Manual
(suitable for PUTF201/202 clamp-on and insertion)
Notice
Please check the shipping list to ensure received goods and ordered goods are same.
Please read the manual carefully before installing and operating.
Please obey all the warnings and requirements before installing, operating and maintaining.
Please operate the products according to the manual strictly. If the users don‟t operate the products as the
manual to result in any consequence, our company will not assume any responsibility.
All the figures are only for reference. Please refer to the real products when operating or consult with
our company.
All products are detected strictly before delivering. If users disassemble the products personally to result
in any loss, our company will not assume any responsibility.
Please read name plate carefully to ensure the flow meter is produced according to your order. Please
check whether the power and voltage on name plate are correct.
Please check the products carefully before installing, please consult with our company if have any
obvious transporting damage.
Please keep the manual properly for your reference anytime.
Please contact our company if the product can‟t work properly or need repaired.
Warning
To avoid damaging users‟ benefits, please read the manual carefully and obey all the warnings and
instructions.
2
PUTF201/202 series ultrasonic flow meter manual
Security Warning
To avoid any financial loss and personal injury, please obey following safety items and operate the products
properly.
① The products are precision measurement instruments, please handle with care to avoid any
damage.
② About AC power supply:
Please do not disassemble the transducer and enclosure in the condition of charging.
Please strictly obey local occupational health and safety. Only professional staff are allowed to
operate.
Please do not touch the plug with wet hands to avoid electric shock and personal injury.
Please do not pull, knot, cut or overbend the electric cable and do not put heavy items on it.
Please do not use the electric cable if insulating layer damaged, avoid personal injury by
electric leakage.
Please avoid to share power with other large-scale equipment. Please adopt power conditioner if
Please install lightning protection device when operating in the lighting stroke properly.
Please do not clean or cover the product with corrosive liquids. Otherwise will reduce the working life
Please carefully operate all exposed thread part which may scratch skin.
Please do not change the transducer cable length, otherwise will affect the features.
Please carefully read the installation requirements and notice about technical parameters before confirming
install position. If beyond the scope of technical parameters, instrument may not work properly or cause other
effect.
3
PUTF201/202 series ultrasonic flow meter manual
CONTENTS
1. SUMMARY ................................................................................................................................................... 1
1.1 Introduction........................................................................................................................................... 1
1.2 Measuring Principle .............................................................................................................................. 1
1.3 Application............................................................................................................................................ 2
1.4 Features ................................................................................................................................................. 2
1.5 Specification ......................................................................................................................................... 3
1.6 Presentation........................................................................................................................................... 4
1.7 Dimension ............................................................................................................................................. 5
4. MENU .......................................................................................................................................................... 18
5. OPERATION INTRODUCTION.............................................................................................................. 40
4
PUTF201/202 series ultrasonic flow meter manual
7.1. Introduction........................................................................................................................................ 59
7.2. Thermal energy meter wiring connection .......................................................................................... 59
7.3. Temperature Range ............................................................................................................................ 60
7.4. Temperature Sensor Mounting Requirements .................................................................................... 60
7.5. Temperature Sensor Mounting Steps ................................................................................................. 60
5
1. Summary
1.1 Introduction
PUF201/202 transit-time ultrasonic flow meter utilizes transit-time principle, including clamp-on and
insertion. When install clamp-on transducer, without flow stop and pipe cutting, simple installation and
convenient for online comparison. Different sizes of transducers can measure different sizes‟ pipe. Insertion
installation effectively solves clap-on flow meter can not accurately measure with scaling pipe and
non-conductive media. Insertion transducer has stop ball valve, without flow stop or pipe cut, easy to install.
User also can mount hoops on unweldable pipe surface. Optional thermal energy measuring function. Widely
applied in processing monitoring, water balance test, district heating balance test, energy efficiency
monitoring etc.
Transit-time ultrasonic flow meter utilizes the difference of ultrasonic pulse forwarding and reversing
flow rate to measure flow. Use a pair of transducers to transmit and receive signal alternately. Transducer
signal travels faster downstream than upstream. Obtain transit-time △t by measuring signal travel time in
downstream and upstream, the average flow velocity can be detained according to △t and velocity V. The
volume flow Q can be calculated out of the flow velocity V and pipe profile S.
V=K*△t
Q=S*V
K: Constant
S: Pipe sectional area
Z Method
V Method
W Method
1
PUTF201/202 series ultrasonic flow meter manual
1.3 Application
Water Utility
Water Conservancy & Resource
College & University
HVAC
Manufacturing
Draining Company
Heating Supply Industry
Petrochemical Industry
1.4 Features
4 lines display velocity, flow rate, total flow and meter status
Clamp-on mounted ,without pipe cutting Or flow stop
Fluid temperature range -40℃~260℃
Built-in date logger is optional
Selecting temperature sensor to measure thermal energy
Insertion flow meter is suitable for measuring DN65-6000 pipeline flow
Bi-directional measurement, flow range from 0.01m/s to 12m/s
2
PUTF201/202 series ultrasonic flow meter manual
1.5 Specification
Transmitter
Transducer
3
PUTF201/202 series ultrasonic flow meter manual
1.6 Presentation
4
PUTF201/202 series ultrasonic flow meter manual
1.7 Dimension
Transmitter
5
PUTF201/202 series ultrasonic flow meter manual
24 5
14 5
10 5
10 5
8 5
24 5
6
PUTF201/202 series ultrasonic flow meter manual
The appointed pipeline surface must be cleaned completely before installing transducer and ensuring
couplant stick well with pipe.
Do not install transducer on elbows, bends or flow cutting pipes (e.g. electromagnetic flow meter),
enable transducer parallel with pipes.
If transducer can‟t parallel with pipes in limite d location, the transducer can be installed vertically
or inclined when no air bubbles in upper part of the pipe.
Use enough couplant to fill the gap between transducer and outer wall of pipe when installing
transducer. Recommend normal temperature couplant in normal temperature and high temperature
couplant in high temperature. Also can use other similar products to replace.
There are 4 kinds of mounting methods “V”, “Z”, “W” and “N”(“W” and “N” method ar e
uncommonly used). Usually V method is used as standard because it provides longer signal path,
therefore having more accurate measurement of flow velocity. It is recommended for pipes sizes
between 25-6000mm.Special attention should be given when mountin g transducers to ensure that
sensors are aligned in direct line and along the center of the pipe and the transducer mounting line
should be aligned.
V method
Z method is recommended when having serious scaling inside of the pipe, entrained air bubb les,
too thick liner etc. Which can induce a lot of noise into signal. Using this method, the signal is
dispersed in a straight path reducing the possibility of noise induction.
Z Method
7
PUTF201/202 series ultrasonic flow meter manual
Upstream transducer connects TRANSMIT, downstream transducer connects RECEIVE. It‟s better
to use exclusive cable.
The appointed pipeline surface must be cleaned completely before installing transducer and ensuring
couplant stick well with pipe.
Notice:
Installing transducer parallel with pipes;
Transducer installing direction.
Note that on horizontal pipe, install transducer at the horizontal axis of pipe sectional area in case
of air bubbles and gravels between transducer and pipe wall.
If transducer can‟t be installed horizontally, then install it vertically or tilt when have no air
bubbles.
Features
Solving the problem that when pipe inside scale or erode seriously, clamp -on transducer
signal is poor and can‟t measure properly.
Can install on cement pipe, cast iron pipe and other non -welding pipe material(need to
customize pipe hoop according pipe outer diameter).
Special trepanning tools can install the transducer with pressure and water, this ensures stable
production and don‟t need to cut off water in future maintenance.
Ultrasonic wave releasing probe touches the measured fluid directly, this improve measuring
accuracy and machine‟s stability.
Compared with electromagnetic flow meter, the insertion transducer is suitable for big
pipeline and it‟s economic and accurate.
Insertion transducer installation needs a certain space, generally the distance between pipe outer
wall and wall is over than 700mm.
Requirements:
8
PUTF201/202 series ultrasonic flow meter manual
1. Installation place
Insertion transducer installation needs bigger space, the distance L1 between pipe ou ter wall and
wall is over than 540mm (welded pipe e.g. carbon steel pipe and stainless steel pipe). Width
W >(D+540×2)mm, L1 over than 700m(unwelded pipe e.g. PE/PVC, cast iron, cement pipe)
W >(D+700×2)mm(as drawing 1).
Drawing 1
Distance
L﹥(D+1000)
Ultrasonic flow meter installation needs certain straight pipe, generally upstream more than 10D,
downstream more than 50D, distance with pump exit or valve more than 30D.
2.7.2 Tool
Installing insertion transducer needs tape, straight screwdriver, cross screwdriver, electric
hand drill, special trepanning tool, monkey spanner and Teflon tape
9
PUTF201/202 series ultrasonic flow meter manual
1. Location
Input the pipe parameters into transmitter, calculating installation distance(recommend to use Z
method) to decide two transducer‟s location, the installation distance is two transducers‟ center
spacing.
Notice: Two transducers must be in same axial plane.
Making location paper: Using a rectangular tape with 4D length(D is pipe diameter), 2D width to
draw a line from edge about 100mm(as drawing 3)
100mm
Twining the location paper on cleaning pipe, need to overlap and align the two side of paper,
which can make line parallel with pipe axle. (as drawing 4)
Pipe axis
Drawing 4
Lengthening the straight line of location paper and drawing a straight line on pipe. (as drawing 5)
Drawing 5
The intersection point of drawing straight line and location paper edge is A. ( as drawing 6)
Drawing 6
A
A
Start from A, measuring the pipe 1/2 perimeter along location paper edge, this point is C, at point
C, drawing a straight line vertical to location paper edge. ( as drawing 7)
Drawing 7
C A
A
10
PUTF201/202 series ultrasonic flow meter manual
Removing location paper, start from C, measuring installation distance L at the drawing straight
lines, then deciding point B. Points A and B are the installation position, e.g. L=280mm (as
drawing 8)
280mm
Drawing 8
•
• •
Drawing 10 1 2 3 4 5 6 7 8
11
PUTF201/202 series ultrasonic flow meter manual
1. Pipeline 2. Ball valve base 3. Special ball valve 4. Location drill 5. 19 super hole
cutter 6. Seal cartridge 7. Drill pipe 8. Electric hand drill
1. Transducer 2. Ball valve base 3. Ball valve 4. Guide screw thread 5. Nut
6. Junction box 7. Signal cable
5. Dimension calculation when transducer stretches into pipe inner wall (as drawing 12)
Transducer‟s length a and pipe wall thickness b are known, transducer length outside of pipe L
also can be measured, just need L = a – b, and c = 0.
Drawing 12 L
Pipe
Wall
C
A
Wire connection
terminal
Shielding
layer
7. Fastening inlet hole nut after connection, then fully filling the junction box by waterproof glue,
finally screwing on the box cover.
Transducer inspection is used for checking whether the installation is proper and can receive
correct, strong ultrasonic signal. And it enables flow meter to operate realibility. By checking
signal strength, transmitter time, time difference and time ratio to ensure whether the installation
is right.
Signal strength (displayed in Window M90) indicates a detected strength of the signal both from upstream and
downstream directions. The relevant signal strength is indicated by numbers from 00.0~99.9. 00.0 represents
no signal detected while 99.9 represent maximum signal strength.
Normally, the stronger the signal strength detected, the instrument will work more reliably, as well as the
more stable the measurement value obtained.
Adjust the transducer to the best position and check to ensure that sonic couplant is applied adequately during
installation in order to obtain the maximum signal strength. System normally requires signal strength over
60.0, which is detected from both upstream and downstream directions. If the signal strength detected is too
low, the transducer installation position and the transducer mounting spacing should be re-adjusted and the
pipe should be inspected again. If necessary, change the mounting to the Z method.
Q value is short for Signal Quality (displayed in Window M90). It indicates the level of the signal detected. Q
value is indicated by numbers from 00~99. 00 represents the minimum signal detected while 99 represent the
maximum. System normally requires signal strength over 60.0.
“Total Time and Delta Time”, which displays in Window No.93, indicates the condition of the installation.
The measurement calculations in the flow meter are based upon these two parameters. Therefore, when
“Delta Time” fluctuates widely, the flow and velocities fluctuate accordingly. This means that the signal
quality detected is too poor, it may be the resulted of poor pipe-installation conditions, inadequate transducer
installation or incorrect parameter input. Generally, “Delta Time” fluctuation should be less than ±20%. Only
when the pipe diameter is too small or velocity is too low, the fluctuation can be wider.
Transit Time Ratio indicates whether the transducer mounting spacing is accurate or not. If the installation is
proper, the normal transit time ratio should be 100±3%. Check it in Window M91. If the transit time ratio is
over 100±3%, it is necessary to check:
(1) Whether the parameters (pipe outside diameter, wall thickness, pipe material, liner, etc.) have been
entered correctly;
13
PUTF201/202 series ultrasonic flow meter manual
(2) Whether the transducer mounting spacing is accordance with the display in Window M25;
(3) Whether the transducer is mounted at the pipe‟s centerline on the same diameter;
(4) Whether the wall thickness is too thick or the pipe mounting is distorted in shape, etc.
2.8.5 Warnings
1)Pipe parameters entered must be RIGHT; Otherwise the flow meter will not work properly.
2)During the installation, apply enough couplants in order to stick the transducer onto the pipe wall. While
checking the signal strength and Q value, move the transducer slowly around the mounting site until the
strongest signal and maximum Q value can be obtained. Make sure that the larger the pipe diameter, the more
the transducer should be moved. Check to be sure the mounting spacing is accordance with the display in
Window M25 and the transducer is mounted at the pipe‟s centerline on the same diameter. Pay special
attention to those pipes that formed by steel rolls (pipe with seams), since such pipe is always irregular. If the
signal strength is always displayed as 0.00, that means there is no signal detected. Thus, it is necessary to
check that the parameters (including all the pipe parameters) have been entered accurately. Check to be sure
the transducer mounting method has been selected properly, the pipe is not worn-out, and the liner is not too
thick. Make sure there is fluid in the pipe or the transducer is not very close to a valve or elbow, and there are
not too many air bubbles in the fluid, etc. With the exception of these reasons, if there is still no signal
detected, the measurement site has to be changed.
3) Make sure that the flow meter is able to run properly with high reliability. The stronger the signal strength
displayed, the higher the Q value reached. The longer the flow meter runs accurately, the higher the reliability
of the flow rates displayed. If there is interference from ambient electromagnetic waves or the signal detected
is too poor, the flow value displayed is not reliable; consequently, the capability for reliable operation is
reduced.
4) After the installation is complete, power on the instrument and check the result accordingly.
Transmitter should be installed in a position that is convenient for maintaining, calibrating, and
watching the LED. The position should be satisfied below conditions:
because sunlight may make the transmitter temperature rise and exceed the maximum value.
14
PUTF201/202 series ultrasonic flow meter manual
──85-265VAC power supply, power lines are connected to 3 groups of terminals(GND, LC, NC)
as shown in drawing.
──24VDC direct current supply, power lines are connected to the terminal DC24V(+, -) as shown
in drawing,(Direct current 12-36V voltage, power > 3w, meter can work properly).
(The series meters can be in common use 220VAC and 24VDC, users can select according to
actual situation)
──T- UP(+,–,GND)and T-DN(+,–,GND)is used for transducers‟ connection. T- UP means
upstream transducer, T-DN means downstream transducer. “+” connect with red signal line, “-”
connect with black signal line, “GND” connect with silvery shielded line.
Two transducers are equal, don‟t need to appoint which one to transmit and receive, if displa y
minus please exchange two transducers‟ connection.
──T1(+, -) and T2(+, -) are used for PT1000 temperature transducer connection, T1
means water inlet temperature transducer, T2 means water outlet temperature transducer.
numbers.
MENU to select a menu. Press this key first, input two menu numbers and then enter the
selected menu. For instance, to input a pipe Outside diameter, press MENU11 keys, where “11” is the
window Address to display the pipe outside diameter. Buzzer sent key has (beep) sound, can use
15
PUTF201/202 series ultrasonic flow meter manual
input the 2-digit window Address code. For instance, to input or check the pipe out side
diameter, just press the MENU11 keys for window Address code 11.
Another method to visit a particular window is to press ∧, ∨ and ENTER keys to scroll the
menu. For instance, if the current window Address code is 66, press ∧ key to enter Window
No.65, press the ∧ again to enter Window No.64; then, press the ∨ key to back Window
There are 3 types of window: 1. Data, e.g. M11, M12. 2. Option, e.g. M14 3. Display, e.g.
M01, M00.
View corresponding parameters by checking data window. If modify the value, directly press
digit key then press ENT. Also can press ENT, then press digit key, finally press ENT again to
confirm.
press ENTER key to confirm the selection. It is possible to press the key 1 to change the
selection and wait until “1. Stainless Steel” is displa yed on the second line of the screen. Then
Generally, press ENT key first if user wants to enter “modify” condition. If the “modify” is
still not possible even after pressing the ENT key, it means that system is locked by a
password. To “Unlock” it, select “Unlock” in Window No. 47 and enter the original password.
The keypad will not respond if the keypad is locked. It only can be unlocked by the entering
16
PUTF201/202 series ultrasonic flow meter manual
original password. Select keypad lock functions in Window No. 48. Please consult factory for
password if necessary.
17
PUTF201/202 series ultrasonic flow meter manual
displayed transducer mounting spacing and the selected mounting method (See installing the
transducers in Part 2).
9. Press the MENU 01 keys to enter Window No.01 to display measureme nt result.
4. Menu
10
Input pipe outer perimeter 63 Analog input AI1 corresponding value range
11 Input pipe outer diameter 64 Analog input AI2 corresponding value range
15 Input pipe material sound speed 66 Set low frequency output flow rate
Original Setting
16 Select lining material type 67 Set high frequency output flow rate
22
Input fluid viscosity 75 #2 alarm lower limit setting
18
PUTF201/202 series ultrasonic flow meter manual
30
Choose Metric/British unit 83 Automatically supply power-off flow switch
31
Choose flow rate unit Thermal measuring 84 Select thermal unit
32
Choose volume unit Select temperature signal
Flow unit setting
40 Input damping factor 93 Display total transit time and delta time
41 Input low velocity eliminate value 94 Display Reynolds number and pipe factor
45 Instrument factor, amend factor +2 +2 Display the last power off date and time
46 Input network address code +2 +3 Display the last power off flow rate
19
PUTF201/202 series ultrasonic flow meter manual
Timing Output
Compatibility communication protocol
50 Data timing output option setting +6
selection
The shortcut to enter display window is pressing the MENU, then enter two digits window number.
MENU00
Positive Volume/Negative Volume/Net Volume/S,Q Value POS 0 m3
This window displays signal strength and quality Q value. NEG 0 m³
NET 0 m³
Signal strength uses 00.0~99.9 to represent. 00.0 means without S=82.6,82.7,Q=96
receiving signal, 99.9 means maximum signal. Normally, signal
MENU01 POS 0 m3
Flow 0 m³
Positive Volume/ Flow/Velocity/S,Q Value Vel 0 m³
This window displays Positive volume/ Flow/Velocity/S,Q S=82.6, 82.7, Q=96
Value
NEG 0 m³
MENU02 Flow 0.0000 m³/h
Vel 15.238 m/s
Negative Volume/ Flow/Velocity/S,Q Value
S=82.6, 82.7, Q=96
This window displays Negative volume/ Flow/Velocity/S,Q
Value
20
PUTF201/202 series ultrasonic flow meter manual
Net 0 m3
MENU03 Flow 0.0000 m3/h
Vel 15.238 m/s
Net Volume/ Flow/Velocity/S,Q Value
S=82.6, 82.7, Q=96
This window displays Net volume/ Flow/Velocity/S,Q Value
MENU04
Date 15-07-18
Date/Time/ Flow rate/S、Q Value Time 12:44:46
Flow –2421.5 m3/h
This window displays Date/Time/ Flow rate/S、Q Value S=82.6, 82.7, Q=96
MENU05
EFR +453.27 GJ/h
Thermal/Net thermal/ Velocity/S、Q Value E.T 0E+0G
This window displays Thermal/Net thermal/Velocity/S、Q Vel 0.0058m3/h
S=82.6,82.6 Q=91
Value.
MENU06
T1= 0.00000
Analog input T1,T2 -12.4900
This window input analog input T1, T2 current and corresponding T2= 0.00000
-12.4900
temperature, pressure value, and depth value
MENU07
Error code
System error code/S、Q Value
---------------------
This window displays working condition and error code. Maybe Date Time Error
have many error code at same time. S=82.6,82.7,Q=96
MENU08
M08
Flow today Net Flow Today
321.45 m3
This window displays total net flow today.
MENU11
This window is used to enter or change the outside (outer) diameter of the pipeline. 20mm to 6000
MENU12
MENU13
Note:Just need to input one of pipe wall thickness and pipe inner diameter.
MENU14
M14
Pipe material Pipe Material
5. PVC
Window for selecting pipe material, familiar pipe materials include:
ultrasound)
4.Copper 9. Others
Users need to input pipe material corresponding sound speed in M15 window after selecting NO. 9
22
PUTF201/202 series ultrasonic flow meter manual
MENU16
M16
Liner material Liner Material
0. No liner
Window for selecting the liner material:
0. No liner 6. Polystyrene
1. Tar Epoxy 7. Polyester
2. Rubber 8. Polyethylene
3. Mortar 9. Ebonite
4. Polypropylene 10. Teflon
5. Polystryol 11. Others
MENU20
M20
Fluid Type Fluid Type
0. Water
Window for selecting fluid type familiar liquids types:
0. Water 8. Other
1. Sea Water 9. Diesel Oil
2. Kerosene 10 .Castor Oil
3. Gasoline 11.Peanut Oil
4. Fuel oil 12.Gasoline #90
5. Crude Oil 13.Gasoline #93
6. Propane at -45℃ 14. Alcohol
7. Butane(0℃) 15. Water(125℃)
MENU23
0. Standard-Middle
1. Plugin type C
2. Standard-Small
3. User type
23
PUTF201/202 series ultrasonic flow meter manual
4. Standard-Large
5. Plug-in B45
If used select user type transducer, must input 4 parameters, about details please contact our
engineers.
MENU24
M24
Transducer installation method
XDCR Mounting
This window is used to select transducer installation method, 0 V
totally 4 kinds:
0. V
2. N(small pipe)
3. W(tiny pipe)
MENU25 M25
XDCR Spacing
Transducer installation spacing 109.677 mm
distance value M25 displays after users input correct parameter setting.
MENU26
working method:
Entry to Save
Entry to Load
Entry to Browse
MENU27
M27
Current Section
24
17671.5 mm2
PUTF201/202 series ultrasonic flow meter manual
MENU28
MENU29
MENU30
0. Metric
1. English
Default is metric.
MENU31
1. Liters (l)
25
PUTF201/202 series ultrasonic flow meter manual
/day (d)
/hour(h)
/minute ( m)
/second (s)
MENU32
MENU33
1. x 0.01
2. x 0.1
3. x 1
4. x 10
5. x 100
6. x 1000
7.x 10000(1E+4)
Default factor is x1
26
PUTF201/202 series ultrasonic flow meter manual
MENU34
M34
Net totalizer Net Totalizer
This window is used for turning on/off the net totalizer. When turn ON
off, The net volume will be not changed in MOO window. Default is
“ON”.
MENU35
Default is “ON”.
MENU 36
Default is “ON”.
M37
MENU37
Totalizer Reset
Totalizer reset NO
Press ENT to select “YES”or“NO”, if press “YES”, will have below selection
None
All
Net Totalizer
POS Totalizer
NEG Totalizer
Energy totalizer
Press ◄, if users want to remove all parameters and restore factory settings.
27
PUTF201/202 series ultrasonic flow meter manual
MENU38
M38
Manual totalizer Manual Totalizer
Manual totalizer is separate, press ENT to start and press ENT again Enter When Ready
MENU39
M39
Language selection
Language
Select Chinese or English English
MENU40
MENU41
M41
Cut-off low flow rate Low flow Cutoff
This window is used for cutting off low flow rate to make it display 0.01 m/s
then system will treat all measuring value as “0” within ±0.01 , usually input 0.01.
MENU42
M42
Zero calibration Set Zero
Press ENT to go
When the fluid is in the static state, the displayed value is called
If use this function when have flow, which will make flow display “0”, can recover by M43
MENU43 M43
Reset Zero
Reset zero
NO
Selecting “YES” to reset the “zero point”.
28
PUTF201/202 series ultrasonic flow meter manual
MENU44
MENU45
M45
Scale factor Scale factor
The Scale Factor is used to modify the measurement results, default 1
is 1.0, please see the calibration data sheet and set this parameter. If
MENU46
MENU47 M47
System Lock
Password protection
unlocked
This window is used for locking system. System will forbid an y
way to unlock is to input correct password. Password consists of number 1 -4. If users forget the
MENU48
M49
MENU49 Comm. Test
DID
29
PUTF201/202 series ultrasonic flow meter manual
MENU50
M50
Logger Option
Logger Option
If select data logger output, please select “ON”, then, press “ENTER” OFF
MENU51
timing output data from this moment. Set **.**.**, it means the timing
output lasts for an infinite time. Press ENT after set the start time.
MENU52 M52
Data Output
Internal data logging direction control: „To RS-232‟.
Only To‘To RS-232’
This menu cannot be changed.
MENU53
used by manufacturer.
MENU54
Current loop Mode Select, Select Current Loop output mode, Use key ∧ or ∨, can select different mode:
30
PUTF201/202 series ultrasonic flow meter manual
MENU55 M55
4mA or 0Ma output Value CL 4mA Output Value
0 m3/h
The flow unit‟s options are the same as that in Menu 31. When the M56 is
MENU56
M56
20mA output Value CL 20mA Output
14400 m3/h
The flow unit‟s options are the same as that in Menu 31.
MENU57
M57
Current loop checkup CL Checkup
This menu is used for checking whether the CL of flow meter is calibrated. ENTER When Ready
MENU58 M58
Current Loop Output display CL Output Display
10.0000 mA
This menu displays current loop output display. If display 10.0000mA, it
MENU60
M60
Set the date and time of the meter. Date 15-06-18
Time 14:28:16
Press ENTER to change it if necessary. This menu is used for modify
system date and time, 24 hour time format incl. Press ENT and modify after
shown “>”.
MENU61
M61
Display Version information and Serial Number Display Version
Ver 8.21
information and Electronic Serial Number (ESN) that are unique for each ESN=159492
31
PUTF201/202 series ultrasonic flow meter manual
MENU62 M62
RS-232C Setup
RS 232 Communication setup: “115200, None”.
9600, None
(Non-RS485, Baud rate, use do not change.)
MENU63
M63
T1 Temperature range of inlet temperature sensor for heat flow application, AI1 Value Range range
0 - 200
wiring terminals is Tin+, Tin-. Press ENTER and use key ∧ or ∨ to
input temperature upper limit and lower limit. Press 0~9 and * to input
MENU64 M64
T2 Temperature range of inlet temperature sensor for heat flow application, AI2 Value Range range
0 - 200
wiring terminals is Tin+, Tin-. Press ENTER and use key ∧ or ∨ to
input temperature upper limit and lower limit. Press 0~9 and * to input
MENU65
M65
Frequency range for the frequency output FO Freq Range
1000-2000
Setup the frequency range for the frequency output. The biggest range is
MENU66
MENU67
M67
Set high frequency output flow rate
High FO Flow Rate
Setup the high frequency output corresponding value of flow rate. 26550 m³/h
32
PUTF201/202 series ultrasonic flow meter manual
MENU70
1. Always ON
2. Lighting For
MENU71 M71
LCD contrast
LCD contrast control.
0
Press ENT, use ▲or ▼to increase or decrease the light.
MENU72
M72
Working timer Working Timer
0000062:54:40
It displays the whole working time since last cleared. It displays hour,
minute, second. It can be cleared by pressing ENTER key, and then select
YES.
MENU73 M73
Alarm #1 Lower limit of flow rate Alarm #1 Low Value
0 m³/h
Set the low value in this menu. In the same condition in menu M78 or M79,
any measuring under this value may cause the OCT or relay alarming.
M74
MENU74
Alarm #1 High Value
Alarm #1 Upper limit of flow rate 14400 m³/h
Set the high value in this menu. In the same condition in menu M78 or M79,
any measuring under this value may cause the OCT or relay alarming.
MENU75
M75
Alarm #2 Low Value
33
0 m³/h
PUTF201/202 series ultrasonic flow meter manual
MENU76
M76
Alarm #2 Upper limit of flow rate
Alarm #2 High Value
Set the high value in this menu. In the same condition in menu M78 or M79,
14400 m³/h
any measuring under this value may cause the OCT or relay alarming.
MENU77
M77
Buzzer setup Buzzer Setup
Key Stroke On
Optional trigger event is below:
0. No Signal
1. Poor Signal
2. Not Ready
3. Reverse Flow
4. AO Over 100%
5. FO Over 120%
6. Alarm #1
7. Alarm #2
8. Batch Control
9. POS Int Pulse
10. NEG Int Pulse
11. NET Int Pulse
12. Energy Pulse
13. ON/OFF Via RS232
14. Fluid Changed
15. Key Stroke ON
16. FO Via RS-232C
17. Not Using
18.
MENU78
0. NO Signal
1. Poor Signal
34
PUTF201/202 series ultrasonic flow meter manual
2. Not Ready
3. Reverse Flow
4. AO Over 100%
5. FO Over 120%
6. Alarm #1
7. Alarm#2
8. Batch Control
13. FO
MENU79
0. NO Signal
1. Poor Signal
2. Not Ready
3. Reverse Flow
4. AO Over 100%
5. FO Over 120%
6. Alarm #1
7. Alarm #2
35
PUTF201/202 series ultrasonic flow meter manual
8. Batch Control
MENU80
1. AI1 Up Edge
3. AI2 Up Edge
5. AI3 Up Edge
7. AI4 Up Edge
MENU81
Batch controller
M81
Batch controller also named batch controller. PUTF201/201 built -in Flow Batch Value
1000 m³
batch controller input signal by keyboard or analog inlet, output
It will enter batch controller display after adjusting the quantitative value in this window.
MENU82
36
PUTF201/202 series ultrasonic flow meter manual
Date totalizer
This window can check the total net accumulation for last 64 M82
days/months/ and last 5 years. Date Totalizer
0. Day
0. Day
1. Month
2. Year
If display “G”, it means the instrument has gain adjustment, maybe it offline in this day.
If display“H”, it means the instrument has poor signal, maybe it is interfered or has problems in
installation. About details please refer to the chapter of “Error information code”
MENU83 M83
Automatic Correction function Auto. Correction
ON
This function can estimate average flow uncounted during the
offline and add the result to the totalizer. Select “ON” to use this
MENU84
M84
Heat unit selection
Heat Flow Unit
Select “GJ” and “KWh” as heat unit. Default heat unit is KWh. 0. GJ
MENU85
0. From AI1,AI2
1. Fixed Difference
MENU86 M86
Thermal capacity Specific Heat Selection
0. GB
37
PUTF201/202 series ultrasonic flow meter manual
MENU87 M87
Heat Flow Totalizer
Heat Totalizer switch
ON
This window is used for turning on/off the heat totalizer.
MENU88
M88
Heat totalizer multiplier factor
Heat Flow Multi.
Available totalizer multiplier factor is 10 -4 --106(E-4--E6). 4. x1 (E0)
MENU89
MENU90
Signal quality Q used 00~99 to represent, 00 means worst quality, 99 means best. Normally signal
MENU91
38
PUTF201/202 series ultrasonic flow meter manual
and the transducers are properly installed, the ratio value should be in the range of 100±3%.
Otherwise the entered parameters and the transducer installation should be checked 。
MENU92
Menu 21.If this value has an obvious difference with the actual fluid
sound speed, pipe parameters entered and the transducer installation should be checked again.
MENU93 M93
Total time, Delta Time
Transit time and delta time
150.43uS
This window is used for detecting transit time(un it uS) of -10.567nS
Normally volatility of delta time should <20%, if >20%, should check transducer installation
MENU94
M94
Reynolds number
Reynolds Number
This window displays the Reynolds number and the pipe factor that 12234.5 0.92435
MENU+0
M+0
Check power-on/off date and time
ON/OFF Time
View power-on/off time and the flow rate in last 64 times.
Enter When Ready
Enter this window, press ENT to display last 64 times power on/off
time and flow rate by pressing ▲.
M+1
MENU+1
Total Work Hours
Flow meter total working time
00012426:35:45
User can view the instrument total working time after delivering
from factory by this function. As shown at the right figure, this
meter total working time is 12426 hours 35 minutes 45 seconds after delivering from factory.
39
PUTF201/202 series ultrasonic flow meter manual
M+2
MENU+2
Last Off Time
Display last power-off date and time.
Date 15-07-17
Time 15:08:59
MENU+3 M+3
Display last power-off flow rate. Last Flow Rate
0. m³/h
MENU +4 M+4
Power-on/off total times ON/OFF Times
Display the meter total power-on/off times after delivering from 182
factory.
MENU+5
Media sound speed changes M+5
The data in this window is a threshold value. Whenever the Velocity Change
estimated sound speed exceeds this threshold, an alarm signal will 1m/s
be generated and can transmitted to buzzer or OCT or relay. This
function can used to produce an alarm or output when fluid material
changes.
MENU+6 M+6
Compatible communication protocol selection: Protocol Selection
0. Protocol10Adxx 0. Protoco10Adxx
1. Protocol1Addc
012
MENU+7
0.355827 GJ/m3
Water temperature and enthalpy
012
This window displays integer part of temperature in AI1, AI2
0.054057 GJ/m3
J/m3 rotoco/0Adxx
J/m3
5. Operation Introduction
that means the output of electric current loop is out of range 100% which is related to the window
No. 56 Increasing the output value in window NO 56, “E” will disappear, if don‟t use electric
If display “Q” it means frequency is over range 120%, need to set No. 69 window. Increase input
value in the window, “Q” will disappear, if don‟t use frequency o utput, can ignore that.
40
PUTF201/202 series ultrasonic flow meter manual
If display “H” it means that receiving ultrasonic signal is bad. About solution please refer to
If display “G” it means the instrument is gain adjusting automatically before measuring. If it‟s
“I” means the instrument can‟t receive ultrasonic wave signal, check whether probe cable
“J” means the instrument hardware has faults. Some software faults are temporary, can try to
power on again. About the details please refer to the chapter “Error information code”.
Second step, transducer A is connected on the upstream connection of m ain board, transducer B is
Third step, view the flow rate value is “+” or “ -” (“+” not display), if it is “+” that means fluid
flow direction is A→B, if it is “-” that means fluid flow direction is B→A.
Using Method: Press MENU30 and ENT in the bottom of left will display„>‟ it means selection
status, Using ▼or ▲to select something needed, then press ENT to confirm.
display„>‟ it means selection status, Using ▼or ▲to select needed flow unit, then press ENT to
select time unit. Same with flow unit selection, finally press ENT to confirm.
Using No. 32 Window to select volume unit, the method is same with flow rate unit selection,
41
PUTF201/202 series ultrasonic flow meter manual
installation.
factor can‟t be too big otherwise flow display will delay, generally the range is 15~30
seconds.This window is data input type, press No. 40 , input time factor, press ENT to confirm.
value, system will treat the flow as “0”. When real flow is “0”, to avoid false accumulation,
generally set the parameter 0.03m/s. When velocity is bigger than the excision value, excision
value is irrelevant with measuring result, have no effect on the measuring result.
Point. Ensure that the pipe is full of liquid and the flow is absolutely stopped. Then run the function in
window M42 by pressing the MENU 42 keys, then press ENTER key and wait until the counter readings
displayed in the top right corner of the screen goes to “00”; thus, the zero set is completed and the instrument
indicates the results automatically through Window No.01. Repeat zero set calibration if it still needs to be
42
PUTF201/202 series ultrasonic flow meter manual
factory calibration value for each instrument prior to shipment from the factory. Run the function in window
M45 by press the MENU 45 keys, then press ENT key. Re-calibration or change the Scale factor may be
necessary on different pipe lines or different applications in order to obtain better accuracy.
5.13.Password Protection
After the instrument lock can also check all the menus, but cannot modify the parameters. System lock is
used to avoid modification of the parameters. System locks in Menu 47 and be locked with 1-4 digits
password. When unlock the menu, use ▲or▼ and then press ENT key.
If the password of system lock or keypad lock is forgotten, please contact the factory.
5.14.Keypad locked
Keypad lock is used to prevent unauthorized keypad operating and make the LCD displayed a certain menu.
The keypad can be locked with 1-8 digits password, and after keypad locked, all keys are invalid, and the
LCD displays the menu that uses entered before enter Menu 48.
If the password of system lock or keypad lock is forgotten, please contact the factory.
interval time and duration. In M50, select [On] and select the output in order, then press “ENTER”.
5.16.4-20mA Output
Possessing a current loop output exceeding an accuracy of 0.1%, the PUTF205 is programmable and
configurable with multiple output modules such as 4 ~20mA or 0~20mA, selected in Window M54. Select in
Window M54. In Window M55, enter a 4mA flow value. Enter the 20mA flow value in Window M56.
For instance, if the flow range in a specific pipe is 0~1000m3/h, enter 0 in Window M55 and 1000 in
Window M56. If the flow ranges from -1000~0~2000m3/h, configure the 20~4~20mA module by selecting
Window M54 when flow direction is not an issue. Enter -1000 in Window M55 and 2000 in Window M56.
When flow direction is an issue, module 0~4~20mA is available. When the flow direction displays as
negative, the current output is in range of 0~4mA, whereas the 4~20mA is for the positive direction. The
output module options are displayed in Window M54. Enter “-1000” in Window M55 and 2000 in Window
43
PUTF201/202 series ultrasonic flow meter manual
M57.
1-~5V.
indicates the high or low flow rate reading. Users can reset the frequency output as well as flow rate per his
requirements.For instance, if a pipe flow range is 0~3000m3/h, the relative frequency output required is
In Window M67 (high limit frequency output flow value), input 3000;
In Window M65 (Select frequency range), Press ENTER, input Low FO frequency 123, Press ∨, input
If user want to output positive volume by relay output, each pulse equals 0.1m³,set as below.
In Window M32, select volume unit, m³
Notice: Please select the suitable totalizer output. If the output is too large or too small, it will influence the
working life and lose the pulse. Recommended velocity is 1~60 pulse/min.
The sound alarm is working by buzzer. In Window M77, select buzzer. The switch alarm is working by OCT
1) No Signal
2) Poor Signal
3) State error
4) Reverse Flow
44
PUTF201/202 series ultrasonic flow meter manual
7) There are 2 alarms, alarm #1 Low Value and alarm #1 High Value. For example, select “Alarm
#1”, set “Alarm #1 Low Value” in Menu 73, and set “Alarm #1 High Value” in Menu 74. Set “Alarm #2 Low
Value” in Menu 75, and set “Alarm #1 High Value” in Menu 76.
For example 1, the setting method that requires the flow meter to generate sound alarm signal when it does
not enter the normal measuring state is, in Window M77, select the second item.
For example 2, requires the relay alarm output works when the flow rate is over 300~1000m³/h, following is
the step.
For example 3, , requires the OCT alarm output works when the flow rate is over 100~500m³/h, requires the
relay alarm output works when the flow rate is over 600~1000m³/h, following is the step.
5.22.OCT output
OCT output is electrically isolated collector level open circuit output.open -shut condition is
programmable.Users can set on/off condition as one of following: alarm signal or totalizer pulse
output etc. The frequency output by OCT, when frequency works, other functions can‟t be used.
Frequency signal is connected by OCT terminal, please pay attention on position and negative
connection.
45
PUTF201/202 series ultrasonic flow meter manual
5.23.Relay Output
Replay output is programmable. Users can set on/off condition as following: alarm signal or
totalizer pulse output etc.(refer to M79).Relay output is connected by RELAY terminal, also ca n be
adopts RAMTRON calendar clips.Press ENT key under MENU60 once modification is needed.
When display„>‟on the left bottom of the LCD, indicates modification status.Use the “.” to skip
over these digits that do not need modification, then press ENT.
Use window M70 to set up backlight. Press ENT key under MENU70, When display„>‟on the left
bottom of the LCD, indicates modification status.Use ▲ or▼ select the setup.
Use M62 to set up parameters. Baud rate and parity parameters can be configured. Using matched
RS-232C/RS-485 converting device to connect to 485 BUS easily. This converting device is
the zero option “Day”, LCD will display as right box. Top left corner “00” -“63”indicates the serial
number. Middle “00-07-21”indicates calendar. Top right corner “--------” indicates working states.
If the status bar only displays “--------”,indicates normal working.If appears other characters,
please
refer to error code. Right value 3412.53 indicates this specific day net flow.
When view monthly flow,press MENU82 ,select the first option“view monthly”.
46
PUTF201/202 series ultrasonic flow meter manual
When view yearly flow, press MENU82 ,select the first option“view yearly”.
5.29.Timer
Window M72 display flow meter working time since reset the totalizer.
When use analog input as controlling signal, inputting over 2mA current signal indicates “1”
Use M80 to select control input signal, use M78(OCT output) or M79 (Relay output), select “8
batch control”, then an output signal is generated on the OCT or replay output.
Start batch controller by using M81 to set the flow quantitative value. Refer to “ Menu table” M80,
M81.
Because our company test strictly all flow meters before delivery.
Before analog input calibration, must operate hardware debugging window. Operation method as
flowing:
Press MENU-53ENT to calibrate current loop output 4mA, use precision ammeter to measure
Using ▲ or ▼ to modify displayed value and stop modify until the current loop value display 4.00,
indicates finish 4mA calibration. Then press ENT to calibrate current loop output 20mA,
calibration method is same as 4mA. Calibration result will be stored in EEPROM automatically
47
PUTF201/202 series ultrasonic flow meter manual
PIN 1 Empty
1 1
PIN 2 RXD Receive 2 2
3 3
PIN 3 TXD Transmit
4 4
PIN 4 DTS 5 5
6 6
PIN 5 Ground 7 7
8 8
PIN 6 DSR 9 9
PIN 7 +5V
PIN 8 Empty
48
PUTF201/202 series ultrasonic flow meter manual
GB(cr) note9
GSM Message communication orderB note9
DUMPnote10 Print buffer zone content in output system ASCII code format
DUMP0 Remove print buffer zone in system
DUMP1 Output print buffer zone content in system(24K) ASCII code format
49
PUTF201/202 series ultrasonic flow meter manual
Note:
0.(cr) means enter,ASCII code value is 0DH. (lf) means newline, ASCII code value is
0AH.
3.1~6 letter means machine‟s status, please refer to error code chapter about the character
8.If there are several flow meters in data network then basic command can‟t operate
separately. Need to add N or W prefix, otherwise all flow meters answer the orders at same
9.Check flow meter parameters by phone messages through GSM module connection.
P can be added before every basic command, it means pass back data has CRC calibration.
(2BH+31H+32H+33H+34H+35H+ 36H+37H+45H+2BH+30H+6DH+33H+20H=(2)F7H)
N Prefix
50
PUTF201/202 series ultrasonic flow meter manual
E.g. If users want to check the velocity of No. 88 flow meter, send order NXDV(CR), The
W Prefix
The method of W: W+ number string address code + basic order, number string range is
0~65535 except for 13(ODH enter), 10(OAH newline), 10(OAH newline), 42(24H*), 38
(26H&).If users want to check 12345 flow meters‟ velocity, send command W12345DV(CR),
& function symbol can add more than 6 basic commands to constitute recombination
command and transmit to flow meter, flow meter answers at same time. E.g. Requiring to pass
back the No.4321 flow meter 1. flow rate, 2. velocity, 3. positive accumulation, 4.thermal
accumulation 5. AI1 analog input current value 6. AI2 analog input value and calibration,
W4321PDQD&PDV&PDI+&PDIE&PBA1&PAI2(CR)
+0.000000E+00m3/d!AC(CR)
+0.000000E+00m/s!88(CR)
+1234567E+0m3 !F7(CR)
+0.000000E+0GJ!DA(CR)
+7.838879E+00mA!59
+3.911033E+01!8E(CR)
same time. Select the compatible protocol 1 or compatible protocol 2 by the menu M+7.
2.Compatible protocol 1 and compatible protocol 2 are compatible with other factory‟s flow meter.
Please do not use the protocol 2 unless replacing.
3. The difference of compatible protocol 1 and compatible protocol 2 is the first one is single
Answer and return message format(ASCII): 8 bytes flow rate (the last two digits are decimals) +12
51
PUTF201/202 series ultrasonic flow meter manual
bytes volume (the last two digits are decimals)+4 bytes signal strength(the last o ne digit is
decimal)+4 bytes signal quality value(the last one digit is decimal)+2 bytes calibration (refer to
4. Calibrating (none)
5.1.3 Command from master to salve, totally 7 formats (0 -6 transfer single byte ASCII)
5.1.3.5 Command 4: obtain salve‟s signal strength and signal quality value
52
PUTF201/202 series ultrasonic flow meter manual
Command: 30
6.1.1 Command+symbol: total 2 bytes, the first byte is the command 0 from master. The second
6.1.2 Flow rate: total 8 bytes (ASCII), the seventh and eighth bytes are decimals.
6.1.3 Volume: total 12 bytes (ASCII), the eleventh and twelfth bytes are decimals.
6.1.4 Check bit: total 2 bytes (ASCII) is the cumulative sum of the formal 22 bytes(decimal
addition), as CRC.
6.1.5 For example, master command is 0, the flow rate symbol is positive, flow rate value is
6.2.1 Command+symbol: total 2 bytes, the first byte is the command 0 from master. The second
6.2.2 Flow rate: total 8 bytes (ASCII), the fifth, sixth, seventh and eighth digits are decimals.
6.2.3 Volume: total 12 bytes (ASCII), the eleventh and twelfth bytes are decimals.
6.2.4 Check bit: total 2 bytes (ASCII) is the cumulative sum of the formal 22 bytes, as CRC. 6.2.5
For example, master command 1 is , the flow rate symbol is positive, flow rate value is
53
PUTF201/202 series ultrasonic flow meter manual
6.3.1 Command+symbol: total 2 bytes, the first byte is the command 2 from master. The second
byte is meaningless.
6.3.2 Positive volume: total 12 bytes (ASCII), the eleventh and twelfth bytes are decim als.
6.3.3 Negative volume: total 12 bytes (ASCII), the eleventh and twelfth bytes are decimals.
6.3.4 Check bit: total 2 bytes (ASCII) is the cumulative sum of the formal 26 bytes, as CRC.
6.3.5 For example, master command 2 is, positive volume is 14368.59 m³. It should transmit 32 ,
30,30,30,30,30,30,31,34,33,36,38,35,39,30,30,30,30,30,31,36,
6.4.1 Command+symbol: total 2 bytes, the first byte is the command 3 from master. The second
byte is meaningless.
6.4.3 Check bit: total 2 bytes (ASCII) is the cumulative sum of the formal 10 bytes, as CRC.
6.4.4 For example, master command 3 is, total time is 4368 min. It should transmit 33,30,30,
6.5.1 Command+symbol: total 2 bytes, the first byte is the command 4 from master. The second
byte is meaningless.
6.5.2 Signal strength and quality value: total 8 byte(ASCII), Signal strength is 4 bytes, the last
byte is decimals.
6.5.3 Check bit: total 2 byte (ASCII) is the cumulative sum of the formal 10 bytes, as CRC.
6.5.4 E.g. Master command is 4, strength value is 88.9, quality value is 17.8. it should
transmit 34,30,30,38,38,39,30,31,37,38,34,35.
54
PUTF201/202 series ultrasonic flow meter manual
meaningless. 2 32H 50 2
3 33H 51 3
6.6.2 Check bit: total 2 bytes(ASCII code)is the
4 34H 52 4
cumulative sum of the formal 2 bytes, as CRC. 5 35H 53 5
6.7.1 6.7.2
6.7.1 Command + symbol, total 2 bytes, the first byte is the command 6 from master. The second
byte is meaningless.
6.7.2 Check bit: total 2 bytes(ASCII code)is the cumulative sum of the formal 2 bytes, as CRC.
6.7.3 This master command is used for deleting slave total time.
Key value coding is used for simulating key on upper computer when conne ction. E.g. input
command“M1”by serial port, same as the key 1 on ultrasonic flow meter keyboard,
MSCOMM1.INPUT=”dqs”+vbcrlf;
55
PUTF201/202 series ultrasonic flow meter manual
2. Use VB to send command, requiring pass back No.4321 flow meter 1. flow rate, 2. velocity, 3.
positive volume, 4.thermal volume 5. AI1 analog input current value 6. AI2 analog input value and
MSCOMM1.INPUT=“W4321PDQD&PDV&PDI+&PDIE&PBA1&PAI2”+VBCRLF;
3. Use VB to send the command to change and set pipe outer diameter(M11 window) as 345mm.
MSCOMM1.INPUT=”M<”+VBCRLF+”M1”+VBCRLF+”M1”+VBCRLF+”M3”+VBCRLF+”M4”
+VBCRLF+”M5”+VBCRLF+”M=”+VBCRLF
In above formula, “M<”means MENU key,“M=”means ENT key,“M1” means “1” key
Blue:RS232 RXD(RS485B)
Register address(hexadecimal)
Register address Data length Description
(40001) 2 Return every-day flow rate
(40003) 4 Return every-hour flow rate
(40005) 6 Return every-minute flow rate
(40007) 8 Return every-second flow rate
(40009) 10 Return velocity
(40011) 12 Return positive volume
(40013) 14 Return negative volume
(40015) 16 Return net volume
(40017) 18 Return thermal volume
(40019) 20 Return every-second thermal flow rate
(40023) 24 Return T1 current value
(40025) 26 Return T2 current value
(40031) 32 Return analog input T1 value
(40033) 34 Return analog input T2 value
(40041) 42 Return electronic serial number
56
PUTF201/202 series ultrasonic flow meter manual
Calibration
Meter Address: One byte, expressed in Hex, the value in Menu 46. For example, M46 shows 22,
Register address of data start: Two bytes, expressed in Hex, and high bit is before low bit. For
example, read data from Flow Velocity, Start Register Address is “00 08”.
Length of Register Address: Two bytes, expressed in Hex, and high bit is before low bit. For
example, read data from Flow rate per day to Net totalizer, the length is “00 10”.
CRC Code: Two bytes, expressed in Hex, and high bit is before low bit.
Velocity 01 03 00 08 00 02 45 C9
Positive volume 01 03 00 0A 00 02 E4 09
Negative volume 01 03 00 0C 00 02 04 08
Net volume 01 03 00 0E 00 02 A5 C8
Meter address, function code, register address, read data area, CRC calibration
Meter Address: One byte, expressed in Hex, the value in Menu 46. For example, M46 is 22, the
57
PUTF201/202 series ultrasonic flow meter manual
Bytes Length of Returning Data Section: One byte, expressed in Hex, every data return 4 bytes.
For example, read every-day and every-hour flow rate, read data byte length is 00 04, return data
Read data area: indicates in Hex, every 4 bytes indicate one data, high byte is front, low byte is
behind.
CRC Code: Two bytes, indicates in Hex, high byte is front, low byte is behind.
58
PUTF201/202 series ultrasonic flow meter manual
7.1. Introduction
PUTF205 serial ultrasonic flow meter(except handheld) has built -in thermal energy measurement
function. The flow meter can automatically calculate fluid energy, display energy flow rate and net
calculate by heat meter according to international standard. M05 Display energy flow rate,net flow
rate。
M85 Temperature signal source selection, input T1 、T2, fixed temperature difference.
Menu 05 displays energy flow rate ,net energy volume. Menu 06 displ ays T1,T2 temperature
value.
59
PUTF201/202 series ultrasonic flow meter manual
When temperature sensor output 1000Ω, temperature is 0℃, and output 1758.56Ω, temperature is
200℃.
User needs to input temperature range 0 -200℃ in menu M63 and M64.
1) Clamp-on temperature sensor is mounted out of pipe and installed in any position. Insertion temperature
2) It is benefit for measuring accuracy, safe and reliable and easily repaired.
3) Insert the sensor to the central position(vertical mount or tilt mount). If the measuring pipe diameter is
4) Temperature sensor cable is provided by manufacturer ,standard cable length is 1.5m, optional 10m.
Take notice of the pipe condition after selecting clamp -on temperature sensor mounting
point.Mounting point pipe must keep clean before clamp the temperature sensor on the surface of
Insertion
Insertion temperature sensor accuracy is higher than clamp -on as the insertion temperature sensor
is directly contacted with measuring media. There are two methods to mount insertion temperature
For weldable pipe(e.g.carbon steel,PVC etc.),weld the ball valve directly on the surface of the
pipe. Do not contain air holes in case of leakage or fracture while welding(clean the pipe surface
For pipes can‟t be directly welded(e.g.iron steel cast,cement etc.),customized pipe hoops are
required(with sealed gasket). Ball valve base are welded on the hoops in advance, tighten the
60
PUTF201/202 series ultrasonic flow meter manual
hoops on the pipe. Please compress the sealing gasket tightly to avoid leaking.
Connect driller sealed gasket with customized ball valve thread, open ball valve aft er tightening.
Push the driller until it touches the surface of pipe. Lock it, switch on the power, start drilling.
Keep low speed, do not fast, slowly drilling in case of drill bit stuck or broken. After opened, push
out the drill rod until the front of drill bit close to the ball valve center. Close ball valve, unload
driller(it‟s better to open a little ball valve,flush away the scarp iron for easy installation of
61
PUTF201/202 series ultrasonic flow meter manual
8. Error Code
Error code reasons and solutions
* No signal received
* Make sure the transducer is close to the
pipe. Use sufficient couplant.
* A poor connection between transducer
* Make sure no rust, stain, oil paint on
and pipe, or too little couplant applied.
the pipe surface. Use flat file to clean the
pipe surface.
*I *The transducer has not been installed
Signal not detected * Check the original settings.
properly.
*Clean the scaling or change the pipe.
Normally, try to change the measuring
* Too much scale formation in the inside
point.
pipe.
* Wait until the liner is saturated and
solidified.
* New Liner
*H * low signal
Low signal strength * The same as above.
* same as above
*H * poor signal
Poor signal quality * The same as above.
* same as above
Current loop value>20mA
* Recheck the setting(see M56 in the
*E ( Ignore it if measuring * 4-20mA current loop output over 100%
manual)or make sure the actual flow is
process don‟t use current * Wrong current loop output setting
too large.
loop)
Frequency output is higher
than setting * Frequency output over 120%, * Recheck the frequency output setting
*Q (Ignore it if measuring * Wrong frequency output setting or (see M66- M69 menu)or make sure
process don‟t use frequency actual flow is too large. actual flow is too large.
output)
* Restart the device and observe the
* Self-checking
*F information displayed on the screen. If
See table 1
the problem persists, contact us.
* Perpetual hardware fault
* Contact us.
Adjusting Gain >S1 *These 4 steps mean that the machine is
going through the gain adjustment
Adjusting Gain >S2
process, preparing for the normal
Adjusting Gain >S3
*G measurement.
Adjusting Gain >S4
Displayed in *If the machine stops on S1 and S2 or
M00,M01,M02,M03 shift between S1 and S2, it means that the
received signal is too low or the waveform
window)
is not so good.
*K Empty pipe, M29 menu If there is fluid in the pipe, input “0”in
No fluid in the pipe or wrong setting
setting the M29 menu.
Remark:Error code *Q,*E doesn‟t influence measurement, only indicate current loop and
62
PUTF201/202 series ultrasonic flow meter manual
63
PUTF201/202 series ultrasonic flow meter manual
64
PUTF201/202 series ultrasonic flow meter manual
65
PUTF201/202 series ultrasonic flow meter manual
Web:www.panda.sh.cn
Tel:+86-21-59863888
Fax:+86-21-59867777