Copy right by Aolong Group,who is the copyright holder. When receiving data carriers, the recipient acquires
the personal, nontransferable and non-exclusive limited license for the use and storage of the software in
conjunction with the hardware supplied by Aolong Group. This documentation may not be copied, duplicated,
reproduced,translated or transferred to electronic media or any other device, completely or partly, without the
-1-
PORTABLE X-RAY UNIT
CONTENTS
3.CONSTRUCTION .................................................................................................... 14
4.INSTALLATION ................................................................................................. - 27 -
5.OPERATION ....................................................................................................... - 27 -
6.NOTES ...............................................................................................................- 34 -
7. MAINTIANANCE ............................................................................................. - 35 -
-2-
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
1. GENERAL INFORMATION
Thank you for purchase “Aolong”brand Industrial NDT X-ray equipment. It is our honor to supply
you the good quality instruments with best after-sale service. “Aolong” Group is an advanced
technological NDT manufacture, and its products are used widely in Petroleum, chemical plant,
2.SPEFICIATION DESCRIPTION
2.1.The feature of Industrial NDT X-ray equipment (with T4777 model console )
3 Has the delaying time function which makes the operator leave the dangerous radioactive
4 The high voltage risen and fallen smoothly which makes the X-ray generator work effectively.
5 According to 1:1 ratio of work and rest automatically, the longest exposed time is 5 minutes.
7 The complete protect measure and the protection function when there is over KV, over mA,
short mA, and over current, over temperature which avoid destroying the whole machine.
9 The design of the structure is more reliable and reasonable and operate easily.
-3-
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
Input Use the dynamotor The power should be more than 5KW
Output H.V.transformer
Humidity ≤85%
Steel A3 (mm)
Focus 600mm
Darkroom Temperature20℃±2℃
Density D≥1.5
Size 365×300×165mm
Working method Rate:1:1 to work and rest, the longest exposed time is 5.0 min preset.
-4-
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
Table 2
225×225×590
XXH-1605 >1.5 50~160 1.0×3.5 30×360° 25×360° 17 12 15
-5-
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
280×280×660
XXH-2005 1.0×3.5 30×360° 25×360° 21.5 24 27
330×330×750
XXH-2505 >2.5 1.0×2.4 30×360° 25×360° 33.5 34 37
340×340×850
-6-
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
Model Output Input Focus Size Beam Angle Max.Penetration Weight of Dimension of
Voltage Generator Generator
-7-
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
XXH-1005 60~160kv 1.2kw 1.0×3.5mm 30×360° 25°×360° 4mm 6mm 11.1kg 190×190×530
XXH-1605 80~200kv 1.5kw 1.0×3.5mm 30×360° 25°×360° 12mm 15mm 15.0kg 225×225×585
XXH-2005 100~250kv 2.0kw 1.0×3.5mm 30×360° 25°×360° 24mm 27mm 21.5kg 285×285×665
XXH-2505 150~300kv 2.5kw 1.0×2.4mm 30×360° 25°×360° 34mm 37mm 33.5kg 320×320×730
XXH-3005 170~300kv 3.0kw 1.0×2.3mm 30×360° 25°×360° 44mm 47mm 40kg 345×345×830
XXH-2005 100~250kv 2.0kw 1.0×5.0mm 30×360° 25°×360° 24mm 27mm 19.5kg 285×285×615
XXH-2505 150~300kv 2.5kw 1.0×5.0mm 30×360° 25°×360° 34mm 37mm 29.5kg 320×320×705
XXH-3005 170~300kv 3.0kw 1.0×5.5mm 30×360° 25°×360° 44mm 47mm 35kg 345×345×715
-8-
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
package
Must leave the factory for Customers within 7 days, or else back to Q.C department for testing again .
9
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
10
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
/ 5 .Pressure meter
/ 6. Electronic component
11
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
Self-check, mutual-check
Vacuum
Unqualified
Qualification
Note:Every working procedures are related to storage and transportation
12
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
l area of Item Item code Wire Item code Item code Wire
Cable no.
7 2.5 2CH5 Seven-core 1 200 Seven-core High voltage coil 450 Red color
1 primary start
14 2.5 2CH3 Seven-core 2 200 Seven-core High voltage coil 450 Blue color
2 primary end
56 1.0 2CH8 Seven-core 3 200 Seven-core High voltage coil 450 Yellow
4 0.4 2CH4 Seven-core 4 200 Seven-core Fan motor lead wire 200 Yellow
4 of barrel color
15 0.4 2CH1 Seven-core 5 200 Seven-core Lead wire of barrel 220 Yellow
44 0.4 2CH6 Seven-core 6 200 Seven-core lead wire of barrel 220 White color
6 temperature relay
color
color
13
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
3.CONSTRUCTION
The whole x-ray unit is composed of console, X-ray generator, power cable, connection cable,
Notice: Don’t decompose the console and the X-ray generator by different models.
tube) together with gas insulation (SF6) are sealed up in aluminum bucked cabinet.The cooling fan
3 Temperature relay.
7 Cooling-fan
The x-ray generator should be placed horizontally and vertically by the two end-rings, which is used
to carry and lift it when getting the operating in the high altitude.
14
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
1. End ring
2. Absorber (head)
3.Cover of fan
4. Fan
5. Hat sink
6.Container of tube
7.Tube inserted
8.Cover area of tube inserted
9.Cover of tube cathode
10.H.V coil
11.3002-2 model iron core
14.Bolt
15.Absorber mat for iron core
16.Cover of machine
17.Absorber(bottom)
18.Pressure gage
19. Wire tie of iron core
20.7 cores wire holder
21. Aerator tank
22.Cover of tube anode
23.Socket of cable
24.Window of tube inserted
1.End loop 2.Center indicator 3. Radiation Window 4.Radiation mark 5.Press vacuum meter
15
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
2 3 4 5
1 2 3 4
D
图3
GX
DF2
Be
4 E 5 6
C
7 14 56 E 44 15 4
3 4 5 6
E
56
58
4 58
15 9
1 2 2CH 2 1
7 8 2CH D
5 6 3
图5
7 4
B
14
44 4
2CH1
1
2CH2
2
A
图7
1 2 3
C 4
图6 2CH6
6
2CH4
4
2CH9
9
DF2
2CH5
5 Title
B
Size Nu mber
Titl e B
- 16 - Size Nu mber
Date: 2 9-Feb -2 0 08
R ev i sio n
File: C:\Do cu men ts
B
2 3 4 DatManual
e: 2Version 2022-FCS-2
9-Feb -2 0 08 5 Sheet o f
File: C :\Do cu men ts an d Sett in gs \Ad min is trato r\桌面\假负载示意图.Dd
Drawn B y: b
3 4 5 6
PORTABLE X-RAY UNIT
- 17 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
3.3 Console
Control unit is the cabinet type construction. Four circuit panels and electronic devices are
composed of it . All operating buttons and switches are equipped on the panel. Power socket,
power switch and earth-terminal are set in the socket box on the side of control unit to make the
operation flexible
mA lamp
KV lamp
Ready lamp
Power lamp
2 KV indicator lamp;
- 18 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
KV board circuit
- 19 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
3 mA indicator lamp;
mA circuit board
- 20 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
- 21 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
Its lighting indicates the console can used to have the exposure work.
It is used to control the exposure time ,resting time& the working status.
- 22 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
- 23 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
200
190 210
120
180 140 150 160 220
110 130
80 130 170
75 85
100 140 170 120 180 230
70 90
110 190
65 95 160
90 150 100 200 240
60 100
150 250
80 160
2005 & 2505 serials Model
1005 & 1605 Serial model
- 24 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
7 H.V On button ;
It is used to choose the right exposure curve according to the object material tested.
- 25 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
3.4.Working theory
The control unit adopts silicon-controlled single-phase bridge rectifier. Rectified voltage changes
into smoothing D.C.voltage through LC filter loop, which changes into unidirectional pulse with
supply of generator, Ma stabilization unit can rise or reduce with filament voltage of tube. Change
the frequency of unidirectional pulse to make tube current stable. Kv adjustment unit can adjust
- 26 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
4.INSTALLATION
1 Verify all parts of unit depending on contents of box after open the packing box. Check all
2 Check the console and the X-ray generator should be the same model, or else the equipment
can’t be used.
3 Turn off power switch after placing control unit and connect power cable. Power supply should
4 Connect one end of earth wire to earth-terminal on the control unit, the other end is
5 Connect between control unit and generator with connecting cable, and ensure that
connected well.
7 X-ray protective installation is provided 2mm thickness lead-plate is protection for operating in
the field. If without conditions, can penetrate when there is no mA within radius 20m around the
8 Finish above, check whether pressure gauge of generator conform to request in technical
5. OPERATION
5.1. Open.
After installation, confirm the connection is right, set the power switch; the LED (2) displayer is on.
After open the machine ,the console do excise automatically firstly, at the same time the LED (2)
- 27 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
1 Preparing indicator lamp ,the H.V indicator lamp and the mA indicator lamp turn off.
1 The LED (2) does not display “Rd” & “--” alternately.
Notice:Once there is any question, the operator should record the error code, and then close the
3 The LED (2) display “XX” (X means 0~5 number )to proof the machine is ready to work.
After the above, the system is ready, the Preparing indicator light is on; the system will get into
Before the exposing, should set the exposing parameter at first (the KV value and the exposing
time).
B. Work out the thickness of the object tested compared with the A3 steel thickness to choose the
C. According to the exposing curve to confirm the KV and the exposing time.
- 28 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
Notice: Because of the difference function of the difference model x-ray generator, the exposing
curve may be not the best reference to the operating, the operator should confirm the best state
Turn the “H.V.seting” switch by the clockwise or anticlockwise, get the KV value position according
to exposing curve. And Turning the switch clockwise will increase, anticlockwise will reduce the KV.
KV is opposite with the model of the equipment, the KV of the console and the X-ray generator is:
120 200
190 210
110 130
75
80
85 180 140 150 160 220
100 140 130 170
70 90
170 120 180 230
65 95
90 150 110 190
60 100 160 100 200 240
80 160
1005 & 1605 Serial model 150 250
2005 & 2505 serials Model
- 29 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
Use the “timer” switch and the “LED (2)” to set the exposing time.
1 Turn the “Timer” switch clockwise or anticlockwise, and the “LED (2)” will increase or reduce
accordingly.
2 The left position of the “LED (2)” is “second” and the right position is 1/10 “minute” is the
multiple Of 6 seconds, for example “4.5” means that the exposing time is 4′30″ .
3 The exposing time of the T4777 model console is from 0.1~5 minutes, and increase or reduce
5.2.2.Start-up exposing
After confirm the KV value and the exposing time is okay ,press“H.V.opening”switch.It is end.
5.2.3.Exposing process
It include 3 start-up ,delaying time, exposing; resting total three (3) steps.
A. Delaying time
After press the “H.V. open” button, in order to make the operator leave the locale easily and safely,
- 30 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
before creating the x-ray, the system will delay 30 seconds automatically. Under the delaying time
phase, the displayer show the time, ,you will hear “tick, tick--------- ” the prompt voice ,the
operator may again press “H.V. opening ” to stop the delaying course ,enter the exposing course
straight .
B. Exposing
The system raises the tube KV, tube mA slowly to the setting value, and adopt the advanced
H.Voltage indicator light of console is blink, means that the x-ray generator is under H.V condition,
few minutes later the “mA indicator lamp ” is blink, means that the system is in exposing Phase.
When the remain time is “0” on the displayer, the system will cut off the H.V, getting into the resting
phase automatically.
If press the “H.V. closing” under the exposing phase, the system will stop the exposing, cut off the
H.V to get into the resting phase, the time is same with the H.V. time, then return to the setting
parameter state.
If there is some fault under the exposing, the console will cut off the H.V. immediately, at the same
time , the LED (2) will display the fault code , the buzzer is ringing non-stop until press the “H.V.
closing” switch..
C. Resting
After the exposing is over (or press the “H.V. close” switch to stop), the system will cut off the H.V.
automatically to rest, the time is same with the H.V. time, the resting phase is forced.
The display show the remain resting time value ,When the resting time is finished, the” Preparing
indicator light” is on, the buzzer have sound prompt, display show the exposing time, return to the
- 31 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
The equipment should training In order to avoid the X-ray generator being damaged or shorten the
life.
Setting the training high voltage based practice use complexion,usually it little high than the high
manual training means the operator trains equipment require to adjust KV advance set knob, the
training time will long ,if the equipment rest last time long. the equipment rests last about 40hs,
over 1 week or 15 days, even more training days needed ,here the training should adding unto 90
minutes.
When x-ray generation crate x-ray under high voltage (no matter under exposing or training), the
system will surveillance the KV. mA.x-ray tube temperature of the x-ray generation, if there is over
- 32 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
mA, no mA, over temperature, the console will cut off the H.V. immediately, at the same time,
display the fault code on the LED (2), the buzzer will work constantly until press the “H.V. CLOSING ”
“HA”:Over mA fault
Under the alarm condition, press the H.V. switch, the system will stop the working to have a rest.
The resting time is the same as the H.V.time, and then returns the exposing parameter setting state.
In order to ensure that the system can work well, the operator should cut off the power supply
under the setting parameter, forbidding to close under connect the X-ray generator with the H.V.
Notice :After the close the machine and should wait for 10 seconds and open it again .Or else it
- 33 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
6. NOTES
1 Please make sure the grounding cable connecting very well .
2 If there are some noise from the X-ray generation inside during rising the H.V., the H.V. skip big
or other not normal things happen, the reason maybe come from x-ray tube discharge.It should
3 The equipment should not work under or over the rated voltage.
4 Working time and rest time rate should be by 1:1,not cut power supply when rest to ensure
that cooling-fan works normally, Must rest the equipment as per the prescript,otherwise will lay
5 Power voltage vibration is large instantaneous, maybe protection circuit works, not belong to
6 Want to lengthen connecting cable; section area of every wire inside cable is in line with
parameter in table.
10-30 1.5
30-50 4.0
50-80 6.0
- 34 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
7. MAINTIANANCE
1 The operator must be familiar with the structure and theory of X-ray machine
2 There must be the instruments and meters,tool and other instruments for inspection circuit
Facilities including: thermostat, vacuum air pump, gas loading device and SF6
3 Be careful for electric shock accident,and there must be safeguard procedures for X-ray harm
4 The equipment is complicated and the operating condition is strict. So in order to lengthen the
The working and storage place where there isn’t corrosive gases and the environment temperature
among-40℃~55℃,the rating humidity of the air ≤85%, the air should keep circulating , and the
The storage place should be clear ,dry and the rating humidity of the air ≤85% .
Keep the environment clean, keep from dust, in order to avoid short circuit.
During the operating, should notice whether the KV or the steady number is normal.
During the operating, set the switch and the button gently.
Should ensure that the cooling-fan works normally. after the operating, the switch
- 35 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
During the moving, avoiding shake and collision, and during the transporting, should
Before open the machine, should check that whether the cooling-fan is normal, in
order to ensure the X-ray generation can be cooled ,or else do not use the machine
x-ray machine If you need measure the console with power supply ,please care for electric
shock.The body and hand can not touch any parts with power open .
2 Firstly you should measure the console and connecting cable .If console and cables are not
3 If there is something wrong with X-ray generator,and the operator wants to measure the it .You
should let the air insulated (SF6) out and make it to -0.1Mpa by vacuums. After staying it for 20
4 Must take the air cooler of X-ray generator down ,then uncover the X-ray tube inserted.
5 Pay attention to it when uncover the X-ray tube inserted,must use the special tools.
6 After the x ray tube is taken off, loosen the screws around the generator barrel, lightly lift up
the barrel. Use an electric iron to weld off the tree lead-wires’ joints
7 Operator should wear radiation-proof clothes and caps, should be aware of the harm arising
by x ray tube’s self-exposure. Prepare a thick cloth bag to sheathe X-ray tube inserted.. Lift the
barrel, observing if there is any color change of X-ray tube inserted, sheathe the X-ray tube inserted
into the thick cloth bag, then weld off the connecting cable.
- 36 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
8 Carefully inspect the X-ray tube inserted if there is unusual color ,the fuse is broken ,the target
surface exits the oxidize trace ,there is a hole punched on the shell of tube.If one really exists,
indicate that there is something wrong with X-ray tube inserted .Must change new one.
9 The X-ray tube inserted replaced must be clear ,no dust.And dry it for 1 hour in the drying
cabinet.
10 Carefully measure the high voltage transformer if burn or puncture ,and if there is odor. If
there is really odor ,and we assure that there is something wrong with high voltage
11 Must cost 48 hours to dry the new high voltage transformer replaced,keep temperature of the
13 Gas inflation for the X-ray generator with SF6 gas, fulfilling it to 0.45Mpa(20℃) .You should
14 Finish gas inflation,make the marks (pressure value),and staying it steady for 12 hours.If
discover reduction voltage,please find the reason and resolve it and then have a test by power
supply.
15 When finish the inspection,operator must have a test for liability ,penetration , sensitivity and
others. Pay attention to accord to the “technology parameter” in manual and be in protective
working environment.
- 37 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
List 1:
- 38 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
HV stops during exposure (removing 3. If there is code PT, check the temperature increase
the failure according to the code of the X-ray generator and cooling fans. Change mA
displayed) board.
4. Please change display circuit board if the above
method is not okay.
1. Check the lamp.
Warning lamp of the X-ray generator
2. Check the cable and plug of the warning lamp.
doesn’t work.
3. Check the connection cable
- 39 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
4 The connect line should be connected correct otherwise will produce others result.
5 The resistance under the fictitious loading between the CH3 with CH5 must be from 10 Ω to
50Ω.
Adjustable power
If change the new board ,you do not still solve the problem . Please find the diode marked by
2744 and 2739,as the picture showing .
- 41 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
Please use the multimeter to the position as the picture show ,and testing the connection of 2
diodes .
If you get the “0”displaying on the screen and hear some warning voice , which mean the
diodes are broken .How to fix them ?
A . Change new one .
B. If you want to use the X-ray units urgently ,you can move this 2 diodes and make the console
work .
POWER
Alarm lamp
No.1connector No.1connector
No.2connector
No.2connector
- 42 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
2. Test the 2 piece of lines of switch ,the multimeter value should be “000”
STEP 1
STEP 2
STEP 3
If the multimeter do not show the “000”, which proof the switch is broken or the lines of
switch is broken.
- 43 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
The shift of
multimeter
should be as
some as the
picture
showing.
- 44 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
3. If you are not sure about it ,you can lease the temperature switch from the tube head and
test it as the picture showing
- 45 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
High voltage transformer primary current sudden The primary connector of HV transformer is
increase and the same time power switch fall destroyed.
off
- 46 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
The pressure of SF6 is below the safe using value Need to re fill the Gas in soon.
cable port must connect with anode(+) of Oscillograph,and other wire from 3 cable port must
normal waveform
Non-normal waveform
2. Faults analysis: It is very difficult to analyze the tube inserted is normal or broken.You should
prepare one ammeter for testing primary current(Use one wire from marked “5” cable port ,and
series connection with a ammeter whose measuring range is 10A for test the DC ). Or you can
connection a ammeter of mA (from marked “8” cable port ,the method is as the same as above but
If the primary current is very low, no mA, which display the lamp of tube inserted is
broken
If the primary current is high,no mA, which display the H.V coil is broken.
If the primary current is very high, mA>5mA and increasing when the KV is raising,which
The better method is that you can use the pliers ammeter to test tube and coil.
- 47 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
Tube inserted:measure the lamp,if it is normal,so display the tube inserted is good.
Coil:firstly look the surface is if it is burned,and smell its odor,finally measure its wire.
- 48 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
The curve line of the relationship between SF6 gap pressure and temperature
0.54
The pressure shown by pressure meter
0.5
0.46
Mpa
0.42
0.38
0.34
-20 0 20 40 60 80
The average temperature of X ray generator(°C); Inflation pressure error: setting time + 0.01Mpa
- 49 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
- 50 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
ISO 9001:2015
- 51 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
ISO 14001:2015
- 52 -
Manual Version 2022-FCS-2
PORTABLE X-RAY UNIT
ISO 45001:2018
- 53 -
Manual Version 2022-FCS-2