6.1 Introduction
This troubleshooting describes the error monitoring and error recovery processing of the analysis
instruments (analyzer and sampler) on the XN-1000/2000 software specifications. The errors are
categorized based on the basic concept in the next section, and error recovery processing is
described.
Analysis error is an error related to result data. This error affects the analysis
Analysis Error
data.
2. On Manual Analysis, the analysis channels which are not affected by errors can be analyzed by
limiting the discrete even though the error occurs. However, because Sampler Analysis follows
“work list setting “, it cannot be analyzed by Sampler Analysis.
6.3.1 Error Recovery processing to be Common to Single Model and Twin Model
Recovery processing on an error occurrence while analyzing by Single Model and/or Twin Model is
classified into “Sampler Analysis continue”, “Sampler Analysis suspend” and “Sampler Analysis
stop. “
When multiple errors occur at the same time, give priority to processing in order of “Sample
Analysis is stop”, “Sampler Analysis suspend”, and “Sampler Analysis continue.” Details of each
processing are as follows:
6.3.2.1 Conditions for continuing the sampler analysis when using Twin Model
On the sampler analysis using Twin Model, if sampler analysis is possible by the other analyzer,
sample analysis is continued, even if one of the analyzers is incapable of analyzing (including
switching to Manual Analysis Mode).
The sampler analysis suspends when all analyzers connecting to the sampler are incapable of
analyzing.
After suspending, when one of the analyzers becomes capable of sample analysis, sampler
analysis restarts.
However, the sampler analysis suspends without depending on the status of analyzer if the
following conditions are congruent with the errors.
1. When error is congruent with the suspending condition by user setting. (sampler ID read
error or the others)
2. When error is congruent with the suspending condition by system setting. (replacement of
common reagents or others)
Also, the sampler analysis and transport suspend without depending on the status of the analyzer if
the following conditions are congruent with the errors.
1. When the error relating to the Hand Unit (including pressure error) occurs.
2. When the sampler mechanism error occurs.
The errors are classified into 4 categories as follows by sources of error monitoring. Depending on
the sampler and analyzer, there are related or unrelated categories to each instrument. (○=related
categories)
Instrument
Category Details of error
Analyzer Sampler
Reanalysis when the data mask error occurred while operating sample analysis.
On the next chapter, describing “re-analysis” as follows: (Changeable by reanalysis rule
setting)
Block: Not reanalyzing whether Repeat error occurs.
Reanalysis
Repeat: Reanalyzing when Block error does not occur.
None: Reanalyzing depending on whether Block / Repeat errors occur or not.
NA: Errors inapplicable reanalysis Judgment.
Table 6-4-2 Processing when errors occurred (describing details on the next chapter)
Purpose of error
To ensure the valve performing.
detection
Method of error
Check 0.25MPa pressure sensor while analysis operating (always)
detection
Conditions of error
When the pressure is less than 0.15Mpa for 3 seconds continuously.
detection
Operation could not be continued due to pressure loss. Remove rack and tubes from
Action Message
analyzer and press [Execute] key. Analyzer initialize will be executed.
Purpose of error
To ensure the solenoid Valve, Valve, and DP performing.
detection
Method of error
Check 0.25MPa pressure sensor when analyzer starts and while operating analysis.
detection
0.25 MPa pressure is outside the required range. Press [Execute] key. A pressure
Action Message
adjustment will be executed.
Purpose of Error
To ensure the Sheath pressure used on the FCM.
Detection
Method of Error
Check 0.16MPa pressure sensor when analyzer starts and while operating analysis.
Detection
0.16 MPa pressure is outside the required range. Press [Execute] key. A pressure
Action Message
adjustment will be executed.
Purpose of Error
To ensure the Sheath pressure used on the RBC/PLT counting, DP, and rinse.
Detection
Method of Error
Check 0.07MPa pressure sensor when analyzer starts and while operating analysis.
Detection
0.07 MPa pressure is outside the required range. Press [Execute] key. A pressure
Action Message
adjustment will be executed.
Purpose of Error To ensure charging volume of dilution sample, draining reaction chamber, DP
Detection operations.
Method of Error
Check -0.04MPa pressure sensor when analyzer starts and while operating analysis.
Detection
-0.04 MPa pressure is outside the required range. Press [Execute] key. A pressure
Action Message
adjustment will be executed.
Purpose of Error To ensure temperatures of WNR Lysing, WDF Lysing, and RET/PLT-F Diluent.
Detection To ensure the reaction of samples.
Method of Error
Check 41°C Reagent Heater Thermister when in ready status and starting analysis.
Detection
41°C reagent heater temperature is outside the required range. Press [Execute] key. The
Action Message
temperature sensor screen will appear.
High:125040
Error Code Judgment Control Unit
Low: 125045
Analyzer Type A
Applicable
Sampler NA
Method of Error
Check 34°C Reagent Heater Thermister when in ready status and starting analysis.
Detection
34°C reagent heater temperature is outside the required range. Press [Execute] key. The
Action Message
temperature sensor screen will appear.
High:125050
Error Code Judgment Control Unit
Low: 125055
The temperatures of WNR reaction chamber, WDF reaction chamber, RET reaction
Meaning
chamber and PLT-F reaction chamber are outside the specified range.
To ensure temperatures of WNR reaction chamber, WDF reaction chamber, RET reaction
Purpose of Error chamber and PLT-F reaction chamber.
Detection
To ensure the stainlyzing reactions of WNR, WDF, RET and PLT-F.
Method of Error
Check 41°C Reaction Thermister when in ready status and starting analysis.
Detection
41°C FCM reaction chamber temperature is outside the required range. Press [Execute]
Action Message
key. The temperature sensor screen will appear.
High:123250
Error Code Judgment Control Unit
Low: 123255
Analyzer Type A
Applicable
Sampler NA
Meaning The temperature of WPC reaction chamber is outside the specified range.
Method of Error
Check 34°C Reaction Thermister when in ready status and starting analysis.
Detection
34°C FCM reaction chamber temperature is outside the required range. Press [Execute]
Action Message
key. The temperature sensor screen will appear.
High:123260
Error Code Judgment Control Unit
Low: 123265
Purpose of Error
To ensure temperatures of FCM detector.
Detection
Method of Error
Check FCM detector thermister when in ready status and starting analysis.
Detection
FCM detector temperature is outside the required range. Press [Execute] key. The
Action Message
temperature sensor screen will appear.
High:121250
Error Code Judgment Control Unit
Low: 121255
Purpose of Error
To ensure temperatures of FCM sheath.
Detection
Method of Error
Check FCM sheath thermister when in ready status and starting analysis.
Detection
FCM sheath temperature is outside the required range. Press [Execute] key. The
Action Message
temperature sensor screen will appear.
High:124110
Error Code Judgment Control Unit
Low: 124115
Meaning The environment temperature of the instrument is outside the specified range.
Purpose of Error Check the temperature of use environment for the instrument.
Detection Check that the environment temperature thermistor is not defective.
Method of Error
Check the Environment Thermister when in ready status and starting analysis.
Detection
Environment temperature is outside the required range. Press [Execute] key. The
Action Message
temperature sensor screen will appear.
High:122510
Error Code Judgment Control Unit
Low: 122515
41°C Reagent Heater Thermistor is defective or 41°C reagent heater thermistor wire is
Meaning
broken.
Purpose of Error
To ensure that 41°C reagent heater thermistor performs properly.
Detection
Method of Error
Check the error information of the driver board in ready status or operating.
Detection
1. Ready Status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.
Analysis cannot be continued because the 41°C reagent heater thermistor wire is broken.
Action Message
Remove the rack and tubes from the analyzer and turn off the analyzer power.
Analyzer Type A
Applicable
Sampler NA
34°C Reagent Heater Thermistor is defective or 41°C reagent heater thermistor wire is
Meaning
broken.
Purpose of Error
To ensure that 34°C reagent heater thermistor performs properly.
Detection
Method of Error
Check the error informationof the driver board in ready status or operating.
Detection
1. Ready Status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.
Analysis cannot be continued because the 34°C reagent heater thermistor wire is broken.
Action Message
Remove the rack and tubes from the analyzer and turn off the analyzer power.
41°C Reaction Chamber Thermistor is defective or 41°C reaction chamber thermistor wire
Meaning
is broken.
Purpose of Error
To ensure that 41°C reaction chamber thermistor performs properly.
Detection
Method of Error
Check the error information of the driver board in ready status or operating.
Detection
1. Ready status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.
Analysis cannot be continued because the 41°C FCM reaction chamber thermistor wire is
Action Message
broken. Remove the rack and tubes from the analyzer and turn off the analyzer power.
Analyzer Type A
Applicable
Sampler NA
34°C Reaction Chamber Thermistor is defective or 34°C reaction chamber thermistor wire
Meaning
is broken.
Purpose of Error
To ensure that 34°C reaction chamber thermistor performs properly.
Detection
Method of Error
Check the error information of the driver board in ready status or operating.
Detection
1. Ready status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.
Analysis cannot be continued because the 34°C FCM reaction chamber thermistor wire is
Action Message
broken. Remove the rack and tubes from the analyzer and turn off the analyzer power.
Analyzer Type A
Applicable
Sampler NA
Meaning FCM Detector Thermistor is defective or FCM detector thermistor wire is broken.
Purpose of Error
To ensure that FCM detector thermistor performs properly.
Detection
Method of Error
Check the error information of the driver board in ready status or operating.
Detection
1. Ready Status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.
Analysis cannot be continued because the FCM detector thermistor wire is broken.
Action Message
Remove the rack and tubes from the analyzer and turn off the analyzer power.
Meaning FCM sheath thermistor is defective or FCM sheath thermistor wire is broken.
Purpose of Error
To ensure that FCM detector thermistor performs properly.
Detection
Method of Error
Check the error information of the driver board in ready status or operating.
Detection
1. Ready Status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.
Analysis cannot be continued because the FCM sheath thermistor wire is broken. Remove
Action Message
the rack and tubes from the analyzer and turn off the analyzer power.
Purpose of Error
To ensure that temperature adjustment of each heater performs properly.
Detection
Method of Error
Check the error information of the driver board in ready status or operating.
Detection
1. Ready Status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.
Meaning APD Thermistor for Temperature Control is defective or its wire is broken.
Purpose of Error
To ensure that APD thermistor for temperature control performs properly.
Detection
Method of Error
Check the error information of the driver board in ready status or operating.
Detection
1. Ready Status:
Conditions of Error When the error occurs for 3 seconds continuously.
Detection 2. Operating:
When the error occurs on timing mentioned in the Timing Chart.
Analysis cannot be continued because the APD thermistor wire is broken. Remove the
Action Message
rack and tubes from the analyzer and turn off the analyzer power.
Purpose of Error
To ensure optimal temperature for analysis
Detection
Method of Error Monitor temperature of each unit, when performing temperature stabilization sequence of
Detection the analyzer.
When the temperature of each unit does not reach the following temperatures within 15
minutes after starting temperature stabilization.
1. 41°C Reaction Chamber: 41°C
Conditions of Error 2. 41°C Reagent Heater: 42°C
Detection
3. 34°C Reaction Chamber (only Type A) : 34°C
4. 34°C Reagent Heater (only Type A): 34°C
5. FCM detector Heater: 42°C
Failed to stabilize temperature in each unit. Remove the rack and tubes from the analyzer
Action Message
and turn off the main analyzer power.
Sampler NA
Purpose of Error
Prevent from in flowing air bubbles into the reagent chamber.
Detection
Method of Error Check the Air Bubble Sensor of CELLPACK DCL Aspiration Line during aspirating
Detection CELLPACK DCL.
Conditions of Error
When detecting air bubbles
Detection
Action Message Press [Execute] key. The reagent replacement screen will appear.
201435
Sampler NA
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the Air Bubble Sensor of CELLSHEATH(C) Aspiration Line during aspirating
Detection CELLPACK DCL.
Conditions of Error
When detecting air bubbles
Detection
Action Message Press [Execute] key. The reagent replacement screen will appear.
Sampler NA
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the Air Bubble Sensor of SULFOLYSER Aspiration Line(SLS) during aspirating
Detection SULFOLYSER.
Conditions of Error
When detecting air bubbles.
Detection
Action Message Press [Execute] key. The reagent replacement screen will appear.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the Air Bubble Sensor of Fluorocell WNR Aspiration Line(WNR-2) during aspirating
Detection Fluorocell WNR.
Conditions of Error
When detecting air bubbles.
Detection
Action Message Press [Execute] key. The reagent replacement screen will appear.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the number of remaining tests which is recorded in RFID Tag of Fluorocell WNR
Detection during aspirating Fluorocell WNR.
Conditions of Error
The number of remaining tests is zero.
Detection
Action Message Open the Fluorocell WNR cover and install new Fluorocell WNR.
None
HELP [Execute]
The error clears when Fluorocell WNR with remaining 1 or number of tests set to the dye
Process
holder.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check Air Bubbles Sensor of Lysercell WDF Aspiration Line(WDF-2) during aspirating
Detection Lysercell WDF.
Conditions of Error The number of Lysercell WDF diaphragm pump operations is over 175 times after the
Detection Float Switch of WDF chamber moves to lower position.
Discrete except DIFF can be analyzed in manual mode. Press [Execute] key. The reagent
Action Message
replacement screen will appear.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the number of remaining tests recorded in RFID Tag of Fluorocell WDF during
Detection aspirating Fluorocell.
Conditions of Error
The number of remaining tests is zero.
Detection
Discrete except DIFF can be analyzed in manual mode. Open the Fluorocell WDF cover
Action Message
and install new Fluorocell WDF.
None.
HELP [Execute]
Process The error clears when Fluorocell WNR with 1 or more tests remaining set to the dye
holder.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check Air Bubbles Sensor of Lysercell WPC Aspiration Line(WPC-2) during aspirating
Detection Lysercell WPC.
Conditions of Error
When detecting air bubbles
Detection
Discrete except WPC can be analyzed in manual mode. Press [Execute] key. The reagent
Action Message
replacement screen will appear.
Analyzer Type A
Applicable
Sampler NA
Analyzable when only CBC, DIFF, RET, and PLT-F are specified
Important Notice
as discrete and also with the manual mode
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the number of remaining tests recorded in RFID Tag of Fluorocell WPC during
Detection aspirating Fluorocell.
Conditions of Error
The number of remaining tests is zero.
Detection
Discrete except WPC can be analyzed in manual mode. Open the Fluorocell WPC cover
Action Message
and install new Fluorocell WPC.
None.
HELP [Execute]
Process The error clears when Fluorocell WPC with 1 or more tests remaining set to the dye
holder.
Sampler NA
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check Air Bubbles Sensor of CELLPACK DFL Aspiration Line (DFL) during aspirating
Detection CELLPACK DFL.
Conditions of Error
When detecting air bubbles
Detection
CBC and DIFF can be analyzed in manual mode. Press [Execute] key. The reagent
Action Message
replacement screen will appear.
Type A / Type B
Analyzer
Applicable (Occurring only with the configuration of RET analysis)
Sampler NA
Analyzable when only CBC, DIFF, PLT-F and WPC are specified
Important Notice
as discrete and also with the manual mode
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the number of remaining tests recorded in RFID Tag of Fluorocell RET during
Detection aspirating Fluorocell RET.
Conditions of Error
The number of remaining tests is zero.
Detection
Discrete except RET can be analyzed in manual mode. Open the Fluorocell RET cover
Action Message
and install new Fluorocell RET.
Type A / Type B
Analyzer
Applicable (Occurring only with the configuration of PLT-F analysis)
Sampler NA
Analyzable when only CBC, DIFF, RET and WPC are specified
Important Notice
as discrete and also with the manual mode
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the number of remaining tests recorded in RFID Tag of Fluorocell PLT during
Detection aspirating Fluorocell PLT.
Conditions of Error
The number of remaining tests is zero.
Detection
Discrete except PLT-F can be analyzed in manual mode. Open the Fluorocell PLT cover
Action Message
and install new Fluorocell PLT.
Purpose of Error
Prompt the user to exchange the reagent.
Detection
Method of Error Check Air Bubbles Sensor of CELLPACK DCL Aspiration Line (DCL) and the double float
Detection switch of CELLPACK DCL reserve tank during aspirating CELLPACK DCL.
Action Message Press [Execute] key. The reagent replacement screen will appear.
Sampler NA
Purpose of Error
Prompt the user to exchange the reagent.
Detection
Method of Error Check Air Bubbles Sensor of CELLPACK DCL Aspiration Line (DCL) and the double float
Detection switch of CELLPACK DCL reserve tank during aspirating CELLPACK DCL.
Action Message Press [Execute] key. The reagent replacement screen will appear.
Purpose of Error
Prompt the user to exchange the reagent.
Detection
Method of Error Check the float switch of SULFOYLYSER reserve tank and the cycle count of
Detection SULFOYLYSER diaphragm pump operations during aspirating SULFOLYSER.
Conditions of Error The SLS chamber float switch is in the lower position and the number of SULFOLYSER
Detection diaphragm pump operations is lower than 60 after the Float Switch goes down.
Action Message Press [Execute] key. The reagent replacement screen will appear.
Purpose of Error
Prompt the user to exchange the reagent.
Detection
Method of Error Check the Float Switch of WNR chamber and the cycle count of Lysercell WNR
Detection diaphragm pump operations during aspirating Lysercell WNR.
Conditions of Error The Float Switch is in the lower position and the number of Lysercell WNR diaphragm
Detection pump operations is lower than 175 after the float switch goes down.
Action Message Press [Execute] key. The reagent replacement screen will appear.
Purpose of Error
Prompt the user to exchange the reagent.
Detection
Method of Error Check the Float Switch of WDF chamber and the cycle count of Lysercell WDF diaphragm
Detection pump operations during aspirating Lysercell WNR.
Conditions of Error The Float Switch is in the lower position and the number of Lysercell WDF diaphragm
Detection pump operations is lower than 175 after the float switch goes down.
Action Message Press [Execute] key. The reagent replacement screen will appear.
Purpose of Error
Prompt the user to exchange the reagent.
Detection
Method of Error Check the Float Switch of WPC chamber and the cycle count of Lysercell WPC diaphragm
Detection pump operations during aspirating Lysercell WPC.
Conditions of Error The Float Switch is in the lower position and the number of Lysercell WPC diaphragm
Detection pump operations is lower than 175 after the float switch goes down.
Action Message Press [Execute] key. The reagent replacement screen will appear.
Purpose of Error
Prompt the user to exchange the reagent.
Detection
Method of Error Check the Float Switch of CELLPACK DFL chamber and the cycle count of CELLPACK
Detection DFL diaphragm pump operations during aspirating CELLPACK DFL.
Conditions of Error The Float Switch is in the lower position and the number of CELLPACK DFL diaphragm
Detection pump operations is lower than 60 after the float switch goes down.
Action Message Press [Execute] key. The reagent replacement screen will appear.
Sampler NA
Purpose of Error
Prevent from in flowing air bubbles into the reagent chamber.
Detection
Method of Error
Check the Air Bubble Sensor of CELLSHEATH(C) Aspiration Line during aspirating c.
Detection
Press [Execute]. The reagent replacement screen will appear. About 8 minutes will be
Action Message
required for CELLSHEATH(C) reagent replacement.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the Float Switch of SULFOLYSER chamber and the cycle count of SULFOLYSER
Detection diaphragm pump operations during aspirating SULFOLYSER.
Action Message Press [Execute]. The reagent replacement screen will appear.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the Float Switch of Lysercell WNR chamber and the cycle count of Lysercell WNR
Detection diaphragm pump operations during aspirating Lysercell WNR.
Action Message Press [Execute]. The reagent replacement screen will appear.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the Float Switch of Lysercell WDF chamber and the cycle count of Lysercell WDF
Detection diaphragm pump operations during aspirating Lysercell WDF.
Action Message Press [Execute]. The reagent replacement screen will appear.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the Float Switch of Lysercell WPC chamber and the cycle count of Lysercell WPC
Detection diaphragm pump operations during aspirating Lysercell WPC.
Action Message Press [Execute]. The reagent replacement screen will appear.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the Float Switch of CELLPACK DFL chamber and the cycle count of CELLPACK
Detection DFL diaphragm pump operations during aspirating CELLPACK DFL.
Action Message Press [Execute]. The reagent replacement screen will appear.
Purpose of Error
To ensure the necessary reagent volume from RU
Detection
Method of Error
Check the reagent remaining volume information from RU.
Detection
Conditions of Error When diluted CELLPACK DST in RU is empty. (Error will not occur when “Ignore RU-20
Detection error” is selected in service setting/
Purpose of Error
To ensure the necessary reagent volume from RU
Detection
Method of Error
Check the reagent information from RU.
Detection
Conditions of Error When reagetnt supply from RU is impossible. (Error will not occur when “Ignore RU-20
Detection error” is selected in service setting/
The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the float-switch in the Reagent Chamber in the Analyzer right after the reagent
Detection replenishing during analyzing.
Conditions of Error
When the float switch is in the lower position.
Detection
Action Message Check the reagent level and press Execute key. Reagent replenishment will be executed.
The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the float-switch in the FCM Sheath Chamber in the analyzer right after the reagent
Detection replenishing during analyzing.
Conditions of Error
When the float switch is in the lower position.
Detection
Action Message Check the reagent level and press [Execute] key. Reagent replacement will be executed.
The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the float-switch in the RBC Sheath Chamber in the analyzer right after the reagent
Detection replenishing during analyzing.
Conditions of Error
When the float switch is in the lower position.
Detection
Action Message Check the reagent level and press [Execute] key. Reagent replacement will be executed.
The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the Air Bubble Sensor of Fluorocell WNR Aspiration Line(WNR-2) and Cover
Detection Sensor of Fluorocell WNR during aspirating Fluorocell WNR.
Action Message Check if sufficient reagent remains and press [Execute] key. Replace the reagent.
The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the Air Bubbles Sensor of the Fluorocell WDF Reagent Line(WDF-1) and the Cover
Detection Sensor of the Fluorocell WDF during aspirating Fluorocell WDF.
Discrete except DIFF can be analyzed in manual mode. Check the reagent level and
Action Message
press [Execute] key. Reagent replacement will be executed.
Analyzer Type A
Applicable
Sampler NA
The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the Air Bubbles Sensor of the Fluorocell WPC Reagent Line(WPC-1) and the
Detection Cover Sensor of the Fluorocell WPC during aspirating Fluorocell WPC.
Discrete except WPC can be analyzed in manual mode. Check the reagent level and
Action Message
press [Execute] key. Reagent replacement will be executed.
Sampler NA
The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the Air Bubble Sensor of the Fluorocell RET Reagent Line(RET) and the Cover
Detection Sensor of the Fluorocell RET during aspirating the Fluorocell RET.
Discrete except RET can be analyzed in manual mode. Check the reagent level and press
Action Message
[Execute] key. Reagent replacement will be executed.
Sampler NA
The reagent in the reagent container cannot be replenished to the instrument with
Meaning
reagent.
Purpose of Error
To ensure the necessary reagent volume for analysis.
Detection
Method of Error Check the Air Bubble Sensor of the Fluorocell PLT Reagent Line(PLT) and the Cover
Detection Sensor of the Fluorocell RET during aspirating the Fluorocell PLT.
Discrete except PLT-F can be analyzed in manual mode. Check the reagent level and
Action Message
press [Execute] key. Reagent replacement will be executed.
Purpose of Error
To ensure that the fluid used in the analysis has been drained.
Detection
Method of Error
Check the float switch position of the waste chamber 1 after the draining is completed.
Detection
Conditions of Error
When the Float Switch is in upper position.
Detection
Check the drain tube and press [Execute] key. Execute the waste chamber drain
Action Message
sequence.
Purpose of Error
To ensure that the fluid used on the analysis has been drained.
Detection
Method of Error
Check the Float Switch position of the waste chamber 2 after the draining is completed.
Detection
Conditions of Error
When the float switch is in lower position.
Detection
Check the drain tube and press [Execute] key. Execute the waste chamber drain
Action Message
sequence.
Meaning Draining from the RBC/HGB Chamber does not operate properly.
Purpose of Error
To ensure draining from the RBC/HGB chamber.
Detection
Check the blank converted values and the following sample converted values when
draining from RBC/HGB chamber.
Method of Error Sample converted value: The converted value of the chamber transmitted light after
Detection draining residual liquid from the RBC/HGB chamber.
Sample converted value: The converted value of the chamber transmitted light after
draining rinse solution from the RBC/HGB chamber.
If any of the following conditions are met, it becomes an error. (When the reaction
chamber drain for maintenance executes, judging with condition 1 only.)
1. Sample converted value 1 < 15000
Conditions of Error
Detection 2. Sample converted value 2 < 15000
3. Sample converted value 1 - Blank Convert value < 12000
4. Sample converted value 2 - Blank Convert value < 12000
Action Message Press [Execute] key. Execute the reaction chamber drain sequence.
Purpose of Error
To ensure draining.
Detection
Method of Error Check the waste tank with the filled sensor (float switch) in ready status and when draining
Detection completed.
In ready status: When the float switch remains in the upper position for continuously 3
seconds
Conditions of Error
Detection Draining complete
d: When the float switch is in upper position.
Action Message Replace the waste container and press [Accept] key.
Purpose of Error
Detect water leak early.
Detection
Method of Error
Check the Water Leak Sensor all the time
Detection
Conditions of Error
When detecting the water leak for continuously 3 seconds
Detection
Analysis cannot be continued because water leak detected. Check the analyzer and turn
Action Message
the main power of the analyzer off.
Purpose of Error
Detect water leak early.
Detection
Method of Error
Check the Water Leak Sensor all the time
Detection
Conditions of Error When “Continue using analyzer when leak is detected” check is OFF and detecting the
Detection water leak for continuously 3 seconds
Action Message Water leak detected in analyzer. Turn off the main power and check the analyzer.
Purpose of Error
To ensure that the water leak sensor works properly.
Detection
Method of Error
Check the water leak sensor error from the driver board always.
Detection
Conditions of Error
When detecting the water leak sensor error for continuously 3 seconds
Detection
Analysis cannot be continued because water leak sns error. Remove the rack and tubes
Action Message
from the analyzer and turn the main power of the analyzer off.
Meaning The motor does not rotate smoothly because the load of WB aspiration pump is too large.
Purpose of Error
To ensure the aspiration / discharge volume of WB samples.
Detection
Check the status of Upper Position Sensor (Photo-Interrupter) when the WB aspiration
Method of Error
motor operation is completed. Only on test operation, check the number of remaining
Detection
pulses.
Conditions of Error
When the upper position sensor (photo-interrupter) does not shade the light.
Detection
Action Message Press [Execute] key. WB aspiration motor test will start.
211000
The motor does not rotate smoothly because the loads of pushing out of samples and a
Meaning
volumetric syringe on the FCM sheath are too large.
Purpose of Error
To ensure the volume of FCM analysis samples.
Detection
Method of Error Check the status of the Photo-Interrupter Home Position and the number of remaining
Detection pulses when sheath motor operation is completed.
(1) When the sheath motor does not shaded the light with the photo-interrupter of the
Conditions of Error home position.
Detection
(2) Remaining pulses > 4800
Action Message Press [Execute] key. Sheath motor test will start.
212010
The motor does not rotate smoothly because the loads of pushing out of samples and a
Meaning
volumetric syringe on the RBC sheath are too large.
Purpose of Error
To ensure the volume of RBC series analysis samples.
Detection
Method of Error Check the status of the Photo-Interrupter Home Position and the number of remaining
Detection pulses when sheath motor operation is completed.
(1) When the sheath motor does not shaded the light with the photo-interrupter of the
Conditions of Error home position.
Detection
(2) Remaining pulses > 4800
Action Message Press Execute key. Sheath motor test will start.
212000
Error Code parameter 1: The number of Judgment Sequence
remaining pulses
The motor does not rotate smoothly because the load of the aspiration unit up-down
Meaning
operation is too large.
Purpose of Error
To ensure the aspiration unit up-down operation.
Detection
Check the status of the Upper Position of the Photo-Interrupter when the aspiration unit
Method of Error
up-down motor ascends. Only when operating with the test mode, check the number of
Detection
remaining pulses while the motor stops.
Conditions of Error When the aspiration unit up-down motor does not shaded the light with the upper position
Detection of the photo-interrupter.
Operation could not be continued because aspiration unit up-down motor stopped.
Action Message Remove rack and tubes from analyzer and press [Execute] key. Analyzer initialize will be
executed.
212020
Error Code parameter 1: The number of Judgment Sequence
remaining pulses
The motor does not rotate smoothly because the load of the aspiration unit left-right
Meaning
operation is too large.
Purpose of Error
To ensure the aspiration unit left-right operation.
Detection
Check the status of the Photo-Interrupter Home Position when the aspiration unit left-right
Method of Error
motor operation is completed. Only when operating with the test mode, check the number
Detection
of remaining pulses when the motor operation completed.
Conditions of Error
When the upper position of the aspiration unit sensor (photo interrupter) is not shaded.
Detection
Action Message Press [Execute] key. Aspiration unit motor test will start.
212021
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Purpose of Error
The sampler analysis stops due to insufficient blood volume.
Detection
Method of Error
Check the blood sensor when checking the tube presence during the sampler analysis.
Detection
Conditions of Error When judging that there are tubes by the Sample Tube Detection Sensor, and also when
Detection the blood volume is insufficient by the Sample Blood Detection Sensor.
Action Message Insufficient blood volume. Check the sample and press [Accept] key.
Meaning The specified volume of blood samples cannot be aspirated during aspirating whole blood.
Purpose of Error
To ensure the WB volume.
Detection
By the blood aspiration sensor, compare between the following conversion value and
blank conversion value while analyzing with WB mode (excluding the background check).
Sample Conversion Value 1:
The value that converted (detected) the turbidity by diluting the piercer tip disposal blood
with EPK before discharging RBC samples
Method of Error
Detection Sample Conversion Value 2:
The value that converted (detected) the turbidity by diluting the whole blood pump
disposal blood with EPK after discharging WBC samples
Sample Conversion Value 3:
The value that converted (detected) the turbidity by diluting the RBC sample blood with
CELLPACK DCL
Insufficient blood volume. Check the sample and press [Accept] key. If the Aspiration
Action Message
Sensor is set to “not used”, this error will not occur.
: 313B037
Meaning The specified volume of blood samples cannot be aspirated during aspirating whole blood.
Purpose of Error
To ensure the WB volume.
Detection
By the blood aspiration sensor, compare between the following conversion value and
blank conversion value during analyzing with WB mode (excluding the blank check).
Method of Error
Detection Sample Conversion Value 2:
The value that converted (detected) the turbidity by diluting the whole blood pump
disposal blood with EPK after discharging WBC samples
Failed to correctly aspirate blood. Check the sample and press [Accept] key. If the
Action Message
Aspiration Sensor is set to “not used”, this error will not occur.
A: 201218
Purpose of Error
To ensure that the blood aspiration sensor operates properly.
Detection
Method of Error By the blood aspiration sensor, compare between the following conversion values and
Detection blank conversion values while analyzing with WB mode (excluding the blank check).
There is a problem in the Aspiration Sensor. Press [Accept]. By changing the Aspiration
Action Message
Sensor setting to “not used”, analysis can be continued.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Purpose of Error
To ensure that the rack on the feed-in rack table sends into the measurement line.
Detection
Method of Error Check the Rack Feed-in Arrival Sensor and the Feed-in Table Rack Detection Sensor
Detection during operating rack feed-in.
When the rack feed-in motor operates a specified number of pulses, the rack is not be
Conditions of Error
detected by the rack feed-in arrival monitoring sensor and the rack is detected by the rack
Detection
feed-in table rack detecting sensor.
Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Purpose of Error To ensure that the rack feed-in unit lever return to the home position after the rack feed-in
Detection operated, and the next rack feed operates properly.
Method of Error Check the Rack Feed-in Home Position Sensor during initializing the rack feed-in
Detection mechanism.
Conditions of Error When the rack feed-in motor operates a specified number of pulses, the rack feed-in
Detection home position sensor is not be shaded.
Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Purpose of Error
To ensure that the rack is moved to the designated location properly.
Detection
Method of Error Compare a specified number of pulses before operation to the Encoder of belt driving
Detection motor after operation while moving the rack.
Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Purpose of Error To ensure that the belt profile of the rack move mechanism returns to the home position
Detection and the next rack operates to move properly.
Method of Error Check the Rack Move Mechanism Home Positio Sensor during initializing the rack move
Detection mechanism.
Conditions of Error When the home position sensor of the rack move mechanism is not shaded as the rack
Detection move motor operated a specified number of pulses.
Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Purpose of Error
To ensure that the rack ejects from the analysis line to the rack ejection table properly.
Detection
Method of Error
Check the Rack Ejection Home Position Sensor during ejecting the rack.
Detection
Conditions of Error When the home position sensor of the rack ejection mechanism is not shaded as the rack
Detection ejection motor operated a specified number of pulses.
Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Purpose of Error To ensure that the lever of the rack ejection mechanism return to the home position and
Detection the rack ejection mechanism is initialized properly.
Method of Error Check the Rack Ejection Home Position Sensor during initializing the rack ejection
Detection mechanism.
Conditions of Error When the home position sensor of the rack ejection mechanism is not shaded when the
Detection rack ejection motor operates a specified number of pulses.
Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Meaning An initialized operation error occurs during the tube presence verification operating.
Purpose of Error To ensure that the tube rotation mechanism returns to the home position, and rack move
Detection operates properly.
Method of Error Check the Tube Presense Verification Home Position Sensor during initializing the tube
Detection presence verification mechanism.
When the home position sensor of the tube presence verification mechanism is not
Conditions of Error
shaded after the tube presence verification mechanism motor operates a specified
Detection
number of pulses.
Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Purpose of Error
To ensure that the feed-in table stopper in upright position.
Detection
Method of Error Check the Feed-in Table Stopper Position Sensor when raising the feed-in table in upright
Detection position.
Conditions of Error When the feed-in table stopper position sensor is not shaded as the feed-in table stopper
Detection motor operates a specified number of pulses.
Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Purpose of Error
To ensure that the rack does not break into the analysis line from the ejection table.
Detection
Method of Error
Check the Ejection Table Stopper Position Sensor after the rack ejection is completed.
Detection
Conditions of Error
When the ejection table stopper position monitoring sensor is shaded.
Detection
Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press [Accept] key.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Meaning The rack exists on the analysis line during operating the analysis line initialization.
Purpose of Error
To ensure that the analysis line initialization operates properly.
Detection
Method of Error Compare the value of a specified number of pulses for operation to the read value by the
Detection encoder on the motor during initializing of the analysis line.
Sampler analysis was stopped because sampler operation error occurred. After checking
Action Message for foreign objects in sampler, remove the rack and tubes from the analyzer and analysis
line and press Accept key.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Meaning The rack is not placed on the rack feed-in table during starting the sampler analysis.
Purpose of Error
Prompt the user to set the rack.
Detection
Method of Error
Check the Feed-in Table Rack Detection Sensor.
Detection
Conditions of Error When the rack is not detected by the rack feed-in table rack detection sensor during
Detection mechanism initialization completion after the sampler analysis starts.
Action Message Press [Accept] key, place rack on feed-in table again, and start sampler analysis.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Purpose of Error
Prompt the user to remove racks from the rack ejection table.
Detection
Method of Error
Check the Ejection Table Rack Full Monitoring Sensor.
Detection
Conditions of Error When the racks are detected by the rack ejection table rack full sensor during starting the
Detection rack ejection operation.
None.
HELP [Execute]
The error clears when the racks were not detected by the rack ejection table rack full
Process
sensor.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Method of Error
None
Detection
Conditions of Error
When the sampler analysis stop operation is completed.
Detection
Completed sampler analysis stop. Press Accept key. Will return to sampler analysis ready
Action Message
state.
Purpose of Error
To ensure that the hand operates at the correct position.
Detection
Method of Error
Check a Hand Position Sensor in a specified timing during sampler analysis.
Detection
It will become an error when the hand position sensor is not shaded in the following timing.
1. Catching a tube from a rack : Hand lower sensor
2. Before mixing operation : Hand upper sensor
Conditions of Error 3. Releasing a tube to a tube holder : Hand middle sensor
Detection 4. Drawing up a tube into an aspiration unit: Hand upper sensor
5. Catching a tube from a tube holder: Hand middle sensor
6. Drawing up a tube : Hand upper sensor
7. Releasing a tube to a rack: Hand lower sensor
Remove the tubes from the analyzer and press [Execute] key. A hand test will be
Action Message
executed.
263004
Parameter 1: remaining
Error Code number of pulses Judgment Sequence
Parameter 2: sensor position
1:lower 2:middle 3:Upper
Detailed Information
HELP Screen Key [Execute]
Purpose of Error
To ensure that catch and release tubes operation works properly.
Detection
Method of Error
Check a Hand Open Sensor in a specified timing during sampler analysis.
Detection
It will become an error when the hand status does not meet the following condition.
1. Before catching a tube from a rack: Hand opened
Conditions of Error 2. After catching a tube from a rack: Hand closed
Detection 3. After releasing the tube to a tube holder : Hand opened
4. After catching the tube from the tube holder: Hand closed
5. After releasing the tube to the rack: Hand middle sensor
Remove the tubes from the analyzer and press [Execute] key. A hand test will be
Action Message
executed.
Purpose of Error
To ensure that the mixing mechanism operates properly.
Detection
Method of Error Check the Mixing Motor Home Position Sensor and the Mixing Tube Detection Sensor in a
Detection specified timing during sampler analysis.
Action Message Remove the tubes from the analyzer and press Execute key. A hand test will be executed.
263005
Parameter 2: sensor position
1: Sensor 29
Error Code 2: Sensor 30 Judgment Sequence
3: Sensor 30
Sensor 29: Mixing motor
Detailed Information home position
Sensor 30: Mixing tube
detection
2015025
Meaning Sample tubes are not placed to a tube holder at the sampler analysis operation.
Purpose of Error
To ensure that tubes are placed to a tube holder correctly.
Detection
Method of Error Check the Sample Tube Detection Sensor in Main Unit when the tube is picked up into the
Detection tube holder.
Conditions of Error
When tubes are not detected.
Detection
Action Message Remove tubes from analyzer and press [Execute] key. A hand test will be executed.
Meaning A tube holder is not placed on the correct position during aspirating the samples
Purpose of Error
To ensure that the WB samples can be aspirated. Avoid the risk of injury to the user.
Detection
Method of Error
Check the Tube Holder Home Position Sensor when picking up the tube holder.
Detection
[Sampler Analysis]
It is an error when matching either of the following conditions:
1.The home position sensor is not shaded when a tube is caught from a rack.
Conditions of Error 2.The home position sensor is shaded when the tube is released to tube holder.
Detection 3.The home position sensor is not shaded when the tube is retracted into the aspiration
unit.
4.The home position sensor is shaded when the tube is caught from the tube holder.
5.The home position sensor is not shaded when the tube is picked up.
Remove tubes from analyzer and press [Execute] key. Tube holder motor test will be
Action Message
executed.
264002
Error Code Parameter 1: remaining Judgment Sequence
number of pulses
Detailed Information HELP Screen Key [Execute]
A tube can't be pulled out from a tube holder when the tube is returned to a rack at
Meaning
the sampler analysis.
Purpose of
To ensure that the tube is picked up from the tube holder properly.
Error Detection
Method of Check the Sample Tube Detection Sensor in Main Unit when the tube holder is is
Error Detection retracted to the Main Unit at the tube return operation.
Conditions of
When the tube is detected.
Error Detection
Action Remove tubes from analyzer and press [Execute] key. A hand test will be
Message executed.
HELP
Execute hand test.
[Execute]
The error clears when the hand test is completed properly.
Process
Judgme
Error Code 264003 Sequence
nt
Tube is remaining in tube holder when switching from manual analysis to sampler
Meaning
analysis.
Purpose of Error
To ensure that there is no sample tube in a tube holder during the sample analysis.
Detection
Method of Error Check the Sample Tube Detection Sensor in Main Unit during switching to Manual mode
Detection or Sampler mode.
Conditions of Error
When detecting a tube.
Detection
Action Message Remove tube from holder and press [Accept] key.
Two tubes have been placed in set-positions of both a normal sample tube and a micro
Meaning
blood tube of a tube holder during manual analysis.
Purpose of Error
To prevent an operator from picking a wrong sample (analysis result).
Detection
Method of Error Check the Sample Tube Detection Sensor in Main Unit and Micro Collection Tube
Detection Detection Sensor during manual analysis.
Conditions of Error
When detecting both of a normal sample tube and a micro collection tube.
Detection
Action Message Place only tube to be analyzed in tube holder and press [Accept] key.
Sample remixing is necessary because reply information for analysis from host computer
Meaning
is delayed on manual analysis.
Purpose of Error
To prevent insufficient sample mixing.
Detection
Method of Error Check the waiting time after transmitting inquiry information to host computer on manual
Detection analysis.
Conditions of Error
When waiting time is over 60 seconds.
Detection
201435
6.5.8.1 Failed to read sample number during sampler analysis (Sampler analysis).
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Meaning The barcode reader failed to read a sample barcode label on a sample test tube.
Purpose of Error
To ensure using sample barcode.
Detection
Method of Error
Check a result of sample number reading during the sampler analysis.
Detection
Conditions of Error
When the sample number was unreadable.
Detection
Error Message Failed to read sample number during sampler analysis(Sampler analysis).
Sampler failed to read sample number during sampler analysis.Check tube barcode label
Action Message
and press [Accept] key.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Meaning The barcode reader failed to read rack barcode label during sampler analysis.
Purpose of Error
To ensure using rack barcode.
Detection
Method of Error
Check a result of rack number reading during the sampler analysis.
Detection
Conditions of Error
When the rack number was unreadable.
Detection
Action Message Check rack barcode label and press [Accept] key.
Meaning The barcode reader failed to read a sample barcode label on a sample test tube.
Purpose of Error
To ensure using sample barcode.
Detection
Method of Error
Check a result of sample number reading during the manual analysis.
Detection
Conditions of Error
When the sample number was unreadable.
Detection
Action Message Check tube barcode label and press [Accept] key.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
An incorrect sample test tube was taken out instead of a correct sample test tube to the
Meaning
analyzer when taking a sample test tube out to the analyzer from a rack.
Purpose of Error
To prevent from picking other sample incorrectly.
Detection
Method of Error To compare the results between read sample number at the sampler and the read sample
Detection number at the analyzer.
Conditions of Error
When the sample numbers do not match.
Detection
Sampler analysis was stopped because sample numbers read in the sampler and in the
Action Message
analyzer do not match. Check for foreign objects in analyzer and press [Accept] key.
Purpose of Error
To prompt a user to input a sample number.
Detection
Method of Error
Check the sample number when checking the tube type in manual analysis.
Detection
Conditions of Error
A micro blood tube is used and the sample number is unspecified.
Detection
Action Message Sample number not specified. Enter sample number in manual analysis screen.
Purpose of Error
Informing a user that no analysis parameter is specified.
Detection
Method of Error Check the replied analysis parameter at analysis information inquiry with using the built-in
Detection BR on manual analysis.
Conditions of Error
When no analysis parameter is specified.
Detection
Action Message No analysis item. Specify analysis item in manual analysis screen.
Purpose of Error
To prompt users that no parameter is specified for analysis.
Detection
Method of Error Check the replied analysis parameter at analysis information inquiry by using the built-in
Detection BR in manual analysis.
Conditions of Error When invalid analysis parameter is specified. (Including invalid analysis parameter which
Detection is specified by reagent related error)
Action Message Invalid analysis item is specified. Specify analysis item in manual analysis screen.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Invalid analysis samples exist on the analysis line with an analyzer operating sampler
Meaning
analysis.
Purpose of Error
To notify a used that the sample cannot be analyzed.
Detection
Method of Error Check the replied analysis parameter at an analysis information inquiry in sampler
Detection analysis.
Invalid analysis item is specified. Check sample analysis item and press [Accept] key.
Action Message
Sample will be skipped and sampler analysis will continue.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Purpose of Error
Samples cannot be analyzed by current status of the analyzer.
Detection
Method of Error
Check a replied analysis parameter at an analysis information inquiry in sampler mode.
Detection
Sample cannot be analyzed because analyzer is not ready for sample analysis. Change
Action Message
analyzer to sampler mode or stop sampler analysis from menu.
None.
HELP [Execute]
The error clears when the analyzer which can analyze a target sample is ready to analyze
Process
in sampler mode.
Purpose of Error
Check that background values are within the range at the background check.
Detection
Method of Error
Check the background values when background check is completed.
Detection
Meaning A variation of count value exceeds the specified range during counting.
Purpose of Error To monitor uneven concentration of samples, unexpected clogging in transducer and so
Detection on.
Method of Error Check the sum of maximum value, minimum value and sampling value after counting
Detection sampling values at a regular interval during analysis
Meaning A variation of count value exceeds the specified range during counting.
Purpose of Error To monitor uneven concentration of samples, unexpected clogging in transducer and so
Detection on.
Method of Error Check the sum of maximum value, minimum value and sampling value after counting
Detection sampling values at a regular interval during analysis
Conditions of Error
Maximum Value – Minimum Value – 100
Detection
The sum of Sampling Value
(The first and last sampling values are NOT included)
Meaning A variation of count value exceeds the specified range during counting.
Purpose of Error To monitor uneven concentration of samples, unexpected clogging in transducer and so
Detection on.
Method of Error Check the sum of maximum value, minimum value and sampling value after counting
Detection sampling values at a regular interval during analysis
2013125
Meaning A variation of count value exceeds the specified range during counting.
Purpose of Error To monitor uneven concentration of samples, unexpected clogging in transducer and so
Detection on.
Method of Error Check the sum of maximum value, minimum value and sampling value after counting
Detection sampling values at a regular interval during analysis
2013125
Analyzer Type A
Applicable
Sampler NA
Meaning A variation of count value exceeds the specified range during counting.
Purpose of Error To monitor uneven concentration of samples, unexpected clogging in transducer and so
Detection on.
Method of Error Check the sum of maximum value, minimum value and sampling value after counting
Detection sampling values at a regular interval during analysis
2013125
Sampler NA
Meaning A variation of count value exceeds the specified range during counting.
Purpose of Error To monitor uneven concentration of samples, unexpected clogging in transducer and so
Detection on.
Method of Error Check the sum of maximum value, minimum value and sampling value after counting
Detection sampling values at a regular interval during analysis
2013125
Sampler NA
Meaning A variation of count value exceeds the specified range during counting.
Purpose of Error To monitor uneven concentration of samples, unexpected clogging in transducer and so
Detection on.
Method of Error Check the sum of maximum value, minimum value and sampling value after counting
Detection sampling values at a regular interval during analysis
2013125
Method of Error Check the clogging signal by the direct current between both electrodes of the sheath flow
Detection detector during analysis.
It becomes an error when clogging value “C” is C >105 in a timing of the timing chart.
C=3.832 x 10 x D x 3.9/256-2.110 x TTD2 x 10-2 + 1.992 x TTD – 3.660 x 1
Conditions of Error
C: degree of clogging (integer, round off the degree of clogging to integer )
Detection
D: A/D converting value of the clogging voltage
TTD: Environment Temperature (**.* ºC)
Action Message Press [Execute] key. RBC detector clog will be removed.
Meaning Air bubbles have been stuck to the RBC detector (pellet).
Purpose of Error
To ensure the RBC/PLT analysis results.
Detection
Method of Error
Check the output pulses with the air bubble monitoring circuit during the analysis.
Detection
Conditions of Error
When sum of the output pulses is over 128.
Detection
Action Message Press [Execute] key. RBC detector clog will be removed.
The count is too low despite the aspiration error is not detected by blood sample
Meaning
aspiration sensor during the analysis.
Purpose of Error
To detect dilution error of the instrument
Detection
Method of Error
Check the analysis result during analysis.
Detection
When the aspiration error is not detected by the blood sample aspiration sensor, and the
analysis results are met the all following conditions.
WBC (WBC-N) < 10.0 [x10^2/µ]
Conditions of Error
Detection RBC < 30 [x 10^4/µL]
HGB < 1.0 [g/dL]
PLT < 2.0 [x10^4/µL]
Abnormally low count because blood sample not aspirated correctly. Check sample and
Action Message
press [Accept] key.
Meaning A/D converted HGB blank value or HGB sample value is out of range.
Purpose of Error
To ensure the HGB blank value and the HGB sample value.
Detection
Method of Error Check the HGB blank value and the HGB sample value during analysis with normal mode
Detection or PD mode.
Action Message Analysis must be repeated. Check analysis result and press [Accept] key.
Purpose of Error
To ensure the RBC channel analysis data.
Detection
Method of Error
Check the analysis result during analyzing.
Detection
Action Message Analysis must be repeated. Check sample mixing state and press [Accept] key.
Purpose of Error
To ensure the PLT channel analysis data.
Detection
Method of Error
Check the analysis result during analyzing.
Detection
Action Message Analysis must be repeated. Check sample mixing state and press [Accept] key.
Purpose of Error
To ensure the WNR channel analysis data.
Detection
Method of Error
Check the analysis result during analyzing.
Detection
Action Message Analysis must be repeated. Check sample mixing state and press [Accept] key.
Purpose of Error
To ensure the WDF channel analysis data.
Detection
Method of Error
Check the analysis result during analyzing.
Detection
Action Message Analysis must be repeated. Check sample mixing state and press [Accept] key.
Analyzer Type A
Applicable
Sampler NA
Purpose of Error
To ensure the WPC channel analysis data.
Detection
Method of Error
Check the analysis result during analyzing.
Detection
Action Message Analysis must be repeated. Check sample mixing state and press [Accept] key.
Purpose of Error
To ensure the RET channel analysis data.
Detection
Method of Error
Check the analysis result during analyzing.
Detection
Action Message Analysis must be repeated. Check sample mixing state and press [Accept] key.
Sampler NA
Purpose of Error
To ensure the PLT-F channel analysis data.
Detection
Method of Error
Check the analysis result during analyzing.
Detection
Action Message Analysis must be repeated. Check sample mixing state and press [Accept] key.
Meaning The analysis parameters results exceed the upper/lower limit of the specified range.
Purpose of Error Notify the users the analysis parameters exceed the upper/lower limit values range and
Detection suspend the sampler analysis.
Method of Error
To compare the analysis result to the panic limit values.
Detection
Conditions of Error
When the analysis results are out of the panic limit values range.
Detection
Action Message Data error occurred. Check analysis result and press [Accept] key.
A: 201218
On the body fluid analysis, an analysis result might affect to the next sample of body fluid
Meaning
analysis with the carry-over because the analysis result is too high.
Purpose of Error
Prompt users to execute a background check before the next analysis.
Detection
Method of Error
Check an analysis result of the body fluid.
Detection
Because the analysis result is high, the body fluid analysis of the next sample may be
Action Message affected. Remove the sample from the sample holder and press [Execute] key. A
background check will be executed.
Method of Error
Check the Front Cover Sensor during the analysis.
Detection
Conditions of Error
When detecting the open cover for 1 second continuously.
Detection
Analysis cannot be continued because front cover was opened during analysis. Remove
Action Message
rack and tubes from analyzer and press [Execute] key. Analyzer initialize will be executed.
Method of Error
Always check the Front Cover Sensor.(except during analysis)
Detection
Conditions of Error
When detecting the open cover for 1 second continuously.
Detection
Purpose of Error
To ensure the data by blocking light from the outside.
Detection
Method of Error
Check the RBC detector cover sensor during analysis and in ready status.
Detection
Conditions of Error In ready status: When detecting the open cover for 1 second continuously.
Detection During analysis: When detecting the open cover on the timing of the timing chart.
Purpose of Error
To ensure the data with blocking light from the outside.
Detection
Method of Error
Check the FCM Detector Cover Sensor during analysis and in ready Status.
Detection
Conditions of Error In ready status: When detecting the open cover for 1 second continuously.
Detection During analysis: When detecting the open cover on the timing of the timing chart.
Purpose of Error
To ensure the data with blocking light from the outside.
Detection
Method of Error
Check always the Fluorocell WNR Cover Sensor.
Detection
Conditions of Error
When detecting the Fluorocell WNR open cover for 1 second continuously.
Detection
Action Message The Fluorocell WNR cover is open. Close the Fluorocell WNR cover.
Purpose of Error
To ensure the data with blocking light from the outside.
Detection
Method of Error
Check always the Fluorocell WDF Cover Sensor.
Detection
Conditions of Error
When detecting the Fluorocell WDF open cover for 1 second continuously.
Detection
Action Message The Fluorocell WDF cover is open. Close the Fluorocell WDF cover.
Analyzer Type A
Applicable
Sampler NA
Purpose of Error
To ensure the data with blocking light from the outside.
Detection
Method of Error
Check always the Fluorocell WPC Cover Sensor.
Detection
Conditions of Error
When detecting the Fluorocell WPC open cover for 1 second continuously.
Detection
Action Message The Fluorocell WPC cover is open. Close the Fluorocell WPC cover.
Purpose of Error
To ensure the data with blocking light from the outside.
Detection
Method of Error
Check always the Fluorocell RET Cover Sensor.
Detection
Conditions of Error
When detecting the Fluorocell RET open cover for 1 second continuously.
Detection
Action Message The Fluorocell RET cover is open. Close the Fluorocell RET cover.
Purpose of Error
To ensure the data with blocking light from the outside.
Detection
Method of Error
Check always the Fluorocell PLT cover sensor.
Detection
Conditions of Error
When detecting the Fluorocell PLT open cover for 1 second continuously.
Detection
Action Message The Fluorocell PLT cover is open. Close the Fluorocell PLT cover.
Meaning Laser drive current value is increasing and expiration of laser life time is approaching.
Purpose of Error To predict the laser diode life time and let the user to contact Sysmex Service
Detection Representative to replace the laser diode.
Compare the initial laser drive current value which is stored the analyzer to the current
Method of Error
laser drive current value at the laser drive current of the each channel during analyzing
Detection
(timing on the timing chart).
Conditions of Error The ratio of the initial laser drive current value to the current laser drive current value is
Detection more than 1.2 times and less than 2.0 times.
Action Message Laser will soon require replacement. Contact a service technician.
Purpose of Error
To ensure the laser power stability.
Detection
Compare the initial laser drive current value which is stored the analyzer to the current
Method of Error
laser drive current value in the laser drive current of each channel during analysis (timing
Detection
on the timing chart).
When the ratio of the initial laser current value and the current laser drive current value is
Conditions of Error 2.0 times or more.
Detection
When the current laser drive current value is over 150mA.
Analysis cannot be continued because laser replacement is required. Remove the rack
Action Message
and tubes from the analyzer and turn the main power of the analyzer off.
Meaning There is a communication error between the analyzer barcode reader and PCB No.30001.
Purpose of Error
To ensure that the barcode reader reads barcode in the analyzer.
Detection
Method of Error Check the connection between the analyzer barcode reader and PCB No.30001 when
Detection starting the analyzer.
Conditions of Error When a connection between the analyzer barcode reader and the PCB No.30001 cannot
Detection be established.
Turn the main power of the analyzer off and on. If error recurs after restarting analyzer,
Action Message
contact a service technician.
Analyzer NA
Applicable
Sampler Single Model / Twin Model
Meaning There is a communication error between the sampler barcode reader and PCB No.30001.
Purpose of Error
To ensure that the barcode reader reads barcode in the sampler.
Detection
Method of Error Check the connection between the sampler barcode reader and PCB No.30001 when
Detection starting the sampler.
Conditions of Error When a connection between the analyzer barcode reader and the PCB No.30001 cannot
Detection be established
Turn the main power of the sampler off and on. If error recurs after restarting analyzer,
Action Message
contact a service technician.
Analyzer NA
Applicable
Sampler Single Model
Purpose of Error
To ensure the communication with the conveyor unit.
Detection
Method of Error Check the communication connection with the conveyor unit when sending a message to
Detection the conveyor unit.
Conditions of Error
When a communication message cannot be sent to the conveyor unit.
Detection
Analyzer NA
Applicable
Sampler Single Model/Twin Model
Communication between the analyzer and the sampler is disconnected during sampler
Meaning
analysis.
Purpose of Error To ensure the communication between the analyzer and the sampler during sampler
Detection analysis.
Method of Error
Check the analyzer communication status during the sampler analysis.
Detection
Conditions of Error
When the sampler cannot communicate with the analyzer during sampler analysis.
Detection
Sampler analysis was stopped because communication with the analyzer was interrupted
Action Message during sampler analysis. Remove the rack and sample tube from the analyzer and
analysis line and press [Accept] key.
Meaning There is a communication error between RFID reader and RFID writer in the analyzer.
Purpose of Error
To ensure the RFID reading in the analyzer.
Detection
Check the response from the RFID reader / writer with the following timing.
1. Starting an analyzer
Method of Error
2. While covers of each dye solution are opened
Detection
3. When the covers of each dye solution are closed.
4. When each dye solution is aspirated.
Conditions of Error
When there is no reply from the RFID reader / writer.
Detection
Turn the main power of the analyzer off and on. If the error occurs again, contact your
Action Message
Sysmex service representative for assistance.
Meaning There is a communication error between the analyzer and the IPU.
Purpose of Error
To ensure the communication between the analyzer and the IPU.
Detection
Method of Error
Check the communication content when the analyzer communicates with the IPU.
Detection
Conditions of Error
When the communication contents is improperly.
Detection
Analysis cannot be continued because communication with the analyzer was error.
Action Message Remove the rack and tubes from the analyzer and turn the main power of the analyzer off
and on.
Purpose of Error
To detect the analysis error by checking time-to-time change.
Detection
Method of Error Executing a quality control by a statistical method weighted average data of the normal
Detection samples as a QC data during plotting to the QC chart.
Conditions of Error
When the data is out of control limit.
Detection
Purpose of Error
To detect the analysis error by checking day-to-day or time-to-time change.
Detection
Method of Error Executing a quality control by a statistical method analysis result of the control blood
Detection sample as a control data during plotting to the QC chart.
Conditions of Error
When the data is out of control limit.
Detection
Purpose of Error
To detect the analysis error by checking day-to-day change.
Detection
Method of Error Executing a quality control by a statistical method analysis result of the control blood
Detection sample as a control data during plotting to the QC chart.
Conditions of Error
When the data is out of control limit.
Detection
Purpose of Error To ensure the precision of QC, by notifying the users that the control blood expiration data
Detection is expired.
Method of Error
Check the control blood lot information (expiration date) when operating QC.
Detection
Conditions of Error
When the control blood expiration date is expired.
Detection
Action Message Enter QC material of new lot and press [Accept] key.
Purpose of Error
To notify the control blood lot registration status.
Detection
Method of Error
Check the control blood lot information when operating QC.
Detection
Conditions of Error
When there is no lot information.
Detection
Purpose of Error
To prompt the users to execute the QC.
Detection
Method of Error
Check that the QC analysis executed at the time of QC execution time set by users.
Detection
Conditions of Error
When QC is not executed within 1 hour.
Detection
Purpose of Error
To detect an instrument error by chcking time-to-time change.
Detection
Method of Error Executing a quality control of sensitivity parameter in the control blood sample analysis
Detection result as a control data by a statistical method during plotting to the QC chart.
Conditions of Error
When the data is out of control limit.
Detection
Purpose of Error
To detect an instrument error by chcking time-to-time change.
Detection
Method of Error Executing a quality control of sensitivity parameter in the control blood sample analysis
Detection result as a control data by a statistical method during plotting to the QC chart.
Conditions of Error
When the data is out of control limit.
Detection
Analyzer Type A
Applicable
Sampler NA
Purpose of Error
To detect an instrument error by chcking time-to-time change.
Detection
Method of Error Executing a quality control of sensitivity parameter in the control blood sample analysis
Detection result as a control data by a statistical method during plotting to the QC chart.
Conditions of Error
When the data is out of control limit.
Detection
Purpose of Error
To detect an instrument error by chcking time-to-time change.
Detection
Method of Error Executing a quality control of sensitivity parameter in the control blood sample analysis
Detection result as a control data by a statistical method during plotting to the QC chart.
Conditions of Error
When the data is out of control limit.
Detection
Purpose of Error
To detect an instrument error by chcking time-to-time change.
Detection
Method of Error Executing a quality control of sensitivity parameter in the control blood sample analysis
Detection result as a control data by a statistical method during plotting to the QC chart.
Conditions of Error
When the data is out of control limit.
Detection
Purpose of Error
To prompt the users to set a correct control blood.
Detection
Method of Error Check the control blood type when the control blood sample numbers were read by a
Detection barcode reader.
Conditions of Error
When the control blood type is different with the current analysis mode.
Detection
Check the Analysis Mode/control blood type and press [Accept] key. Change Analysis
Action Message
Mode and repeat analysis.
None.
HELP [Execute]
Process The error clears when [Accept] is pressed. Skip a subject sample analysis during sampler
analysis.
Purpose of Error
To prompt the users to replace piercer.
Detection
Method of Error Check the cycle count of the piercer operation before start up the analyzer and during
Detection analysis.
Conditions of Error
Numbers of piercer operation cycle count> 120000 times
Detection
Turn off analyzer power and replace piercer. After replacing piercer, reset piercer count in
Action Message
counter screen of maintenance menu.
Purpose of Error
Prompt the users for cleaning.
Detection
Method of Error Check the cycle count of analysis after the previous cleaning performance during
Detection aspirating sample.
Conditions of Error
Numbers of cycle count after previous cleaning cycle count > 1000 times
Detection
Action Message Regular cleaning is necessary for proper use of the analyzer.
Purpose of Error
Prompt the users to cleaning.
Detection
Method of Error
Check the cycle count of analysis after the previous cleaning during aspirating sample.
Detection
Conditions of Error
Numbers of cycle count after previous cleaning > 1000 times
Detection
Regular cleaning is necessary for proper use of the analyzer. Press [Execute] key. The
Action Message
cleaning screen will appear.
Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection
Compare the current date with the expiration date set in reagent replacement before the
following timing:
Method of Error 1. before background check
Detection
2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
Conditions of Error
When reagent has expired.
Detection
Action Message Press [Execute] key. The reagent replacement screen will appear.
Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection
Compare the current date with the expiration date set in reagent replacement before the
following timing:
Method of Error 1. before background check
Detection
2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
Conditions of Error
When reagent has expired.
Detection
Action Message Press [Execute] key. The reagent replacement screen will appear.
Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection
Compare the current date with the expiration date set in reagent replacement before the
following timing:
Method of Error 1. before background check
Detection
2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
Conditions of Error
When reagent has expired.
Detection
Action Message Press [Execute] key. The reagent replacement screen will appear.
Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection
Compare the current date with the expiration date set in reagent replacement before the
following timing:
1. before background check
Method of Error
Detection 2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
4. When Fluorocell WDR is set.
Conditions of Error
When reagent has expired
Detection
Action Message Open the Fluorocell WNR cover and install new Fluorocell WNR.
Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection
Compare the current date with the expiration date set in reagent replacement before the
following timing:
Method of Error 1. before background check
Detection
2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
Conditions of Error
When reagent has expired.
Detection
Action Message Press [Execute] key. The reagent replacement screen will appear.
Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection
Compare the current date with the expiration date set in reagent replacement before the
following timing:
1. before background check
Method of Error
Detection 2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
4. when FRUOROCELL WDF is set
Conditions of Error
When reagent has expired.
Detection
Action Message Open the Fluorocell WDF cover and install new Fluorocell WDF.
Analyzer Type A
Applicable
Sampler NA
Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection
Compare the current date with the expiration date set in reagent replacement before the
following timing:
Method of Error 1. before background check
Detection
2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
Conditions of Error
When reagent has expired.
Detection
Action Message Press [Execute] key. The reagent replacement screen will appear.
Analyzer Type A
Applicable
Sampler NA
Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection
Compare the current date with the expiration date set in reagent replacement before the
following timing:
1. before background check
Method of Error
Detection 2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
4. when Fluorocell WPC is set
Conditions of Error
When reagent has expired.
Detection
Action Message Open the Fluorocell WPC cover and install new Fluorocell WPC.
Analyzer Type A
Applicable
Sampler NA
Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection
Compare the current date with the expiration date set in reagent replacement before the
following timing:
Method of Error 1. before background check
Detection
2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
Conditions of Error
When reagent has expired.
Detection
Action Message Press [Execute] key. The reagent replacement screen will appear.
Sampler NA
Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection
Compare the current date with the expiration date set in reagent replacement before the
following timing:
1. before background check
Method of Error
Detection 2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
4. when Fluorocell RET is set
Conditions of Error
When reagent has expired.
Detection
Action Message Open the Fluorocell RET cover and install new Fluorocell RET.
Sampler NA
Purpose of Error
To prevent to exceed the reagent expiration date as a reagent control function.
Detection
Compare the current date with the expiration date set in reagent replacement before the
following timing:
1. before background check
Method of Error
Detection 2. before sample aspiration in manual analysis
3. before collecting the sample tube during sampler analysis
4. when Fluorocell PLT is set
Conditions of Error
When reagent has expired.
Detection
Action Message Open the Fluorocell PLT cover and install new Fluorocell PLT.
Purpose of Error
To prompt resume the analyzer to ready status before an analysis.
Detection
Method of Error Check the elapsed time after the analyzer is in the manual ready status or the sampler
Detection ready status.
Conditions of Error
5 hours elapsed.
Detection
Purpose of Error To prevent installing an incorrect dye, or notify that the incorrect dye is installed for
Detection preventing analyzing with the incorrect reagent.
Reading the cartridge RFID tag placed in the Fluorocell WNR holder in the following
timing, and then checking the read reagent code.
Method of Error 1. When starting the analyzer.
Detection
2. When dye cartridge is installed in the Fluorocell WNR holder.
3. When closing the Fluorocell WNR cover.
Conditions of Error When reading reagent code of the RFID tag is different from the reagent code which
Detection should be placed in the relevant dye holder.
Purpose of Error To prevent installing an incorrect dye, or notify that the incorrect dye is installed for
Detection preventing analyzing with the incorrect reagent.
Reading the cartridge RFID tag placed in the Fluorocell WDF holder in the following
timing, and then checking the read reagent code.
Method of Error 1. When starting the analyzer.
Detection
2. When dye cartridge is installed in the Fluorocell WDF holder.
3. When closing the Fluorocell WDF cover.
Conditions of Error When reading reagent code of the RFID tag is different from the reagent code which
Detection should be placed in the relevant dye holder.
Purpose of Error To prevent installing an incorrect dye, or notify that the incorrect dye is installed for
Detection preventing analyzing with the incorrect reagent.
Reading the cartridge RFID tag placed in the Fluorocell WPC holder in the following
timing, and then checking the read reagent code.
Method of Error 1. When starting the analyzer.
Detection
2. When dye cartridge is installed in the Fluorocell WPC holder.
3. When closing the Fluorocell WPC cover.
Conditions of Error When reading reagent code of the RFID tag is different from the reagent code which
Detection should be placed in the relevant dye holder.
Sampler NA
Purpose of Error To prevent installing an incorrect dye, or notify that the incorrect dye is installed for
Detection preventing analyzing with the incorrect reagent.
Reading the cartridge RFID tag placed in the Fluorocell RET holder in the following timing,
and then checking the read reagent code.
Method of Error 1. When starting the analyzer.
Detection
2. When dye cartridge is installed in the Fluorocell RET holder.
3. When closing the Fluorocell RET cover.
Conditions of Error When reading reagent code of the RFID tag is different from the reagent code which
Detection should be placed in the relevant dye holder.
Sampler NA
Purpose of Error To prevent installing an incorrect dye, or notify that the incorrect dye is installed for
Detection preventing analyzing with the incorrect reagent.
Reading the cartridge RFID tag placed in the Fluorocell PLT holder in the following timing,
and then checking the read reagent code.
Method of Error 1. When starting the analyzer.
Detection
2. When dye cartridge is installed in the Fluorocell PLT holder.
3. When closing the Fluorocell PLT cover.
Conditions of Error When reading reagent code of the RFID tag is different from the reagent code which
Detection should be placed in the relevant dye holder.
Purpose of Error
Ensure the necessary volume of reagent for analysis.
Detection
Reading the cartridge RFID tag placed in the Fluorocell WNR holder in the following
timing, then checking a response from the RFID tag.
Method of Error
Detection 1. When starting the analyzer
2. When closing the Fluorocell WNR cover.
Conditions of Error
No response from RFID tag
Detection
Action Message Open the Fluorocell WNR cover and install Fluorocell WNR.
Purpose of Error
Ensure the necessary volume of reagent for analysis.
Detection
Reading the cartridge RFID tag placed in the Fluorocell WDF holder in the following
timing, then checking a response from the RFID tag.
Method of Error
Detection 1. When starting the analyzer
2. When closing the Fluorocell WDF cover.
Conditions of Error
No response from RFID tag
Detection
Action Message Open the Fluorocell WDF cover and install Fluorocell WDF.
Analyzer Type A
Applicable
Sampler NA
Purpose of Error
Ensure the necessary volume of reagent for analysis.
Detection
Reading the cartridge RFID tag placed in the Fluorocell WPC holder in the following
timing, then checking a response from the RFID tag.
Method of Error
Detection 1. When starting the analyzer
2. When closing the Fluorocell WPC cover.
Conditions of Error
No response from RFID tag
Detection
Action Message Open the Fluorocell WPC cover and install Fluorocell WPC.
Sampler NA
Purpose of Error
Ensure the necessary volume of reagent for analysis.
Detection
Reading the cartridge RFID tag placed in the Fluorocell RET holder in the following timing,
then checking a response from the RFID tag.
Method of Error
Detection 1. When starting the analyzer
2. When closing the Fluorocell RET cover.
Conditions of Error
No response from RFID tag
Detection
Action Message Open the Fluorocell RET cover and install Fluorocell RET.
Sampler NA
Purpose of Error
Ensure the necessary volume of reagent for analysis.
Detection
Reading the cartridge RFID tag placed in the Fluorocell PLT holder in the following timing,
then checking a response from the RFID tag.
Method of Error
Detection 1. When starting the analyzer
2. When closing the Fluorocell PLT cover.
Conditions of Error
No response from RFID tag
Detection
Action Message Open the Fluorocell PLT cover and install Fluorocell PLT.
Purpose of Error Preventing deteriorating a dye from refilling dye into dye cartridge and preventing
Detection analyzing with deteriorated dye.
Reading the RFID tag of a cartridge installed in the Fluorocell WNR holder in the following
timing, and comparing the usage status stored in the RFID tag with number of remaining
Method of Error tests.
Detection
1. When starting the analyzer.
2. When closing the FRUOLOCELL WNR cover.
It will become an error when the following condition formulas are met:
CQC+{(IQC/10) x ROB} > IQC
Conditions of Error
1. The number of remaining tests stored in RFID : CQC (Current Quantity Count)
Detection
2. The number of the total tests stored in RFID: IQC (Initial Quantity Control)
3. The number of the Read Only Blocks in use status check block: ROB
Action Message Install new Fluorocell WNR in the Fluorocell WNR holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell WNR with the correct number of use status check-
blocks is installed in the Fluorocell WNR holder and closing the dye cover.
Purpose of Error Preventing deteriorating a dye from refilling dye into dye cartridge and preventing
Detection analyzing with deteriorated dye.
Reading the RFID tag of a cartridge installed in the Fluorocell WNR holder in the following
timing, and comparing the usage status stored in the RFID tag with number of remaining
Method of Error tests.
Detection
1. When starting the analyzer.
2. When closing the FRUOLOCELL WDF cover.
It will become an error when the following condition formulas are met:
CQC+{(IQC/10) x ROB}>IQC
Conditions of Error
1. The number of remaining tests stored in RFID : CQC (Current Quantity Count)
Detection
2. The number of the total tests stored in RFID: IQC (Initial Quantity Control)
3. The number of the Read Only Blocks in use status check block: ROB
Discrete except DIFF can be analyzed in manual mode. Install new Fluorocell WDF in the
Action Message
Fluorocell WDF holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell WDF with the correct number of use status check-
blocks is installed in the Fluorocell WDF holder and closing the dye cover.
Analyzer Type A
Applicable
Sampler NA
Analyzable when only CBC, DIFF, RET, and PLT-F are specified
Important Notice
as discrete and also with manual mode.
Purpose of Error Preventing deteriorating a dye from refilling dye into dye cartridge and preventing
Detection analyzing with deteriorated dye.
Reading the RFID tag of a cartridge installed in the Fluorocell WPC holder in the following
timing, and comparing the usage status stored in the RFID tag with number of remaining
Method of Error tests.
Detection
1. When starting the analyzer.
2. When closing the FRUOLOCELL WPC cover.
It will become an error when the following condition formulas are met:
CQC+{(IQC/10) x ROB}>IQC
Conditions of Error
1. The number of remaining tests stored in RFID : CQC (Current Quantity Count)
Detection
2. The number of the total tests stored in RFID: IQC (Initial Quantity Control)
3. The number of the Read Only Blocks in use status check block: ROB
Discrete except WPC can be analyzed in manual mode. Install new Fluorocell WPC in the
Action Message
Fluorocell WPC holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell WPC with the correct number of use status check-
blocks is installed in the Fluorocell WPC holder and closing the dye cover.
Sampler NA
Analyzable when only CBC, DIFF, PLT-F, and WPC are specified
Important Notice
as discrete and also with manual mode.
Purpose of Error Preventing deteriorating a dye from refilling dye into dye cartridge and preventing
Detection analyzing with deteriorated dye.
Reading the RFID tag of a cartridge installed in the Fluorocell RET holder in the following
timing, and comparing the usage status stored in the RFID tag with number of remaining
Method of Error tests.
Detection
1. When starting the analyzer.
2. When closing the FRUOLOCELL RET cover.
It will become an error when the following condition formulas are met:
CQC+{(IQC/10) x ROB}>IQC
Conditions of Error
1. The number of remaining tests stored in RFID : CQC (Current Quantity Count)
Detection
2. The number of the total tests stored in RFID: IQC (Initial Quantity Control)
3. The number of the Read Only Blocks in use status check block: ROB
Discrete except RET can be analyzed in manual mode. Install new Fluorocell RET in the
Action Message
Fluorocell RET holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell RET with the correct number of use status check-
blocks is installed in the Fluorocell RET holder and closing the dye cover.
Sampler NA
Analyzable when only CBC, DIFF, RET, and WPC are specified
Important Notice
as discrete and also with manual mode.
Purpose of Error Preventing deteriorating a dye from refilling dye into dye cartridge and preventing
Detection analyzing with deteriorated dye.
Reading the RFID tag of a cartridge installed in the Fluorocell PLT holder in the following
timing, and comparing the usage status stored in the RFID tag with number of remaining
Method of Error tests.
Detection
1. When starting the analyzer.
2. When closing the FRUOLOCELL PLT cover.
It will become an error when the following condition formulas are met:
CQC+{(IQC/10) x ROB}>IQC
Conditions of Error
1. The number of remaining tests stored in RFID : CQC (Current Quantity Count)
Detection
2. The number of the total tests stored in RFID: IQC (Initial Quantity Control)
3. The number of the Read Only Blocks in use status check block: ROB
Discrete except PLT-F can be analyzed in manual mode. Install new Fluorocell PLT in the
Action Message
Fluorocell PLT holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell RET with the correct number of use status check-
blocks is installed in the Fluorocell RET holder and closing the dye cover.
Read the cartridge RFID tag placed in the Fluorocell WNR holder in the following timing,
and compare the read UID from RFID tag with the stored encrypted UID in the tag.
Method of Error
Detection 1. When starting the analysis
2. When closing the Fluorocell WNR cover.
Conditions of Error
The UID from RFID tag and the stored encrypted UID do not match.
Detection
Action Message Install normal Fluorocell WNR in the Fluorocell WNR holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell WNR with the correct UID is installed in the Fluorocell
WNR holder.
Read the cartridge RFID tag placed in the Fluorocell WDF holder in the following timing,
and compare the read UID from RFID tag withthe stored encrypted UID in the tag.
Method of Error
Detection 1. When starting the analysis
2. When closing the Fluorocell WDF cover.
Conditions of Error
The UID from RFID tag and the stored encrypted UID do not match.
Detection
Action Message Install normal Fluorocell WDF in the Fluorocell WDF holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell WDF with the correct UID is installed in the Fluorocell
WDF holder.
Analyzer Type A
Applicable
Sampler NA
Read the cartridge RFID tag placed in the Fluorocell WPC holder in the following timing,
and compare the read UID from RFID tag withthe stored encrypted UID in the tag.
Method of Error
Detection 1. When starting the analysis
2. When closing the Fluorocell WPC cover.
Conditions of Error
The UID from RFID tag and the stored encrypted UID do not match.
Detection
Action Message Install normal Fluorocell WPC in the Fluorocell WPC holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell WDF with the correct UID is installed in the Fluorocell
WPC holder.
Sampler NA
Read the cartridge RFID tag placed in the Fluorocell RET holder in the following timing,
and compare the read UID from RFID tag withthe stored encrypted UID in the tag.
Method of Error
Detection 1. When starting the analysis
2. When closing the Fluorocell RET cover.
Conditions of Error
The UID from RFID tag and the stored encrypted UID do not match.
Detection
Action Message Install normal Fluorocell RET in the Fluorocell RET holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell RET with the correct UID is installed in the Fluorocell
RET holder.
Sampler NA
Read the cartridge RFID tag placed in the Fluorocell PLT holder in the following timing,
and compare the read UID from RFID tag withthe stored encrypted UID in the tag.
Method of Error
Detection 1. When starting the analysis
2. When closing the Fluorocell PLT cover.
Conditions of Error
The UID from RFID tag and the stored encrypted UID do not match.
Detection
Action Message Install normal Fluorocell PLT in the Fluorocell PLT holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell PLT with the correct UID is installed in the Fluorocell
PLT holder.
Meaning The remaining volume of dye cannot be written in the RFID tag.
Method of Error
Check the response from the RFID tag when operating writing-access to RFID
Detection
Conditions of Error
No response from the RFID tag.
Detection
Action Message Install Fluorocell WNR with normal RFID added in the Fluorocell WNR holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell WNR with correct RFID is installed in the Fluorocell
WNR holder.
Meaning The remaining volume of dye cannot be written in the RFID tag.
Method of Error
Check the response from the RFID tag when operating writing-access to RFID
Detection
Conditions of Error
No response from the RFID tag.
Detection
Discrete except DIFF can be analyzed in manual mode. Install Fluorocell WDF with
Action Message
normal RFID added in the Fluorocell WDF holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell WDF with correct RFID is installed in the Fluorocell
WDF holder.
Analyzer Type A
Applicable
Sampler NA
Analyzable when only CBC, DIFF, RET, and PLT-F are specified
Important Notice
as discrete and also with the manual mode
Meaning The remaining volume of dye cannot be written in the RFID tag.
Method of Error
Check the response from the RFID tag when operating writing-access to RFID
Detection
Conditions of Error
No response from the RFID tag.
Detection
Discrete except WPC can be analyzed in manual mode. Install Fluorocell WPC with
Action Message
normal RFID added in the Fluorocell WPC holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell WPC with correct RFID is installed in the Fluorocell
WPC holder.
Sampler NA
Analyzable when only CBC, DIFF, PLT-F, and WPC are specified
Important Notice
as discrete and also with the manual mode
Meaning The remaining volume of dye cannot be written in the RFID tag.
Method of Error
Check the response from the RFID tag when operating writing-access to RFID
Detection
Conditions of Error
No response from the RFID tag.
Detection
Discrete except RET can be analyzed in manual mode. Install Fluorocell RET with normal
Action Message
RFID added in the Fluorocell RET holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell RET with correct RFID is installed in the Fluorocell
RET holder.
Sampler NA
Analyzable when only CBC, DIFF, RET, and WPC are specified
Important Notice
as discrete and also with the manual mode
Meaning The remaining volume of dye cannot be written in the RFID tag.
Method of Error
Check the response from the RFID tag when operating writing-access to RFID
Detection
Conditions of Error
No response from the RFID tag.
Detection
Discrete except PLT-F can be analyzed in manual mode. Install Fluorocell PLT with
Action Message
normal RFID added in the Fluorocell PLT holder.
None.
HELP [Execute]
Process The error clears when the Fluorocell PLT with correct RFID is installed in the Fluorocell
PLT holder.
Description
Sensor No.
(SA-10/SA-20)
S1 Not Used
S2 Not Used
S4 Rack full
S7 Blood monitoring
Right sampler
pool return
Analysis line prevention
Left sampler pool
(Rack horizontal-feed unit) stopper
return prevention
stopper
6HQVRU Right sampler pool
Left sampler pool Left sampler pool (Rack feed-in table)
rack full sensor
6HQVRU
201362
7 FEEDER_ASSY NO.85 Rack conveyor belt (rear) home position KB1242-AA14LF 22837068
8 FEEDER_ASSY NO.85 Rack conveyor belt (rear) home position KB1240-AA23LF 22837054
9 FEEDER_ASSY NO.85 Rack conveyor belt (front) home position KB1242-AA14LF 22837068
10 FEEDER_ASSY NO.85 Rack conveyor belt (front) home position KB1240-AA23LF 22837054
DRIVE MECH_ASSY
16 Sample tube monitoring KI1235-AALF 22837023
NO.84
SA-10 MAIN
18 Rack full KR1207-AALF 22837071
COMPLETE
2 FEEDER_ASSY NO.32 Stopper position monitor for Supply Rack KI1235-AALF 22837023
7 FEEDER_ASSY NO.46 Rack conveyor belt (rear) home position KB1242-AA14LF 22837068
8 FEEDER_ASSY NO.46 Rack conveyor belt (rear) home position KB1240-AA23LF 22837054
9 FEEDER_ASSY NO.46 Rack conveyor belt (front) home position KB1242-AA14LF 22837068
10 FEEDER_ASSY NO.46 Rack conveyor belt (front) home position KB1240-AA23LF 22837054
DRIVE MECH_ASSY
16 Sample tube monitoring KI1235-AALF 22837023
NO.84
SA-10 MAIN
18 Rack full KR1207-AALF 22837071
COMPLETE
Sampler
Unidirectional Sampler Analysis Stop 3
Measurement
Purpose of Error To ensure that the rack on the feed-in rack table sends into the
Detection measurement line.
Check the Rack Feed-in Arrival Sensor and the Feed-in Table Rack
Method of Error
Detection Sensor
Detection
during operating rack feed-in.
When the rack feed-in motor operates a specified number of pulses, the
Conditions of Error
rack feed-in arrival monitoring sensor is not shaded and the the rack
Detection
feed-in table rack detecting sensor is not shaded.
Action Message Remove the rack and tubes from the tube holder and analysis line and
Screen press the [Execute]. The rack feed-in mechanism will be initialized.
HELP [Execute] The rack feed-in mechanism will be initialized. The error clears when
Process the process above is completed.
Show Detailed
None
Procedure Key
2015037
Sampler
Unidirectional Sampler Analysis Stop 3
Measurement
To ensure that the rack feed-in unit lever return to the home position
Purpose of Error
after the rack feed-in operated, and the next rack feed operates
Detection
properly.
Check the Rack Feed-in Home Position Sensor during initializing the
Method of Error
rack feed-in
Detection
mechanism.
When the rack feed-in motor operates a specified number of pulses, the
Conditions of Error
rack feed-in
Detection
home position sensor is not be shaded.
Action Message Remove the rack and tubes from the tube holder and analysis line and
Screen press the [Execute]. The rack feed-in mechanism will be initialized.
HELP [Execute] The rack feed-in mechanism will be initialized. The error clears when
Process the process above is completed.
Show Detailed
None
Procedure Key
Sampler
Unidirectional Sampler Analysis Stop 3
Measurement
Purpose of Error
To ensure that the rack is moved to the designated location properly.
Detection
Method of Error
Check the Rack Move Sensor (Right, Left), when rack is moving.
Detection
Error is checked when both Move sensor (Right and Left) are not
shaded for more than 50 ms during the process from starting the
shifting motor operation to the completely returning to the initial position.
System monitors at a constant timing (every 0.01 second).
Error is checked when Move sensor (Right) or Move sensor (Left) is not
shaded during the process from the finishing rack shift operation to next
Conditions of Error rack shift operation. System monitors at a constant timing (every 0.1
Detection second).
Except, while from hand picks up test tube though hand returns test
tube to rack (with performing mixing and analysis).
Error is checked when Tube Position Sensor (right or left) is shaded at
the timing during the interrupt READY. System monitors at a constant
timing (every 0.1 second).
Action Message Remove the rack and tubes from the tube holder and analysis line and
Screen press the [Execute]. The rack move mechanism will be initialized.
HELP [Execute] The rack move mechanism will be initialized. The error clears when the
Process process above is completed.
Show Detailed
None
Procedure Key
Sampler
Unidirectional Sampler Analysis Stop 3
Measurement
Method of Error Check the Move Home Position Sensor during initializing the move
Detection mechanism.
When the rack move motor operates a specified number of pulses, the
Conditions of Error
rack move
Detection
home position sensor is not be shaded.
Action Message Remove the rack and tubes from the tube holder and analysis line and
Screen press the [Execute]. The rack move mechanism will be initialized.
HELP [Execute]
The error clears when the process above is completed.
Process
Show Detailed
None
Procedure Key
Sampler
Unidirectional Sampler Analysis Stop 3
Measurement
Purpose of Error To ensure that the rack ejects from the analysis line to the rack ejection
Detection table properly.
Method of Error Check the Rack Ejection Home Position Sensor during ejecting the
Detection rack.
Conditions of Error When the rack eject motor operates a specified number of pulses, the
Detection rack eject home position sensor is shaded.
Action Message Remove the rack and tubes from the tube holder and analysis line and
Screen press the [Execute].
HELP [Execute] To perform Initializing rack eject mechanism The error clears when the
Process process above is completed.
Show Detailed
None
Procedure Key
Sampler
Unidirectional Sampler Analysis Stop 3
Measurement
To ensure that the lever of the rack ejection mechanism return to the
Purpose of Error
home position and
Detection
the rack ejection mechanism is initialized properly.
Method of Error Check the Rack Ejection Home Position Sensor during ejecting the
Detection rack.
Conditions of Error When the rack eject motor operates a specified number of pulses, the
Detection rack eject home position sensor is not shaded.
Action Message Remove the rack and tubes from the tube holder and analysis line and
Screen press the [Execute]. The rack move mechanism will be initialized.
HELP [Execute] To perform Initializing rack eject mechanism The error clears when the
Process process above is completed.
Show Detailed
None
Procedure Key
Sampler
Unidirectional Sampler Analysis Stop 3
Measurement
Meaning The rack exists on the analysis line when analysis starts.
Purpose of Error
To ensure the sampler analysis starts properly.
Detection
Method of Error
System checks blood volume sensor and tube detecting sensor.
Detection
Error is checked when the tube detection sensor detects the tube when
starting the sample analysis.
Conditions of Error Error is checked when the sample volume sensor detects blood for the
Detection tube when starting the sample analysis
Error is checked when tube move sensor (right or left) is not shaded
when starting the sampler analysis.
Action Message Rack detected on analysis line at start of sampler analysis. Remove rack from
Screen analysis line and press [Accept].
HELP [Execute]
None. The error clears when [Accept] key is pressed.
Process
Show Detailed
None
Procedure Key
Sampler
Unidirectional Sampler Analysis Stop 3
Measurement
To ensure the rack is located on the measurement line and the tube can
Purpose of Error
be returned to the original position in the rack after the sample is
Detection
aspirated.
Method of Error System checks move Sensor(right and left). When error occurs, hand
Detection clipper does not descend.
Error is checked when Move sensor (Right) or Move sensor (Left) is not
Conditions of Error shaded during the process from the finishing rack shift operation to next
Detection rack shift operation. System monitors at a constant timing (every 0.1
second).
Action Message Press [Accept], place rack on feed-in table again, and start sampler
Screen analysis.
HELP [Execute]
None. The error clears when [Accept] key is pressed.
Process
Show Detailed
None
Procedure Key
Sampler
NA (Occurs only before sampler analysis starts.)
Measurement
Meaning Rack is not placed on rack feed-in table when sampler analysis starts.
Purpose of Error
To urge operator to set the racks in the right rack pool.
Detection
Method of Error
System checks Rack Feed In Sensor or Rack Feed In Lever Sensor.
Detection
Conditions of Error Error is checked when rack feed in lever sensor is shaded after rack
Detection feed in operation starts.
Action Message Press [Accept], place rack on feed-in table again, and start sampler
Screen analysis.
HELP [Execute]
The error clears when [Accept] key is pressed.
Process
Show Detailed
None
Procedure Key
Sampler
Unidirectional Sampler Analysis Stop 3
Measurement
Purpose of Error
Prompt the user to remove racks from the rack ejection table.
Detection
Method of Error Prompt the user to remove racks from the rack ejection table. When
Detection error occurs, rack is not shifted. It resumes when error is recovered.
Error is checked when the rack full sensor detects the rack when
Conditions of Error starting the sampler mode analysis.
Detection When the racks are detected by the rack ejection table rack full sensor
during starting the rack ejection operation.
Action Message
Remove racks from rack ejection table.
Screen
None. The error clears when the racks were not detected by the rack
HELP [Execute]
ejection table rack full
Process
sensor.
Show Detailed
None
Procedure Key