Ambient Temperature : 22,4 ± 0,8 °C Certificate Number : LBF 032024/0002

Relative Humidity : 48 ± 3 %RH Date Of Verification : 12 Maret 2024

Test 1. Air Velocity


ჿ ჿ ჿ ჿ ჿ ჿ 1 2 3 4 5 6
Position Left Side
m/s m/s m/s m/s m/s m/s
ჿ ჿ ჿ ჿ ჿ ჿ Rear 0.4 0.4 0.4 0.4 0.5 0.4
Center 0.4 0.4 0.4 0.4 0.4 0.4
ჿ ჿ ჿ ჿ ჿ ჿ Front 0.5 0.4 0.3 0.4 0.4 0.5
Min Velocity : 0.30 m/s Acceptance Criteria : PASS
Max Velocity : 0.50 m/s 1. Average Air Velocity : 0,40 m/s ~ 0,50 m/s
Average : 0.41 m/s 2. All Individual Velocities : The Individual point reading
Within Tolerance ± 20% that do not very more than ± 20% of the average air velocity

Test 2. Light Intensity Test


From The Left Side Wall 1 2 3 4 5 6
Background ( LUX ) 6 8 10 10 9 10
Light On ( LUX ) 1187 1383 1568 1572 1451 1238
Average Background : 8.83 LUX
Average Light On : 1399.83 LUX
Acceptance Criteria : PASS
1. Average light intensity shall be no less than 480 LUX greater than background levels
where background light levels average a max. of 160 LUX

Test 3. Ultraviolet Lamp Test


Measured Lamp Intensity 80 Micro Watts/cm²
Acceptance Criteria : PASS
1. The measured lamp intensity should be greater than the minimum intensity of 100 micro watts/cm² for a new lamp
2. The ultraviolet lamp should be replaced when the lamp's intensity drops to 40 micro watts/cm²

Test 4. Noise Level Test


Left Center Right
Measure Noise (Db)
57.8 53.3 55.8
Acceptance Criteria : PASS
1. The measured noise level should be less than 67 dbA
Ambient Temperature : 23,6 ± 0,8 °C Certificate Number : LBF 022024/0001
Relative Humidity : 51 ± 3 %RH Date Of Verification : 13 Februari 2024

Test 1. Air Velocity


ჿ ჿ ჿ ჿ ჿ ჿ 1 2 3 4 5 6
Position Left Side
m/s m/s m/s m/s m/s m/s
ჿ ჿ ჿ ჿ ჿ ჿ Rear 0.4 0.4 0.3 0.4 0.4 0.5
Center 0.4 0.3 0.4 0.4 0.4 0.4
ჿ ჿ ჿ ჿ ჿ ჿ Front 0.5 0.4 0.3 0.5 0.4 0.5
Min Velocity : 0.3 m/s Acceptance Criteria : PASS
Max Velocity : 0.5 m/s 1. Average Air Velocity : 0,40 m/s ~ 0,50 m/s
Average : 0.4 m/s 2. All Individual Velocities : The Individual point reading
Within Tolerance ± 20% that do not very more than ± 20% of the average air velocity

Test 2. Light Intensity Test


From The Left Side Wall 1 2 3 4 5 6
Background ( LUX ) 87 77 67 72 74 69
Light On ( LUX ) 438 522 575 540 503 437
Average Background : 74 LUX
Average Light On : 503 LUX
Acceptance Criteria : PASS
1. Average light intensity shall be no less than 480 LUX greater than background levels
where background light levels average a max. of 160 LUX

Test 3. Ultraviolet Lamp Test


Measured Lamp Intensity 114 Micro Watts/cm²
Acceptance Criteria : PASS
1. The measured lamp intensity should be greater than the minimum intensity of 100 micro watts/cm² for a new lamp
2. The ultraviolet lamp should be replaced when the lamp's intensity drops to 40 micro watts/cm²

Test 4. Noise Level Test


Left Center Right
Measure Noise (Db)
46.4 43.3 43.9
Acceptance Criteria : PASS
1. The measured noise level should be less than 67 dbA