Professional Documents
Culture Documents
EZ-200
Instruction Manual
Ver.1.43
Table of Contents
Precautions・・・・・・・・・・・・・・・・・・・3
Measuring Aw of Samples・・・・・・・・・・・・13
Maintenance ・・・・・・・・・・・・・・・・・22
Dialog boxes・・・・・・・・・・・・・・・・・・36
Contact Details・・・・・・・・・・・・・・・・・47
WATER ACTIVITY METER EZ-200
Aw) of food and other materials. Use this meter only for this application.
- Temperature 15-35℃.
A place where temperature fluctuation is small (free from direct sunlight, free from air from an
* It is recommended that the water activity meter be placed in a thermostat. By reducing the
temperature difference inside and outside the thermostat chamber, it is possible to reduce
the temperature change during opening and closing of the thermostat chamber.
A place where the humidity is not drastically higher than the Aw of a sample to be measured.
* For example, when you measure a sample of 0.50 Aw, it is recommended that the water
activity meter be placed in a location where humidity does not exceed 50%RH (equivalent of
0.50 Aw).
Note
Temperature affects the Aw. The Aw cannot be measured correctly in a location where
temperature fluctuates.
This water activity meter is calibrated at 25ºC. Although this water activity meter can be used at
the temperature range of 15 to 35ºC, for more accurate measurement, it is recommended that
1
WATER ACTIVITY METER EZ-200
Bottom of
the sample
chamber
Hinge
Fan
Touch panel
DC plug
Inlet air filters DC power jack AC cable AC adapter
2
WATER ACTIVITY METER EZ-200
Precautions
Danger
This symbol indicates that improper use of this water activity meter may cause a
possibly dangerous situation that could result in injury or severe property damage.
- Do not insert the adapter into a socket of other than 100 to 240 V (rated voltage).
- Improper use of this water activity meter may cause fire or a failure.
Top lid
- The top lid of the water activity meter is designed to be heavy and to have a slightly loosened
hinge, since the top lid completely seals the sample chamber. Use both hands when opening
and closing the top lid of the water activity meter, and be careful not to get your hands
caught.
Fan
- Be careful not to get your hair, etc. caught in the fan.
- Improper use of this water activity meter may result in injury or a failure.
3
WATER ACTIVITY METER EZ-200
Caution
This symbol indicates that improper use of this water activity meter may cause damage
or failure to the meter or a decrease in the accuracy of Aw measurement results.
- This water activity meter is a precision device. Application of strong shocks or disassembly
- Application of strong shocks on the touch panel may break the panel.
- If the packing that seals the sample chamber tightly is dirty, scratched, damaged, or
abnormalities such as contact failure to the bottom of the sample chamber, it cannot seal
the sample chamber completely. As a result, the elapsed time until completion of the
measurement (hereinafter called the measurement time) may be extended, and the
replace it (see page 31). Apply the supplied grease on the packing to seal the sample
chamber completely and check the contact condition of packing daily (see page 6).
- Be careful not to let samples (such as saturated salt solution) adhere to the fan. If any sample
- Be careful not to let samples adhere to the sensor. Do not move the water activity meter
while any sample are placed in the sample chamber. It may break down or the sensor may
deteriorate if any sample adheres in the sensor. If the sensor is covered with any sample, the
- Leaving any sample to stand as they are in the sample chamber for a long time will promote
- Acid gas causes deterioration of the sensor. Measure the Aw of saturated salt solution daily
(see page 24) to check the status of the sensor. Check there is no abnormality in packing
saturated salt solution and the theoretical value (see page 25). Use of an optional sensor
protection filter (see page 45) is recommended if the sensor is deteriorated significantly.
4
WATER ACTIVITY METER EZ-200
- Be careful not to apply strong shocks on the sensor when replacing the sensor or applying
the sensor protection filter. If the sensor is scratched, bent, cut, or otherwise damaged, the
Sensor board
Humidity sensor
Temperature
sensor
- If the cable connected to the board is subjected to a load, such as pulling or twisting, during
- In the USB port, USB flash derive is for exclusive use. Do not be connected to the PCs with a
USB cable.
・Avoid using thinner, organic solvents, strongly acidic substances, and other substances for
cleaning. Using these chemicals may cause a failure, damage to the exterior, rust, and
Note
This water activity meter is designed to perform measurement under the condition where the
humidity (Aw) inside the sample chamber is changed from low to high. Particularly for a
sample of approximately 0.60 Aw or less, drying the inside of the sample chamber with silica
5
WATER ACTIVITY METER EZ-200
stabilized before taking measurement. In addition, configure the necessary settings for the
measurement.
1. Place the water activity meter in a location where temperature fluctuation is small, to stabilize
Note
2. Open the top lid to confirm that the packing attached to the sample chamber resin, and the
bottom of the sample chamber that contacts with the packing are not dirty. Apply grease on
the packing (grease is applied when delivered). The property quantity of grease is 0.1 g (the
one cutting with the small medicine spoonful of the attachment). Closes and press the top lid,
then recognize the grease stuck in bottom of the sample chamber. The state that the 2-3 mm
6
WATER ACTIVITY METER EZ-200
Danger
Use both hands when opening and closing the top lid of the water activity meter, and be
Caution
If the packing that seals the sample chamber tightly is abnormal, it cannot seal the sample
chamber completely. As a result, the measurement time may be extended, and the
3. Connect the AC cable to the AC adapter of the power adapter for EZ-200. After inserting the
AC cable into the socket and inserting the DC plug into the jack on the back of EZ-200, turn
AC adapter
DC plug
AC cable
Danger
Use only the power adapter for EZ-200.
4. After the opening screen is displayed for four seconds, display changes to the measurement
screen. The measurement screen displays the measured values of Aw and temperature.
7
WATER ACTIVITY METER EZ-200
1.43
5. Leave the water activity meter to stand for 30 minutes with the top lid closed to complete the
warm up.
6. Touch the [Setting] button on the measurement screen to move to the setting screen.
Furthermore, move to various setting screens to configure necessary settings for the
measurement (this can be done while warming up). Touch the [Back] button to return to the
setting screen from various setting screens. Turning the power OFF does not reset the setting
details.
8
WATER ACTIVITY METER EZ-200
Aw indication
Touch the [Aw indication] button on the setting screen to move to the [Aw indication] screen.
The water activity meter shows the Aw value with two decimal places. The rounding method of
the third decimal place can be selected from the following three methods: rounded, rounded
The factory default is "rounded," and the screen displays "rounded." Touch the "rounded
screen.
The date and time is displayed in the following format: the last two digits of A.D./Month/Day
To change the value, touch the part that you want to change to display the numeric keypad
input screen. Enter the value with the numeric keys and touch the [ENT] key to complete the
change in value.
9
WATER ACTIVITY METER EZ-200
Buzzer
To move to the buzzer setting screen, touch the [Buzzer] button on the setting screen.
Set the ring time (0 to 99 seconds) and the buzzer volume (with four levels from OFF, Small,
Medium, to Large).
The factory default of the ring time is 5 seconds and that of the ring volume is Medium.
To change the value, touch the value of the ring time to display the numeric keypad input
screen.
10
WATER ACTIVITY METER EZ-200
Language
To move to the language change screen, touch the [Language] button on the setting screen.
Measurement Mode
To move to the measurement mode setting screen, touch the [Meas. mode] button on the
setting screen.
The factory default of the measurement mode is Normal; [Normal] is indicated on the button,
Switch to the dedicated mode when using the sensor protection filter (optional).
Touch the [Normal] button to indicate [Sensor protection filter] in the button. Switching is
For the application method of the sensor protection filter and the measurement method, see
page 45.
11
WATER ACTIVITY METER EZ-200
The currently set measurement mode can be checked in the measurement screen.
That is the preparation of the water activity meter. Carry out the Aw measurement of saturated
salt solution (see page 24) before taking the Aw measurement of the samples if necessary. If
there is a discrepancy between the result of Aw measurement of saturated salt solution and
the theoretical value, check there is no abnormality in packing and calibrate the Aw (see page
25).
12
WATER ACTIVITY METER EZ-200
Measuring Aw of Sample
1. Prepare sample.
Close the lid and store until the container reaches the same temperature as the water activity
meter.
2. Remove the lid of the sample container and open the top lid of the water activity meter. After
placing the sample in the sample chamber, close the top lid of the water activity meter.
Danger
Use both hands when opening and closing the top lid of the water activity meter, and be
Caution
Incomplete sealing of the sample chamber may extend the measurement time and reduce the
measurement accuracy.
Note
For a sample with 0.60 Aw or less, it is recommended to dry the inside of the sample chamber
with silica gel before taking measurement. (See page 17 "How to dry inside the sample
13
WATER ACTIVITY METER EZ-200
regardless of Aw.
3. To start the measurement, touch the [START] button on the measurement screen.
The [START] button is replaced with the [STOP] button, and the elapsed time (*) counter will
start.
The displays of the Aw and the temperature (both of them are measured values) blink during
the measurement.
Note
There is a tendency that the higher the Aw of the sample, the longer the elapsed time (*) until
reasons include reduced sealing of the sample chamber, unstable temperature during the
measurement, or change in the status of the sensor (status where calibration or sensor
replacement is required).
To stop the measurement, touch the [STOP] button during the measurement. When the [STOP]
button is touched, the screen displays "Measurement Stop" together with the measured values
14
WATER ACTIVITY METER EZ-200
The data when the measurement is stopped are not stored in the history.
To check the state of change in the Aw (measured value) and the temperature (measured
value) from the graph during the measurement, touch the [Graph] button on the measurement
screen during the measurement or after completing the measurement (see page 20).
4. The buzzer signals the end of measurement (the buzzer does not sound if "Ring time of the
buzzer to inform completing the measurement" is set to "0 seconds" or "Buzzer volume" is set
to "OFF"). On the screen, a message is displayed as "Measurement Completed." along with the
measured Aw value and temperature. The elapsed time counter also stops.
button.
page 18).
15
WATER ACTIVITY METER EZ-200
Note
This water activity meter provides a correction function to measure the Aw with high accuracy
and in a short time, and displays the Aw value that is corrected when the measurement is
completed. The temperature displayed is the measured value when the measurement is
completed.
graph data in a USB flash drive, save the data before taking the next measurement (see page
20).
5. When touched, the [RESET] button is replaced with the [START] button to allow measurement
of the next sample. The elapsed time is cleared and shows “0m: 0s.” The displays of Aw and
temperature return to the measured values, and the measurement result (corrected Aw value)
will be cleared.
6. To measure the next sample, remove the sample already measured from the sample chamber
7. Make sure to remove sample from the sample chamber when finished with measurements and
* You can extend the life of the sensor by storing the silica gel in the sample chamber with the
Caution
Leaving any sample to stand as they are in the sample chamber for a long time will promote
16
WATER ACTIVITY METER EZ-200
regardless of Aw
1) Put silica gel in the sample container. With the lid closed, store it under the same
2) Place silica gel in the sample chamber, close the top lid of the water activity meter, and dry
inside the sample chamber in accordance with the "Drying conditions" in the table below.
3) Open the top lid of the water activity meter to remove the silica gel. Place sample and then
4) Touch the [START] button on the measurement screen to take the measurement.
Note
The moisture absorption capacity of silica gel is reduced if dropping of Aw is slow even with
the silica gel placed in the sample chamber. Replace the silica gel with new silica gel.
Among silica gels supplied as accessories, the green one turns to pink and then to orange as it
absorbs moisture. The time to replace silica gel is when the green silica gel turns to pink.
17
WATER ACTIVITY METER EZ-200
data sets and save them). The data saved in the USB flash drive can be managed in a personal
computer. Insert the USB flash drive into the USB port located in the back of the main unit, and
Saving the data on the USB flash drive is completed when a dialog box, "Saved successfully to
Caution
In the USB port, USB flash derive is for exclusive use. Do not be connected to the PCs with a USB
cable.
18
WATER ACTIVITY METER EZ-200
A folder "SAMP01" is created in the USB flash drive. The history data are saved in CSV file format
(P20)
19
WATER ACTIVITY METER EZ-200
them). When the [Clear] button is touched on the history screen, a dialog box, "All data will be
every 8 seconds during the measurement (the data saved are referred to as log data). The screen
is replaced with the graph display screen by touching the [Graph] button on the measurement
screen during the measurement or after completing the measurement. The change in the Aw
(measured value) and the temperature (measured value) during the measurement can be
checked from the graph. Only one measurement of the log data can be saved in this water
activity meter. The log data are reset at the time of taking the next measurement.
20
WATER ACTIVITY METER EZ-200
The log data can be saved on a USB flash drive for each measurement. Insert the USB flash drive
into the USB port, and touch the [Save] button on the graph display screen. Saving the data on
the USB flash drive is completed when a dialog box, "Saved successfully to the USB flash drive.",
appears.
Caution
In the USB port, USB flash derive is for exclusive use. Do not be connected to the PCs with a USB
cable.
A folder "SAMP02" is created in the USB flash drive. The log data are saved in CSV file format in
21
WATER ACTIVITY METER EZ-200
Maintenance
Prepare saturated salt solution
Prepare four types of saturated salt solution as described in the table below.
Aw of Weight of the Weight of
Reagent of saturated salt solution* saturated salt reagent water
No.
(accessories) solution (g/sample (g/sample
(25.0℃) container) container)
Magnesium chloride hexahydrate
C-1 0.328 12.3 1.2
MgCl2・6H2O
Magnesium nitrate hexahydrate
C-2 0.529 10.6 1.2
Mg(NO3)2・6H2O
Sodium chloride
C-3 0.753 10.0 2.2
NaCl
Ammonium dihydrogenphosphate
C-4 0.930 9.6 2.6
NH4H2PO4
* Guaranteed Reagent
Add water (purified water, ion exchanged water, or distilled water) to each reagent, and mix well.
The reagent is in a proper state when it becomes like sorbet and most of the reagent appears
over the liquid level (slightly covered with liquid). Adjust the weights of reagent and water
appropriately.
Before using the saturated salt solution, close the lid of the container and leave it to stand as is
for over an hour under the environment of 24.0 to 26.0ºC, since the Aw of just prepared
The saturated salt solution absorbs moisture if it is placed under an environment where the
humidity is higher than the Aw theoretical value, and dries if it is placed under an environment
where the humidity is lower than the Aw theoretical value. Thus, make sure that the lid of the
22
WATER ACTIVITY METER EZ-200
container is closed firmly when storing it. If the saturated salt solution has absorbed moisture or
- For moisture absorbed saturated salt solution, empty supernatant fluid, add a small amount of
the reagent, mix well, and leave to stand as is for an hour or more.
- For dried saturated salt solution, add water (add the reagent if necessary), mix well, and leave to
23
WATER ACTIVITY METER EZ-200
Measure the Aw of saturated salt solution (daily maintenance)
For checking the status of the sensor, routine measurement of the Aw of the saturated salt
solution is recommended.
Measure in an environment where the sample chamber temperature (panel display temperature)
is 24.0 ° C to 26.0 ° C. Measure the Aw without applying the sensor protection filter (optional).
Check there is no abnormality in packing and calibrate the Aw if there is a discrepancy between
the result of the Aw measurement and the theoretical value listed in the table below.
For example, if your allowable accuracy is ±0.01 Aw, calibrate the Aw when the discrepancy
between the result of the measurement and the theoretical value is ±0.02 Aw or greater.
Theoretical values of Aw of each saturated salt solution at a
temperature between 24.0 and 26.0ºC
Temperature C-1 C-2 C-3 C-4
(℃) MgCl2 Mg(NO3)2 NaCl NH4H2PO4
24.0 0.331 0.535 0.755 0.932
24.1 0.331 0.534 0.755 0.932
24.2 0.331 0.534 0.755 0.932
24.3 0.330 0.533 0.755 0.932
24.4 0.330 0.532 0.754 0.931
24.5 0.330 0.532 0.754 0.931
24.6 0.329 0.531 0.754 0.931
24.7 0.329 0.531 0.754 0.931
24.8 0.329 0.530 0.753 0.930
24.9 0.328 0.529 0.753 0.930
25.0 0.328 0.529 0.753 0.930
25.1 0.327 0.529 0.753 0.930
25.2 0.327 0.528 0.753 0.929
25.3 0.327 0.528 0.752 0.929
25.4 0.327 0.528 0.752 0.929
25.5 0.326 0.527 0.752 0.929
25.6 0.326 0.527 0.752 0.928
25.7 0.326 0.527 0.752 0.928
25.8 0.325 0.527 0.752 0.928
25.9 0.325 0.526 0.752 0.927
26.0 0.325 0.526 0.751 0.927
The values in the above are corrected values by taking the temperature characteristics of the
sensor into consideration.
This table lists the values with three decimal places, since the water activity meter shows the
value with two decimal places, and offers three rounding methods that you can choose from,
"rounded," "rounded down," and "rounded up."
24
WATER ACTIVITY METER EZ-200
1) Place the water activity meter in a location where temperature fluctuation is small, to stabilize
2) Among the four types of saturated salt solutions, prepare the one that is close to the Aw value
of the sample (prepare all four types if the Aw value of the sample is unknown). Put the
saturated salt solution in the sample container, and store it under the same temperature
3) Measure the Aw of the saturated salt solution by the same method as the Aw measurement of
For the saturated salt solutions C-1 and C-2, before taking the measurement, dry the inside the
The Aw cannot be measured correctly if the status of the sensor changes. Calibrate the Aw if
there is a discrepancy between the result of the Aw measurement (daily maintenance) of the
To calibrate the Aw, use the four types of saturated salt solutions, C-1, C-2, C-3, and C-4
(calibrate four points). To measure the sample of the limited Aw range, use two or three types of
the saturated salt solutions as listed in the table below in accordance with the range of Aw
25
WATER ACTIVITY METER EZ-200
Calibrate the Aw in an environment where the sample chamber temperature (panel display
1) Place the water activity meter in a location where temperature fluctuation is small, to stabilize
3) Touch the [Maint.] button on the measurement screen to move to the maintenance screen.
Touch the [Aw cal.] button on the maintenance screen to move to the Aw calibration screen. The
Aw calibration screen displays the voltage value of the humidity sensor (a measured value, the
unit is V) along with the measured value of the Aw and the temperature (the higher the Aw, the
higher the voltage value). The voltage values during the calibration carried out so far are
recorded in "Initial" and "Current." "Initial" is the voltage value before delivery or at the
calibration after replacing the sensor. "Current" is the voltage value at the latest calibration.
26
WATER ACTIVITY METER EZ-200
<Maintenance screen>
4) Place the saturated salt solution of C-1 in the sample chamber, close the top lid of the water
activity meter, and leave it to stand as is for an hour or more until the Aw becomes stable (the
5) Touch the [START] button on the Aw calibration screen to take the measurement.
6) The buzzer sounds when the measurement is completed at the elapsed time of 1m 00ss
(measurement time of 1 minute) (the buzzer does not sound if "Ring time of the buzzer to
inform completing the measurement" is set to "0 seconds" or "Buzzer volume" is set to "OFF").
The screen displays the measured values of the Aw, the voltage value, and the temperature
The data of the measurement result on the Aw calibration screen are not saved in the history.
27
WATER ACTIVITY METER EZ-200
7) Touching the calibration button of [C1 0.328], voltage corrected based on the voltage value
8) Touch the [RESET] button to calibrate C-2, C-3, and C-4 in the same manner. Time to leave the
saturated salt solutions of C-3 and C-4 to stand is two hours or more.
Replace the sensor, including the sensor board. The replacement sensor board is sold as an
optional accessory (see page 47). It is recommended that the packing be replaced at the time of
replacing the sensor (see page 31). Use the new packing included in the replacement sensor
board.
After replace the sensor and packing, confirm the contact condition of the packing, perform
A thermometer for calibration is needed for calibrating the temperature. It is recommended that
28
WATER ACTIVITY METER EZ-200
2) Open the top lid and remove grease adhered to the packing of the sample chamber resin.
3) Turn the sample chamber resin counterclockwise, and remove the sensor board from the top
lid (putting on rubber gloves will make it easy if it cannot be turned with bare hands).
Caution
If the cable connected to the board is subjected to a load, such as pulling or twisting, the cable
may be broken.
Caution
Be careful not to apply strong impact on the sensor.
29
WATER ACTIVITY METER EZ-200
5) Remove the cables from the sensor board. Pull the cables in the B direction with pressing
6) Insert the cables into the new board. Insert the connector in the C direction until it clicks.
7) Fit the connector part with the sensor board rotation preventive groove and fix the sensor
Make sure to calibrate the temperature and the Aw after replacing the sensor.
30
WATER ACTIVITY METER EZ-200
hardening, is found on the packing. The replacement packing is sold as an optional accessory
1) Remove the deteriorated packing (if it cannot be removed with hands, using tweezers helps
remove it).
2) Clean the groove where the packing is attached using waste cloth, and make sure that no
4) Apply plenty of grease on the groove side of the packing to attach the packing.
5) Push the packing on the grooves side with the fingers as a whole and fix it securely.
Check the contact condition of packing not only after replacing packing but as routine
maintenance.
1) Remove grease and dirt on the packing and the bottom of the sample chamber that
3) Close the top lid slowly with both hands (do not press downwards), if the trace of grease has
a uniform width on the bottom of the sample chamber, the contact is normal. If there is an
abnormality, check again whether the sensor board and packing are securely fixed. If
abnormality cannot be corrected, check with the contact details (see page 47) because
inspection is necessary.
4) If packing contact is normal, apply normal amount of grease. The property quantity of
grease is 0.1 g (the one cutting with the small medicine spoonful of the attachment). Closes
and press the top lid, then recognize the grease stuck on bottom of the sample chamber. The
31
WATER ACTIVITY METER EZ-200
Make sure that the sensor board is fixed securely and confirm there is no abnormality in
packing.
Calibrate the temperature without applying the sensor protection filter (optional).
meter.
be placed in a thermostat.
3) When the warm up is completed, touch the [Temp. cal.] button on the maintenance screen
4) When the numerical value of the calibration thermometer is touched, the numeric keypad
input screen is displayed and you can change the numerical value within the range of 24.0 to
26.0.
6) Check that temperature of "Panel display" and the temperature indicated by the
thermometer become the same (±0.1℃), and leave it to stand as is for 10 minutes.
32
WATER ACTIVITY METER EZ-200
7 ) Confirm that the temperatures indicated by "Panel display" and the temperature indicated
become the same. The calibration is completed if the temperature difference is within
±0.1ºC.
If the temperature difference is ±0.2ºC or greater, calibrate the temperature again from step
4). If the temperature difference does not become within ±0.1ºC even after steps 4) to 7 are
carried out three times, it is suspected that the temperature changes of the environment are
large. Take measures such as shielding the water activity meter and the thermometer that
becomes the standard with wind protection, or turn the power of the air conditioner off
temporarily.
33
WATER ACTIVITY METER EZ-200
After replacing the sensor, make sure to perform Aw calibration after calibrating the
temperature. Calibrate the Aw without applying the sensor protection filter (optional).
1) Clear the "Initial" voltage value displayed on the Aw calibration screen. A dialog box, "Clear
initial” should be operated only in the cal. after replacing the sensor. Is it OK if the initial
values are reset?", appears by touching the "Clear initial" button. Touch the [OK] button to
2) Calibrate four points by the same method as the Aw calibration for daily maintenance (see
page 25). In the calibration after clearing the initial, the same voltage value is recorded in
34
WATER ACTIVITY METER EZ-200
water activity meter. Take out the inlet air filter and wash it in water regularly. Turn the power of
the water activity meter OFF when taking out the inlet air filter. Dry the filter with air sufficiently,
cloth dampened with water or neutral detergent diluted with water to wipe off dirt.
Caution
Avoid using thinner, organic solvents, strongly acidic substances, and other substances for
cleaning.
35
WATER ACTIVITY METER EZ-200
Dialog boxes
This water activity meter is designed to display various dialog boxes.
From this section, dialog boxes other than those described above are explained.
<Measurement screen>
(1) "History memory has been exhausted. The oldest data will be deleted. Is that OK?"
This dialog box appears when you touch the [START] button while 100 history data sets are
saved. Touch the [OK] button to take measurement. The oldest data set is deleted when the
If not deleting the history data, this dialog box appears every time you touch the [START]
button. Once the history data are deleted, this dialog box does not appear until the number of
36
WATER ACTIVITY METER EZ-200
This dialog box appears when you move to the measurement screen without calibrating the
Aw and touch the [START] button after carrying out "Clear initial" on the Aw calibration screen.
Calibrate the Aw (calibrate four points) on the maintenance screen. Touch the [OK] button to
This dialog appears when you touch the [Save] button without inserting the USB flash drive.
Insert the USB flash drive into the USB port before touching the [Save] button. Touch the [OK]
Caution
In the USB port, USB flash derive is for exclusive use. Do not be connected to the PCs with a USB
cable.
37
WATER ACTIVITY METER EZ-200
This dialog box appears if a writing error occurs when saving the data in the USB flash drive.
Make sure that the USB flash drive has space, insert the USB flash drive again, and touch the
[Save] button. If the same dialog box appears, use a different USB flash drive. Touch the [OK]
This dialog box appears if an error other than (3) or (4) occurs when saving the data in the USB
flash drive. For example, this window appears when the USB flash drive is removed from the
USB port before saving is completed. Reinsert the USB flash drive, and touch the [Save] button
again. If the same dialog box appears, use a different USB flash drive. Touch the [OK] button to
38
WATER ACTIVITY METER EZ-200
Aw calibration screen
(6) "Cal. should be done under the condition at 24.0~26.0ºC in the sample chamber."
This dialog box appears if the temperature is not between 24.0 and 26.0ºC when you touched
Set the temperature in the sample chamber between 24.0 and 26.0ºC, and then take the
measurement again. Touch the [OK] button to close this dialog box.
restarting."
This dialog box appears when the voltage value detected during the measurement changes
more than a certain value. Extend the time to leave it to stand as is, and take the measurement
again, since the humidity (Aw) in the sample chamber may not be stabilized. Touch the [OK]
39
WATER ACTIVITY METER EZ-200
This dialog box appears if there is more than a given difference between the measured voltage
value and the initial voltage value when you touch any of the [C-1 0.328], the [C-2 0.529], the
[C-3 0.753], or the [C-4 0.930] button. This screen allows you to check whether the touched
This dialog box appears when touching the same calibration button again after touching the
[OK] button in (8). This window also appears if there is a given range of difference between the
measured voltage value and the initial voltage value regardless of the above.
40
WATER ACTIVITY METER EZ-200
A dialog box, "Check the condition of packing.", appears in order to determine whether the
abnormality is found in the packing, a dialog box, "The sensor should be replaced.", appears
(see page 28 for replacing the sensor). If an abnormality is found in the packing, a dialog box,
"The packing should be put right before restarting.", appears. Take the measurement again
after the packing put right. Touch the [OK] button to close this dialog box.
41
WATER ACTIVITY METER EZ-200
This dialog box appears when touching the [Back] button without calibrating the Aw after
carrying out "Clear initial" (see page 34). The Aw cannot be measured without calibrating the
Aw.
This window also appears when touching any of the [C-1 0.328], the [C-2 0.529], the [C-3
0.753], or the [C-4 0.930] calibration button without measuring the saturated salt solution.
Touch the calibration button after measuring the saturated salt solution.
This dialog box appears when touching any of the [C-1 0.328], the [C-2 0.529], the [C-3 0.753],
or the [C-4 0.930] calibration button while stopping measurement of the saturated salt by
Touch the calibration button after measuring the saturated salt solution.
42
WATER ACTIVITY METER EZ-200
Alarm Display
The measurement screen or the Aw calibration screen displays "Measurement Completed" when
the measurement is competed normally. If, however, any error occurs, an alarm listed in the
Contents of abnormality
Alarm
and action after the occurrence of the alarm
It is in a state where a temperature change that exceeds ±1.0ºC
is found after taking the measurement and the measurement is
suspended. Check the following items:
- Whether the water activity meter is placed in a location where
Err. Change of temp.
the temperature change is small.
(Normal ±1.0ºC) - Whether the temperature of the water activity meter is stable.
- Whether the time for warming up is sufficiently secured.
- Whether the temperature of the sample is the same as that of
the water activity meter (within ±1.0ºC)
Measurement time exceeds 30 minutes, measurement is
stopped. Check the following items.
·- Whether the water activity meter is placed in a location where
the temperature change is small.
Over measurement time - Whether the temperature of the water activity meter is stable.
- Whether the time for warming up is sufficiently secured.
- Whether the temperature of the sample is the same as that of
the water activity meter (within ±1.0ºC)
- Whether the packing is normal.
It is in a state where the measurement cannot be taken or the
measurement is suspended because the temperature is not
within 15 to 35ºC. Adjust the temperature of the environment, or
Err. Environmental temp.
move the water activity meter to a location where the
(Normal 15~35ºC) temperature is between 15 and 35ºC. Take the measurement
again after the temperature of the water activity meter becomes
stable.
43
WATER ACTIVITY METER EZ-200
It is in a state where the measurement cannot be taken or the
measurement is suspended because the Aw is not within 0.10 to
0.98. The former case may occur because the inside of the
sample chamber is dried excessively by using silica gel. In this
Err. Meas. Range
case, place the sample in the sample chamber and touch the
(Normal 0.10~0.98 Aw) [START] button after the Aw indicates 0.10 or greater. If there is
no problem with the drying method, the measurement cannot
be taken because the Aw of the sample is out of the
measurement range.
It is in a state where the temperature is not within 24.0 to 26.0ºC
on the Aw calibration screen after the measurement is taken, and
Err. Cal. temp.
the measurement is suspended.
(Normal 24.0~26.0ºC) Take the measurement again under an environment where the
temperature of the sample chamber is between 24.0 and 26.0ºC.
It is in a state where the measurement cannot be taken because
the sensor board and the cable are not connected, or the cable is
Sensors are not broken. Make sure that the connector of the cable is inserted
securely. If the screen displays the same alarm again, the cable
connected.
may be broken. Repair is needed. See page 47 for contact details
and contact us.
44
WATER ACTIVITY METER EZ-200
How to Attach
the Sensor Protection Filter
Acid gas causes deterioration of the sensor. Use of the sensor protection filter is recommended if
the sensor is deteriorated significantly. The sensor protection filter is sold as an optional
2) Open the top lid of the water activity meter to attach the filter to the sensor as shown in the
pictures below (using tweezers helps in attaching it. The power switch can be ON or OFF.):
Caution
Be careful not to apply strong impact on the sensor.
3) Set the Meas. mode to "Sensor protection filter" (see page 11) to measure the sample. When
the sensor protection filter is attached, the measurement time becomes longer than when the
Refer to “How to dry inside the sample chamber using silica gel”(see page 17) for drying
45
WATER ACTIVITY METER EZ-200
Product Specifications
Product name Water Activity Meter
Model EZ-200
AKA-24016(ATS040T-A240)
Output:DC24V
Dimensions 165(W)×130(H)×270(D)mm
Weight 4.3kg
If any reagent of saturated salt solution for calibration finished up, purchase the
guaranteed Reagent.
46
WATER ACTIVITY METER EZ-200
Optional accessories
Sample container 10 units
Grease 50 g
for calibration)
Contact Details
http://www.freund.co.jp/english/
47