Title:
Calibration Procedure of
Microwave Digestion System
Revision:
Prepared on 01/27/2011
Copy no: 1
Prepared by:
Aida Wong Assistance Manager Signature: Date: 01/27/2011
Approved by:
Karen Chung Division Manager Signature: Date: 01/27/2011
/conversion/tmp/activity_task_scratch/716791244.doc
Calibration Procedure
AMENDMENT RECORD
Page 2 of 4
Calibration Procedure
1. Calibration Interval
1.1 Energy Test, Turntable rotation rate, Temperature sensor and Time check - Every 3 months.
2. Reference Document
2.1 ETHOS 1 User Manual
2.2 Tips & Techniques for Milestone Microwave Lab Stations
2.3 USEPA 3051 – Microwave Assisted Acid Digestion of Sediments, Sludges, Soils and Oils
2.4 USEPA 3052 – Microwave Assisted Acid Digestion of Siliceous and Organically Based
Matrices
3. Apparatus
3.1 Reference thermocouple
3.2 Beaker
3.3 Timer
3.4 General lab glassware
4. Calibration
4.1 Energy Test – to check the microwave power emission.
4.1.1 Turn on the microwave digestion system and warm up about 15 minutes.
4.1.2 Use 1 L of D.I. water under 12 ℃ heating up for 60 seconds at 1000 W.
4.1.3 Measure the temperatures before (Ti) and after (Tf) heating up and calculate the power
output.
4.1.4 Refer to page 113, ETHOS 1 User Manual for the details of Energy Test.
4.1.5 Repeat the test twice and the limit of acceptance is ± 15 %.
Page 3 of 4
Calibration Procedure
4.4.3 Start the calibration timer (3.3) together when the program starts.
4.4.4 Record the time when the program stops.
4.4.5 Repeat the step 4.4.1 – 4.4.4 twice. Obtain an average value.
4.4.6 The acceptance criteria of the average value is ±1 second of the desired time (1800 s)
- End -