Professional Documents
Culture Documents
JAN-1186
ECDIS
ROUTE PLANNING SYSTEM
SERVICE MANUAL
Cautions for High Voltage
High voltages, ranging from several hundreds to tens of thousands of volts, are used in
electronic apparatus, such as radio and radar instruments. These voltages are totally
harmless in most operations. However, touching a component inside the unit is very
dangerous. (Any person other than authorized service engineers should not service,
inspect, or adjust the unit.)
High voltages on the order of tens of thousand volts are most likely to cause instant
deaths from electrical shocks. At times, even voltages on the order of several hundred volts
could lead to electrocution. To defend against electrical shock hazards, don't put your hand
into the inside of apparatus. When you put in a hand unavoidably in case of urgent, it is
strongly suggested to turn off the power switch and allow the capacitors, etc. to discharge
with a wire having its one end positively grounded to remove residual charges. Before you
put your hand into the inside of apparatus, make sure that internal parts are no longer
charged. Extra protection is ensured by wearing dry cotton gloves at this time. Another
important precaution to observe is to keep one hand in your pocket at a time, instead of
using both hands at the same time.
It is also important to select a secure footing to work on, as the secondary effects of
electrical shock hazards can be more serious. In the event of electrical shocks, disinfect the
burnt site completely and obtain medical care immediately.
i
Method of First-Aid Treatment
(1) Refrain from touching the patient carelessly as a result of the accident; the first-aider
could suffer from electrical shocks by himself or herself.
(2) Turn off the power calmly and certainly, and move the patient apart from the cable
gently.
(3) Call or send for a physician or ambulance immediately, or ask someone to call doctor.
(4) Lay the patient on the back, loosening the necktie, clothes, belts and so on.
(6) Opening the patient's mouth, remove artificial teeth, cigarettes, chewing gum, etc. if
any. With the patient's mouth open, stretch the tongue and insert a towel or the like into
the mouth to prevent the tongue from being withdrawn into the throat. (If the patient
clenches the teeth so tight that the mouth won't open, use a screwdriver or the like to
force the mouth open and then insert a towel or the like into the mouth.)
(7) Wipe off the mouth to prevent foaming mucus and saliva from accumulating.
ii
☆Treatment to Give When the Patient Has a
Pulse Beating but Has Ceased to Breathe
*Performing mouth-to-mouth artificial respiration - Fig. 1
(1) Bend the patient's face backward until it is directed to look back. (A pillow may be
placed under the neck.)
(2) Pull up the lower jaw to open up the airway. (to spread the airway)
(3) Pinching the patient's nose, breathe deeply and blow your breath into the patient's
mouth strongly, with care to close it completely. Then, move your mouth away and
take a deep breath, and blow into his or her mouth. Repeat blowing at 10 to 15 times
a minute (always with the patient's nostrils closed).
(4) Continue artificial respiration until natural respiration is restored.
(5) If the patient's mouth won't open easily, insert a pipe, such as one made of rubber or
vinyl, into either nostril. Then, take a deep breath and blow into the nostril through
the pipe, with the other nostril and the mouth completely closed.
(6) The patient may stand up abruptly upon recovering consciousness. Keep the patient
lying calmly, giving him or her coffee, tea or any other hot drink (but not alcoholic
drink) to keep him or her warm.
iii
☆Treatment to Give When the Patient Has No
Pulse Beating and Has Ceased to Breathe
*Performing cardiac massage - Fig. 2
If the patient has no pulse beating, with the pupils open and no heartbeat being heard,
the patient has a cardiac arrest and requires immediate artificial respiration. Continue
this until a medical specialist arrives, and follow his or her directions after that.
(1) Putting one hand on about the lower one third of the patient's ribs and the other hand
over the back of the first, with your elbow fully stretched (with bended elbow, you
can’t press to the extent the patient’s ribs are depressed), apply your body weight to
the hands to press the patient's body until it is depressed about 2 cm (Repeat this
about 50 times a minute). (Cardiac massage.)
(2) If only one first-aider is available, perform a cardiac massage about 15 times and
then give mouth-to-mouth artificial respiration 2 times. Repeat this sequence. If two
first-aiders are available, while one person performs a cardiac massage 15 times,
and the other should give mouth-to-mouth artificial respiration 2 times. Repeat this
sequence (combined cardiac massage and mouth-to-mouth artificial respiration
method).
(3) Check the patient's pupils and feel the pulse from time to time. When the pupils are
restored to normal and the pulse begins to beat regularly, stop treating and keep the
patient calm while giving him or her coffee, tea or any other hot drink to keep him or
her warm while watching him or her carefully.
① ②
③ ④
iv
PREFACE
This Service Manual describes the maintenance of JAN-2000/1186
ECDIS/Route Planning System. For the best operation and performance
result of JAN-2000/1186 ECDIS/Route Planning System, please read and
understand this manual thoroughly before the maintenance.
v
● Before Operation ●
Pictorial Indication
Various pictorial indications are included in this manual and are shown
on the component equipment so that you can operate them safely and
correctly and prevent any danger to you and / or to other persons and
any damage to your property during operation. Such indications and
their meanings are as follows.
Please understand them before you read this manual:
Warning label
There is a warning label on the top cover of the equipment.
Do not try to remove, break or modify the label.
vi
● Precautions Upon Equipment Operation ●
The power of the display unit (NWZ-173-E) is not turned off even when the
power of the processing unit (NDC-2000) is turned off.
Before connecting/removing the power cable of the display unit, make sure to
turn off the breaker of the main power line. Otherwise, an electric shock or
failure may occur.
Before replacing the fan filter, make sure to turn off the power and confirm
the fan stops.
Disconnect the cables before pulling out the base fixing the hard disk. If the
disk is forcibly pulled, the cables and the connectors may be damaged.
Before restoring system by the [Recovery Boot CD], read the description to
fully understand the function of the [Recovery Boot CD].
If you perform restoration with abnormal disk configuration, the equipment
may malfunction.
vii
CONTENTS
CONTENTS
1. CONFIGURATION.....................................................................1-1
1.1 Components............................................................................................................................................ 1-1
1.1.1 JAN-2000........................................................................................................................................... 1-1
1.1.2 JAN-1186 ........................................................................................................................................... 1-1
1.2 Optional Accessories ............................................................................................................................. 1-2
1.2.1 JAN-2000........................................................................................................................................... 1-2
1.2.2 JAN-1186 ........................................................................................................................................... 1-3
1.3 Outline Drawing ...................................................................................................................................... 1-4
1.4 Specifications ......................................................................................................................................... 1-9
1.4.1 Specifications of NDC-2000 Processor Unit ................................................................................. 1-9
1.4.2 Specifications of Optional Function .............................................................................................. 1-9
1.4.3 Composition Unit ........................................................................................................................... 1-10
1.4.3.1 Detail Specifications of Unit....................................................................................................... 1-12
2. MAINTENANCE ........................................................................2-1
2.1 Maintenance of Fan Filter in Power Supply Unit ................................................................................. 2-1
2.2 Setting...................................................................................................................................................... 2-2
2.2.1 Setting of NDC-2000 ........................................................................................................................ 2-2
2.2.1.1 Confirmation and Setting of IP Address..................................................................................... 2-2
2.2.2 Setting of “Sensor Set” ................................................................................................................... 2-7
2.2.2.1 Displaying the “Sensor Set” ........................................................................................................ 2-7
2.2.2.2 Setting Sensors............................................................................................................................. 2-9
2.2.2.3 Confirming Sensor...................................................................................................................... 2-14
2.2.2.4 Setting “Data Source” ................................................................................................................ 2-15
2.2.2.5 Setting of Priority........................................................................................................................ 2-16
2.2.2.6 Saving the Setting....................................................................................................................... 2-17
2.2.2.7 Setting of Alarm Relay Set ......................................................................................................... 2-19
2.2.2.8 Setting of Internal Port (CQD-2121) .......................................................................................... 2-20
2.2.2.9 Setting of Soft ware/Data path ................................................................................................... 2-21
2.2.2.10 Confirmation of “Monitor size” ............................................................................................... 2-23
2.2.2.11 Confirmation of Processing Unit setting ................................................................................ 2-24
2.2.3 Updating Software ......................................................................................................................... 2-25
2.2.3.1 Stop the ECDIS Software............................................................................................................ 2-25
2.2.3.2 Installing Software ...................................................................................................................... 2-27
2.3 Self-diagnostic Function...................................................................................................................... 2-32
2.4 Replacing Hard Disk............................................................................................................................. 2-35
2.5 Initializing HDD ..................................................................................................................................... 2-42
2.6 Confirming Disk Configuration ........................................................................................................... 2-44
2.6.1 Confirming Drive Assignment ...................................................................................................... 2-44
2.6.2 Restoring Drive Assignment......................................................................................................... 2-47
2.7 Setting of Virtual Memory ................................................................................................................. 2-53
2.7.1 Confirmation of pagefile.sys and Deletion ............................................................................... 2-53
2.8 Replacing Silicon Disk ......................................................................................................................... 2-59
2.9 Replacing CMH-2193 Add-in Serial Board ......................................................................................... 2-61
2.10 Replacing CBD-1818 Power Supply Unit ......................................................................................... 2-64
2.10.1 Removing Power Supply Unit..................................................................................................... 2-64
2.10.2 Installation of Power Supply Unit............................................................................................... 2-69
2.11 Replacing Backup Battery (CR-2032) ............................................................................................... 2-73
2.12 Replacing DVD Drive.......................................................................................................................... 2-78
2.13 Replacing Motherboard ..................................................................................................................... 2-85
2.14 Replacing Front Panel (CCK-969) ..................................................................................................... 2-91
2.15 Shape list of Screw............................................................................................................................. 2-93
2.16 System Recovery by “Recovery CD” ............................................................................................... 2-95
2.16.1 System Recovery Procedure ...................................................................................................... 2-95
2.16.2 Setting Resolution ....................................................................................................................... 2-98
DOC No. 7ZPNA4153
CONTENTS
4. APPENDICES ...........................................................................4-1
A) Exploded Diagram................................................................................................................................ 4-1
B) Parts List ............................................................................................................................................... 4-2
C) List of Replacement Parts ................................................................................................................... 4-4
D) About Parts Regularly Exchanged ..................................................................................................... 4-4
E) Operation Panel Key Code List........................................................................................................... 4-5
F) Software Directory Structure............................................................................................................... 4-6
G) Password List....................................................................................................................................... 4-7
1. Configuration
1.1 Components
A list of components is shown below.
1.1.1 JAN-2000
●JAN-2000 Basic Components
NAME TYPE Q’TY REMARKS
Processing Unit NDC-2000 1
LCD Monitor NWZ-173-EA/NWZ-147-AC 1 19-inch LCD/18.1-inch LCD
Operation Unit NCE-7721C 1
1.1.2 JAN-1186
●JAN-1186 Basic Components
NAME TYPE Q’TY REMARKS
Processing Unit NDC-2000 1
19” Monitor
LCD Monitor HP L1950 19” TFT/ 1 LCD Monitor is a commercial one. It will
be change in an equivalent.
17” Monitor
LCD Monitor HP L1750 17” TFT 1 LCD Monitor is a commercial one. It will
be change in an equivalent.
PC keyboard is a commercial one. It will
PC keyboard G84-4400LPBGB-2 1
be change in an equivalent.
1-1
1 Configuration
1.2.1 JAN-2000
1-2
1 Configuration
1.2.2 JAN-1186
HP ProCurve
HUB Switches 2312 With 1
MS630
1-3
1 Configuration
MASS: 10.6 kg
Unit: mm
NDC-2000 Processor Unit
1-4
1 Configuration
1-5
1 Configuration
1-6
1 Configuration
1-7
1 Configuration
1-8
1 Configuration
1.4 Specifications
1.4.1 Specifications of NDC-2000 Processor Unit
No. Function Description Remarks
1 System record Silicon Disk Drive -40℃~+85℃ Temperature
medium Capacity:1GB warranty
2 Data record Hard Disk Drive
medium Capacity:80GB
3 External media 1 DVD-ROM Drive
Reading medium:DVD-ROM/CD-ROM
4ch(4800bps/38400bps)
The RS422 input is converted into 4ch can be added in the option, and
4 NMEA Signal line RS232C and it inputs and outputs to it is possible to correspond up to
the processor unit. total 8ch.
・ Input/Output:4ch/3ch
SXGA(1280×1024) It is possible to output up to UXGA
5 Monitor output RGB Analog output by the setting. However, it depends
on the resolution of the monitor.
6 USB 4ch is set up in front I/O.
(open to user:2ch)
7 LAN 10/100M-Base 1ch The data receiving is possible from
1000M-Base 1ch OSV-DAU by UDP.
8 PS/2 2ch for mouse/keyboard
AGP slot The video outlet is not used for
9 Expansion slot 1 Expanding to the AGP slot in the onboard.
mechanism is improper.
PCI-X slot
10 Expansion slot 2 Expanding to the PCI-X slot in the
mechanism is improper.
Especially, there is no card of the
PCI slot 1 schedule for use.
11 Expansion slot 3 Only a full PCI card is scheduled to be When the customer uses this slot
able to be mounted. with the simplified chart display, the
support of JRC is off the subject.
12 Expansion slot 4 PCI slot 2 Mounting slot of PCI board for
Only a full PCI card can be mounted. serial expansion option.
13 Alarm System alarm:Output from operation CPU Fan failed , CPU Temp rising
panel unit and etc.
1-9
1
Keyboard PS/2
J1005 (Keyboard゙)
Operation Unit
(Trackball)
J1003 PS/2
(Mouse)
GPS
GYRO TB1002 J1008 COM1 A2
AIS (NMEA) SATA0
Configuration
CDD-717
K.B.
HDD
COM2
J1015 SATA1
Relay Signal (IN)
TB1003
Battery DC24V
TB3001-1 IDE1 CDD-716
SDD
AC Power Fail (OUT) TB3001-2 GPO
J1006 AUDIO
1.4.3 Composition Unit
1-10
USB4
e-Token
LAN1
LINE LOAD
LAN2
A B
C D
CMH-2193
A5
Add-in Serial Board PCI1
CB1 ATX CBD-1818
(Option) OUT AC Power Supply
Circuit Breaker Not use PCI2 POWER
NAF
J2002 VIDEO IN
Monitor Unit
(RGB)
J2001
A7 AC85V~AC264V
CCK-969 FPNL
Front Panel Circuit
接続先は使用機種により異なります。
※*The connection destination is different depending on the use model.
1 Configuration
System Configuration
NDC-2000 Component Unit
ITEM UNIT MODEL Q’TY REMARKS
Processor unit μATX motherboard CMC-1323 1 PFU product
(NDC-2000)
CPU Included in 1 Intel Pentium M1.6G
CMC-1323
Main Memory Included in 1 512MB
CMC-1323
Silicon disk drive CDD-716 1
1-11
1 Configuration
I/F Specifications
Item Content
Expansion bus slot PCI PCI-X 1slot, PCI 2 slots
AGP 1slot
I/O IDE ATA100×2 ports
S-ATA 2 ports
LAN 10Base-T/100Base-TX/1000Base-T 1 port
10Base-T/100Base-TX 1 port
Audio Line Out, Line In, MIC In, CD In, AUX In
USB 4 ports
Video output 1 port
LVDS 1 port
Serial 4 ports (D-sub×3,Pin-Header×1)
Parallel 1 port (Pin-Header×1)
Keyboard PS/2×1 port
Mouse PS/2×1 port
1-12
1 Configuration
1-13
1 Configuration
1-14
1 Configuration
Fuse
ACL
~ Filter Rectifier PFC
Smoothing
Inverter
ACN Circuit
FG
+5V SB
Auxiliary power
GND
Rectification -12V
Smoothing
Voltage
Detection Delay Circuit
PW R_OK
Circuit
Voltage Detection
Photo coupler
Err or Amplification
1-15
1 Configuration
1-16
1 Configuration
//Function//
1) AC Power supply input
2) DC12V/5V drive
3) I/O terminal stand for Serial 4ch
4) I/O terminal stand when Serial Expansion card is mounted
5) RS422/232Cconversion for 8ch
6) LED lighting for Serial I/O confirmation
7) Connector mounting for keyboard connection with JAN-701
8) Audio amplifier circuit
1-17
1 Configuration
Circuit
Breaker
PFAIL_N.C
PFAIL_COM
PU_PWALM Operation
AC Power Fail AC Power Fail Circuit N_ALM_PS_OPE Panel
PFAIL_N.O Unit
PFAIL_ACK+
AC Power Fail PFAIL_ACK-
ACK
CON1_N.C
CON1_COM Relay Contact RelayOut1(GPO18)
External CON1_N.O Output Circuit 1
Equipment MASK(GPO20)
CON2_N.C
CON2_COM Relay Contact
General-purpose RelayOut2(GPO19)
contact point output CON2_N.O Output Circuit 2
CON3+
CON3- Relay Contact
Output Circuit 3 RelayOut3(GPO21)
Motherboard
CON1_IN+
General-purpose CON1_IN(GPIO37)
CON1_IN- Contact Input Circuit 1
contact point Input RESET(GPO23)
1-18
1 Configuration
//Function//
Circuit that outputs three signals of PWR SW、RESET SW、HDD LED from mother board to processor
unit. It start/reset/confirms the operation with this circuit in the processor unit.
PWR_SW
Keyboard
PWR_SW
RESET
R
PWR_LED
R
HDD_LED
1-19
2 Maintenance
2. Maintenance
2.1 Maintenance of Fan Filter in Power Supply Unit
The fan filter installed on the rear side of NDC-2000 processing unit needs to be cleaned periodically.
Please confirm the POWER switch in the back of NDC-2000 is turned off
whenever fan filters are exchanged, and the rotation of the fan has
stopped.
POWER switch
1) Remove the cover by opening the edges of the ▲ sides of the cover to outside.
2) Slap the sponge filter inside the cover by hands to dean off the dust.
Cover
NDC-2000 unit
FAN Sponge filter
DOC No. 7ZPNA4153
2-1
2 Maintenance
2.2 Setting
2.2.1 Setting of NDC-2000
Note:
This setting is need for sharing the Route between ECDIS and NDC-2000
“85101
Password dialog
[Close] button
2-2
2 Maintenance
(2) Left-click the [Start] button on task bar and select the “Control Panel”.
2-3
2 Maintenance
(4) Right-click “Local Area Connection Intel(R) PRO/1000MT” and select “Properties”.
2-4
2 Maintenance
For JAN-2000
No1 unit
Item Main Sub
IP Address 192.168.60.59 192.168.61.59
Subnet Mask 255.255.255.0
Default gateway 192.168.60.255 192.168.61.225
No2 unit
Item Main Sub
IP Address 192.168.60.65 192.168.61.65
Subnet Mask 255.255.255.0
Default gateway 192.168.60.255 192.168.61.225
For JAN-1186
No1 unit
Item Main Sub
IP Address 192.168.60.156 192.168.61.156
Subnet Mask 255.255.255.0
Default gateway 192.168.60.255 192.168.61.225
No2 unit
Item Main Sub
IP Address 192.168.60.157 192.168.61.157
Subnet Mask 255.255.255.0
Default gateway 192.168.60.255 192.168.61.225
2-5
2 Maintenance
(7) No.1 unit don’t need to set up the IP Address, because factory setting is done.
In case of No.2 unit, need to set up the IP Address. Refer to (6) setting.
IP Address setting is complete, left-click [OK] button.
IP Address
Subnet Mask
OK button
2-6
2 Maintenance
“8510”
Password dialog
[Close] button
2-7
2 Maintenance
(2) The following dialog displays in the right-hand side of the “Startup Menu”. Left-click
[Configuration Set] button.
Sub menu
2-8
2 Maintenance
Sensor Name
2-9
2 Maintenance
Interface Type
2-10
2 Maintenance
Local
Note:
Please set “Not Use” for sensor that isn’t used in ECDIS/RPS.
2-11
2
Table 1 JAN-2000 Setting
Cable Connecti Connection Sensor Use/ Interface Local Baud Data Parity Stop Buffer Verify Time
Connected from on Port Name Not Use Type Port Rate Length Bits Size Checksum Out
Equipment Equipment Terminal
No.1 DGPS TX+ TB1002 COM1_RX GPS1 Use Full 1 4800 8 None 1 1024 ON 15
+
"No.1 ECDIS").
TX- COM1_RX-
Maintenance
NSK Unit 1 (Red) COM2_TX GYRO I/F Use ― 2 9600 8 Even 2 1024 OFF 6
+
3 (Yellow) COM2_RX-
No.2 DGPS TX+ COM3_RX GPS2 Use Full 3 4800 8 None 1 1024 ON 15
+
TX- COM3_RX-
Sounder TX+ COM5_RX Depth Use ― 5 4800 8 None 1 1024 ON 15
+
TX- COM5_RX-
AIS TX+ COM6_RX AIS Use ― 6 38400 8 None 1 4096 ― 90
+
2-12
TX- COM6_RX-
TT1 TX+ COM7_RX TT1(NMEA) Use ― 7 4800 8 None 1 1024 ON 15
(NMEA) +
TX- COM7_RX-
TT2 TX+ COM8_RX TT2(NMEA) Use ― 8 4800 8 None 1 1024 ON 15
(NMEA) +
TX- COM8_RX-
Operation P1003 J1003 ― KEYBOAR Use NCE-7721 4 4800 8 None 1 1024 ― 0
Panel Unit D
TX- COM5_RX-
The followings are a set table to receive the sensor on ECDIS. Set the equipment where SRB is
Table 1 JAN-1186 Setting
Cable Connectio Connection Sensor Use/ Interface Local Baud Data Parity Stop Buffer Verify Time
Connected from n Port Name Not Use Type Port Rate Length Bits Size Checksum Out
Equipment Equipment Terminal
No.1 DGPS TX+ TB1002 COM1_RX+ GPS1 Use Full 1 4800 8 None 1 1024 ON 15
TX- COM1_RX-
Digitizer TX+ COM2_RX+ YEOMAN Use ― 2 4800 8 None 1 1024 OFF 0
TX- COM2_RX-
Printer LAN - LAN Printer Use ― ― ― ― ― ― ― ― ―
(1000Base)
2-13
2
Maintenance
(1) When the "Communication" [Recv Test] button of the [Sensor Set] dialog is left-clicked, the Line
Monitor dialog is displayed. Receiving data in the terminal board set to select the input
destination of the sensor data by "Select Port", and to left-click the [Start] button can be
confirmed.
Receiving data
2-14
2 Maintenance
Set it when
“VDR(JRC)" is in
"Use"
Set it when
“AIS(NMEA)-TX" is in "Use".
2-15
2 Maintenance
Note:
・Set all Primary to None when you do not receive the data from SRB.
For JAN-2000
Select all None for JAN-2000.
Note:
・A usual setting is None. Set the device name of JAN-701B/901B to 1st Primary only
when it is connected with JAN-701B/901B by LAN.
・ Set [Optional Sensor] only when SRB2 is mounted on JAN-701B/901B.
・ Note that the signal cannot be received when the setting of Priority is mistaken.
2-16
2 Maintenance
[Set] button
[Export] button
2-17
2 Maintenance
(3) When [Save As] dialog starts up, left-click the [OK] button.
(File name is ”SensorData○○○○●●◎◎△△▲▲□□.ini”.)
Note:
・ When saving changed setting, be sure to click a [Set] button.
2-18
2 Maintenance
2-19
2 Maintenance
(1) Set the alarm from among “Alarm Name" of each terminal name.
(2) After setting the alarm, left-click the [OK] button or the [Apply] button under the right of the
[AlarmRelay Out Setting] dialog.
[OK] button :Save the setting and close [Alarm Relay Out Setting] dialog.
[Apply] button :Save the setting.
2-20
2 Maintenance
(1) Move the mouse cursor on “Startup Menu”, and right-click to display “Password” dialog, right-click
once again, and software keyboard displays.
(2) Enter password [8510] by using the software keyboard.
“8510”
Password display
Start up display
Software keyboard
2-21
2 Maintenance
2-22
2 Maintenance
(1) Confirm "NWZ-173(19inch)" has been selected with Display Type on the Configuration screen, click
[OK], and shut the screen.
2-23
2 Maintenance
Note:
・ When check off the “GPIO”, the output of relay contact don’t function.
Don’t change the “GPIO setting”.
・ When check off “EmbedWare setting”, HardWare Alarm don’t function.
Don’t change the “EmbedWare setting”.
2-24
2 Maintenance
[Exit] button
2-25
2 Maintenance
[OK] button
2-26
2 Maintenance
(1) Move the mouse cursor onto “Startup Menu”, and then right-click to display “Password”
dialog.
(2) Right-click again on the “Password” dialog to display software keyboard.
(3) Enter password [85100] by using the software keyboard, and left-click the [Ent] button.
Software keyboard
(4) Confirm the number of "Maintenance No." because “Software Information" is displayed.
1) Confirm the
“Maintenance No.”.
2) Left-click [Close]
button.
(5) Left-click [Close] button to end “System Information”, and “Startup Menu” displays.
(6) Insert “Update CD” into the DVD drive.
2-27
2 Maintenance
DVD drive
(7) Move the mouse cursor on “Startup Menu”, and right-click to display “Password” dialog.
(8) On “Password” dialog, right-click once again, and software keyboard displays.
(9) Enter password [8510] by using the software keyboard, and then left-click [ENT] button.
(10) The new dialog displays to the right of the “Startup Menu”. Left-click [Software Update]
button.
2-28
2 Maintenance
(11) Confirm that “Software Update” path displays in the upper dialog box, and then left-click [OK]
button. (When the path does not display, confirm that the “Update CD” is inserted in the DVD
Drive.)
(12) Confirm the software version to be updated, and then left-click [Yes] button.
(13) Place a check in the checkbox of the “Single HDD”, and then set up as following.
2-29
2 Maintenance
(15) Select “Display type” by referring to the table below (“Display Type Reference Table”), and
left-click [OK] button.
2-30
2 Maintenance
(18) Move the mouse cursor on “Startup Menu”, and right-click to display “Password” dialog.
(19) On “Password” dialog, right-click once again, and software keyboard displays.
(20) Enter password [85100] by using the software keyboard, and left-click [Ent] button.
(21) After “System Information” displays, confirm that “Maintenance No.” is same as the version
of step 12). Upon confirming, left-click [Close] button.
2-31
2 Maintenance
“0000”
Password display
[Close] button
2-32
2 Maintenance
(3) The system file is automatically collected. Then, the text file will be displayed after the self
diagnostic is completed. Save by selecting the saving destination.
NOTE: Do not close the window while self-diagnosing.
Self-diagnosing
Text file
(4) Select [File] from Text file menu and left-click [Save as…].
2-33
2 Maintenance
(6) Close the window of the text file after the self-diagnosis ends.
The screen of the self- diagnosis file disappears, and a start-up screen is displayed.
Startup screen
2-34
2 Maintenance
Before replacing Hard Disk, make sure to turn off the breaker of the rear
power switch. Otherwise, an electric shock or a malfunction may occur.
The hard disk used in this system has a limited service life, so that they should better be replaced
periodically. When the power supply is on constantly, nominal life is two years. In case of abnormal
condition or outliving, replace a hard disk with the following procedure.
※ After replacement, it takes several minutes for starting.
※ The shape of screw is replaced, is indicated by (Type*). When miss it’s shape, refer to “2.15
Shape list of screw”.
POWER switch
(2) Remove the top cover and the front cover of NDC-2000.
Top cover
2-35
2 Maintenance
Disconnect the following cables before pulling out the base fixing the hard disk.
When the disk is forcibly pulled, the cables and the connectors may be
damaged.
S-ATA cable
S-ATA power cable
S-ATA cable
S-ATA power cable
2-36
2 Maintenance
(3) Remove all cables that are connected with I/F terminal board (CQD-2174).
LCD POWER CABLE
AC IN
J1011(Breaker)
J1012(Breaker)
SENSOR DATA
J1006(To Audio)
(4) After remove the cable that is connected with I/F terminal board (CQD-2174), unscrew the 5
points(Type3) and removing the insert base.
Insert base
2-37
2 Maintenance
HD SATA cable
(6) Pull out the data cable (SATA) on the front cover side under the motherboard.
Unscrew (Type2)
2-38
2 Maintenance
(8) Slide the base fixing the hard disk to the front side by 3 cm and remove the connector of silicon
disk and hard disk power cable, and then remove the base.
Hard disk
2-39
2 Maintenance
(9) Remove the four screws fixing the hard disk from the rear side, and replace the hard disk with
new one. After replacing, fix the hard disk with the four screws by the reverse procedure of
removing, and insert the base into the initial position.
Unscrew (Type1)
There are two type of Hard disk that one is fixed only data cable and another is fixed data cable
and power cable.
In case of the hard disk that is fixed data cable and power cable, it is necessary to insert the
spacer between hard disk and base. When fixing the hard disk on the base, insert the spacer
attached between hard disk and base, and fix by the attached screw.
When fixing the data cable only fixed, it isn’t need to insert the spacer.
2-40
2 Maintenance
(10) When the new hard disk is not initialized, initialize it by referring to 2.5 “Initializing HDD”.
(11) After replacement of HDD、refer to”2.6 Confirming Disk Configuration”, and confirm the disk
configuration.
(12) Connect the cable through under motherboard. The cable is fixed with motherboard.
(13) Connect the relay connector of silicon disk and hard disk power cable.
(14) Fix the plate by two screws.
(15) Close the top cover by the reverse procedure of removing, fixing I/F terminal board and connecting
the cable.
2-41
2 Maintenance
Password display
[Close] button
(2) Left-click the [Hard Disk Initialization] button from startup display.
2-42
2 Maintenance
2-43
2 Maintenance
2-44
2 Maintenance
“5963”
Password display
[Close] button
Windows explorer
2-45
2 Maintenance
Drive type
Drive letter
2-46
2 Maintenance
When mismatch
Normal assignment
2-47
2 Maintenance
Restoration method
(1) After the “Startup” menu is displayed, move the cursor onto it and right-click to display the
“PASSWORD” dialog box. Right-click to start the Software keyboard. Input the password
“5963”and left-click [OK] button. Left-click [Close] button to close software keyboard.
“5963”
Password display
[Close] button
Windows explorer
2-48
2 Maintenance
Manage
Disk Management
2-49
2 Maintenance
(6) Right-click “E:HDD” and left-click “Change Drive Letter and Paths…”.
Right-click E: drive
[Change…] button
2-50
2 Maintenance
Select D: drive
(10) After confirming D drive has been selected, click [OK] button.
[OK] button
2-51
2 Maintenance
[Yes] button
(12) Confirm the drive was changed from E drive to D drive. Then, click Exit from the File menu to
end.
Select Exit
2-52
2 Maintenance
pagefile.sys
(2) When you confirm pagefile.sys, click the System icon from Control Panel.
System icon
2-53
2 Maintenance
(3) Select “Advanced” TAB and left-click [Setting] button of “Performance” menu.
Advanced tab
[Settings] button
(4) Select “Advanced” TAB and left-click [Change] button of “Virtual memory”.
Advanced tab
[Change] button
2-54
2 Maintenance
Select C:
(6) Select “No paging file “ with radio button of “Paging file size for selected drive”.
No paging file
2-55
2 Maintenance
[Set] button
[OK] button
(8) When the following message windows are displayed, left-click [OK] button.
[OK] button
2-56
2 Maintenance
[OK] button
[OK] button
(11) When the confirmation of restart message is displayed, left-click [Yes] button.
[Yes] button
2-57
2 Maintenance
(12) After restarting the unit, the system configuration is changed. Confirm that pagefile.sys
disappeared.
(13) Right-click C: drive and open property. Confirm that the disk space is over 100MB.
2-58
2 Maintenance
Before replacing Silicon Disk, make sure to turn off the breaker of the rear
power switch. Otherwise, an electric shock or a malfunction may occur.
POWER switch
(2) Refer to “2.4 Replacing Hard Disk” and remove the base with procedure from (1) to (4).
(3) Remove the four screws of silicon disk that is fixed on base.
Unscrew (Type7)
2-59
2 Maintenance
(4) Confirm the jumper setting of the new silicon disk is set for Master.
MASTER
(5) Fix the silicon disk with the four screws by the reverse procedure of removing, and connect the
cable. Then insert the base into the initial position and fix the base with the two screws.
(6) Refer “6.16 Restoration by Recovery Boot CD” and recover the system from image CD.
2-60
2 Maintenance
Before replacing Add-in Serial Board, make sure to turn off the breaker of the
rear power switch. Otherwise, an electric shock or a malfunction may
occur.
POWER switch
2-61
2 Maintenance
(2) Remove the top cover and the front cover of NDC-2000.
Top cover
(4) Remove the attached cable connected with J1009 connector of I/F terminal board (CQD-2174)
and the add-in serial board connector.
2-62
2 Maintenance
(5) Remove the screws fixing the add-in serial board, and remove the add-in serial board.
Unscrew (Type2)
PCI slot
2-63
2 Maintenance
Before starting the exchange of the CBD-1818 power supply unit, turn off the
breaker in the main power supply line connected with NDC-2000 without fail.
Otherwise, an electric shock or a malfunction may occur.
2-64
2 Maintenance
Unscrew 4 points
Top cover
2-65
2 Maintenance
(4) Remove all the cables connected with the terminal board (CQD-2174).
LCD POWER CABLE
AC IN
J1011(Breaker)
J1012(Breaker)
SENSOR DATA
J1006(To Audio)
(5) Remove five screws on the terminal board and remove the base with I/F terminal board
CQD-2174.
Base
2-66
2 Maintenance
(6) Remove DVD power cable, motherboard power cable and HDD power cable.
(7) Remove the six screws on both sides and the nine screws on the bottom side of NDC-2000 to
remove the backside board.
Unscrew right and left 6 points (Type4)
2-67
2 Maintenance
(8) Cut the cable tie and remove the AC cable connecting to the power supply unit, and the fan filter
and two screws on the rear side of NDC-2000.
Remove 2 screws(Type5).
Cable tie
Fan filter
Cover AC cable
Caution:The fan of power supply unit can't replace only a fan because the fan unites with
power supply unit. Please replace the power supply unit with FAN.
2-68
2 Maintenance
(1) Fix the cover with 2 screws and fit the FAN filter. Fix 2 screws (Type5)
Fan filter
Cover
(2) Insert the AC power supply cable under the L angle plate.
Cable tie
2-69
2 Maintenance
(4) Fix the rear plate with 6 screws and fix the base with 9 screws.
(5) Connect DVD power cable, motherboard power cable and HDD power cable as following.
Motherboard power cable
2-70
2 Maintenance
(6) When I/F terminal board fix on base, pay attention as follow points.
Caution:
・ Install the connector carefully and don't touch a CPU fan.
・ Install it not to interpose the cable or the flat cable for the I/F terminal cable.
Flat cable
(7) Fix the base of terminal I/F board circuit (CQD-2174) with the screw in five points.
Base
2-71
2 Maintenance
(8) When you fix the base with the screw, connect all removed cables with terminal I/F board circuit
(CQD-2174).
AC IN
J1011(Breaker)
J1012(Breaker)
SENSOR DATA
J1006(To Audio)
2-72
2 Maintenance
Before starting the exchange of the backup batteries, turn off the breaker in
the main power supply line connected with NDC-2000 without fail.
Otherwise, an electric shock or a malfunction may occur.
When removing the connector and cable, make sure that doesn't touch a CPU
fan. The cable will touch and CPU fan may be damaged.
The coin battery is used to maintain information on the setting of BIOS, the date, and time in this
device. In the battery longevity, five years are standards. When the setting at BIOS and time is not
correctly saved after the power supply of this device is turned on, it is necessary to change a battery.
※The screw removed when exchanging it has been described shape to "(Type*)". Refer to “2.15
Shape list of screw" when you do not understand the screw at the installation.
2-73
2 Maintenance
(1) Turn off the main power breaker that is connected with NDC-2000.
Top cover
2-74
2 Maintenance
(3) Remove all the cables connected with the terminal board (CQD-2174).
LCD POWER CABLE
AC IN
J1011(Breaker)
J1012(Breaker)
SENSOR DATA
J1006(To Audio)
(4) Remove five screws on the terminal board and remove the I/F terminal board CQD-2174 with
the base.
Base
2-75
2 Maintenance
(5) Replace the battery in the battery holder on the motherboard CMC-1323 with a new one.
Motherboard CMC-1323
(6) When I/F terminal board fix on base, pay attention as follow points.
Caution:
・ Install the connector carefully and don't touch a CPU fan.
・ Install it not to interpose the cable or the flat cable for the I/F terminal cable.
Flat cable
2-76
2 Maintenance
(7) Install I/F terminal board (CQD-2174) on the base and fix 5 points with screw.
Base
(8) When you fix the base with the screw, connect all removed cables with terminal I/F board circuit
(CQD-2174).
LCD POWER CABLE
AC IN
J1011(Breaker)
J1012(Breaker)
SENSOR DATA
J1006(To Audio)
J1004(To CCK-969)
2-77
2 Maintenance
Before starting the exchange of DVD drives, turn off the breaker in the
main power supply line connected with NDC-2000 without fail.
Otherwise, an electric shock or a malfunction may occur.
DVD drive
(1) Turn off the main power breaker that is connected with NDC-2000.
2-78
2 Maintenance
(2) Remove the top cover and the front cover of NDC-2000.
Top cover
(3) Remove all the cables connected to the terminal board (CQD-2174).
AC IN
J1011(Breaker)
J1012(Breaker)
SENSOR DATA
J1006(To Audio)
2-79
2 Maintenance
(4) Remove five screws on the terminal board and remove the I/F terminal board CQD-2174 with
the base.
Base
2-80
2 Maintenance
(7) Remove 4 screws from front panel and 2 screws from left side. Take off DVD drive.
2-81
2 Maintenance
(9) Unscrew 4 points of right and left. Remove DVD drive from the base.
PCB
Unscrew (Type6)
※When replace DVD drive, remove the PCB from rear side.
(10) After replacing, install the DVD drive by the reverse procedures.
※Pay attention the direction of flat cable.
2-82
2 Maintenance
(11) After replace the DVD drive, pay attention as follow points.
Caution:
・ Install the connector carefully and don't touch a CPU fan.
・ Install it not to interpose the cable or the flat cable for the I/F terminal cable.
Flat cable
(12) Install I/F terminal board (CQD-2174) on the base and fix 5 points with screw.
Base
2-83
2 Maintenance
(13) When you fix the base with the screw, connect all removed cables with terminal I/F board
circuit (CQD-2174).
AC IN
J1011(Breaker)
J1012(Breaker)
SENSOR DATA
J1006(To Audio)
2-84
2 Maintenance
Before starting the exchange of the mother board, turn off the breaker
in the main power supply line connected with NDC-2000 without fail.
Otherwise, an electric shock or a malfunction may occur.
(1) Turn off the main power breaker that is connected with NDC-2000.
(2) Remove the top cover and the front cover of NDC-2000.
Top cover
2-85
2 Maintenance
(3) Remove all the cables connected to the terminal board (CQD-2174).
AC IN
J1011(Breaker)
J1012(Breaker)
SENSOR DATA
J1006(To Audio)
J1004(To CCK-969)
(4) Remove five screws on the terminal board and remove the I/F terminal board CQD-2174 with
the base.
Base
2-86
2 Maintenance
(5) Remove DVD power cable, motherboard power cable and HDD power cable.
Motherboard power
(6) Refer to “2.12 Replace DVD drive” and remove DVD drive unit.
(7) Remove the mouse cable, COM cable and LAN cable from front panel.
COM cable
LAN cable
RGB cable
Mouse cable
(8) Refer to “2.9 Replacing CMH-2193 Add-in serial board” and remove serial board.
(9) Remove the 7 cables are connected with motherboard.
2-87
2 Maintenance
N.C
IDE2
IDE1
Front
COM4 W1003
HDD W1006
FRONT W1001
N.C
COM4 HDD
W1003 W1006
N.C
FRONT
W1001
※ Exchange each mother boards because memory and CPU have combined with the mother board.
2-88
2 Maintenance
(12) After replacing the mother board, connect the base of DVD drive unit and terminal I/F board
circuit (CQD-2174) by the opposite procedure noting the following point.
Caution:
・ Install the connector carefully and don't touch a CPU fan.
・ Install it not to interpose the cable or the flat cable for the I/F terminal cable.
Flat cable
(13) Install I/F terminal board (CQD-2174) on the base and fix 5 points with screw.
Base
2-89
2 Maintenance
(14) When you fix the base with the screw, connect all removed cables with terminal I/F board
circuit (CQD-2174).
AC IN
J1011(Breaker)
J1012(Breaker)
SENSOR DATA
J1006(To Audio)
2-90
2 Maintenance
Before starting the exchange of front panels, turn off the POWER
switch in the back of NDC-2000. Otherwise, an electric shock or a
malfunction may occur.
The screw removed when exchanging it has been described shape to "(Type*)". Refer to “2.15
Shape list of screw" when you do not understand the screw at the installation.
(1) Turn off the main power breaker that is connected with NDC-2000.
POWER switch
(2) Remove the top cover and the front cover of NDC-2000.
Top cover
2-91
2 Maintenance
(3) Remove the cable and the screw, and exchange front panels.
Front switch
Remove cable
Unscrew (Type3)
Front of processing unit (NDC-2000)
Refer to the following figure for pulling of the cable and install the front panel.
2-92
2 Maintenance
There are 9 kinds of screw in all removed when the unit of NDC-2000 is exchanged.
Refer to the following table and install a screw based on the mentioned shape of " (Type*)"
by the explanation of each unit.
2-93
2 Maintenance
2-94
2 Maintenance
Before restoring system by the [Recovery Boot CD], read the description to
fully understand the function of the [Recovery Boot CD].
When you perform restoration with abnormal disk configuration, the
equipment may malfunction.
The system file including the NDC-2000 simplified chart software is saved in the ATA flash disk.
NDC-2000 will not start when the system file in the flash disk is damaged. In this case, the damaged
system file may be restored and started by the following procedure “System Image Deployment by
Recovery Boot CD” describing how to restore the system file into the factory-set status
Caution:
• The ATA flash disk cannot be restored when it is physically damaged.
• When the hard disk is broken, the equipment cannot start normally even after system
restoration.
• When the software has been version upped or the individual sensor setting has been changed
after delivery, those changes will be erased by restoration.
2-95
2 Maintenance
---------------------------------------------------------
Disk information
---------------------------------------------------------
Disk:1 (977.48MB) M:[1024MB ATA Flash Disk] S:[A151186248F001523525]
1.1 (977.45MB) Active Volume NTFS Primary
Disk:2 (76.69GB) M:[HDS728080PLA380] S:[PFDB32S6TU3W4M]
2.1 (76.69GB) Volume NTFS Primary
( 2.46MB) Unused Space Primary
---------------------------------------------------------
.
.
---------------------------------------------------------
NDC-2000 Recovery Boot CD
ver.1.0
---------------------------------------------------------
Restore Image file from CD to System Drive(Disk:1)
.
1) Remove Boot CD.
2) Insert CD-R "NDC-2000 Backup Image" printed.
3) Wait for a few seconds.
4) Left-click to continue.....
NOTE:
• The ATA flash disk and the hard disk may subject to change in the future. In this case,
the model name of each disk will differ from the one written above.
• When Disk1 and 2 are inversely displayed, keep pressing the POWER switch of
NDC-2000 to turn off the power. Then turn the power on again, and after the JRC logo
appears press the [Ctrl] + [C] keys of the keyboard. After confirming the Disk
configuration is restored, perform the following procedure.
• When either or neither of Disk1/2 is displayed, the breakdown of the drive etc. are
thought. Do not execute the following operation in that case until the breakdown etc. are
solved.
• Though it is written in the label of Recovery Boot CD as S-Chart, there is no problem,
and use it as it is, please.
2-96
2 Maintenance
(4) Press the eject button of the DVD drive to eject the startup disk “Recovery Boot CD “.
(5) Insert the “Recovery Image CD” into the DVD drive.
(6) Wait until the LED stops flickering, and left-click the mouse or trackball.
(7) The “GHOST” will start and the system will automatically be copied from the “Recovery Image
CD” to the ATA flash disk.
_______________________________________________________________________________
(9) Turn on the power again and confirm the system starts normally.
System restoration is now completed.
NOTE: After the system is restored by using the image file, restore the allocation of the drive
referring to "2.6.2 Restoring Drive Assignment" when the allocation of the drive is
mismatched (E:HDD drive,Q:DVD drive).
2-97
2 Maintenance
(1) Left-click [Start] button and select [Control Panel] from [Start menu].
Control Panel
Display
Settings
2-98
2 Maintenance
(4) Select [Screen resolution] and [Color quality] from following table. Left-click [OK] button.
Screen resolution
Color quality
2-99
3 Trouble shooting
3. Trouble Shooting
When the unit is not working properly, check the following points and carry out the appropriate repair
before you ask for repair.
When the following remedies do not rectify the problem, or a location not mentioned in the following
tables is not in order, contact the nearest JRC office or agent, and ask for repair.
Nothing is displayed on The power switch is not turning Turn on the power switch.
screen. on.
The video cable is not Connect it correctly referring to
connected. Installation Manual.
The unit does not start after The IDE cable of the silicon Connect the cable correctly,
pressing the power switch on disk or the power cable is not and start again.
the front side of the connected.
processing unit.
The screen display is Error in initial setting. In the case of NWZ-173-E,
deviated. keep pressing the brilliance
control to adjust the display
automatically.
3-1
3 Trouble shooting
Depth value is not displayed. The echo sounder is not turned Turn on the echo sounder.
on.
The cable is not connected to Check the cable connection
the processing unit from echo between echo sounder and
sounder. processing unit.
The main unit has Make a contact to JRC.
malfunctioned.
ARPA is not displayed. The [ARPA] key is not pressed. Press the [ARPA] key.
3-2
3 Trouble shooting
The brightness does not The lighting control circuit does Make a contact to JRC.
change even if the not work.
[BRILLIANCE] knob is turned. The cable is damaged. Replace the cable.
When the [ ] (Navigation & Hard disk failure Replace the hard disk by
Planning) button is referring to “2.6 Confirming
left-clicked in the Startup Disk Configuration”.
menu, the message “Path is
incorrect or Master Drive is
failed” is displayed.
3-3
4 Appendices
4. Appendices
A) Exploded Diagram
4-1
4 Appendices
B) Parts List
4-2
4 Appendices
4-3
4 Appendices
Please consult shop that purchases it, agency or our each branch, the office, and the liaison office about
the exchange of consumables parts.
4-4
4 Appendices
4-5
4 Appendices
D:
¥ECDIS_DATA¥:Chart information, alarm information, and log information are stored.
¥dynamic_ini¥
¥AlmList¥
¥CM93¥
¥CM93_3¥
¥Index¥
¥ARCS¥
¥S57¥
¥Log¥
¥Alm¥
¥Own¥
¥Rot¥
¥Tgt¥
¥Nvtx¥
¥Route¥
¥Tmp¥
¥Ucht¥
¥Screen_Shot¥
¥User¥
¥Vdr¥
¥TGT_Track¥
¥OWN TRACK¥
4-6
4 Appendices
G) Password List
4-7
JAN-701B/901B
ECDIS
取扱説明書
URL http://www.jrc.co.jp
Marine Service Department
Telephone : +81-3-3492-1305
Facsimile : +81-3-3779-1420
e-mail : tmsc@jrc.co.jp
AMSTERDAM Branch
Telephone : +31-20-658-0750
Facsimile : +31-20-658-0755
e-mail : service@jrcams.nl
SEATTLE Branch
Telephone : +1-206-654-5644
Facsimile : +1-206-654-7030
e-mail : service@jrcamerica.com
CODE No.7ZPNA4153 01ETM ISO 9001, ISO 14001 Certified
AUG. 2009 Edition 1 JRC Printed in Japan