TECHNICAL MOTIVATION:
THE CONNECTION TUBE BETWEEN THE ELECTROVALVE (EV1) AND THE DILUTER HAS
BEEN CHANGED IN ORDER TO GUARANTEE A GREATER DEGREE OF RESISTANCE AND
RELIABILITY IN TIME.
DETAILS:
Procedure:
3. Insert and tighten by hand the pipe fitting (Ref.2 – Fig.2) in the diluter’s tap.
4. Insert the pipe fitting (Ref. 1 – Fig.2) in the tap of the electro-valve (EV1) and tighten it
by using a 6 mm monkey wrench.
5. Insert the tube (Ref. 3 – Fig. 2) between Ref. 1 and Ref. 2 in Fig. 2.
8. Select the “Probe Wash Test” button to check the right hydraulic packing of the sampling
path.
Warning: In order to maintain the high level of precision , it is absolutely necessary that ONLY the
standard tube provided by Assel (Code: 90-01253-00) be used.
B 3
(µP1) (µP1)
C
1
D A 2
2 2
(µP2) (µP2)
1 1
Fig.1 Fig.2
A 2
A A
ENCLOSED: ----------------