Anesthesia Machine
AX 900

Technical Specification
Physical Characteristics
Size 689 mm × 800 mm × 1400 mm
Weight 128kg
Maximum Bearing
Weight 210kg Operational
Screen Size: 15” + 8” TFT touch screen dimensions 472mm×248mm
Resolution 1024 × 768 Dimensions with
Handrail Length 750mm Additional Accessory 472mm×248mm×380mm
Caster wheel 4 wheels, central baking; brakes Interface:
optional; size 5” 2 USB interfaces
Operation Environment RJ45 network interface
Working Temp 10~40°C 4 auxiliary power output
Humidity ≤93% AC power interface
Power Supply 100-240V~, 50/60Hz±1Hz Equal-potential grounding terminal
Battery Type Rechargeable Lithium-ion battery DB9 interface
Battery Capacity 7000 mAh, 11.1VDC VGA interface
Battery Recharging Drawers
Time 6 hours for charging; Top: Size:462mm×287mm×141mm
Battery backup 3 hours for continuous working Bearing Weight: 1Kg
Waveforms Pressure-time; Flow rate-time; Bottom: Size: 437mm×287mm×245mm
Spirometry Capacity-time; EtCO2 concentration Bearing Weight: 3Kg
Pressure-volume; Flow-volume; Gas-bag Sway Brace
Pressure- flow Length: 425mm
Top Plate Height: 240mm
Maximum supporting Features
capacity 20kg Anesthesia process Open, semi closed, closed circuit
Operational Patients Adult, pediatric
dimensions 508mm×313mm Working Mode Manual, Mechanical, Standby
Dimensions with Compliance Compliance Correction
Additional Accessory 508mm×313mm×380mm Configuration Possibility of configuration
Workbench observation
Maximum Heating Available
supporting capacity 20kg Bypass Available
Oxygen sensor Available
Optional Negative pressure drainage; Isolation Tpause (Inspiratory
transformer; AGSS; BIS; pause) OFF, 5~60% of inspiratory time
MASIMO EtCO2 (sidestream); Tinsp (Inspiratory 0.2~5s
MASIMO EtCO2 (mainstream); time)
MASIMO AG (sidestream); Respironics Pinsp (Inspiratory
EtCO2 (mainstream); Respironics pressure) 5~70 cmH2O
EtCO2 (Sidestream); MASIMO AG+O2 PEEP OFF, 3~30 cmH2O
(sidestream); Artema AG; Artema Trigger pressure -20~-1 cmH2O
AG+O2; Optimal flow indication; Trigger window 5~90%
Anesthetic usage monitoring; CPB Trigger flow rate 0.2~15 L/ min
Flush oxygen (Rapid
Ventilator Specification Oxygenation) 25~75 L/ min
Ventilation Modes Inspiratory stop level 5~80%
VCV/VC Volume-Controlled Ventilation with Tslope (Pressure
tidal volume compensation slope) 0~2.0s
PCV/VPC Pressure Control Ventilation Ventilator Monitoring Ranges
SIMV-VC, SIMV-PC Synchronized Intermittent TV (Inspiratory tidal
Mandatory Ventilation volume) 0~3000 mL
PRVC Pressure Regulated Volume Control TV (expiratory tidal
PSV/ CPAP Pressure Support Ventilation volume) 0~3000 mL
Others Manual and automatic ventilation MV (Per-minute
Optional SIMV-PRVC, PSV Pro ventilation amount) 0~100 L/min
Ventilation principle Chronometric, volumetric and FiO2 (Oxygen
barometric concentration) 18~100%
Ventilation Electronically controlled& Paw (Airway
pneumatically driven pressure) -20~120 cmH2O
Driven gas Electronic selective air or O2 PEEP 0~70 cmH2O
Breathing circuit Ppeak
volume 1000 ml + bag (Airway pressure) -20~120 cmH2O
Ventilator Setting ranges Pmean
Tidal volume range 15 ~1500 mL (Mean pressure) -20~120 cmH2O
MV (Per-minute Pplat
ventilation amount) 0~100 L/min (Platform pressure) 0~120 cmH2O
Plimit (pressure) 10~100 cmH2O I.E. (Inspiratory-
RR expiratory ratio) 4:1~1:12
3~60 cmH2O RR 0~120 bpm
f (Respiratory Rate) 4~100 bpm Compliance 0~300 mL/cmH2O
I.E. (Inspiratory Resistance 0~600 cmH2O/(s/L)
Expiratory ratio) 4:1~1:10 NMT
Apnea I.E. 4:1~1:8 stimulus output Pulse Width: 200us;
Apnea time 10~30s accuracy: ±10%;
Apnea pressure 3~60 cmH2O Stimulate current range: 0mA~80mA;
Freq. Min. (Min. Step size 5mA;
frequency for apnea- 2-60 bpm Accuracy: ±5% or ±2mA( whichever is
ventilation) greater)
Maximum skin impedance: 5KΩ Accuracy:
ST mode ST-Ratio: 0%~200%; SEV: 0~1%: ± 0.15%; 1~5%: ±0.2%;
Measurement interval: manual, 1s, 5~8%: ±0.4%;
10s, 20s, 30s DES: 0~1%: ± 0.15%; 1~5%: ±0.2%;
TOF mode TOF-Count:0~4 5~10%: ±0.4%;10~15%: ± 0.6%;
TOF-Ratio: 0%~200% 15~18%: ±1%;
TOF-T1%:0%~150% ISO, ENF, HAL: 0~1%: ±0.15%;
Measurement interval: Manual, 12s, 1~5vol %: ±0.2%;
15s, 20s, 30s, 1min, 2min, 5min, N2O: ± (2% + 2% of the reading)
10min, 15min, 30min, 60min O2: 0~25%: ±1%; 25~80%: ±2%;
PTC mode PTC-Count: 0~20; 80~100%: ±3%;
Measurement interval: Manual; CO2: 0~15%: ± (0.2% + 2% of the
DBS mode reading); 15~25%: unspecified
DBS-Count: 0~2; Artema AG SEV: 0~8%
DBS-Ratio: 0~200%; DES: 0~18%
Measurement interval: Manual, 15s, HAL/ ISO/ ENF: 0~5%
20s, 30s, 1min, 2min, 5min, 10min, N2O: 0~100%
15min, 30min, 60min O2: 0~100%
TET mode Physiological rigidity mode, provides CO2: 0~10%
50Hz and 0.2ms pulse stimulation Anesthesia depth
Block recovery Off、1、2、3、4、5%、10%、 BIS 0.0~100.0
20%、30%、40%、50%、60%、 SQI 0.0~100.0%
70%、80%、90%、100% EMG 0~100dB
EtCO2 ESR 0.0~100.0%
MASIMO EtCO2 0~190mmHg, 0~25% (at 760mmHg) Ventilator Performance
(sidestream); Accuracy: ± (0.3%+4% of reading). Inlet Pressure range 0.28~0.6 MPa
MASIMO EtCO2 0~190mmHg, 0~25% (at 760mmHg) Peak gas flow >100 L/min
(mainstream) Accuracy: ± (0.3%+4% of reading). Flow range 1~100 L/min
Respironics EtCO2 0~150mmHg, 0~19.7% Flow compensation
(sidestream); (at 760mmHg) rage 200 mL/min to 18 L/min
Accuracy: 0~5.3%: ±0.3%; Inspiratory flow Maximum inspiratory flow shall not
5.4~9.2%: ±5% of reading; be smaller than 100L/min when gas
9.3~13.2%: ±8% of reading; supply pressure is 280KPa.
13.3~19.7%: ±10% of reading;
Respironics EtCO2 0~150mmHg, 0~19.7% (at 760mmHg) Pressure limitation Controlled by the electronic relief
(mainstream) Accuracy: 0~5.3%: ±0.3%; Controlling means for valve fitted inside the ventilator;
5.4~9.2%: ±5% of reading; ventilator Controlled by the mechanical relief
9.3~13.2%: ±8% of reading; valve fitted inside the ventilator.
13.3~19.7%: ±10% of reading;
AG Ventilator accuracy
MASIMO AG SEV: 0~25% Control accuracy
DES: 0~25% TV 15~60 ml: ±10ml;
HAL/ ISO/ ENF: 0~25% 60~210 ml: ±15ml;
N2O: 0~100% 210~1500 ml: ±7% of set value.
O2: 0~100% PCV Inspiratory pressure: ±2.5cmH2O or
CO2: 0~25% (0~190mmHg)
±7% of set value, whichever is greater; Others: undefined.
greater. PEEP 0 cmH2O~70 cmH2O: ±2.0 cmH2O or
Limiting pressure: ±2.5cmH2O or ±7% ± 4% of set value, whichever is
of set value, whichever is greater. greater; Others: undefined.
PEEP: OFF: undefined; 3~30cmH2O: Pmean -20 cmH2O~120 cmH2O: ±2.0 cmH2O
±2.0cmH2O, or ±8% of set value, or ± 4% of setting value, whichever is
whichever is the greater. greater; Others: undefined.
Supporting pressure: ±2.5cmH2O or Pplat 0 cmH2O~120 cmH2O: ±2.0 cmH2O or
±7% of set value, whichever is greater. ± 4% of set value, whichever is
Apnea pressure: ±2.5cmH2O or ±7% greater; Others: undefined.
of set value, whichever is greater. RR ±1 bpm or ±5% of set value,
Trigger pressure: ±2.0cmH2O. whichever is greater.
RR ±1 bpm or ±5% of set value, I.E. 2: 1~1: 4: ±10% of reading value;
whichever is the greater. 4: 1~2: 1 and 1: 4~1: 12: ±25% of
I.E. 2: 1~1: 4: ±10% of reading value; setting value; Others: undefined.
Other ranges: ±25% of reading value. MV 0 L/min~30 L/min: ±1 L/min or ±15%
Apnea I.E. 2: 1~1: 4: ±10% of set value; of set value, whichever is greater; >30
Other ranges: ±25% of set value. L/min: undefined.
Tpause In the range of 20%~60%: ±15% of set Compliance 0 ml/cmH2O~250 ml/cmH2O: ±0.5
value; ml/cmH2O or ± 15% of reading value,
Other ranges: undefined. whichever is greater.
Inspiratory time ±0.2s Other ranges: undefined.
Inspiratory pause ±15% of set value; Resistance 0 cmH2O/(L/s) ~20 cmH2O/(L/s): ±
Trigger window ±10% 10 cmH2O/(L/s);20
Trigger flow rate ±1 L/ min cmH2O/(L/s)~500 cmH2O/(L/s):±
Inspiratory stop level ±10% 50% of reading value; Other ranges:
O2/ N2O/ Air flow 10~100% of the full scale: ±10% of the undefined.
control reading value. Oxygen sensor ±3%
Total flow control Air balance gas: ≤±3% O2/ N2O/ Air flow 10~100% of the full scale: ±10% of the
N2O balance gas: ≤±3% control reading value. Other ranges:
Backup flow control Pure Oxygen flow rate is 0~10 L/min: undefined.
≤±10%; Others: undefined. Total flow control Air balance gas: ≤±3%
Auxiliary flow control 10~100% of the full scale: ±10% of the N2O balance gas: ≤±3%
reading value. Other ranges: Backup flow control Pure Oxygen flow rate is 0~10 L/min:
undefined. ≤±3%; Others: undefined.
Monitoring accuracy Auxiliary flow control 10~100% of the full scale: ±10% of the
TV 0~60ml: ±10 ml; 60ml ~ 3000ml: ± reading value. Other ranges:
(expiratory) 20ml or ± 7% of reading value, undefined.
whichever is greater; Others: Alarm Settings
undefined. Tidal volume High: 5~1600 ml
TV 60ml ~ 3000ml: ± 20ml or ± 7% of Low: 0 ~1595 ml
(Inspiratory) reading value, whichever is greater; MV High: 2~100L/min
Others: undefined. Low: 0 ~98L/min
Paw -20 cmH2O~120 cmH2O: ±2.0 cmH2O FiO2 High: 20~105%
or ± 4% of set value, whichever is Low: 18 ~ 103%
Ppeak High: 2 ~100cmH2O Accuracy ± (4% of full scales reading + 4% of
Low: 0 ~98cmH2O actual reading)
Apnea alarm Two (2) triggering conditions are Adjustable Pressure Limiting (APL)valve
satisfied simultaneously: Range 1~75 cmH2O
1. Airway pressure is continuously Touch indication >30 cmH2O
lower than (PEEP +3) cmH2O for more Accuracy: ±1.0 cmH2O
than 30 seconds. Minimum opening
2. Expiratory tidal volume is pressure 0.3 cmH2O (dry), 0.5 cmH2O (humid)
continuously lower than 10ml for Breathing Circuit Parameters
more than 30 seconds. Compliance ≤4mL/100Pa
Increase the set values of tidal volume Automatically compensates for
and respiratory frequency, or set it to compression loss within the breathing
Manual/spontaneous mode. circuit in mechanical mode
Alarm Audible and visual alarm; Volume of CO2 2000ml
Alarm access Easy access by shortcut canister
Flow meters
Type Electronic flow meter Feature Heated at 134 degree, removable,
N2O range 0 ~15 L/min easy to dismantle and sterilize
Air range 0 ~15 L/min
O2 range 0.2 ~15 L/min Gas Monitoring
Total flow control Air balance gas: 21~100% Carbon Dioxide (CO2) Modules
N2O balance gas: 25~100% Type Mainstream ETCO2, Sidestream
Total flow range 0.2 ~18 L/min ETCO2
Backup flow control 0 ~15 L/min Method Infrared absorption
Proportional device Equipped with a safety system to Display Numeric and curve displayed in
ensure an O2 concentration of at least screen
25% Alarm delay 1~10s (step size: 1s)
Gas Supply Sweep 6.25 mm/s,12.5 mm/s
Pipeline gasses O2, N2O, Air Anesthetic Agent (AG) Module
Backup Maximum sound
gas-cylinder gasses O2, N2O, Air pressure for low
Pipeline gas alarm 79dB
connection NIST Measurement type Side stream
Backup cylinder Module type Phasin ISA AG module
connection YOKE-CGA Accuracy ±10ml/min or ±10%, whichever is
Inlet pressure range 280~600 kPa greater
Filter 60-80µm Monitored CO2, N2O, AA, MAC, Paramagnetic O2
Features Switch easily to the other gas without parameters and BIS
interrupting the ventilation Active AGSS
Auxiliary gas supply O2& O2, Air mixture gas Feature High flow, low vacuum
Size 535 mm×120 mm×155 mm
Breathing System Specification Weight 2.2 kg
System Pressure Gauge Applies ISO 80601-2-13 and YY 0635-2
Range -20~100 cmH2O Pressure relief device Atmospheric pressure compensation
port
Connector ISO9170-2 or BS6834 standard Automatic Anesthesia machine able to automatic
connector recognition recognize halogenated gases
Flow of suction 50-80L/min
Resistance 0.75KPa ,75L/min Power (No isolation transformer)
Filter Stainless steel mesh, with pore size of External AC power supply
60~100μm Input voltage 100~240 V~/ 100~120V~
ACGO Input current 3.5~8.5 A/8.5 A
Connector Taper coaxial fitting of 22mm Input frequency 50/60 Hz
(outside) and 15 (inside) Leakage current < 500μA
Back pressure Auxiliary output supply
generated at the Output voltage 100~240 V~/ 100~120V~
rear end of Output frequency 50/60 Hz
anesthesia
vaporizer and the Power (With isolation transformer)
front-end of ACGO External AC power supply
≦2kPa
during rapid Input voltage 100~120 V~/ 220~240V~
oxygen variation Input current 3.5 A/8.5 A
Input frequency 50/60 Hz
Flush O2 Leakage current < 500μA
100% fast oxygen Auxiliary output supply
Vaporizer Output voltage 100~120 V~/ 220~240V~
Brand Drager and Penlon available Output frequency 50/60 Hz
Locking Two vaporizers with interlocking
system