SERVICE NEWS N°08 Date 21.05.

2003

INSTRUMENT MODEL : LIASYS NEW

X INFO X MODIFICATION Mandatory SPARE PARTS


X Suggested

SUBJECT: DILUTER FITTINGS

INSTRUMENTS CONCERNED : X ALL FROM S/N:

TECHNICAL MOTIVATION:

THE CONNECTION TUBE BETWEEN THE ELECTROVALVE (EV1) AND THE DILUTER HAS
BEEN CHANGED IN ORDER TO GUARANTEE A GREATER DEGREE OF RESISTANCE AND
RELIABILITY IN TIME.

DETAILS:

Description: Kit for EV-Diluter connection ; Code: 99-00906-00

Procedure:

1. Unscrew and remove Ref. A and C in Fig. 1.

2. Unscrew Ref. D in Fig. 1.

3. Insert and tighten by hand the pipe fitting (Ref.2 – Fig.2) in the diluter’s tap.

4. Insert the pipe fitting (Ref. 1 – Fig.2) in the tap of the electro-valve (EV1) and tighten it
by using a 6 mm monkey wrench.

5. Insert the tube (Ref. 3 – Fig. 2) between Ref. 1 and Ref. 2 in Fig. 2.

6. Enter in the “Diagnostic” program

7. Select the “Diluter” folder

8. Select the “Probe Wash Test” button to check the right hydraulic packing of the sampling
path.

AMS S.r.l. Doc.Code Rev. Page 1 of 3


Via E.Barsanti 17/A - 00012 GUIDONIA (ROMA) MD 28 00
This document can not be reproduced, distributed or utilized in any other way without authorization of AMS - ANALYZER MEDICAL SYSTEM S.r.l.
SERVICE NEWS N°08 Date 21.05.2003

Warning: In order to maintain the high level of precision , it is absolutely necessary that ONLY the
standard tube provided by Assel (Code: 90-01253-00) be used.

* Drawing Ref. 1: E.V. (Rinse) Fitting Code: C190-01254-00

* Drawing Ref. 2: Diluter Fitting Code: C101-01222-00

* Drawing Ref. 3: Tygon Tubing Code: 90-01253-00

B 3

(µP1) (µP1)

C
1
D A 2

2 2
(µP2) (µP2)

1 1

Fig.1 Fig.2

A 2

A A

NOTE: THE ABOVE MENTIONED MODIFICATION HAS ALREADY BEEN MADE


STARTING FROM S/N: 0304106 AND THEREAFTER.

AMS S.r.l. Doc.Code Rev. Page 2 of 3


Via E.Barsanti 17/A - 00012 GUIDONIA (ROMA) MD 28 00
This document can not be reproduced, distributed or utilized in any other way without authorization of AMS - ANALYZER MEDICAL SYSTEM S.r.l.
SERVICE NEWS N°08 Date 21.05.2003

ENCLOSED: ----------------

Editing Approval

Enzo ZATTINI Antonio GAGLIARDUCCI

AMS S.r.l. Doc.Code Rev. Page 3 of 3


Via E.Barsanti 17/A - 00012 GUIDONIA (ROMA) MD 28 00
This document can not be reproduced, distributed or utilized in any other way without authorization of AMS - ANALYZER MEDICAL SYSTEM S.r.l.

You might also like