By changing the Aspiration
Action Message
Sensor setting to “not used”, analysis can be continued.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 161100 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-79 March 2015


6.5.6 Sampler Operation Errors

6.5.6.1 Rack feed-in error

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Hardware Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 3
Measurement Twin Model: Sampler Analysis Stop 3

Important Notice None

Meaning A rack feed-in error occurs.

Purpose of Error
To ensure that the rack on the feed-in rack table sends into the measurement line.
Detection

Method of Error Check the Rack Feed-in Arrival Sensor and the Feed-in Table Rack Detection Sensor
Detection during operating rack feed-in.

When the rack feed-in motor operates a specified number of pulses, the rack is not be
Conditions of Error
detected by the rack feed-in arrival monitoring sensor and the rack is detected by the rack
Detection
feed-in table rack detecting sensor.

Error Message Rack feed-in error

Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 253030 Judgment Sequence

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-80 March 2015


6.5.6.2 Rack feed-in home position error

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Hardware Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 3
Measurement Twin Model: Sampler Analysis Stop 3

Important Notice None

Meaning A rack feed-in mechanism initialized error occurs.

Purpose of Error To ensure that the rack feed-in unit lever return to the home position after the rack feed-in
Detection operated, and the next rack feed operates properly.

Method of Error Check the Rack Feed-in Home Position Sensor during initializing the rack feed-in
Detection mechanism.

Conditions of Error When the rack feed-in motor operates a specified number of pulses, the rack feed-in
Detection home position sensor is not be shaded.

Error Message Rack feed-in home position error

Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 253031 Judgment Sequence

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-81 March 2015


6.5.6.3 Rack move error(Front belt)

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Hardware Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 3
Measurement Twin Model: Sampler Analysis Stop 3

Important Notice None

Meaning Rack moving error occurs

Purpose of Error
To ensure that the rack is moved to the designated location properly.
Detection

Method of Error Compare a specified number of pulses before operation to the Encoder of belt driving
Detection motor after operation while moving the rack.

Conditions of Error when the following condition is met:


Detection Encoder Value – Specified Number of Pulse > ± 20

Error Message Rack move error(Front belt)

Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 254030 Judgment Sequence

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-82 March 2015


6.5.6.4 Rack move mechanism initialization error(Front belt)

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Hardware Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 3
Measurement Twin Model: Sampler Analysis Stop 3

Important Notice None

Meaning Rack move mechanism initialization error occurs.

Purpose of Error To ensure that the belt profile of the rack move mechanism returns to the home position
Detection and the next rack operates to move properly.

Method of Error Check the Rack Move Mechanism Home Positio Sensor during initializing the rack move
Detection mechanism.

Conditions of Error When the home position sensor of the rack move mechanism is not shaded as the rack
Detection move motor operated a specified number of pulses.

Error Message Rack move mechanism initialization error(Front belt)

Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 254031 Judgment Sequence

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-83 March 2015


6.5.6.5 Rack ejection error

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Hardware Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 3
Measurement Twin Model: Sampler Analysis Stop 3

Important Notice None

Meaning Rack ejection error occurs.

Purpose of Error
To ensure that the rack ejects from the analysis line to the rack ejection table properly.
Detection

Method of Error
Check the Rack Ejection Home Position Sensor during ejecting the rack.
Detection

Conditions of Error When the home position sensor of the rack ejection mechanism is not shaded as the rack
Detection ejection motor operated a specified number of pulses.

Error Message Rack ejection error

Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 255050 Judgment Sequence

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-84 March 2015


6.5.6.6 Rack ejection home position error

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Hardware Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 3
Measurement Twin Model: Sampler Analysis Stop 3

Important Notice None

Meaning Initializing operation error of the rack ejection lever occurs.

Purpose of Error To ensure that the lever of the rack ejection mechanism return to the home position and
Detection the rack ejection mechanism is initialized properly.

Method of Error Check the Rack Ejection Home Position Sensor during initializing the rack ejection
Detection mechanism.

Conditions of Error When the home position sensor of the rack ejection mechanism is not shaded when the
Detection rack ejection motor operates a specified number of pulses.

Error Message Rack ejection home position error

Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 255051 Judgment Sequence

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-85 March 2015


6.5.6.7 Tube presence verification home position error

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Hardware Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 3
Measurement Twin Model: Sampler Analysis Stop 3

Important Notice None

Meaning An initialized operation error occurs during the tube presence verification operating.

Purpose of Error To ensure that the tube rotation mechanism returns to the home position, and rack move
Detection operates properly.

Method of Error Check the Tube Presense Verification Home Position Sensor during initializing the tube
Detection presence verification mechanism.

When the home position sensor of the tube presence verification mechanism is not
Conditions of Error
shaded after the tube presence verification mechanism motor operates a specified
Detection
number of pulses.

Error Message Tube presence verification home position error

Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 259001 Judgment Sequence

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-86 March 2015


6.5.6.8 Feed-in table stopper position error

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Hardware Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 3
Measurement Twin Model: Sampler Analysis Stop 3

Important Notice None

Meaning An error occurs in the feed-in table stopper.

Purpose of Error
To ensure that the feed-in table stopper in upright position.
Detection

Method of Error Check the Feed-in Table Stopper Position Sensor when raising the feed-in table in upright
Detection position.

Conditions of Error When the feed-in table stopper position sensor is not shaded as the feed-in table stopper
Detection motor operates a specified number of pulses.

Error Message Feed-in table stopper position error

Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 253032 Judgment Sequence

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-87 March 2015


6.5.6.9 Ejection table stopper position error

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Hardware Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 3
Measurement Twin Model: Sampler Analysis Stop 3

Important Notice None

Meaning An error occurs in the ejection table stopper position.

Purpose of Error
To ensure that the rack does not break into the analysis line from the ejection table.
Detection

Method of Error
Check the Ejection Table Stopper Position Sensor after the rack ejection is completed.
Detection

Conditions of Error
When the ejection table stopper position monitoring sensor is shaded.
Detection

Error Message Ejection table stopper position error

Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 255052 Judgment Sequence

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-88 March 2015


6.5.6.10 Sampler belt error

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Hardware Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 3
Measurement Twin Model: Sampler Analysis Stop 3

Important Notice None

Meaning The rack exists on the analysis line during operating the analysis line initialization.

Purpose of Error
To ensure that the analysis line initialization operates properly.
Detection

Method of Error Compare the value of a specified number of pulses for operation to the read value by the
Detection encoder on the motor during initializing of the analysis line.

Conditions of Error When the following condition is met:


Detection Encoder Value – a specified number of pulses > ± 50

Error Message Ejection table stopper position error

Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press Accept key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 422102 Judgment Sequence

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-89 March 2015


6.5.6.11 Rack not placed on feed-in table.

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler
NA (Occurring only before sample analysis starts.)
Measurement

Important Notice None

Meaning The rack is not placed on the rack feed-in table during starting the sampler analysis.

Purpose of Error
Prompt the user to set the rack.
Detection

Method of Error
Check the Feed-in Table Rack Detection Sensor.
Detection

Conditions of Error When the rack is not detected by the rack feed-in table rack detection sensor during
Detection mechanism initialization completion after the sampler analysis starts.

Error Message Rack not placed on feed-in table.

Action Message Press [Accept] key, place rack on feed-in table again, and start sampler analysis.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 422022 Judgment Sequence

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-90 March 2015


6.5.6.12 Ejection table is full

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Suspend

Important Notice None

Meaning The rack ejection table is full with racks.

Purpose of Error
Prompt the user to remove racks from the rack ejection table.
Detection

Method of Error
Check the Ejection Table Rack Full Monitoring Sensor.
Detection

Conditions of Error When the racks are detected by the rack ejection table rack full sensor during starting the
Detection rack ejection operation.

Error Message Ejection table is full

Action Message Remove racks from rack ejection table.

None.
HELP [Execute]
The error clears when the racks were not detected by the rack ejection table rack full
Process
sensor.

Error Code 422012 Judgment Sequence

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-91 March 2015


6.5.6.13 Completed sampler analysis stop

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds NA Reanalysis NA


Process When Error
Occurred Sampler
NA (Occurring only when the sampler analysis stops)
Measurement

Important Notice None

Meaning Sampler analysis stop operation is completed.

Purpose of Error Detecting error is not the purpose. .


Detection Purpose to notice the user that the sampler analysis stop operation is complete.

Method of Error
None
Detection

Conditions of Error
When the sampler analysis stop operation is completed.
Detection

Error Message Completed sampler analysis stop

Completed sampler analysis stop. Press Accept key. Will return to sampler analysis ready
Action Message
state.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 422200 Judgment Sequence

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-92 March 2015


6.5.7 Hand Unit and Tube Holder Error

6.5.7.1 Hand up-down error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔


Alarm Sounds Analysis Stop Reanalysis NA
Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 2
Measurement Twin Model: Sampler Analysis Stop 2

Important Notice None

Meaning The hand is placed incorrect position.

Purpose of Error
To ensure that the hand operates at the correct position.
Detection

Method of Error
Check a Hand Position Sensor in a specified timing during sampler analysis.
Detection

It will become an error when the hand position sensor is not shaded in the following timing.
1. Catching a tube from a rack : Hand lower sensor
2. Before mixing operation : Hand upper sensor
Conditions of Error 3. Releasing a tube to a tube holder : Hand middle sensor
Detection 4. Drawing up a tube into an aspiration unit: Hand upper sensor
5. Catching a tube from a tube holder: Hand middle sensor
6. Drawing up a tube : Hand upper sensor
7. Releasing a tube to a rack: Hand lower sensor

Error Message Hand up-down error

Remove the tubes from the analyzer and press [Execute] key. A hand test will be
Action Message
executed.

HELP [Execute] Execute a hand test.


Process The error clears when the hand test is completed properly.

263004
Parameter 1: remaining
Error Code number of pulses Judgment Sequence
Parameter 2: sensor position
1:lower 2:middle 3:Upper
Detailed Information
HELP Screen Key [Execute]

Maintenance Menu Hand Test

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-93 March 2015


6.5.7.2 Hand open/close error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔


Alarm Sounds Analysis Stop Reanalysis NA
Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 2
Measurement Twin Model: Sampler Analysis Stop 2

Important Notice None

Meaning The hand open/close does not operate correctly.

Purpose of Error
To ensure that catch and release tubes operation works properly.
Detection

Method of Error
Check a Hand Open Sensor in a specified timing during sampler analysis.
Detection

It will become an error when the hand status does not meet the following condition.
1. Before catching a tube from a rack: Hand opened
Conditions of Error 2. After catching a tube from a rack: Hand closed
Detection 3. After releasing the tube to a tube holder : Hand opened
4. After catching the tube from the tube holder: Hand closed
5. After releasing the tube to the rack: Hand middle sensor

Error Message Hand open/close error

Remove the tubes from the analyzer and press [Execute] key. A hand test will be
Action Message
executed.

HELP [Execute] Execute a hand test.


Process The error clears when the hand test is completed properly.

Error Code 263008 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Hand Test

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-94 March 2015


6.5.7.3 Mixing error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔


Alarm Sounds Analysis Stop Reanalysis NA
Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 2
Measurement Twin Model: Sampler Analysis Stop 2

Important Notice None

Meaning The mixing mechanism does not operate properly.

Purpose of Error
To ensure that the mixing mechanism operates properly.
Detection

Method of Error Check the Mixing Motor Home Position Sensor and the Mixing Tube Detection Sensor in a
Detection specified timing during sampler analysis.

It is an error when the following condition is met.


1. The mixing motor home position sensor is not shaded at initial mixing operation
Conditions of Error
2. The mixing motor home position sensor is shaded at mixing operation
Detection
3. The Mixing Tube Detection Sensor remains shading at mixing operation
4. The Mixing Tube Detection Sensor remains not shading at mixing operation

Error Message Mixing Error

Action Message Remove the tubes from the analyzer and press Execute key. A hand test will be executed.

HELP [Execute] Execute a hand test.


Process The error clears when the hand test is completed properly.

263005
Parameter 2: sensor position

1: Sensor 29
Error Code 2: Sensor 30 Judgment Sequence
3: Sensor 30
Sensor 29: Mixing motor
Detailed Information home position
Sensor 30: Mixing tube
detection

HELP Screen Key [Execute]

Maintenance Menu Hand Test

Show Detailed Procedure Key None

2015025

XN-1000/2000 S/M 6-95 March 2015


6.5.7.4 Tube pickup error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔


Alarm Sounds Analysis Stop Reanalysis NA
Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 2
Measurement Twin Model: Sampler Analysis Stop 2

Important Notice None

Meaning Sample tubes are not placed to a tube holder at the sampler analysis operation.

Purpose of Error
To ensure that tubes are placed to a tube holder correctly.
Detection

Method of Error Check the Sample Tube Detection Sensor in Main Unit when the tube is picked up into the
Detection tube holder.

Conditions of Error
When tubes are not detected.
Detection

Error Message Tube pickup error

Action Message Remove tubes from analyzer and press [Execute] key. A hand test will be executed.

HELP [Execute] Execute a hand test.


Process The error clears when the hand test is completed properly.

Error Code 264001 Judgment Sequence

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Hand Test

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-96 March 2015


6.5.7.5 Tube holder move error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 2
Measurement Twin Model: Sampler Analysis Stop 2

Important Notice None

Meaning A tube holder is not placed on the correct position during aspirating the samples

Purpose of Error
To ensure that the WB samples can be aspirated. Avoid the risk of injury to the user.
Detection

Method of Error
Check the Tube Holder Home Position Sensor when picking up the tube holder.
Detection

[Sampler Analysis]
It is an error when matching either of the following conditions:
1.The home position sensor is not shaded when a tube is caught from a rack.
Conditions of Error 2.The home position sensor is shaded when the tube is released to tube holder.
Detection 3.The home position sensor is not shaded when the tube is retracted into the aspiration
unit.
4.The home position sensor is shaded when the tube is caught from the tube holder.
5.The home position sensor is not shaded when the tube is picked up.

Error Message Tube holder move error

Remove tubes from analyzer and press [Execute] key. Tube holder motor test will be
Action Message
executed.

HELP [Execute] Execute a tube holder motor test.


Process The error clears when the tube holder motor test is completed properly.

264002
Error Code Parameter 1: remaining Judgment Sequence
number of pulses
Detailed Information HELP Screen Key [Execute]

Maintenance Menu Tube Holder Motor Test

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-97 March 2015


6.5.7.6 Tube return error
Analyzer Type A / Type B
Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When
Error Occurred Sampler Single Model: Sampler Analysis Stop 2
Measurement Twin Model: Sampler Analysis Stop 2

Important Notice None

A tube can't be pulled out from a tube holder when the tube is returned to a rack at
Meaning
the sampler analysis.

Purpose of
To ensure that the tube is picked up from the tube holder properly.
Error Detection

Method of Check the Sample Tube Detection Sensor in Main Unit when the tube holder is is
Error Detection retracted to the Main Unit at the tube return operation.

Conditions of
When the tube is detected.
Error Detection

Error Message Tube return error

Action Remove tubes from analyzer and press [Execute] key. A hand test will be
Message executed.

HELP
Execute hand test.
[Execute]
The error clears when the hand test is completed properly.
Process

Judgme
Error Code 264003 Sequence
nt

Detailed HELP Screen Key [Execute]


Information
Maintenance Menu Hand Test

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-98 March 2015


6.5.7.7 Tube remains in tube holder

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔


Alarm Sounds Warning Reanalysis NA
Process When Error
Occurred Sampler
NA
Measurement

Important Notice None

Tube is remaining in tube holder when switching from manual analysis to sampler
Meaning
analysis.

Purpose of Error
To ensure that there is no sample tube in a tube holder during the sample analysis.
Detection

Method of Error Check the Sample Tube Detection Sensor in Main Unit during switching to Manual mode
Detection or Sampler mode.

Conditions of Error
When detecting a tube.
Detection

Error Message Tube remains in tube holder

Action Message Remove tube from holder and press [Accept] key.

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 264004 Judgment Sequence

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-99 March 2015


6.5.7.8 Two tubes are in tube holder

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None

Two tubes have been placed in set-positions of both a normal sample tube and a micro
Meaning
blood tube of a tube holder during manual analysis.

Purpose of Error
To prevent an operator from picking a wrong sample (analysis result).
Detection

Method of Error Check the Sample Tube Detection Sensor in Main Unit and Micro Collection Tube
Detection Detection Sensor during manual analysis.

Conditions of Error
When detecting both of a normal sample tube and a micro collection tube.
Detection

Error Message Two tubes are in tube holder

Action Message Place only tube to be analyzed in tube holder and press [Accept] key.

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 264005 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-100 March 2015


6.5.7.9 The sample must be remixed.

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None

Sample remixing is necessary because reply information for analysis from host computer
Meaning
is delayed on manual analysis.

Purpose of Error
To prevent insufficient sample mixing.
Detection

Method of Error Check the waiting time after transmitting inquiry information to host computer on manual
Detection analysis.

Conditions of Error
When waiting time is over 60 seconds.
Detection

Error Message The sample must be remixed

Action Message Re-mix the sample and press [Accept].

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 269013 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

201435

XN-1000/2000 S/M 6-101 March 2015


6.5.8 Sample Number / Rack Number Error

6.5.8.1 Failed to read sample number during sampler analysis (Sampler analysis).

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis Block (fixed)


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None

Meaning The barcode reader failed to read a sample barcode label on a sample test tube.

Purpose of Error
To ensure using sample barcode.
Detection

Method of Error
Check a result of sample number reading during the sampler analysis.
Detection

Conditions of Error
When the sample number was unreadable.
Detection

Error Message Failed to read sample number during sampler analysis(Sampler analysis).

Sampler failed to read sample number during sampler analysis.Check tube barcode label
Action Message
and press [Accept] key.

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 446020 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-102 March 2015


6.5.8.2 Failed to read rack number

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis Block (fixed)


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None

Meaning The barcode reader failed to read rack barcode label during sampler analysis.

Purpose of Error
To ensure using rack barcode.
Detection

Method of Error
Check a result of rack number reading during the sampler analysis.
Detection

Conditions of Error
When the rack number was unreadable.
Detection

Error Message Failed to read rack number

Action Message Check rack barcode label and press [Accept] key.

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 446040 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-103 March 2015


6.5.8.3 Failed to read sample number during sampler analysis.

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis Block (fixed)


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

Meaning The barcode reader failed to read a sample barcode label on a sample test tube.

Purpose of Error
To ensure using sample barcode.
Detection

Method of Error
Check a result of sample number reading during the manual analysis.
Detection

Conditions of Error
When the sample number was unreadable.
Detection

Error Message Failed to read sample number during sampler analysis.

Action Message Check tube barcode label and press [Accept] key.

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 446111 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-104 March 2015


6.5.8.4 Positive ID check error

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Hardware error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis Block (fixed)


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 2
Measurement Twin Model: Sampler Analysis Stop 2

Important Notice None

An incorrect sample test tube was taken out instead of a correct sample test tube to the
Meaning
analyzer when taking a sample test tube out to the analyzer from a rack.

Purpose of Error
To prevent from picking other sample incorrectly.
Detection

Method of Error To compare the results between read sample number at the sampler and the read sample
Detection number at the analyzer.

Conditions of Error
When the sample numbers do not match.
Detection

Error Message Positive ID check error

Sampler analysis was stopped because sample numbers read in the sampler and in the
Action Message
analyzer do not match. Check for foreign objects in analyzer and press [Accept] key.

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 446112 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-105 March 2015


6.5.8.5 Sample number not input

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware error

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis Block (fixed)


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None

Meaning A sample number is unspecified on manual analysis.

Purpose of Error
To prompt a user to input a sample number.
Detection

Method of Error
Check the sample number when checking the tube type in manual analysis.
Detection

Conditions of Error
A micro blood tube is used and the sample number is unspecified.
Detection

Error Message Sample number not input

Action Message Sample number not specified. Enter sample number in manual analysis screen.

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 511010 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-106 March 2015


6.5.9 Order Error

6.5.9.1 Analysis item not specified

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None

Meaning No analysis paramete is specified on manual analysis.

Purpose of Error
Informing a user that no analysis parameter is specified.
Detection

Method of Error Check the replied analysis parameter at analysis information inquiry with using the built-in
Detection BR on manual analysis.

Conditions of Error
When no analysis parameter is specified.
Detection

Error Message Analysis item not specified

Action Message No analysis item. Specify analysis item in manual analysis screen.

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 511012 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-107 March 2015


6.5.9.2 Invalid analysis item is specified

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None

Meaning Invalid analysis parameter is specified on manual analysis.

Purpose of Error
To prompt users that no parameter is specified for analysis.
Detection

Method of Error Check the replied analysis parameter at analysis information inquiry by using the built-in
Detection BR in manual analysis.

Conditions of Error When invalid analysis parameter is specified. (Including invalid analysis parameter which
Detection is specified by reagent related error)

Error Message Invalid analysis item is specified

Action Message Invalid analysis item is specified. Specify analysis item in manual analysis screen.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 511014 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-108 March 2015


6.5.9.3 Invalid analysis item is specified (Sampler analysis)

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Suspend

Important Notice None

Invalid analysis samples exist on the analysis line with an analyzer operating sampler
Meaning
analysis.

Purpose of Error
To notify a used that the sample cannot be analyzed.
Detection

Method of Error Check the replied analysis parameter at an analysis information inquiry in sampler
Detection analysis.

An invalid analysis parameter is specified.


Conditions of Error
Detection ( When there are other samples can be analyzed other than above conditions, these
samples will have higher priority. )

Error Message Invalid analysis item is specified (Sampler analysis)

Invalid analysis item is specified. Check sample analysis item and press [Accept] key.
Action Message
Sample will be skipped and sampler analysis will continue.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 511016 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-109 March 2015


6.5.9.4 No analyzer is ready.

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Suspend

Important Notice None

Meaning There is no analyzer which can analyze samples.

Purpose of Error
Samples cannot be analyzed by current status of the analyzer.
Detection

Method of Error
Check a replied analysis parameter at an analysis information inquiry in sampler mode.
Detection

When an analysis parameter is specified which cannot be analyzed by current status of


Conditions of Error the analyzer such as one of the analyzers is in manual mode.
Detection ( When there are other samples can be analyzed other than above conditions, these
samples will have higher priority. )

Error Message No analyzer is ready.

Sample cannot be analyzed because analyzer is not ready for sample analysis. Change
Action Message
analyzer to sampler mode or stop sampler analysis from menu.

None.
HELP [Execute]
The error clears when the analyzer which can analyze a target sample is ready to analyze
Process
in sampler mode.

Error Code 511017 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-110 March 2015


6.5.10 Data Analysis Error

6.5.10.1 Background check error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None

A measurement error of an analysis result becomes large because background value is


Meaning
high.

Purpose of Error
Check that background values are within the range at the background check.
Detection

Method of Error
Check the background values when background check is completed.
Detection

When each parameter is over the following value:


1. WBC-N 1.0 [x102/uL ]
2. WBC-D 1.0 [x102/uL]
3. WBC-P 1.0 [x102/uL]
Conditions of Error
4. RBC-I 2.0 [x104/uL]
Detection
5. HGB 0.1 [g/dL]
6. PLT-I 1.0 [x104/uL]
7. PLT-F 1.0 [x104/uL]
8. PLT-O 1.0 [x104/uL]

Error Message Background check error

Action Message Press [Execute] key. Auto rinse will start.

Execute auto rinse.


HELP [Execute]
Process The error clears when the background value becomes lower than the condition of error
detection after auto rinse.

Error Code 411010 Judgment Control Unit

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Auto Rinse

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-111 March 2015


6.5.10.2 RBC sampling error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/Suspend

Important Notice None

Meaning A variation of count value exceeds the specified range during counting.

Purpose of Error To monitor uneven concentration of samples, unexpected clogging in transducer and so
Detection on.

Method of Error Check the sum of maximum value, minimum value and sampling value after counting
Detection sampling values at a regular interval during analysis

It is an error when matching the following formula.

Maximum Value – Minimum Value – 150


Conditions of Error
Detection
The sum of Sampling Value

(The first and last sampling values are NOT included)

Error Message RBC sampling error

Action Message None

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 413020 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-112 March 2015


6.5.10.3 PLT sampling error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/Suspend

Important Notice None

Meaning A variation of count value exceeds the specified range during counting.

Purpose of Error To monitor uneven concentration of samples, unexpected clogging in transducer and so
Detection on.

Method of Error Check the sum of maximum value, minimum value and sampling value after counting
Detection sampling values at a regular interval during analysis

It is an error when matching the following formula.

Conditions of Error
Maximum Value – Minimum Value – 100
Detection
The sum of Sampling Value
(The first and last sampling values are NOT included)

Error Message PLT sampling error

Action Message None

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 413030 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-113 March 2015


6.5.10.4 WNR sampling error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/Suspend

Important Notice None

Meaning A variation of count value exceeds the specified range during counting.

Purpose of Error To monitor uneven concentration of samples, unexpected clogging in transducer and so
Detection on.

Method of Error Check the sum of maximum value, minimum value and sampling value after counting
Detection sampling values at a regular interval during analysis

It is an error when matching the following formula.

Maximum Value – Minimum Value – 250


Conditions of Error X 100 > 3.0
Detection The sum of Sampling Value

(The first and last sampling values are NOT included)

Error Message WNR sampling error

Action Message None

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 413120 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

2013125

XN-1000/2000 S/M 6-114 March 2015


6.5.10.5 WDF sampling error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/Suspend

Important Notice None

Meaning A variation of count value exceeds the specified range during counting.

Purpose of Error To monitor uneven concentration of samples, unexpected clogging in transducer and so
Detection on.

Method of Error Check the sum of maximum value, minimum value and sampling value after counting
Detection sampling values at a regular interval during analysis

It is an error when matching the following formula.


Maximum Value – Minimum Value – 200
Conditions of Error X 100 > 1.5
Detection The sum of Sampling Value

(The first and last sampling values are NOT included)

Error Message WDF sampling error

Action Message None

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 413070 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

2013125

XN-1000/2000 S/M 6-115 March 2015


6.5.10.6 WPC sampling error

Analyzer Type A
Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/Suspend

Important Notice None

Meaning A variation of count value exceeds the specified range during counting.

Purpose of Error To monitor uneven concentration of samples, unexpected clogging in transducer and so
Detection on.

Method of Error Check the sum of maximum value, minimum value and sampling value after counting
Detection sampling values at a regular interval during analysis

It is an error when matching the following formula.

Maximum Value – Minimum Value – 250


Conditions of Error X 100 > 1.5
Detection The sum of Sampling Value

(The first and last sampling values are NOT included)

Error Message WPC sampling error

Action Message None

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 413130 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

2013125

XN-1000/2000 S/M 6-116 March 2015


6.5.10.7 RET sampling error

Type A /Type B ( Occurring only with the configuration of RET


Analyzer
Applicable analysis)

Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/Suspend

Important Notice None

Meaning A variation of count value exceeds the specified range during counting.

Purpose of Error To monitor uneven concentration of samples, unexpected clogging in transducer and so
Detection on.

Method of Error Check the sum of maximum value, minimum value and sampling value after counting
Detection sampling values at a regular interval during analysis

It is an error when matching the following formula.

Maximum Value – Minimum Value – 400


Conditions of Error X 100 > 3.5
Detection The sum of Sampling Value

(The first and last sampling values are NOT included)

Error Message RET sampling error

Action Message None

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 413090 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

2013125

XN-1000/2000 S/M 6-117 March 2015


6.5.10.8 PLT-F sampling error

Type A /Type B ( Occurring only with the configuration of PLT-F


Analyzer
Applicable analysis)

Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/Suspend

Important Notice None

Meaning A variation of count value exceeds the specified range during counting.

Purpose of Error To monitor uneven concentration of samples, unexpected clogging in transducer and so
Detection on.

Method of Error Check the sum of maximum value, minimum value and sampling value after counting
Detection sampling values at a regular interval during analysis

It is an error when matching the following formula.


Maximum Value – Minimum Value – 250
Conditions of Error X 100 > 3.0
Detection The sum of Sampling Value

(The first and last sampling values are NOT included)

Error Message PLT-F sampling error

Action Message None

HELP [Execute] None


Process The error clears when [Accept] key is pressed.

Error Code 413140 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

2013125

XN-1000/2000 S/M 6-118 March 2015


6.5.10.9 RBC detector clog

Analyzer Type A /Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/Suspend

Important Notice None

Meaning The RBC detector is clogged.

Purpose of Error To ensure the sheath flow.


Detection To ensure the RBC / PLT analysis results.

Method of Error Check the clogging signal by the direct current between both electrodes of the sheath flow
Detection detector during analysis.

It becomes an error when clogging value “C” is C >105 in a timing of the timing chart.
C=3.832 x 10 x D x 3.9/256-2.110 x TTD2 x 10-2 + 1.992 x TTD – 3.660 x 1
Conditions of Error
C: degree of clogging (integer, round off the degree of clogging to integer )
Detection
D: A/D converting value of the clogging voltage
TTD: Environment Temperature (**.* ºC)

Error Message RBC detector clog

Action Message Press [Execute] key. RBC detector clog will be removed.

HELP [Execute] Execute to remove RBC detector clog.


Process The error clears when RBC detector clog completed removing properly.

Error Code 311240 Judgment Control Unit

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu RBC detector clog removing

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-119 March 2015


6.5.10.10 Bubbles in RBC detector

Analyzer Type A /Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/Suspend

Important Notice None

Meaning Air bubbles have been stuck to the RBC detector (pellet).

Purpose of Error
To ensure the RBC/PLT analysis results.
Detection

Method of Error
Check the output pulses with the air bubble monitoring circuit during the analysis.
Detection

Conditions of Error
When sum of the output pulses is over 128.
Detection

Error Message Bubbles in RBC detector

Action Message Press [Execute] key. RBC detector clog will be removed.

HELP [Execute] Execute to remove RBC detector clog.


Process The error clears when RBC detector clog completed removing properly.

Error Code 311910 Judgment Control Unit

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu RBC detector clog removing

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-120 March 2015


6.5.10.11 Low count error

Analyzer Type A /Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis Block


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None

The count is too low despite the aspiration error is not detected by blood sample
Meaning
aspiration sensor during the analysis.

Purpose of Error
To detect dilution error of the instrument
Detection

Method of Error
Check the analysis result during analysis.
Detection

When the aspiration error is not detected by the blood sample aspiration sensor, and the
analysis results are met the all following conditions.
WBC (WBC-N) < 10.0 [x10^2/µ]
Conditions of Error
Detection RBC < 30 [x 10^4/µL]
HGB < 1.0 [g/dL]
PLT < 2.0 [x10^4/µL]

Error Message Low count error

Abnormally low count because blood sample not aspirated correctly. Check sample and
Action Message
press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 416010 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-121 March 2015


6.5.10.12 HGB error

Analyzer Type A /Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

Meaning A/D converted HGB blank value or HGB sample value is out of range.

Purpose of Error
To ensure the HGB blank value and the HGB sample value.
Detection

Method of Error Check the HGB blank value and the HGB sample value during analysis with normal mode
Detection or PD mode.

It is an error when matching either the following conditions.


1. Blank convert value < 50
Conditions of Error
2. Sample convert value < 12000
Detection
3. Sample convert value – blank convert value < -50
4. Sample convert value – blank convert value > 3600

Error Message HGB error

Action Message Analysis must be repeated. Check analysis result and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 415000 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-122 March 2015


6.5.10.13 RBC channel error

Analyzer Type A /Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

Meaning An error occurs in RBC channel.

Purpose of Error
To ensure the RBC channel analysis data.
Detection

Method of Error
Check the analysis result during analyzing.
Detection

Conditions of Error It is an error when the following condition is met.


Detection RBC > 9999 [x10^4 / µL]

Error Message RBC channel error

Action Message Analysis must be repeated. Check sample mixing state and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 414020 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-123 March 2015


6.5.10.14 PLT channel error

Analyzer Type A /Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

Meaning An error occurs in PLT channel.

Purpose of Error
To ensure the PLT channel analysis data.
Detection

Method of Error
Check the analysis result during analyzing.
Detection

Conditions of Error It is an error when the following condition is met.


Detection PLT-I > 9999 [x10^3 / µL]

Error Message PLT channel error

Action Message Analysis must be repeated. Check sample mixing state and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 414030 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-124 March 2015


6.5.10.15 WNR channel error

Analyzer Type A /Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

Meaning An error occurs in WNR channel of FCM. (Lysing defective)

Purpose of Error
To ensure the WNR channel analysis data.
Detection

Method of Error
Check the analysis result during analyzing.
Detection

Conditions of Error It is an error when the following condition is met.


Detection TNC (TNC-N) > 9999.9 [x10^2 / µL]

Error Message WNR channel error

Action Message Analysis must be repeated. Check sample mixing state and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 414070 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-125 March 2015


6.5.10.16 WDF channel error

Analyzer Type A /Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

Meaning An error occurs in WDF channel.

Purpose of Error
To ensure the WDF channel analysis data.
Detection

Method of Error
Check the analysis result during analyzing.
Detection

Conditions of Error It is an error when the following condition is met.


Detection TNC-D > 9999.9 [x10^2 / µL]

Error Message WDF channel error

Action Message Analysis must be repeated. Check sample mixing state and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 414050 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-126 March 2015


6.5.10.17 WPC channel error

Analyzer Type A
Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

Meaning An error occurs in WPC channel.

Purpose of Error
To ensure the WPC channel analysis data.
Detection

Method of Error
Check the analysis result during analyzing.
Detection

Conditions of Error It is an error when the following condition is met.


Detection TNC-P > 9999.9 [x10^2 / µL]

Error Message WPC channel error

Action Message Analysis must be repeated. Check sample mixing state and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 414080 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-127 March 2015


6.5.10.18 RET channel error

Analyzer Type A (Occurring only with the configuration of RET analysis)


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

Meaning An error occurs in RET channel.

Purpose of Error
To ensure the RET channel analysis data.
Detection

Method of Error
Check the analysis result during analyzing.
Detection

It is an error when matching either of the following conditions.


1. RBC-0 > 9999 [x10^4/µL]
2. RBC-0/RBC > 1.5
Conditions of Error 3. RBC-0 / RBC < 0.5
Detection 4. RBC-X ≥ 60[ch]
5. RBC-X ≤ and RBC-0 ≥ 30 [x10^4 / µL]
However, conditions 2 and 3 will not become errors when the following condition is met.
RBC-0 < 30 [x10^4/µL] and RBC < 30 [x10^4 / µL]

Error Message RET channel error

Action Message Analysis must be repeated. Check sample mixing state and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 414040 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-128 March 2015


6.5.10.19 PLT-F channel error

Type A /Type B (Occurring only with the configuration of PLT-F


Analyzer
Applicable analysis)

Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue/ Suspend

Important Notice None

Meaning An error occurs in PLT-F channel.

Purpose of Error
To ensure the PLT-F channel analysis data.
Detection

Method of Error
Check the analysis result during analyzing.
Detection

Conditions of Error It is an error when the following condition is met.


Detection PLT-F > 99999 [x10^2 / µL]

Error Message PLT-F channel error

Action Message Analysis must be repeated. Check sample mixing state and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 414090 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-129 March 2015


6.5.10.20 < Data Errors >

Analyzer Type A /Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis (✔)

Alarm Sounds (Analysis Stop) Reanalysis Block


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None

Meaning The analysis parameters results exceed the upper/lower limit of the specified range.

Purpose of Error Notify the users the analysis parameters exceed the upper/lower limit values range and
Detection suspend the sampler analysis.

Method of Error
To compare the analysis result to the panic limit values.
Detection

Conditions of Error
When the analysis results are out of the panic limit values range.
Detection

Error Message < Data Errors >

Action Message Data error occurred. Check analysis result and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 417020 Judgment Analysis Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

A: 201218

XN-1000/2000 S/M 6-130 March 2015


6.5.10.21 Analysis result is high

Analyzer Type A /Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 2
Measurement Twin Model: Sampler Analysis Stop 2

Important Notice None

On the body fluid analysis, an analysis result might affect to the next sample of body fluid
Meaning
analysis with the carry-over because the analysis result is too high.

Purpose of Error
Prompt users to execute a background check before the next analysis.
Detection

Method of Error
Check an analysis result of the body fluid.
Detection

When the following conditions are met.


Conditions of Error
WBC-BF > 10.00 [x 10^2/uL]
Detection
RBC-BF > 100.0 [x 10^4/uL]

Error Message Analysis result is high

Because the analysis result is high, the body fluid analysis of the next sample may be
Action Message affected. Remove the sample from the sample holder and press [Execute] key. A
background check will be executed.

HELP [Execute] Execute background check.


Process The error clears when executing an analysis operation (including background check).

Error Code 417025 Judgment Analysis Unit

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-131 March 2015


6.5.11 Cover Error

6.5.11.1 Front cover open error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 2
Measurement Twin Model: Sampler Analysis Stop 2

Important Notice To stop the sequence of the analysis unit immediately.

Meaning The front cover opens during analysis.

Purpose of Error To ensure the WB sample aspiration.


Detection To avoid a risk of injury to the users.

Method of Error
Check the Front Cover Sensor during the analysis.
Detection

Conditions of Error
When detecting the open cover for 1 second continuously.
Detection

Error Message Front cover open error

Analysis cannot be continued because front cover was opened during analysis. Remove
Action Message
rack and tubes from analyzer and press [Execute] key. Analyzer initialize will be executed.

HELP [Execute] Restart the analysis unit.


Process The error clears when restarting the analysis unit.

Error Code 421580 Judgment Control Unit

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-132 March 2015


6.5.11.2 Front cover is open

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔


Alarm Sounds Analysis Stop Reanalysis NA
Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

Meaning The front cover is open.

Purpose of Error To ensure the WB sample aspiration.


Detection To avoid a risk of injury to the users.

Method of Error
Always check the Front Cover Sensor.(except during analysis)
Detection

Conditions of Error
When detecting the open cover for 1 second continuously.
Detection

Error Message Front cover is open

Action Message Front cover is open. Close cover.

HELP [Execute] None.


Process The error clears when the cover sensor detects the cover is closed.

Error Code 421581 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-133 March 2015


6.5.11.3 RBC detector cover is open.

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

Meaning The RBC detector cover is open.

Purpose of Error
To ensure the data by blocking light from the outside.
Detection

Method of Error
Check the RBC detector cover sensor during analysis and in ready status.
Detection

Conditions of Error In ready status: When detecting the open cover for 1 second continuously.
Detection During analysis: When detecting the open cover on the timing of the timing chart.

Error Message RBC detector cover is open.

Action Message RBC detector cover is open. Close cover.

HELP [Execute] None.


Process The error clears when the cover sensor detects the cover is closed.

Error Code 421020 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-134 March 2015


6.5.11.4 FCM detector cover is open.

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Repeat


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

Meaning The FCM detector cover is open.

Purpose of Error
To ensure the data with blocking light from the outside.
Detection

Method of Error
Check the FCM Detector Cover Sensor during analysis and in ready Status.
Detection

Conditions of Error In ready status: When detecting the open cover for 1 second continuously.
Detection During analysis: When detecting the open cover on the timing of the timing chart.

Error Message FCM detector cover is open.

Action Message FCM detector cover is open. Close cover.

HELP [Execute] None.


Process The error clears when the cover sensor detects the cover is closed.

Error Code 421110 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-135 March 2015


6.5.11.5 Fluorocell WNR cover is open

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

Meaning Fluorocell WNR cover is open

Purpose of Error
To ensure the data with blocking light from the outside.
Detection

Method of Error
Check always the Fluorocell WNR Cover Sensor.
Detection

Conditions of Error
When detecting the Fluorocell WNR open cover for 1 second continuously.
Detection

Error Message Fluorocell WNR cover is open

Action Message The Fluorocell WNR cover is open. Close the Fluorocell WNR cover.

HELP [Execute] None.


Process The error clears when the cover sensor detects the cover is closed.

Error Code 421583 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-136 March 2015


6.5.11.6 Fluorocell WDF cover is open

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

Meaning Fluorocell WDF cover is open

Purpose of Error
To ensure the data with blocking light from the outside.
Detection

Method of Error
Check always the Fluorocell WDF Cover Sensor.
Detection

Conditions of Error
When detecting the Fluorocell WDF open cover for 1 second continuously.
Detection

Error Message Fluorocell WDF cover is open

Action Message The Fluorocell WDF cover is open. Close the Fluorocell WDF cover.

HELP [Execute] None.


Process The error clears when the cover sensor detects the cover is closed.

Error Code 421584 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-137 March 2015


6.5.11.7 Fluorocell WPC cover is open

Analyzer Type A
Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

Meaning Fluorocell WPC cover is open

Purpose of Error
To ensure the data with blocking light from the outside.
Detection

Method of Error
Check always the Fluorocell WPC Cover Sensor.
Detection

Conditions of Error
When detecting the Fluorocell WPC open cover for 1 second continuously.
Detection

Error Message Fluorocell WPC cover is open

Action Message The Fluorocell WPC cover is open. Close the Fluorocell WPC cover.

HELP [Execute] None.


Process The error clears when the cover sensor detects the cover is closed.

Error Code 421585 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-138 March 2015


6.5.11.8 Fluorocell RET cover is open

Analyzer Type A / Type B (Occurring by the configurations of only RET)


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

Meaning Fluorocell RET cover is open

Purpose of Error
To ensure the data with blocking light from the outside.
Detection

Method of Error
Check always the Fluorocell RET Cover Sensor.
Detection

Conditions of Error
When detecting the Fluorocell RET open cover for 1 second continuously.
Detection

Error Message Fluorocell RET cover is open

Action Message The Fluorocell RET cover is open. Close the Fluorocell RET cover.

HELP [Execute] None.


Process The error clears when the cover sensor detects the cover is closed.

Error Code 421586 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-139 March 2015


6.5.11.9 Fluorocell PLT cover is open

Analyzer Type A / Type B (Occurring by the configurations of only PLT-F)


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None

Meaning Fluorocell PLT cover is open

Purpose of Error
To ensure the data with blocking light from the outside.
Detection

Method of Error
Check always the Fluorocell PLT cover sensor.
Detection

Conditions of Error
When detecting the Fluorocell PLT open cover for 1 second continuously.
Detection

Error Message Fluorocell PLT cover is open

Action Message The Fluorocell PLT cover is open. Close the Fluorocell PLT cover.

HELP [Execute] None.


Process The error clears when the cover sensor detects the cover is closed.

Error Code 421587 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-140 March 2015


6.5.12 Laser Error

6.5.12.1 Laser life

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None.

Meaning Laser drive current value is increasing and expiration of laser life time is approaching.

Purpose of Error To predict the laser diode life time and let the user to contact Sysmex Service
Detection Representative to replace the laser diode.

Compare the initial laser drive current value which is stored the analyzer to the current
Method of Error
laser drive current value at the laser drive current of the each channel during analyzing
Detection
(timing on the timing chart).

Conditions of Error The ratio of the initial laser drive current value to the current laser drive current value is
Detection more than 1.2 times and less than 2.0 times.

Error Message Laser life

Action Message Laser will soon require replacement. Contact a service technician.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 321050 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-141 March 2015


6.5.12.2 Laser output error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask ✔ Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis Block (fixed)


Process When Error
Occurred Sampler Single Model: Sampler Analysis Stop 1
Measurement Twin Model: Sampler Analysis Stop 1

Important Notice None.

Meaning The laser power exceeds the controlling range.

Purpose of Error
To ensure the laser power stability.
Detection

Compare the initial laser drive current value which is stored the analyzer to the current
Method of Error
laser drive current value in the laser drive current of each channel during analysis (timing
Detection
on the timing chart).

When the ratio of the initial laser current value and the current laser drive current value is
Conditions of Error 2.0 times or more.
Detection
When the current laser drive current value is over 150mA.

Error Message Laser output error

Analysis cannot be continued because laser replacement is required. Remove the rack
Action Message
and tubes from the analyzer and turn the main power of the analyzer off.

HELP [Execute] None.


Process The error clears when turned the main power of the analyzer off.

Error Code 321060 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-142 March 2015


6.5.13 System Error

6.5.13.1 Analyzer barcode reader communication error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler Single Model: N/A
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None.

Meaning There is a communication error between the analyzer barcode reader and PCB No.30001.

Purpose of Error
To ensure that the barcode reader reads barcode in the analyzer.
Detection

Method of Error Check the connection between the analyzer barcode reader and PCB No.30001 when
Detection starting the analyzer.

Conditions of Error When a connection between the analyzer barcode reader and the PCB No.30001 cannot
Detection be established.

Error Message Analyzer barcode reader communication error

Turn the main power of the analyzer off and on. If error recurs after restarting analyzer,
Action Message
contact a service technician.

HELP [Execute] None.


Process The error clears when turned the main power of the analyzer off.

Error Code 446110 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-143 March 2015


6.5.13.2 Sampler barcode reader communication error

Analyzer NA
Applicable
Sampler Single Model / Twin Model

Error Category Hardware Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None.

Meaning There is a communication error between the sampler barcode reader and PCB No.30001.

Purpose of Error
To ensure that the barcode reader reads barcode in the sampler.
Detection

Method of Error Check the connection between the sampler barcode reader and PCB No.30001 when
Detection starting the sampler.

Conditions of Error When a connection between the analyzer barcode reader and the PCB No.30001 cannot
Detection be established

Error Message Sampler barcode reader communication error

Turn the main power of the sampler off and on. If error recurs after restarting analyzer,
Action Message
contact a service technician.

HELP [Execute] None.


Process The error clears when turned the main power of the sampler off.

Error Code 446010 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-144 March 2015


6.5.13.3 Conveyor communication error

Analyzer NA
Applicable
Sampler Single Model

Error Category Hardware Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler
Single Model : Sampler Analysis Suspend
Measurement

Important Notice None.

Meaning The communication with the conveyor unit is disconnected.

Purpose of Error
To ensure the communication with the conveyor unit.
Detection

Method of Error Check the communication connection with the conveyor unit when sending a message to
Detection the conveyor unit.

Conditions of Error
When a communication message cannot be sent to the conveyor unit.
Detection

Error Message Conveyor communication error

Action Message Communicate with conveyor.

HELP [Execute] None.


Process The error clears when the connection with the conveyor unit is ensured.

Error Code 449510 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-145 March 2015


6.5.13.4 Communication error during sampler analysis.

Analyzer NA
Applicable
Sampler Single Model/Twin Model

Error Category Hardware Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model : Sampler Analysis Stop 2
Measurement Twin Model: Sampler Analysis Stop 2

Important Notice None.

Communication between the analyzer and the sampler is disconnected during sampler
Meaning
analysis.

Purpose of Error To ensure the communication between the analyzer and the sampler during sampler
Detection analysis.

Method of Error
Check the analyzer communication status during the sampler analysis.
Detection

Conditions of Error
When the sampler cannot communicate with the analyzer during sampler analysis.
Detection

Error Message Communication error during sampler analysis.

Sampler analysis was stopped because communication with the analyzer was interrupted
Action Message during sampler analysis. Remove the rack and sample tube from the analyzer and
analysis line and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 449015 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-146 March 2015


6.5.13.5 RFID communication error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model : NA
Measurement Twin Model: Sampler Analysis Continue / Suspend

Important Notice None.

Meaning There is a communication error between RFID reader and RFID writer in the analyzer.

Purpose of Error
To ensure the RFID reading in the analyzer.
Detection

Check the response from the RFID reader / writer with the following timing.
1. Starting an analyzer
Method of Error
2. While covers of each dye solution are opened
Detection
3. When the covers of each dye solution are closed.
4. When each dye solution is aspirated.

Conditions of Error
When there is no reply from the RFID reader / writer.
Detection

Error Message RFID communication error

Turn the main power of the analyzer off and on. If the error occurs again, contact your
Action Message
Sysmex service representative for assistance.

HELP [Execute] None.


Process The error clears when turning off the main power of the analyzer.

Error Code 449025 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-147 March 2015


6.5.13.6 Instrument communication error

Analyzer Type A / Type B


Applicable
Sampler Single Model / Twin Model

Error Category Hardware Error

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model : Sampler Analysis Stop 3
Measurement Twin Model: Sampler Analysis Stop 3

Important Notice None.

Meaning There is a communication error between the analyzer and the IPU.

Purpose of Error
To ensure the communication between the analyzer and the IPU.
Detection

Method of Error
Check the communication content when the analyzer communicates with the IPU.
Detection

Conditions of Error
When the communication contents is improperly.
Detection

Error Message Instrument communication error

Analysis cannot be continued because communication with the analyzer was error.
Action Message Remove the rack and tubes from the analyzer and turn the main power of the analyzer off
and on.

HELP [Execute] None.


Process The error clears when turning off the main power of the analyzer.

Error Code 449035 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-148 March 2015


6.5.14 QC Error

6.5.14.1 < X-barM control error >

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning An abnormality is detected in X-barM limit.

Purpose of Error
To detect the analysis error by checking time-to-time change.
Detection

Method of Error Executing a quality control by a statistical method weighted average data of the normal
Detection samples as a QC data during plotting to the QC chart.

Conditions of Error
When the data is out of control limit.
Detection

Error Message < X-barM control error >

Action Message Check QC chart and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 461110 Judgment IPU

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key Showing the QC procedure

XN-1000/2000 S/M 6-149 March 2015


6.5.14.2 < L-J Control Error >

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning L-J limit error occurs.

Purpose of Error
To detect the analysis error by checking day-to-day or time-to-time change.
Detection

Method of Error Executing a quality control by a statistical method analysis result of the control blood
Detection sample as a control data during plotting to the QC chart.

Conditions of Error
When the data is out of control limit.
Detection

Error Message < L-J Control Error >

Action Message Check QC chart and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 461160 Judgment IPU

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key Showing the QC procedure

XN-1000/2000 S/M 6-150 March 2015


6.5.14.3 < X-bar control error >

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning X-bar limit error occurs.

Purpose of Error
To detect the analysis error by checking day-to-day change.
Detection

Method of Error Executing a quality control by a statistical method analysis result of the control blood
Detection sample as a control data during plotting to the QC chart.

Conditions of Error
When the data is out of control limit.
Detection

Error Message < X-bar control error >

Action Message Check QC chart and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 461150 Judgment IPU

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key Showing the QC procedure

XN-1000/2000 S/M 6-151 March 2015


6.5.14.4 < Control has expired. >

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning Control blood expiration date is expired.

Purpose of Error To ensure the precision of QC, by notifying the users that the control blood expiration data
Detection is expired.

Method of Error
Check the control blood lot information (expiration date) when operating QC.
Detection

Conditions of Error
When the control blood expiration date is expired.
Detection

Error Message < Control has expired. >

Action Message Enter QC material of new lot and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 464010 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key Showing the QC procedure

XN-1000/2000 S/M 6-152 March 2015


6.5.14.5 < Control is not entered. >

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning New lot information of the control blood is not entered.

Purpose of Error
To notify the control blood lot registration status.
Detection

Method of Error
Check the control blood lot information when operating QC.
Detection

Conditions of Error
When there is no lot information.
Detection

Error Message < Control is not entered. >

Action Message Enter QC material in IPU and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 464110 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key Showing the QC procedure

XN-1000/2000 S/M 6-153 March 2015


6.5.14.6 < QC not executed. >

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning It is a timing of executing the QC

Purpose of Error
To prompt the users to execute the QC.
Detection

Method of Error
Check that the QC analysis executed at the time of QC execution time set by users.
Detection

Conditions of Error
When QC is not executed within 1 hour.
Detection

Error Message < QC not executed. >

Action Message Execute QC and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 461170 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key Showing the QC procedure

XN-1000/2000 S/M 6-154 March 2015


6.5.14.7 <WNR Scattergram sensitivity error>

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning CQ error in the sensitivity parameter occurs.

Purpose of Error
To detect an instrument error by chcking time-to-time change.
Detection

Method of Error Executing a quality control of sensitivity parameter in the control blood sample analysis
Detection result as a control data by a statistical method during plotting to the QC chart.

Conditions of Error
When the data is out of control limit.
Detection

Error Message <WNR Scattergram sensitivity error>

Action Message Check the scattergram and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 417601 Judgment IPU

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-155 March 2015


6.5.14.8 <WDF Scattergram sensitivity error>

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning CQ error in the sensitivity parameter occurs.

Purpose of Error
To detect an instrument error by chcking time-to-time change.
Detection

Method of Error Executing a quality control of sensitivity parameter in the control blood sample analysis
Detection result as a control data by a statistical method during plotting to the QC chart.

Conditions of Error
When the data is out of control limit.
Detection

Error Message <WDF scattergram sensitivity error>

Action Message Check the scattergram and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 417602 Judgment IPU

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-156 March 2015


6.5.14.9 <WPC Scattergram sensitivity error>

Analyzer Type A
Applicable
Sampler NA

Error Category Analysis Error

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning CQ error in the sensitivity parameter occurs.

Purpose of Error
To detect an instrument error by chcking time-to-time change.
Detection

Method of Error Executing a quality control of sensitivity parameter in the control blood sample analysis
Detection result as a control data by a statistical method during plotting to the QC chart.

Conditions of Error
When the data is out of control limit.
Detection

Error Message <WPC Scattergram sensitivity error>

Action Message Check the scattergram and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 417603 Judgment IPU

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-157 March 2015


6.5.14.10 <RET Scattergram sensitivity error>

Analyzer Type A / Type B ( Occurring by the configurations of only RET)


Applicable
Sampler NA

Error Category Analysis Error

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning CQ error in the sensitivity parameter occurs.

Purpose of Error
To detect an instrument error by chcking time-to-time change.
Detection

Method of Error Executing a quality control of sensitivity parameter in the control blood sample analysis
Detection result as a control data by a statistical method during plotting to the QC chart.

Conditions of Error
When the data is out of control limit.
Detection

Error Message <RET Scattergram sensitivity error>

Action Message Check the scattergram and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 417604 Judgment IPU

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-158 March 2015


6.5.14.11 <PLT-F Scattergram sensitivity error>

Analyzer Type A / Type B ( Occurring by the configurations of only PLT-F)


Applicable
Sampler NA

Error Category Analysis Error

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning CQ error in the sensitivity parameter occurs.

Purpose of Error
To detect an instrument error by chcking time-to-time change.
Detection

Method of Error Executing a quality control of sensitivity parameter in the control blood sample analysis
Detection result as a control data by a statistical method during plotting to the QC chart.

Conditions of Error
When the data is out of control limit.
Detection

Error Message <PLT-F Scattergram sensitivity error>

Action Message Check the scattergram and press [Accept] key.

HELP [Execute] None.


Process The error clears when [Accept] key is pressed.

Error Code 417605 Judgment IPU

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-159 March 2015


6.5.14.12 <Check Analysis Mode>

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Analysis Error

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning A control blood set on the current analysis mode is incorrect.

Purpose of Error
To prompt the users to set a correct control blood.
Detection

Method of Error Check the control blood type when the control blood sample numbers were read by a
Detection barcode reader.

Conditions of Error
When the control blood type is different with the current analysis mode.
Detection

Error Message <Check Analysis Mode>

Check the Analysis Mode/control blood type and press [Accept] key. Change Analysis
Action Message
Mode and repeat analysis.

None.
HELP [Execute]
Process The error clears when [Accept] is pressed. Skip a subject sample analysis during sampler
analysis.

Error Code 419060 Judgment Control Unit

HELP Screen Key [Accept]


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-160 March 2015


6.5.15 User Maintenance / Warning

6.5.15.1 Piercer replacement is required.

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None.

Meaning It's time to replace the piercer.

Purpose of Error
To prompt the users to replace piercer.
Detection

Method of Error Check the cycle count of the piercer operation before start up the analyzer and during
Detection analysis.

Conditions of Error
Numbers of piercer operation cycle count> 120000 times
Detection

Error Message Piercer replacement is required.

Turn off analyzer power and replace piercer. After replacing piercer, reset piercer count in
Action Message
counter screen of maintenance menu.

HELP [Execute] None.


Process The error clears when numbers of the piercer operation are reset.

Error Code 471610 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key Showing the piercer replacement procedure

XN-1000/2000 S/M 6-161 March 2015


6.5.15.2 Cleaning is required.

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Continue
Measurement Twin Model: Sampler Analysis Continue

Important Notice None.

Meaning It is a time to execute cleaning.

Purpose of Error
Prompt the users for cleaning.
Detection

Method of Error Check the cycle count of analysis after the previous cleaning performance during
Detection aspirating sample.

Conditions of Error
Numbers of cycle count after previous cleaning cycle count > 1000 times
Detection

Error Message Cleaning is required.

Action Message Regular cleaning is necessary for proper use of the analyzer.

HELP [Execute] Cleaning screen appears


Process The error clears when cleaning completed properly.

Error Code 471140 Judgment Control Unit

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Cleaning

Show Detailed Procedure Key Showing the cleaning procedure

XN-1000/2000 S/M 6-162 March 2015


6.5.15.3 Cleaning is required (Warning)

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Continue / Susupend

Important Notice None.

Meaning It is a time to execute cleaning.

Purpose of Error
Prompt the users to cleaning.
Detection

Method of Error
Check the cycle count of analysis after the previous cleaning during aspirating sample.
Detection

Conditions of Error
Numbers of cycle count after previous cleaning > 1000 times
Detection

Error Message Cleaning is required (warning)

Regular cleaning is necessary for proper use of the analyzer. Press [Execute] key. The
Action Message
cleaning screen will appear.

HELP [Execute] Cleaning screen appears


Process The error clears when cleaning completed properly.

Error Code 471150 Judgment Control Unit

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Cleaning

Show Detailed Procedure Key Showing the cleaning procedure

XN-1000/2000 S/M 6-163 March 2015


6.5.15.4 CELLPACK DCL has expired.

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: Sampler Analysis Suspend
Measurement Twin Model: Sampler Analysis Suspend

Important Notice None.

Meaning The reagent has expired.

Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection

Compare the current date with the expiration date set in reagent replacement before the
following timing:
Method of Error 1. before background check
Detection
2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis

Conditions of Error
When reagent has expired.
Detection

Error Message CELLPACK DCL has expired.

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] Reagent replacement screen appears


Process The error clears when replacing reagent completed properly.

Error Code 472020 Judgment Control Unit

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key Reagent replacement procedure

XN-1000/2000 S/M 6-164 March 2015


6.5.15.5 SULFOLYSER has expired

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning The reagent has expired.

Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection

Compare the current date with the expiration date set in reagent replacement before the
following timing:
Method of Error 1. before background check
Detection
2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis

Conditions of Error
When reagent has expired.
Detection

Error Message SULFOLYSER has expired

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] Reagent replacement screen appears


Process The error clears when replacing reagent completed properly.

Error Code 472022 Judgment Control Unit

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key Reagent replacement procedure

XN-1000/2000 S/M 6-165 March 2015


6.5.15.6 Lysercell WNR has expired

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning The reagent has expired.

Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection

Compare the current date with the expiration date set in reagent replacement before the
following timing:
Method of Error 1. before background check
Detection
2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis

Conditions of Error
When reagent has expired.
Detection

Error Message Lysercell WNR has expired

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] Reagent replacement screen appears


Process The error clears when replacing reagent completed properly.

Error Code 472040 Judgment Control Unit

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key Reagent replacement procedure

XN-1000/2000 S/M 6-166 March 2015


6.5.15.7 Fluorocell WNR has expired

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning The reagent has expired.

Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection

Compare the current date with the expiration date set in reagent replacement before the
following timing:
1. before background check
Method of Error
Detection 2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
4. When Fluorocell WDR is set.

Conditions of Error
When reagent has expired
Detection

Error Message Fluorocell WNR has expired.

Action Message Open the Fluorocell WNR cover and install new Fluorocell WNR.

HELP [Execute] None.


Process The error clears when installing new Fluorocell WNR to the dye holder properly.

Error Code 472041 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key Reagent replacement procedure

XN-1000/2000 S/M 6-167 March 2015


6.5.15.8 Lysercell WDF has expired

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning The reagent has expired.

Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection

Compare the current date with the expiration date set in reagent replacement before the
following timing:
Method of Error 1. before background check
Detection
2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis

Conditions of Error
When reagent has expired.
Detection

Error Message Lysercell WDF has expired

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] Reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 472024 Judgment Control Unit

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key Reagent replacement procedure

XN-1000/2000 S/M 6-168 March 2015


6.5.15.9 Fluorocell WDF has expired

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning The reagent has expired.

Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection

Compare the current date with the expiration date set in reagent replacement before the
following timing:
1. before background check
Method of Error
Detection 2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
4. when FRUOROCELL WDF is set

Conditions of Error
When reagent has expired.
Detection

Error Message Fluorocell WDF has expired

Action Message Open the Fluorocell WDF cover and install new Fluorocell WDF.

HELP [Execute] None


Process The error clears when within expiration date Fluorocell WDF set to the dye holder.

Error Code 472025 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key Reagent replacement procedure

XN-1000/2000 S/M 6-169 March 2015


6.5.15.10 Lysercell WPC has expired

Analyzer Type A
Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning The reagent has expired.

Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection

Compare the current date with the expiration date set in reagent replacement before the
following timing:
Method of Error 1. before background check
Detection
2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis

Conditions of Error
When reagent has expired.
Detection

Error Message Lysercell WPC has expired

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] Reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 472042 Judgment Control Unit

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key Reagent replacement procedure

XN-1000/2000 S/M 6-170 March 2015


6.5.15.11 Fluorocell WPC has expired

Analyzer Type A
Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning The reagent has expired.

Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection

Compare the current date with the expiration date set in reagent replacement before the
following timing:
1. before background check
Method of Error
Detection 2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
4. when Fluorocell WPC is set

Conditions of Error
When reagent has expired.
Detection

Error Message Fluorocell WPC has expired

Action Message Open the Fluorocell WPC cover and install new Fluorocell WPC.

HELP [Execute] None


Process The error clears when within expiration date Fluorocell WDF set to the dye holder.

Error Code 472043 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key Reagent replacement procedure

XN-1000/2000 S/M 6-171 March 2015


6.5.15.12 CELLPACK DFL has expired

Analyzer Type A
Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning The reagent has expired.

Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection

Compare the current date with the expiration date set in reagent replacement before the
following timing:
Method of Error 1. before background check
Detection
2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis

Conditions of Error
When reagent has expired.
Detection

Error Message CELLPACK DFL has expired.

Action Message Press [Execute] key. The reagent replacement screen will appear.

HELP [Execute] Reagent replacement screen appears.


Process The error clears when reagent is replaced properly.

Error Code 472044 Judgment Control Unit

HELP Screen Key [Execute]


Detailed Information
Maintenance Menu Reagent replacement

Show Detailed Procedure Key Reagent replacement procedure

XN-1000/2000 S/M 6-172 March 2015


6.5.15.13 Fluorocell RET has expired

Type A / Type B (Occurring by the configurations of only RET


Analyzer
Applicable analysis)

Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning The reagent has expired.

Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection

Compare the current date with the expiration date set in reagent replacement before the
following timing:
1. before background check
Method of Error
Detection 2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
4. when Fluorocell RET is set

Conditions of Error
When reagent has expired.
Detection

Error Message Fluorocell RET has expired

Action Message Open the Fluorocell RET cover and install new Fluorocell RET.

HELP [Execute] None


Process The error clears when within expiration date Fluorocell RET set to the dye holder.

Error Code 472045 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key Reagent replacement procedure

XN-1000/2000 S/M 6-173 March 2015


6.5.15.14 Fluorocell PLT has expired

Type A / Type B (Occurring by the configurations of only PLT-F


Analyzer
Applicable analysis)

Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds (Analysis Stop) Reanalysis NA


Process When Error
Occurred Sampler Single Model: (Sampler Analysis Suspend)
Measurement Twin Model: (Sampler Analysis Suspend)

Important Notice None.

Meaning The reagent has expired.

Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection

Compare the current date with the expiration date set in reagent replacement before the
following timing:
1. before background check
Method of Error
Detection 2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
4. when Fluorocell PLT is set

Conditions of Error
When reagent has expired.
Detection

Error Message Fluorocell PLT has expired

Action Message Open the Fluorocell PLT cover and install new Fluorocell PLT.

HELP [Execute] None


Process The error clears when within expiration date Fluorocell RET set to the dye holder.

Error Code 472046 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key Reagent replacement procedure

XN-1000/2000 S/M 6-174 March 2015


6.5.15.15 Press Start SW

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Maintenance

Data Mask NA Invalid Analysis NA

Alarm Sounds NA Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None.

Meaning Analyzer is in sleep-mode status.

Purpose of Error
To prompt resume the analyzer to ready status before an analysis.
Detection

Method of Error Check the elapsed time after the analyzer is in the manual ready status or the sampler
Detection ready status.

Conditions of Error
5 hours elapsed.
Detection

Error Message Press Start SW

Action Message None

HELP [Execute] None


Process The error clears when the rinse operation for the analyzer completed properly.

Error Code 479200 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-175 March 2015


6.5.16 Dye Holder Error

6.5.16.1 Wrong reagent installed in Fluorocell WNR holder

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None.

Meaning An incoreect dye reagent is installed in the dye holder.

Purpose of Error To prevent installing an incorrect dye, or notify that the incorrect dye is installed for
Detection preventing analyzing with the incorrect reagent.

Reading the cartridge RFID tag placed in the Fluorocell WNR holder in the following
timing, and then checking the read reagent code.
Method of Error 1. When starting the analyzer.
Detection
2. When dye cartridge is installed in the Fluorocell WNR holder.
3. When closing the Fluorocell WNR cover.

Conditions of Error When reading reagent code of the RFID tag is different from the reagent code which
Detection should be placed in the relevant dye holder.

Error Message Wrong reagent installed in Fluorocell WNR holder

Action Message Install Fluorocell WNR in the Fluorocell WNR holder.

HELP [Execute] None.


Process The error clears when Fluorocell WNR is installed in the Fluorocell WNR holder.

Error Code 135201 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-176 March 2015


6.5.16.2 Wrong reagent installed in Fluorocell WDF holder

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None.

Meaning An incoreect dye reagent is installed in the dye holder.

Purpose of Error To prevent installing an incorrect dye, or notify that the incorrect dye is installed for
Detection preventing analyzing with the incorrect reagent.

Reading the cartridge RFID tag placed in the Fluorocell WDF holder in the following
timing, and then checking the read reagent code.
Method of Error 1. When starting the analyzer.
Detection
2. When dye cartridge is installed in the Fluorocell WDF holder.
3. When closing the Fluorocell WDF cover.

Conditions of Error When reading reagent code of the RFID tag is different from the reagent code which
Detection should be placed in the relevant dye holder.

Error Message Wrong reagent installed in Fluorocell WDF holder

Action Message Install Fluorocell WDF in the Fluorocell WDF holder.

HELP [Execute] None.


Process The error clears when Fluorocell WDF is installed in the Fluorocell WDF holder.

Error Code 133151 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-177 March 2015


6.5.16.3 Wrong reagent installed in Fluorocell WPC holder

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None.

Meaning An incoreect dye reagent is installed in the dye holder.

Purpose of Error To prevent installing an incorrect dye, or notify that the incorrect dye is installed for
Detection preventing analyzing with the incorrect reagent.

Reading the cartridge RFID tag placed in the Fluorocell WPC holder in the following
timing, and then checking the read reagent code.
Method of Error 1. When starting the analyzer.
Detection
2. When dye cartridge is installed in the Fluorocell WPC holder.
3. When closing the Fluorocell WPC cover.

Conditions of Error When reading reagent code of the RFID tag is different from the reagent code which
Detection should be placed in the relevant dye holder.

Error Message Wrong reagent installed in Fluorocell WPC holder

Action Message Install Fluorocell WPC in the Fluorocell WPC holder.

HELP [Execute] None.


Process The error clears when Fluorocell WPC is installed in the Fluorocell WPC holder.

Error Code 135231 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-178 March 2015


6.5.16.4 Wrong reagent installed in Fluorocell RET holder

Type A / Type B (Occurring only with the configuration of RET


Analyzer
Applicable analysis)

Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None.

Meaning An incoreect dye reagent is installed in the dye holder.

Purpose of Error To prevent installing an incorrect dye, or notify that the incorrect dye is installed for
Detection preventing analyzing with the incorrect reagent.

Reading the cartridge RFID tag placed in the Fluorocell RET holder in the following timing,
and then checking the read reagent code.
Method of Error 1. When starting the analyzer.
Detection
2. When dye cartridge is installed in the Fluorocell RET holder.
3. When closing the Fluorocell RET cover.

Conditions of Error When reading reagent code of the RFID tag is different from the reagent code which
Detection should be placed in the relevant dye holder.

Error Message Wrong reagent installed in Fluorocell RET holder

Action Message Install Fluorocell RET in the Fluorocell RET holder.

HELP [Execute] None.


Process The error clears when Fluorocell RET is installed in the Fluorocell RET holder.

Error Code 135211 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-179 March 2015


6.5.16.5 Wrong reagent installed in Fluorocell PLT holder

Type A / Type B (Occurring only with the configuration of PLT-F


Analyzer
Applicable analysis)

Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None.

Meaning An incoreect dye reagent is installed in the dye holder.

Purpose of Error To prevent installing an incorrect dye, or notify that the incorrect dye is installed for
Detection preventing analyzing with the incorrect reagent.

Reading the cartridge RFID tag placed in the Fluorocell PLT holder in the following timing,
and then checking the read reagent code.
Method of Error 1. When starting the analyzer.
Detection
2. When dye cartridge is installed in the Fluorocell PLT holder.
3. When closing the Fluorocell PLT cover.

Conditions of Error When reading reagent code of the RFID tag is different from the reagent code which
Detection should be placed in the relevant dye holder.

Error Message Wrong reagent installed in Fluorocell PLT holder

Action Message Install Fluorocell PLT in the Fluorocell PLT holder.

HELP [Execute] None.


Process The error clears when Fluorocell PLT is installed in the Fluorocell PLT holder.

Error Code 135221 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-180 March 2015


6.5.16.6 Fluorocell WNR is not installed

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None.

Meaning The dye is not installed in the dye holder.

Purpose of Error
Ensure the necessary volume of reagent for analysis.
Detection

Reading the cartridge RFID tag placed in the Fluorocell WNR holder in the following
timing, then checking a response from the RFID tag.
Method of Error
Detection 1. When starting the analyzer
2. When closing the Fluorocell WNR cover.

Conditions of Error
No response from RFID tag
Detection

Error Message Fluorocell WNR is not installed

Action Message Open the Fluorocell WNR cover and install Fluorocell WNR.

HELP [Execute] None.


Process The error clears when a dye with RFID is installed in the Fluorocell WNR holder.

Error Code 135202 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-181 March 2015


6.5.16.7 Fluorocell WDF is not installed

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None.

Meaning The dye is not installed in the dye holder.

Purpose of Error
Ensure the necessary volume of reagent for analysis.
Detection

Reading the cartridge RFID tag placed in the Fluorocell WDF holder in the following
timing, then checking a response from the RFID tag.
Method of Error
Detection 1. When starting the analyzer
2. When closing the Fluorocell WDF cover.

Conditions of Error
No response from RFID tag
Detection

Error Message Fluorocell WDF is not installed

Action Message Open the Fluorocell WDF cover and install Fluorocell WDF.

HELP [Execute] None.


Process The error clears when a dye with RFID is installed in the Fluorocell WDF holder.

Error Code 133152 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-182 March 2015


6.5.16.8 Fluorocell WPC is not installed

Analyzer Type A
Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None.

Meaning The dye is not installed in the dye holder.

Purpose of Error
Ensure the necessary volume of reagent for analysis.
Detection

Reading the cartridge RFID tag placed in the Fluorocell WPC holder in the following
timing, then checking a response from the RFID tag.
Method of Error
Detection 1. When starting the analyzer
2. When closing the Fluorocell WPC cover.

Conditions of Error
No response from RFID tag
Detection

Error Message Fluorocell WPC is not installed

Action Message Open the Fluorocell WPC cover and install Fluorocell WPC.

HELP [Execute] None.


Process The error clears when a dye with RFID is installed in the Fluorocell WPC holder.

Error Code 135232 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-183 March 2015


6.5.16.9 Fluorocell RET is not installed

Type A / Type B ( Occurring by the configurations of only RET


Analyzer
Applicable analysis)

Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None.

Meaning The dye is not installed in the dye holder.

Purpose of Error
Ensure the necessary volume of reagent for analysis.
Detection

Reading the cartridge RFID tag placed in the Fluorocell RET holder in the following timing,
then checking a response from the RFID tag.
Method of Error
Detection 1. When starting the analyzer
2. When closing the Fluorocell RET cover.

Conditions of Error
No response from RFID tag
Detection

Error Message Fluorocell RET is not installed

Action Message Open the Fluorocell RET cover and install Fluorocell RET.

HELP [Execute] None.


Process The error clears when a dye with RFID is installed in the Fluorocell RET holder.

Error Code 135212 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-184 March 2015


6.5.16.10 Fluorocell PLT is not installed

Type A / Type B ( Occurring by the configurations of only PLT-F


Analyzer
Applicable analysis)

Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None.

Meaning The dye is not installed in the dye holder.

Purpose of Error
Ensure the necessary volume of reagent for analysis.
Detection

Reading the cartridge RFID tag placed in the Fluorocell PLT holder in the following timing,
then checking a response from the RFID tag.
Method of Error
Detection 1. When starting the analyzer
2. When closing the Fluorocell PLT cover.

Conditions of Error
No response from RFID tag
Detection

Error Message Fluorocell PLT is not installed

Action Message Open the Fluorocell PLT cover and install Fluorocell PLT.

HELP [Execute] None.


Process The error clears when a dye with RFID is installed in the Fluorocell PLT holder.

Error Code 135222 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-185 March 2015


6.5.16.11 Fluorocell WNR has already been used

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None.

Meaning The dye cartridge has already been spent.

Purpose of Error Preventing deteriorating a dye from refilling dye into dye cartridge and preventing
Detection analyzing with deteriorated dye.

Reading the RFID tag of a cartridge installed in the Fluorocell WNR holder in the following
timing, and comparing the usage status stored in the RFID tag with number of remaining
Method of Error tests.
Detection
1. When starting the analyzer.
2. When closing the FRUOLOCELL WNR cover.

It will become an error when the following condition formulas are met:
CQC+{(IQC/10) x ROB} > IQC
Conditions of Error
1. The number of remaining tests stored in RFID : CQC (Current Quantity Count)
Detection
2. The number of the total tests stored in RFID: IQC (Initial Quantity Control)
3. The number of the Read Only Blocks in use status check block: ROB

Error Message Fluorocell WNR has already been used

Action Message Install new Fluorocell WNR in the Fluorocell WNR holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell WNR with the correct number of use status check-
blocks is installed in the Fluorocell WNR holder and closing the dye cover.

Error Code 135203 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-186 March 2015


6.5.16.12 Fluorocell WDF has already been used

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error Sampler
Occurred NA
Measurement

Analyzable when only CBC, RET, and PLT-F are specified as


Important Notice
discrete and also with manual mode.

Meaning The dye cartridge has already been spent.

Purpose of Error Preventing deteriorating a dye from refilling dye into dye cartridge and preventing
Detection analyzing with deteriorated dye.

Reading the RFID tag of a cartridge installed in the Fluorocell WNR holder in the following
timing, and comparing the usage status stored in the RFID tag with number of remaining
Method of Error tests.
Detection
1. When starting the analyzer.
2. When closing the FRUOLOCELL WDF cover.

It will become an error when the following condition formulas are met:
CQC+{(IQC/10) x ROB}>IQC
Conditions of Error
1. The number of remaining tests stored in RFID : CQC (Current Quantity Count)
Detection
2. The number of the total tests stored in RFID: IQC (Initial Quantity Control)
3. The number of the Read Only Blocks in use status check block: ROB

Error Message Fluorocell WDF has already been used

Discrete except DIFF can be analyzed in manual mode. Install new Fluorocell WDF in the
Action Message
Fluorocell WDF holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell WDF with the correct number of use status check-
blocks is installed in the Fluorocell WDF holder and closing the dye cover.

Error Code 133153 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-187 March 2015


6.5.16.13 Fluorocell WPC has already been used

Analyzer Type A
Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error Sampler
Occurred NA
Measurement

Analyzable when only CBC, DIFF, RET, and PLT-F are specified
Important Notice
as discrete and also with manual mode.

Meaning The dye cartridge has already been spent.

Purpose of Error Preventing deteriorating a dye from refilling dye into dye cartridge and preventing
Detection analyzing with deteriorated dye.

Reading the RFID tag of a cartridge installed in the Fluorocell WPC holder in the following
timing, and comparing the usage status stored in the RFID tag with number of remaining
Method of Error tests.
Detection
1. When starting the analyzer.
2. When closing the FRUOLOCELL WPC cover.

It will become an error when the following condition formulas are met:
CQC+{(IQC/10) x ROB}>IQC
Conditions of Error
1. The number of remaining tests stored in RFID : CQC (Current Quantity Count)
Detection
2. The number of the total tests stored in RFID: IQC (Initial Quantity Control)
3. The number of the Read Only Blocks in use status check block: ROB

Error Message Fluorocell WPC has already been used

Discrete except WPC can be analyzed in manual mode. Install new Fluorocell WPC in the
Action Message
Fluorocell WPC holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell WPC with the correct number of use status check-
blocks is installed in the Fluorocell WPC holder and closing the dye cover.

Error Code 135233 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-188 March 2015


6.5.16.14 Fluorocell RET has already been used

Type A / Type B (Occurring by only the configuration of RET


Analyzer
Applicable analysis)

Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error Sampler
Occurred NA
Measurement

Analyzable when only CBC, DIFF, PLT-F, and WPC are specified
Important Notice
as discrete and also with manual mode.

Meaning The dye cartridge has already been spent.

Purpose of Error Preventing deteriorating a dye from refilling dye into dye cartridge and preventing
Detection analyzing with deteriorated dye.

Reading the RFID tag of a cartridge installed in the Fluorocell RET holder in the following
timing, and comparing the usage status stored in the RFID tag with number of remaining
Method of Error tests.
Detection
1. When starting the analyzer.
2. When closing the FRUOLOCELL RET cover.

It will become an error when the following condition formulas are met:
CQC+{(IQC/10) x ROB}>IQC
Conditions of Error
1. The number of remaining tests stored in RFID : CQC (Current Quantity Count)
Detection
2. The number of the total tests stored in RFID: IQC (Initial Quantity Control)
3. The number of the Read Only Blocks in use status check block: ROB

Error Message Fluorocell RET has already been used

Discrete except RET can be analyzed in manual mode. Install new Fluorocell RET in the
Action Message
Fluorocell RET holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell RET with the correct number of use status check-
blocks is installed in the Fluorocell RET holder and closing the dye cover.

Error Code 135213 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-189 March 2015


6.5.16.15 Fluorocell PLT has already been used

Type A / Type B (Occurring by only the configuration of PLT-F


Analyzer
Applicable analysis)

Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Analysis Stop Reanalysis NA


Process When Error Sampler
Occurred NA
Measurement

Analyzable when only CBC, DIFF, RET, and WPC are specified
Important Notice
as discrete and also with manual mode.

Meaning The dye cartridge has already been spent.

Purpose of Error Preventing deteriorating a dye from refilling dye into dye cartridge and preventing
Detection analyzing with deteriorated dye.

Reading the RFID tag of a cartridge installed in the Fluorocell PLT holder in the following
timing, and comparing the usage status stored in the RFID tag with number of remaining
Method of Error tests.
Detection
1. When starting the analyzer.
2. When closing the FRUOLOCELL PLT cover.

It will become an error when the following condition formulas are met:
CQC+{(IQC/10) x ROB}>IQC
Conditions of Error
1. The number of remaining tests stored in RFID : CQC (Current Quantity Count)
Detection
2. The number of the total tests stored in RFID: IQC (Initial Quantity Control)
3. The number of the Read Only Blocks in use status check block: ROB

Error Message Fluorocell PLT has already been used

Discrete except PLT-F can be analyzed in manual mode. Install new Fluorocell PLT in the
Action Message
Fluorocell PLT holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell RET with the correct number of use status check-
blocks is installed in the Fluorocell RET holder and closing the dye cover.

Error Code 135223 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-190 March 2015


6.5.16.16 Cannot recognize Fluorocell WNR information

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None

Meaning Dye information installed in the dye holder cannot be recognized.

Purpose of Error 1. To monitor the remaining volume of the reagent.


Detection 2. To check the RFID tag damage.

Read the cartridge RFID tag placed in the Fluorocell WNR holder in the following timing,
and compare the read UID from RFID tag with the stored encrypted UID in the tag.
Method of Error
Detection 1. When starting the analysis
2. When closing the Fluorocell WNR cover.

Conditions of Error
The UID from RFID tag and the stored encrypted UID do not match.
Detection

Error Message Cannot recognize Fluorocell WNR information

Action Message Install normal Fluorocell WNR in the Fluorocell WNR holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell WNR with the correct UID is installed in the Fluorocell
WNR holder.

Error Code 135204 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-191 March 2015


6.5.16.17 Cannot recognize Fluorocell WDF information

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None

Meaning Dye information installed in the dye holder cannot be recognized.

Purpose of Error 1. To monitor the remaining volume of the reagent.


Detection 2. To check the RFID tag damage.

Read the cartridge RFID tag placed in the Fluorocell WDF holder in the following timing,
and compare the read UID from RFID tag withthe stored encrypted UID in the tag.
Method of Error
Detection 1. When starting the analysis
2. When closing the Fluorocell WDF cover.

Conditions of Error
The UID from RFID tag and the stored encrypted UID do not match.
Detection

Error Message Cannot recognize Fluorocell WDF information

Action Message Install normal Fluorocell WDF in the Fluorocell WDF holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell WDF with the correct UID is installed in the Fluorocell
WDF holder.

Error Code 133154 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-192 March 2015


6.5.16.18 Cannot recognize Fluorocell WPC information

Analyzer Type A
Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None

Meaning Dye information installed in the dye holder cannot be recognized.

Purpose of Error 1. To monitor the remaining volume of the reagent.


Detection 2. To check the RFID tag damage.

Read the cartridge RFID tag placed in the Fluorocell WPC holder in the following timing,
and compare the read UID from RFID tag withthe stored encrypted UID in the tag.
Method of Error
Detection 1. When starting the analysis
2. When closing the Fluorocell WPC cover.

Conditions of Error
The UID from RFID tag and the stored encrypted UID do not match.
Detection

Error Message Cannot recognize Fluorocell WDF information

Action Message Install normal Fluorocell WPC in the Fluorocell WPC holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell WDF with the correct UID is installed in the Fluorocell
WPC holder.

Error Code 135234 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-193 March 2015


6.5.16.19 Cannot recognize Fluorocell RET information

Type A / Type B (Occurring by only the configuration of RET


Analyzer
Applicable analysis)

Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None

Meaning Dye information installed in the dye holder cannot be recognized.

Purpose of Error 1. To monitor the remaining volume of the reagent.


Detection 2. To check the RFID Tag damage.

Read the cartridge RFID tag placed in the Fluorocell RET holder in the following timing,
and compare the read UID from RFID tag withthe stored encrypted UID in the tag.
Method of Error
Detection 1. When starting the analysis
2. When closing the Fluorocell RET cover.

Conditions of Error
The UID from RFID tag and the stored encrypted UID do not match.
Detection

Error Message Cannot recognize Fluorocell RET information

Action Message Install normal Fluorocell RET in the Fluorocell RET holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell RET with the correct UID is installed in the Fluorocell
RET holder.

Error Code 135214 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-194 March 2015


6.5.16.20 Cannot recognize Fluorocell PLT information

Type A / Type B (Occurring by only the configuration of PLT-F


Analyzer
Applicable analysis)

Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis NA

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None

Meaning Dye information installed in the dye holder cannot be recognized.

Purpose of Error 1. To monitor the remaining volume of the reagent.


Detection 2. To check the RFID tag damage.

Read the cartridge RFID tag placed in the Fluorocell PLT holder in the following timing,
and compare the read UID from RFID tag withthe stored encrypted UID in the tag.
Method of Error
Detection 1. When starting the analysis
2. When closing the Fluorocell PLT cover.

Conditions of Error
The UID from RFID tag and the stored encrypted UID do not match.
Detection

Error Message Cannot recognize Fluorocell PLT information

Action Message Install normal Fluorocell PLT in the Fluorocell PLT holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell PLT with the correct UID is installed in the Fluorocell
PLT holder.

Error Code 135224 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-195 March 2015


6.5.16.21 Fluorocell WNR RFID tag error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Warning Reanalysis NA


Process When Error
Occurred Sampler
NA
Measurement

Important Notice None

Meaning The remaining volume of dye cannot be written in the RFID tag.

Purpose of Error 1. To monitor the remaining volume of the reagent.


Detection 2. To check the RFID tag damage.

Method of Error
Check the response from the RFID tag when operating writing-access to RFID
Detection

Conditions of Error
No response from the RFID tag.
Detection

Error Message Fluorocell WNR RFID tag error

Action Message Install Fluorocell WNR with normal RFID added in the Fluorocell WNR holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell WNR with correct RFID is installed in the Fluorocell
WNR holder.

Error Code 135206 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-196 March 2015


6.5.16.22 Fluorocell WDF RFID tag error

Analyzer Type A / Type B


Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Warning Reanalysis NA


Process When Error Sampler
Occurred NA
Measurement

Analyzable when only CBC, RET, and PLT-F are specified as


Important Notice
discrete and also with the manual mode

Meaning The remaining volume of dye cannot be written in the RFID tag.

Purpose of Error 1. To monitor the remaining volume of the reagent.


Detection 2. To check the RFID tag damage.

Method of Error
Check the response from the RFID tag when operating writing-access to RFID
Detection

Conditions of Error
No response from the RFID tag.
Detection

Error Message Fluorocell WDF RFID tag error

Discrete except DIFF can be analyzed in manual mode. Install Fluorocell WDF with
Action Message
normal RFID added in the Fluorocell WDF holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell WDF with correct RFID is installed in the Fluorocell
WDF holder.

Error Code 133156 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-197 March 2015


6.5.16.23 Fluorocell WPC RFID tag error

Analyzer Type A
Applicable
Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Warning Reanalysis NA


Process When Error Sampler
Occurred NA
Measurement

Analyzable when only CBC, DIFF, RET, and PLT-F are specified
Important Notice
as discrete and also with the manual mode

Meaning The remaining volume of dye cannot be written in the RFID tag.

Purpose of Error 1. To monitor the remaining volume of the reagent.


Detection 2. To check the RFID tag damage.

Method of Error
Check the response from the RFID tag when operating writing-access to RFID
Detection

Conditions of Error
No response from the RFID tag.
Detection

Error Message Fluorocell WPC RFID tag error

Discrete except WPC can be analyzed in manual mode. Install Fluorocell WPC with
Action Message
normal RFID added in the Fluorocell WPC holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell WPC with correct RFID is installed in the Fluorocell
WPC holder.

Error Code 135236 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-198 March 2015


6.5.16.24 Fluorocell RET RFID tag error

Type A / Type B (Occurring by only the configuration of RET


Analyzer
Applicable analysis)

Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Warning Reanalysis NA


Process When Error Sampler
Occurred NA
Measurement

Analyzable when only CBC, DIFF, PLT-F, and WPC are specified
Important Notice
as discrete and also with the manual mode

Meaning The remaining volume of dye cannot be written in the RFID tag.

Purpose of Error 1. To monitor the remaining volume of the reagent.


Detection 2. To check the RFID tag damage.

Method of Error
Check the response from the RFID tag when operating writing-access to RFID
Detection

Conditions of Error
No response from the RFID tag.
Detection

Error Message Fluorocell RET RFID tag error

Discrete except RET can be analyzed in manual mode. Install Fluorocell RET with normal
Action Message
RFID added in the Fluorocell RET holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell RET with correct RFID is installed in the Fluorocell
RET holder.

Error Code 135216 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-199 March 2015


6.5.16.25 Fluorocell PLT RFID tag error

Type A / Type B (Occurring by only the configuration of PLT-F


Analyzer
Applicable analysis)

Sampler NA

Error Category Caution

Data Mask NA Invalid Analysis ✔

Alarm Sounds Warning Reanalysis NA


Process When Error Sampler
Occurred NA
Measurement

Analyzable when only CBC, DIFF, RET, and WPC are specified
Important Notice
as discrete and also with the manual mode

Meaning The remaining volume of dye cannot be written in the RFID tag.

Purpose of Error 1. To monitor the remaining volume of the reagent.


Detection 2. To check the RFID tag damage.

Method of Error
Check the response from the RFID tag when operating writing-access to RFID
Detection

Conditions of Error
No response from the RFID tag.
Detection

Error Message Fluorocell PLT RFID tag error

Discrete except PLT-F can be analyzed in manual mode. Install Fluorocell PLT with
Action Message
normal RFID added in the Fluorocell PLT holder.

None.
HELP [Execute]
Process The error clears when the Fluorocell PLT with correct RFID is installed in the Fluorocell
PLT holder.

Error Code 135226 Judgment Control Unit

HELP Screen Key [Accept] (grayout)


Detailed Information
Maintenance Menu None

Show Detailed Procedure Key None

XN-1000/2000 S/M 6-200 March 2015


6.6 SA-10/SA-20 Sensors

6.6.1 Sensor List (Sampler Sonsor Display)

Description
Sensor No.
(SA-10/SA-20)
S1 Not Used

S2 Not Used

S3 Sample tube monitoring

S4 Rack full

Left sampler pool return


S5
prevention stopper

S6 Rack feed-in arrival monitoring

S7 Blood monitoring

S8 Right sampler pool detecting

Analysis line termination


S9
monitoring

S10 Not Used

S11 Not Used

S12 Not Used

S13 Not Used

S14 Not Used

S15 Not Used

S16 Not Used

Analysis line termination


monitoring sensor Sample tube monitoring sensor Tube rotation mechanism
6HQVRU 
6HQVRU 

Rack feed-in arrival


Blood monitoring sensor monitoring sensor
6HQVRU 
6HQVRU 

Right sampler
pool return
Analysis line prevention
Left sampler pool
(Rack horizontal-feed unit) stopper
return prevention
stopper
6HQVRU  Right sampler pool
Left sampler pool Left sampler pool (Rack feed-in table)
rack full sensor
6HQVRU 


Right sampler pool


rack detecting sensor
6HQVRU 

201362

XN-1000/2000 S/M 6-201 March 2015


6.6.2 SA-10 Sensor List (Other than Sampler Sonsor Display)
No. Unit Assy Description Sensor Code No.
1 FEEDER_ASSY NO.34 Rack feed-in arm home position KI1235-AALF 22837023

Stopper position monitor sensor for


2 FEEDER_ASSY NO.34 KI1235-AALF 22837023
Supply Rack

3 FEEDER_ASSY NO.85 Rack ramaining monitor (transmitter) KB1242-AA14LF 22837068

4 FEEDER_ASSY NO.34 Rack ramaining monitor (receiver) KB1240-AA23LF 22837054

5 FEEDER_ASSY NO.85 Rack feed-in arrival monitor (transmitter) KB1242-AA14LF 22837068

6 FEEDER_ASSY NO.85 Rack feed-in arrival monitor (receiver) KB1240-AA23LF 22837054

7 FEEDER_ASSY NO.85 Rack conveyor belt (rear) home position KB1242-AA14LF 22837068

8 FEEDER_ASSY NO.85 Rack conveyor belt (rear) home position KB1240-AA23LF 22837054

9 FEEDER_ASSY NO.85 Rack conveyor belt (front) home position KB1242-AA14LF 22837068

10 FEEDER_ASSY NO.85 Rack conveyor belt (front) home position KB1240-AA23LF 22837054

11 FEEDER_ASSY NO.85 Blood monitoring KB1242-AA14LF 22837068

12 FEEDER_ASSY NO.85 Blood monitoring KB1242-AA22LF 22837103

13 FEEDER_ASSY NO.85 Analysis line termination monitoring KB1242-AA14LF 22837068

14 FEEDER_ASSY NO.85 Analysis line termination monitoring KB1240-AA23LF 22837054

15 FEEDER_ASSY NO.85 Return provention stopper position KI1235-AALF 22837023

DRIVE MECH_ASSY
16 Sample tube monitoring KI1235-AALF 22837023
NO.84

17 FEEDER_ASSY NO.35 Rack feed-out slideder home position KI1235-AALF 22837023

SA-10 MAIN
18 Rack full KR1207-AALF 22837071
COMPLETE

XN-1000/2000 S/M 6-202 March 2015


6.6.3 SA-20 Sensor List (Other than Sampler Sensor Display)
No. Unit Assy Description Sensor Code No.
1 FEEDER_ASSY NO.32 Rack feeding arm home position detect KI1235-AALF 22837023

2 FEEDER_ASSY NO.32 Stopper position monitor for Supply Rack KI1235-AALF 22837023

3 FEEDER_ASSY NO.46 Rack ramaining monitor (transmitter) KB1242-AA14LF 22837068

4 FEEDER_ASSY NO.32 Rack ramaining monitor (receiver) KB1240-AA23LF 22837054

5 FEEDER_ASSY NO.46 Rack feed-in arrival monitor (transmitter) KB1242-AA14LF 22837068

6 FEEDER_ASSY NO.46 Rack feed-in arrival monitor (receiver) KB1240-AA23LF 22837054

7 FEEDER_ASSY NO.46 Rack conveyor belt (rear) home position KB1242-AA14LF 22837068

8 FEEDER_ASSY NO.46 Rack conveyor belt (rear) home position KB1240-AA23LF 22837054

9 FEEDER_ASSY NO.46 Rack conveyor belt (front) home position KB1242-AA14LF 22837068

10 FEEDER_ASSY NO.46 Rack conveyor belt (front) home position KB1240-AA23LF 22837054

11 FEEDER_ASSY NO.46 Blood monitoring KB1242-AA14LF 22837068

12 FEEDER_ASSY NO.46 Blood monitoring KB1242-AA22LF 22837103

13 FEEDER_ASSY NO.46 Analysis line termination monitoring KB1242-AA14LF 22837068

14 FEEDER_ASSY NO.46 Analysis line termination monitoring KB1240-AA23LF 22837054

15 FEEDER_ASSY NO.46 Return provention stopper position KI1235-AALF 22837023

DRIVE MECH_ASSY
16 Sample tube monitoring KI1235-AALF 22837023
NO.84

17 FEEDER_ASSY NO.35 Rack feed-out slideder home position KI1235-AALF 22837023

SA-10 MAIN
18 Rack full KR1207-AALF 22837071
COMPLETE

XN-1000/2000 S/M 6-203 March 2015


6.7 Sampler (Unidiractional Sampler SA-01) Operation Errors

6.7.1 Rack Feed In error

Application Main Unit NA

Sampler Unidirectional Sampler (SA-01)

Error Category Hardware Error

Process When Error NA


Data Mask NA Invalid Analysis
Occurred

Alarm Sounds 3 Reanalysis NA

Sampler
Unidirectional Sampler Analysis Stop 3
Measurement

Important Notice None

Meaning A rack feed-in error occurs.

Purpose of Error To ensure that the rack on the feed-in rack table sends into the
Detection measurement line.

Check the Rack Feed-in Arrival Sensor and the Feed-in Table Rack
Method of Error
Detection Sensor
Detection
during operating rack feed-in.

When the rack feed-in motor operates a specified number of pulses, the
Conditions of Error
rack feed-in arrival monitoring sensor is not shaded and the the rack
Detection
feed-in table rack detecting sensor is not shaded.

Error Messages Rack Feed-in Error:

Action Message Remove the rack and tubes from the tube holder and analysis line and
Screen press the [Execute]. The rack feed-in mechanism will be initialized.

HELP [Execute] The rack feed-in mechanism will be initialized. The error clears when
Process the process above is completed.

Detailed Information Error Code 253000 Judgment Sequence

HELP Screen Key [Run]

Maintenance Menu None

Show Detailed
None
Procedure Key

2015037

XN-1000/2000 S/M 6-204 March 2015


6.7.2 Rack feed-in home position error

Application Main Unit NA

Sampler Unidirectional Sampler (SA-01)

Error Category Hardware Error

Process When Error


Data Mask NA Invalid Analysis NA
Occurred

Alarm Sounds 3 Reanalysis NA

Sampler
Unidirectional Sampler Analysis Stop 3
Measurement

Important Notice None

Meaning A rack feed-in mechanism initialized error occurs.

To ensure that the rack feed-in unit lever return to the home position
Purpose of Error
after the rack feed-in operated, and the next rack feed operates
Detection
properly.

Check the Rack Feed-in Home Position Sensor during initializing the
Method of Error
rack feed-in
Detection
mechanism.

When the rack feed-in motor operates a specified number of pulses, the
Conditions of Error
rack feed-in
Detection
home position sensor is not be shaded.

Error Messages Rack feed-in home position error

Action Message Remove the rack and tubes from the tube holder and analysis line and
Screen press the [Execute]. The rack feed-in mechanism will be initialized.

HELP [Execute] The rack feed-in mechanism will be initialized. The error clears when
Process the process above is completed.

Detailed Information Error Code 253010 Judgment Sequence

HELP Screen Key [Run]

Maintenance Menu None

Show Detailed
None
Procedure Key

XN-1000/2000 S/M 6-205 March 2015


6.7.3 Rack move error

Application Main Unit NA

Sampler Unidirectional Sampler (SA-01)

Error Category Hardware Error

Process When Error


Data Mask NA Invalid Analysis NA
Occurred

Alarm Sounds 3 Reanalysis NA

Sampler
Unidirectional Sampler Analysis Stop 3
Measurement

Important Notice None

Meaning Rack moving error occurs

Purpose of Error
To ensure that the rack is moved to the designated location properly.
Detection

Method of Error
Check the Rack Move Sensor (Right, Left), when rack is moving.
Detection

Error is checked when both Move sensor (Right and Left) are not
shaded for more than 50 ms during the process from starting the
shifting motor operation to the completely returning to the initial position.
System monitors at a constant timing (every 0.01 second).
Error is checked when Move sensor (Right) or Move sensor (Left) is not
shaded during the process from the finishing rack shift operation to next
Conditions of Error rack shift operation. System monitors at a constant timing (every 0.1
Detection second).
Except, while from hand picks up test tube though hand returns test
tube to rack (with performing mixing and analysis).
Error is checked when Tube Position Sensor (right or left) is shaded at
the timing during the interrupt READY. System monitors at a constant
timing (every 0.1 second).

Error Messages Rack move error

Action Message Remove the rack and tubes from the tube holder and analysis line and
Screen press the [Execute]. The rack move mechanism will be initialized.

HELP [Execute] The rack move mechanism will be initialized. The error clears when the
Process process above is completed.

Detailed Information Error Code 252001 Judgment Sequence

HELP Screen Key [Run]

Maintenance Menu None

Show Detailed
None
Procedure Key

XN-1000/2000 S/M 6-206 March 2015


6.7.4 Rack move home position error

Application Main Unit NA

Sampler Unidirectional Sampler (SA-01)

Error Category Hardware Error

Process When Error


Data Mask NA Invalid Analysis NA
Occurred

Alarm Sounds 3 Reanalysis NA

Sampler
Unidirectional Sampler Analysis Stop 3
Measurement

Important Notice None

Meaning A rack move mechanism initialized error occurs.

To ensure the rack shift lever to be properly returned to the initial


Purpose of Error
position after shifting operation and the shifting operation for the next
Detection
rack can be properly started.

Method of Error Check the Move Home Position Sensor during initializing the move
Detection mechanism.

When the rack move motor operates a specified number of pulses, the
Conditions of Error
rack move
Detection
home position sensor is not be shaded.

Error Messages Rack move home position error

Action Message Remove the rack and tubes from the tube holder and analysis line and
Screen press the [Execute]. The rack move mechanism will be initialized.

HELP [Execute]
The error clears when the process above is completed.
Process

Detailed Information Error Code 254010 Judgment Sequence

HELP Screen Key [Run]

Maintenance Menu None

Show Detailed
None
Procedure Key

XN-1000/2000 S/M 6-207 March 2015


6.7.5 Rack ejection error

Application Main Unit NA

Sampler Unidirectional Sampler (SA-01)

Error Category Hardware Error

Process When Error


Data Mask NA Invalid Analysis NA
Occurred

Alarm Sounds 3 Reanalysis NA

Sampler
Unidirectional Sampler Analysis Stop 3
Measurement

Important Notice None

Meaning Rack ejection error occurs.

Purpose of Error To ensure that the rack ejects from the analysis line to the rack ejection
Detection table properly.

Method of Error Check the Rack Ejection Home Position Sensor during ejecting the
Detection rack.

Conditions of Error When the rack eject motor operates a specified number of pulses, the
Detection rack eject home position sensor is shaded.

Error Messages Rack ejection error

Action Message Remove the rack and tubes from the tube holder and analysis line and
Screen press the [Execute].

HELP [Execute] To perform Initializing rack eject mechanism The error clears when the
Process process above is completed.

Detailed Information Error Code 255000 Judgment Sequence

HELP Screen Key [Run]

Maintenance Menu None

Show Detailed
None
Procedure Key

XN-1000/2000 S/M 6-208 March 2015


6.7.6 Rack ejection home position error

Application Main Unit NA

Sampler Unidirectional Sampler (SA-01)

Error Category Hardware Error

Process When Error


Data Mask NA Invalid Analysis NA
Occurred

Alarm Sounds 3 Reanalysis NA

Sampler
Unidirectional Sampler Analysis Stop 3
Measurement

Important Notice None

Meaning A rack eject mechanism initialized error occurs.

To ensure that the lever of the rack ejection mechanism return to the
Purpose of Error
home position and
Detection
the rack ejection mechanism is initialized properly.

Method of Error Check the Rack Ejection Home Position Sensor during ejecting the
Detection rack.

Conditions of Error When the rack eject motor operates a specified number of pulses, the
Detection rack eject home position sensor is not shaded.

Error Messages Rack ejection home position error

Action Message Remove the rack and tubes from the tube holder and analysis line and
Screen press the [Execute]. The rack move mechanism will be initialized.

HELP [Execute] To perform Initializing rack eject mechanism The error clears when the
Process process above is completed.

Detailed Information Error Code 255010 Judgment Sequence

HELP Screen Key [Run]

Maintenance Menu None

Show Detailed
None
Procedure Key

XN-1000/2000 S/M 6-209 March 2015


6.7.7 Sampler belt error

Application Main Unit NA

Sampler Unidirectional Sampler (SA-01)

Error Category Hardware Error

Process When Error


Data Mask NA Invalid Analysis NA
Occurred

Alarm Sounds 3 Reanalysis NA

Sampler
Unidirectional Sampler Analysis Stop 3
Measurement

Important Notice None

Meaning The rack exists on the analysis line when analysis starts.

Purpose of Error
To ensure the sampler analysis starts properly.
Detection

Method of Error
System checks blood volume sensor and tube detecting sensor.
Detection

Error is checked when the tube detection sensor detects the tube when
starting the sample analysis.
Conditions of Error Error is checked when the sample volume sensor detects blood for the
Detection tube when starting the sample analysis
Error is checked when tube move sensor (right or left) is not shaded
when starting the sampler analysis.

Error Messages Sampler belt error

Action Message Rack detected on analysis line at start of sampler analysis. Remove rack from
Screen analysis line and press [Accept].

HELP [Execute]
None. The error clears when [Accept] key is pressed.
Process

Detailed Information Error Code 422100 Judgment Sequence

HELP Screen Key Check

Maintenance Menu None

Show Detailed
None
Procedure Key

XN-1000/2000 S/M 6-210 March 2015


6.7.8 Rack Removed

Application Main Unit NA

Sampler Unidirectional Sampler (SA-01)

Error Category Caution

Process When Error


Data Mask NA Invalid Analysis NA
Occurred

Alarm Sounds 2 Reanalysis NA

Sampler
Unidirectional Sampler Analysis Stop 3
Measurement

Important Notice None

The rack on the measurement line is moved while a tube is clamped by


Meaning
the hand clipper from the rack.

To ensure the rack is located on the measurement line and the tube can
Purpose of Error
be returned to the original position in the rack after the sample is
Detection
aspirated.

Method of Error System checks move Sensor(right and left). When error occurs, hand
Detection clipper does not descend.

Error is checked when Move sensor (Right) or Move sensor (Left) is not
Conditions of Error shaded during the process from the finishing rack shift operation to next
Detection rack shift operation. System monitors at a constant timing (every 0.1
second).

Error Messages Rack Removed

Action Message Press [Accept], place rack on feed-in table again, and start sampler
Screen analysis.

HELP [Execute]
None. The error clears when [Accept] key is pressed.
Process

Detailed Information Error Code 422030 Judgment Sequence

HELP Screen Key Check

Maintenance Menu None

Show Detailed
None
Procedure Key

XN-1000/2000 S/M 6-211 March 2015


6.7.9 Rack not placed on feed-in table

Application Main Unit NA

Sampler Unidirectional Sampler (SA-01)

Error Category Caution

Process When Error


Data Mask NA Invalid Analysis NA
Occurred

Alarm Sounds 2 Reanalysis NA

Sampler
NA (Occurs only before sampler analysis starts.)
Measurement

Important Notice None

Meaning Rack is not placed on rack feed-in table when sampler analysis starts.

Purpose of Error
To urge operator to set the racks in the right rack pool.
Detection

Method of Error
System checks Rack Feed In Sensor or Rack Feed In Lever Sensor.
Detection

Conditions of Error Error is checked when rack feed in lever sensor is shaded after rack
Detection feed in operation starts.

Error Messages Rack not placed on feed-in table

Action Message Press [Accept], place rack on feed-in table again, and start sampler
Screen analysis.

HELP [Execute]
The error clears when [Accept] key is pressed.
Process

Detailed Information Error Code 422020 Judgment Sequence

HELP Screen Key Check

Maintenance Menu None

Show Detailed
None
Procedure Key

XN-1000/2000 S/M 6-212 March 2015


6.7.10 Ejection table is full

Application Main Unit NA

Sampler Unidirectional Sampler (SA-01)

Error Category Caution

Process When Error


Data Mask NA Invalid Analysis NA
Occurred

Alarm Sounds 2 Reanalysis NA

Sampler
Unidirectional Sampler Analysis Stop 3
Measurement

Important Notice None

Meaning The rack ejection table is full with racks.

Purpose of Error
Prompt the user to remove racks from the rack ejection table.
Detection

Method of Error Prompt the user to remove racks from the rack ejection table. When
Detection error occurs, rack is not shifted. It resumes when error is recovered.

Error is checked when the rack full sensor detects the rack when
Conditions of Error starting the sampler mode analysis.
Detection When the racks are detected by the rack ejection table rack full sensor
during starting the rack ejection operation.

Error Messages Ejection table is full

Action Message
Remove racks from rack ejection table.
Screen

None. The error clears when the racks were not detected by the rack
HELP [Execute]
ejection table rack full
Process
sensor.

Detailed Information Error Code 422010 Judgment Sequence

HELP Screen Key [Accept] (grayout)

Maintenance Menu None

Show Detailed
None
Procedure Key

XN-1000/2000 S/M 6-213 March 2015