You are on page 1of 10

SPESIFIKASI ALAT LABORATORIUM BSL 3 SPESIFIKASI ALAT LABORATORIUM BSL 3 ADENDUM II

No. NAMA BARANG SPESIFIKASI No. NAMA BARANG SPESIFIKASI


1 Double Door Autoclave for BSL 3 1 Double Door Autoclave for BSL 3
A Spesification A Spesification
a. Chamber Shape Round or Rectangular a. Chamber Shape Round or Rectangular
b. Chamber capacity/ volume (liters) 150 - 200 b. Chamber capacity/ volume (liters) 150 - 200
c, Power Consumtion (W) : less than 15000 c, Power Consumtion (W) : less than 15000
d. Sterilization Chamber & Door Materials Stainless steel 1.4404/AiSi316L d. Sterilization Chamber & Door Materials Stainless steel 1.4404/AiSi316L
e. Housing Autoclave Materials minimal Stainless steel 1.4301/ SS304 e. Housing Autoclave Materials minimal Stainless steel 1.4301/ SS304
f. Autoclave Door Hinged door/ Automatic sliding door f. Autoclave Door Hinged door/ Automatic sliding door
g. Lock Safety Yes, suitable for BSL 3 g. Lock Safety Yes, suitable for BSL 3
h. Temperature range ( C) up to 140 or higher h. Temperature range ( C) up to 130 or higher
i. Pressure range (bar absolute pressure) up to 3.5 or higher i. Pressure range (bar absolute pressure) up to 3 or higher
j. Weight (kg) : Less than 900 j. Weight (kg) : Less than 900
Eksternal Dimensions (HxWxD) (mm) Eksternal Dimensions (HxWxD) (mm)
k. : max 2000x max 1000x max1200 k. : max 2000x max 1200x max1200
(does not include base frame) (does not include base frame)
l. Controller : Touchscreen l. Controller : Touchscreen
m. Alarm systems : YES m. Alarm systems : YES
n. Bioseal/ biological sealing flange : YES n. Bioseal/ biological sealing flange : YES

B Includes of : : - Sterilizable exhaust air filter B Includes of : : - Sterilizable exhaust air filter
- RO Water purification system - RO Water purification system
- Vacuum system - Vacuum system
- Steam generator - Steam generator
- Rapid cooling system - Rapid cooling system
- LAN or USB for data transfer - LAN or USB for data transfer
- Software for monitoring/ analyzing - Software for monitoring/ analyzing
- Document IQ, OQ - Document IQ, OQ
- Original Manual Book - Original Manual Book
- Operational Training - Operational Training
- Certificate of Origin - Certificate of Origin
- Minimum 1 year guarantee - Minimum 1 year guarantee

2 AUTOCLAVE 2 AUTOCLAVE
A Spesification A Spesification
a Capacity : Minimum 50 L a Capacity : Minimum 50 L
b. Power source : 220 V, 50/60Hz b. Power source : 220 V, 50/60Hz
c. Power consumption : Max 2.0 kW c. Power consumption : Max 2.0 kW
d. Sterilization temp. range : 105-135 °C d. Sterilization temp. range : 105-135 °C
e. Max allowed pressure : 0.26 Mpa e. Max allowed pressure : 0.26 Mpa
g. Temp. display range : 45 to 135 °C or wider g. Temp. display range : 45 to 135 °C or wider
h. Chamber material : stainless steel (SUS304) h. Chamber material : stainless steel (SUS304)
i. Sterilization timer : 1-250 minutes or longer i. Sterilization timer : 1-250 minutes or longer

B Features: B Features:
- Selective mode : : sterilization of agar (liquid sterilization with - Selective mode : : sterilization of agar (liquid sterilization with
- Safety device/warning alarm : - Lack of water prevention device - Safety device/warning alarm : - Lack of water prevention device
- Interlock mechanism - Interlock mechanism
- Pressure safety valve - Pressure safety valve
- Over pressure power cut-off - Over pressure power cut-off
- Over temperature power cut-of - Over temperature power cut-of
- Included Standard Acessoris : Stainless steel wire basket - Included Standard Acessoris : Stainless steel wire basket

C Include of : - Integrated printer C Include of : - Integrated printer


- Operational Training & Installation - Operational Training & Installation
- Minimum 1 year guarantee - Minimum 1 year guarantee
- Original manual book - Original manual book
- Certificate of Origin - Certificate of Origin
3 INKUBATOR 3 INKUBATOR
A Applications A Applications
1. Research, pharmaceutics, medicine and food chemistry. 1. Research, pharmaceutics, medicine and food chemistry.

B Specification B Specification
1 Volume Min 50 L 1 Volume Min 50 L
2 Doors - Inner glass door 2 Doors - Inner glass door
- Outer door - Outer door
- Lockable door handle. - Lockable door handle.
3 Internal Material Stainless steel 3 Internal Material Stainless steel
4 Shelves Stainless steel 4 Shelves Stainless steel
5 Convection Natural convection 5 Convection Natural convection
6 Temperature Range at least 5 above ambient temperature to 70 6 Temperature Range at least 5 above ambient temperature to 70
7 Setting accuracy temperature 0.1 °C 7 Setting accuracy temperature 0.1 °C
8 Power Supply 220/230 V, 50/60 Hz 8 Power Supply 220/230 V, 50/60 Hz
9 Controller Display Single/ LED/ LCD display 9 Controller Display Single/ LED/ LCD display
10 Alarm Available 10 Alarm Available
11 Timer Available 11 Timer Available
12 Data Logging Ethernet interface/ USB port 12 Data Logging Ethernet interface/ USB port
13 Temperature sensor Pt 100 13 Temperature sensor Pt 100
14 Temperature Control According to DIN 12880 14 Temperature Control According to DIN 12880
15 15
C Include of : - Operational Training & Installation C Include of : - Operational Training & Installation
- Minimum 1 year guarantee - Minimum 1 year guarantee
- Original manual book - Original manual book
- Certificate of Origin - Certificate of Origin

4 WORKSTATION FOR 7 PIPETTES 4 WORKSTATION FOR 7 PIPETTES


- Compatible with Socorex Acura Manual 825 - Compatible with Socorex Acura Manual 825

Mengetahui, 7 Juni 2022


Pejabat Pembuat Komitmen

Bertha Lolo Lukita, S.Si, Apt., M.Farm


SPESIFIKASI CONTAINMENT BSL-3

JUMLAH HARGA
NO DESKRIPSI MERK & NOMOR KATALOG VOLUME SATUAN HARGA SATUAN (RP)
(RP)

I ENCLOSURE CONTAINMENT BSL-3 vendor yang berpengalaman sejenis 69 m2 IDR -

1 Set Ante Room 5 m2


1 Set BSL-3 Lab Suite 33 m2
1 Set BSL-3 Storage (Specimen Repository) 5 m2
1 Set R Sterilasi & Dekontaminasi 20 m2
1 Set R. Kontrol & Observasi 6 m2

Spesifikasi Ruang Containment:


Perbedaan tekanan antar zona (differential pressure) -12.5 ~ 25 Pa Pa
Tekanan di Weather Vestibule +20 Pa
Tekanan netral di R Sterilisasi & Dekontaminasi; R Kontrol & Observasi 0 Pa
Luas Containment BSL-3 69 m2
Volume ruang Containment BSL-3 193,2 m3
Pencahayaan 400-500 lux
ACH minimum (air change rate per hour) 15 ACH
Suhu 23˚C + 1 23 + 1 ˚C
JUMLAH HARGA
NO DESKRIPSI MERK & NOMOR KATALOG VOLUME SATUAN HARGA SATUAN (RP)
(RP)

RH (Relative humidity) 65% + 10


Supply dan exhaust dengan filtrasi, HEPA filter(Scan tested ≥ 99.99%)
Udara luar yang ditreatment (filtrasi dan kelembaban)

Lingkungan Containment BSL-3 dikendalikan secara manual dari panel kontrol dibantu dengan BMS (Building Management System) hanya untuk monitor sistem containment antara lain:
Sistem AHU, suhu, tekanan, ACH, smoke detector, CCTV, sistem pintu dan secure access dengan sistem alarm secara remote melalui SMS.
o
Sistem tata udara terdiri dari sepasang single pass AHU yang mampu mengendalikan suhu dengan ketat (+1 C), RH (+10), differential air pressure , dan ACH.
Fasilitas BSL-3 ini didukung genset yang dilengkapi dengan ATS (automatic transfer switch ).
Electrical requirement: 380, 3ph
Emergency Power Back-up 150KVA 3ph
Main Control Panel BSL-3
Sub Main Control Panel BSL-3

SISTEM ENCLOSURE RUANG KEDAP UDARA -

a. Ketahanan terhadap Tekanan sampai - 500 Pa -


b. Material
Insulated panel system ( sistem pintu dengan dan kusen berikut penggantung dan
- v
asesoris lainnya yang sesua)
- tahan terhadap proses dekontaminasi & pembersihan rutin/sering
c. Dinding
- Tebal minimal 50 mm -
d. Plafon
minimal 100 mm (dapat diinjak
- Tebal
untuk perawatan)
e. Lantai
Sistem heavy duty dengan vinyl berikut persiapannya atau epoxy resin 2 komponen
-
bebas nonilfenol
- lantai self-levelling dengan kemiringan ke arah pembuangan
- Halus, tidak licin dan tanpa sambungan
- Anti microbial
- Tahan Kimia
- Tahan gesekan
f. Pintu
- Jumlah pintu sesuai sistem panel enclosure 5 Nos

- Dilengkapi vision panel dan asesoris

g. Jendela -
- Jumlah jendela sesuai sistem panel enclosure 8 Nos -
- Insulated double glazing -
h. Emergency Shower
- Terpasang di dinding atau plafon (instalasi dan pengkedapan) 1 Nos
i. Portable Fire Extinguisher, 3.5kg c/w storage box, sealed off 3 Nos
j. Handwash basin c/w automated faucet 2 Nos -
k. Gas System (CO2 ): 1 Nos -
CO2 gas tubing with control valves, header pipes, standby and working tanks, storage
- -
rack and safety valves, and sealing off
l. Installation System for Passthrough Autoclave: -
- Bioseal
- Electrical connection 3ph*400 V/50/60Hz, 32A CEEKON - socket
- Drain DN 50, HT-Pipe with drain trap, height max. 300mm
JUMLAH HARGA
NO DESKRIPSI MERK & NOMOR KATALOG VOLUME SATUAN HARGA SATUAN (RP)
(RP)

Demineralized water 3/4" pipe, stop valve external thread, pressure 0,5-2 bar,
-
according to BIN EN 285
Cold water supply, 3/4" pipe, stop valve external thread, pressure min. 2.5-5 bar,
- o
max. calcium hardness 10 , 1,783 mmoL/L
Compressed air, 1/2" pipe, stop valve, internal thread, 5-7 bar, according as ISO
-
08573-1:2010 category 4

ELEKTRIKAL & ELEKTRONIK -

Sistem Elektrikal, Elektronik, Penerangan, Plumbing, CCTV, LAN, Telepon, dan Sensor; tidak termasuk Building Management System

a. Electrical Panels
c/w feeder cable to Panel Room (lokasi R Panel di gedung eksisting) -

b. HVAC Control Panel


c/w feeder cable, VSD contoller for AHU & both exhaust fans and interlocking system with -
hardwiring for Supply and Exhaust line ensuring no possitive pressure
c. LED Lighting (Illumination 400-500lux)
c/w wiring, switches, sealing off -

d. Electrical outlets
c/w wiring, sealing off -

e. CCTV monitor camera c/w cabling installations, data storage, NVR, 22 in monitor, UPS and
sealing off -

- Jumlah 12 Nos

5 (inside BSL-3 unit) ; 7


- Lokasi
(outside/perimeter BSL-3)

f. LAN outlets
c/w wiring inside the lab and switch hub panel inside Control Room and sealing off -

- Jumlah 30 Nos

18 (inside BSL-3 unit) ; 12


- Lokasi (outside BSL-3 containment
zone)

- Integration to building's system by others

g.
Telephone outlets
-
c/w wiring inside the lab and terminal box panel inside Control Room

- Jumlah 8 Nos

- Lokasi 8 (inside BSL-3 unit)

- Integration to Campus' system by others

h.
Smoke detectors
-
c/w wiring inside the lab to the closest detector point and sealing off

- Integration to Campus' system by others

i. Secure Access System & Interlocking Door system -

i controller + power supply + battery


JUMLAH HARGA
NO DESKRIPSI MERK & NOMOR KATALOG VOLUME SATUAN HARGA SATUAN (RP)
(RP)

accessories:
ii
reader, release button, emergency break glass, EM lock, bracket, Emergency button

iii software & installation charge, consumable

iv test-com & training

SISTEM TATA UDARA: -

The BSL-3 HVAC System :


- Custom Air Handling Unit c/w cabling & piping installations, silencer, humidity control, electric heater, fresh air filtration bank with maximum 95% efficiency filter media and filter media
replacement for 1 year
- Supply Air Volume c/w concaved perforated stainless steel 304 air grille and manual low leak stainless steel volume controler
- Supply Air GI Ductwork c/w transition reducers from units, plenum boxes, packing seal, insulation, support & hanger, label and all must be performing in such thus sound level inside the lab
is below 58 dB
- Exhaust Air System c/w redundant exhaust fans, minimum exit velocity of 2000 fpm, exhaust stack, non return dampers, cabling installation, side access HEPA Filter with stainless steel 304
housing and decon ports, maximum efficiency 99.97% @0.3 micron and filter media replacement for 1 year
- Exhaust Air System c/w redundant exhaust fans, minimum exit velocity of 2000 fpm, exhaust stack, non return dampers, cabling installation, side access HEPA Filter with stainless steel 304
housing and decon ports, maximum efficiency 99.97% @0.3 micron and filter media replacement for 1 year
- Exhaust Air Volume c/w flat perforated stainless steel 304 air grille and manual low leak stainless steel volume control
- Exhaust Thimble Duct powder coat steel hood c/w manual low leak stainless steel volume control
- Exhaust Air GI Ductwork c/w transition reducers from units, plenum boxes, packing seal, insulation, support & hanger, label and all must be performing in such thus sound level inside the
lab is below 58 dB
- Air tight dampers c/w reducers and installation
- Constant air volume valves for both supply and exhaust lines c/w accessories
- Lab decon ports (optional)

II SISTEM CONTAINMENT BSL-3 - Peralatan Laboratorium IDR -

Incubator 1 set IDR IDR -


Applications: Research, pharmaceutics, medicine, and food chemistry

Specifications:
Volume: minimum 50L
Doors: - Inner glass door
- Outer door
- Lockable door handle
Internal material: Stainless steel material
Shelves: Stainless steel
Convection: Natural convection
Temperature range: o
at least 5 above ambient temperature to 70 C
Setting accuracy temperature: o
0.1 C
Power supply: 220/230 V, 50/60 Hz
Controller display: Single/ LED/ LCD display
Alarm: Available
Timer: Available
Data Logging: Ethernet interface/ USB port
Temperature sensor: Pt 100
Temperature control: According to DIN 12880
Include: - Operational Training & Installation
- Minimum 1 year guarantee
- Original manual book
- Certificate of Origin
Double door Autoclave 1 set IDR IDR -
Termasuk instalasi dan koordinasi
JUMLAH HARGA
NO DESKRIPSI MERK & NOMOR KATALOG VOLUME SATUAN HARGA SATUAN (RP)
(RP)

Specification:
Chamber Shape: Square
Chamber capacity/volume (liters): 150-200
Power consumption (W): less than 15000
Sterilization Chamber & Door Materials: Stainless steel 1.4404/AiSi316L
Housing Auotclave Materials: Stainless steel 1.4301/SS304
Autoclave door: Hinged door/Automatic sliding door
Lock Safety: Yes, suitable for BSL-3
o up to 140 or higher
Temperature range ( C):
Pressure range (bar absolute pressure): up to 3.5 or higher
Weight (kg): Less than 900kg
External dimensions (HxWxD) mm max 2000x1000x1200 (does not include base
frame)
Controller: touchscreen
Alarm system: YES
Includes: Sterilizable exhaust air filter
RO Water purification system
Vacuum system
Steam generator
Rapid cooling system
Network and USB port for data transfer
Software for monitoring/analyzing
Document IQ, OQ
Original Manual Book
Operational Training
Certificate of Origin
Minimum 1 year guarantee
GLP Kit Micropipette 0 set IDR IDR -

Workstation For Socorex Pipettes 0 set IDR IDR -


- Compatible with Socorex Acura Manual 825

Autoclave 0 set IDR IDR -


Specification:
Capacity: minimum 50L
Power source: 220V, 50/60Hz
Power consumption: Max 2.0 kW
Sterilisation temp. range: 105-135oC
Max. allowed pressure 0.26 Mpa
Temp. display range 45 to 135 °C or wider
Chamber material stainless steel (SUS304)
Sterilization timer 1-250 minutes or longer
Features:
Selective mode: sterilization of agar (liquid sterilization with
warming), sterilization of liquid, sterilization
of solid/medical instrument
Safety device/warning alarm: - Lack of water prevention device
- Interlock mechanism
- Pressure safety valve
- Over pressure power cut-off
- Over temperature power cut-off
Included Standard Acessoris: Stainless steel wire basket
Include: - Integrated printer
- Operational Training & Installation
- Minimum 1 year guarantee
- Original manual book
- Certificate of Origin

III SISTEM CONTAINMENT BSL-3 - Mebel Laboratorium IDR -


JUMLAH HARGA
NO DESKRIPSI MERK & NOMOR KATALOG VOLUME SATUAN HARGA SATUAN (RP)
(RP)

a. The Pass through box is a static interlocking pass-box that serves as a “Safe Barrier” in 2 set IDR - -
transfer of materials and/or working tools between two environments with different level
of containment, minimizing the passage of duct and air-borne particles that are present in
the environment and in the objects passing through.
Passthrough Box (Air Tight with interlock & intercom (approx. 400x400x450mm) on 1 set
-
containment unit
Passthrough Box (Air Tight) with interlock & intercom (approx.600x600x600mm) at 1 set
-
BSL-3 Lobby
SS 304
- Material

b. Lab Bench 14,4 m' IDR - IDR -


Materials
- Non-porous Phenolic Resin lab table top with backsplash (per detailed drawings)

Custom steel frame with powdercoated finished to withstand cleaning disinfectant


-
agents and

- Able to withstand 200-250 kg equipments.


Installations must be sealed off in accordance to laboratory guildelines tightness
c. Lab Sink 2 set IDR - IDR -
Materals
Chemical resistance Polypropylene sink with anti siphon bottle trap c/w accessories
-
and sealing off in accordance to laboratory guidelines tightness

- Laboratory faucet c/w one way wrist operated handle, antibackflow and shutoff valve

f. Gowning Rack 2 set IDR - IDR -

Materials
Custom steel cabinet with powdercoated finish to withstand cleaning disinfectant
- agents, installations must be sealed off in accordance to laboratory guildelines
tightness

- Provide with Include charging sockets for PAPR charging and space for PPE storage

IV BSL-3 SUPPORT SYSTEM IDR -

IV.1 BUILDING MANAGEMENT SYSTEM 1 set IDR - IDR -


a. Building Management System to monitor system:
- Workstation/Printer for BMS
b. Security System
- Security System Platform
- Dedicated Server ( upgradeable)
- IR Cameras
- WEB500 I/O Point software controls plattform
- CoppertreeCube-S-SQL
- I/O Modules
- Sensors
- Display for Temperature, RH and Pressure Differentials, AHU flowrate, exhaust and door
indicator
Compatible with BSl-3 System

IV.2 DIESEL GENERATOR SET 1 set IDR - IDR -


JUMLAH HARGA
NO DESKRIPSI MERK & NOMOR KATALOG VOLUME SATUAN HARGA SATUAN (RP)
(RP)

Capacity: 150KVA
Type: Silent
System 380, 3 Phase
Frequency 50Hz
w/ Automatic Transfer Switch
Accessories: Engine Instrument Panel, Electronic Governor, Isolation Mounting, Vibration
Mounting, Unit c/w Skid, Standard mounted Radiator, Dry Type Air Cleaner, Flexible Exhaust
Connection.
Supply and installation system netral unit genset and grounding genset include cable NYA 70
mm² c/w pvc cables, earth bar, test clamp, earthing pits, clamp pipe control box + cover
include other equipment.
Termasuk alat bantu, Tangki solar, pompa tangan, Solar untuk uji, Testing & Commissioning
Training
Manual book
Garansi

IV.3 DUMBWAITER 1 set IDR - IDR -

- Capacity: 150/200kg
- Type: Floor opening
- Car dimension (inside) minimum 850x900x1200mm
- Motor 1,5 KW
- Terpasang, termasuk alat bantu,Testing & Commissioning

V OTHER PROVISIONS IDR -

TRAINING BSL-3 Use & Maintenance


V.1 1 ls IDR - IDR -
Pelatihan untuk minimal 10 orang personel didampingi tenaga ahli Biosafety

Materi pelatihan:
•Biosafety & Biosecurity principles and practices
•Microorganism classification based on Risk Group
•Biorisk Management
•Biological Safety Cabinet (BSC)
•Laboratory Acquired Infection (LAI)
•Biosecurity & Dual Use Research of Conduct (DURC)
•Engineering in Biosafety - Operation & Maintenance of BSL-3
•Waste Treatment
•Chemical Safety
•IBC & BSO
•Handling and transportation
•SOP
•Emergency Response Plan

V.2 VALIDATION OF BSL-3 DESIGN & CONSTRUCTION 1 ls IDR - IDR -

Validation for Biosafety & Biosecurity in High Containment Facilities


Menggunakan alat ukur yang sudah dikalibrasi, didampingi tenaga ahli bersertifikat NSF 49

V.3 MONITORING & ROUTINE MAINTENANCE 1 tahun IDR - IDR -

Monitoring serta perawatan dan penggantian spare part rutin minimal untuk pre-filter dan
medium filter selama setahun

V.4 POWERED AIR PURIFYING RESPIRATOR 2 unit IDR - IDR -


Mengetahui, 7 Juni 2022
Pejabat Pembuat Komitmen

Bertha Lolo Lukita, S.Si, Apt., M.Farm

You might also like