2. Connect the temperature sensor Even if the power is turned off in the HOLD state, the hold
Connect the temperature sensor to the connector (for sensor data (measured value) is held. If the power is turned on
connection) at the top of the mainframe. again, the thermometer starts from the HOLD state.
However, if the batteries are replaced, the hold data and
state are released.
3.2 Power ON/OFF Peak hold & Bottom hold functions
Power ON This function is used to always store, then display the maximum
Press the POWER key. The following displays appear in order. (peak) or minimum (bottom) measured value while following up
measured value changes.
Pressing the POWER key again turns off the power. Hold action
start
When the automatic power-OFF function is used, the
power is automatically turned off if the preset time is over.
(Factory set value: 3 minutes.) Display updating Display updating
2 IM350DP01-E5
[Bottom hold] Details of screen in User setting mode
Measured value (PV) Select each screen by pressing the HOLD key, then select the
Displayed temperature during setting details by pressing the POWER key. The details thus set
Hold action the operation of bottom hold are established just when the next setting screen is selected by
start the HOLD key.
Bottom hold User setting mode
Bottom hold °C/°F selection *
Peak/bottom hold *
°C
Automatically
Measured value display mode
Temperature display .H.ZZ POWER
.H.__
Press and hold the
HOLD key for
23.0 Peak hold
(Factory set value)
Bottom hold
°C
5 seconds or more. HOLD
.U.@C
Release the HOLD key.
The “°C/°F selection” screen appears in .h.oF POWER
.h.on
User setting mode.
Refer to Details of screen in Advanced hold function OFF Advanced hold function ON
User setting mode (Factory set value)
HOLD
It is impossible to enter User setting mode from Measured * The parameter being selected is displayed at each first setting
value display mode. display.
IM350DP01-E5 3
Display Description Solution
6. SPECIFICATIONS
ooo °C
Over-scale
Measured temperature
Check connecting state of
sensor.
Input type: Thermocouple type K (JIS/IEC)
is above the Measuring Confirm that the sensor or
[Flashing] range high limit value. wire is not broken. Influence of external resistance:
Underscale Approx. 0.1 µV/Ω
uuu Measured temperature Measuring range: −200 to +1200 °C
°C is below the Measuring −328 to +1999 °F
[Flashing] range low limit value. −199.9 to +199.9 °C (°F) *
* Display range with the resolution of 0.1°C (°F).
The resolution out of this range is 1 °C (°F).
Self-diagnostic error
Sampling cycle: Approx. 0.3 seconds
If two or more errors occur simultaneously, the total summation
of these error codes is displayed. Action of input break: Upscale (B.O display lights)
Display Description Solution Display: LCD (TN type)
Display accuracy: ± (|0.2 % of indicated value| + 1 digit) or
Adjusted data error Turn OFF then restart the
E 1 °C
Adjusted data range is instrument or remove,
Within ± 2 °C (4 °F)
Less than the value whichever is greater
abnormal. insert the batteries and
[Flashing] restart. However, for less than
−100 °C (−150 °F): Within ± 4 °C (8 °F)
Data back-up error When the operation returns
E 2 °C
Error in reading/ to normal upon restoring
the instrument, noise may
Resolution: 1 °C (°F) or 0.1 °C (°F)
writing of EEPROM Power supply: AA Size (LR6 or R6) dry batteries, 1.5 V,
[Flashing] cause an error. 2 pcs.
Check the noise level near Battery life: Approx. 1000 hours
A/D conversion error
E 4 °C
Response signal from
the instrument.
Check for the external noise
(Total of operating hours)
[When using alkaline dry batteries in the ambient
A/D converter is
[Flashing] source. If an error occurs temperature near 23 °C]
abnormal.
after the power is turned on However, the battery service life varies
Temperature again, the DP-350C must depending on the operating condition and
compensation error be repaired or replaced. environment. Therefore, the above service
Ambient temperature Please contact RKC sales life period is only for reference.
exceeds 90 °C. office or the agent. Power consumption: 5 mW (Mean value)
Allowable ambient temperature:
Problem and solution 0 to 50 °C (32 to 122 °F)
Allowable ambient humidity:
Problem Probable cause Solution
45 to 85 % RH
No power is No battery is loaded. Load a battery. Ambient operating: Free from no corrosive gases, atmosphere
turned on. no large amount of dust and a splash of
(No display The battery with less Change both batteries at
service life is used. the same time.
water.
appears)
Weight: Approx. 140 g (Including dry batteries)
Measured The battery with less Change both batteries at
Dimensions: 145 × 52 × 25 mm (H ×W × D)
temperature service life is used. the same time.
display is Noise source is Separate the noise source
abnormal present near the from the instrument.
instrument.
Proper sensor is not Use the specified sensor. 7. BINDING THE HAND STRAP
used.
Sensor insertion Check whether sensor is Binding the hand strap to the DP-350 as follows.
depth is insufficient. inserted loosely. If yes, fully 2
1
insert the sensor
Sensor insertion Insert the sensor at the
position is not specified location.
appropriate.
The display CPU overrun Keep the thermometer
does not away from the noise
change generation source, then
No key take the following
operation is measures.
accepted Remove the batteries.
When requiring the strap for replacement, please contact
Load new batteries after
RKC sales office or the agent.
a lapse of more than 10
seconds.
Press the POWER key to
turn on the power.
If the thermometer does not The first edition: FEB. 1997
operate normally even after The fifth edition: NOV. 2010 [IMQ01]
the above measures are
taken, please contact RKC
sales office or the agent. ® RKC INSTRUMENT INC.
Replaceable dry batteries: HEADQUARTERS:16-6, KUGAHARA 5-CHOME, OHTA-KU
TOKYO 146-8515 JAPAN
Alkaline (AA Size [LR6], 1.5 V, 2 pcs.) or
PHONE: 03-3751-9799 (+81 3 3751 9799)
Manganese (AA Size [R6], 1.5 V, 2 pcs.)
E-mail: info@rkcinst.co.jp
FAX: 03-3751-8585 (+81 3 3751 8585)
IM350DP01-E5 NOV. 2010