T4777 Model Console For

Portable X-ray Equipment Of XXG serial Type, XXH serial

Type and XXQ serial Type.

Copy right by Aolong Group,who is the copyright holder. When receiving data carriers, the recipient acquires

the personal, nontransferable and non-exclusive limited license for the use and storage of the software in

conjunction with the hardware supplied by Aolong Group. This documentation may not be copied, duplicated,

reproduced,translated or transferred to electronic media or any other device, completely or partly, without the

prior written approval.

-1-
PORTABLE X-RAY UNIT

CONTENTS

1. GENERAL INFORMATION .................................................................................... - 3 -

2.SPEFICIATION DESCRIPTION ................................................................................ - 3 -

3.CONSTRUCTION .................................................................................................... 14

4.INSTALLATION ................................................................................................. - 27 -

5.OPERATION ....................................................................................................... - 27 -

6.NOTES ...............................................................................................................- 34 -

7. MAINTIANANCE ............................................................................................. - 35 -

8.MESUREEMTN AND REPAIR ...............................................................................- 36 -

9.FAILURE AND SOLUITON ................................................................................ - 38 -

10.CE & ISO CERTIFICATE ...................................................................................... - 50 -

-2-
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

1. GENERAL INFORMATION
Thank you for purchase “Aolong”brand Industrial NDT X-ray equipment. It is our honor to supply

you the good quality instruments with best after-sale service. “Aolong” Group is an advanced

technological NDT manufacture, and its products are used widely in Petroleum, chemical plant,

pressure vessel, Boiler, metallurgy, shipbuilding and so on.

2.SPEFICIATION DESCRIPTION
2.1.The feature of Industrial NDT X-ray equipment (with T4777 model console )

1 Anti-jamming strong which is suitable for the dynamo working together.

2 It can choose the expose parameter accurately by setting the press-button.

3 Has the delaying time function which makes the operator leave the dangerous radioactive

operating room when the X-ray unit generate the X-rays.

4 The high voltage risen and fallen smoothly which makes the X-ray generator work effectively.

5 According to 1:1 ratio of work and rest automatically, the longest exposed time is 5 minutes.

6 Warm up the X-ray unit by manual operation.

7 The complete protect measure and the protection function when there is over KV, over mA,

short mA, and over current, over temperature which avoid destroying the whole machine.

8 Exposing time and error code is displayed by counter.

9 The design of the structure is more reliable and reasonable and operate easily.

-3-
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

2.2Technical parameter (Table 1)

Item Technical Parameter

The whole machine power Power supply Signal 50Hz, AC220V±10%V

Input Use the dynamotor The power should be more than 5KW

Tube current 5mA

The whole machine power Voltage vibration of ±1%

Output H.V.transformer

Cooling method Forced air cooling

X-ray generator Insulated method SF6 gas insulated

working pressure 0.35~0.45Mpa6

Working condition Temperature -30℃~+40℃

Humidity ≤85%

Sea-level elevation 1000 m

Steel A3 (mm)

Focus 600mm

Condition for max. Exposing time 5min

penetration Film TIANJIN Ⅲ model film

Darkroom Temperature20℃±2℃

Density D≥1.5

Size 365×300×165mm

T4777 Console Weight 11.2kg

Time error 5min±5%

Working method Rate:1:1 to work and rest, the longest exposed time is 5.0 min preset.

-4-
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

Table 2

Technical Specification of The whole serials X-ray units

X-ray tube X-ray generator


Max.penetration
X-ray tube Beam angle
Model Out Power (KW) Focus size (mm)
Voltage (KVP) Size (mm) Weight (kg)
(mm) Cone target Level target
Cone Level

XXH-1005 1.0×3.5 30×360° 25×360° 190×190×530 12 4 6


>1.2 30~100
XXQ-1005 0.8×0.8 40+5° 190×190×530 12 8

XXQ-1605 0.8×0.8 40+5° 17 19

225×225×590
XXH-1605 >1.5 50~160 1.0×3.5 30×360° 25×360° 17 12 15

XXG-1605 0.8×0.8 40+5° 225×225×570 19 19

XXQ-2005 >2 50~200 1.5×1.5 40+5° 230×230×610 18 30

-5-
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

280×280×660
XXH-2005 1.0×3.5 30×360° 25×360° 21.5 24 27

XXG-2005 2.0×2.0 40+5° 280×280×620 24 30

XXQ-2505 2.0×2.0 40+5° 35 40

330×330×750
XXH-2505 >2.5 1.0×2.4 30×360° 25×360° 33.5 34 37

XXG-2505 50~250 320×320×650 31 40


2.0×2.0 40+5°

XXQ-3005 >3 2.3×2.3 40+5° 45 50

340×340×850

XXH-3005 100~300 1.0×2.3 30×360° 25×360° 40 44 47

XXG-3005 2.5×2.5 40+5° 340×340×680 36.5 50

-6-
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

Table 3 Directional Portable X-ray unit(with ceramic X-ray tube)

Model Output Input Focus Size Beam Max.Penetration Weight of Dimension of


Voltage Angle Generator Generator

XXG-1605 80~160kv 1.5kw 0.8×0.8mm 40+5° 19mm 14.5kg 225×225×550


mm

XXG-2005 100~200kv 2.0kw 2.0×2.0mm 40+5° 30mm 18.0kg 285×285×615


mm

XXG-2505 150~250kv 2.5kw 2.0×2.0mm 40+5° 40mm 30.5kg 320×320×640


mm

XXG-3005 170~300kv 3.0kw 2.5×2.5mm 40+5° 50mm 36.5kg 345×345×670


mm

XXG-3505 180~350kv 3.4kw 2.8×3.0mm 40+5° 55-60mm 45kg 345×345×800


mm

Table 4 Directional Portable X-ray unit(with glass X-ray tube)

Model Output Input Focus Size Beam Angle Max.Penetration Weight of Dimension of
Voltage Generator Generator

XXQ-1005 60~160kv 1.5kw 0.8×0.8mm 40° 8mm 11.1kg 190×190×530


mm
XXQ-1605 80~200kv 2.0kw 0.8×0.8mm 40° 19mm 15.2kg 225×225×585
mm

XXQ-2005 100~250kv 2.5kw 2.0×2.0mm 40+5° 30mm 23.0kg 285×285×665


mm

XXQ-2505 150~300kv 3.0kw 2.0×2.0mm 40+5° 40mm 35.0kg 320×320×730


mm

XXQ-3005 170~300kv 1.5kw 2.3×2.3mm 40+5° 50mm 45.5kg 345×345×830


mm

XXQ-3505 180~350kv 3.4kw 2.8×3.0mm 40+5° 55-60mm 48kg 345×345×800


mm

-7-
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

Table 5 Panoramic Portable X-ray unit (with glass X-ray tube)

Beam Angle Max. Penetration Dimension of


Output Voltage Weight of
Model Input Focus Size Level Generator
Cone Target Cone Target Level Target Generator
Target (mm)

XXH-1005 60~160kv 1.2kw 1.0×3.5mm 30×360° 25°×360° 4mm 6mm 11.1kg 190×190×530

XXH-1605 80~200kv 1.5kw 1.0×3.5mm 30×360° 25°×360° 12mm 15mm 15.0kg 225×225×585

XXH-2005 100~250kv 2.0kw 1.0×3.5mm 30×360° 25°×360° 24mm 27mm 21.5kg 285×285×665

XXH-2505 150~300kv 2.5kw 1.0×2.4mm 30×360° 25°×360° 34mm 37mm 33.5kg 320×320×730

XXH-3005 170~300kv 3.0kw 1.0×2.3mm 30×360° 25°×360° 44mm 47mm 40kg 345×345×830

Table 6 Panoramic portable X-ray flaw unit(with ceramic X-ray tube)

Beam Angle Max. Penetration Dimension of


Output Voltage Weight of
Model Input Focus Size Level Generator
Cone Target Cone Target Level Target Generator
Target (mm)

XXH-2005 100~250kv 2.0kw 1.0×5.0mm 30×360° 25°×360° 24mm 27mm 19.5kg 285×285×615

XXH-2505 150~300kv 2.5kw 1.0×5.0mm 30×360° 25°×360° 34mm 37mm 29.5kg 320×320×705

XXH-3005 170~300kv 3.0kw 1.0×5.5mm 30×360° 25°×360° 44mm 47mm 35kg 345×345×715

-8-
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

2.3 manufacturing Procedure

material purchased and tested

Console assembly Generator assembly Cable assembly accessories

Whole machine Testing by QC. Department

machine aging machine tested


machine working 8 hours
without stop
qualified

package

Warehousing for storage

Must leave the factory for Customers within 7 days, or else back to Q.C department for testing again .

9
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

2.4 The components of whole X-ray unit


2.4.1 Please see the following pictures as the reference .

Picture 1 & Picture 2

10
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

2.4.2 The components list

Console including Generator including

1.main machine part 1.main machine part

2. Circuit board 2 .high voltage coil

3. Electronic components 3 .iron core

4. X-ray tube inserted

/ 5 .Pressure meter

/ 6. Electronic component

Cable components including Specification Quality

1. low voltage cable (8 cores or 10 cores) 25 meters 1 nos.

2. power cable (3 cores) 10 meters 1 nos.

3. warning lamp and cable (2 cores) 5 meters 1 nos.

4. earth cable and earth pole 5 meters 1 nos.

Accessories including Quality

1. fuse (2A or 5A)


5 nos.
2. lead letter
1 box
3.IQI
2 nos.
4 .dark bag
5 kits
5. screen intensifier
5 kits
6. accessory bag
1 nos.

11
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

2.5 Flow block of console assembly

Weld circuit Clean & Console Console Console Qualification


board Testing assembly debugging aging

Flow block of Generator assembly


Preparation before assembly

Material claim Material inspection

Preparation before assembly

marking line grinding clean dry

High voltage transformer assembly Fault of high voltage coil

The dry of high voltage transformer


The assembly of The assembly of end
generator barrel cap
The assembly of The assembly of fan
dust cover cover

Fault of x ray tube


The assembly of high voltage generator

Self-check, mutual-check

Vacuum
Unqualified

Fill with sulfur fluoride

Experiment Not qualified Fault judgment Repair

Qualification
Note:Every working procedures are related to storage and transportation

12
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

2.6 The lines connecting details

Sectiona Connection point 1 Connection point 2

l area of Item Item code Wire Item code Item code Wire
Cable no.

gumme code length length Remark


d wire ( mm ) ( mm )
( mm²)

7 2.5 2CH5 Seven-core 1 200 Seven-core High voltage coil 450 Red color

1 primary start

14 2.5 2CH3 Seven-core 2 200 Seven-core High voltage coil 450 Blue color

2 primary end

56 1.0 2CH8 Seven-core 3 200 Seven-core High voltage coil 450 Yellow

3 secondary start color

4 0.4 2CH4 Seven-core 4 200 Seven-core Fan motor lead wire 200 Yellow

4 of barrel color

15 0.4 2CH1 Seven-core 5 200 Seven-core Lead wire of barrel 220 Yellow

5 Fan motor color

44 0.4 2CH6 Seven-core 6 200 Seven-core lead wire of barrel 220 White color

6 temperature relay

E 0.4 Seven-core lead wire of barrel 220 Double

7 temperature relay color

E 1.0 2CH9 Seven-core 7 200 Seven-core End cap 220 Double

7 combination earth color

58 0.4 2CH2 3CH 1 100 Yellow

color

4 0.4 2CH4 3CH2 100 Yellow

color

13
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

3.CONSTRUCTION

The whole x-ray unit is composed of console, X-ray generator, power cable, connection cable,

accessories and so on.

Notice: Don’t decompose the console and the X-ray generator by different models.

3.1. X-ray generator

X-ray generator is assembling construction-ray tube, H.V.transformer (Including filament winding of

tube) together with gas insulation (SF6) are sealed up in aluminum bucked cabinet.The cooling fan

and heat-sink is mounted in the end of X-ray generator as cooling device.

1 X-ray tube inserted

2 H.V.transformer (including filament winding)

3 Temperature relay.

4 Gas pressure gauge

5 Connecting cable socket

6 Alarm lamp socket

7 Cooling-fan

The x-ray generator should be placed horizontally and vertically by the two end-rings, which is used

to carry and lift it when getting the operating in the high altitude.

14
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

3.2Drawing of X-ray generator.

1. End ring
2. Absorber (head)
3.Cover of fan
4. Fan
5. Hat sink
6.Container of tube
7.Tube inserted
8.Cover area of tube inserted
9.Cover of tube cathode
10.H.V coil
11.3002-2 model iron core
14.Bolt
15.Absorber mat for iron core
16.Cover of machine
17.Absorber(bottom)
18.Pressure gage
19. Wire tie of iron core
20.7 cores wire holder
21. Aerator tank
22.Cover of tube anode
23.Socket of cable
24.Window of tube inserted

The drawings of generator(No. 2)

1.End loop 2.Center indicator 3. Radiation Window 4.Radiation mark 5.Press vacuum meter

6.alarm lamp jack 7. Connect cable jack 8. nameplate

15
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

2 3 4 5

1 2 3 4

D
图3
GX
DF2

Be

4 E 5 6
C

(1) (2) (3) (7) (6) (5) (4)

7 14 56 E 44 15 4
3 4 5 6

E
56
58
4 58
15 9
1 2 2CH 2 1
7 8 2CH D
5 6 3

图5
7 4
B
14
44 4

2CH1
1

2CH2
2
A

图7
1 2 3
C 4

图6 2CH6
6
2CH4
4
2CH9
9
DF2
2CH5
5 Title

B
Size Nu mber
Titl e B
- 16 - Size Nu mber
Date: 2 9-Feb -2 0 08
R ev i sio n
File: C:\Do cu men ts
B
2 3 4 DatManual
e: 2Version 2022-FCS-2
9-Feb -2 0 08 5 Sheet o f
File: C :\Do cu men ts an d Sett in gs \Ad min is trato r\桌面\假负载示意图.Dd
Drawn B y: b

3 4 5 6
PORTABLE X-RAY UNIT

3.3 Working theory drawings.

- 17 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

3.3 Console

Control unit is the cabinet type construction. Four circuit panels and electronic devices are

composed of it . All operating buttons and switches are equipped on the panel. Power socket,

power switch and earth-terminal are set in the socket box on the side of control unit to make the

operation flexible

T4777 console circuit (Picture 1)

LED(2) displayer indicator

“Time set” switch

mA lamp
KV lamp
Ready lamp
Power lamp

1 Power Indicator lamp;

Its lighting indicates the console can expose the X-ray.

2 KV indicator lamp;

Its lighting indicates the X-ray generator is working.

- 18 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

KV board circuit

- 19 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

3 mA indicator lamp;

Its lighting indicates the X-ray generator is working well.

mA circuit board

- 20 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

- 21 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

4 READY indicator lamp;

Its lighting indicates the console can used to have the exposure work.

5 LED(2) dis-player indicator;

It is used to control the exposure time ,resting time& the working status.

Displaying circuit board (1)&(2)

- 22 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

- 23 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

6 H.V setting switch;

It is used to set the exposure parameter.

200
190 210
120
180 140 150 160 220
110 130
80 130 170
75 85
100 140 170 120 180 230
70 90
110 190
65 95 160
90 150 100 200 240
60 100
150 250
80 160
2005 & 2505 serials Model
1005 & 1605 Serial model

3005 Serial model

- 24 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

7 H.V On button ;

It is used to operate the exposing, training or close the machine

8 H.V. Off button;

It is used to stop the exposure and working of the whole machine.

9 Exposing Curve panel; (Please see the following pictures as reference )

It is used to choose the right exposure curve according to the object material tested.

- 25 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

3.4.Working theory

The control unit adopts silicon-controlled single-phase bridge rectifier. Rectified voltage changes

into smoothing D.C.voltage through LC filter loop, which changes into unidirectional pulse with

changeable frequency through silicon-controlled carrier loop, sent to H.V.pulse transformer as

supply of generator, Ma stabilization unit can rise or reduce with filament voltage of tube. Change

the frequency of unidirectional pulse to make tube current stable. Kv adjustment unit can adjust

continuously to meet with the different materials testing.

- 26 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

4.INSTALLATION
1 Verify all parts of unit depending on contents of box after open the packing box. Check all

parts which are not destroyed or broken .

2 Check the console and the X-ray generator should be the same model, or else the equipment

can’t be used.

3 Turn off power switch after placing control unit and connect power cable. Power supply should

conform to the request in “Technical parameter”

4 Connect one end of earth wire to earth-terminal on the control unit, the other end is

connected to damp place or specials earth wore.

5 Connect between control unit and generator with connecting cable, and ensure that

connected well.

6 Connect alarm lamp on generator to show generating X-ray.

7 X-ray protective installation is provided 2mm thickness lead-plate is protection for operating in

the field. If without conditions, can penetrate when there is no mA within radius 20m around the

center focus of generator.

8 Finish above, check whether pressure gauge of generator conform to request in technical

parameter, if not please do not to turn it on in order to avoid damaging generator.

5. OPERATION
5.1. Open.

After installation, confirm the connection is right, set the power switch; the LED (2) displayer is on.

5.1.1 .Check automatically

After open the machine ,the console do excise automatically firstly, at the same time the LED (2)

displays “Rd” & “--” alternately.

- 27 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

The following state is normal during the machine working

1 Preparing indicator lamp ,the H.V indicator lamp and the mA indicator lamp turn off.

2 The H.V.open button and H.V.close button can not be used.

The following state is not normal during the machine working

1 The LED (2) does not display “Rd” & “--” alternately.

2 The Buzzer is on non-stop.

Notice:Once there is any question, the operator should record the error code, and then close the

machine . Please contract with the manufacture directly .

If system is testing well, about after 30s, will display:

1 Preparing indicator lamp is on ;

2 The Buzzer is on (one or two).

3 The LED (2) display “XX” (X means 0~5 number )to proof the machine is ready to work.

5.2 .Exposing operating

After the above, the system is ready, the Preparing indicator light is on; the system will get into

setting parameter state.

5.2.1. setting exposing parameter

Before the exposing, should set the exposing parameter at first (the KV value and the exposing

time).

Setting the exposing time follow the step:

A. Confirm the material and its thickness of the object tested .

B. Work out the thickness of the object tested compared with the A3 steel thickness to choose the

right exposure curve .

C. According to the exposing curve to confirm the KV and the exposing time.

- 28 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

Notice: Because of the difference function of the difference model x-ray generator, the exposing

curve may be not the best reference to the operating, the operator should confirm the best state

for your equipment according the operating condition.

(1).How to adjust the KV value

Turn the “H.V.seting” switch by the clockwise or anticlockwise, get the KV value position according

to exposing curve. And Turning the switch clockwise will increase, anticlockwise will reduce the KV.

KV is opposite with the model of the equipment, the KV of the console and the X-ray generator is:

1005 model: 60~100 KV 1605 model: 80~160 KV

2005 model: 100~200 KV 2505 model: 150~250 KV

3005 model: 170~300 KV 3505 model: 180~350 KV

120 200
190 210
110 130
75
80
85 180 140 150 160 220
100 140 130 170
70 90
170 120 180 230
65 95
90 150 110 190
60 100 160 100 200 240
80 160
1005 & 1605 Serial model 150 250
2005 & 2505 serials Model

- 29 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

3005 Serial model

(2). How to adjust the exposing time

Use the “timer” switch and the “LED (2)” to set the exposing time.

1 Turn the “Timer” switch clockwise or anticlockwise, and the “LED (2)” will increase or reduce

accordingly.

2 The left position of the “LED (2)” is “second” and the right position is 1/10 “minute” is the

multiple Of 6 seconds, for example “4.5” means that the exposing time is 4′30″ .

3 The exposing time of the T4777 model console is from 0.1~5 minutes, and increase or reduce

by 0.1 minute step.

5.2.2.Start-up exposing

After confirm the KV value and the exposing time is okay ,press“H.V.opening”switch.It is end.

5.2.3.Exposing process

It include 3 start-up ,delaying time, exposing; resting total three (3) steps.

A. Delaying time

After press the “H.V. open” button, in order to make the operator leave the locale easily and safely,

- 30 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

before creating the x-ray, the system will delay 30 seconds automatically. Under the delaying time

phase, the displayer show the time, ,you will hear “tick, tick--------- ” the prompt voice ,the

operator may again press “H.V. opening ” to stop the delaying course ,enter the exposing course

straight .

B. Exposing

The system raises the tube KV, tube mA slowly to the setting value, and adopt the advanced

controlling method to adjust.

H.Voltage indicator light of console is blink, means that the x-ray generator is under H.V condition,

few minutes later the “mA indicator lamp ” is blink, means that the system is in exposing Phase.

When the remain time is “0” on the displayer, the system will cut off the H.V, getting into the resting

phase automatically.

If press the “H.V. closing” under the exposing phase, the system will stop the exposing, cut off the

H.V to get into the resting phase, the time is same with the H.V. time, then return to the setting

parameter state.

If there is some fault under the exposing, the console will cut off the H.V. immediately, at the same

time , the LED (2) will display the fault code , the buzzer is ringing non-stop until press the “H.V.

closing” switch..

C. Resting

After the exposing is over (or press the “H.V. close” switch to stop), the system will cut off the H.V.

automatically to rest, the time is same with the H.V. time, the resting phase is forced.

The display show the remain resting time value ,When the resting time is finished, the” Preparing

indicator light” is on, the buzzer have sound prompt, display show the exposing time, return to the

state of setting the exposing parameter automatically .

- 31 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

5.3. Training operating

The equipment should training In order to avoid the X-ray generator being damaged or shorten the

life.

Setting the training high voltage based practice use complexion,usually it little high than the high

voltage of exposing.the equipment is not setting automatic training,only manual training.

manual training means the operator trains equipment require to adjust KV advance set knob, the

training time will long ,if the equipment rest last time long. the equipment rests last about 40hs,

over 1 week or 15 days, even more training days needed ,here the training should adding unto 90

minutes.

5.4. Alarm protection

When x-ray generation crate x-ray under high voltage (no matter under exposing or training), the

system will surveillance the KV. mA.x-ray tube temperature of the x-ray generation, if there is over

- 32 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

mA, no mA, over temperature, the console will cut off the H.V. immediately, at the same time,

display the fault code on the LED (2), the buzzer will work constantly until press the “H.V. CLOSING ”

switch ,the following is the meaning of the fault code :

“HU”:Over high voltage fault

“HA”:Over mA fault

“LA”: Less mA fault

“PT”: Over temperature fault

Under the alarm condition, press the H.V. switch, the system will stop the working to have a rest.

The resting time is the same as the H.V.time, and then returns the exposing parameter setting state.

5.5. Turn off the machine

In order to ensure that the system can work well, the operator should cut off the power supply

under the setting parameter, forbidding to close under connect the X-ray generator with the H.V.

( during exposing or training, exclude the pause or the rest time ).

Notice :After the close the machine and should wait for 10 seconds and open it again .Or else it

will damage the component.

- 33 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

6. NOTES
1 Please make sure the grounding cable connecting very well .

2 If there are some noise from the X-ray generation inside during rising the H.V., the H.V. skip big

or other not normal things happen, the reason maybe come from x-ray tube discharge.It should

stop operating immediately.

3 The equipment should not work under or over the rated voltage.

4 Working time and rest time rate should be by 1:1,not cut power supply when rest to ensure

that cooling-fan works normally, Must rest the equipment as per the prescript,otherwise will lay

waste the equipment.

5 Power voltage vibration is large instantaneous, maybe protection circuit works, not belong to

the fault of equipment, continue to operation.

6 Want to lengthen connecting cable; section area of every wire inside cable is in line with

parameter in table.

Lengthen (m) Section area (mm2)

10-30 1.5

30-50 4.0

50-80 6.0

- 34 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

7. MAINTIANANCE
1 The operator must be familiar with the structure and theory of X-ray machine

2 There must be the instruments and meters,tool and other instruments for inspection circuit

and other parts.

 Instrument and meters including : oscillograph, theostat, stabilized voltage supply,

multimeter, vacuum tester, Leak Detector for halogen.

 Tool including: necessary electronic meters , locksmith and so on.

 Facilities including: thermostat, vacuum air pump, gas loading device and SF6

3 Be careful for electric shock accident,and there must be safeguard procedures for X-ray harm

4 The equipment is complicated and the operating condition is strict. So in order to lengthen the

equipment, must give the best maintenance.

The working and storage place where there isn’t corrosive gases and the environment temperature

among-40℃~55℃,the rating humidity of the air ≤85%, the air should keep circulating , and the

floor should keep dry.

5 The maintenance of console

 The storage place should be clear ,dry and the rating humidity of the air ≤85% .

 Avoid acute shake and collision.

 Keep the environment clean, keep from dust, in order to avoid short circuit.

 During the operating, should notice whether the KV or the steady number is normal.

 During the operating, set the switch and the button gently.

 Should ensure that the cooling-fan works normally. after the operating, the switch

should set on the minimum.

6 The maintenance of the X-ray generation

- 35 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

 During the moving, avoiding shake and collision, and during the transporting, should

be upright transport or storage .

 The cable plug should be keep clean, dry.

 Forbidding working over 0.35p or lower than -30℃.

 Before open the machine, should check that whether the cooling-fan is normal, in

order to ensure the X-ray generation can be cooled ,or else do not use the machine

8. MESUREEMTN AND REPAIR


1 Before the measurement, the operator must assure that there is no damage of the surface of

x-ray machine If you need measure the console with power supply ,please care for electric

shock.The body and hand can not touch any parts with power open .

2 Firstly you should measure the console and connecting cable .If console and cables are not

broken,then connect the main machine to adjust the X-ray generator.

3 If there is something wrong with X-ray generator,and the operator wants to measure the it .You

should let the air insulated (SF6) out and make it to -0.1Mpa by vacuums. After staying it for 20

hour,you should open it.

4 Must take the air cooler of X-ray generator down ,then uncover the X-ray tube inserted.

5 Pay attention to it when uncover the X-ray tube inserted,must use the special tools.

6 After the x ray tube is taken off, loosen the screws around the generator barrel, lightly lift up

the barrel. Use an electric iron to weld off the tree lead-wires’ joints

7 Operator should wear radiation-proof clothes and caps, should be aware of the harm arising

by x ray tube’s self-exposure. Prepare a thick cloth bag to sheathe X-ray tube inserted.. Lift the

barrel, observing if there is any color change of X-ray tube inserted, sheathe the X-ray tube inserted

into the thick cloth bag, then weld off the connecting cable.

- 36 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

8 Carefully inspect the X-ray tube inserted if there is unusual color ,the fuse is broken ,the target

surface exits the oxidize trace ,there is a hole punched on the shell of tube.If one really exists,

indicate that there is something wrong with X-ray tube inserted .Must change new one.

9 The X-ray tube inserted replaced must be clear ,no dust.And dry it for 1 hour in the drying

cabinet.

10 Carefully measure the high voltage transformer if burn or puncture ,and if there is odor. If

there is really odor ,and we assure that there is something wrong with high voltage

transformer .Must change new one.

11 Must cost 48 hours to dry the new high voltage transformer replaced,keep temperature of the

drying cabinet at 100℃

12 Must cost 2 hours to vacuum the X-ray generator

13 Gas inflation for the X-ray generator with SF6 gas, fulfilling it to 0.45Mpa(20℃) .You should

change the Mpa value according to the temperature.

14 Finish gas inflation,make the marks (pressure value),and staying it steady for 12 hours.If

discover reduction voltage,please find the reason and resolve it and then have a test by power

supply.

15 When finish the inspection,operator must have a test for liability ,penetration , sensitivity and

others. Pay attention to accord to the “technology parameter” in manual and be in protective

working environment.

- 37 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

9. FAILURE AND SOLUITON


9.1 Failure & Solutions of the Portable X-ray unit

List 1:

Failure Phenomenon Solution

1.Check all of power supplying connecting.


2.Check the connecting of the power cable and cable
plugs.
No display of start power controller 3. Check the fuse of console.
4.Check drop-out line of single-chip microcomputer (SCM)
and other plugs of circuit board.
5. Change display board of the console.
1. Change KV control board. 2. Change mA control board.
Power switch turns off in soon or in the
3. Change the semi-controlled bridge.
few seconds power when open the
4.Change the silicon controlled
machine. rectifier.(SCR4,KK.20/800Model)

1.Please use another console to judge the problem.


2. If the x-ray generator is broken,please send it back to the
manufacture directly or replace the new one.
3. Test whether the console is good. (details see test
The console displays normally but the
connect diagram)
machine has the protection action and
4. If the console failure has been defined, following
then the high voltage cut off. solutions can be used.
a. Replace the mA board
b. Replace KV board
c. Replace silicon controlled rectifier.

- 38 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

1. If there is code HU, change new KV board.


The console works well after power
2. If there is code HA, LA or PT error code, change new
starting, but it can’t withstand the high
mA board.
voltage.
3. Change display board.

9.2 Failure & Solutions of the Portable X-ray Unit


List 2:

Failure Phenomenon Solution

The display window shows jumbled 1.Change display board.


code after high power starts. 2.Change mA board.

Time can’t be set the exposure time. 1.Change display board.


2.Check the connecting between the
displaying board with Time-preset switch.

No audio signal/indication after finish 1.Change display board.


the exposure working. 2.check the buzzer and its connecting .

High Voltage can’t stop automatically


Change display board.
when exposure is finished

1. If there is code HU, please change KV circuit board.


2. If there is code HA, LA, change mA circuit board.

HV stops during exposure (removing 3. If there is code PT, check the temperature increase

the failure according to the code of the X-ray generator and cooling fans. Change mA

displayed) board.
4. Please change display circuit board if the above
method is not okay.
1. Check the lamp.
Warning lamp of the X-ray generator
2. Check the cable and plug of the warning lamp.
doesn’t work.
3. Check the connection cable

- 39 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

1. Fans are broken.


Cooling fans of the X-ray generator
2. Connect cable doesn’t connect very well or cable is
doesn’t work.
damaged.

Console experiment connection drawing

1 CH3,CH5,CH6,CH9,CH8 show the socket derivation foot(jack) of joins cable of controller.

2 The adjustable power range is 0V~15V。

3 The adjustable power can be replaced by 9V cascade battery.

4 The connect line should be connected correct otherwise will produce others result.

5 The resistance under the fictitious loading between the CH3 with CH5 must be from 10 Ω to

50Ω.

Load resistance 20 Ω allow pass 10A

current(may use 2KW electric cooker)

Adjustable power

4.5How to change silicon controlled rectifier.


- 40 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

Welding it according to the pointed location.

4.6 How to solve the L.A problem of console


Please see the following picture 1 ,you will see the mA board .Please unload the mA board ,and

change new one .

If change the new board ,you do not still solve the problem . Please find the diode marked by
2744 and 2739,as the picture showing .

- 41 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

Please use the multimeter to the position as the picture show ,and testing the connection of 2
diodes .

If you get the “0”displaying on the screen and hear some warning voice , which mean the
diodes are broken .How to fix them ?
A . Change new one .
B. If you want to use the X-ray units urgently ,you can move this 2 diodes and make the console
work .

4.7How to test the lead door interlock switch (PT Problem )


Lead door interlock

POWER

Alarm lamp

Lead door interlock

No.1connector No.1connector

No.2connector
No.2connector

1. Make sure to install the interlock switch as picture showing.

- 42 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

Make sure these 2 parts


are connected very well

2. Test the 2 piece of lines of switch ,the multimeter value should be “000”
STEP 1

STEP 2

Test this 2 lines by multimeter

,which should show the “000”

STEP 3
If the multimeter do not show the “000”, which proof the switch is broken or the lines of
switch is broken.

4. 8 How to test the temperature switch


1. Make sure the multimeter has the right shift on the 200ohm ,as the picture showing

- 43 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

The shift of
multimeter
should be as
some as the
picture
showing.

2. Test the no.6 and no.9 line as the picture showing.

The value should be “000”as the picture showing

- 44 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

3. If you are not sure about it ,you can lease the temperature switch from the tube head and
test it as the picture showing

And then test it by multimeter as picture showing

- 45 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

The value may be “000”or”0000”which proofs the switch is okay.

9.3 Failure & Solutions of the Portable X-ray Detector


List 3:

Failure Phenomenon Solution

High voltage transformer primary current sudden The primary connector of HV transformer is
increase and the same time power switch fall destroyed.
off

High voltage transformer primary current sudden


HV coil is damaged.
increase and the same time power switch fall off

High voltage primary current is high and increases


HV coil is damaged.
by and by KV increases.

Filament of X-ray tube is disconnected or


HV primary current is low and no tube current
filament of X-ray tube is damaged.

HV primary current is high and tube current is


over 5mA. The current increases with KV value X-ray tube is destroyed
increases.

High voltage transformer primary current high ,


HV coil is damaged.
without tube current

- 46 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

The pressure of SF6 is below the safe using value Need to re fill the Gas in soon.

1.Oscillograph connecting and operation


Choose two cable core (marked “3”,”5 “) to connect with Oscillograph by two wires, the wire from 5

cable port must connect with anode(+) of Oscillograph,and other wire from 3 cable port must

connect with anode (-) of Oscillograph.

DS1052Oscillograph band:RIGOL, Model:DS1052

normal waveform

Non-normal waveform

2. Faults analysis: It is very difficult to analyze the tube inserted is normal or broken.You should

prepare one ammeter for testing primary current(Use one wire from marked “5” cable port ,and

series connection with a ammeter whose measuring range is 10A for test the DC ). Or you can

connection a ammeter of mA (from marked “8” cable port ,the method is as the same as above but

the measuring range of ammeter is greater than 20mA)

If the primary current is very low, no mA, which display the lamp of tube inserted is

broken

If the primary current is high,no mA, which display the H.V coil is broken.

If the primary current is very high, mA>5mA and increasing when the KV is raising,which

display the tube inserted is broken.

The better method is that you can use the pliers ammeter to test tube and coil.

- 47 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

Tube inserted:measure the lamp,if it is normal,so display the tube inserted is good.

The picture as the reference.

Coil:firstly look the surface is if it is burned,and smell its odor,finally measure its wire.

Measure the primary current of coil

- 48 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

4.4Familiar Failure & Solutions of the Portable X-ray Detector


List 4:

Failure Phenomenon Solution

Using Halogen leak detector to find out leak


SF6 leakage of X-ray generator
point and seal it.

The curve line of the relationship between SF6 gap pressure and temperature

0.54
The pressure shown by pressure meter

0.5

0.46
Mpa

0.42

0.38

0.34
-20 0 20 40 60 80
The average temperature of X ray generator(°C); Inflation pressure error: setting time + 0.01Mpa

- 49 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

10. CE & ISO CERTIFICATE

- 50 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

ISO 9001:2015

- 51 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

ISO 14001:2015

- 52 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT

ISO 45001:2018

- 53 -
Manual Version 2022-FCS-2

You might also like