Anesthesia Machine
AX 700

Technical Specification Waveforms: Pressure-time; Flow


rate-time; Capacity-time; ET EtCO2
Physical Characteristics
concentration
Size 689 mm × 800 mm × 1400 mm
Loops: Pressure-volume; Flow-
Weight 125 kg volume; Pressure- flow

Maximum Bearing Top Plate


Weight
210 kg Maximum supporting
capacity
Screen Size: 12.1” TFT touch screen 20kg

Resolution 800 × 600 Operational


dimensions
Handrail Length 750mm 508mm×313mm

Caster wheel 4 wheels 5”, central baking; brakes Dimensions with


optional Additional Accessory
508mm×313mm×380mm
Optional module slot 3
Workbench
number
Maximum
Operation Environment
supporting capacity 20kg
Working Temp 10~40°C

Humidity ≤93%

Power Supply 100-240V~, 50/60Hz±1Hz

Battery Type Rechargeable Lithium-ion battery

Battery Capacity 4400mAh, 11.1VDC

Battery Recharging
Time
Maximum 4 hours for charging

Battery backup trace 2 hours for continuous working


ACGO Available

Oxygen sensor Available

Operational Optional Negative pressure drainage; Isolation


dimensions transformer; AGSS; BIS;
472mm×248mm
MASIMO EtCO2 (sidestream);
Dimensions with
MASIMO EtCO2 (mainstream);
Additional Accessory
472mm×248mm×380mm MASIMO AG (sidestream);
Respironics EtCO2 (mainstream);
Interface:
Respironics EtCO2 (Sidestream);
USB port MASIMO AG+O2 (sidestream);
Artema AG; Artema AG+O2; CPB
RJ45

4 auxiliary power output


Ventilator Specification
AC power interface
Ventilation Modes
Equal-potential grounding terminal
VCV/VC Volume-Controlled Ventilation with
DB9 interface tidal volume compensation
Drawers PCV/VPC Pressure Control Ventilation
Top: Size:462mm×287mm×141mm SIMV-VC, SIMV-PC Synchronized Intermittent
Bearing Weight: 1Kg Mandatory Ventilation

Size: 437mm×287mm×245mm PSV/ CPAP Pressure Support Ventilation


Bottom:
Bearing Weight: 3Kg Others Manual and automatic ventilation

Gas-bag Sway Brace Optional PRVC, SIMV-PRVC, PSVPro

Length: 400mm Ventilation principle Chronometric, volumetric and


barometric
Height: 240mm
Ventilation Electronically controlled&
Features
pneumatically driven
Anesthesia process Open, semi closed, closed circuit
Driven gas O2(air: optional)
Patients Adult, pediatric
Breathing circuit
Mode Manual, Mechanical, Standby volume
1000 ml + bag
Compliance Compliance Correction
Ventilator Setting ranges
Configuration Possibility of configuration
observation Monitoring parameter Tidal volume, Inspiratory, expiratory
flow, minute volume, frequency,
Heating Available pressure (Pmean, Pplat, Ppeak,
PEEP), Oxygen, CO2, N2O and
Bypass Available
halogenated expiratory
concentration, Pressure, oxygen, Positive End Expiratory Pressure (PEEP)
CO2, N2O and Halogen numerical
Type Integrated, electronically controlled
values, compliance and patient
resistance Range 0~70 cmH2O

Tidal volume range 15 ~1500 mL Ventilator Monitoring Ranges

MV (Per-minute TV (Inspiratory tidal


ventilation amount) volume)
0~100 L/min 0~3000 mL

Pressure range TV (expiratory tidal


volume)
(limit) 10~100 cmH2O 0~3000 mL

PPS (Pressure MV (Per-minute


ventilation amount)
support level) 3~60 cmH2O 0~100 L/min

f (breathing freq) 4~100 bpm FiO2

(I: E) Inspiratory (Oxygen concentration) 18~100%

/Expiratory ratio 4:1~1:10 Airway pressure -20~120cmH2O

Apnea I: E 4:1~1:8 PEEP 0~70cmH2O

Apnea time 10~30s Ppeak

Apnea pressure 3~60 cmH2O (Airway pressure) -20~120 cmH2O

Freq. Min. (Min. Pmean


frequency for apnea-
2-60 bpm (Mean pressure) -20~120 cmH2O
ventilation)
Pplat
Tip/Te (Inspiratory
pause) (Platform pressure) 0~120 cmH2O
OFF, 5~60% of inspiratory time
I: E (Inspiratory-
Inspiratory time 0.2~5s
expiratory ratio)
4:1~1:12
Inspiratory pressure 5~70 cmH2O
Freq (Respiratory rate) 0~120 bpm
PEEP OFF, 3~30 cmH2O
Compl (Compliance) 0~300 mL/cmH2O
Trigger pressure -20~-1 cmH2O
Resistance 0~600 cmH2O/(s/L)
Trigger window 5~90%
EtCO2
Trigger flow 0.2~15 L/ min
MASIMO EtCO2 0~190mmHg, 0~25% (at 760mmHg)
Flush oxygen 25~75 L/ min
(sidestream);
Accuracy: ± (0.3%+4% of reading).
Inspiratory stop level 5~80%
MASIMO EtCO2 0~190mmHg, 0~25% (at 760mmHg)
Pressure slope 0~2.0s
(mainstream)
Accuracy: ± (0.3%+4% of reading). Artema AG SEV: 0~8%
DES: 0~18%
Respironics EtCO2 0~150mmHg, 0~19.7% (at
HAL/ ISO/ ENF: 0~5%
(sidestream); 760mmHg)
N2O: 0~100%
Accuracy: 0~5.3%: ±0.3%;
O2: 0~100%
5.4~9.2%: ±5% of reading;
CO2: 0~10%
9.3~13.2%: ±8% of reading;
Anesthesia depth
13.3~19.7%: ±10% of reading;
BIS 0.0~100.0
Respironics EtCO2 0~150mmHg, 0~19.7% (at
(mainstream) 760mmHg) SQI 0.0~100.0%

Accuracy: 0~5.3%: ±0.3%; EMG 0~100dB

5.4~9.2%: ±5% of reading; ESR 0.0~100.0%

9.3~13.2%: ±8% of reading; Ventilator Performance

13.3~19.7%: ±10% of reading; Pressure range at inlet 0.28~0.6 MPa

AG Peak gas flow >100 L/min

MASIMO AG SEV: 0~25% Flow valve range 1~100 L/min


DES: 0~25%
Flow compensation
HAL/ ISO/ ENF: 0~25%
rage
200 mL/min to 18 L/min
N2O: 0~100%
Inspiratory flow Maximum inspiratory flow shall not
O2: 0~100%
be smaller than 100L/min when gas
CO2: 0~25% (0~190mmHg) supply pressure is 280KPa.

Accuracy: Range of flow valve 3~100 L/min

SEV: 0~1%: ± 0.15%; 1~5%: ±0.2%; Pressure limitation Controlled by the electronic relief
5~8%: ±0.4%; valve fitted inside the ventilator;
Controlling means for
DES: 0~1%: ± 0.15%; 1~5%: ±0.2%; ventilator Controlled by the mechanical relief
5~10%: ±0.4%;10~15%: ± 0.6%; valve fitted inside the ventilator.
15~18%: ±1%;

ISO, ENF, HAL: 0~1%: ±0.15%;


Ventilator Accuracy
1~5vol %: ±0.2%;
N2O: ± (2% + 2% of the reading) Control accuracy

O2: 0~25%: ±1%; 25~80%: ±2%; TV 15~60 ml: ±10ml;


80~100%: ±3%;
60~210 ml: ±15ml;
CO2: 0~15%: ± (0.2% + 2% of the
210~1500 ml: ±7% of set value.
reading); 15~25%: unspecified
PCV Inspiratory pressure: ±2.5cmH2O or Backup flow control Pure Oxygen flow rate is 0~10 L/min:
±7% of set value, whichever the <±3%; Others: undefined.
greater.
Auxiliary flow control 10~100% of the full scale: ±10% of
Limiting pressure: ±2.5cmH2O or the reading value. Other ranges:
±7% of set value, whichever the undefined.
greater.
Monitoring accuracy
PEEP: OFF: undefined; 3~30cmH2O:
TV 0~60ml: ±10 ml; 60ml ~ 3000ml: ±
±2.0cmH2O, or ±8% of set value,
20ml or ± 7% of reading value,
whichever is the greater. (expiratory)
whichever is greater; Others:
Supporting pressure: ±2.5cmH2O or undefined.
±7% of set value, whichever the
TV 60ml ~ 3000ml: ± 20ml or ± 7% of
greater.
reading value, whichever is greater;
(Inspiratory)
Apnea pressure: ±2.5cmH2O or ±7% Others: undefined.
of set value, whichever the greater.
Paw -20 cmH2O~120 cmH2O: ±2.0
Trigger pressure: ±2.0cmH2O. cmH2O or ± 4% of set value,
whichever is greater; Others:
Freq ±1 bpm or ±5% of set value,
undefined.
whichever is the greater.

I: E 2: 1~1: 4: ±10% of reading value; PEEP 0 cmH2O~70 cmH2O: ±2.0 cmH2O or


± 4% of set value, whichever is
Other ranges: ±25% of reading value. greater; Others: undefined.

Apnea I: E 2: 1~1: 4: ±10% of set value; Pmean -20 cmH2O~120 cmH2O: ±2.0

Other ranges: ±25% of set value. cmH2O or ± 4% of setting value,


whichever is greater;
Tpause In the range of 20%~60%: ±15% of
set value; Others: undefined.

Other ranges: undefined. Pplat 0 cmH2O~120 cmH2O: ±2.0 cmH2O


or ± 4% of set value, whichever is
Inspiratory time ±0.2s
greater; Others: undefined.
Trigger window ±10%
Freq ±1 bpm or ±5% of set value,
Trigger flow rate ±1 L/ min whichever is the greater.

Inspiratory stop level ±10% I: E 2: 1~1: 4: ±10% of reading value;

O2/ N2O/ Air flow 10~100% of the full scale: ±10% of 4: 1~2: 1 and 1: 4~1: 12: ±25% of
control the reading value. Other ranges: setting value; Others: undefined.
undefined.
MV 0 L/min~30 L/min: ±1 L/min or ±15%
Total flow control Air balance gas: ≤±3% of set value, whichever is greater;
Others: undefined.
N2O balance gas: ≤±3%
Compliance 0 ml/cmH2O~250 ml/cmH2O: ±0.5
ml/cmH2O or ± 15% of reading 1. Airway pressure is continuously
value, whichever is greater; lower than (PEEP +3) cmH2O for
more than 30 seconds.
Other ranges: undefined.
2. Expiratory tidal volume is
Resistance 0 cmH2O/(L/s) ~20 cmH2O/(L/s): ±10
continuously lower than 10ml for
cmH2O/(L/s);
more than 30 seconds.
20 cmH2O/(L/s) ~500 cmH2O/(L/s):
Increase the set values of tidal
±50% of reading value;
volume and respiratory frequency, or
Other ranges: undefined. set it to Manual/spontaneous mode.

Oxygen sensor ±3% Alarm Audible and visual alarm;

O2/ N2O/ Air flow 10~100% of the full scale: ±10% of Alarm access Easy access by shortcut
control the reading value. Other ranges:
Flow meters
undefined.
Type Electronic flow meter
Total flow control Air balance gas: ≤±3%
N2O range 0 ~10 L/min
N2O balance gas: ≤±3%
Air range 0 ~10L/min
Backup flow control Pure Oxygen flow rate is 0~10 L/min:
≤±3%; Others: undefined. O2 range 0.2 ~10 L/min

Auxiliary flow control 10~100% of the full scale: ±10% of Total flow control Air balance gas: 21~100%
the reading value. Other ranges:
N2O balance gas:
undefined.
25~100%
Alarm Settings
Total flow range 0.2 ~15 L/min
Tidal volume
Gas Supply
(expiratory) High: 5~1600 ml
Pipeline gasses O2, N2O, Air
Low: 0 ~1595 ml
backup
MV High: 2~100L/min
gas-cylinder gasses O2, N2O, Air
Low: 0 ~98L/min
Pipeline gas
Inspired oxygen High: 20~105% connection
NIST
Low: 18 ~ 103%
backup cylinder
Ppeak High: 2 ~100cmH2O
connection YOKE-CGA
Low: 0 ~98cmH2O
Pressure range at inlet 280~600 kPa
Apnea alarm Two (2) triggering conditions are
Filter 60-80um
satisfied simultaneously:
Features Switch easily to the other gas
without interrupting the ventilation
Auxiliary gas supply O2 Display Numeric and curve displayed in
screen

Alarm delay 1~10s (step size: 1s)


Breathing Circuit Specification
Sweep 6.25 mm/s,12.5 mm/s
System Pressure Gauge
Anesthetic Agent (AG) Module
Range -20~100 cmH2O
Maximum sound
Accuracy ± (4% of full scales reading + 4% of
pressure for low alarm
actual reading)

Adjustable Pressure Limiting (APL)valve 79dB

Range 1~75 cmH2O Measurement type Side stream

Tactile knob indication Module type Phasin ISA AG module


at
>30 cmH2O Accuracy ±10ml/min or ±10%, whichever is
greater
Accuracy: ±1.0 cmH2O
Monitored parameters CO2, N2O, AA, MAC, Paramagnetic
Minimum opening
O2 and BIS
pressure 0.3 cmH2O (dry), 0.5 cmH2O (humid)
Active AGSS
Breathing Circuit Parameters
Feature High flow, low vacuum
Compliance ≤4mL/100Pa
Size 535mm×120mm×155mm
Automatically compensates for
Weight 2.2kg
compression loss within the
breathing circuit in mechanical mode Applies ISO 80601-2-13 and YY 0635-2

Breathing Frequency 4 – 100 bpm Pressure relief device

Volume of CO2 canister 2000ml Atmospheric pressure compensation


port

Connector ISO9170-2 or BS6834 standard


Feature Heated at 134 degree, removable,
connector
easy to dismantle and sterilize
Flow of suction 50-80L/min

Resistance 0.75KPa ,75L/min


Gas Monitoring
Filter Stainless steel mesh, with pore size
Carbon Dioxide (CO2) Modules
of 60~100μm
Type Mainstream ETCO2, Sidestream
ACGO
ETCO2
Connector Taper coaxial fitting of 22mm
Method Infrared absorption
(outside) and 15 (inside)
Back pressure generated at the rear end of Input frequency 50/60 Hz
anesthesia vaporizer and the front-end of ACGO
Leakage current < 500μA
during quick oxygen charging ≦2kPa
Auxiliary output supply
Flush O2
Output voltage 100~240 V~/ 100~120V~
100% fast oxygen
Output frequency 50/60 Hz
Vaporizer

Brand Drager and Penlon available


Power (With isolation transformer)
Locking Two vaporizers with interlocking
External AC power supply
system
Input voltage 100~120 V~/ 220~240V~
Automatic recognition Anesthesia machine able to
automatic recognize halogenated Input current 3.5 A/8.5 A
gases
Input frequency 50/60 Hz

Leakage current < 500μA


Power (No isolation transformer)
Auxiliary output supply
External AC power supply
Output voltage 100~120 V~/ 220~240V~
Input voltage 100~240 V~/ 100~120V~
Output frequency 50/60 Hz
Input current 3.5~8.5 A/8.5 A

*Notice: Specifications subject to changes without prior

notice. All rights reserved by Comen