You are on page 1of 24

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles

for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Designation: D1414 − 22

Standard Test Methods for


Rubber O-Rings1
This standard is issued under the fixed designation D1414; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the U.S. Department of Defense.

1. Scope Retraction at Lower Temperatures (TR Test)


1.1 These test methods describe the procedures for deter- D1415 Test Method for Rubber Property—International
mining the physical properties of O-rings and changes in these Hardness
properties due to aging. D2240 Test Method for Rubber Property—Durometer Hard-
ness
1.2 The values stated in SI units are to be regarded as the D3767 Practice for Rubber—Measurement of Dimensions
standard. The values given in parentheses are for information D4483 Practice for Evaluating Precision for Test Method
only. Standards in the Rubber and Carbon Black Manufacturing
1.3 This standard does not purport to address all of the Industries
safety concerns, if any, associated with its use. It is the 2.2 Aerospace Standard:
responsibility of the user of this standard to establish appro- AS568A O–ring Sizes3
priate safety, health, and environmental practices and deter-
mine the applicability of regulatory limitations prior to use. 3. Terminology
1.4 This international standard was developed in accor- 3.1 Definitions:
dance with internationally recognized principles on standard- 3.1.1 O-ring—a rubber seal of homogeneous composition
ization established in the Decision on Principles for the molded in one piece to the configuration of a torus with circular
Development of International Standards, Guides and Recom- cross section. The O-ring is used as a dynamic or static seal
mendations issued by the World Trade Organization Technical usually installed in a machined groove.
Barriers to Trade (TBT) Committee.
4. Significance and Use
2. Referenced Documents
4.1 These test methods provide acceptable procedures for
2.1 ASTM Standards:2 quality control purposes and for the determination of engineer-
D297 Test Methods for Rubber Products—Chemical Analy- ing characteristics.
sis
D395 Test Methods for Rubber Property—Compression Set 5. General Methods
D412 Test Methods for Vulcanized Rubber and Thermoplas- 5.1 Except as otherwise specified in these test methods for
tic Elastomers—Tension rubber O-rings, the following ASTM test methods applicable in
D471 Test Method for Rubber Property—Effect of Liquids general to vulcanized rubber, shall be complied with as
D573 Test Method for Rubber—Deterioration in an Air required and are hereby made a part of these test methods:
Oven 5.1.1 Tension Test—Test Methods D412.
D865 Test Method for Rubber—Deterioration by Heating in 5.1.2 Compression Set—Test Methods D395.
Air (Test Tube Enclosure) 5.1.3 Low Temperature—Test Method D1329.
D1329 Test Method for Evaluating Rubber Property— 5.1.4 Density—Test Methods D297.
5.1.5 Fluid Aging—Test Method D471.
1
5.1.6 Heat Aging—Test Methods D865 and D573.
These test methods are under the jurisdiction of ASTM Committee D11 on
Rubber and Rubber-like Materials and are the direct responsibility of Subcommittee
5.1.7 Hardness—Test Methods D2240 and D1415.
D11.37 on Coated Fabrics, Rubber Threads and Seals. 5.2 In case of conflict between the provisions of these test
Current edition approved Oct. 1, 2022. Published December 2022. Originally
approved in 1956. Last previous edition approved in 2015 as D1414 – 15. DOI:
methods and those of detailed procedures herein, the latter
10.1520/D1414-22. shall take precedence.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
3
Standards volume information, refer to the standard’s Document Summary page on Standardization Documents Order Desk, Bldg. 4 Section D, 700 Robbins Ave.,
the ASTM website. Philadelphia, PA 19111-5094, Attn: NPODS.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States

1
D1414 − 22
6. Test Conditions spool one full revolution for each 150 mm (6.0 in.) of travel of
6.1 The temperature of the testing room shall be 23 6 2°C the power-driven grips or (2) lubricating the contact surface of
(73.4 6 3.6°F). The specimens to be tested shall be kept in this the spools with castor oil. Grips for testing straight specimens
room for a minimum of 30 min previous to the time of testing. may be of any type provided breaks do not occur in a portion
For referee tests the minimum time for conditioning shall be 16 of the specimen that has been at any time in contact with the
h. In cases where this temperature cannot be obtained, the grips. For specimens smaller than 25 mm (1 in.) in inside
actual temperature shall be reported. diameter, spools of smaller diameter and closer approach may
be employed.
7. Dimensional Measurements 8.2 Test Specimen:
7.1 Thickness: 8.2.1 The test specimen shall consist of an entire O-ring
7.1.1 The thickness or cross section shall be measured by a except in the following cases: (1) when the diameter is too
micrometer in accordance with Practice D3767, Procedure A. large to permit a break within the limits of travel of the testing
7.1.2 The O-ring specimen shall be measured at four points machine, (2) when the specimen has been previously cut open
equally distributed around the circumference in both the radial to permit an aging test, and (3) when the specimen is unaged
and axial direction. An average reading shall be used for but is to be compared with an aged specimen as described in
calculation. (2). For these cases a specimen cut from a ring shall be
7.2 Diameter—A stepped cone or optical comparator may employed.
be used for measuring the inside diameter of small rings. The 8.2.2 Specimens that exhibit obvious flaws at any point on
diametric intervals of the stepped cone shall not exceed 2 % of the surface, such as might result from imperfect molding or
the diameter to be measured (see Fig. 1). flash removal, shall not be tested.
8.3 Procedure for Determination of Tensile Strength, Ulti-
8. Tension Testing
mate Elongation, and Tensile Stress:
8.1 Testing Machine— The testing machine shall conform to 8.3.1 Bring the grips close enough together so the specimen
the requirements specified in Section 3 of Test Methods D412, can be installed without stretching. Set the machine for a speed
with the exception of the grips. Grips for testing rings shall of 500 6 50 mm (20 6 2 in.)/min, set the chart recorder, and
consist of ball-bearing spools at least 9 mm (0.35 in.) in engage the clutch. (If a recorder is not used, note the spool
diameter, and be capable of being brought within 19 mm (0.75 center-to-center distance continually by means of a suitable
in.) center-to-center distance at closest approach. Stresses scale). The stressed inside circumference, L, is equal to twice
within the specimen shall be minimized by (1) rotating one the distance between spools plus the circumference of one
spool. Record the breaking force value, F, at the elongation
specified for tensile stress and at the time of rupture. Measure
the center-to-center distance between the spools at rupture to
the nearest 2.5 mm (0.1 in.) and record as the value for D.
8.3.2 For the application of bench marks to segmented
specimens as described in 8.2, hold the piece as straight as
possible being careful not to apply excessive stress. Follow the
procedure described in 8.3.1. Discard results for any specimen
that breaks at a point that has at any time contacted the grip,
and test an additional specimen. A suitable extensiometer may
be employed for elongation determination.
8.4 Calculations:
8.4.1 Tensile Strength—Calculate the tensile strength as
follows:
8.4.1.1 For O-rings on spool grips:
T 5 F/A (1)

where:
T = tensile strength,
F = breaking force, and
A = twice the cross-sectional area calculated from axial
thickness, W, as follows:
A 5 π W 2 /2 5 1.57 W 2
(2)

therefore:
2
T 5 F/1.57 W (3)

FIG. 1 Stepped Cone Gage 8.4.1.2 For straight or single-strand specimens:

2
D1414 − 22
A 5 0.785 W 2
(4) 8.6.1 The report shall include the following:
T 5 F/0.785 W 2
(5) 8.6.1.1 Results calculated as in 8.4,
8.6.1.2 Date of test,
8.4.2 Ultimate Elongation—Calculate the ultimate elonga- 8.6.1.3 Temperature of test room if other than provided for
tion as follows: in Section 6,
8.4.2.1 For rings: 8.6.1.4 Type of testing machine used, including diameter of
Ultimate elongation, % 5 @ ~ 2D1G 2 C ! /C # 3 100 (6) spool grip if other than standard, and speed, if other than 8.3 6
where: 0.8 mm (20.0 6 2 in.)/min, and
8.6.1.5 Type of specimen (ring or straight).
D = distance between centers of the spool grips at the time of
rupture of specimen,
G = circumference of one spool (spool diameter × 3.14), and 9. Tension Testing Precision and Bias
C = inside circumference of the specimen (or inside 9.1 These precision statements have been prepared in ac-
diameter × 3.14). cordance with Practice D4483. Refer to this practice for
8.4.2.2 For straight or single-strand specimen cut from an terminology and other statistical details.
O-ring: 9.2 Precision—The precision results in this precision and
Ultimate elongation, % 5 @ ~ L 2 L bias section give an estimate of the precision of these test
o ! /L o # 3 100 (7)
methods with the materials (rubbers, carbon blacks, etc.) used
where: in the interlaboratory program described in Table 1A and Table
L = observed distance between bench marks of stressed 1B. The precision parameters should not be used for accep-
specimen, and tance or rejection testing of any group of materials without
Lo = original distance between bench marks. documentation that they are applicable to those particular
8.4.3 Tensile Stress— Calculate the tensile stress as in the materials and the specific testing protocols of these test
case of tensile strength except the stress is reported at a methods.
predetermined elongation below the ultimate elongation. Cal- 9.3 A type 1 inter-laboratory precision program for tensile
culate the center-to-center distance of the spool grips, D, where strength was conducted during the periods listed with the
the prescribed elongation, E, occurs, as follows: number of laboratories listed and the o-ring materials listed in
D 5 1/2 @ ~ EC/100! 1C 2 G # (8) Table 1A. In each laboratory, two single tests were performed
NOTE 1—This value of tensile stress at a predetermined elongation is on the o-rings by 2 different technicians on 2 different days for
based upon inside circumference and differs from that determined by Test a total of four replicates. A test result is the value obtained from
Methods D412, which is based on the mean circumference. a single determination. Both repeatability and reproducibility
8.5 Characteristics of Sample Tested—The median of the represent short-term (daily) testing conditions. Acceptable
values for three specimens shall be taken as the characteristics difference values were not measured.
of the sample tested, except that under the following conditions 9.4 Tensile Strength Repeatability—The relative repeatabil-
the median of the values for five specimens (two additional) ity limits, (r), of this test method are as shown in Table 1A for
shall be used: each of the materials listed in the table. The difference between
8.5.1 If one or more values do not meet the specified two single test results (or determinations) found on identical
requirements when testing for compliance with specifications. test material under the repeatability conditions prescribed for
8.5.2 If referee tests are being made. this test will exceed the repeatability value on an average of not
8.6 Report: more than once in 20 determinations in the normal and correct

TABLE 1 A Precision Parameters for Test Methods D1414, Tensile Strength (Type 1 Precision), Size 2-214 O-rings
Units MPa
Number of
Material Period Laboratories Mean Level Sr r (r) SR R (R)
(M/V)A
FVMQ; MIL- Sep 2020 6(0/0) 7.1 0.66 1.87 26.3 0.70 1.99 27.9
DTL-25988C;
T1, C1, G60
FVMQ; MIL- Oct ’19; Feb 7(0/1) 8.2 0.24 0.68 8.2 0.74 2.09 25.3
DTL-25988C; ’20
T1, C3, G75
FKM; April 2021 3(0/0) 12.1 1.98 5.61 46.4 2.22 6.30 52.0
AMS7259E
NBR; AMS-P- Sep-Dec 2018 10(1/0) 12.2 0.45 1.28 10.5 0.54 1.54 12.6
5315B
FKM; AMS7287 May-Jun 2019 8(1/1) 12.5 0.80 2.28 18.3 0.80 2.28 18.3
Average 10.4
Pooled Values 1.03 2.91 27.9 1.18 3.33 31.9
A
M=Mean outliers; V=Variability outliers

3
D1414 − 22
TABLE 1 B Precision Parameters for Test Methods D1414, Ultimate Elongation (Type 1 Precision), Size 2-214 O-rings (continued)
Units Percent
Number of
Material Period Laboratories Mean Level Sr r (r) SR R (R)
(M/V)A
FKM; April 2021 3(0/0) 165.2 22.61 63.97 38.7 23.54 66.63 40.3
AMS7259E
FVMQ; MIL- Oct ’19; Feb 7(1/1) 176.9 8.98 25.41 14.4 11.72 33.17 18.8
DTL-25988C; ’20
T1, C3, G75
FKM; AMS7287 May-Jun 2019 8(1/0) 233.8 19.07 53.97 23.1 19.07 53.97 23.1
NBR; AMS-P- Sep-Dec 2018 10(1/1) 248.9 7.82 22.13 8.9 16.87 47.75 19.2
5315B
FVMQ; MIL- Sep 2020 6(0/0) 285.6 21.94 62.10 21.7 27.23 77.05 27.0
DTL-25988C;
T1, C1, G60
Average 222.1
Pooled Values 17.31 48.98 22.1 20.40 57.74 26.0
A
M=Mean outliers; V=Variability outliers

operation of the method. Two single test results that differ by percent and relative precision is a percent, the expression of
more than the appropriate value from Table 1A or Table 1B relative precision would be calculated as a percent of a percent.
must be suspected of being from different populations and This can be difficult to apply when evaluating the testing
some appropriate action taken. precision. Therefore, absolute precision is preferred for prac-
NOTE 2—Appropriate action may be an investigation of the test method tical applications.
procedure or apparatus for faulty operation or the declaration of a
significant difference in the two materials, samples, etc., which generated 9.9 Ultimate Elongation Repeatability—The absolute re-
the two test results. peatability limits, r, of this test method are as shown in Table
9.5 Tensile Strength Reproducibility—The relative repro- 1B for each of the materials listed in the table. The difference
ducibility limits, (R), of this test method are as shown in Table between two single test results (or determinations) found on
1A for each of the materials listed in the table. The difference identical test material under the repeatability conditions pre-
between two single and independent test results found by two scribed for this test will exceed the repeatability value on an
operators working under the prescribed reproducibility condi- average of not more than once in 20 determinations in the
tions in different laboratories on identical test material will normal and correct operation of the method. Two single test
exceed the reproducibility value on an average of not more results that differ by more than the appropriate value from
than once in 20 determinations in the normal and correct Table 1B must be suspected of being from different populations
operation of the method. Two single test results produced in and some appropriate action taken.
different laboratories that differ by more than the appropriate NOTE 4—Appropriate action may be an investigation of the test method
procedure or apparatus for faulty operation or the declaration of a
value from Table 1A must be suspected of being from different significant difference in the two materials, samples, etc., which generated
populations and some appropriate investigative or technical/ the two test results.
commercial action taken.
NOTE 3—Appropriate action may be an investigation in each laboratory 9.10 Ultimate Elongation Reproducibility—The absolute
of the test method procedure or apparatus for faulty operation or the reproducibility limits, R, of this test method are as shown in
declaration of a significant difference in the two materials, samples, Table 1B for each of the materials listed in the table. The
laboratories, etc., which generated the two test results. Technical/ difference between two single and independent test results
commercial actions are mutually resolved by the laboratories. found by two operators working under the prescribed repro-
9.6 Bias—Bias is the difference between a test value and a ducibility conditions in different laboratories on identical test
reference value. However, for this test method, bias has not material will exceed the reproducibility value on an average of
been determined. not more than once in 20 determinations in the normal and
9.7 A type 1 inter-laboratory precision program for ultimate correct operation of the method. Two single test results
elongation was conducted during the periods listed with the produced in different laboratories that differ by more than the
number of laboratories listed and the o-ring materials listed in appropriate value from Table 1B must be suspected of being
Table 1B. In each laboratory, two single tests were performed from different populations and some appropriate investigative
on the o-rings by 2 different technicians on 2 different days for or technical/commercial action taken.
a total of four replicates. A test result is the value obtained from NOTE 5—Appropriate action may be an investigation in each laboratory
of the test method procedure or apparatus for faulty operation or the
a single determination. Both repeatability and reproducibility declaration of a significant difference in the two materials, samples,
represent short-term (daily) testing conditions. Acceptable laboratories, etc., which generated the two test results. Technical/
difference values were not measured. commercial actions are mutually resolved by the laboratories.
9.8 Relative versus absolute repeatability and reproducibil- 9.11 Bias—Bias is the difference between a test value and a
ity limits. For ultimate elongation, relative precision is statis- reference value. However, for this test method, bias has not
tically preferred. However, since the measurement units are in been determined.

4
D1414 − 22
10. Tension Set Test 11.3 Test Specimen— The specimen shall be a suitable
10.1 Procedure: section cut from the sample O-ring. A52-mm (2-in.) section has
10.1.1 Stretch the O-ring specimens to 100 % elongation been found to be satisfactory on sample O-rings of 17-mm
and hold at this strain for 10 min. At the end of this period (0.67-in.) or larger inside diameter. On smaller O-rings, a
release the specimen and allow it to recover for 10 min. At the section of approximately 3 mm (1⁄8 in.) shall be cut from the
end of this recovery period, measure the inside diameter with sample and discarded.
a step gage reference in 7.2 and shown in Fig. 1. NOTE 6—Testing a whole O-ring for compression set is always avoided
10.1.2 Apply the strain by (1) stretching the specimen over because air trapped inside the O-ring results in a different set of conditions
a tapered cone which has a step at the required 100 % on the inside and outside surface of the specimen.
elongation, or (2) by means of a tension test machine stopping 11.4 Procedure—Measure the actual specimen to within
the machine at 100 % elongation, or ( 3) by the use of any 0.025 mm (0.001 in.) and calculate the spacer to be 75 % of
special fixture capable of holding the O-ring over two spools at this thickness, to within 0.025 mm (0.001 in.). Close the plates,
constant distance between spool centers. employing a sufficient number of bolts to prevent deflection
10.2 Calculation— Calculate tension set as follows: upon heating. At the end of the heating period, release the
plates as soon as possible and allow the specimens to cool to
Tension set, % 5 @ ~ L 1 2 L o ! /L o # 3 100 (9) room temperature under the conditions described in Test
where: Methods D395.
L1 = diameter after testing, and
Lo = original diameter. 12. Compression Set Test Precision and Bias
12.1 These precision statements have been prepared in
10.3 Straight specimens cut from large O-rings may be
accordance with Practice D4483. Refer to this practice for
tested and calculated in accordance with Test Methods D412.
terminology and other statistical details.
11. Compression Set Test 12.2 Precision—The precision results in this precision and
11.1 Nature of Test— The compression set of O-ring speci- bias section give an estimate of the precision of this test
mens shall follow the general procedures of Method B of Test method with the materials (rubbers, carbon blacks, etc.) used in
Methods D395, with the following exceptions: the interlaboratory program described in Table 2. The precision
11.2 Apparatus: parameters should not be used for acceptance or rejection
11.2.1 Spacers—The spacers shall be adjusted to within testing of any group of materials without documentation that
0.025 mm (0.01 in.). This can be done by employing shim they are applicable to those particular materials, test
stock of various thickness. conditions, and the specific testing protocols of this test
11.2.2 Measurement— The cross section before and after method.
test shall be measured in the same marked location with a 12.3 A type 1 inter-laboratory precision program for com-
ball-point (tubing) micrometer or other method as described in pression set was conducted during the periods listed with the
7.1. number of laboratories listed with the o-ring materials listed
TABLE 2 Precision Parameters for Test Methods D395, Method B, Compression Set (Type 1 Precision), Size 2-214 O-rings
Units Percent
Number of
Test Condi-
Material Period Laboratories Mean Level Sr r (r) SR R (R)
tions
(M/V)A
FVMQ; MIL- Sep 2020 Air; 23.9°C; 5(1/1) 8.4 1.54 4.37 52.0 1.54 4.37 52.0
DTL-25988C; 70 h
T1, C1, G60
FKM; May-Jun 2019 Air; 200°C; 8(1/1) 9.0 0.47 1.33 14.8 0.67 1.88 20.9
AMS7287 22 h
FKM; Apr 2017 Air; 200°C; 8(0/0) 9.7 1.02 2.87 29.6 1.53 4.32 44.5
AMS7276H 22 h
FVMQ; MIL- Oct ’19; Feb Air; 175°C; 5(0/0) 10.9 1.28 3.61 33.1 1.75 4.96 45.5
DTL-25988C; ’20 22 h
T1, C3, G75
FVMQ; MIL- Oct ’19; Feb Air; 23.9°C; 5(0/1) 12.3 1.33 3.77 30.6 2.79 7.88 64.0
DTL-25988C; ’20 70 h
T1, C3, G75
FVMQ; MIL- Sep 2020 Air; 175°C; 5(0/0) 18.8 3.59 10.16 54.2 3.62 10.25 54.6
DTL-25988C; 22 h
T1, C1, G60
FKM; April 2021 Air; 200°C; 3(0/0) 19.6 2.11 5.98 30.6 2.73 7.72 39.4
AMS7259E 22 h
NBR; AMS-P- Sep-Dec 2018 Air; 70°C; 10(1/0) 20.7 1.11 3.14 15.2 1.22 3.46 16.7
5315B 70 h
Average 13.7
Pooled Values 1.79 5.07 37.1 2.18 6.17 45.1
A
M=Mean outliers; V=Variability outliers

5
D1414 − 22
and the test conditions listed in Table 2. In each laboratory, two from different populations and some appropriate investigative
single tests were performed on the o-rings by 2 different or technical/commercial action taken.
technicians on 2 different days for a total of four replicates. A NOTE 8—Appropriate action may be an investigation in each laboratory
test result is the value obtained from a single determination. of the test method procedure or apparatus for faulty operation or the
declaration of a significant difference in the two materials, samples,
Both repeatability and reproducibility represent short-term laboratories, etc., which generated the two test results. Technical/
(daily) testing conditions. Acceptable difference values were commercial actions are mutually resolved by the laboratories.
not measured.
12.7 Bias—Bias is the difference between a test value and a
12.4 Relative versus absolute repeatability and reproducibil- reference value. However, for this test method, bias has not
ity limits. For compression set, relative precision is statistically been determined.
preferred. However, since the measurement units are in percent
and relative precision is a percent, the expression of relative 13. Compression Set in Liquids
precision would be calculated as a percent of a percent. This 13.1 Compression set may be conducted immersed in any
can be difficult to apply when evaluating the testing precision. liquid provided a suitable container is available. The container
Therefore, absolute precision is preferred for practical appli- needs to be large enough to hold the plates and withstand the
cations. vapor pressure of the liquid at test temperature.
12.5 Compression Set Repeatability—The absolute repeat- 13.2 The specimen is prepared in accordance with 11.3, and
ability limits, r, of this test method are as shown in Table 2 for at the termination of the heating period the plates are released
each of the materials and conditions listed in the table. The immediately and the specimens are allowed to cool to room
difference between two single test results (or determinations) temperature for 30 min in a fresh cool portion of the test liquid.
found on identical test material under the repeatability condi- Measurement and reporting are the same except the type of
tions prescribed for this test will exceed the repeatability value liquid is listed in the report.
on an average of not more than once in 20 determinations in the
normal and correct operation of the method. Two single test 14. Compression Set Test in Lubricants Precision and
results that differ by more than the appropriate value from Bias
Table 2 must be suspected of being from different populations 14.1 These precision statements have been prepared in
and some appropriate action taken. accordance with Practice D4483. Refer to this practice for
NOTE 7—Appropriate action may be an investigation of the test method terminology and other statistical details.
procedure or apparatus for faulty operation or the declaration of a
significant difference in the two materials, samples, etc., which generated 14.2 Precision—The precision results in this precision and
the two test results. bias section give an estimate of the precision of this test
12.6 Compression Set Reproducibility—The absolute repro- method with the materials (rubbers, carbon blacks, etc.) used in
ducibility limits, R, of this test method are as shown in Table the interlaboratory program described in Table 3. The precision
2 for each of the materials and conditions listed in the table. parameters should not be used for acceptance or rejection
The difference between two single and independent test results testing of any group of materials without documentation that
found by two operators working under the prescribed repro- they are applicable to those particular materials, test
ducibility conditions in different laboratories on identical test conditions, and the specific testing protocols of this test
material will exceed the reproducibility value on an average of method.
not more than once in 20 determinations in the normal and 14.3 A type 1 inter-laboratory precision program for com-
correct operation of the method. Two single test results pression set in lubricants was conducted during the periods
produced in different laboratories that differ by more than the listed with the number of laboratories listed with the o-ring
appropriate value from Table 2 must be suspected of being materials listed and the test conditions with lubricants listed in

TABLE 3 Precision Parameters for Test Methods D395, Method B, Compression Set (Type 1 Precision), Size 2-214 O-rings
Units Percent
Number of
Test Condi-
Material Period Laboratories Mean Level Sr r (r) SR R (R)
tions
(M/V)A
FKM; April 2021 AMS 3023 at 2(0/0) 12.4 1.62 4.59 37.1 1.62 4.59 37.1
AMS7259E 200°C for
70 h
FVMQ; MIL- Oct ’19 AMS 3021 at 4(0/0) 29.8 3.25 9.19 30.9 5.18 14.67 49.3
DTL-25988C; 175°C for
T1, C3, G75 70 h
FVMQ; MIL- Sep 2020 AMS 3021 at 4(0/1) 31.1 2.53 7.15 23.0 8.37 23.67 76.2
DTL-25988C; 150°C for
T1, C1, G60 70 h
Average 24.4
Pooled Values 2.55 7.23 29.6 5.76 16.30 66.8
A
M=Mean outliers; V=Variability outliers

6
D1414 − 22
Table 3. In each laboratory, two single tests were performed on being from different populations and some appropriate inves-
the o-rings by 2 different technicians on 2 different days for a tigative or technical/commercial action taken.
total of four replicates. A test result is the value obtained from NOTE 10—Appropriate action may be an investigation in each labora-
a single determination. Both repeatability and reproducibility tory of the test method procedure or apparatus for faulty operation or the
represent short-term (daily) testing conditions. Acceptable declaration of a significant difference in the two materials, samples,
laboratories, etc., which generated the two test results. Technical/
difference values were not measured. commercial actions are mutually resolved by the laboratories.
14.4 Relative versus absolute repeatability and reproducibil- 14.7 Bias—Bias is the difference between a test value and a
ity limits. For compression set in lubricants, relative precision reference value. However, for this test method, bias has not
is statistically preferred. However, since the measurement units been determined.
are in percent and relative precision is a percent, the expression
of relative precision would be calculated as a percent of a 15. Low-Temperature Test
percent. This can be difficult to apply when evaluating the
testing precision. Therefore, absolute precision is preferred for 15.1 Nature of Test— The TR-10 or Temperature Retraction
practical applications. Test shall be in accordance with Test Method D1329 with the
exceptions given in 15.2 to 15.4.
14.5 Compression Set in Lubricants Repeatability—The
absolute repeatability limits, r, of this test method are as shown 15.2 Test Specimen— The test specimen shall be an O-ring
in Table 3 for each of the materials and conditions listed in the of 1.5 to 3.8-mm (0.06 to 0.15-in.) cross-sectional diameter and
table. The difference between two single test results (or of sufficient length to reach from clamp to clamp when the ring
determinations) found on identical test material under the is cut and held in straight position.
repeatability conditions prescribed for this test will exceed the 15.3 Procedure—Stretch the mounted specimen 50 % as
repeatability value on an average of not more than once in 20 standard. If a greater or lesser stretch is used, report the actual
determinations in the normal and correct operation of the stretch.
method. Two single test results that differ by more than the
appropriate value from Table 3 must be suspected of being 15.4 Report—The report shall include the following:
from different populations and some appropriate action taken. 15.4.1 Testing elongation, in percent, and
NOTE 9—Appropriate action may be an investigation of the test method 15.4.2 Temperature at which the specimen retracts 10 %,
procedure or apparatus for faulty operation or the declaration of a designated as TR-10 and at other points, such as 30, 50, or
significant difference in the two materials, samples, etc., which generated 70 %, or all, if required.
the two test results.
14.6 Compression Set in Lubricants Reproducibility—The 16. Low-Temperature (TR10) Test Precision and Bias
absolute reproducibility limits, R, of this test method are as
shown in Table 3 for each of the materials and conditions listed 16.1 These precision statements have been prepared in
in the table. The difference between two single and indepen- accordance with Practice D4483. Refer to this practice for
dent test results found by two operators working under the terminology and other statistical details.
prescribed reproducibility conditions in different laboratories 16.2 Precision—The precision results in this precision and
on identical test material will exceed the reproducibility value bias section give an estimate of the precision of this test
on an average of not more than once in 20 determinations in the method with the materials (rubbers, carbon blacks, etc.) used in
normal and correct operation of the method. Two single test the interlaboratory program described in Table 4. The precision
results produced in different laboratories that differ by more parameters should not be used for acceptance or rejection
than the appropriate value from Table 3 must be suspected of testing of any group of materials without documentation that

TABLE 4 Precision Parameters for Test Method D1329, TR10, Retraction at Lower Temperature (Type 1 Precision), Size 2-214 O-rings
Units °C
Number of
Material Period Laboratories Mean Level Sr r (r) SR R (R)
(M/V)A
FVMQ; MIL- Oct ’19; Feb 4(0/1) -63.1 0.29 0.82 -1.3 2.10 5.95 -9.4
DTL-25988C; ’20
T1, C3, G75
FVMQ; MIL- Sep 2020 2(0/0) -60.5 0.38 1.08 -1.8 1.40 3.96 -6.6
DTL-25988C;
T1, C1, G60
NBR; AMS-P- Sep-Dec 2018 6(1/1) -45.0 0.16 0.46 -1.0 0.93 2.64 -5.9
5315B
FKM; AMS7287 May-Jun 2019 6(1/1) -29.9 0.19 0.54 -1.8 0.31 0.88 -2.9
FKM; April 2021 2(0/0) -19.2 1.70 4.82 -25.1 2.07 5.85 -30.5
AMS7259E
Average -43.5
Pooled Values 0.80 2.26 -5.2 1.52 4.31 -9.9
A
M=Mean outliers; V=Variability outliers

7
D1414 − 22
they are applicable to those particular materials, test tory of the test method procedure or apparatus for faulty operation or the
conditions, and the specific testing protocols of this test declaration of a significant difference in the two materials, samples,
laboratories, etc., which generated the two test results. Technical/
method. commercial actions are mutually resolved by the laboratories.
16.3 A type 1 inter-laboratory precision program for low 16.6 Bias—Bias is the difference between a test value and a
temperature test (TR10) was conducted during the periods reference value. However, for this test method, bias has not
listed with the number of laboratories listed with the o-ring been determined.
materials listed in Table 4. In each laboratory, two single tests
were performed on the o-rings by 2 different technicians on 2 17. Relative Density Test
different days for a total of four replicates. A test result is the 17.1 Test Specimen— Any O-ring, O-rings, or segment of an
value obtained from a single determination. Both repeatability O-ring may be considered as the specimen. The minimum mass
and reproducibility represent short-term (daily) testing condi- of the specimen shall be 0.005 N (0.5 gf). In the case of very
tions. Acceptable difference values were not measured. small O-rings, two or more pieces may be used as the specimen
16.4 Low Temperature (TR10) Repeatability—The absolute to accomplish this total minimum mass.
repeatability limits, r, of this test method are as shown in Table 17.2 Procedure—Employ the procedure outlined in Test
3 for each of the materials and conditions listed in the table. Methods D297, density hydrostatic method.
The difference between two single test results (or determina-
tions) found on identical test material under the repeatability NOTE 13—Extreme accuracy is required in making relative density
conditions prescribed for this test will exceed the repeatability determinations on small O-rings 0.001 N (1 g or less). The surface tension
on the suspending wire is critical; this wire should be of the smallest
value on an average of not more than once in 20 determinations diameter corrosion-resistant wire that will support the specimen weight.
in the normal and correct operation of the method. Two single This surface tension on the wire and the elimination of small air bubbles
test results that differ by more than the appropriate value from on the surface of the specimen can be reduced by the addition of a few
Table 4 must be suspected of being from different populations drops of wetting agent to the water in the weighing beaker.
and some appropriate action taken.
NOTE 11—Appropriate action may be an investigation of the test
18. Relative Density Test Precision and Bias
method procedure or apparatus for faulty operation or the declaration of a 18.1 These precision statements have been prepared in
significant difference in the two materials, samples, etc., which generated accordance with Practice D4483. Refer to this practice for
the two test results.
terminology and other statistical details.
16.5 Low Temperature (TR10) Reproducibility—The abso-
18.2 Precision—The precision results in this precision and
lute reproducibility limits, R, of this test method are as shown
bias section give an estimate of the precision of these test
in Table 4 for each of the materials and conditions listed in the
methods with the materials (rubbers, carbon blacks, etc.) used
table. The difference between two single and independent test
in the interlaboratory program described in Table 5. The
results found by two operators working under the prescribed
precision parameters should not be used for acceptance or
reproducibility conditions in different laboratories on identical
rejection testing of any group of materials without documen-
test material will exceed the reproducibility value on an
tation that they are applicable to those particular materials and
average of not more than once in 20 determinations in the
the specific testing protocols of these test methods.
normal and correct operation of the method. Two single test
results produced in different laboratories that differ by more 18.3 A type 1 inter-laboratory precision program for tensile
than the appropriate value from Table 4 must be suspected of strength was conducted during the periods listed with the
being from different populations and some appropriate inves- number of laboratories listed and the o-ring materials listed in
tigative or technical/commercial action taken. Table 5. In each laboratory, two single tests were performed on
NOTE 12—Appropriate action may be an investigation in each labora- the o-rings by 2 different technicians on 2 different days for a

TABLE 5 Precision Parameters for Test Methods D297, Density (Type 1 Precision), Size 2-214 O-rings
Units Mg/cm3
Number of
Material Period Laboratories Mean Level Sr r (r) SR R (R)
(M/V)A
NBR; AMS-P- Sep-Dec 2018 10(0/1) 1.2665 0.00268 0.00758 0.5983 0.00584 0.01652 1.3040
5315B
FVMQ; MIL- Sep 2020 5(1/0) 1.4783 0.00362 0.01024 0.6925 0.00362 0.01024 0.6925
DTL-25988C;
T1, C1, G60
FVMQ; MIL- Oct ’19; Feb 8(0/0) 1.5029 0.00346 0.00979 0.6512 0.00604 0.01709 1.1372
DTL-25988C; ’20
T1, C3, G75
FKM; AMS7287 May-Jun 2019 9(1/1) 1.7654 0.00524 0.01483 0.8402 0.00941 0.02663 1.5083
FKM; April 2021 3(0/0) 1.8406 0.00483 0.01365 0.7419 0.00483 0.01365 0.7419
AMS7259E
Average 1.5707
Pooled Values 0.00407 0.01153 0.7339 0.00625 0.01769 1.1264
A
M=Mean outliers; V=Variability outliers

8
D1414 − 22
total of four replicates. A test result is the value obtained from insert as described, the O-ring shall be cut into segments. The
a single determination. Both repeatability and reproducibility specimen shall be a segment 90 6 2 mm (3.5 6 0.08 in.) in
represent short-term (daily) testing conditions. Acceptable length. All tests for hardness, tension, elongation, and volume
difference values were not measured. change shall be made in triplicate. The same three specimens
18.4 Relative Density Repeatability—The relative repeat- may be used for all tests with hardness and volume determi-
ability limits, (r), of this test method are as shown in Table 5 for nations made prior to stress-strain tests. In the case of liquids
each of the materials listed in the table. The difference between that are volatile at room temperature, the specimen must be
two single test results (or determinations) found on identical quickly placed back into the test liquid after hardness and
test material under the repeatability conditions prescribed for volume test and prior to tension testing.
this test will exceed the repeatability value on an average of not 19.4 Procedure:
more than once in 20 determinations in the normal and correct 19.4.1 Suspend three specimens at one level in the test tube
operation of the method. Two single test results that differ by by the use of corrosion-resistant wire; separate the specimens
more than the appropriate value from Table 5 must be by bending small loops in the wire or by the use of corrosion-
suspected of being from different populations and some appro- resistant washers. For whole O-rings 6 mm (0.25 in.) in cross
priate action taken. section and larger, place one specimen in each test tube. For
NOTE 14—Appropriate action may be an investigation of the test cross-section sizes less than 4 mm (0.16 in.) place three
method procedure or apparatus for faulty operation or the declaration of a specimens in each test tube.
significant difference in the two materials, samples, etc., which generated
19.4.2 Suspend the specimens vertically so that 25 mm (1.0
the two test results.
in.) of liquid is between the lower extremity of the specimen
18.5 Relative Density Reproducibility—The relative repro- and the bottom of the tube. Add enough liquid to cover the
ducibility limits, (R), of this test method are as shown in Table specimens to a depth of 25 mm (1.0 in.) over the upper
5 for each of the materials listed in the table. The difference extremity of the specimens.
between two single and independent test results found by two
operators working under the prescribed reproducibility condi- NOTE 16—For liquid mixtures, where the volume percent of the most
minor constituent is known, the liquid volume of that constituent shall be
tions in different laboratories on identical test material will 15 times the volume of the test piece.
exceed the reproducibility value on an average of not more For liquid mixtures where the volume percent of the most minor
than once in 20 determinations in the normal and correct constituent is not known, then that volume is assumed to be 1 %, and the
operation of the method. Two single test results produced in volume of the test liquid will be 1500 times the volume of the test pieces.
different laboratories that differ by more than the appropriate If the above procedure is used, operators should be aware that the test
tube specified in Test Method D471 may not be adequate and alternative
value from Table 5 must be suspected of being from different containers and heating system may be required.
populations and some appropriate investigative or technical/ NOTE 17—A problem that can occur with large whole specimens is
commercial action taken. cracking at the point of maximum bend at the upper or lower portion, or
NOTE 15—Appropriate action may be an investigation in each labora- both, of the specimen. When cracking is observed upon removal from the
tory of the test method procedure or apparatus for faulty operation or the oil, or when breaks occur at these points during tension tests, a retest shall
declaration of a significant difference in the two materials, samples, be made on segments cut from an O-ring.
laboratories, etc., which generated the two test results. Technical/ 19.4.3 Make volume change determinations on a minimum
commercial actions are mutually resolved by the laboratories.
specimen weight of 0.005 N (0.5 gf). This may be accom-
18.6 Bias—Bias is the difference between a test value and a plished by using two or more pieces for the specimen in the
reference value. However, for this test method, bias has not case of very small O-rings (see Note 16).
been determined.
20. Immersion Test Precision and Bias
19. Immersion Test
20.1 These precision statements have been prepared in
19.1 Nature of Test— For determination of the changes in accordance with Practice D4483. Refer to this practice for
physical properties after liquid immersion, the definitions, terminology and other statistical details.
standard test liquid descriptions, and general technique shall be
20.2 Precision—The precision results in this precision and
in accordance with Test Method D471. The detailed procedure
bias section give an estimate of the precision of these test
for O-rings, outlined in this section however, shall take
methods with the materials (rubbers, carbon blacks, etc.) used
precedence in case of conflict.
in the interlaboratory program described in Table 6A through
19.2 Apparatus: Table 6H. The precision parameters should not be used for
19.2.1 The apparatus specified in Test Method D471 for acceptance or rejection testing of any group of materials
volatile liquids shall be used. without documentation that they are applicable to those par-
19.2.2 Standard Heating Device shall be an aluminum block ticular materials and the specific testing protocols of these test
heater or oil bath heater. The temperature shall be maintained methods.
within 6 1°C (1.8°F). 20.3 A type 1 inter-laboratory precision program for volume
19.3 Test Specimens— The test specimen shall consist of an change in lubricants was conducted during the periods listed
entire O-ring when the outside diameter is small enough to with the number of laboratories listed with the o-ring materials
permit installation in a test tube with the fluid level adjusted in the test conditions and lubricants listed in Table 6A. In each
accordance with 19.4.2. For sizes too large in circumference to laboratory, two single tests were performed on the o-rings by 2

9
D1414 − 22
TABLE 6 A Precision Parameters for Test Method D471, Volume Change, Effect of Liquids (Type 1 Precision), Size 2-214 O-rings
Units Percent
Material Period Test Condi- Number of Mean Level Sr r (r) SR R (R)
tions Laboratories
(M/V)A
FVMQ; MIL- Oct ’19 AMS3021 at 4(1/1) 7.2 0.59 1.68 23.4 0.59 1.68 23.4
DTL-25988C; 175°C for
T1, C3, G75 70 h
FVMQ; MIL- Sep 2020 AMS3021 at 4(0/1) 14.0 0.56 1.58 11.3 3.77 10.68 76.6
DTL-25988C; 150°C for
T1, C1, G60 70 h
FKM; April 2021 AMS 3023 at 2(0/0) 14.0 1.41 4.00 28.6 1.41 4.00 28.6
AMS7259E 200°C for
70 h
FKM; May-Jun 2019 AMS3085 at 6(0/1) 16.8 0.59 1.67 10.0 1.69 4.78 28.4
AMS7287 200°C for
70 h
Average 13.0
Pooled Values 0.87 2.46 18.9 2.20 6.24 48.1
A
M=Mean outliers; V=Variability outliers

different technicians on 2 different days for a total of four from different populations and some appropriate investigative
replicates. A test result is the value obtained from a single or technical/commercial action taken.
determination. Both repeatability and reproducibility represent NOTE 19—Appropriate action may be an investigation in each labora-
short-term (daily) testing conditions. Acceptable difference tory of the test method procedure or apparatus for faulty operation or the
declaration of a significant difference in the two materials, samples,
values were not measured. laboratories, etc., which generated the two test results. Technical/
20.4 Relative versus absolute repeatability and reproducibil- commercial actions are mutually resolved by the laboratories.
ity limits. For volume change in lubricants, relative precision is 20.7 Bias—Bias is the difference between a test value and a
statistically preferred. However, since the measurement units reference value. However, for this test method, bias has not
are in percent and relative precision is a percent, the expression been determined.
of relative precision would be calculated as a percent of a 20.8 A type 1 inter-laboratory precision program for volume
percent. This can be difficult to apply when evaluating the change in fuels was conducted during the periods listed with
testing precision. Therefore, absolute precision is preferred for the number of laboratories listed with the o-ring materials the
practical applications. test conditions and fuels listed in Table 6B. In each laboratory,
20.5 Volume Change in Lubricants Repeatability—The ab- two single tests were performed on the o-rings by 2 different
solute repeatability limits, r, of this test method are as shown technicians on 2 different days for a total of four replicates. A
in Table 6A for each of the materials listed in the table. The test result is the value obtained from a single determination.
difference between two single test results (or determinations) Both repeatability and reproducibility represent short-term
found on identical test material under the repeatability condi- (daily) testing conditions. Acceptable difference values were
tions prescribed for this test will exceed the repeatability value not measured.
on an average of not more than once in 20 determinations in the 20.9 Relative versus absolute repeatability and reproducibil-
normal and correct operation of the method. Two single test ity limits. For volume change in fuels, relative precision is
results that differ by more than the appropriate value from statistically preferred. However, since the measurement units
Table 6A must be suspected of being from different populations are in percent and relative precision is a percent, the expression
and some appropriate action taken. of relative precision would be calculated as a percent of a
NOTE 18—Appropriate action may be an investigation of the test
method procedure or apparatus for faulty operation or the declaration of a
percent. This can be difficult to apply when evaluating the
significant difference in the two materials, samples, etc., which generated testing precision. Therefore, absolute precision is preferred for
the two test results. practical applications.
20.6 Volume Change in Lubricants Reproducibility—The 20.10 Volume Change in Fuels Repeatability—The absolute
absolute reproducibility limits, R, of this test method are as repeatability limits, r, of this test method are as shown in Table
shown in Table 6A for each of the materials listed in the table. 6B for each of the materials listed in the table. The difference
The difference between two single and independent test results between two single test results (or determinations) found on
found by two operators working under the prescribed repro- identical test material under the repeatability conditions pre-
ducibility conditions in different laboratories on identical test scribed for this test will exceed the repeatability value on an
material will exceed the reproducibility value on an average of average of not more than once in 20 determinations in the
not more than once in 20 determinations in the normal and normal and correct operation of the method. Two single test
correct operation of the method. Two single test results results that differ by more than the appropriate value from
produced in different laboratories that differ by more than the Table 6B must be suspected of being from different populations
appropriate value from Table 6A must be suspected of being and some appropriate action taken.

10
D1414 − 22
TABLE 6 B Precision Parameters for Test Method D471, Volume Change, Effect of Liquids (Type 1 Precision), Size 2-214 O-rings (con-
tinued)
Units Percent
Number of
Test Condi-
Material Period Laboratories Mean Level Sr r (r) SR R (R)
tions
(M/V)A
FKM; April 2021 Fuel B at 2(0/0) 1.0
AMS7259E 23°C for 70 h
NBR; AMS-P- Sep-Dec 2018 Fuel A at 8(1/1) 2.4 0.43 1.21 49.8 0.57 1.61 66.3
5315B 23°C for 72 h
FKM; May-Jun 2019 Fuel B at 6(0/0) 2.8 0.42 1.20 43.1 0.52 1.47 52.8
AMS7287 23°C for 70 h
FVMQ; MIL- Oct ’19 AMS2629 at 6(1/0) 4.3 0.43 1.21 27.9 0.59 1.66 38.4
DTL-25988C; 23.9°C for
T1, C3, G75 22 h
FVMQ; MIL- Sep 2020 AMS2629 at 6(1/1) 6.5 0.38 1.07 16.4 0.49 1.39 21.4
DTL-25988C; 23.9°C for
T1, C1, G60 22 h
NBR; AMS-P- Sep-Dec 2018 Fuel B at 9(1/1) 30.0 0.76 2.15 7.2 1.31 3.71 12.4
5315B 23°C for 72 h
Average 7.8
Pooled Values 0.50 1.42 18.1 0.76 2.15 27.5
A
M=Mean outliers; V=Variability outliers

NOTE 20—Appropriate action may be an investigation of the test laboratories, etc., which generated the two test results. Technical/
method procedure or apparatus for faulty operation or the declaration of a commercial actions are mutually resolved by the laboratories.
significant difference in the two materials, samples, etc., which generated
the two test results. 20.12 Bias—Bias is the difference between a test value and
a reference value. However, for this test method, bias has not
20.11 Volume Change in Fuels Reproducibility—The abso-
been determined.
lute reproducibility limits, R, of this test method are as shown
in Table 6B for each of the materials listed in the table. The 20.13 A type 1 inter-laboratory precision program for hard-
difference between two single and independent test results ness change in lubricants was conducted during the periods
found by two operators working under the prescribed repro- listed with the number of laboratories listed with the o-ring
ducibility conditions in different laboratories on identical test materials the test conditions and lubricants listed in Table 6C.
material will exceed the reproducibility value on an average of In each laboratory, two single tests were performed on the
not more than once in 20 determinations in the normal and o-rings by 2 different technicians on 2 different days for a total
correct operation of the method. Two single test results of four replicates. A test result is the value obtained from a
produced in different laboratories that differ by more than the single determination. Both repeatability and reproducibility
appropriate value from Table 6B must be suspected of being represent short-term (daily) testing conditions. Acceptable
from different populations and some appropriate investigative difference values were not measured.
or technical/commercial action taken.
NOTE 21—Appropriate action may be an investigation in each labora- 20.14 Hardness Change in Lubricants Repeatability—The
tory of the test method procedure or apparatus for faulty operation or the absolute repeatability limits, r, of this test method are as shown
declaration of a significant difference in the two materials, samples, in Table 6C for each of the materials listed in the table. The

TABLE 6 C Precision Parameters for Test Method D471, Hardness Change, Effect of Liquids (Type 1 Precision), Size 2-214 O-rings
(continued)
Units Percent
Number of
Test Condi-
Material Period Laboratories Mean Level Sr r (r) SR R (R)
tions
(M/V)A
FKM; April 2021
AMS 3023 at 2(0/0) -7.6 2.07 5.86 -76.9 2.07 5.86 -76.9
AMS7259E 200°C for
70 h
FVMQ; MIL- Sep 2020 AMS3021 at 4(0/1) -7.6 0.90 2.54 -33.6 2.37 6.71 -88.9
DTL-25988C; 150°C for
T1, C1, G60 70 h
FVMQ; MIL- Oct ’19 AMS3021 at 4(0/0) -5.0 1.36 3.84 -77.4 2.05 5.79 -116.7
DTL-25988C; 175°C for
T1, C3, G75 70 h
FKM; May-Jun 2019 AMS3085 at 6(0/1) -4.4 1.82 5.14 -117.3 2.06 5.82 -132.9
AMS7287 200°C for
70 h
Average -6.1
Pooled Values 1.60 4.53 -73.8 2.14 6.06 -98.8
A
M=Mean outliers; V=Variability outliers

11
D1414 − 22
difference between two single test results (or determinations) short-term (daily) testing conditions. Acceptable difference
found on identical test material under the repeatability condi- values were not measured.
tions prescribed for this test will exceed the repeatability value
20.18 Hardness Change in Fuels Repeatability—The abso-
on an average of not more than once in 20 determinations in the
lute repeatability limits, r, of this test method are as shown in
normal and correct operation of the method. Two single test
Table 6D for each of the materials listed in the table. The
results that differ by more than the appropriate value from
difference between two single test results (or determinations)
Table 6C must be suspected of being from different populations
found on identical test material under the repeatability condi-
and some appropriate action taken.
tions prescribed for this test will exceed the repeatability value
NOTE 22—Appropriate action may be an investigation of the test
method procedure or apparatus for faulty operation or the declaration of a on an average of not more than once in 20 determinations in the
significant difference in the two materials, samples, etc., which generated normal and correct operation of the method. Two single test
the two test results. results that differ by more than the appropriate value from
20.15 Hardness Change in Lubricants Reproducibility— Table 6D must be suspected of being from different popula-
The absolute reproducibility limits, R, of this test method are tions and some appropriate action taken.
as shown in Table 6C for each of the materials listed in the NOTE 24—Appropriate action may be an investigation of the test
table. The difference between two single and independent test method procedure or apparatus for faulty operation or the declaration of a
significant difference in the two materials, samples, etc., which generated
results found by two operators working under the prescribed the two test results.
reproducibility conditions in different laboratories on identical
test material will exceed the reproducibility value on an 20.19 Hardness Change in Fuels Reproducibility—The ab-
average of not more than once in 20 determinations in the solute reproducibility limits, R, of this test method are as
normal and correct operation of the method. Two single test shown in Table 6D for each of the materials listed in the table.
results produced in different laboratories that differ by more The difference between two single and independent test results
than the appropriate value from Table 6C must be suspected of found by two operators working under the prescribed repro-
being from different populations and some appropriate inves- ducibility conditions in different laboratories on identical test
tigative or technical/commercial action taken. material will exceed the reproducibility value on an average of
NOTE 23—Appropriate action may be an investigation in each labora- not more than once in 20 determinations in the normal and
tory of the test method procedure or apparatus for faulty operation or the correct operation of the method. Two single test results
declaration of a significant difference in the two materials, samples, produced in different laboratories that differ by more than the
laboratories, etc., which generated the two test results. Technical/ appropriate value from Table 6D must be suspected of being
commercial actions are mutually resolved by the laboratories.
from different populations and some appropriate investigative
20.16 Bias—Bias is the difference between a test value and or technical/commercial action taken.
a reference value. However, for this test method, bias has not NOTE 25—Appropriate action may be an investigation in each labora-
been determined. tory of the test method procedure or apparatus for faulty operation or the
declaration of a significant difference in the two materials, samples,
20.17 A type 1 inter-laboratory precision program for hard- laboratories, etc., which generated the two test results. Technical/
ness change in fuels was conducted during the periods listed commercial actions are mutually resolved by the laboratories.
with the number of laboratories listed with the o-ring materials
20.20 Bias—Bias is the difference between a test value and
the test conditions and fuel listed in Table 6D. In each
a reference value. However, for this test method, bias has not
laboratory, two single tests were performed on the o-rings by 2
been determined.
different technicians on 2 different days for a total of four
replicates. A test result is the value obtained from a single 20.21 A type 1 inter-laboratory precision program for ten-
determination. Both repeatability and reproducibility represent sile strength change in lubricants was conducted during the

TABLE 6 D Precision Parameters for Test Method D471, Hardness Change, Effect of Liquids (Type 1 Precision), Size 2-214 O-rings
(continued)
Units Percent
Number of
Test Condi-
Material Period Laboratories Mean Level Sr r (r) SR R (R)
tions
(M/V)A
FVMQ; MIL- Oct ’19 AMS2629 at 5(0/1) -4.6 0.89 2.52 -55.4 1.50 4.24 -93.1
DTL-25988C; 23.9°C for
T1, C3, G75 22 h
FVMQ; MIL- Sep 2020 AMS2629 at 5(1/0) -3.1 0.70 1.99 -65.2 1.07 3.02 -98.9
DTL-25988C; 23.9°C for
T1, C1, G60 22 h
FKM; April 2021 Fuel B at 2(0/0) -0.5 1.29 3.65 -730.7 1.80 5.10 -1020.4
AMS7259E 23°C for 70 h
FKM; May-Jun 2019 Fuel B at 6(1/1) 0.0 1.06 3.00 -7503.1 1.06 3.01 -7520.5
AMS7287 23°C for 70 h
Average -2.0
Pooled Values 1.01 2.86 -140.3 1.39 3.94 -193.5
A
M=Mean outliers; V=Variability outliers

12
D1414 − 22
periods listed with the number of laboratories listed with the laboratories, etc., which generated the two test results. Technical/
o-ring materials the test conditions and lubricants listed in commercial actions are mutually resolved by the laboratories.
Table 6E. In each laboratory, two single tests were performed 20.24 Bias—Bias is the difference between a test value and
on the o-rings by 2 different technicians on 2 different days for a reference value. However, for this test method, bias has not
a total of four replicates. A test result is the value obtained from been determined.
a single determination. Both repeatability and reproducibility 20.25 A type 1 inter-laboratory precision program for ten-
represent short-term (daily) testing conditions. Acceptable sile strength change in fuels was conducted during the periods
difference values were not measured. listed with the number of laboratories listed with the o-ring
20.22 Tensile Strength Change in Lubricants materials the test conditions and lubricants listed in Table 6F.
Repeatability—The absolute repeatability limits, r, of this test In each laboratory, two single tests were performed on the
method are as shown in Table 6E for each of the materials o-rings by 2 different technicians on 2 different days for a total
listed in the table. The difference between two single test of four replicates. A test result is the value obtained from a
results (or determinations) found on identical test material single determination. Both repeatability and reproducibility
under the repeatability conditions prescribed for this test will represent short-term (daily) testing conditions. Acceptable
exceed the repeatability value on an average of not more than difference values were not measured.
once in 20 determinations in the normal and correct operation
20.26 Tensile Strength Change in Fuels Repeatability—The
of the method. Two single test results that differ by more than
absolute repeatability limits, r, of this test method are as shown
the appropriate value from Table 6E must be suspected of
in Table 6F for each of the materials listed in the table. The
being from different populations and some appropriate action
difference between two single test results (or determinations)
taken.
found on identical test material under the repeatability condi-
NOTE 26—Appropriate action may be an investigation of the test
method procedure or apparatus for faulty operation or the declaration of a tions prescribed for this test will exceed the repeatability value
significant difference in the two materials, samples, etc., which generated on an average of not more than once in 20 determinations in the
the two test results. normal and correct operation of the method. Two single test
20.23 Tensile Strength Change in Lubricants results that differ by more than the appropriate value from
Reproducibility—The absolute reproducibility limits, R, of this Table 6F must be suspected of being from different populations
test method are as shown in Table 6E for each of the materials and some appropriate action taken.
listed in the table. The difference between two single and NOTE 28—Appropriate action may be an investigation of the test
method procedure or apparatus for faulty operation or the declaration of a
independent test results found by two operators working under significant difference in the two materials, samples, etc., which generated
the prescribed reproducibility conditions in different laborato- the two test results.
ries on identical test material will exceed the reproducibility
20.27 Tensile Strength Change in Fuels Reproducibility—
value on an average of not more than once in 20 determinations
The absolute reproducibility limits, R, of this test method are
in the normal and correct operation of the method. Two single
as shown in Table 6F for each of the materials listed in the
test results produced in different laboratories that differ by
table. The difference between two single and independent test
more than the appropriate value from Table 6E must be
results found by two operators working under the prescribed
suspected of being from different populations and some appro-
reproducibility conditions in different laboratories on identical
priate investigative or technical/commercial action taken.
NOTE 27—Appropriate action may be an investigation in each labora- test material will exceed the reproducibility value on an
tory of the test method procedure or apparatus for faulty operation or the average of not more than once in 20 determinations in the
declaration of a significant difference in the two materials, samples, normal and correct operation of the method. Two single test

TABLE 6 E Precision Parameters for Test Method D471, Tensile Strength Change, Effect of Liquids (Type 1 Precision), Size 2-214
O-rings (continued)
Units Percent
Number of
Test Condi-
Material Period Laboratories Mean Level Sr r (r) SR R (R)
tions
(M/V)A
FVMQ; MIL- Oct ’19 AMS3021 at 4(0/0) -54.9 6.74 19.09 -34.8 6.74 19.09 -34.8
DTL-25988C; 175°C for
T1, C3, G75 70 h
FVMQ; MIL- Sep 2020 AMS3021 at 4(0/0) -40.2 17.04 48.21 -120.0 18.56 52.53 -130.7
DTL-25988C; 150°C for
T1, C1, G60 70 h
FKM; May-Jun 2019 AMS3085 at 6(0/0) -14.0 9.04 25.59 -183.4 9.91 28.04 -200.9
AMS7287 200°C for
70 h
FKM; April 2021 AMS 3023 at 2(0/0) -9.0 23.34 66.05 -733.9 34.76 98.38 -1093.1
AMS7259E 200°C for
70 h
Average -29.5
Pooled Values 15.51 43.89 -148.8 20.60 58.28 -197.5
A
M=Mean outliers; V=Variability outliers

13
D1414 − 22
TABLE 6 F Precision Parameters for Test Method D471, Tensile Strength Change, Effect of Liquids (Type 1 Precision), Size 2-214
O-rings (continued)
Units Percent
Number of
Test Condi-
Material Period Laboratories Mean Level Sr r (r) SR R (R)
tions
(M/V)A
FVMQ; MIL- Sep 2020 AMS2629 at 6(1/0) -14.5 17.09 48.37 -333.3 17.09 48.37 -333.3
DTL-25988C; 23.9°C for
T1, C1, G60 22 h
FVMQ; MIL- Oct ’19 AMS2629 at 6(0/1) -13.7 6.83 19.33 -140.8 8.50 24.05 -175.1
DTL-25988C; 23.9°C for
T1, C3, G75 22 h
FKM; May-Jun 2019 Fuel B at 6(1/0) -10.0 8.26 23.39 -235.1 8.26 23.39 -235.1
AMS7287 23°C for 70 h
FKM; April 2021 Fuel B at 2(0/0) -1.6 32.38 91.63 -5638.6 35.92 101.65 -6255.6
AMS7259E 23°C for 70 h
Average -10.0
Pooled Values 19.07 53.98 -542.3 20.75 58.73 -590.0
A
M=Mean outliers; V=Variability outliers

results produced in different laboratories that differ by more 20.30 Relative versus absolute repeatability and reproduc-
than the appropriate value from Table 6F must be suspected of ibility limits. For ultimate elongation change in lubricants,
being from different populations and some appropriate inves- relative precision is statistically preferred. However, since the
tigative or technical/commercial action taken. measurement units are in percent and relative precision is a
NOTE 29—Appropriate action may be an investigation in each labora- percent, the expression of relative precision would be calcu-
tory of the test method procedure or apparatus for faulty operation or the lated as a percent of a percent. This can be difficult to apply
declaration of a significant difference in the two materials, samples,
laboratories, etc., which generated the two test results. Technical/ when evaluating the testing precision. Therefore, absolute
commercial actions are mutually resolved by the laboratories. precision is preferred for practical applications.
20.28 Bias—Bias is the difference between a test value and 20.31 Ultimate Elongation Change in Lubricants
a reference value. However, for this test method, bias has not Repeatability—The absolute repeatability limits, r, of this test
been determined. method are as shown in Table 6G for each of the materials
20.29 A type 1 inter-laboratory precision program for ulti- listed in the table. The difference between two single test
mate elongation change in lubricants was conducted during the results (or determinations) found on identical test material
periods listed with the number of laboratories listed with the under the repeatability conditions prescribed for this test will
o-ring materials the test conditions and lubricants listed in exceed the repeatability value on an average of not more than
Table 6G. In each laboratory, two single tests were performed once in 20 determinations in the normal and correct operation
on the o-rings by 2 different technicians on 2 different days for of the method. Two single test results that differ by more than
a total of four replicates. A test result is the value obtained from the appropriate value from Table 6G must be suspected of
a single determination. Both repeatability and reproducibility being from different populations and some appropriate action
represent short-term (daily) testing conditions. Acceptable taken.
difference values were not measured. NOTE 30—Appropriate action may be an investigation of the test

TABLE 6 G Precision Parameters for Test Method D471, Elongation Change, Effect of Liquids (Type 1 Precision), Size 2-214 O-rings
(continued)
Units Percent
Number of
Test Condi-
Material Period Laboratories Mean Level Sr r (r) SR R (R)
tions
(M/V)A
FVMQ; MIL- Oct ’19 AMS3021 at 4(0/0) -46.1 8.49 24.03 -52.1 8.49 24.03 -52.1
DTL-25988C; 175°C for
T1, C3, G75 70 h
FVMQ; MIL- Sep 2020 AMS3021 at 4(0/0) -22.7 15.20 43.00 -189.3 16.96 48.01 -211.4
DTL-25988C; 150°C for
T1, C1, G60 70 h
FKM; April 2021 AMS 3023 at 2(0/0) -0.1 19.22 54.38 -43507.2 19.22 54.38 -43507.2
AMS7259E 200°C for
70 h
FKM; May-Jun 2019 AMS3085 at 6(1/0) 11.8 14.24 40.31 341.3 14.24 40.31 341.3
AMS7287 200°C for
70 h
Average -14.3
Pooled Values 14.79 41.86 -292.9 15.26 43.20 -302.3
A
M=Mean outliers; V=Variability outliers

14
D1414 − 22
method procedure or apparatus for faulty operation or the declaration of a in the normal and correct operation of the method. Two single
significant difference in the two materials, samples, etc., which generated test results that differ by more than the appropriate value from
the two test results.
Table 6H must be suspected of being from different popula-
20.32 Ultimate Elongation Change in Lubricants tions and some appropriate action taken.
Reproducibility—The absolute reproducibility limits, R, of this NOTE 32—Appropriate action may be an investigation of the test
test method are as shown in Table 6G for each of the materials method procedure or apparatus for faulty operation or the declaration of a
listed in the table. The difference between two single and significant difference in the two materials, samples, etc., which generated
the two test results.
independent test results found by two operators working under
the prescribed reproducibility conditions in different laborato- 20.36 Ultimate Elongation Change in Fuels
ries on identical test material will exceed the reproducibility Reproducibility—The absolute reproducibility limits, R, of this
value on an average of not more than once in 20 determinations test method are as shown in Table 6H for each of the materials
in the normal and correct operation of the method. Two single listed in the table. The difference between two single and
test results produced in different laboratories that differ by independent test results found by two operators working under
more than the appropriate value from Table 6G must be the prescribed reproducibility conditions in different laborato-
suspected of being from different populations and some appro- ries on identical test material will exceed the reproducibility
priate investigative or technical/commercial action taken. value on an average of not more than once in 20 determinations
NOTE 31—Appropriate action may be an investigation in each labora- in the normal and correct operation of the method. Two single
tory of the test method procedure or apparatus for faulty operation or the test results produced in different laboratories that differ by
declaration of a significant difference in the two materials, samples, more than the appropriate value from Table 6H must be
laboratories, etc., which generated the two test results. Technical/
commercial actions are mutually resolved by the laboratories.
suspected of being from different populations and some appro-
priate investigative or technical/commercial action taken.
20.33 Bias—Bias is the difference between a test value and NOTE 33—Appropriate action may be an investigation in each labora-
a reference value. However, for this test method, bias has not tory of the test method procedure or apparatus for faulty operation or the
been determined. declaration of a significant difference in the two materials, samples,
laboratories, etc., which generated the two test results. Technical/
20.34 A type 1 inter-laboratory precision program for ulti- commercial actions are mutually resolved by the laboratories.
mate elongation change in fuels was conducted during the
20.37 Bias—Bias is the difference between a test value and
periods listed with the number of laboratories listed with the
a reference value. However, for this test method, bias has not
o-ring materials the test conditions and lubricants listed in
been determined.
Table 6H. In each laboratory, two single tests were performed
on the o-rings by 2 different technicians on 2 different days for 21. Heat Aging Test
a total of four replicates. A test result is the value obtained from 21.1 Procedure—The general procedures for heat aging
a single determination. Both repeatability and reproducibility O-rings shall be in accordance with Test Method D573, with
represent short-term (daily) testing conditions. Acceptable the following exception: When required by the material
difference values were not measured. specification, or when agreed upon by the purchaser and the
20.35 Ultimate Elongation Change in Fuels Repeatability— seller, the procedure shall be in accordance with Test Method
The absolute repeatability limits, r, of this test method are as D865. The report shall state which method was employed.
shown in Table 6H for each of the materials listed in the table.
The difference between two single test results (or determina- 22. Heat Aging Test Precision and Bias
tions) found on identical test material under the repeatability 22.1 These precision statements have been prepared in
conditions prescribed for this test will exceed the repeatability accordance with Practice D4483. Refer to this practice for
value on an average of not more than once in 20 determinations terminology and other statistical details.

TABLE 6 H Precision Parameters for Test Method D471, Elongation Change, Effect of Liquids (Type 1 Precision), Size 2-214 O-rings
(continued)
Units Percent
Number of
Test Condi-
Material Period Laboratories Mean Level Sr r (r) SR R (R)
tions
(M/V)A
FVMQ; MIL- Oct ’19 AMS2629 at 6(0/0) -4.2 9.41 26.64 -640.3 11.27 31.90 -766.6
DTL-25988C; 23.9°C for
T1, C3, G75 22 h
FVMQ; MIL- Sep 2020 AMS2629 at 6(1/0) -2.8 15.20 43.03 -1515.1 15.20 43.03 -1515.1
DTL-25988C; 23.9°C for
T1, C1, G60 22 h
FKM; May-Jun 2019 Fuel B at 6(0/0) 4.1 10.70 30.29 741.7 10.70 30.29 741.7
AMS7287 23°C for 70 h
FKM; April 2021 Fuel B at 2(0/0) 7.3 15.01 42.47 585.8 15.01 42.47 585.8
AMS7259E 23°C for 70 h
Average 1.1
Pooled Values 12.84 36.34 3355.5 13.21 37.38 3451.8
A
M=Mean outliers; V=Variability outliers

15
D1414 − 22
22.2 Precision—The precision results in this precision and correct operation of the method. Two single test results
bias section give an estimate of the precision of these test produced in different laboratories that differ by more than the
methods with the materials (rubbers, carbon blacks, etc.) used appropriate value from Table 7A must be suspected of being
in the interlaboratory program described in Table 7A through from different populations and some appropriate investigative
Table 7D. The precision parameters should not be used for or technical/commercial action taken.
acceptance or rejection testing of any group of materials NOTE 35—Appropriate action may be an investigation in each labora-
without documentation that they are applicable to those par- tory of the test method procedure or apparatus for faulty operation or the
ticular materials and the specific testing protocols of these test declaration of a significant difference in the two materials, samples,
laboratories, etc., which generated the two test results. Technical/
methods. commercial actions are mutually resolved by the laboratories.
22.3 A type 1 inter-laboratory precision program for hard- 22.6 Bias—Bias is the difference between a test value and a
ness change in air was conducted during the periods listed with reference value. However, for this test method, bias has not
the number of laboratories listed with the o-ring materials and been determined.
the test conditions listed in Table 7A. In each laboratory, two
single tests were performed on the o-rings by 2 different 22.7 A type 1 inter-laboratory precision program for tensile
technicians on 2 different days for a total of four replicates. A strength change in air was conducted during the periods listed
test result is the value obtained from a single determination. with the number of laboratories listed with the o-ring materials
Both repeatability and reproducibility represent short-term and the test conditions listed in Table 7B. In each laboratory,
(daily) testing conditions. Acceptable difference values were two single tests were performed on the o-rings by 2 different
not measured. technicians on 2 different days for a total of four replicates. A
test result is the value obtained from a single determination.
22.4 Hardness Change in Air Repeatability—The absolute Both repeatability and reproducibility represent short-term
repeatability limits, r, of this test method are as shown in Table (daily) testing conditions. Acceptable difference values were
7A for each of the materials listed in the table. The difference not measured.
between two single test results (or determinations) found on
identical test material under the repeatability conditions pre- 22.8 Tensile Strength Change in Air Repeatability—The
scribed for this test will exceed the repeatability value on an absolute repeatability limits, r, of this test method are as shown
average of not more than once in 20 determinations in the in Table 7B for each of the materials listed in the table. The
normal and correct operation of the method. Two single test difference between two single test results (or determinations)
results that differ by more than the appropriate value from found on identical test material under the repeatability condi-
Table 7A must be suspected of being from different populations tions prescribed for this test will exceed the repeatability value
and some appropriate action taken. on an average of not more than once in 20 determinations in the
NOTE 34—Appropriate action may be an investigation of the test normal and correct operation of the method. Two single test
method procedure or apparatus for faulty operation or the declaration of a results that differ by more than the appropriate value from
significant difference in the two materials, samples, etc., which generated Table 7B must be suspected of being from different populations
the two test results.
and some appropriate action taken.
22.5 Hardness Change in Air Reproducibility—The abso- NOTE 36—Appropriate action may be an investigation of the test
lute reproducibility limits, R, of this test method are as shown method procedure or apparatus for faulty operation or the declaration of a
in Table 7A for each of the materials listed in the table. The significant difference in the two materials, samples, etc., which generated
difference between two single and independent test results the two test results.
found by two operators working under the prescribed repro- 22.9 Tensile Strength Change in Air Reproducibility—The
ducibility conditions in different laboratories on identical test absolute reproducibility limits, R, of this test method are as
material will exceed the reproducibility value on an average of shown in Table 7B for each of the materials listed in the table.
not more than once in 20 determinations in the normal and The difference between two single and independent test results

TABLE 7 A Precision Parameters for Test Method D573, Hardness Change, Deterioration in an Air Oven (Type 1 Precision), Size 2-214
O-rings
Units Percent
Number of
Test Condi-
Material Period Laboratories Mean Level Sr r (r) SR R (R)
tions
(M/V)A
FVMQ; MIL- Oct ’19 Air at 225°C 4(0/0) 0.8 0.63 1.78 224.3 1.50 4.24 534.0
DTL-25988C; for 70 h
T1, C3, G75
FVMQ; MIL- Sep 2020 Air at 200°C 4(0/0) 2.0 0.67 1.90 93.0 0.93 2.62 128.3
DTL-25988C; for 70 h
T1, C1, G60
FKM; April 2021 Air at 270°C 2(0/0) 3.6 1.86 5.26 145.2 1.86 5.26 145.2
AMS7259E for 70 h
Average 2.2
Pooled Values 1.20 3.39 157.4 1.48 4.19 194.3
A
M=Mean outliers; V=Variability outliers

16
D1414 − 22
TABLE 7 B Precision Parameters for Test Method D573, Tensile Strength Change, Deterioration in an Air Oven (Type 1 Precision), Size
2-214 O-rings (continued)
Units Percent
Number of
Test Condi-
Material Period Laboratories Mean Level Sr r (r) SR R (R)
tions
(M/V)A
FVMQ; MIL- Oct ’19 Air at 225°C 4(0/0) -21.2 7.61 21.53 -101.5 7.61 21.53 -101.5
DTL-25988C; for 70 h
T1, C3, G75
FKM; April 2021 Air at 270°C 2(0/0) -5.6 17.70 50.10 -902.7 25.83 73.10 -1317.1
AMS7259E for 70 h
FVMQ; MIL- Sep 2020 Air at 200°C 4(1/0) -2.4 11.62 32.88 -1360.8 11.62 32.88 -1360.8
DTL-25988C; for 70 h
T1, C1, G60
Average -9.7
Pooled Values 12.99 36.76 -377.9 16.93 47.92 -492.6
A
M=Mean outliers; V=Variability outliers

found by two operators working under the prescribed repro- 22.12 Relative versus absolute repeatability and reproduc-
ducibility conditions in different laboratories on identical test ibility limits. For ultimate elongation change in air, relative
material will exceed the reproducibility value on an average of precision is statistically preferred. However, since the measure-
not more than once in 20 determinations in the normal and ment units are in percent and relative precision is a percent, the
correct operation of the method. Two single test results expression of relative precision would be calculated as a
produced in different laboratories that differ by more than the percent of a percent. This can be difficult to apply when
appropriate value from Table 7B must be suspected of being evaluating the testing precision. Therefore, absolute precision
from different populations and some appropriate investigative is preferred for practical applications.
or technical/commercial action taken. 22.13 Ultimate Elongation Change in Air Repeatability—
NOTE 37—Appropriate action may be an investigation in each labora- The absolute repeatability limits, r, of this test method are as
tory of the test method procedure or apparatus for faulty operation or the
declaration of a significant difference in the two materials, samples, shown in Table 7C for each of the materials listed in the table.
laboratories, etc., which generated the two test results. Technical/ The difference between two single test results (or determina-
commercial actions are mutually resolved by the laboratories. tions) found on identical test material under the repeatability
22.10 Bias—Bias is the difference between a test value and conditions prescribed for this test will exceed the repeatability
a reference value. However, for this test method, bias has not value on an average of not more than once in 20 determinations
been determined. in the normal and correct operation of the method. Two single
test results that differ by more than the appropriate value from
22.11 A type 1 inter-laboratory precision program for ulti-
Table 7C must be suspected of being from different populations
mate elongation change in air was conducted during the
and some appropriate action taken.
periods listed with the number of laboratories listed with the NOTE 38—Appropriate action may be an investigation of the test
o-ring materials and the test conditions listed in Table 7C. In method procedure or apparatus for faulty operation or the declaration of a
each laboratory, two single tests were performed on the o-rings significant difference in the two materials, samples, etc., which generated
by 2 different technicians on 2 different days for a total of four the two test results.
replicates. A test result is the value obtained from a single 22.14 Ultimate Elongation Change in Air Reproducibility—
determination. Both repeatability and reproducibility represent The absolute reproducibility limits, R, of this test method are
short-term (daily) testing conditions. Acceptable difference as shown in Table 7C for each of the materials listed in the
values were not measured. table. The difference between two single and independent test

TABLE 7 C Precision Parameters for Test Method D573, Elongation Change, Deterioration in an Air Oven (Type 1 Precision), Size
2-214 O-rings (continued)
Units Percent
Number of
Test Condi-
Material Period Laboratories Mean Level Sr r (r) SR R (R)
tions
(M/V)A
FVMQ; MIL- Oct ’19 Air at 225°C 4(0/0) -16.9 7.46 21.11 -125.0 7.96 22.53 -133.4
DTL-25988C; for 70 h
T1, C3, G75
FVMQ; MIL- Sep 2020 Air at 200°C 4(0/0) -4.5 9.31 26.35 -587.3 9.31 26.35 -587.3
DTL-25988C; for 70 h
T1, C1, G60
FKM; April 2021 Air at 270°C 2(0/0) 0.0 11.24 31.82 127289.2 11.24 31.82 127289.2
AMS7259E for 70 h
Average -7.1
Pooled Values 9.47 26.79 -376.5 9.60 27.17 -381.9
A
M=Mean outliers; V=Variability outliers

17
D1414 − 22
results found by two operators working under the prescribed normal and correct operation of the method. Two single test
reproducibility conditions in different laboratories on identical results that differ by more than the appropriate value from
test material will exceed the reproducibility value on an Table 7D must be suspected of being from different popula-
average of not more than once in 20 determinations in the tions and some appropriate action taken.
normal and correct operation of the method. Two single test NOTE 40—Appropriate action may be an investigation of the test
results produced in different laboratories that differ by more method procedure or apparatus for faulty operation or the declaration of a
significant difference in the two materials, samples, etc., which generated
than the appropriate value from Table 7C must be suspected of the two test results.
being from different populations and some appropriate inves-
tigative or technical/commercial action taken. 22.19 Weight Loss in Air Reproducibility—The absolute
NOTE 39—Appropriate action may be an investigation in each labora- reproducibility limits, R, of this test method are as shown in
tory of the test method procedure or apparatus for faulty operation or the Table 7D for each of the materials listed in the table. The
declaration of a significant difference in the two materials, samples, difference between two single and independent test results
laboratories, etc., which generated the two test results. Technical/ found by two operators working under the prescribed repro-
commercial actions are mutually resolved by the laboratories.
ducibility conditions in different laboratories on identical test
22.15 Bias—Bias is the difference between a test value and material will exceed the reproducibility value on an average of
a reference value. However, for this test method, bias has not not more than once in 20 determinations in the normal and
been determined. correct operation of the method. Two single test results
22.16 A type 1 inter-laboratory precision program for produced in different laboratories that differ by more than the
weight loss in air was conducted during the periods listed with appropriate value from Table 7D must be suspected of being
the number of laboratories listed with the o-ring materials and from different populations and some appropriate investigative
the test conditions listed in Table 7D. In each laboratory, two or technical/commercial action taken.
single tests were performed on the o-rings by 2 different NOTE 41—Appropriate action may be an investigation in each labora-
tory of the test method procedure or apparatus for faulty operation or the
technicians on 2 different days for a total of four replicates. A declaration of a significant difference in the two materials, samples,
test result is the value obtained from a single determination. laboratories, etc., which generated the two test results. Technical/
Both repeatability and reproducibility represent short-term commercial actions are mutually resolved by the laboratories.
(daily) testing conditions. Acceptable difference values were 22.20 Bias—Bias is the difference between a test value and
not measured. a reference value. However, for this test method, bias has not
22.17 Relative versus absolute repeatability and reproduc- been determined.
ibility limits. For weight loss in air, relative precision is
statistically preferred. However, since the measurement units 23. Hardness Test
are in percent and relative precision is a percent, the expression 23.1 Nature of Test— It is difficult to obtain reliability in
of relative precision would be calculated as a percent of a testing hardness of O-rings due to the curved surface and
percent. This can be difficult to apply when evaluating the relatively small cross section thickness. With certain precau-
testing precision. Therefore, absolute precision is preferred for tions and proper technique the following test methods may be
practical applications. employed. It is recommended that for O-rings of cross-
22.18 Weight Loss in Air Repeatability—The absolute re- sectional thickness less than 6 mm (0.25 in.) the microtester
peatability limits, r, of this test method are as shown in Table specified in Test Method D1415 be employed.
7D for each of the materials listed in the table. The difference 23.2 Procedure:
between two single test results (or determinations) found on 23.2.1 Test Method D2240—This procedure was developed
identical test material under the repeatability conditions pre- to test rubber articles of 6-mm (0.25-in.) thickness minimum
scribed for this test will exceed the repeatability value on an and employs an indentor point 2.5 mm (0.1 in.) in length. Use
average of not more than once in 20 determinations in the caution (1) to center the specimen so the point of the indentor

TABLE 7 D Precision Parameters for Test Method D573, Weight Loss, Deterioration in an Air Oven (Type 1 Precision), Size 2-214
O-rings (continued)
Units Percent
Number of
Test Condi-
Material Period Laboratories Mean Level Sr r (r) SR R (R)
tions
(M/V)A
FVMQ; MIL- Oct ’19 Air at 225°C 4(0/1) 0.034 0.134 0.380 1125.3 0.551 1.560 4623.0
DTL-25988C; for 70 h
T1, C3, G75
FVMQ; MIL- Sep 2020 Air at 200°C 4(0/1) 0.073 0.029 0.082 111.7 0.449 1.271 1737.5
DTL-25988C; for 70 h
T1, C1, G60
FKM; April 2021 Air at 270°C 2(0/0) 0.275 0.206 0.583 212.2 5.236 14.82 5387.9
AMS7259E for 70 h
Average 0.127
Pooled Values 0.143 0.405 317.9 3.051 8.63 6782.1
A
M=Mean outliers; V=Variability outliers

18
D1414 − 22
reads on the center or thickest portion, and (2) to apply just NOTE 43—Values obtained with the micro hardness tester may differ
enough pressure to bring the base of the instrument in close from durometer values due to the difference in methods. The surface effect
from aging or curing is more pronounced with the micro tester.
contact with the specimen without compressing the specimen.
A holding fixture may be employed which will provide (1) 24. Hardness Test Precision and Bias
limitation of travel of the durometer head and (2) a self- 24.1 These precision statements have been prepared in
centering feature to eliminate alignment difficulties. A fixture accordance with Practice D4483. Refer to this practice for
that accomplishes these requirements is shown in Fig. 2. Any terminology and other statistical details.
fixture or design principle that will accomplish the stipulations
of (1) and (2) above may be employed. Do not make hardness 24.2 Precision—The precision results in this precision and
determinations on the parting or “flash” line. bias section give an estimate of the precision of these test
methods with the materials (rubbers, carbon blacks, etc.) used
NOTE 42—The durometer as described in Test Method D2240 is not in the interlaboratory program described in Table 8A through
recommended for testing O-rings with cross section of less than 6 mm
Table 8C. The precision parameters should not be used for
(0.25 in.). If hardness measurements are made with this instrument on
O-rings of smaller cross section, the values obtained may be different from acceptance or rejection testing of any group of materials
those obtained on a standard specimen. Fixtures as shown in 23.2.2 should without documentation that they are applicable to those par-
be employed for such determinations. For O-rings of less than 3-mm ticular materials and the specific testing protocols of these test
(0.12-in.) cross section, the fixture should be designed, double depth, to methods.
hold vertically two cross-section thicknesses of the specimen. This is to
increase the total thickness to greater than the length of the indentor point. 24.3 A type 1 inter-laboratory precision program for durom-
The total depth of the groove must be adjusted to a depth that will reduce eter hardness type A was conducted during the periods listed
within 62 the hardness of a standard specimen of the same material. with the number of laboratories listed and with the o-ring
23.2.2 Test Method D1415—This test method, using a micro materials listed in Table 8A. In each laboratory, two single tests
tester (see 3.3 of Test Method D1415) permits the testing of were performed on the o-rings by 2 different technicians on 2
very small specimens to 1-mm (0.04-in.) minimum thickness. different days for a total of four replicates. A test result is the
This test method is recommended for determining the hardness value obtained from a single determination. Both repeatability
of O-rings with a cross-sectional thickness of 6 mm (0.25 in.) and reproducibility represent short-term (daily) testing condi-
and smaller. tions. Acceptable difference values were not measured.

FIG. 2 Holding Fixture for Durometer in Hardness Test

19
D1414 − 22
TABLE 8 A Precision Parameters for Test Method D2240, Type A Durometer Hardness (Type 1 Precision), Size 2-214 O-rings
Durometer
Units
Units
Number of
Material/ Speci-
Period Laboratories Mean Level Sr r (r) SR R (R)
fication
(M/V)A
FVMQ; MIL- Sep 2020 4(0/0) 63.8 0.45 1.29 2.0 4.06 11.49 18.0
DTL-25988C;
T1, C1, G60
NBR; AMS-P- Sep-Dec 2018 9(0/1) 66.6 0.42 1.20 1.8 2.69 7.61 11.4
5315B
FKM; AMS7287 May-Jun 2019 9(0/1) 72.6 0.55 1.56 2.1 2.07 5.87 8.1
FVMQ; MIL- Oct ’19; Feb 7(0/0) 78.8 0.43 1.23 1.6 2.24 6.33 8.0
DTL-25988C; ’20
T1, C3, G75
FKM; April 2021 3(0/0) 86.9 1.44 4.09 4.7 1.44 4.09 4.7
AMS7259E
Average 73.7
Pooled Values 0.77 2.18 3.0 2.65 7.50 10.2
A
M=Mean outliers; V=Variability outliers

24.4 Durometer Hardness Type A Repeatability—The abso- not more than once in 20 determinations in the normal and
lute repeatability limits, r, of this test method are as shown in correct operation of the method. Two single test results
Table 8A for each of the materials listed in the table. The produced in different laboratories that differ by more than the
difference between two single test results (or determinations) appropriate value from Table 8A must be suspected of being
found on identical test material under the repeatability condi- from different populations and some appropriate investigative
tions prescribed for this test will exceed the repeatability value or technical/commercial action taken.
on an average of not more than once in 20 determinations in the NOTE 45—Appropriate action may be an investigation in each labora-
normal and correct operation of the method. Two single test tory of the test method procedure or apparatus for faulty operation or the
results that differ by more than the appropriate value from declaration of a significant difference in the two materials, samples,
Table 8A must be suspected of being from different populations laboratories, etc., which generated the two test results. Technical/
commercial actions are mutually resolved by the laboratories.
and some appropriate action taken.
NOTE 44—Appropriate action may be an investigation of the test 24.6 Bias—Bias is the difference between a test value and a
method procedure or apparatus for faulty operation or the declaration of a reference value. However, for this test method, bias has not
significant difference in the two materials, samples, etc., which generated
the two test results. been determined.
24.5 Durometer Hardness Type A Reproducibility—The ab- 24.7 A type 1 inter-laboratory precision program for durom-
solute reproducibility limits, R, of this test method are as eter hardness type M was conducted during the periods listed
shown in Table 8A for each of the materials listed in the table. with the number of laboratories listed and with the o-ring
The difference between two single and independent test results materials listed in Table 8B. In each laboratory, two single tests
found by two operators working under the prescribed repro- were performed on the o-rings by 2 different technicians on 2
ducibility conditions in different laboratories on identical test different days for a total of four replicates. A test result is the
material will exceed the reproducibility value on an average of value obtained from a single determination. Both repeatability

TABLE 8 B Precision Parameters for Test Method D2240, Type M Durometer Hardness (Type 1 Precision), Size 2-214 O-rings (contin-
ued)
Durometer
Units
Units
Number of
Material/ Speci-
Period Laboratories Mean Level Sr r (r) SR R (R)
fication
(M/V)A
FVMQ; MIL- Sep 2020 4(1/1) 70.6 0.43 1.22 1.7 1.51 4.27 6.0
DTL-25988C;
T1, C1, G60
NBR; AMS-P- Sep-Dec 2018 10(1/1) 74.0 0.41 1.16 1.6 4.58 12.95 17.5
5315B
FKM; AMS7287 May-Jun 2019 7(1/1) 78.9 0.33 0.93 1.2 4.12 11.67 14.8
FVMQ; MIL- Oct ’19; Feb 6(1/1) 81.0 0.52 1.48 1.8 3.62 10.23 12.6
DTL-25988C; ’20
T1, C3, G75
FKM; April 2021 2(0/0) 91.5 0.68 1.92 2.1 1.24 3.52 3.8
AMS7259E
Average 79.2
Pooled Values 0.49 1.38 1.7 3.31 9.37 11.8
A
M=Mean outliers; V=Variability outliers

20
D1414 − 22
and reproducibility represent short-term (daily) testing condi- materials listed in Table 8C. In each laboratory, two single tests
tions. Acceptable difference values were not measured. were performed on the o-rings by 2 different technicians on 2
24.8 Durometer Hardness Type M Repeatability—The ab- different days for a total of four replicates. A test result is the
solute repeatability limits, r, of this test method are as shown value obtained from a single determination. Both repeatability
in Table 8B for each of the materials listed in the table. The and reproducibility represent short-term (daily) testing condi-
difference between two single test results (or determinations) tions. Acceptable difference values were not measured.
found on identical test material under the repeatability condi- 24.12 Relative versus absolute repeatability and reproduc-
tions prescribed for this test will exceed the repeatability value ibility limits. For international hardness type M, relative
on an average of not more than once in 20 determinations in the precision is statistically preferred. However, since the measure-
normal and correct operation of the method. Two single test ment units are in percent and relative precision is a percent, the
results that differ by more than the appropriate value from expression of relative precision would be calculated as a
Table 8B must be suspected of being from different populations percent of a percent. This can be difficult to apply when
and some appropriate action taken. evaluating the testing precision. Therefore, absolute precision
NOTE 46—Appropriate action may be an investigation of the test is preferred for practical applications.
method procedure or apparatus for faulty operation or the declaration of a
significant difference in the two materials, samples, etc., which generated 24.13 International Hardness Type M Repeatability—The
the two test results. absolute repeatability limits, r, of this test method are as shown
24.9 Durometer Hardness Type M Reproducibility—The in Table 8C for each of the materials listed in the table. The
absolute reproducibility limits, R, of this test method are as difference between two single test results (or determinations)
shown in Table 8B for each of the materials listed in the table. found on identical test material under the repeatability condi-
The difference between two single and independent test results tions prescribed for this test will exceed the repeatability value
found by two operators working under the prescribed repro- on an average of not more than once in 20 determinations in the
ducibility conditions in different laboratories on identical test normal and correct operation of the method. Two single test
material will exceed the reproducibility value on an average of results that differ by more than the appropriate value from
not more than once in 20 determinations in the normal and Table 8C must be suspected of being from different populations
correct operation of the method. Two single test results and some appropriate action taken.
produced in different laboratories that differ by more than the NOTE 48—Appropriate action may be an investigation of the test
appropriate value from Table 8B must be suspected of being method procedure or apparatus for faulty operation or the declaration of a
significant difference in the two materials, samples, etc., which generated
from different populations and some appropriate investigative
the two test results.
or technical/commercial action taken.
NOTE 47—Appropriate action may be an investigation in each labora- 24.14 International Hardness Type M Reproducibility—The
tory of the test method procedure or apparatus for faulty operation or the absolute reproducibility limits, R, of this test method are as
declaration of a significant difference in the two materials, samples, shown in Table 8C for each of the materials listed in the table.
laboratories, etc., which generated the two test results. Technical/
commercial actions are mutually resolved by the laboratories.
The difference between two single and independent test results
found by two operators working under the prescribed repro-
24.10 Bias—Bias is the difference between a test value and ducibility conditions in different laboratories on identical test
a reference value. However, for this test method, bias has not material will exceed the reproducibility value on an average of
been determined. not more than once in 20 determinations in the normal and
24.11 A type 1 inter-laboratory precision program for inter- correct operation of the method. Two single test results
national hardness type M was conducted during the periods produced in different laboratories that differ by more than the
listed with the number of laboratories listed and with the o-ring appropriate value from Table 8C must be suspected of being

TABLE 8 C Precision Parameters for Test Method D1415, Type M International Hardness (Type 1 Precision), Size 2-214 O-rings (contin-
ued)
Units IRHD Units
Number of
Material/ Speci-
Period Laboratories Mean Level Sr r (r) SR R (R)
fication
(M/V)A
FVMQ; MIL- Sep 2020 4(0/0) 56.3 1.20 3.39 6.0 1.79 5.06 9.0
DTL-25988C;
T1, C1, G60
NBR; AMS-P- Sep-Dec 2018 9(1/1) 62.9 0.39 1.10 1.8 0.88 2.50 4.0
5315B
FKM; AMS7287 May-Jun 2019 7(1/0) 70.1 0.42 1.19 1.7 0.63 1.79 2.6
FVMQ; MIL- Oct ’19; Feb 5(0/1) 77.5 0.63 1.79 2.3 2.27 6.43 8.3
DTL-25988C; ’20
T1, C3, G75
FKM; April 2021 1(0/0) 87.7 0.44 1.25 1.4
AMS7259E
Average 70.9
Pooled Values 0.69 1.94 2.7 1.54 4.37 6.2
A
M=Mean outliers; V=Variability outliers

21
D1414 − 22
from different populations and some appropriate investigative from a tube may give slightly different results. If an alternative
or technical/commercial action taken. method is used it must be noted in the report.
NOTE 49—Appropriate action may be an investigation in each labora- 25.3.2 The preforms shall be placed in the shrinkage test
tory of the test method procedure or apparatus for faulty operation or the mold which is maintained at curing temperature 62°C
declaration of a significant difference in the two materials, samples, (63.6°F) in a suitable press. The mold shall be closed quickly
laboratories, etc., which generated the two test results. Technical/
commercial actions are mutually resolved by the laboratories. and brought to a pressure equal to a minimum of 3.4 MPa (500
psi) across the mold for the optimum curing time of the
24.15 Bias—Bias is the difference between a test value and material being tested. This curing time and temperature shall
a reference value. However, for this test method, bias has not be noted in the report. Curing time and especially temperature
been determined. are very critical in shrinkage consideration.
25.3.3 After curing is complete the specimens shall be
25. O-Ring Mold Shrinkage Test allowed to rest for a minimum of 16 h at a temperature of 23
25.1 Nature of Test— This test method involves the molding 6 2°C (73.4 6 3.6°F). Flash removal shall be done individu-
of various rubber materials in cavities of standard dimensions ally and carefully so as not to disturb the basic dimensions of
representative of typical O-ring configurations. The specimens the specimen. An abrasive paper or sand-stick of fine grade is
are then measured for inside diameter and cross-sectional usually employed for the operation. Excess flash on the inside
thickness within 0.025 mm (0.001 in.) to determine actual diameter must be avoided since this will alter the inside
mold shrinkage as a means of process control. diameter measurement.
25.3.4 Specimen rings obtained must be free of all major
25.2 Apparatus:
defects, such as nonfills, backrind, etc., as these factors may
25.2.1 Shrinkage Mold—Most critical is the shrinkage affect true shrinkage determination. Care must be exercised to
mold. Great care must be taken in both the manufacture and ensure proper closing of the mold to prevent erroneous
maintenance of this mold. Care must be taken not to dent, cross-section shrinkage results.
scratch, or otherwise damage the cavity. The mold cavities
must be kept clean. Harsh mechanical or chemical methods 25.4 Procedure:
that would dull the “cut off” or increase the cavity size must be 25.4.1 Measure the cross section by the use of tubing
avoided. The surface finish must be the best commercial micrometers or other method as described in 7.1. Determine
practice. actual shrinking by subtracting the average or mean cross
25.2.2 Calipers and Micrometers—Cavity measurement section of the specimen from the actual mold cavity size.
shall be made by the use of vernier calipers and depth Measurement of the cross section in inside diameter, outside
micrometers as follows: diameter, and side-to-side axes is desirable. This provides a
25.2.2.1 Align the edges of the vernier with the inside check on the accuracy of the mold cavity radius and true
diameter of the cavity “with magnification.” Take a reading roundness of cross section. Some materials, such as silicones,
that is the outside diameter of the cavity. may create an imbalance on the two cross-section axes due to
inherent shrinkage characteristics, thus producing an out-of-
25.2.2.2 Open the calipers to align the outside diameter of
round cross section.
the cavity. This gives the width of the cross section.
25.4.2 Measure inside diameter by the use of a suitable step
25.2.2.3 Use depth micrometers to check the depth of the gage as shown in Fig. 1, or a suitable taper gage (see 7.2). One
cavity. Check each half for depth. The total of the two halves or more gages may be necessary to cover a wide range of
gives the total depth of the cavity. shrinkage. Take care not to force the specimen on the step
25.2.2.4 Dimensions for the cavity of a recommended test gage. Use only enough pressure to be certain the inside
size are 3.6-mm (0.142-in.) cross section by 25-mm (1.0-in.) diameter is in intimate contact with the gage. Determine actual
inside diameter. linear shrinkage of the inside diameter by subtracting the
NOTE 50—Other test sizes varying the cross section and inside diameter average or mean inside diameter value of the specimens from
may be preferred and are permitted. This size is recommended merely as the actual mold cavity inside diameter. The resultant value
the median standard cross section in a diameter easily molded and divided by the mold inside diameter and multiplied by 100 is
measured. Some laboratories may prefer a larger size, the important the percent shrinkage. An alternative method of expression is
consideration being that the seller and the manufacturer use the same or
comparable size for control work. It has been found important to use a test in units such as millimetres per millimetre or inches per inch.
mold cross-section size as close as possible to the cross-section size of the 25.4.3 An alternative method of measuring the inside diam-
production mold being considered. eter and cross section is the shadowgraph or comparator. This
25.3 Test Specimen: instrument is very accurate, but if the O-ring is unsupported the
25.3.1 Preparation of preforms is by the die-cut method diameter must be taken in two directions and averaged. To
unless otherwise specified. Sample stock shall be prepared by prevent any out-of-roundness, a special cone can be used in
sheeting on a standard laboratory mill to a thickness of conjunction with the comparator. The comparator is also used
approximately 20 to 25 % greater than the cavity depth. to determine roundness of cross section by measuring thin
Milling time should be the minimum required to obtain a slices of the cross-sectional area.
smooth sheet. The sheet shall be allowed to rest for 30 to 60 25.5 Report—The report shall include the following:
min after which preforms shall be cut out with a suitable die. 25.5.1 Method of preforming.
Other types of preforming, such as a spliced extrusion or slice 25.5.2 Number of specimens measured, and

22
D1414 − 22
25.5.3 Average shrinkage, expressed as a percentage. 26.4.2 Remove the cover and place the individual speci-
mens and test panels into the open chamber for 1 h of
26. Corrosion Test preconditioning.
26.1 This test is intended to give a fast and relatively 26.4.3 After this 1 h preconditioning, assemble the speci-
reproducible rating of the corrosion potential for an O-ring. mens with panels as shown in Fig. 3. Apply sufficient pressure
to cause a deflection of 15 % of the original cross-sectional
26.2 Apparatus:
diameter of the specimens.
26.2.1 The apparatus shall consist of any chamber that is
26.4.4 Place the assembled fixture on a watchglass and
corrosion-proof, can be sealed, and is of convenient size for the
insert it into the test chamber.
work load.
26.4.5 Cover the chamber and allow it to stand for 96 h at
NOTE 51—A laboratory desiccator jar is commonly used. 23 6 2°C (73.4 6 3.6°F).
26.2.2 The test panels unless otherwise specified shall be 26.4.6 Remove the specimens, wash the panels with toluene
made of SAE-1020 steel. The panels shall be made in accor- or other suitable solvent system, and wipe with a lint-free cloth.
dance with the drawings in Fig. 3. If superficial rust or adhering particles from the specimen are
26.2.3 Relative humidity of 100 % is maintained by the use still present, a light buffing with crocus cloth, steel, or brass
of a solution of 85 % distilled water and 15 % USP glycerin. wool is permitted.
This solution should be added to a depth equal to at least 20 % 26.5 Compare the results to the examples shown in Fig. 4.
of the depth of the chamber. The rating shall be nil, moderate, or severe.
26.2.4 A corrosion-proof shelf shall be employed to hold the
27. Precision and Bias
fixtures close above but not touching the solution.
27.1 The precision and bias tables and statements for each
26.3 Test Specimen— The test specimens shall consist of
test method are given in a precision section following the
two segments 38 mm (1.5 in.) in length, cut from the sample
procedural section for each test method. See the section for the
O-ring. If the sample O-ring is smaller than 38 mm in
test method of interest for the applicable precision and bias
circumference, two whole O-rings may be used.
information.
26.4 Procedure:
26.4.1 Clean the panels with toluene or other solvent system 28. Keywords
that will produce a clean dry surface free from film. 28.1 aging tests; o-rings; physical properties

FIG. 3 Corrosion Test Panels

23
D1414 − 22

FIG. 4 Examples of Corrosion Test Results

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

24

You might also like