\m] MS9-5s
(MS9-3s) Automatic Full
Digital Cell Counter
INSTALLATION MANUAL
Installation Manual
2011-09
d MELET SCHLOESING Laboratoires - 9 Chaussee Jules Cesar - Evolic 402 - 95520 OSNY FRANCE
- Tel. : ( + 33) 01 30 75 30 00 - FAX : ( + 33) 01 30 73 17 80 I
S
LABORATOIRES
Page 2 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
IMPORTANT
A SEE PAGE 54 THE LICENSE AGREEMENT
BEFORE INSTALLING OR USING THE
INSTRUMENT.
A
IMPORTANT
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1
CONTENT Page
UNPACKING ................................................................................................................. 5
LABORATOIRES
Transport box......................................................................................................... 6
Right View.............................................................................................................. 16
INSTALLATION.............................................................................................................. 25
Cables ................................................................................................................... 28
MS-Card® .............................................................................................................. 38
MS-Pack® ............................................................................................................. 40
MS9-5s®................................................................................................................ 52
MS9-3s®................................................................................................................ 53
UNPACKING
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1
TRANSPORT BOX
Notes :
LABORATOIRES
Accessories
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1
Notes :
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 8 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1
THE INSTRUMENT
The instrument is packed in the box with two knife at this stage.
plastic pockets and one accessories box which © 1 plastic pocket which contains
includes CD-ROMs, reagent tubings, a power - 1 blank CD-RW.
cord, blood tube adaptors and two keys.
O A mouse,
© A mouse pad,
© A power cord,
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 10 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1
©©
!
EXTERNAL VIEW
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 1 12 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
FRONT VIEW
Notes :
LABORATOIRES
LABORATOIRES
RIGHT VIEW
O An upper cover,
© A DVD-R/RW drive,
Notes :
LABORATOIRES
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 16 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
LEFT VIEW
O An upper cover,
Notes :
LABORATOIRES
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 18 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
BACK VIEW
Notes :
LABORATOIRES
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 20 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
CONNECTION PANEL
© Do not use:
© Parallel printer connector LPT1, The keyboard must not be plugged on the
connection panel, but on a specific plug
© The CK1-E acquisition board, located under the front cover of the MS9-5
© The MS-Card reader on FK1-S fluidic board, (MS9-3) (See 'Cables' section).
LABORATOIRES -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1
o ©
©
©
o ©
e ©
©
©
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 22 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
! BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1
INSTALLATION
LABORATOIRES
INSTALLATION ADVICE
Environment
Choose a suitable place to install the MS9-5s (MS9-3s):
- Put it on a flat and solid surface to ensure a correct ventilation of the MS9-5s (MS9-3s).
- Plan a location on the left for the reagents pack (MS-Pack® or Batch®).
Eo-Diff is only for MS9-5s.
- Plan at least 20 cm of free space on the right for the opening of the DVD-R/RW recorder.
- Select an environment free of dust.
- Favor a working environment at nearly 20-25°C. Do not install the instrument and the
reagents beside a window exposed to the sun or near a radiator for example.
DVD-R/RW
Wast
e
Normalized
earthed line
LABORATOIRES
LABORATOIRES
Plan an earthed line voltage connection. A good earthing is required for a correct use of the
MS9-5s or MS9-3s.
A
IMPORTANT
It is VERY DANGEROUS for the MS9-5s or MS9-3s electronic
parts, the analysis results and the safety of the operator to connect
the MS9-5s or MS9-3s Analyzer on a main power supply which
does not have any ground (Earth). The Phases, Neutral and Ground
(3rd connector on the plug) need to be really supplied from a
normalized plug certified by a professional installer. Not only for the
security of the operator, the need and quality of the ground value
gives also the MS9-5s or MS9-3s analyzers the zero voltage
reference calibration used for hematology analysis. No Guarantee
and/or Warranty could be applied in case of a MS9- 5s or MS9-3s
not properly connected to a normalized ground and a safe electric
installation.
Fuses
F 2 A-250 V for 200-240
V F 4 A-125 V for 100-
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING
120 V Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 27
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 28 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
Notes
CABLES
The first stage of the MS9-5s (MS9-3s)
installation is to connect cables as follows:
LABORATOIRES
_____________________________________________________________________________________________________________________ DS
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 30 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
INTERNAL PREPARATION
LABORATOIRES
©,©
____________________________________________________________________________________________tf
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 32 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
REAGENT TUBINGS
The use of a MS-Pack® with wrong reagent tubings could generate fluidic errors.
LABORATOIRES -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 33
LABORATOIRES
BP 508 - OSNY
- 95528
CERGY-
PONTOISE
Cedex /
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 35
REAGENT
TUBINGS
Installation
The reagent tubings must be connected as
follows:
0Black Waste
© Orange Lyse Eo-Diff® (MS9-5s)
(D Yellow Lyse Acti-Diff®
© White Isoflux®
© Red Hemoref®
® Blue Transflux®
® Green Air transfer (no tubing)
Notes :
LABORATOIRES
BP 508 - OSNY
- 95528
CERGY-
PONTOISE
Cedex /
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 38 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 39
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 40 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
MS-CARD®
Notes :
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 41
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 42 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
MS-PACK®
.../...
Notes :
- Lyse Eo-Diff® is only for MS9-5s
- To install a MS9-5s Batch or a MS9- 3s
Batch, see the 'MS9-5s(-3s) Batch :
Installation' section.
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 43
IE1
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires
MS-PACK®
© Pull one side of the cap of the Isoflux® bottle,
then pull it up entirely.
.../...
Notes :
LABORATOIRES
HUM
MMWI HWMMkf
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 46 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
MS-PACK®
© Plug the Lyse Acti-Diff® reagent (©: © Check that the colors of the reagent tubings
yellow carton precut area and cap with fittings correspond to the reagent bottles:
HUM: human lysing reagent, or white Notes :
with yellow ring with VET: vet lysing - During the opening of the Lyse Acti-
reagent), then the Transflux® (©: blue Diff®, Transflux® and Hemoref® bottles,
carton precut area and cap), the a colored ring stays on each bottle and
Hemoref® (©: red carton precut area allows the identification of the reagents.
and cap), then with the Lyse Eo-Diff® It is recommended to leave it in place.
(MS9-5s) (©: 1 separate bottle to be The caps of the Lyse Acti-Diff® and the
connected to the orange tubing fitting). Lyse Eo-Diff® reagents can be white.
They are identifiable with the color of
Plug the reagents ONE BY ONE to their rings.
avoid any confusion. - Lyse Eo-Diff® is only for MS9-5s
© Take the left tubing (black fitting and very
short flexible tubing) on a waste
container.
.../...
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 48 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
MS-PACK®
A
To have the instrument work properly, the
reagents must be at the same level as the
MS9-5s (MS9-3s).
Notes :
LABORATOIRES
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73
17 80
Page : 50 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoires
MS9-5s(-3s)
The MS9-5s(-3s) Batch installation is
BATCH Installation
Notes : (the instructions are written on the
described as follows: bottle), then normally resume the
O MS-Card® : reagents installation (see 'MS-
Open the Lyse Acti-Diff® carton box: the Pack®1 section).
MS-Card® protective holder is in the If by lack of space it would not be
closing flap. possible to have all the elements of the
Batch on the same level,
As with the MS-Pack® take the card out - It is possible to have some bottles
and insert it in the MS-Card® reader placed at a lower level:
located in the middle of the connection - Isoflux®,
panel (See 'MS-Card®' section). - Transflux®,
- Hemoref®.
© Tubing kit adaptation: - A long priming sequence must be
If necessary refer to the 'Reagent tubings programmed in this case at the
: Batch adaptation' section to adapt the MS9-5s (MS9-3s) startup.
tubing kit to the MS9-5s (MS9-3s) Batch - It introduces risks of higher
as shown on the picture. consumption.
- Please contact our technical
© Installation : service for further information.
Dilute the Hemoref® reagent of the Batch
A
t?
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73
17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 51
©
*^*<
U II
tf
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1
POWER SUPPLY
If these values are not reached it is fundamental and strictly necessary to request a better main
power installation by a professional. Earthing the MS9- 5s or the MS9-3s is a basic need for
safety, g o o d r e s u l t s a n d M a n u f a c t u r e r guarantee/warranty application.
Notes :
LABORATOIRES -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
LABORATOIRES
Technical
Hybrid stepper motors sequential d i lutor with m icro-control ler management
Specific 60 (im chamber for RBC and THR
Specific 80 (im chamber for WBC
Specific 80 im chamber for eosinophils
Built-in PC Computer:
- 1.6 GHz INTEL ATOM N270 or compatible processor
- Hard-disk storage up to 50 000 tests or more*
- Alphanumeric PC mini-keyboard
- Built-in LCD TFT graphic color high resolution display (SVGA-800x600)
- Touch screen
- Fast Ethernet lan adapter 10/100 Base TX
- RS232 ports mono and/or bi-directional transmission
- 1 parallel port, 3 USB ports 2.0
compliant Analyses back-up and user-
manual on CD-ROM
Windows XP® Multitask application
Quality control: Fully in compliance with international standards the MS9-5s
quality control Program include Jenning’s graph
Dimension : Width 340 mm, Height 420 mm, Depth 370 mm Weight: 20kg
Power supply: 100-240Volt / 50-50 Hz 300VA (automatic commutation)
Reagents: Diluent:ISOFLUX;Detergent:TRANSFLUX;Lysingreagent:ACTI-DIFF;HEMOREF;
Specific lysing reagent EO-DIFF Eosinophils
MAIN FEATURES MS9-3s®
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
Page : 56 MS9-5s® MS9-3s® Installation Manual / 2011-09 MELET SCHLOESING Laboratoi res
Technical
Hybrid stepper motors sequential dilutor with micro-controller
management Specific 60 im chamber for RBC and THR Specific 80 (im
chamber for WBC Built-in PC Computer:
- VIA 733 MHz or compatible processor
- Hard-disk storage up to 50 000 tests or more*
- Alphanumeric PC mini-keyboard
- Built-in LCD TFT graphic color high resolution display (VGA-640x480)
- Touch screen
- Fast Ethernet lan adapter 10/100 Base TX
- RS232 ports mono and/or bi-directional transmission
- 1 parallel port, 1 USB ports 1.1 compliant Analyses back-up
and user-manual on CD-ROM
Windows 98SE® application
Quality control: Fully in compliance with international standards the MS9-3s
quality control Program include Jenning’s graph
Dimension : Width 340 mm, Height 420 mm, Depth 370 mm Weight: 20kg
Power supply: 100-240Volt / 50-50 Hz 300VA (automatic commutation)
Reagents:Diluent:ISOFLUX;Detergent:TRANSFLUX;Lysingreagent:ACTI-DIFF;HEMOREF;
MELET SCHLOESING Laboratoires ’SOFTWARE &
MS-Card’ LICENSE AGREEMENT
IMPORTANT - READ BEFORE INSTALLING OR USING
This document is a LEGAL AGREEMENT between you and MELET SCHLOESING Laboratoires. Do not use or load any of your new
in vitro diagnostic MSL Instrument's Software and/or any associated materials as MS-Card (the "reagent card") until you have carefully
read the following terms and conditions. By loading and/or using the Software and/or using the MS-Card used with your new MSL
instrument, you agree to be bound to the terms of this License Agreement. If you do not wish to so agree, DO NOT install or use the
Software or any other associated materials as MS-Card.
MANUFACTURER. The "Manufacturer" word used herein includes, (i) companies having originally created your analyser's Software and/ or any associated
materials as the MS-Card (electronic card which supervises the reagent quality and includes owner's codes), (ii) companies licensed to manage any rights and/or
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80
MELET SCHLOESING Laboratoires MS9-5s® MS9-3s® Installation Manual / 2011-09 Page : 1 1
copyrights linked to any Intellectual property used in the machine and/or in any associated material as the MS-Card, (iii) individual who has any rights on any of the
Intellectual property used in the machine and/or in any associated material as the MS-Card,
MSL INSTRUMENTS. The "MSL Instruments" words used herein include Instrument(s), (i) manufactured by MELET SCHLOESING Laboratoires - France and/or
its affiliates and/or subsidiaries and distributed directly or indirectly (by mean of local Distributors), (ii) manufactured by MELET SCHLOESING Laboratoires - France
and/or its affiliates and/or subsidiaries for other companies under an OEM agreement.
LICENSE. You may copy and/or install any Manufacturer's Software onto the single built-in computer (or computer board) of your new MSL instrument for your
personal, non-commercial use, and you may NOT make any back-up copy of the Software. You may use the MS-Card on your MSL instrument to run it and profit
of its reagent quality management BUT you may NOT make any back-up and/ or copy and/or duplicate and/or compatible material of this card.
OWNERSHIP OF SOFTWARE AND COPYRIGHTS. Title to all copies of the Software remains with the Manufacturer or its suppliers. The Software is copyrighted
and protected by the European laws and other countries, and international treaty provisions. You may not remove any copyright notices from the Software. The
Manufacturer may make changes to the Software, and/or to items referenced therein as the MS-Card, at any time without notice, but is not obligated to support or
update the Software. Except as otherwise expressly provided, the Manufacturer grants no express or implied right under its patents, copyrights, trademarks, or
other intellectual property rights. You may transfer the Software and/or any associated materials as MS-Card only if the recipient agreesto be fully bound
by these
terms and if you retain no copies of the Software.
LIMITED MEDIA WARRANTY. If the Software has been delivered directly by the Manufacturer (or one of its Distributors) on physical media, the Manufacturer
warrants the media to be free from material physical defects for a period of ninety (90) days after delivery by the Manufacturer (or one of its Distributors). If such a
defect is found, return the media to the Manufacturer or your local Distributor for replacement or alternate delivery of the Software as the Manufacturer may select.
The limited media warranty is the same for the MS-Card if the MS-Card has not been already in use.
EXCLUSION OF OTHER WARRANTIES EXCEPT AS PROVIDED ABOVE, THE SOFTWARE IS PROVIDED "AS IS" WITHOUT ANY EXPRESS OR IMPLIED
WARRANTY OF ANY KIND INCLUDING WARRANTIES OF MERCHANTABILITY, NONINFRINGEMENT, OR FITNESS FOR A PARTICULAR PURPOSE. The
Manufacturer does not warrant or assume responsibility for the accuracy or completeness of any information, text, graphics, links or other items contained within
the Software and/or the MS-Card.
MANUFACTURER LIMITATION OF LIABILITY. IN NO EVENT SHALL THE MANUFACTURER OR ITS SUPPLIERS BELIABLE FOR
ANY DAMAGES WHATSOEVER (INCLUDING, WITHOUT LIMITATION, LOST PROFITS, BUSINESS INTERRUPTION, OR LOST INFORMATION) ARISING
OUT OF THE USE OF OR INABILITY TO USE THE SOFTWARE, EVEN IF THE MANUFACTURER OR ITS SUPPLIERS HAVE BEEN ADVISED OF THE
POSSIBILITY OF SUCH DAMAGES. SOME JURISDICTIONS PROHIBIT EXCLUSION OR LIMITATION OF LIABILITY FOR IMPLIED WARRANTIES OR
CONSEQUENTIAL OR INCIDENTAL DAMAGES, SO THE ABOVE LIMITATION MAY NOT APPLY TO YOU. YOU MAY ALSO HAVE OTHER LEGAL RIGHTS
THAT VARY FROM JURISDICTION TO JURISDICTION.
USER LIABILITY. IN THE EVENT YOU DO NOT RESPECT DIRECTLY OR INDIRECTLY AND/OR, INFRINGES PARTIALLY OR ENTIRELY THE
MANUFACTURER'S INDUSTRIAL PROPERTY -AND/OR ANY OTHER COMPANY'S INDUSTRIAL PROPERTY USED BY ANY OF MSL INSTRUMENTS-
INCLUDING SOFTWARE COPYRIGHT, PATENTS, TRADE MARKS, LOGO, MS-CARDS TECHNOLOGY OR OTHERS, OR, HELP OR PARTICIPATES WITH A
THIRD PARTY NOT TO RESPECT THE SAME MANUFACTURER'S INDUSTRIAL PROPERTIES -AND/OR ANY OTHER COMPANY'S INDUSTRIAL PROPERTY
USED BY ANY OF MSL INSTRUMENTS- AS ABOVE MENTIONED, INCLUDING DISASSEMBLING THE SOFTWARE OF ANY OF MSL INSTRUMENTS AND/OR
ANY KIND OF OTHER REVERSE-ENGINEERING ACTION, YOU ARE ENTIRELY LIABLE TO THE MANUFACTURER FOR ANY SPECIAL DAMAGES,
INCLUDING LOST OF PROFITS, LOST SAVINGS OR OTHER INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND.ANY FAILURE TO RESPECT
THE MANUFACTURER'S INDUSTRIAL PROPERTY SHALL BE DEEMED AS A CONTRACT VIOLATION AND A MATERIAL BREACH OF THIS AGREEMENT.
APPLICABLE LAWS. Claims arising under this Agreement shall be governed by the laws of your country and/or countries of Europe, excluding its principles of
conflict of laws and the United Nations Convention on Contracts for the Sale of Goods. You may not export the Software and/or any associated materials as MS-
Cards in violation of applicable export laws and regulations.
LABORATOIRES
BP 508 - OSNY - 95528 CERGY-PONTOISE Cedex / FRANCE Tel.: +33(0)1 30 75 30 00 Fax.: +33(0)1 30 73 17 80