Professional Documents
Culture Documents
University of Balamand
Laboratory report 1
Experiment 1
Maria F.Hajjar
Presented to:
Maria Ch. Moujaes Dr. Hala Chamieh
Mariam A. Hamze
Karam N. Al Aryan 1
Company Your Company Name (C) Copyright (Print Date) All Rights Reserved
Logotype
Fatima Kataya
TABLE OF CONTENTS
1. Background information
1.1. What is glucose
1.2. Metabolism
1.3. Chemical structure
1.4. Normal general range in blood
1.5. Associated diseases
3. Materials required
3.1. Laboratory equipments
3.2. Specimen used
3.3. Reagents used
4. Methodology
4.1. Usage of micropipette
4.2. Tubes
4.3. Incubation
5. Results
5.1. Calculating blood glucose levels
5.2. Spectrophotometer usage
5.3. Calculating the absorbance
2
Company Your Company Name (C) Copyright (Print Date) All Rights Reserved
Logotype
1. Background information
1.1.What is glucose
Glucose is the most important carbohydrate fuel in the body. In the fed state, the
majority of circulating glucose comes from the diet; in the
fasting state, gluconeogenesis and glycogenolysis maintain
glucose concentrations. Very little glucose is found in the
diet as glucose; most is found in more complex
carbohydrates that are broken down to monosaccharides
though the digestive process.
1.2. Metabolism:
Your body is designed to keep the level of glucose in your blood constant. Beta cells in your pancreas monitor
your blood sugar level every few seconds. When your blood glucose rises
after you eat, the beta cells release insulin into your bloodstream. Insulin
acts like a key, unlocking muscle, fat, and liver cells so glucose can get
inside them.
In type 1 diabetes, your body doesn't have enough insulin. The immune system
attacks and destroys cells of the pancreas, where insulin is made.
In type 2 diabetes, the cells don't respond to insulin like they should. So the
pancreas needs to make more and more insulin to move glucose into the cells.
Eventually, the pancreas is damaged and can't make enough insulin to meet the
body's needs.
3
Company Your Company Name (C) Copyright (Print Date) All Rights Reserved
Logotype
2. Purposes & objectives
2.1. Objective & Purposes
The purpose of this experiment is to determine the amount of glucose levels in human serum
with a quantitative calorimetric enzymatic method in vitro
4
Company Your Company Name (C) Copyright (Print Date) All Rights Reserved
Logotype
3. Materials required
3.1. Laboratory Equipments
Test tubes
Two Micropipettes
Cuvettes
Tube rack
Spectrophotometer
Serum heparinized
Buffer reagent:
5
Company Your Company Name (C) Copyright (Print Date) All Rights Reserved
Logotype
-Phosphate buffer reagent (ph= 7)
Standard solution:
4. Methodology
Steps to follow when using a micropipette:
Select the volume and set the tip
Press and hold the plunger at the first stop
Place the tip in the liquid and slowly release the plunger
Pose for a seconde and then move the tip and inserted into the delivery vessel
Press the plunger to the second stop and pose for two seconds
Remove the tip and release the plunger
Inject the tip into the waste container
6
Company Your Company Name (C) Copyright (Print Date) All Rights Reserved
Logotype
Tube 1: Blank Tube 2: standard
2-standard - 10 µl -
3-blood,serum
- - 10µl
sample
Mix your tubes well
Incubate all the tubes for 15
minutes at 37 degree Celsius
After the incubation, the tubes were analyzed using a
spectrophotometer to measure the absorbance of each one of them at a
specific wavelength
The blank absorbance was substracted from the sample
Measure the absorbance of the sample
Measure the absorbance of the standard
Use the values of the absorbances calculated to calculate the blood sugar concentrations in the
sample tube, using a calibration curve based on the standard absorbance values
Record your results
7
Company Your Company Name (C) Copyright (Print Date) All Rights Reserved
Logotype
The results were interpreted according to the established guidelines for normal and abnormal blood
sugar levels.
Any discrepancies or unexpected results were noted and further investigation was conducted as
needed.
2. Set the wavelength dial to 600 nm. This is a wavelength of light that gets absorbed readily
3. With no vial in the spectrophotometer, the light path is closed. Therefore no light is transmitted, or in
other words there is an infinite absorption.
4. Now fill a cuvette (a small test tube that is 12 mm in diameter and 100 mm in length) with Blank 1.
5. Insert the cuvette containing Blank 1 into the sample chamber. As you do the light path will be opened.
This blank solution does not contain any Grape Kool-Aid, and so the absorbance should be set to zero.
Use the right knob to set the absorbance to zero.
6. Next fill a cuvette with Grape Kool-Aid test solution. Insert it into the spectrophotometer and record its
absorbance in question number one of your assignment.
5. Results
Calculations :
0.117× 100
Formula : 0.223 = 50 mg / dl of glucose in blood serum ( hypoglycemic)
The amount of glucose obtained in the serum sample was less than the normal range : 75-
110mg/dl
8
Company Your Company Name (C) Copyright (Print Date) All Rights Reserved
Logotype
The glucose concentration in the serum should’ve been more than 65 mg/dl, since that the serum
was obtained from man
In the discussion/interpretation part we will discuss why, and what was the reason for this error
and false calculations
Why was the amount of glucose calculated false? What is the error
that was done in the lab that has lead to a false result?
Adjusting the volume on micropipettes
1. During extraction of the human blood serum, we might have not correctly extracted the
right amount of volume which is 10µl
2. We might have not released the thumb knob correctly.
3. We might have not released slowly the serum collected in our tube, which will lead to
bubbles formation, hence falsifying our results
Our results showed 50 mg/dl of glucose in the human serum, however the right answer should be
within the normal range.
9
Company Your Company Name (C) Copyright (Print Date) All Rights Reserved
Logotype
Why did we use micropipettes not normal pipettes?
Micropipettes are the most important and latest technical tool for the laboratories. There are so many
different types of laboratories and in all of them people have to work with liquids and semi liquids. For
getting perfect volume of them they need something which can carry the material in perfect volume and
can release it at any required place.
10
Company Your Company Name (C) Copyright (Print Date) All Rights Reserved
Logotype
11
Company Your Company Name (C) Copyright (Print Date) All Rights Reserved
Logotype