GEMINI GXL

Installation Technical Manual

4535 671 38811


Revision B

This document and the information contained in it is proprietary and confidential information of Philips Medical Systems (“Philips”) and may not be
reproduced, copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips
Legal Department. Use of this document and the information contained in it is strictly reserved for current Philips personnel and Philips customers
who have a current and valid license from Philips for use by the customer’s designated in-house service employee on equipment located at the
customer’s designated site. Use of this document by unauthorized persons is strictly prohibited. Report violation of these requirements to the Philips
Legal Department at 22100 Bothell Everett Highway, Bothell, WA 98021. This document must be returned to Philips when the user is no longer
licensed and in any event upon Philips’ first written request.
Philips Medical Systems © 2015
Philips Medical Systems
2015 Koninklijke Philips Electronics N. V. All Rights Reserved.

Warranty Disclaimer
PHILIPS PROVIDES THIS DOCUMENT WITHOUT WARRANTY OF ANY KIND, EITHER IMPLIED OR EXPRESSED, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

Limitation of Liability
Philips has taken care to ensure the accuracy of this document. However, Philips assumes no liability for errors or omissions and reserves the right to make changes without
further notice to any products herein to improve reliability, function, or design. Philips may make improvements or changes in the product(s) or program(s) described in this
document at any time.

4535 671 38811 Rev.B Philips Medical Systems Confidential and Proprietary Information 2
Symbol descriptions

Attention symbol. Radiation warning symbol.

Laser warning symbol. Biohazard warning symbol.

Magnetism warning symbol. Projectile warning symbol.

Electrical warning symbol.

Revision History

Revision Date Comments

A May 2005 Initial Release. ECO 2010903.

B June 2015 E-059201 Add Fixed Mobile Site

4535 671 38811 Rev.B Philips Medical Systems Confidential and Proprietary Information 3
Safety Information
To the User of This Manual
The user of this manual is directed to read and carefully review the instructions, warnings and cautions contained herein prior to beginning installation or service
activities. While you may have previously installed or serviced equipment similar to that described in this manual, changes in design, manufacture or procedure may have
occurred which significantly affect the present installation or service.

Installation and Environment


Except for installations requiring certification by the manufacturer per federal standards, see that a radiation protection survey is made by a qualified expert in
accordance with NCRP 102, section 7, as revised or replaced in the future. Perform a survey after every change in equipment, workload, or operating conditions which
might significantly increase the probability of persons receiving more than the maximum permissible dose equivalent.

Diagnostic Imaging Systems - Mechanical-electrical Warning


All of the moveable assemblies and parts of this equipment should be operated with care and routinely inspected in accordance with the manufacturer’s
recommendations contained in the equipment manuals.
Only properly trained and qualified personnel should be permitted access to any internal parts. Live electrical terminals are deadly; be sure line disconnects are opened
and other appropriate precautions are taken before opening access doors, removing enclosure panels, or attaching accessories.
Do not under any circumstances, remove the flexible high tension cables from the x-ray tube housing or high tension generator and/ or the access covers from the
generator until the main and auxiliary power supplies have been disconnected. Failure to comply with the above may result in serious or fatal bodily injuries to the
operator or those in the area.

Electrical-grounding Instructions
The equipment must be grounded to an earth ground by a separate conductor. The neutral side of the line is not to be considered the earth ground. On equipment
provided with a line cord, the equipment must be connected to properly grounded, three-pin receptacle. Do not use a three-to-two pin adapter.

Diagnostic Imaging Systems - Radiation Warning


X-ray and Gamma-rays are dangerous to both operator and others in the vicinity unless established safe exposure procedures are strictly observed.
The useful and scattered beams can produce serious or fatal bodily injuries to any persons in the surrounding area if used by an unskilled operator. Adequate precautions
must always be taken to avoid exposure to the useful beam, as well as to leakage radiation from within the source housing or to scattered radiation resulting from the
passage of radiation through matter.
Those authorized to operate, participate in or supervise the operation of the equipment must be thoroughly familiar and comply completely with the current established
safe exposure factors and procedures described in publications, such as: Subchapter J of Title 21 of the Code of Federal Regulations, "Diagnostic X-ray Systems and
Their Major Components", and the national council on radiation protection (NCRP) no. 102, "Medical X-ray And Gamma-ray Protection For Energies Up To 10
Mevequipment Design and Use", as revised or replaced in the future.
Those responsible for planning of x-ray and gamma-ray equipment installations must be thoroughly familiar and comply completely with NCRP no. 49, “Structural
Shielding Design and Evaluation For Medical Of X-rays and Gamma-rays of E nergies Up to 10 Mev”, as revised and replaced in the future. Failure to observe these
warnings may cause serious or fatal bodily injuries to the operator or those in the area.
4535 671 38811 Rev.B Philips Medical Systems Confidential and Proprietary Information 4
Table of Contents
System Power Up/Down ........................................................................................................11
System Power Up.................................................................................................................................11
System Power Off ................................................................................................................................15
Restart Computers ...............................................................................................................................22

Introduction ..........................................................................................................................26

Equipment, Parts, and Tools.................................................................................................27

Pre-Installation Guide and Site Preparation .........................................................................29


General Information..............................................................................................................................29
Safety Information ...............................................................................................................................32
Radiation Safety...................................................................................................................................33
General Requirements ........................................................................................................................35
Fixed Site Requirements......................................................................................................................37
Mobile Requirements ...........................................................................................................................41
GEMINI System Power Quality Requirements.....................................................................................42
Power Closet Environmental Requirements.........................................................................................43
Gantry Power Conditioner and/or Gantry UPS..........................................................................43
Gantry Isolation Transformer.....................................................................................................44
Control Room Power Requirements.....................................................................................................45
Power Box Description.........................................................................................................................45

Unpacking and Transporting System Components ..............................................................49

Uninterruptable Power Supply Setup....................................................................................50

Floor Template Preparation ..................................................................................................51


Mobile Floor Template..........................................................................................................................51
Overview....................................................................................................................................51

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 5
Table of Contents
Fixed Site Floor Template
....................................................................................................................54 Overview
..............................................................................................................................................54 Floor
Template Layout .........................................................................................................................55 GSU
Floor Template Preparation.........................................................................................................58
Patient Table Floor Template Preparation ...........................................................................................60

Gantry Separation Unit Leveling ...........................................................................................62


Overview ..............................................................................................................................................62
Mobile Cabling......................................................................................................................................71

Patient Table Mechanical Alignment and Leveling ...............................................................72


Overview ..............................................................................................................................................72

Pouring the Micorox Compound-GSU ...................................................................................83


Overview ..............................................................................................................................................83
Pouring the Micorox Compound – GSU...............................................................................................84

Anchoring the GSU ...............................................................................................................90

PET Gantry Mounting ............................................................................................................96


Overview ..............................................................................................................................................96
Positioning the PET Gantry..................................................................................................................96

CT Gantry Mounting ............................................................................................................104


Positioning the CT Gantry
..................................................................................................................104

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 6
Table of Contents
Gantry Mechanical Alignment .............................................................................................106
Overview ............................................................................................................................................106
Testing/Calibrating the Laser and Mirror............................................................................................107
Calibrating the Laser ...............................................................................................................107
Calibrating the Mirror...............................................................................................................110
Installing the Laser and Mirror Assemblies ........................................................................................113
Verifying PET Gantry Alignment to Rails.................................................................................115
CT Gantry Parallelism (Relative to the PET Gantry) .........................................................................116
Center Axis Alignment.............................................................................................................118
Alternate CT Leveling Procedure ............................................................................................120
System Retest .........................................................................................................................121

Electrical Interconnect........................................................................................................122
Fixed Site Wall-Mounted Panel..........................................................................................................123
Mobile Wall-Mounted Panel ...............................................................................................................124
TEAL Power Unit................................................................................................................................125
GSU Electrical Connections...............................................................................................................130
Mobile GSU Electrical Connections ...................................................................................................137
CT Gantry Electrical Connections ......................................................................................................139
PET Gantry Electrical Connections....................................................................................................154
GEMINI Patient Table Electrical Connections....................................................................................160
Mobile Patient Table Connections......................................................................................................167
Mobile CT Gantry Connections
..........................................................................................................169
Mobile CT Cabling...................................................................................................................172
Mobile PET Gantry Floor Locks .........................................................................................................175

Inbore Lifter Installation .....................................................................................................180


Overview ............................................................................................................................................180

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 7
Table of Contents
Inbore Lifter Installation......................................................................................................................181
Inbore Lifter Electrical Connect ...............................................................................................182
Mobile Cabling....................................................................................................................................184
Control Room Set-Up ..........................................................................................................185
Computer Room Preparation .............................................................................................................188
PET View Workstation Set-up ................................................................................................195
CT Recon Electrical Connections............................................................................................201

Final Patient Table Alignment .............................................................................................206


Overview ............................................................................................................................................206

Pouring the Micorox Compound – Patient Table ................................................................207


Overview ............................................................................................................................................207

Anchoring the Patient Table ...............................................................................................212

Calibrations and Performance Verification .........................................................................219


Calibration Procedures.......................................................................................................................219

PET Performance Verification Procedures .........................................................................219

CT Laser Registration .........................................................................................................219

GEMINI System Networking ................................................................................................220

Diagnostic X-Ray System Assembly Report .......................................................................221

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 8
Table of Contents
Packaging and Shipping Returnables .................................................................................222
PET Gantry Roll-Lifts..........................................................................................................................222
CT Gantry Swivel Casters..................................................................................................................222
GSU
Transport ...................................................................................................................................222
Patient Table Transport......................................................................................................................223
Clean Up ............................................................................................................................................223
GEMINI Customer Acceptance and Image Alignment .........................................................223

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 9
GEMINI GXL Installation Technical Manual System Power Up/Down

System Power Up/Down


The Power Up/Down procedures are required after the system is installed. Due to the frequent need for these
procedures, this section appears in the front of this manual for your convenience.

CAUTION System Power Up and Power Down steps must but performed in the proper sequence. Failure to comply will
result in system failure.

System Power Up
Use this procedure to turn on the GEMINI System when all the power to the system was turned off. If you are only
restarting the computers, then refer to the procedure for Restart Computers. Before turning on the system, verify all
phantoms and phantom holders are removed from the Patient Table.
1 Turn on the main circuit breaker and wall power switch.
2 If the gantries to your system are equipped with the optional UPS system, turn on the gantry UPS system.
a Press the EPO button.
After turning on the main power, the UPS beeps and the Alarm indicator on the UPS control panel flashes.
b Wait approximately 15 seconds after turning on the power.
c Then press and hold the Inverter Start button until the UPS system displays the message “System
Normal”.
1

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 10
GEMINI GXL Installation Technical Manual System Power Up/Down

1. EPO 2. Inverter Start


3. Inverter Stop 4. Audible Alarm Cancel
Control Panel for Gantry UPS
3 Turn on the power to the PET gantry.
At the rear of the PET gantry, turn on the power switches in the following order:

Switch A B C

A - Main

B - Reset
Push button to reset
circuit breaker

C - Gantry

D - Computer F E D

E - MCS

F - HV
PET Gantry Power Distribution Panel
4 Turn on the power from the Uninterruptible Power Supply (UPS) to the computer workstations.
At the front panel of the UPS system, press and hold the ON/TEST button until you hear a beep.
5 Turn on the power to the Sun computer and display monitor for the PET Workstation.
6 a Open the front panel on the PET Workstation computer.
b Turn the power switch that is next to the On/Standby button to On.
c Press and hold the On/Standby button to start up the computer.
Allow approximately three minutes for the computer to start up.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 11
GEMINI GXL Installation Technical Manual System Power Up/Down

Off On
Position Position

Push to start

PET Reconstruction Computer


PET Reconstruction Computer Power Switches
7 Logon to the PET Workstation.

a Type patient as the user name.

b Type the password.

c Press Return.

8 At the CIRS Computer Cabinet, turn on the two upper CIRS Computers.
a At the top computer in the cabinet, press and hold the power button until the power indicator light turns on.
b At the middle computer, press and hold the power button until the indicator light turns on.
9 Wait three minutes before proceeding.
NOTE It is imperative to wait the full three minutes before proceeding.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 12
GEMINI GXL Installation Technical Manual System Power Up/Down

10 Turn on the bottom CIRS computer.


Press and hold the power button until the indicator light turns on.

CAUTION Do not press the button at the bottom of the CIRS computer cabinet unless instructed to do so by a
Philips Service Specialist.
11 Turn on the power to the Dell computer for the Acquisition Workstation. a Press the power button on the Dell
Computer to turn on the power.
b If necessary, turn on the power to the Acquisition Workstation display monitors.
When the computer start up process begins, allow approximately 90 seconds for the process to complete.
The Log In window appears.
12 Start up the GEMINI operating software.
a At the Log In window, enter pet as the required username.
A password is not required.
b Press Enter.
Wait while the computer starts up.
The CT computer system connects to the PET computer system. Typically the system requires three to five
minutes for the computers to turn on.
If it appears the computers are not turning on or establishing connections, allow ten minutes for the start up
and connection to complete before initiating troubleshooting procedures.
The system displays an error message if the computers do not establish a connection. If an error occurs, refer to
Restart Computers.
13 When the Keyswitch Prompt is displayed on the Acquisition Workstation, turn the keyswitch on the side of the
Scan Control Panel to On and then release the key to the Standby position.
The Table Homing window appears.
NOTE Before starting the Table Homing procedure, verify that the CT tube is not spinning and the phantom holder is not
attached to the table.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 13
GEMINI GXL Installation Technical Manual System Power Up/Down

14 At the Table Homing window, click Start to home the Patient Table.
Wait while the system homes the Patient Table. When the procedure is complete, the Table Homing window
closes.
The Separation System Homing window appears.
15 At the Separation System Homing window, click Enable to start the Gantry Separation System homing
sequence.
The system prompts you to close the Gantry Separation System.
16 Go to the Gantry Separation System keypad and press the Close and Enable buttons and completely close the
gantries.
The Closed indicator light on the keypad turns on when the system is fully closed.

Gantry Separation System Closed Indicator


16 The system is ready for use and displays the appropriate Desktop on the workstation monitors.
Perform the required Daily Quality Control and Calibration procedures as established by your institution.
Do not attempt to perform procedures before performing Daily Quality Control and calibration. If you attempt to
perform procedures before completing quality control and calibration, the system may not function as expected.
For example, the laser lights on the PET gantry are not available until the System Initialization procedure is
performed.

System Power Off


It is recommended that you keep the power on the system at all times. Use the following procedure in those
situations where it is necessary to turn off the GEMINI System.
While it is not required that you restart the CT Computing System daily, it is recommended that you shut down the
system at least weekly, or more frequently if your site deems it necessary.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 14
GEMINI GXL Installation Technical Manual System Power Up/Down

CAUTION Unexpected power interruptions can damage the computers and the software. Do not turn off the electric power
to the computer until you have shut down the computers.
1 Close all applications and open windows on the workstations.
2 Shut down the PET acquisition computer.
a At the Acquisition Workstation, select Home and then select Utilities.
b Select PET Shutdown from the Misc menu displayed in the Task Information box.
3 At the prompt, type Yes and then press Enter.
Wait until the system displays a message to indicate PET shutdown is complete.
4 Shut down the PET Workstation computer.
a Click anywhere on the desktop of the PET Workstation to display the Main Menu and click Exit.
b At the Login window, click Options > Command Line Login.
c Press Return.
d At the Solaris Login box, type shut.
e Press Return.
f When prompted, type in the password.
g Press Return.
h Answer any additional prompts.
5 When the system displays message Program Terminated OK, turn off power to the PET Workstation
computer and monitor.
a Open the PET Workstation computer cabinet.
b Open the front panel on the PET Workstation computer.
c Press and hold the On/Standby button to shutdown the computer.
d Turn the power switch that is next to the On/Standby button to Off.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 15
GEMINI GXL Installation Technical Manual System Power Up/Down

Off On
Position Position

Push for
Standby

PET Reconstruction Computer


PET Reconstruction Computer Power Switches
6 Turn the keyswitch on the Scan Control Panel to the Off position to turn off the CT Gantry.

Off

The gantry turns off in approximately 1 minute.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 16
GEMINI GXL Installation Technical Manual System Power Up/Down

NOTE If the Keyswitch Prompt window displays, ignore the prompt to turn the keyswitch to On.
7 Turn off the power to the CIRS computers.
a Open the CIRS Computer Cabinet.
b For each computer in the CIRS cabinet, press the power button and hold it until the power indicator light
turns off, approximately 5 seconds.
It is not critical to turn off the computers in a particular sequence.

CAUTION Do not press the bottom button in the CIRS computer cabinet unless instructed to do so by a Philips Service
Specialist.

CIRS Computer Power Switches


8 Log off the Acquisition Workstation computer.
a Select Home on the Desktop, and then click Logout from the Scan Tool Box.
The system displays a prompt to confirm you want to logout.
b Click Yes to confirm logout.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 17
GEMINI GXL Installation Technical Manual System Power Up/Down

The Logon Information window opens.


9 Click Shutdown.
10 Select Shutdown again and then click OK.
Wait as the computers automatically shut down and turn off.
11 Turn off the Workstation Uninterruptible Power Supply (UPS) to the computers.
At the front panel of the UPS system, press and hold the Off button until you hear a beep.

1. Workstation UPS Off Switch

Workstation UPS Front Panel


The power from the UPS to the power outlets is off. The batteries continue to charge and the cooling fan
remains On until you turn off the Main Power Supply in Step 13.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 18
GEMINI GXL Installation Technical Manual System Power Up/Down

12 Turn off power to the PET Gantry.


Go to the rear of the PET gantry and turn off the power using the switches at the rear of the system. Turn off
the switches in the order specified.

Switch E D

A - HV

B - MCS

C - Computer

D - Gantry

A B C

E - Main

PET Gantry Power Distribution Panel


13 Turn off the power supply to the system
• If your system is not equipped with the optional UPS for the gantries, then go to the main circuit breaker box
and turn off the main circuit breaker for the gantry systems.
• If your system is equipped with the optional UPS for the gantries, use the UPS control panel to shut down
the UPS to the gantries.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 19
GEMINI GXL Installation Technical Manual System Power Up/Down

1. Chloride UPS Control Panel

Chloride UPS for the Gantry Systems


a Push and hold the Inverter Stop button on the UPS control panel until Inverter Indicator light flashes..
1

1. EPO 2. Inverter Start


3. Inverter Stop 4. Audible Alarm Cancel
Control Panel for Gantry UPS
b Press the EPO button.
c The UPS alarm beeps and the indicator light flashes.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 20
GEMINI GXL Installation Technical Manual System Power Up/Down

d Go to the main circuit breaker box and turn off the main circuit breaker for the gantry systems.

CAUTION To avoid damage to the system, after turning off the main circuit breaker, do not turn on the power until
you have waited at least 30 seconds.
e Press the Alarm button on the UPS control panel to silence the alarm.

Restart Computers
Use this procedure as needed to restart both the CT and the PET computer systems on the GEMINI System
without turning off the power to the gantries.

Restart CT and PET Computer Systems


1 Close all applications and open windows on the workstations.
2 Shut down the PET acquisition computer.
a At the Acquisition Workstation, select Home and then select Utilities
b Select PET Shutdown from the Misc menu displayed in the Task
Information box.
3 At the prompt on the Acquisition Workstation, type yes to confirm shutdown of the system and press Enter.
Wait until the system displays a message to indicate shutdown is complete.
The shutdown procedure may take up to ten minutes. Do not continue until the system displays a prompt to
continue.
NOTE Once you have shutdown the PET acquisition computer, you cannot resume operations until you have turned off and
turned back on the power to the PET gantry and restarted the CT computer system.
4 Shut down the PET Workstation computer.
a Click anywhere on the desktop to display the Main Menu and click
Exit.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 21
GEMINI GXL Installation Technical Manual System Power Up/Down

b b At the Login window, click Options > Command Line Login.

c c Press Return.

d At the Solaris Login box, type shut.


e Press Return.
f When prompted, type in the password.
g Press Return.
h Answer any additional prompts.
Wait for the system to display the message Program Terminated OK.
5 Turn off the power to the PET Workstation computer and monitor.
a Open the front panel on the PET Workstation computer.
b Press and hold the On/Standby button to shutdown the computer.
6 Turn off the power to the PET acquisition computer by turning off the Computer switch on the PET Gantry
Power Distribution Panel at the rear of the PET Gantry.
Wait one minute.
7 Turn on the power to the PET Workstation computer and display monitor.
a Open the front panel on the PET Workstation computer.
b Verify the power switch that is next to the On/Standby button is set to On.
c Press and hold the On/Standby button to start up the computer.
Allow approximately three minutes for the computer to start up.
8 Log on to the PET reconstruction computer.
a Enter patient as the user name.
b Enter the password.
c Press Return.
9 Restart the PET acquisition computer by turning on the Computer switch at the PET Gantry Power Distribution
Panel.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 22
GEMINI GXL Installation Technical Manual System Power Up/Down

10 Turn off the power to the CIRS computers.


a Open the CIRS Computer Cabinet.
b For each computer in the CIRS cabinet, press the power button and hold it until the power indicator light
turns off, approximately seconds.
It is not critical to turn off the computers in a particular sequence.
NOTE Do not turn off the power switch at the bottom of the CIRS Computer Cabinet
11 Close all open applications on the Acquisition Workstation.
12 Log off the Acquisition Workstation computer.
a Select Home on the Desktop, and then click Logout from the Scan Tool Box.
The system displays a prompt to confirm you want to logout.
b Click Yes to confirm logout.
The Logon Information window opens.
13 Click Shutdown.
14 Select Shutdown and then click OK.
Wait as the computers automatically shut down and turn off.
NOTE It is imperative that you wait the full three minutes before proceeding.
15 Turn on the power to the Dell computer for the Acquisition Workstation.
If necessary, turn on the power to the Acquisition Workstation display monitors.
When the computer start up process begins, allow approximately 90 seconds for the process to complete.
The Log In window appears.
16 At the Log In window, type pet as the required username.
A password is not required.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 23
GEMINI GXL Installation Technical Manual System Power Up/Down

17 Press Enter or click OK.


Wait as the CT Acquisition computer starts up. Allow at least three minutes for the computers to start up.
The CT computers connect to the PET computer system. Typically the system requires three to five minutes for
the computers to turn on.
If it appears that computers are not turning on or establishing connections, allow ten minutes for the start up and
connection to complete before initiating troubleshooting. The system displays an error message if the computers
do not establish a connection.
18 If the Acquisition Workstation displays the Keyswitch Prompt, turn the keyswitch on the side of the Scan
Control Panel to On and then release the key to the Standby position.
19 If the system displays the Table Homing window, click Start to home the patient table.
The system must know the home position of Patient Table before performing any test procedures.
20 If the system displays the Separation System Homing window, click Enable to home the Gantry Separation
System. The system prompts you to close the Gantry Separation Unit.
21 If required, close the Gantry Separation System. At the Gantry Separation System keypad, press the Close
and Enable buttons to completely close the gantries.
The Closed indicator light on the keypad turns on when the system is fully closed.

Gantry Separation System Closed Indicator


22 The system displays the appropriate Desktop on the workstation monitors.
Perform the required Daily Quality Control and Calibration procedures as established by the site.
Do not attempt to perform procedures before performing Daily Quality Control and calibration. If you attempt to
perform procedures before completing quality control and calibration, the system may not function as
expected. For example, the laser lights on the PET gantry are not available until the System Initialization
procedure is performed.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 24
GEMINI GXL Installation Technical Manual
Introduction

Introduction
This manual contains the procedure for properly installing the components of the GEMINI System in both Fixed,
Mobile and Fixed Mobile Sites. The GEMINI System is a hybrid PET/CT System, capable of performing sequential
PET and CT studies on a patient. The integral patient table can keep the patient relatively motion free, and
images obtained in this manner may later be viewed as multi-modality images with both anatomical and functional
information overlaid. A drive and alignment system referred to as the Gantry Separation Unit, or GSU, allows the
PET portion of GEMINI to move rearward enabling unobstructed CT imaging and improved patient comfort.

It is recommended that you read through the entire manual once before performing any work to familiarize
yourself with the procedures as they are described.
Equipment, Parts, and Tools

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 25
GEMINI GXL Installation Technical Manual

Equipment, Parts, and Tools


GEMINI Installation Kit (P/N 4535 679 45401, Supplied with System)
Description QTY P/N
Floor Leveling Template Assy, Separation Unit 1 4535 679 41191
Floor Leveling Template Assy, Front Table 1 4535 679 41201
Floor Leveling Template Assy, Rear Table 1 4535 679 42611
Shims, 0.003” Thick, PET Detector 8 4535 679 45471
Shims, 0.005” Thick, PET Detector 8 4535 679 37751
Shims, 0.015” Thick, PET Detector 8 4535 679 45481
Shims, 0.030” Thick, PET Detector 8 4535 679 37741
Grout, Micorox Xtra–Fluid 25 lb.and Stirring Paddle 3 306481
Barrier Plate, Floor Pour 22 104487
Drill Bit, Hilti 18 mm Dia. 1 306460
Anchor, Hilti #HSL M12/25mm (Qty. 15) 18 306457
Petroleum Jelly, 1 lb. 1 305053
Paint Roller, 4” Wide 1 4535 679 50411
Center Punch 1 4535 679 34761
Cable, SC Fiber Duplex 100 ft. 1 711439
Duct Tape, Gray Cloth 1 305020
Paddle 1

PET Gantry Transporter 1

CT Gantry Transporter 1

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 26
GEMINI GXL Installation Technical Manual

GSU Transporter 1

Shipping Cart, Table 1 4535 679 48031


Grout Hardener for 306481 3 4535 679 54851
Equipment, Parts, and Tools

Service Installation Kit (P/N 4535 679 50261 Recyclables, Ordered by Installers, Shipped Separately from System)

Description QTY P/N

Feeler Gauge Set 26 leaves, 3” L 1 4535 679 50391

Precision Machinists Level 1 4535 679 47871

Digital Level 1 4535 679 50341

Ruler, 24” Aluminum 1 4535 679 50381

Fixture Assembly, Laser Alignment 1 4535 679 47781

Fixture Assembly, Mirror Alignment 1 4535 679 47791

Short Couch Alignment Plate 1 4535 679 69211

1/2” Drive Socket Wrench, 16” Length 1 4535 679 50401

Calipers, Plastic 1 4535 679 51941

Metal Locator, Electronic 1 4535 679 48101

Drill, Hilti Hammer Style 1 100789

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 27
GEMINI GXL Installation Technical Manual
Tools Required but not Supplied

Description QTY P/N

Standard FSE Tool Kit 1 4535 662 72391

CT Service Tool Kit 1 4535 662 84931

Electric Drill 1/2” Chuck 1

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 28
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Pre-Installation Guide and Site Preparation


Personnel Required Time Required

1 Field Service Engineer 4 Hours

General Information
Prior to moving any system components or beginning any service work, it is important that you are aware of the
specific installation requirements of your facility. Each Fixed Site installation must be performed in accordance with
the Site Planning Guide, and conditional items will vary from one site to another. The installation for Mobile
Sites (Fixed location or trailer) will have a few variances which are described in this manual. Review the
following and determine which of the conditions may apply to your specific installation. This will help you follow the
procedure presented in this manual, and ensure that you can perform the installation in a timely and efficient
manner.
• Discuss any special instructions or requests made by the Facility Administrators with the local Philips Service
Specialist.
• Identify and retrieve any additional equipment that will be required prior to beginning the installation procedure.
• Determine the minimum vertical clearance of the ceilings and door ways. You may have to remove the top
cover set from the gantries before moving them in order to gain additional clearance.
• Evaluate how much working space you will have in the room where the GEMINI system is to be installed. This is
very important and will affect the installation procedure order of operations. For example: in some rooms, if the
patient table is installed before the CT gantry is brought into the room, there may not be enough clearance
around the table for the CT gantry to be positioned. As a result, some installations may need to have all
equipment placed in them first.
• Survey the room floor for levelness. This is critical for proper operation of the GEMINI system. Floor levelness
shall not exceed 1/2” (13 mm) when measured between any two peripheral points within the area of installation.
It is the customer’s responsibility to correct the floor if the results of the floor measurements do not meet these
specifications.
• Ensure that the floor will carry the system loads. Extra soft floor coverings may cause the settling of the system,
resulting in a change to the gantry alignments.
• Due to the weight of the GEMINI gantries it is recommended that they be positioned in the room before leveling
the GSU and patient couch. This will prevent errors caused by floor deflection from affecting the installation.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 29
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

• Note that when servicing the Patient Table, the GEMINI Patient Table does not utilize a Table Safety Brake,
unlike earlier Philips medical equipment you may have serviced.
Additionally, you will be asked to refer to other supporting documentation throughout this manual:
• GEMINI Site Planning Guide
• GEMINI Field Replaceable Parts Technical Manual (P/N 4535 671 38821)
• GEMINI Calibration Technical Manual(P/N 4535 671 38781)
• GEMINI Performance Technical Manual(P/N 4535 671 38791)
• UPS Installation Technical Manual (P/N 4535 670 12221)
• GEMINI Battery Pack Extender Field Installation Procedure (P/N 4535 679 56791)
• Tripp Lite UPS Owner’s Manual (supplied with UPS)
• Tripp Lite External Battery Pack Owner’s Manual (supplied with Battery Pack)
• Teal Line Transformer Technical Manual (supplied with Transformer)
The GEMINI System Installation Overview flow chart on the following page has been provided to give an overview
of the major procedural steps you will follow during the installation process.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 30
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 31
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Safety Information
Handling Heavy Loads
Always wear the proper clothing (safety boots and gloves) when handling heavy or awkward loads. Ensure that
correct lifting methods are used to avoid physical injury. Follow all procedures carefully and use mechanical
devices or several FSEs to handle extremely heavy or awkward loads.

WARNING ONE PERSON SHOULD NOT ATTEMPT TO LIFT ANY PART OR ASSEMBLY
GREATER THAN 25 LBS. (11 KG) TO A HEIGHT ABOVE THE WAIST. DO NOT ALLOW ONE PERSON TO
ATTEMPT TO LIFT ANY PART OR ASSEMBLY THAT APPEARS TO BE UNREASONABLE IN WEIGHT OR
PHYSICAL SIZE.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 32
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Radiation Safety
Radioactive Sources
The GEMINI system requires seven radioactive calibration sources:
• Six Na-22 quality control and calibration sources. These sources supersede the Germanium line source
(SOURCE-GE-0.5) shipped with earlier scanners. Sodium has a longer half-life than Germanium, and
should therefore only require replacement every 2 years.
• One Na-22 quality control and calibration source.
For domestic shipments, the sources are drop-shipped from the source vendor: Isotope Products Laboratories
(IPL). For export, the sources must be purchased separately from IPL’s local distributor. Ordering information is
provided at the time of sale.

Item Source-Na-22 (Qty. 1) Source-NA-22 (Qty. 6)

Activity (nominal) (0.1 mCi) 3.7 MBq (0.01mCi)

Housing material High-strength plastic

Housing thickness 1.2 mm (0.050 inch)

Exit window material Plastic

Exit window thickness 1.2 mm (0.050 inch)

Mechanical Drawing SOURCE-NA-0.15

Isotope Products
Manufacturer Laboratories (IPL)

Manufacturer’s part number GF-0227 0.10 mCi

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 33
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Applicable Regulations
Radiation safety considerations in the design of the GEMINI are driven by US Nuclear Regulatory Commission
(NRC) rules. The NRC specifies that the maximum annual dose of radiation permitted for workers in the United
States is 5000 mrem per year (NRC Title 10 Part 20 (1-1-95)). This corresponds to an average dose of
approximately 2.5 mrem/hour. For gamma radiation, 1 mrem=1 mR=1 mrad=10 USv.
The National Council on Radiation Protection and Measurement (NCRP) recommends the adoption of “ALARA: As
Low As Reasonably Achievable” as the operating policy of radiation safety programs. The NCRP specifies that the
maximum acceptable level of emitted radiation from a medical device with its source in a stowed (non-scanning)
condition is 200 mrad/hr at 5 cm and 2 mrad/hr at 1 m (NCRP Report no. 102 “Medical Equipment Design,
Performance and Use”).

Radiation Considerations
The isotopes used in the GEMINI emit a higher radiation than the isotopes used in a single photon imaging; 511
keV vs. 140 keV for Tc-99m. This means that special care must be taken in the layout of the PET suite.
The GEMINI exam room should be close to the radiopharmaceutical dispensing area and a patient waiting area.
The patient waiting area is a room where the patient can wait during the uptake period, i.e. after the activity has
been injected. These three areas should be a reasonable distance from offices and Nuclear Medicine scanning
rooms.
Customers should consult a Radiation Safety Person (RSP) to determine proper safety requirements.
Patient Dose
In a PET study, the patient is exposed to radiation:
• Internal exposure from the injected dose of a radiopharmaceutical
In keeping with the principle of ALARA, minimizing the dose administered to the patient was an important
consideration in the design of the GEMINI system. This has been accomplished by reducing the injected dose.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 34
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

General Requirements
Recommended Exam Room Size 26' x 16' (7.9248 m x 4.8768 m)

Recommended Control/Equipment Room


Size 12' x 16' (3.6576 m x 4.8768 m)

Minimum Exam Room Size 24' x 14' (7.3152 m x 4.2672 m)

Minimum Control/Equipment Room Size 8' x 14' (2.4384 m x 4.2672 m)

Scanner Dimensions 20' 1" x 6' 10 1/2" (6.1214 m x 2.0955 m)

Total System Weight 8325 lbs. (3776 kg), excluding power conditioning equipment

Temperature Requirements 18 - 24° C (64 - 75° F) Temp. changes not to exceed 5deg F/hr.

Humidity Requirements 30% - 60% non-condensing

26,000 BTU/hr. (20,000-CT / 6,000-PET)


Heat Load: Exam Room 4,000 BTU/hr. (PET and CT workstations and misc.)
Add 5,000 BTU/hr. for Recon Cabinet
Heat Load: Control Room Does not include power regulating devices.

Floor Loading (Floor levelness not to vary in


excess of 1/2” within the footprint of the Unless the floor is reinforced concrete slab poured on grade, a
system Licensed Structural Engineer must evaluate the floor capacity.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 35
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Power Requirements
Mains

Type:
Three Phase
Volatage:
480VAC rms + 10%
Frequency:
50 or 60 Hz, nominal
Sag/Surge Regulation:
5%
Distribution Transformer:
100Kva, minimum

Shielding Requirements Must be determined by Radiation Supervisor

Dedicated analog telephone line required for


Modem Line remote support

F-18 - Amount as required for patient doses

Radioactivity Licence Requirements Na-22 - 10 mCi (3.7 MBq, Qty. 6);IPL#MMS05-22-10U


Na-22 -100 mCi (.037 MBq, Qty. 1); IPL#GF-0227

The magnetic field may not exceed 1.0 Gauss at any point on
Environmental Requirement the system.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 36
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Fixed Site Requirements

EXAM ROOM = 416 SQUARE FEET (38.2 sq.m.)


TOTAL SUITE = 608 SQUARE FEET (55.9 sq.m.)

Figure 1: Recommended Room Layout (Fixed and Fixed Mobile Sites)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 37
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

EXAM ROOM = 336 SQUARE FEET (30.9 sq.m.)


TOTAL SUITE = 455 SQUARE FEET (41.8 sq.m.)

Figure 2: Minimum Room Layout (Fixed and Fixed Mobile Sites)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 38
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Figure 3: Preferred Raceway Layout

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 39
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Figure 4: Preferred Junction Box/Conduit Layout

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 40
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Mobile Requirements
GEMINI System Power Quality Requirements

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 41
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Branch Power Recommended 112.5kVa

Maximum Power Required 80kVa

Nominal Input Power (+/– 10%) 480VAC, 3-Phase WYE, Grounded Neutral

Line Frequency Range 47 to 63Hz

Room Disconnect or Circuit Breaker 100 Amps

Phase-to-phase Impedance 0.200 Ohms

% Load Voltage Regulation ±10% Total at System

% Voltage Drop Allowed in Cable 1%

Instantaneous Current 135 Amps

Voltage Drop in Cable 4.8 V

Phase to Phase Voltage Balance 5%

Line Voltage Impulse, (Transient) (<1ms) 1200Vpk line to line

Line Voltage Swell (>1 cycle, <3 sec) Swell to 140% of the line V

Line Voltage Sag (>1 cycle, <3 sec) Sag to 90% of the line V

Voltage Dropout Allowed (% nom. V for x 70% <100ms, 0% <8ms


sec)
Total Harmonic Voltage Distortion <10% of Maximum Load

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 42
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Power Closet Environmental Requirements

Gantry Power Conditioner and/or Gantry UPS


Power closet recommended minimum * 4’-0” x 10’-0” (1220mm) x 3050mm)

Unrestricted access to UPS cabinets (front: 12” (300mm) one side: 36” (910mm)

Total system weight Approx. 4,000 lbs. (1812kg) (battery and elec. Cabinet)

Temperature requirements 85 F (30 C) maximum

Humidity requirements 5% - 95% non-condensing

Heat load 5,500 btu/hr (typical)

Audible noise < 65 dBA @ 1 meter

Unless the floor is reinforced concrete slab on grade, a licensed


Floor loading structural engineer must evaluate the floor capacity

Power requirements to UPS


Mains
type:
Voltage: Three phase Delta plus protective earth
480V –20% nominal VAC rms
Frequency: Sag/ (other voltage configurations available as req.)
Surge regulation: 50 or 60 Hz, nominal
Distribution 5%
transformer: 100 KVA, minimum
* It is required that the Power Conditioner and UPS be located in an environmentally controlled equipment closet.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 43
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Recommended location is in an area outside of the Gemini Suite.


Gantry Isolation Transformer
Cable entry/ exit Rear

Total system weight Approx. 720lbs. (326kg)

Temperature requirements 100 F (38 C) maximum

Humidity requirements 5% - 95% non-condensing

Heat load 1950 btu/hr (typical)

Audible noise < 50 dBA @ 1 meter

Power
requiremen
ts Mains
type: Three phase Delta plus protective earth
Voltage: 200V – 500V
Frequency: 50 or 60 Hz, nominal

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 44
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Wire size from Distribution Source or Power Conditioning Unit to Room Main Disconnect Switch or Circuit
Breaker Box:
0–50 ft. (15.2 M) 2
4 #2 (25 mm )

51–100 ft. (30.5 M) 2


4 #1 (35 mm )

101–200 ft. (61 M) 2


4 #00 (70 mm )
Wire size from Room Main Disconnect Switch or Circuit Breaker Box to CT Scanner Gantry:
25 ft. (7.6 M) Maximum 2
4 #4 (35 mm )
All of the above listed conductors include a dedicated (earth) ground wire the same size as the power conductors.

Control Room Power Requirements


The CT and PET Consoles are powered by a UPS with 220V, 50/60Hz, 2kVa (1600W minimum).

Power Box Description


Main Disconnect: 600V, 3 phase, 3 pole, 100A breaker
Circuit Breaker used will be UL listed or recognized, rated 100A maximum. To ensure compliance with IEC 60328,
verify that the circuit breaker is not marked with a Greek symbol. Look for VDE, TUV, MEMKO, SEMKO, DEMKO,
or FIMKO certification mark.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 45
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Remote on/off switch: 120V, single-phase

Figure 6: System Cabling

Note: Mobile Sites do not


have a System UPS but rather an Acquisition UPS only.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 46
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Figure 7: Incoming Power Schematic

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 47
GEMINI GXL Installation Technical Manual Pre-Installation Guide and Site Preparation

Figure 8: Cabling Diagram

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 48
GEMINI GXL Installation Technical Manual
Unpacking and Transporting System Components

Unpacking and Transporting System Components


The physical offloading, unpacking, and transport of the system from receiving area to scan room is the
responsibility of a third-party company contracted to perform these specific duties. However, the lead FSE should
be present to supervise these activities and ensure that the system components are handled properly. For your
reference, some useful information is included below.
• Determine whether or not the top covers will have to be removed from the PET gantry. Both gantries of the
GEMINI system require a minimum 48” (1220 mm) door opening to pass straight through the doorway. Overall
system height is 81.5” (2060 mm) with the covers on and the CT baseplate lifted 1 3/4” (45 mm) off the floor.
Passage under a standard 80” (2030 mm) door requires that the top covers be removed. Refer to the GEMINI
Field Replaceable Parts Technical Manual (P/N 4535 671 38821) for further information on cover set
removal and replacement. Should it be necessary to remove the top covers, a step ladder may be needed.
• Depending on the site specific installation details and instructions, it may be required that masonite or similar
material in the path of the gantry be laid out to protect the facility floors from damage. Refer to the GEMINI GXL
Site Planning Guide for further information.
• Crating panels, plastic wrapping, cardboard, and foam should be removed from the equipment and set aside or
discarded. This is also the responsibility of the third-party contractor.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 49
GEMINI GXL Installation Technical Manual Floor Template Preparation
Uninterruptable Power Supply Setup

Uninterruptable Power Supply Setup


The 2.2kVa Tripp Lite UPS and Battery Pack should be installed at this time. Refer to the following Technical
Manuals for reference during installation and set up of the UPS and Battery Pack:
• UPS Installation Technical Manual (P/N 4535 670 12221) • GEMINI Battery Pack Extender Field
Installation Procedure (P/N 4535 679 56791) • Tripp Lite UPS Owner’s Manual (supplied with UPS) • Tripp
Lite External Battery Pack Owner’s Manual (supplied with Battery Pack)

NOTE The Tripp Lite UPS and Battery Pack should be serviced by a trained Tripp Lite Field Technician.
Reference the products’ Owner’s Manual. Failure to comply will void the product warranty.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 50
GEMINI GXL Installation Technical Manual Floor Template Preparation

Floor Template Preparation

Floor Template Used for Mobile and Fixed Mobile

Overview
Mylar floor templates are provided to designate the GSU and patient table locations within the trailer.

PROCEDURE

1 Ensure that the work area is clean, and remove any non-essential equipment from the area.
2 Locate and identify the GSU and Patient Table floor templates that are shipped with the system.
3 Establish the centerline of the system on the floor (using a non-permanent writing utensil) to properly orient the
floor templates.
NOTE Acceptable approaches include (but are not limited to) laser lines, and a string or cable used to make reference
marks on the floor. Chalk lines can be difficult to lay out and are easily misread so they are not recommended.

HINT Use a string and pencil to mark a centerline the entire length of the room. Stretch the string tight and mark the
floor using a pencil.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 51
GEMINI GXL Installation Technical Manual Floor Template Preparation

6’ Min.
(1.83 m)
to wall

4 Place the GSU and Patient Table templates on the floor utilizing the established centerline.
5 Tape the mylar templates to the floor.
a Check the layout and positioning measurements again. Ensure that
the mylar templates are in the correct position relative to the room
using the centerline on the floor as a guide.
b Tape the four corners and both sides of the GSU template down.

c Tape the four corners and sides of the Patient Table template
down.

6 Draw a line on template perimeters.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 52
GEMINI GXL Installation Technical Manual Floor Template Preparation

7 Remove templates and cut wood flooring on the perimeter line to


expose floor plating for the GSU and Couch mounting plates.
8 Thoroughly clean cutout areas before inserting mounting plates.
9 Insert the mounting plates into the cut areas to ensure a proper
fit. Mark the mounting holes and then remove the mounting plates.
10 Drill mounting holes through metal floor. Additional information
on the steel floor plate and fasteners can be found in 453567451491.
11 Insert and bolt the GSU and Couch mounting plates.
12 Proceed to the “Gantry Separation Unit Leveling” and “Patient
Table Mechanical Alignment and Leveling” Sections.
Ethafoam floor templates and Micorox Compound Pouring are not
required for the Mobile Units.

Couch Mounting Plate

NOTE See the “Mobile Specific” procedure at the end of the “Gantry Separation Unit Leveling” section to install the “Lock/Unlock
Switch to GSU Cable” (4535 679 71841).

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 53
GEMINI GXL Installation Technical Manual Floor Template Preparation

Fixed Site Floor Template


Personnel Required Time Required

2 Field Service Engineers 6 Hours (3 Lead and 3 Assistant)

Overview
Mylar and ethafoam floor templates are provided to designate the GSU and patient table locations within the room.
In addition to providing positioning information, the templates provide an area in which the Micorox compound can
be poured into once the GSU and Patient Table have been leveled. Refer to the GEMINI Site Planning Guide for
more information and then complete the following procedure to prepare the for more information and then complete
the following procedure to prepare the floor templates. It is important to note that some room sizes will require all of
the equipment (both gantries, patient table, etc.) to be in the room before laying out and preparing the floor
template. Before securing the templates to the floor, verify that will be sufficient clearance to bring the equipment
into the room without obstruction.
Bare concrete is always preferable to any type of floor covering, as soft floor coverings such as linoleum tile can
take weeks to settle under the weight of the gantries. This will make it very difficult to level and align the system. If
you are installing over a soft floor covering, it is recommended that you return and re-align the system after 30
days.
Floor Template Layout
PROCEDURE

1 Ensure that the room floor is level within 1/2” (13 mm), measured between any two peripheral points in the
room.

Entire Floor Area 1/2” (13 mm)


Allowable
Levelness
Tolerance

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 54
GEMINI GXL Installation Technical Manual Floor Template Preparation

CAUTION The 13 mm floor level tolerance is the total acceptable error. This includes all tolerances. For
example, 13.1 mm is not acceptable. Failure to comply with this specification will result in an
improperly installed and aligned system.
2 Ensure that the work area is clean, and remove any non-essential equipment from the room.
3 Locate and identify the GSU and Patient Table floor templates that are shipped with the system.
4 Establish the centerline of the system on the site floor (using a non-permanent writing utensil) to properly orient
the floor templates. Set the initial installation location as specified in the Site Planning Guide.

NOTE Floor tile seams are not straight enough to satisfactorily align the templates; do not use them.
Acceptable alternative approaches include (but are not limited to) laser lines, and a string or cable used to make
reference marks on the floor. Chalk lines can be difficult to lay out and are easily misread so they are not
recommended.
Note the distance between the two floor templates, measured from the inside edge of the foam, will be
62 13/16" (1.595 m).
HINT Use a string and pencil to mark a centerline the entire length of the room. Stretch the string tight and mark the
floor using a pencil.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 55
GEMINI GXL Installation Technical Manual Floor Template Preparation

6’ Min.
(1.83 m)
to wall

13/16
(1581MM)

5 Place the GSU and Patient Table templates on the floor utilizing the established centerline.
6 Verify that the GSU mounting holes are clear of rebar using an electronic metal locator (P/N 4535 679 48101).
• If rebar is encountered, use the alternate mounting holes.
The GSU mounting hole locations (shown on the GSU floor template) must be checked for rebar as the alternate
mounting holes are not staggered but in-line with one another.
NOTE All electronic metal locators come with instructions. It is strongly recommended that you read and understand
them prior to operating the device.

CAUTION The GSU mounting hole locations must be checked for the presence of rebar prior to affixing the
template in place on the site floor. Failure to comply may result in one or both sets of mounting
holes being obstructed by rebar.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 56
GEMINI GXL Installation Technical Manual Floor Template Preparation

7 Tape the mylar and foam templates to the floor.


a Check the layout and positioning measurements. Ensure that the mylar templates
are in the correct position relative to the room using the centerline on the floor as a
guide.
b Tape the four corners and both sides of the GSU template down.
c Tape the four corners and sides of the Patient Table template down.
The foam surrounds the areas in which the leveling compound will be poured.

Mylar and Foam Patient Table


Floor Template

Duct Tape
(Four Corners)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 57
GEMINI GXL Installation Technical Manual Floor Template Preparation

GSU Floor Template Preparation


PROCEDURE

1 Tape the edges around the cut-outs to the bare floor.


The cut-outs prevent the leveling screws from tearing the
mylar once the GSU has been set down onto the template.
Taping the edges of the cut-outs down prevents the grout
leaking onto the floor.

Sections of Tape Placed Around the Outside Edge

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 58
GEMINI GXL Installation Technical Manual Floor Template Preparation

2 Apply a thin layer of petroleum jelly to the foam strips


where indicated.
a Use latex gloves and apply the petroleum jelly Ethafoam Strips
manually to the foam strips along the entire length of
the edges indicated by the arrows.
• Ethafoam (outermost) strips: Apply petroleum jelly
to the inside edges only.
Open Cell Foam Strips
• Open cell foam (center) strips: Apply petroleum
jelly to the top and inside edges (towards the
ethafoam strips) only.
b Maintain even coverage and avoid applying an
excessive amount in any one section.
In the image shown, the horizontal arrows indicate the
inside edges of the foam, and the vertical arrows indicate
the top of the foam. The petroleum jelly will prevent the
leveling compound from sticking to the foam wherever it is
applied. This allows the foam to be removed once the leveling compound has cured.

CAUTION Failure to properly cover the ethafoam with petroleum jelly where indicated will cause the foam
to bond to the leveling compound. The foam will be extremely difficult to remove and will be
visually unappealing.
NOTE A small amount of petroleum jelly applied to each of the foam seams will help reduce the potential for leaks.

Patient Table Floor Template Preparation


PROCEDURE

1 Apply a layer of petroleum jelly to the foam.


a Use latex gloves and apply the petroleum jelly manually to the foam strips along the entire length of the edges
indicated by the arrows.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 59
GEMINI GXL Installation Technical Manual Floor Template Preparation

• Ethafoam (taller) strips: Apply petroleum jelly to the


inside edges only. Open Cell
• Open cell foam (shorter) strip: Apply petroleum jelly Foam Strip
to the top and inside edge (towards the ethafoam
strips) only. b Maintain even coverage and avoid
applying an excessive amount in any one section.
c Repeat for the remaining end of the patient table
template.
Ethafoam
Strips

CAUTION Failure to properly cover the ethafoam with petroleum jelly where indicated will cause the foam to
bond to the leveling compound. The foam will be extremely difficult to remove and will be visually
unappealing.
In the image shown, the horizontal arrows indicate the inside edges of the foam, and the vertical arrows indicate
the top of the foam. The petroleum jelly will prevent the leveling compound from sticking to the foam wherever it is
applied. This allows the foam to be removed once the leveling compound has cured.
NOTE A small amount of petroleum jelly applied to each of the foam seams will help reduce the potential
for leaks.
2 Cut-out the leveling screw holes from the patient table floor template.
This will prevent the leveling screws from tearing the mylar.
3 Tape the edges of the leveling screw cut-outs to the floor.
This will prevent the epoxy from leaking past the template and onto the floor.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 60
GEMINI GXL Installation Technical Manual Gantry Separation Unit Leveling

Gantry Separation Unit Leveling

Personnel Required Time Required

1 Field Service Engineers 2 Hours

Overview
In the following steps, you will position the GSU relative to the floor template and then level it prior to installation.
However, to ensure accuracy, the site floor must pre-loaded with the weight of the CT and PET gantries placed
roughly in their installed position before any adjustments are made to the leveling screws. When leveling the
GSU, the Patient Table must be set to the same height as the GSU. Due to floor irregularities, the GSU may have
to be placed higher or lower to achieve the proper alignment with the Patient Table.

Figure 9: PET & CT Gantries Placed Into Position: Floor Preloaded

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 61
GEMINI GXL Installation Technical Manual Gantry Separation Unit Leveling

Leveling the Gantry Separation Unit

PROCEDURE

1 Roll the GSU unit into the work area.


a Remove any packing and/or shipping material from the separation unit.
b Position the GSU alongside the GSU floor template.
2 Inspect the GSU baseplate.
a Verify that adhesive backed mylar has been installed GSU
on the bottom of the GSU.
b Verify that the mylar has been cut out of the viewing
holes.
Spot Faced Non-Spot
c Trim away the mylar from the viewing holes if necessary. Anchor Hole Faced Viewing
The viewing holes are the holes along the sides of the GSU (Mylar Not Cut Hole (Mylar Cut
baseplate that do not have the spot face. (The spot faced holes Out Here) Out Here)
are the anchor holes.) The leveling compound must be able to
rise to the top of the leveling holes.

CAUTION Ensure that both the GSU baseplate and Micorox compound are at room temperature before pouring.
Failure to do so may compromise the leveling procedure. Additionally, the GSU could become warped as
a result of the excessive temperature differential.
3 Coat the leveling screw threads with petroleum jelly prior to installing the GSU.
a Remove the three leveling screws from the GSU if not already done.
b Apply petroleum jelly to the entire length of the screw threads. Re-install the leveling screws into the GSU,
but do not allow them to extend below the bottom of the GSU.
This will ensure that the screws can be removed once the leveling compound has cured.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 62
GEMINI GXL Installation Technical Manual Gantry Separation Unit Leveling

4 Position the separation unit over the floor template.


a Roll the separation unit over the floor template.
b Align the GSU to the template using the inside edges of the ethafoam.
Position the GSU so that it will rest inside the ethafoam when lowered.
5 Coat the inside diameters of the GSU viewing holes
with petroleum jelly. GSU
This will ensure that the leveling compound does not bond to
the GSU baseplate.

Spot Faced Non-Spot Faced


Anchor Hole Viewing Hole
(Petroleum Jelly Here)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 63
GEMINI GXL Installation Technical Manual Gantry Separation Unit Leveling

6 Lower the GSU onto the floor template. GSU In Position


a Lower the three leveling screws so that the
weight of the GSU is supported by the screws and
not the transporter wheels.
b Remove the shipping support braces.
c Remove the transporter wheels.
d Raise the leveling screws evenly to lower the
GSU until it rests completely on the inner foam
strips.

CAUTION Ensure that the GSU compresses both of the foam strips located in the center of the floor template. If
the separation unit does not make contact with and compress the center strips, a leak will occur when
the leveling epoxy is poured.
7 Position the PET gantry over the GSU.
a Raise the PET gantry casters to the maximum height.
b Position the PET gantry over the GSU at a point approximating the middle range of the GSU travel.
(See Figure 10 for more detail.)
Prior to leveling the GSU it is recommended that both the PET and CT gantries are located in their installed
positions relative to the GSU and the Patient Table. This is critical as the site floor must be preloaded to
maximum deflection prior to leveling the system components.

CAUTION Failure to locate the PET and CT gantries in their installed positions relative to the GSU and the Patient
Table may result in an installation that does not allow the Patient Table to be properly calibrated.
8 Position the CT gantry between the GSU and the Patient Table.
• Roll the CT gantry into position between the GSU and the Patient Table. (See Figure 10 for more detail.)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 64
GEMINI GXL Installation Technical Manual Gantry Separation Unit Leveling

CAUTION Ensure that the gantries are as close to their installed positions as possible before performing any
leveling adjustments. If the site floor is not preloaded with the weight of the gantries before the
GSU and Patient Table are leveled, the system alignment will not be accurate.

Figure 10: PET & CT Gantries Placed Into Position: Floor Preloaded

NOTE Whenever instructed to use the precision level, it is important that you mark the exact location on which you have
placed it using a pencil. The level must be placed in the exact same location each time it is used to yield accurate and
reliable readings.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 65
GEMINI GXL Installation Technical Manual Gantry Separation Unit Leveling

9 Level the GSU in the X-axis.


Precision Level
a Place the precision level across the X-axis, or width, of the PET
mounting plate.
b Note and mark the location of the level so that future
measurements may be made from the exact same location.
The goal is to level the GSU to the same height as the Patient Table
while keeping it as low to the ground as possible. The leveling
X-Axis
compound requires a 1/4” minimum pour thickness, and the
compressed thickness of the center foam strips ensures that this is
met. A gap between the bottom of the GSU and foam strips will result
in a leak when pouring the leveling compound. Therefore, the leveling
must be accomplished without the presence of a gap.
c Tighten the corresponding leveling screw(s) in one-quarter turn
increments until the plate is level across the width of the PET
mounting plate.

Center Marks
NOTE Any adjustment of the leveling screw(s) that results in no movement of the precision level bubble indicates that the
leveling screws are not in contact with the floor.
NOTE
The axis is considered level when the bubble reads ±2 divisions from centered on the scale. Note that the scale has
the center position accentuated, and the bubble can be off two divisions in either direction.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 66
GEMINI GXL Installation Technical Manual Gantry Separation Unit Leveling

PET Side
GSU Leveling Screw Adjustments

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 67
GEMINI GXL Installation Technical Manual Gantry Separation Unit Leveling

10 Level the GSU in the Y-axis.


a Place the precision level on the left side of the GSU rail as Precision Level
shown.

Y-Axis

b Tighten the corresponding leveling screw(s) in one-quarter


turn increments until the level bubble is centered as shown.
Tightening a leveling screw will raise that side or end of the unit,
while loosening a leveling screw will lower it.

Center Marks

NOTE Any adjustment of the leveling screw(s) that results in no movement of the precision level bubble indicates that
the leveling screws are not in contact with the floor.

NOTE The axis is considered level when the bubble reads ±2 divisions from centered on the scale. Note that the scale
has the center position accentuated, and the bubble can be off two
divisions in either direction.
11 Re-check the GSU for level in the X-axis.
a Place the precision level across the width of the plate to insure that
Precision Level it is still level.

“X” Axis

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 68
GEMINI GXL Installation Technical Manual Gantry Separation Unit Leveling

b Repeat step 10, if necessary.


The level must be placed in the same location as it was previously measured for a consistent reading. It is
possible that the Y-axis leveling may have altered your initial adjustments.

NOTE If it is necessary to make multiple adjustments with the leveling screws, be aware of the height of the
separation unit in relation to the floor template foam. If you raise the separation unit too high in the attempt to
level it, you may create a gap between the bottom of the unit and the foam which will allow the leveling
compound to leak out.

12 Re-check the GSU for level in the Y-axis.


Precision Level

Y-Axis

• Place the level along the length of the left side GSU rail and
check that the rail is still level. Repeat step 11, if necessary.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 69
Center Marks
GEMINI GXL Installation Technical Manual Gantry Separation Unit Leveling

The level must be placed in the same location as it was previously measured for a consistent reading. It is
possible that the X-axis leveling may have altered your initial adjustments.

NOTE If it is necessary to make multiple adjustments with the leveling screws, be aware of the height of the separation
unit in relation to the floor template foam and the patient table. If you raise the separation unit too high in the
attempt to level it, you may create a gap between the bottom of the unit and the foam which will allow the
leveling compound to leak out.

Mobile Cabling
To ensure easy access, prior to positioning the gantries, install the Lock/Unlock Switch to GSU Cable at this time.

PROCEDURE

1 From the GEMINI Mobile Parts Kit, obtain the Lock/Unlock


Sw to GSU Cable (4535 679 71841).
2 From the front end of the GSU (table side), feed cable
through the cabling channel as shown.
Connector end facing table.
3 Refer to Page 138 for cabling connections.

Cabling
Channel

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 70
GEMINI GXL Installation Technical Manual Patient Table Mechanical Alignment and Leveling

Patient Table Mechanical Alignment and Leveling

Personnel Required Time Required

1 Field Service Engineer 2 Hours

Overview
In the following steps, you will position the Patient Table and then level it prior to installation. The Patient Table
and the GSU must be set to the same plane. The Patient Table Alignment Fixture will be used to help accomplish
this.
The Patient Table Alignment Fixture offers no automatic left /right positioning reference. Any left/right centering
must use the centerline scribed onto the floor for reference. Use the access holes in the fixture for viewing the
centerline. The fixture ensures proper critical spacing between the gantry and the table and also allows the
proper vertical orientation (planar offset) to be established.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 71
GEMINI GXL Installation Technical Manual Patient Table Mechanical Alignment and Leveling

Preparing the Patient Table

PROCEDURE

1 Place the Patient Table Alignment Fixture on the front of the CT


Set Screw
gantry.
2 Lower the set screw to hold up the CT-end of the fixture.
3 Center the Alignment Fixture on the centerline of the system.

Patient Table
Alignment Fixture
Front of CT Gantry

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 72
GEMINI GXL Installation Technical Manual Patient Table Mechanical Alignment and Leveling

WARNING THE TABLE INSTALLATION CART IS DESIGNED TO ALLOW THE TABLE


ASSEMBLY TO BE RAISED AND TILTED TO ALLOW TRANSPORT THROUGH AN
ELEVATOR OF SHALLOW DEPTH. THE INSTALLATION CART IS NOT MEANT TO
TRANSPORT THE TABLE IN THE RAISED POSITION. THE TABLE MUST BE LOWERED DURING ALL
OTHER PHASES OF INSTALLATION AND TRANSPORT. FAILURE TO COMPLY MAY RESULT IN
EQUIPMENT DAMAGE, SERIOUS PHYSICAL INJURY, OR DEATH.

WARNING DO NOT ATTEMPT TO MOVE THE PATIENT TABLE WITHOUT THE ASSISTANCE
OF ANOTHER FSE. THE TABLE IS HEAVY AND COULD BECOME UNSTABLE WHILE BEING MOVED.
FAILURE TO COMPLY MAY RESULT IN SERIOUS INJURY.

4 Roll the Patient Table into the room.


5 Maneuver the Patient Table into approximate position, about 3 feet from the CT gantry.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 73
GEMINI GXL Installation Technical Manual Patient Table Mechanical Alignment and Leveling

NOTE Use the centerline of the system and


the witness marks on the Patient Table
baseplate as the guide for positioning
the Patient Table Transport.

Patient Table Baseplate


Alignment Fixture

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 74
GEMINI GXL Installation Technical Manual Patient Table Mechanical Alignment and Leveling

6 Lift the end of the Alignment Fixture that is closest to the Patient Table and push the Table Transport so that
the Patient Table Baseplate is under the Alignment Fixture.

CT Gantry Alignment Fixture

Patient Table Baseplate

NOTE It may be necessary to lower the Table Transport to allow the Alignment Fixture to rest on the Patient Table
baseplate. If this becomes necessary, do not allow any part of the Patient Table to contact the floor as this will
make moving the Patient Table very difficult.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 75
GEMINI GXL Installation Technical Manual Patient Table Mechanical Alignment and Leveling

7 Seat the mounting pins of the Alignment fixture in the mounting holes of the Patient Table baseplate.

Alignment Fixture

NOTE Ensure that the Mounting Pins are seated firmly in the mounting holes of the Patient Table baseplate.
This will allow the Patient Table to be properly mechanically aligned and levelled.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 76
GEMINI GXL Installation Technical Manual Patient Table Mechanical Alignment and Leveling

Leveling the Patient Table


8 Locate the three table leveling screws.
9 Remove each screw, thoroughly coat the threads
with petroleum jelly, and then reinstall.
The table has three leveling screws threaded into
the base plate. Two of the screws are located at the
head end of the table, just inboard of the vertical
drive assembly on the left and right sides. The third
leveling screw can be found behind the foot end
vertical drive assembly, at the center of the base
plate.

Foot End Table Leveling Head End Table Leveling


Screw (Qty. 1) Screws (Qty. 2)

CAUTIO Protective shims prevent the table leveling screws from causing damage to the scan
room floor. Ensure that the protective shims are placed properly beneath the table leveling
screws before allowing them to make contact with the floor. Failure to comply may cause
permanent damage to the scan room floor.
10 Insert protective shims underneath the Patient Table baseplate, between the bottom of the leveling screws
and the floor.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 77
GEMINI GXL Installation Technical Manual Patient Table Mechanical Alignment and Leveling

Patient Table Baseplate


Front

Table Leveling Screw (Qty. 3)

Table Floor Shims (Qty. 3)

NOTE The shims used under the leveling screws are universal; the slot serves no purpose in this application.

NOTE It will be necessary to utilize a long screwdriver or similar tool to slide the shim under the front center leveling
screw.
11 Lower the Patient Table transport so that the entire weight of the Patient Table is on the three leveling screws.
12 Remove the Patient Table transport from the Patient Table.
13 Lower the Patient Table, using the leveling screws, to approximately 1/4” (6 mm) above the floor.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 78
GEMINI GXL Installation Technical Manual Patient Table Mechanical Alignment and Leveling

NOTE Whenever instructed to use the Precision Level, it is important to mark the exact location on which it is placed using
a pencil. The level must be placed in the exact same location each time it is used to yield accurate and reliable
readings.
14 Level the Patient Table in the y-axis.
a Place the precision level lengthwise (approximately Precision Label Patient Table Baseplate
centered along the table baseplate.
b Mark the outline of the precision level on the table
baseplate with a pencil. This will ensure that the level
is placed in the exact same position for future
measurements.
c Adjust the leveling screws, using quarter-turn
increments.

Y-Axis

d Continue adjusting the Patient Table until the


Precision Level indicates level, ±1 division.

Center Marks

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 79
GEMINI GXL Installation Technical Manual Patient Table Mechanical Alignment and Leveling

15 Level the Patient Table in the x-axis.


a Push the Patient Table Top Pallet forward slightly. Precision Level and Table Alignment
b Place the GEMINI Table Alignment Fixture (p/n 4535 679 Fixture
69971) across the table rails in the x-axis.
c Place the Precision Level on the GEMINI Table Alignment
Fixture Pins.
d Note and mark the exact location of the level for future
measurements.
e Adjust the two front leveling screws as necessary until the
precision level bubble is centered. (X-adjustments are
made solely with the two rear leveling screws.)

X-Axis

NOTE The table is now level in both axes. However, the height in relation to the gantries is not yet set. The alignment
fixture must be used to establish table height and position requirements.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 80
GEMINI GXL Installation Technical Manual Patient Table Mechanical Alignment and Leveling

16 Level the Alignment Fixture.


a Place the precision level along the length of the
alignment fixture (front to back) as close to the center
of the alignment fixture as possible.
b Note and mark the location of the level so that
future measurements may be made from the exact
same location
c Adjust all three of the table leveling screws an
equal amount until the level bubble is centered (±1
division).
17 Repeat steps13 through 15 as necessary to maintain
level in each of the positions.

Alignment Fixture

NOTE It is important when repeating the steps, to ensure that the Precision Level is placed in the exact same location as
it was placed for the previous measurement. This will ensure an accurate measurement for levelness.

NOTE Once all three levelness checks are satisfactory, the Patient Table can now be considered mechanically
level. Proceed to Electrical Interconnect.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 81
GEMINI GXL Installation Technical Manual Pouring the Micorox Compound-GSU

Pouring the Micorox Compound-GSU

Overview
The following steps outline the procedure for mixing and pouring the Micorox compound. The Micorox is a self-
leveling two-component epoxy compound used to fill the volume between the floor and the leveled GSU. You will
be required to work against the cure time of the Micorox compound and as such, must make certain you are ready
to use it and will be able to work continuously without interruption. Exercise caution when mixing and pouring the
compound to prevent splashing, as damage to the floor, walls, and cabinets may occur. Wear disposable gloves at
all times and follow the manufacturer's specific instructions for cleaning and skin contact should that occur. Once
the Micorox compound has fully cured, the GSU can be anchored to the site floor.

Personnel Required Time Required

1 Hour to perform procedure


1 Field Service Engineer 8 Hours minimum for Micorox to cure

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 82
GEMINI GXL Installation Technical Manual Pouring the Micorox Compound-GSU

Pouring the Micorox Compound – GSU


PROCEDURE

18 Locate and retrieve the following in preparation of mixing and pouring the Micorox compound:
• Electric drill motor with 1/2” chuck
• (2) 25 lb. (6.8kg) cans of Micorox epoxy base
• (2) 1 qt. (0.95L) cans of Micorox hardener
• Disposable gloves
• Mixing paddle
• Plastic, paper, or cardboard sheets
• Tape
19 Protect the floor around the separation unit.
HINT Tape paper, cardboard, or plastic sheet to the floor around the outside of the separation unit. Use as little tape
as possible, to avoid leaving tape or adhesive behind on the floor of the facility.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 83
GEMINI GXL Installation Technical Manual Pouring the Micorox Compound-GSU

20 Mix the two components of the floor pour epoxy, one pail at a time.

Mixing the Epoxy Base and Hardener

NOTE The epoxy base and hardener must be at room temperature prior to use.
a Place one 15 lb. (6.8 kg) pail of epoxy base on a piece of heavy cardboard or similar protection.
b Install the mixing paddle in the drill chuck.
Each side of the GSU will require approximately one pail of the Micorox compound. However, for ease of pouring,
it is recommended that only one pail be mixed at a time.

CAUTION Do not mix the epoxy while the pail is sitting directly on the floor. When the two components of
the epoxy are mixed, an exothermic reaction occurs which will heat the metal pail and may
damage the floor of the facility. Place a piece of cardboard or similar material under the pail to
protect the floor.
c Put on a pair of latex gloves and remove the lids from one pail of epoxy base and one container of
hardener.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 84
GEMINI GXL Installation Technical Manual Pouring the Micorox Compound-GSU

d Pour the entire contents of the hardener container into the epoxy base pail.

e Insert the mixing paddle and blend for a minimum of 10 minutes.

21 Set the mixer aside.


a Remove the mixer from the pail slowly and place it on a piece of cardboard or paper.
b Do not allow any of the epoxy to drip onto the separation unit or the surrounding floor outside of the
template.
• If any spillage does occur, immediately clean it up with rags or paper towels.

CAUTION Exercise caution so that the epoxy does not drip on the floor. If dripping or spilling does occur,
ensure that you clean it up immediately. Failure to do so may cause permanent damage to the
flooring.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 85
GEMINI GXL Installation Technical Manual Pouring the Micorox Compound-GSU

22 Begin pouring the mixed Micorox into one side of the GSU floor
template.
a Start at either end and slowly pour the epoxy into the floor
template at one of the pour zones. Check for leaks while GSU Pour Zones
pouring; these can be plugged using a small amount of
petroleum jelly.
b Monitor the viewing holes along the side of the GSU. As the area
in which you are pouring begins to fill, move to the center of the
template, and then finally to the other end of the template.
c Alternate between the two end and center pour zones until the
Micorox is approximately 1/8" (3 mm) from the top edge of the
viewing holes. Do not overfill.
d While pouring one side, instruct another FSE to begin preparing
and mixing another pail of the epoxy.

CAUTION Exercise caution so that the epoxy does not drip on the floor. If dripping or spilling does occur,
ensure that you clean it up immediately. Failure to do so may cause permanent damage to the
flooring.
The GSU floor template has three troughs that protrude from each side. These troughs are called pour zones.
These pour zones will be removed once the Micorox has cured later in the procedure. It will be necessary to
move the PET gantry slightly to access the center pour zones.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 86
GEMINI GXL Installation Technical Manual Pouring the Micorox Compound-GSU

23 Fill the remaining side of the GSU floor template.


a Start at either end and slowly pour the epoxy into
the floor template at one of the pour zones.
b Monitor the viewing holes along the side of the
GSU. As the area in which you are pouring begins
to fill, move to the center of the template, and then
finally to the other end of the template.
c Alternate between the two end and center pour
zones until the Micorox is approximately 1/8" (3 mm)
from the top edge of the viewing holes. Do not
overfill.
d Allow the Micorox in both sides of the GSU to settle out for twenty to thirty minutes undisturbed. If additional
epoxy is required, it should be poured at this time.
Use the second pail of Micorox to fill the remaining side of the GSU.

CAUTION Exercise caution so that the epoxy does not drip on the floor. If dripping or spilling does occur,
ensure that you clean it up quickly. Failure to do so may cause permanent damage to the flooring.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 87
GEMINI GXL Installation Technical Manual Pouring the Micorox Compound-GSU

24 Insert the pour zone barrier strips.


a Coat both sides of the barrier strips with petroleum jelly.
b Insert one of the barrier strips into each side of the floor
template as shown.
Pour Zones
The barrier strips allow the six pour zones to be removed
Barrier Strips
from the sides of the GSU once the Micorox has cured.

CAUTION Ensure that all partially used cans of Micorox are


kept well ventilated and away from combustibles.
The exothermic reaction produces significant fumes
and heat. Partially used cans should be placed on a
concrete floor or similar until the compound has
fully cured. Failure to comply may result in damage
to the site floors and/or equipment.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 88
GEMINI GXL Installation Technical Manual Anchoring the GSU

Anchoring the GSU

Personnel Required Time Required

1 Field Service Engineer 1 Hours


1 Retrieve the following equipment in preparation for drilling the floor
anchor holes.
• Hammer Drill
• Masonry Drill Bit
• Vacuum
• Blow Bulb
• Hammer Hammer Drill With Depth
• 3/4” Wrench Marking Tape
• Floor Anchor Bolts
2 Insert and secure the masonry drill bit into the drill chuck.
• If the drill you are using is so equipped, set the drill depth gauge to 4 1/2” (11.5 cm) from the top of the
spot face holes in the GSU.
• If the drill you are using is not equipped with a depth gauge, measure 4 1/2” (11.5 cm) from the tip of the
drill bit, and place a piece of tape around the bit at that point.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 89
GEMINI GXL Installation Technical Manual Anchoring the GSU

3 Drill two anchor holes through the GSU into the floor.
a Select two opposite corners of the GSU in which to install the first two floor
anchors.
b Drill through the selected spot faced holes into the floor using the drill and
masonry bit. Keep in mind there is a bare minimum anchor requirement for
the GSU, which stipulates that there be no less than four anchors per side
and no two consecutive holes without anchors.

NOTE If you hit rebar while drilling, select the next nearest spot faced hole and use it to set the floor anchor into. Ensure
that the substitute anchor hole complies with the bare minimum anchor requirement explained above.
c Keep the vacuum on and positioned near the drill bit while drilling to extract dust and debris.
d Continue to drill until the tape on the drill bit is even with the top of the GSU spot faced hole or the depth
gauge indicates 4 1/2” (11.5 cm).
4 Install two anchor bolts into the drilled holes.
a Use the blow bulb and vacuum again prior to installing the anchor bolts to ensure that both of the drilled
holes are clean and free of debris.
b Insert the anchor bolts into the holes, and tap them down flush with the separation unit using a
hammer.
c Tighten the anchor bolts using a 3/4” socket or wrench.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 90
GEMINI GXL Installation Technical Manual Anchoring the GSU

WARNING FAILURE TO THOROUGHLY REMOVE DEBRIS FROM THE ANCHOR HOLES WILL
PREVENT THE ANCHORS FROM BEING INSTALLED AT THE SPECIFIED DEPTH. IF
THE ANCHORS ARE NOT INSTALLED AT THE SPECIFIED DEPTH, THE EQUIPMENT WILL NOT BE
PROPERLY SECURED TO THE FLOOR AND SERIOUS INJURY OR DEATH MAY RESULT.
5 Pull the barrier strips up out of the epoxy on both sides of the separation unit.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 91
GEMINI GXL Installation Technical Manual Anchoring the GSU

6 Remove the six pour zones from the outside of the floor
template.
a Using a hammer and screwdriver, tap and/or pry the
six rectangular fill areas away from the sides of the
floor template. b Discard the pieces of epoxy and
foam as they are removed.

Pour Zone (Cured)

CAUTION When removing the pour zones from the rest of the GSU floor template, use caution that the tools
you are using do not damage the floor. Failure to exercise caution may result in significant damage
to the facility flooring.
7 Clean up the edges of the cured Micorox slab.
• Trim any excess foam or jagged epoxy from the base of the
separation unit using a razor knife.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 92
GEMINI GXL Installation Technical Manual Anchoring the GSU

The object is to make the outside edge of the poured epoxy look as clean and finished as possible.

Cured Micorox Slab


8 Drill the remaining GSU anchor holes.
Installation of the GSU floor anchors requires that a minimum of four
anchors be installed per side. Under no circumstances should two
consecutive anchor locations be without floor anchors installed. a
Select the spot faced holes along both sides of the GSU.
b Using the drill and masonry bit, drill through the selected spot
faced holes into the floor.
NOTE If you hit rebar while drilling, select the next nearest spot faced
hole and use it to set the floor anchor into.
c Keep the vacuum on and positioned near the drill bit while drilling
to extract dust and debris.
d Pause frequently, remove the drill bit from the hole(s), and use the
blow bulb and vacuum to keep the inside of the hole(s) free of
debris.
e Continue to drill until the tape on the drill bit is even with the top of the separation unit spot face or the depth
gauge indicates 4 1/2” (11.5 cm).

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 93
GEMINI GXL Installation Technical Manual Anchoring the GSU

9 Install the remaining anchor bolts.


a Use the blow bulb and vacuum again prior to installing the
anchor bolts to ensure that each drilled hole is clean and
free of debris.
b Insert the anchor bolts into the holes, and tap them down
flush with the GSU using a hammer.
c Tighten the anchor bolts using a 3/4” wrench.
Installation of the GSU floor anchors requires that a minimum of
four anchors be installed per side. Under no circumstances
should two consecutive anchor locations be without floor
anchors installed.
10 Clean up the work area around the separation unit.
a Pick up the paper and cardboard used to protect the floor
around the GSU.
b Remove any tape left on the floor, and vacuum up any
remaining dust or debris.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 94
GEMINI GXL Installation Technical Manual PET Gantry Mounting

PET Gantry Mounting

Overview
The proper mounting of the PET gantry to the GSU is the first step of a procedure that will result in the alignment
of the PET and CT gantries. The frame of the PET gantry has a certain amount of deflection under the weight of
the detector and shims must be used to compensate for this. The PET gantry must be installed so that it is
perpendicular to the linear bearings of the GSU.

Personnel Required Time Required

2 Field Service Engineers 4 Hours (3 Lead and 1 Assistant)


Positioning the PET Gantry
PROCEDURE

1 Roll the PET gantry into position over the separation unit mounting plate.
• Move the PET gantry so that the two 1/2” (13 mm) diameter holes at the
bottom of the gantry base are positioned over the two pins in the GSU
mounting plate.

NOTE The PET gantry cannot be raised high enough to clear the GSU drive motor
assembly. Therefore, the PET gantry must be rolled into position over the GSU from the front as shown.
2 Lower the PET gantry onto the locating pins. Lower Each Side of the Gantry

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 95
GEMINI GXL Installation Technical Manual PET Gantry Mounting

onto the Locating Pins


• Slowly lower both sides of the gantry until it is resting

on the mounting plate locating pins.

Locating Pins
(Qty. 2)

3 Remove the Roll-Lifts from the gantry.


a Loosen and remove each of the four bolts from the
Roll-Lifts.
b Secure the bolts back into the Roll-Lift forks for return shipping.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 96
GEMINI GXL Installation Technical Manual PET Gantry Mounting

4 Install the PET gantry to GSU mounting screws.


• Loosely install one of the eight supplied 1/2-13 x
1 1/4” hex head screws into each of the holes
but do not tighten.
On the underside of the GSU mounting plate
there are eight clearance holes, two on each of
the four mounting pads. The eight 1/2-13 x 1 1/4”
hex head screws (Grade 8) are inserted through
the GSU mounting plate up into the PET base
plate.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 97
GEMINI GXL Installation Technical Manual PET Gantry Mounting

5 Measure the angle of offset at the GSU.


Precision Level

a Place the precision level at a fixed location on the solid GSU


rail (not the Encoder rail).

CAUTION Do not place any magnetic objects on the encoder rail. Magnetic objects will damage or interfere
with the proper operation of the encoder.
b Place the appropriate thickness of feeler gauge between the precision level and the GSU rail until the
precision level indicates level.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 98
GEMINI GXL Installation Technical Manual PET Gantry Mounting

6 Leveling the PET gantry.


a Retrieve the precision level and place it against the PET gantry front
plate (near the center of the bore).
b Adjust the PET Gantry Leveling Bolts until the precision level
indicates level.
c Mark the front of the gantry with a marker or piece of tape indicating
the point where you placed the level.

NOTE It is important that you mark the location where you placed the level on the PET gantry front plate.
Subsequent checks of the gantry using the digital level must be done in the same location.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 99
GEMINI GXL Installation Technical Manual PET Gantry Mounting

d Adjust the Left and Right PET Gantry Leveling


Bolts simultaneously until the precision level
indicates level.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 100
GEMINI GXL Installation Technical Manual PET Gantry Mounting

7 Level the PET Gantry relative to the GSU.


a Place the feeler gauge determined in the previous step between the precision level and the PET
Gantry Front Plate at the location previously marked
with tape. PETGantry
b Adjust the Left and Right PET Gantry Leveling Bolts
simultaneously until the precision level indicates Precision Level
level. 8 Install shims between the PET and GSU
mounting plate:
Feeler Gauge
a Once the PET gantry is level measure the gap
between the GSU mounting plate and the base of
the PET gantry using feeler gauges.

Feeler Gauge

Precision Level

Precision Level

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 101
GEMINI GXL Installation Technical Manual PET Gantry Mounting

NOTE The illustration on this page is shown greatly exaggerated for clarity. The goal is to make the PET gantry
perpendicular to the linear bearing ways.
b Insert the shims between the bottom of the PET gantry and the top of the mounting plate; equal sets on
both the left and right sides of the gantry. The PET gantry must also be level from side to side, to ensure it
is level in both directions.
c Ensure that the shims are installed under the PET gantry as far as they will go.
d Lower the PET Gantry, using the Left and Right Leveling Bolts simultaneously, until all weight is off of the
Leveling Bolts and on the shims.
Use the plastic calipers to measure the proper amount of shims. Two equal sets of shims are required, one for
the left and right sides of the gantry.
Install enough shims to take up the space between the PET Gantry and the GSU mounting plate.
9 Re-check the PET gantry for level relative to the GSU.
a Tighten the eight 1/2-13 x 1 1/4” hex head screws under the GSU mounting plate.
b Verify that the PET gantry remains level relative to the GSU after the eight 1/2-13 x 1 1/4” hex head screws
have been tightened. If it remains level, proceed to the next step, c. If it does not, you will have to go back to
Step 8 and add or subtract the appropriate amount of shims.
c Remove the two PET Gantry Leveling Bolts.
NOTE Only check the gantry for level when the eight mounting screws located under the GSU mounting plate have been installed
and tightened securely. This ensures that the shims are completely compressed.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 102
GEMINI GXL Installation Technical Manual
CT Gantry Mounting

CT Gantry Mounting
Personnel Required Time Required

2 Field Service Engineers 5 Man-Hours (4 Lead and 1 Assistant)

Positioning the CT Gantry


PROCEDURE

1 Locate and identify the two 3/4” (19 mm) diameter pins at the front of the GSU.
2 Locate and identify the two 3/4” (19 mm) diameter holes in 3/4” (19 mm)
the CT gantry base plate. Diameter Pins
These holes correspond with the two ¾” (19mm) diameter pins at the front of
the GSU

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 103
GEMINI GXL Installation Technical Manual
3 Position the CT gantry over the GSU.
CT Gantry
a Align the pair of 3/4” (19 mm) diameter holes in
the CT gantry base plate with the corresponding
pins in the GSU.
b Lower the CT gantry until the bottom of the
gantry base plate is approximately 1/8” (3 mm)
below the top of the 3/4” (19 mm) diameter pins.

GSU 3/4” (19 mm)


Diameter Holes

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 104
GEMINI GXL Installation Technical Manual

Gantry Mechanical Alignment

Personnel Required Time Required

2 Field Service Engineers 2 Hours

Overview
The PET and CT gantries must be mechanically aligned before installation can be completed. To check and verify
this alignment, a laser and mirror are used to indicate the position and angles of the gantries.
To ensure maximum safety and alignment accuracy, be sure to read through the entire procedure before
completing the steps as instructed.
Additionally, the laser and mirror must be checked for accuracy using the calibration procedures as described in
Testing/Calibrating the Laser and Mirror prior to use.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 105
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

Testing/Calibrating the Laser and Mirror


You can determine if the laser and mirror assemblies require calibration by observing the shape drawn by the
beam as the assemblies are rotated. If any shape other than a circle is drawn, then there is likely a problem with
debris allowing the assemblies to rock, or there is debris between the components and the mounting surfaces.
Therefore, it is important that the assemblies be placed on a clean and sturdy surface to prevent rocking.

Calibrating the Laser


For the most accurate results, find a place where the ceiling is 10 feet or higher from the laser source. The farther
the distance the more accurate the adjustment will be.
1 Place the laser assembly on the floor, a table, or countertop.
Ensure that the surface is clean and free of debris.
It is important that the surface is level enough to prevent the assembly from rocking back and forth. Dirt and or
debris under the assembly may cause it to rock and affect the pattern drawn by the beam.
NOTE It is common to place these components on a table or countertop during calibration. If someone were to lean on
the table or countertop during calibrations they may compromise the calibration process and the adjustments will
need to be repeated.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 106
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

2 Remove the three mounting screws, noting the


installation height of the screws before removal. Also
note the associated hardware.
Once the three mounting screws have been removed,
the laser mounting plate is free to swivel on the
gimbal.

WARNING DO NOT LOOK DIRECTLY INTO THE LASER WHEN IT IS POWERED ON. FAILURE TO COMPLY MAY RESULT IN
SERIOUS EYE INJURY AND VISION IMPAIRMENT.
3 Turn the laser on. Observe the beam projected onto the ceiling.
NOTE The laser is a battery powered device and will quickly deplete the batteries if left on. To prevent the laser from
becoming non-functional, always turn the laser off when not using it.
4 Verify that the projected beam displays a round dot on the ceiling.
• Carefully spin the laser around on the mounting plate and observe the dot on the ceiling.
Adjust the three screws until the beam projected on the ceiling remains in a fixed position despite the rotation
of the laser assembly as described in Step 5.
5 To adjust the laser:
a Locate the adjusting hardware.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 107
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

NOTE The adjusting screws have a relatively small amount of travel. Should you bottom out any adjusting screws the
results will be unpredictable. This could include, but is not limited to, adjusting a screw without having an effect
on the beam position.
b As you rotate the laser assembly it will draw a circle on the ceiling. Once you determine where the beam is
stop rotating the laser and tighten the screw farthest away from where the beam is positioned.
As you continue to tighten the correct screws the circle diameter will get smaller and smaller until it is zero.
6 Reinstall the laser onto the assembly using the original hardware.
Install the hardware so that each screw is approximately the same height as it was before it was removed.
NOTE If mirror calibration is required, do not reinstall the laser assembly at this time, as the laser is used to
calibrate the mirror.

Calibrating the Mirror


PROCEDURE

1 Unscrew the mirror cap and then remove the cap and lense.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 108
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

CAUTION Do not touch the mirror surface. Fingerprints, dirt, or grease will prevent the laser from reflecting
properly.
2 Place the mirror assembly on the floor or a table.
• Ensure that the surface is clean and free of debris.
It is important that the surface is level enough to prevent the assembly from rocking back and forth. Dirt and or
debris under the assembly may cause it to rock and affect the pattern drawn by the beam.
NOTE It is common to place these components on a table or countertop during calibration. If someone were
to lean on the table or countertop during calibrations they may compromise the calibration process and the
adjustments will need to be repeated.
3 Position the laser above the mirror so that the beam (once turned on) will be directed towards the center of the
mirror.
A stack of books or similar device works well. The laser should be oriented so that the beam (once turned on) is
as close to the center of the mirror as possible.
4 Ensure that the distance between the mirror and wall is a minimum of 8’ (2.4 m). Move the laser and mirror if
necessary to achieve this distance.
The accuracy of this adjustment increases with the distance of the mirror to the wall. There should be no less
than 8’ (2.4 m), but greater than 8’ (2.4 m) is preferred.
5 Remove the three mounting screws from the mirror, noting the installation height of the screws before removal.
Also note the associated hardware.
Once the three screws have been removed, the mirror to will be free to rotate on the gimbal.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 109
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

6 Turn the laser on.

Mirror Adjustment
Screws (Qty. 3)

Mirror Adjustment
Screws (Qty. 3)

NOTE The laser is a battery powered device and will quickly deplete the batteries if left on. To prevent the laser from
becoming non-functional, always turn the laser off when not using it.
7 Rotate the mirror carefully while
noting the position of the beam on
the wall.
When properly calibrated, this beam will
remain in a fixed location despite the
rotary motion. Recalibration is
necessary if the beam moves in a
circular fashion on the wall while the
mirror is being rotated.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 110
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

8 Choose one of following


If the dot: Then go to:

Moves in a circular fashion The mirror must be calibrated. Proceed to Step 9.

The mirror is adjusted properly. Proceed to Installing the Laser and Mirror
Remains steady Assemblies

9 Rotate the mirror noting the size and location of the circular path the beam creates.
10 Tighten the adjusting hardware until the beam moves into the center of the circular path.
11 Repeat the adjustments until the beam remains at a single point despite the rotating mirror assembly.
12 Proceed to Installing the Laser and Mirror Assemblies once the mirror assembly has been calibrated.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 111
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

Installing the Laser and Mirror Assemblies


1 Open the front upper cover on the PET gantry.
2 Install the Laser Assembly to the PET Gantry using the 1/4 in. dowel pins and the 1/4-20 x 1” SHCS with Lock
and Flat washers.
Ensure that the laser has been turned off. Install hand tight only. Do
not over-torque.

NOTE The laser is a battery powered device and will quickly deplete the batteries if left on. To prevent the laser from
becoming non-functional, always turn the laser off when not in use.
3 Lower (close) the front upper PET gantry cover.
4 Open the rear upper cover on the CT gantry.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 112
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

5 Install the mirror assembly to the CT Gantry using


the 5/16” dowel pins and the 1/4-20 x 1” SHCS with Mirror Assembly
Lock and Flat washers.
6 Lower (close) the rear upper cover on the CT
gantry.
Install hand tight only. Do not over-torque.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 113
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

Verifying PET Gantry Alignment to Rails


PROCEDURE

1 Turn on laser and note beam location.

WARNING DO NOT LOOK DIRECTLY INTO THE LASER WHEN IT IS POWERED ON. FAILURE TO COMPLY MAY RESULT IN
SERIOUS EYE INJURY AND VISION IMPAIRMENT.
Leave the mirror lens on.
2 Move the PET Gantry on the GSU throughout the entire range of travel.
If the laser beam projected onto the mirror Assembly moves vertically more than 1 mm (.04”), then the PET Gantry
has not been shimmed properly (relative to the PET vertical frame plate). Refer to PET Gantry Mounting for further
information. Horizontal motion is allowed at this time.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 114
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

CT Gantry Parallelism (Relative to the PET Gantry)


PROCEDURE

1 Move the PET Gantry to the open position.


2 Remove the “puck” from the mirror assembly to expose the entire mirror.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 115
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

3 Level the CT Gantry base plate so that the laser beam is reflected back onto itself.

a Adjust the CT gantry until the laser is reflected back onto


itself within 1 mm (.04”) in all directions on the Laser
Assembly face.
Use the four CT Gantry Shipping Wheels for initial leveling.
The CT Gantry is initially positioned slightly low from final
position. Any leveling will raise the Gantry slightly up off of
the five leveling pads.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 116
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

Center Axis Alignment


PROCEDURE

1 Replace the lens in the mirror assembly.


The center hole in the lens is the Laser target.
2 Adjust the height of the CT Gantry until the laser is
within 1 mm (.04”) vertically of the “puck” center hole.
Adjust the four CT Gantry Shipping Wheels, adjusted
equally to maintain parallelism, to attain the correct laser
height.
3 Adjust the left to right level of the CT Gantry until the
laser is within 1 mm (.04”) horizontally of the “puck”
center hole.

NOTE Lowering the right side of the CT Gantry will move the CT Gantry Center Point to the right. Raising the right side of the
CT Gantry will move the Center Point to the left. The opposite is true for the left side.

Adjust the CT Gantry Shipping Wheels in left or right pairs to attain the correct horizontal laser placement.
When raising or lowering one side of the CT Gantry, the Center Point will also be raised or lowered. An overall
height adjustment may be necessary after any angular adjustments have been made.
4 Tighten the five Gantry Jack Screws, located in the CT Gantry base plate, until they make contact with the
floor.
Do not over-tighten the Jack Screws as this can affect the Gantry level.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 117
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

CAUTION Do not allow the CT gantry to initiate rotor spin or attempt to turn the rotor by hand unless the
front center leveling pad has been lowered first. The adjusting screw will not clear the rotor if the
front center leveling pad is left in the raised position and damage to the CT subassemblies will
result.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 118
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

Alternate CT Leveling Procedure


PROCEDURE

1 Move the PET gantry to the open position.


2 Lower the CT gantry to the lowest position.
Use the CT gantry wheel adjusters to lower the gantry to the lowest possible position.
3 Turn the laser on.
4 Place a bubble level on the laser mirror mounting plate.
5 Raise the CT gantry while trying to maintain level.
Use the CT gantry wheel adjusters to raise the gantry, ensuring that the bubble level reads level. As the gantry
moves upwards, the laser dot will begin to move closer to the hole in the center of the mirror "puck".
6 Stop raising the CT gantry as soon as the laser dot is visible on the same horizontal plane as the hole in the
mirror "puck".
The laser dot should be on the same level as the hole in the center of the mirror "puck", but will likely be either
to the left or right of center.
7 Adjust the sides of the CT gantry as follows:
a To move the laser dot to the left, adjust both right side wheels.
b To move the laser dot to the right, adjust both left side wheels.
Depending on the position of the laser dot, you will have to adjust either both left side or both right side gantry
wheels. Once you have selected which side, adjust the wheels until the laser dot is aligned horizontally with
the hole in the center of the mirror "puck".
8 Adjust the CT gantry to vertically center the laser dot with the hole in the mirror "puck".
• Raise either BOTH front or BOTH rear wheels until the laser dot is centered with the hole in the center of
the mirror "puck".

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 119
GEMINI GXL Installation Technical Manual Gantry Mechanical Alignment

System Retest
PROCEDURE

1 Position the PET Gantry in the open position.


2 Turn on the laser.
The laser beam must reflect off of the hole in the mirror Assembly “puck” back to the laser within 2 mm (.080”)
in all directions.
3 Move the PET Gantry to the closed position.
The laser beam must reflect off of the hole in the mirror Assembly “puck” back to the laser within 1 mm (.040”)
in all directions.
4 Repeat the appropriate sections of the alignment procedure if the above specifications are not met. If the
specifications have been met, proceed to Electrical Interconnect.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 120
GEMINI GXL Installation Technical Manual Electrical Interconnect

Electrical Interconnect

WARNING HAZARDOUS VOLTAGE PRESENT. FAILURE TO INSURE THAT POWER IS


TURNED OFF AT THE SOURCE CAN RESULT IN SERIOUS BODILY INJURY,
DEATH, AND SERIOUS EQUIPMENT DAMAGE. TURN OFF THE ELECTRIC POWER SUPPLY AT THE
SOURCE BEFORE PERFORMING ANY OF THE PROCEDURES DETAILED IN THIS SECTION.

Personnel Required Time Required

2 Field Service Engineers 14 Hours (10 Lead and 4 Assistant)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 121
GEMINI GXL Installation Technical Manual Electrical Interconnect

Fixed and Fixed Mobile Sites Wall-Mounted Panel


1 Apply the appropriate colored tape to both ends of the corresponding
incoming power cables.
2 Complete the incoming power to wall-mounted panel connections.
• Connect the three 480V cables to the corresponding terminals.
• Connect the ground wire to the proper terminal.

480 V cables (Marked Brown, Orange


and Yellow

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 122
GEMINI GXL Installation Technical Manual Electrical Interconnect

Mobile Wall-Mounted Panel


1 Complete the incoming power to wall-mounted
panel connections.
• Connect the three 480V cables to the
corresponding terminals as shown.
• Connect the ground wire to the proper
terminal as shown.

480V cables Ground


Connections

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 123
GEMINI GXL Installation Technical Manual Electrical Interconnect

TEAL Power Unit


1 Loosen and remove the screws securing the access cover to the top
of the TEAL Power Unit.Access Panel

TEAL Power Unit

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 124
GEMINI GXL Installation Technical Manual Electrical Interconnect

2 Route the Power Cable bundle from the


wallbox to the back of the TEAL Power Unit. Wallbox Power Cables
3 Bring the power cables and ground wire up
to the Input connection points on the inside
of the TEAL cabinet.

TEAL Power Unit

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 125
GEMINI GXL Installation Technical Manual Electrical Interconnect

4 Connect the Power Cables from the


wallbox to the Input connection points
inside the TEAL Power Unit.

Input Power Connection

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 126
GEMINI GXL Installation Technical Manual Electrical Interconnect

5 Route one end of the Main Power Conduit Main Power Conduit
to the back of the TEAL Power Unit.
6 Bring the power cables and ground wire up to
the Output connection points on the inside of
the TEAL cabinet.

TEAL Power Unit

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 127
GEMINI GXL Installation Technical Manual Electrical Interconnect

7 Connect the Power Cables from the Main


Power Conduit to the Output connection
points inside the TEAL Power Unit.

Main Power Conduit


Connection Points

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 128
GEMINI GXL Installation Technical Manual Electrical Interconnect

GSU Electrical Connections


PROCEDURE

1 Connect the optional Scan room "Door


Interlock" wiring to TB30 on the GSU
baseplate (normally closed contacts).

GSU Baseplate

TB30
Suite Door
Switch Wiring

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 129
GEMINI GXL Installation Technical Manual Electrical Interconnect

2 Connect the X-Ray On wiring harness to TB20 on the GSU baseplate.

TB20

X-Ray On Wiring Harness

GSU Baseplate

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 130
GEMINI GXL Installation Technical Manual Electrical Interconnect

3 Connect the dual fiber optic cable from the PET Server to the Bulkhead Bracket mounted to the GSU
baseplate.

Dual Fiber Optic Cable

GSU Baseplate

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 131
GEMINI GXL Installation Technical Manual Electrical Interconnect

4 Connect the 15-pin Acquisitor cable from the


CT Recon cabinet to the Bulkhead
Bracket mounted to
the GSU baseplate.

15-Pin Acquisitor Cable (Located


Under the Dual Fiber Optic Cable

GSU Baseplate

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 132
GEMINI GXL Installation Technical Manual Electrical Interconnect

5 Connect the CAT5 EKG cable from the CT Host cabinet to the GSU Bulkhead Bracket.

CAT5 EKG Cable Connection

GSU Baseplate

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 133
GEMINI GXL Installation Technical Manual Electrical Interconnect

6 Connect the 98 ft. (30 m) single fiber optic


cable from the CT Recon cabinet to the
GSU Bulkhead Bracket. Single Fiber Optic Cable

GSU Baseplate

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 134
GEMINI GXL Installation Technical Manual Electrical Interconnect

7 Connect the 50-pin CanBus cable from the CT Host cabinet to the Bulkhead Bracket mounted to the GSU
baseplate.

50-pin CanBus Cable

GSU Baseplate

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 135
GEMINI GXL Installation Technical Manual Electrical Interconnect

Mobile GSU Electrical Connections


Mobile Units require a Table/Docking I/F Block Assembly to be added to the GSU along with several additional
cables. These parts can be obtained from the GEMINI Mobile Parts Kit (4535 679 71311).

PROCEDURE
Mounting Hardware
1/4-20 x 3/8” SHCS w/ Lockwashers
1 Using the mounting hardware supplied, TB60
TB40
mount the Table/Docking I/F Block
Assembly (4535 679 71901) as shown.
2 Connect the Right and Left
“Gantry Docking Switch Harnesses” (4535
679 71831) wiring to TB40 and TB60 on
the Table/Docking I/F Block Assembly.
a Right Side Docking Switch
• Black wire to TB40-5A
• Red wire to TB40-6A
• Black wire to TB60-3A
• White wire to TB60-4A b Left
Side Docking Switch
• Black wire to TB40-1A
• Red wire to TB40-2A Right Side Left Side
• Black wire to TB60-1A
• White wire to TB60-2A Gantry Docking Table/Docking
Switch Harness I/F Block Assembly

(Right Side)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 136
GEMINI GXL Installation Technical Manual Electrical Interconnect

3 Connect the “Lock/Unlock Sw to GSU Cable” TB50 TB60


TB40
(4535 679 71841) wiring to TB40, TB50, and
TB60 on the Table/Docking I/F Block Assembly.
This cable was inserted through the cabling
channel at the end of the “Gantry Separation Unit
Leveling” procedure on Page 71.
• Black wire to TB40-3A
• Red wire to TB40-4A
• Black wire to TB50-1A
• Green wire to TB50-2A
• Black wire to TB60-1B
• White wire to TB60-4B
The two connectors at the opposite end of this
harness are marked as followed:
• TBL SW (see Page 168)
Lock/Unlock Sw to GSU Cable

• J15 (see Page 184)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 137
GEMINI GXL Installation Technical Manual Electrical Interconnect

4 Connect the “Pallet Unlocked LED” coach


wiring to TB50 on the Table/Docking I/F Block
Assembly.
• Red wire to TB50-1B
• White wire to TB50-2B
5 Connect the “Pallet Locked LED” coach TB50
wiring to TB40 on the Table/Docking I/F Block
Assembly.
• Red wire to TB40-1BTB40
• White wire to TB40-6B

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 138
GEMINI GXL Installation Technical Manual Electrical Interconnect

CT Gantry Electrical Connections

WARNING MAIN POWER MUST BE DISCONNECTED AND LOCKED OUT PRIOR TO


HANDLING ANY OF THE WIRE BUNDLES OR COMPLETING ANY CONNECTIONS. FAILURE TO
COMPLY MAY RESULT IN DEATH, SERIOUS INJURY, AND SERIOUS EQUIPMENT DAMAGE.
PROCEDURE

1 Route the 440 VAC Main Power cables


and the two 220 VAC power cables under
the PET Gantry, through GSU trough, to
the back of the CT Gantry.

440 VAC Main Power and 220 VAC


(UPS and CT Recon) Power Cables

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 139
GEMINI GXL Installation Technical Manual Electrical Interconnect

2 At the rear of the CT Gantry, connect the cables


from the Main Power Conduit to the Main Power Ground Lugs
Terminal Block and Ground Lugs, ensuring that
the incoming power cables correspond to the
gantry cables already attached to the terminal
block.

Main Power
Terminal Block

Main Power
Ground Lugs
Terminal Block

480V cables
(Marked Brown,
Orange, Yellow)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 140
GEMINI GXL Installation Technical Manual Electrical Interconnect

3 Connect the Control Room UPS Power


Cable to the following points at the back of
Point 4 TB372
Point 8 the CT gantry:
• Brown wire to the 6A fuse F373
• Blue wire to TB372, point 4
• Green/Yellow wire to TB372, point 8

F373 CT Console
Power Cable

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 141
GEMINI GXL Installation Technical Manual Electrical Interconnect

4 Connect the CT Recon Power Cable to the


Point 5 Point 9 TB372 following points at
the back of the CT gantry:
• Brown wire to the 10A fuse F374
• Blue wire to TB372, point 5
• Green/Yellow wire to TB372, point 9

F374 CT Recon
Power Cable

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 142
GEMINI GXL Installation Technical Manual Electrical Interconnect

5 Connect the Suite Door wiring harness plug


(Cable Track 4) from TB30 on the GSU
Baseplate to J37 on the Stargate
Distribution Board.
6 Connect the X-Ray On wiring harness
(Cable Track 4) from TB20 on the GSU
Baseplate to TB20 at the rear of the CT
gantry.

J37 Stargate Distribution Board

TB20

X-Ray On Wiring Harness

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 143
GEMINI GXL Installation Technical Manual Electrical Interconnect

7 Connect the dual fiber optic cable (Cable Stargate Distribution Board
Track 4) from the GSU Bulkhead Bracket to
the 8-port hub.

Dual Fiber Optic


Cable Connection 8-Port Hub

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 144
GEMINI GXL Installation Technical Manual Electrical Interconnect

8 Connect the single fiber optic cable


(Cable Track 4) from the GSU Bulkhead
Bracket to the Optical Stator Module at
the rear of the CT gantry.

Single Fiber Optic Cable Optical Stator Module

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 145
GEMINI GXL Installation Technical Manual Electrical Interconnect

9 Connect the 50-pin CanBus cable (Cable


Track 4) from the GSU Bulkhead Bracket to
connector J16 on the GHOST board.

50-pin CanBus Cable

GHOST Board

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 146
GEMINI GXL Installation Technical Manual Electrical Interconnect

10 Connect the Acquisitor cable (Cable Track


4) from the GSU Bulkhead Bracket to the
GHOST board at P27.

Acquisitor Cable (P27)

GHOST Board

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 147
GEMINI GXL Installation Technical Manual Electrical Interconnect

11 Connect the EKG CAT5 cable (Cable


Track 4) from the GSU Bulkhead Bracket
Rear of CT Gantry
to the I-Box USB Extender (Xmitter Module).

I-Box USB Extender

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 148
GEMINI GXL Installation Technical Manual Electrical Interconnect

12 Connect the following three cables from


the Outer Cable Take-Up to the Stargate
Distribution Board Assembly:
• J31 SGG I/F to connector J31
• J35 SGG I/F to connector J35
• J39 SGG I/F to connector J39
The Stargate Distribution Board
Assembly is located at the rear of the CT
gantry.

Outer Cable Take-Up

J39 J35 J31

Stargate Distribution Board

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 149
GEMINI GXL Installation Technical Manual Electrical Interconnect

13 Connect the 230VAC cable set ALLEGRO AC POWER from


the Outer Cable Take-Up to the Terminal Block next to the Green/Yellow Wire
Blue Wire
Main Power Terminal Block.
• Brown wire to Fuse Holder
• Blue wire to Ground Block
• Green/Yellow wire to ground lug
14 Connect the data cable HUB 1 from the Outer Take-Up Cable
to the 8-port Hub connector 1SW.
The 8-Port Hub is mounted to the rear of the CT gantry base
underneath the Stargate Distribution Board.
15 Connect the data cable from the Patient Table to the #8 port of
Brown Wire Allegro Power
the Hub. Cable

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 150
GEMINI GXL Installation Technical Manual Electrical Interconnect

16 Ensure that the two switches are set to the Two Switches set to “No Cross” and “Full Duplex” positions.
Stargate “Full Duplex” & “No
Distribution Board Cross” Patient Table

Dual-Fiber 1SW 8-Port Hub


Cable Connection

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 151
GEMINI GXL Installation Technical Manual Electrical Interconnect

17 Connect the following four cables from the Middle


Cable Take-Up to the Stargate Distribution Board:
• J6 SGG I/F to connector J6
• J2 SGG I/F to connector J2
• J43 SGG I/F to connector J43
• J45 SGG I/F to connector J45

Middle Cable Take-Up

J2 J45 J43 J6 (Not Shown)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 152
GEMINI GXL Installation Technical Manual Electrical Interconnect

Connect the following two cables from


the Middle Cable Take-Up to the GHOST GHOST Board
Board:
• J24 GHOST to connector J24
• J23 GHOST to connector J23
The GHOST Board is located on the RH
side of the CT gantry.

J24 J23

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 153
GEMINI GXL Installation Technical Manual Electrical Interconnect

19 Connect the following six cables from the


Inner Cable Take Up to the Stargate
Distribution Board:
• J9 SGG I/F to connector J9
• J29 SGG I/F to connector J29
• J23 SGG I/F to connector J23
• J47 SGG I/F to connector J47
• J10 SGG I/F to connector J10

Inner Cable Take-Up

J23 J29 J10 J9 J47

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 154
GEMINI GXL Installation Technical Manual Electrical Interconnect

20 Connect the CT Recon ground wire to the CT gantry base plate CT Gantry Base Plate ground stud located at
the rear of the CT gantry. Ground Stud

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 155
GEMINI GXL Installation Technical Manual Electrical Interconnect

PET Gantry Electrical Connections


PROCEDURE

1 Locate and identify the PDU assembly.


The PDU is located at the rear of the PET gantry. PET Rear Gantry
(Left Side)

2 Locate and identify the Anderson connector located


near the PDU.

PET Gantry (Rear)

3 Connect the PET AC Power wiring harness (Outer


Cable Take-Up) from the CT gantry to the Anderson
connector.

Outer Cable Take-Up

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 156
GEMINI GXL Installation Technical Manual Electrical Interconnect

4 Locate and identify the Allegro Distribution Board


Assembly.
The Allegro Distribution Board Assembly is located at
the front of the PET Gantry.

Allegro Distribution Board


Assembly

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 157
GEMINI GXL Installation Technical Manual Electrical Interconnect

5 Connect the following cables from the Outer Cable Take-Up to the Allegro Distribution Board:
• J22 ALG I/F to connector J22
• J20 ALG I/F to connector J20
• J18 ALG I/F to connector J218 J22

J20

J18

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 158
GEMINI GXL Installation Technical Manual Electrical Interconnect

6 Locate and identify the Acquisition Computer CPU.


The Acquisition Computer CPU is located at the rear of
the PET Gantry.

PET Gantry (Rear View) Acquisition Computer CPU

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 159
GEMINI GXL Installation Technical Manual Electrical Interconnect

7 Connect the data cable PC from the Outer


Cable Take-Up to the Acquisition
Computer CPU.

Acquisition Computer CPU


Data Cable “PC” Connection

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 160
GEMINI GXL Installation Technical Manual Electrical Interconnect

8 Connect the cables from the Cable Take-Up


to the Allegro Distribution Board as shown. J15 J25 J14 J3

J23

J24

J9

J6

J16

J12

J8

J29 J2 J5
9 Locate and identify the PIB Board at the rear of the Gantry.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 161
GEMINI GXL Installation Technical Manual Electrical Interconnect

10 Connect the cable JC1 PIB PCB from the Middle Cable Take-Up to the PIB board connector JC1

JC1

PET Gantry (Rear View) PIB Board

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 162
GEMINI GXL Installation Technical Manual Electrical Interconnect

GEMINI Patient Table Electrical Connections


PROCEDURE

1 Locate and identify the CT to Patient Table Wiring


Harnesses.

CT to Patient Table
Wiring Harness #2

CT to Patient Table
Wiring Harness #1

2 Route the wiring harnesses to the Patient Table and


into the wiring trough.

Patient Table CT to Patient Table


Wiring Trough Wiring Harnesses

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 163
GEMINI GXL Installation Technical Manual Electrical Interconnect

3 Locate and identify the GEMINI Couch Control Board Assembly.


The GEMINI Couch Control Board Assembly is located on the RH Side of the Patient Table.
4 Connect the cables from Wiring P724
Harness #1 to the GEMINI Couch
Control Board Assembly at the
following connections:
• Cable J701 (L11925 -
Couch Base Bd.) to
connector P701.
• Cable J702 (L11925 -
Couch Base Bd.) to
connector P702.
• Cable P724 (L11924 -
Couch Power Bd.) to
connector P724.

P701 P702

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 164
GEMINI GXL Installation Technical Manual Electrical Interconnect

5 Connect the cables from Wiring P11 P23


Harness #2 to the GEMINI Couch
Control Board Assembly at the
following connections:
• Cable P23 (4535 679 26421 -
GCCB) to connector P23.
• Cable P12 (GCCB) to
connector P12.
• Cable J14 (4535 679 26461 -
SGG/IF) to connector P11.
• Cable P18 (4535 679 26451 -
GCCB) to connector P18.

P12 P18

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 165
GEMINI GXL Installation Technical Manual Electrical Interconnect

6 Connect the two ground wires labeled PE1 (L22524 - Couch) and PE2 (L23524 - Couch) from Wiring Harness
#1 to the ground lug in the Patient Table base.
Patient Table (LH Side)

PE1 and PE2 Connected


to Ground Lug

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 166
GEMINI GXL Installation Technical Manual Electrical Interconnect

7 Connect J746 (L11903 - ACCU to Couch Cable) to its 6-pin


mate. Terminal Block

The mate for J746 is a wiring harness connected to the


terminal block.

J746

8 Connect J743 (L22525 - 24V Patient Table Cable) to its


4-pin mate. Patient Table
Switch Assembly
The mate for J743 is a wiring harness connected to the J743
Patient Table Switch Assembly.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 167
GEMINI GXL Installation Technical Manual Electrical Interconnect

9 Locate and identify the Patient Table PC Assembly


The Patient Table PC Assembly is located on the RH Side of the Patient Table.
.

Patient Table PC Assembly

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 168
GEMINI GXL Installation Technical Manual Electrical Interconnect

10 Connect the data cable marked ARGOS to the Patient Table PC Assembly.

Patient Table PC Assembly Data Cable Argos

Mobile Patient Table Connections


In addition to the standard Patient Table connections, Mobile sites require the “Table Lock/Unlock Sw Cable”
connection to the “Mobile Table Lock Lever Assembly”. These parts can be obtained from the GEMINI Mobile
Parts Kit (4535 679 71311).

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 169
GEMINI GXL Installation Technical Manual Electrical Interconnect

PROCEDURE
Lock Unlock Mobile Table
1 Connect the Table Lock/Unlock Sw Cable (4535 679 Lock Lever Assembly
Switch Switch
71861) to the Lock and Unlock switches.
a Lock Switch
• Black wire to COM
• Red wire to NC b Unlock Switch
• Black wire to COM
• Green wire to NO
The mate for TBL SW (4 pin connector) is the
Lock/Unlock Switch Cable (4535 679 71851) which is
defined later in this procedure.
COM NC COM NO

2 Using the mounting hardware, attach the


Mounting Plate and Table Lock Lever Assembly to the Patient Table. Mounting Mounting
Plate Hardware

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 170
GEMINI GXL Installation Technical Manual Electrical Interconnect

3 Dress the Table Lock/Unlock Sw Cable as


shown.

Table Lock/Unlock Sw Cable

4 Run the cable along side of the table cabling down to table trough. Dress cable with tie wraps.
5 Connect the Table Lock/Unlock Sw Cable
(4535 679 71861) TBL SW to the Lock/
Unlock Sw Cable (4535 679 71851) 4-pin
mate.
6 Feed Lock/Unlock Sw Cable through the
bottom of the CT Gantry back to the
Inbore Lifter. Connect the TBL SW
connector to the 4-pin mate of cable 4535
679 71841.
T
a
b
l
e

L
Lock/Unlock Sw Cable

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 171
GEMINI GXL Installation Technical Manual Electrical Interconnect
Mobile CT Gantry Connections
Cabling Disconnects
Mobile requires some disconnecting of cables on the CT Gantry.

PROCEDURE

1 Locate the Breaker Panel on the right side of Breaker


Panel
the CT Gantry.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 172
GEMINI GXL Installation Technical Manual Electrical Interconnect

Unmount
Breaker
Panel

2 Unmount the Breaker panel.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 173
GEMINI GXL Installation Technical Manual Electrical Interconnect
3 From the AC Control Unit, remove the
STATOR PWR SP Breaker
blue wire from TB302-2. CB301
4 Disconnect and remove the opposite end
of the blue wire from the Tilt Drive
Breaker (CB305-LD).
5 Disconnect and remove the black jumper
wire (120VAC) from both the Tilt Drive
Breaker (CB305-LN) to the Stator Power
Supply breaker (CB301-LN).

TILT DRIVE Breaker Remove Blue Wire


CB305 TB302-2

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 174
GEMINI GXL Installation Technical Manual Electrical Interconnect
6 Under the CT Gantry upper front cover,
disconnect the Disconnect Brown Wire brown
wire from the Blue Switch Utility Xfmr Power
from SW-1
Interrupt (SW-1).
7 Fold back and tie wrap the brown wire to the
harness.

8 From the rear bottom of the CT Gantry,


disconnect the brown wire from the Utility/
Console Xfmr Primary Controller (TB371-8B).
9 Fold back and tie wrap the brown wire.

Disconnect Brown Wire from


TB371-8B

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 175
GEMINI GXL Installation Technical Manual Electrical Interconnect
Mobile CT Cabling
In addition to the standard CT Gantry connections, Mobile sites require additional cabling. These parts can be
obtained from the GEMINI Mobile Parts Kit (4535 679 71311).

PROCEDURE
1 Remove any existing connections
from the TILT DRIVE Breaker
(CB305).
2 Install the Utility Power Interrupt
Cable (4535 679 71721).
3 CB305-LD and CB305-LN
connectors to the TILT DRIVE
Breaker (CB305).
4 Run the TB371-8B wire end
through the access hole in the
gantry as shown.

TILT DRIVE Breaker TB371-8B wire end (4535 679 71721)


CB305 through access hole

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 176
GEMINI GXL Installation Technical Manual Electrical Interconnect
5 At the rear bottom of the CT Gantry, connect
the brown wire end (TB371-8B) to the Utility/
Console Xfmr Primary Controller.

TB371-8B

6 Run the SW-1 connector up to the Blue Switch Utility SW-1


Xfmr Power Interrupt (SW-1) which is located under
the CT Gantry upper front cover.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 177
GEMINI GXL Installation Technical Manual Electrical Interconnect
7 Mount the Breaker Panel to the gantry.
8 Relabel TILT DRIVE breaker with new Breaker
label “UTILITY XFMR PWR”. Panel

Add new Label


UTILITY
XFMR PWR

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 178
GEMINI GXL Installation Technical Manual Electrical Interconnect
Mobile and Fixed Mobile Sites PET Gantry Floor Locks
Mobile requires that the PET Gantry be locked in place at all times, except for maintenance reasons. The following
procedure describes the installation of the Mobile Gantry Floor Lock Assembly (4535 679 71931). This installation
is required on both sides of the PET Gantry.

PROCEDURE

1 The Mobile Gantry Block must be mounted


to the coach steel sub-flooring.
Proper positioning of the Gantry Block must
first be established before routing the floor
surface.
2 The Locked Position of the PET Gantry is to
be 7/16” back from the Closed-Position.
Push the PET Gantry all the way up against
the CT Gantry. Mark the position, and then
Mobile Locking Mobile
move the PET Gantry away from the CT Gantry Block Screw Gantry Bracket
Gantry by 7/16”.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 179
GEMINI GXL Installation Technical Manual Electrical Interconnect
3 Assemble the Locking Screw by inserting the
Locking Screw through the Spring Washer, Locking
Screw
Spring, Gantry Bracket and into the threaded
Switch Activator.
Spring
Washer

Spring

Switch Activator Mobile


(not shown) Gantry Bracket
4 Align the Mobile Gantry Block to the Locking Screw by
temporarily mounting the Mobile Gantry Bracket.
In order to fit the Gantry Block beneath the Gantry
Bracket, it is necessary not to install the Spacers and
Flat Washers at this time and do not tighten the
Gantry Bracket mounting scews.

Spacer and Mobile


Flat Washers Gantry Bracket

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 180
GEMINI GXL Installation Technical Manual Electrical Interconnect
5 With the Gantry Bracket mounted, slide the Gantry Block into position. Insert the Locking Screw into the
Gantry Block to ensure proper alignment.
6 With the Gantry Block in place, mark the perimeter for routing.
7 Remove the Floor Lock Assembly from the gantry and also move Gantry Block.
8 Route the coach floor down to the steel sub-flooring.

9 Thoroughly clean the routed area and place the


Gantry Block back into position.
10 Mark the position for the two Pull Pin holes.
11 Remove the Gantry Block and drill the Pull Pin
holes.
12 Set the Gantry Block in place with the Pull Pins
inserted.
13 Drill pilot holes for the mounting screws.

Mounting Holes Pull Pin Holes


(Qty. 2) (Qty. 2)
4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 181
GEMINI GXL Installation Technical Manual Electr
14 Remove the Gantry Block. Drill and tap the steel floor for the mounting hardware.

15 Flip the Gantry Block over and mount the


Plunger Switches and Switch Cover.
16 The Gantry Docking Switches Harness (4535
679 71831) should be connected to the
Plunger Switches as followed:
a Lock Plunger Switch
• Black wire to COMMON
• Red wire to NORM OPEN
b Sense Plunger Switch
• Black wire to COMMON
• White wire to NORM OPEN

Gantry Docking
Switch Cover Plunger Switches
Switches Harness

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information
GEMINI GXL Installation Technical Manual Electrical Interconnect

17 The coach floor must also be routed for the Gantry Docking switches Harness to sit flush with the floor
surface.

18 With the Gantry Block in place, dress the harness


to the GSU and mark the harness for routing.
19 Move the Gantry Block and harness out of
position and route the coach floor.
Ensure that the harness routing is
narrow enough to keep the harness
in place when inserted.
20 Thoroughly clean the routed areas and place the
Gantry Blocks and harnesses into position.
21 Mount the Gantry Blocks to the coach floor using
Dress the Harness the mounting screws.
From the Locking Blocks

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 183
GEMINI GXL Installation Technical Manual Electrical Interconnect

22 Mount of the Mobile Gantry Bracket to the bottom of the PET


Gantry.
Ensure the bracket is mounted level with the
floor via the Spacer and the number of Flat
Washers used.
23 Lock the PET Gantry in place.

Spacer and Mobile


Flat Washers
Gantry Bracket

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 184
GEMINI GXL Installation Technical Manual Inbore Lifter Installation

Inbore Lifter Installation

Personnel Required Time Required

1 Field Service Engineer 1 Hours

Overview
This section will detail the proper method of installing and wiring the Inbore Lifter. It is recommended that the
entire procedure be read through once before proceeding with the installation.

Inbore Lifter

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 185
GEMINI GXL Installation Technical Manual Inbore Lifter Installation

Inbore Lifter Installation


PROCEDURE

2 R Inbore Lifter
e
m
o
v
e

t
h
e

p
a
c Table Lifter
k Mounting
a Blocks (Qty. 2)
g
i Install and
n Tighten SHCS

3 Install and tighten the four screws with washers securing


the lifter to the mounting blocks.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 186
GEMINI GXL Installation Technical Manual Inbore Lifter Installation

Inbore Lifter Electrical Connect


1 Connect the wiring harness from J1 of the J10
Inbore Lifter MMI board to connector J10 of
the Stargate Distribution Board.

Stargate Distribution Board

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 187
GEMINI GXL Installation Technical Manual Inbore Lifter Installation

2 Connect the wiring harness from J2 of the J9


Inbore Lifter MMI board to connector J9 of
the Stargate Distribution Board.

Stargate Distribution Board

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 188
GEMINI GXL Installation Technical Manual Inbore Lifter Installation

Mobile and Filed Mobile Cabling


In addition to the standard Inbore Lifter connections, Mobile sites require the “Lock/Unlock Sw to GSU Cable”
connection to the Inbore Lifter MMI Board.

PROCEDURE

1 Connect the “Lock/Unlock Sw to GSU


Cable” (4535 679 71841) from the GSU J15
“Table Docking I/F Block Assembly” to
connector J15 of the Inbore Lifter MMI
Board.

Inbore Lifter MMI Board

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 189
GEMINI GXL Installation Technical Manual Control Room Set-Up

Control Room Set-Up

Personnel Required Time Required

2 Field Service Engineer 1 Hours (8 Lead and 3 Assistant)


The GEMINI standard system configuration is a stand-alone system which can be easily modified for networking
and remote access. The system consists of two workstations located in the Control Room. The workstation
assemblies are comprised of the acquisition workstations for both the PET and CT.
“Remote workstation” is the term used to describe additional workstations connected to the GEMINI networked
system configuration. The remote workstation, also referred to as the “doctor’s remote viewing station”, consists of
a workstation and monitor. The GEMINI system is capable of hosting multiple remote workstations. Remote
workstations can only interface with one GEMINI system.

WARNING HAZARDOUS VOLTAGE PRESENT. USE CARE WHEN INSTALLING THE


CONTROL ROOM COMPONENTS. DO NOT APPLY POWER TO THE CONTROL
ROOM COMPONENTS UNTIL ALL CONNECTIONS ARE MADE. FAILURE TO COMPLY MAY RESULT IN
ELECTRICAL SHOCK AND DAMAGE TO SENSITIVE ELECTRONIC COMPONENTS.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 190
GEMINI GXL Installation Technical Manual Control Room Set-Up

Typical Control Room Workstation Setup

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 191
GEMINI GXL Installation Technical Manual Control Room Set-Up

220 vac

220 vac
GSU

GND

IN

Single Fiber

15 pin cable Acq


CT
Dual
Fiber Cat 5
Recon
Host
Rack
50 p Can

Cat 5
Local Net

Cat 5

Cat 5
P3
Gantry Lan
Giga
Stargate 2
port port bit
USB
2 3 Conv
Net Switch
CT
USB J2 Scan
1 Box

Hosp Giga
Net bit

La
Pet CT n
Hosp
Net
Server Host 1

IN

U
P
IN IN
220 vac
S
220
vac

Computer Interconnect

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 192
GEMINI GXL Installation Technical Manual Control Room Set-Up

Computer Room Preparation


PROCEDURE

1 Verify the placement of all computers, monitors, and options with the customer.
2 Place all Control Room components according to the customer requirements.
3 Install the UPS system.
HINT Refer to the UPS Installation Technical Manual (P/N 4535 670 12221).
4 Verify that a local network connection and analog telephone line are active in the room.
5 Install the remote UPS shut-off box on a convenient wall location.
6 Place and install any auto-injector computer.
7 Place and install any LAP laser computer.

CT Workstation Set-Up

PROCEDURE

1 Place the DAT tape drive, optical drive, and CD-ROM drive on top of the tower case.
2 Remove the rear cover from the CT Host tower case.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 193
GEMINI GXL Installation Technical Manual Control Room Set-Up

3 Connect AC power from the UPS to the CT Host


computer.

CT Host Computer
(Rear View)

220 VAC Power Connection

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 194
GEMINI GXL Installation Technical Manual Control Room Set-Up

4 Connect the two AC power cables from the Monitor Power Cables (Qty. 2)
2 monitors to the CT Host computer Power
Distribution Unit.

Power Distribution Unit


Power Cables (Qty. 2)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 195
GEMINI GXL Installation Technical Manual Control Room Set-Up

5 Connect the keyboard cable to the KB


connection on the CT Host computer.
6 Connect the mouse to the MOUSE
connection on the CT Host computer.

Mouse Connection

Keyboard Connection

CT Host Computer (Rear View)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 196
GEMINI GXL Installation Technical Manual Control Room Set-Up

7 Connect the cables from the two CT monitors


to the back of the CT Host computer.

CT Monitor Cables (Qty. 2)

CT Host Computer (Rear View)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 197
GEMINI GXL Installation Technical Manual Control Room Set-Up

8 Connect the CT Box Assembly to the CT Box (Stargate) J2 Connector.

CT Host Computer (Rear View)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 198
GEMINI GXL Installation Technical Manual Control Room Set-Up

9 Connect the 50-pin CanBus cable from the CT gantry to


the back of the CT Host computer.

CT Gantry Connection

10 Connect the CAT5 cable from the GSU Bulkhead Bracket to the
USB Extender (Xcvr Module) at the back of the CT Host
computer.

USB Extender CAT5 Cable (from


(Xcvr Module) Bulkhead Bracket)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 199
GEMINI GXL Installation Technical Manual Control Room Set-Up

11 Connect the data cable from the CT


Recon to the LAN2 port of the CT Host.
12 Connect the Building Network data cable
to the LAN1 port of the CT Host.

LAN2 LAN1

CT Host Computer (Rear View)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 200
GEMINI GXL Installation Technical Manual Control Room Set-Up

PETView Workstation Set-up


PROCEDURE

1 Remove the rear cover from the PET Server computer cabinet.
2 Connect the power cable from the UPS to the PET Server Power Strip.

PET Server
Power Cable (Rear View)
From UPS

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 201
GEMINI GXL Installation Technical Manual Control Room Set-Up

3 Connect the Data cable from the CT Ethernet Switch


Host cabinet to Port 3 of the Ethernet
Switch.
4 Connect the Dual Fiber cable from the
GSU Bulkhead Bracket to Fiber Port FX
of the Ethernet Switch.

CT Host Data Cable Dual Fiber Cable

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 202
GEMINI GXL Installation Technical Manual Control Room Set-Up

5 Connect the customer supplied analog telephone line to the Line In port.

Customer Supplied
Telephone Line

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 203
GEMINI GXL Installation Technical Manual Control Room Set-Up

6 Connect the PET Server Keyboard to the


USB 0 port.
7 Connect the PET Server Mouse to the
USB 1 port.

PET Server Keyboard PET Server Mouse

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 204
GEMINI GXL Installation Technical Manual Control Room Set-Up

8 Connect the 19” Flat Panel PET Display monitor to the Video port.

9 Connect the power cord from the 19” Flat Panel PET Display to the
Power Strip.
10 Connect the power cord from the CD R/W Drive (Optional) to the Power
Strip.
11 Connect the power cord from the Optical Drive to the Power Strip.
12 Connect the provided SCSI cable from the optical drive to the DAT tape
drive.
13 Connect the provided SCSI cable from the DAT tape drive to the CD-
ROM drive.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 205
GEMINI GXL Installation Technical Manual Control Room Set-Up

14 Connect the provided SCSI cable from the PET Server computer to the optical drive.

CT Recon Electrical Connections


1 Connect the 220 VAC CT Recon Power Cable to
the CT Recon at the connector labeled Power In
P8.

Power In P8

CT Recon (Rear View)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 206
GEMINI GXL Installation Technical Manual Control Room Set-Up

2 Connect the single fiber optic cable from


the GSU Bulkhead Bracket to the back
of the CT Recon at connector DS5.

Connector DS5

CT Recon (Rear View)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 207
GEMINI GXL Installation Technical Manual Control Room Set-Up

3 Connect the 15-pin Acquisitor cable


from the GSU Bulkhead Bracket to the
back of the CT Recon at ACQ J1.

15-Pin Acquisitor Cable Connection

CT Recon (Rear View)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 208
GEMINI GXL Installation Technical Manual Control Room Set-Up

4 Connect the ground wire from the CT


gantry base plate to the back of the CT
Recon at the ground stud GND Sc.

GND Sc.

CT Recon (Rear View)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 209
GEMINI GXL Installation Technical Manual Control Room Set-Up

5 Connect the data cable from the CT


Host computer to the back of the CT
Recon at the Host Rack connection.

Host Rack Connection

CT Recon (Rear View)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 210
GEMINI GXL Installation Technical Manual
Final Patient Table Alignment

Final Patient Table Alignment

Overview
Before the Patient Table grout is poured and the table anchored, the alignment of the Patient Table in relation to
the gantries must be checked. This procedure instructs how to check the Patient Table alignment.
1 Extend the Patient Table into the
PET gantry bore.

2 Establish the centerline of the Patient Table Upper Pallet.


3 Obtain a measurement from two opposite, fixed points of the PET gantry bore.
4 Ensure that the centerline of the Patient Table Upper Pallet is exactly half-way between the fixed
points mentioned in step 3.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 211
GEMINI GXL Installation Technical Manual Pouring the Micorox Compound – Patient Table
NOTE If the Patient Table is not centered in the PET gantry bore, center the Patient Table and refer to the Patient Table
Mechanical Alignment and Leveling section for details on aligning the Patient Table.

Fixed Site Pouring the Micorox Compound – Patient Table

Overview
The following steps outline the procedure for mixing and pouring the Micorox compound. The Micorox is a
selfleveling two-component epoxy compound used to fill the volume between the floor and the leveled Patient
Table. You will be required to work against the cure time of the Micorox compound and as such, must make certain
you are ready to use it and will be able to work continuously without interruption. Exercise caution when mixing and
pouring the compound to prevent splashing, as damage to the floor, walls, and cabinets may occur. Wear
disposable gloves at all times and follow the manufacturer's specific instructions for cleaning and skin contact
should that occur. Once the Micorox compound has fully cured, the Patient Table can be anchored to the site floor.

Personnel Required Time Required

1 Hour to perform procedure


1 Field Service Engineer 8 Hours minimum for Micorox to cure
PROCEDURE

1 Prepare the following tools and material are ready for the mixing and pouring of the Micorox compound:
• Electric drill motor with 1/2” chuck
• (2)25 lb. (6.8 kg) cans of Micorox epoxy base
• (2) 1 qt. (0.95L) cans of Micorox hardener
• Disposable gloves
• Mixing paddle
• Plastic, paper, or cardboard sheets
• Tape

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 212
GEMINI GXL Installation Technical Manual Pouring the Micorox Compound – Patient Table
2 Protect the floor around the Patient Table. a Tape paper, cardboard, or plastic sheet to the floor around the
outside of the Patient Table.
Use as little tape as possible, to avoid leaving tape or adhesive behind on the floor of the facility.
3 Mix the two components of the floor pour epoxy, one pail at a time.
Each end of the Patient Table template will require approximately
one pail of the Micorox compound. However, for ease of pouring, it
is recommended that only one pail be mixed at a time.

NOTE The epoxy base and hardener must be at room temperature prior to use.
a Place one 25 lb. (6.8 kg) pail of epoxy base on a piece of heavy cardboard or similar floor protection.
b Install the mixing paddle in the drill chuck.

CAUTION Do not mix the epoxy while the pail is sitting directly on the floor. When the two components of
the epoxy are mixed, an exothermic reaction occurs which will heat the metal pail and may
damage the floor of the facility. Place a piece of cardboard or similar under the pail to protect the
floor.
c Put on a pair of latex gloves and remove the lids from one 25 lb.
pail of epoxy base and one 1 qt. container of hardener.
d Pour the entire contents of the hardener into the epoxy base.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 213
GEMINI GXL Installation Technical Manual Pouring the Micorox Compound – Patient Table
e Insert the mixing paddle and blend for a minimum of 10
minutes.

4 Set the mixer aside.

a Remove the mixer from the pail slowly and place it on a piece of cardboard or paper.
b Do not allow any of the epoxy to drip onto the separation unit or the surrounding floor outside of the
template.
• If any spillage does occur, quickly clean it up with rags or paper towels.

CAUTION Exercise caution so that the epoxy does not drip on the floor. If dripping or spilling does occur, ensure
that you clean it up quickly. Failure to do so may cause permanent damage to the flooring.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 214
GEMINI GXL Installation Technical Manual Pouring the Micorox Compound – Patient Table
5 Begin pouring the mixed Micorox into one end of the Patient Table
floor template.
The Patient Table template has two troughs that protrude from each
side. These troughs are called pouring zones. These pour zones Patient Table
Pour Zones
will be removed once the Micorox has cured later in the procedure.
a Start at either end and slowly pour the epoxy into the floor
template at one of the pour zones. Check for leaks while pouring;
these can be plugged using a small amount of petroleum jelly.
b Monitor the viewing holes in the Patient Table base plate. As the
area in which you are pouring begins to fill, move to the pour
zone on the other side of the table end. Alternate between the
two pour zones and continue filling until the Micorox is
approximately 1/8" (3 mm) from the top of the viewing holes. Do
not overfill.
c Instruct another FSE to begin preparing and mixing another pail of the epoxy if more than one-half of a can
was used in the previous pour.
d Repeat step b for the remaining end of the Patient Table floor template.
e Allow the Micorox in both sides of the Patient Table to remain undisturbed for twenty to thirty minutes.
Re-check the Micorox level and add more if required.

CAUTION Exercise caution so that the epoxy does not drip on the floor. if dripping or spilling does occur, ensure that you
clean it up quickly. Failure to do so may cause
permanent damage to the flooring.
6 Insert the pour zone barrier strips. Pour Zone
Barrier Strip
a Coat both sides of the barrier strips with petroleum jelly.
b Insert two barrier strips into each corner of the floor template
as shown.
Pour Zone

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 215
GEMINI GXL Installation Technical Manual Pouring the Micorox Compound – Patient Table
The barrier strips allow the four pour zones to be removed from the sides of the Patient Table.
7 Allow the leveling epoxy to fully cure overnight.
• Clear the room of all loose hardware and tools, and
properly dispose of the material containers before
leaving the facility.

NOTE You may notice that the GSU or table changes levelness slightly after the Micorox has been poured.
This is normal and is not significant.
Plan to resume the installation the following day. The Micorox requires eight hours to fully cure.

CAUTION Ensure that all partially used cans of Micorox are kept well ventilated and away from
combustibles. The exothermic reaction produces significant fumes and heat. Partially used cans
should be placed on a concrete floor or similar until the compound has fully cured. Failure to
comply may result in damage to the site floors and/or equipment.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 216
GEMINI GXL Installation Technical Manual Anchoring the Patient Table

Anchoring the Patient Table

Personnel Required Time Required

1 Field Service Engineer 1 Hour


PROCEDURE

1 Retrieve the following equipment in preparation for drilling the floor anchor holes.
• Hammer Drill
• Masonry Drill Bit
• Vacuum
• Blow Bulb
• Hammer
• 3/4” Wrench
• Floor Anchor Bolts
2 Insert and secure the masonry drill bit into the drill chuck; if not already done.
• If the drill you are using is so equipped, set the drill depth gauge to 4 1/2” (11.5 cm) from the top of the
spot face holes in the GSU.
• If the drill you are using is not equipped with a depth gauge, measure 4 1/2” (11.5 cm) from the tip of the
drill bit, and place a piece of tape around the bit at that point.
3 Drill two anchor holes through the patient table into the floor.
To provide adequate clearance for the drill under the outer edges of the table, it may be necessary to raise the table.
a Identify and select one primary anchor location in each end of the table (head and foot) in
which to install the first two floor anchors.
NOTE If you hit rebar while drilling, select the next nearest spot faced hole and use it to set the floor anchor into.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 217
GEMINI GXL Installation Technical Manual Anchoring the Patient Table

b Drill through the selected holes into the floor using the drill and masonry bit. c Keep the
vacuum on and positioned near the drill bit while drilling to extract dust and debris.
d Pause frequently, remove the drill bit from the hole(s), and use the blow bulb to keep the inside of the hole(s)
free of debris.
e Continue to drill until the tape on the drill bit is even with the top of the hole in the patient table base or the
depth gauge indicates 4 1/2” (11.5 cm).

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 218
GEMINI GXL Installation Technical Manual Anchoring the Patient Table

4 Install two anchor bolts into the drilled holes.


a Use the blow bulb and vacuum again prior to installing the anchor bolts
to ensure that both holes you drilled are clean and free of debris.
b Insert the anchor bolts into the holes, and tap them down flush with the
table base plate using a hammer.
c Tighten the anchor bolts using a 3/4” wrench.
5 Pull the barrier strips up out of the epoxy and set them aside.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 219
GEMINI GXL Installation Technical Manual Anchoring the Patient Table

6 Remove the four pour zones from the outside of the floor template.
a Using a hammer and screwdriver, tap and/or pry the four rectangular
fill areas away from the sides of the floor template.
b Discard the pieces of epoxy and foam as they are removed.

Pour Zone

CAUTION When removing the pour zones from the Patient Table floor template,
use caution that the tools you are using do not damage the floor. Failure to exercise caution may
result in significant damage to the facility flooring.
7 Clean up the edges of the cured Micorox slab.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 220
GEMINI GXL Installation Technical Manual Anchoring the Patient Table

a Trim any excess foam or jagged epoxy


from the base of the separation unit using a
razor knife.
The object is to make the outside edge of the
poured epoxy look as clean and finished as
possible.

Cured Micorox Slab Trimmed and


Cleaned Edge

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 221
GEMINI GXL Installation Technical Manual Anchoring the Patient Table

8 Drill the remaining patient table anchor holes.


a Select the remaining primary anchor locations.
b Using the drill and masonry bit, drill through the selected
primary holes into the floor.

Anchor Locations Anchor Locations


(Primary and (Primary and Secondary)
Secondary

NOTE If you hit rebar while drilling, select the closest secondary anchor location and use it to set the floor anchor.
c Keep the vacuum on and positioned near the drill bit while drilling to extract dust and debris.
d Pause frequently, remove the drill bit from the hole(s), and use the blow bulb and vacuum to keep the inside
of the hole(s) free of debris.
e Continue to drill until the tape on the drill bit is even with the top of the separation unit spot face or the depth
gauge indicates 4 1/2” (11.5 cm).
Table anchoring will include a total of six anchors. One anchor will be installed at each of the four table corners in
either the primary or secondary locations. Two additional anchors will be installed inboard of the foot end lifting
device at both sides of the patient table. Primary and secondary anchor locations are provided in case you hit
rebar when drilling.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 222
GEMINI GXL Installation Technical Manual Anchoring the Patient Table

9 Install the remaining anchor bolts.


a Use the blow bulb and vacuum again prior to installing the anchor
bolts to ensure that the drilled holes are clean and free of debris.
b Insert the anchor bolts into the holes, and tap them
down flush with the table base plate using a
hammer.
c Tighten the anchor bolts using a 3/4” socket wrench.
10 Clean up the work area around the patient table.
a Pick up the paper and cardboard used to protect the
floor around the Patient Table.
b Remove any tape left on the floor, and vacuum up any remaining
dust or debris.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 223
GEMINI GXL Installation Technical Manual
Calibrations and Performance Verification

Calibrations and Performance Verification

Calibration Procedures
Once the GEMINI has been mechanically installed and powered up, the system must be tested for proper
operation before completing the remainder of the installation. To verify proper operation and ensure performance
standards are met, a series of calibrations and performance verification procedures will be performed. Follow the
work flow charts and timeline below, and refer to the GEMINI Calibration Technical Manual (P/N 4535 671
38781) for the detailed instructions on how to perform each calibration and test. Once each calibration or
performance verification procedure is complete, refer back to these charts for the next procedure.

PET Performance Verification Procedures


After all of the calibration procedures have been completed, the system must undergo performance verification
testing. Use the following work flow chart as a guide and refer to the GEMINIGXL Calibration Technical Manual
(P/N 4535 671 38781) for the detailed instructions on how to perform each verification procedure.

CT Laser Registration
Refer to the GEMINIGXL System Laser Alignment procedure located in the GEMINI Calibration Technical
Manual(P/N 4535 671 38781) for information regarding the CT lasers.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 224
GEMINI GXL Installation Technical Manual

GEMINI System Networking

Personnel Required Time Required

2 Field Service Engineers 3 Man-Hours (1 Lead and 2 Assistant)


The PET and CT systems will have the proper settings for the networking values for the CT and PET system
computers set during the manufacturing process. The patient table will come with its proper default values as well.
This procedure describes how to check, and only if necessary, change the system networking. The default values
should not be changed on the GEMINI private network. The only changes required will be for site assigned IP
addresses. These would include the following:
• The second Ethernet card in the PETview server.
• The site assigned IP addresses for the CT host.
• The hospital router, if necessary.
• The extra PETview workstation., if ordered
• The extra Brilliance workstation., if ordered.
NOTE The CT Host computer has three IP addresses; one for the Recon computer, one for the GEMINI private network,
and one for the hospital backbone (building) network. The PETview workstation has two IP addresses; one for the
GEMINI private network and one for the hospital backbone (building) network. For DICOM configuration, the CT
Host computer and the PETview workstation should be set to the hospital network. Any external DICOM devices
should be configured to communicate to the site-assigned external IP's given to the CT Host and PETview
workstation.
Diagnostic X-Ray System Assembly Report

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 225
GEMINI GXL Installation Technical Manual

Diagnostic X-Ray System Assembly Report


Personnel Required Time Required

2 Field Service Engineers 6 Man-Hours (4 Lead and 2 Assistant)


Once the GEMINI system has been installed, it is necessary that you complete and submit the following forms.
These are required anytime a diagnostic X-ray system has been installed and when a part has been replaced
during service.

For Installations performed within the U.S.:


It is a requirement that FDA Form (2579) Report of Assembly of a Diagnostic X-Ray System, be filled out and
submitted to your local office. The local service office may have additional forms that need to be completed
depending on state & local codes.

For International Installations:


Contact your local service office for proper paperwork and forms that may need to be completed and submitted to
the appropriate authorities.
Packaging and Shipping Returnables

Packaging and Shipping Returnables


It is important to remember that the cost of each set of GEMINI transporters remains charged to each Service
Zone until the transporters are returned to the factory. Authorized Return tags will be shipped with each domestic
GEMINI system.
Personnel Required Time Required

2 Field Service Engineers 6 Man-Hours (4 Lead and 2 Assistant)

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 226
GEMINI GXL Installation Technical Manual

PET Gantry Roll-Lifts


PROCEDURE

1 Disassemble the PET Gantry modified Roll-Lifts.


2 Prepare the modified Roll-Lifts for shipping.

CT Gantry Swivel Casters


PROCEDURE

1 Gather the CT Gantry Swivel Casters.


2 Prepare the CT Gantry Swivel Casters for shipping.

GSU Transport
PROCEDURE

1 Disassemble the GSU Transport, if not already done.


2 Prepare the GSU Transport for shipping.
Patient Table Transport
PROCEDURE

1 Disassemble the Patient Table Transport.


2 Prepare the Patient Table Transport for shipping.

4535 671 38811 Rev. B Philips Medical Systems Confidential and Proprietary Information 227
GEMINI GXL Installation Technical Manual

Clean Up
PROCEDURE

1 Ensure that the entire work area is clean and free of debris.
2 Perform a tool inspection to ensure that no tools will be left behind or lost.
3 Ensure that you have provided the customer with all of the necessary documentation, including the Diagnostic
X-Ray System Assembly Report, Performance Verification Worksheets, and any additional paperwork from
subassembly vendors.

GEMINI Customer Acceptance and Image Alignment


After completing the Calibrations and Performance Verification, you must next perform the Customer Acceptance
and Image Alignment procedure. The Customer Acceptance and Image Alignment procedure tests the combined
imaging ability of the GEMINI system and is performed anytime an existing system has been serviced or a new
system is installed.
The Customer Acceptance and Image Alignment procedure is located in the GEMINIGXL Calibration Technical
Manual(P/N 4535 671 38891).