PQ Test Protocol MPA Sample Compartment 20231211 210221
PQ Test Protocol MPA Sample Compartment 20231211 210221
Maximum 100% Line Deviation: 0.5 Measured 100% Line Deviation: 0.25
Energy Test
________________
Page 1 of 5 Page Signature
OVP - PQ Test Protocol
Signal to Noise Test, Spectra
100% Line Test, Spectra
Energy Test, Spectra
101.0
99.5 100.0 100.5
Transmittance [%]
99.0
1 aa9c1054-74b5-4fcf-ac3f-a7f3299c6973
2 9da99bb5-4550-40f2-82d3-26bd21a5314d
3 acea8d16-340d-427b-9854-3afc43f8b8bd
4 acd86fce-bece-4300-878e-0a3a6574dfea
5 3cd4bf6c-06c1-41f4-af34-dba45b9a2f59
________________
Page 2 of 5 Page Signature
OVP - PQ Test Protocol
Interferogram Peak Test, Spectrum
0.25
0.15
0.05
-0.05
-0.15
aa9c1054-74b5-4fcf-ac3f-a7f3299c6973
________________
Page 3 of 5 Page Signature
OVP - PQ Test Protocol
Wavenumber Accuracy Test BRM 2065 (35.4 Deg. C, Peak is T. corr.), Spectrum
10245.69
0.8
Absorbance Units
0.4 0.6
0.2
f25f36ff-7a3b-45cc-b45c-07578cd28bd5
________________
Page 4 of 5 Page Signature
OVP - PQ Test Protocol
Photometric Reproducibility Test - Glass Filter A, Spectrum
25
Transmittance [%]
10 15 20
5
b5e0e224-8a09-49d9-bf5c-d3b1b15b23d9
________________
Page 5 of 5 Page Signature