Specification: AX 900

SHENZHEN COMEN MEDICAL INSTRUMENTS CO.,LTD


No.2 of FIYATA Timepiece Building, Nanhuan Avenue, Gongming
sub-district, Guangming New District, Shenzhen, P.R. China
Tel: +86-755-26408879
Fax:+86-755-26431232
Email: info@szcomen.com
Web: www.comen.com

2020 V 1.2
Anesthesia Machine
AX 900

Technical Specification
Physical Characteristics
Size 689 mm × 800 mm × 1400 mm
Weight 128kg
Maximum Bearing Operational
Weight 210kg dimensions 472mm×248mm
Screen Size: 15” + 8” TFT touch screen Dimensions with
Resolution 800 × 600 Additional Accessory 472mm×248mm×380mm
Handrail Length 750mm Interface:
Caster wheel 4 wheels 5”, central baking; brakes 2 USB interfaces
optional RJ45 network interface
Operation Environment 4 auxiliary power output
Working Temp 10~40°C AC power interface
Humidity ≤93% Equal-potential grounding terminal
Power Supply 100-240V~, 50/60Hz±1Hz DB9 interface
Battery Type Rechargeable Lithium-ion battery VGA interface
Battery Capacity 7000 mAh, 11.1VDC Drawers
Battery Recharging Top: Size:462mm×287mm×141mm
Time 6 hours for charging; Bearing Weight: 1Kg
Battery backup Trace 3 hours for continuous working Bottom: Size: 437mm×287mm×245mm Bearing
Waveforms: Pressure-time; Flow rate- Weight: 3Kg
time; Capacity-time; ET Gas-bag Sway Brace
EtCO2 concentration; EEG Length: 425mm
Loops: Pressure-volume; Flow-volume; Height: 240mm
Pressure- flow Features
Top Plate Anesthesia process Open, semi closed, closed circuit
Maximum supporting Patients Adult, pediatric
capacity 20kg Working Mode Manual, Mechanical, Standby
Operational Compliance Compliance Correction
dimensions 508mm×313mm Configuration Possibility of configuration observation
Dimensions with Heating Available
Additional Accessory 508mm×313mm×380mm Bypass Available
Workbench Oxygen sensor Available
Maximum Optional Negative pressure drainage; Isolation
supporting capacity 20kg transformer; AGSS; BIS;
MASIMO EtCO2 (sidestream); MASIMO Apnea pressure 3~60 cmH2O
EtCO2 (mainstream); MASIMO AG Freq. Min. (Min.
(sidestream); Respironics EtCO2 frequency for apnea- 2-60 bpm
(mainstream); Respironics EtCO2 ventilation)
(Sidestream); MASIMO AG+O2 Tip/Te (Inspiratory
(sidestream); Artema AG; Artema pause) OFF, 5~16% of inspiratory time
AG+O2; Optimal flow indication; Tinsp (Inspiratory 0.2~5s
Anesthetic usage monitoring; CPB time)
Pinsp (Inspiratory
Ventilator Specification pressure) 5~70 cmH2O
Ventilation Modes PEEP OFF, 3~30 cmH2O
VCV/VC Volume-Controlled Ventilation with Trigger pressure -20~-1 cmH2O
tidal volume compensation Trigger window 5~90%
PCV/VPC Pressure Control Ventilation Trigger flow rate 0.2~15 L/ min
SIMV-VC, SIMV-PC Synchronized Intermittent Flush oxygen (Rapid
Mandatory Ventilation Oxygenation) 25~75 L/ min
PRVC Pressure Regulated Volume Control Inspiratory stop level 5~80%
PSV/ CPAP Pressure Support Ventilation Tslope (Pressure
Others Manual and automatic ventilation slope) 0~2.0s
Optional SIMV-PRVC, PSV Pro Ventilator Monitoring Ranges
Ventilation principle Chronometric, volumetric and TV (Inspiratory tidal
barometric volume) 0~3000 mL
Ventilation Electronically controlled& TV (expiratory tidal
pneumatically driven volume) 0~3000 mL
Driven gas Electronic selective air or O2 MV (Per-minute
Breathing circuit ventilation amount) 0~100 L/min
volume 1000 ml + bag FiO2 (Oxygen
Ventilator Setting ranges concentration) 18~100%
Monitoring Tidal volume, Inspiratory, expiratory Paw (Airway
parameter flow, minute volume, frequency, pressure) -20~120 cmH2O
pressure (Pmean, Pplat, Ppeak, PEEP), PEEP 0~70 cmH2O
Oxygen, CO2, N2O and halogenated Ppeak
expiratory concentration, Pressure, (Airway pressure) -20~120 cmH2O
oxygen, CO2, N2O and Halogen Pmean
numerical values, compliance and (Mean pressure) -20~120 cmH2O
patient resistance Pplat
Tidal volume range 15 ~1500 mL (Platform pressure) 0~120 cmH2O
MV (Per-minute I.E. (Inspiratory-
ventilation amount) 0~100 L/min expiratory ratio) 4:1~1:12
Plimit (pressure) 10~100 cmH2O Rate
PPS (pressure (Respiratory rate) 0~120 bpm
support level) 3~60 cmH2O Compliance 0~300 mL/cmH2O
f (Respiratory Rate) 4~100 bpm Resistance 0~600 cmH2O/(s/L)
I.E. (Inspiratory EtCO2
Expiratory ratio) 4:1~1:10 MASIMO EtCO2 0~190mmHg, 0~25% (at 760mmHg)
Apnea I.E. 4:1~1:8 (sidestream); Accuracy: ± (0.3%+4% of reading).
Apnea time 10~30s
MASIMO EtCO2 0~190mmHg, 0~25% (at 760mmHg) Flow valve range 1~100 L/min
(mainstream) Accuracy: ± (0.3%+4% of reading). Flow compensation
Respironics EtCO2 0~150mmHg, 0~19.7% rage 200 mL/min to 15 L/min
(sidestream); (at 760mmHg) Inspiratory flow Maximum inspiratory flow shall not be
Accuracy: 0~5.3%: ±0.3%; smaller than 100L/min when gas
5.4~9.2%: ±5% of reading; supply pressure is 280KPa.
9.3~13.2%: ±8% of reading;
13.3~19.7%: ±10% of reading; Pressure limitation Controlled by the electronic relief
Respironics EtCO2 0~150mmHg, 0~19.7% (at 760mmHg) Controlling means for valve fitted inside the ventilator;
(mainstream) Accuracy: 0~5.3%: ±0.3%; ventilator Controlled by the mechanical relief
5.4~9.2%: ±5% of reading; valve fitted inside the ventilator.
9.3~13.2%: ±8% of reading;
13.3~19.7%: ±10% of reading; Ventilator accuracy
AG Control accuracy
MASIMO AG SEV: 0~25% TV 15~60 ml: ±10ml;
DES: 0~25% 60~210 ml: ±15ml;
HAL/ ISO/ ENF: 0~25% 210~1500 ml: ±7% of set value.
N2O: 0~100% PCV Inspiratory pressure: ±2.5cmH2O or
O2: 0~100% ±7% of set value, whichever is greater.
CO2: 0~25% (0~190mmHg)
Limiting pressure: ±2.5cmH2O or ±7%
Accuracy:
of set value, whichever is greater.
SEV: 0~1%: ± 0.15%; 1~5%: ±0.2%;
PEEP: OFF: undefined; 3~30cmH2O:
5~8%: ±0.4%; ±2.0cmH2O, or ±8% of set value,
DES: 0~1%: ± 0.15%; 1~5%: ±0.2%; whichever is the greater.
5~10%: ±0.4%;10~15%: ± 0.6%;
Supporting pressure: ±2.5cmH2O or
15~18%: ±1%; ±7% of set value, whichever is greater.
ISO, ENF, HAL: 0~1%: ±0.15%;
Apnea pressure: ±2.5cmH2O or ±7% of
1~5vol %: ±0.2%;
set value, whichever is greater.
N2O: ± (2% + 2% of the reading) Trigger pressure: ±2.0cmH2O.
O2: 0~25%: ±1%; 25~80%: ±2%;
Rate ±1 bpm or ±5% of set value, whichever
80~100%: ±3%; is the greater.
CO2: 0~15%: ± (0.2% + 2% of the I.E. 2: 1~1: 4: ±10% of reading value;
reading); 15~25%: unspecified
Other ranges: ±25% of reading value.
Artema AG SEV: 0~8% Apnea I.E. 2: 1~1: 4: ±10% of set value;
DES: 0~18%
Other ranges: ±25% of set value.
HAL/ ISO/ ENF: 0~5% Tip/Te 20%~60%: ±15% of set value;
N2O: 0~100%
Other ranges: undefined.
O2: 0~100%
Inspiratory time ±0.2s
CO2: 0~10%
Inspiratory pause ±15% of set value;
Anesthesia depth
Trigger window ±10%
BIS 0.0~100.0
Trigger flow rate ±1 L/ min
SQI 0.0~100.0%
Inspiratory stop level ±10%
EMG 0~100dB
O2/ N2O/ Air flow 10~100% of the full scale: ±10% of the
ESR 0.0~100.0%
control reading value.
Ventilator Performance
Total flow control Air balance gas: ≤±3%
Inlet Pressure range 0.28~0.6 MPa
N2O balance gas: ≤±3%
Peak gas flow 100 L/min plus fresh gas
Backup flow control Pure Oxygen flow rate is 0~10 L/min: Backup flow control Pure Oxygen flow rate is 0~10 L/min:
≤±10%; Others: undefined. ≤±3%; Others: undefined.
Auxiliary flow control 10~100% of the full scale: ±10% of the Auxiliary flow control 10~100% of the full scale: ±10% of the
reading value. Other ranges: reading value. Other ranges:
undefined. undefined.
Monitoring accuracy Alarm Settings
TV 0~60ml: ±10 ml; 60ml ~ 3000ml: ± Tidal volume High: 5~1600 ml
(expiratory) 20ml or ± 7% of reading value, Low: 0 ~1595 ml
whichever is greater; Others: MV High: 2~100L/ml
undefined. Low: 0 ~98L/ml
TV 60ml ~ 3000ml: ± 20ml or ± 7% of FiO2 High: 20~105%
(Inspiratory) reading value, whichever is greater; Low: 18 ~ 103%
Others: undefined. Ppeak High: 2 ~100cmH2O
Paw -20 cmH2O~120 cmH2O: ±2.0 cmH2O Low: 0 ~98cmH2O
or ± 4% of set value, whichever is Apnea alarm 20s; with error of ±3%
greater; Others: undefined. Alarm Audible and visual alarm;
PEEP 0 cmH2O~70 cmH2O: ±2.0 cmH2O or ± Alarm access Easy access by shortcut
4% of set value, whichever is greater; Flow meters
Others: undefined. Type Electronic flow meter
Pmean -20 cmH2O~120 cmH2O: ±2.0 cmH2O N2O range 0 ~15 L/min
or ± 4% of setting value, whichever is Air range 0 ~15 L/min
greater; Others: undefined. O2 range 0.2 ~15 L/min
Pplat 0 cmH2O~120 cmH2O: ±2.0 cmH2O or Total flow control Air balance gas: 21~100%
± 4% of set value, whichever is greater; N2O balance gas: 25~100%
Others: undefined. Total flow range 0.2 ~18 L/min
Rate ±1 bpm or ±5% of set value, whichever Standby flow control 0 ~15 L/min
is greater. Proportional device Equipped with a safety system to
I.E. 2: 1~1: 4: ±10% of reading value; ensure an O2 concentration of at least
4: 1~2: 1 and 1: 4~1: 12: ±25% of 25%
setting value; Others: undefined. Gas Supply
MV 0 L/min~30 L/min: ±1 L/min or ±15% of Pipeline gasses O2, N2O, Air
set value, whichever is greater; >30 Standby
L/min: undefined. gas-cylinder gasses O2, N2O, Air
Compliance 0 ml/cmH2O~250 ml/cmH2O: ±0.5 Pipeline gas
ml/cmH2O or ± 15% of reading value, connection NIST
whichever is greater; Standby cylinder
Other ranges: undefined. connection PISS
Resistance 0 cmH2O/(L/s) ~20 cmH2O/(L/s): ±10 Inlet pressure range 280~600 kPa
cmH2O/(L/s);20 cmH2O/(L/s)~500 Filter 60-80µm
cmH2O/(L/s):±50% of reading Features Switch easily to the other gas without
value; Other ranges: undefined. interrupting the ventilation
Oxygen sensor ±3% Auxiliary gas supply O2& O2, Air mixture gas
O2/ N2O/ Air flow 10~100% of the full scale: ±10% of the
control reading value. Other ranges: Breathing System Specification
undefined. System Pressure Gauge
Total flow control Air balance gas: ≤±3% Range -20~100 cmH2O
N2O balance gas: ≤±3%
Accuracy ± (4% of full scales reading + 4% of Pressure relief device Atmospheric pressure compensation
actual reading) port
Adjustable Pressure Limiting (APL)valve Connector ISO9170-2 or BS6834 standard
Range 1~75 cmH2O connector
Tactile knob Flow of suction 50-80L/min
indication at >30 cmH2O Resistance 0.75KPa ,75L/min
Accuracy: ±1.0 cmH2O Filter Stainless steel mesh, with pore size of
Minimum opening 60~100μm
pressure 0.3 cmH2O (dry), 0.5 cmH2O (humid) ACGO
Breathing Circuit Parameters Connector Taper coaxial fitting of 22mm (outside)
Compliance ≤4mL/100Pa and 15 (inside)
Automatically compensates for Back pressure generated at the rear end of
compression loss with in the breathing anesthesia vaporizer and the front-end of ACGO
≦2kPa
circuit in mechanical mode during quick oxygen charging
Volume of CO2 The leakage shall not be greater than Flush O2
canister 50 ml/min at 3 kPa 100% fast oxygen
Water Trap 7mL, easy to be disassembled Vaporizer
Feature Heated at 134 degree, removable, easy Brand Drager and Penlon available
to dismantle and sterilize Locking Two vaporizers with interlocking
system
Gas Monitoring Automatic Anesthesia machine able to automatic
Carbon Dioxide (CO2) Modules recognition recognize halogenated gases
Type Mainstream ETCO2, Sidestream ETCO2
Method Infrared absorption Power (No isolation transformer)
Display Numeric and curve displayed in screen External AC power supply
Alarm delay 1~10s (step size: 1s) Input voltage 100~240 V~/ 100~120V~
Sweep 6.25 mm/s,12.5 mm/s Input current 3.5~8.5 A/8.5 A
Anesthetic Agent (AG) Module Input frequency 50/60 Hz
Maximum sound Leakage current < 500μA
pressure for low Auxiliary output supply
alarm 79dB Output voltage 100~240 V~/ 100~120V~
Measurement type Side stream Output frequency 50/60 Hz
Module type Phasin ISA AG module
Accuracy ±10ml/min or ±10%, whichever is Power (With isolation transformer)
greater External AC power supply
Monitored CO2, N2O, AA, MAC, Paramagnetic O2 Input voltage 100~120 V~/ 220~240V~
parameters and BIS Input current 3.5 A/8.5 A
Active AGSS Input frequency 50/60 Hz
Feature High flow, low vacuum Leakage current < 500μA
Size 535 mm×120 mm×155 mm Auxiliary output supply
Weight 2.2 kg Output voltage 100~120 V~/ 220~240V~
Applies ISO 80601-2-13 and YY 0635-2 Output frequency 50/60 Hz

*Notice: Specifications subject to changes without prior


notice. All rights reserved by Comen