Professional Documents
Culture Documents
Purpose:
The purpose of this SOP is to outline the procedure for running the Prothrombin Time (PT) test using
TEClot PT-S reagent on the TECO Coatron M2 instrument.
2. Materials:
d. Use plasma within 4 hours; otherwise, store frozen and thaw just prior to use.
4. Reagent Preparation:
a. Reconstitute the TEClot PT-S Reagent with high purity water as follows:
- A0230-010: 2 mL
- A0230-040: 4 mL
- A0230-100: 10 mL
b. Let the reagent stand at room temperature with occasional swirling for at least 15 minutes.
c. Place the reagent into the instrument and let it incubate for an additional 15 minutes.
d. Swirl the reagent before each testing. A mixing bar can be used for this purpose on Coatron
instruments.
Note:
Unopened reagents are stable until the expiration date shown on the label when stored at 2°-8°C.
Opened reagent stability:
37ºC: 8 hours
9. Expected Results:
note : results are influenced by instruments, technique, calibration etc. Each laboratory is recommended
to establish its own range on the specific instrument used.