Professional Documents
Culture Documents
Qualification
Unit description
Precisa balance BJ 2100D
Location
Unit identification
Name Inventory No.
Type BJ 2100D Weighing Range 2100 g
Serial No. Readability 0.1 g
Mains supply Mains frequency 50-60 Hz
Nom. power supply Administrator
ok not ok removed
The balance corresponds to above data
Components delivered
ok not ok removed
Components delivered correspond to manual
Log book
A log book was opened Yes No
Installation
ok not ok removed
Location corresponding to manual
Assemble the in-use cover for the display
Fix draftshield
Install weighing pan support and weighing pan
Levelling
ok not ok removed
Instrument fitted with a levelling bubble
1/2
Installation
Qualification
Configuration
ok not ok removed
Configuration set according to requirements
Balance calibration
ok not ok removed
Balance calibrated
Final comments
The installation has been performed correctly. The balance is ready for use.
By signing this, the user confirms delivery and installation of the balance in accordance with the sales
agreement.
2/2
SOP: OQ BJ 2100D SOP :
Page :
Effective Date:
Prerequisites
The balance should be switched on for at least 1 hours before the start of the test. The location must
be calm and free from vibration. Clean the balance and adjust it. Acclimatize test weights to
surroundings.
1. Tare
2. Place the test weight (2000 g) on the centre of the weighing pan, print or note the value.
3. Determine the difference between the test weight calibration report and the display.
4. Acceptance criteria: Difference between nominal value and display 0.5 g
5. If the deviation is more than the maximum permissible error, a Precisa service engineer must
perform a span adjustment.
Repeatability
Maximum permissible error: 0.2 g
Test weight: 2000 g
Eccentricity Test
Maximum permissible error: 0.5 g
Test weight: 800 g
1. Tare
2. Place the test weight (800 g) on the centre of the weighing pan, print or note the value.
3. Shift the weight to the next corner as indicated in the diagram, print or note the value.
4. Repeat for the other 3 corners and again for the centre
5. Acceptance criteria: largest value – smallest value 0.5 g
Linearity Test
Maximum permissible error: 0.5 g
Test weights: 500 g, 1000 g, 1500 g, 2000 g
1/1
Operational
Qualification
Unit Identification
Name Inventory No.
Type BJ 2100D Weighing Range 2100 g
Serial No. Readability 0.1 g
Initial Situation
installation repair maintenance
Span Calibration
Test weight: 2000 g maximum permissible error: 0.5 g
Repeatability Test
Test weight: 2000 g maximum permissible standard deviation: 0.2 g
Eccentricity Test
Test weight: 800 g maximum permissible error: 0.5 g
Position Display
1
2 Wmax-Wmin:
3
4
5 within specification
6 not within specification
Linearity Test
Test weights: 500 g, 1000 g, 1500 g, 2000 g maximum permissible error: 0.5 g
within specification
not within specification
1/2
Operational
Qualification
Type
Class
Serial -No.
Weights and Measures approved until
Training Certificate
Name Date Signature Name Instructor Signature
Instructor
Final Comments
2/2
Performance
Qualification
Unit Identification
Name Inventory No.
Type BJ 2100D Weighing Range 2100 g
Serial No. Readability 0.1 g
The balance should be switched on for at least 1 hour before the start of the test. The location must be
calm and free from vibration. Clean the balance and adjust it. Acclimatize test weights to surroundings.
1/1