\m] MS9-5s
(MS9-3s) Automatic Full
Digital Cell Counter
INSTALLATION MANUAL

Indication for Use :


The MS9-5s® (MS9-3s®) is an automatic hematology analyzer used to quantify formed elements
(erythrocytes, leukocytes, thrombocytes) in blood and perform a three-part (MS9-3s®) or a five-
part (MS9-5s®) leukocyte differential cell count (3 PART : Lymphocytes, Monocytes,
Granulocytes / 5 PART : Lymphocytes, Monocytes, Neutrophiles, Eosinophiles, Basophiles).

Installation Manual
2011-09

d MELET SCHLOESING Laboratoires - 9 Chaussee Jules Cesar - Evolic 402 - 95520 OSNY FRANCE
- Tel. : ( + 33) 01 30 75 30 00 - FAX : ( + 33) 01 30 73 17 80 I
S
LABORATOIRES
Page 2 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

IMPORTANT
A SEE PAGE 54 THE LICENSE AGREEMENT
BEFORE INSTALLING OR USING THE
INSTRUMENT.

A
IMPORTANT

SEE PAGE 27 FOR GUARANTEE AND


WARRANTY RESTRICTIONS DEPENDING
ON THE INSTALLATION OF THE
INSTRUMENT.

Issued on: 2009-05-07


Last revision on: 2011-09-19

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

CONTENT Page

UNPACKING ................................................................................................................. 5

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE


Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 4 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

Transport box......................................................................................................... 6

Unpacking the MS9-5s® or MS9-3s® .................................................................... 8

The instrument ....................................................................................................... 10

EXTERNAL VIEW ........................................................................................................... 13

Front view ............................................................................................................... 14

Right View.............................................................................................................. 16

Left view ................................................................................................................ 18

Back view .............................................................................................................. 20

The connection panel ............................................................................................. 22

INSTALLATION.............................................................................................................. 25

Installation advice .................................................................................................. 26

Cables ................................................................................................................... 28

Internal preparation ................................................................................................. 30

Reagent tubings ...................................................................................................... 32

Reagent tubings : Installation ............................................................................ 34

Reagent tubings : MS9-5s(MS9-3s) Batch adaptation ....................................... 36

MS-Card® .............................................................................................................. 38

MS-Pack® ............................................................................................................. 40

MS9-5s® (MS9-3s®) Batch : Installation ................................................................ 48

Power supply ......................................................................................................... 50

MAIN FEATURES ......................................................................................................... 52

MS9-5s®................................................................................................................ 52

MS9-3s®................................................................................................................ 53

LICENSE AGREEMENT ................................................................................................ 54

UNPACKING
LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

TRANSPORT BOX

The transport box is made to prevent any kind


of shock.

O Two fast opening manual locks on the front


side give easy access to the inside of the
box.

© Unlock the two latches.

© Open the lid to access to the MS9-5s or MS9-


3s and accessories.

Notes :

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE


Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 6

Accessories

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

UNPACKING THE MS9-5s


(MS9-3s)
O Take out the plastic pocket with the manuals
which can be found on the right side of the
instrument.

© To extract the MS9-5s or MS9-3s from the


transport box, tilt it backwards holding the
bottom at the right of the front part and the
top left side of the back part, as shown on
the picture.

© Take out the keyboard and accessories box.

Store the transport box in a safe place for further


use, once all the elements have been taken out.

Notes :

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 8 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

THE INSTRUMENT

The instrument is packed in the box with two knife at this stage.
plastic pockets and one accessories box which © 1 plastic pocket which contains
includes CD-ROMs, reagent tubings, a power - 1 blank CD-RW.
cord, blood tube adaptors and two keys.

Take off with the hands the protection plastic


film which covers the instrument. Never use a Notes

Check the accessories as follows:

O A mouse,

© A compact alphanumeric keyboard in a


plastic pocket,

© A mouse pad,

© A plastic pocket which contains :


- 6 (5 for MS9-3s) reagent
tubings,
- 2 (1 for MS9-3s) caps in a small
plastic pocket,

© A power cord,

© A plastic pocket which contains :


- 1 Installation Manual,
- 1 Quick Tour (or User's) Manual,

© A plastic pocket which contains :


- 4 blood clot filters,
- 4 whole blood tube adaptors,
- 2 keys for the opening and closing
of the MS9-5s (MS9-3s) front
cover,

© 1 CD-R containing the electronic


manuals of the MS9-5/-5s (MS9-3/-3s),

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 10 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

©©
!
EXTERNAL VIEW

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 1 12 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

FRONT VIEW

A front description of the MS9-5s (MS9-3s)


shows:

O A large polychrome back-lighted SVGA


LCD screen,

© A plastic front cover,

© A whole blood tube holder which is also the


analysis starting switch place,

© A compact alphanumeric keyboard.

Notes :

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.:


+33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.:


+33(0)1 30 73 17 80
Page : 1 14 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

RIGHT VIEW

A right profile description of the MS9-5s (MS9-


3s) shows:

O An upper cover,

© A space for the whole blood sampling


needle,

© A whole blood tube holder which is also the


analysis starting switch place,

© A DVD-R/RW drive,

© A lock for the opening and closing of the


upper cover (key provided with the
instrument).

Notes :

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE


Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 16 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

LEFT VIEW

A left profile description of the MS9-5s (MS9-


3s) shows:

O An upper cover,

© A multi connection panel,

© 6 (5 for MS9-3s) plastic Luer lock fittings


with caps on for the reagents tubings
connection (reagent inlets),

© 1 exhaust tubing for air transfer.

Notes :

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.:


+33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 18 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

BACK VIEW

A back side description of the MS9-5s (MS9-


3s) shows:

O Hinges for the upper front cover,

© Internal fixation holes,

© Electric main power supply plug with ON/OFF


switch and fuse holder,

© An air flow filter.

Notes :

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.:


+33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 20 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

CONNECTION PANEL

The multi connection panel description shows:

O External VGA video monitor connector,

© PS/2 connector for mouse,

© Do not use:

Use only for technical purpose. Notes :

© 3 USB 2.0 connectors,

© Parallel printer connector LPT1, The keyboard must not be plugged on the
connection panel, but on a specific plug
© The CK1-E acquisition board, located under the front cover of the MS9-5
© The MS-Card reader on FK1-S fluidic board, (MS9-3) (See 'Cables' section).

© DB9 serial connector: COM1,

© LAN RJ45 connector (10/100 Mb/s


Ethernet),

© DB9 serial connector: COM2.

ONLY CONNECT PERIPHERAL INSTRUMENTS WHICH ARE COMPLIANT WITH REF.:


CEI 950, CEI 1010-1.

LABORATOIRES -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

o ©
©
©

o ©
e ©
©
©

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 22 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

! BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

INSTALLATION

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30


00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires

INSTALLATION ADVICE

Environment
Choose a suitable place to install the MS9-5s (MS9-3s):
- Put it on a flat and solid surface to ensure a correct ventilation of the MS9-5s (MS9-3s).
- Plan a location on the left for the reagents pack (MS-Pack® or Batch®).
Eo-Diff is only for MS9-5s.
- Plan at least 20 cm of free space on the right for the opening of the DVD-R/RW recorder.
- Select an environment free of dust.
- Favor a working environment at nearly 20-25°C. Do not install the instrument and the
reagents beside a window exposed to the sun or near a radiator for example.

DVD-R/RW

Wast
e
Normalized
earthed line

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.:


+33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires voltage
MS9-5s® MS9-3s® plug
Installation Manual / 2011-09 Page : 1 1

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30


00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires

Main power supply

Plan an earthed line voltage connection. A good earthing is required for a correct use of the
MS9-5s or MS9-3s.

A disturbed line voltage or a low quality earthing


will give bad analysis results.

A
IMPORTANT
It is VERY DANGEROUS for the MS9-5s or MS9-3s electronic
parts, the analysis results and the safety of the operator to connect
the MS9-5s or MS9-3s Analyzer on a main power supply which
does not have any ground (Earth). The Phases, Neutral and Ground
(3rd connector on the plug) need to be really supplied from a
normalized plug certified by a professional installer. Not only for the
security of the operator, the need and quality of the ground value
gives also the MS9-5s or MS9-3s analyzers the zero voltage
reference calibration used for hematology analysis. No Guarantee
and/or Warranty could be applied in case of a MS9- 5s or MS9-3s
not properly connected to a normalized ground and a safe electric
installation.

Main power supply:


110-220 V / 50-60 Hz 300 VA
Automatic switching voltage

Fuses
F 2 A-250 V for 200-240
V F 4 A-125 V for 100-

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING
120 V Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 27

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 28 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

Notes
CABLES
The first stage of the MS9-5s (MS9-3s)
installation is to connect cables as follows:

0,0,© Connection of the alphanumeric


keyboard under the front panel of the
instrument with the PS/2 female - DIN
male adaptor.
Respect the right DIN plug position when
connecting.

© Connection of the mouse.

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex /


FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30
73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 29

_____________________________________________________________________________________________________________________ DS
LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 30 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

INTERNAL PREPARATION

Some internal preparation must be done


before taking any further steps:

O Put one of the keys from the accessories box in


the lock on the right side of the MS9-5s (MS9-
3s).

© Open the upper front cover by turning the key


counterclockwise.

© Unscrew the plastic caps of the reagent fittings


(on the bottom left front part of the MS9-5s or
MS9-3s) in a counterclockwise direction and
take them off.

Store the caps in a safe place for a further


use.

© Take off the 4 transportation scotch tape pieces


(arrows on the picture) inside the instrument.
Notes :

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex /


FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30
73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 31

©,©

____________________________________________________________________________________________tf
LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 32 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

REAGENT TUBINGS

The reagent tubings are used to link the


reagent pack to the MS9-5s (MS9-3s). They
are provided in a plastic pocket (©) which
contains :

© 1 kit of 6 (5 for MS9-3s) tubings equipped


with caps and flexible tubings which allow
the MS-Pack® to be plugged to the MS9.

© A small plastic pocket labeled "Batch 1000"


with:
MS9-5s: Two additional caps (1 orange
© and 1 blue ©) with flexible tubings to
adapt the reagent tubings to a MS9-5s
Batch.
MS9-3s: One additional cap (1 blue ©)
with flexible tubing to adapt the reagent
tubing to a MS9-3s Batch.

Each cap has a flexible tubing whose length


fits the corresponding reagent bottle.

The use of a MS-Pack® with wrong reagent tubings could generate fluidic errors.

LABORATOIRES -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 33

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex /


FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73
17 80
Page : 34 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

LABORATOIRES

BP 508 - OSNY
- 95528
CERGY-
PONTOISE
Cedex /
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 35

REAGENT
TUBINGS
Installation
The reagent tubings must be connected as
follows:

O Make sure that the protective caps of the


Luer locks of the MS9-5s (MS9-3s) have
been removed (take them off if
necessary).

© Install by screwing in a clockwise direction


each tubing on the Luer locks according
to its corresponding color.

© Check the reagent tubings color


positions from the front side to the back
side as follows:

0Black Waste
© Orange Lyse Eo-Diff® (MS9-5s)
(D Yellow Lyse Acti-Diff®
© White Isoflux®
© Red Hemoref®
® Blue Transflux®
® Green Air transfer (no tubing)

Notes :

- Lyse Eo-Diff is only for MS9-5s.


- Each cap is equipped with a flexible
tubing that will soak into the reagent.
The flexible tubings must therefore be
laid on a clean surface as long as they
are not in their respective bottles in

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex /


FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73
17 80
Page : 36 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

order to avoid any contamination.

LABORATOIRES

BP 508 - OSNY
- 95528
CERGY-
PONTOISE
Cedex /
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 38 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

REAGENT TUBINGS MS9-


5s(-3s) Batch adaptation
The reagent tubings kit allows the connection MS-Pack®,
of a reagent pack to the MS9-5s (MS9-3s). - big transparent caps 4 MS9-5s or
There are two reagents packagings which MS9-3s Batch.
have different size: the MS-Pack® and the Notes :
MS9-5s Batch (MS9-3s Batch).
Each cap is equipped with a flexible
The Transflux® cap (blue) and the Lyse Eo-
tubing that will soak into the reagent.
Diff® cap (orange)(MS9-5s) must be adapted
The flexible tubings must therefore be
according to the packaging. Their size
laid on a clean surface as long as they
indicates to which type they correspond:
are not in their respective bottles in
- small caps 4 order to avoid any contamination.

When replacing a MS-Pack® with a MS9-5s


Batch, proceed as follows:
(For MS9-3s Batch, skip steps 1 and 2)

O Unscrew the Lyse Eo-Diff® cap from the


Lyse Eo-Diff® tubing (orange).

© Extract the big transparent cap with an


orange ring from the pocket labeled
'Batch 1000' (in the tubing kit pocket) and
screw it on the Lyse Eo-Diff® tubing.

© Repeat stages 1 to 2 with the Transflux® cap


(blue).

© With a dry piece of non-fluffy cloth clean the


flexible tubings of both removed caps,
then put them in the pocket labeled
'Batch 1000'.
Store the pocket in a safe place.

© Check that the caps match the color of the


reagent tubings (see 'Reagent tubings',
'Installation', ©).

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 39

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 40 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

MS-CARD®

The reagent pack or batch is controlled by a


smart card called 'MS-Card®'.

O Open the MS-Card® carton pocket placed


on the side of the MS-Pack®.

© The plastic MS-Card® holder appears.

© Take the plastic MS-Card® holder out and


close the MS-Card® carton pocket.

© Insert the MS-Card® in its place in the


middle of the connection panel.

The MS-Card® must be inserted


following the direction of the arrow drawn
on the card itself and with the printed side
on top as shown on the picture.

Notes :

The MS-Card® of the MS9-5s Batch or


MS9-3s Batch (1000 cycles) is located in
the carton pocket of the Lyse Acti- Diff®
reagent (See 'MS9-5s(-3s) Batch :
Installation section).

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 41

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 42 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

MS-PACK®

The reagent pack installation is described as


follows:

® Check that the diameters of the caps


connected to the tubings correspond to
the reagent bottles that you have (default
packaging: MS-Pack®):
© Blue Transflux®
© Red Hemoref®
© White Isoflux®
© Yellow Lyse Acti-Diff®
© Orange Lyse Eo-Diff® (MS9-5s)
© Black Waste
(See 'Reagent tubings : MS9-5s(-3s)
Batch adaptation' section if necessary).

O Open the white carton precut area on top of


the MS-Pack®, corresponding to the
diluent (Isoflux®).

© The transparent cap of the Isoflux® bottle


appears, sealed with a ring.

© Remove the ring.

.../...

Notes :
- Lyse Eo-Diff® is only for MS9-5s
- To install a MS9-5s Batch or a MS9- 3s
Batch, see the 'MS9-5s(-3s) Batch :
Installation' section.

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 43

IE1
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires

MS-PACK®
© Pull one side of the cap of the Isoflux® bottle,
then pull it up entirely.

Do not open the reagent bottle before pulling it


up, otherwise there will be an
important leak of diluent.

© Unscrew the cap of the bottle and take it off.

© Take the cap of the Isoflux® reagent tubing


(white fittings) and clean its flexible tubing
with a non-fluffy piece of cloth and
distilled water.

© Insert the flexible tubing of the Isoflux®


tubing (white) in the bottle and screw the
cap.

.../...

Notes :

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.:


+33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 45

HUM
MMWI HWMMkf

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 46 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

MS-PACK®
© Plug the Lyse Acti-Diff® reagent (©: © Check that the colors of the reagent tubings
yellow carton precut area and cap with fittings correspond to the reagent bottles:
HUM: human lysing reagent, or white Notes :
with yellow ring with VET: vet lysing - During the opening of the Lyse Acti-
reagent), then the Transflux® (©: blue Diff®, Transflux® and Hemoref® bottles,
carton precut area and cap), the a colored ring stays on each bottle and
Hemoref® (©: red carton precut area allows the identification of the reagents.
and cap), then with the Lyse Eo-Diff® It is recommended to leave it in place.
(MS9-5s) (©: 1 separate bottle to be The caps of the Lyse Acti-Diff® and the
connected to the orange tubing fitting). Lyse Eo-Diff® reagents can be white.
They are identifiable with the color of
Plug the reagents ONE BY ONE to their rings.
avoid any confusion. - Lyse Eo-Diff® is only for MS9-5s
© Take the left tubing (black fitting and very
short flexible tubing) on a waste
container.

Orange : Lyse Eo-Diff® (MS9-5s)


Yellow : Lyse Acti-Diff®
White : Isoflux®
Red : Hemoref®
Blue : Transflux®

.../...

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE


Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 47

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 48 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

MS-PACK®

© When the reagent pack installation is


complete, put the pack on the left of the
MS9-5s (MS9-3s) at the same level, and
the waste container at a lower level than
the MS9-5s (MS9-3s).

A
To have the instrument work properly, the
reagents must be at the same level as the
MS9-5s (MS9-3s).

Notes :

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE


Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73
17 80
Page : 50 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

MS9-5s(-3s)
The MS9-5s(-3s) Batch installation is
BATCH Installation
Notes : (the instructions are written on the
described as follows: bottle), then normally resume the
O MS-Card® : reagents installation (see 'MS-
Open the Lyse Acti-Diff® carton box: the Pack®1 section).
MS-Card® protective holder is in the If by lack of space it would not be
closing flap. possible to have all the elements of the
Batch on the same level,
As with the MS-Pack® take the card out - It is possible to have some bottles
and insert it in the MS-Card® reader placed at a lower level:
located in the middle of the connection - Isoflux®,
panel (See 'MS-Card®' section). - Transflux®,
- Hemoref®.
© Tubing kit adaptation: - A long priming sequence must be
If necessary refer to the 'Reagent tubings programmed in this case at the
: Batch adaptation' section to adapt the MS9-5s (MS9-3s) startup.
tubing kit to the MS9-5s (MS9-3s) Batch - It introduces risks of higher
as shown on the picture. consumption.
- Please contact our technical
© Installation : service for further information.
Dilute the Hemoref® reagent of the Batch

To have the instrument work properly the Batch


reagents should be at the same level as the MS9-
5s (MS9-3s) (®>.

A
t?
LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73
17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 51

©
*^*<

U II

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30


00 Fax.: +33(0)1 30 73 17 80
Page : 52 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

tf
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

POWER SUPPLY

Connection of the power supply cord:


First check the quality of the current on the wall
plug:
Phase...Neutral : 220 V or 110 V +/- 10%
NeutraL.Terre : 1 V +/- 1 V Phase...Terre :
= Phase...Neutral

If these values are not reached it is fundamental and strictly necessary to request a better main
power installation by a professional. Earthing the MS9- 5s or the MS9-3s is a basic need for
safety, g o o d r e s u l t s a n d M a n u f a c t u r e r guarantee/warranty application.

If an external printer is used with the MS9-5s


(MS9-3s), both devices (MS9-5s(-3s) & printer)
must be connected on the same multi-plug.

Once the MS9-5s or the MS9-3s is plugged


refer to the MS9-5s User's guide or MS9-3s
User's guide.

Notes :

LABORATOIRES -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00


Fax.: +33(0)1 30 73 17 80
Page : 54 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
Page : 55 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires

MAIN FEATURES MS9-5s®


33 parameters with 5-part-differential:
WBC; Lym (%,#); Mon (%,#); Neu (%,#); Eo (%,#); Ba (%,#); IC (%;#); AL (%,#); RBC; Hb;
MCV. Hct; MCH. MCHC. RDW-CV. RDW-SD. iRBC%; Macro RBC%; THR; MPV; Pct; PDW;
Mode; Median; iTHR%; Macro THR%
3 histograms and 1 scattergram:
Histograms: RBC, PLT, Histogram Eo.
Scattergram: for WBC.
Digital pre-programmed banks:
Male, Female, Child, New-Born.

Performances Linearity Precision113 Carry-over


GB 0 - 120 103/pl < 2% < 1%
GR 0 - 8 106/|jl 2% 1,5%
PLT 0 - 1000(2)103/jl < 5% < 2%
Hgb 0 - 24 g/dl < 2% < 1%
VGM < 2%
Hct 0 - 70 %
(1)
Precision are given for the normal range of each parameter.
(2)
Up to 1500 10>l with P.R.P.

Technical
Hybrid stepper motors sequential d i lutor with m icro-control ler management
Specific 60 (im chamber for RBC and THR
Specific 80 (im chamber for WBC
Specific 80 im chamber for eosinophils
Built-in PC Computer:
- 1.6 GHz INTEL ATOM N270 or compatible processor
- Hard-disk storage up to 50 000 tests or more*
- Alphanumeric PC mini-keyboard
- Built-in LCD TFT graphic color high resolution display (SVGA-800x600)
- Touch screen
- Fast Ethernet lan adapter 10/100 Base TX
- RS232 ports mono and/or bi-directional transmission
- 1 parallel port, 3 USB ports 2.0
compliant Analyses back-up and user-
manual on CD-ROM
Windows XP® Multitask application
Quality control: Fully in compliance with international standards the MS9-5s
quality control Program include Jenning’s graph

Dimension : Width 340 mm, Height 420 mm, Depth 370 mm Weight: 20kg
Power supply: 100-240Volt / 50-50 Hz 300VA (automatic commutation)

Reagents: Diluent:ISOFLUX;Detergent:TRANSFLUX;Lysingreagent:ACTI-DIFF;HEMOREF;
Specific lysing reagent EO-DIFF Eosinophils
MAIN FEATURES MS9-3s®
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 56 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoi res

25 parameters with 3-part-differential:


WBC; Lym (%,#); Mon (%,#); Gr (%,#); RBC; Hb; MCV; Hct; MCH; MCHC; RDW-CV;
RDW-SD; iRBC%; MacroRBC%; THR; MPV; Pct; PDW; Mode; Median; iTHR%; MacroTHR% 3
histograms: WBC, RBC, PLT.
Digital pre-programmed banks:
Male, Female, Child, New-Born.

Performances Linearity Precision01 Carry-over


GB 0 - 100 1 03/pl < 2% < 1%
GR 0 - 7,5 106/|jl < 2% < 1,5%
PLT 0 - 1000(2)103/jl < 5% < 2%
Hgb 0 - 24 g/dl < 2% < 1%
VGM 2%
Hct
(1)
0 - 70 %
Precision are given for the normal range of each parameter.
(2)
Up to 1500 1 03/^l with P.R.P.

Technical
Hybrid stepper motors sequential dilutor with micro-controller
management Specific 60 im chamber for RBC and THR Specific 80 (im
chamber for WBC Built-in PC Computer:
- VIA 733 MHz or compatible processor
- Hard-disk storage up to 50 000 tests or more*
- Alphanumeric PC mini-keyboard
- Built-in LCD TFT graphic color high resolution display (VGA-640x480)
- Touch screen
- Fast Ethernet lan adapter 10/100 Base TX
- RS232 ports mono and/or bi-directional transmission
- 1 parallel port, 1 USB ports 1.1 compliant Analyses back-up
and user-manual on CD-ROM
Windows 98SE® application
Quality control: Fully in compliance with international standards the MS9-3s
quality control Program include Jenning’s graph

Dimension : Width 340 mm, Height 420 mm, Depth 370 mm Weight: 20kg
Power supply: 100-240Volt / 50-50 Hz 300VA (automatic commutation)

Reagents:Diluent:ISOFLUX;Detergent:TRANSFLUX;Lysingreagent:ACTI-DIFF;HEMOREF;
MELET SCHLOESING Laboratoires ’SOFTWARE &
MS-Card’ LICENSE AGREEMENT
IMPORTANT - READ BEFORE INSTALLING OR USING
This document is a LEGAL AGREEMENT between you and MELET SCHLOESING Laboratoires. Do not use or load any of your new
in vitro diagnostic MSL Instrument's Software and/or any associated materials as MS-Card (the "reagent card") until you have carefully
read the following terms and conditions. By loading and/or using the Software and/or using the MS-Card used with your new MSL
instrument, you agree to be bound to the terms of this License Agreement. If you do not wish to so agree, DO NOT install or use the
Software or any other associated materials as MS-Card.

MANUFACTURER. The "Manufacturer" word used herein includes, (i) companies having originally created your analyser's Software and/ or any associated
materials as the MS-Card (electronic card which supervises the reagent quality and includes owner's codes), (ii) companies licensed to manage any rights and/or

LABORATOIRES

BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1

copyrights linked to any Intellectual property used in the machine and/or in any associated material as the MS-Card, (iii) individual who has any rights on any of the
Intellectual property used in the machine and/or in any associated material as the MS-Card,

MSL INSTRUMENTS. The "MSL Instruments" words used herein include Instrument(s), (i) manufactured by MELET SCHLOESING Laboratoires - France and/or
its affiliates and/or subsidiaries and distributed directly or indirectly (by mean of local Distributors), (ii) manufactured by MELET SCHLOESING Laboratoires - France
and/or its affiliates and/or subsidiaries for other companies under an OEM agreement.

LICENSE. You may copy and/or install any Manufacturer's Software onto the single built-in computer (or computer board) of your new MSL instrument for your
personal, non-commercial use, and you may NOT make any back-up copy of the Software. You may use the MS-Card on your MSL instrument to run it and profit
of its reagent quality management BUT you may NOT make any back-up and/ or copy and/or duplicate and/or compatible material of this card.

OWNERSHIP OF SOFTWARE AND COPYRIGHTS. Title to all copies of the Software remains with the Manufacturer or its suppliers. The Software is copyrighted
and protected by the European laws and other countries, and international treaty provisions. You may not remove any copyright notices from the Software. The
Manufacturer may make changes to the Software, and/or to items referenced therein as the MS-Card, at any time without notice, but is not obligated to support or
update the Software. Except as otherwise expressly provided, the Manufacturer grants no express or implied right under its patents, copyrights, trademarks, or
other intellectual property rights. You may transfer the Software and/or any associated materials as MS-Card only if the recipient agreesto be fully bound
by these
terms and if you retain no copies of the Software.

LIMITED MEDIA WARRANTY. If the Software has been delivered directly by the Manufacturer (or one of its Distributors) on physical media, the Manufacturer
warrants the media to be free from material physical defects for a period of ninety (90) days after delivery by the Manufacturer (or one of its Distributors). If such a
defect is found, return the media to the Manufacturer or your local Distributor for replacement or alternate delivery of the Software as the Manufacturer may select.
The limited media warranty is the same for the MS-Card if the MS-Card has not been already in use.

EXCLUSION OF OTHER WARRANTIES EXCEPT AS PROVIDED ABOVE, THE SOFTWARE IS PROVIDED "AS IS" WITHOUT ANY EXPRESS OR IMPLIED
WARRANTY OF ANY KIND INCLUDING WARRANTIES OF MERCHANTABILITY, NONINFRINGEMENT, OR FITNESS FOR A PARTICULAR PURPOSE. The
Manufacturer does not warrant or assume responsibility for the accuracy or completeness of any information, text, graphics, links or other items contained within
the Software and/or the MS-Card.

MANUFACTURER LIMITATION OF LIABILITY. IN NO EVENT SHALL THE MANUFACTURER OR ITS SUPPLIERS BELIABLE FOR
ANY DAMAGES WHATSOEVER (INCLUDING, WITHOUT LIMITATION, LOST PROFITS, BUSINESS INTERRUPTION, OR LOST INFORMATION) ARISING
OUT OF THE USE OF OR INABILITY TO USE THE SOFTWARE, EVEN IF THE MANUFACTURER OR ITS SUPPLIERS HAVE BEEN ADVISED OF THE
POSSIBILITY OF SUCH DAMAGES. SOME JURISDICTIONS PROHIBIT EXCLUSION OR LIMITATION OF LIABILITY FOR IMPLIED WARRANTIES OR
CONSEQUENTIAL OR INCIDENTAL DAMAGES, SO THE ABOVE LIMITATION MAY NOT APPLY TO YOU. YOU MAY ALSO HAVE OTHER LEGAL RIGHTS
THAT VARY FROM JURISDICTION TO JURISDICTION.

USER LIABILITY. IN THE EVENT YOU DO NOT RESPECT DIRECTLY OR INDIRECTLY AND/OR, INFRINGES PARTIALLY OR ENTIRELY THE
MANUFACTURER'S INDUSTRIAL PROPERTY -AND/OR ANY OTHER COMPANY'S INDUSTRIAL PROPERTY USED BY ANY OF MSL INSTRUMENTS-
INCLUDING SOFTWARE COPYRIGHT, PATENTS, TRADE MARKS, LOGO, MS-CARDS TECHNOLOGY OR OTHERS, OR, HELP OR PARTICIPATES WITH A
THIRD PARTY NOT TO RESPECT THE SAME MANUFACTURER'S INDUSTRIAL PROPERTIES -AND/OR ANY OTHER COMPANY'S INDUSTRIAL PROPERTY
USED BY ANY OF MSL INSTRUMENTS- AS ABOVE MENTIONED, INCLUDING DISASSEMBLING THE SOFTWARE OF ANY OF MSL INSTRUMENTS AND/OR
ANY KIND OF OTHER REVERSE-ENGINEERING ACTION, YOU ARE ENTIRELY LIABLE TO THE MANUFACTURER FOR ANY SPECIAL DAMAGES,
INCLUDING LOST OF PROFITS, LOST SAVINGS OR OTHER INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND.ANY FAILURE TO RESPECT
THE MANUFACTURER'S INDUSTRIAL PROPERTY SHALL BE DEEMED AS A CONTRACT VIOLATION AND A MATERIAL BREACH OF THIS AGREEMENT.

APPLICABLE LAWS. Claims arising under this Agreement shall be governed by the laws of your country and/or countries of Europe, excluding its principles of