You are on page 1of 6

Installation

Qualification

Unit description
Precisa balance BJ 2100D

Location

Unit identification
Name Inventory No.
Type BJ 2100D Weighing Range 2100 g
Serial No. Readability 0.1 g
Mains supply Mains frequency 50-60 Hz
Nom. power supply Administrator

ok not ok removed
The balance corresponds to above data

Components delivered

ok not ok removed
Components delivered correspond to manual

Log book
A log book was opened Yes No

Location of log book:

Installation
ok not ok removed
Location corresponding to manual
Assemble the in-use cover for the display
Fix draftshield
Install weighing pan support and weighing pan

Electrical power supply


ok not ok removed
Connection carried out according data on type plate

Levelling
ok not ok removed
Instrument fitted with a levelling bubble

1/2
Installation
Qualification

Configuration
ok not ok removed
Configuration set according to requirements

Changes compared with factory configuration/Print out of user configuration:

Balance calibration
ok not ok removed
Balance calibrated

Result of the IQ-Test protocol


If all test are passed with OK or removed, it is assured that the balance is installed according to the
installation guidelines of the manufacturer

Final comments

The installation has been performed correctly. The balance is ready for use.

Name: ______________________ Function: _________________________

Date: ______________________ Signature: _________________________

By signing this, the user confirms delivery and installation of the balance in accordance with the sales
agreement.

Name: ______________________ Function: _________________________

Date: ______________________ Signature: _________________________

2/2
SOP: OQ BJ 2100D SOP :
Page :
Effective Date:

Standard Operating Procedure for the OQ of Precisa Balance BJ 2100D

Prerequisites
The balance should be switched on for at least 1 hours before the start of the test. The location must
be calm and free from vibration. Clean the balance and adjust it. Acclimatize test weights to
surroundings.

Span Calibration (Sensitivity Adjustment)


Maximum permissible error: 0.5 g
Test weight: 2000 g

1. Tare
2. Place the test weight (2000 g) on the centre of the weighing pan, print or note the value.
3. Determine the difference between the test weight calibration report and the display.
4. Acceptance criteria: Difference between nominal value and display  0.5 g
5. If the deviation is more than the maximum permissible error, a Precisa service engineer must
perform a span adjustment.

Repeatability
Maximum permissible error: 0.2 g
Test weight: 2000 g

1. Tare the balance


2. Place the test weight (2000 g) on the weighing pan, print or note the value.
3. Lift off test weight, tare.
4. Repeat Steps 2 and 3 for the next 9 weighing procedure.
5.Acceptance criteria: standard deviation  0.2 g

Eccentricity Test
Maximum permissible error: 0.5 g
Test weight: 800 g

1. Tare
2. Place the test weight (800 g) on the centre of the weighing pan, print or note the value.
3. Shift the weight to the next corner as indicated in the diagram, print or note the value.
4. Repeat for the other 3 corners and again for the centre
5. Acceptance criteria: largest value – smallest value  0.5 g

Linearity Test
Maximum permissible error: 0.5 g
Test weights: 500 g, 1000 g, 1500 g, 2000 g

1. Tare the balance


2. Load test weight 1, print or note value
3. Repeat for the other 3 weights
4. Acceptance criteria: all Differences between nominal weight and display  0.5 g

1/1
Operational
Qualification

Unit Description Location


Precisa balance

Unit Identification
Name Inventory No.
Type BJ 2100D Weighing Range 2100 g
Serial No. Readability 0.1 g

Initial Situation
 installation  repair  maintenance

description of location  calm  air conditioned


 unstable  weighing bench

Span Calibration
Test weight: 2000 g maximum permissible error: 0.5 g

Nominal Weight Display Difference  within specification


 not within specification

Repeatability Test
Test weight: 2000 g maximum permissible standard deviation: 0.2 g

Measuring Display Measuring Display


1 6
2 7 standard deviation:
3 8
4 9  within specification
5 10  not within specification

Eccentricity Test
Test weight: 800 g maximum permissible error: 0.5 g

Position Display
1
2 Wmax-Wmin:
3
4
5  within specification
6  not within specification

Linearity Test
Test weights: 500 g, 1000 g, 1500 g, 2000 g maximum permissible error: 0.5 g

nominal weight Display Difference

 within specification
 not within specification

1/2
Operational
Qualification

Verified with Weight Set

Type
Class

Serial -No.
Weights and Measures approved until

Training Certificate
Name Date Signature Name Instructor Signature
Instructor

Final Comments

The balance meets all requirements of Design Qualification

Name: ______________________ Function: _________________________

Date: ______________________ Signature: _________________________

2/2
Performance
Qualification

Unit Description Location


Precisa balance

Unit Identification
Name Inventory No.
Type BJ 2100D Weighing Range 2100 g
Serial No. Readability 0.1 g

User’s SOP for Performance Test


Define the reference weights and whether a single point test is appropriate or
whether several points have to tested:
test frequency
application

weight 1 weight class 1


acceptance limit 1 weight certificate 1

weight 2 weight class 2


acceptance limit 2 weight certificate 2

weight 3 weight class 3


acceptance limit 3 weight certificate 3

The balance should be switched on for at least 1 hour before the start of the test. The location must be
calm and free from vibration. Clean the balance and adjust it. Acclimatize test weights to surroundings.

SOP for Performance Test Results


Date Weight 1 Weight 2 Weight 3 Satisfactory Name Signature

1/1