Instruction for Viewing the Service Manual
This manual is divided into Section 1: System, Section 2: D Module, Section 3: P Module, Section 4: E
Module, Section 5:Core/ISE, Section 6:Service PC(E170), Section 7:Up Grade.
Section 1 System
Chapter 1: Specification of the Unit
Chapter 2: Installation
Chapter 3: Trouble Shooting
Chapter 4: Maintenance
Chapter 5: Circuit Diagram
Section 2 D Module(D2400)
Section 3 P Module(P800)
Section 5 Core/ISE
Section 2 to Section 3 and Section 5 contains specification of each of the modules
(Chapter 1), Flow Path System (Chapter 2), Mechanical System (Chapter 3) and
Electrical System (Chapter 4). Operation that requires special adjustments and
notes have diagrams and instruction, so please read carefully before working on
them.
Section 4 E Module(E170)
Chapter 1: Specifications
Chapter 2: Flow Path System
Chapter 3: Mechanics
Chapter 4: Electronics
Chapter 5: Adjustment Specifications On E-170 Mechanical System
Chapter 6: Software Adjustment Manual
Section 7 Up Grade
Chapter 1: Hybrid modular retrofit manual
Chapter 2: Module extension procedure for Hybrid Modular
TABLE OF CONTENTS
Section 1 System
Section 2 D Module
Section 3 P Module
Section 4 E Module
Section 5 Core/ISE
Section 6 Service PC
Section 7 Up Grade
TABLE OF CONTENTS
System
Chapter 1 Specifications
Chapter 2 Installation
Chapter 3 Troubleshooting
Chapter 4 Maintenance
Section Page
1.1 Configuration Diagram of MODULAR (DPE) ....................................S1-1
CHAPTER 1 SPECIFICATIONS
Items Specifications
Method Discreet method of simultaneous analysis for multiple tests according to
analyzer module assembly (combinations).
Sample Cup Hitachi cups (standard y minute amount *) and blood-collecting tubes
(Note): Minute cups can be used only with ISE and P800.
Specify at the screen when using minute cups.
Method of Sample Loading Rack loading method for the Hitachi Racks (5 sample racks).
Number of Batches for Racks 300 samples (30 rack batch trays x2)
Rack Processing Method Distribution method in which the racks are captured by the various analyzer
modules.
Number of Items for Analysis Maximum of 100 items (according to configuration of the Analysis Modules)
Photometric (86 tests) + Calculation Test (8 tests) + Blood Serum Indexes (3
tests) + Electrolyte (3 tests)
Assay Method 1-Point End, 2-Point End, 3-Point Rate, 3-Point, Rate A, Rate B
Calibration Linear, K-factor, Isozyme, Non-linear methods.
Non-linear function = Maximum of 6 points
4-parameter Logit-Log, 5-parameter Logit-Log, 5-parameter exponential
function, spline function, polygonal line working curve
Calculation Functions Calculation between tests and data correction
Monitoring Functions Such as reaction process monitoring, data review, working curve and
calibration rates.
Quality Control Realtime quality control (binary method), quality control for samples within a
day and from a different day
Re-testing Function Available (automatic re-test, manual re-test)
Output Register Analysis results (monitor type, report type), reaction process monitoring,
work sheet, re-test list, reproducibility check, calibration load list, calibration
trace, alarm trace, screen hard copy, etc.
Maximum number of data Sample data (general, stat): 10000 samples
memorized Quality controlled samples: 60 types x 150 points
Reaction process data: 10000 tests
Printer Dot matrix 80 digits
CRT Model 17 color CRT. touch screen included
System I/F RS232C
Ground Special No. 3 type, ground resistance under 10 Ω.
Ambient temperature range 15~32°C (fluctuation during measurement was ±2°C)
∗E Module : 18~32 °C (± ±2 °C per hour)
Ambient humidity range 45~85%RH (must not have condensation)
Item Specifications
Operation mode/design Discrete, random access
Throughput Up to 170 tests per hour-module
Detection channels 2 channels per module (2 units of detection unit are incorporated)
Sample pipetting volume 10 to 50 µL
Sample liquid level detection Available
Reagent pipetting volume Assay; 50 to 80 µL, beads; 30 to 50 µL, and dilution; 50 to 190 µL
Reagent empty sensing Available
Reaction container Assay cup
Reaction-required volume The largest reaction volume of an assay cup is 160 µL.
Incubation disk A turntable disk, which can hold 54 assay cups.
Sample carrier Hitachi standard racks
Sample containers Hitachi standard cups, and test tubes (described separately)
Sample types Plasma, serum, and urine
Assay protocol 18-min assay and 27-min assay are available.
Automatic dilution Sample dilution is available.
Assay calibration Available
Instrument calibration Available (service function)
Quality control utility Available
Assay reagent Rack-pack type
Diluent Like Elecsys2010
Pretreatment Like Elecsys2010
Rack-pack reagent capacity The reagent disk can hold 25 packs (assay reagent, diluent, and
pretreatment).
Continuous loading of Not available
rack-packs
Special purpose SAP, BlankCell, and CellCheck (for assay performance test)
reagent for service
System reagent ProCell and CleanCell, Pre Clean M
System reagent container 2 bottles each of ProCell and CleanCell for 1000 tests/bottle.
Bottle switching is available.
System cleaning solution An adapter exclusive for Sysclean is used. Syswash for Elecsys2010
is not allowed to use.
Beads mixing Available
Disposable Dispensing-exclusive tips and assay cups are used (like
Elecsys2010).
Disposable capacity A total of 11 magazines, each of which can hold 1000 pieces each of
dispensing-exclusive tips and assay cups, is available for their
successive delivery.
Sample ID One-dimensional bar code is used.
Rack ID Available
Reagent ID Two dimensional bar code is used.
Item Specifications
Reagent disk temperature Available (20 ± 3 °C)
control
System reagent temperature Available (28 ± 2 °C at Preheat)
control (PreCell and CleanCell)
System reagent temperature Available (20 ± 1 °C at EWS Preheat)
control (Pre Clean M)
System cleaning solution Not available
temperature control
Incubator temperature control Available (37.0 ± 0.3 °C)
Detection unit temperature Available (28.0 ± 0.3 °C)
control
Temperature and humidity 18 to 32 °C and 45 to 85%
limits for satisfactory
performance
Temperature and humidity 5 to 35 °C and 30 to 95%
limits for successful operation
Temperature and humidity -20 to 75 °C and 5 to 95%
limits during storage and
transport
Allowed fluctuations in ±2 °C per hour
temperature for successful
operation
Solid waste 2 waste receptacle for used dispensing tips and assay cups
Liquid waste Common to the core
Water supply Common to the core
Power supply Power distribution with fuse free circuit breaker(32A,three phase),
AC400/230±10%(3/N/PE-three phase)50/60Hz
Power consumption 2 kVA (the core included)
Dimensions 1200 (W) × 800 (D) × 1200 (H) mm
Mass 380 kg (NET)
Heat dissipation (BTU) Described separately
Clot detection Available
Noise 65 dB or less during operation
55 dB or less while being idle
102 DEE 2740 4800 1100 1650 About 1710 8.0 110 Less than 65 About 17640
030 PPP 2400 4800 1100 1200 About 1300 8.0 150 Less than 65 About 19800
021 PPE 1770 4800 1100 1200 About 1430 8.0 130 Less than 65 About 18720
012 PEE 1140 4800 1100 1200 About 1560 8.0 110 Less than 65 About 17640
003 EEE 510 4800 1100 1200 About 1690 8.0 90 Less than 65 About 16560
400 DDDD 9600 6000 1100 1650 About 2200 10.0 200 Less than 65 About 25200
310 DDDP 8000 6000 1100 1650 About 2050 10.0 200 Less than 65 About 25200
301 DDDE 7370 6000 1100 1650 About 2180 10.0 180 Less than 65 About 24120
220 DDPP 6400 6000 1100 1650 About 1900 10.0 200 Less than 65 About 25200
211 DDPE 5770 6000 1100 1650 About 2030 10.0 180 Less than 65 About 24120
202 DDEE 5140 6000 1100 1650 About 2160 10.0 160 Less than 65 About 23040
121 DPPE 4170 6000 1100 1650 About 1880 10.0 180 Less than 65 About 24120
112 DPEE 3540 6000 1100 1650 About 2010 10.0 160 Less than 65 About 23040
103 DEEE 2910 6000 1100 1650 About 2140 10.0 140 Less than 65 About 21960
040 PPPP 3200 6000 1100 1200 About 1600 10.0 200 Less than 65 About 25200
Service Manual
Chapter 1.2.6
031 PPPE 2570 6000 1100 1200 About 1730 10.0 180 Less than 65 About 24120
022 PPEE 1940 6000 1100 1200 About 1860 10.0 160 Less than 65 About 23040
013 PEEE 1310 6000 1100 1200 About 1990 10.0 140 Less than 65 About 21960
004 EEEE 680 6000 1100 1200 About 2120 10.0 120 Less than 65
About 20880
Ver.1.0-May.2002
Service Manual
Chapter 1.2.6
Section 1 Modular Analytics/System Service Manual
2 1 0
ID No. 10 –th 10 –th 10 –th
(Rack Kind) digit digit digit
Stop Position
Detection
Hole
Item Combination
Types of Test Tubes φ13x75, φ13x100
φ16x75, φ16x100
Hitachi standard cup, micro cup
Combination of test tubes and Standard cup +Test tube (φ16x75, φ16x100)
sample cups µ Cup +Test tube (φ16x75, φ16x100)
However, µ Cup cannot be used for D-module and E-module.
④ Label Dimensions
17mm
Min. margin 5mm
Max. label length 85mm
Barcode Symbol
Max. 75mm
50mm
Min. margin 5mm
5mm
Min. 12mm
Min. 16mm
Table S1.4-1 Parts Where Reagent May Come In Contact or Pass Through
a Reagent cup e Stirrer
b Cell f Waste solution drain
c Rack g Rack transport line
d Pipetter Nozzle h Area surrounding these
(3) Others
c There are customers that prefer to have the cover for the line system removed, however, rack
transfer speed for model 7600 is very fast and it is dangerous, so please do not remove the cover.
d This is constructed so that parts can be replaced without removing the modules. Take precautions
not to damage the units, wiring or piping near by when replacing parts.
e In case analyzer problems should occur, please have on hand, minimum of following items in order to
investigate the cause.
Cautionary matters concerning the safety instructions are indicated with direction words such
as ”DANGER” , ”WARNING” , ”CAUTION” and the safety alert symbol as follows.
c d, k 11
f
e
16
f l
13, 15, 20
19
19
21 22 j i
(12): Top of the Printer
14
17, 18
g,h
(4) D-Module
h, i
4, 13, 14
5
10
8, 12
11
15
16
3
1
9
2
17
(5) P-Module
11
7
6
13
3
4 5
12
10
1
14
2
(7) E Module
Section Page
2.1 INSTALLATION REQUIREMENT .................................................................... S2-1
2.1.1 Shipping Conditions .................................................................................... S2-1
2.1.2 Environmental Conditions .......................................................................... S2-1
2.1.3 Installation Location .................................................................................... S2-2
2.1.4 Site Requirements ....................................................................................... S2-4
CHAPTER 2 INSTALLATION
ITEM CONDITIONS
Transport -20~75°C
Temperature
Humidity 5~95%
Vibrations ≤2G
Drop ≤30G
Impact ≤30G
Note) Possible damage may occur if exposed to environments that exceed the following conditions.
NOTE: A grounding terminal with the grounding resistance of 10Ω or less must be available not far from the
equipment.
D ra in P o w e r S u p p ly
(500) P u re W a te r S u p p ly
(300)
O p e ra tio n 1100
Loader M o d u le 1 U n lo a d e r
U n it
(1000)
Fig. S2.1-1
1100
Loader Module 1 Module 2 Unloader
Operation
(1000)
Fig. S2.1-2
1100
Loader Module 1 Module 2 Module 3 Unloader
Operation
(1000)
Fig. S2.1-3
1100
Load Module 1 Module 2 Module 3 Module 4 Unloader
Operation
(1000)
Fig. S2.1-4
For the purpose of maintaining safety when installing the device, a 50cm or more space is required in the rear
area, 1m or more in the front, and 50cm or more on both the left and right sides.
Installation diagram, connection diagram and specifications for grounding are shown below; please refer to
those for determining the installation area.
5 Refrigerator Arrange for a refrigerator with an appropriate capacity for the storage of
reagents and samples.
7 Drain opening Prepare a drain opening within 5m from drain opening in the back of the unit
with a diameter that is 50mm or more and height from the bottom of 100mm or
less.
Check the items purchased by referring to the purchase order and making entries in the detailed
statement.
2.2.2 Unpacking
(4) E Module
1. At Transportation
Insert the fork lift in the allowable dimension range of 720 mm as shown in Fig.S2.2-1.
It is allowed to insert the folk lift only from the front position of the instrument as illustrated below.
When using the fork lift, take sufficient care not to damage the external covers.
(940)
Fork lift
(1140)
(80)
(Allowable dimension
range of fork lift insertion)
1-2) When moving the instrument on its casters, use the four corners of the instrument
frame (approx. 100 mm) as shown in Fig.S2.2-2.
Do not apply a force to the cover posts.
2. At Installation/Demounting
When installing the instrument onto the core line or demounting the instrument from it, use the handles
indicated in Fig.S2.2-3.
Do not apply a force to the cover posts as noted in Fig.S2.2-2.
Handle
Handle
Fig.S2.2-3 At Installation/Demounting
Printer
CRT
Keyboard
PC Rack
PC
CRT Cable
Touch Screen
Power Cable
Cable
Fig. S2.2-5 CRT power cable connections on the back of the CRT
Printer
Power Cable
Touch Panel
Mouse
Printer
Keyboard
CRT
e Adjust the height of Main Line and the perpendicularity of the line, then secure the joints with 4
bolts with hexagon holes.
(Notes)
The Main line should be 0.5-1.5mm higher than the C2 line and both D and J line are straight.
Do not move the connection plate for the joints since it has been adjusted at the factory.
d After the No.3 module line, the assemblies are the same as the No.2 module line.
(5) Unloader
Connect the Unloader vertically against the module line by making the same adjustments as the Loader.
Level by turning the adjuster to the Unloader. After adjusting the height and perpendicularity of the
Unloader with the module line, secure it to the connector with 4 bolts.
Beam Sensor
Unloader
Loader
G-J1
G-J2
D-J2
Belt
Unloading
Loading
GND for
Various
G1 Unit Cables
Connector D-J2 J-36
J-33
DC Power
AC Power Cable
DC Power Cable
(Module Side)
Signal Cable
Fig. S2.2-22 Wiring under The Main Line and The Inceased Line
① Attach the L pipes for both diluted and concentrated wasted water drain.
② Remove the wasted water drain cup fixed the main line or increased lines.
③ Remove the part of the vacuum pump unit for transportation.
④ Set the level the reaction disk and roughly adjust by turning the adjuster casters so that the center
of the L-line is approximately 791mm. (Should be leveled)
⑤ Connect the Analyzer module to the appropriate line base by matching it with the guide pin.
⑥ Find the level, then adjust by using the adjustment jig so that the distance between the main line
and L-line is 91-94mm and level; secure by lowering the 4 adjusters. (Adjuster casters are
raised.)
⑦ Check to see that the space between the Rack and the Guide at the stopping position of the Rack
at the E-Handling position (C2, line feed END position) is 2~3mm.
⑧ Adjust by grabbing the rack with the E-Line. When it is moved from the D-Line to the L1-Line
manually, the space between the bottom of the Rack and L1 is 1~2mm when an adjustment jig is
used.
⑨ Grab the Rack with the F-line and make sure that the rack overlap the sensor of F-D9 from 9 to
12mm when it is moved manually from the L2 line to the D-line. And then space between the
bottom of the Rack and D line is 1~2mm when an adjustment jig is used.
⑩ Fill the Raction Bath with deionized water. Adjust the water level by turning the adjuster caster
while measuring with a scale the four points; front, back, left and right of the position of water in the
Reaction Bath.
⑪ Check again to see that the difference in the water level in the Reaction Bath is within 1mm.
⑫ Secure D-module and No.1 module main line with connection bolts. (Bolt with hex hole x2)
⑬ Recheck the relationship between the L-line and the D-line of g and h by grabbing the Rack with
the E/F-lines.
⑭ Recheck the relationship between the L-line and the D-line of g and h by grabbing the Rack with
the E/F-lines.
D Module P Module
(2) Signals
LON-J1
Main line 2 CM-J11 Main line 1
CM-J11
HUB1 HUB2
Port Destination Port Destination
1 Core 1 1 M4
2 Core 2 2 M5
3 Operation PC 3 M6
4 M1 4 Not Used
5 M2 5 Not Used
6 M3 6 Not Used
7 HUB2 Port 7 For SVP
8 For Debugging 8 HUB1 Port
E J4
E J5
C-Line Shutter
E J3
E J4
E-J4
E-J5
Fig. S2.2-30 Wiring Between the Units (No.2 Module and later)
(2) Insert the beads mixer paddle into the BM Joint (P/N 768-3509). Be certain that the paddle is fully,
approximately 17 mm, inserted until the stopper position in the BM Joint .
(3) Use a hex wrench to fasten two M3 hexagon socket head screws that fix the beads mixer paddle.
It should be better to fasten one screw completely before the other.
(4) Verify by manually turning the BM Joint that the beads mixer paddle is not eccentric. If eccentric, adjust
the installation so that the paddle is not eccentric. If the paddle were left eccentric, reagent bubbles
would be produced, which cause erroneous sensing of the reagent level.
M3 hexagon
socket head
screws (2)
M3 hex wrench
BM Joint
(P/N 768-3509)
∗ In the same manner as for the beads mixer paddle, mount the reagent probe and sample probe also
(separately packed).
After carry out the “ RESET” on the E170, attach the Beads Mixer Paddle
as the direction of arrow part in Figure E2.2-33.
Don’t rotate the shaft of the Beads Mixer at the Home position in the meantime.
Arm head
Holder
To be secured with W point
screws (M3 × 3) (2 points)
Fig. S2.2-33
After mounting the sample nozzle, carry out LDD voltage adjustment referring to Section 4 (Chapter 4.7).
4. Measurement Cell
(1) You had better shake a reference electrode to be installed a few times against crystallization of KCl
powder filled inside the electrode. Then, install the reference electrode in the detection unit.
Measurement cell
Reference electrode
(2). Open the lid of the detection unit, and remove cushioning material that fix the tube for sipper, the tube
for sipper syringe, and the magnet arm, which are affixed before factory shipment for transport
protection.
NOTICE: Turn off the power before opening the lid of the detection unit; the photomultiplier would be
destroyed if the lid is opened with the high voltage being applied.
(3) Remove two screws that fix the cell-mounting plate, and draw out the cell plate toward you.
(4) Fasten two M4 × 16 screws that fix the measurement cell. (The two screws are secured by a tape
inside the unit before factory shipment.)
Remove two screws that fix the cell-mounting Fasten two M4 × 16 screws that fix the
plate, and draw out the cell plate toward you. measurement cell.
(5) Verify and ensure that there is no play between the cell frame and the measurement cell.
A B
B-B
Screw M4 × 16
PMT
Cell frame
Measurement cell
(6) Run the lead wires from the measurement cell not to be entangled with the magnet arm when it is in
action.
(7) Connect and tighten both the tube for sipper and the tube for sipper syringe.
A loose connection of the tube can cause flow irregularities.
GND
Power HOT
Supply
COMMON
N L G
Main Unit
Fig. S2.2-36 Drain Water Receptacle (No.2 Module) Fig. S2.2-37 Drain Water to Outside
[ISE Unit]
c For the concentrated waste solution, insert one end of the drain tube to ISE drain nipple located at the
back of the Loader.
d Connect the tube for the low-concentration waste solution to No.1 Module drain water receptacle (on
the low-concentration waste solution side).
b) P-Module
① Front of the Module
Detergent1: High-Alkaline D
Detergent2: Hydrogen
② Reagent Refrigerator
1D1, 2D1: High-Alkaline D
1D2, 2D2: High-Calnon
2D3 : Hydrogen
c) E-Module
① Set each ProCell (two bottles) at position PC1/PC2 (on the lower front of the E170 unit).
② Set each CleanCell (two bottles) at position CC1/CC2 (on the lower front of the E170 unit).
③ Set each PreClean (two bottles) at Pos 1/2 (on the left front cover of the E170 unit).
④ Set magazines (up to 12 magazines) in the magazine storage section.
⑤ Set two waste material boxes in the solid waste box.
⑥ Set the PC reservoir and the CC reservoir in place.
(2) Place deionized water in the supply water tank and the reaction baths.
(3) Turn ON the power to the Operation area.
(4) Turn ON the main breaker after turning ON the module breaker.
(7) Press the buttons PC1/PC2, CC1/CC2, PreClean 1/2, and Solid Waste Box 1/2 of E-Module.
Functions of the PC1/PC2, CC1/CC2, PreClean 1/2, and Solid Waste Box 1/2 buttons:
c Select “Mechanical Check” from “Check” on the Maintenance Screen in the Utility Job and press START.
d Select an “Analyzer Unit” for this operation as an operating module and press START after entering the
[Confirmation Items]
P-Module
a) The tip of the probe must in the center of the S-cup and not touch the bath in the Rinsing Bath.
(S probe only)
b) The tip of the probe must be cushioned within the cells. (S probe only)
c) There must be sufficient rinsing water for probes in the rinsing bath.
d Rinsing Mechanism
e Stirrer
b) There must be sufficient rinsing water for stir sticks in the rinsing bath.
D-Module
a) The tip of the probe must be in the center of the Cell, S-Cup; it must not hit the rinsing bath.
c) There must be sufficient rinsing water for the probes in the rinsing bath.
d Rinsing Mechanism
e Stirrer
b) There must be sufficient rinsing water for stir stick in the rinsing bath.
b) There must be sufficient rinsing water for reagent nozzle in the rinsing bath.
(2.5) Adjustment of water volume of the rinse unit and confirmation of vacuum pressure
Execute the mechanism check and adjust the water volume wit the valves
Check to make sure that there is no alarm related to LSS being generated with the P-Module cell rinsing
c Start the operation without reagent in the tube and make sure that the alarm is generated for all LSS.
d Fill each bottle with deionized water and let it prime sufficiently; then with tubes filled with water start the
e Start the analysis without reagent in the tube and confirm that the alarm is generated for all LSS.
c Execute “Disk Check” from “Check” on the Maintenance screen in the Utility Job.
c Execute “BCR Check” from “Check” on the Maintenance screen in the Utility Job.
e Press the Print button, select “Print Cumulative Information”, and press the view button.
e Lamp (both P and D): Check the water pressure by pressing the cold water tube.
Check adjustment in each mechanisms according to the procedural steps shown below.
For details, refer to Section 4-6/4-6 - Adjustment Procedures.
Magazine drawer
Incubator
Beads Mixer
Cap Opener
S.Probe
Gripper
Sipper
W2 Gripper
W2 Sipper
Take the following checkup procedures after making sure that each unit has been adjusted properly.
① Check that each ProCell (two bottles) is set properly at position PC1/PC2 (on the lower front of the E170
unit).
② Check that each CleanCell (two bottles) is set properly at position CC1/CC2 (on the lower front of the
E170 unit).
③ Check that each PreClean (two bottles) is set properly at Pos 1/2 (on the left front cover of the E170 unit).
④ Check that at least six magazines (up to 12 magazines) are set properly in the magazine storage section.
⑤ Check that two waste material boxes are set properly in the solid waste box.
⑥ After completion of inventory check, open the "Reagent Overview" window on the control PC and check
the remaining amount of each reagent.
① On the control PC, select "Beads Mixer Wash" and specify "5" as a repeat count.
Then, carry out "Beads Mixer Wash".
Check that washing water is supplied from the beads mixer washing bath at the time of operation.
② On the control PC, select "Pipetter Air Purge" and specify "5" as a repeat count.
Then, carry out "Pipetter Air Purge".
Check that reagent pipetter inside washing water, reagent pipetter outside washing water, and sample
pipetter inside washing water are supplied at the time of operation.
③ On the control PC, select "Sipper Wash" and specify "5" as a repeat count. Then, carry out "Sipper
Wash".
Check that washing water is supplied from the shipper washing bath at the time of operation.
④ On the control PC, select "System Reagent Wash" and specify "5" as a repeat count. Then, carry out
"System Reagent Wash".
Check that washing water is fed from the PC/CC reservoir supply nozzle at the time of operation.
⑤ On the control PC, select "W2 Sipper Air Purge" and specify "5" as a repeat count.
Then, carry out "W2 Sipper Air Purge".
Check that washing water is fed from the tip of the W2 sipper probe at the time of operation.
① On the control PC, select "System Reagent" for "Reagent Prime" and specify "5" as a repeat count. Then,
carry out "Reagent Prime".
Check that PC/CC is fed from the PC/CC reservoir supply nozzle at the time of operation.
② On the control PC, select "W2 Sipper" for "Reagent Prime" and specify "5" as a repeat count.
Then, carry out "Reagent Prime".
Check that PreClean is fed from the tip of the W2 sipper probe at the time of operation.
2 Operation Check
According the following procedure, carry out "Mechanism Check" to ensure that the mechanism operations
are normal.
① Set ten empty racks on the rack entry part located on the core.
② On the control PC, select "Mechanism Check" for "Check".
③ Select "All", and carry out "Mechanism Check".
④ After one hour of operation, click the STOP button on the control PC window to stop the instrument.
⑤ Check that no alarm has occurred during the operation.
At the time of installation, carry out analytical performance check according to the procedure mentioned below.
Before taking this procedure, make sure that the Service PC is connected with the instrument.
1) MC Preparation
3) PMT HV Adjustment
5) BlankCell Calibration
6) Assay Calibration
(1) MC Preparation
① Check that ProCell/CleanCell is set at a predetermined position on the instrument and that the remaining
amount of ProCell/CleanCell is more than 10% per module.
② On the Service PC, select "MC Preparation" for "Maintenance" and then select channel "Detection Unit
1&2". Specify "50" as a repeat count, and click the "Start" button.
③ At the end of MC preparation, check that the "Abnormal MC Condition" alarm 318-001 or 318-003 is not
indicated. At this step, ignore the "Abnormal MC Condition" alarms 318-002 and 318-004.
④ If the "Abnormal MC Condition" alarm 318-001 or 318-003 is indicated at step (3), check whether the
measurement cell connection in the detection unit is disconnected or not.
If it is disconnected, reconnect the measurement cell connector and then carry out "MC Preparation" 50
times.
⑤ If the measurement connector is not disconnected, just carry out "MC Preparation" 50 times.
If the "Abnormal MC Condition" alarm 318-001 (Ch 1) or 318-003 (Ch 2) is indicated at the end of
execution, replace the measurement cell with a new one.
① Make sure that ProCell/CleanCell is set at a predetermined position on the instrument and that the
remaining amount of ProCell/CleanCell is more than 10% per module.
② On the Service PC, select "System Volume Check" for "Service" and then select channel "Detection Unit
1&2". Then, click the Start button.
③ At the end of system volume check, check that the "System Volume Check Error" alarm 321-001, 002,
003 or 004 is not indicated.
④ If the alarm 321-001, 002, 003 or 004 is indicated, check whether the joint parts of the sipper probe, tube
and detection unit are secured properly.
① Check that ProCell/CleanCell is set at a predetermined position on the instrument and that the remaining
amount of ProCell/CleanCell is more than 10% per module.
② Set the BlankCell rack pack at position 1 on the reagent disc. Make sure that a barcode label for E170 is
attached on the BlankCell rack pack.
③ Close the reagent disc cover, and carry out "Reagent Registration".
④ On the control PC, open the "Reagent/Status" window and check that the BlankCell rack pack has been
registered.
⑤ Make sure that the magazine (Tip/Vessel) is set at a predetermined position on the instrument and that
more than 32 pieces remain per module.
⑥ On the Service PC, select "PMT HV Adjustment" for "Service" and then select channel "Detection Unit
1&2". Then, click the Start button.
⑦ At the end of PMT HV adjustment, make sure that the "PMT HV Adjustment Error" alarm 320-001 or 002
is not indicated.
⑧ If the alarm 320-001 or 002 is indicated, select "System Parameter" from the "DB List" on the "Utility"
window of the Service PC corresponding to the module in which the alarm has occurred.
Then, click the Cook button located at the lower right of the window. Then, the Excel will be launched.
Select the "PMT HV Adjust Sheet", and check the results of BCR2 measurement at PMT HV Adjustment.
① Check that ProCell/CleanCell is set at a predetermined position on the instrument and that the remaining
amount of ProCell/CleanCell is more than 20% per module.
② Check that PreClean (washing buffer) is set at a predetermined position on the instrument and that the
remaining amount of PreClean is more than 10% per module.
③ Check that the magazine (Tip/Vessel) is set at a predetermined position on the instrument and that more
than 130 pieces remain per module.
④ Referring to the table shown below, set the reagent rack packs on the reagent disc.
⑤ Referring to the table shown below, set the samples on the APC rack (banana rack).
For these samples, use the Hitachi standard cups. On the APC rack, position 1 is located on the front
side and position 5 is located on the rear side.
∗ As to TSH Cal. 1 and Cal. 2, the entire volume contained in the calibrator bottle should be put into each
standard cup.
⑥ On the Service PC, open the "Maintenance" window and select "Service" on it.
Then, select "Assay Performance Check" and click the Perform… … button.
⑦ The Assay Performance Check window will then open. On this window, select the "AM Test" and "TSH
Test" check boxes and then select channel "Detection Unit 1&2".
Then, click the Start button.
⑧ After completion of Assay Performance Check, carry out data processing using the data conversion Excel
file stored in the Service PC.
① Check that ProCell/CleanCell is set at a predetermined position on the instrument and that the remaining
amount of ProCell/CleanCell is more than 10% per module.
② Set the BlankCell rack pack at position 1 on the reagent disc. Make sure that a barcode label of E170 is
attached on the BlankCell rack pack.
③ Close the reagent disc cover, and carry out "Reagent Registration".
④ On the control PC, open the "Reagent/Status" window and check that the BlankCell rack pack has been
registered.
⑤ Make sure that the magazine (Tip/Vessel) is set at a predetermined position on the instrument and that
more than 12 pieces remain per module.
⑥ On the Service PC, select "BlankCell Calibration" for "Service" and then select channel "Detection Unit
1&2". Then, click the Start button.
⑦ At the end of BlankCell Calibration, make sure that the "BlankCell Error" alarm 319-001 or 003 is not
indicated.
⑧ If the alarm 319-001 or 003 is indicated, select "System Parameter" from the "DB List" on the "Utility"
window of the Service PC corresponding to the module in which the alarm has occurred.
Then, click the Cook button located at the lower right of the window.
Then, the Excel will be launched. Select the "BlankCell Sheet", and check the results of BCR1/BCR2
measurement for deviation from any standard reference.
After completion of BlankCell Calibration, carry out any required assay calibration.
Performance test on the E170 immunoassay module will be carried out in the following sequence.
However, the entire procedures are not always required. Refer to each section when the particular
procedure is required.
PMT HV Adjustment
The system volume check will be carried out when one of the following situations takes place.
• After installation
• When the measurement cell is replaced
• When either the sipper tube is replaced or the tube junction is disconnected
Maintenance
Maintenance Utility
Utility
Select
Select
④ Select “System Volume Check,” No. 8 of Maintenance Items, on the right of the window, and then touch
or click Select button on the bottom of the window.
⑤ Select a target module and channels(s) to execute the system volume check, and then touch or click
Execute button on the bottom of the window; the equipment replaces the reservoir PC, and begins the
system volume check.
Moduleの選択
Module
の選択 selection
Channel選択
Channel
選択 selection
Execute
Execute
② Click the Maintenance tab to open a window similar to that shown below.
③ Select “Service” in the Category field and “System Volume Check” in the Item field.
④ Either double-click the “System Volume Check” or click Perform button on the bottom of the window.
⑤ Then, a window similar to that shown below appears. Select a target channel to execute the system
volume check, and click Start button on the bottom of the window; the equipment replaces the reservoir
PC, and begins the system volume check.
2 PMT HV Adjustment
NOTE: The following check must be Performed with new ProCell M bottles.
The PMT HV adjustment will be carried out when one of the following situations takes place.
• After installation
• When the PMT is replaced
• When the measurement cell is replaced
② Place a blank cell rack pack on an empty position on the reagent disk, and then close the cover of the
reagent disk. Verify that the blank cell rack pack is registered on the Setting window of Reagent job
when the reagent disk scan is carried out subsequently.
⑤ Select “PMT HV Adjustment,” No. 7 of Maintenance Items, on the right of the window, and then touch or
click Select button on the bottom of the window.
⑥ Select a target module and channels(s) to execute the PMT HV adjustment, and then touch or click
Execute button on the bottom of the window; the equipment carries out the pre-operation and then the
PMT HV adjustment.
ModuleModuleの選択
selectionの選択
Channel selection
Channelの選択
の選択
Execute
Execute
⑦ PMT HV adjustment is carried out as follows: Obtain the mean of three consecutive measurement
values of BC R2. Then compare the mean with the target value specified in the bar code information
of the blank cell rack pack whether the mean satisfies the target value ± 5%. If the requirement is not
met, the PMT high voltage will be automatically modified and the measurement procedure will be
repeated until the requirement is satisfied.
⑧ However, if the requirement is not satisfied for five consecutive repetitions, an alarm message of “PMT
HV Adjustment Fail” (320-001) will be displayed, and the adjustment will be terminated.
⑨ When the alarm message of “PMT HV Adjustment Fail” (320-001) is displayed, examine the tubing of
the measurement cell in the detection unit, and carry out “MC Preparation” 50 times. Then, carry out
the PMT HV adjustment again.
② Place a blank cell rack pack on an empty position on the reagent disk, and then close the cover of the
reagent disk. Verify that the blank cell rack pack is registered on the Setting window of Reagent job
when the reagent disk scan is carried out subsequently.
④ Select “Service” in the Category field and “PMT HV Adjustment” in the Item field.
⑤ Either double-click the “PMT HV Adjustment” or click Perform button on the bottom of the window.
⑥ Select a target module and channels(s) to execute the PMT HV adjustment, and then touch or click
Start button on the bottom of the window; the equipment carries out the pre-operation and then the
PMT HV adjustment.
NOTE: The following check must be Performed with new ProCell M bottles.
The assay performance check will be carried out when one of the following situations takes place.
• After installation
• When the measurement cell is replaced
• When the PMT is replaced
• When either the sipper tube is replaced or the tube junction is disconnected
The Service PC only is available for the assay performance check. Follow the steps given below.
• AM Test
② Place each kind of reagent on the reagent disk according to the requirement given in Table S2.4-1
below, and close the cover of the reagent disk.
③ Place each sample on the sample rack according to the requirement given in Table S2.4-2 below by
referring to the sample location diagram for assay performance check shown in Fig.S2.4-8.
NOTE: Beads buffer and BCR2 are to get from RD(rack pack) position iSAP and BlankCell.
Position
Position 5 5
L-Line Sampling
L-Line Sampling position
position
Washing
Washing Station
station
Incubator position
Incubator position (No.54)
(no. 54)
Tip waste position1
Tip waste position 1
Tip waste position 2
Tip waste position 2
Incubator
Incubator
Position11
Position
Stripper
Stripper plate
plate
Tip buffer position
Tip Buffer position
⑤ Select “Service” in the Category field and “Assay Performance Check” in the Item field.
⑥ Either double-click the “Assay Performance Check” or click Perform button on the bottom of the
window.
⑦ Then, check the “AM Test” check-box among APC Test field on the Assay Performance Check
window (Fig.S2.4-9), which appears subsequently.
⑧ On the same window, select target channel(s) to execute the assay performance check and click OK
button; the equipment carries out the pre-operation and then the assay performance check.
⑨ When the assay performance check is complete, the assay results will be saved into “Result.dat” file in
“C¥/E-Mosys/pc/database/module1/pattern1” directory. Notice that the “module1” corresponds to the
first module (module1), the second module (module2), and so forth viewed from the core module.
⑩ Table S2.4-3 shows the test items, number of replicates, and result calculation formulae.
Table S2.4-4 shows for each test item the requirement specifications and countermeasures when the
corresponding requirement is not satisfied. If the AM test results do not satisfy the requirements, take
appropriate actions according to Table S2.4-4.
Number of
Test Item Calculation Formula
Replicates
Sensitivity — {(Ru-complex concentration contained in BCR2 solution) ×
-12
(2-SD value of BCR1 photometric values) × 3 × 10 }/(mean
of BCR2 photometric values - mean of BCR1 photometric
values)
BCR1/ (BCR2 5 Mean of BCR1 photometric values/(mean of BCR2
normalized) photometric values/target value of BCR2 photometric values)
R2 10 Mean of BCR2 photometric values
luminescence
BCR2-D 10
iSAP 3
Carry-over 3 (Mean of three AB photometric values after three cycles of
alternate measurements of iSAP and AB - mean of BCR1
photometric values)/(mean of three iSAP photometric values
after three cycles of alternate measurements of iSAP and AB)
• AM2 Test
② Place each kind of reagent on the reagent disk according to the requirement given in Table S2.4-5
below, and close the cover of the reagent disk.
③ Place each sample on the sample rack according to the requirement given in Table S2.4-6 below by
referring to the sample location diagram for assay performance check.
⑤ Select “Service” in the Category field and “Assay Performance Check” in the Item field.
⑥ Either double-click the “Assay Performance Check” or click Perform button on the bottom of the
window.
⑦ Then, check the “AM2 Test” check-box among APC Test field on the Assay Performance Check
window (Fig. S2.4-10), which appears subsequently.
⑧ On the same window, select target channel(s) to execute the assay performance check and click OK
button; the equipment carries out the pre-operation and then the assay performance check.
⑨ When the assay performance check is complete, the assay results will be saved into “Result.dat” file in
“C¥/E-Mosys/pc/database/module1/pattern1” directory. Notice that the “module1” corresponds to the
first module (module1), the second module (module2), and so forth viewed from the core module.
⑩ If the AM2 test results do not satisfy the requirements, take appropriate actions according to Table
S2.4-4.
• AM3 Test
② Place each kind of reagent on the reagent disk according to the requirement given in Table S2.4-7
below, and close the cover of the reagent disk.
③ Place each sample on the sample rack according to the requirement given in Table S2.4-8 below by
referring to the sample location diagram for assay performance check.
⑤ Select “Service” in the Category field and “Assay Performance Check” in the Item field.
⑥ Either double-click the “Assay Performance Check” or click Perform button on the bottom of the
window.
⑦ Then, check the “AM3 Test” check-box among APC Test field on the Assay Performance Check
window (Fig. S2.4-11), which appears subsequently.
⑧ On the same window, select target channel(s) to execute the assay performance check and click OK
button; the equipment carries out the pre-operation and then the assay performance check.
⑨ When the assay performance check is complete, the assay results will be saved into “Result.dat” file in
“C¥/E-Mosys/pc/database/module1/pattern1” directory. Notice that the “module1” corresponds to the
first module (module1), the second module (module2), and so forth viewed from the core module.
⑩ If the AM3 test results do not satisfy the requirements, take appropriate actions according to Table
S2.4-4.
• TSH Test
② Place each kind of reagent on the reagent disk according to the requirement given in Table S2.4-9
below, and close the cover of the reagent disk.
③ Place each sample on the sample rack according to the requirement given in Table S2.4-10 below by
referring to the sample location diagram for assay performance check.
⑤ Select “Service” in the Category field and “Assay Performance Check” in the Item field.
⑥ Either double-click the “Assay Performance Check” or click Perform button on the bottom of the
window.
⑦ Then, check the “TSH Test” check-box among APC Test field on the Assay Performance Check
window (Fig. S2.4-12), which appears subsequently.
⑧ On the same window, select target channel(s) to execute the assay performance check and click OK
button; the equipment carries out the pre-operation and then the assay performance check.
⑨ When the assay performance check is complete, the assay results will be saved into “Result.dat” file in
“C¥/E-Mosys/pc/database/module1/pattern1” directory. Notice that the “module1” corresponds to the
first module (module1), the second module (module2), and so forth viewed from the core module.
⑩ Table S2.4-11 shows the test items, number of replicates, and result calculation formulae.
Table S2.4-12 shows for each test item the requirement specifications and countermeasures when the
corresponding requirement is not satisfied. If the TSH test results do not satisfy the requirements, take
appropriate actions according to Table S2.4-12.
Number of
Test Item Calculation Formula
Replicates
TSH LDL (9) — 2-SD value of TSH Cal. 1 (results of the second through
the tenth)/tangent
TSH Cal. 1/ 8 Mean of TSH Cal. 1 photometric values (mean of results
(BCR2 of the second through the tenth)/(mean of BCR2
normalized) photometric values/target value of BCR2 photometric
values)
TSH Cal. 1/Cal.2 8:(Without PW)
5:(With PW)
TSH 1st value 8 The first photometric value of TSH Cal. 1 / (mean of TSH
High Cal. 1 photometric values of the second through the tenth)
Tangent — {(Mean of TSH Cal. 2 photometric values) - (mean of TSH
Cal. 1 photometric values of the second through the
tenth)}/{(TSH Cal. 2 concentration) - (TSH Cal. 1
concentration)}
Tangent/TSH — Tangent/mean of TSH Cal. 1 photometric values
Cal. 1
(cont’d)
Test Item Requirements Countermeasures
TSH LDL (9) ≦ 0.005 1. If the CV values of TSH Cal. 1 are high, the junction
(µLU/mL) between the sipper and the tube is suspected.
Replace them, if needed.
2. If the TSH Cal. 1 photometric values are high, reagents
or calibrators are suspected. Check their serviceable
life or lot numbers. Replace them, if needed.
3. If the TSH Cal. 2 photometric values are low, reagents
or calibrators are suspected. Check their serviceable
life or lot numbers. Replace them, if needed.
TSH 1st value ≦ 8.0 % 1. If the first photometric value of TSH Cal. 1 is high,
High reagents or calibrators are suspected. Check their
serviceable life or lot numbers. Replace them, if
needed.
2. It is suspected that the electrode in the measurement
cell is unstable. Carry out “MC Preparation” 50 times,
and then, carry out the TSH test again.
3. Replace the measurement cell.
TSH Cal. 1 CV ≦ 5.0% The junction between the sipper and the tube is suspected.
(With PW) Replace them, if needed.
TSH Cal. 2 CV ≦ 3.0% 1. If the CV is high, examine whether or not the
(With PW) measurement cell is inclined or the magnet is off the
normal position.
2. If the photometric value is low, examine whether the
electrode cable is not entangled at the magnet-housing
hole in the measurement cell.
3. Reagents or calibrators are suspected.
Check their serviceable life or lot numbers.
Replace them, if needed.
Cal.1 Factor ≧ 0.5
(With PW)
Cal.2 Factor ≧ 0.7
(With PW)
• AM+TSH Test
② Place each kind of reagent on the reagent disk according to the requirement given in Table S2.4-13
below, and close the cover of the reagent disk.
Table S2.4-13 Requirement on Reagent Placement on the Reagent Disk
③ Place each sample on the sample rack according to the requirement given in Table S2.4-14 below by
referring to the sample location diagram for assay performance check.
⑤ Select “Service” in the Category field and “Assay Performance Check” in the Item field.
⑥ Either double-click the “Assay Performance Check” or click Perform button on the bottom of the
window.
⑦ Then, check the “AM+TSH Test” check-box among APC Test field on the Assay Performance Check
window (Fig. S2.4-13), which appears subsequently.
⑧ On the same window, select target channel(s) to execute the assay performance check and click OK
button; the equipment carries out the pre-operation and then the assay performance check.
⑨ When the assay performance check is complete, the assay results will be saved into “Result.dat” file in
“C¥/E-Mosys/pc/database/module1/pattern1” directory. Notice that the “module1” corresponds to the
first module (module1), the second module (module2), and so forth viewed from the core module.
⑩ If the AM+TSH test results do not satisfy the requirements, take appropriate actions according to Table
S2.4-4/S2.4.11/S2.4.12.
NOTE: The following check must be Performed with new ProCell M bottles.
• After installation
• When the measurement cell is replaced
• When the PMT is replaced
② Place a blank cell rack pack on an empty position on the reagent disk, and then close the cover of the
reagent disk. Verify that the blank cell rack pack is registered on the Setting window of Reagent job
when the reagent disk scan is carried out subsequently.
⑤ Select “Blank Cell Calibration,” No. 6 of Maintenance Items, on the right of the window, and then touch
or click Select button on the bottom of the window.
. Select a target module and channels(s) to execute the blank cell calibration, and then touch or click
Execute button on the bottom of the window; the equipment carries out the pre-operation and then the
blank cell calibration. Leave the default comparison limit values intact.
Moduleの選択
Module selectionの選択
Channel selection
Channelの選択
の選択 Comparison
Comparison Limit Limit
Execute
Execute
② Place a blank cell rack pack on an empty position on the reagent disk, and then close the cover of the
reagent disk. Verify that the blank cell rack pack is registered on the Setting window of Reagent job
when the reagent disk scan is carried out subsequently.
④ Select “Service” in the Category field and “Black Cell Calibration” in the Item field.
⑤ Either double-click the “Black Cell Calibration” or click Perform button on the bottom of the window.
⑥ Select a target module and channels(s) to execute the blank cell calibration, and then touch or click
Start button on the bottom of the window; the equipment carries out the blank cell calibration.
BlankCell calibration has following four criteria. BlankCell calibration is failed and flagged when at least
one of the following quality criteria is judged as not OK.
(Mean value of (median/target) on other channels) – ((median/target) on that channel) × 100 >
comparison limit [%]
In case there is no valid result for any detection channels on that modular system, following formula is
applied.
2.5 Tentative Procedure for Shut-Down of Elecsys MODULE when it is not Used for
a while a Week or so (drafted)
2.5.1. When the Elecsys MODULAR System is not Used for a while a Week or so
Before the entire Elecsys MODULAR System is shut down, the following procedure is recommended in
order to get rid of any troubles afterward.
2. Make sure that no tips/vessels are mounted on the incubator and anywhere on the system.
If any, dispose of them.
3. Make sure that no magazines are mounted on the magazine fixer. If any, remove them.
4. Remove magazines from the magazine lifter. Make sure that tips/vessels mounted on the magazine
have no foreign objects, dirt or water. Properly preserve them to avoid being deformed.
5. Remove used magazines from the magazine receiver, and dispose of them.
6. Remove solid waste bags, and dispose of them with wasted tips/vessels as well.
10. Place four PC/CC bottles fulfilled with distilled water for each on the positions where ProCells and
CleanCells are set.
11. Place two bottles fulfilled with distilled water into the position where the PreCleans are set.
16. Perform ten(10) “Sipper Air Purge” for each channel in “Maintenance” screen.
19. Carry out all maintenance items according to E170 Operator's Manual.
Shut down the main power of the entire Elecsys MODULAR System according to an usual
shut-down procedure.
To start up the system after a several-day shut-down, the following procedure is recommended.
1. Carry out a maintenance according to Operator’s Manual for C-MODULAR, such as washing of Water
Supply Tank and Drain/Supply water in Core.
2. Remove each bottle placed on the positions of ProCells, CleanCells and PreCleans beforehand.
3. Place each bottle for ProCell, CleanCell and PreClean to specified positions on the system.
5. Place new solid waste bags into the solid waste boxes.
7. After carrying out step 1. to 6. above in the Elecsys MODULAR system, the system switch on.
NOTE: If the system is in stand-by status, carry out the next procedures.
14. Check LLD voltage of sample pipetter and sippers according to E170 service manual afterward.
NOTE: If the LLD voltage of the sample pipetter is out of specified range, readjust it and perform 50(fifty)
“Mechanisms Check“ according to E170 service manual..
5) The following Window will open after all the files have been read (approximately 4 min), click “OK.”
6) Confirm the contents of the Log File. Check to see that “Transfer complete” is displayed after “Put
XXX” (refer to supplement for details regarding XXX). Then close this window.
7) Select “YES,” after confirming that all the transfers have been completed correctly.
8) When the installation is completed remove the installation CD and select “YES” to shut down the PC.
9) Wait until “It’s now safe to turn off your computer” is displayed.
10) After Pull out the CD, press the Orange button on the device. Wait 2 minutes, then turn off the power.
Turn the power back ON with the Green button and startup again.
11) After restarting, confirm that HD check is executed and the sum value is correct.
[Supplement]
Connected to 172.18.38.230.
ftp> ftp>
put CCK
200 Port set okay
150 Opening BINARY mode data connection
226 Transfer complete
779337 bytes sent in 2.38 seconds (326.90 Kbytes/sec)
ftp> ftp>
put CC
200 Port set okay
150 Opening BINARY mode data connection
226 Transfer complete
779301 bytes sent in 2.49 seconds (312.60 Kbytes/sec)
ftp> ftp>
put INSTALL
200 Port set okay
150 Opening BINARY mode data connection
226 Transfer complete
1268 bytes sent in 0.02 seconds (63.40 Kbytes/sec)
ftp> ftp>
put SUSCC1
200 Port set okay
150 Opening BINARY mode data connection
226 Transfer complete
209 bytes sent in 0.00 seconds (209000.00 Kbytes/sec)
ftp> ftp>
ftp>
put update.bat
ftp> ftp>
put CC.map
200 Port set okay
150 Opening BINARY mode data connection
226 Transfer complete
155792 bytes sent in 0.46 seconds (337.94 Kbytes/sec)
ftp> ftp>
ls
200 Port set okay
150 Opening ASCII mode data connection
ABS.DAT
ALM.DAT
ALM_BUF.DAT
CALIB.DAT
CMD.DAT
ETHERCMM.DAT
HOST_CMM.DAT
ISEQCDT1.DAT
ISEQCDT2.DAT
LON_CMM.DAT
MAINT.DAT
PA.DAT
PCPA.DAT
QCCOM.DAT
QCDT1.DAT
QCDT2.DAT
QCDT3.DAT
QCDT4.DAT
QCDT5.DAT
QCDT6.DAT
REAG.DAT
SMPLST.DAT
SPA.DAT
TM.DAT
TM_BUF.DAT
TSRS.DAT
ZPA.DAT
SMP.BTR
CATRACE1.BTR
CATRACE2.BTR
CATRACE3.BTR
CATRACE4.BTR
CATRACE5.BTR
CATRACE6.BTR
CARES1.BTR
CARES2.BTR
CARES3.BTR
CARES4.BTR
CARES5.BTR
CARES6.BTR
CATS.BTR
CAPRETS1.BTR
CAPRETS2.BTR
CAPRETS3.BTR
CAPRETS4.BTR
CABTLTS1.BTR
CABTLTS2.BTR
CABTLTS3.BTR
CABTLTS4.BTR
NONID.BTR
MIN_SBAR.BTR
RBAR_AU1.BTR
RBAR_AU2.BTR
RBAR_AU3.BTR
RBAR_AU4.BTR
RBAR_AU5.BTR
RBAR_AU6.BTR
RID_AU1.BTR
RID_AU2.BTR
RID_AU3.BTR
RID_AU4.BTR
RID_AU5.BTR
RID_AU6.BTR
RID_CORE.BTR
ABS_NE.BTR
ABS_C.BTR
SUSCCX
SUSCCK
CC
CCK
CCK.SYM
CCKK
CCKS
CCKD
CC.SYM
CC1
CC1.SYM
CC2
CC2.SYM
CCF
CCF.SYM
INSTALL
SYSVER
SUM.TXT
SUSCC1
SUSCC2
SUSCCF
UPDATE.BAT
UPDATE.TXT
CC.MAP
CC1.MAP
CC2.MAP
CCF.MAP
CCK.MAP
L
ftp> quit
221 Bye...see you later
1. C drive is assumed to have the original settings from (Acct.) for it to work properly.
(All the files must be in C:\sd0a; and ftp server must be operating and its home directory must be in
C:\)
2. The following points must be in place for all the computers for Core, ISE and Analyzers for the
ECPU500.
➀ Voltage adjustments and electrical wiring are done correctly.
➁ Dip SW for ECPU500 board is set correctly.
➂ Dip SW for HMCNT board is set correctly.
➃ Cable for Ethernet is connected correctly.
➄ Cable for LON is connected correctly.
➅ Boot parameter is the initial factory settings.
➆ With all console cables for Core, ISE, and Analyzer computers for the ECPU500 connected an
output from the Boot ROM of the ECPU500 has been confirmed.
[Procedure]
(1) With the power OFF, install HD’s jumper pin to the new HD set to SCSI ID=6(Refer to C4.5.3.17
section or Service Memo). Then remove the SCSI cable from the Core and connect it to the ISE.
(2) Turn the power ON. Wait until the Initial Operation screen on the PC comes up.
(3) Complete the MoSys.exe in the following manner:
➀ Open “Windows NT Security” with Ctrl+Alt+Delete.
➁ Click “Task Manager” to start.
➂ Select “Process,” select “MoSys.exe,” then click “Finish Process.”
➄ Click the “x” button on the upper right side of the screen and close.
(4) Change PC’s IP Address to “172, 18, 38, 230” in the following manner:
➀ Open the “Control Panel” by selecting (Start) - (Settings) - (Control Panel) on the taskbar.
If “Connection Failure” occurs, wait 30 seconds and press “OK” to reconnect. Repeat 2~3 times until
the connection is made.
➅ When the connection is made “[VRT500]->” will be displayed; type (diskInit “/sd0a”).
Make sure that the following response is returned.
[ “/sd0a” Initialized
value = 0 = 0x0 ]
➆ Click the “x” button on the upper right side of the screen and end the “telnet.”
➇ Click the “x” button on the upper right side of the screen and close the “Command Prompt.”
(6) Start the Batch File for the Core HD installation in the following manner.
➀ After Select (Start) - (Run) on taskbar, type “ e:\svc_tool\stupcore\hdinit.”, then push ‘OK’.
➁ When “Press Any Key” appears on the screen, press the “Spacebar”.
➂ Click the “x” button on the upper right side of the screen and close the Explorer.
(7) Restore PC’s IP Address to “172, 18, 38, 237” in the following manner:
➀ Open the “Control Panel” by selecting (Start) - (Settings) - (Control Panel) on the taskbar.
2.2.1 After the computer is restarted, the display has changed from the character-based screen to the
Windows-style screen. Click Exit at Working in Comfort Window, the PC will Reboot again.
2.2.2 The PC will start Hewlett-Packard Windows NT Setup window automaticlly after reboot.
2.2.3 When Software License Agreement Window appears, Click I Agree.
2.2.4 Click Next at Window Windows NT Setup.
2.2.5 When name and Organization appears:
3.3 Select the tag of Performance, Click the Change button of Virtual Memory.
3.4 For Page File Size for Selected Drive, Change Initial Size [MB] to 200 and change Maximum
Size[MB] to 250, then click the set button.
3.5 Click the OK button of the Window Virtual Memory, and Click the Close button to close the
window of Properties. When the below window appears, click the No button.
4.2 Click the button of Display Type , the Window below appears.
4.4 Click Have Disk button on the Window Change Display, the Install From Disk window appears.
Input d:\
\hpdrive\ \nt1.70j at Copy manufacturer’s files from, and Click the OK button.
\video\
4.5 When the below Change Display window appears, click the OK button.
4.6 When the below Third-party Drivers window appears, click the Yes button
4.7 When Setup asks you for Installing driver, click OK button at next window.
4.8 Close all other windows except System settings Change window. Then Click Yes button on below
Window. Computer will be restarted.
4.9 After the computer restart up, login the Windows NT. Click the right button of the mouse on the
desktop where no icons are. Select Properties from the menu. Display Settings Window
appears.
4.10 Follow the steps below to select the options shown in the figure below:
For Color Palette, Select 65536 Colors
For Desktop Area: Select 1024 by 768 pixels
There is no change for font size and Refresh Frequency.
Click OK button to finish the setting.
4.11 When Display Settings Window appears, Click OK button to finish the change. And confirm if the
display have changed as setting.
5.3.2 When Click then manufacture and model... appears, select Fujitsu for Manufactures and
Fujitsu DL 3300 for Printers. Then click Next for continue.
5.3.3 When Type in the name of printer. When... appears, click Next button for continue.
5.3.4 When Indicate whether this printer will be ...appears, click Next button for continue.
5.3.5 When After your printer is installed, you can... appears, click radio button NO, then Click
Finish button to continue.
5.3.6 When Insert Disk window appears, insert Windows NT 4.0 CD-ROM to e: drive and wait until
the Windows NT CD-ROM Setup window appears. Click × to close the window of Setup
and click OK button on Insert Disk window.
5.3.7
W
h
e
n
5.4 Click the right button of the mouse on Fujitsu DL 3300 icon and select Document Defaults... from
menu. Fujitsu DL 3300 Default Document Properties window appears.
5.5 Select Gray Scale for Color Appearance by Clicking radio button. And Click OK button to close the
window of Fujitsu DL 3300 Default Document Properties .
5.6 Click the right button of the mouse on Fujitsu DL 3300 icon and select Properties from menu.
Fujitsu DL 3300 Properties window appears.
5.7 Click the tag of Device Settings . Next double click Halftone Setup... at the middle of the Windows.
Device Color/Halftone Properties windows appears.
5.8 Drug the button to change Pixel Diameter to 194.3%, then click OK button. It will return to Fujitsu
DL 3300 Properties window. Click OK button to continue.
6.1.2 Click Change bottom at the tag of Identification. 「Identification Changes] Window
appears.
6.1.3 For Workgroup, Chage it from WORKGROUP to MOSYS, and Click OK bottom.
Next Click OK bottom at Network Configuration Window. It trun back to the window of
Network.
6.2.2 Select NetBEUI at Network Protocols list and Click Remove... bottom.
When the warning message appears as bellow, Click Yes bottom. NetBEUI will be Deleted
from the list.
6.2.3 Do the same as 6.2.2, NWLink IPX/SPX Compertible Transport and NWLink
NetBIOS will be deleted from the list at the same time.
6.3.2 Select TCP/IP Protocols from the list and Click OK bottom.
6.3.3 TCP/IP Setup Window appears as follow, then Click NO bottom to continue.
6.3.4 Input C: \i386 at Windows NT Setup window and click Continue bottom.
6.3.5 After all files were copied, Click close bottomto back to Network window.
6.3.6 Input 172.18.38.237 for IP Address, 255.255.255.0 for Subnet Mask, 172.18.38.1 for
Default Gateway at Microsoft TCP/IP Protocol window.
6.3.7 Click Yes bottom at Network Setting Change window and PC will reboot.
7.2 Click the right button of the mouse on the Network Neighborhood icon of the desktop and select
Properties from the menu. Network window appears.
7.3 Click Add button on Services tag at window Network, Select Network Service window appears,
Select Microsoft Peer Web Server of Network Service and click OK button.
7.4 When Internet Information Server--Files Needed window appears, insert Windows NT CD-ROM
to e: drive, input e:\
\i386 for installed from:, then click OK button.
7.5.3 When Microsoft Peer Web Services Setup window appears, click Yes button.
7.6.2 When FTP Service Properties for cu window appears, Change 1000 to 10 for Maximum
Connections at Service tag, remove the check of Allow Anonymous Connecting.
7.6.3 When below Internet Service Manager window appears, click Yes button.
7.6.4 Double click C: at tag Directories of window Microsoft Internet Service Manager ,input C:
\ for Directory: at Directory Properties window , and click OK button.
7.6.5 Click Logging tag, and check Monthly for Automatically open new log, then click OK
button.
9. Setup touchpanel
9.1 Make a back up of Hardware Profiles
9.1.1 Click the right button of the mouse on the My Computer icon of the desktop and select
Properties from the menu. System Properties window appears.
9.1.2 Click the tag of Hardware Profiles and select Original Configuration of Available
Hardware Profiles, then click Copy... button, Copy Profile Window appears.
9.1.3 Input No Touch for To: and click OK button.
9.1.4 Change Wait for user selection for ... of Multiple Hardware Profiles from 30 to 1.
9.2.4 When Choose Destination Location window appears, click Next button
9.2.5 When MonitorMouse for Windows NT Instration Option window appears, click Next
button.
9.2.8 When Touch Screen Setup window appears, Check PC-Bus for Controller Type and
change AccuTouch2201 to IntelliTouch40×
×5. Then Click OK button.
9.3.4 When red circle mark appears at upleft the window, touch it by finger, and the mark will
move to downright, touch it again, next the mark move to upright , touch it by finger.
10.2 Press PageDown key to find Install Service Pack(in blue Character), and double click on it.
10.3 When Confirm File Open window appears, Click Open button to continue.
10.9 When below window appears, remove Windows NT CD-ROM from e: drive and Click OK button to
restart Computer.
11.5 Only ck ODBC Core and Access ODBC Driver for Options, click Install button.
11.6 Files copy to harddisk. When below window appears, click OK button.
11.7 Remove CD-ROM from e: drive , Shutdown the computer and restart it.
12. Install MODULAR System software
12.1 Login Windows NT after computer restart.
12.2 Insert MODULAR system Install CD-ROM to e: drive.
12.3 CD-ROM will run automatically and MODULAR System window appears. Click Setup button.
12.4 Files copy to harddisk.(It will take about 4 minutes) When Information window appears, Click OK
button.
12.10 When User Manager window appears, select Hitachi for User Name and double click it.
12.11 Only check User cannot change password and Password never expires, and click OK button.
12.14 When Shut Down window appears, remove CD-ROM from e: drive and press Enter to Restart
computer.
13.1.3 When Task Manager Warning window appears, click Yes button. And click × to close
Windows NT Task Manager window.
13.2.3 Select Start Menu Programs tag and Click Remove button. Remove Shortcuts/Folders
window appears.
13.2.4 Select Accessories on Remove Shortcuts/Folders window and click Remove button to
delete. When Confirm Folder Delete window appears, Click Yes button.
13.2.5 Delete the below two programs just like what was done at 13.2.4.
Command Prompt, Windows NT Explorer
13.2.6 Click Close button at window Remove Shortcuts/Folders to close the window.
13.2.7 Click OK button to close Taskbar Properties window.
13.3.3 Delete below two icons just like what was done at 13.3.2.
Inbox
My Briefcase
13.3.4 Click the right button of the mouse on the desktop where is no icon and select [Arrange
Icon]-[Auto Arrange] from menu. Icon will be auto arranged.
13.5.2 Click Screen Saver tag, and Select Marquee Display for Screen Saver. Change Wait: to
10 minutes.
13.5.3 Click Settings button at Display Properties window. Marquee Setup window appears.
13.5.4 Setup as follow at Marquee Setup window
Position: Random
Background Color: (Black)
Text: Please touch MOUSE to leave Screen Saver.
13.5.5 Click Format Text... button at Marquee Setup window, Format Text window appears.
14.2 When Repair Disk Utility window appears, insert a new Floppy Disk to a: drive and click Create
Repair Disk button.
14.3 When below Repair Disk Utility window appears, Click OK button.
14.4 When below Repair Disk Utility window appears, Click OK button.
15.4 Delete
program will
run. When
DELCMD
window
appears, click
OK button to
confirm finish.
Since the E170 module does not contain a hard disk component, it is required to save adjustment data and
system parameter data onto a floppy disk.
1. File Structure
Prepare a floppy disk for adjustment of each module, and indicate a serial number of the module on a label
of the floppy disk. Then, write the following data to the root directory of the floppy disk (read the data from it).
(1) Writing
(2) Reading
Based on: FD file specifications V3.0 for hybrid modular mechanism adjustment
Section Page
3.1 ALARM LIST .......................................................................................S3-1
3.1.1 D Module S3-1
3.1.2 P-Module ..........................................................................................................S3-189
3.1.3 E-Module ..........................................................................................................S3-296
3.1.4 Core-Module ....................................................................................................S3-370
3.1.5 ISE Module .......................................................................................................S3-439
3.1.6 PC .....................................................................................................................S3-456
2 1 2 Abnormal Stirrer 1 Movement Stirrer 1 does not reach its highest position when ascending at the reaction cell. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the UDP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the U/D motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
4 1 4 Abnormal Stirrer 1 Movement Stirrer 1 does not reach the rinse bath position when moving toward the rinse bath. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the HP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the Rotation motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
5 1 5 Abnormal Stirrer 1 Movement Stirrer 1 does not move to the reaction disk. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the Cell Position Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the Rotation motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
6 1 6 Abnormal Stirrer 1 Movement When reset, stirrer 1 does not return to the rinse bath (home) position. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the HP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the Rotation motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
7 1 7 Abnormal Stirrer 1 Movement When reset, stirrer 1 does not leave the rinse bath (home) position. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the HP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the Rotation motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
8 1 8 Abnormal Stirrer 1 Movement Stirrer 1 does not move in its highest position. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the UDP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the U/D motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
9 1 9 Abnormal Stirrer 1 Movement Stirrer 1 does not leave its highest position when descending at the rinse bath. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the UDP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the U/D motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
10 1 10 Abnormal Stirrer 1 Movement Stirrer 1 does not leave its highest position when descending at the reaction cell. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the UDP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the U/D motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
11 2 1 Abnormal Stirrer 2 Movement Stirrer 2 does not reach its highest position when ascending at the rinse bath or the first time a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the UDP Sensor 1. Check the Sensor
after reset. b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the U/D motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
12 2 2 Abnormal Stirrer 2 Movement Stirrer 2 does not reach its highest position when ascending at the reaction cell. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the UDP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the U/D motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
14 2 4 Abnormal Stirrer 2 Movement Stirrer 2 does not reach the rinse bath position when moving toward the rinse bath. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the HP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the Rotation motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
15 2 5 Abnormal Stirrer 2 Movement Stirrer 2 does not move to the reaction disk. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the Cell Position Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the Rotation motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
16 2 6 Abnormal Stirrer 2 Movement When reset, Stirrer 2 does not return to the rinse bath (home) position. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the HP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the Rotation motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
17 2 7 Abnormal Stirrer 2 Movement When reset, Stirrer 2 does not leave the rinse bath (home) position. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the HP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the Rotation motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
18 2 8 Abnormal Stirrer 2 Movement Stirrer 2 does not move in its highest position. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the UDP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the U/D motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
19 2 9 Abnormal Stirrer 2 Movement Stirrer 2 does not leave its highest position when descending at the rinse bath. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the UDP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the U/D motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
20 2 10 Abnormal Stirrer 2 Movement Stirrer 2 does not leave its highest position when descending at the reaction cell. a. Remove any obstructions from the stirrer 1 mechanism pathway. 1. Defective the UDP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the U/D motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
21 3 1 Abnormal Rinse Unit Movement The Rinse Unit does not detect its highest position when ascending. a. Check the rinse unit and the reaction disk. 1. Defective the UDP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the U/D motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 4. Check the DipSW setting and replace the PCBs.
S3 - 1
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
22 3 2 Abnormal Rinse Unit Movement The rinse mechanism does not descend into the reaction cell from its highest position. a. Check the rinse unit and the reaction disk. 1. Defective the UDP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the U/D motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 4. Check the DipSW setting and replace the PCBs.
23 3 3 Abnormal Rinse Unit Movement The rinse mechanism does not stop at its highest position. a. Check the rinse unit and the reaction disk. 1. Defective the UDP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the U/D motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. Resume operation; if alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 4. Check the DipSW setting and replace the PCBs.
41 5 1 Reaction Disk Positioning Failure Reaction disk does not detect its stop position. a. If reaction disk has recently been washed, check for water droplets on bottom 1. Defective the UDP Sensor 1. Check the Sensor
of disk. Dry thoroughly. 2. Defective the motor 2. Replace the motor
b. Check for water droplets on the detectors below the reaction disk. Dry 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
thoroughly. 4. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 4. Check the DipSW setting and replace the PCBs.
c. Resume operation; if alarm recurs, call Technical Support.
42 5 2 Reaction Disk Positioning Failure Reaction disk does not stop at the correct position. a. If reaction disk has recently been washed, check for water droplets on bottom 1. Defective the UDP Sensor 1. Check the Sensor
of disk. Dry thoroughly. 2. Defective the motor 2. Replace the motor
b. Check for water droplets on the detectors below the reaction disk. Dry 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
thoroughly. 4. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 4. Check the DipSW setting and replace the PCBs.
c. Resume operation; if alarm recurs, call Technical Support.
43 5 3 Reaction Disk Positioning Failure Reaction disk does not detect its home position when reset. a. If reaction disk has recently been washed, check for water droplets on bottom 1. Defective the HP Sensor 1. Check the Sensor
of disk. Dry thoroughly. 2. Defective the motor 2. Replace the motor
b. Check for water droplets on the detectors below the reaction disk. Dry 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
thoroughly. 4. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 4. Check the DipSW setting and replace the PCBs.
c. Resume operation; if alarm recurs, call Technical Support.
44 5 4 Reaction Disk Positioning Failure First reaction cell does not stop at the correct position when reset. a. If reaction disk has recently been washed, check for water droplets on bottom 1. Defective the HP Sensor 1. Check the Sensor
of disk. Dry thoroughly. 2. Defective the motor 2. Replace the motor
b. Check for water droplets on the detectors below the reaction disk. Dry 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
thoroughly. 4. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 4. Check the DipSW setting and replace the PCBs.
c. Resume operation; if alarm recurs, call Technical Support.
45 5 5 Reaction Disk Positioning Failure When reaction disk rotates, the sample probe, reagent probe, stirrer, or cell rinse a. If other alarms are present, correct them first. 1. Defective the UDP Sensor on the Serum Sampling 1. Check the Sensor
mechanism is not set at the highest point, at the reaction cell. b. If reaction disk has recently been washed, check for water droplets on bottom Mecha, or the Stirrer Mecha. or the Rinse mecha. 2. Replace the motor
of disk. Dry thoroughly. 2. Defective the motor 3. Check whether signal wire breaks or not.
c. Check for water droplets on the detectors below the reaction disk. Dry 3. Signal Wire breaks 4. Check the DipSW setting and replace the PCBs.
thoroughly. 4. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB.
d. Resume operation; if alarm recurs, call Technical Support.
51 6 1 Abnormal Sample Probe Movement Sample probe does not ascend to the highest point when leaving any position other than the a. Remove any obstructions from the sample probe’s pathway. 1. Defective the UDP Sensor 1. Check the Sensor
reaction cell or the first time after reset. b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. If alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
52 6 2 Abnormal Sample Probe Movement Sample probe does not ascend to the highest point when leaving the reaction cell. a. Remove any obstructions from the sample probe’s pathway. 1. Defective the UDP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. If alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
53 6 3 Abnormal Sample Probe Movement Sample probe descends abnormally at any position other than at reaction cell. a. Remove any obstructions from the sample probe’s pathway. 1. Defective the probe position over the sample cup or the 1. Check the position of the probe, the sample cup or the rinse bath.
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the rinse bath. 2. Check the Sensor
check 10 times. Observe operation. 2. Defective the Abn. Des. Sensor. 3. Check the mirror.
c. If alarm recurs, call Technical Support. 3. Defective the mirror. 4. Check whether signal wire breaks or not.
4. Signal wire breaks. 5. Check the DipSW setting and replace the PCBs.
5. Defective the the DS2 PCB or HMC2 PCB.
54 6 4 Abnormal Sample Probe Movement The sample probe descends abnormally into the reaction cell. a. Remove any obstructions from the sample probe’s pathway. 1. Defective the probe position over the cell. 1. Check the position of position.
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the Abn. Des. Sensor. 2. Check the Sensor
check 10 times. Observe operation. 3. Defective the mirror. 3. Check the mirror.
c. If alarm recurs, call Technical Support. 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the the DS2 PCB or HMC2 PCB. 5. Check the DipSW setting and replace the PCBs.
55 6 5 Abnormal Sample Probe Movement Sample probe does not descend at any position other than at reaction cell. a. Remove any obstructions from the sample probe's pathway. 1. Defective the UDP Sensor 1. Check the Sensor
b. Inspect probe where it is attached to the sample probe arm. The probe is 2. Defective the motor 2. Replace the motor
spring-mounted and should travel freely up and spring down to its original position 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
when released. Check sample probe for distortion. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify probe wire connection.
d. Check sample cup for distortion.
e. Touch Utility/Maintenance/Maintenance. Execute S.Probe Adjust (Horiz). Place
adjustment tools on rack and Execute S. Probe Adjust (Vert).
f. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the
check 10 times. Observe operation.
g. If alarm recurs, call Technical Support.
56 6 6 Abnormal Sample Probe Movement Sample probe does not descend into the reaction cell. a. Remove any obstructions from the sample probe's pathway. 1. Defective the UDP Sensor 1. Check the Sensor
b. Inspect probe where it is attached to the sample probe arm. The probe is 2. Defective the motor 2. Replace the motor
spring-mounted and should travel freely up and spring down to its original position 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
when released. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify probe wire connection.
d. Touch Utility/Maintenance/Maintenance. Execute S.Probe Adjust (Horiz).
During the horizontal probe adjust, when the sample probe is entering the dilution
vessel, ensure that the probe does not come in contact with the ISE nozzles.
e. Resume operation; if alarm recurs, replace the sample probe following the
instructions in the Maintenance Guide.
f. If alarm recurs, call Technical Support.
57 6 7 Abnormal Sample Probe Movement Sample probe 1 and 2 continues to descend abnormally. Abnormal descent detector a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor 1. Check the Sensor
remains ON. check 10 times. Observe operation. 2. Defective the motor 2. Replace the motor
b. Inspect probe where it is attached to the sample probe arm. The probe is 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
spring-mounted and should travel freely up and spring down to its original position 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
when released.
c. Verify probe wire connection.
d. Inspect cup for proper height.
e. Check sample cup for distortion.
f. Place the adjustment tool on a rack. Touch Utility/Maintenance/Maintenance.
Execute S.Probe Adjust (Vert).
g. Resume operation; if alarm recurs, call Technical Support.
S3 - 2
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
58 6 8 Abnormal Sample Probe Movement Sample probe cannot detect the correct cell position as it rotates to the reaction disk. a. Clean the tip of the sample probe. 1. Defective the probe position over the sample cup or the 1. Check the position of the probe, the sample cup or the rinse bath.
b. Verify sample probe wire connections. rinse bath. 2. Check the Sensor
c. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the Abn. Des. Sensor. 3. Check the mirror.
check 10 times. Observe operation. 3. Defective the mirror. 4. Check whether signal wire breaks or not.
d. Resume operation; if alarm recurs, replace the sample probe following the 4. Signal wire breaks. 5. Check the DipSW setting and replace the PCBs.
instructions in the Maintenance Guide. 5. Defective the the DS2 PCB or HMC2 PCB.
e. If alarm recurs, call Technical Support.
59 6 9 Abnormal Sample Probe Movement Sample probe does not leave reaction cell position when trying to rotate toward another a. Remove any obstructions from the sample probe’s pathway. 1. Defective the HP Sensor 1. Check the Sensor
position. b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. If alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
60 6 10 Abnormal Sample Probe Movement Sample probe 3 and 4 continues to descend abnormally. Abnormal descent detector a. Remove any obstructions from the sample probe’s pathway. 1. Defective the UDP Sensor 1. Check the Sensor
remains ON. b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. If alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
61 6 11 Abnormal Sample Probe Movement Liquid level detector is not reset when probe descending. a. Clean the tip of the sample probe. 1. Touch the probes. 1. Align the probe position.
b. Check that the DI water quality resistance is > 1.5 megaohms. 2. Defective the DS2 PCB or the HMC2. 2. Check the DipSW setting and replace the PCBs.
c. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the
check 10 times. Observe operation.
d. Resume operation; if alarm recurs, call Technical Support.
62 6 12 Abnormal Sample Probe Movement Sample probe continues to descend abnormally or Probe adjustment resulted in failure a. Set adjustment tool on rack correctly and place rack in the loader. 1. Defective way of vertical probe adjustment. 1. Put the jig on the rack directly.(No tube)
during vertical probe adjustment. b. Touch Utility/Maintenance/Maintenance/Probe Adjust/Select. Execute a vertical 2. Defective the LLD. 2. Check whether signal wire break or not.
sample probe adjust. 3. Defective the DS2 PCB or HMC2 PCB. 3. Check the DipSW setting and replace the PCBs.
c. Resume operation; if alarm recurs, call Technical Support.
63 6 13 Abnormal Sample Probe Movement Sample Probe does not rotate in its highest position. a. Remove any obstructions from the sample probe’s pathway. 1. Defective the UDP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. If alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
64 6 14 Abnormal Sample Probe Movement Sample probe does not move when it tries to rotate from its reset (home) position. a. Remove any obstructions from the sample probe’s pathway. 1. Defective the HP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. If alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
65 6 15 Abnormal Sample Probe Movement Sample probe does not detect its reset (home) position. a. Remove any obstructions from the sample probe’s pathway. 1. Defective the HP Sensor 1. Check the Sensor
b. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 2. Defective the motor 2. Replace the motor
check 10 times. Observe operation. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
c. If alarm recurs, call Technical Support. 4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
101 10 1 Sample Short There is insufficient sample volume in the specified sample container on rack. a. Add sample to the sample container. 1. Defective the sample volume 2. Defective the cup and 1. Check the sample volume.
b. If there is sufficient sample in the sample container, clean the tip of the sample tube size 2. Replace the cup or tube recommended by Hitachi.
probe and verify that the probe wire is properly connected. 3. Defective the LLD 3. Check whether signal wire breaks or not.
c. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 4. Defective the U/D motor 4. Replace the motor.
check 10 times. Observe operation. 5. Defective the DS2 PCB or the HMC2 PCB. 5. Check the DipSW setting and replace the PCBs.
d. Resume operation; if alarm recurs, call Technical Support.
131 13 1 Abnormal Sample Syringe 1 Movement Sample syringe 1 does not ascend to the highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor. 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor. 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal wire breaks. 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
132 13 2 Abnormal Sample Syringe 1 Movement Sample syringe 1 will not descend from the highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor. 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor. 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal wire breaks. 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
141 14 1 Abnormal Sample Syringe 2 Movement Sample syringe 2 does not ascend to the highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor. 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor. 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal wire breaks. 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
142 14 2 Abnormal Sample Syringe 2 Movement Sample syringe 2 will not descend from the highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor. 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor. 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal wire breaks. 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
151 15 1 Abnormal Sample Syringe 3 Movement Sample syringe 3 does not ascend to the highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor. 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor. 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal wire breaks. 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
S3 - 3
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
152 15 2 Abnormal Sample Syringe 3 Movement Sample syringe 3 will not descend from the highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor. 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor. 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal wire breaks. 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
161 16 1 Abnormal Sample Syringe 4 Movement Sample syringe 4 does not ascend to the highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor. 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor. 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal wire breaks. 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
162 16 2 Abnormal Sample Syringe 4 Movement Sample syringe 4 will not descend from the highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor. 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor. 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal wire breaks. 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
171 17 1 Abnormal Detergent Syringe Movement Detergent Syringe does not ascend to the highest position. . Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the check 1. Defective the UDP Sensor. 1. Check the Sensor
10 times. Observe operation. 2. Defective the motor. 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal wire breaks. 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
172 17 2 Abnormal Detergent Syringe Movement Detergent Syringe will not descend from the highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor. 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor. 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal wire breaks. 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
181 18 1 Multichange Valve 1 Multichange Valve 1 does not detect its home stop position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
check 10 times. Observe operation. 2. Defective the HP Sensor. 2.Check the Sensor.
b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
182 18 2 Multichange Valve 1 Multichange Valve 1 does not detect its dispensing stop position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
183 18 3 Multichange Valve 1 Multichange Valve 1 does not stop at correct position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
184 18 4 Multichange Valve 1 During reset, Multichange Valve 1 does not leave its home position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
check 10 times. Observe operation. 2. Defective the HP Sensor. 2.Check the Sensor.
b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
185 18 5 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger =Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = -- b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
186 18 6 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse > 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
187 18 7 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse > 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
188 18 8 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
S3 - 4
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
980 18 100 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 1 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1001 18 101 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 1 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1002 18 150 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 1 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1003 18 151 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 2 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1004 18 200 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 2 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1005 18 201 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 2 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1006 18 202 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 2 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1007 18 250 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 2 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1008 18 251 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 2 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1009 18 252 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1010 18 300 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1011 18 301 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1012 18 302 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1013 18 303 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not ensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1014 18 350 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1015 18 351 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
S3 - 5
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1016 18 352 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1017 18 353 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 4 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1018 18 400 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 4 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
S3 - 6
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1032 18 504 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1033 18 505 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1034 18 550 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1035 18 551 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1036 18 552 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1037 18 553 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1038 18 554 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1039 18 555 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1040 18 600 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1041 18 601 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1042 18 602 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1043 18 603 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1044 18 604 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1045 18 605 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1046 18 606 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1047 18 650 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
S3 - 7
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1048 18 651 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1049 18 652 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1050 18 653 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1051 18 654 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1052 18 655 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1053 18 656 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1054 18 700 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1055 18 701 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1056 18 702 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1057 18 703 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1058 18 704 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1059 18 705 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1060 18 706 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1061 18 707 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 7
1062 18 750 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1063 18 751 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
S3 - 8
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1064 18 752 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1065 18 753 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1066 18 754 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1067 18 755 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1068 18 756 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1069 18 757 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 7
1070 18 800 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1071 18 801 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1072 18 802 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1073 18 803 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1074 18 804 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1075 18 805 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1076 18 806 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1077 18 807 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 7
1078 18 808 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 8
1079 18 850 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
S3 - 9
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1080 18 851 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1081 18 852 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1082 18 853 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1083 18 854 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1084 18 855 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1085 18 856 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1086 18 857 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 7
1087 18 858 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 8
1088 18 900 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1089 18 901 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1090 18 902 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1091 18 903 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1092 18 904 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1093 18 905 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1094 18 906 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1095 18 907 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 7
S3 - 10
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1096 18 908 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 8
1097 18 909 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 9
1098 18 950 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1099 18 951 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1100 18 952 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1101 18 953 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1102 18 954 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1103 18 955 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1104 18 956 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1105 18 957 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 7
1106 18 958 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 8
1107 18 959 Multichange Valve 1 *Asia Only Multichange Valve 1 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 1 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 9
19 1 Multichange Valve 2 Multichange Valve 2 does not detect its home stop position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
check 10 times. Observe operation. 2. Defective the HP Sensor. 2.Check the Sensor.
b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
191 19 2 Multichange Valve 2 Multichange Valve 2 does not detect its dispense stop position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
192 19 3 Multichange Valve 2 Multichange Valve 2 does not stop at the correct position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
193 19 4 Multichange Valve 2 During reset, Multichange Valve 2 does not leave its home position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
check 10 times. Observe operation. 2. Defective the HP Sensor. 2.Check the Sensor.
b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
S3 - 11
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
194 19 5 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = -- b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
195 19 6 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse > 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
196 19 7 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse > 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
197 19 8 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
1108 19 100 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 1 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1201 19 101 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 1 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1202 19 150 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 1 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1203 19 151 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 2 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1204 19 200 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 2 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1205 19 201 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 2 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1206 19 202 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 2 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1207 19 250 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 2 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1208 19 251 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 2 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1209 19 252 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1210 19 300 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1211 19 301 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
S3 - 12
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1212 19 302 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1213 19 303 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1214 19 350 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1215 19 351 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1216 19 352 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 3 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1217 19 353 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 4 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1218 19 400 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 4 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1219 19 401 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 4 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1220 19 402 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 4 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1221 19 403 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 4 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1222 19 404 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 4 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1223 19 450 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 4 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1224 19 451 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 4 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1225 19 452 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 4 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1226 19 453 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 4 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1227 19 454 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
S3 - 13
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1228 19 500 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1229 19 501 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1230 19 502 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1231 19 503 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1232 19 504 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1233 19 505 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1234 19 550 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1235 19 551 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1236 19 552 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1237 19 553 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1238 19 554 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 5 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1239 19 555 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1240 19 600 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1241 19 601 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1242 19 602 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1243 19 603 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
S3 - 14
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1244 19 604 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1245 19 605 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1246 19 606 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1247 19 650 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1248 19 651 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1249 19 652 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1250 19 653 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1251 19 654 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1252 19 655 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 6 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1253 19 656 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1254 19 700 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1255 19 701 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1256 19 702 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1257 19 703 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1258 19 704 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1259 19 705 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
S3 - 15
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1260 19 706 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1261 19 707 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 7
1262 19 750 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1263 19 751 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1264 19 752 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1265 19 753 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1266 19 754 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1267 19 755 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1268 19 756 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 7 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1269 19 757 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 7
1270 19 800 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1271 19 801 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1272 19 802 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1273 19 803 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1274 19 804 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1275 19 805 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
S3 - 16
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1276 19 806 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1277 19 807 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 7
1278 19 808 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 8
1279 19 850 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1280 19 851 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1281 19 852 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1282 19 853 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1283 19 854 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1284 19 855 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1285 19 856 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1286 19 857 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 8 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 7
1287 19 858 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 8
1288 19 900 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1289 19 901 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1290 19 902 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1291 19 903 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
S3 - 17
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1292 19 904 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1293 19 905 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1294 19 906 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1295 19 907 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 7
1296 19 908 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 8
1297 19 909 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = ON b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 9
1298 19 950 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 0
1299 19 951 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 1
1300 19 952 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 2
1301 19 953 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 3
1302 19 954 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 4
1303 19 955 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 5
1304 19 956 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 6
1305 19 957 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 7
1306 19 958 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 8
1307 19 959 Multichange Valve 2 *Asia Only Multichange Valve 2 does not stop correctly. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the MSV. 1. Clean up or replace the MSV.
Stop Trigger = Not Sensor check 10 times. Observe operation. 2. Defective the Count Sensor. 2.Check the Sensor.
Sensor Status = OFF b. Resume operation; if alarm recurs, call Technical Support. 3. Defective the motor. 3. Replace the motor
Remaining Pulse = 0 4. Signal wire breaks. 4. Check whether signal wire breaks or not.
Start Limit Counter = 9 5. Defective the DS2 PCB, the HMC1 PCB or DR1-D PCB. 5. Check the DipSW setting and replace the PCBs.
Remaining Limit Counter = 9
S3 - 18
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
20 1 Abnormal Reagent 1 Nozzle Movement Reagent 1 Nozzle Movement does not reach reaction cell position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the Cell Position Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the F/B motor 2. Replace the motor
b. Resume operation; if alarm recurs, call Technical Support. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
201 20 2 Abnormal Reagent 1 Nozzle Movement Reagent 1 Nozzle Movement does not reach at rinse bath position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the HP Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the F/B motor 2. Replace the motor
b. Resume operation; if alarm recurs, call Technical Support. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
202 20 3 Abnormal Reagent 1 Nozzle Movement During reset, Reagent 2 Nozzle Movement does not leave its home position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the HP Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the F/B motor 2. Replace the motor
b. Resume operation; if alarm recurs, call Technical Support. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
203 21 1 Abnormal Reagent 2 Nozzle Movement Reagent 2 Nozzle Movement does not reach at reaction cell position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the Cell Position Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the F/B motor 2. Replace the motor
b. Resume operation; if alarm recurs, call Technical Support. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
211 21 2 Abnormal Reagent 2 Nozzle Movement Reagent 2 Nozzle Movement does not reach at rinse bath position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the HP Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the F/B motor 2. Replace the motor
b. Resume operation; if alarm recurs, call Technical Support. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
212 21 3 Abnormal Reagent 2 Nozzle Movement During reset, Reagent 2 Nozzle Movement does not leave its home position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the HP Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the F/B motor 2. Replace the motor
b. Resume operation; if alarm recurs, call Technical Support. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
4. Defective the DS2 PCB, the HMC2 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
213 22 1 Abnormal Reagent 1 Syringe 1 Movement Reagent syringe R1-1 does not ascend to its highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
221 22 2 Abnormal Reagent 1 Syringe 1 Movement Reagent syringe R1-1 will not descend from its highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
222 23 1 Abnormal Reagent 1 Syringe 2 Movement Reagent syringe R1-2 does not ascend to its highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
231 23 2 Abnormal Reagent 1 Syringe 2 Movement Reagent syringe R1-2 will not descend from its highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
232 24 1 Abnormal Reagent 2 Syringe 1 Movement Reagent syringe R2-1 does not ascend to its highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
241 24 2 Abnormal Reagent 2 Syringe 1 Movement Reagent syringe R2-1 will not descend from its highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
242 25 1 Abnormal Reagent 2 Syringe 2 Movement Reagent syringe R2-2 does not ascend to its highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
S3 - 19
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
251 25 2 Abnormal Reagent 2 Syringe 2 Movement Reagent syringe R2-2 will not descend from its highest position. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the UDP Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor 2. Replace the motor
b. If the sample pipettor was recently serviced, verify that retaining screw is not 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
over tightened and that the seals are installed correctly. 4. Defective the DS2 PCB, the HMC3 PCB or DR3 PCB. 4. Check the DipSW setting and replace the PCBs.
c. Verify that the sample pipettor displacement rod is not bent.
d. Touch Utility/Maintenance/Maintenance/Air Purge/Select. Observe the sample
syringe for descending movement.
e. Resume operation; If alarm recurs, call Technical Support.
252 26 1 Reagent Changeover Valve Reagent Changeover Valve is not set on reagent side. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the Reagent side Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor 2. Replace the motor
b. Resume operation; if alarm recurs, call Technical Support. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
4. Defective the RY1 PCB, DS2 PCB, the HMC3 PCB or 4. Check the DipSW setting and replace the PCBs.
DR3 PCB.
261 26 2 Reagent Changeover Valve Reagent Changeover Valve is not set on dispensing nozzle side. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the dispensing nozzle side Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor 2. Replace the motor
b. Resume operation; if alarm recurs, call Technical Support. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
4. Defective the RY1 PCB, DS2 PCB, the HMC3 PCB or 4. Check the DipSW setting and replace the PCBs.
DR3 PCB.
262 26 3 Reagent Changeover Valve During reset, Reagent Changeover Valve is not set on reagent side. a. Touch Utility/Maintenance/Check/Mechanisms Check/Select. Execute the 1. Defective the Reagent side Sensor 1. Check the Sensor
check 10 times. Observe operation. 2. Defective the motor 2. Replace the motor
b. Resume operation; if alarm recurs, call Technical Support. 3. Signal Wire breaks 3. Check whether signal wire breaks or not.
4. Defective the RY1 PCB, DS2 PCB, the HMC3 PCB or 4. Check the DipSW setting and replace the PCBs.
DR3 PCB.
273 28 1 Inc. Water Temperature Error Water temperature in the incubation bath exceeds 40 degrees C. a. Ensure that the refrigerator condenser filter on the front of instrument is not 1. Defective the thermistor. 1. Replace the thermistor.
clogged with dust or debris. 2. Defective the heater. 2. Replace the heater.
b. Ensure that the instrument cooling fans on the back of the instrument are not 3. Defective the TEMP-M PCB. 3. Check the DipSW setting and replace the PCBs.
clogged with dust or debris. 4. Defective the magnet pump. 4. Check the magnet pump.
c. Ensure that the room temperature is 15 to 32 degrees C. 5. Defective the SV on the cooling unit. 5. Check the SV.
d. Verify that the incubation bath water is circulating and that water level is 6. Defective the cooling unit 6. Check the cooling unit.
normal. 7. Defective the tubing 7. Check whether the tubes are not crushed.
e. Touch Utility/Maintenance/Maintenance/Inc. Water Exchange/Select. Execute
the incubation bath exchange.
f. Monitor the System Overview screen to make sure the temperature returns to
37 degrees C.
g. Resume operation; if alarm recurs, call Technical Support.
281 28 2 Inc. Water Temperature Error Water temperature in the incubation bath is outside a range of 36.5 to 37.5 degrees C. a. Ensure that the refrigerator condenser filter on the front of instrument is not 1. Defective the thermistor. 1. Replace the thermistor.
clogged with dust or debris. 2. Defective the heater. 2. Replace the heater.
b. Ensure that the instrument cooling fans on the back of the instrument are not 3. Defective the TEMP-M PCB. 3. Check the DipSW setting and replace the PCBs.
clogged with dust or debris. 4. Defective the magnet pump. 4. Check the magnet pump.
c. Ensure that the room temperature is 15 to 32 degrees C. 5. Defective the SV on the cooling unit. 5. Check the SV.
d. Verify that the incubation bath water is circulating and that water level is 6. Defective the cooling unit 6. Check the cooling unit.
normal. 7. Defective the tubing 7. Check whether the tubes are not crushed.
e. Touch Utility/Maintenance/Maintenance/Inc. Water Exchange/Select. Execute
the incubation bath exchange.
f. Monitor the System Overview screen to make sure the temperature returns to
37 degrees C.
g. Resume operation; if alarm recurs, call Technical Support.
282 29 1 Inc. Water Level Too Low Incubation bath water level is too low. a. Ensure that the deionized water supply is on. 1. Defective the level balance of D module. 1. Check the level balance
b. Check that an adequate supply of Hitergent is located behind the left front 2. Defective the level sensor. 2. Check whether the sensor is not bent or get rusty and check sensor's
door. 3. Defective the manual cock. connector.
c. Touch Utility/Maintenance/Maintenance/Inc. Water Exchange/Select. Execute 4. Defective the SV32. 3. Check the manual cock
the incubation bath exchange. 4. Replace the SV32.
d. Resume operation; if alarm recurs, call Technical Support.
291 30 1 Inc. Water Exchange Interval Exceeded More than 24 hours have elapsed since the last incubation bath water exchange. a. Touch Utility/Maintenance/Maintenance/Inc. Water Exchange/Select. Execute 1. No execute the incubation bath water exchange for more 1. Execute the incubation bath water exchange .
the incubation bath exchange. than 24 hours.
301 35 1 Power Supply Unit > 70 degrees C The power supply unit's temperature is 70 degrees C or higher. a. Ensure that the instrument cooling fans on back of instrument are clean and 1. Defective the cooling fan on the PCB's rack. 1. Check whether the fan is moved.
operational. 2. Defective the thermo-sensor. 2. Replace the sensor.
b. Ensure that the refrigerator condenser filter on front of instrument is not
clogged with dust or debris.
c. Ensure that room temperature is 15 to 32 degrees C.
d. Resume operation; if alarm recurs, call Technical Support.
351 36 1 Water Supply Stop from CORE Deionized Water supply from CORE stops. a. Ensure that the deionized water supply is on. 1. Defective the water pressure. 1. Check whether deionized water supplier is on.
b. Ensure that the external water supply pressure is 15-25 psi. 2. Check the water pressure
c. Ensure that the external supply water flow rate is sufficient for your analyzer’s
module configuration.
d. Clean the inlet water filter.
e. Resume operation; if alarm recurs, call Technical Support.
361 37 1 Low Pressure in Vacuum Tank The vacuum pump is not supplying enough negative pressure. a. Ensure that there are no air leaks around the fittings on the vacuum tank. The 1. Defective the SV40 or SV41. 1. Check the SVs.
vacuum tank is located behind the left front door of the module. 2. Defective the vacuum pump. 2. Check the vacuum pump.
b. Ensure that the vacuum negative pressure is -0.02 MPa to - 0.05 MPa during 3. Defective the tube 3. Check whether the tube has the holes.
operation. The gauge is visible through the slotted metal to the left of the vacuum 4. Defective the vacuum sensor. 4. Replace the vacuum sensor.
tank. 5. Defective the DS2 PCB or the HMC2 PCB. 5. Replace the PCBs
c. Resume operation; if alarm recurs, call Technical Support.
371 38 1 Liquid in Vacuum Tank Liquid has been detected in the vacuum tank. a. Verify that drain hose to waste container is properly positioned and not 1. Defective the SV30 or SV 31 1. Check the the SVs
pinched. 2. Defective the liquid-full sensor. 2. Check the liquid-full sensor.
b. Verify that main drain tubing from instrument is not pinched. 3. Defective the DS2 PCB or the HMC2 PCB. 3. Replace the PCBs.
c. If liquid exists in vacuum tank, manually drain the tank. The vacuum tank is
located behind the left front door of the module.
d. Resume operation; if alarm recurs, call Technical Support.
391 40 1 Photometer Inner Lamp > 3.3 Abs During normal operation, the cell blank absorbance measurement (Inner Cell Ring) for one a. Verify that the incubation bath is full. If not, touch 1. Defective the lamp. 1. Replace the lamp.
wavelength is greater than 3.3 ABS more than 2 cycles. Utility/Maintenance/Maintenance/Inc. Water Exchange/Select. Execute the 2. Defective the LogADC PCB, the BCR-DIST PCB or the 2. Replace the PCBs.
incubation bath exchange. DS2 PCB. 3. Replace the cells.
b. Ensure that incubation bath is not cloudy and that there is no excessive 3. Defective the cell. 4. Clean up the incubation bath window.
foaming. If either condition exists, clean the incubation bath. 4. Defective the incubation bath window. 5. Replace the photometer.
c. Ensure that the lamp lead wires are not touching and that the wires are 5. Defective the photometer
securely fastened.
d. Verify proper photometer lamp positioning.
e. Touch Utility/Maintenance/Maintenance/Photometer Check/Select. Execute the
photometer check.
f. Resume operation; if alarm recurs, replace photometer lamp according to the
procedure in the Maintenance Guide.
S3 - 20
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
401 40 2 Photometer Inner Lamp > 3.3 Abs During normal operation, the cell blank absorbance measurement (Inner Cell Ring) for all a. Verify that the incubation bath is full. If not, touch 1. Defective the lamp. 1. Replace the lamp.
wavelengths is greater than 3.3 ABS 2 consecutive times. Utility/Maintenance/Maintenance/Inc. Water Exchange/Select. Execute the 2. Defective the LogADC PCB, the BCR-DIST PCB or the 2. Replace the PCBs.
incubation bath exchange. DS2 PCB. 3. Replace the cells.
b. Ensure that incubation bath is not cloudy and that there is no excessive 3. Defective the cell. 4. Clean up the incubation bath window.
foaming. If either condition exists, clean the incubation bath. 4. Defective the incubation bath window. 5. Replace the photometer.
c. Ensure that the lamp lead wires are not touching and that the wires are 5. Defective the photometer
securely fastened.
d. Verify proper photometer lamp positioning.
e. Touch Utility/Maintenance/Maintenance/Photometer Check/Select. Execute the
photometer check.
f. Resume operation; if alarm recurs, replace photometer lamp according to the
procedure in the Maintenance Guide.
402 40 3 Photometer Inner Lamp > 3.3 Abs During normal operation, the cell blank absorbance measurement (Inner Cell Ring) for all a. Verify that the incubation bath is full. If not, touch 1. Defective the lamp. 1. Replace the lamp.
wavelengths is greater than 3.3 ABS 10 consecutive times. Utility/Maintenance/Maintenance/Inc. Water Exchange/Select. Execute the 2. Defective the LogADC PCB, the BCR-DIST PCB or the 2. Replace the PCBs.
incubation bath exchange. DS2 PCB. 3. Replace the cells.
b. Ensure that incubation bath is not cloudy and that there is no excessive 3. Defective the cell. 4. Clean up the incubation bath window.
foaming. If either condition exists, clean the incubation bath. 4. Defective the incubation bath window. 5. Replace the photometer.
c. Ensure that the lamp lead wires are not touching and that the wires are 5. Defective the photometer
securely fastened.
d. Verify proper photometer lamp positioning.
e. Touch Utility/Maintenance/Maintenance/Photometer Check/Select. Execute the
photometer check.
f. Resume operation; if alarm recurs, replace photometer lamp according to the
procedure in the Maintenance Guide.
403 41 1 Photometer Outer Lamp > 3.3 Abs During normal operation, the cell blank absorbance measurement (Outer Cell Ring) for one a. Verify that the incubation bath is full. If not, touch 1. Defective the lamp. 1. Replace the lamp.
wavelength is greater than 3.3 ABS more than 2 cycles. Utility/Maintenance/Maintenance/Inc. Water Exchange/Select. Execute the 2. Defective the LogADC PCB, the BCR-DIST PCB or the 2. Replace the PCBs.
incubation bath exchange. DS2 PCB. 3. Replace the cells.
b. Ensure that incubation bath is not cloudy and that there is no excessive 3. Defective the cell. 4. Clean up the incubation bath window.
foaming. If either condition exists, clean the incubation bath. 4. Defective the incubation bath window. 5. Replace the photometer.
c. Ensure that the lamp lead wires are not touching and that the wires are 5. Defective the photometer
securely fastened.
d. Verify proper photometer lamp positioning.
e. Touch Utility/Maintenance/Maintenance/Photometer Check/Select. Execute the
photometer check.
f. Resume operation; if alarm recurs, replace photometer lamp according to the
procedure in the Maintenance Guide.
411 41 2 Photometer Outer Lamp > 3.3 Abs During normal operation, the cell blank absorbance measurement (Outer Cell Ring) for all a. Verify that the incubation bath is full. If not, touch 1. Defective the lamp. 1. Replace the lamp.
wavelengths is greater than 3.3 ABS 2 consecutive times. Utility/Maintenance/Maintenance/Inc. Water Exchange/Select. Execute the 2. Defective the LogADC PCB, the BCR-DIST PCB or the 2. Replace the PCBs.
incubation bath exchange. DS2 PCB. 3. Replace the cells.
b. Ensure that incubation bath is not cloudy and that there is no excessive 3. Defective the cell. 4. Clean up the incubation bath window.
foaming. If either condition exists, clean the incubation bath. 4. Defective the incubation bath window. 5. Replace the photometer.
c. Ensure that the lamp lead wires are not touching and that the wires are 5. Defective the photometer
securely fastened.
d. Verify proper photometer lamp positioning.
e. Touch Utility/Maintenance/Maintenance/Photometer Check/Select. Execute the
photometer check.
f. Resume operation; if alarm recurs, replace photometer lamp according to the
procedure in the Maintenance Guide.
412 41 3 Photometer Outer Lamp > 3.3 Abs During normal operation, the cell blank absorbance measurement (Outer Cell Ring) for all a. Verify that the incubation bath is full. If not, touch 1. Defective the lamp. 1. Replace the lamp.
wavelengths is greater than 3.3 ABS 10 consecutive times. Utility/Maintenance/Maintenance/Inc. Water Exchange/Select. Execute the 2. Defective the LogADC PCB, the BCR-DIST PCB or the 2. Replace the PCBs.
incubation bath exchange. DS2 PCB. 3. Replace the cells.
b. Ensure that incubation bath is not cloudy and that there is no excessive 3. Defective the cell. 4. Clean up the incubation bath window.
foaming. If either condition exists, clean the incubation bath. 4. Defective the incubation bath window. 5. Replace the photometer.
c. Ensure that the lamp lead wires are not touching and that the wires are 5. Defective the photometer
securely fastened.
d. Verify proper photometer lamp positioning.
e. Touch Utility/Maintenance/Maintenance/Photometer Check/Select. Execute the
photometer check.
f. Resume operation; if alarm recurs, replace photometer lamp according to the
procedure in the Maintenance Guide.
413 42 1 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
501 42 2 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 21
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
502 42 3 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
503 42 4 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
504 42 5 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
505 42 6 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
506 42 7 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
507 42 8 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
508 42 9 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 22
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
509 42 10 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
510 42 11 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
511 42 12 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
512 42 13 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
513 42 14 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
514 42 15 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
515 42 16 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 23
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
516 42 17 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
517 42 18 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
518 42 19 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
519 42 20 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
520 42 21 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
521 42 22 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
522 42 23 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 24
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
523 42 24 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
524 42 25 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
525 42 26 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
526 42 27 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
527 42 28 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
528 42 29 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
529 42 30 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 25
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
530 42 31 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
531 42 32 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
532 42 33 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
533 42 34 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
534 42 35 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
535 42 36 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
536 42 37 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 26
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
537 42 38 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
538 42 39 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
539 42 40 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
540 42 41 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
541 42 42 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
542 42 43 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
543 42 44 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 27
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
544 42 45 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
545 42 46 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
546 42 47 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
547 42 48 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
548 42 49 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
549 42 50 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
550 42 51 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 28
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
551 42 52 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
552 42 53 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
553 42 54 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
554 42 55 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
555 42 56 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
556 42 57 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
557 42 58 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 29
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
558 42 59 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
559 42 60 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
560 42 61 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
561 42 62 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
562 42 63 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
563 42 64 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
564 42 65 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 30
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
565 42 66 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
566 42 67 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
567 42 68 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
568 42 69 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
569 42 70 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
570 42 71 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
571 42 72 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 31
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
572 42 73 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
573 42 74 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
574 42 75 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
575 42 76 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
576 42 77 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
577 42 78 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
578 42 79 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 32
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
579 42 80 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
580 42 81 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
581 42 82 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
582 42 83 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
583 42 84 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
584 42 85 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
585 42 86 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 33
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
586 42 87 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
587 42 88 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
588 42 89 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
589 42 90 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
590 42 91 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
591 42 92 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
592 42 93 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 34
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
593 42 94 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
594 42 95 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
595 42 96 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
596 42 97 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
597 42 98 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
598 42 99 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
599 42 100 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 35
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
600 42 101 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
601 42 102 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
602 42 103 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
603 42 104 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
604 42 105 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
605 42 106 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
606 42 107 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 36
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
607 42 108 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
608 42 109 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
609 42 110 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
610 42 111 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
611 42 112 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
612 42 113 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
613 42 114 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 37
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
614 42 115 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
615 42 116 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
616 42 117 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
617 42 118 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
618 42 119 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
619 42 120 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
620 42 121 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 38
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
621 42 122 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
622 42 123 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
623 42 124 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
624 42 125 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
625 42 126 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
626 42 127 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
627 42 128 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 39
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
628 42 129 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
629 42 130 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
630 42 131 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
631 42 132 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
632 42 133 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
633 42 134 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
634 42 135 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 40
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
635 42 136 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
636 42 137 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
637 42 138 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
638 42 139 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
639 42 140 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
640 42 141 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
641 42 142 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 41
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
642 42 143 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
643 42 144 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
644 42 145 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
645 42 146 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
646 42 147 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
647 42 148 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
648 42 149 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 42
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
649 42 150 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
650 42 151 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
651 42 152 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
652 42 153 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
653 42 154 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
654 42 155 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
655 42 156 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 43
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
656 42 157 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
657 42 158 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
658 42 159 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
659 42 160 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
660 42 161 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
661 42 162 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
662 42 163 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 44
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
663 42 164 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
664 42 165 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
665 42 166 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
666 42 167 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
667 42 168 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
668 42 169 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
669 42 170 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 45
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
670 42 171 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
671 42 172 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
672 42 173 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
673 42 174 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
674 42 175 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
675 42 176 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
676 42 177 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 46
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
677 42 178 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
678 42 179 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
679 42 180 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
680 42 181 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
681 42 182 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
682 42 183 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
683 42 184 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 47
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
684 42 185 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
685 42 186 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
686 42 187 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
687 42 188 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
688 42 189 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
689 42 190 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
690 42 191 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 48
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
691 42 192 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
692 42 193 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
693 42 194 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
694 42 195 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
695 42 196 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
696 42 197 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
697 42 198 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 49
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
698 42 199 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
699 42 200 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
700 42 201 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
701 42 202 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
702 42 203 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
703 42 204 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
704 42 205 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 50
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
705 42 206 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
706 42 207 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
707 42 208 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
708 42 209 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
709 42 210 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
710 42 211 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
711 42 212 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 51
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
712 42 213 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
713 42 214 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
714 42 215 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
715 42 216 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
716 42 217 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
717 42 218 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
718 42 219 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 52
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
719 42 220 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
720 42 221 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
721 42 222 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
722 42 223 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
723 42 224 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
724 42 225 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
725 42 226 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 53
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
726 42 227 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
727 42 228 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
728 42 229 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
729 42 230 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
730 42 231 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
731 42 232 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
732 42 233 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 54
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
733 42 234 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
734 42 235 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
735 42 236 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
736 42 237 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
737 42 238 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
738 42 239 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
739 42 240 Inner Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 55
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
740 43 1 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
741 43 2 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
742 43 3 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
743 43 4 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
744 43 5 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
745 43 6 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
746 43 7 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 56
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
747 43 8 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
748 43 9 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
749 43 10 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
750 43 11 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
751 43 12 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
752 43 13 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
753 43 14 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 57
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
754 43 15 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
755 43 16 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
756 43 17 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
757 43 18 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
758 43 19 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
759 43 20 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
760 43 21 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 58
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
761 43 22 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
762 43 23 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
763 43 24 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
764 43 25 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
765 43 26 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
766 43 27 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
767 43 28 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 59
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
768 43 29 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
769 43 30 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
770 43 31 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
771 43 32 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
772 43 33 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
773 43 34 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
774 43 35 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 60
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
775 43 36 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
776 43 37 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
777 43 38 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
778 43 39 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
779 43 40 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
780 43 41 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
781 43 42 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 61
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
782 43 43 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
783 43 44 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
784 43 45 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
785 43 46 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
786 43 47 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
787 43 48 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
788 43 49 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 62
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
789 43 50 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
790 43 51 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
791 43 52 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
792 43 53 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
793 43 54 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
794 43 55 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
795 43 56 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 63
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
796 43 57 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
797 43 58 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
798 43 59 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
799 43 60 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
800 43 61 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
801 43 62 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
802 43 63 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 64
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
803 43 64 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
804 43 65 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
805 43 66 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
806 43 67 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
807 43 68 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
808 43 69 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
809 43 70 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 65
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
810 43 71 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
811 43 72 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
812 43 73 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
813 43 74 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
814 43 75 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
815 43 76 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
816 43 77 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 66
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
817 43 78 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
818 43 79 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
819 43 80 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
820 43 81 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
821 43 82 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
822 43 83 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
823 43 84 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 67
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
824 43 85 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
825 43 86 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
826 43 87 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
827 43 88 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
828 43 89 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
829 43 90 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
830 43 91 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 68
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
831 43 92 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
832 43 93 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
833 43 94 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
834 43 95 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
835 43 96 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
836 43 97 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
837 43 98 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 69
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
838 43 99 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
839 43 100 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
840 43 101 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
841 43 102 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
842 43 103 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
843 43 104 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
844 43 105 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 70
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
845 43 106 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
846 43 107 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
847 43 108 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
848 43 109 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
849 43 110 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
850 43 111 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
851 43 112 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 71
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
852 43 113 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
853 43 114 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
854 43 115 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
855 43 116 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
856 43 117 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
857 43 118 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
858 43 119 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 72
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
859 43 120 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
860 43 121 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
861 43 122 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
862 43 123 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
863 43 124 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
864 43 125 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
865 43 126 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 73
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
866 43 127 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
867 43 128 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
868 43 129 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
869 43 130 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
870 43 131 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
871 43 132 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
872 43 133 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 74
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
873 43 134 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
874 43 135 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
875 43 136 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
876 43 137 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
877 43 138 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
878 43 139 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
879 43 140 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 75
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
880 43 141 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
881 43 142 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
882 43 143 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
883 43 144 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
884 43 145 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
885 43 146 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
886 43 147 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 76
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
887 43 148 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
888 43 149 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
889 43 150 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
890 43 151 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
891 43 152 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
892 43 153 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
893 43 154 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 77
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
894 43 155 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
895 43 156 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
896 43 157 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
897 43 158 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
898 43 159 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
899 43 160 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
900 43 161 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 78
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
901 43 162 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
902 43 163 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
903 43 164 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
904 43 165 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
905 43 166 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
906 43 167 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
907 43 168 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 79
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
908 43 169 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
909 43 170 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
910 43 171 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
911 43 172 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
912 43 173 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
913 43 174 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
914 43 175 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 80
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
915 43 176 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
916 43 177 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
917 43 178 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
918 43 179 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
919 43 180 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
920 43 181 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
921 43 182 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 81
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
922 43 183 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
923 43 184 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
924 43 185 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
925 43 186 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
926 43 187 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
927 43 188 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
928 43 189 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 82
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
929 43 190 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
930 43 191 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
931 43 192 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
932 43 193 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
933 43 194 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
934 43 195 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
935 43 196 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 83
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
936 43 197 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
937 43 198 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
938 43 199 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
939 43 200 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
940 43 201 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
941 43 202 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
942 43 203 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 84
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
943 43 204 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
944 43 205 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
945 43 206 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
946 43 207 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
947 43 208 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
948 43 209 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
949 43 210 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 85
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
950 43 211 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
951 43 212 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
952 43 213 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
953 43 214 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
954 43 215 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
955 43 216 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
956 43 217 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 86
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
957 43 218 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
958 43 219 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
959 43 220 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
960 43 221 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
961 43 222 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
962 43 223 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
963 43 224 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 87
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
964 43 225 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
965 43 226 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
966 43 227 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
967 43 228 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
968 43 229 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
969 43 230 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
970 43 231 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 88
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
971 43 232 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
972 43 233 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
973 43 234 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
974 43 235 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
975 43 236 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
976 43 237 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
977 43 238 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 89
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
978 43 239 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
wavelength by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
979 43 240 Outer Cell Blank out of Limits (2X) The ADC value of the cell blank measurement of one cell (Outer Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for two consecutive measurements. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1520 44 1 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1521 44 2 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1522 44 3 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1523 44 4 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1524 44 5 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 90
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1525 44 6 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1526 44 7 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1527 44 8 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1528 44 9 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1529 44 10 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1530 44 11 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1531 44 12 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 91
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1532 44 13 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1533 44 14 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1534 44 15 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1535 44 16 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1536 44 17 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1537 44 18 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1538 44 19 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 92
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1539 44 20 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1540 44 21 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1541 44 22 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1542 44 23 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1543 44 24 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1544 44 25 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1545 44 26 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 93
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1546 44 27 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1547 44 28 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1548 44 29 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1549 44 30 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1550 44 31 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1551 44 32 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1552 44 33 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 94
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1553 44 34 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1554 44 35 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1555 44 36 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1556 44 37 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1557 44 38 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1558 44 39 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1559 44 40 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 95
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1560 44 41 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1561 44 42 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1562 44 43 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1563 44 44 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1564 44 45 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1565 44 46 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1566 44 47 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 96
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1567 44 48 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1568 44 49 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1569 44 50 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1570 44 51 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1571 44 52 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1572 44 53 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1573 44 54 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 97
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1574 44 55 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1575 44 56 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1576 44 57 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1577 44 58 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1578 44 59 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1579 44 60 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1580 44 61 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 98
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1581 44 62 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1582 44 63 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1583 44 64 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1584 44 65 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1585 44 66 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1586 44 67 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1587 44 68 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 99
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1588 44 69 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1589 44 70 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1590 44 71 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1591 44 72 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1592 44 73 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1593 44 74 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1594 44 75 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 100
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1595 44 76 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1596 44 77 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1597 44 78 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1598 44 79 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1599 44 80 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1600 44 81 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1601 44 82 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 101
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1602 44 83 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1603 44 84 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1604 44 85 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1605 44 86 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1606 44 87 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1607 44 88 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1608 44 89 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 102
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1609 44 90 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1610 44 91 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1611 44 92 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1612 44 93 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1613 44 94 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1614 44 95 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1615 44 96 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 103
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1616 44 97 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1617 44 98 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1618 44 99 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1619 44 100 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1620 44 101 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1621 44 102 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1622 44 103 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 104
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1623 44 104 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1624 44 105 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1625 44 106 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1626 44 107 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1627 44 108 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1628 44 109 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1629 44 110 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 105
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1630 44 111 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1631 44 112 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1632 44 113 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1633 44 114 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1634 44 115 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1635 44 116 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1636 44 117 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 106
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1637 44 118 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1638 44 119 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1639 44 120 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1640 44 121 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1641 44 122 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1642 44 123 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1643 44 124 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 107
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1644 44 125 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1645 44 126 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1646 44 127 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1647 44 128 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1648 44 129 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1649 44 130 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1650 44 131 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 108
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1651 44 132 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1652 44 133 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1653 44 134 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1654 44 135 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1655 44 136 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1656 44 137 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1657 44 138 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 109
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1658 44 139 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1659 44 140 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1660 44 141 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1661 44 142 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1662 44 143 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1663 44 144 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1664 44 145 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 110
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1665 44 146 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1666 44 147 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1667 44 148 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1668 44 149 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1669 44 150 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1670 44 151 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1671 44 152 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 111
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1672 44 153 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1673 44 154 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1674 44 155 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1675 44 156 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1676 44 157 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1677 44 158 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1678 44 159 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 112
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1679 44 160 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1680 44 161 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1681 44 162 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1682 44 163 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1683 44 164 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1684 44 165 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1685 44 166 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 113
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1686 44 167 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1687 44 168 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1688 44 169 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1689 44 170 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1690 44 171 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1691 44 172 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1692 44 173 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 114
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1693 44 174 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1694 44 175 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1695 44 176 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1696 44 177 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1697 44 178 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1698 44 179 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1699 44 180 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 115
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1700 44 181 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1701 44 182 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1702 44 183 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1703 44 184 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1704 44 185 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1705 44 186 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1706 44 187 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 116
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1707 44 188 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1708 44 189 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1709 44 190 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1710 44 191 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1711 44 192 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1712 44 193 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1713 44 194 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 117
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1714 44 195 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1715 44 196 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1716 44 197 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1717 44 198 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1718 44 199 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1719 44 200 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1720 44 201 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 118
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1721 44 202 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1722 44 203 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1723 44 204 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1724 44 205 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1725 44 206 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1726 44 207 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1727 44 208 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 119
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1728 44 209 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1729 44 210 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1730 44 211 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1731 44 212 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1732 44 213 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1733 44 214 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1734 44 215 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 120
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1735 44 216 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1736 44 217 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1737 44 218 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1738 44 219 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1739 44 220 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1740 44 221 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1741 44 222 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 121
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1742 44 223 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1743 44 224 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1744 44 225 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1745 44 226 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1746 44 227 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1747 44 228 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1748 44 229 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 122
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1749 44 230 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1750 44 231 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1751 44 232 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1752 44 233 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1753 44 234 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1754 44 235 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1755 44 236 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 123
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1756 44 237 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1757 44 238 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1758 44 239 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1759 44 240 Inner Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1760 45 1 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1761 45 2 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1762 45 3 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 124
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1763 45 4 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1764 45 5 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1765 45 6 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1766 45 7 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1767 45 8 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1768 45 9 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1769 45 10 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 125
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1770 45 11 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1771 45 12 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1772 45 13 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1773 45 14 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1774 45 15 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1775 45 16 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1776 45 17 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 126
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1777 45 18 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1778 45 19 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1779 45 20 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1780 45 21 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1781 45 22 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1782 45 23 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1783 45 24 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 127
AlarmCode AlarmType AlarmNo AlarmMessage AlarmDesc AlarmRemedy Cause Remedy
1784 45 25 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1785 45 26 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1786 45 27 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1787 45 28 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1788 45 29 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1789 45 30 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
1790 45 31 Outer Cell Blank out of Limits (3X) The ADC value of the cell blank measurement of one cell (Inner Cell Ring) differs from the a. Ensure that there is no excessive foaming or particles in the incubation bath. 1. Defective the cell. 1. Replace the cell
reference value of the cell blank measurement (taken from maintenance) for one wavelength b. Verify that the reaction cell is not contaminated or cracked. 2. Bubbles or foreign bodys 2. Check the Hitergent volume(3.5-7.0ml) and remove the foreign bodies.
by more than 0.1 ABS for three consecutive measurements and more. c. Wipe the outside of the reaction cells with gauze moistened with incubation 3. Defective the window. 3. Check whether the window is fixed sufficiently and water leakage.
Subcode 1-240 indicates the cell number. bath water. 4. Defective the LogADC PCB, the BCRDIST PCB or ,the 4. Replace the LogADC PCB.
NOTE: This cell is not used for analysis. d. Touch Utility/Maintenance/Maintenance/ Cell Wash/Select. Execute a Cell DS2 PCB.
Wash.
e. Resume operation. If alarm recurs, ensure that there is an adequate amount of
rinse water from the rinse mechanism.
f. Resume operation; if alarm recurs, touch Utility/Maintenance/Maintenance/Cell
Blank/Select. Execute a cell blank. If cells are out of specification, replace the
reaction cells according to the procedure in the Maintenance Guide.
S3 - 128