Professional Documents
Culture Documents
STANDARD 20486
First edition
2017-12
Reference number
ISO 20486:2017(E)
© ISO 2017
Provided by IHS Markit under license with ANSI
ISO 20486:2017(E)
ii
Provided by IHS Markit under license with ANSI
© ISO 2017 – All rights reserved
ISO 20486:2017(E)
Contents Page
Foreword .......................................................................................................................................................................................................................................... v
1 Scope ................................................................................................................................................................................................................................. 1
2 Normative references ...................................................................................................................................................................................... 2
3 Terms and definitions ..................................................................................................................................................................................... 2
4 Nominal leakage rates .................................................................................................................................................................................... 3
5 Classification o f leaks ...................................................................................................................................................................................... 3
5.1 Permeation leak ..................................................................................................................................................................................... 3
5.2 Conductance leaks ............................................................................................................................................................................... 3
5.2.1 Capillary leak ...................................................................................................................................................................... 3
5.2.2 Aperture leak (orifice) ................................................................................................................................................ 4
5.2.3 Compressed powder leak ......................................................................................................................................... 4
6 Calibration by comparison ......................................................................................................................................................................... 4
6.1 Methods A, As, B and B s .................................................................................................................................................................... 4
6.2 Applicability o f comparison methods ................................................................................................................................. 4
6.3 Preparation of leaks and apparatus ...................................................................................................................................... 5
6.3.1 Leak detector ...................................................................................................................................................................... 5
6.3.2 Connection to the leak detector .......................................................................................................................... 5
6.3.3 Temperature accommodation .............................................................................................................................. 7
6.4 Measurement............................................................................................................................................................................................ 7
6.4.1 Set-up ......................................................................................................................................................................................... 7
6.4.2 General measurement sequence ........................................................................................................................ 7
6.5 Evaluation for methods A, As, B and B s (Comparison) ......................................................................................... 8
6.5.1 Determination of leakage rate .............................................................................................................................. 8
6.5.2 Influence factors to measurement uncertainty ..................................................................................... 9
7 Volumetric calibration................................................................................................................................................................................. 10
7.1 Direct flow (Method C) .................................................................................................................................................................. 10
7.1.1 General................................................................................................................................................................................... 10
7.1.2 Equipment .......................................................................................................................................................................... 10
7.1.3 Preparation of leaks and apparatus ............................................................................................................. 10
7.1.4 Measurement ................................................................................................................................................................... 11
7.1.5 Evaluation for Method C (direct flow measurement) ................................................................... 13
7.2 Leak measurement under water (Method D) ............................................................................................................ 14
7.2.1 General................................................................................................................................................................................... 14
7.2.2 Equipment .......................................................................................................................................................................... 14
7.2.3 Preparation of leaks and apparatus ............................................................................................................. 14
7.2.4 Measurement ................................................................................................................................................................... 15
7.2.5 Evaluation for Method D ........................................................................................................................................ 16
7.2.6 Influence factors to measurement uncertainty .................................................................................. 17
7.3 Calibration by (volumetric) gas meter (Method E) .............................................................................................. 17
7.3.1 General................................................................................................................................................................................... 17
7.3.2 Equipment .......................................................................................................................................................................... 18
7.3.3 Preparation of leaks and apparatus ............................................................................................................. 18
7.3.4 Measurement ................................................................................................................................................................... 18
7.3.5 Evaluation for Method E (gas meter) .......................................................................................................... 18
7.3.6 Influence factors to measurement uncertainty .................................................................................. 19
7.4 Calibration by pressure change in a known volume (Method F) .............................................................. 19
7.4.1 General................................................................................................................................................................................... 19
7.4.2 Preparation of leaks and apparatus ............................................................................................................. 20
7.4.3 Measurement ................................................................................................................................................................... 22
7.4.4 Special situation in vacuum chambers ...................................................................................................... 23
7.4.5 Evaluation for Method F (pressure change) .......................................................................................... 25
7.4.6 Influence factors to measurement uncertainty .................................................................................. 25
7.5 C alib ratio n by vo lume change at co ns tant p res s ure (M etho d G) ............................................................. 26
7.5.1 Equipment .......................................................................................................................................................................... 26
7.5.2 Preparation of leaks and apparatus ............................................................................................................. 26
7.5.3 Measurement ................................................................................................................................................................... 26
7.5.4 Evaluation for Method G (volume change at constant pressure) ........................................ 27
8 G e n e ra l i n fl u........................................................................................................................................................................................... 28
e n c e s
9 Report ........................................................................................................................................................................................................................... 28
10 Labelling of reference leaks ................................................................................................................................................................... 29
11 Handling of reference leaks ................................................................................................................................................................... 29
11.1 General ........................................................................................................................................................................................................ 29
1 1 .2 Permeatio n leaks (no rmally with res ervo ir fitted the leak o utlet) ......................................................... 29
1 1 .3 C o nductance leaks (no rmally witho ut res ervo ir) .................................................................................................. 29
Annex A (informative) Calculation of leakage rate decrease due to tracer gas depletion in
the reservoir .......................................................................................................................................................................................................... 30
Bibliography ............................................................................................................................................................................................................................. 32
iv
Provided by IHS Markit under license with ANSI
© ISO 2017 – All rights reserved
ISO 20486:2017(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work o f preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters o f
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
di fferent types o f ISO documents should be noted. This document was dra fted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some o f the elements o f this document may be the subject o f
patent rights. ISO shall not be held responsible for identi fying any or all such patent rights. Details o f
any patent rights identified during the development o f the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is in formation given for the convenience o f users and does not
constitute an endorsement.
For an explanation on the voluntary nature o f standards, the meaning o f ISO specific terms and
expressions related to con formity assessment, as well as in formation about ISO's adherence to the
World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT) see the following
URL: www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 135, Non-destructive testing,
Subcommittee SC 6, Leak testing.
a) Calibration by comparison
2 Normative references
T he fol lowi ng do c u ments are re ferre d to i n the te xt i n s uch a way that s ome or a l l o f thei r content
con s titute s re qu i rements o f th i s do c u ment. For date d re ference s , on ly the e d ition c ite d appl ie s . For
undate d re ference s , the l ate s t e d ition o f the re ference d do c u ment (i nclud i ng a ny amend ments) appl ie s .
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https://www.iso.org/obp
— IEC Electropedia: available at http://www.electropedia.org/
3.1
unknown leak
leak having a stable and repeatable leakage rate of known order of magnitude that can be determined
b y c a l ibration
3.2
reference leak
c a l ibrate d le a k wh ich may b e u s e d to c a l ibrate a no ther le a k
N o te 1 to entr y: T he u ncer ta i nty o f the re ference le a k i s lower th a n the re qu i re d u ncer ta i nty o f the le a k to b e
calibrated.
3.3
calibration
s e t o f op eration s wh ich e s tabl i sh , under s p e ci fie d cond ition s , the rel ation sh ip b e twe en le a kage rate
va lue s repre s ente d b y an u n known le a k and the corre s p ond i ng known va lue s o f the le a kage rate
N o te 1 to entr y: I n the c a s e o f c a l ibratio n b y comp a ri s on , the known va lue s o f the le a kage rate a re repre s ente d b y
a reference leak.
N o te 2 to entr y: N o r m a l l y, the re s u lt o f a c a l ib ration i s gi ven a s the le a kage rate va lue for the re ference le a k with
a s ta nd a rd u ncer ta i nty.
3.4
nominal leakage rate
le a kage rate o f a le a k c a lc u late d for s p e ci fie d re ference cond ition s
N o te 1 to entr y: I n le a k de te c tion , le a kage rate s a re co m mo n l y gi ven i n u n its o f pV-th ro ughput ( P a· m 3/s, mbar l/s,
Std cm3 /m i n) . T he s e a re o n l y a pre c i s e me a s u re o f ga s flow i f the temp eratu re i s given a nd kep t con s ta nt. F low
u n its s uch a s m a s s flow ( g/y) or mol a r flow (mol/s) a re s ome ti me s u s e d to o verco me th i s pro b lem .
5 C l a s s i fi c a ti o n o f l e a ks
appre c iable temp eratu re co e fficient (approxi mately 3 , 5 %/K ) . Typic a l le a kage rate s are i n the range
(approxi mately 0 , 3 %/K ) but e as i ly blo cks i f no t ha nd le d with c a re . Typic a l le a kage rate s a re gre ater
than 10 Pa·m3/s. −7
5 . 2 . 2 A p e r tu r e l e a k ( o r i fi c e )
Orifices are seldom used as re ference leaks in practice, as they are di fficult to manu facture and even
more prone to blocking than capillaries.
NOTE Critical flow orifices are a form o f aperture leak that is commonly found in industry, but are out o f the
scope of this document.
5.2.3 Compressed powder leak
This type o f leak uses metal powder compressed into a tube. They are usually o ffered without reservoir.
They are used for routine check o f the sensitivity o f leak detectors but they are not stable enough to be
used as calibrated leaks. Their suitability depends on how well controlled the storage and operating
conditions are, and on the required uncertainty.
6 Calibration by comparison
For leaks with exhaust into the atmosphere and measurement by sni ffing gas, more conditions shall be
controlled. These are:
— the background level of tracer gas shall be as low as possible and as stable as possible;
— the total gas flow rate o f the sni ffer shall be high enough to take up the total tracer gas flow out o f
the leak;
— the aspiration o f the sni ffer (the coupling to the leak exhaust) shall be o f suitable geometry to make
sure that the atmospheric gas flow across the leak exhaust takes up the whole tracer gas flow from
the leak opening.
As a consequence, the measurement uncertainty is appreciably higher for sni ffer leaks than for
vacuum leaks.
Methods by comparison are there fore applicable but not pre ferable for the calibration o f sni ffer leaks
(with exhaust to atmosphere).
6.3 Preparation of leaks and apparatus
Key
1 transfer device (leak detector)
2 test port
3 rig
4 leaks to be calibrated and reference leak
5 ho o ds fo r thermal s tab ility
I n the c a s e o f s n i ffer le a ks , the con ne c tion to the le a k de te c tor s n i ffi ng tip i s made b y an adap ter wh ich
ma ke s a tight conne c tion to the le a k outle t and enable s atmo s pheric ai r to b e conti nuou s ly s ucke d
Figure 2
acro s s the le a k exh au s t via an ai r i n le t (s e e Key item 3 i n ) , s o that the whole le a k ga s flow
i s ta ken up b y the s n i ffer tip . T he ai r i n le t op en i ng s ha l l no t th ro ttle the fre e flow o f ai r to mai ntai n
Key
1 gasket 4 test leak with leak opening
2 sniffer tip 5 adapter body with gasket
3 air inlet 6 sniffer opening with cross-wise slot
Figure 3 — Example for a coupling adapter for sniffer leaks
6.4.1 Set-up
It is important to ensure that the effective pumping speed at the leak detector inlet for vacuum leaks,
respectively the sni ffer gas flow for overpressure leaks, is not changed during the measurements.
I f possible, either with the leak detector or in an auxiliary device, a long averaging time may be used to
decrease the statistical measurement uncertainty.
All the measurement instruments should be adjusted in such a way that they give nearly full-scale
deflections for the biggest leak.
6.4.2 General measurement sequence
Generally, each reading shall be obtained only a fter the signal o f the trans fer instrument has stabilized.
A su fficient number o f readings shall be taken to achieve the lowest possible statistical uncertainty.
is as follows:
a) z ero s igna l de term i nation: all va lve s clo s e d for vac uu m le a ks , re s p e c tively s n i ffer tip in pu re
αref, αu a re the temp eratu re co e ffic ients o f the re ference a nd un known le a k, re s p e c ti vely;
Δ Tref Tu
, Δ are the departures of the temperature of the leaks from the reference temperature of
the re ference and u n known le a k, re s p e c tively.
NOTE 1 The readings (Rref and Ru ) a re o b ta i ne d from the le a k de te c tor d i s p l ay a s the d i fference o f the o utp ut
NO TE 2 T he temp eratu re co e ffic ient o f the re ference le a k i s no r m a l l y s tate d . I f the temp eratu re co e ffic ient o f
temperature difference between each reference leak and its calibration reference temperature is
Formula (2) holds. See Figure 4.
con s idere d . I n th i s c as e, the s i mpl i fie d
R u − R1
Q u = ( Q2 − Q1 ) + Q1 (2)
R 2 − R1
where
Q u , Q1 , Q2 are le a kage rate s o f the u n known le a k and the re ference le a ks 1 and 2 , re s p e c tively;
Ru , R 1 , R 2 are re ad i ngs for the u n known le a k and the re ference le a ks 1 and 2 , re s p e c tively.
Key
A reference leak 1
B unknown leak
C reference leak 2
R reading
Q leakage rate
NOTE Q1 is the leak with the smaller rate, Q2 the leak with the larger rate. Qu, the leakage rate of the
unknown leak to be calibrated, lies between these two known leaks.
Figure 4 — Two-point calibration of a leak
6 . 5 . 2 I n fl u e n c e fa c to r s to m e a s u re m e n t u n c e r ta i n ty
7 Volumetric calibration
7 . 1 D i re c t fl o w ( M e th o d C )
7.1.1 General
This method is applicable to conductance leaks in the range of 10 −5 Pa·m3/s (~0,006 Std cm3/min) to
0,2 Pa·m3/s (~100 Std cm3/min). Leaks from 10 −6 Pa·m3/s (~0,000 6 Std cm3/min) to 10 −5 Pa·m3/s
(~0,006 Std cm3/min) can be calibrated, but with a rather large uncertainty. In that range, i f a suitable
re ference leak is available, methods A or B should be employed to give lower uncertainty.
Two types o f pressure and flow conditions shall be considered:
— flow from pressure to atmosphere (see 7.1.4.1);
— flow from atmosphere to vacuum (see 7.1.4.2).
The third possible flow condition, pressure to vacuum, cannot be measured with Method C (i f this is
required, a calibration with tracer gas according Method A or B with a suitable MSLD or a calibration
according to Method F or G needs to be made.)
7.1.2 Equipment
To calibrate a leak by measurement o f capillary flow according to Method C described in 7.1.4, a
calibrated glass tube (pre ferably with a suitable vent valve at one end, see Figures 5 and 6 ) is necessary.
An indicator fluid (normally water with some sur factant added or special oils) is used to produce the
measurement slug in the capillary.
To measure the time of slug movement, a timer or stopwatch is needed. Instruments based on the timed
movements o f a film in a tube are also available, e.g. a bubble flow meter.
As conductance leak elements normally have no tracer gas reservoir, a separate tracer gas supply is
needed or calibration may be per formed with filtered, oil free and dry atmospheric air.
7.1.3 Preparation of leaks and apparatus
Atmosphere to vacuum
In this case, the leak outlet is connected to a vacuum and the inlet to the capillary tube. The capillary
inlet is open to atmospheric pressure (see Figure 6).
If standard leakage rates are required, the outlet absolute pressure shall be less than 100 Pa.
7.1.4 Measurement
Key
1 gas pressure from reservoir 7 atmosphere
2 unknown leak 8 indicating slug
3 optional pump connection (to initialize the slug position) 9 calibration marks
4 optional vent valve 10 connecting tubes
5 trailing edge of slug 11 zero space
6 inside diameter capillary tube
Key
1 vacuum 7 atmosphere
2 unknown leak 8 indicating slug
3 gas fill connection 9 calibration marks
4 filling valve 10 connecting tubes
5 leading edge of slug 11 zero space
6 inside diameter capillary tube
Figure 6 — Set-up for Method C measurement: atmosphere to vacuum
7 . 1 . 5 E va l u a ti o n f
o r M e th o d C ( d i r e c t fl o w m e a s u re m e n t)
∆V
Q pV = p TA ×
∆t
(3)
where
is the pV-throughput at test condition;
Q pV
p TA is the test pressure at leak exit;
Pressure-to-atmosphere measurements shall be corrected for the vapour pressure of the indicating
f
l iqu id, f
or the vi s co s ity o f f the tracer ga s a nd a l s o or the i n fluence o pre s s u re a nd temp eratu re . For the
nominal leakage rate, reference pressure and temperature are given in Clause 4. Formula (4) assumes
lam i nar flow i n a rou nd pip e (H agen-Poi s eu i l le) . For e ach typ e o f le a k, it sh a l l b e che cke d whe ther
th i s a s s ump tion i s appl ic able . O ther wi s e, it i s ne ce s s ar y to c a lc u l ate with a n ade quate formu la or an
empirical correction.
T he nom i na l le a kage rate i s c a lc u late d by Formula (4):
η ref p ref , in 2 − p ref ,out 2 p test ,out × Tref p test ,out
Q N = Q pV × × × ×
η test p test , in 2 − p test ,out 2 p ref ,out × Ttest p test ,out − ( p v,ref − p v,test )
(4)
where
Q N is the nominal leakage rate;
Q pV is the pV-throughput at test condition according to Formula (3);
η ref i s the vi s co s ity o f ga s at re ference cond ition s;
Atmosphere to vacuum measurement need not to be corrected for vapour pressure because, in this
situation, gas with high humidity is passing the leak.
The nominal leakage rate is given by:
η ref p ref , in 2 − p ref ,out 2 p test ,out × Tref
Q N = Q pV × × ×
η test p test , in 2 − p test ,out 2 p ref ,out × Ttest
(5)
7 . 1 . 5 . 2 I n fl u e n c e f
a c to r s to m e a s u re m e n t u n c e r ta i n ty
7.2.1 General
This method is applicable to conductance leaks in the range of 0,2 Pa·m3/s (~100 Std cm3/min) up to
around 8 Pa·m3/s (~5 000 Std cm3/min).
Calibration from atmosphere to vacuum is not possible with this method.
7.2.2 Equipment
Leaks, which are too large to be calibrated by Method C can be calibrated by catching the escaping gas
under water in a calibrated volume.
A measuring glass with calibrated volume, a stopwatch, temperature measuring equipment for ambient
temperature and water temperature, a test pressure sensor and a sensor for ambient pressure are
necessary.
As conductance leaks normally have no tracer gas reservoir, a separate tracer gas supply is needed for
calibration.
7.2.3 Preparation of leaks and apparatus
The measuring glass is filled with water (see Figure 7). The water shall have the same temperature as
the ambient temperature. The leak is connected to the gas supply and the flow o f the leak is conducted
via a hose into the measuring glass. The time to fill a certain volume with gas shall be recorded.
I f the inlet pressure o f the leak is low (<50 kPa), the hydrostatic pressure is not negligible and the
average pressure shall be considered and checked at the beginning of the calibration. This is achieved
by replacing some o f the water in the measuring glass with gas via the unknown leak. When the water
in the measuring glass has been replaced by hal f o f the planned measuring gas volume, the gas supply
valve shall be closed. After a few seconds the pressure will equilibrate and the pressure gauge at the
leak inlet shows the difference between ambient pressure and the average exhaust pressure at the
unknown leak. This pressure plus the ambient pressure is the outlet pressure, p test,out, of the leak and
Formula (6).
sh a l l b e u s e d i n the c a lc u lation b y
Fu r thermore, s ati s fy the re qu i rements for the prep a ration accord i ng to M e tho d C .
Key
1 test pressure 5 water basin
2 pressure gauge 6 measuring glass
3 unknown leak 7 water level
4 connecting hose 8 shut-off valve
Figure 7 — Set-up for Method D measurement
7.2.4 Measurement
a) T he pre s s u re s en s or s ha l l b e place d d i re c tly at the entra nce o f the re ference le a k (s e e Figure 8);
b) the i n le t pre s s u re at the re ference le a k sh a l l b e adj u s te d with exhau s t to a mbient;
c) water temp eratu re and ambient temp eratu re sha l l b e the s ame with i n a tolerance o f ±1 K;
e) the hose shall be inserted into the measuring glass and the time shall be measured until a certain
volume of gas is accumulated inside the measuring glass (see Figure 9);
f) rep e at s tep s b) to e) at le a s t th re e ti me s . T he rep e atabi l ity of the me a s urements sha l l be
with i n ± 2 % .
Key
1 test pressure 4 connecting hose
2 pressure gauge 5 water basin
3 unknown leak 6 measuring glass
Figure 8 — Method D: Start condition of measurement
Key
1 test pressure 4 connecting hose
2 pressure gauge 5 water basin
3 unknown leak 6 measuring glass
Figure 9 — Method D: End condition of measurement
e ach typ e o f re ference le a k, it s ha l l b e che cke d whe ther th i s c a lc u lation i s p erm itte d . O ther wi s e, it i s
where
QN is the nominal leakage rate;
QpV is the pV-throughput at test condition according to Formula (3);
η ref i s the vi s co s ity o f ga s at re ference cond ition s;
7 . 2 . 6 I n fl u e n c e f
a c to r s to m e a s u re m e n t u n c e r ta i n ty
— test time;
— test volume;
— ambient pressure;
— temperature of the water;
— ambient temperature;
— vapour pressure;
— test pressure;
— s olubi l ity o f the ga s i nto the l iqu id .
T he u ncer tai nty s ha l l b e c a lc u l ate d accord i ng the com mon gu idel i ne s (s e e [ 2]).
7.3 Calibration by (volumetric) gas meter (Method E)
7.3.1 General
This method is applicable to conductance leaks in the range of 2 000 Std cm3/min up to around
100 000 Std cm3/min.
7.3.2 Equipment
If the leaks are so large that it is not possible to calibrate them using Methods C or D, it is possible to use
a calibrated gas meter. The gas flow from the re ference leak (outlet) shall be led into the gas meter. The
test time shall be minimum 1,5 min.
As equipment, a calibrated gas meter, a stopwatch, temperature measuring equipment for ambient
temperature and a sensor for ambient pressure are necessary.
NOTE As a gas meter measures volume flow (e.g. m 3/h) the total pressure is needed to obtain pV-throughput.
As conductance leaks normally have no tracer gas reservoir, a separate tracer gas supply is needed or
calibration may be per formed with filtered, oil free and dry atmospheric air.
7.3.3 Preparation of leaks and apparatus
The re ference leak shall be fitted directly at the entrance o f the gas meter. The pressure sensor shall be
close to the entrance of the reference leak.
Furthermore, the requirements for the preparation according to Method C shall be observed.
7.3.4 Measurement
a) The pressure sensor shall be directly at the entrance o f the re ference leak;
b) the re ference leak shall be mounted directly at the inlet o f the gas meter;
c) the time shall be measured until a certain volume has passed the gas meter;
d) the measured volume should be larger than 1,5 × of the nominal volume of the reference leak;
e) repeat step c) at least three times. The repeatability o f the measurements should be within ±2 %.
7.4.1 General
This method is an alternative to Methods A and B and it can be used for most types o f non-condensing,
non-corrosive gases.
The leakage rate is calculated by the pressure rise in a known volume over time. By using di fferent
volumes and different times, it is possible to adapt to different leakage rates. This method is usable
for leaks with or without a gas reservoir. It is also possible to measure a gas flow from over-pressure
to ambient pressure, from ambient pressure to vacuum, from over-pressure to vacuum or from over-
pressure to another over-pressure.
As equipment, a pressure sensor, a stopwatch, a chamber with a known volume and a gas supply for all
re ference leaks without own gas reservoir is necessary. I f it is necessary to use other pressures inside
the known volume, a vacuum pump or a separate gas supply with dry and clean gas is also necessary.
The leak unit of 1 mbar·l/s means a pressure change of 1 mbar in 1 l volume per second. This is the
same as a leakage rate of 1 bar cm3/s = 0,987 Std cm3/s.
The leakage rate for those leaks can be specified in Std cm 3/min, in mbar·l/s or in Pa·m 3/s.
Using this method, it is possible to determine a leakage rate by a pressure change in a known volume.
There are three options for calibration with this method:
1) gas flow through the calibrated leak from pressure (absolute pressure) to vacuum in the test
chamber;
2) gas flow through the calibrated leak from positive pressure to ambient pressure in the test
chamber;
3) gas flow through the calibrated leak from pressure (absolute pressure) to a lower pressure in the
test chamber.
© ISO 2017 – All rights reserved 19
Provided by IHS Markit under license with ANSI
ISO 20486:2017(E)
The accuracy o f the calibration depends on the test volume, on the test time and under certain
conditions also on physical e ffects (e.g. desorption from the chamber walls). It can be a time-consuming
calibration and uncertainty decreases with the test time. This type o f calibration is pre ferred for leaks
in the range between 10 −7 Pa·m3/s to 10 −3 Pa·m3/s.
The advantage o f this method is that it can be used for nearly all gases and a mass spectrometer is
not needed.
NOTE A deeper evaluation of this method can be found in [3].
7.4.2 Preparation of leaks and apparatus
Independent from the calibration method, the first step is to determine the exact inside volume o f the
test chamber. This can be done by two methods:
1) connection of the test chamber to a known volume;
2) injecting a known gas volume to the test chamber.
It is important that the valve between the chamber and the known volume does not have an internal
volume itsel f. A change o f the volume, which is generated by the valve, shall be taken into account.
During the volume check, the valve to the calibrated leak shall be open so that the dead volume between
leak and valve o f the re ference leak is included in the calculation. There should be no flow through the
calibrated leak during the volume check.
The volume check shall be done 3 times and shall be used in the calculation o f measurement uncertainty.
In a test with vacuum chambers, there is the e ffect o f pressure rise by outgassing from the sur faces.
Before a calibration, this effect of outgassing shall be determined and be subtracted from the result.
The system shall be thermally insulated against rapid ambient temperature changes.
It shall be ensured that the test chamber is free of leaks.
7.4.2.1 Volume determination of test chamber by connecting a known volume
A known gas volume, V2 , with pressure, p2 , is connected to the test chamber of unknown volume, V1 ,
and pressure, p1 , by a valve (which has no internal volume). Opening the valve generates a total volume,
V3 , with a uniform pressure, p3 .
The leak tightness o f the overall system shall be ensured. See Figure 10.
Key
1 test chamber V1 with pressure p1 5 vacuum pump
2 pressure sensor 6 reference leak
3 venting valve 7 dead volume VD
4 known gas volume V2 with pressure p2
Figure 10 — Set-up for Method F measurement by connection of the test chamber to a
known volume
Key
1 test chamber V1 with pressure p1 5 vacuum pump
2 pressure sensor 6 reference leak
3 venting valve 7 connecting valve
4 known gas volume V2 with pressure p2
Figure 11 — Set-up for Method F measurement by injecting a known gas volume to the test
chamber
7.4.3 Measurement
After the volume of the test chamber has been determined, the leak to be calibrated shall be connected
to the test chamber via the connecting valve. The gas from the unknown leak causes an increase of
pressure in the test chamber over time. The pressure increase in the test chamber can be calculated to
yield a lea kage rate .
f
e ach typ e o le a k, it s ha l l b e che cke d whe ther th i s c a lc u lation i s p erm itte d . O ther wi s e, it i s ne ce s s ar y to
T he le a kage rate o f the le a k to b e c a l ibrate d at atmo s pheric pre s s u re i s given b y Formula (10):
∆p × V1 η ref p ref , in 2 − p ref ,out 2 p test ,out × Tref
QN =
t
× × ×
η test p test , in 2 − p test ,out 2 p ref ,out × Ttest
(10)
where
Q N is the nominal leakage rate;
Δ p is the pressure change in the chamber during test time;
V 1 is the volume of the test chamber;
t is the test time;
η ref i s the vi s co s ity o f ga s at re ference cond ition s;
7.4.4.1 General
If the chamber is under vacuum, it is possible that the inner surfaces are outgassing. This outgassing
f f f
i n fluence s the te s t re s u lt a nd it i s there ore ne ce s s ar y to de term i ne the amou nt o de s orbi ng gas b e ore
outgassing is repeatable.
Key
Δp pressure increase by outgassing
p total pressure
t measurement time
Figure 12 — Method F: pressure increase by outgassing process
A fter the volume o f the test chamber has been determined, the test system shall be set to start
conditions and the leak to be calibrated shall be connected to the test chamber.
The leakage rate of the unknown leak causes a pressure rise inside the test chamber volume, V1 . The
leakage rate o f the unknown leak is calculated by this pressure rise in the test chamber volume, V1 ,
corrected for the outgassing of the test chamber. See Figure 13.
Key
1 pressure increase by outgassing + leakage rate
2 pressure increase by outgassing
p total pressure
t measurement time
Figure 13 — Method F: pressure increase by outgassing process and leakage rate
o f the le a k to b e s u re that there i s the corre c t pre s s u re . T he u ncer ta i nty o f th i s manome ter sha l l b e a l s o
ta ken i nto accou nt i n the c a lc u lation o f the u ncer tai nty o f the u n known le a k.
— test time;
— volume of the test chamber;
— ambient temperature;
— test pressure at reference leak;
— u ncer tai nty o f pre s s u re s en s or (i f pre s ent) at u n known le a k;
T he u ncer tai nty s ha l l b e c a lc u l ate d accord i ng the com mon gu idel i ne s (s e e [ 2]).
To c a lc u late the u ncer ta i nty o f the volu me o f the te s t ch amb er a l l de ad volu me s , e s p e ci a l ly tho s e
7.5.1 Equipment
The system setup is shown in Figure 14 . There is a connection flange for the leak to be calibrated and
two valves for venting to atmospheric pressure. A diaphragm gauge measures the differential pressure.
The change in volume ΔV is made with a calibrated drive to which a piston is connected. The piston has
a vacuum tight seal to the chamber of calibration.
Key
1 leak to be calibrated 4 piston drive
2 differtial pressur sensor 5 volume V
3 valve V1 6 valve V2
Figure 14 — Schematic of the pΔV calibration system
7.5.3 Measurement
Once the leak is connected and the valves shut, the pressure in the calibration volume starts to increase.
As soon as the di fferential pressure rises above a certain level, a change o f volume, ΔV, is made by the
piston. As a consequence of the increase in volume, the pressure drops below the zero differential
pressure level but starts to increase again due to the gas emitted from the leak. When the pressure
reaches the original level, another increase in volume is performed. This procedure of the volume
increment is repeated at least 20 times. The pressure data shall be recorded. See Figure 15.
Key
t time in s
Y differential pressure in hPa
1 valve operations
2 cycles
I n b o th c a s e s , the s e t-up sh a l l a l low for 2 0 con s e c uti ve e qua l volu me cha nge s , Δ V , during one movement.
Measurement of calibration pressure, p 0 , s ha l l be done with a c a l ibrate d gauge, typic a l ly clo s e to
atmospheric pressure.
T he ti me i nter va l, t
Δ , b e twe en two me as u rements sh a l l be eva luate d by averagi ng ab out
20 measurements of time intervals between zero differential pressure crossings. The scatter of these
i nter va l s i s mo s tly de term i ne d b y the uncer ta i nty o f the d i fferenti a l pre s s ure gauge .
7.5.4.1 General
A determination of the leakage rate at the calibration temperature requires four quantities:
— absolute pressure, p;
— V;
change i n volume, Δ
— t;
ti me, Δ
∆V
q= ⋅p
∆t 0
(12)
where
Δ V is the periodic volume change;
Δ t is the time interval between two measurements;
p 0 is the reference pressure held constant during the measurement (close to atmospheric pressure).
7 . 5 . 4 . 2 I n fl u e n c e f
a c to r s to m e a s u re m e n t u n c e r ta i n ty
T he random u ncer tai ntie s ca n no t be c a lc u late d but sh a l l be de term i ne d by a s tati s tic a l ana lys i s
(rep e atabi l ity ana lys i s) o f the s c atter o f re ad i ngs o f the d i fferent i n s tru ments u s e d or more e as i ly b y a
nu mb er o f re s u lts eva luate d with a l l s ys tematic i n fluence s held as con s tant as p o s s ible .
T he re s u lt o f s uch a rep e atabi l ity ana lys i s i s a s ample o f me a s u rement re s u lts , wh ich typic a l ly have a
norma l d i s tribution o f va lue s . Fol lowi ng s tand ard me tho d s o f uncer ta i nty eva luation, the u ncer ta i nty
— testing time
— te s ti ng volume (u ncer ta i nty o f the pi s ton and l i ne a r u n it)
— ambient temperature
— te s ti ng pre s s u re (uncer ta i nty o f the di fferenti a l pre s s ure s en s or, s tabi l ity o f re ference pre s s u re)
T he flow th rough conduc tance le a ks (c api l la r y ori fice typ e) i s i n fluence d b y vi s co s ity and the vi s co s ity
i s a l s o i n fluence d b y pre s s u re . T here fore, th i s i n fluence s ha l l b e ta ken i nto account when con s ideri ng
— heat from the hand on the apparatus before start of the measurement;
— evaporation cooling of the surface of the test volume (Method D);
— expansion refrigeration;
— flow re s i s tance o f ga s me ter ( pre s s u re lo s s i n s ide ga s me ter) .
Known i n fluence s s hou ld b e m i n i m i z e d b y appropri ate me tho d s and a re a l i s tic as s e s s ment o f the
re s u lti ng uncer ta i nty s hou ld b e done and adde d to the c a lc u late d me a s urement u ncer tai nty.
9 Report
The calibration report of a reference leak shall contain the following information:
— typ e o f le a k ( p erme ation or conduc ta nce) ;
— typ e o f tracer ga s;
— measured leakage rate (in throughput units: Pa·m3/s, mbar·l/s or Std cm3/min);
— inlet and outlet pressure, for which the measured leakage rate is valid;
— nominal leakage rate (1 000 mbar, 20 °C, in throughput units: Pa·m3/s, mbar·l/s or volume units:
Std cm3/min);
— u ncer tai nty o f me a s urement (re s u lt o f c a lc u lation) ;
— calibration temperature;
— e xp e c te d de cre as e o f le a kage rate [due to deple tion o f tracer ga s re s er voi r, i f a ny (%/a) s e e Annex A];
— u s e d mea s u rement e qu ipment (i nclud i ng the s p e ci fie d tolera nce/uncer ta i nty o f th i s e qu ipment) ;
— date of calibration;
— method used, according to this standard (A, As , B, B s , C, D, E, F or G);
— u n ique identi fic ation for the re ference le a k;
11.1 General
To prevent breakage of the leak element, which is often made from glass or quartz, calibrated leaks
shou ld a lways b e hand le d with c are . I f da mage i s s u s p e c te d , the le a k sha l l b e re - che cke d for corre c t
1 1 . 2 P e r m e a ti o n l e a ks ( n o r m a l l y w i th re s e r vo i r fi tte d th e l e a k o u tl e t)
For p erme ation le a ks , wh ich norma l ly have ver y s ma l l le a kage rate s , it i s mo s t i mp or ta nt duri ng s torage
space in front of the closed valve or in the elastomeric seal of the valve.
11.3 Conductance leaks (normally without reservoir)
Conductance leaks shall be kept in an environment free of dust particles and moisture, which is best
ach ieve d b y clo s i ng thei r outle t b y either a fla nge or a l l-me ta l va lve . T h i s do e s no t prevent tracer ga s
Annex A
(informative)
Calculation of leakage rate decrease due to tracer gas depletion in
the reservoir
dp x
Q x = −V
dt
(A.1)
On the other hand, Qx can be factorized into two functions, C and f( p x(t), pamb). C contains the time-
i ndep endent f( p x(t), p amb) describes the dependence on pressures of the
ge ome tr y, materi a l , e tc .
throughput. pamb deno te s the ambient pre s s u re at wh ich the tracer ga s flows .
Q x = Cf ( p p ) x
,
amb
(A.2)
Inserting Formula (A.2) into Formula (A.1) yield s a d i fferenti a l formu l a who s e s olution de s c rib e s the
time dependence of Qx . For a fi rs t orientation o f the de cre a s e o f the le a kage rate, the fi rs t deri vative o f
Q x to t at t = 0 is a good approximation.
The time dependence, f( p x(t), pamb), can be expressed for one of the following cases.
a) Permeation type leak f( p x , p amb) = p x – p amb
(mole cu la r flow) :
2 p x ,0 Q x ,0
Q x = Q x ,0 1 − t (A.5)
(
V p x2 ,0 − p amb
2
)
F low i nto vac uu m s i mpl i fie s Formula (A.5) to:
2 Q x ,0
Q x = Q x ,0 1 − t
Vp x ,0
(A.6)
Bibliography
[1] Arai K., Yoshida H., H irata M., Akimichi H., Kobata T. Stability tests of standard leaks
for the calibration by a comparison method Original Research Article. Measurement. 2012, 45
pp. 2441–2444
[2] JCGM 100:2008, Evaluation of measurement data — Guide to the expression of uncertainty in
measurement (GUM)
[3] Arai K., & Yoshida H. Primary flow meter for calibrating a sni ffer test leak artefact by a
pressure rise method. Metrologia. •••, 51 (5) p. 522
ICS 19.100
Price based on 32 pages
© ISO 2017 – All rights reserved
Provided by IHS Markit under license with ANSI