※The device is being used for X-ray purpose. Point 200HD


PNX.UM.033.0(C) (2013.04.18) CONTENTS

CONTENTS
1. OVERVIEW ............................................................................................................................................. 3

2. FEAUTRES AND PURPOSE .................................................................................................................. 4

1) FEATURES .........................................................................................................................................................................4

2) PERPOSE ...........................................................................................................................................................................5

3. STRUCTURE............................................................................................................................................ 5

4. SAFETY .................................................................................................................................................... 6

5. PRECAUTIONS ....................................................................................................................................... 7

6. SPECIFICATIONS ................................................................................................................................. 11

7. COMPOSITION .................................................................................................................................... 14

1) CT & PANORAMA ..................................................................................................................................................... 14

2) CONTROL PANEL ....................................................................................................................................................... 16

3) MEASURE CONDITION ............................................................................................................................................ 18

8. HOW TO USE....................................................................................................................................... 21

9. MAINTENANCE ................................................................................................................................... 39

1) USE, STORAGE, MANAGEMENT .......................................................................................................................... 39

2) MAINTENANCE, WARRANTY ................................................................................................................................ 40

3) WASTE/DELIVER ......................................................................................................................................................... 42

10. INFORMATION.................................................................................................................................. 43

User’s Manual Point 200HD 2 / 46


PNX.UM.033.0(C) (2013.04.18) 1. OVERVIEW
PointNix Co., Ltd provides this printed User Manual as a guide for the
operation of Point 200HD, the High Definition HD Digital Panorama
manufactured by PointNix Co., Ltd. This document is designed to help
the users for better understanding of the equipment in order to obtain
satisfactory results from the use of the equipment.

Point Point 200HD, the High Definition HD Digital Panorama, is


representing the latest technology in X-ray Digital Panorama field. Point
Point 200HD features superior imaging quality, and is designed to
provide the X-ray professional with an intuitive, easy-to-use interface. It
converts X-ray information directly to digital signal to acquire high
definition images for X-ray purposes. It is consisted of X-Ray tube, High
Voltage Power supply, Panorama detector assembly, Head Position,
Main frame, Control Panel, Image-taking stand, Base plate , Column,
Imaging controlling device(PC, Monitor), and CDX-View Software(Ver.
1.93).

Point Point 200HD is an outcome of high technology that reflects the


next generation’s needs for X-ray Digital Panorama. The use of new
material called Cadmium Telluride (CdTe) to the Panorama sensor makes
this X-ray equipment acquire clear and accurate images without having
image distortion problems. Point Point 200HD maintains a low radiation
dosage due to its high absorptance capacity.

User’s Manual Point 200HD 3 / 46


PNX.UM.032.0(C) (2013.03.20) 3. STRUCTURE

1) Features

① High resolution HD sensor

② Secured X-ray exposure

③ Clear and accurate images due to CsI sensor

④ Low radiation dosage due to high absorptance

⑤ Scanning and receiving images at the same time

⑥ Centralized membrane switch

⑦ Integrated software program, CDX-View


◆ CDX-View: This is a tool of digital diagnosis system which implements

all images from Panorama, Cephalo, CT, Intra Oral Camera,


Digital Camera, and Intra Oral Sensor etc. This is
represented as an integrated data system of X-ray clinics
on Network use.

2) Purpose
The equipment is to acquire tomographic images of teeth, jaw, oral structure
and panoramic etc. by transforming x-ray exposed signals to digital signals for
diagnosis purposes.

User’s Manual Point 200HD 4 / 46


PNX.UM.032.0(C) (2013.03.20) 3. STRUCTURE

■ Dimension : 870(W) X 1020(L) X 1,620(2,260) mm

■ Weight : 150 kg

User’s Manual Point 200HD 5 / 46


PNX.UM.032.0(C) (2013.04.15) 4. SAFETY

1) Symbols

Symbols Description
Direct risk of death, serious adverse reaction,
serious damage to the properties, or fire.

Indirect risk of death, serious adverse reaction,


serious damage to the properties, or fire.

Indicate the dangers of risk of serious and slight


injuries, partial damages to the properties or data
loss.

Warning, Electrical shock

Radiography exposure indicator

Indication mark for protective earth

Type B applied part

Indication mark for radiate position beam

User’s Manual Point 200HD 6 / 46


PNX.UM.032.0(E) (2013.04.15) 5. PRECAUTIONS

1) Direction for the use

Please read carefully to prevent misoperation due to electricity


damage and electromagnetic waves.
(1) Be sure to pull the plug out the device before cleaning the device or
not using for long time.
(2) Always keep clean the device by wiping with soft and dry towel.
(3) Maintain clean and dry around X-ray room and around the device.
(4) Caution for fire from a sparkler to wood piece, sawdust, dissolve
material.
(5) Do not place electric plug near hot machine or equipment.
(6) This device should be used at facility where protection equipment
against X-ray is provided.
(7) Modification by unauthorized person is prohibited. Only trained,
qualified and authorized engineers/technicians are allowed to
service this device.
(8) Staff must use device in mentioned order at manual and be well-
informed of the content of manual for patient’s safety.
(9) Before use, check the device’s appearance, connected PC,
connected cable and etc.
(10) Staff must be able to communicate with patient until scanning is
finished.
(11) Turn off the power when device is not using for more than 6 hours.
(12) Do not put any object over the device during the usage.
(13) Only qualified engineer is able to inspect and maintenance/repair of
the device.

User’s Manual Point 200HD 7 / 46


PNX.UM.032.0(E) (2013.04.15) 5. PRECAUTIONS
(14) Please use humidity-resistance plug when the device is using
outside. This outlet for outside have to have Ground Fault
Interrupter (G.F.I) function and this type of circuit breaker prevents
user’s electric shock.
(15) Extension code must use which is grounded and use when it is
necessary.
(16) Caution for deterioration in quality and damage of product due to the
usage of near the high intensity electric and magnetic field.
(17) We recommend taking 6 minutes of rest between scanning.
(18) Please contact agent or head office if an error occurs during the
device usage.

This product is safe product which passed Certification Test of Electro


Magnetic Interference and Tolerance

2) Safety Regulations

(1) Keep the device away from water or humidity.


(2) Make sure patient is not moving while scanning.
(3) If there is a doubt the oil is leaking from X-Ray tube, please turn off
the device immediately and contact to agent or head office.
(4) Always equip fire distinguisher and staff must well-informed of its
use.
(5) PC Hardware and the device must be installed at least 2 meter
away from each other to avoid unnecessary contact of patient to PC
Hardware while operating.

User’s Manual Point 200HD 8 / 46


PNX.UM.032.0(E) (2013.04.15) 5. PRECAUTIONS
(6) When you clean the device, make sure to turn off the device and
disconnect all cables between the device and PC. Be careful not to
spill anything into the device.

3) Danger

(1) This device’s position beam can cause impaired vision. Do not aim
position beam direct to patient’s eyes and guide patient not to look at
the position beam.
(2) Do not put anything in radius of the device movement. That can
cause a danger or damage of device during scanning.
(3) Do not place flammable materials or volatile materials near the
device.
(4) Stop the device by pressing emergency switch when it occurs
problem during scanning.

4) X-ray Protective Policy

(1) Staff must obligatorily provide protection material such as lead apron
to patient when they are scanning image.
(2) Frequent exposed to X-Ray can cause cancer and various diseased.
So staff must protect themselves by lead apron or lead protect.
(3) Staff must keep the distance at least 2meters (7ft.) from focus and
X-Ray to hand exposure switch.
(4) The device must locate inside of shield room and staff should keep
eyes on scanning through outside window.

User’s Manual Point 200HD 9 / 46


PNX.UM.032.0(E) (2013.04.15) 5. PRECAUTIONS
(5) Person who needs special attention such as children, pregnant
woman, elderly and unhealthy patient must counsel with medical
attendant before scanning.
(6) During the scanning, staff must keep check the device and patient for
in case of emergency.
(7) During the scanning, stop the device immediately by pressing
emergency button if it works unusual.
(8) Patients and staffs are prohibited long-time expose to X-Ray and take
enough rest because excessive use of X-Ray can cause various
diseases.
(9) After scanning, it should wait 6minutes of cool time for next scanning.

The law relate to Radiation Safety and Protect is different for each
Country. User must observe the Law of Radiation Safety and Protect for
their own country.

User’s Manual Point 200HD 10 / 46


PNX.UM.032.0(E) (2013.04.15) 6. SPECIFICATIONS

No. Name Technical Specifications


■ X-ray Beam: Cone Beam
■ kV: 60~90kV
■ mA: 4~15mA
■ Focal spot: 0.5 X 0.5 mm
■ Target angle: 5 °
X-ray Tube ■ Inherent filtration: 0.8mmAl or higher
D-054SB
(equivalent at 50kV)
(Option1)
■ Cooling method: oil immersed(Max. 60℃) and
convection oil cooling
■ Anode heat storage capacity: 35kJ
■ Mas anode heat dissipation: 250W
■ Exposure time: Max. 20sec
1
■ X-ray Beam: Cone Beam
■ kV: 60~90kV
■ mA: 4~10mA
■ Focal spot: 0.5 X 0.5 mm
■ Target angle: 5 °
X-ray Tube ■ Inherent filtration: 0.8mmAl or higher
PDM90P
(equivalent at 50kV)
(Option2)
■ Cooling method: oil immersed(Max. 60℃) and
convection oil cooling
■ Anode heat storage capacity: 35kJ
■ Mas anode heat dissipation: 250W
■ Exposure time: Max. 20sec

User’s Manual Point 200HD 11 / 46


PNX.UM.032.0(E) (2013.04.15) 6. SPECIFICATIONS

No. Name Technical Specifications


■ Sensor Material : Direct Emission Super AMOLED
CdTe, 750 ㎛ thick
■ Active Area(mm) : 150 x 4.8
■ Pixel number of active area : 3125 x 100
■ Pixel size : 48 ㎛
Panorama ■ Resolution :10.4 lp/mm
2
Detector ■ Dynamic Range(dB) : ≥72
■ Frame Rate : 200fps
■ Data Bits per pixel : 12bit
■ Data I/O: GigE, Optical Link, or Camera Link
■ Command I/O : 3 TTL triggers
■ Power Supply Requirement : D +5V/1A, A +5V/1A

■ Version: Ver. 1.93


■ OS: Windows XP or higher
■ CPU: Dual Core Pentium 2.4 GHz or higher
■ Supported GPU: NVidia GPU 1GB or higher
CDX-
3 Soft ware ■ RAM : 3G or higher
VIEW
■ HDD : 500GB free hard disk space or higher
■ Network : Gigabit Ethernet Network card
■ Display : 32bit color display or higher
■ Supported resolution : 1,280 X 1,024 or higher

■ OS : Windows XP or higher
■ CPU : Intel Core i5-2400 or higher
Imaging ■ GPU : NVidia GPU 1GB or higher
control ■ RAM : 3G or higher
4 PC
device ■ HDD : 500GB free hard disk space or higher
(Option) ■ Network : Gigabit Ethernet Network card
■ Display : 32bit color display or higher
■ Resolutions : 1,280 X 1,024 or higher

User’s Manual Point 200HD 12 / 46


PNX.UM.032.0(E) (2013.04.15) 6. SPECIFICATIONS

No. Name Technical Specifications


■ OS : Windows XP or higher
■ CPU : Intel Core i5-2400 or higher
Imaging ■ GPU : NVidia GPU 1GB or higher
control ■ RAM : 3G or higher
5 PC
device ■ HDD : 500GB free hard disk space or higher
(Option) ■ Network : Gigabit Ethernet Network card
■ Display : 32bit color display or higher
■ Resolutions : 1,280 X 1,024 or higher

■ Screen size : True color 19 inch


■ Viewing angle : 170° Horizontal, 170° Vertical
Imaging ■ Brightness : up to 250 nits (cd/m2)
control
6 Monitor ■ Signal input : VGA, DVI-I
device
■ Contrast rate : 500 : 1
(Option)
■ Recommended resolution : 1,280 X 1,024 or higher
■ Height adjustment : Vertical tilt range of -5°~+25°

User’s Manual Point 200HD 13 / 46


PNX.UM.032.0(E) (2013.04.15) 7. COMPOSITION

1) Panorama

8
3
9
4 10
11 0
5 0

12
0
13
0
6

User’s Manual Point 200HD 14 / 46


PNX.UM.032.0(E) (2013.04.15) 7. COMPOSITION

No. Name Functions


1 Main Frame Rotation driving part of the rotation arm

2 Rotation Unit Rotation part following to the teeth locus while scanning

3 Detector part Image detector included (detachable)

4 Mirror Patient’s position fixing mirror


5 Control panel Operating control panel

6 Column 2 Support Column 3

7 Base plate Support and fix the whole body of the device

8 Column 3 Support Main frame part (UP/DOWN column)

9 X-Ray unit Contains the X-ray tube refrigerated by oil

10 Bite position Help patient’s tooth stable

11 Chin position Help patient’s chin stable

12 Image filming Support Control panel, Position, Patient’s handle


stand
Patient’s holding are to be properly positioned for
13 Handle
scanning

Emergency switch button- Stop device automatically by pressing


switch when unexpected danger situation

User’s Manual Point 200HD 15 / 46


PNX.UM.032.0(E) (2013.04.15) 7. COMPOSITION

2) Control Panel

No. Name Functions


Basic mode when power is on and also use this button to
ⓐ Panorama mode
select the Panorama mode

ⓑ TMJ mode Select the TMJ mode

ⓒ Cephalo mode Select the Cephalo mode

ⓓ CT mode Select the CT mode

User’s Manual Point 200HD 16 / 46


PNX.UM.032.0(E) (2013.04.15) 7. COMPOSITION

No. Name Functions


Preparation process before taking an image.
Fix the patient’s position. Rotation part, detector part and
ⓔ READY
collimator part are moving depends on scanning mode.
(LED is on when it is ready to scan)

ⓕ X-ray lamp Indicator the means images are on the process of taking

ⓖ STOP lamp Indicator the means image taking is completed

Help to adjust the patient’s horizontal and vertical


positions.
ⓗ Position beam
(Automatically turned off in 30 seconds, and it is not
operating while scanning)

kV control key-Adjust kV based on patient and indicate


ⓘ kV
error mark

mA control key-Adjust mA based on patient and indicate


ⓙ mA
error mark

ⓚ Time Adjust image taking time and indicate error mark

Adjust numbers of ⓘⓙ and ⓚ.



Up, Down Move column to upward or downward. (Possible to

operate also in Ready condition.)

Import pre-set value of kV and mA for each ⓘ, ⓙ and


ⓝ ⓚ button which are set based on each patient’s info
ⓞ Memory mode body type(ⓝ-Children, ⓞ-Woman, ⓟ-Man) information.
ⓟ (Each button contains 2 types of memory function, and
can change function by one press.)

ⓠ Image taking Move rotation arm back or forth while image scanning
ⓡ location of
(ⓠ-Protrusion, ⓡ-Normal, ⓢ-Depression)
ⓢ rotation unit

User’s Manual Point 200HD 17 / 46


PNX.UM.032.0(E) (2013.04.15) 7. COMPOSITION

3) Measure Condition

User’s Manual Point 200HD 18 / 46


PNX.UM.032.0(E) (2013.04.15) 7. COMPOSITION

SID location

Condition SID(mm)

PA , Tmj 505

Condition SCAN TIME

PANORAMA 17s

Tmj 11s

Condition Rotation Angle

PANORAMA
270°
Tmj

Data combination

This is a high-voltage generator of Intermittent


load type, it is the maximum available tube 90kV 10mA
voltage, when X-ray is exposing.
This is a high-voltage generator of Intermittent
load type, it is the maximum available tube 65kV 15mA / 90kV 10mA
current, when X-ray is exposing.
This is a high-voltage generator of Intermittent
load type, it is the maximum available electric 65kV 15mA / 90kV 10mA
power, when X-ray is exposing.

User’s Manual Point 200HD 19 / 46


PNX.UM.032.0(E) (2013.04.15) 7. COMPOSITION

Continuous operation of intermittent load


MODE kA mA SEC
Man 70 10 17
Pano Woman 66 8 17
Child 66 5 17
Man 70 10 11
TMJ Woman 66 8 11
Child 66 5 11

User’s Manual Point 200HD 20 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

1) Preparations before use


(1) Turn the power switch on and connect the power supply of X-Ray
control device. Be sure to check the normal functioning of X-Ray
exposure part.
(2) At least 30 minutes of warming up time is required before start
taking images.
(3) Read this manual carefully to fully notice the instructions for correct
usage.
(4) Make sure the normal functioning of this device.
(5) Ground connection (earth) is requested to be set correctly.
(6) Users are expected to wear a protective lead apron to protect them
from X-Ray exposure.
(7) Place a cover for measurement accessories

2) How to use the PANORAMA

(1) Turn the power switch ON.


(Rotation area is moving back to the normal location.)

Power switch

User’s Manual Point 200HD 21 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

(2) Select the MODE (Panorama, TMJ) upon your needs.


(Once power is ON, it is basically set in Ready which means
preparation for scanning is completed. After a scanning, we assigned
30 seconds to set Ready automatically, and if you want to cancel this,
press Ready button again.)

Check if Ready LED is ON


Panorama MODE

TMJ MODE

(3) Adjust values of kV and mA.


Once adjustment is completed, buttons are using for
UP/DOWN movement of the column.

 Check if Ready LED is ON :


Buttons are using for adjustment
kV, mA
 Ready LED is off :
Buttons are using for UP/DOWN
Movement of the column

① Press kV or mA buttons.

② Use buttons to adjust values.

User’s Manual Point 200HD 22 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

③ Press kV or mA button.
< -Child>, < -Woman>, < -Man>Memory function
Memory function uses pre-saved kV and mA values so additional
information change is not required for another scanning process.
Select pre-saved kV and mA. These values are selected and
saved after considering patient’s type.

(4) Press button.


(5) Walk the patient to the device to the point where chinrest base is
located. Patient’s chin should be positioned correctly on the chinrest
base.

Chinrest base

(6) Adjust height of column with button


(7) Locate the patient and place chin gently to Chin Position. Then
adjust Head Position’s screw to make patient’s face verticality.

Head Position
Movable for 30mm

Chin Position

(8) Press button to set the laser beam to be located in the middle.

Laser

User’s Manual Point 200HD 23 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

(9) Adjust column one more time.


(10) Instruct the patient to hold the bite guide between maxillary central
incisor and mandibular central incisor

Bite

(11) Grasp the handle tightly and lean back at an angle of 10 degrees.
(Feet are located half step ahead.)

(12) Use buttons in Control Panel such as ① -Protrusion, ② -

Normal, ③ -Depression. Scanning location of rotation arm is


changed depends on the type of sinus.

① ② ③

(13) Press button in CDX-VIEW program and start scanning


after pressing Hand switch.
(14) Analyze, save and print the acquired images.
(15) Shut down the program and disconnect the power supply. All
indicators on Control Touch Panel will be turned off.

User’s Manual Point 200HD 24 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

3) How to use the Program

(1) New patient registration


When the user runs CDX-VIEW program, the user can get the same
window as (Picture 1). Furthermore, a new patient named with
“Nobody” will be selected without any acquired images. For
registering new patient, press button (②). Then, the user can get
the same window as (Picture 2).Among the patient’s information,
name and identified chart number must be recorded. Brief medical
records can be registered on “Diagnosis”, and if patient’s
information is needed to be edited, press button (③) or selects[edit]-
[patient information].

<Picture 1> <Picture 2>

User’s Manual Point 200HD 25 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE
(2) Scanning the images

① Panorama Scanning
From the CDX-View <Picture #1>,, select Panorama images and
then press “Acquire” button(①), or click double click any empty
space ( , ②), the screen will be showed up as <Picture2> and the
user can take a new Panorama image. Also the user can save an
acquired image as <Picture3>.

<Picture1>

When the user scans a new image, the screen will be showing up
as<Picture 3>. In the case of TMJ mode, expect for an acquired
image, the window will be the same as Panorama.

User’s Manual Point 200HD 26 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

<Picture 3-1: Acquired image processing>


A. Original: The user can get the initial or original images
Before applying high defined image function.
B. High definition: After selecting acquired image, values for
sensitivity are adjustable for highly defined images
compare to originals. Also, saving the changed
images is available.
C. Close: When user selects close function, the current
Acquired images are not saved and screen becomes to
original window.
D. Save and Close: Save the currently acquired image and go
back to the initial screen. After saving the image,
main screen will show the latest acquired image.
The previous screen remains at previous image
screen.

User’s Manual Point 200HD 27 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

<Picture 3-2: Saving an acquired image>

User’s Manual Point 200HD 28 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

② TMJ (Temporo-Mandibular Joint)Scanning

A. From the CDX-View program, as <Picture 1> click the


“Panorama image” tab, and then select the Panorama selecting
window.

<Picture 1>
B. When user clicks mouse’s right button, the screen will beturned
up as <Picture 2>. At this stage, click “Acquiringnew images
(TMJ mode)”.

<Picture 2>

User’s Manual Point 200HD 29 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

C. When the program runs just as the following <Picture 3>, press
hand switch for scanning. TMJ scanning will be completed by
scanning patient twice. For the first scanning, ask patient to
close his/her mouth, and ask patient to open his/her mouth for
the second scanning.

<Picture 3: Left:Direct Emission Sensor>

User’s Manual Point 200HD 30 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

D. When all the scanning process is finished, selects “Save and


Close” for saving patient’s data.

E. After the scanning process, the user can get the window same
as the <Picture 4>.

<Picture 4>

User’s Manual Point 200HD 31 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

③ Image processing-Panorama
(Same methods for both Panorama)

A. The user can access to the image processing as <Picture 1>,


either double clicking the specific pictures (①), or selects
[Image]-[Image processing] from the menu bar (②).

<Picture 1>
B. Displaying the image

User’s Manual Point 200HD 32 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE
Either displays the image information as <Picture 2> or shows
Preview (①-1), FMS (①-2), and History Window (①-3).

<Picture 2: Image information and each supporting windows>


■ Preview Window:
Shows the location for the current image.When the image is
displayed by one screen, selects the total area.
■ FMS Window:
Shows other images of the current patient.(Intra oral camera,
panorama etc.)
■ History Window:
Shows previous images of currently selected image on FMS
window.

User’s Manual Point 200HD 33 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

C. Brightness adjustment
The image can be adjustable for supporting judgment of the
image by film color.

■ : Reading and judging the image that has the

same effect as reading on the view box.

<Picture 3-1: Panorama image with film color >

■ : Coloring on the acquired image.

<Picture 3-2: Panorama image with coloring >

User’s Manual Point 200HD 34 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

D. Flash
Image processing can be partly applied. The result of the image
processing can process image more in detail like <Picture 4>
depends on the flash size and flash zoom.

■ : According to its methods, partial

image processing is available.


a. Normal: Image without any processing.
b. Equlize: Partly equalized image processing.
c. Sharpen: Partly high defined image processing.
d. HiPass: Partly high frequency image processing.
e. Blur: Partly smoothing image processing.
f. Median: Partly removing the noises image processing.

■ : Depends on the size of indication, it

partly specifies an available area for image processing


(minimum 100 to maximum 400).

■ : Depends on the zoom rate of

indication, it partly enlarges the image processing (minimum


1 to maximum 4 times).

User’s Manual Point 200HD 35 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

<Picture 4 – Method: Sharpen, flash size: 300, flash zoom: 2X>


E. Commonly used image processing menu

<Picture 5>
Above <Picture 5> shows that the entire imaging process menu
is the same in CDX-View expect for “drawing” and “implant”
menu.

Color reversing.
(Black to the white, white to the black)

Highlight the color contrast by widen color ranges.

Flipping for left/right reverse.

User’s Manual Point 200HD 36 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

Flipping for up/down reverse.

90° rotation to the left.

90° rotation to the right.

180° rotation.

Image sharpening by drawing a sharp dividing line.

Dividing line is more sharpen compare to high definition


(sharpening) filter.

Increase the intensity of sharpening effect higher than


sharpening filter.

Smooth image by making vague image look in general.

Reduce noise of the image.

F. Implant
An implant simulation, such as drawing and other efficient
functions are available in CDX-view program before
implementing the implant surgery.

<Picture 6-1>

User’s Manual Point 200HD 37 / 46


PNX.UM.032.0(E) (2013.04.15) 8. HOW TO USE

a. : Selects the implant that is subjected to be used and


starts the simulation. Types of implant, diameter, length and
rotations are selectable and changeable with using the
menu bar (③, ④, ⑤, ⑥).

b. : Selects the Crown that is subjected to be used


and starts the simulation. Types of Crown, diameter, length
and rotations are selectable and changeable with using the
menu bar (③, ④, ⑤, ⑥).

c. :Selects the types of implant or crown.

d. : Changes the diameter of implant


or crown.

e. : Changes the length of implant or


crown.

f. : Rotates the implant or crown. The


image can be rotated by clicking mouse‟s left button at the
edge of the object.

<Picture 6-2: Image that has remarked for implant and crown>

User’s Manual Point 200HD 38 / 46


PNX.UM.032.0(E) (2013.04.15) 9. MAINTENANCE

1) Use, Storage, Management


(1) Use and Environmental requirements
① Measurement spaces for patient: To properly adjust the device
and patient's positioning, the installation of inner parts of the
device has to keep at least 200cm (W) & 150 cm distances.

② Width of door: Minimum 150cm

③ Height: minimum 250cm

④ Floor: It has to support at least 500 Kg/m

⑤ Operation Requirements
Choose a site falling within the following temperature and humidity
ranges.
■ Temperature: 10 ~ 40℃
■ Humidity: 95%

⑥ Storage Requirements
■ Temperature: -10 ~ 50℃
■ Humidity: 20 ~ 95%
■ Atmospheric pressure: 600~1060Hpa
(2) Maintenance after use
① The device should be cleaned for the next use. Contaminated
materials are wiped with chemical solvents.
Allowed chemical solvents for disinfection
: Acetylene, Butyl Alcohol, Ethanol 96%, Methanol, Soap
Restricted chemical solvents for disinfection
: Benzene, Acetone, Acetic Ether

User’s Manual Point 200HD 39 / 46


PNX.UM.032.0(E) (2013.04.15) 9. MAINTENANCE

2) Maintenance, Warranty

(1) Maintenance
This device is designed to improve customer satisfaction with
supreme functions. To makre sure its safe operation, we recomment
a periodic inspection at least for one time per year.

① User managment
If problems are occurred, please immediate contact your local
distributor of PointNix Co., Ltd for further instructions.

② User inspection
User is asked to inspect these details in below every
month.(Monthly inspection is required for this information in below.)
■ Check if Ready button should be ON within two minuted after
power is connected.
■ Check signal sound is normally working while scanning.
■ Check all lampes in control panel are ON.
■ Check all buttons in control panel are working properly.

User’s Manual Point 200HD 40 / 46


PNX.UM.032.0(E) (2013.04.15) 9. MAINTENANCE
(2) Warranty
PointNix Co., Ltd guarantees that the device is free from the
manufacturing problems for a year from shipping date. For X-ray
tube, PointNix will provide the warranty for 1 year or 10,000 times of
scanning which comes first. Warranty period is valid for one year
from the shipping date. Any repair work or replacement that is
occurred after this period is not considered as warranty and
additional expenses are charged for these works after warranty
period.
Warranty shall not apply to natural disasters (including fire,
earthquake and other acts of GOD), the neglect, misuse or other-
than-ordinary use of the device, and attempted repair or alteration.

User’s Manual Point 200HD 41 / 46


PNX.UM.032.0(E) (2013.04.15) 9. MAINTENANCE

3) Waste/Deliver

(1) Waste
This device is basically designed to saftely use and dispose of
device to reduce the environmental pollution. Except X-ray
tube and such a several part, exterior and interior part is
possible to recycle as following use.
All the part or component that inculdes dangerous substance
must be seperatly disposed by regulating laws.

Main part need Waste Dangerous,


Part Recycle
to dispose disposal Separation
Aluminum,
Frame, Cover o
Plastic
Motor o
Board o
Cable Cop, Iron o
Wood, Board,
Package o
Paper
X-RAY Tube o
Sensor head Please return to PointNixhead office.
Other parts o
Ask detail instructions to your local distributor.

(2) Ways to be delivered/transported


Air, Cargo, Ocean

User’s Manual Point 200HD 42 / 46


PNX.UM.032.0(E) (2013.04.15) 10. INFORMATION
■ Product Name
x-ray system, panoramic/tomographic, digital

■ Model
Point 200HD

■ Input Voltage
100~120/220-240V~

■ Input Frequency
50/60Hz

■ Input Current
15/8A
■ Fuse
T25AL 250V~
■ Filter

X-ray own filter 0.8mmAl


Aluminum
Minimum total filter 2.7mmAl

User’s Manual Point 200HD 43 / 46


PNX.UM.032.0(E) (2013.04.15) 10. INFORMATION

■ Maximum allowed rating


2.0KVA
■ Device Type
Fixed Type
■ Operating mode
Discontinuous behavior
■ Apparent resistance
<1Ω
■ Applied standard
IEC 60601-1
IEC 60601-1-3
IEC 60601-2-7
IEC 60601-2-28
IEC 60601-2-32
IEC 60601-1-2
■ Class
According to the type of protecting against electric shock
CLASS Ⅰ EQUIPENT
According to the degree of protecting against electric shock
TYPE B APPLIED PART
■ EC Representative
DTF TECHNOLOGY Srl
ADDRESS : Via Castel Morrone 18, 20219 Milano, Italy

TEL : +39-02-84893641, Fax : +39- 02-8471894

User’s Manual Point 200HD 44 / 46


PNX.UM.032.0(E) (2013.04.15) 10. INFORMATION

Copyright ⓒ 2013
All rights reserved by PointNix Co., Ltd.

The information contained in this manual, the copyright laws


of the Republic of Korea, and computer programs are protected
by the Act. Contents of the manual in any form without prior
written consent of the manufacturer to modify, change, or
cannot reproduce.

Information of the manual can be changed without consent of


the user according to improvement and complement of the
product. For more information about the product or manual, or
any questions, please contact our nearest dealer or head office.

User’s Manual Point 200HD 45 / 46


MANUFACTURE:
PointNix Co., Ltd.

ADDRESS:
#302~303, 3F,
ACETWIN TOWER 1
285, Digital-Ro, Guro-Gu,
Seoul, Korea

HOMEPAGE:
http://www.pointnix.com

E-MAIL:
info@pointnix.com

TEL:
02-854-8315

FAX:
02-839-4973