How to Connect an

Incubator’s CO2 Tank


INSTRUCTIONS
Internal Tool Only.

MCO-230AIC Series | MCO-170AIC Series

1 2

1.1 - Remove the side panel 2.1 - Retrieve your two stage regulator*
1.2 - Turn the power switch ON
*Note: We recommend using this type of regulator to deliver con-
1.3 - Allow the incubator to run to 37°C before sistent 5psi pressure to the incubator. Two stage regulators are
connecting the CO2 supply. the best option for incubators because they provide a consistent
pressure as the tank is emptying.

3 C

A B

CO2 Tank with Two Stage Regulator


3.1 - Make sure to use the fiber washer (included
with the regulator) to avoid leaks where it VALVE NAME
CLOCKWISE COUNTERCLOCK-
DIRECTION WISE DIRECTION
connects to the tank
A
3.2 - Important: CLOSE the outlet valve completely Outlet Valve Closes the valve Opens the valve
& OPEN pressure regulator* B Tank
Closes the valve Opens the valve
Shut-off Valve
*Note: Over pressurizing is a common issue when changing
C
tanks on the CO2 incubators and can cause the lines to pop off Pressure Decreases the
the incubator. Increases the pressure
Regulator pressure

powered by
HOW TO CONNECT AN INCUBATOR’S CO2 TANK
Models: MCO-230AIC | MCO-170AIC

4 5

4.1 - Use the tubing supplied with the incubator or a 5.1 - Connect the tubing to the tank and the inlet on
tube with 1/4” inside diameter the incubator

6 7

6.1 - Secure tubing the clips (included in the 7.1 - Open the tank & adjust the pressure regulator
incubator kit) so that the gauge reads 5psi
6.2 - Press the clips together so that the tubing is 7.2 - Open the outlet valve all the way to make sure
less likely to pop off the CO2 gas is being delivered to your incubator

Watch us on YouTube!
For more information, please watch our YouTube
video on How to Connect an Incubator’s CO2 Tank.

https://www.youtube.com/watch?v=qDafZv_83tM

Important: Every time you change the tank make


sure to follow this process to avoid blowing off tub-
ing which is a very common issue when connecting
CO2 gas to incubators.