Professional Documents
Culture Documents
http://www.shimadzu.com/an/literature/gc/sic114141.html
Introduction
The Barrier Ionization Discharge (BID) detector generates a
17.7 eV helium plasma that ionizes almost all compounds
except Neon. A newly designed quartz dielectric chamber
allows for a lower discharge current and higher operating
temperature. The BID is a universal detector with sensitivity
greater than 100 times that of a TCD. It is an ideal detector
for trace levels of permanent gas, water, volatile fatty acids
and light hydrocarbons.
Two customized GC applications are presented. Two
6-port valves are used to inject, trap and release
permanent gases. A 4-port switching valve was used to
direct high concentration sample components to a TCD.
Compounds
Principle of Ionization Reaction
2
New Applications Using GC BID Detector
Structure of BID
He
Quartz glass tube
Helium plasma
(Dielectric) Three electrodes are placed on the quartz tube. High
voltage is applied to electrodes, and plasma is
Low frequency generated
power supply
Plasma generation
zone
Structure of BID
He
He carrier TCD Impossible to detect
TCD Analysis of O2,N2,CH4, CO
Ar carrier He
TCD Analysis of H2 MTN-1 FID
ppm levels of CO and CO2 are detected
3
New Applications Using GC BID Detector
N2
O2
600000
400000
300000
CARBON DIOXIDE
CARBON MONOXIDE
ACETYLENE,ETHYLENE
200000
100000
METHANE
ETHANE
H2
-100000
0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 min
Carbon Carbon
Hydrogen Oxygen Nitrogen Methane Ethylene Ethane Acetylene
Monoxide Dioxide
Concentration
(ppm) 100 2500 16900 100 100 400 100 100 100
4
New Applications Using GC BID Detector
N-BUTANE
PROPANE
ETHANE
N-PENTANE
150000
125000
N-HEXANE
METHANE
100000
75000
50000
25000
0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5 min
Concentration
(ppm) 100 100 400 100 100 100
5
New Applications Using GC BID Detector
O2
O2
N2
3500000
100000
CO2
3000000 90000
N2
80000
2500000 70000
C2H4/C2H6
60000
2000000
50000
1500000 40000
C2H2
CO2
30000
C2H4/C2H6
1000000
20000
CH4
C2H2
500000 10000
CH4
CO
CO
H2
0
0
-10000
2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0 22.5 min 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0 22.5 min
Conclusions
• The new Shimadzu GC BID is a universal, sensitive, and rugged detector that can be used in a variety of applications.