Professional Documents
Culture Documents
Chapter: 3
Bio-electric Amplifiers
1. Bio-electric Amplifier
Bio-electric Amplifiers (or Bio-potential Amplifiers) also called Bio-Amplifiers are specifically
designed for processing of Bio-electric signals as they are low in amplitude.
Bio-electric signals are generally low in amplitude. Amplifiers that are specifically designed for
processing this type of Bio-potentials are known as Bio-potential Amplifiers. The outputs from
such amplifiers are used for analysis and they show up as ECG, EMG or other Bio-electric
waveforms. These amplifiers typically process voltages but in some cases they process current.
The amplifier provides high impedance, high CMRR and thereby minimizes loading effects. This
is the vital functionality of Bio-potential Amplifiers. For Biomedical applications, Bio-amplifiers
must meet the below mentioned requirements in order to work incessantly.
To measure Bio-potential, electrodes are placed on Human skin as shown in the Fig. 1.1. The
signals from the Electrodes pass on to the Amplifier stage. Amplifier helps in minimizing,
eliminating most of the signals interfering with the measurement of Bio-potentials and final
readout is obtained.
1
Biomedical Electronics- Bio-electric Amplifiers
2
Biomedical Electronics- Bio-electric Amplifiers
Isolation Amplifier
Optical Isolator
Chopper Amplifier
Differential Amplifier
Operational Amplifier
Instrumentation Amplifier
3
Biomedical Electronics- Bio-electric Amplifiers
In the Isolation Amplifier stage, galvanic decoupling of the patient from the measuring
equipment is served. It prevents galvanic currents from deteriorating Signal to Noise Ratio
(SNR) and provides safety to the patient from electrical hazards. Transformer, Optical or
Capacitive Couplers are used in Analog Isolation Amplifiers, to transmit signal through the
isolation barrier. On the other hand, Digital Isolation Amplifiers use Voltage and frequency
converter to digitize the signal before it is transmitted.
Recording of the Bio-potentials in the last stage is done with electrical systems which produce
strong electrical and magnetic fields. Hence the system is capacitively coupled and the current
flows to the ground electrode.
4
Biomedical Electronics- Bio-electric Amplifiers
2. Carrier Amplifier
Carrier amplifier is a direct-current amplifier in which the dc input signal is filtered by a low-
pass filter, then used to modulate a carrier so it can be amplified conventionally as an
alternating-current signal; the amplified dc output is obtained by rectifying and filtering the
rectified carrier signal.
3. Isolation Amplifier
These amplifiers are also known as Pre-amplifier isolation circuits (Symbol shown in Fig. 3.1). It
provides electrical isolation for the safety of the patient. It prevents accidental shocks and
increases the input impedance of the patient’s monitoring system.
5
Biomedical Electronics- Bio-electric Amplifiers
Isolation Barrier
Demodulator
Output amplifier
4. Optical Isolator
An opto-isolator (also called an opto-coupler, photo-coupler, or optical isolator) is
an electronic component that transfers electrical signals between two isolated circuits by using
light. Opto-isolators prevent high voltages from affecting the system receiving the
signal. Commercially available opto-isolators withstand input-to-output voltages up to 10 kV and
voltage transients with speeds up to 25 kV/μs.
A common type of opto-isolator consists of an LED and a phototransistor in the same opaque
package. Other types of source-sensor combinations include LED-photodiode, LED-LASCR,
6
Biomedical Electronics- Bio-electric Amplifiers
and lamp-photo-resistor pairs. Usually opto-isolators transfer digital (on-off) signals, but some
techniques allow them to be used with analog signals.
Fig. 4 – Schematic diagram of an opto-isolator showing source of light (LED) on the left,
dielectric barrier in the center, and sensor (phototransistor) on the right.
5. Chopper Amplifier
Noise and drift are the two major issues encountered when recording Bio-potentials. Noise is
caused by the movement of the patient or due to the recording device. A DC Amplifier hits a
sudden peak in the output when the input is zero. A Chopper Amplifier samples the problem of
drift in DC amplifiers.
The chopper amplifier helps to remove the noise and dc drift in high-gain amplifiers which are
used for very weak bio-potentials.
7
Biomedical Electronics- Bio-electric Amplifiers
6. Differential Amplifier
These are used to amplify the difference between the voltages applied to its inputs.
8
Biomedical Electronics- Bio-electric Amplifiers
Amplifiers built using either FET’s (Field Effect Transistors) or BJT’s (Bipolar Junction
Transistors).
7. Operational Amplifier
These are multistage amplifiers which are interconnected and occupy minimal space even though
it consists of many Transistors, Resistors, and FET’s. They are available in the form of an
Integrated Circuit (IC).
8. Instrumentation Amplifier
An Instrumentation amplifier (Ins-amp) is basically a differential amplifier, but with high input
impedance so that it can faithfully reproduce voltage signals from any source. The idea is that
there should not be significant voltage drop in the output impedance of the source, almost all the
source voltage should be dropped across the input impedance of the amplifier. A differential
amplifier is a basic circuit to pick up and amplify a small signal in presence of a relatively large
9
Biomedical Electronics- Bio-electric Amplifiers
background noise. However, it is only possible if the noise is the same at two electrical points,
which carry the signal, with respect to a third point. That is the noise should be common to both
the input terminals. To clarify, suppose a small signal, of the order of a few mV, appears
between two electrical terminals. Now if an external noise with respect to a third terminal
appears on both the above two terminals with equal magnitude and phase then it is possible to
eliminate the noise. Instrumentation amplifiers are used in industry as well as in bioelectrical
measurements.
It consists of 3-amplifiers in the circuit. The input to the amplifier is the output from the
Transducer. A non-inverting amplifier is connected to each of the input of the Differential
Amplifier. Non-inverting amplifiers are the ones on the left side of the diagram. The non-
inverting amplifiers together form the input of the instrumentation amplifier.
The third op-amp is called the difference amplifier and is the output of the instrumentation
amplifier. The difference between the two input signals forms the output V out. V1 and V2 are the
inputs to Op-amp 1 and Op-amp 2 respectively. A typical application of these amplifiers includes
biomedical applications such as Bio-potential Amplifier. High gain and high impedance are
attained using these amplifiers which are crucial in medical instruments to determine the health
condition of an individual.
10
Biomedical Electronics- Bio-electric Amplifiers
Here the output is proportional to the voltage difference between the two inputs. The amplifier
has two stages. The 1st stage has a differential input and also has a differential output (double
ended input and double ended output) while the 2nd stage has a differential input and a single
ended output.
+Vcc
U1:A
8
3
V1
1
2
4
R1 10k
10k
-Vcc R4 -Vcc
10k U2:A
4
R2 BAT1
2.2k 2 Vout 6V
1
3
8
R5 TL072
R3 GND
10k 10k BAT2
6V
+Vcc
U1:B
R7
4
4.7k
6
7
-Vcc (-12 V)
5
V2
8
TL072 RV1
10K
11
Biomedical Electronics- Bio-electric Amplifiers
A1 1 2 R1 R2 (2)
In figure 1, R1 =R3=10K and R2=2.2K
The 2nd stage has a differential input and single ended output given by,
A2 R6 R4 (4)
It is assumed that the common-mode gain is zero and R6 R4 ( R7 RV 1) R5
Ad A1 A2 (1 2 R1 R2 )( R6 R4 ) (5)
If the common-mode gain, Acm is not zero then,
V1 V2
Vcm (7)
2
8.4. Introducing Non-zero Common Mode Gain
The common mode input voltage V cm is given by the average of the two input voltages V1 and
V2, i.e., as given in Eq. 7,
V1 V2
Vcm
2
If the common mode voltage gain Acm is not zero then we will have a contribution from V cm as
well, as given in Eq.6. In Eq.6, if the differential input voltage is zero, i.e., if V2 – V1 = 0, then
12
Biomedical Electronics- Bio-electric Amplifiers
Problem-3.8.1: Find the overall gain of an Instrumentation Amplifier (Fig. 8.2) if the following
resistor values are used: R1 = 10 kΩ, R2 = 500 Ω, R4 = 10 kΩ, and R6 = 100 kΩ.
Solution: According to Fig. 8.2 we get that, the 1st stage gain,
A1 1 2 R1 R2
nd
And the 2 stage gain,
A2 R6 R4
Now we get that the overall gain of the Instrumentation Amplifier is:
Ad A1 A2 (1 2 R1 R2 )( R6 R4 )
Problem-3.8.2: Find the overall gain of an Instrumentation Amplifier (Fig. 8.1) if the following
resistor values are used: R1 = 15 kΩ, R2 = 1 kΩ, R3 = 20 kΩ, and R4 = 200 kΩ.
Solution: According to Fig. 8.1 we get that, the 1st stage gain,
A1 1 2 R1 R2
nd
And the 2 stage gain,
A2 = R4/R3
Now we get that the overall gain of the Instrumentation Amplifier is:
Ad A1 A2 (1 2 R1 R2 )( R4 R3 )
13