Professional Documents
Culture Documents
Pacing leads
Model 2088TC
User's Manual
CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician.
WARNING: This product can expose you to chemicals including ethylene oxide, which is known to the State of California to
cause cancer and birth defects or other reproductive harm. For more information, go to www.P65Warnings.ca.gov.
™ Indicates a trademark of the Abbott group of companies.
‡ Indicates a third-party trademark, which is property of its respective owner.
Pat. http://www.abbott.com/patents
© 2023 Abbott. All Rights Reserved.
Device Description
The Tendril™ STS Model 2088TC leads are bipolar, steroid-eluting, active fixation implantable leads with
Optim™ insulation that can be placed in the right atrium, right ventricle, or left bundle branch area for
pacing and sensing. Certain Tendril STS Model 2088TC lead lengths have been tested for use in the MRI
environment and are designated MR Conditional. See MRI Safety Information (page 2).
The active fixation leads use an extendable-retractable helix at the tip. The design of the tip aids visibility
under fluoroscopy. The leads’ insulation tubing is treated with Fast-Pass™ coating to creates a lubricious
surface to facilitate insertion.
Features of Tendril STS Model 2088TC leads include:
▪ Soft Tip Header — reduces lead tip pressure
▪ Optim™ insulation — a silicone-polyurethane copolymer
▪ Tantalum Marker Ring — to facilitate optimal lead positioning
▪ Active Fixation — features a rotating, extendable/retractable helix for secure anchoring
▪ Steroid Elution — Dexamethasone sodium phosphate (DSP) is slowly released through the tip electrode
upon contact with the body fluid. The drug is intended to promote low chronic stimulation threshold by
suppressing the local inflammatory response to a foreign body. The target dose of DSP in the monolithic
controlled device (MCRD) is 460 micrograms.
Indications
Tendril™ STS leads are indicated for use in combination with a compatible pacemaker, implantable
cardioverter defibrillator (ICD) or cardiac resynchronization therapy (CRT-P/CRT-D) device to provide sensing
and pacing for the management of chronic symptomatic bradycardia and various atrioventricular conduction
abnormalities in patients who experience syncope, presyncope, fatigue, or disorientation due to
arrhythmia/bradycardia, or any combination of these symptoms. The Tendril STS leads are implanted
transvenously in either the right atrium, the right ventricle or the left bundle branch area.
For indications for the pacing system, please refer to the applicable pulse generator manual.
Intended Use
Tendril™ STS Model leads are bipolar, steroid eluting, active fixation implantable leads intended for use with
an implantable pulse generator to provide long-term cardiac pacing and sensing in either the right atrium,
and/or the right ventricle. They are also intended for long-term sensing and pacing in the left ventricular
bundle branch area as an alternative to right ventricular pacing.
Table 1. Lead accessories and their intended purpose
Accessories Intended Use
Suture sleeve Protect the lead from damage when it is secured to the venous entry site.
Vein pick Lift and dilate the vein at the lead entry site.
Lead cap Insulate and protect the lead connector when it is not connected to a
device.
Helix Locking Tool Extend and retract the helix of an active-fixation lead.
NOTE: These and other accessories may be ordered separately. For more information, contact your local
Abbott Medical representative.
1
Intended Clinical Benefits
Tendril™ STS low voltage transvenous pacing leads provide sensing and pacing for the therapeutic
management of chronic symptomatic bradyarrhythmia and various atrioventricular conduction
abnormalities when used in combination with a compatible pulse generator system to relieve symptoms of a
slow heart rate and restore normal blood circulation.
Tendril™ STS pacing leads, when used in combination with a compatible pulse generator system, provide the
following clinical benefit:
▪ sensing and pacing for the therapeutic management of chronic symptomatic bradyarrhythmia and
various atrioventricular conduction abnormalities.
46 MR Conditional
52 MR Conditional
58 MR Conditional
65 Untested
100 Untested
Contraindications
Tendril™ STS Model 2088TC leads are contraindicated:
▪ in the presence of tricuspid atresia (if the lead is to be positioned in the right ventricle or left bundle
branch area).
▪ for patients with mechanical tricuspid valves (if the lead is to be positioned in the right ventricle or left
bundle branch area).
▪ in patients who are expected to be hypersensitive to a single dose of one milligram of dexamethasone
sodium phosphate.
For contraindications for the pacing system, please refer to the applicable pulse generator manual.
Package Contents
Each package contains product literature and the following sterile components:
▪ One lead
▪ One tip retainer (to be removed prior to implant)
▪ One radiopaque suture sleeve attached to lead
2
▪ Accessories
– One vein pick
– Two clip-on tools
– Stainless steel stylets with knob colors designating degree of firmness and shape. See Stylet Color
Codes (page 18).
The non-pyrogenic label is applicable to the lead and introducer catheter only and is not applicable to any
other component contained in the lead package.
3
Maintaining Device Effectiveness
Device Storage
▪ Store the lead and accessories at room temperature between 68°F and 77°F (20°C and 25°C). Excursions
are permitted between 59°F and 86°F (15°C and 30°C). During transportation and handling, the lead and
accessories can be exposed to temperature excursions between 23°F and 122°F (-5°C and 50°C) that last
up to 24 hours.
▪ The lead package has been sterilized with ethylene oxide for direct introduction of the inner tray into
the surgical field.
▪ Before the package is opened, inspect it visually for any damage that may have compromised sterility.
Sterilization Instructions
The package contents have been sterilized with ethylene oxide before shipment. The lead and accessories
are for single use only and are not intended to be resterilized.
▪ If the sterile package has been compromised, contact Abbott Medical.
Lead Implantation
▪ Check the "Use‑By" date on the package label. Do not implant a lead if the "Use-By" date has expired.
▪ Before opening the lead package, confirm that the lead is compatible with the device to be implanted.
▪ Lead implantation should be performed only when proper emergency facilities for cardioversion and/or
defibrillation are available.
▪ The manipulation of any and all hardware while in the vascular system should only be performed under
continuous fluoroscopic monitoring.
▪ During this procedure it is advisable to also have echocardiographic equipment available.
▪ If subclavian venipuncture is used for lead introduction, it is important to insert the lead as lateral as
possible during entry of the lead into the vein.
▪ If removal of the suture sleeve is necessary, use caution when sliding over the electrode rings as lead
damage can occur.
▪ Failure to use the suture sleeve to secure the lead may result in lead dislodgment or in damage to the
lead’s insulation and/or conductor coil.
▪ Pay close attention to the handling of the helix extension/retraction mechanism before and during
implantation.
Conformance to Standards
The lead complies with IS-1 connector standard ISO 5841-3.
Testing the Mechanical Operation of the Helix with the Clip-On Tool
For information on using the clip-on tool, see Securing the Tip (page 9).
6
Using the Lead Introducer
If a lead introducer is used, follow the instructions provided with the introducer.
CAUTION:
▪ If using a percutaneous lead introducer with a hemostasis valve, make sure the valve allows for
appropriate passage of the lead without damaging the lead body.
▪ Be certain the vein pick does not puncture the insulation of the lead. This could prevent proper
lead function.
▪ Do not use excessive force while inserting the stylet.
▪ When subclavian venipuncture is used for lead introduction, it is important to insert the lead as
lateral as possible during entry of the lead into the vein.
▪ Avoid positioning the lead so that it becomes sharply bent or subjected to tension.
▪ Do not grip the lead with surgical instruments.
▪ Do not leave a lead unconnected in a patient unless the lead is capped.
7
10. Ask the patient to breathe deeply and check that the lead keeps its J-shape.
11. Ask the patient to cough to ensure that the electrode is securely anchored.
8
Figure 5. Opening the clip-on tool
1. Insert lead into notch
4. After the fixation site has been selected, hold the lead body stationary in one hand and turn the Helix
Locking Tool in the direction of the arrow marked “EXTEND HELIX” (clockwise).
NOTE: See Technical Specifications (page 16) for the approximate number of turns required. On the
fluoroscopic image, the helix will be extended beyond the marker ring.
5. PSA cables may be attached in the zones depicted below.
6. If necessary, retract the helix by holding the lead body stationary in one hand and turning the Helix
Locking Tool in the direction marked “RETRACT HELIX” (counterclockwise).
10
Figure 9. Helix Locking Tool retraction
11
Figure 10. LBBAP lead placement
WARNING: Excessive torque and/or tip pressure may cause acute trauma to the endocardium and/or
septum, including possible perforation, which may result in temporarily high or low impedance, or high
capture threshold values. Monitor torque transfer, as it may vary from patient to patient.
CAUTION: During implant procedure, if septal perforation is observed, stop rotating the lead body and
follow the next section Acute removal or repositioning of the Lead (page 13).
CAUTION: Without the HLT, the helix will inadvertently be retracted when the lead body is rotated
clockwise. Check for helix retraction periodically on fluoroscopy and when an increase in impedance is
observed.
CAUTION: If torque is not transferring effectively as the lead body is being rotated, do not continue to
rotate the lead body. Further rotations can result in damage to the helix or make it difficult to remove.
Reposition the catheter and reattempt burrowing.
12
NOTE: It is also important to minimize the friction within the catheter by keeping it free of too many
kinks, turns, sticky contrast etc. This is important to allow torque to be transmitted properly.
Intraoperative Measurements
NOTE: The stylet should be removed before testing the lead for mechanical stability or making final
intraoperative measurements.
It is important to verify stimulation threshold and sensing capability during implantation. A pacing system
analyzer (PSA) is recommended for these electrical measurements.
A low threshold value and high intracardiac signals are signs that the lead has been positioned satisfactorily.
WARNING: A pacing lead inserted into the heart presents a direct, low-impedance pathway for current
flow to the myocardium. Use only battery-powered test equipment for electrical measurements.
Recommended Values
If the initial measurements are different from those in the Recommended values table (page 13), it is best to
wait a while and then repeat the measurements. If the values do not stabilize at an acceptable level it may
be necessary to alter the position of the electrode tip.
Table 4. Recommended values measured with a PSA
Atrium Ventricle1
Acute Stimulation Threshold ≤1.5 V ≤1.0 V
at default pulse width
13
Table 4. Recommended values measured with a PSA
Atrium Ventricle1
Acute Sensing Amplitude ≥2 mV ≥5 mV
Bipolar Pacing Impedance 250-2000 Ω 250-2000 Ω
1
includes LBBAP
1. Vein
2. Fascia
3. Lead body
4. Suture sleeve
14
CAUTION:
▪ If an accessory suture sleeve is used, it should be used as a replacement to the existing suture
sleeve on the lead and should not be used in conjunction with the existing suture sleeve. The use
of multiple suture sleeves to secure the lead is not recommended as it may result in damage to the
lead.
▪ Use the suture sleeve to distribute the tension created by the suture. Failure to use the suture
sleeve may result in damage to the lead’s insulation or conductor coil.
▪ If removal of the suture sleeve is necessary, use caution when sliding over the electrode rings as
lead damage can occur.
▪ Suture sleeve sticking can occur. If this occurs, carefully twist the sleeve off the ring toward the
connector pin; pulling the suture sleeve when it is positioned over the electrode ring may cause a
tear in the lead body near the electrode ring.
▪ Do not tie the suture around the suture sleeve and lead too tightly, as this may result in excessive
stress applied to the lead body.
If an accessory suture sleeve is used:
1. Carefully open the slit in the sleeve and position the sleeve on the lead body.
2. Follow the procedure above to suture and secure the sleeve.
1
Furman S, Hayes DL, Holmes DR: A Practice of Cardiac Pacing. 3rd ed. New York: Futura Publishing, Inc.; 1993:286-289.
Belott, PH, Reynolds, DW. Permanent Pacemaker and Implantable Cardioverter-Defibrillator Implantation. In: Ellenbogen
KA, Kay GN, Wilkoff BL, eds. Clinical Cardiac Pacing and Defibrillation. 2nd ed. Philadelphia, PA: WB Saunders;
1995:613-615.
15
▪ UDI number
3. Fill out the corresponding fields on the Patient Identification Card:
▪ Lead Model & Length
▪ Implant Location of the lead (e.g., RA, RV, LBBAP)
Patient Follow-up
Give the completed card to the patient. This will serve as a temporary card until Abbott Medical mails a
permanent card directly to the patient. To obtain a replacement card if a patient loses or damages their
card, contact Abbott Medical Technical Support (page 16).
Lead Extraction
CAUTION:
▪ Lead extraction carries with it clinical risk. If a pacing lead must be removed due to infection or
other serious reason, great care should be exercised.
▪ A pacing lead explanted for any reason should never be implanted in another patient.
If the lead or any portion of it is extracted, return it to the manufacturer.
It is generally recommended that a chronically implanted endocardial pacing lead not be repositioned
except in special circumstances.
Explantation
After use, this product may be a potential biohazard. If the lead or any portion of it is extracted, handle it
according to local regulations. Clean the explanted device with disinfectant and return it to Abbott Medical
for investigation and safe disposal. For safety reasons, it is recommended that all used leads be enclosed in a
protective cover.
Please complete an Out of Service/Explant/Patient Death form and return it to Abbott Medical with the
explanted device. Whenever possible, send along a printout of the programmed settings of the pulse
generator.
Technical Support
Abbott Medical maintains 24‑hour phone lines for technical questions and support:
▪ 1 818 362 6822
▪ 1 800 722 3774 (toll-free within North America)
▪ + 46 8 474 4147 (Sweden)
▪ + 61 2 9936 1200 (Australia)
▪ medical.abbott/manuals
For additional assistance, call your local Abbott Medical representative.
Any serious incident should be reported to Abbott Medical and the FDA.
Technical Specifications
2
Abbott Medical IS‑1 lead connectors are compatible with connector cavities that conform to the international connector
standard ISO 5841‑3.
3
Compatible with introducers such as those manufactured by Daig and Pressure Products.
16
Table 5. Nominal Specifications
Maximum diameter of lead 1.9 mm
body
Fixation mechanism active helix
Electrode configuration
Tip helix
Ring cylindrical
Electrode spacing
Tip to ring 10 mm
Electrode surface area
Tip 7.0 mm2
Ring 15.3 mm2
Electrode length
Tip (extended) 2 mm
Electrical resistance
Tip to connector pin 46 cm: 14-24Ω; 52 cm: 16-28Ω; 58 cm: 19-31Ω;
65 cm: 21-35Ω; 100 cm: 35-57Ω
Ring to connector ring 46 cm: 29-49Ω; 52 cm: 33-55Ω; 58 cm: 37-61Ω;
65 cm: 42-68Ω; 100 cm: 63-105Ω
Materials
Conductors multifilar MP35NLT
Connectors stainless steel
Lead body insulation Optim™ 4 insulation with exterior Fast-Pass™ coating
Suture sleeve silicone rubber
Tip and ring electrodes titanium nitride-coated platinum iridium alloy
Soft tip silicone rubber
Steroid-eluting plug silicone rubber with 460 micrograms of dexamethasone sodium phosphate
(DSP)
Stylet stainless steel
Lead cap silicone rubber
4
Optim is a silicone-polyurethane copolymer.
5
Lead length, stylet configuration, and differences in anatomy may cause variations in the number of rotations required
to extend the helix.
6
Not to exceed 35 rotations.
7
Not to exceed 35 rotations.
8
Not to exceed 35 rotations.
17
Stylet Color Codes
Table 6. Stylet color codes
Knob Color Description Diameter
Green Soft straight (20 mm taper) 0.014 in./0.35 mm
Light green Extra-soft straight (40 mm taper) 0.014 in./0.35 mm
Yellow Firm straight 0.015 in./0.38 mm
Red Extra firm straight 0.016 in./0.41 mm
Green, white dot Soft (20 mm taper) 0.014 in./0.35 mm
Symbols
The symbols below and harmonized symbols may be found on the product or product label. For
harmonized symbols, refer to the Universal Symbols Glossary at medical.abbott/manuals.
Symbol Description
Accessories
Product literature
Product of USA
Product of Malaysia
Patient identification
Date
Physician telephone
Lead model/length
Location of implant
18
Symbol Description
Double sterile barrier
Pacing lead
MR Conditional
19
abbott.com
2023-12
ARTEN600320887 A
*600320887*