OTY-12001(2.0)

How to initialize the CODE 39 (without checking) sample barcode for BS-400
I. If the CODE 39 sample barcode has fixed digits, you could do as follow:
1. Barcode initialization
Run the Debug software, and choose “Bar code”“Sample Bar code”, and then click the “initialize”, as the
picture below:

Fig1. Barcode initialization


Note: If "Bar code initialization failure" pops up in this procedure, please check cable connection between the
barcode reader and the Three Disk Drive Board.
2. Shake hands
This step is for checking the communication is OK or not. After clicking "shake hands", if the instruction area
shows the result as "{<K100,5,1,0,1>}", it means the communication succeeds. If not, please check the cables
and try again.

No.: ZSH-12075 Version:1.0


Bio-Chemistry Global Technical Support Dept.

Fig2. Shake hands

3. Preset the Symbology and the length in the Operation Software.

Fig3. Preset the Symbology and the length in the Operation Software.

4. Scan the sample barcode.

No.: ZSH-12075 Version:1.0


Bio-Chemistry Global Technical Support Dept.

Fig4. Scan the sample barcode

No.: ZSH-12075 Version:1.0


Bio-Chemistry Global Technical Support Dept.
OTY-12001(2.0)

II. If the CODE 39 sample barcodes have different digits, you could do as follow:
1. Set the Format and Symbology of the barcode in the Operation Software.

Fig5. Set the Format and the Symbology

2. Log off the operation software.

3. Run the debug software, and send the instruction of initialization.

Fig5. Send the instruction directly

No.: Version:1.0
Bio-chemistry Global Technical Support Dept.
OTY-12001(2.0)
4. Save the instruction.

Fig6. Save the setting.


5. Run the Operation Software and scan the sample barcodes, then the CODE 39 sample barcodes
with different digits (without checking) will be displayed.

Fig7. The CODE 39 sample Barcode with 5 digits (without checking) is scanned

Fig8. The CODE 39 sample Barcode with 7 digits (without checking) is scanned too
NOTE:
1) The letters in the instruction“<K470,1,0,0,0,0,0,1>” and “<Z>” are capital letters, not lowercase letters;
2) After setting according to above, only CODE 39 sample barcode could be scanned for this BS-400. If you
want to scan the other kind of barcode, you could execute the initialization again as the Fig.1.above, because
these two statuses could not coexist.