You are on page 1of 3

Journal of Analytical Toxicology, Vol.

21, October 1997

TechnicalNote [

A Simple SpectrophotometricAssayfor the


Measurement of Soluble Silica in Water
Barbara R. Manno 1, Imad K. Abukhalaf2,, and Joseph E. Manno 1,3
Departments of 1Psychiatry,2Medicine, and 3EmergencyMedicine, Louisiana State UniversityMedical Center,

Downloaded from https://academic.oup.com/jat/article-abstract/21/6/503/735411 by guest on 31 July 2019


School of Medicine, 1501 Kings Highway, P.O. Box 33932, Shreveport, Louisiana 71130-3932

Introduction with this procedure to prevent silicate contamination from


the glassware. Therefore, all solutions were prepared in plastic
The National Committee for Clinical Laboratory Standards containers (polystyrene or polypropylene). Polystyrene test
(NCCLS, 1) recommends that Type I water used in clinical tubes were used in the preparation of standard curves and
laboratories must be virtually silica (SiO2) free (< 0.05 rag/L). testing of unknowns.
The recommendation has been adopted by the Collegeof Amer- Sodium metasilicate was dissolved in water to give a stock
ican Pathologists (CAP) with the inclusion of a question silicate concentration of 1.0 mg/mL from which a seven-point
regarding soluble silicate in water in CAP Laboratory Accred- standard curve covering a concentration range of 0-0.1 mg/L
itation Program checklists (2). These requirements are based was constructed (TableI). Standard solutions and water samples
upon silica's direct interference with spectrophotometric were used at room temperature (approximately20-25~ To 5.0
measurements at specific wavelengths of the light spectrum. rnL of each working silicate standard solution and unknown(s),
Therefore, its presence in water not only interferes with trace 125 pL of Molybdate 3 Reagent (catalog # 1995-37, Hach Silica
metal and electrolyte analyses but also could adversely affect Testing Kit) was added. After vortex mixing, the tubes were al-
many enzymatic determinations such as those procedures lowed to stand for 4 rnin. Then 125 pL of Hach Citric Acid
using alcohol dehydrogenase and various phosphatases (1). Reagent (catalog# 22542-37) was added to each tube, and the re-
A commercial silica testing kit (catalog # 22550-00, Hach action was allowedto proceed for 1 rain. TWo-hundredand fifty
Silica Test Kit, Hach, Loveland, CO) was evaluated. Kit specifi- microliters of the Hach Amino Acid F Suspension (catalog #
cations indicated that it measured silica concentrations in the 22538-69, AAFS)was then added to each tube. The AAFSmust
range of 0-1 mg/L. Measurement of the color intensity of the be prepared immediately before use by suspending the contents
resulting silicate-molybdate complex required visual compar- of one AAFSreagent pillow in 1.0 mL of distilled water. Prepa-
ison of the developed color to a color disc directed toward an ration of one pillowaccording to these instructions providedsuf-
artificial or natural light source and depended on color com- ficient reagent for four tubes for testing. Tubeswere immediately
parison by the naked eye. Hach Silica Testing Kit reagents were vortex mixed and absorbance measured at 803 nm using a
used in the development of a modified procedure for the de- Hitachi U-2000 spectrophotometer (San Jos~, CA) using a water
tection and more accurate quantitation of low concentrations blank.
(< 0.05 rag/L) of soluble silica in water. Final quantitative results are calculated using the following
formula (3):
Cu = (AuC,) + (As-Au)

Materials and Methods Cu = Concentration of the unknown (mg/L)


Au = Absorbance of sample without added silicate
Reagents were purchased individually from the Hach Com- Cs = Concentration of the standard (rag/L)
pany. Sodium metasilicate nonahydrate (Na203Sio9H20, min- As = Absorbance of sample with added silicate
imum 98%, FW 284.2) used for the preparation of the stock
standard silicate solution was purchased from Sigma Chemical
(catalog # S-4392, St. Louis, MO). Results and Discussion
It is important that use of borosilicate glassware is avoided

* PresentAddress:MorehouseSchoolof Medicine, Departmentof Pharmacologyand


Silica occurs in soluble and colloidal forms. Colloidal silica
Toxicology,720 WestviewDrive, Atlanta, GA 30310. removal from water may be achieved by either ion exchange

Reproduction (photocopying) of editorial content of this journal is prohibited without publisher's permission. 503
Journal of Analytical Toxicology, Vol. 21, October 1997

chromatography or more effectively by dis-


Table I. Details of the Standard Curve Construction for the Quantitation of tillation or the combination of both sys-
Soluble Silica in Water
tems. Soluble silica may be eliminated from
Concentrationof 0.1 mg/t Molybdate 3 Citric acid F Amino acid F
water by the use ion exchange and micro-
silica Water Sodium reagent reagent suspension* filtration or their combination.
(rag/L) (mL) metasilicate(mL) (pL) (pl) (pL) Both soluble silica and phosphates co-
exist in most primary water supplies. Both
Blank 5.0 0 0 0 0 react with ammonium molybdate at acidic
pH to form heteropolyacids. Colloidal silica
0.02 4.0 1.0 125 125 250
is extremely difficult to measure because it
0.04 3.0 2.0 125 125 250
is unreactive to mdybdate. Molybdophos-
0.05 2.5 2.5 125 125 250 phoric acid, the product of the phosphate-
0.06 2.0 3.0 125 125 250 molybdate reaction, is labile to citric acid
hydrolysis, but its silica counterpart, molyb-

Downloaded from https://academic.oup.com/jat/article-abstract/21/6/503/735411 by guest on 31 July 2019


0.08 1.0 4.0 125 125 250
0.10 0.0 5.0 125 125 250 dosilicic acid, is resistant to this hydrolysis.
An obvious concern with the method would
Unknown 5.0 0.0 125 125 250
be the probability of phosphate interference
* Vortex mix the Amino Acid F suspension before adding it to the test tube. Vortex mix the test tube before in a spectrophotometric method. Phosphate
moving to the next one, and so on.
interference was negligible in this proce-
dure for two reasons. First, excess amounts
of citric acid were added to destroy the
0.04 molybdophosphoric acid. Second, because
this complex absorbs light in the yellow
X range of the spectrum and molybdosilicic
acid absorbs light in the blue range, reading
0.035 the absorbance of the reaction mixture at
803 nm would negate potential phosphate
interference.
A method referenced by NCCLS (1) for
0.03 X
the measurement of soluble silica in water
IlL is similar to the water silicate method de-
scribed in Standard Method for the Exami-
0.025
nation of Water and Waste Water (4). The
methods are tedious and time consuming
X
for a variety of reasons. Some of these
1
X 0 reagents have relatively short shelf lives, as
0.02 short as two weeks in some instances. The
use of at least seven chemicals and acids,
.<
X some of which have very specific storage
conditions, is required. Large amounts of
0.015 g
stock material are needed, and a series
of standards is specified, each in a volume of
100 mL of water. Because water testing is
conducted on a periodic basis, the com-
0.01
0 plexity of the method and the lability of the
g reagents translate to increased expenditure,
analyst time, specific chemical storage re-
0.005 quirements, and hazardous waste disposal.
Although the Hach Silica Test Kit with
visual color comparison may be useful for
field work for the estimation of high con-
centrations of silicates in water, its use at
0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09 0.1 0.11 low silicate concentrations required for
Soluble silicate concentration (rag/L) clinical laboratory applications was limited.
It was virtually impossible to reliably com-
Figure 1. Scatter plot of replicate absorbance readings (n = 5) at 803 nm wavelength at six con- pare the intensities of two blue colors with
centrations with the spectrophotometer set using a water blank. the naked eye at low silicate concentrations
(< 0.07 mg/L).

504
Journal of Analytical Toxicology,Vol. 21, October 1997

testing to meet Type I specifications.


Table !!. Exampleof ResultsObtained usingTwo Spectrophotometric
Silicate Methods Replacement of all cartridges in the system
resulted in no measurable soluble silicate in
Before system cartridge After system cartridge water from the purification system.
replacement replacement Analytical issues associated with practical
quantitative analysis of water for soluble
Sample source Silicate method I' Silicate method II r Silicate method I' Silicate method II t silica at very low concentrations have been
(mg/mL) (mg/mL) (mg/L) (mg/L)
addressed. It was found that the Hach Silica
Tap water 295 300 - _ Test Kit with visual color comparison was
not sufficiently sensitive at soluble silica
PostMilli-Q system - 0.069 0.005 0.00o concentrations below 0.07 mg/mL. The
* Reference 3.
reagents supplied with the kit when used
t Method described herein. with a spectrophotometer at a wavelength
of 803 nm rather than the visual filter

Downloaded from https://academic.oup.com/jat/article-abstract/21/6/503/735411 by guest on 31 July 2019


The modification resulted in a simple spectrophotometric reading device overcomes the limitations of the method. There
procedure for the determination of soluble silica concentra- are several other issues, however, surrounding the NCCLS
tions in water. Because soluble silica in the water supply of guidelines themselves which are ambiguous and that must be
certain geographical locations is high, CAP requires that labo- addressed. The NCCLS definition for "source water" was un-
ratories test their water soluble silica content to determine if it clear, that is, water supplied to or water exiting the purification
exceeds the recommended limit of___0.05 mg/L. system.
The spectrophotometric method described was found to be It is recommended that water that exits the final water
linear over the concentration range of 0-0.5 mg/L. Data, how- purification system be selected for soluble-silica testing be-
ever, are only included in the range of 0-0.1 mg/mL because we cause this will be the water that is used in the laboratory, and
focused our work around the maximum allowable silicate con- laboratories that outsource their soluble-silica water testing
centrations of 0.05 mg/mL (Figure 1). Linear regression ana- ascertain the analytical methodology used by the reference
lysis of data (5) from replicate standard curves (n = 5) de- laboratories, including the limit of detection and limit of quan-
scribed the curve ~ = m x + b) as absorbance = 0.3466X + titation of the method.
1.3286E-03 with a correlation coefficient of 0.996.
Derivation of the final quantitative results is accomplished by
calculations based upon measurement of small concentrations
of ions in solutions (standard addition method) as is done in References
atomic absorption spectroscopy (3). This approach is necessary
because the water used in all solutions used in the procedure
1. National Committee for Clinical Laboratory Standards. Prepara-
as well as the testing of the source water is essentially from the tion and Testingof Reagent Water in the Clinical Laboratory. 2nd
same source. It also accommodates the measurement of very ed. Approved Guideline. NCCLS document C3-A2. NCCLS,
low concentrations of target analytes in solution. Villanova, PA, 1991.
Our laboratory waterpurification system consists of a Culligan 2. College of American Pathologists (CAP), Laboratory Accreditation
mixed bed ion exchange and carbon purification system (Cul- Program, Northfield, IL, 1996.
3. I. Rubeska and B. Moldan. Atomic Absorption Spectrometry.CRC
ligan Technologies, Northbmok, IL) in tandem with a MilliQ Press, Cleveland, OH, 1967.
purification system consisting of one Super C and two ion 4. A.E. Greenberg, R.R. Trussell, and L.S. Clesceri, Eds. Standard
exchange cartridges and a 0.22-t~mfilter (Milli-Q,Bedford,MA). Methods for the Examination of Water and Waste Water, 16th ed.
Table II shows examples of results using both methods before American Public Health Association, Washington, D.C., 1985.
and after system maintenance. It was found that water produced 5. R.T.Tallarida and R.B. Murray. Manual of Pharmacologic Calcu-
lations with Computer Programs. Springer-Verlag, New York,
before replacement of the single Super C and two ion exchange 1981.
cartridges in the MilliQ System but after replacement of the Manuscript received March 25, 1997;
Culligan cartridges still passed resistivity and bacterial culture revision accepted June 13, 1997.

505

You might also like