6.12 Others.....................................................................................................................................................39
6.12.1 Voltage Low Limit ......................................................................................................................39
6.12.2 CPU Failure (“Number”) ............................................................................................................40
[B]
Barcode Scanner Driver Error ..........................................................................................................................23
Barcode Scanner Error .....................................................................................................................................23
[C]
CPU Failure (“Number”)....................................................................................................................................40
Change Sample Rack.......................................................................................................................................24
Check Control Expiry ........................................................................................................................................28
Check Reagent Expiry ......................................................................................................................................28
Check Reagent Volume 1.................................................................................................................................38
Check Syringe Unit ...........................................................................................................................................20
[E]
Enter Work List .................................................................................................................................................38
Exist same Sample No. ....................................................................................................................................26
[H]
HC ACK Code Error..........................................................................................................................................32
HC ACK Time Out ............................................................................................................................................33
HC CTS Time Out.............................................................................................................................................32
HC Communication Error..................................................................................................................................32
HC ETX Time Out.............................................................................................................................................33
HC Off Line .......................................................................................................................................................31
HC Reception Count Error ................................................................................................................................32
HC STX Time Out.............................................................................................................................................33
HC Transmission Count Error ..........................................................................................................................32
[I]
Instruction Not Found in HC .............................................................................................................................33
Insufficient Reaction Tube ................................................................................................................................38
Insufficient Reagent (Holder No.) .....................................................................................................................16
Interrupt by Mechanical Stop ............................................................................................................................36
Invalid Settings..................................................................................................................................................29
[L]
Light Shield Open .............................................................................................................................................37
Load New Sample.............................................................................................................................................26
[N]
No Paper...........................................................................................................................................................30
No Reaction Tube.............................................................................................................................................17
No Sample (Rack Position)...............................................................................................................................14
[P]
Pressure Pump Error ..........................................................................................................................................1
Printer Error ......................................................................................................................................................31
Probe Crash......................................................................................................................................................18
[Q]
QC Data Error ...................................................................................................................................................29
QC Limit Error...................................................................................................................................................34
QC Limit Error (STOP)......................................................................................................................................34
[S]
Sampling Error (Position)..................................................................................................................................15
Serial Illegal Error .............................................................................................................................................33
Set Sample Rack ..............................................................................................................................................21
Set more than two points ..................................................................................................................................34
Standard Curve Error........................................................................................................................................35
Stored Data Error..............................................................................................................................................29
Syringe Error.....................................................................................................................................................19
[T]
Temp. Sensor Error/(Position) ............................................................................................................................6
Temperature Error (High)/Cooler, Temperature Error (Low)/Cooler ..................................................................5
Temperature Error (High)/Detector, Temperature Error (Low)/Detector ............................................................4
Temperature Error (High)/Incubator, Temperature Error (Low)/Incubator .........................................................3
Temperature Error (High)/Room, Temperature Error (Low)/Room ....................................................................6
Tube Catch Error (Position) ..............................................................................................................................12
Tube Release Error (Position) ..........................................................................................................................13
Turned Off During Operation ............................................................................................................................37
[V]
Vacuum Pump Error ...........................................................................................................................................2
Voltage Low Limit..............................................................................................................................................39
[W]
Waste Bottle is full ..............................................................................................................................................7
[X]
X Axis Home Position Error ................................................................................................................................8
[Y]
Y Axis Home Position Error ................................................................................................................................9
[Z]
Z Axis Down Error.............................................................................................................................................11
Z Axis Home Position Error ..............................................................................................................................10
Pipette Crash
Sensor
HOST
Sampler Cover Paper Empty
Open Sensor Sensor
Sysmex Tube
CA-500
SERIES Detection
Sensor
STAT
Barcode Reader
Home Position Sensor
Monitor Timing: At time of each analysis start and/on rinse start, just before opening SV
2
Monitor Method: Monitoring 225 g/cm r 10% by the comparator
CA-500’s Action: Emergency stop
-141V peak
Figure 6-1-1: Normal Oscilloscope Wave
Monitor Timing: At time of rinse sequence is started and the sample probe has moved to the rinse cup
Monitor Method: The liquid sensor checks the error when rinse liquid has collected in the rinse cup.
CA-500’s Action: Emergency stop
NOTE: Use high input impedance device. Usually normal Digital Volt Meter input impedance
is higher than 1 Mega : and Oscilloscope is higher than 10 Mega :.
3. When the voltage between pin 4 and 5 of CN8 (PCB No. 4086) shows 12 V, the harness of Peltier
heater unit is failed. Otherwise, the PCB No. 4086 is failed.
R R
E
Monitor Timing: When the probe has been apart from liquid surface after sample or reagent aspiration
Monitor Method: The capacitance liquid level detection sensor monitors the probe has not been apart from
the sample or reagent.
CA-500’s Action: In the case of a sample, the instrument stops. In the case of a position of reagent, the
concerned analysis will not be done. The related analysis parameter of reagent will not be
done thereafter. In the case of rinse reagent, however, emergency stop is made.
Monitor Timing: Just after the catching operation at reaction tube rack
Monitor Method: Tube detection sensor monitors a reaction tube.
CA-500’s Action: Interruption
* If sample discharged in the tube rack or on the reagent stage, clean them.
* PCB No. 2134 latches Crash signal each time. So, confirm the above by pressing the liquid level sensor
reset switch in the service mode.
Monitor Timing: When turning on power supply and changing the preset
Monitor Method: CPU communicate with Bar Code Scanner
CA-500’s Action: Warning
Monitor Timing: When the analysis is re-started after interrupted to load new sample
Monitor Method: Sample rack sensor is monitored.
CA-500’s Action: Warning
Coagulation
Error code Status Action
ERROR0001 Temperature error occurred when Refer to 6.2 Temperature Errors.
(Temp Error) sample was being analyzed.
ERROR0002 Detected coagulation curve shows Reanalyze the sample to make
(Slight Coagulation) extremely small change. comprehensive judgment, taking sample,
reagent, etc. into account.
ERROR0004 Coagulation did not finish within Reanalyze the sample to make
(Analysis Time preset detection time or within Auto comprehensive judgment, taking sample,
Over) Extension Mode detection time. reagent, etc. into account. Or conduct
another test with prolonged detection time.
ERROR0008 1. Curve shows a drop in some area.
(Coag. Curve Error) 2. Curve shows too steep in some
areas.
3. As buffer was cool, clotting time
was extended by the influence of
air bubbles. (Fbg)
ERROR0016 Turbidity was too high to analyze. Reanalyze the sample by pre-dilution.
(Turbidity Level
Over)
ERROR0032 Clotting could not be detected. Reanalyze the sample after checking
(No coagulation) samples and reagents (storage status blood
collecting method, etc.) for any errors, and
make comprehensive judgment.
ERROR0064 Due to any problem before photo Reanalyze the sample after preparing
(Measurement detection, such as insufficient sample required amount of plasma.
Error) plasma, instrument error, insufficient
reagent, Mechanical Stop Switch
operation, analysis could not be done.
ERROR0100 The detection time at 50% point is Reanalyze the sample after checking
(Range Over) exceeding the preset time, i.e. samples and reagents (storage status,
ERROR0200 coagulation is drastic. blood collecting method, etc.) for any errors,
(Fbg Range Over) and make comprehensive judgment.
Chromogenic/Immunoassy
Error code Status Action
ERROR0401 Transmitted Light was too low. Check the reagent, lamp, etc. and reanalyze.
(Trans Light Low)
ERROR0402 Transmitted Light was too high. Check the reagent, lamp, etc. and reanalyze.
(Trans Light High)
ERROR0404 Reaction Curve had no linearity. Reanalyze the sample.
(No Linearity)
ERROR0408 Reaction Curve changed to the Check the setting, reagent and reanalyze.
(Reaction Curve Error) opposite direction.
ERROR0416 Antigen became range over (with Reanalyze the sample by re-dilution, etc.
(Range Over) Immunoassy only).
ERROR0528 Polynomial adjustment could not Check the setting, reagent and reanalyze.
(No Polynomial be performed (when using
Adjustment) VlinIntegral mode only).
+5
GND
Monitor Timing: At time of analysis start, during analysis, and rising mode.
Monitor Method: Photo interrupter monitors the light shield cover is open or closed by the light shield.
CA-500’s Action: Emergency stop
No. 6
1. Main program failure of PCB No. 7015
No. 6
1. Main program failure